RX Gnatus Manual
RX Gnatus Manual
RX Gnatus Manual
INSTRUCTIONS OF USE
EQUIPMENT:
Technical name: Odontological X -Ray
Trade name: Timex 70 E - X-Ray
Brand: GNATUS
Trade models:
• Timex 70 E X-Ray Mobile Column
• Timex 70 E X-Ray Wall
• Timex 70 E X-Ray Pantographic Mobile Column
• Timex 70 E X-Ray Pantographic Wall
• Timex 70 E X-Ray Pantographic Fixed Column
Manufacturer/ Distribuitor:
GNATUS - EQUIPAMENTOS MÉDICO-ODONTOLÓGICOS LTDA.
Rod. Abrão Assed , Km 53+450m - Cx. Postal 782 CEP 14097-500
Ribeirão Preto - S.P. - Brasil
Fone +55 (16) 2102-5000 - Fax +55 (16) 2102-5001
C.N.P.J. 48.015.119/0001-64 - Insc. Est. 582.329.957.115
www.gnatus.com.br - [email protected]
Technical Duties: Gilberto Henrique Canesin Nomelini
CREA-SP: 0600891412
ATTENTION
For greater safety:
• Read and understand all the instructions contained in these
Instructions for Use before installing or operating this Equipment.
Note: These Instructions for Use must be read by all the operators
of this Equipment.
2
INDEX
PRESENTATION OF MANUAL ............................................................................ 02
IDENTIFICATION OF EQUIPMENT..................................................................... 05
-Dear Customer ................................................................................................. 05
-Identification .................................................................................................... 05
-Principles and bases applied to the functioning of the product.................................. 06
-Description of Equipment ................................................................................... 06
-Indication of Equipment ..................................................................................... 06
INSTALLATION OF EQUIPMENT
-Dimensions and positioning Chassis command box X-Ray Wall Model......................... 28
OPERATION OF EQUIPMENT
-Instructions of transmitter assembly operation (how to move the cylinder head) ........ 30
-Operating instructions......................................................................................... 30
-Operating Instructions “Smaller colimator rectangular kit and extension with
colimator cone kit”.............................................................................................. 31
-Limitation and indication of bunch extension of x radiation ..................................... 31
-Precaution to be observed before 1st application of load ......................................... 31
-Bunch radiation characteristics ............................................................................ 32
-Radiographic techniques letters ........................................................................... 33
-Procedure for development with recent chemical reagent ........................................ 33
-Radiographic testing .......................................................................................... 34
3
INDEX
-Precommendations for preserving the equipment ................................................... 36
-Precautions and warnings “during the use” of equipment ........................................ 36
-Precautions and warnings “after” the use of equipment .......................................... 37
-Precautions and warnings during the “cleaning and disinfection” of equipment ........... 37
-Precautions in case of alteration in the functioning of equipment ............................. 37
-Precautions to be adopted against foreseeable or uncommon risks, related to the
deactivation and abandoning of equipment ............................................................ 37
4
Description of the Equipment
Dear Customer
Congratulations. You have made a good choice when you decided to buy a GNATUS
QUALITY product comparable to the best products available in the World. This manual is a
general presentation of your product and it will give you important details to help you to
solve possible problems.
Please, read it and keep this with you.
Identification
Technical name: Odontological X -Ray
Trade name: Timex 70 E - X-Ray
Trade models:
• Timex 70 E X-Ray Mobile Column
• Timex 70 E X-Ray Wall
• Timex 70 E X-Ray Pantographic Mobile Column
• Timex 70 E X-Ray Pantographic Wall
• Timex 70 E X-Ray Pantographic Fixed Column
Brand: GNATUS
5
Description of the Equipment
Principles and bases applied to the functioning of the product
Photomultipliers set, which are generated in X-rays tubes that transform light energy
into electric energy, and then a circuit set forms the image, in order to diagnose or to guide
evasive medical procedures and dental treatments.
Description of Equipment
This device is for intraoral radiography of patient’s dentition with the purpose of making
a diagnostic.
It is a unit of X- Ray for odontological use, with nominal tension of 70kVp and current
in the tube of
7,0 mA. It has a centesimal digital timer, particularly developed for using with digital
radiographic sensors, it reduces the time of radiation exposure and also it is indicated for
conventional radiographic films.
