Quality Assurance in Radiation Therapy: Simulators & Ancillary Equipment
Quality Assurance in Radiation Therapy: Simulators & Ancillary Equipment
2
Radiographic Simulators
rotating
gantry
localizing laser
(not shown)
patient support
assembly
3
Radiographic Simulators
X-ray source with CBCT capability
adjustable collimation
field light
rotating
gantry
patient support
assembly
5
Radiographic Simulators QA:
Reports, Recommendations, & Guidelines
• AAPM:
– Report 46, Task Group 40, “Comprehensive QA for
Radiation Oncology” (1994)
– http://www.aapm.org/pubs/reports/RPT_46.PDF
6
Radiographic Simulators QA:
Recommended Frequency & Tolerances (AAPM)
essentially the
same as linac
7
AAPM Report 46, Task Group 40, “Comprehensive QA for Radiation
Oncology” (1994)
CT Simulators
External lasers for water bath for
marking origin/isocenter CT contrast injector thermoplastic masks
8
CT Simulation Process
9
Simulation Process at CT
• Patient aligned in
treatment position
• Immobilization
prepared
• Laser origin marked
(location of lasers
during CT)
• CT image acquired
• Isocenter marked
(optional)
• Patient setup
instructions recorded
10
CT Simulation Setup Examples:
laser location
Marked
(often fiducials placed for CT)
Immobilization
details noted
11
CT Simulator: Components
• Bore / CT apparatus
• Internal/external localizing lasers
• Patient support (flat table top to mimic treatment
table)
12
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
QA of CT Simulators:
Reports, Recommendations, & Guidelines
• AAPM:
– Report 46, Task Group 40, “Comprehensive QA for
Radiation Oncology” (1994)
• http://www.aapm.org/pubs/reports/RPT_46.PDF
– Report 83, Task Group 66, “Quality assurance for CT and
the CT simulation process” (2003)
• http://www.aapm.org/pubs/reports/RPT_83.pdf
– Report 39, Task Group 2, “Specification and Acceptance
Testing of Computed Tomography Scanners” (1993)
• http://aapm.org/pubs/reports/RPT_39.pdf
13
CT Simulators:
AAPM TG66 Recommended QA
• Radiation Safety: shielding Imaging tests:
survey • CT number accuracy
• Radiation dosimetry: CTDI • Image noise
• Lasers: alignment with imaging • In plane spatial integrity
planes • Field uniformity
• Tabletop: • Electron density to CT
– alignment with imaging planes conversion
– indexing & position
• Spatial resolution
• Gantry tilt accuracy
• Contrast resolution
• Scan localization
• CT dosimetry:
– dose from CT scan (CTDI)
– radiation profile width
– sensitivity profile width
• Generator tests
14
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: Laser Alignment
distance between external lasers and
the gantry (& imaging plane) is typically
• Three sets of lasers: fixed at 50cm
16
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: Laser Alignment
17
CT Simulator QA: Laser Motion
18
CT Simulator QA: Tabletop
20
CT Simulator QA: Gantry Tilt
• Scan range is
defined using a
scout image
• Verify actual
scanned volume
corresponds to
requested scan
volume
• Also verify radiation
& sensitivity profile
22
CT Simulator QA: Radiation Profile
23
CT Simulator QA: Sensitivity Profile
• evaluates “post-patient”
collimation (it is a function of
pre- and post- patient
collimation)
• defines actual width of imaged
slice
• excessive sensitivity profile
width can lead to loss of
resolution in longitudinal
direction
• excessively narrow sensitivity
profile can result in increased
quantum noise
• measurement:
– use inclined metal ramp
– length of ramp in image slice can
be used to calculate slice
thickness
24
Catphan manual
CT Simulator QA: Sensitivity Profile
25
Catphan 700 manual
CT Simulator QA: CTDI
Helical:
26
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: CTDI
• 2 phantoms utilized:
– head:
• length=15cm
• diameter=16cm
– body
• length=15cm
