Adventurer+ Analytical and Precision Balances Service Manual
Adventurer+ Analytical and Precision Balances Service Manual
ADVENTURER+
ANALYTICAL AND
PRECISION BALANCES
SERVICE MANUAL
Ohaus Corporation 19A Chapin Road, P.O. Box 2033, Pine Brook, NJ 07058-2033 (973) 377-9000
SERVICE MANUAL
Adventurer Balances
Ohaus Corporation
19A Chapin Road
P.O. Box 2033
Pine Brook, NJ 07058-2033
The information contained in this manual is believed to be accurate at the time of publication, but Ohaus
Corporation assumes no liability arising from the use or misuse of this material. Reproduction of this
material is strictly prohibited.
OHAUS
DOCUMENT NUMBER
83020790
THIS DOCUMENT CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION OF
PART NUMBER
OHAUS CORP. THE CONTENTS HEREOF MAY NOT BE USED NOR REPRODUCED IN
TITLE
RECORD OF CHANGES
REV. DATE ECN BY DESCRIPTION
A 08/04/09 N/A NO INITIAL DRAFT
TABLE OF CONTENTS
1 INTRODUCTION ..................................................................................................................................... 6
1.1 INTRODUCTION .............................................................................................................................. 6
1.2 SERVICE FACILITIES...................................................................................................................... 6
1.3 TOOLS AND TEST EQUIPMENT REQUIRED ................................................................................ 7
1.3.1 Special Tools ............................................................................................................................ 7
1.3.2 Standard Tools and Test Equipment...................................................................................... 7
1.4 TEST MASSES REQUIRED............................................................................................................. 7
1.5 SPECIFICATIONS ............................................................................................................................ 8
2 THEORY OF OPERATION ..................................................................................................................... 9
2.1 BALANCE OPERATION .................................................................................................................. 9
2.2 OVERVIEW OF THE CONTROLS ................................................................................................... 9
2.3 OVERVIEW OF THE DISPLAY INDICATORS............................................................................... 10
2.4 TURNING ON THE BALANCE....................................................................................................... 10
2.5 MENU SETUP ................................................................................................................................ 11
2.6 TRANSDUCER DESCRIPTION ..................................................................................................... 12
2.7 BLOCK DIAGRAM CIRCUIT DESCRIPTION ................................................................................ 12
2.7.1 Main PC Board........................................................................................................................ 12
2.7.2 Cell PC Board ......................................................................................................................... 13
2.7.3 RS232 Board........................................................................................................................... 13
3 TROUBLESHOOTING .......................................................................................................................... 15
3.1 TROUBLESHOOTING ................................................................................................................... 15
3.2 DIAGNOSTIC GUIDE ..................................................................................................................... 15
3.2.1 Diagnosis ................................................................................................................................ 15
3.3 ERROR CODES ............................................................................................................................. 16
4 MAINTENANCE PROCEDURES.......................................................................................................... 17
4.1 PREVENTIVE MAINTENANCE...................................................................................................... 17
4.1.1 Preventive Maintenance Checklist ....................................................................................... 17
4.2 TESTING ........................................................................................................................................ 17
4.2.1 Operational Test..................................................................................................................... 17
Segment Display Test ......................................................................................................................... 17
4.2.2 Calibration .............................................................................................................................. 18
4.2.3 Performance Tests - Precision and Analytical Balances ................................................... 20
4.3 DISASSEMBLY/ASSEMBLY PROCEDURES............................................................................... 27
4.3.1 Disassembling the Balance................................................................................................... 27
4.3.2 Re-assembling the Balance .................................................................................................. 28
4.4 REPLACEMENT OF MAJOR COMPONENTS .............................................................................. 29
4.4.1 Transducer Removal - Precision .......................................................................................... 29
4.4.2 Transducer Disassembly/Cleaning/Alignment - Precision ................................................ 29
4.4.3 Transducer Removal - Analytical ......................................................................................... 34
4.4.4 Main Printed Circuit Board (PCB) Replacement - Precision and Analytical Balances .... 36
4.4.5 RS232 Printed Circuit Board Replacement - Precision and Analytical Balances ............ 36
4.4.6 Membrane Switch Replacement - Precision and Analytical Balances.............................. 36
5 AR Series Service Tool ....................................................................................................................... 38
5.1 Run ARServiceToolDealer.exe .................................................................................................... 38
1 INTRODUCTION
1.1 INTRODUCTION
This service manual contains the information needed to perform routine maintenance and service on the Ohaus
Adventurer+ Analytical and Precision balances. Balances are rated from 65 grams to 4200 grams capacity.
