0% found this document useful (0 votes)
151 views82 pages

Quality Assurance in Radiation Therapy: Simulators & Ancillary Equipment

The document discusses quality assurance for various medical imaging and treatment planning devices used in radiation therapy, including radiographic simulators, CT simulators, and image guidance systems. It covers the components, purpose, and recommended quality assurance procedures and guidelines for these devices as outlined in reports from the American Association of Physicists in Medicine.

Uploaded by

nefzi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
151 views82 pages

Quality Assurance in Radiation Therapy: Simulators & Ancillary Equipment

The document discusses quality assurance for various medical imaging and treatment planning devices used in radiation therapy, including radiographic simulators, CT simulators, and image guidance systems. It covers the components, purpose, and recommended quality assurance procedures and guidelines for these devices as outlined in reports from the American Association of Physicists in Medicine.

Uploaded by

nefzi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 82

Quality Assurance in Radiation Therapy:

Simulators & Ancillary Equipment

ICPT School on Medical Physics for Radiation Therapy


Justus Adamson PhD
Assistant Professor
Department of Radiation Oncology
Duke University Medical Center
Topics

Overview & Quality Assurance of:


•  Radiographic Simulators
•  CT-Simulators
•  MV image guidance
•  kV image guidance
•  Patient support systems
•  Patient immobilization devices

2
Radiographic Simulators

field light X-ray source with


adjustable collimation &
source-to-axis distance

rotating
gantry
localizing laser
(not shown)

Imager with adj.


distance to axis

patient support
assembly

3
Radiographic Simulators
X-ray source with CBCT capability
adjustable collimation

field light
rotating
gantry

patient support
assembly

Imager with adj.


distance to axis
localizing laser
(not shown)
4
Radiographic Simulators:
Components & Purpose
•  Components:
–  Imaging source & detector
–  Localizing lasers
–  Optical distance indicator
–  Field light
–  Patient support assembly
•  Purpose: to reproduce the geometric conditions of
the radiation therapy equipment
–  Should be subject to the same mechanical checks as linear
accelerators
–  Image quality should be checked following guidelines for
diagnostic radiography

5
Radiographic Simulators QA:
Reports, Recommendations, & Guidelines
•  AAPM:
–  Report 46, Task Group 40, “Comprehensive QA for
Radiation Oncology” (1994)
–  http://www.aapm.org/pubs/reports/RPT_46.PDF

6
Radiographic Simulators QA:
Recommended Frequency & Tolerances (AAPM)

essentially the
same as linac

7
AAPM Report 46, Task Group 40, “Comprehensive QA for Radiation
Oncology” (1994)
CT Simulators
External lasers for water bath for
marking origin/isocenter CT contrast injector thermoplastic masks

CT control room CT bore Internal CT lasers flat patient


support table

8
CT Simulation Process

1.  CT Scan, patient positioning, immobilization


1.  similar to diagnostic CT
2.  added requirements of localization (lasers) &
immobilization
2.  Treatment planning & CT simulation
1.  performed in treatment planning system: contouring,
isocenter placement, selection of treatment geometry,
documentation
3.  Treatment setup
1.  setup at machine according to instructions from CT-
simulation & treatment planning

9
Simulation Process at CT

•  Patient aligned in
treatment position
•  Immobilization
prepared
•  Laser origin marked
(location of lasers
during CT)
•  CT image acquired
•  Isocenter marked
(optional)
•  Patient setup
instructions recorded

10
CT Simulation Setup Examples:

laser location
Marked
(often fiducials placed for CT)
Immobilization
details noted

11
CT Simulator: Components

•  Bore / CT apparatus
•  Internal/external localizing lasers
•  Patient support (flat table top to mimic treatment
table)

12
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
QA of CT Simulators:
Reports, Recommendations, & Guidelines
•  AAPM:
–  Report 46, Task Group 40, “Comprehensive QA for
Radiation Oncology” (1994)
•  http://www.aapm.org/pubs/reports/RPT_46.PDF
–  Report 83, Task Group 66, “Quality assurance for CT and
the CT simulation process” (2003)
•  http://www.aapm.org/pubs/reports/RPT_83.pdf
–  Report 39, Task Group 2, “Specification and Acceptance
Testing of Computed Tomography Scanners” (1993)
•  http://aapm.org/pubs/reports/RPT_39.pdf

