Unidentified Alarm Occurrence / Run CPU Address Error: Point of Detection Application

This document describes unidentified alarm occurrences or run CPU address errors on TCD 2 and TCD 3. For TCD 2, it occurs when unidentified codes are read by the alarm interrupt more than 3 times without an address error. For TCD 3, it occurs when the unidentified address accessed by the DOC-200 PCB is detected more than 3 times. It provides steps to check the +5V voltage, and potentially replace components like the GAL, EPLD, or FLASH MEMORY chip on the DOC-200 PCB to resolve the issues.

Uploaded by

Quang Pham Duy
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
13K views1 page

Unidentified Alarm Occurrence / Run CPU Address Error: Point of Detection Application

This document describes unidentified alarm occurrences or run CPU address errors on TCD 2 and TCD 3. For TCD 2, it occurs when unidentified codes are read by the alarm interrupt more than 3 times without an address error. For TCD 3, it occurs when the unidentified address accessed by the DOC-200 PCB is detected more than 3 times. It provides steps to check the +5V voltage, and potentially replace components like the GAL, EPLD, or FLASH MEMORY chip on the DOC-200 PCB to resolve the issues.

Uploaded by

Quang Pham Duy
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Point of ● Run CPU ● DI120~150

TCD 2,3 Unidentified Alarm Occurrence / Run CPU Address Error Application
Detection ○ Speed CPU ● DI180~420

[Meaning] ■TCD 2: When un identified codes are read [Symptoms]


by ALARM Interrupt(IR1 of 82C59A) 3 or more consecutive times. 1.Check ALARM LED(LED3) of DOC-200 PCB flickering status.
[When an alarm interrupt occurs without an address error.] ● Normal →OFF
● Abnormal(Address ERROR) →ON
■TCD 3: When unidentified address by address map of
DOC-200 PCB access is detected 3 or more consecutive times. [Actions to take]
1. Check +5[V] voltage.
Must be at least +4.8[V].
[Diagram] 1) +5[V] : Connector MG/1, MG/6
2) GND : TP16(Located just above Connector MG)

2. As to TCD 2, GAL(U30), 82C59A(U26), or EPLD(U11) of


DOC-200 PCB may be failed.
: Replace GAL first. If TCD 2 error still exists, replace EPLD.
* Be careful not to damage socket during EPLD
replacement.(Be careful for electrostatics.)

3. As to TCD 3, DOC-200 PCB FLASH MEMORY(U12, U13) or


GAL inside program of the FLASH MEMORY may be failed.
Replace FLASH MEMORY.
Alarm Interrupt * Be careful not to damage or bend pins while replacing
read_alarm_code() FLASH MEMORY.(Be careful for electrostatics.)
4. If TCD 2 or 3 error exists even after above actions
SIS_ADDR_ERROR ERR_ADDR(TCD 3)
have been taken, replace DOC-200 PCB.
system_down()

Count >= 3
INC
Count

usr_reboot()

TCD 2

Undefined_Alarm_Count >3
TCD 3

INC
Undefined_Alarm_Count