MA-3000 NIC-600-2231-11 Instruction Manual PDF
MA-3000 NIC-600-2231-11 Instruction Manual PDF
MA-3000 NIC-600-2231-11 Instruction Manual PDF
NIC-600-2231-11
Table of Contents
1. INTRODUCTION ................................................................................................................ 1
3. SPECIFICATIONS .............................................................................................................11
4. SYSTEM INSTALLATION................................................................................................ 12
5. MEASUREMENT .............................................................................................................. 23
6. MAINTENANCE ............................................................................................................... 42
7. MA3WIN............................................................................................................................ 72
8. CONTACT ........................................................................................................................118
NIC-600-2231-11 1. Introduction
1. Introduction
For the safe use of this product, please observe the following precaution.
In no event shall Nippon Instruments Corporation, Inc. be liable for any direct,
indirect, punitive, incidental, special consequential damages, to property or life,
whatsoever arising out of or connected with the use or misuse of our products.
Please refer to the attached documents for the optional reagent dispenser and sample
changer.
1
NIC-600-2231-11 1. Introduction
Do not directly measure the sample decomposed with nitric acid. The
Warning
activated carbon in the exhaust gas filter may generate heat. Dilute the
sample until the concentration of nitric acid falls to 5wt% or less before
the measurement.
Do not disassemble or modify the product. This may cause a
Caution
malfunction.
Use the product after consulting the manual. Improper operation
may cause malfunction or bodily injury.
Comply with the range of specifications. Failure may occur with
misuse.
Do not operate with wet hands.
Maintenance of the heater must be carried out after the power has
Warning
been turned OFF and the heater has cooled down to close to normal
room temperature or it could result in burns.
Shut OFF the power supply before staring the maintenance work.
Caution
Otherwise electric shock may occur.
Be sure to do degreasing cleaning the O-ring before use. If degreasing
cleaning is not carried out, there is a possibility that the sample heating
tube will come off during measurement.
2
NIC-600-2231-11 1. Introduction
It is strongly recommended to carefully read this manual before using the system in order
to ensure that its full performance can be utilized.
Mercury/MA-3000 System
3
NIC-600-2231-11 2. Operating principle
2. Operating Principle
2.1. Operation
The sample heating furnace (H-1) heats samples to evaporate any mercury compounds
contained in them. The decomposing furnace (H-2A, H-2B) then decomposes the mercury
compounds in the gases generated in H-1 to remove any interfering ingredients from
them.
The mercury collector (H-3) collects the atomized mercury gas in the form of gold
amalgam, removes any interfering ingredients, and condenses and purifies the mercury.
After the heat decomposition sequence has completed, the mercury collection tube is
heated to liberate the mercury again as mercury gas, and the absorbance is then
measured.
Cold-vapor atomic absorption at a wavelength of 253.7nm is used in the detection.
Pressure regulator
Mercury Activated
collecting tube carbon 1
Lamp Photoelectric
tube SIG1 Exhaust Heating
Light absorption vaporization
cell 2 Exhaust
Tray rack Light absorption
cell SIG2
Photoelectric
tube REF
Activated
carbon 2
Drain tank
Silencer
Flow Diagram
4
NIC-600-2231-11 2. Operating principle
2.2.1. Front
⑤
④
②
5
NIC-600-2231-11 2. Operating principle
③ ①
②
⑤
⑥
④
6
NIC-600-2231-11 2. Operating principle
2.2.3. Rear
②
①
⑤
③ ④ ⑥
⑦
⑧
⑨
7
NIC-600-2231-11 2. Operating principle
④
③
⑤ ①
⑨ ⑦
8
NIC-600-2231-11 2. Operating principle
③
① ②
⑤ ⑦
⑥
⑧ ⑨
9
NIC-600-2231-11 2. Operating principle
10
NIC-600-2231-11 3.Specifications
3. Specifications
11
NIC-600-2231-11 4. System installation
4. System installation
4.1. Conditions for installation
4.1.1. Environment for installation
12
NIC-600-2231-11 4. System installation
* If there is any risk of generation of any harmful gasses such as halogen, sulfur and
selenium during decomposition by heating, please be sure to vent the exhaust gas.
* Use a >4mm tube to connect the exhaust gas tube. Be noted that accurate
measurements will be impossible if the tube is too thin or bended.
Power cable
Oxygen tube
13
NIC-600-2231-11 4. System installation
*IF the OS language setting is not English (United States), please set as follows.
Control Panel – Clock, Language, and Region – Change date, time, or number formats
Format: English (United States)
(1) Insert the supplied CD-ROM. (The installer will start automatically.)
* If any other application programs are running on Windows, close them.
(2) The “Welcome to the MA3Win Setup Wizard”
window will open. Click Next.
14
NIC-600-2231-11 4. System installation
15
NIC-600-2231-11 4. System installation
If MACROMEDIA FLASH PLAYER is not installed on PC, install it from the supplied
CD-ROM.
In Explorer, open the folder “Macromedia Flash Player” and execute ”install_flash_player_
_active_x.msi” from the CD-ROM.
16
NIC-600-2231-11 4. System installation
Use the supplied USB-Serial converter REX-USB60F to control the MA-3000 from a PC
without a serial port.
* Operation is guaranteed only when REX-USB60F USB-Serial Converter is used.
Windows7
Execute “USB60F_Setup(Windows7).exe” located in
“USBDriver(REX-USB60FverXX)Windows7” of the supplied CD-ROM.
* Complete the following setup procedures before connecting the
REX-USB60F to the PC.
17
NIC-600-2231-11 4. System installation
The RS232C port needs to be set to enable the communication. Follow the steps below to
set the port.
18
NIC-600-2231-11 4. System installation
(4) Select the communication port checked in Step (1) and click OK to apply the setting.
(5) Close MA3Win and restart it for the setting to take effect.
* The port number changes depending on the port to which the USB-Serial
Converter is connected. Note that the setting needs be made again if the
converter is connected to another port.
19
NIC-600-2231-11 4. System installation
Windows7
To upgrade MA3Win, first uninstall it and then install the upgraded version.
20
NIC-600-2231-11 4. System installation
4.3.7. How to upgrade the version of ROM built into the MA-3000
Follow the steps below to upgrade the version of the ROM built into the MA-3000.
Select [Run]-[System]-[Firmware Update] from the menu. The following Firmware Update
window will appear.
21
NIC-600-2231-11 4. System installation
22
NIC-600-2231-11 5. Measurement
5. Measurement
Enter measurement
conditions
Start measurement
circuit
Heating collection tube / 7 Clean the
PURGE2 15sec.
Measuring absorbance measurement circuit
8 Measurement MEASURE 30sec.+7sec.
9 Clean the mercury
collection tube MEAS PURGE 30sec.
Cooling Cool H1
10 Cool H1 and H3 COOL 20 sec.
Boat unloading
End
23
NIC-600-2231-11 5. Measurement
5.2. Samples
Warning
· Do not directly measure the sample decomposed with nitric acid. The activated
carbon in the exhaust gas filter may generate heat. Dilute the sample until the
concentration of nitric acid falls to 5wt% or less before the measurement.
24
NIC-600-2231-11 5. Measurement
The concentration of a mercury standard solution tends to change significantly after being
prepared, which can cause variations in the calibration curve or low reproducibility of a
measurement. L-cysteine should therefore be added to the dilution water of a mercury
standard solution to control the change in concentration.
