Trace Quantities of Sulfur in Light Liquid Petroleum Hydrocarbons by Oxidative Microcoulometry
Trace Quantities of Sulfur in Light Liquid Petroleum Hydrocarbons by Oxidative Microcoulometry
Trace Quantities of Sulfur in Light Liquid Petroleum Hydrocarbons by Oxidative Microcoulometry
Designation: D 3120 – 06
1
D 3120 – 06
NOTE 1—Take care not to use excessive stirring and possibly damage
the electrodes with the stir bar. The creation of a slight vortex is adequate.
2
D 3120 – 06
FIG. 2 Flow Diagram for Typical Coulometric Apparatus for Trace Sulfur Determination
injecting the sample from a microlitre syringe at a constant rate 7.7 Oxygen—High purity grade (HP),5 used as the reactant
between 0.5 to 1.0 mL/s is required (see Note 2). gas. (Warning—Oxygen vigorously accelerates combustion.)
NOTE 2—Take care not to introduce the sample too fast into the
7.8 Gas Regulators—Two-stage gas regulators shall be used
oxidation zone of the furnace and overload the combustion capacity of the for the oxygen and inert carrier gas.
pyrolysis tube. Program the sample inlet system to deliver the sample at 7.9 Cell Electrolyte Solution—Dissolve 0.5 g of potassium
a sufficiently controlled and repeatable rate to prevent any incomplete iodide (KI) and 0.6 g of sodium azide (NaN3) in approximately
combustion by-products (coke or soot) from forming at the exit of the 500 mL of high-purity water, add 6 mL of glacial acetic acid
pyrolysis tube. (CH3COOH), and dilute to 1000 mL or follow the manufac-
6.12 Balance—With a weighing precision of 60.01 mg. turer’s specifications.
NOTE 3—Take care to store bulk quantities of the electrolyte in a dark
7. Reagents and Materials
place. It is recommended to prepare fresh electrolyte at least every three
7.1 Purity of Reagents—Reagent grade chemicals shall be months.
used in all tests. Unless otherwise indicated, it is intended that
7.10 Sodium Azide (NaN3), fine granular. (Warning—
all reagents shall conform to the specifications of the Commit-
Toxic. Causes eye and skin irritation. Explosive.)
tee on Analytical Reagents of the American Chemical Society,
where such specifications are available.4 Other grades may be 7.11 Potassium Iodide (KI), fine granular.
used, provided it is first ascertained that the reagent is of 7.12 Potassium Chloride (KCl), fine granular. Used for the
sufficiently high purity to permit its use without lessening the 1M Ag/AgCl double junction reference electrode.
accuracy of the determination. 7.13 Potassium Nitrate (KNO3), fine granular. Used for the
7.2 Purity of Water—Unless otherwise indicated, references 1M Ag/AgCl double junction reference electrode.
to water shall be understood to mean reagent water conforming 7.14 Iodine (I), 20 mesh or less, for saturated reference
to Specification D 1193, Type II and III. electrode.
7.3 Quartz Wool—Grade fine. 7.15 Toluene, Xylenes, Isooctane—Reagent grade. (Other
7.4 Acetic Acid (CH3COOH)—Glacial acetic acid with solvents similar to those occurring in the samples being
specific gravity = 1.05. (Warning—Poison. Corrosive. Com- analyzed are acceptable.) A solvent blank correction is required
bustible. May be fatal if swallowed. Causes severe burns. due to the inherent sulfur present in the solvents used for
Harmful if inhaled.) standard preparation and sample dilution. (Warning—
7.5 Phosphoric Acid (85 % w/w)—Orthophosphoric acid Flammable solvents. Harmful if inhaled. Vapors may cause
(H3PO4). (Warning—Poison. Corrosive. May be fatal if swal- flash fire.)
lowed. Causes severe burns.) NOTE 4—The use of solvents with non-detectable levels of sulfur
7.6 Inert Gas—Argon or helium, high purity grade (HP),5 relative to the sulfur content in the sample can make the solvent blank
used as carrier gas. (Warning—Compressed gas under high correction unnecessary.
pressure. Gas reduces oxygen available for breathing.)
7.16 Dibenzothiophene—FW 184.26, 17.399 % (mass/
mass) S (see Note 4).
7.17 n-Butyl Sulfide—FW 146.29, 21.92 % (mass/mass) S
4
Reagent Chemicals, American Chemical Society Specifications, American (see Note 4).
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
listed by the American Chemical Society, see Analar Standards for Laboratory 7.18 Thionaphthene (Benzothiophene)—FW 134.20,
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia 23.90% (mass/mass) S (see Note 4).
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,
MD. NOTE 5—A correction for chemical impurity can be applied if deemed
5
High-purity grade gas has a minimum purity of 99.995 %. necessary.