13
CT Simulators:
AAPM TG66 Recommended QA
•  Radiation Safety: shielding Imaging tests:
survey •  CT number accuracy
•  Radiation dosimetry: CTDI •  Image noise
•  Lasers: alignment with imaging •  In plane spatial integrity
planes •  Field uniformity
•  Tabletop: •  Electron density to CT
–  alignment with imaging planes conversion
–  indexing & position
•  Spatial resolution
•  Gantry tilt accuracy
•  Contrast resolution
•  Scan localization
•  CT dosimetry:
–  dose from CT scan (CTDI)
–  radiation profile width
–  sensitivity profile width
•  Generator tests

14
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: Laser Alignment
distance between external lasers and
the gantry (& imaging plane) is typically
•  Three sets of lasers: fixed at 50cm

wall mounted overhead sagittal


gantry lasers lasers (external) laser (external)

gantry lasers should accurately external lasers should be


identify scan plane accurately spaced from imaging
& should be parallel and plane
orthogonal with scan plane & & should be parallel and
orthogonal with scan plane &
intersect center
intersect at a point co-incident
15
with center
CT Simulator QA: Laser Alignment

16
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: Laser Alignment

17
CT Simulator QA: Laser Motion

•  External lasers can often be


shifted to mark a new isocenter
after CT
•  Laser motion should be
accurate, linear, & reproducible

18
CT Simulator QA: Tabletop

•  (often) has ability to


“register” immobilization
devices at specified
positions
•  Necessary criteria:
–  should mimic treatment
table
–  flat, level, orthogonal
–  similar sag properties
–  motion indicators & table
position should be
accurate & reproducible
interlocking immobilization devices 19
CT Simulator QA: Tabletop

ensure table is level


relative to both gravity &
imaging plane

verify accuracy of longitudinal


& vertical table motion

20
CT Simulator QA: Gantry Tilt

•  Many CT gantries can be


tilted for diagnostic scans
–  not typically used for CT
simulations
•  Accuracy of tilt (especially
at 0˚) should be verified
–  TG66 recommendation:
verify with film
•  Suggestion: mark the
external laser position on
the scanner with the
gantry level
21
CT Simulator QA: Scan Localization
(from Scout Image)

•  Scan range is
defined using a
scout image
•  Verify actual
scanned volume
corresponds to
requested scan
volume
•  Also verify radiation
& sensitivity profile
22
CT Simulator QA: Radiation Profile

•  evaluates “pre-patient” collimation


•  dose profile prior to detector collimation
•  excessively wide radiation profile can result in
unnecessary patient dose
•  excessively narrow radiation profile can result in
increased quantum noise
•  measurement: full width at half maximum of
exposed film (measured for each slice thickness)

23
CT Simulator QA: Sensitivity Profile

•  evaluates “post-patient”
collimation (it is a function of
pre- and post- patient
collimation)
•  defines actual width of imaged
slice
•  excessive sensitivity profile
width can lead to loss of
resolution in longitudinal
direction
•  excessively narrow sensitivity
profile can result in increased
quantum noise
•  measurement:
–  use inclined metal ramp
–  length of ramp in image slice can
be used to calculate slice
thickness

24
Catphan manual
CT Simulator QA: Sensitivity Profile

25
Catphan 700 manual
CT Simulator QA: CTDI

•  CTDI: Integrated dose


(along z-axis) from one
CTDIw = 2/3 CTDI100(surface) + 1/3
axial CT slice CTDI100(center)
•  CTDI100mm is what is
measured in practice N=number of simultaneous axial
•  CTDI is usually ~2x higher scans per rotation (multi-slice CT)
at the surface Axial:
•  Dose Length Product
(DLP): defines total energy
absorbed by a scanned
volume: I=table increment T=thickness of
per axial scan one axial scan

Helical:

26
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: CTDI

•  2 phantoms utilized:
–  head:
•  length=15cm
•  diameter=16cm
–  body
•  length=15cm
•  diameter=32cm
–  holes for chamber:
•  central hole
•  4-8 periphery holes