(Preparation procedure)
* The L-cysteine solution and the mercury standard solutions prepared can be
used for 6 months if they are stored in a cool dark place.
* The stability of the solutions will deteriorate as they age or if kept anywhere
warm. Store in a cool dark place and quality control by, for example, labeling
it with the date prepared.
* Use the same measuring flask if possible when re-preparing the standard
solutions again. If a new flask is used, ensure to thoroughly clean it with
(1+10) nitric acid. Particularly when preparing a standard solution of 1ppm or
less, care should be taken to avoid contamination from the flask by
thoroughly cleaning with acid the flask itself along with the plug.
* When constructing the calibration curve of 20000ng or greater, put about
100μL of L-cysteine solution into the sample boat and directly add drops of
the mercury standard solution of 1000ppm (mercury chloride).
25
NIC-600-2231-11 5. Measurement
<Daily Work>
Accurate measurement is achieved by the sample boats being heat-treated before
measurement to remove any trace quantities of mercury. To heat-treat the sample boats,
repeat measurements on the MA-3000, or heat them in an electric furnace at 750°C for
around 3 hours.
Sample boats must be cooled down after the blank treatment and stored in a hermetically
sealed container.
* Use different boats for high and low concentration samples. When a liquid sample
is measured, in particular, mercury penetrating in a boat seeps out and the
outputted measurement value becomes higher than the actual value.
With measurements for some types of samples such as highly volatile substances, the
mercury contained in them can evaporate during the interval of the sample being placed
on the sample changer to starting measurement. Additive B is sometimes added to
prevent this phenomenon.
Additive B absorbs any mercury in the environment during the manufacturing process and
storage. Use of any additive B that has not undergone blanking treatment can lead to
falsely high mercury content in the measurement, thus making blank treatment of the
additive necessary before being used.
Heat the additive in a crucible at 750°C for about 3 hours as the blank treatment. Any
additive that has undergone the blank treatment must be stored in a hermetically sealed
container or used immediately after having cooled down or it will absorb mercury from the
atmosphere, thereby increasing its blank value.
26
NIC-600-2231-11 5. Measurement
Click the shortcut icon on the desktop, or click [Start] and select [NIC] – [MA3Win] –
[MA3Win] from [All programs].
The icons are indicated in gray if RS-232C port does not exist.
Connect the USB serial converter REX-USB60F to the USB
port, and set the port number again.
27
NIC-600-2231-11 5. Measurement
Open the main cock of the oxygen gas cylinder and adjust the pressure regulator on the
rear panel of the MA-3000 to 0.05 through 0.07 MPa.
Gas supply, leak, and operation of the valves from V1 through V3 are checked. The check
proceeds automatically but the diagnostic software automatically closes if any problem is
detected. In that case, turn OFF the power once, correct the problem and then turn ON
the power again.
When the MA-3000 is powered ON and the start-up operation of the sample changer is
completed, “Check of flow path” window opens. Select Yes and the check of flow path
starts.
(Refer to details of the flow of the flow path check in the next section.)
If you do not wish to execute the check of flow path, select No and then the following
dialog box appears.
If you wish to heat the heaters, click OK to start heating.
* Once you cancel the check of flow path, the "Check of flow path” window is not
displayed until theMA-3000 main unit is powered ON again.
28
NIC-600-2231-11 5. Measurement
* If the mass flow controller's zero point is displaced due to insufficient adjustment, a
failure may be caused. If the MFC does not indicate zero when the flow rate is zero,
adjust zero point using a zero point adjustment trimmer of the MFC.
(6) When all the checks are successfully completed, the heaters automatically start the
heating.
29
NIC-600-2231-11 5. Measurement
5.6.5. Conditioning
If the check of flow path is successfully completed, the conditioning process starts. The
temperature of each heater is increased with oxygen so that the system is purged to
remove moisture and other impurities adhering on the sample heating tube.
* Each heater is heated up to the temperature set for measurement. The set
temperature of the heater H1 is 150˚C and that of H3 is 500˚C. Click Stop if you wish
to stop the conditioning.
* If there is no abnormality, be sure to perform conditioning. Otherwise, the sample
heating tube absorbs moisture, it may be obstructed or break.
(1) The liquid in the drain tank can be discharged by turning the roller of
the roller clamp downward. Collect the waste water in an appropriate
container and dispose of it in the correct manner.
(2) Close the roller clamp.
* Do not apply any excessive force or you may break the indentations on
the clamp.
30
NIC-600-2231-11 5. Measurement
5.7. Measurement
The blank of the system is checked. Perform the blank measurement until the values are
stabilized because mercury adhering to the flow path is released when the measurement
is started.
Click the icon to perform the purge measurement or register the blank check on the
table as in the normal measurement. If the boat number is set to 0, it is possible to check
only the system blank.
After the measurement, if the sample boat is contaminated, it should be cleaned with
neutral detergent, dried, and then heat-treated in the same manner as above.
Click the icon to open “Measurement Condition” screen and set the flow rate for
measurement, the standard heating mode (method), the standard check and the blank
check.
31
NIC-600-2231-11 5. Measurement
When the blank check functions, a line of BlankCheck is inserted below the corresponding
sample line.
32
NIC-600-2231-11 5. Measurement
Click the icon to open “Table Condition.” Select the type of sample, unit and other
conditions. The settings can be changed after the measurement. (For details, refer to
7.2.4.2 Table condition)
Analysis mode:
Select either PEAK (Peak
height) or AREA (Peak area).
Usually AREA is selected and
PEAK should be selected in the
case the peak is so small that
the variation becomes large if
AREA is used.
Type of sample:
Select either Solid or Liquid,
which changes only the
displayed unit and items but not
affect the method.
* The measurement method is set as “STD” initially. When using other CRM (especially
solid CRM), select proper measurement method from measurement condition.
33
NIC-600-2231-11 5. Measurement
Place the standard samples in the special sample boats side by side on a boat tray and
then set on a rack.
34
NIC-600-2231-11 5. Measurement
Select [Run]-[Measurement start] from the menu or click the icon to start a
measurement.
As the measurement starts, the system
asks you the name of the file in which
the measurement data should be stored.
Set the file name.
Click Save and then the measurement
starts.
The following message will appear when the system has completed measurement of all the
samples whose boxes were checked.
To stop a measurement, select [Run] – [Stop] from the menu or click icon.
This will result in measurement ceasing after the current measurement has completed.
In this state, if [Run] – [Emergency] from the menu or the icon is clicked, the system will
immediately switch to the cooling sequence, and then returns to READY mode on completion
of the cooling process.
35
NIC-600-2231-11 5. Measurement
Click the icon and check the calibration curve using the “Calibration curve” window
on the analysis screen.
Check respective calibration curves for
LOW and HIGH.
The sensitivity of each system is different and the “WARNING” may be displayed
depending on the values set during the software installation. Set the values with reference
to the data of the shipping inspection and the measured values at the time of delivery or
after the maintenance.
* X (employ or ignore the data) is automatically entered according to the height of PEAK.
Change the X as necessary.
(LOW is adopted when ABS is 0.3 or less, while HIGH is adopted when ABS is 0.3 more.)
36
NIC-600-2231-11 5. Measurement
Place the sample in a sample boat. If the sample is liquid, measure it with a micropipette
or if it is a solid, with a balance.
