0% found this document useful (0 votes)
94 views59 pages

CS-1 Program V2.00R04 Version Upgrade Instructions

This document provides instructions for upgrading the version of a CS-1 program from V1.10R01 or earlier to V2.00R04. The instructions include preparing for the upgrade, performing a new installation of the CS-1 program if needed, and upgrading the existing program. Key steps involve backing up files, upgrading the CS-1 program software, updating definition files and registry settings, and checking for any other necessary updates depending on the existing version.

Uploaded by

phcalvinsxe
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
94 views59 pages

CS-1 Program V2.00R04 Version Upgrade Instructions

This document provides instructions for upgrading the version of a CS-1 program from V1.10R01 or earlier to V2.00R04. The instructions include preparing for the upgrade, performing a new installation of the CS-1 program if needed, and upgrading the existing program. Key steps involve backing up files, upgrading the CS-1 program software, updating definition files and registry settings, and checking for any other necessary updates depending on the existing version.

Uploaded by

phcalvinsxe
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 59

CS-1 Program V2.

00R04

Version Upgrade Instructions

1st Version June 02, 2005


Approved Examined
Made by
by by
Contents

1. Preparation..................................................................... 2

2. New Installation ............................................................. 4

3. Version upgrade of existing program ............................. 5

4. Portable Mode............................................................... 51
->If ”Portable Mode” is set, this procedure must be performed.
5. Appendix ....................................................................... 53

-1-
1. Preparation
- When version upgrading the CS-1, version upgrade of the software of the reader (For
REGIUS170, V1.00R22 or later (MCB), V1.00R09 or later(SCB_CF)) will be also required.
For REGIUS190, if the factory default version is V1.00R21 or later (MCB), this upgrade is
not necessary.
Please have Update-kit2 : Ver1.00R09 [R170/R190 Firmware Upgrade Instructions] ready.
- For a new CS-1 installation, have [CS-1 Program V1.10R03] ready.

Version of each device should be combined with the version shown in the [Current System]
or the [Version upgraded System] of the following chart.

*If any CS-1 or REGIUS170 has been version upgraded in the existing system, the version of
all the devices has to correspond to the version of [Version upgraded System] of the chart.
*For the version upgrade procedures, start with the CS-1 that include JM.
Image reading is possible in a tentative mixed system of old and new. However ,under the
mixed environment, compatibility cannot be partially retained. Immediately version
upgrade all the devices (On the day or day after).
Current System Version upgraded System

Version upgrade

CS-1 to V1.10R03 CS-1 V2.00R01 or later

Version upgrade

REGIUS 170 REGIUS 170


To V1.00R18(MCB) V1.00R22(MCB) or later
To V1.00R04(SCB_CF) V1.00R07(SCB_CF) or later
When setting up

CS-3 V2.00R01 or later

REGIUS 190 V1.00R21(MCB)or later


V1.00R07(SCB_CF)or later

-2-
*1) If a SUPER DISK is not write-protected, “RmtGroup.Dbg” file will be created.
This file can be deleted as it is only used for installation.
*2) A version up CD can be copied to the CS-1 via the network.
In that case, create a folder (For example, the name of the folder is “CD”) and
copy all on the CD to the folder created.
However, delete the copied files after version upgrading.
*3) For a new installation, installation Date and installation Time displayed on the
System Tool won’t be updated by installing the CS-1 program V2.00R04.

-3-
2. New installation
Follow the procedures in this manual to newly install a CS-1 and a CS-3 at the same time.

2.1 Follow the new installation instructions [CS-1 Program V1.10R03] to newly install a CS-1.

2.2 Follow 3. Version upgrade of existing program to version upgrade the CS-1 in this manual.

-4-
3. Version upgrade of existing program

Follow the procedures below to upgrade the version of the CS-1 program for an existing
CS-1. Procedures differ depending on the version of the existing CS-1 program.

3.0 Preparation................................................................................................... 6
Æ Verification is required irrespective of the version of the existing CS-1
program.
3.1 Version upgrade of the CS-1 program.......................................................... 7
Æ This procedure is mandatory irrespective of the version of the existing
CS-1 program.
3.2 Updating of definition file/registry ............................................................ 17
Æ This procedure is mandatory irrespective of the version of the existing
CS-1 program.
3.3 Canceling of printer no-connection provisional setting............................. 20
Æ This procedure allows setup when the existing CS-1 program is V1.00R03T2 or
earlier and the printer no-connection provisional setting is enabled.
3.4 Version upgrade of process parameters/exam tags ................................... 22
Æ This procedure is mandatory irrespective of the version of the existing
CS-1 program.
3.5 Correction of process parameter of shoulder joints.................................. 42
Æ This procedure is mandatory the existing CS-1 program is V1.10R01T1.
3.6 Input control cancellation of Latin characters .......................................... 44
Æ This procedure should be performed only when the existing CS-1 program is
V1.00R03E (V1.00R03T3) and Latin1 characters are used in European
countries.
3.7 Notes on LCD display gradient .................................................................. 45
Æ Verification is mandatory the existing CS-1 program is V1.10R00 or earlier.
3.8 Notes on manufacturer name to be output to HOST ................................ 46
Æ This procedure should be performed when the existing CS-1 program is
V1.00R04 or earlier.
3.9 Notes on Study Description to be output to HOST ...................................... 47
Æ This procedure should be performed when the existing CS-1 program is
V1.10R02E or earlier.
3.10 Editing and notes for image output gradient .......................................... 48
Æ This procedure is mandatory irrespective of the version of the existing
CS-1 program.
3.11 Setting up the HIPAA functions (Security function) ................................ 50
Æ This procedure is required only when security functions that
satisfy the HIPAA Law are necessary. If the functions are already set up, no
further setup is required.

-5-
3.0 Preparation
The contents of the files overwritten or initialized by the version upgrade should be taken
down. Perform the prior preparation to the measures for handling image density errors
resulting from SG value computation errors.

3.0.1 Backing up the display LUT


This procedure should be performed when the existing CS-1 program is
fromV1.00R00T1 to V1.10R01T1.
Copy the folder C:¥ ¥Konica¥CS-1¥Env¥Manage¥Crt_Lut to the desktop using
Windows Explorer.
If the CS-1 program is V1.00 and a display luminance calibration is executed, a
backup procedure should be executed to convert the display LUT in 3.2 Definition
file/Registry.