3
D 3120 – 06
TABLE 1 Recommended Preparation of Working Standards
Working Standard Concentration, Stock Standard Concentration, Volume Stock Standard, Total Volume Working Standard,
µg-S/mL µg-S/mL mL mL
500 1000 25 50
50 1000 5 100
10 500A 2 100
A
Working standard.
7.19 Sulfur, Standard Solution (approximately 1000 mg-S/ 8.2 Consult current OSHA regulations and supplier’s Mate-
mL)—Prepare a stock solution by accurately weighing approxi- rial Safety Data Sheets (MSDS) for materials used in this test
mately 0.5748 g of dibenzothiophene or 0.4652 g of n-butyl method.
sulfide or 0.4184 g of thionaphathene into a tared 100 mL, type
A volumetric flask. Dilute to volume with a selected solvent. 9. Sampling
This stock can then be further diluted to prepare sulfur working 9.1 Obtain a test sample in accordance with Practice D 4057
and calibration standards as outlined in Tables 1-3 (see Notes or Practice D 4177. To preserve volatile components which are
6-8). in some samples, do not uncover samples any longer than
necessary. Samples should be analyzed as soon as possible
A 3 B 3 C 3 106
µg2S/mL 5 (3) after collection from bulk supplies to prevent loss of sulfur, or
100 mL
contamination due to exposure or contact with sample con-
where: tainer.
A = grams of standard. 9.1.1 (Warning—Samples collected at temperatures below
B = weight of fraction sulfur (S) in standard. room temperature can undergo expansion and rupture the
C = weight of fraction standard purity. container. For such samples, do not fill the container to the top;
7.20 Sulfur, Standard Working Solution—Prepare the work- leave sufficient air space above the sample to allow room for
ing standards using standard type A volumetric pipet as expansion.)
outlined in Table 1. 9.2 If the test sample is not used immediately, thoroughly
NOTE 6—Working and calibration standards prepared on a regular basis
mix the container prior to taking a test specimen.
recommended, depending upon frequency of use and age. Stock solutions
10. Preparation of Apparatus
typically have a useful life of about 3 months.
NOTE 7—Calibration standards can be prepared and diluted on a 10.1 Carefully insert the quartz pyrolysis tube into the
mass/mass basis, when calculation results are adjusted to accommodate furnace, and connect the oxygen and inert carrier gas lines.
them. 10.2 Connect the boat drive system or syringe drive system
NOTE 8—Stock, working and calibration standards from commercial to the pyrolysis tube.
sources can be used if checked for accuracy and can meet the performance
criteria of this test method.
10.3 Perform a leak check according to the manufacturer’s
instructions.
7.21 Quality Assurance (QA) Samples—Samples that are 10.4 Assemble the titration cell, and add all required solu-
liquid petroleum materials or standards products of known tions, including cell electrolyte, according to the manufactur-
sulfur content that were not used in the generation of the er’s instructions.
apparatus’ calibration curve. These (QA) samples are to be 10.5 Connect the titration cell to the apparatus according to
used to check the validity of the testing process as described in the manufacturer’s instructions.
Section 11. An ample supply of QA sample material should be 10.6 Adjust the flow of gases and set the furnace tempera-
available for the intended period of use and shall also be tures and instrumental parameters according to the manufac-
homogenous and stable under the anticipated storage time and turer’s instructions.
conditions. 10.7 If using a boat inlet system, pre-bake the sample boats
used for the analysis per the manufacturer’s instructions.
8. Hazards
8.1 High temperature is employed in this test method. Extra 11. Calibration and Standardization
care shall be exercised when using flammable materials near 11.1 Following the manufacturer’s recommended proce-
the furnace. dures, set the operational parameters of the apparatus for
4
D 3120 – 06
TABLE 3 Recommended Preparation of Calibration Standards
Concentration Working Standard, Volume of Working Standard, Total Volume Calibration Standard, Concentration Calibration Standard,
µg-S/mL mL mL µg-S/mL
10 2 100 0.2
10 5 100 0.5
50 2 100 1.0
50 5 100 2.5
50 10 100 5.0
500 5 100 25
500 10 100 50
1000A 25 100 250
500 ... ... 500
1000A ... ... 1000A
A
Standard stock solution.
5
D 3120 – 06
pyrolysis tube. Increase the residence time for the boat in the 12. Quality Assurance (QA)
oxidation zone of the furnace if coke or soot is observed on the 12.1 Confirm the performance of the instrument of the test
boat after the analysis is finished. procedure by analyzing a quality control (QC) sample after
11.5.5 Syringe Inlet System—For direct injection, carefully each calibration and at least each day of use thereafter.
insert the syringe needle into the inlet of the combustion tube 12.2 When QC/QA protocols are already established in the
and attach the syringe to the driving mechanism. Allow time testing facility, these can be used when they confirm the
for the sample residues to be volatilized from the needle reliability of the test result.