* When an additive is used, put it into a boat before weighing the sample.
37
NIC-600-2231-11 5. Measurement
38
NIC-600-2231-11 5. Measurement
Method Standard
Type Method amount of
Number sample
Organism (Solid)
1 Solid sample 200mg Ordinary solid samples.
Fish meat/Plant
Select [Run] – [Measurement start] from the menu or click the icon to start a
measurement.
39
NIC-600-2231-11 5. Measurement
Review the measurement results and determine whether to use the data or not.
* As for the analysis range, the data less than 10ng are automatically applied for LOW
range and the data greater than 10ng are applied for HIGH range. An analysis range
can be changed and is performed in the Range column.
* If the recovery of the standard solution after the measurement is 90% or less (≤ 90%),
the sensitivity might decrease due to poor decomposition of the sample. Try to reduce
the amount of sample or reconsider the method.
* When the sensitivity decreases, repeated measurements of water may recover the
sensitivity. The sensitivity can be recovered and checked at the same time by
repeating the measurements of the standard solution of around 200μL.
To print the measurement results, select [File (F)] – [Print] from the menu, or click the
icon. (For details, refer to 7.2.1.6 Print)
Number of
profiles per
line
● The width of each column synchronizes with the setting in the table screen. Adjust the
width of column on the table screen if the column seems too narrow when printed. A
column is not printed if its width is set to 0.
40
NIC-600-2231-11 5. Measurement
5.7.6. Termination
Caution
*If you don’t turn-on MA-3000 for a long time, the sample heating tube might be
obstructed by moisture absorption.
・If you don’t turn-on MA-3000 more than 2 weeks, please turn on the MA-3000 power
switch and do conditioning. (Conditioning process start automatically after checking the
flow path)
・Otherwise, please dismount the sample heating tube and store it in a desiccator to
prevent moisture absorption. (Refer to 6.4.2 Dismounting the sample heating tube and
6.4.3 Mounting the sample heating tube)
(2) Close the Windows and turn off the power of PC and Printer.
(3) Power off MA-3000.
(4) Discharge liquid of a drain tank.
(5) Close the main cock of the Oxygen gas cylinder.
41
NIC-600-2231-11 6. Maintenance
6. Maintenance
Sample changer All boats must be retained without falling out. Contact our service
when the problem is
not solved by the
adjustment.
Unusual noise No noise is generated while the sample Grease the drive
changer is running. mechanism.
Blank value Repeat purge measurements, and the peak 6.3.2 Outlying blank
value at Low range must be at most 0.001. value
Calibration curve When the calibration curve is drawn for the 6.3.1 Degradation of
standard solution, the correlation coefficient peak sensitivity
must be at least 0.999.
Sensitivity Gradient of the calibration curve must be within 6.3.1 Degradation of
±30% of the inclination at delivery. peak sensitivity
Carry over The blank value immediately after Replace the sample
measurement of a standard solution must be at heating tube/mercury
most 2% of the mercury amount measured just collection tube.
before it.
Accuracy The value of a standard solution measured Review the method and
after an actual sample measurement must be the amount of sample
within ±5%.
42
NIC-600-2231-11 6. Maintenance
Annual inspection
Inspection criteria Criteria Corrective action
O-rings at sample inlet No cracks on the O-rings Replace the O-ring with a
new one.
Sample inlet fork No cracks on the fork Replace the sample inlet.
Drive unit linear guide No rust on the operating shafts Grease the shafts.
Sample heating tube Catalyst retained in its place without Replace the sample heating
displacement tube.
Mercury collection tube No contamination on the quartz tube, Replace the mercury
and no discoloration of quartz wool. collection tube.
Absorption cell No contamination Tear-down and clean or
replace the cell
Pressure regulator No variation in pressure Replace the pressure
regulator.
Flow rate No flow variation exceeding 0.02 Replace the mass flow
L/min. controller.
Detector output Must be 4±0.5V in the maintenance Wipe off any dirt on the
(Ref) window of MA3Win. lamp/lens etc.,
Adjust optical systems.
Detector output Must be 4±0.5V in the maintenance Clean the cell unit or replace
(SIG1, SIG2) window of MA3Win. it. Wipe off any dirt on the
cell window and
photoelectric tube
Air pump tube (Option) Needs to be replaced every year. Replace.
43
NIC-600-2231-11 6. Maintenance
STAND-BY xx:xx Appears for 20 minutes after turning the power ON.
The measurement data are not reliable during the count
down because the heaters and optical systems are not
stable.
The heater (%) interlock has occurred and Displayed when the heater interlock has occurred. Contact
the heater (%) turned OFF. our service if there is any odor. When no deficiency is
Turn ON the power again if there is no observed in the appearance, turn OFF the power and then
problem found. turn it ON again, and check if the error continues.
THERE IS A TRAY IN This warning is generated when there is any boat tray left
There is a tray in the tray evacuation rack. on the tray evacuation rack. Remove the tray and confirm
Remove the tray. that there are all the boats on the tray. If not, the sample
boats are left on the inlet shaft. Turn OFF the power of the
system and open the front cover to remove the sample
boats.
A tray is loaded into a tray rack which is under Do not load a tray into a tray rack which is under
measurement. Remove the tray. measurement. If the warning message is displayed when
the tray is not loaded, it is considered to be caused by the
influence of outer light or a failure of the sensor.
The number of measurements has exceeded Displayed when the number of measurements has
2000. It is recommended to replace the exceeded 2000.
sample heating tube. It is recommended to replace the sample heating tube
because its performance may be degraded. Be sure to
reset the counter after the replacement. (Maintenance
=>GLP=>MEAS CNT CLEAR)
There is a problem with the case fan. The rotation of heat exhaust fan is not detected. If the fan
If the case fan is not rotating, stop the use of is not rotating, the temperature in the unit will rise and the
the system. system may be damaged.
The heater (%) does not reach the pre-heat The heater does not reach the set temperature. When the
temperature. set temperature is reached, the message automatically
disappears.
44
NIC-600-2231-11 6. Maintenance
If any of the following errors occurs during a measurement, the system will cease
measurement and switch to the cooling sequence. Once the temperatures of heaters H1
and H3 has cooled, the system will stop.
45
NIC-600-2231-11 6. Maintenance
11 HEATER (H5) BROKEN The error is issued when the temperature of H5 does
The temperature of H1 does not rise to not reach 90% of the pre-heating temperature 30
required pre-heating temperature minutes after turning on the instrument.
Turn OFF the power once and turn it ON again and,
if the error repeats, contact our service.
12 HEATER (RH1) BROKEN The error message is issued when the temperature
The temperature of the 1st mercury collector of the heater does not reach 400˚C at the end of
tube of RH-MA3 does not rise. Check the ATOMIZE process.
connection and power of RH-MA3. Check the power cable and the switch of RH-MA3,
and, if no problem is observed with them, contact our
service.
13 HEATER (RH2) BROKEN The error message is issued when the temperature
The temperature of the 2nd mercury collector of the heater does reach 400˚C at the end of
tube of RH-MA3 does not rise. Check the ATOMIZE process.
connection and power of RH-MA3. Check the power cable and the switch of RH-MA3,
and, if no problem is observed with them contact our
service.
46
NIC-600-2231-11 6. Maintenance
If any of the following errors occurs during a measurement, the system will immediately
cease measurement and stop.