3.0.2 Since the following files will be updated as a result of the version upgrade, take note
of any changes and re-set them after completing the version upgrade.
This procedure should be performed when the existing CS-1 program is
fromV1.00R00T1 to V1.10R01T1.
The following files may be modified if non-standard information is obtained and output
from and to RIS.
-C:¥Konica¥CS-1¥Env¥System¥RisInDicomData1.ini
-C:¥Konica¥CS-1¥Env¥System¥RisInDicomData2.ini
-C:¥Konica¥CS-1¥Env¥System¥RisOutDicomData1.ini
-C:¥Konica¥CS-1¥Env¥System¥RisOutDicomData2.ini

The following files may be modified if information other than that available as
standard on the service tool is used for name printing.
-C:¥Konica¥CS-1¥Env¥System¥StampFormE.ini
-C:¥Konica¥CS-1¥Env¥System¥StampFormJ.ini
-C:¥Konica¥CS-1¥Env¥System¥StampFormO.ini

3.0.3 Since setup of the printer-linked overlay of the HOST definition file is cleared, if it
has been modified, take note of any changes and re-set them after completing the
version upgrade.
This procedure should be performed when the existing CS-1 program is
fromV1.00R00T1 to V1.10R2T1.

-6-
3.1 Version upgrade of the CS-1 program
Procedures differ depending on the HIPAA function setup as switched over to the REGIUS
service screen.
For HIPAA Mode Off, setup should be conducted following Sections 3.1.1.1 to 3.1.1.8.
For HIPAA Mode On, follow the procedures in Sections 3.1.2.1 to 3.1.2.5.

3.1.1 HIPAA Mode Off

3.1.1.1 Boot the CS-1. After booting, the following screen will be displayed. Note, however,
that a different screen may be displayed when an “Exam List” is selected for the
screen format, but similar operations to those described here can be performed.

Version of V1.10R01 or earlier Version of V1.10R01 or later

3.1.1.2 Click Konica REGIUS, and the following screen will be displayed (V1.10R01 or
earlier)
Click KONICA MINOLTA, and the following screen will be displayed (V1.10R01 or
later.)

-7-
3.1.1.3 Click Utility, and the following screen will be displayed.

3.1.1.4 Enter 5678, and then click GO. The following screen will be displayed.

3.1.1.5 Click Exit, and the following screen will be displayed.

-8-
3.1.1.6 Click Yes, and the following screen will be displayed.

3.1.1.7 Click Shutdown, and the following screen will be displayed.

3.1.1.8 Æ Go to Section 3.1.3 to continue operations.

-9-
3.1.2 HIPAA Mode ON

3.1.2.1 Boot the CS-1. After booting, the following Login screen will be displayed.
Enter “5246” and “5678” in the Login and Password boxes, respectively, and click Enter.

3.1.2.2 When login is successful, the following screen will be displayed. Note, however, that if
an “Exam List” is selected for the screen format, a different screen will be displayed
but similar operations to those described hereafter can be conducted.

Version of V1.10R01 or earlier Version of V1.10R01 or later

- 10 -
3.1.2.3 Click Konica REGIUS, and the following screen will be displayed (V1.10R01
or earlier).
Click KONICA MINOLTA, and the following screen will be displayed
¥ (V1.10R01 or later).

3.1.2.4 Click Shutdown to display the following screen.

3.1.2.5 Æ Go to Section 3.1.3 to continue operations.

- 11 -
3.1.3 Click YES to display the following screen.

3.1.4 Click Service Tool, and the following screen will be displayed.

3.1.5 Enter 5678, and click Go. The following screen will be displayed.

- 12 -
3.1.6 Click File Import, and the following screen will be displayed.

3.1.7 Click”…”button and select Desktop. When “CD” is selected, the following screen will
be displayed.*See 1. Preparation *2.

3.1.8 Select PcProg2.00R4T2.lzh, and the following screen will be displayed.

- 13 -
3.1.9 Click OK. The file for the version upgrade will be copied from the directory “CD” to the
hard disk, and the execution status is displayed in a dialog box. After the file is copied,
the following screen will be displayed

3.1.10 Click Upgrade, and the following screen will be displayed.

* The display may be different depending on the status of the version upgrades made in the past.
- 14 -
3.1.11 Choose Ver2.00R4T2,and click OK.
The file for the version upgrade will be copied into the program folder.
After the file is copied, the following screen will be displayed.

3.1.12 Click Exit, and the following screen will be displayed.

3.1.13 Click Yes, and the following screen will be displayed.

- 15 -
3.1.14 Click Windows Desktop. The CS-1 program will stop and return to the
Windows Desktop screen.

- 16 -
3.2 Updating the definition file/registry
3.2.1 Using Windows Wxplorer, refer to the Desktop and open the folder where the installer
was copied.(Example: “CD”).

3.2.2 Open VersionUP_CS-1 folder.


Select the folder of the current version.

- 17 -
■ In the case of selecting V1.00R00-V1.10R03 folder

3.2.3 Execute the following files by clicking on the icon depending on the existing CS-1
program installed.

V1.00R00_ModifyConfig.bat
……… Choose this file to upgrade from V1.00R00E

V1.00R03-R04T3_ModifyConfig.bat
……… Choose this file for V1.00R03E (V1.00R03T3) to V1.00R04T3.

V1.00R04T4-R04T5_ModifyConfig.bat
……… Choose this file for V1.00R04T4E to V1.00R04E (V1.00R04T5)

V1.10R00_ModifyConfig.bat
……… Choose this file for V1.10R00T1E.

V1.10R02_ModifyConfig.bat
……… Choose this file for V1.10R02T1E

V1.10R03_ModifyConfig.bat
……… Choose this file for V1.10R03

3.2.4 A command prompt will appear and updating of the definition file/registry is started.
Screen and message that may be displayed depend on the existing CS-1program installed.
-Updating the registry
Click Yes Æ Click OK

- 18 -
-On the command prompt, “Hit Any Key or Press any key to continue… “
will be displyaed.
Click the Enter key.

-“Item Not-defined” warning dialog


Click OK.

*This is a notification that a tag not used in the installed version has been detected,
which can be neglected.

When this is completed, go to 3.2.5.

■ For selection of V2.00R00-V2.00R01folder,


Execute Versionup.bat.
When “Press any key to continue…” is displayed, press any key to end.
When this is completed, go to 3.2.5.

3.2.5 Reboot the CS-1 by clicking the[Start] menu – [Shutdown] – [Reboot].


3.2 Completes updating early part of definition file/registry
Also end of 3.4 later part of definition file/registry need to be updated. Do not fail
to update at that time.
Note, however, that if you have executed the update of the definition file/registry two
or more times, be sure to execute the procedure described in “Important” below
before rebooting the CS-1.

Important: When you update the definition file/registry two or more times, the display LUT
will be incorrect. If you have done this, restore the display LUT from the backup
in “Backing up the Display LUT” in Section 3.0.1 and update the definition
file/registry again.