27
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: CTDI
chamber exposure calibration factor (R/C)

converts exposure in air to


electrometer correction (C/Rdg) absorbed dose in medium. (0.94
cGy/R for muscle @ Eeff=70keV)
temperature & pressure
correction

28
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator QA: Generator Tests

•  Tests include:
–  peak potential (kVp)
–  half value layer (HVL)
–  mAs linearity
–  mAs reproducibility
–  time accuracy
–  (possible focal spot size)
•  Measurement
preferences:
–  Non-invasive
measurement preferred
–  Performed with kV tube
“parked”

29
CT Simulators:
AAPM TG66 Recommended QA
Component   Frequency   Tolerance  
Radiation safety survey   initially   regulatory limits  
Patient dosimetry from CT annually & after component ±20% manufacturer
(CTDI)   replacement   specs  
daily/monthly & after laser
Laser alignment   adjustment   ±2mm  
Table: orientation relative to
imaging plane   monthly & after laser adjustment  ±2mm  
Table: vertical & long. motion   monthly   ±1mm  
Table: indexing & position   annually   ±1mm  
Gantry tilt accuracy   annually   ±1˚  
Scan localization   annually   ±1mm  
Radiation profile width   annually   manufacturer specs  
Sensitivity profile width   semi-annually   ±1mm  
initially & after component
Generator tests   replacement   manufacturer specs  
30
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulators:
AAPM TG66 Recommended Imaging Tests
•  CT number accuracy
•  Image noise
•  In plane spatial integrity
•  Field uniformity
•  Electron density to CT conversion
•  Spatial resolution
•  Contrast resolution

31
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
CT Simulator Imaging QA:
CT# accuracy & in high contrast
plane spatial integrity resolution

low contrast
resolution uniformity & noise

32
CT Simulators:
AAPM TG66 Recommended (Imaging) QA

Imaging Test   Frequency   Tolerance  


daily / monthly / annually
CT number accuracy   (less to more comprehensive)  0 ± 5 HU for water  
Image noise   daily   manufacturer specs  
In plane spatial integrity   daily / monthly   ±1mm  
monthly (most common kVp),
Field uniformity   annually all kVps   within ±5HU  
Electron density to CT
number conversion   annually & after calibration   consistent with baseline  
Spatial resolution   annually   manufacturer specs  
Contrast resolution   annually   manufacturer specs  

33
AAPM Report 83, Task Group 66, “Quality assurance for computed-tomography simulators and the
computed tomography simulation process” (2003)
MV Image Guidance:

MV (portal) imager
34
MV Image Guidance: 2D Imaging

Digitally Reconstructed
Radiograph (DRR) MV Projection Image

35
MV (Portal) Imaging

Why do megavoltage portal images have such poor


contrast compared to diagnostic images?

•  Compton effect has weak Z


dependence, very little
differential absorption
•  diagnostic: photoelectric
dominates
•  MV: Compton dominates
•  Scattered photons +
secondary electrons -> not
easily removed
•  Large penumbra: geometric +
phantom scatter

36
3D MV Imaging: MVCT (Tomotherapy)

37
MV CBCT (Linac): kV CT (diagnostic)
vs MV CBCT
kV CT (diagnostic) MV CBCT

5-15 cGy 38
kV Image Guidance:

