MULTIX Swing
AX
Disposal Instructions
System
Allprohibited.
is
© Siemens,
documents2005
Offenders
may only be
willused
be liable
for rendering
for damages.
services
All other
on Siemens
rights are
Healthcare
reserved.Products. Any document in electronic form may be printed once. Copy and distribution of electronic documents and hardcopies
© Siemens, 2005
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.
2005
2
10093895
Copyright
“© Siemens, 2005“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA,
Siemens Healthcare Diagnostics Inc. - USA and/or Siemens Healthcare Diagnostics Prod-
ucts GmbH - Germany.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.
Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Disposal Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General Information 0
Graphics
The illustrations and drawings may differ slightly from the system shipped.
Disposal Instructions 0
Hazardous Materials
Hazardous substances are written in "italics" in the following chapter. They have to be dis-
posed of by an authorized disposal firm in compliance with industry standards.
Safety Information
When performing installation and disposal, the product-specific safety information con-
tained in the documents, as well as the general safety information contained in the System
Manual must be observed.
Preparation
• If necessary for disassembly/transport, move the unit into an appropriate position.
• Completely switch off the system.
WARNING Ensure that the system and the on-site power supply are
completely de-energized prior to beginning deinstallation or
disposal procedures.
¹ The system and on-site power supply have to be
secured against switch-on.
• Remove the grid (1/Fig. 1 / p. 6) from the Bucky wall stand; lead.
• Remove the collimator (2/Fig. 1 / p. 6).
• Remove the housing.
Fig. 2:
• Remove the focus-near plates on the top of the collimator (2/Fig. 2 / p. 7); lead.
Fig. 3:
• Remove the grid (1/Fig. 3 / p. 8) from the Bucky wall stand; lead.
Fig. 5: Fig. 6:
• Remove the cover from the single-tank X-ray tube (1/Fig. 6 / p. 9) and disconnect the
cable connections.
• Remove the single-tank X-ray tube from the mounted component (e.g. the unit)
(2/Fig. 6 / p. 9).
• Remove the oil pressure switch (3/Fig. 6 / p. 9) and drain the oil into a suitable con-
tainer (approx. 12 liters).
Remove the shielding cone (4/Fig. 6 / p. 9), the shielding flange (5/Fig. 6 / p. 9) and the
shielding disk (6/Fig. 6 / p. 9) from the single-tank X-ray tube.
• Remove the cylinder-head screws from the single-tank X-ray tube (7/Fig. 6 / p. 9).
• Lift off the cover (8/Fig. 6 / p. 9) from the tank on the single-tank X-ray tube.
• Remove the internal shielding (9/Fig. 6 / p. 9), the radiation protective tube
(10/Fig. 6 / p. 9) and the radiation protective cap (11/Fig. 6 / p. 9).
• Remove the X-ray tube.
Fig. 7:
Fig. 8:
• Generator Control Console and Organ Attachment (Option)
Overview of Components 0
Disassembly 0
• Remove the lithium battery from the motherboard in the image system PC.
• If applicable, remove the grid from the detector; lead.