- 19 -
3.3 Canceling printer no-connection provisional setting
When no printer is connected, the CS-1 program V1.00R03T2 or earlier images are totally
blacked out on the Operation screen. Therefore, we have provisionally set the printer
connection to “Registered” to avoid this problem. Since measures have implemented for
handling the problem in V1.00R03T3 and later, you are requested to cancel the provisional
measures and enable the measures implemented in V1.00R03T3 and later

3.3.1 Canceling the provisional setting in V1.00R03T2. Service tool → Console → System
Info → System(Host,Printer). Select Pseudo Registered for Printer 1, and Save&Exit.

- 20 -
3.3.2 Service Tool→Console→Host. Select the host to be registered, set all items related to
printer registration(Sync.) to ON(Only Triming should be set OFF as it is pending),
and click Save&Exit.

*Triming function is pending.

- 21 -
3.4 Version upgrade of process parameters/exam tags
[Outline]
This section describes correction for the parameter of frequency processing related to QC,
the setting of fixed exposure field for mammogram and etc., 24x30 mammography, and
added output size designation for 15x30 plates.
*15x30 plates are planed to be launched after December 2004.
*If start operating Dmax3.6 in mammogram for the first time from this version. Use not
“Mammogram 3.6“ exam tags, but ”Mammogram” exam tags (Also use “Mammogram” for
Image process parameter.)

[Applicable Versions]
Table 1 provides an outline of corrections and versions that should be upgraded accordingly

Table 1

Part, Processing Outlined Measures Applicable Versions


Mammogram Change in positioning of mammary All versions of V1.00R04T4 or earlier
DMAX3.0 (horizontal flip).
Mammogram All versions of V1.10R00 or earlier
Change in rotation information (for image processing)
DMAX3.0,3.6
Mammogram All versions of V1.10R04T5 or earlier
Addition of parameters for radiograpy
radiography
FIX Change in parameters for image processing method All versions of V1.00R04T4 or earlier
QC Change in parameters for image processing for QC All versions of V1.00R04T4 or earlier
LINEAC Change in numbers for image processing parameters All versions of V1.00R04T4 or earlier
STICHING Addition of parameters for stitching entire leg radiograpy All versions of V1.00R04T4 or earlier
2D correction Addition of parameters for 2D correction All versions of V1.10R00 or earlier
Expiration Change in mismatch of set name All versions of V1.00R04T4 or earlier
Overall Head Accuracy improvement in recognition processing of All versions of V1.10R00 or earlier
divided pattern
Shoulder Joint Eliminated mismatch in the user tool on new shoulder V1.10R01T1
joints
QC Change in parameters of frequency processing All versions of V1.10R02 or earlier
Mammo,FIX,TEST Correction to be able to use fixed exposure area in All versions of V1.10R02 or earlier
different shape
Pneumoconiosis H Processing OFF( Only Maker Master) All versions of V1.10R02 or earlier
All parts Addition of output size designation of 24x30 Mammo, All versions of V1.10R02 or earlier
15x30

Hereinafter, the method for upgrading the version will be described.

[Outline of version upgrade procedures]


Version upgrade of image processing parameter/exam tags should be performed to both
REGIUS Master and User Master
However, Part of REGIUS Master and USER Master can be modified with script. Correct
referring to the following procedures. If manual correction is needed in User Master, the
method of correction will be described later. Perform according to need.

- 22 -
3.4.1[Backing up the existing system]
Procedures differ depending on how the HIPAA function is set. When the HIPAA mode is off,
setup should be performed according to procedures stated in Sections 3.4.1.1.1 to 3.4.1.1.5.
When the HIPAA mode is on, follow the procedures in Sections 3.5.1.2.1 to 3.5.1.2.8.

3.4.1.1 HIPAA Mode Off


3.4.1.1.1 Boot the CS-1. After booting, the following screen will be displayed. Note, however,
that a different screen will be displayed when an “Exam List” is selected for
the screen format but similar operations to those described here can be
conducted.

3.4.1.1.2 Click KONICA MINOLTA, and the following screen will be displayed.

- 23 -
3.4.1.1.3 Click Utility, and the following screen will be displayed.

3.4.1.1.4 Enter 5678 and click GO. The following screen will be displayed.

3.4.1.1.5 Click Backup.


Æ Go to Section 3.4.1.3 to continue operations.

- 24 -
3.4.1.2 HIPAA Mode On
3.4.1.2.1 Boot the CS-1. After booting, the following login screen will be displayed.
Enter “5246” and “5678” in the Login and Password boxes, respectively, and click
Enter.

3.4.1.2.2 When login is successful, the following screen will be displayed. Note, however, that if
an “Exam List” is selected for the screen format, a different screen will be displayed,
similar operations to those described here can be conducted.

- 25 -
3.4.1.2.3 Click KONICA MINOLTA, and the following screen will be displayed.

3.4.1.2.4 Click Shutdown, and the following screen will be displayed.

3.4.1.2.5 Click YES, and the following screen will be displayed.

- 26 -
3.4.1.2.6 Click Service Tool, and the following screen will be displayed.

3.4.1.2.7 Enter 5678 and click Go.

3.4.1.2.8 Click Backup.


Æ Go to Section 3.4.1.3 to continue operations.

- 27 -
3.4.1.3 Choose System on the following screen. Then, choose User Exam Tag and Process
Parameter under the backup tab and save them.

3.4.1.4 When the save is completed, select Exit in the following screen.

3.4.1.5 Select Exit again in the following screen.

- 28 -
3.4.1.6 Click Exit, and the following screen will be displayed.
HIPPA Mode ON HIPPA Mode OFF

3.4.1.7 Click Shutdown, and the following screen will be displayed.

3.4.1.8 Click YES, and the following screen will be displayed. Click Windows Desktop to
display the desktop.

- 29 -
3.4.2 [Upgrading version of the REGIUS Master, and updating definition file (later)]
First, make sure the desktop screen is displayed. Then perform the following procedures.
Execute the following file to upgrade the REGIUS Master version and update definition
file (later).

1) Refer to the following files by using windows explorer.

2) Open the folder for the existing version.

3)Double-click on ParameterUpdate.bat.

Note1:In case of responding wrongly to the window prompt, go to (7) ‘Please wait
for about 5 seconds.’ Message will be displayed. Then double click on
‘PUscriptE.vbs. to go back to procedure (4)-(6). After completing (6), press
ENTER key to go to (7).
Note2:Be sure to backup the script before executing it. Once the upgrade has been
completed, the script cannot be restored.

4) Confirm the version upgrade on the initial screen.


