PTS 2000
PTS 2000
PTS 2000
0
...................................................
User’s Manual
Operators Manual
....................................................................
Copyright
© Puritan-Bennett Corporation 1999, 2003. All rights reserved.
The information contained in this manual is the sole property of
Puritan Bennett and may not be reproduced or copied without the permission of
the Company.
This manual may be revised or replaced by Puritan Bennett at any time and
without notice. Users always should ensure that they have the most current
version applicable to their equipment. While the information set forth
herein is believed to be accurate, it is not a substitute for the exercise of
professional judgement.
The equipment described in this document should only be used by trained
professionals. Puritan Bennett’s sole responsibility with respect to such
equipment, and its use, is as stated in the limited warranty provided with
such equipment.
Nothing in this manual shall limit or restrict in any way Puritan Bennett’s right
to revise or otherwise change or modify the equipment described herein,
without notice. In the absence of an express, written agreement to the contrary,
Puritan Bennett has no obligation to furnish any such revision, changes,
or modifications to the owner or user of the equipment described herein.
Manufactured for Authorized Representative
Puritan-Bennett Corporation Tyco Healthcare UK LTD
4280 Hacienda Drive 154 Fareham Road
Pleasanton, CA 94588 USA Gosport PO13 0AS, U.K.
Toll Free: 1.800.635.5267
.......
Warning
Indicates a condition that can endanger operator or
service person.
Caution
Indicates a condition that can damage the equipment.
NOTE:
Indicates points of particular emphasis that make operation of
the PTS 2000 more efficient or convenient.
Warning
Possible explosion hazard if used in the presence of
flammable anesthetics.
Warning
ELECTRICAL SHOCK HAZARD: Only qualified and trained
personnel are authorized to service the PTS 2000 or replace
the unit’s O2 sensor. Use only factory-approved parts and
procedures to service this instrument. Contact Puritan Bennett
service personnel for further information.
.......
Caution
Always use dry gas; humidified gas will affect the accuracy of
flow measurements.
Caution
Always use the Re/Flex or D/Flex filter assembly (or equivalent)
in-line in the high-flow inlet hose; if the filter is not used
correctly, performance can be impaired.
Caution
Always cap the flowmeter inlets and outlets when the
instrument is transported or not in use.
Warning
Grounding reliability can be achieved only when the equipment
is connected to a receptacle marked “HOSPITAL GRADE.”
Serialization
The PTS 2000 test instrument has a unique serial number
that is used to track calibration dates and identify the unit’s
year of manufacture.
Customer assistance
For further assistance contact your local Puritan Bennett
representative.
.......
.......
Section 7 Appendix
7.1 Replacing O2 sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
.......
Section 7 Appendix
Figure 7-1. Location of oxygen sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
.......
Section 7 Appendix
.......
.......
V-00060
V-00049
NOTE:
See Section 5 for specification and part number information.
V-00050
Item Description
Symbol Meaning
.......
V-00045
V-00054
1 2 3
7 6 5 V-00044
1. Contains O2 sensor
.......
V-00052
NOTE:
Standard default unit of measure appear as slpm. Other
measurements of lpm (liters per minute), cfm (cubic feet
per minute), and scfm (standard cubic feet per minute) may
also be selected.
.......
Caution
To protect the flow sensors from contamination and
condensation when not in use, the protective caps should be
fully seated over the high flow and low flow inlet ports.
1 2 3 4 5
V-00054
+5 VDC 1A
Caution
The barometric port must always be open to atmosphere;
any obstruction can damage the pressure sensors or result in
an invalid reading.
.......
2
1
V-00053
1. HIGH FLOW EXHAUST: exhaust port for the high flow inlet
(shown with protective cap off).
2. LOW FLOW EXHAUST: exhaust port for low flow inlet
(shown with protective cap off).
Caution
To protect flow sensors from contamination and
condensation when not in use, the protective caps should be
fully seated over the high flow and low flow exhaust ports.
1.2.5 Filters
Because PTS 2000 flow sensors are susceptible to contamination
and condensation, two reusable filters are provided with the
PTS 2000: the Re/Flex filter is installed in-line with the high flow
inlet port, and the low flow input filter is installed in-line with
the low flow inlet port. These filters eliminate contaminants
from the gas flow before it passes through the flow transducers.
The reusable Re/Flex filter is designed for repeated autoclaving,
while the D/Flex filter must be discarded after use. Both filters
have barbed connections at both ends and require a silicone
coupling at the output port which connects to the tester.
An arrow on the top each filter indicates air flow direction.
These components are identified in Figure 1-10.
1. Re/Flex filter
2. D/Flex filter
3. low flow filter
4. couplings
.......
Re/Flex
filter
D/Flex
filter
Low flow
filter V-00009
Caution
• When high and low inlet/exhaust ports are not in use,
fit protective caps over the ports to protect from
outside contaminates.
.......
V-00055
NOTE:
Each time the PTS 2000 power switch is turned off and on,
the pressure transducers are automatically reset, setting the
baseline to zero. (For additional information see Section 4.)
1.2.7 Cleaning
Clean exterior surfaces of PTS 2000 and power supply with a
clean cloth using isopropyl alcohol, hydrogen peroxide (3%), or
ammonia (15%).
.......