The content of this manual is contained in five chapters.
Chapter 1 Introduction - Contains information regarding service facilities, tools and test equipment,
measuring masses, and specifications.
Chapter 2 Theory of Operation - Contains information on the mechanical and electronic functions of the
balance.
Chapter 5 AR Series Service Tool – Describe how to use the service tool.
Chapter 6 Drawings and Parts Lists - Contains exploded views of Adventurer balances and Transducers
identifying all serviceable components with parts lists.
Before servicing the balance, you should be familiar with the Instruction Manual which is packed with every
balance.
1.5 SPECIFICATIONS
Complete specification for the Ohaus adventurer balances are listed in Table 1-2. When a balance has been
serviced, it must meet the specifications listed in the table. Before servicing the balance, determine what
specifications are not met.
2 THEORY OF OPERATION
Exploded view drawings are included in chapter 5 which identifies all mechanical components of the balance. The
descriptions in this manual refer to the name of the components identified on the exploded view drawing with the
numbered callout.
With power applied, press , all segments will appear briefly followed by a software revision number (when
plugged in first time) or after a power interruption and then * 0.00g. Allow 20 minutes warm-up time. When the
balance is first turned on, it can be used to weigh in grams or tare items without setting the menus.
Leveling
5 Indicates leveling position of the balance
indicator
6 Leveling feet Used to level the balance
Power ON
Press , all segments will appear briefly followed by a software revision number (when plugged in first time)
or after a power interruption and then * 0.00g. Allow 20 minutes warm-up time. When the balance is first turned on,
it can be used to weigh in grams or tare items without setting the menus.
Power OFF
To turn the balance OFF, press and hold until the display indicates OFF then release.
The setup menu permits you to activate available weighing units, parts counting, linearity calibration, keystroke
tone and print parameters. To enter the menu, press and hold . To advance through the menus,
press . To enter a selected menu, press . After you make selections in any menu, you may
continue to make additional changes in other menus before exiting and saving. To exit a menu and return to
The mechanical zero of the system is moved to the approximate middle of the measuring range by the Counter
Weight. Displacement of the Position Sensor Vane from the normal position results in a current change in the
Force Coils. The Force Coils are located in the field of the permanent Magnets which are mounted on the
Transducer Base. This means that with a 50% load on the Pan, the Ratio Beam is in equilibrium and the sum of
the forces from the Force Coils is zero. The temperature sensitivity of the Permanent magnets is compensated in
the electronics by a Temperature Sensor which is physically mounted on the magnet assembly.
provides 12Volts dc. The main PC board contains a diode bridge and 4 voltage regulators which provide
regulated voltages. The 3.3 Volts is provided to the digital circuitry and the 10 Volts dc and -12 Volts are for
operating the balance, and 5 Volts is for backlit.
MELSI2
MELSI2 is a digital integrated circuit that contains the digital circuitry needed to operate a two-channel charge
balance deltasigma A/D converter.
Microprocessing Unit
The LPC2138 microcontrollers is based on a 32/16 bit ARM7TDMI-S™ CPU with real-time emulation and
embedded trace support, that combines the microcontroller with 512 kB of embedded high speed Flash memory.