39
kV Based IGRT

40
William Beaumont Hospital, 2002
Conventional CT

CBCT

~2 cGy
kV Image Guidance:
3D image registration

43
MV & kV Image Guidance Systems:
Reports, Recommendations, & Guidelines
•  AAPM:
–  Task Group 142, “Quality assurance of medical accelerators” (2009)
•  http://www.aapm.org/pubs/reports/RPT_142.pdf
–  Task Group 104, “The Role of In-Room kV X-Ray Imaging for Patient
Setup and Target Localization” (2009)
•  http://www.aapm.org/pubs/reports/RPT_104.pdf
–  Task Group 179, “QA for IGRT utilizing CT-based technologies” (2012)
•  http://www.aapm.org/pubs/reports/RPT_179.pdf
–  Task Group 58, “Clinical use of electronic portal imaging” (2001)
•  http://www.aapm.org/pubs/reports/RPT_75.pdf
–  Task Group 148, “QA for helical tomotherapy” (2010)
•  http://www.aapm.org/pubs/reports/RPT_148.pdf
–  Task Group 75, “Management of imaging dose during IGRT” (2007)
•  http://www.aapm.org/pubs/reports/RPT_95.pdf
–  Task Group 23, “The measurement, reporting, and management of
radiation dose in CT” (2008)
–  Task Group 179, “QA for IGRT utilizing CT-based technologies” (2012)
•  Islam et. al., “Patient dose from kV CBCT imaging in radiation
therapy” (2006)
–  http://dx.doi.org/10.1118/1.2198169
44
3D (Tomographic) IGRT:

45
3D IGRT Geometric Calibration

geometric calibration of kV x-
ray imaging system relative to
MV

“flexing” of detector is
corrected as a function of
gantry angle

46
3D IGRT Geometric Calibration

daily check of geometric accuracy


absolute flex

residual flex
~0.25mm

47
Imaging Panel Calibration

48
3D CBCT:
panel shift to achieve larger field of view

49
3D CBCT: Bowtie Filter(s)
half fan full fan

50
kV/MV Image Guidance Routine QA:

Planar (2D) Imaging: 3D Imaging


•  Collision interlocks •  Collision interlocks
•  Positioning / Repositioning •  Positioning / Repositioning
•  Imaging & Treatment •  Imaging & Treatment
Coordinate Coincidence Coordinate Coincidence
•  Scaling •  Geometric Distortion
•  Spatial Resolution •  Spatial Resolution
•  Contrast •  Contrast
•  Uniformity & Noise •  HU Constancy
•  Imaging Dose •  Uniformity & Noise
•  Beam quality / energy (kV) •  Imaging Dose
51
AAPM Task Group 142, “Quality Assurance of Medical Accelerators” (2009)
kV/MV Image Guidance QA:
Positioning / Repositioning
•  More important than
image quality in
IGRT setting!
•  Basic functionality
test for image
guidance
–  setup phantom
–  image
–  shift based on
imaging
–  verify shift

52
kV/MV Image Guidance QA:
Imaging & Treatment Coordinate Coincidence
device for 2D imaging:
•  Each imaging system
has its own
coordinate system
•  Correlation with
delivery coordinate MV image:
system through a kV image:
calibration process
•  Example: verify kV
isocenter coincidence
with MV isocenter
–  image same setup
with MV & kV
device for
3D imaging:
53
Scaling
2D

•  Ensure accurate
image scaling
•  Performed by using
imaging software to
measure known
3D
distances in each
axis
•  For 2D: distance from
source is important
due to magnification

54
2D IGRT QA:
Contrast & Spatial Resolution

kV phantom: MV phantom:

spatial resolution

contrast resolution orientation of


phantom relative
to detector array
matters!

55
2D IGRT QA:
Uniformity & Noise

kV uniformity MV uniformity

56
3D IGRT Image Quality Tests:
Similar to diagnostic CT
CT# accuracy: CT# accuracy:
diagnostic CT kV CBCT

CT#s are relative for


Often CT#s are calibrated
CBCT due to large
using a phantom scan to
proportion of scatter in
match the expected values.
projection images
Accurate CT# in phantom may
not translate to accurate CT#s
in a patient!
57
3D IGRT Image Quality Tests:
Similar to diagnostic CT
high contrast resolution: high contrast resolution:
diagnostic CT kV CBCT

high contrast resolution will


be dependent on the imaging
protocol

58
3D IGRT Image Quality Tests:
Similar to diagnostic CT
low contrast resolution low contrast resolution
diagnostic CT kV CBCT

poorer low contrast resolution uniformity & noise uniformity & noise
for CBCT (expected) diagnostic CT kV CBCT

CT#s monitored relative to a


baseline (due to expected
differences from truth)
59
MV IGRT Imaging Dose:

•  Imaging is done with the treatment beam hence


dose can be directly calculated using treatment
planning system / hand calculation
•  Exception: some linear accelerators have a lower
energy (1MV / 2.5MV) used only for imaging

