ValveLink VL2000
ValveLink VL2000
Form 5723
August 2001
Using
ValveLink VL2000 Series
Software with HART
FIELDVUE Instruments
D102801X012
www.Fisher.com
FIELDVUE, ValveLink, and Fisher are marks owned by Fisher Controls International, Inc., a
business of Emerson Process Management. The Emerson logo is a trademark and service mark of
Emerson Electric Co. HART is a mark owned by the HART Communications Foundation.
FOUNDATION fieldbus is a mark owned by the Fieldbus Foundation. All other marks are the
property of their respective owners.
While this information is presented in good faith and believed to be accurate, Fisher Controls does not
guarantee satisfactory results from reliance upon such information. Nothing contained herein is to be
construed as a warranty or guarantee, expressed or implied, regarding the performance,
merchantability, fitness or any other matter with respect to the products, nor as a recommendation to
use any product or process in conflict with any patent. Fisher Controls reserves the right, without
notice, to alter or improve the designs or specifications of the products described herein.
Comment Form
ValveLink Software User Guide
The ValveLink Software User guide is intended to provide the basic information you
need to understand ValveLink software. Detailed information about ValveLink software
is provided in ValveLink software online help.
Please give us your feedback to help us improve this manual.
Did you use this manual to: Yes No
Obtain an overview of ValveLink software
features? _____ _____
Learn about the use of ValveLink software
menus, buttons, and windows? _____ _____
Learn basic information about Customizing,
Calibrating, Tag Reports, Diagnostics, and _____ _____
Database Operations?
______________________________________________________________________
______________________________________________________________________
:_____________________________________________________________________
______________________________________________________________________
VL2000 Series
Yes No
May we contact you about your comments? _____ _____
Name________________________________________________________
Company_____________________________________________________
Phone________________________________________________________
Date__________________________________________________________
Thank you for your comments.
Fold and mail this form to Fisher Controls International, Inc.
___________________________________________
Name______________________________________
Company___________________________________
Address____________________________________
Marshalltown, IA 50158
VL2000 Series
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
What is ValveLink software . . . . . . . . . . . . . . . . . . . . . 1–2
About this Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
Organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–5
How To Use This Guide . . . . . . . . . . . . . . . . . . . . . . . . 1–7
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–8
VL2000 Series
ValveLink Software Online Help . . . . . . . . . . . . . . . . . 4–16
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
Working With Calibration . . . . . . . . . . . . . . . . . . . . . . . . 7–2
Calibrating Analog Inputs . . . . . . . . . . . . . . . . . . . . . . . 7–6
Auto Travel Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 7–8
Manual Travel Calibration . . . . . . . . . . . . . . . . . . . . . . . 7–12
Travel Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . 7–13
Pressure Sensor Calibration . . . . . . . . . . . . . . . . . . . . . 7–14
VL2000 Series
Batch Runner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9–1
The Batch Runner Tool . . . . . . . . . . . . . . . . . . . . . . . . . 9–2
Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10–1
Creating Tag Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . 10–2
Using the AuditLog Viewer . . . . . . . . . . . . . . . . . . . . 10–4
AuditLog Toolbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10–6
Printing the AuditLog . . . . . . . . . . . . . . . . . . . . . . . . . 10–11
Glossary
Index
VL2000 Series
This page intentionally left blank.
VL2000 Series
Section 1 Introduction
11
D Managing tags.
D Calibrating an instrument.
D Setting up an instrument.
D Solving problems.
See the ValveLink VL2000 Series Software Installation Guide for
instructions on how to install ValveLink software and how to connect
ValveLink software to FIELDVUE instruments.
Service/Support Directory
To contact technical support, refer to the support package that
accompanied your ValveLink software which includes Fisher
Controls International, Inc., address, telephone and fax numbers.
Or find them online at:
www.fisher.com
and select Support.
Software Updates
For updates refer to the Foundation support package that
accompanied your ValveLink software. If you subscribe to
Foundation Support, the latest ValveLink VL2000 Series software
updates are available online:
www.valvelink.com
and select Updates.
22
PC
FV031–M
Figure 2-1. Attaching the ValveLink Hardlock.