(needle blank). Once the baseline has been reestablished, 12.3 When there is not QC/QA protocol established in the
immediately start the analysis. Remove the syringe once the testing facility, Appendix X1 can be used as the QC/QA
sample has been syringe once the sample has been completely system.
injected and the analysis is finished.
NOTE 15—To avoid overloading the combustion capacity of the system, 13. Procedure
injection rates between 0.5-1.0 µL/s are recommended. 13.1 Obtain a test sample using the procedure described in
11.5.5.1 Reduce the sample size, rate of injection, or both, Section 9. The sulfur concentration in the test sample shall be
for the direct injection the test sample into the furnace if coke less than the concentration of the highest standard and greater
or soot is observed on the exit end of the pyrolysis tube. than the concentration of the lowest standard used in the
11.6 Calibrate the instrument using one of the following two calibration. If required, a dilution should be performed on
techniques: either a weight or volume basis.
11.6.1 Automatic Internal Calibration—Measure the cali- 13.1.1 Gravimetric Dilution, (mass/mass)—Record the
bration standards and solvent blank using one of the procedures mass of the test sample and the total mass of the sample and
described in 11.3–11.5.5. Analyze the solvent blank and each solvent.
standard a minimum of three times. Calibrate the apparatus per 13.1.2 Volumetric Dilution, (mass/volume)—Record the
the manufacturer’s instructions. Use either a first or second mass of the test sample and the total volume of the test sample
order linear regression and use this calibration curve to and solvent.
determine the amount of sulfur in the unknown sample. The 13.2 Flush the sampling syringe several times with the
system’s performance shall be checked each day of use. (See unknown sample. Determine the sulfur concentration using one
Section 12.) of the procedures described in 11.3–11.5.5.
11.6.2 Manual Calibration Curve—Measure the solvent 13.3 Inspect the combustion tube and other flow path
blank and calibration standards using one of the procedures components to verify complete oxidation of the test sample.
described in 11.3–11.5.5. Analyze the solvent blank and each 13.3.1 Boat Inlet Systems—Reduce the boat drive introduc-
calibration standard a minimum of three times and record their tion rate, sample size, or both, if coke or soot is observed on the
absolute values in either µg or ng sulfur. Subtract the averaged exit end of the pyrolysis tube. Increase the residence time for
blank value from each calibration standard and construct a the boat in the oxidation zone of the furnace if coke or soot is
curve plotting the mass of sulfur measured for each calibration observed on the boat after the analysis is finished.
standard (y-axis) verses the theoretical mass of sulfur injected. 13.3.2 Syringe Inlet Systems—Reduce the sample size, rate
Perform a first or second order linear regression and use the of injection, or both, for the direct injection the test sample into
calibration curve to determine the amount of sulfur in the the furnace if coke or soot is observed on the exit end of the
unknown sample. The system’s performance shall be checked pyrolysis tube.
each day of use. (See Section 12.) 13.3.3 Cleaning and Recalibration—Clean any coked or
sooted parts per the manufacturer’s instructions. Assemble and
NOTE 16—Not all of the sulfur in the sample exits the oxidation zone of
the furnace as titratable SO2. In the strongly oxidative conditions of the leak check the apparatus per the manufacturer’s instructions
pyrolysis tube, some of the sulfur is also converted to SO3, which does not after any cleaning or adjustment. Repeat instrument calibration
react with the titrant. Accordingly, sulfur standards appropriate to the prior to reanalysis of test samples.
sample’s boiling range and sulfur type is recommended to guarantee 13.4 To obtain one result measure each test sample solution
adequate calibration. Recoveries less than 75% are to be considered three times and calculate the average value.
suspect. If recoveries below 75% are observed, it is recommended the 13.5 If the sample size is determined by volume, density
operator check the coulometric system, measuring parameters, and oper-
ating techniques are in proper order. If the apparatus is being operated
values needed for calculations are to be measured (using Test
properly, recoveries between 75 to 90% are to be expected. Method D 1298, Test Method D 4052, or equivalent) at the
temperature at which the sample was tested. If the sample size
11.7 If the fraction of sulfur converted to SO2(recovery
was determined by weight, density values are not required.
factors) drops below 75% of the standard solutions, prepare
fresh standards to check if the standards have been altered. 14. Calculation
Procedural details should be reviewed if a low recovery factors
persist. 14.1 Calculate the sulfur concentration in the test sample in
11.8 Cleaning and Recalibration—Clean any coked or mg/kg (ppm) as follows:
sooted parts per the manufacturer’s instructions. Assemble and C 3 1000
Sulfur, mg/kg ~ppm! 5 M 3 K or (5)
leak check the apparatus per the manufacturer’s instructions
after any cleaning or adjustment. Repeat instrument calibration C 3 1000
Sulfur, mg/kg ~ppm! 5 V 3 D (6)
prior to reanalysis of test samples.
