Lab 1. Understanding The Work of Microscopes: Campbell Biology, Reece Et Al, 2016) )
Lab 1. Understanding The Work of Microscopes: Campbell Biology, Reece Et Al, 2016) )
2021
INTRODUCTION
Imagine you work in a clinical laboratory that specializes in diagnosing cancer.
Every day, you examine biopsies (tissue specimens) from patients and look for any
signs of disease. The biopsies shown in Figure 2.1 have just been shipped to your lab
for analysis; how can you tell whether these specimens appear normal or abnormal?
Since cells are too small to be seen with the naked eye, you rely on a microscope to
magnify these specimens for analysis. (The cells in Figure 2.1 have already been
magnified with a microscope.) Once the specimens are in focus, you draw on your
knowledge of typical and atypical cell structure to detect any signs of disease. Within
a matter of minutes, you conclude that the first patient’s biopsy appears normal.
However, the second patient’s biopsy contains cells that appear to be cancerous.
Throughout the course of this semester, you will use microscopes to view cells,
and other biological matter in greater detail. During this lab, you will learn the proper
way to handle and focus a compound light microscope, in theory.
DISSECTING MICROSCOPES
A dissecting microscope shows the three-dimensional surface of a specimen in
greater detail. The objects viewed through a dissecting microscope—organs, insects,
leaves, etc.—are generally too large or too thick to be viewed through a compound
microscope. As the name implies, dissections can also be performed under this
microscope, when desired. The magnification ranges of a dissecting microscope
(typically 5x-50x) is much lower than the magnification range of a compound
microscope. For this reason, cells cannot be distinguished in most cases. The basic
parts of a dissecting microscope are labeled on Figure 2.2, and the steps for proper
usage are summarized next to the figure.
8. View your specimen by looking through the oculars from this point forward. Using the
course focus knob, move the slide away from the objective until the image
comes into rough focus.
9. Slowly turn the fine focus knob and bring the image into sharp focus. Only use
the fi ne focus knob to sharpen images from this point forward.
10. Scan the microscope slide in a pattern similar to the one shown below. If your
microscope has a mechanical stage, use the stage control knobs to move the
slide.
11. Center the object of interest before moving to a higher level of magnification.
12. Turn the nosepiece and position the next objective (example: 10x objective)
directly above the light beam traveling through the slide. Since your
microscope is parfocal, the image should still be in focus and require little (if
any) adjustment. Bring the image into sharp focus using the fi ne focus knob, if
needed.
13. Center the object of interest before moving to a higher level of magnification.
14. Turn the nosepiece and position the next objective (example: 40x objective)
directly above the light beam traveling through the slide. Bring the image into
sharp focus using the fine focus knob, if needed. The 100x objective lens (if
present) will not be used during Lab 2.
15. When you are finished observing the microscope slide, turn the nosepiece and
bring the lowest power objective (example: 4x objective) back into the viewing
field.
16. Remove the microscope slide from the stage. If you are using a mechanical
stage, move the slide away from the objectives prior to removal.
17. Turn off the light source, unplug the microscope, and wrap its cord around the
cord holder (if present) or base. Use lens paper to clean the microscope lenses.
18. Using two hands, carry the microscope back to its storage cabinet.
The magnification power of the ocular lens is written on the eyepiece, while the
magnification power of each objective lens is written on the side of the objective. The
total magnification of a specimen can be calculated using the following equation:
For instance, a specimen is magnified one hundred times (100x) when the 10x ocular
lens is used in conjunction with the 10x objective lens.
DEPTH OF FIELD
Since compound microscopes create two-dimensional images, only one plane (or
layer) of a specimen is in sharp focus at any given time. The thickness of this plane—
the depth of field—is dependent on the magnification power being used.
ELECTRON MICROSCOPE
Until recently, the resolution barrier prevented cell biologists from using
standard light microscopy when studying organelles, the membrane-enclosed
structures within eukaryotic cells. To see these structures in any detail required the
development of a new instrument. In the 1950s, the electron microscope was
introduced to biology. Rather than focusing light, the electron microscope (EM)
focuses a beam of electrons through the specimen or onto its surface. Resolution is
inversely related to the wavelength of the light (or electrons) a microscope uses for
imaging, and electron beams have much shorter wavelengths than visible light.
Modern electron microscopes can theoretically achieve a resolution of about 0.002 nm,
though in practice they usually cannot resolve structures smaller than about 2 nm
across. Still, this is a 100-fold improvement over the standard light microscope.
The scanning electron microscope (SEM) is especially useful for detailed study
of the topography of a specimen. The electron beam scans the surface of the sample,
usually coated with a thin film of gold. The beam excites electrons on the surface, and
these secondary electrons are detected by a device that translates the pattern of
LAB MANUAL | PRINCIPLES OF BIOLOGY I | SEMESTER 1 | 2020.2021
electrons into an electronic signal sent to a video screen. The result is an image of the
specimen’s surface that appears three-dimensional.
The transmission electron microscope (TEM) is used to study the internal
structure of cells. The TEM aims an electron beam through a very thin section of the
specimen, much as a light microscope aims light through a sample on a slide. For the
TEM, the specimen has been stained with atoms of heavy metals, which attach to
certain cellular structures, thus enhancing the electron density of some parts of the cell
more than others. The electrons passing through the specimen are scattered more in
the denser regions, so fewer are transmitted. The image displays the pattern of
transmitted electrons. Instead of using glass lenses, both the SEM and TEM use
electromagnets as lenses to bend the paths of the electrons, ultimately focusing the
image onto a monitor for viewing.
Electron microscopes have revealed many subcellular structures that were
impossible to resolve with the light microscope. But the light microscope offers
advantages, especially in studying living cells. A disadvantage of electron microscopy
is that the methods used to prepare the specimen kill the cells. Specimen preparation
for any type of microscopy can introduce artifacts, structural features seen in
micrographs that do not exist in the living cell.