MiSeq® System

User Guide

FOR RESEARCH USE ONLY

ILLUMINA PROPRIETARY
Part # 15027617 Rev. O
Catalog # SY-410-9001DOC
September 2014
This document and its contents are proprietary to Illumina, Inc. and its affiliates ("Illumina"), and are intended solely for the contractual use of its
customer in connection with the use of the product(s) described herein and for no other purpose. This document and its contents shall not be used or
distributed for any other purpose and/or otherwise communicated, disclosed, or reproduced in any way whatsoever without the prior written
consent of Illumina. Illumina does not convey any license under its patent, trademark, copyright, or common-law rights nor similar rights of any
third parties by this document.

The Software is licensed to you under the terms and conditions of the Illumina Sequencing Software License Agreement in a separate document. If
you do not agree to the terms and conditions therein, Illumina does not license the Software to you, and you should not use or install the Software

The instructions in this document must be strictly and explicitly followed by qualified and properly trained personnel in order to ensure the proper
and safe use of the product(s) described herein. All of the contents of this document must be fully read and understood prior to using such product
(s).

FAILURE TO COMPLETELY READ AND EXPLICITLY FOLLOW ALL OF THE INSTRUCTIONS CONTAINED HEREIN MAY RESULT IN
DAMAGE TO THE PRODUCT(S), INJURY TO PERSONS, INCLUDING TO USERS OR OTHERS, AND DAMAGE TO OTHER PROPERTY.

ILLUMINA DOES NOT ASSUME ANY LIABILITY ARISING OUT OF THE IMPROPER USE OF THE PRODUCT(S) DESCRIBED HEREIN
(INCLUDING PARTS THEREOF OR SOFTWARE) OR ANY USE OF SUCH PRODUCT(S) OUTSIDE THE SCOPE OF THE EXPRESS WRITTEN
LICENSES OR PERMISSIONS GRANTED BY ILLUMINA IN CONNECTION WITH CUSTOMER'S ACQUISITION OF SUCH PRODUCT(S).

FOR RESEARCH USE ONLY

© 2014 Illumina, Inc. All rights reserved.

Illumina, 24sure, BaseSpace, BeadArray, BlueFish, BlueFuse, BlueGnome, cBot, CSPro, CytoChip, DesignStudio, Epicentre, GAIIx, Genetic
Energy, Genome Analyzer, GenomeStudio, GoldenGate, HiScan, HiSeq, HiSeq X, Infinium, iScan, iSelect, ForenSeq, MiSeq, MiSeqDx,
MiSeq FGx, NeoPrep, Nextera, NextBio, NextSeq, Powered by Illumina, SeqMonitor, SureMDA, TruGenome, TruSeq, TruSight, Understand
Your Genome, UYG, VeraCode, verifi, VeriSeq, the pumpkin orange color, and the streaming bases design are trademarks of Illumina, Inc. and/or
its affiliate(s) in the U.S. and/or other countries. All other names, logos, and other trademarks are the property of their respective owners.

This software contains the SeqAn Library, which is licensed to Illumina and distributed under the following license:

Copyright © 2010, Knut Reinert, FU Berlin, All rights reserved. Redistribution and use in source and binary forms, with or without modification,
are permitted provided that the following conditions are met:

1 Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.

2 Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the
documentation and/or other materials provided with the distribution.

3 Neither the name of the FU Berlin or Knut Reinert nor the names of its contributors may be used to endorse or promote products derived
from this software without specific prior written permission.

THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS "AS IS" AND ANY EXPRESS OR IMPLIED
WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE COPYRIGHT HOLDER OR CONTRIBUTORS BE LIABLE FOR ANY
DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO,
PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER
CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR
OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.