6
D 3120 – 06
7
D 3120 – 06
sulfur gasoline and diesel sample types analyzed by eight 15.1.2.1 Case I—The difference between two single and
laboratories. A practical limit of quantitation (PLOQ) of '3 independent results obtained by different operators working in
µg/g S, was determined for the gasoline sample types and '5 different laboratories on identical test material would, in the
µg/g S for diesel sample types. long run, in the normal and correct operation of the test
15.1.1 Repeatability: method, exceed the following values in 1 case in 20, where x
15.1.1.1 Case I—The difference between two test results = the average of two test results.
obtained by the same operator with the same apparatus under
R 5 0.5793 X ~0.7901! (8)
constant operating conditions on identical test material would,
in the long run, in the normal and correct operation of the test 15.1.2.2 Case II—Gasoline samples between approximately
method, exceed the following values in only in 1 case in 20, 3 and 100 mg/g and diesel samples between approximately 5
where x = the average of the of two test results. and 40 µg/g were analyzed. The reproducibility values deter-
mined were as follows:
r 5 0.2802 X ~0.7901! (7)
Gasoline: 0.1470(X + 7.4502)1.3226
15.1.1.2 Case II—Gasoline samples between approximately Diesel: 0.5152(X + 0.65758)
where: X is the average of duplicate results in µg/g
3 and 100 µg/g and diesel samples between approximately 5
and 40 µg/g were analyzed. The repeatability values deter- 15.2 Bias—Three National Institute of Standards and Tech-
mined were as follows: nology (NIST) Standard Reference Materials (SRM) were
Gasoline: 0.03194(X + 7.4502)1.3226 analyzed to determine the bias. These samples were gasoline
Diesel: 0.08520 (X + 0.65758) SRMs 2298 (4.6 µg/g S) and 2299 (13.6 µg/g S), and diesel
where: X is the average of duplicate results in µg/g
SRM 2723a (10 µg/g S). There was no significant bias for
15.1.2 Reproducibility: SRMs 2299 and 2723a. However, there was a bias of -1.25 for
SRM 2298. 8
8
16. Keywords
Supporting data have been filed at ASTM International Headquarters and may
be obtained by requesting Research Report RR:D02–1547. 16.1 light hydrocarbons; microcoulometry; sulfur
APPENDIXES
(Nonmandatory Information)
X1.1 Confirm the performance of the instrument or the test test method, the frequency of QC testing is dependent of the
procedure by analyzing a quality control sample. criticality of the quality being measured, the demonstrated
stability of the testing process, and customer requirements.
X1.2 Prior to monitoring the measurement process, the user Generally, a QC sample is analyzed each testing day with
of the test method needs to determine the average value and
routine samples. The QC frequency should be increased if a
control limits of the QC sample (see Practice D 6299 and MNL
large number of samples are routinely analyzed. However,
79).
when it is demonstrated that the testing is under statistical
X1.3 Record the QC results and analyze by control charts control, the QC testing frequency may be reduced. The QC
or other statistically equivalent techniques to ascertain the sample precision should be checked against the ASTM test
statistical control status of the total testing process (see Practice method precision to ensure data quality.
D 6299 and MNL 79). Any out -of-control data should trigger
investigation for root cause(s). The results of this investigation X1.5 It is recommended that, if possible, the type of QC
may, but not necessarily, result in instrument e-calibration. sample that is regularly tested be representative of the material
routinely analyzed. An ample supply of QC sample material
X1.4 In the absence of explicit requirements given in the
should be available for the intended period of use, and shall be
homogeneous and stable under the anticipated storage condi-
9
ASTM MNL 7, Manual on Presentation of Data Control Chart Analysis, 6th tions. See Practice D 6299 and MNL 79 for further guidance on
ed., available from ASTM International, W. Conshohocken, PA 19428. QC and control charting techniques.
8
D 3120 – 06
9
D 3120 – 06
SUMMARY OF CHANGES
Subcommittee D02.03 has identified the location of selected changes to this standard since the last issue
(D 3120–03a) that may impact the use of this standard.
(1) Changed Figs. X2.1 and X2.2 to Fig. 1 and Fig. 2, (2) Moved original paragraphs X2.1 and X2.2 to 6.1 and 6.2,
respectively. respectively.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org).
10