■Few questions will be asked in the followings to proceed the processing in
Mammogram appropriately. If not using Mammogram, select “YES” for all the

- 30 -
questions below to end the script.
- After choosing Yes, start your work at Step (5) below. (Selecting ‘Yes’ changes the
parameter of QC back to normal value)
- However, if you do not wan change the parameter setting because of clients request or
other reason, choose No to terminate the process.

5) Check the current version, and then confirm that the upgrade is only for newly added
parameters or it will include previous updates.
-If the current version is V1.10R00 or later, choose Yes. Only updated part of image
processing parameter after V1.10R00 will be corrected. Start your work at Step (7)
below.
-If the current version is V1.10R00 or earlier, choose No, and start your work at Step
(6) below.

6) Confirm whether the Exam tags of the Mammogram Dmax3.6 are used or not.
- When the Exam tags for the Mammogram Dmax3.6 are used, be sure to choose Yes.
The Exam keys and imaging parameters for the Dmax3.6 will not be modified, and
modifications of other Exam keys will be initiated.
Thereafter, continue your work at Step (7).
- If the Exam tags for the Mammogram Dmax3.6 are not used, choose No.
- When Cancel is chosen, those Exam tags will not be updated.
(Updating of other Exam tags including up to QC and LINEAC are not cancelled.)

- 31 -
To avoid incorrect settings, confirm whether the image processing parameters for the
Mammogram Dmax3.6 is updated or not.
- When the Exam tags for the Mammogram Dmax3.6 are not used, be sure to choose
Yes. The image processing parameters for the Mammogram Dmax3.6 will be updated.
(The work procedures become redundant when the upgrade is conduced from
V.1.00R04T5, but presents no problem.)
- When Exam tags for the Mammogram Dmax3.6 are used, choose No.

If No is chosen, the following message will be displayed, and the addition of Exam tags
for the Mammogram Dmax3.6, as well as updating of the image processing parameters
will be cancelled. Click OK.

7) After a while, a command prompt screen will appear. When a message(‘Please wait for
about 5 seconds’) is displayed, press Enter.
* It takes a few minutes for the message above to be displayed. This is not an error.

8)“Parameter Convert Completed” will be displayed.


Press Enter to end.

9) Reboot the CS-1, selecting [Start] menu – [Shut down] – [Reboot].

- 32 -
3.4.3 [Version upgrade for the User Master]

■Take the following procedures only when using the target body parts.
3.4.3.1 Mammogram DMAX3.6
Change positioning of mammary (horizontal flip). Table 2 shows Exam tags to be changed.

Table 2

Classification Group Exp. Cond. REGIUS MASTER Code


MAMMOGRAM CC (R) Cr-Ca #101010802100610000
MAMMOGRAM CC (R) Ca-Cr #101010802100620000
MAMMOGRAM CC (L) Cr-Ca #101010803200610000
MAMMOGRAM CC (L) Ca-Cr #101010803200620000
MAMMOGRAM Mag. View Mag. View #101010804030000000
MAMMOGRAM Mag CC-R Cr-Ca #101010805130610000
MAMMOGRAM Mag CC-R Ca-Cr #101010805130620000
MAMMOGRAM Mag CC-L Ca-Cr #101010806230610000
MAMMOGRAM Mag CC-L Cr-Ca #101010806230620000
MAMMOGRAM MLO (R) MLO #101010808100670000
MAMMOGRAM MLO (R) LMO #101010808100680000
MAMMOGRAM MLO (L) MLO #101010809200670000
MAMMOGRAM MLO (L) LMO #101010809200680000
MAMMOGRAM Mag MLO-R MLO #101010811130670000
MAMMOGRAM Mag MLO-R LMO #101010811130680000
MAMMOGRAM Mag MLO-L MLO #101010812230670000
MAMMOGRAM Mag MLO-L LMO #101010812230680000
MAMMOGRAM ML (R) LMO #101010814100630000
MAMMOGRAM ML (L) LMO #101010815200630000

According to need, version upgrade (Change) the User Master.


The version upgrade of the User Master allows changes in positioning of mamma
(horizontal flip). If the changes have been done, or there is no problems on the current
position(in the case that Mammpgrapy is not used), this procedure is not needed.
If version upgrading the User Master, create new exam tags, referring to 3.4.3.5 Detailed
Addition Procedures.

- 33 -
3.4.3.2 Mammary ductgrapy
This section describes the method for adding exam tags for contrast study. Table 3 shows the
exam tags to be added.

Table 3
Classification Group Exp. condition Serial No. (# + 18 digits)
Mammogram Ducto. CC #101010816000610000
Mammogram 3.6 Ducto. CC 3.6 #101010816000610100
Mammogram Ducto. MLO #101010817000670000
Mammogram 3.6 Ducto. MLO3.6 #101010817000670100

Add exam tags to the User Master by referring to Section 3.4.3.5. Detailed addition
procedures.
Version upgrading is not necessary when exposure exam keys for contrast study of mammary
gland radiography are not used.

3.4.3.3 LINEAC
This section describes the method for adding exam tags and image processing parameters to
the LINEAC. Table 4 shows the exam tags to be added, and Table 5 shows the image
processing parameters (which, however, have been already added by executing the
above-stated procedures in [Version upgrade for the REGIUS Master]).

Table 4
Classification Group Exp. condition Serial No. (# + 18 digits)
LINACGRAPH Linacgraph Linacgraph #101011830041000000
LINACGRAPH Linacgraph LG. Head #101011830041000100
LINACGRAPH Linacgraph LG. Chest #101011830041000200
LINACGRAPH Linacgraph LG. Pelv. #101011830041000300
LINACGRAPH Linacgraph LG. Extr. #101011830041000400

Table 5
Process Name Serial No. (# + 18 digits)
Linacgraphy #1012010000
LG. Head #1012020000
LG. Chest #1012030000
LG. Pelv. #1012040000
LG. Extremity #1012050000

Add exam tags to the User Master by referring to Section 3.4.3.5 Detailed addition procedures.
Version upgrading is not necessary when exposure exam tags related to the LINEAC are not
used.

- 34 -
3.4.3.4 Adding new exam tags to stitching imaging parameters
This section describes the method for adding new exam tags to stitching imaging parameters.
Table 6 shows the exam tags to be added, and Table 7 shows the imaging parameters (which,
however, have been already added by executing the procedures in Section 3.4.2 [Version
upgrade for the REGIUS Master]).