Read Before Using this Product


This Product, and its use and disposition, is subject to the following terms and conditions. If Purchaser does not agree to these
terms and conditions then Purchaser is not authorized by Illumina to use this Product and Purchaser must not use this Product.

ii Part # 15027617 Rev. O


1 Definitions. "Application Specific IP" means Illumina owned or controlled intellectual property rights that pertain to
this Product (and use thereof) only with regard to specific field(s) or specific application(s). Application Specific IP
excludes all Illumina owned or controlled intellectual property that cover aspects or features of this Product (or use
thereof) that are common to this Product in all possible applications and all possible fields of use (the "Core IP").
Application Specific IP and Core IP are separate, non-overlapping, subsets of all Illumina owned or controlled intellectual
property. By way of non-limiting example, Illumina intellectual property rights for specific diagnostic methods, for
specific forensic methods, or for specific nucleic acid biomarkers, sequences, or combinations of biomarkers or
sequences are examples of Application Specific IP. "Consumable(s)" means Illumina branded reagents and consumable
items that are intended by Illumina for use with, and are to be consumed through the use of, Hardware.
"Documentation" means Illumina's user manual for this Product, including without limitation, package inserts, and any
other documentation that accompany this Product or that are referenced by the Product or in the packaging for the Product
in effect on the date of shipment from Illumina. Documentation includes this document. "Hardware" means Illumina
branded instruments, accessories or peripherals. "Illumina" means Illumina, Inc. or an Illumina affiliate, as applicable.
"Product" means the product that this document accompanies (e.g., Hardware, Consumables, or Software). "Purchaser"
is the person or entity that rightfully and legally acquires this Product from Illumina or an Illumina authorized dealer.
"Software" means Illumina branded software (e.g., Hardware operating software, data analysis software). All Software is
licensed and not sold and may be subject to additional terms found in the Software's end user license agreement.
"Specifications" means Illumina's written specifications for this Product in effect on the date that the Product ships from
Illumina.
2 Research Use Only Rights. Subject to these terms and conditions and unless otherwise agreed upon in writing by an
officer of Illumina, Purchaser is granted only a non-exclusive, non-transferable, personal, non-sublicensable right under
Illumina's Core IP, in existence on the date that this Product ships from Illumina, solely to use this Product in Purchaser's
facility for Purchaser's internal research purposes (which includes research services provided to third parties) and solely
in accordance with this Product's Documentation, but specifically excluding any use that (a) would require rights or a
license from Illumina to Application Specific IP, (b) is a re-use of a previously used Consumable, (c) is the disassembling,
reverse-engineering, reverse-compiling, or reverse-assembling of this Product, (d) is the separation, extraction, or
isolation of components of this Product or other unauthorized analysis of this Product, (e) gains access to or determines
the methods of operation of this Product, (f) is the use of non-Illumina reagent/consumables with Illumina's Hardware
(does not apply if the Specifications or Documentation state otherwise), or (g) is the transfer to a third-party of, or sub-
licensing of, Software or any third-party software. All Software, whether provided separately, installed on, or embedded
in a Product, is licensed to Purchaser and not sold. Except as expressly stated in this Section, no right or license under
any of Illumina's intellectual property rights is or are granted expressly, by implication, or by estoppel.
Purchaser is solely responsible for determining whether Purchaser has all intellectual property rights that are
necessary for Purchaser's intended uses of this Product, including without limitation, any rights from third
parties or rights to Application Specific IP. Illumina makes no guarantee or warranty that purchaser's specific
intended uses will not infringe the intellectual property rights of a third party or Application Specific IP.
3 Regulatory. This Product has not been approved, cleared, or licensed by the United States Food and Drug
Administration or any other regulatory entity whether foreign or domestic for any specific intended use, whether
research, commercial, diagnostic, or otherwise. This Product is labeled For Research Use Only. Purchaser must ensure it
has any regulatory approvals that are necessary for Purchaser's intended uses of this Product.
4 Unauthorized Uses. Purchaser agrees: (a) to use each Consumable only one time, and (b) to use only Illumina
consumables/reagents with Illumina Hardware. The limitations in (a)-(b) do not apply if the Documentation or
Specifications for this Product state otherwise. Purchaser agrees not to, nor authorize any third party to, engage in any of
the following activities: (i) disassemble, reverse-engineer, reverse-compile, or reverse-assemble the Product, (ii) separate,
extract, or isolate components of this Product or subject this Product or components thereof to any analysis not expressly
authorized in this Product's Documentation, (iii) gain access to or attempt to determine the methods of operation of this

