US-751 Service Manual: Ito Co., Ltd. 1st Edition
US-751 Service Manual: Ito Co., Ltd. 1st Edition
Service Manual
Clicking the blue colored character in this page can enter the page
where detail information is written.
Index
Product outline Specification(General) Specification (Ultrasound)
Adjustments
2. Assembly / Disassembly
Lower Case
US-751 Table of Contents
Contents Page
1 Product Outline 1
4 Specification 3
5 Used material 7
6 General specification 8
15 Spare parts 20
16 Trouble shooting 21
17 Error messages 25
18 Disassembly procedure 26
1. Outline
US-751 is a dual frequency ultrasound therapy device designed for overseas use only.
The standard probe is the L-type ultrasound probe which delivers the two frequency
outputs of 1MHz and 3MHz.
Gel and the L-type probe are also supplied as the standard accessories and the
the S-type probe and the EST cable are provided as the option.
1
US-751 Specification (1)
Electric specification
Power supply voltage AC100 - 240v +/-10%
Open error:
Generated when probe is disconnected during
output.
Probe error:
Error function Generated when abnormal connection of probe is
detected.
System error:
Generated when the main CPU found abnormal
communication with boards.
2
US-751 Specification (1)
The output adjustment is available through encoder and menu by 0.05W/cm2 step.
・Transducer: Installed inside of the probe head. The ultrasound probe (L) and (S) is
selected according to the parts of therapy.
When output, the LCD window, the front panel LED and the probe
LED indicates the output.
・Timer: 1-30 minutes (set by minutes step. Deviation: 30 minutes +/-1 minutes)
*If timer setting is changed during operation, counting will restart from
the beginning.
3
05W/cm2 step.
L) and (S) is
4
US-751 Specification (2)
・Other functions
A) Auto contact function
During ultrasound therapy, the impedance change of the transducer is used to judge the
contact/no contact (or in water/in air) of the probe head with the tissue. In the contact
(in water) state, the probe performs output as is set. In the no contact (in air) state, output
goes into standby mode and weakens the output slightly.
When output, the LCD window, the front panel LED and the probe LED indicate output. .
Also, therapy stops if output standby status lasts for 3 minutes.
5
US-751 Specification (3)
ITEM SPECIFICATION
Dimension 90mm(W) x 23mm(D) x 93mm(H) (Including leg: 96mm)
Weight About 3.0kg
In accordance with the design standard of ultrasound therapy
Intensity
device for business purposes "84E-0019"
Device ……… 5years
Accessary……3 years (except consumable items such as therapeutic
Mechanical life
gel)
Consumable.. therapeutic gel
6
US-751 Used material
1. Raw material
7
US-751 General information
Operation theory
US-751 is a ultrasound pain relief device utilizing thermal and non thermal
physiologic reaction of the tissue which is caused by the delivered ultrasound of
the device.
US-751 generates the two frequency of 1MHz and 3MHz of which their level are
controlled by the level controller circuit and drives the transducer to inject the
ultrasound vibration to treat the affected tissue.
The power of each circuit are supplied from commercial AC which is converted
to DC in the power regulator.
CPU controls output level, timer, error detection, display control, and display
functions.
LCD and LED display the status of manipulation of touch panel, switches,
and encoder.
User can select the ultrasound probe from S-type or L-type according to the
treatment part.
8
US-751 Notes of Operation and Usage
1. Usage
(1) Parameter setting.
LCD touch panel is used to select, change, or set the output parameters.
Note: The low frequency electrode pad is electrically plus (+) and the
ultrasound probe is minus (-).
Please maintain the low frequency electrode pad distance from the
ultrasound probe head.
3. After use
(1) Clean the probe head and keep it in a dry place.
