Configuration Software AmeTrim User Manual US/ESP
Configuration Software AmeTrim User Manual US/ESP
1. INTRODUCTION .............................................................................. 4
3 2013-07-01 128160 03
1. Introduction
This manual contains operating instructions for:
’
JOFRA WINDOWS ADJUSTMENT PROGRAM
AMETRIM
Gydevang 32-34
DK-3450 Allerød
Denmark
Tel.: +45 48 16 80 00
Fax: +45 48 16 80 80
128160 03 2013-07-01 4
2. General information
You are advised to return the JOFRA instruments to AMETEK
Denmark A/S or an accredited laboratory at least once a year for
calibration.
Technical assistance
Please contact the distributor, if you require technical
assistance.
RTC series
PTC series
ATC series
DTI 100/1000
ASM-80x B
5 2013-07-01 128160 03
2.2 PCs, minimum hardware requirements:
Ametrim can be run direct from the USB for the ATC, DTI and ASM
instruments however the software requires JOFRACAL software
supplied on the same USB memory stick, to be installed before
Ametrim can be used together with RTC or PTC calibrators.
If you want to install the program manually, the USB also contains
a SETUP.EXE file in the Jofracal folder.
128160 03 2013-07-01 6
2.5 Setup Printer
If you are unsure how to use these settings, refer to your Windows
Help.
7 2013-07-01 128160 03
3. Starting the Ametrim software.
Note…
Before starting this software, the PC and the calibrator must
be connected together and the calibrator switched on (see
section 4.1).
The calibrator must not be performing any tasks like switch
test, autostep or workorders. That means that the calibrator
must be in the main menu before starting the software.
Use the normal Windows procedure to start the AmeTrim
software. If you are unsure of how to start software
programs, refer to your Windows Help
Caution…
Connect the USB serial cable provided to the USB device
port on the side of the calibrator and to a USB port on the
PC.
For communication via TCPIP connect the Ethernet cable to
the connection also on the side of the RTC calibrator.
128160 03 2013-07-01 8
3.3 Ametrim - Main menu.
The displayed device types on the left depend on the method and
type of comport selected.
9 2013-07-01 128160 03
Press “Connect” to start communications.
Note…
If no RTC appears in the list check it is properly connected
and Windows has registered a new hardware device.
This is required the very first time you connect a particular
RTC to a PC. Then press the button to refresh the list.
Note. Please refer to the RTC user manual regarding setting the IP
address on the actual calibrator.
128160 03 2013-07-01 10
4. AMETRIM – RTC/PTC/ATC main menu.
Note…
When software is used to control the calibrator, the
calibrator’s keyboard is disabled and the display indicates
that the calibrator is remote controlled.
11 2013-07-01 128160 03
Depending on the instrument type connected the following processes
may be activated or not:
128160 03 2013-07-01 12
4.1 Temperature Adjustment
This enables you to check the accuracy of the heat source and the
internal sensor in the calibrator.
This can be done in 3 ways:
13 2013-07-01 128160 03
Fig. 1
128160 03 2013-07-01 14
Click a radio button to select the scenario that should be
used, depending on which calibrated temperature reference
that is available.
Click a radio button to select the desired temperature units.
Ensure that the precision thermometer is positioned correctly
and ready to measure the temperatures.
Confirm that the calibration should be started.
Note.
An option exists for the RTC-700 the maximum temperature
can be set to 660C.
15 2013-07-01 128160 03
When using the external reference sensor input on the
calibrator, or when using a DTI, the calibration and
adjustment will now run automatically.
128160 03 2013-07-01 16
Click OK. The READ temperature measured by the calibrator
and the TRUE value are entered in the table.
The calibrator now starts working towards the next
temperature level.
Repeat these steps until READ and TRUE values have been
entered for each temperature step.
Note…
Downloading new coefficients to the calibrator overwrites
the existing ones already resident in the calibrator.
Note…
Before calibrating/adjusting the inputs the calibrator must
have been turned on for at least half an hour with the SET
temperature “Not Activated” (the initial mode when turned
on).