Fixed column with the option of mobile column with 4-caster base; super steady columns,
painted in epoxy at 250º C, and option of base for mounting on the wall built in steel, painted
in epoxy and covered by a cover in polystyrene of high impact.
It is composed of type-pantographic arm (when pantographic model), which allows
better reach and use in many positions.
Tube (ampoule), with focal point of 0,8 x 0,8mm, filtration with aluminum equivalence
of 3,22 mm, cylindrical guide made in radiopaque polymer in order to avoid secondary
radiations, wound completely immersed in special oil.
Manual trigger at a distance of 5m. Duly tested by legal body, respecting the rules in effect
of radiological protection and elaborated by National Nuclear Energy Commission - CNEN.
The test method for measuring tension parameters of mean peak in the x- ray tube
(kVp), mean current in the x-ray tube (mA), application time of load on x-ray tube(s) and
product current X time in the x-ray time (mAs), adopted is the following:
It is used for measuring the device Dynalyzer III digital display, connected to a high
tension unit. This unit consists of a resistive tension divisor of 1:20.000.
For odontological x-ray equipment it is used an adapter system in order to make a
connection of the device Dynalyzer III. “In order to check the results obtained, it is connected
to high tension unit The digital storage oscilloscope 2230 produced by Tektronix which gives
the possibility to store an electrical signal to which is subject the x-ray tube, thus allowing
an inspection of mean peak in the x-ray tube and the application time of load. Such results
can be compared with those obtained from Dynalyzer III”.
ISO 9001/2000 and ISO 13485/2003 Quality System, assuryng the products are
manufactured under standart procedures.
Products manufactured in agreement with RDC-59 - ANVISA (Sanitary Surveillance
National Agency).
Indication of Equipment
This equipment is for dental use use only. It must be operated and utilized by specialized
professional (certified professional, according to the legislation of the country) and following
the instructions of the manual. The operation of the equipment required, for the professional,
the utilization of correct instruments and it should to be in perfect conditions of the use, and
to protect the professional, the patients and others, in the eventual danger situation.
6
MODULES, ACCESSORIES, OPTIONS AND MATE-
RIALS OF CONSUMPTION
14 15
03
4A 06
03
02
01
08
05 03
02 07 5A
06 04
07
02
05
08
01 06
07
01
09
08 09
11
10
10
12
13
7
MODULES, ACCESSORIES, OPTIONS AND MATE-
RIALS OF CONSUMPTION
14 15
4B 3A 3A 4A 06
01
02 08
05
02 07 5A
06
07 01
09
08
11
12
10
8
TECHNICAL SPECIFICATIONS
Technical features
General
Model
• Timex 70 E X-Ray Mobile Column
• Timex 70 E X-Ray Wall
• Timex 70 E X-Ray Pantographic Mobile Column
• Timex 70 E X-Ray Pantographic Wall
• Timex 70 E X-Ray Pantographic Fixed Column
9
TECHNICAL SPECIFICATIONS
Technical features
General
Nominal Focus Point Value
(0,8 x 0,8mm) positioned in relation to the reference axis of the XR tube in compliance
with IEC336/1982)
Tube Conditioning
At iddle position the Cone should always be directed downwards
kVp tube (average)
70 kVp (for max. tube current 7.0 mA)
kVp (average) beam peak
70 kVp +/- 10%
Net weight Gross weight
Timex 70 E X-Ray Mobile Column
50 kg 64,5 kg
Net weight Gross weight
Timex 70 E X-Ray Wall
25 kg 29 kg
Net weight Gross weight
Timex 70 E X-Ray Pantographic Mobile Column
61,5 kg 75,5 kg
Net weight Gross weight
Timex 70 E X-Ray Pantographic Wall
31 kg 34,5 kg
Net weight Gross weight
Timex 70 E X-Ray Pantographic Fixed Column
35,6 kg 41 kg
Power Supply
Nominal Supply Type
Model Tension Rate Freq Consumption
Installation
127V~ 125V~ + 4% 60Hz 10A Mobile
220V~ 220V~ + 4% 60Hz 6A Mobile
Timex 70 E
Mobile Column 220V~ 220V~ + 4% 50Hz 6A Mobile
230V~ 230V~ + 4% 50Hz 5,5A Mobile
240V~ 240V~ + 4% 50Hz 5A Mobile
10
TECHNICAL SPECIFICATIONS
Technical features
Power Supply
Nominal Supply Type
Model Tension Rate Freq Consumption
Installation
127V~ 125V~ + 4% 60Hz 10A Fixed
220V~ 220V~ + 4% 60Hz 6A Fixed
Timex 70 E Wall 220V~ 220V~ + 4% 50Hz 6A Fixed
230V~ 230V~ + 4% 50Hz 5,5A Fixed
240V~ 240V~ + 4% 50Hz 5A Fixed
11
TECHNICAL SPECIFICATIONS
Technical features
Cooling - Sb
The device presents protection against excessive heating of the tube.