A 128-bit wide memory interface and an unique accelerator architecture enable 32-bit code execution at
maximum clock rate.
RTC unit
The LPC2138 include the RTC function. A rechargeable circuitry is designed for the RTC as power supply.
Parameter Storage
The EEPROM is a 2048 x 8 bit electrically erasable and programmable memory which stores some information of
production.
TXD, DTR and DSR are provided and go to a driver/receiver IC located on the RS232 board.
3 TROUBLESHOOTING
3.1 TROUBLESHOOTING
This section of the manual specifies problem areas of the balance which can occur. information is contained to
isolate specific problems using Table 3-1, Diagnostic Guide, and Table 3-2, Error Codes. Follow all directions
step by step. Make certain that the work area is clean and use care when handling components of the balance.
3.2.1 Diagnosis
1. Isolate and identify the symptom.
2. Refer to Table 3-1 Diagnostic guide and locate the symptom.
3. Follow the suggested remedies in the order that they appear.
4. Perform the indicated checks, or see the appropriate section of the manual.
5. Repair or replace the defective section of the balance.
NOTE:
If more than one symptom is observed, it is necessary to approach one
area at a time, and also remember that the symptoms may be interrelated.
In the event that erratic or fluctuating weight readings are observed, it is necessary to isolate the problem to either
the mechanical area or the electronic area of the balance.
Ohaus Corporation
19A Chapin Road
P.O. Box 2033
Pine Brook, NJ 07058-2033 USA
Tel: 973-377-9000
Fax: 973-593-0359
In the United States call toll free, 800-526-0659 between 8:00 a.m. and 6:00 p.m. EST.
Error 8.4 Weight reading below underload limit. Put platform on balance.
Error 9.5 Calibration information is error Send the balance back to Ohaus and repaired.
Repeat to connect or disconnect power.
Error 5.3 Can not read EEPROM datum on loadcell.
If can't be remedy, send back to Ohaus.
4 MAINTENANCE PROCEDURES
CAUTION
DO NOT USE CHEMICAL CLEANERS OR SOLVENTS OF ANY TYPE.
SOME CLEANERS ARE ABRASIVE AND MAY AFFECT THE FINISH OF
THE BALANCE.
4.2 TESTING
Before servicing the Adventurer balance, an operational test and various performance tests should be made to
ascertain whether or not the balance meets specifications. Turn the balance on and allow it warm up for at least
one hour before performing these tests. Make sure the test area is free from drafts and the surface that the
balance rests on is level and vibration free. The masses used for the performance tests and adjustments must
meet or exceed ASTM Class 1 Tolerance.
4.2.1 Operational Test
1. Connect a functioning Power Adapter to the balance power receptacle assembly located at the rear of the
balance.
2. Plug the Power Cord into a suitable power source.
Segment Display Test
Turn the balance on, all segments are enabled and displayed breifly, then the model number of the balance
followed by a software revision number. This is a segment display test. Figure 4-1 is a full display test.
1. With the balance turned ON, press and hold until MENU is displayed in the LCD main area
(CALIBRATION is display in auxiliary area). Then release the switch.
2. Press , CAL is displayed in the main area (SPAN is displayed in auxiliary area), then press
again, --0-- momentarily displayed in main area (BUSY is display in auxiliary area).
Do not disturb the balance when –0-- is displayed. Incorrect calibration may result.
3. After several seconds, the value of span calibration mass is displayed in the main area (ADD WEIGHT is
displayed in the auxiliary area). Place the indicated mass on the pan. BUSY is displayed and flickered in the
auxiliary area.
4. After several seconds, --0-- is displayed in the main area (CLEANPAN is displayed in the auxiliary area). It
means remove the mass.
5. Remove the calibration mass from the pan. BUSY is momentarily displayed and CAL DONE is displayed in
the auxiliary area. Then the balance returns to the weighing mode.
1. With the balance turned ON, press and hold until MENU is displayed in the LCD main area
(CALIBRATION is display in auxiliary area). Then release the switch.
