Error Analysis Chemistry 141 - Fall 2016: Random and Systematic Error
Error Analysis Chemistry 141 - Fall 2016: Random and Systematic Error
Error Analysis Chemistry 141 - Fall 2016: Random and Systematic Error
Error and uncertainty may seem synonymous with trivial mistakes in the lab, but they are
actually well-defined aspects of any numerical measurement in a laboratory experiment.
A number reported without consideration of precision tells an incomplete story, thus a goal in the
General Chemistry laboratory is to start thinking about the value and precision of any numerical
result—whether it is a laboratory measurement, a survey result, or a sports statistic.
The use of significant digits in calculations is covered in the textbook and is used in the lecture
portion of General Chemistry. Significant digits are an approximate approach to treat precision,
and they provide rules for addition and multiplication.
This guide will be sufficient for your work in General Chemistry. More information can be found
in John R. Taylor’s An Introduction to Error Analysis.*
One goal for lab work will be controlling the two types of experimental error: systematic
error and random error.
Systematic error arises from a flaw in experimental design or equipment and can be
detected and corrected. This type of error leads to inaccurate measurements of the true value.
The best way to check for systematic error is to use different methods to perform the same
measurement.
Random error is always present and cannot be corrected. It has to do with the precision of
measurements in laboratory and is the statistical uncertainty in the last digits of the precision.
An example of random error is that which arises from reading a burette, which is somewhat
subjective and therefore varies at each reading. Note: “Error” and “uncertainty” are
sometimes used interchangeably to mean “random error.” The phrase “error in a
measurement” is synonymous with “uncertainty in a measurement.”
One aim will be to eliminate systematic error and minimize random error to obtain a high
degree of both accuracy and precision. A goal of the General Chemistry laboratory will be to
practice thinking about the largest contributors to both types of error in our experiments.
CH141 Laboratory Error Analysis Fall 2016 page 2 of 9
Systematic error is corrected for in the lab procedure. Understanding random error comes
from repeated measurements to give a set of replicas. The statistics of the set of replicas
gives a way to understand the value and error in measurement. The statistical tools to be used
are the Mean, the Standard Deviation, and the Standard Deviation of the Mean.
The standard deviation of the mean (SDOM), , is also called the estimated standard
error. If the full experiment was repeated (all replicas), the spread in the average results
should be related to the standard deviation of the mean:
Using Excel
Here the discussion is limited to the mean, the standard deviation, and the standard deviation of
the mean. For example, when five different volumetric flasks were each used to measure 50.0
mL of water at 25 ˚C and then each quantity of water was weighed, the following masses in
grams were determined: 49.897, 49.938, 49.599, 49.965, 49.771. A spreadsheet was used to
calculate the mean, standard deviation and standard deviation of the mean as shown in Figure 1.
CH141 Laboratory Error Analysis Fall 2016 page 3 of 9
Figure 1. The actual appearance of the spreadsheet used to do the calculations is shown to
the left and the same spreadsheet appears on the right with the equations used to calculate
the quantities shown instead of the numerical result of those calculations.
The number of decimal places shown for the mean, SD, and SDOM were adjusted with the
buttons in the toolbar and using the SDOM to decide the last decimal place to report.
Remember the %error is equal to (SDOM/mean) *100.
Reporting Error
Whenever possible report numerical results with error. The quantity we’ll use to estimate
measurement error in General Chemistry is the SDOM, which will be reported with one
significant digit. This precision of the error will determine the precision of the measurement
(e.g., indicates the last decimal place to the right to report).
Significance
In General Chemistry lab, the word “significant” doesn’t always mean “important.” The
calculated error bars will be used to gauge the significance of measurements. If two measured
values are contained within each other’s error bars, the measured values are the same—with no
significant difference between them. If two measured values and their error bars are well
separated, then the two values are unique, and their difference is “significant.” There is a grey
area between those two extremes, and some results can be inconclusive. A full understanding of
the significance of inconclusive results requires other statistical tests such as a t-test. More
information can be found on page 150 of Taylor.1
Once a measurement has been made some dimensional analysis needs to be done. There are easy
rules to follow for multiplication and addition:
Multiplication/Division
Multiplication and division scale a quantity, so these operations also scale the error. The
mass of water in the example above can be used to calculate the density of water at 25 ˚C,
by dividing the mass by the volume. The absolute error will be divided by the volume as
well:
1 g
49.8 ± 0.1 g ✕ = 0.996 ± 0.002
50.0 mL mL
Addition/Subtraction
Addition and subtraction just shift a quantity, so they don’t change the error. For instance,
if the temperature of the room was measured to be 25.3 ± 0.3 ˚C, and that value was then
converted to Kelvin by adding 273.15 to the measured temperature, as shown below. The
absolute error is unchanged because Kelvin and Celsius have the same scale, just a
different origin.