A new image can be performed only after a time equivalent of 30 times the trigger time,
necessary time for natural cooling and return of normal functions.
Ex: Trigger time = 0,06 seconds.
Cooling time = 0,06seconds X 30 = 1,8 seconds.
Thermal protection
The device has a safety mechanism against temperature elevation of the transmitter
assembly.
A thermal protection turn the device’s functions off if the internal temperature of the
transmitter assembly overpasses the allowed limit , damaging the internal components of
the assembly.
Overvoltage Protection
The device has a safety mechanism for overvoltage in device power supply that does
not allow trigger when power supply overpasses approximately 3 Volts of specified limits.
Undervoltage Protection
The device has a safety mechanism for undervoltage in device power supply that does
not allow trigger when power supply it goes inferior the 10 V~ of specified limits.
12
TECHNICAL SPECIFICATIONS
Technical features
Trigger
The button must be continuously pressed until the end of transmission, monitored by
a sound signal (beep), that is, if the button activation is interrupted, the transmission will
also be automatically interrupted , by indicating at display “A4”.
13
TECHNICAL SPECIFICATIONS
Radiation quality
Ingress of quality Parametes for charge exposure
equivalence 0,32s
FILTERS Tube crystal Tension
1,26 mm Al 70 kVp ± 10%
Insulating oil, column 17mm Current
0,64 mm Al 7 mA ± 1,5 mA
Window Power
0,32 mm Al 0,49 kW
Additional filter (aluminum) Product current time
1,00 mm Al 2,24 mAs
TOTAL FILTRATION = 3,22 mm Al Highest energy applied in intermittent
mode during 1 h
170 KJ
Note:
Filtration with thickness equivalency of aluminum.
- Using as method an x-ray tube tension of 70 kVp and 2.5 mmAl of semiconducting
reductor.
-The method using the measurement of the first semireductor layer was in narrow beam
conditions, with the x-ray equipment operating with x-ray tube tension value of 70 kVp with
anodic current of 7mA and exposure time of 1.0 second.
14
TECHNICAL SPECIFICATIONS
IMPORTANT: Information referring to x-ray dosage, radiation leak can be found in the
conformity report attached to this manual.
NOTE:
- The serial number of the dome, beam limiter and radiation source is the same of the
equipment, and it is stated in the product’s label.
Transmitter assembly
06 05 04
03 02 01
10 09 08 07
Standards applied:
This product was tested and approved as per the standards:
IEC 60601-1: (1988); IEC 60601-2-7: (1998); IEC 61000-4-3: (2006);
Amendment 1: (1993); IEC 60601-2-28: (1993); IEC 61000-4-4: (2004);
Amendment 2: (1995); IEC 60601-2-32: (1994); IEC 61000-4-5: (2005);
IEC 60601-1-2: (2001); CISPR 11, edição 3.1 (1999); IEC 61000-4-6: (2006);
IEC 60601-1-3: (1994); IEC 61000-4-2: (2001); IEC 61000-4-11: (2004);
NBR-IEC série 601-1 Equipamento Eletromédico - Parte 1: Prescrições gerais para
segurança;
EN 980:2003 (Ed. 2) - Graphical symbols for use in the labelling of medical devices;
ISO 14971: 2007
ISO 9687: 1993 - Dental equipment - graphical symbols;
ISO 7494 - 1:2004 - Dental Units - Port 1: General requeriments and test methods
ISO 13485:2003
ISO 780:1997
ISO 11144:1995
15
TECHNICAL SPECIFICATIONS
Dimensions (mm)
Model: Mobile Column
16
TECHNICAL SPECIFICATIONS
Dimensions (mm)
Model: Wall
17
TECHNICAL SPECIFICATIONS
Dimensions (mm)
Model: Pantographic Mobile Column
18
TECHNICAL SPECIFICATIONS
Dimensions (mm)
Model: Pantographic Fixed Column
19
TECHNICAL SPECIFICATIONS
Dimensions (mm)
Model: Pantographic Wall
20
TECHNICAL SPECIFICATIONS
Packing symbols
Maximum stacking: Packing to be transported and
It determines the maximum / or stored avoiding humidity,
quantity of boxes which can be rains and wet floor.
stacked during transportation
and storage “as per packaging”.