Message Description and corrective action
14 BC MOTOR-X TIME OUT ERROR An error has occurred when the home position
The movement of X axis motor does not sensor of the horizontal shaft moved.
complete. Turn off the system, and remove Turn OFF the power and open the front cover to
the tray on the X axis. check it.
If the system cannot be recovered, contact our
service.
15 BC MOTOR-Y TIME OUT ERROR] An error has occurred when the home position
The movement of Y axis motor does not sensor of the vertical shaft moved.
complete. Turn off the system, and remove Turn OFF the power and open the front cover to
the tray on the Y axis. grease the linear guide and the trapezoidal screw
thread.
If the system cannot be recovered, contact our
service.
16 BC MOTOR-Z TIME OUT ERROR An error has occurred when the home position
Turn off the system, and remove the boat on sensor of the inlet shat moved.
the boat handling poker. Turn OFF the power and open the front cover to
grease the linear guide and the trapezoidal screw
thread.
If the system cannot be recovered, contact our
service.
17 BC LIFT TIME OUT ERROR An error occurred when the lift went up/down.
The movement of lift motor does not Turn OFF the power, open the front door, and adjust
complete. Turn off the system, and remove the sample changer with BC ADJUST. (Refer to
the boat on the lift. 6.4.7Adjusting the sample changer.)
If the system cannot be recovered, contact our
service.
18 NO TRAY!! Open the front cover to check. If a tray is not placed
The try is not placed on the arm correctly. Turn on the arm correctly due to displacement of the
off the system, and check the tray sensor of sample changer, adjust the sample changer. (Refer
the arm. to 6.4.7 Adjusting the sample changer)
If the system cannot be recovered, contact our
service.
19 BC MOTOR-Z ERROR!! The error message is displayed when stepping out
The movement of the inlet shaft motor does of the motor is detected.
not complete. Turn off the system, and Contact our service.
remove the boat on the boat handling poker.
20 THE RETURNING OF TRAY FAILED!! A tray may be stopped on the way. Turn OFF the
The sample boat tray cannot be correctly set power, open the front cover and remove the tray. If
in the rack. Turn off the system, and remove the error repeats, adjust the sample changer.
the tray on the arm.
21 LIFT SENSOR ERROR Contact our service.
An error has occurred in the lift sensor.
22 HEATER(H1) TEMP ABNORMAL] A failure of the fan or infrared temperature sensor.
The temperature of H1 does not decrease to Contact our service.
pre-heat temperature.
23 HEATER(H2A) The error is issued when the temperature is above
H2A Overheating Detected. 660 deg C
Contact our service.
24 HEATER(H2B) The error is issued when the temperature is above
H2B Overheating Detected. 390 deg C
Contact our service.
25 HEATER(H3) The error is issued when the temperature is above
H3 Overheating Detected. 800 deg C
Contact our service.
47
NIC-600-2231-11 6. Maintenance
48
NIC-600-2231-11 6. Maintenance
When any of the following messages is displayed, the measurement sequence stops
temporarily and, if the problem is corrected, the sequence restarts.
Message Description and corrective action
Close the door of tray rack. If the door of tray rack is opened during the transfer
of a tray, the tray cannot be transferred correctly.
Close the door during the transfer of a tray.
There is a tray in the return position. A tray is detected on the rack to which a tray is to be
Remove the tray. returned. Remove a tray if any.
The sensor on the tray rack may malfunction due to
the influence of an intensive light. Ensure that the
rack is not directly exposed to outer light.
There is a tray on the tray evacuation rack. A tray is detected on the tray evacuation rack when a
Remove the tray. tray is returned there. Remove a tray if any.
The sensor on the tray rack may malfunction due to
the influence of an intensive light. Ensure that the
rack is not directly exposed to outer light.
The heater H-2 has a thermostat to protect it from overheating. A thermostat being
activated will result in the heater not working. In this case, press the central button on the
thermostat to reset it.
49
NIC-600-2231-11 6. Maintenance
6.3. Troubleshooting
When the sensitivity does not recover by repeated measurements with a standard solution,
check the flow chart as below.
START
NO
Check for leak Find any leak and repair.
YES
NO
Does the method work Check the method.
without any problem?
YES
YES
YES
NO
Have collection tube and joints been Replace the sample heating
cleaned or replaced? tube. Clean the collection
tube/joint
YES
NO
Has the heating tube been replaced Replace the sample
with a new one? heating tube.
YES
YES NO
Has the peak
sensitivity been
restored?
When the blank value does not decrease from an extremely large value (for example
several nanograms) even if the sample boats are washed or the blank measurements are
repeated, check the flow chart as below.
* Use a sample boat suitable for high/low concentration depending on the purpose.
START NO
YES
NO
Is the boat new? Repeat measurement with
a new boat.
YES
NO
Is the blank value high even if measured
with “TEST” method?
Clean the sample inlet.
YES
YES
YES
NO
Does the peak value decrease to 0.001 or
less when the blank measurement is
repeated?
YES
Contact our
service.
END
51
NIC-600-2231-11 6. Maintenance
6.4. Maintenance
If the check fails, turn OFF the power and check for loosened joints, insufficient
tightening of the sample inlet screw, or damage on flow path.
Check the tubes only after the power switch has been turned
Warning OFF and the heater has cooled down to room temperature or it
could result in burns.
52
NIC-600-2231-11 6. Maintenance
Carry out maintenance of the heated sections only after the power
Warning switch has been turned OFF and the heater has cooled down to
room temperature or it could result in burns.
53
NIC-600-2231-11 6. Maintenance
54
NIC-600-2231-11 6. Maintenance
55
NIC-600-2231-11 6. Maintenance
No space
56
NIC-600-2231-11 6. Maintenance
57
NIC-600-2231-11 6. Maintenance
58
NIC-600-2231-11 6. Maintenance
Turn OFF the power and open the top cover and
then the front cover.
59
NIC-600-2231-11 6. Maintenance
Carry out maintenance on the optical box only after having turned
the power switch OFF and allowing it to cool down to normal room
Caution
temperature or it could result in burns as the box keeps the
constant temperature of approx. 150°C during operation.
60
NIC-600-2231-11 6. Maintenance
61
NIC-600-2231-11 6. Maintenance
62
NIC-600-2231-11 6. Maintenance
If the transfer error occurs in the sample changer, adjust the position.
Open Maintenance screen, select “BC” tab and click BC Adjust.
63
NIC-600-2231-11 6. Maintenance
Tray No.: Enter the number of the tray onto which the
boat is loaded.
Boat No.: Enter the boat number.
►: Advances by 1 movement
◄: Returns by 1 movement
►►: Advances to the next step.
◄◄: Returns to the previous step.
① Axis-Y Down Move the arm to the designated
tray.
64
NIC-600-2231-11 6. Maintenance
65
NIC-600-2231-11 6. Maintenance
66
NIC-600-2231-11 6. Maintenance
If any unusual noise is heard when the sample changer is running, apply grease to where
the noise is generated.
Apply grease to the horizontal shaft, vertical shaft, sample inlet shaft, slide guides of the
lift, and trapezoid screw threads.
* Use only the grease specified for the system. Any other type of grease may not adhere
very well and thus cause failure of the equipment.
67
NIC-600-2231-11 6. Maintenance
Re-adjust the optical system if A/D FULL or LAMP BROKEN error occurs.