Note
To install ValveLink software on a computer running
Windows NT, you must have Administrator privileges.
Note
When installation is complete, remove the ValveLink
hardlock and store it in a safe place. It provides your
backup license.
Note
If you do not have a HART modem or FIELDVUE
instrument available, proceed to Step 5.
Modem
E0350 / IL
Figure 2-3. Instrument Connections.
Click the Start button. From the Start menu select Pro-
grams>ValveLink>ValveLink.
Step 6: Login
Click OK.
Type a name for the new security group, then click the
OK button.
Click on the new user name to highlight it. Click the list
box arrow to the right of the Security Group box and se-
lect the desired security group. Click OK.
MODEM SYMBOL
VALVE SYMBOL
Note
If you do not see a valve symbol, you may not be
connected to a FIELDVUE instrument. Recheck Step 3
and Step 4, then right click on the HART modem
symbol and select Scan for New.
33
Note
For full access to ValveLink features you must
discontinue use of the default name and password.
Create security groups, and assign new user names
and passwords.
Note
Before deleting a security group, make sure the group
does not have any user names assigned to it. If you
delete a security group to which user names are
assigned, those users will be able to access the
ValveLink program but will not be authorized to perform
any of the privileges controlled through security groups.
If you have system privileges you may assign users specific kinds of
access by adding and deleting user names, changing user pass-
words, and changing a user’s security group.
D Plant Area
D Manufacturer
D Valve Style
D Packing Type
Or any names meaningful in your facility.
You can also establish selections which further define each group.
You assign individual tags to specific group selections.
Tags that are not assigned to any group appear as unassigned in the
database.
When the tag you want to delete appears in the Tags Found box,
highlight it then click the Delete Tag button at the bottom of the Tag
Management window.
Click OK to confirm that you want to delete the tag when you see the
message, ”Delete tag, and all its associated datasets in the
ValveLink database?”.
ValveLink software checks for duplicate tags. If you are trying to im-
port a tag name that already exists, you are presented with several
options.
Copy a file containing tag information and datasets into the
ValveLink database.
To import tag data, from the menu bar select:
Tag>Import Tag Data (See figure 3-10).
If you cannot select Import Tag, talk to your ValveLink administrator
for security authorization. In the File Name box, identify the file you
want to import. Click OK.
Use tag groups to narrow the tags listed in the Database Tags box.
To add to the Tags to Export box, select tags from the Database
Tags box and click the Add button.
To remove a tag from the Tags to Export box, select it then click the
Remove button.
In the Datasets to Include in the Export File area of the window, se-
lect the types of datasets you want to include with the exported tags.
In the Export File Name box, enter the file name and directory where
ValveLink should store data, for example, A:\TAGS.EXP.
Click the Export Tags button. When you finish exporting tags, click
the Close button.
Now you have completed assignment of specific User and Group
information which allows ValveLink software to monitor and capture
performance data.
If you have the Application Setup privilege, you may set preferences.
From the menu bar select:
Customize ValveLink>Preferences.
There are eight tabs containing choices for setting up your
instrument. Select the Help button on each page and ValveLink
software online help gives detailed instructions.
Communication Page
ValveLink software communicates with instruments through a HART
modem or HART multiplexers. Click the
Preferences>Communication tab to select the method of
communication.
Multiplexers Page
ValveLink software can support up to 32 multiplexers on each
network. If you have fewer multiplexers wired to the network,
however, when doing a network scan, you can tell the ValveLink
software to scan for a lesser number of multiplexer addresses. You
also can establish whether the ValveLink software functions as the
primary master or the secondary master. To set multiplexer
parameters, click the Preferences>Multiplexers tab.
Network Alert Scan Page
If the ValveLink software is communicating over a HART multiplexer
network, you can run Network Alert Scan to scan the instruments on
the network for current alert conditions. To set alert options, click on
the Preferences>Network Alert Scan tab.
Diagnostics Page
To set the collection interval for diagnostic tests, click on the
Preferences>Diagnostics tab.
Units Page
To select the units of measure, select: Preferences >Units tab.
44
Figure 4-1. ValveLink Software Title Bar, Menu Bar, and Toolbar.