Product symbols
Careful: It indicates an important Turned on
instruction for the operation of the
product. Not following it can cause
dangerous malfunctioning.
Turned off
21
TECHNICAL SPECIFICATIONS
Product symbols
01
02
22
TECHNICAL SPECIFICATIONS
Content of accessible and non-accessible demarcations
03
04
06
05
23
TECHNICAL SPECIFICATIONS
Content of accessible and non-accessible demarcations
07 11
08
09 10
24
TECHNICAL SPECIFICATIONS
Content of accessible and non-accessible demarcations
Description Fixation local
01 - Product identification tag Cabinet cover
02 - Product identification tag Cylinder head
03 - Ades. tag. Pantographic Arm Cabinet cover
04 - Ades. tag. Adjustable arm Cabinet cover
05 - Ades tag. radiation Cylinder head
06 - Ades tag. CE 0499 Packing
07 - Ades tag. identification Tension Cylinder head Cylinder head
08 - Ades tag. electro medic safety seal Packing and Cylinder head
09 - Product identification tag X-Ray control
10 - Ades tag. - warning Arm articulations and cabinet
cover
11 - Ades tag. – tape removal Arm
25
TECHNICAL SPECIFICATIONS
Tube letters with technical characteristics
26
TECHNICAL SPECIFICATIONS
Tube letters with technical characteristics
27
INSTALLATION OF EQUIPMENT
Dimensions and positioning
Chassis command box X-Ray Wall Model
Recommendations for pre-installation preparation of X-Ray
Wall model:
In order to assure the perfect functioning of wall model x-ray,
we recommend that the services of pre-installation be performed
by professionals duly qualified.
Any damages that may occur to the device, caused by fault
at pre-installation will not be covered by warranty.
“For more information, refer to pre-installation manual”.
Electrical
passage box
Distance
from floor
to first hole
28
INSTALLATION OF EQUIPMENT
The installation of this equipment requires specialized technical
assistance (Gnatus).
- This equipment shall only be able to be unpacked and installed by a Gnatus authorized
technician, under penalty of losing the warranty, as only (s)he has the information, suitable
tools and training required to execute this task.
- Gnatus bears no responsibility for damages or accidents caused by poor installation
executed by a technician not authorized by Gnatus.
- Only after the equipment has been installed and duly tested by the authorized technician
representing Gnatus, will it be ready to start work operations.
29
OPERATION OF EQUIPMENT
Operating Instructions of Transmitter Assembly How to move
the cylinder head
- When moving the cylinder head, grasp the cylinder (02) and back part (01) as shown
in the picture.
Operating Instructions
Only after an authorized technician have properly installed
and tested the equipment, it will be ready to work.
Verify if the equipment is connected to the main grid.
Turn on the general switch (09) and a green light will 01
turn on.
Choose the time scale from 0 to 3.2 seconds according
to the work to be performed. “See radiographic technique
chart on page 33” and set it through the buttons (04 and
05) located on the control.
When the patient is properly prepared for the 02
radiography:
Remove the support control, press the shutter-button 03
(06) (Beep beeping) and keep it pressed.
The “yellow” LED must be lighted during radiation 04
emission. 06
Note: Turn off the general key (09) whenever the 05
equipment is not in use. 07
Always keep the remote control in its holder.
08
Note: If the messages appear on the display:
• A1 •A2
• A3 •A4 09
•Sb •A5
NOTE:
The value indicated on the digital command display refers to the exposure time. To
obtain the charge application time, add 0.2 seconds (pre-warming) to any value indicated
on the display.
- As it is a pulse device and depends on the wiring frequency, it is not possible to provide
time of 0,063s belonging to geometric series within the range of ± 10%.