Open “Run"-“Maintenance” and click Optical Adjust button.
Optical Adjust screen opens, where the gain SIG1, SIG2, and REF can be adjusted.
(Adjust to ZERO=0V, SPAN=4V)
Maintenance
Exit
Click AUTO, and the span and zero of each detector are automatically adjusted.
The power of the lamp is turned OFF when the zero adjustment is performed. Wait at least
20 minutes before carrying out the measurement of a sample after the zero adjustment.
68
NIC-600-2231-11 6. Maintenance
Carry out the zero adjustment of the mass flow controller if the flow rate is not zero even
when the carrier gas is not flowing.
Maintenance
Open “Run”-“Maintenance”, and click the button.
69
NIC-600-2231-11 6. Maintenance
(1) Remove the Teflon tube connected to the Pharmed tube of the air pump.
(2) Grip the cover (blue) of the air pump and turn it about 30 degrees
counterclockwise. The hook fixing the cover will be released and the cover will
come off. Pull the cover out toward you.
(3) Turn over the cover, remove the internal rotor, and then the Pharmed tube.
70
NIC-600-2231-11 6. Maintenance
(4) Push a new Pharmed tube into the cover while turning the rotor.
* Engage the INSULOKs at both ends of the tube using the hooks inside the cover.
Ensure to engage the INSULOKs with
the hooks inside the cover when
inserting the tube.
(5) Put the hooks of the cover in the hole of the air pump, and install the cover by
turning it slightly clockwise.
(6) Install the Teflon tube to the right-side Pharmed tube of the air pump.
71
NIC-600-2231-11 7. MA3Win
7. MA3Win
This software has two functions; controlling the MA-3000 and processing the
measurement data.
All the operations of the MA-3000 can be controlled using the software.
Control window
Displays measurement sequence in animation.
72
NIC-600-2231-11 7. MA3Win
Profile window
The profiles of the measurement results can be viewed.
(1) The profile can be magnified or shrunk with respect to the vertical axis.
(2) Different data can be viewed by shifting the screen to the left or right.
(3) The number of data items to be displayed can be specified. The profile can
also be magnified or shrunk with respect to the time axis.
*The scale of the profiles set on this screen applies to the scale used in printing.
Calibration curve window
The window displays the current calibration curve.
The “CALIB” tab displays the constructed calibration curve.
The “Result” tab displays the verification results of the calibration curve.
* The verification result of the calibration curve is used to evaluate the accuracy of
the constructed cubic calibration curve. The standard value is plotted on the
horizontal axis while the amount of mercury obtained by the constructed cubic
calibration curve is plotted on the vertical axis, and the correlation coefficient
between them is taken as an index of the accuracy.
Table window
The measurement results can be viewed in this window.
73
NIC-600-2231-11 7. MA3Win
Title bar: The title of the software and the file being currently edited can be viewed.:
Menu bar: Functions shown are displayed. (Refer to 7.2 List of menu commands)
Tool bar: The functions you often use are listed in the form of icons.
11
12
13
14
15
① "Display Change" Refer to 7.2.5.1 Display change (CTL), 7.2.5.2 Display change
(ANA)”
② “Start" Refer to “7.2.3.1Measurement start”
③ “Stop" Refer to “7.2.3.2Stop”
④ "Purge" Refer to “7.2.3.4 Purge”
⑤ "Emergency" Refer to “7.2.3.3 Emergency”
⑥ "Method Set" Refer to 7.2.3.8 Measurement condition”
⑦ "New" Refer to 7.2.1.1 New project (N)”
⑧ "Open" Refer to 7.2.1.2 Open (D)”
⑨ "Save as." Refer to 7.2.1.3 Save as (A)”
⑩ "Print Preview" Refer to 7.2.1.5 Print Preview (V)”
⑪ "Print" Refer to 7.2.1.6 Print (P)”
⑫ "Statistics" Refer to 7.2.4.3 Statistics (S)”
⑬ "Table Condition" Refer to 7.2.4.2 Table condition”
⑭ "PEAK/INTE" Refer to 7.2.4.2 Table condition”
⑮ "Drift Correction" Refer to 7.2.4.2 Table condition”
74
NIC-600-2231-11 7. MA3Win
Samples are registered and the results of measurement can be viewed in the Table
window.
The Table window includes a standard sample table and an unknown sample table.
Calibration curves are created in the standard sample table, and unknown samples are
analyzed in the unknown sample table. 50 standard samples and 200 unknown samples
can be registered in respective tables.
The two tables can be switched by clicking the tabs at the bottom of the window.
* The items to be displayed change according to the settings made in table conditions.
* The width of column can be arbitrarily changed. (When the width of a column has
been set to zero, select the columns on both sides of it and widen it so that it can be
re-displayed.)
Measurement
Check the samples to be measured.
Click the box to enter a checkmark and click again to clear it.
Only samples with checkmarks will be measured
Sample Number
Enter the tray number of the sample changer.
The sample changer allows random access and hence any number you want can be
entered.
If “0” is entered, the system will carry out a purge measurement without inserting a
boat.
Date
The time when the measurement is completed is registered after the measurement.
Remarks
Any comment can be entered.
75
NIC-600-2231-11 7. MA3Win
Sample name
Select STD when using the measurement data to construct a calibration
curve, or CV when calculating a quantitative lower limit or checking system
operation
Standard value
Enter the standard sample (mercury) amount.
“STD INPUT” window appears when the “Standard”
cell is clicked. Enter the standard concentration
(temperature in the case of the standard gas), the
injection volume, and then the calculated value of the
amount of mercury.
The unit of the standard concentration can be
changed by clicking the unit button. When you want to input the standard value
(amount of mercury) directly, press ESC and the “STD INPUT” window disappears.
LOW
Indicates the measurement result of LOW range.
HIGH
Indicates the measurement result of HIGH range.
PEAK or AREA
The cell indicates a measured PEAK value (peak height) or AREA value (area).
[ON/OFF] indicates whether or not drift correction has been activated.
Dev[%], Deviation
The deviation between the measured value and standard value is displayed as a
percentage.
76
NIC-600-2231-11 7. MA3Win
Method
The heating conditions can be individually set for each
sample with the measurement of unknown samples. Select
an option from the pull-down menu* Detailed settings of
the method can be made from “Run“ –“Method setting”
Name
Enter any sample name.
Click the arrow on the right of a cell to open the pull-down
menu on which special names and names added by
executing “Table” – “Registration of names” are displayed.
Special names
CHECK STD
Used to check the condition (sensitivity) of the unit when measuring unknown
sample.
When a measurement is performed using the sample name of “CHECK STD,” it can
be discontinued if the result is out of the range. (Refer to 7.2.3.8Measurement
condition.)
For measurements under “CHECK STD,” use a standard solution in a sample boat.
R Blank
If an additive reagent is used in the measurement of unknown samples, the amount
of mercury contained in it may affect the measurement results. To compensate for
that, the mercury amount contained in the reagent is measured as R.Blank and used
to make adjustments for it.
Select “R.Blank” in the Name column, and perform the measurement of the reagent
in the same way as an unknown sample. The value of R.Blank is then automatically
subtracted from the measured value (mercury amount and concentration) of an
unknown sample. (For the details, refer to 7.6 R.Blank Calculation method.)
77
NIC-600-2231-11 7. MA3Win
SVOL
Enter the sample volume or weight into this cell.