2. Press , CAL is displayed in the main area (SPAN is displayed in auxiliary area), then
The following performance tests are used to evaluate the balance operation before and after repairs. Each
balance tested must meet the requirements specified in each test as well as the specifications listed in Table 1-2
depending upon the model. Before proceeding with the following tests, all the procedures starting with paragraph
4.2 must have been accomplished on the Balance first.
5B. An error in the Off-Center Load Adjustment can affect the results of the Precision Test. Inspect and
correct if necessary. See Off-Center Load Test.
Platform. Press the button to return the reading to zero. Move the mass halfway to the rear of the
Platform and note the reading. Move the mass halfway between the center and the left edge front of the Platform
and note the reading. Repeat this test for the right edge position and note the reading. Move the mass halfway to
the front of the Platform and note the reading. Note any differences in the displayed weight reading at all positions.
Maximum allowable change is per Table 4-2 for each of the listed balances four positions.
Transducer Adjustment
This adjustment requires that the balance Top Cover be removed and the Transducer Shield removed to gain
access to the Transducer for adjustments. Refer to Figures 4-4 and 5-1 for location of components called out in
this procedure.
NOTE: The top cover connector plug may be used to turn the balance on. The pins are numbered from 5 - 1
starting at the left. Shorting pins 5 and 1 turns the unit on and tares. Shorting pins 4 and 1 turns the unit off.
See Figure 4-7, page 4-14 for pin connections.
2. Locate the four screws holding the Transducer Shield in place. Remove the screws.
3. Remove the Transducer Shield. The Transducer is now exposed for adjustments.
4. Reinstall the Platform on 1mg balance and Sub Platform and Weighing Platform on 10/100mg balance onto
the exposed Transducer shaft.
NOTE:
Do not attempt these adjustments on the balance unless the balance is free
from drafts and is level. The balance is very sensitive and adjustments will
be affected.
5. Plug the power cord into the balance and turn the balance on.
6. Place 1/2 of the balance capacity or greater on the rear of the Platform.
7. Tare the balance. Move mass to right front and left front of the Platform as shown in Figure 4-3 and note
the readings. Correct the worst error first by turning Adjusting Screws on the Transducer shown in Figure 4-4.
8. Repeat the Off-Center Load Test and adjustments until balance is within specifications.
NOTE:
If the Off-Center Load Adjustment cannot be completed, refer to
thePrecision Test/Adjustment.
Place 1/2 of the balance capacity in the center of the Platform. Press the button to return the reading to
zero. Move the mass halfway (between the center and the edge) to the front of the Platform. Note any differences
in the displayed weight reading. Repeat this test for the back, left, and right positions. Maximum allowable change
in displayed weight readings
is + 5 counts for each of the four positions. If the displayed weight readings change more than 5 counts, proceed
to the following adjustment. To correct an off-center load error, proceed as follows:
Adjustment
1. Refer to paragraph 4.3.1, page 4-10 and open the balance.
2. Remove the two adhesive hole plugs on the transducer shield to gain access to the off center load
adjustment screws.
3. Reinstall the Platform.
NOTE:
Do not attempt these adjustments on the balance unless the balance is free
from drafts. The Balance is very sensitive and adjustments will be affected.
4. Plug the power cord into the balance and turn the Balance on.
5. Place 1/2 of the Balance capacity on the center of the Platform.
6. Press the >O/T< button to return the displayed weight to 0g. Slide mass front, back, left and
right to Platform edge. Note displayed values. Correct any errors by turning Cornerload Adjusting Screws
shown in Figure 4-5 (adjust the Hex Nut portion and ensure center screw does notturn).
and is in accordance with the tolerances listed in Tables 4-2 and 4-3. The masses used to perform this test must
meet or exceed ASTM Class 1 Tolerance.
NOTE:
The balance must pass the Off-Center Load test and Precision Test before
the Centered Load Test is performed.