Figure 4-2 shows the title bar. The title bar contains the name of the
application on the left side and the window control buttons on the
right.
When you select a device from the connection view pane, the tag
displays in the title bar. After a tag is opened the specific ValveLink
software function is also named in the title bar.
Menu Bar
Figure 4-3 shows the menu bar. Menus provide quick and easy
access to all of the ValveLink software features. To select from a
menu, click an item to open a list of options. Click an option to select
it.
Status Bar
This section explains the menus on the menu bar and the
commands you can access from each menu. Menus provide quick
and easy access to all of the ValveLink software features. To select
from a menu, click an item to open a list of options. Click an option to
select it. Select Help on the ValveLink software menu bar, then
Search For Help On, and type in the subject or menu command to
receive more information.
Step 4: Scroll through the terms listed until you find the word
you’re looking for.
WARNING
Changes to the Instrument setup may cause
changes in the output pressure or valve travel.
To avoid personal injury or property damage,
be sure the instrument and valve are isolated
from the process.
Note
The Setup Wizard does not read the existing
configuration of DVC5000 or DVC6000 Series
FIELDVUE Instruments.
Note
The performance tuner is not available for instrument
levels AC and HC.
D The performance tuner also does not detect or adjust for flow
restricting accessories placed between the digital valve controller
and the actuator. This includes devices such as:
D Solenoid valves
D Travel Alerts
D Travel Control
D Pressure
D Tuning
D Notes
For DVC6000 Series instruments with the ESD option, select from
one of the following tabs on the DVC6000-ESD Detailed Setup win-
dow:
D Initial Setup
D Inputs
D Cycle Counter Alert
D Deviation and Other Alerts
To view the data on a page, click the tab for that page to bring it to
the front.
The color of the page tabs indicate the status of the instrument data
and dataset data. Until information is read from the instrument, the
page tab remains gray. After the instrument information on a page is
read from the instrument, and if the instrument data matches the
data in the dataset, the tab for that page turns blue. On pages where
the data read from the instrument is different from the dataset data,
the page tab is yellow.
Each page, except the Notes page, in the Detailed Setup window
has two columns: one for instrument data and one for the ValveLink
dataset. The instrument data is the setup data polled from the instru-
ment. The ValveLink dataset is the data retrieved from the ValveLink
software database. These columns allow you to easily compare the
setup data in the instrument with the data in your database.
Working with Datasets
Database data is stored in datasets. You can have many datasets
for the same instrument. When you work with setup data, you can
choose to save the last polled information from the instrument in a
dataset. When you save a dataset, the dataset is given a name that
includes the date and time corresponding to when the data was
saved. You can also add a description of up to 40 characters to the
name. To add the description, go to the Notes page by clicking on
the Notes tab. Type your description in the Dataset Description box
Note
Downloading the Cycle Counter, instrument clock, and
Travel or Pressure Accumulator setup data with the
Download/Upload ALL Parameters arrows requires a
check in the checkbox on the General page in the
Preferences window. (From the menu bar, select
Customize ValveLink>Preferences>General tab.)
There may be times when the instrument data does not match the
dataset data, but still the download (left pointing) arrows are dim.
Note
If one type of master takes an instrument Out Of
Service, the same type must put it In Service. For
example, if a device set up as a primary master takes
an instrument Out Of Service, a device setup as a
primary master must be used to place the instrument In
Service.
Note
When the instrument is Out Of Service, the output will
not track the input. A temporary means of process
control may be necessary before taking the instrument
out of service.
Note
To remove the instrument protection requires installing
a jumper across the instrument AUX+ and AUX–
terminals. Refer to the instrument instruction manual for
jumper details.
D Trim
D Actuator
D Reference
D Notes
To view the data on a page, click the tab for that page to bring it to
the front.
Before setting up valve specifications, you may want to establish
defaults for units of measure by selecting:
Customize ValveLink>Preferences.
77
WARNING
Calibration may result in a sudden change in valve
position and loss of process control. DO NOT
continue until the instrument and the associated
valve are isolated from the process.
Before beginning calibration, open the instrument tag and check the
tag information on the status bar to verify:
D The instrument is connected.