( IEC 60601-2-7 : 29.1.106 E)
30
OPERATION OF EQUIPMENT
Operating Instructions “Smaller colimator rectangular kit and
extension with colimator cone kit”
By applying a slight effort, press the cover (01) against the adapter (02) observing the
alignment of openings (A) with catches (B) “click system”.
After assembling cover (01) and adaptor (02), attach the assembly mounted in the
complementary cone (03), afterwards attach to the cylinder head (04).
The attachment between the couplings shall be performed with rotation movements,
and the alignment between locking cams shall be observed (C).
02 02 C
B B
C C
C C
01 01
A A
WARNING
The operator shall keep a bigger distance between the distal extreme of limiting bunch
device and patient’s surface.
31
OPERATION OF EQUIPMENT
Radiation bunch characteristics
The pictures below present radiation bunch characteristics with and without the use of
extension whit colimador cone.
The extension whit colimador cone is used to apply the parallelism technique which
gives less distortion to the image generated due to x-ray from film/sensor presented less
inclination related to reference axis.
NOTE: If the extension whit colimador cone is not used, it does not affect the
safety level to patient during use of this device.
32
OPERATION OF EQUIPMENT
Radiography technical letter
Standard table for exposure time with type E film in adult individuals using the bisector
periapical technique.
DEP - Skin penetration dose (mGy)
DEP - 3,5 mGy - Standard level established by ordinance 453 (D.O.U. 103 02/06/98) for
a periapical exame of typical adult patient and film of group E. (Extaspeed/Agfa M-4)
Film D (Ultra-speed/Agfa M-2) double the time;
Children, con-sider 2/3 of time;
Angle with the vertical.
Superior Jaw
Incisor +40º 0,32 - 0,40 1,1 - 1,5 0,06 0,10
Canine +45º 0,40 - 0,50 1,5 - 2,0 0,08 0,16
Pre-molar +30º 0,40 - 0,50 1,5 - 2,0 0,08 0,16
Molar +20º 0,50 - 0,64 2,0 - 2,6 0,10 0,24
Lower Jaw
Incisor -15º 0,32 - 0,40 1,1 - 1,5 0,06 0,10
Canine -20º 0,32 - 0,40 1,1 - 1,5 0,06 0,10
Pre-molar -10º 0,32 - 0,40 1,1 - 1,5 0,06 0,10
Molar -5º 0,40 - 0,50 1,5 - 2,0 0,08 0,16
Bitewing
Anterior +8º 0,32 - 0,40 1,1 - 1,5 0,06 0,10
Posterior +8º 0,32 - 0,40 1,1 - 1,5 0,06 0,10
Oclusal
Upper Jaw +60º 0,50 - 0,63 2,0 - 2,6 --- ---
Lower Jaw -70º 0,40 - 0,50 1,5 - 2,0
--- ---
33
OPERATION OF EQUIPMENT
Radiography test
After the assembly of the equipment the authorized
technician must carry out a radiography test, using
the materials (film, dark room, processor and fixator) 23
supplied by the client, following the instructions
below:
24
Place the head directed downwards (90o position)
next to an horizontal surface (aprox. 50 mm).
Centralize the film (25) utilizing for such the
auxilliary points (24) existing on the Collimator Cylinder 25
(23), put over the film (25), a metallic object (26). (Ex.:
clips or coin) as in the figure below:
26
Select a time scale of 0.32 seconds.
Activate the button “Start” (06) (see pg. 30).
NOTE:
The material (clips, coins, film, etc) will not be
contaminated, and can be taken out after trigger.
Develop the film at Dark Room with the following
times:
- 1 minute at development;
- 1 minute at fixation.
Afterwards, check if radiography is according to indicated
standard.
In case of differences in comparison, see page 40
Radiography carried out with “fault”.
• Kodak Film. If it is not possible to test by the authorized technician
• Ektaspeed Type (because of lack of material) the proprietary will be
It is not provided with the product responsible for it before initiating operations.
Caution:
Never test the device with the cylinder aimed to you, or people that are next to the place
Always do the test with the cylinder aimed downwards, by keeping a minimum distance
of 2m.