25.3 ± 0.3 ˚C + 273.15 = 298.4 ± 0.3 K
CH141 Laboratory Error Analysis Fall 2016 page 5 of 9
!
Figure 2. The mass of water at 25 ˚C increases linearly with volume, giving a slope of
0.994 ±0.002 g/mL and a y-intercept of 0.06 ± 0.08 g.
Where did ±0.002 g/mL and ± 0.08 g in the figure caption come from? While it’s good to give
the R2-value to give a sense for the precision of the fit a better strategy is to use Excel’s LINEST
function to calculate the errors in these parameters.
8. The Excel table below shows exactly what the LINEST generates for output. The top two
rows of the LINEST function output contain the value and error for the slope and intercept.
These data were needed to make the Figure 2 caption above. Please make sure that the r2
reported for the slope is consistent with the value obtained from the equation for the line in
your graph. If you wish, you can look up the meanings of the other quantities reported by
LINEST as they are beyond the scope of this course and document.
Reference
*Taylor, J. R. An Introduction to to Error Analysis: The Study of Uncertainties in Physical
Measurements, 2nd Ed.; University Science Books: Sausalito, CA, 1997.
CH141 Laboratory Error Analysis Fall 2016 page 7 of 9
• Systematic Errors: Without any changes in the procedure, systematic errors are repeated if
the experiment is repeated. Systematic errors have a biased effect on the final results; systematic
errors make the final result high or low, but not both. Instrument calibration errors are examples
of systematic errors. Environmental effects can also be causes of systematic error, for example a
change in lab temperature changing the calibration of a balance or the volume of a flask. An
example of a systematic error from the CaCO3 precipitation experiment is that small particles
pass through the glass frits in a Gooch crucible, making the final precipitate mass too small.
Systematic errors affect the accuracy of the final results.
A given measurement can contribute to both random and systematic error. Non-integer
measurements always contribute to the random error. For example, a miscalibrated balance is a
source of random and systematic error. Systematic errors are often corrected by completing a
determination using a different method or by comparing results among different laboratories.
• Student Mistakes: Student mistakes are just student mistakes; they are neither random nor
systematic errors. Examples in this category are spills, misreading a device such as a burette,
misinterpretation of the procedure, incorrect handling of a micro-pipettor, and forgetting to rinse
out a beaker when doing a quantitative transfer. These errors are known and easily preventable,
if the experiment is repeated. Systematic errors occur with each repetition of the experiment,
assuming no changes in instrumentation. Mistakes should be noted in the Results section of your
report as mistakes.
recorded to the nearest 0.01 mL. However, visually estimating the volume to better than ±0.02
mL is difficult. Consequently the precision of the volume delivered by the burette is poorer than
±0.02 mL, since two readings are necessary. Correspondingly in the titration example, the
volume delivered by the buret at best is 15.67 ± 0.03 mL, or three significant figures.
Correspondingly, the final unknown concentration is officially known to three significant
figures. The conclusion is that the precision is determined primarily by the random error in the
burette readings and pipette. The random error in the other volume and mass determinations are
not consequential.
Systematic Measurement Errors: Every measurement is a potential source of systematic error.
However, with thoughtful construction of the procedure many measurements can be discounted
as significant sources of systematic error. So while the calibration of the glassware and the
balance used in a titration experiment are technically sources of systematic error, these errors are
easily avoided. The calibration of the balances is periodically checked using a registered
calibration mass. Standard volumetric glassware is certified by the manufacturer through
calibration against National Institute of Standards and Technology (NIST) traceable procedures.
In a titration experiment the only significant systematic errors are in the purity of the reagents
and the visual determination of the end point. The purity of the reagents also includes absorption
of moisture from the ambient air. So reagents that are susceptible to atmospheric moisture
absorption are usually kept in low humidity desiccators. In a titration, the primary systematic
error is the endpoint determination. The difference between the equivalence point and the
measured end point is called the titration error. A visual end point is always slightly beyond the
equivalence point because of the necessity of seeing the color change by eye. The result is that
the volume of titrant delivered is too large, giving a larger final concentration than the true value.
The conclusion is that the accuracy is determined primarily by systematic error in the end point.
Example Lab Report Section on Error Analysis
The discussion, above, gives the complete thought process for determining the most important
errors in the experiment. The section in the lab report that presents your conclusions is
disappointingly short, by comparison. For the titration example:
The precision is dominated by the random error of the volume readings of the burette and
volumetric pipette. The other volumetric glassware contribute insignificant random error.
The accuracy is determined by the systematic error in the visual detection of the end point.
The visual end point is at a volume larger than the equivalence point, giving a higher final
result than the true concentration value.