D The Instrument Protection is set to None.
Canceling Calibration
If calibration stops abruptly, for example, if the instrument loses
power, ValveLink software notifies you with a message. Because
ValveLink software calibrates an instrument in steps, a failure does
not stop the entire calibration procedure. Rather, the current
calibration step remains visible. From the system prompt, you can
click the Cancel button to stop calibrating or click the Next button to
retry the step.
If you click the Cancel button, either from the system prompt or from
the Calibration progress bar, ValveLink prompts, “Calibration
canceled”. The instrument will be restored to its original state.
Most Least
WARNING
Calibration may result in a sudden change in valve
position and loss of process control. DO NOT
continue until the instrument and the associated
valve are isolated from the process.
Instrument
Figure 7-3. Example Analog Input Calibration Setup (HART Modem shown)
WARNING
Calibration may result in a sudden change in valve
position and loss of process control. DO NOT
continue until the instrument and the associated
valve are isolated from the process.
Note
Auto calibration adjusts the instrument to apply a light
seat load. To ensure a full shutoff seat load, configure
the Travel Cutoff Low to 0.5%.
Figure 7-4. Example Auto Travel Calibration Setup (DVC5000 Series with HART Modem Shown)
user interaction is required. ValveLink automatically calibrates the
instrument, using valve and actuator stops as the 0% and the 100%
calibration points. If calibration is for a sliding-stem valve actuator,
continue to Step 1.
Step 1: Select the desired crossover adjustment. There are
three choices for adjusting the crossover: Manual, Last,
or Default. Manual is the recommended choice. It per-
mits you to manually set the crossover. If you select
Manual, go to Step 2.
If you select Last, this is the last step. The crossover
setting currently stored in the instrument from the last
travel calibration is used, followed by a series of actions
that will automatically calibrate the instrument. Use this
selection if you cannot use manual, such as when you
cannot see the valve.
If you select Default, this is also the last step. A value for
the crossover is written to the instrument, followed by a
series of actions that will automatically calibrate the
instrument. Use this selection only as a last resort.
WARNING
Calibration may result in a sudden change in valve
position and loss of process control. DO NOT
continue until the instrument and the associated
valve are isolated from the process.
The travel sensor is normally adjusted at the factory and should not
require adjustment. However, if the travel sensor has been replaced,
adjust the travel sensor by performing the following procedure. See
the “Maintenance” section, of the instrument instruction manual for
travel sensor replacement procedures.
WARNING
Calibration may result in a sudden change in valve
position and loss of process control. DO NOT
continue until the instrument and the associated
valve are isolated from the process.
WARNING
Calibration may result in a sudden change in valve
position and loss of process control. DO NOT
continue until the instrument and the associated
valve are isolated from the process.
Instrument
1
NOTE:
1 FOR DVC6000 SERIES INSTRUMENTS, CONNECT THE PRESSURE GAUGE TO OUTPUT A OR
B OR SUPPLY DEPENDING ON THE PRESSURE SENSOR BEING CALIBRATED.
Note
Diagnostic tests permit you to run interactive tests that
move the valve, poll for data, and display the data
graphically.
Some diagnostic tests may not be available with your installed con-
figuration of the software. To see which tests are available, see the
ValveLink Help screen, About ValveLink.
Diagnostic checks allow you to display and monitor an instrument’s
condition.
D Checks
Note
Before running any diagnostic test which moves the
valve, change the instrument’s mode to Out of Service.
When you finish running tests, change the mode back
to In Service. To change from In Service to Out of
Service, or vice versa, click the Instrument Mode button
on the toolbar, or select:
Instrument Setup>Mode>Instrument.
Setting Up for Trending Before you can begin trending, you must
specify the tags of the instruments that will
be communicating the data you wish to
To identify the instruments from which to trend.
trend data, select :
Click the Setup button. From the Network
Network menu>Network Scan>Network Scan Setup window, select the tags to
Alert Scan window. scan. Select the Help button for detailed
instructions.
The Trend List If the network is being actively scanned,
you can see a list of all tags currently
trended.