Table 6
Classification Group Exp. condition Serial No. (# + 18 digits)

STITCHING Venography AP #101012823043100000


STITCHING Venography AP Upper #101012823043100100
STITCHING Venography AP Middle #101012823043100200
STITCHING Venography AP Lower #101012823043100300
STITCHING Venography LAT #101012823043130000
STITCHING Venography LAT Upper #101012823043130100
STITCHING Venography LAT Middle #101012823043130200
STITCHING Venography LAT Lower #101012823043130300

Table 7
Process Name Serial No. (# + 18 digits)
Venography AP U. #1012230000
Venography AP M. #1012240000
Venography APU L. #1012250000
Venography LAT U. #1012260000
Venography LAT M. #1012270000
Venography LAT L. #1012280000

- 35 -
3.4.3.4.1Add exam tags to the User Master by referring to the procedures stated below.
Version upgrading is not necessary when the Exam tags for stitching, total lower leg
radiography, are not used

Boot the CS-1 application program.

- Choose User Master Edit on the User Tool, choose 170 for Reader (modality of extract
conditions), and then choose Stitching for Body Part.

-Add the required exam tags from the exam tags that were added anew to the REGIUS Master
based on Table 6 shown above. For a new installation, the default exam tags registered with
the User Master are six shaded tags as shown on the right-hand side.

- 36 -
-Next, place the exam tags registered with the User Master in order on the Routine screen.
Choose Screen Customization on the User Tool, and then choose Stitching.
Then, place the items added here from the REGIUS Master in order, and create the exam tags.
(Move the exam tags to the desired location by dragging and dropping them. To change an
exam tag/group name, right-click the name.)

- 37 -
- Creating set keys
According to the facility’s operating method, create the required set keys for stitching
exposure in the following ways:

Facility where connection processing is performed with the REGIUS-IM


Æ A set of only three (or two) “ * Middle” views.
Facility where connection processing is not used.
Æ A set of three “ *Upper”, “ *Middle” and “ *Lower” views, or a set of two
“ *Middle” and “ *Lower” views.

To execute a set of three views, etc., save the same exam tags in the same layout.
For the second time and thereafter, you will be prompted for the exam tag name, and
therefore, you are requested to change the name. (A yellow underline will appear beneath
the layout.)
To confirm the details of an exam tag, left-click the layout.

- 38 -
3.4.3.5 Detailed Addition Procedures
Procedures differ depending on the HIPAA Function Setting.
- HIPPA Mode OFF: Set up the following (2) to (5) referring to 3.5.1.1.1 to 3.5.1.1.2
- HIPPA Mode ON: Set up the following (1) to (5), referring to 3.5.1.2.1 to 3.5.1.2.3.

1) Click Utility.

2)Enter 1234, and click Go.

- 39 -
3) Select User Master Edit on the User Tool, Select 170 for Reader and MAMMO for body part
in exam tag.

4) Add necessary exam tags, referring to REGIUS Master or Table 2


(Select CC(R)3.6 Cr-Ca from REGIUS Master and click the green arrow in the middle to add
it.)

- 40 -
5) Arrange exam tags registered in the User Master on the Examination Check Screen.
Select Screen Customization from the User Tool, then MAMMO.

Create exam tags, and arrange items added this time from REGIUS Master.(Move the items to
the places to be arranged by left-drug. To change exam tags name/group name, right-click on
the place where you would like to change.

- 41 -
3.5 Correction of process parameter of shoulder joints
This procedure is not necessary as it is integrated into 3.4 Version upgrade of process
parameters/exam tags.

Corrections (Table 1) should be taken, referring to the following procedures.

Table 1

Process Name Serial No (#+10 digits)


Shoulder AP #1011210000
Shoulder AP #2011210000

1) Select Process on the Service tool.

2) Select Shoulder AP(Cassette), then Property.

- 42 -
Check if values are as follows on page 3/5.
[Check items]
1.Density L → No Use
2.Density H → Use

4) Check Shoulder AP(Upright)


,referring to the procedures above.

- 43 -
3.6 Input control cancellation of Latin characters
3.6.1 Versions earlier than V1.00R03E prohibit entry of Latin characters in a letter string
input to the CS-1. You must cancel the restriction with the following procedures, since
an update of the program to V1.00R04T5 will not cancel it.
The restricted items are:
Patient Name
Requesting Physician
Referring Physician
Recording Physician
Physician of Reading Study
Performing Physician
Other Patient name

3.6.2 From REGIUS service screen, Open Service Tool → Console → Layout → Patient,then
choose all the restricted items from the list and Save&Exit.

When Patient Name is selected,

3.6.3 Service Tool → Console → Layout → Additional and do the same as 3.6.2

3.6.4 Service Tool → Console → Layout → Routine and do the same as 3.6.2

- 44 -
3.7 Notes on LCD display gradient
As default display LUT has been improved in the CS-1 program V1.10R01. if the version of
the program is earlier than V1.10R00 and the default display LUT has been used, it will be
overwritten by the improved default display LUT. For reference ,the version earlier than
V1.10R00 has the following characters:

-V1.00R00 to V1.00R04
Æ Display is darker across the screen, and areas of low density are flat.
-V1.10R00
Æ Display is lighter across the screen, and display quality of the lung field in
the chest image is inferior to V1.00.

While using this version, if you want to use the default display LUT, whose version is earlier
than V1.10R00 for any reason, you have to adjust the settings with the following procedures:

3.7.1 From the windows explore, refer to Drive A (SUPER DISK) or the desktop, then open
the folder (Example “CD”) in which the installer was copied.

3.7.2 Open “Monitor Correction LUT” folder

3.7.3 Install the default display LUT of the version earlier than V1.10R00.
Execute V1.00.EXE or V1.10.EXE

− V1.00R00 to V1.00R04 compatible

− V1.10R00 compatible

3.7.4 When installation status finishes, it is completed.

- 45 -
3.8 Notes on manufacturer name to be output to HOST
As a result of changes in the corporate name due to the integration of KONICA and MINOLTA,
you have to change the manufacturer name output to HOST to “KONICA MINOLTA” from
“KONICA.”
However, by the change of the manufacturer name, normal operation may not be carried out
depending on the device connected to HOST.
After confirmation that Image sent to HOST is normal, change the manufacturer name to
“KONICA MINOLTA”.
Method for changing the manufacturer name
Change the Manufacturer box on the System Info to show KONICAMINOLTA.”

- 46 -
3.9 Notes on Study Description to be output to HOST
For CS-1 Program V1.10R02E or later, the value on Study Description (0008,1030) to be
output to HOST has been revised.

