Calibration of Thermometers: de La Salle University
Calibration of Thermometers: de La Salle University
Calibration of Thermometers: de La Salle University
Experiment # 1
Calibration of Thermometers
Calibration of Thermometers
Kimberley B. Cellona, Lloyd Ross M. Dar, Mona Lyndsay C. Javier,
Maristela Ann F. Lopez, Elijah Jeremie D. Soriano
Department of Chemical Engineering, Gokongwei College of Engineering, De La Salle University,
2401 Taft Avenue, Manila 1004, Philippines
T 1−T 2
B= (2)
P1−P2
Figure 2. Setup for the comparison of temperature readings for
Mercury thermometers and the dial thermometer.
Where B is the external pressure coefficient, expressed
in F/psia, T 1 and T 2are the temperatures read at P1 and
o
II. METHODOLGY
2.2 Equipment Figure 3. Block calibrator and setup for calibration above
This experiment involved four set-ups: stream bath, boiling temperature of water.
ice bath, temperature-controlled water bath, and a
block calibrator. Three mercury thermometers, a dial 2.3 Experimental Procedure
thermometer, and a thermocouple were used to
measure the temperature in the different set-ups 2.3.1 Calibration Based on Physical Properties of
Pure Substances
An amount of ice was placed in a beaker and
cold water was added. The initial temperature
of the thermometer was noted before
Cruz, Dela Cruz, Francisco(your last names only) Dr. Lawrence P. Belo
[2]
submerging it into the ice bath (Note: The bub thermometer was also recorded with the
must not touch the surface of the beaker.). The thermocouple’s temperature. This procedure
temperature was recorded every 15 seconds was repeated at 10 oC intervals until 200 oC.
until the thermometer reaches a constant
reading. The constant reading is the melting
point reading for the particular thermometer. In the analysis of the data, it is noted that the
The whole procedure was repeated for all measurement of the length of Mercury immersed in
temperature reading devices. the liquid was not conducted in the procedure, and
therefore Eq. 1 cannot be used. Further, [4] also
An amount of water was placed in the steam illustrates that the correction factors obtained from
bath apparatus. The apparatus was heated applying the equation are minimal, and thus for this
using a Bunsen burner. Once the water is experiment will be assumed negligible. Lastly, since
boiling, the initial temperature of the the experiment was conducted in atmospheric
thermometer was noted begore placing the pressure, Eq. 2 will also not be used.
thermometer on top of the boiling water (Note:
The bub must not be submerged underwater.).
The temperature was recorded every minute III. RESULTS & DISCUSSION
until the thermometer reaches a constant
reading. The constant reading is the boiling 3.1 Calibration based on Physical Properties of Pure
point reading for the particular thermometer. Substances
The whole procedure was repeated for al With expected readings of 0oC and 100oC for the
temperature reading devices. melting point and boiling point of water under
atmospheric pressure, the five thermometers
2.3.2 Comparison of Temperature Readings with examined exhibited different readings from the same
a Standard Instrument steam bath and ice bath setups, thus highlighting the
In this part of the experiment, the need for calibration. Fig. 4 shows a bar chart
thermocouple connected to the Fluke digital illustrating the melting and boiling points observed
monitor was considered as the standard for each device, and their difference from the 100 oC
instrument. The thermometers and and 0oC line.
thermocouple were hanged using a support
and were submerged in water contained in a Melting Point
beaker. The beaker was then heated until the 120
98.9 98.2 100 99.5 99.6
Fluke digital monitor reading was 100 oC. The 100
temperature reading of the other thermometers 80
were then recorded. The procedure was
60
repeated for every 5 oC drop on the Fluke
Temperature (oC)
40
digital monitor reading until the reading
reaches 60 oC. 20
0.1 0.0 0.0 2.0 0.0
0
Mecury Thermom...
Mecury Thermom...
Mecury Thermom...
Dial Thermometer
Thermocouple
Temperature (oC)
280
Table 1. Summary of accuracy of thermometers according to
melting point and boiling point of water at 1 atm. 220
90
attributed to the loss in calibration of the thermometer
with use over time. In addition, human bias may still be 80
considered as a contributing factor, however, given the
70
reproducibility of the readings from both directions, may
be considered minimal. 60
50
100
40
Temperature Reading (oC)
90 40 50 60 70 80 90 100
0.00
40.00 55.00 70.00 85.00 100.00 210
Standard Temperature (oC)
190
Temperature Reading (oC)
REFERENCES
Cruz, Dela Cruz, Francisco(your last names only) Dr. Lawrence P. Belo
[10]
APPENDICES
A. Raw Data
Table 7. Using block calibrator for temperatures above the boiling point of water.
Block Mercury Error Thermocouple Error
Calibrator (oC) thermometer (oC) (%) (oC) (%)
100 102 2.00 100 0
110 112 1.82 110.3 0.273
120 122 1.67 120 0
130 132 1.54 130 0
140 142 1.43 140 0
150 152 1.33 150 0
160 162 1.25 160.4 0
170 172 1.18 170 0
180 182 1.11 180 0
190 194 2.11 190 0
200 202 1.00 200 0
B. Sample Calculations
Percent error for thermocouple at standard temperature 110oC in calibration with block calibrator
|110℃−110.3 ℃|
Error= ×100=0.273 %
110 ℃
Cruz, Dela Cruz, Francisco(your last names only) Dr. Lawrence P. Belo
[12]
Cruz, Dela Cruz, Francisco(your last names only) Dr. Lawrence P. Belo
[13]