To see the trend list, from the menu bar,
select: For each tag listed, the time last updated
is displayed, along with the most recent
Network>Trend List. travel, input, pressure, and temperature
data. This list is constantly updated as
data is received from the instruments.
Viewing Trends To view trend data for an instrument:
Open an instrument’s tag, right click the
Select: tag and select Trend from the drop-down
menu, click the Trend button on the tool-
Diagnostics>Trend>Travel Activity. bar, or Diagnostics>Trend.
HART instruments have an alert record that can store alerts from
any of the enabled alert groups:
D Valve Alerts
D Failure Alerts
D Miscellaneous Alerts
The alert record also includes the date and time (from the instrument
clock) the alert occurred.
To view the alert record, select:
Diagnostics menu>Instrument Alert Record.
The recorded alerts and the time and date the first alert occurred is
displayed.
99
With Batch Runner you can setup the ValveLink software to auto-
matically run diagnostic tests, calibrate, or upload configuration data
to multiple valves with a user specified routine.
During a turn-around or production change over, you can upload
configurations, run the Performance Tuner to optimize tuning, or
even reset the instrument clock without any interaction by personnel.
Batch runner reduces redundancy and increases consistency by
allowing the user to set up a batch once, then repeatedly run it for
different groups of valves.
1010
From the Tag Management window, you can print a report that
contains selected information for all the tags listed in the Tags Found
box. whether they are highlighted or not. You can use one of several
methods to print the report.
D From the menu bar select :
Tag>Open>Modify Tag.
D From the Tag menu. In the Tag Management window, enter
into the Tag or Partial Tag field the tag or several characters of those
tags you want to appear on the report. If necessary, use Tag Groups
to limit the tags in the Tags Found box. When the list of tags is
correct, click the Print List button at the bottom of the window.
D From the toolbar, click the Tag icon. In the Tag Management
window, enter into the Tag or Partial Tag field the tag or several
characters of those tags you want to appear on the report. If
necessary, use Tag Groups to limit the tags in the Tags Found box.
When the list of tags is correct, click the Print List button at the
bottom of the window.
The Print Tags Report window is shown in figure 10-1.
You can select the tag information that appears on the report by
checking the box to the left of the item.
Note
Selecting all the users in the list box is not identical to
no filter. Some log entries do not have users associated
with them, and a blank user is not listed in the list box.
Selecting all the users will display all log entries except
those with no user name.
1111
This section contains basic information about Working with the DataSync
Program, including:
After you log in, you must choose the two databases with which you
wish to work.
D The source database includes the tags you wish to copy.
Once you have selected the databases, you must select the tags
with which you wish to work. As shown in figure 11-2, DataSync
shows you all of the tags in the source database. Select only the
tags you want to move to the destination by using the ⇐, ⇒, All ⇐,
and All ⇒ buttons.
D To move a group of tags to the Selected tags box, highlight
them in the Tags in source database box and click the ⇒ button.
D To move all of the tags to the Selected tags box, click the All ⇒
button.
D To remove a group of tags from the Selected tags box, highlight
them in the Selected tags box and click the ⇐ button.
D To remove all of the tags from the Selected tags box, click the
⇐ All button.
On the left side of the screen, you can see the group names with a
drop-down list of the subgroups in each group. If you select a
subgroup from a drop-down list, only tags in that subgroup will be
displayed in the Tags in source database box.
DataSync can work with databases on two computers that are con-
nected by a parallel or serial cable. The following describes how to
use the Direct Cable Connection program that comes with Windows.
If you do not see Direct Cable Connection on the Accessories menu,
it is not installed. Refer to Windows Help for information on installing
missing Windows components.
The computer running DataSync is the Guest computer. The other
computer is the Host computer.
Step 1: On the Host computer, make sure the drive and/or folder
containing the ValveLink database is shared.
Step 2: To make a folder shared, right click on it and select
Sharing.
Step 3: Refer to figure 11-4. On the Sharing tab, select Shared
As and enter the folder name in the box labeled Share
Name
Step 4: Click OK to close the properties window.
Step 5: Connect the two computers with a null modem serial or
parallel cable.
Step 6: On both computers, run Direct Cable Connection by
selecting: Start>Programs.