-V1.00R00E to V1.10R01E -> Exam Tag Key Name


-V1.10R02E to -> Requested Procedure Step(0032,1060)

For V1.10R02E or later, when outputting Exam Tag Key Name, rewrite the INI file as below.
- Subject INI file

C:¥Konica¥CS-1¥Env¥System¥HostStorage1.ini HOST 1
C:¥Konica¥CS-1¥Env¥System¥HostStorage2.ini HOST 2
C:¥Konica¥CS-1¥Env¥System¥HostStorage3.ini HOST 3
C:¥Konica¥CS-1¥Env¥System¥HostStorage4.ini HOST 4

- Contents to be edited
Change 0008,1030=0 to [STORAGE]0008,1030=1
Add ; to the line head of 0008,1030=L0,1,1I,0032,1060 in [STORAGE ADDDATA] and
change to ;0008,1030=L0,1,1I,0032,1060

[STORAGE]
; 0:OFF 1:ON
0008,1030=1 ; Study Description
0018,1030=1 ; Protocol Name
0028,0030=1 ; Pixel Spacing
0018,1405=1 ; S Val (Relative X-ray Exposure [ON]/[OFF]) (When this item doesn't ・・・・・ value 1)
0028,1050=2047 ; Window Center (When this item doesn'nt exist,it works as value 2047)
0028,1051=4096 ; Window Width (When this item doesn'nt exist,it works as value 4096)

[STORAGE ADDDATA]
; Additional Dataset for Storage (CR/DX)
; Please check [HOST info]-[Dataset Extension] in service tool.
;
;0002,0010 =PN,1,1,I,0010,0010 ; Patient Name
;0002,0020 =LO,1,1,I,0010,0020 ; Patient ID
0040,0253 =SH, 1,1,I,0020,0010 ; Performed Procedure Step ID Study ID
0040,0244 =DA, 1,1,I,0008,0020 ; Performed Procedure Step Start Date
0040,0245 =TM, 1,1,I,0008,0030 ; Performed Procedure Step Start Time
0040,0254 =LO, 1,1,I,0040,0007 ; Performed Procedure Step Description
0040,0260 =SQ,32,1,I,0040,0008 ; Performed Protocol Code Sequence
;0008,1030 =LO, 1,1,I,0032,1060 ; Study Description(2004/01/26)

[STORAGE MWM ADDDATA]


; Additional Dataset for MWM
; Outputting under (0040,0275) Request Attributes Sequence
; Left side = Outputting Tag Number
; Right side=VR and Tag Number in CS-1 header(I:Get form Image Header,F:Output following data)
;
; Available only in MWM
;
0040,1001 =SH, 1,1,I,0040,1001 ;
0040,0009 =SH, 1,1,I,0040,0009 ;
0040,0007 =LO, 1,1,I,0040,0007 ;
0040,0008 =SQ,32,1,I,0040,0008 ;

- 47 -
3.10 Editing and notes on image output gradient
This procedure is mandatory irrespective of the version of the existing CS-1 program.
The version upgrade to V2.00R03 and the following setting enable image output to printer to
have a better gradient. Additionally, this allows Dmax4.0 output of DRYPRO 793 and allows
to initialize gradient corresponding to light box with different luminance and environments
with different brightness.
(When version upgrading to V2.00R03, LUT has been changed to that for P Value.)

3.10.1 Open Printer INFO of Service tool.


For PRINTER1 to 3, in the case of DRYPRO 752/771/793, set Presentation
LUT=ON、Photometric=Monochrome2、Image Processing=P-Values.
(If it has been set, it is not necessary to change them)

- 48 -
3.10.2 Open HOST INFO of Service tool.
For HOST 1 to 4, set Image Processing=P-Val.Presentation(others).
( If it has been set, it is not necessary to change it.)

- 49 -
3.11 Setting up the HIPAA functions (Security function) (V1.00R04
and later)
3.11.1 Open System Info in Service Tool, and tick HIPAA Mode to enable it.
The HIPAA mode is enabled after rebooting the CS-1.

3.11.2 After booting, the following Login screen will be displayed.


Enter “5246” and “5678” in the Login and Password boxes, respectively.
Click ENTER.

By referring to the Service Manual, click Operator on User Tool, and enter the
login/passwords for all operators. At this time, be sure to designate at least one
person with Administrator rights for Group Membership. For all registered
engineers, be sure to set a login name and a password for each one.

- 50 -
4 Portable Mode (2.00R01 and later)
Refer to Portable Mode installation Manual for detailed installation procedures(When
installing program to handy terminal, designate the newest version.)

1. For new installation and version upgrade of the existing program


If there is a PDA (Portable Mode) in a facility and standardJM and backupJM have
been set, Execute PDA – CASIO -> ExecDBSettingFiles -> SetupExecDB.bat following
“3.1.2 Executing the “SetupExecDB” file” of CS1 Portable Mode Installation Manual.

For new installation


On V2.00R04 Version upgrade disk, there is the version upgrade file for portable mode
in the following location.
- Prepare “CS-1/CS-3 Program V2.00R04” and open “PDA-CASIO”.

PDA-CASIO

-Select HTProg_V100R01.lzh and create a folder named “REGIUS” on the desktop of the
CS-1/3 and copy HTProg_V1.00R01.lzh to “REGIUS” on the desktop.
-Click Service tool -> Adjustment -> PDA tool -> Upgrade and
designate ”HTProg_V100R01.lzh”in “Regius folder on the desktop.

-When the version upgrade is completed, delete “REGIUS” folder.

*1) CS-1/CS-3 program and the issue of handy terminal program may not synchronize.

- 51 -
*2) Even if Portable Mode has been set for the CS-1 on a site, when version upgrading the
CS-1 from V1.XX to V2.00R00 or later for the first time, it is necessary to execute the
“SetupExecDB” file” again. ( For details of the setting method of “SetupExecDB” , refer to
CS1 Portable Mode Installation Manual)

- 52 -
5 Appendix
5.1 List of CS-1 Program Changes
V1.00R01T1
- Periodic noise on printed images
Facility: Municipal Muroran General Hospital
- Changes in exam tags of User Tool and exam tags of Individual Property Edit Tools are not reflected.
Facility: Showa General Hospital

- MTF measurements for the Service Tool cannot be entered.

V1.00R01T2
- Defective image display Facility: Municipal Hino General Hospital
- Infinite error loops when MWM/MPPS operations/studies are initiated
simultaneously. Facility: Municipal Muroran General Hospital
- Redundant issuance of Study ID for MWM/MPPS operations
Facility: Municipal Muroran General Hospital
V1.00R01T3
- Captured images are not displayed (re-writing of JM configuration by backing up/restoring system
information). Facility: Aoyama Hospital

V1.00R01T4
- Error update failure of Reader status display Facility: Aoyama Hospital
- Acquisition failure of processing parameters of image adjustment for the simple/detailed screen of User Tool.
Facility: Municipal Muroran General Hospital
- Captured images are not displayed (provisional treatment by turning power off/on right after releasing
cassette). Facility: Sato Hospital

V1.00R01T5
- Dealing with magnetic cards at East Kobe Hospital.
Facility: Higashi Kobe Hospital
- Dealing with mammogram (detection of bonding margin position of chest-wall side plate.)

