Caballero 0730 PDF
Caballero 0730 PDF
Caballero 0730 PDF
• Validation of Process
– Provide overview for process validation
– Outline the components of a process validation
– Provide a specific example of a process validation
2
When do we do it?
• Applies to new or significantly changed
equipment or processes that affect
donor or patient safety, or the safety,
purity or potency of products.
• Examples include:
– Equipment used during product processing
– Transport procedures, including transport
carrier
– New or revised processing procedures
3
Equipment Validation or Qualification
• Phase I
– Equipment Installation Qualification (IQ)
• Identify the equipment
• Installation/Maintenance/Calibration
requirements verified with Biomed
• SOP written to include maintenance /calibration
requirements
• Equipment added to Preventive maintenance
list
• Supply vendor qualification check completed
• Supplies added to order list
• Individual(s) responsible for completing this
phase
4
Equipment Validation or Qualification
Validation Checklist
Instrument: _____________________ SN: _________________
5
Equipment Validation or Qualification
• Phase II
– Operation Qualification (OQ)
• Operational Variable/Critical control points
identified
• Does the instrument function as described by
manufacturer?
• SOP written
• SOP specifies expected outcomes
• Training and competency records completed
• Individual(s) responsible for completing this
phase
6
Equipment Validation or Qualification
Validation Checklist
Instrument: _____________________ SN: _________________
7
Equipment Validation or Qualification
• Phase III
– Performance Qualification
• Does the equipment function correctly and
consistently for the intended application (Mock
products)
• Monitoring process in place
• SOP written
8
Equipment Validation or Qualification
Validation Checklist
Instrument: _____________________ SN: _________________
9
Equipment Validation or Qualification
• Example of Equipment Validation-
Steps to follow:
– Identify equipment used to be implemented
or improved.
10
Writing The Validation Plan
• Validation Title
11
Example of Control Rate Freezer Validation
Validation Plan
Type: (IQ) Installation Qualification (OQ) Operation Qualification (PQ) Performance Qualification
I. PURPOSE OF VALIDATION
To prospectively ensure that the MRV controller performs as expected after installation and prior to
freezing of patient products.
12
Equipment Validation or Qualification
• Responsibility assignment
13
Equipment Validation or Qualification
III. RESPONSIBILITY ASSIGNMENT- Document name in table (Signatures follow plan and results).
Validation Plan written by: Lizette Caballero, MT
Validation Plan reviewed by: Susan Ingersoll, Ph.D.
Validation Plan approved by (medical director): Vijay Reddy, M.D., Ph.D.
Installation Performed by: Robert Conine, Cryo Associates Representative
Validation performed by: Anginett Batista, Lizette Caballero
Validation Results evaluated by: Lizette Caballero, MT
Validation Results reviewed by: Lizette Caballero, MT
Validation Results approved by (medical director): Vijay Reddy, M.D., Ph.D.
Validation Results approved by (QA Manager): Marie Fuentes- Rivera, RN
14
Equipment Validation or Qualification
15
Equipment Validation or Qualification
16
Equipment Validation or Qualification
Three test runs will be performed using conditions similar to a stem cell product prior to using the
freezer to cryopreserve any patient products and 20 patient products will be prospectively analyzed to
ensure that the freezer is working properly after installation.
17
Equipment Validation or Qualification
C.3. Performance Qualification
Three test runs will be performed prior to using the freezer to cryopreserve any patient products and 20
patient products will be prospectively analyzed to ensure that the freezer is working properly after
installation.
Part I. Perform three test runs following this procedure:
A. Collect peripheral blood using 5 EDTA tubes from a donor.
B. Prepare a buffy coat with the EDTA blood sample by centrifuging for 5 minutes at 1000
RPM’s.
C. Save red cells and plasma.
D. Label two cryovials with: Donor’s name, Date of testing, test sample number.
E. Make a hole in one of the cryovial’s lid to allow chamber probe placement.
F. Using a 12 X 75 plastic tube mix:
1. buffy coat cells from step B.
2. 0.4ml plasma from step C.