Step 7: From the programs menu, select Accessories and Direct
Cable Connection. If Direct Cable Connection is not
installed, install it from the Windows CD-ROM by run-
ning Add/Remove Programs from the Control Panel.
Step 8: If Direct Cable Connection has not been run on the com-
puters, a connection wizard will take you through the
process of setting up the connection. If Direct Cable
Connection has been run, the computer shows you the
current settings, as shown in figure 11-5.
Step 9: If changes are required, click on Change to set up the
connection on both computers.
Step 10: On the Host computer, select Host and the correct port
for the connection.
Step 11: On the Guest computer, select Guest and the correct
port for the connection.
Step 12: If the Direct Cable Connection does not automatically try
to connect, click Listen on the Host computer and Con-
nect on the Guest computer.
Step 14: After the computers connect, a window should pop up,
as shown in figure 11-4, on the Guest computer that
displays the shared folders on the Host. Right click the
folder you are going to use and select Map Network
Drive. Select a drive letter and click OK.
Help! Can’t Transfer If you’re running Query & Match from the Network menu,
Data in ValveLink will not transfer data between an instrument and
Query & Match. a database tag if the transfer arrows are dimmed rather
than black
If you can’t transfer data:
Check Security Do you have security autho-
Authorization. rization to transfer configura-
tion data? Ask your Valve-
Link administrator.
Check Do the instrument’s mode
Instrument Mode and and protection settings
Protection. match parameter require-
ments? Check the Instru-
ment Mode and Instrument
Protection requirements for
the parameter you are trying
to transfer.
Help! Can’t Upload or To poll an instrument for configuration parameters or to
Download upload data, the instrument must be connected to and
Configuration. communicating with the ValveLink software. On the Config-
uration window, if the Read Inst button is dimmed or if all of
the Download/Upload arrows are dimmed, the instrument is
probably not connected.
If you can’t poll the instrument or upload or download data:
Check Network Check tag information on
Connection. the toolbar and verify
ValveLink is connected to
and communicating with the
instrument.
Help! My Batch If you can’t run an operation using Batch Runner from the
Operations Won’t Run. Tools menu:
Create a Tag. Does the tag exist in the da-
tabase?
Check Network Check tag information on
Connection. the status bar and verify
ValveLink is connected to
and communicating with the
instrument.
Check Instrument Mode. Some operations require the
instrument mode to be Out
of Service. Check tag in-
formation on the status bar
and verify the Instrument
Mode is set to Out of Ser-
vice.
Check Instrument Verify the Instrument
Protection. Protection is set to Configu-
ration or None.
Check Instrument Level. The instrument may not
support the operation you
are trying to run.
Bench Set
The pressure range required to move
Actuator Return the actuator from its upper stop to the
In a piston actuator, the pressure valve’s rated travel without any valve
opposing the stroke pressure (typically frictional forces present.
the pressure below the piston).
boot or boot up
The process of initializing software to
alert point hardware to bring the software to a
The values set to trigger an alert. known state.
EU free time
Acronym: Engineering Units Percent of time that the microprocessor
is idle. A typical value is 25%. The
actual value depends on the number of
feedback arm functions in the instrument that are
The mechanical connection between enabled, and on the amount of
the valve stem linkage and the communication currently in progress.
instrument’s travel sensor.
gain
feedback connection The ratio of output change to input
Identifies to the FIELDVUE instrument change.
how the feedback linkage is connected
to the actuator.
hardware revision
Revision number of the Fisher Controls
feedback signal instrument hardware. The physical
Indicates to the instrument the actual components of the instrument are
position of the valve. The travel sensor defined as the hardware.
provides the feedback signal to the
instrument printed wiring board
assembly. A mechanical linkage HART
connects the travel sensor to the valve HART stands for Highway Addressable
stem or shaft. Remote Transducer.
Instrument level
Determines the functions available for
Import the FIELDVUE instrument.
Bringing in tag data from an export file
generated by another ValveLink
program. Instrument mode
Determines if the instrument should
respond to its analog input signal. There
are two instrument modes: In service
and Out of Service.