V1.00R01T6
- Change in the number of output LUTs for the Service Tool and Luminance Adjustment Tool.
- Change in bit map of the initial screen.
- Dealing with suspension of MWM/MPPS operations/studies

V1.00R02T1
- Improvement to 12-bit display image (handling pseudo contour of LINEAC images).
- Abnormal lossless JPEG of HOST output (handling 3Z’s HOST)
- Change of default screen keyboard for new retrieval screen (default for Patient ID is changed to ABC.)

V1.00R03T1
- Addition of X-ray generation linking function (via ID-680X).

V1.00R03T2
- Output position is changed improperly. Facility: Sapporo Century Hospital, etc.
- Application error for 2-image/2-frame preview display (depending on the mask size of E-processing and
F-processing) Facility: Okinawa Kaiho Hospital
- Disappearance of REGIUS170 captured images attributed to CS-1 program.
Facility: Hyogo Prefectural Nishinomiya Hospital

V1.00R03T3
- Character style error of data on Reader display (REGIUS350/550 runs away on transmission of Chinese
characters) Facility: Aoyama Hospital
- Print error due to change in exam tags Facility: Osaka Gyoumeikan Hospital

- 53 -
- Images cannot be transferred to Toshiba server (study dates are cleared.)
Facility: Chibana Clinic, etc.
- Abnormal interlocking of tube position/radiation field aperture at X-ray signal interlocking type B.
Facility: St. Johns
- Disappearance of REGIUS350/550 captured images (when X-ray generator switch is pressed while images
are being captured). Facility: First Nagoya Hospital
- Abnormal output method of User Tool Exam Tag Key Property Edit.
Facility: First Mizushima Hospital, etc.
- Abnormal marker size Facility: Fukushima Prefectural Ohno Hospital
- Private data violates DICOM standards. Facility: Fukushima Prefectural Ohno Hospital
- Images cannot be transferred to Yokogawa server (characters other than those specified for VR regulations.
- Abnormal trimming position of HOST-output images.
- Measures for application error of radiation field recognition processing procedure 5.
- Measures for application error of division pattern recognition processing.
- Strengthening of interlock with ID-680SM.
- Handling Dmax3.6 mammogram (handling Dmax3.6 image processing/printing; handling chest wall side
darkening process)
- Dealing with darkening of areas beyond radiation field.
- Dealing with no printer connection status.
- Dealing with HOST-output information of User Tool Exam Tag Key Property Edit.
- Dealing with JM backup.
- Addition of status-operation function of JM Job (deletion of a job when REGIUS170 malfunctions).
- Review of methods for printer interlock and application of image processing of HOST.
- Interlocking with REGIUS-IM (processing parameters are acquired when CS-1 is booted).
- Addition of parameters for stitching.

V1.00R04T1
- Abnormal reading sensitivity setup. Facility: Tenri Hospital
- Abnormal restoration of patient DB
- Addition of ASCII kana conversion method (similar conversion to REGIUS350 by substitution of INI file).
- Abnormal darkening of circular radiation field.
- Dealing with Latin1-ready character set.
- Dealing with HIPAA.

V1.00R04T2
- Disappearance of images when switching to backup JM.

V1.00R04T3
- Assorted information transmitted to HOST is not output, and engineer names are not output on name print.
- Abnormal image density due to incorrect SG value computation.
Facility: Kyushu Medical Center

V1.00R04T4
- Images cannot be switched even if the Main/Sub of Detailed Picture Quality Adjustment Screen of User Tool
is switched.

- Change in notations of items on System Information Tool of Service Tool (HIPAA Æ Security).

V1.00R04T5
- Distortion of REGIUS170 captured images Facility: Tenri Hospital
- Abnormal screensaver cancellation when receiving order from RIS.
Facility: Kurobe City Hospital
- Abnormal notification of de-touch completion when the number of exposures is zero (0).
Facility: International Medical Center of Japan
- No issuance of Affective SOP Instance UID for MWM/MPPS operations.
Faclility:New Hampshire
- Hatching of OK button on Engineer Name Selection Screen.
- Abnormal focus control of Engineer Name Registration Screen for HIPAA.

- 54 -
- Abnormal first year of Japanese-style year for name printing.
- Change of output mode judgment when switching to backup for Standard 75X and Backup 711.
- Abnormal image display of User Tool image adjustment simple/detailed screen.
- Improvement of DL/DH adjustments of User Tool image adjustment simple/detailed screen.
- Abnormal pointing detection on User Tool picture quality adjustment/exposure conditions screen.
- Addition of study ID automatic issuance form check of System Information Tool of Service Tool.
- Abnormal save/restore of password storage files for HIPAA of Backup/Restore Tool in Service Tool.
- Application error in thinning-out image bit conversion tool of Version Upgrade Tool for existing
installation.
- Changes in setup default values as required by KMS.
- Abnormal routine for computing histogram of front chest processing.
- Defective setting of ROI position for side thoracic vertebra processing (domain beyond human body).
- Addition of search range setting of human body for mammogram.
- Abnormal masking shape for darkening processing.
- Change of processing parameter for LINEAC frequency enhancing level to 0.70.
- Change of processing parameter FIX method to zero (0).
- Change of rotation rate of processing parameters for MAMMO (3.6) to 321.
- Addition of stitching radiography series for the processing parameter.
- Addition of AJS Dr.View/RIS interlocking function.
- Addition of remote maintenance function.
V1.10R00T1
- Addition of portable mode function. Applied to a limited number of monitor facilities.
- Addition of 2D correction function. Applied to the limited number of monitor facilities.
- Addition of interlocking function with SANYO Medicom.
- Strengthening of AJS Dr.View/RIS interlock function.
- Addition of P-value image output function.
- Strengthening of screen control functions.
- Handling of DRYPRO771.
- Addition of relative X-ray dose to HOST output.
- Addition of ASCII items of patient names on new patient retrieval screen.
- Error that exposure-associated information and study date of substituted images become blank.
- Disturbance in HOST-output images in automatic direction recognition 270 degrees and output orientation of
horizontal FIXED Facility: Kojima City Hospital
- Addition of a new abdomen processing function.
- Addition of rotation control flags for mammogram processing.
- Addition of density adjustment range restrictions for User Tool detailed image adjustment screen.
- Abnormal out-of-range reference of division pattern recognition processing.
- Error due to out-of-range reference of radiation field recognition processing.
- Abnormal shutter value displayed.
- Abnormal darkening processing for mammogram chest-wall side
Facility: Aoyama Hospital