3. 0.4mlPlasma-Lyte A
4. 0.2ml DMSO
G. Run WBC (using SOP# 193.455 “Cell-Dyn Analyzer Operation” and viability using SOP# 193.406
“Viability Test procedure”)
H. Transfer mix from step F to 2 cryovials from step D (1.0ml each).
I. Insert Chamber probe trough hole in lid of cryiovial containing cells and freezing media.
J. Place both cryovials inside chamber
K. Start cryopreservation procedure following SOP 193.433 “Control Rate Freezer Operation”.
L. Move sample without probe to Liquid Nitrogen Tank. Discard 2nd cryovial (the one with the hole
on the lid).
M. Save print out of cryopreservation graph.
N. Repeat steps 1-10 using two more donors.
O. Keep cryovial for at least 24 hours before testing.
P. After 24 hrs thaw cryovial using waterbath and perform:
1. Viability test
2. WBC counts
18
Equipment Validation or Qualification
• List of data/records to be collected
Examples include:
– New or revised SOPs, labels and forms
– Training and competency records
– Checklists
– Quality Control and outcome
measurements
19
Equipment Validation or Qualification
20
Equipment Validation or Qualification
• Establish Acceptance criteria for each
data/record collected
– Must be defined and measurable
• List of references
21
Equipment Validation or Qualification
C. Establish acceptance criteria.
E.1. Installation Qualification
1. Instrument turn on and off.
2. Control Panel, Displays indicators working properly.
3. Alarm sound is working properly when activated.
4. Controller pass calibration testing performed by Cryo Associates Rep.
5. Printer works properly
Part II.
1. WBC recovery within 20% of pre cryopreservation value.
2. Post thawed viability result >50%.
3. ANC engraftment < 12 days and PLT engraftment < 15.
22
Validation Results
DATE(S) OF VALIDATION:
VI. VALIDATION RESULTS
A. List SOPs, personnel, equipment, supplies, or procedure steps that deviated from or
were added to original validation plan. State reason for change and document approval
of change.
VII. CONCLUSION
A. Validation data evaluation and determination of acceptance.
Data must meet pre-determined acceptance criteria.
B. Comments/Actions.
23
FDA Guidance for Industry Process Validation: General
Principles and Practices
24
Example of Process Validation
25
Process Validation
26
Process Validation
27
Process Validation
• Establish Testing Conditions:
– Preparation of the Inoculums (Staphylococcus epidermidis
and Candida albicans)
• Dilute each culture to 0.5 O.D. Dilute the bacteria with sterile
saline and the fungus with sterile water. This 0.5 dilution
represents approximately 108 CFU/ml
• Dilute 1:100 to give 106 CFU/ml
• Dilute to 3 X 103 CFU/ml by adding 3ul of the 108 dilution to 1
ml of diluent
28
Process Validation
• Draw up 3ml of the remaining blood in 3cc syringe
and inoculate the negative control bottles with 1.5ml
of blood
29
Process Validation
• Make up freezing media using supernatant
plasma, plasma Lyte A and DMSO
30
Process Validation
• Data to be Collected: reports of growth from
both bacterial and fungal cultures
• Expected Results:
– The negative control will no grow any microbial or
fungal culture
– The No DMSO control will grow both bacteria and
fungus
– The 10% and 20% DMSO test samples will grow
the bacteria and fungus the blood was inoculated
with.
31
Process Validation
2.5
1.5
S. epidermidis
Days
C. albicans
1
0.5
0
0% 10% 20%
DMSO Concentration
32
Process Validation
33
Process Validation
34
Summary
• Perform validation according to
Validation Plan
• Data/records evaluation and
determination of acceptance must meet
pre-determined test criteria
• Ensure new or revised SOPs, forms or
labels are in place for use
• Ensure completion of training and
competency records
• Ensure ongoing monitoring is in place
35
Contact Information
Lizette Caballero
MT(ASCP)
2501 North Orange Ave Ste 786
Orlando, FL 32817
407-303-2442
[email protected]
36
37