ADDSTD
Used to determine the addition recovery of an unknown sample. Enter the amount of
mercury added to the sample.
MOIST
Enter the moisture content in the cell to compensate for moisture. If the cell was blank,
the calculation will be made with assumption that the value is zero. It is displayed
when “Moisture Content” is “ON.”
PEAK or AREA
Peak or Area and HIGH range and LOW range is outputted after measurement.
MEAS
The amount of mercury calculated by the calibration curve from PEAK or AREA value
is outputted.
CONC
Concentration is outputted.
Concentrat ion (Measured mercury value - R.BLANK )
Sample amount x (100 - Moisture content ratio) 100
Recovery
Recovery rate is outputted
Recovery Measured value - Sample amount x Average concentration of sample
Standard addition
Measurement results with the same sample name are used in the calculation.
The recovery rate is calculated from the concentration obtained from the results
whose standard addition cells are blank.
⑪ Range
Selects whether the LOW data or the HIGH data should be employed.
78
NIC-600-2231-11 7. MA3Win
Executed to clear all the data displayed and start new measurement.
Clear MEMO
If “Clear MEMO” is checked, the content in “MEMO” will also be cleared when the data
are cleared.
79
NIC-600-2231-11 7. MA3Win
80
NIC-600-2231-11 7. MA3Win
Page
Set whether to print page numbers.
Start page No.
Set the start page number.
81
NIC-600-2231-11 7. MA3Win
Memo Set : Print the items in “Memo” that were set in the measurement
conditions.
Calibration : Print the calibration curve as a graph.
Standard
Table : Print the table of standard samples.
Statistics : Print the statistical results of standard samples.
Profile : Print the profiles of standard samples.
The settings of the scale and the number of profiles printed in one
line are the same as those for the table of the standard sample.
Set the respective scales for HIGH and LOW ranges.
Sample
Table : Print the table of unknown samples.
Statistics : Print the statistical results of unknown samples.
Profile : Print the profiles of unknown samples.
The setting is the same as that for the table of unknown sample.
Method : Print the measurement condition.
Measure comp : Print the measurement completion time for each sample.
time
82
NIC-600-2231-11 7. MA3Win
Copies selected data to the clipboard. Any data in the selected cells are copied to the
clipboard when the table window is active. When the profile window is active, any image
data are copied to the clipboard
Cuts the data in selected cells. PEAK/AREA, HG and CONC cannot be cut.
* The values of SVOL, CVOL, and DVOL turn to “1” when cut
Pastes data copied to the clipboard to cells. Click cells in which the data are to be pasted,
and then execute the command.
* No data can be pasted in the PEAK / AREA, HG or CONC columns.
Removes a line(s). Click the line number to select the entire line and execute the
command.
Inserts a blank line. Click the line number to select the entire line and execute the
command. When multiple lines are selected, blank lines in the same number are inserted.
83
NIC-600-2231-11 7. MA3Win
Receives sample amount data from an electronic balance and writes it into the sample
amount cell.
* The function is effective only when the type of sample is set to “Solid” in the table.
* Balances available include the MS series, ML series and ME series of Mettler-Toledo.
The balance can be set by selecting System Menu - Balance Port Set.(Refer to
7.2.4.9 Balance communication settings (B))
84
NIC-600-2231-11 7. MA3Win
85
NIC-600-2231-11 7. MA3Win
*Backup file
Every time the file is automatically saved, a backup file with the same name and the
extension “MA3DEPOSIT” is created. (Example: TEST.MA3 TEST.
MA3DEPOSIT)
If a file has been erroneously edited and saved, it is possible to restore the data by
reading the backup file after changing its extension to “MA3.”
* However, if the measurement is continued with the same file name, the file is saved
based on the erroneously edited content.
The backup files continue to increase if not deleted intentionally. Periodically delete
those that are not necessary.
7.2.3.2. Stop
The measurement stops once the current measurement has been completed (the shortcut
key: F8).
7.2.3.3. Emergency
7.2.3.4. Purge
Any mercury content remaining in the system is purged before measurement starts. The
result of the purge measurement is not recorded in the Table.
7.2.3.5. Reset
The system terminates all the sequences and return to READY Status.
86
NIC-600-2231-11 7. MA3Win
Sets the heating conditions (Atomize1 through 4) of the sample heating furnace H1.
Several conditions have been preset as factory defaults, but it is also possible to change
them or create new conditions.
A heating sequence consists of 4 stages, i.e. the moisture removal stage and 3
temperature raising stages.
Atomize1: Remove any moisture and interfering substances with low boiling points.
Atomize2 through 4: Decompose mercury compound in the sample.
Heat temp:
Set the highest heating
temperature.
Heat time:
Set the heating time.
Flow:
Set the flow during heating.
Start duty:
Set the duty cycle when heating
starts.
Max duty:
Set the maximum duty cycle.
* The temperature control will only
work if the temperature exceeds the
set Heat temp.
Slope time:
Set the time interval from Start duty to
Max duty.
Method Number: Select the method number to display the details of the method settings.
Save : Save the settings of the method after having been edited.
* The settings are saved for each method instead of all at once.
* Ensure to save settings every time changes are made.
87
NIC-600-2231-11 7. MA3Win
7.2.3.7. Starting the measurement and turning OFF the heater after the
measurement
Start the measurement and turn OFF the heater after the measurement.
The heater can be automatically turned on again after it is turned OFF.
88
NIC-600-2231-11 7. MA3Win
7.2.3.8.1 General
STD Method
Sets the method used for the measurement of standard solution.
Flow rate
Sets the flow rate during the measurement. (Selected from 0.1L/min, 0.2L/min, or
0.5L/min.)
* There is a limitation in the change of the flow rate because the condition must be the
same to maintain the consistency of the date in the same table.
● Once the measurement is carried out, the flow rate cannot be changed. (It can
be changed when a new measurement is carried out.)
● It cannot be changed even in a new measurement if ABS remains in it.
STD check
Sets the condition of standard check.
STD [ng] Enter the amount of mercury of the standard sample to be measured.
Recovery Rate Set the deviation of recovery rate, which should be determined to be
[100%±] abnormal for measurements under CHECK STD.
Abnormal Select to “Discontinue” or “Continue” the subsequent measurement in
condition the case the recovery rate falls below the set value.
89
NIC-600-2231-11 7. MA3Win
90
NIC-600-2231-11 7. MA3Win
* When the blank value is measured after the measurement of a sample, about 1% of
the immediately previous detection value is detected as historic record.
* Using the default setting, Blank check may be frequently carried out and the
measurement may take a long time to complete when only samples of which
concentration is high are measured. In such a case, change the threshold level and
recovery level.
ABS(LOW) ABS(HIGH)*
0.1ng 0.004 10ng 0.001
1ng 0.02 100ng 0.01
5ng 0.1 1,000ng 0.1
10ng 0.2 5,000ng 0.4
20ng 0.4 10,000ng 0.6
50ng 1.0 20,000ng 0.9
91
NIC-600-2231-11 7. MA3Win
Check the tubes only after the power switch has been
Warning turned OFF and the heater has cooled down to room
temperature or it could result in burns.
92
NIC-600-2231-11 7. MA3Win
7.2.3.11. Maintenance
The movement and status of the hardware components get monitored on the screen.
Used for maintenance or when a system failure has occurred.