Input characteristic
The relationship between the ranged
travel and ranged input. Possible values ISA
include linear, equal percentage, and
Acronym: Instrument Society of
quick opening.
America
leak class
Input current Defines the allowable leakage of a valve
The current signal from the control when it is closed. Leak class numbers
system that serves as the analog input are listed in two standards: ANSI/FCI
to the instrument. 70-2-1991 and IEC 534-4 (1986)
parallel
measured current Simultaneous: said of data transmission
The output signal as measured by a on two or more channels at the same
measurement loop and displayed on the time.
test screen.
plant area
The collection of equipment in a plant
that has common manufacturing
strategies and alarm strategies.
memory
A computer’s storage for programs and
data. Most computers and positioner plot
microprocessors have both internal and A graph that plots positioner input signal
external memories: use of internal (pressure) versus travel and provides
memories is faster, but the capacity of analysis of positioner calibration and
internal memories is more limited. performance.
resolution
primary master The least interval between two adjacent
discrete details, which can be,
Masters are communicating devices. A
distinguished one from the other.
primary master is a communicating
device typically permanently wired into
the instrument in the field. You can response time
establish either a communicating device The measured time required for a valve,
in the control room or a PC running or component of, to respond to a given
ValveLink software as the primary position.
master for an instrument. In contrast, a
hand-held communicator is a secondary
master because it is not permanently ROM
wired into the instrument. Acronym: Read-Only Memory
seat load
The contact force between the seat and
RAM valve plug when the valve is at its
Acronym: Random Access Memory closed position.
software
random access memory (RAM) Microprocessor or computer programs
A type of semiconductor memory. A and routines that a user can change.
user can read from and write to a RAM,
electrically, as often as desired. spring rate
The force change per unit change in
length.
read-only memory (ROM)
A memory in which information is stored stroking time
permanently. A user can examine ROM The time, in seconds, required to move
contents as often as desired but cannot the valve from its fully open position to
change the contents. fully closed, or vice versa.
A udit Log
Filtering D
Applying Multiple Filters, 10-9
Database Tags
by Date and Time, 10-9 Copying, 3-11
by Event, 10-9 Deleting, 3-11
by Tag, 10-8 Exporting Data, 3-13
by User, 10-8 Importing Data, 3-12
R emoving Filters, 10-9 Modifying, 3-10
P rinting, 10-11
Datasets, 6-4
Sorting, 10-10
T ransfer ing Data, 6-6
T oolbar, 10-6
V iewing, 10-4 DataSync
Destination Database, 11-3
Direct Cable Connection, 11-14
Operation, 11-7
B Setting Preferences, 11-13
Source Database, 11-3
Batch Runner, 9-2 T ag Selection, 11-5
Detailed Setup, 6-2
Diagnostic Tests, 8-2
C Cancelling,
L ow Power
8-6
Warning, 8-7
R unning, 8-6
Calibrating
Units of Measure, 8-7
Analog Inputs, 7-6
Canceling Calibration, 7-3
L ow Power Warning, 7-4 F
P res ure Sensor, 7-14
R estoring Factory Calibration, 7-5 F actory Calibration, Restoring , 7-5
T ravel
A uto,
Manual,
7-8
7-12
H
T ravel Sensor, 7-13 HAR T Modem, Instal ing, 2-4
T
P
T ag Groups, 3-9
P asswords, User, 3-6
T ag Reports, 10-2
P erformance Tuner, 5-4 T ags
P references Database, 3-8
A uditL og page, 3-16 HAR T, 3-8
Communication page, 3-15 T echnical Support, 1-8
Diagnostics page, 3-15
ESD page, 3-17 T oolbar
General page, 3-16 A uditL og Viewer, 10-6
HAR T Modem page, 3-16 V alveLink, 4-3
L anguage page, 3-17 T rending
Modbus page, 3-16 Setup, 8-9
Multiplexers page, 3-15 T ravel History, 8-10
Network Alert Scan page, 3-15 T rend List, 8-9
Units page, 3-15 V iewing, 8-9
Fisher
Marshalltown, Iowa 50158 USA
Cernay 68700 France
Sao Paulo 05424 Brazil
Singapore 128461
www.Fisher.com