This test is used to determine the linearity of the balance throughout its operating range.
Tables 4-4 and 4-5 lists the suggested masses to be used for checking linearity on each Balance model.
Before beginning:
• Perform calibration.
Preliminary Weighings
Using the Load 3 and Load 4 weight values (Table 4-3 or 4-4), perform some preliminary weighing and make sure the
display returns to zero after weighing.
1. Place Load 3 on the center of the Platform, read the displayed weight, then remove Load 3. If the display
Test
1. Place the reference mass on the center of the Platform and record the exact value (all decimal places)
displayed.
2. Remove the reference mass and verify that the display returns to zero. If it does not, disregard the reading,
press to rezero the display and take that reading again. See note above.
3. Repeat steps 1 and 2 five times and determine the Mean of the weight readings:
Mean = Sum of Readings/5
Record the Mean as Mean(0)
+0.00 mg
Cumulative Deviations are less than 210g balance tolerance of +0.2
If the cumulative deviations are within the tolerance, the balance is linear. If the balance proves to be non-linear,
retest the Center Load Test of the balance. If the balance does not meet the specifications of the Center Load
Test, the balance must be serviced.
This procedure is used when it is required to replace either the front Panel Membrane Switch, RS232 PC Board,
Main PC Board, or to gain access to the Transducer for removal and parts replacement and alignment.
Figure 4-6.
NOTE:
Visually inspect the Transducer Assembly for bent, cracked, or distorted
Flexures. Each Flexure should be perfectly straight. If it is determined that a
Flexure requires replacement, refer to paragraph 4.4.2 for the Upper Flexure
Arm and Lower Flexure Arm removal procedure.
The Arm Flexure and Ratio Flexures may each be replaced independent of one another and without
disassembling the entire Transducer. To clean between the Magnet Assembly and the Force Coil, remove the
Top and Bottom Flexure Arms and Center Poast with the Load Flexure still attached (Load Flexure must be
disconnected from the Ratio Beam first). To proceed, refer to Ratio Beam Removal/Installation Instructions
4.4.2.4, starting with step 3.
For the following procedures, the Transducer should be removed from the Unit Base as described previously.
1. Install the Test Fixture on the front of the Transducer using the six screws supplied with the fixture.
2. Remove and replace Flexures as required. Ratio Flexures and Load Flexure are accessible.
NOTE: In the next step, be careful as to the color and location of the wires removed. There is one Green wire,
one Brown wire and Brown and Green wires twisted together. It is suggested that a diagram be made of the
wire positions which you can refer back to when resoldering.
4. Unsolder the four Wires which connect the Ratio Beam to the Terminal Block mounted on the top of the
Load Cell Frame. The resistance measurement from the Brown/Green wire to the individual Green wire or the
individual Brown wire should be 460 ohms. The resistance reading from any wire to the frame should be open
or infinity.
5. Using a piece of tape, tape the position Sensor Board on the top and side to the frame. This is done so that
the Sensor Board can be accurately positioned back in place. Use a razor blade to cut the tape between the
board and frame. When the board is replaced, you can line up the tape edges for horizontal and vertical
positioning. Remove the three screws holding the position Sensor Board from the Transducer Frame. Two of
the screws have washers, these are tightened when replaced.
NOTE: The one screw on the left side of the position Sensor Board is not to be tightened when replaced, this
screw is a guide only.
6. Remove the two small screws holding the Magnet Cover in place.
7. The Ratio Beam can now be removed by lifting straight up.
8. Clean the air gap in the Magnet and also check the Force coil for any signs of debris. Clean if necessary.
(Adhesive tape turned back on itself is very helpful when trying to remove debris in the Magnet Assembly.)
9. Reinstall the Ratio Beam in the reverse order. Make sure the position Sensor Board is properly aligned
then replace and tighten the two screws on the right side of the board with the washers. DO NOT TIGHTEN
the single screw on the left side of the board, this screw is a guide only.
