Operation of Sterility Testing Isolators and Validation Issues
Operation of Sterility Testing Isolators and Validation Issues
http://www.pharmamicroresources.com/
https://www.pharmig.org.uk
Sterility testing concerns
Introduction Introduction to isolators
Isolator decontamination cycles
Biological indicators
Validation of isolator cycles
Key operational decisions
Maintaining isolators
Major concern with the sterility test is the risk of false positives.
These can lead to expensive and lengthy failure
investigations.
Proving a false positive is extremely difficult.
X
X
A false positive is a microorganism that has been transferred
through cross-contamination by personnel or from the test
environment.
the isolator is
sealed from the
outside
environment.
To maintain the barrier with the The maintenance of isolators is important, especially
external environment, isolators in ensuring the minimization of leakage.
consist of either a flexible film Isolators are normally tested biannually for HEPA
filter issues.
for the outer wall or have a solid
wall envelope. Regular checks for leakage and pressure differential,
often with alarms in place.
VHP
• Reliant on a high vapour phase concentration - achieved by initial dehumidification of the air.
• The process uses 30-35% hydrogen peroxide liquid which is flash evaporated to achieve the
required vapour phase concentration.
• The process is always stopped before the dew point is reached - a “Dry” process.
Biodecontamination # 3
HPV
• The process also flash evaporates 30-35% hydrogen peroxide to create a vapour.
• No initial dehumidification.
• The process is brought beyond the dew point to saturated vapour conditions.
iHP
The isolator system consists of two solid wall Type 1 positive pressure isolators linked by a rapid
gassing port.
The gassing port and the isolators are attached to a gas generator.
Images of the isolator examined – taken by Tim Sandle
Introducing isolator
decontamination cycles
Pre-cycle Disinfection
Prior to final product items (and other material that is not sterilized and contained within double
wrapping) being loaded into the isolator or gassing port, items should be disinfected.
A rationale for the inclusion of this step is that gaseous disinfection requires a pre-cleaning step.
The target of the gaseous disinfection process is surface bio-decontamination and does not match
the performance of defined penetrative sterilization processes detailed in pharmacopeia.
Surface gaseous disinfection requires a starting condition wherein a surface is visibly clean of soiling.
Visible contamination or soiling may occlude microorganisms from the disinfection process.
There are some key steps to ensure an optimal hydrogen peroxide vapor disinfection process:
2. Development of the gas concentration in the target environment to saturated vapor conditions, past
dew point, and transition into liquid phase.
1. With this case study: at saturated vapor conditions, the gas concentration can hold no more
molecules, thus, the process of condensation formation and disinfection agent surface
deposition starts.
3. Re-evaporation of the surface condensate and removal of residual gas to target endpoint
concentration.
2. The generator then produces hydrogen peroxide vapor by passing aqueous hydrogen peroxide over
a vaporizer.
1. The gassing phase is validated to deliver a disinfection agent dose volume that will reach
required in the target area and on target surfaces.
3. The vapor is circulated at a programmed concentration in the air and held for a set period of time
(dwell phase).
4. After the hydrogen peroxide vapor has circulated in the enclosed space for a pre-defined period of
time, it is circulated and broken down into water and oxygen to safe levels, at 1 parts per million
(ppm).
ISOLATOR Cycle
Validation of isolator cycles
With this case study, the gassing cycles were validated and developed
using biological indicators.
Biological indicators
Hydrogen peroxide is a surface disinfectant; therefore, it is important that the biological indicators
are presented to ensure maximum surface area exposure.
To place in the isolator the envelope must be folded back so that the biological indicator does not
lying flat on a surface.
Kapton tape can be used.
After the exposure of each biological indicator, the discs must be transferred into tryptone soya
broth and incubated at 55-60 C for a minimum of 7 days.
Temperature range suitable for the recovery of thermophilic bacteria.
Biological indicators
An important part of the OQ and PQ stages is with the development of load patterns for sterility
testing.
For this: The different sterility test load combinations were mapped out.
These were evaluated for the smallest and largest configurations by assessing the different
load types (this was assessed by calculating the surface area and by an assessment of the
degree of absorbent material).
OQ data demonstrated that the maximum load was the worst-case.
Once assessed, the maximum load was evaluated against biological indicators using
different cycle parameters.
A cycle that achieved total biological indicator kill was chosen as a base cycle and performed
on three occasions.
Routine cycle parameters were then established by incorporating an ‘overkill’ for gassing agent
volume and dwell time to ensure a robust cycle.
During performance qualification each different load type was assessed using biological
indicators and the established routine cycle.
As a minimum this is ordinarily the largest size for each load type.
Requalification is to verify that the isolator system and gassing port continue to operate as
expected; that is to demonstrate that the gassing system continues to kill a known population of
resistant spores (a ‘kill’ or ‘no kill’ test).
In relation to this study proposal, the decision taken was to re-qualify the entire isolator system
using biological indicators on an annual basis.
Frequency of decontamination
There will be occasions when the inside of the isolator should be
cleaned and disinfected, such as following a spillage incident.
Environmental monitoring
Many isolators are held within a cleanroom or a controlled environment.
The isolator described in this study protocol was held within an EU GMP Grade D/ISO 14644
class 9 (‘in operation’) cleanroom.
Where a cleanroom is used, the cleanroom will have to meet various operational parameters
itself. These parameters will include:
Positive pressure differential to the outside corridor
Controlled temperature and humidity.
High efficiency particulate air (HEPA) (with filters of 99.995% efficiency)
Controlled air-changes
Particles monitored in the dynamic state
Room subject to viable environmental monitoring room falls under a routine maintenance
plan including classification.
If the time falls outside the set range, then the gassing cycle will probably be ineffective.
The cycle parameters of peroxide injection rates and phase times were initially developed to
ascertain the required gas concentration profile during the qualification.
Gassing dwell time is set during the cycle development stage of OQ for each different cycle
Positive pressure relative to the isolator room is required to prevent the ingress of
contamination from the outside environment into the enclosed clean zone.
The minimum specification of +20 Pa is the lowest setting to prevent this occurrence and to
leave some safety margin.
The maximum specification of 250 Pa relates to the integrity of the isolator chamber.
Positive pressure
From a starting pressure of 150 Pa, a timed sequence of monitoring leakage over a 90 second
period is undertaken. By the end of the 90 seconds the pressure must not fall by more than 25 Pa.
Although unidirectional airflow is not required for a sterility testing isolator, there is a theoretical
argument that the airflow should demonstrably remove contamination.
Air speed should be 0.45 meters per second (ms-1) at the filter face (±20%)
A set number of air changes is necessary for the dilution and removal of any potential
contaminants from the isolator environment.
Particle levels:
0.5 µm = 3, 520 counts per cubic meter 5.0 µm = 20 counts per cubic metre
Further reading:
Sandle, T. and Tours, N. (2013). Validation and Operation of a
Sterility Testing Isolator: a Study Proposal, Journal of Validation
Technology, Vol. 19, No.1
Summary
Pharmaceutical Microbiology:
http://www.pharmamicroresources.com/
Pharmig: https://www.pharmig.org.uk