• diameter=32cm
– holes for chamber:
• central hole
• 4-8 periphery holes
27
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: CTDI
chamber exposure calibration factor (R/C)
28
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: Generator Tests
• Tests include:
– peak potential (kVp)
– half value layer (HVL)
– mAs linearity
– mAs reproducibility
– time accuracy
– (possible focal spot size)
• Measurement
preferences:
– Non-invasive
measurement preferred
– Performed with kV tube
“parked”
29
CT Simulators:
AAPM TG66 Recommended QA
Component
Frequency
Tolerance
Radiation safety survey
initially
regulatory limits
Patient dosimetry from CT annually & after component ±20% manufacturer
(CTDI)
replacement
specs
daily/monthly & after laser
Laser alignment
adjustment
±2mm
Table: orientation relative to
imaging plane
monthly & after laser adjustment
±2mm
Table: vertical & long. motion
monthly
±1mm
Table: indexing & position
annually
±1mm
Gantry tilt accuracy
annually
±1˚
Scan localization
annually
±1mm
Radiation profile width
annually
manufacturer specs
Sensitivity profile width
semi-annually
±1mm
initially & after component
Generator tests
replacement
manufacturer specs
30
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulators:
AAPM TG66 Recommended Imaging Tests
• CT number accuracy
• Image noise
• In plane spatial integrity
• Field uniformity
• Electron density to CT conversion
• Spatial resolution
• Contrast resolution
31
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator Imaging QA:
CT# accuracy & in high contrast
plane spatial integrity resolution
low contrast
resolution uniformity & noise
32
CT Simulators:
AAPM TG66 Recommended (Imaging) QA
33
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
MV Image Guidance:
MV (portal) imager
34
MV Image Guidance: 2D Imaging
Digitally Reconstructed
Radiograph (DRR) MV Projection Image
35
MV (Portal) Imaging
36
3D MV Imaging: MVCT (Tomotherapy)
37
MV CBCT (Linac): kV CT (diagnostic)
vs MV CBCT
kV CT (diagnostic) MV CBCT
5-15 cGy 38
kV Image Guidance:
39
kV Based IGRT
40
William Beaumont Hospital, 2002
Conventional CT
CBCT
~2 cGy
kV Image Guidance:
3D image registration
43
MV & kV Image Guidance Systems:
Reports, Recommendations, & Guidelines
• AAPM:
– Task Group 142, “Quality assurance of medical accelerators” (2009)
• http://www.aapm.org/pubs/reports/RPT_142.pdf
– Task Group 104, “The Role of In-Room kV X-Ray Imaging for Patient
Setup and Target Localization” (2009)
• http://www.aapm.org/pubs/reports/RPT_104.pdf
– Task Group 179, “QA for IGRT utilizing CT-based technologies” (2012)
• http://www.aapm.org/pubs/reports/RPT_179.pdf
– Task Group 58, “Clinical use of electronic portal imaging” (2001)
• http://www.aapm.org/pubs/reports/RPT_75.pdf
– Task Group 148, “QA for helical tomotherapy” (2010)
• http://www.aapm.org/pubs/reports/RPT_148.pdf
– Task Group 75, “Management of imaging dose during IGRT” (2007)
• http://www.aapm.org/pubs/reports/RPT_95.pdf
– Task Group 23, “The measurement, reporting, and management of
radiation dose in CT” (2008)
– Task Group 179, “QA for IGRT utilizing CT-based technologies” (2012)
• Islam et. al., “Patient dose from kV CBCT imaging in radiation
therapy” (2006)
– http://dx.doi.org/10.1118/1.2198169
44
3D (Tomographic) IGRT:
45
3D IGRT Geometric Calibration
geometric calibration of kV x-
ray imaging system relative to
MV
“flexing” of detector is
corrected as a function of
gantry angle
46
3D IGRT Geometric Calibration
residual flex
~0.25mm
47
Imaging Panel Calibration
48
3D CBCT:
panel shift to achieve larger field of view
49
3D CBCT: Bowtie Filter(s)
half fan full fan
50
kV/MV Image Guidance Routine QA:
52
kV/MV Image Guidance QA:
Imaging & Treatment Coordinate Coincidence
device for 2D imaging:
• Each imaging system
