Performance Outline: EC4-250-NO Nitric Oxide Electrochemical Sensor

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

EC4-250-NO

Nitric Oxide Electrochemical Sensor

PERFORMANCE OUTLINE
Range ................................................................. 0 - 250 ppm
Maximum overload................................................. 1000 ppm
Sensitivity............................................ 320 - 480 nA/ppm NO
Zero in air at 20 °C...............................-0.06 to +4.5 ppm NO
Resolution...........................................................0.5 ppm NO
Linearity ...................................................linear across range
Response time, t90 ....................................................... <35 s
Long-term output drift.................................<2% signal/month
Warranty .................................................................... 2 years

OPERATING CONDITIONS
Temperature range ...........................................-20 to +50 °C
Operating humidity................15 - 90% RH (non-condensing)
Pressure range .................................. 90 - 110 kPa (nominal)
Operating circuit............... see Electrochemical Toxic Sensor
Application Note
Applied potential .....................................................+300 mV
Storage life........... 6 months in original packaging (0 - 25 °C)

PHYSICAL CHARACTERISTICS
Label colour ...................................................................silver
Weight.................................................................. 5 g approx.
Dimensions ...........................................................see outline
This device is designed to be RoHS compliant.

CROSS-SENSITIVITY DATA
GAS Concentration Sensor
(ppm) Response
(As ppm NO)
Sulfur Dioxide 5 0
Nitrogen Dioxide 5 1.5
Carbon Monoxide 300 0
Hydrogen 15 -1.5

Important Note
Do not solder to the connector pins as this may damage the
sensor and thereby invalidate the warranty.
Details on recommended connector pins can be found in the
Frequently Asked Questions within the Gas Sensor section
of the SGX website.

SGX Sensortech(IS) Ltd Registered in England No.08067077 A1A-764713 Version 3, October 2009
TEMPERATURE DEPENDENCE
EC4-250-NO Graph of Zero vs Temperature

12

10

8
Zero / ppm NO

6
Average
+95% conf
-95% conf.
4

0
-30 -20 -10 0 10 20 30 40 50 60

-2
Temperature / Degrees Celsius

EC4-250-NO Graph of Sensitivity vs Temperature

140

120
% of Sensitivty ar 20 Deg C

100

80
Average
+95% conf
-95% conf.
60

40

20

0
-30 -20 -10 0 10 20 30 40 50 60
Temperature / Degrees Celsius

SGX Sensortech(IS) Ltd Registered in England No.08067077 A1A-764713 Version 3, October 2009

You might also like