9
US-751 Dimension and Parts name (1)
3 Output control VR
4 Stop button
6 Power SW
9 LED
10 Power filter
10
US-751 Dimension and Parts name (2)
No. Name
1 Plug
2 Code
mm
3 Connector
No. Name
1 Transducer
2 Case
3 Cable
4 Plug
5 Code protector
6 Output indicator
7 Core filter
mm
No. Name
1 Transducer
2 Case
3 Cable
4 Plug
5 Code protector
6 Output indicator
7 Core filter
mm
11
US-751 Dimension and Parts name (3)
1 Output connector
2 Code
3 Core filter
4 Core filter
5 Rectangular connector
(2) Ultrasound probe (L)
6 Φ2 plug
1 Output connector
2 Code
3 Core filter
4 Rectangular connector
5 Φ2 plug
12
US-751 Overall Block Diagram
No.9 board
No.6 board
US-751 Block Diagram (+B Line No.1 card)
CN3
U6 U1
1 Touch panel
DC Reg DC Reg
2
TP26: +3.3VPLL
LCD Controller LRC_VccPLL
LCD Backlight LCD Controller ULC_VccPLL
TP38: +3.8V TP28: +3.3V
TP29: +2.5V
U10 U7 U8
DC Reg DC Reg DC Reg LCD Controller Vcc
TP42: +1.2V
U11
DC Reg LCD Controller Vcc
DIPSW
U2 Vcc
LCD Controller Vcc
LCD Controller IO Vcc
LCD Controller Aux Vcc
U9: RES
LCD
14
US-751 Block Diagram (+B Line No.2 Card)
Sound SW
15
US-751 Block Diagram (+B Line No.4Card)
TP61
Pulse shaping
DC Fan U4,5,10,11
Under voltage
Pulse
Phase Cont.: U1, U2, U3, U6, U9, U12,
Shaping:
U13,
U4, U5, U10,
+V
RY1,2,3,4 Coil
CPU
CPU Vref
16
US-751 Block Diagram (Signal Path No.1 Card)
+3.3v +3.3v
U4
RAM
U2
Flash ROM
+3.3v LCD
U9 14
15
17
18
17
US-751 Block Diagram (Signal Path No.1 Card)
+3.3v TP4
STOP 1 CMOMBI_DE
TP5
2
+15v TP19
3 ES-520 IN
LED1 LED2 TP20
TR1:B COMB TP24 4 ES-320 IN
TP81 5 SU520SW-A
U1
+15v
TR2:R
CPU SU520SW-B
TR3:G EST
+3.3v
U2
U3
Audio
U7
RESET
U8
18
US-751 Block Diagram (Signal Path No.1 Card)
Switching Regulator
Power Source
I DATA
Q20, Q28,vQ25, Q28,
U27
Q29, Q30
Switching Regulator
V Data
CPU
U26, U28, U29, U21,
U31
1MHz /
P DET
U22, U20
U12, U13, U9
19
US-751 Block Diagram (Signal Path No.5 Card)
9
L probe
10
TP4
S probe
TP15
20
US-751 Spare Parts
7 LCD 070280
21
US-751 Troubleshooting (No Power)
No Power (1)
Check +15 at
Step 1 TP52 of No.2
card.
No
+15v exists.
OK
OK
Check TP25 of No.1 Change No.2 card.
Step 4
card.
No
+15v exists.
OK
Check TP-28 of No.1
Step 5
card. Check harness connection
between No.2 and No.1 card.
OK
No
+3.3v exists.
OK
No
+1.2v exists.
OK
Change LCD
22
US-751 Troubleshooting
No
+15v exists.
OK
Check TP-28 of No.1
Step 2 card. Check harness connection
between No.2 and No.1 card.
OK
No
+3.3v exists.
OK
No
+1.2v exists.
OK
Change LCD
23
US-751 (No output)
No output
Change Probe
with new one.
No
Change probe.
Check TP-61
No
+15v exists.
Ye
Check harness connection
Check TP12 or between No.4 and No.2 card.
TP22.
No
Change probe
Change No.4 card.
24
US-751
Confirm that
output is normal.
No
Output is normal.
Yes
Go to Troubleshooting
(No Output)
Change the probe..
25
US-751 Error messages
Open error.
Indicates that probe was disconnected from the output terminal during treatment.
Probe error.
Indicates that the connected probe has problem. Probe change is required.
System error.
Indicates 15v DET abnormality, fan rotation abnormality or circuit misalignment
of control board.
Indicates abnormality of main fan rotation.
Indicates communication error between DCN and US board or between DCN and
SW board.
26
US-751 Disassembly procedure
No.2 card
27
US-751 (Lower case)
EMC insulation
DC Fan
No.8 (EMC)
Speaker card
No.4 (US)
card
Power supply
unit
US Fan
No.2 card
28
US-751 (Disassembly No2 & No.8 card)
No.2 Card
NC1 NC9
NC10
NC2
NC12 NC3
NC7
No.8 Card
CN3 CN4
CN1
CN2
29
US-751 (Disassembly No.4 card)
No.4 Card
30
US-750 (Disassembly Power supply unit)
Remove the four screws , disconnect TB1 and TB2 to take out
the power supply unit.
Grounding point
of the power
supply unit.
31
US-751 (Disassembly upper case 1)
32
US-751 (Disassembly upper case 2)
33
US-751 (Disassembly upper case 3)
34
Ultrasound Power Measurement
Probe holder
(Differs by probe) Probe mounting plate (Differs by probe)
35