MiSeq System User Guide


iii
Product, or (iv) transfer to a third-party, or grant a sublicense, to any Software or any third-party software. Purchaser
further agrees that the contents of and methods of operation of this Product are proprietary to Illumina and this Product
contains or embodies trade secrets of Illumina. The conditions and restrictions found in these terms and conditions are
bargained for conditions of sale and therefore control the sale of and use of this Product by Purchaser.
5 Limited Liability. TO THE EXTENT PERMITTED BY LAW, IN NO EVENT SHALL ILLUMINA OR ITS
SUPPLIERS BE LIABLE TO PURCHASER OR ANY THIRD PARTY FOR COSTS OF PROCUREMENT OF
SUBSTITUTE PRODUCTS OR SERVICES, LOST PROFITS, DATA OR BUSINESS, OR FOR ANY INDIRECT,
SPECIAL, INCIDENTAL, EXEMPLARY, CONSEQUENTIAL, OR PUNITIVE DAMAGES OF ANY KIND ARISING
OUT OF OR IN CONNECTION WITH, WITHOUT LIMITATION, THE SALE OF THIS PRODUCT, ITS USE,
ILLUMINA'S PERFORMANCE HEREUNDER OR ANY OF THESE TERMS AND CONDITIONS, HOWEVER
ARISING OR CAUSED AND ON ANY THEORY OF LIABILITY (WHETHER IN CONTRACT, TORT
(INCLUDING NEGLIGENCE), STRICT LIABILITY OR OTHERWISE).
6 ILLUMINA'S TOTAL AND CUMULATIVE LIABILITY TO PURCHASER OR ANY THIRD PARTY ARISING OUT
OF OR IN CONNECTION WITH THESE TERMS AND CONDITIONS, INCLUDING WITHOUT LIMITATION,
THIS PRODUCT (INCLUDING USE THEREOF) AND ILLUMINA'S PERFORMANCE HEREUNDER, WHETHER
IN CONTRACT, TORT (INCLUDING NEGLIGENCE), STRICT LIABILITY OR OTHERWISE, SHALL IN NO
EVENT EXCEED THE AMOUNT PAID TO ILLUMINA FOR THIS PRODUCT.
7 Limitations on Illumina Provided Warranties. TO THE EXTENT PERMITTED BY LAW AND SUBJECT TO THE
EXPRESS PRODUCT WARRANTY MADE HEREIN ILLUMINA MAKES NO (AND EXPRESSLY DISCLAIMS
ALL) WARRANTIES, EXPRESS, IMPLIED OR STATUTORY, WITH RESPECT TO THIS PRODUCT,
INCLUDING WITHOUT LIMITATION, ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS FOR A
PARTICULAR PURPOSE, NONINFRINGEMENT, OR ARISING FROM COURSE OF PERFORMANCE,
DEALING, USAGE OR TRADE. WITHOUT LIMITING THE GENERALITY OF THE FOREGOING, ILLUMINA
MAKES NO CLAIM, REPRESENTATION, OR WARRANTY OF ANY KIND AS TO THE UTILITY OF THIS
PRODUCT FOR PURCHASER'S INTENDED USES.
8 Product Warranty. All warranties are personal to the Purchaser and may not be transferred or assigned to a third-party,
including an affiliate of Purchaser. All warranties are facility specific and do not transfer if the Product is moved to
another facility of Purchaser, unless Illumina conducts such move.
a Warranty for Consumables. Illumina warrants that Consumables, other than custom Consumables, will conform to
their Specifications until the later of (i) 3 months from the date of shipment from Illumina, and (ii) any expiration
date or the end of the shelf-life pre-printed on such Consumable by Illumina, but in no event later than 12 months
from the date of shipment. With respect to custom Consumables (i.e., Consumables made to specifications or
designs made by Purchaser or provided to Illumina by, or on behalf of, Purchaser), Illumina only warrants that the
custom Consumables will be made and tested in accordance with Illumina's standard manufacturing and quality
control processes. Illumina makes no warranty that custom Consumables will work as intended by Purchaser or for
Purchaser's intended uses.
b Warranty for Hardware. Illumina warrants that Hardware, other than Upgraded Components, will conform to its
Specifications for a period of 12 months after its shipment date from Illumina unless the Hardware includes Illumina
provided installation in which case the warranty period begins on the date of installation or 30 days after the date it
was delivered, whichever occurs first ("Base Hardware Warranty"). "Upgraded Components" means Illumina
provided components, modifications, or enhancements to Hardware that was previously acquired by Purchaser.
Illumina warrants that Upgraded Components will conform to their Specifications for a period of 90 days from the
date the Upgraded Components are installed. Upgraded Components do not extend the warranty for the Hardware
unless the upgrade was conducted by Illumina at Illumina's facilities in which case the upgraded Hardware shipped
to Purchaser comes with a Base Hardware Warranty.