17 2013-07-01 128160 03
In order to calibrate the inputs the following calibrated signals must be
used:
RTD inputs:
1000Ω (800 to 1200Ω)
2700Ω (2300 to 2900Ω)
mA inputs:
0 mA (-1 to +1 mA)
10 mA (9 to 11 mA)
20 mA (18 to 22 mA)
V Inputs:
0V (-1 to +1 V)
5V (3 to 7 V)
10 V (9 to 11 V)
mV inputs:
0 mV (-1 to +1 mV)
35 Mv (25 to 45 mV)
70 mV (63 to 77 mV)
128160 03 2013-07-01 18
RTC Nominal values Accepted range
RTD and Reference Inputs:
18Ω (15 to 22Ω)
345Ω (330 to 350Ω)
DLC inputs
0 mV (-1 to +1 mV)
RTD 400 inputs:
18Ω (15 to 22Ω)
345Ω (330 to 350Ω)
RTD 4000 inputs:
18Ω (15 to 22Ω)
3450Ω (3330 to 3500Ω)
mA inputs:
0 mA (-1 to +1 mA)
20 mA (18 to 22 mA)
V Inputs:
0V (-1 to +1 V)
10 V (9 to 11 V)
mV inputs:
0 mV (-1 to +1 mV)
70 mV (63 to 77 mV)
19 2013-07-01 128160 03
PTC Nominal values Accepted range
RTD and Reference Inputs:
18Ω (15 to 22Ω)
345Ω (330 to 350Ω)
RTD 400 inputs:
18Ω (15 to 22Ω)
345Ω (330 to 350Ω)
RTD 4000 inputs:
18Ω (15 to 22Ω)
3450Ω (3330 to 3500Ω)
mA inputs:
0 mA (-1 to +1 mA)
20 mA (18 to 22 mA)
mV inputs:
0 mV (-1 to +1 mV)
70 mV (63 to 77 mV)
Cold Junction compensation: T-type thermocouple and a
reference junction, for
example an ice bath.
128160 03 2013-07-01 20
Principle sketch
Fig. 2
If at any point you do not wish to continue, click Stop. The software
exits this option and returns to the Main Menu.
By clicking the Calibration Date button, you can view the last
calibration dates for each of the inputs.
21 2013-07-01 128160 03
To adjust and calibrate the current Input values
Each input has several pre-set input values. Ensure that the
reference signal source is adjusted correctly when prompted.
For Cold Junction compensation, ensure the corresponding
V value for the calibrated junction temperature is entered
(0V @ 0C).
Note…
There is no reporting facility in AmeTrim for the RTC/PTC.
128160 03 2013-07-01 22
4.3 Axial uniformity adjustment. (RTC/PTC series).
Note…
For RTC-158/250 axial uniformity adjustment must not be
carried out when the calibrator is filled with liquid.
Adjustment must be carried out when using an insert.
23 2013-07-01 128160 03
RTC-156/157/158/159/250, PTC-125/155/350 scenario
Note…
Use AMETEK 127587 temperature sensor (recommended)
or similar with short measuring zone for axial uniformity
adjustment.
128160 03 2013-07-01 24
PTC-660 scenario
Note…
Use STS-150 A966 temperature sensor (recommended) or
similar with short measuring zone for axial uniformity
adjustment.
25 2013-07-01 128160 03
RTC-700 scenario
Note…
Use STS-200 A 970 temperature sensor (recommended) or
similar for axial uniformity adjustment.
128160 03 2013-07-01 26
Reference sensor
27 2013-07-01 128160 03
Select “ITS-90” or “Callendar van Dusen coefficients” and
type in the coefficients from the Calibration Certificate
supplied with the reference sensor.
Note…
These coefficients are not saved. Therefore, they must be
typed in and downloaded when they need to be used.
Default values
This option downloads a set of default values, which are stored in the
AmeTrim-ATC/DTI software.