10. Solder the wires back to the Terminal Block removed in step 4. Make sure the wires are properly located.
11. Reinstall the Top and Bottom Flexure Arms removed in steps 2 and 3.
12. Tighten all fasteners and reinstall the Transducer in the Base.
13. Conduct Performance Tests.
NOTE: When facing the Sensor Board, you will notice there are three sets of wires connected to the board.
The three wires on the left side of the board are connected magnet assembly in the Ratio beam. Make sure
these wires are marked to go back in the same position when resoldered. The two white wires at the top of the
board are thermistor wires and the polarity is not important. The two wires located at the bottom right hand
side of the board must be replaced back in there original position. Mark these wires. Make a little diagram of
all wire positions which you can refer back to when resoldering.
5. Unsolder all wires noting there locations.
6. Remove the three screws holding the position Sensor Board from the Transducer Frame. Two of the
screws have washers, these are tightened when replaced.
NOTE: The one screw on the left side of the position Sensor Board is not to be tightened when replaced, this
screw is a guide only.
4.4.3.1 Upper Flexure Arm and Lower Flexure Arm Removal/Installation - Analytical
CAUTION
Before the Upper or Lower Flexure Arms can be removed, the Center Post
which contains the Load Flexure must be secured to the base of the
Transducer with two screws removed from the Bottom Mounting Plate. This
is necessary to prevent further damage from occuring when trying to
remove and replace individual Flexures.
1. Remove the three screws which secure Bottom Mounting Plate to Transducer Base.
2. At the front of the Transducer Assembly, insert the two Shims between the Center Post and the Transducer
Base with the open slotted end in first, see Figure 4-9. The slots should be aligned so that the screw holes in
the Center post to the Transducer Base are visible.
3. Take the two Screws removed from the Platform Mounting Assembly and install them in the Center Post
through the slots in the Shims. Finger tightens the screws. This will hold the Upper and Lower Flexures Arms
in place.
4. Carefully loosen the two Screws located at the rear of the Upper Flexure Arm and the rear of the Lower
Flexure Arm. These are the screws which hold the Flexures in place. Do not touch the two rear most screws
with the nuts on top as these are the adjustment screws.
5. Fully tighten the two screws which hold the Center Post in place.
6. At the top of the Transducer Assembly, remove the two Screws and Washers which secure the Platform
Support to the Center Post.
NOTE:
Do not remove the Upper or Lower Flexure Arms unless there is evidence of
damage to one or more of the Flexure Links.
7. To remove the Upper Flexure Arm from the Transducer Assembly, remove the four Screws and Washers,
one at each Flexure point.
8. To remove the Lower Flexure Arm from the Transducer Assembly, remove the four Screws and Washers,
one at each Flexure point.
9. Examine each Flexure Link. Each one must be perfectly straight. Replace any Flexure Link which is bent,
twisted, cracked or deformed in any manner.
10. To replace a defective Flexure Link, remove the Screw and Washer which secures the Flexure Link to the
Upper or Lower Flexure Arm. Replace as required.
11. To install Upper and or Lower Flexure Arms, perform steps 1 through 9 in reverse order.
Note: The reference resistor of the Position Sensor PC Board must be transferred to the replacement
Position Sensor PC Board.
9. Install the replacement Sensor Board. Replace all screws previously removed. The Sensor Board contains
an optical sensor which must be positioned in the center of travel of the Ratio Beam when the balance is
operated. If the Sensor Board is positioned to low or to high, the balance will not function properly.
10. Resolder all wires previously removed.
11. Re-install the Transducer in the base assembly.
12. Connect the Transducer cable back to the Sensor Board.
13. Connect the Top Cover membrane switch cable to the Main PC Board. The Top Cover should be lying in
front of the Base not on top. If you are unable to connect the cable properly, see Figure 4-7 which is the
connector on the main PC board. You can operate the balance directly from the connector by momentarily
shorting out the pins as indicated to turn it on or off.