Caution
The PTS 2000 is designed for use only with the power supply
provided. Do not attempt to operate the PTS 2000 using any
other power source.
V-00056
Warning
Position all hoses and wires so they do not present a hazard
to the people working around them or interfere with the
operation of the tester and the device being tested.
Low flow
filter
Re/Flex Coupling
filter
V-00057
V-00058
V-00059
V-00060
V-00061
NOTE:
Each time the PTS 2000 power switch is turned off and on,
the pressure transducers are automatically zeroed.
Do not apply pressure to low or high pressure ports
during powerup.
NOTE:
If PTS 2000 has not been maintained in an environment
close to normal room temperature, allow additional time
for warmup.
NOTE:
The unit will not display duplicate parameters on a
test screen.
.......
NOTE:
The trigger level is a global value. Changing it in one
location on one test screen will change it in every location
on all test screens.
.......
VOLUME L mL ft3
NOTE:
The REMOTE OPERATION mode requires RS-232 interface.
.......
V-00063
V-00062
V-00066
2.5.4.5 Volume
The VOLUME test function measures air, air/O2 mix, and O2.
V-00064
NOTE:
Before performing a high pressure measurement, the high
pressure baseline should be reset. See Section 4.1.1.
.......
V-00065
2.5.4.8 Barometric
BAROMETRIC is an informational prompt only and displays the
atmospheric barometric pressure.
2.5.4.9 Oxygen
The OXYGEN test function measures O2 content of a gas at the
high flow port.
.......
2.5.4.12 E time
E TIME is a shortened expression for Expiratory Time. E TIME is
the duration in seconds that the measure of HIGH FLOW is
continually less than, or equal to, the set threshold level. The
starting point of E TIME is considered valid when at least 250
msec of HIGH FLOW data is measured equal to, or below, the
threshold level. E TIME is measured in seconds (sec).
Auxiliary menu 3 3
...................................................................
3.1 Purpose and functions
The AUXILIARY MENU is accessed via the UP or DOWN ARROW
keys from the start-up screen. This menu provides edit options
for units of measure, and provides access to a variety of
diagnostic tests. Once accessed, five menu items will appear:
• CALIBRATE OXYGEN
• FLOW COMPARE
• DEFAULT TESTS
• CLEAR ALL TESTS
• EDIT MODE ACCESS
Press ENTER to activate AUXILIARY MENU. The active menu
item will flash on LCD screen, and applicable test may be
initiated. If another active menu item is desired, press UP or
DOWN ARROW key.
OXYGEN CALIBRATION
Press ↵ to start.
Press ESC to cancel.
NOTE:
There will be a delay until the PTS 2000 registers a steady
oxygen sensor reading. The screen will continue to display a
message indicating that the unit is acquiring 21% O2. Once
this occurs, the PTS 2000 will display a message
SEND 100% O2.
.......
V-00067
FLOW COMPARE
High Low
x.xx slpm
Low Flow
x.xxx slpm
xx % Error
No
Yes
No
Yes
.......
No
Yes
Secondary functions 4 4
...................................................................
4.1 Clearing data
This section provides instructions for clearing data entered into
the PTS 2000: zeroing pressure transducers, clearing peak
pressure values, and clearing breathing parameters.
NOTE:
Each time the PTS 2000 power switch is turned off and on,
the pressure transducers are automatically zeroed. Do not
apply pressure to the low or high pressure ports during
power up.
NOTE:
If pressure and breathing parameters or peak values are both
displayed, pressure will be zeroed and values will be cleared.
Caution
Do not use the PTS 2000 if the unit has been dropped, or if
liquid penetrates the unit.
.......
Repacking graphic
V-00068
N/A 1 1/ -inch
8 flow quick-connect coupling (female)
Specification Range
Input frequency 50 to 60 Hz
Specification Range
Specification Range
Specification Range
Specification Range
Specification Range
NOTE:
Performance specifications are subject to change without
notice and apply only to units that are used, maintained,
and calibrated according to the procedures described
in this manual.
Caution
Only the manufacturer’s cables or an equivalent double-
shielded, host interface cable should be used with this
equipment. Other types of cables may violate FCC rules and
regulations. Also, changes or modifications to the electronics
or enclosure of this product must be expressly approved by
Puritan Bennett; otherwise, the user’s authority to operate
the equipment may be voided by the FCC.
Glossary of terms 6 6
...................................................................
Term Description
ac alternating current
ATM atmosphere
bar bars
dc direct current
L liters
mbar milibars
mL mililiters
PC personal computer
Appendix 7 7
...................................................................
7.1 Replacing O2 sensor
The oxygen sensor should be replaced every two years or as
often as necessary. The sensor’s life depends on the operating
environment. The sensor’s life span is reduced if used in an
environment of higher temperatures or one that is oxygen rich.
To change the oxygen sensor in the PTS 2000 (version 4.0
firmware), follow these instructions:
1. Turn off power to PTS 2000 and unplug from power source.
2. Remove six small screws that fasten top cover of PTS 2000.
Lift off top cover.
3. Locate oxygen sensor.
NOTE:
O2 sensor is located behind front panel as shown in
Figure 7-1.
.......
Front panel
V-00048
Caution
When installing the O2 sensor, do not overtighten. This
could break the O-ring seal on sensor and cause high flow
flowmeter to leak.