Minco: MINCO Temptran Field and Bench Calibration
Minco: MINCO Temptran Field and Bench Calibration
Applies to Temptran models: TT110, TT111, TT115, TT160, TT176, TT210, TT211, and
TT220.
Depending upon accuracy requirements, two calibration procedures are described in this
manual. The first method described improves the accuracy of the Temptran at a single
temperature within the temperature range, while possibly degrading accuracy at the zero
and/or span temperatures.1 Accuracies of 0.2% of span or better are possible, but are very
dependent on the accuracies of the equipment and measurement techniques used.
The second method, bench calibration, will enable you to calibrate your Temptran to within
0.2% of span over the full Temptran range. This method takes into account the basic accuracy
of the Temptran, and its warm-up drift. Total system accuracy is the sum of the Temptran's
accuracy and the RTD's2 accuracy. The RTD's accuracy must include the RTD's tolerance
and drift specifications. However, calibrating a Temptran to a specific RTD effectively removes
RTD tolerance from system accuracy specifications, resulting in a better system accuracy.
In the calibration instructions below, be aware that getting good accuracy is very difficult.
BE ABSOLUTELY POSITIVE THAT THE TEMPTRAN NEEDS CALIBRATION. The
accuracy specifications of voltmeters, current meters, and temperature meters may be as poor
as ±0.5% of reading, which is five times worse than the Temptran's ±0.1% factory calibration
accuracy. It is best to calibrate using instruments whose accuracies are five to ten times better
than that of the Temptran's. This should minimize the cumulative errors.
1
The zero and span temperatures are, respectively, the lowest and highest temperatures of the
Temptran range. They are denoted as TZERO and TSPAN. These have a different meaning from the term
temperature span. Temperature span is the number of degrees, or the difference, between the span
and zero temperatures.
2
RTD – Resistance Temperature Detector. An RTD is a resistor whose resistance changes with
temperature.
Equipment needed:
• Portable temperature meter.
• A Hewlett Packard 3468A 5½ Digit multimeter or equivalent.
• A small screwdriver to make adjustments to the zero and span screws.
Procedure:
1. Connect a current meter in series with the 4-20 mA loop. See Figure 1.
2. Measure the temperature at the sensor and calculate what the current from the transmitter
should be using the following formula:
Notes:
1) The zero pot's adjustment is limited to about ±5% of span.
2) A change in the Zero setting affects the whole temperature range; that is, if you
increase the Zero by 2 degrees, you have increased every other reading by
approximately 2 degrees.
Equipment needed:
• A small screwdriver to make adjustments to the zero and span screws.
• Two decade boxes with milliohm resolution.
• A Hewlett Packard 3468A 5½ Digit multimeter or equivalent.
• DC power supply.
• RSENSE resistor, see text for accuracy and value.
Procedure:
2
eliminate tolerance and temperature drift errors. The resistor should have a tolerance
of ±0.01% and a temperature coefficient (TC) of 10 ppm/°C or less. Use the 4-wire
measurement mode (Kelvin connection) on your ohmmeter to verify that the resistor is
within the range of 100Ω ± 0.010Ω (99.990 to 100.010Ω).
b. It is best to use a supply voltage that the Temptran will see during normal operation
(typically, 24 VDC). This will minimize self-heating and supply voltage effects.
However, make sure there is sufficient supply voltage so the current loop functions
properly.
c. Set each decade box to one of the two resistances printed on the side of the Temptran.
Measure the resistance from the ends of the leadwires attached to the decade box
using a Kelvin connection. Adjust the decade boxes as necessary to achieve the
correct resistances. The setting numbers may not match the exact resistance, so
make a note of the box settings if necessary.
Notes:
1) Models TT110, TT111, TT115, TT160, TT176, TT210, TT211, and TT220 use nominal
RTD resistances. The two resistance values are printed on the Temptran's label or
found in a resistance vs. temperature table available from Minco. When using nominal
(ideal) resistance figures, your calibration ignores the interchangeability tolerance from
one RTD to the next.
2) Models TT150, TT151, TT155, TT200, TT676, TT710, TT711, and TT720 use each
RTD's unique zero and span resistances which are printed on the Temptran's label.
Because each transmitter of this type has been factory-calibrated to one specific RTD
sensor, the RTD's are not interchangeable. Make sure to keep the same transmitter
with the same sensor. To assist you, a matching number has been placed on both the
Temptran and the RTD.
Option:
Instead of using the resistances printed on the side of the Temptran, you can measure the
resistance of the RTD in a temperature bath at the Zero and Span temperatures. Then,
use these resistances where you would have used the label resistances. The temperature
of the bath and the resistance readings of the RTD's must be made carefully and exactly.
3. Switch to the decade box which is providing the 4 mA (Zero) resistance and adjust the
Temptran's Zero adjustment screw. The location of the two screws is marked on the label
on the side of the transmitter. Adjust the screw until the voltmeter reads 0.40000 ±0.00320
volts (0.39680 to 0.40320 volts).
4. Switch to the decade box which is providing the 20 mA (Span) resistance and adjust the
Temptran's Span adjustment screw until the voltmeter reads:
2.00000 ±0.00320 volts (1.99680 to 2.00320 volts).
5. Repeat these steps two or three times until unit is calibrated at both points.
6. Give the unit a light rap on the table to seat the pots. Check both the zero and span again;
repeat calibration as necessary
3
Figure 1 Wiring Diagram