14. Place the Platform in place on top of the Transducer. Apply power to the balance and turn the balance on.
15. If the Sensor Board is improperly positioned, too high or too low, the display indicates Error 8.4. Adjust the
Sensor Board up and down until the display indicates other than 8.4. Then tighten the two screws. After
completing this adjustment, turn the balance off and remove the Platform.
16. Replace the Transducer Shield over the Transducer and secure with four screws.
17. Refer to paragraph 4.3.2, page 4-11 and re-assemble the balance.
18. Conduct Performance Tests.
4.4.4 Main Printed Circuit Board (PCB) Replacement - Precision and Analytical Balances
Refer to Figure 4-5 and 5-1 for this procedure.
1. Refer to paragraph 4.3.1, page 4-10 and open the balance.
2. On the Main PC Board, disconnect the Load Cell Cable, RS232 Cable, Power Cable and the cable from the
Membrane Switch on the front panel.
3. Remove the four screws on the Main PC Board which secure it to the Base.
4. Install replacement Main PC Board in the reverse order as removed. Check the LCD display on the new
board and remove protective plastic covering if in place.
5. Refer to paragraph 4.3.2, page 4-11 and re-assemble the balance.
6. After assembly, verify that the balance meets all specifications. Perform all tests.
4.4.5 RS232 Printed Circuit Board Replacement - Precision and Analytical Balances
The RS232 PC Board is located at the rear of the balance. Refer to Figure 4-5 and Figure 6-5 for this procedure.
1. Refer to paragraph 4.3.1, page 4-10 and open the balance.
2. At the rear of the balance, remove the two hex head nuts which secure the RS232 Board to the base.
3. Remove the small mounting plate and lift out the RS232 board.
4. Disconnect the RS232 Cable from the RS232 PC Board.
5. Install replacement RS232 Board in the reverse order as removed.
6. Refer to paragraph 4.3.2, page 4-11 and re-assemble the balance.
7. After assembly, verify that the balance meets all specifications. Perform all tests.
5. Peel off the adhesive backing from the new Membrane Switch, then place the cable from the switch
through the hole in the Top Cover. Carefully position the new Membrane Switch on the Top Cover, starting at
the bottom edges and roll the Membrane Switch upwards into its final position. Smooth down the switch until it
is completely in position on the Top Cover.
6. Place the Top Cover into position next to the base and connect the Membrane Switch cable to the
connector on the Main PC Board.
7. Refer to paragraph 4.3.2, page 4-11 and re-assemble the balance.
8. After assembly, verify that the balance meets all specifications. Perform all tests.
Accord the setting of scale to set service tool COM PORT setting. The default setting is 2400, 7, n, 2, none.
6.1 DRAWINGS
This section of the manual contains exploded views for the Aventurer Precision and Analytical balances. The
exploded view drawings are designed to identify the parts which can be serviced on the balance in the field.
NOTE:
In all cases where a part is replaced, the balance must be thoroughly
checked after the replacement is made. The balance MUST meet the
parameters of all applicable specifications in this manual.
If further technical information is needed, please contact:
Ohaus Corporation
19A Chapin Road
P.O. Box 2033
Pine Brook, NJ 07058-2033 USA
Tel: 973-377-9000
Fax: 973-593-0359
In the United States call toll free, 800-526-0659 between 8:00 a.m. and 6:00
p.m. EST.
Figure 6-1 Overall Exploded View of Precision Balance without drift shield
Table 6-1 Precision Balance (without Drift Shield) Replacement Parts List
Figure 6-2 Exploded View of Precision Transducer (without drift shield mode).
Figure 6-3 Overall Exploded View of Precision Balance (with drift shield).
Table 6-3 Precision Balance (with Drift Shield) Replacement Parts List
Figure 6-4 Exploded View of Precision Transducer (with drift shield mode).