60
kV IGRT Imaging Dose & Beam Quality

•  2D: many meters are •  3D:


available to measure –  Farmer chamber in
kVp, HVL, exposure to cylindrical phantom
skin, etc. (CTDI phantom)
–  Nk calibration factor

61
kV/MV Image Guidance:
Recommendations for Daily QA

daily:
functionality &
geometric
accuracy

62
kV/MV Image Guidance:
Recommendations for Monthly QA

monthly: geometric
+ image quality

63
kV/MV Image Guidance:
Recommendations for Annual QA

annual: geometry,
imaging dose,
beam quality

64
Recommended QA & Tolerances

65
Patient Support Systems:

QA
•  Geometric:
–  accuracy of table index, &
patient re-positioning
–  accuracy of couch angle
–  pitch & roll accuracy (for 6-
degree capable tables)
•  Dosimetric:
–  increased skin dose
–  reduced tumor dose
–  altered dose distribution

66
Weight Limits:

Bore

 Weight Limit
Diameter Image FOV (cm)!
# (Lbs)!
(cm)!
Exact couch# 500#
Truebeam 500 (440 end)
Brainlab Robotic Couch# 275#
Brainlab Non-Robotic# 350#
TrueBeam STX (Perfect Pitch) 440
Linac (CBCT)# 440# 45 or 25 (~ 15 in S/I)#
Simulator# 600#
CT Simulator (GE OptimaCT580RT)# 500# 80# 65#
CT Simulator (Siemens)# 660# 78# 50 (78 ext. recon)#
PET/CT Simulator (Siemens)# 500# 78# 50 (78 ext. recon)#
CT Simulator (Phillips)# 650# 85# 60 (70 ext. recon)#
MR (GE)# 350# 60# 48#
MRI (GE OptimaMR45W)# 500# 70# 50#
Know your machine limits, & keep available for reference
67
Couchtop Dosimetric Considerations:
Increased Skin Dose

increased surface dose

68
AAPM Task Group 176, “Dosimetric effects caused by couch tops and immobilization devices” (2014)
Couchtop Dosimetric Considerations:
Attenuation (at Depth)

attenuation can be
considerable even for
carbon fiber couches (this
one has a foam core)

the amount of
attenuation can vary
based on geometry

69
Patient Support Systems:
Couch Attenuation

•  Couch attenuation can be managed by:


–  using transmission factor in hand calculation
–  modeling the couch in the treatment planning
system
–  do nothing (when attenuation is negligible)

couch models in the treatment planning system


HU values
can be
specified

70
Patient Support Systems:
Couch Attenuation

•  Some tables
have adjustable
support bars with
high attenuation!
•  Take care to
make sure the
beam doesn’t
enter through
them

71
AAPM Task Group 176, “Dosimetric effects caused by couch tops and immobilization devices” (2014)
Immobilization Devices:
Dosimetric Considerations
•  Can be accounted for manually (using attenuation
factor) or within the planning system
•  Within TPS:
–  include in CT at time of simulation
•  Measurements:
–  attenuation point measurements: ion chamber at depth in
phantom
–  surface dose measurements:
•  extrapolation chamber
•  plane parallel chamber
•  OSL/TLD
•  Film

72
Immobilization Devices:
Measured Transmission Factors:

73
Immobilization Devices

74
AAPM Task Group 176, “Dosimetric effects caused by couch tops and immobilization devices” (2014)
Immobilization Devices

•  Avoid entrance
through devices
when possible

75
AAPM Task Group 176, “Dosimetric effects caused by couch tops and immobilization devices” (2014)
THANK YOU

76
Additional Notes:

•  CT Simulators often include tools for motion


management
–  4DCT
•  MIP
•  AIP
•  Min-IP
–  Breath hold CT

77
x-ray off (table translation)

4DCT

end expiration

end inspiration

78
4DCT Breathing Signal

79
4DCT Reconstruction

Maximum Average Minimum


Intensity Intensity Intensity
Individual Projection Projection Projection
phases (MIP) (AIP) (MinIP) Color AIP

80
4DCT

3D Free
Breathing (FB) 4DCT AIP 4DCT MIP

81
Gating:

82

You might also like