V1.10R01T1
- Portable mode (to be applied for all versions)
- FIX processing (measures to enable switching between low and high sensitivity of REGIUS170)
Facility: Tenri Hospital
- Improvement of 2-frame output stamps with different processing.
- Handling GSPS of HOST.
- Handling P-value image output of printer.
- Embedding of reverse mark.
- Restoration processing in reader unit of User Tool, Service Tool exam keys and processing parameter.
- Handling of change of corporate name.
- Without MWM polling.
- Calibration tool for display.
- Error display “At Study in Other X-ray Room” when selection multiple body parts in a study.
Facility: Yokohama Seamen’s Hospital
- Abnormal trimming position adjustment of printer format tab of Detailed Image Screen.
- 55 -
Facility: Fujii Hospital
- Corruption of image bitmap.
- Improvement to prohibit booting of User Utility due to memory leaks, etc.
Facility: B&W Hospital
-Image processing
1. Correction following failure modification of Mammogram: 150. Corrections were made to threshold value of
mammary region, reversing of ROI coordinate, etc.
2. Error treatment for a case where mammary gland region is very small and normal treatment cannot be expected
was added.
3. Darkening processing: Defective recognition of polygons and rotational failure were corrected.
4. Radiation field recognition processing: Abnormal display of overlay tab of Detailed Image Screen. Correction of
routine that may cause discontinuance at image edges, routine that re-detects oblique radiation field sides, and
routines that may cause out-of-range reference in the detection of applicable circular radiation fields.
5. Division pattern recognition processing: Addition of a routine whose recognition accuracy has been improved
for four limb bones.
6. Processing to change ROI sizes: Processing to enlarge/reduce ROI size in vertical and horizontal directions at a
ratio based on the recognition result.
7. Shoulder joint (ipp): For shoulder joint images, parameters were corrected so that the shoulder joints can be
acknowledged stably even if the heart, chin, etc. are exposed.
8. Mammogram (ipp): Image processing parameters were corrected so that an object identifying position is
determined based on image rotation information to avoid a failure that an object position on the image is
wrongly acknowledged due to existence of a marker or radiation filed edges.
9. Division pattern recognition (ipp): Parameters of division pattern recognition processing for head were changed
to new dedicated processing parameters.
10. Child, infant (exp): Method for notating a child, infant, etc. was corrected for a unified procedure.
11. Exposure exam tag: Group name was changed from Konica to Service.

- Abnormal display of overlay tab of Detailed Image Screen


Facility: Aoyama Hospital
- Abnormal correction of capture size when a substitution image is destroyed.
Facility: Seirei Yokohama Hospital
- Improvement of default display LUT

V1.10R02T1
- Correction of failure that the patient information exposured by REGIUS350 has been replaced to the patient
information of the previous examination.
Facility: National International Iryou C
- Correction of failure that CS-1 becomes inoperative when switching over to Backup JM
- Addition of Backup and Restore processing of LUT for display
- Addition of restoration process in a lump sum of User tool/ Service tool exposure parameters
- Correction of failure in preset H processing
- Shoulder joints : Abnormal parameter in image processing
- Abnormal processing in masking Facility: Nagoya Daiichi Red Cross Hospital

V1.10R03T1
- Expansion of DICOM character set. (Dealt with 3/4 level )
- Launch cm-cassette in domestically
- Dealt with 2 kinds of magnetic cards.
-Addition of FTP Gateway function
-Addition of Mammogram sub scan uniformity correction function
-Redundant Study ID Facility: Hiroshima City Hospital
-Abnormal reading in REGIUS5350/550 Facility: Heart Life Hospital
-Image Processing: Change of algorithm
1.Darkening Process: Dealing with abnormal masking areas when ROI fixed mode is selected.
Dealing with abnormal masking areas when rotating with non-ROI fixed mode.
Improvement of accuracy
2.Front Chest: Enhancement of errors measures applicable to flat images
- 56 -
3.Head: Accuracy improvement of removal of exposure areas without object.
4.G4 Change exposure area recognition method from none to method 5
5.C-spine AP,LAT Addition of ROI size change process
6.Exposure area recognition processing :Improvement to avoid extreme process failure in method 3 and 5.
Addition of fixed function of exposure areas in method 5.
-Image Processing: Changes of parameters
1.Deletion of TEST Calibration 350/550 from the display set.
2.Output size is set for 24x30 in Image size for all the exam tags.
3. Exam tags for pneumoconiosis has been changed from HF processing to F processing.
4. Addition of 15x30 Image size to all the exam tags.
5.Change of exposure recognition area method in Mammo, Mammo Mag. View, Fix and G4 process parameter
6.Correction unused outlying parameters of QC processing.

V2.00R00T2

- Improvement in HOST printer interlocking and related function


- Improvement in updating process of the registered patient DB of local patient DB
- Improvement in PN component group (Domestic Version)
- Improvement in PN component (Overseas Version)
- Function of 2D correction of REGIUS350/550
- DICOM communication log
- Add study name in view item of Examination List Screen.
- Extend the group name view length of exposure condition
- Banned future DOB input
*)Warning message appears twice when receiving future DOB information from IDS/RIS
-Dealing with 15x30 cm cassette
- Image processing Change of algorithm
1.Image Processing – Improvement in darkening adjustment function
2.Image processing – Accuracy improvement in removal of foreign substances of knee joints
3.Image processing – Correction of abnormal G Value of stomach flat image
-Change of image processing parameter
1. Image processing – Addition of LUT explanation

V2.00R01T1

- Addition of printer trimming interlocking function


- Dealing with REGIUS 550
- Addition of Grid deletion function

V2.00R03T1

- Addition of image share function


- Addition of annotation function
- Addition of Stitching linking function
- Addition of statistics and analysis function

V2.00R04T2

-DICOM MG Storage Function added


-Corrected malfunction of designated reader keep on sleep mode when power is on for a long time
-Corrected malfunction of wrongly display the status of registration list
-Corrected stiching malfunction (1) that left and right movement is unequual
-Corrected stiching malfunction(2) that white line is misplaced for rataion
-Corrected MPPPS tag (modified connection error with GE-made RIS, SOP instance UID is not issued, N-Set set
Character Set mistakenly)
-Corrected misrecognition as standing position when Decubitus reader device is connected to one devive
-Malfunction between REGIUS 370 and 570 erase parameter
-Corrected malfunction of displayed 7/10 on the page position off exposure information screen

- 57 -
-Additional function for stamping (Capitalize fist letter)

- 58 -

You might also like