7.2.3.11.1 MA
A/D
H1[degC] H1 temperature
H2A [degC] H2A temperature
H2B [degC] H2B temperature
H3 [degC] H3 temperature
H4 (PIPE)[degC] Temperature of piping heater
H5 (OPT)[degC] Temperature of optical system heater.
H6 (RH1)[degC] Temperature of heating furnace (Optional)
H7 (RH2) Temperature of heating furnace RH2 (Optional)
H8 (RSV) Not used
FLOW1[L/min] Flow of mass flow controller
LAMP [V] Lamp voltage
CAP [V] Lamp heater voltage
93
NIC-600-2231-11 7. MA3Win
I/O
V1 Turns ON/OFF the solenoid valve 1.
V2 Turns ON/OFF the solenoid valve 2.
V3 Turns ON/OFF the solenoid valve 3.
V4 Turns ON/OFF the three directional solenoid valve 4.
V5 Turns ON/OFF the three directional solenoid valve 5.
FAN1 Turns ON/OFF the cooling fan of sample heating furnace H1.
FAN3 Turns ON/OFF the cooling fan of mercury collecting furnace H3.
BUZZER Turns OFF/ON the buzzer.
LED Turns ON/OFF the LEDs on the board.
P2 Turns ON/OFF pump P2 for reducing vaporization (Optional)
FAN6 (RH1) Turns ON/OFF cooling fan 1 for RH-MA (Optional)
FAN7 (RH2) Turns ON/OFF cooling fan 2 for RH-MA (Optional)
FAN8 (YOBI) Not used
FLOW [L/min] Flow test of mass flow controller
HEATER
For the test of H1 (Enter temperature). Limited up to 10% of
H1 the capability for protection purpose.
For the test of H2A (Enter temperature). *Do not set any
H2A value higher than the default.
For the test of H2B (Enter temperature). *Do not set any
H2B value higher than the default.
For the test of H3 (Enter temperature). *Do not set any value
H3 higher than the default.
For the test of H4 (Enter temperature). *Do not set any value
H4 (PIPE) higher than the default.
For the test of H5 (Enter temperature). *Do not set any value
H5 (OPT) higher than the default.
H6 (RH1) Controls the H1 temperature of the collection tube heating
furnace PH-MA. (Optional)
H7 (RH2) Controls the H2 temperature of the collection tube heating
furnace PH-MA. (Optional)
H8 (YOBI) Not used
The heaters are tested using the temperature entered in the left column and the duty
entered in the right column.
94
NIC-600-2231-11 7. MA3Win
Sig/Ref
Sig1 [V] Output voltage of the signal from “LOW” side detector
Ref [V] Output voltage of the signal from the Ref. lamp detector
Sig2 [V] Output voltage of the signal from “HIGH” side detector
Lamp OFF Optical lamp OFF
Lamp ON Optical lamp ON
Optical Adjust Adjustment of voltage of Sig1, Ref, Sig2
Exit
95
NIC-600-2231-11 7. MA3Win
7.2.3.11.2 BC
SENSOR
LIFT-UP Status of the upper limit sensor of the lift
LIFT-DOWN Status of the lower limit sensor of the lift
TRAY Status of the tray detection sensor on the arm
BOAT Status of the boat detection sensor
DOOR OPEN Status of the door open sensor
REFUGE TRAY BOX Status of the refuge tray box sensor
AXIS=Z IN0 Z axis sensor
SPARE1 Not used
TRAY SENSOR
TRAY01 to 10 Status of the tray detection sensors in the rack
96
NIC-600-2231-11 7. MA3Win
BC ADJUST
For any sample boat, the action of the sample changer can be tested.
Run an action
Return before an action
Run a step
Return before a step
7.2.3.11.3 ETC
97
NIC-600-2231-11 7. MA3Win
7.2.3.11.4 SC5
Used for testing the SC-5 sample changer for reducing vaporization
Item Function
RD Turns ON/OFF each pump.
Get the SC-5 information (VERSION, SERIAL). Click
INFORMATION “REQUEST”.
Click “READ”. Turns ON/OFF H2SO4 pump. Get the location
OFFSET information (TABLE, ARM, LIFT).
98
NIC-600-2231-11 7. MA3Win
7.2.3.11.5 GLP
Item Function
LAMP ACTIVE TIME Indicates the accumulated active period of the lamp.
ABS Hg Indicates the accumulated amount of mercury
measured.
MEAS CNT Indicates the accumulated number of measurement.
99
NIC-600-2231-11 7. MA3Win
7.2.3.11.6 Parameter
100
NIC-600-2231-11 7. MA3Win
-Print example-
101
NIC-600-2231-11 7. MA3Win
The calibration curve, mercury amount, and concentration are calculated based on the
conditions set in this window. The conditions can be changed even after a measurement
has taken place.
PEAK/INTE
Selects Analysis mode from PEAK (peak analysis) or AREA (area analysis).
Sample
Selects the type of Sample from “Liquid” or “Solid”
Calibration Curve
Selects the type of calibration curve for each of the Low and High measurement ranges.
Three options are available for use as the type of calibration curve: the normal curve
calculated using the least square method (+b), the curve calculated using the least
square method and passing through the blank (+Blank), and the curve calculated using
the least square method and passing through zero (+0).
Selecting the normal curve (+b) with normal measurements and the curve passing
through the blank (+Blank) with measurements of trace quantities of mercury close to
the blank value enable the best measurement results.
102
NIC-600-2231-11 7. MA3Win
Unit
Sets the units of standard mercury amount, sample amount, mercury amount, and
concentration.
The selectable unit varies depending on the type of sample.
Others
Decimal Places
Sets the decimal places of the concentration to be displayed.
Drift Correction
Selects whether or not to apply drift correction to the base line
In normal use, “Yes” should be selected. Cancel the drift correction for analysis only
when checking the condition of the system.
Minus Display
When a calculated concentration/amount of mercury is a negative, selects to display
whether the value with minus sign or zero instead.
Moisture Content
Selects whether to display moisture correction or not.
* Can only be selected when the type of sample is set to “Solid”.
Statistic
Selects Mercury amount or Concentration as the object of the statistical calculations.
103
NIC-600-2231-11 7. MA3Win
104
NIC-600-2231-11 7. MA3Win
Sets the RS232C port used for communication with the MA-300
105
NIC-600-2231-11 7. MA3Win
Language
Selects Japanese or English as the display language.
Measurement method
Selects the measurement method from the options listed below.
If REDUCTION-AMALGAM or REDUCTION-DIRECT is selected, also select an
optional unit.
HEATING VAPORIZE
RH
REDUCTION-AMALGAM
REDUCTION-DIRECT
Optional Unit
* Any new settings with “Language” and “Measurement method” only take effect after
the software has been restarted. The software automatically restarts after any
changes to the settings.
106
NIC-600-2231-11 7. MA3Win
Limit (%)
The equation of the calibration curve changes to red and WARNING is displayed when
the deviation between the back-calculated value and the set value is out of this range.
107
NIC-600-2231-11 7. MA3Win
* The baud-rate must match the baud rate of the balance to be connected.
108
NIC-600-2231-11 7. MA3Win
Upgrade the version of the ROM built into the MA-3000. (Refer to 4.3.7
109
NIC-600-2231-11 7. MA3Win
Displays the contents of sequence log files and error history files.