128160 03 2013-07-01 28
4.4 Factory defaults ( RTC/PTC )
29 2013-07-01 128160 03
4.5 Update calibration dates
The page allows the user to update the calibration dates in the
RTC/PTC calibrator.
NOTE! For PTC series only the Heat source date is actual.
128160 03 2013-07-01 30
5. Using DTI with the AmeTrim
Select “DTI” and “Comport” in the “Device Quick Connect” dialog box
and press “Connect”.
Caution…
1. Ensure that both the PC and the DTI are switched
off at the mains. Failure to do so may result in your
equipment being damaged.
2. Connect the serial cable provided to the “RS 232” port on
the back of the DTI and to the COM port on the PC.
3. Switch on the PC and the DTI.
31 2013-07-01 128160 03
DTI main menu.
128160 03 2013-07-01 32
Reference sensor
33 2013-07-01 128160 03
Use the mouse to position the pointer in the boxes in the
table.
Note…
These coefficients are not saved. Therefore, they must be
typed in and downloaded when they need to be used.
Default values
This option downloads a set of default values, which are stored in the
AmeTrim-ATC/DTI software.
128160 03 2013-07-01 34
5.3 RTD input adjustment
50 ohm resistor
350 ohm resistor
DTI-1000 B :
12 ohm resistor
85 ohm resistor
35 2013-07-01 128160 03
Connect the reference resistors to the input channels and enter the
resistor values. Press “Start measurement”, wait until the readings for
“New low reference value” and “New high reference value” have
stabilized (approx. 2 minutes) and press “Lock new values” to stop
the measurement. The locked values can now be downloaded to the
DTI or cancelled.
Calibration date
128160 03 2013-07-01 36
5.4 Analog output adjustment
This option allows you to adjust the analog output of the DTI-1000.
For adjusting the analog output you need a voltmeter 0V to 5V range
with 0.1mV resolution and an accuracy of 0.1 mV.
After entering the DTI analog output adjustment window, press “Start”
to begin the adjustment procedure.
37 2013-07-01 128160 03
Measure the output voltages of the two analog outputs, enter the
values and press “Next”.
128160 03 2013-07-01 38
Repeat the measurement of the output voltage for high output, enter
the values and press “Next”.
39 2013-07-01 128160 03
6. Using ASM with the AmeTrim
Note…
Before starting this software, the PC and the ASM-80x
must be connected together and the ASM-80x switched on.
Caution…
1. Ensure that both the PC and the ASM-80x are switched
off at the mains. Failure to do so may result in your
equipment being damaged.
2. Connect the serial cable provided to the “RS 232” port on
the front of the ASM-80x and to the COM port on the PC.
3. Switch on the PC and the ASM-80x.
128160 03 2013-07-01 40
As the software starts, it detects the type of instrument connected to
the PC and reads its serial number.
This information plus the COM port the calibrator is connected to is
displayed at the bottom of the AMETRIM window.
41 2013-07-01 128160 03
6.2 Run calibration/adjustment
Note…
Before calibrating/adjusting the inputs, the ASM-80x must
have been turned on for at least half an hour.
128160 03 2013-07-01 42
The number and type of electrical inputs to be checked is user-
definable. You can also define if you want to check the inputs by just
measuring them, adjust and calibrate or to measure, adjust and then
calibrate the electrical inputs.
Principle sketch
If at any point you do not wish to continue, click Stop. The software
exits this option and returns to the Main Menu.
43 2013-07-01 128160 03
By clicking the Calibration dates button, you can view the
last calibration date.
128160 03 2013-07-01 44
Type a unique name for the measurements. Click As found
and ensure that the date is correct. This information will
appear in a measurement report displayed on screen.
Note…
These results are not saved electronically. If you require a
record of these measurements, click Print while it is
displayed.
Note…
If you select Cold Junction compensation, either you must
also select the mV (TC) Input or be sure that the mV(TC)
range is calibrated.
45 2013-07-01 128160 03
Connect the reference signal source to the appropriate
channel (when prompted).