34
PRECAUTIONS, RESTRICTIONS AND WARNINGS
Transportation and storage
This equipment must be transported and stored observing the following directions:
- Avoid falls and impacts;
- Keep it dry, do not expose it to rain, water drops or wet floor;
- Keep it away from water and direct sunlight, and in it original wrapping;
- Don’t move it over irregular surfaces, protect it from rain and observe the maximum
stack quantity specified in the packaging;
- Transportation and storage temperature range: -12°C to 50°C;
- Transportation and storage relative humidity range: 0°C to 90°C;
- Atmospheric pressure range: 500hPa to 1060hPa (375 mmHg to 795 mmHg).
35
PRECAUTIONS, RESTRICTIONS AND WARNINGS
These information also make part of the Manual of Installation and
Maintenance of the equipment that can be found with the authorized
Gnatus technician.
36
PRECAUTIONS, RESTRICTIONS AND WARNINGS
37
CORRECTIVE AND PREVENTIVE MAINTENANCE
AND PRESERVATION
Additional procedures for reuse
The equipment can be reused in undetermined, i.e. unlimited, quantities, only needing
to be cleaned and disinfected.
Disinfection
Use clean and soft cloth dampened in alcohol 70% to disinfection of the equipment.
Never use corrosive disinfectants or solvents.
Cleaning
The cleaning procedure below should be executed at the start of the
working day and after each patient.
Always turn off the main switch before executing the procedures of
daily maintenance.
WARNING:
- In order to prevent risks and damages to equipment, make
sure that the liquid does not enter into the unit.
- The application of other solvent-based cleaning products or
sodium hypochloride isn’t recommended, because they may damage
the equipment.
38
CORRECTIVE AND PREVENTIVE MAINTENANCE
AND PRESERVATION
Preventive Maintenance
The equipment should be calibrated routinely, as per the legislation in force in the
country.
But never with a period exceeding 3 years.
In order to protect your equipment, seek Gnatus technical assistance for periodic revisions
of preventive maintenance.
Corrective Maintenance
If the equipment has any abnormality, check if the problem is related to any of the items
listed in the item Unforeseen Events (situation, cause and solution).
If it is not possible to solve the problem, turn off the equipment, and request Gnatus
technical assistance.
39
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS
Upon coming across any problem in operation, follow the instructions below to
check and repair the problem, and/or get in touch with your representative.
-At radiography appe- -Wrong positioning of cylin- -Radiograph using the parallelism
ars a semi circle. der. technique, using for that auxiliary
lines of collimator cylinder.
-Sb -Protection against excessi- -Wait for the correct cooling time to
ve tube heating. reestablish the normal functions –
see page 12 – cooling item.
40
WARRANTY OF EQUIPMENT
This equipment is covered by the warranty terms counting from the date of installation,
as specified below; provided that the defect has occurred in normal conditions of use and
that the equipment has not remained stored for more than 06 months counting from the
issue date of the sales document until the date of the actual installation.
- WARRANTY TERMS: Verify the guarantee certificate;
- LOSS OF THE WARRANTY:
A) Attempt to repair using an inadequate tool or by unauthorized technicians;
B) Installation of the equipment by an unauthorized technician;
C) Damage arising from inappropriate storage or signs of infringement;
D) Incorrect use of the equipment;
E) Use of a cleaning product not indicated by the factory;
F) Falls or blows which the equipment may undergo or lack of observation of an compliance
with the guidelines of the Owner’s Manual, which was delivered with the present document,
together with the equipment. Repair or replacement of parts during the warranty period
shall not extend the validity term of their warranty.
- This warranty doe snot exempt the customer from paying the service charge for the visit
and the travel expenses of the technician, except when the customer sends the equipment
to execute the maintenance inside the establishment of the technical assistance.
“Consumer Defense Code - art. 50, unique paragraph”.
- The Warranty Certificate comes with the product and must be filled in upon the date
of installation by the Gnatus Authorized Technician.
- Queries and information: GNATUS Help Desk (+55) 16 2102-5000.
- Check the warranty term attached to this manual.
FINAL CONSIDERATIONS
The most important aspect related to equipment care is that concerning spare parts.
To guarantee the life span of your equipment, use only original Gnatus spare parts.
They are sure to follow the technical specifications and standards required by Gnatus.
We must also point out to you our chain of authorized dealers. Only dealers that make
part of this chain will be able to keep your equipment constantly new for they count on
technical assistants who have been trained and on spedific tools for the correct maintenance
of your equipment.
Doubts and information: GNATUS Call center (55-16) 2102-5000.
41
42
43