110
NIC-600-2231-11 7. MA3Win
111
NIC-600-2231-11 7. MA3Win
7.2.6. Help
7.2.6.1. About…
112
NIC-600-2231-11 7. MA3Win
《y=ax+b》
a = ( ∑yx * n - ∑y * ∑x ) / ( ∑x^2 * n - ∑x^2 )
b = ( ∑y - a * ∑x ) / n
《y=ax+Blank》
a = ( ∑yx - (Blank * ∑x) ) / ∑x^2
b = Blank
《y=ax+0》
a = ∑yx / ∑x^2
b=0
《y=ax^3+bx^2+cx+d》
tx3yp= ∑x^3y u1= (tx3yp*n - ty1p*tx3p) s1= (u1*u9 - u3*u6)
tx2yp= ∑x^2y u2= (tx2yp*n - ty1p*tx2p) s2= (u2*u9 - u3*u8)
tx1yp= ∑xy u3= (tx1yp*n - ty1p*tx1p) s3= (u4*u9 - u6*u6)
ty1p = ∑y u4= (tx6p *n - tx3p*tx3p) s4= (u5*u9 - u8*u6)
tx6p = a∑x^6 u5= (tx5p *n - tx3p*tx2p) s5= (u7*u9 - u8*u8)
tx5p = a∑x^5 u6= (tx4p *n - tx3p*tx1p)
tx4p = a∑x^4 u7= (tx4p *n - tx2p*tx2p)
tx3p = a∑x^3 u8= (tx3p *n - tx2p*tx1p)
tx2p = b∑x^2 u9= (tx2p *n - tx1p*tx1p)
tx1p = c∑x
a= ( s1*s5 - s2*s4 ) / ( s3*s5 - s4*s4 )
b= ( s1 - a * s3 ) / s4
c= ( u1 - a * u4 - b * u5 ) / u6
d= ( tx3yp - a * tx6p - b * tx5p - c * tx4p ) / tx3p
113
NIC-600-2231-11 7. MA3Win
《y=ax^3+bx^2+cx+Blank》
tx3yp= ∑x^3y s1= (tx3yp * tx2p) - (tx1yp * tx4p)
tx2yp= ∑x^2y s2= (tx2yp * tx2p) - (tx1yp * tx3p)
tx1yp= ∑xy s3= (tx6p * tx2p) - (tx4p * tx4p)
tx6p = a∑x^6 s4= (tx5p * tx2p) - (tx4p * tx3p)
tx5p = a∑x^5 s5= (tx3p * tx2p) - (tx1p * tx4p)
tx4p = a∑x^4 s6= (tx4p * tx2p) - (tx3p * tx3p)
tx3p = a∑x^3 s7= (tx2p * tx2p) - (tx1p * tx3p)
tx2p = b∑x^2
tx1p = c∑x
a= ( (s1*s6- s2*s4) - Blank(s5*s6 - s7*s4) ) / ( s3*s6 - s4*s4 )
b= ( s1 - Blank*s5 - a*s3 ) / s4
c= ( tx3yp - (a * tx6p) - (b * tx5p) - (Blank * tx3p) ) / tx4p
d= Blank
《y=ax^3+bx^2+cx+0》
tx3y1p= ∑x^3y s1= (tx3yp * tx2p) - (tx1yp * tx4p)
tx2y1p= ∑x^2y s2= (tx2yp * tx2p) - (tx1yp * tx3p)
tx1y1p= ∑xy s3= (tx6p * tx2p) - (tx4p * tx4p)
tx6p = ∑x^6 s4= (tx5p * tx2p) - (tx3p * tx4p)
tx5p = ∑x^5 s5= (tx4p * tx2p) - (tx3p * tx3p)
Tx4p = ∑x^4
Tx3p = ∑x^3
Tx2p = ∑x^2
a= ( s1*s5 - s2*s4 ) / ( s3*s5 - s4*s4 )
b= ( s1 - a*s3 ) / s4
c= ( tx3yp - (a * tx6p) - (b * tx5p) ) / tx4p
d= 0
114
NIC-600-2231-11 7. MA3Win
Mercury amount
Thermal vaporization
Mercury amount = Measured value
* If R.BLANK is specified,
Mercury amount = Measured value – Measured value of R.BLANK
Reducing vaporization
If the unit of sample amount is “none”,
Mercury = None
If the unit of sample amount is not “none”,
Mercury amount = Measured value x (Constant volume / Dispensed
volume)
Concentration
Thermal vaporization
Concentration = Mercury amount / Sample amount
Reducing vaporization
If the sample amount is “none”,
Concentration = Measured value
If the sample amount is not “none”,
Concentration = Mercury amount / Sample amount
If the moisture correction is valid (i.e. the type of Sample is set to “Solid” and
Moisture Content is set to “Yes”),
Concentration = Concentration x 100 / (100 – Moisture content [%])
115
NIC-600-2231-11 7. MA3Win
Example
Name SVOL ADDSTD PEAK MEAS CONC RECOVE
[mg] [ng] [ng] [ppm] RY[%]
Sample1 40 0.004005 48.7061 1.218 -
Sample1 30 0.003719 45.2043 1.507 -
Sample1 35 0.003927 47.7510 1.364 -
Sample1 30 40.000 0.006713 81.2528 2.708 100.9
* In measuring a sample with large variation, the variation of the recovery of the added
standard solution/substance becomes larger as the added amount is smaller. Take
the variation of the sample into consideration when determining the amount of
standard solution/substance to be added.
116
NIC-600-2231-11 7. MA3Win
When “R.Blank" is selected for a name of a sample, the R.Blank value is automatically
subtracted from the measurement results of an unknown sample (the amount and
concentration of mercury).
It is possible to set two or more R.Blank values by adding a letter after the character string
of R.Blank when changing the type of the additive or reset the R.Blank value at a certain
interval.
The R.Blank values with the same name obtained from a sample are averaged, which is
used for the compensation for the measurement value of the sample.
Sample Measured
Concentration
No. Sample Name Volume Value
[μg/L]
[μL] [ng]
1 Sample1 100 0.050 0.500
2 Sample2 100 0.051 0.510 No compensation
3 Sample3 100 0.052 0.520
4 RBlank1 0.001
5 RBlank1 0.000
6 Sample4 100 0.005 0.043 Concentration calculated
by subtracting the
7 Sample5 100 0.051 0.503 average of RBlank1 from
the measured value.
8 Sample6 100 0.052 0.513
9 Sample7 100 0.050 0.493
10 RBlank2 0.006
11 RBlank2 0.005
12 Sample8 100 0.050 0.445 Concentration calculated
by subtracting the
13 Sample9 100 0.051 0.455 average of RBlank2 from
the measured value.
14 RBlank3 0.010
15 Sample10 100 0.050 0.400 Concentration calculated
by subtracting RBlank3
16 Sample11 100 0.012 0.020 from the measured value.
17 RBlank1 0.001
18 Sample12 100 0.051 0.503
19 Sample13 100 0.052 0.513 Concentration calculated
by subtracting the
20 Sample14 100 0.053 0.523 average of RBlank1 from
the measured value.
21 Sample15 100 0.054 0.533
117
NIC-600-2231-11 8. Contact
8. Contact
Technical Center
14-8 Akaoji-cho, Takatsuki-shi, Osaka 569-1146, Japan
Phone: +81-72-693-6820
Fax: +81-72-694-0663
118