Each input has several pre-set input values. Ensure that the
reference signal source is adjusted correctly when
prompted.
Note…
These results are not saved electronically. If you require a
record of these measurements, click Print while the report
is displayed.
128160 03 2013-07-01 46
When you are finished viewing or printing the report, click
Close. This returns you to the first dialogue in the
Temperature adjustment option.
To calibrate (as found), adjust and calibrate (as left) the ASM-80x
Note…
If you select Cold Junction compensation, either you must
also select the mV (TC) Input or be sure that the mV(TC)
range is calibrated.
Each input has several pre-set input values. Ensure that the
reference signal source is adjusted correctly when
prompted.
47 2013-07-01 128160 03
When all the selected inputs have been measured, adjusted
and re-calibrated a new dialogue appears.
Note…
These results are not saved electronically. If you require a
record of these measurements, click Print while the report
is displayed.
128160 03 2013-07-01 48
7. Using ATC with the AmeTrim
Caution…
1. Ensure that both the PC and the calibrator are switched
off at the mains. Failure to do so may result in your
equipment being damaged.
2. Connect the serial cable provided to the “RS232” port on
the front of the calibrator and to the COM port on the
PC.
3. Switch on the PC and the calibrator.
49 2013-07-01 128160 03
Note…
When software is used to control the calibrator, the
calibrator’s keyboard is disabled and the display indicates
that the calibrator is remote controlled.
F1 F2 F3 F4 F5
128160 03 2013-07-01 50
7.2 Adjusting and calibrating the instrument
51 2013-07-01 128160 03
7.3 Temperature Adjustment
Temperature Adjustment
This enables you to check the accuracy of the heat source and
the internal sensor in the calibrator.
This can be done in 3 ways:
Using the external reference sensor on the ATC (B-version) to
measure the TRUE temperature, AmeTrim-ATC can perform a
fully automatic calibration and adjustment of the internal
sensor. The calibrated reference sensor should then be
connected to the reference input.
Using a Digital Temperature Indicator (DTI) AmeTrim-ATC can
read the TRUE temperature from the DTI and also make a
fully automatic calibration and adjustment of the internal
sensor. The DTI must be connected to a vacant serial COM
port. See fig. 3.
Using any other reference instrument, AmeTrim-ATC will set
the temperature sequence and the TRUE values can be
entered as they occur as manual inputs.
128160 03 2013-07-01 52
Sensor
DTI
TS
N
U ME
TR
INS
N
IO
AT
I BR
C AL
F5
F4
F3
F2 9
F1 8
6
7
5
3
4
C 2
ES
.
1
i 0
_
Computer
RS232 cable
Fig. 3
53 2013-07-01 128160 03
Click a radio button to select the scenario that should be
used, depending on which calibrated temperature reference
that is available.
Click a radio button to select the desired temperature units.
Click a radio button to select whether the calibration only
should be performed as an “As Found” which means without
adjusting the internal sensor, or if it should be both as an “As
found” measurement followed by and adjustment of the
sensor and a final “As Left” calibration after the adjustment.
After the calibration measurement reports on both
calibrations can be printed.
Enter the ambient temperature and the operator initials.
These are to be used on the printed measurement reports.
128160 03 2013-07-01 54
Ensure that the precision thermometer is positioned correctly
and ready to measure the temperatures.
Confirm that the calibration should be started.
Information on the last calibration date for the reference input
is read from the calibrator and given to help you ensuring an
adequate calibrated reference.
55 2013-07-01 128160 03
When using the external reference sensor input on the
calibrator, or when using a DTI, the calibration and
adjustment will now run automatically.
128160 03 2013-07-01 56
the new dialogue, type in the correct “TRUE temperature”
value read from the external precision thermometer.
Repeat these steps until READ and TRUE values have been
entered for each temperature step.
Note…
If you want a record of the measured TRUE and READ
values, you must click Print now. These values are not saved
once you return to the AmeTrim Main Menu.
