0% found this document useful (0 votes)
135 views2 pages

Seculife ps100 DB - GB PDF

The SECULIFE PS100 is a patient simulator that allows for 12-lead ECG simulations. It has four selectable ECG waveforms with constant QRS duration and six performance testing characteristics. The PS100 has 10 universal patient lead connectors, displays four heart rates, and one amplitude. It is powered by a 9V battery or optional battery eliminator and measures less than 1 lb.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
135 views2 pages

Seculife ps100 DB - GB PDF

The SECULIFE PS100 is a patient simulator that allows for 12-lead ECG simulations. It has four selectable ECG waveforms with constant QRS duration and six performance testing characteristics. The PS100 has 10 universal patient lead connectors, displays four heart rates, and one amplitude. It is powered by a 9V battery or optional battery eliminator and measures less than 1 lb.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

SECULIFE | PS100

PATIENT-SIMULATOR
3-349-767-03
2/5.17

SECULIFE PS100 for testing patient monitors

• 3 sine, 2 square and 2 triangle selectable curves


• Indicator for residual voltage of the battery (in %)
• Patient simulation with 4 waveforms
• ECG with 30, 60, 120 and 240 BPM
• With constant QRS duration
• Leak test realisable

Specification DUT picture


The SECULIFE PS100 is a microprocessor-based patient
simulator. The user has four ECG characteristics with con-
stant QRS duration and six characteristics to test the de-
vice performance.

ECG-Simulations
The 10 Universal Patient Lead Connectors allow for 12
lead ECG simulations with independent outputs. AHA and
IEC color-coded labels are located on the face of the unit
to aid in connecting the corresponding AHA and IEC Pa-
tient Leads.

AHA-Label IEC-Label Description


RA R Right Arm
LA L Left Arm
RL N Right Leg
LL F Left Leg
V1 C1 V Leads (V1-V6)
V2 C2 (U.S. and Canada)
V3 C3 Also referred to as pericardi- Purchase parts
V4 C4 al, precordial or unipolar 1 Operating Instruction
V5 C5 Chest Leads (C1-C6) 1 Transport case
V6 C6 (International) 1 Battery
1 Power supply (model U.S.)
1 Power supply (model EU)

GMC-I Messtechnik GmbH


SECULIFE | PS100
PATIENT-SIMULATOR

Technical Data

Device:
Construction 184.4 x 113.3 x 38.4 mm/ABS Plastic
Weight ≤ 1 Lbs (0.45 kg)
Faceplate Lexan/Back printed
Operating Range 15 to 40 °C (59 to 104 °F)
Storage Range -20 to 65 °C (-4 to 149 °F)

General
Display 11 LED
Backlight No
Lead Test Terminals Yes
RS232 No
Power 1 x 9 V Battery
Battery Eliminator Yes (Option)

Functions:
ECG-NSR Yes Pacemaker No
Leads 10 Arrhythmias No
Rates 4 Blood Pressure No
Amplitudes 1 Temperature No
QRS Interval Adult
ST Segment Elevation No
ECG-Performance Yes
Sine Waves 3
Square Waves 2
Triangle Waves 1
Pulse Waves -
Respiration No
Rate -
Baseline -
Delta Impedance -
NIPB No
Rates -
Pressure Output -
Special Modes No

Ordering information
Description Type Order number
Patient-Simulator SECULIFE PS100 M695L

For more information on accessories, see


• www.gossenmetrawatt.com

Edited in Germany • Subject to change without notice • A PDF version is available on the Internet

Phone +49 911 8602-111


GMC-I Messtechnik GmbH Fax +49 911 8602-777
Südwestpark 15 E-Mail [email protected]
90449 Nürnberg•Germany www.gossenmetrawatt.com