has its own
coordinate system
• Correlation with
delivery coordinate MV image:
system through a kV image:
calibration process
• Example: verify kV
isocenter coincidence
with MV isocenter
– image same setup
with MV & kV
device for
3D imaging:
53
Scaling
2D
• Ensure accurate
image scaling
• Performed by using
imaging software to
measure known
3D
distances in each
axis
• For 2D: distance from
source is important
due to magnification
54
2D IGRT QA:
Contrast & Spatial Resolution
kV phantom: MV phantom:
spatial resolution
55
2D IGRT QA:
Uniformity & Noise
kV uniformity MV uniformity
56
3D IGRT Image Quality Tests:
Similar to diagnostic CT
CT# accuracy: CT# accuracy:
diagnostic CT kV CBCT
58
3D IGRT Image Quality Tests:
Similar to diagnostic CT
low contrast resolution low contrast resolution
diagnostic CT kV CBCT
poorer low contrast resolution uniformity & noise uniformity & noise
for CBCT (expected) diagnostic CT kV CBCT
60
kV IGRT Imaging Dose & Beam Quality
61
kV/MV Image Guidance:
Recommendations for Daily QA
daily:
functionality &
geometric
accuracy
62
kV/MV Image Guidance:
Recommendations for Monthly QA
monthly: geometric
+ image quality
63
kV/MV Image Guidance:
Recommendations for Annual QA
annual: geometry,
imaging dose,
beam quality
64
Recommended QA & Tolerances
65
Patient Support Systems:
QA
• Geometric:
– accuracy of table index, &
patient re-positioning
– accuracy of couch angle
– pitch & roll accuracy (for 6-
degree capable tables)
• Dosimetric:
– increased skin dose
– reduced tumor dose
– altered dose distribution
66
Weight Limits:
Bore
Weight Limit
Diameter Image FOV (cm)!
# (Lbs)!
(cm)!
Exact couch# 500#
Truebeam 500 (440 end)
Brainlab Robotic Couch# 275#
Brainlab Non-Robotic# 350#
TrueBeam STX (Perfect Pitch) 440
Linac (CBCT)# 440# 45 or 25 (~ 15 in S/I)#
Simulator# 600#
CT Simulator (GE OptimaCT580RT)# 500# 80# 65#
CT Simulator (Siemens)# 660# 78# 50 (78 ext. recon)#
PET/CT Simulator (Siemens)# 500# 78# 50 (78 ext. recon)#
CT Simulator (Phillips)# 650# 85# 60 (70 ext. recon)#
MR (GE)# 350# 60# 48#
MRI (GE OptimaMR45W)# 500# 70# 50#
Know your machine limits, & keep available for reference
67
Couchtop Dosimetric Considerations:
Increased Skin Dose
68
AAPM Task Group 176, “Dosimetric effects caused by couch tops and immobilization devices” (2014)
Couchtop Dosimetric Considerations:
Attenuation (at Depth)
attenuation can be
considerable even for
carbon fiber couches (this
one has a foam core)
the amount of
attenuation can vary
based on geometry
69
Patient Support Systems:
Couch Attenuation
70
Patient Support Systems:
Couch Attenuation
• Some tables
have adjustable
support bars with
high attenuation!
• Take care to
make sure the
beam doesn’t
enter through
them
71
AAPM Task Group 176, “Dosimetric effects caused by couch tops and immobilization devices” (2014)
Immobilization Devices:
Dosimetric Considerations
• Can be accounted for manually (using attenuation
factor) or within the planning system
• Within TPS:
– include in CT at time of simulation
• Measurements:
– attenuation point measurements: ion chamber at depth in
phantom
– surface dose measurements:
• extrapolation chamber
• plane parallel chamber
• OSL/TLD
• Film
72
Immobilization Devices:
Measured Transmission Factors:
73
Immobilization Devices
74
AAPM Task Group 176, “Dosimetric effects caused by couch tops and immobilization devices” (2014)
Immobilization Devices
• Avoid entrance
through devices
when possible
75
AAPM Task Group 176, “Dosimetric effects caused by couch tops and immobilization devices” (2014)
THANK YOU
76
Additional Notes:
77
x-ray off (table translation)
4DCT
end expiration
end inspiration
78
4DCT Breathing Signal
79
4DCT Reconstruction
80
4DCT
3D Free
Breathing (FB) 4DCT AIP 4DCT MIP
81
Gating:
82