Note…
Downloading new coefficients to the calibrator overwrites
the existing ones already resident in the calibrator.
57 2013-07-01 128160 03
Click Done to exit the Temperature adjustment option.
Note…
Before calibrating/adjusting the inputs the calibrator must
have been turned on for at least half an hour with the SET
temperature “Not Activated” (the initial mode when turned
on).
128160 03 2013-07-01 58
In order to calibrate the inputs the following calibrated signals must be
used:
59 2013-07-01 128160 03
Cold Junction compensation requires a special setup, using a
calibrated temperature source, for example an ice bath. This and the
measurement principles are shown briefly below:
TS
EN
UM
TR
NS
NI
TIO
I BRA
C AL
F5
F4
F3
F2 9
F1 8
6
7 5
3
4
ES
C 2 .
1 0
i _
Principle sketch
Fig. 4
If at any point you do not wish to continue, click Stop. The software
exits this option and returns to the Main Menu.
128160 03 2013-07-01 60
In AmeTrim-ATC main menu, click Input adjustment.
By clicking the Calibration Date button, you can view the last
calibration dates for each of the inputs.
61 2013-07-01 128160 03
Type a unique name for the measurements. Click As found
and ensure that the date is correct. This information will
appear in a measurement report displayed on screen.
Note…
These results are not saved electronically. If you require a
record of these measurements, click Print while it is
displayed.
Note…
If you select Cold Junction compensation, either you must
also select the mV (TC) Input or be sure that the (TC) Input
is calibrated.
128160 03 2013-07-01 62
Connect the reference signal source to the appropriate
input(s) (when prompted).
Each input has several pre-set input values. Ensure that the
reference signal source is adjusted correctly when
prompted.
Note…
These results are not saved electronically. If you require a
record of these measurements, click Print while the report is
displayed.
63 2013-07-01 128160 03
When you are finished viewing or printing the report, click
Close. This returns you to the first dialogue in the
Temperature adjustment option.
This option enables you to view the calibration values currently loaded
in the calibrator for the reference sensor as well as enter and
download values for new reference sensors. In addition to the
equipment already described in sections 7.1, you also require the
calibration certificate for the new reference sensor.
In AmeTrim-ATC main menu, click Reference sensor.
128160 03 2013-07-01 64
To view current values in the Calibrator
Click Done when you are finished viewing the values. This
returns you to the AmeTrim-ATC Main Menu.
Default values
This option downloads a set of default values, which are stored in the
AmeTrim-ATC software.
65 2013-07-01 128160 03
7.6 Managing DTI sensor coefficients
This dialog enables you to type in coefficient values for the reference
sensor connected to the DTI
Note…
These coefficients are not saved. Therefore, they must be typed
in and downloaded when they need to be used.
When you are satisfied with all the values, click Download.
Default values can be selected by using the Default button.
128160 03 2013-07-01 66
AMETEK Sensors, Test & Calibration
A business unit of AMETEK Measurement &
Calibration Technologies Division offering the
following industry leading brands for test and
calibration instrumentation.
Crystal Pressure
Digital pressure gauges and calibrators that are
accurate, easy-to-use and reliable. Designed for
use in the harshest environments; most products
carry an IS, IP67 and DNV rating.
www.ametekcalibration.com
United Kingdom USA Singapore
Tel +44 (0)1243 833 302 Florida Tel +65 6484 2388
[email protected] Tel +1 (800) 527 9999 [email protected]
[email protected]
France China
Tel +33 (0)1 30 68 89 40 California Shanghai
[email protected] Tel +1 (800) 444 1850 Tel +86 21 5868 5111
[email protected]
Germany Beijing
Tel +49 (0)2159 9136 510 India Tel +86 10 8526 2111
[email protected] Tel +91 22 2836 4750 [email protected]
[email protected]
Denmark
Tel +45 4816 8000
[email protected]
Information in this document is subject to change without notice. ©2016, by AMETEK, Inc., www.ametek.com. All rights reserved.