iv Part # 15027617 Rev. O


c Exclusions from Warranty Coverage. The foregoing warranties do not apply to the extent a non-conformance is
due to (i) abuse, misuse, neglect, negligence, accident, improper storage, or use contrary to the Documentation or
Specifications, (ii) improper handling, installation, maintenance, or repair (other than if performed by Illumina's
personnel), (iii) unauthorized alterations, (iv) Force Majeure events, or (v) use with a third party's good not provided
by Illumina (unless the Product's Documentation or Specifications expressly state such third party's good is for use
with the Product).
d Procedure for Warranty Coverage. In order to be eligible for repair or replacement under this warranty Purchaser
must (i) promptly contact Illumina's support department to report the non-conformance, (ii) cooperate with Illumina
in confirming or diagnosing the non-conformance, and (iii) return this Product, transportation charges prepaid to
Illumina following Illumina's instructions or, if agreed by Illumina and Purchaser, grant Illumina's authorized repair
personnel access to this Product in order to confirm the non-conformance and make repairs.
e Sole Remedy under Warranty. Illumina will, at its option, repair or replace non-conforming Product that it
confirms is covered by this warranty. Repaired or replaced Consumables come with a 30-day warranty. Hardware
may be repaired or replaced with functionally equivalent, reconditioned, or new Hardware or components (if only a
component of Hardware is non-conforming). If the Hardware is replaced in its entirety, the warranty period for the
replacement is 90 days from the date of shipment or the remaining period on the original Hardware warranty,
whichever is shorter. If only a component is being repaired or replaced, the warranty period for such component is
90 days from the date of shipment or the remaining period on the original Hardware warranty, whichever ends later.
The preceding states Purchaser's sole remedy and Illumina's sole obligations under the warranty provided
hereunder.
f Third-Party Goods and Warranty. Illumina has no warranty obligations with respect to any goods originating
from a third party and supplied to Purchaser hereunder. Third-party goods are those that are labeled or branded
with a third-party's name. The warranty for third-party goods, if any, is provided by the original manufacturer.
Upon written request Illumina will attempt to pass through any such warranty to Purchaser.
9 Indemnification.
a Infringement Indemnification by Illumina. Subject to these terms and conditions, including without limitation,
the Exclusions to Illumina's Indemnification Obligations (Section 9(b) below), the Conditions to Indemnification
Obligations (Section 9(d) below), Illumina shall (i) defend, indemnify and hold harmless Purchaser against any
third-party claim or action alleging that this Product when used for research use purposes, in accordance with these
terms and conditions, and in accordance with this Product's Documentation and Specifications infringes the valid
and enforceable intellectual property rights of a third party, and (ii) pay all settlements entered into, and all final
judgments and costs (including reasonable attorneys' fees) awarded against Purchaser in connection with such
infringement claim. If this Product or any part thereof, becomes, or in Illumina's opinion may become, the subject of
an infringement claim, Illumina shall have the right, at its option, to (A) procure for Purchaser the right to continue
using this Product, (B) modify or replace this Product with a substantially equivalent non-infringing substitute, or
(C) require the return of this Product and terminate the rights, license, and any other permissions provided to
Purchaser with respect this Product and refund to Purchaser the depreciated value (as shown in Purchaser's official
records) of the returned Product at the time of such return; provided that, no refund will be given for used-up or
expired Consumables. This Section states the entire liability of Illumina for any infringement of third party
intellectual property rights.
b Exclusions to Illumina Indemnification Obligations. Illumina has no obligation to defend, indemnify or hold
harmless Purchaser for any Illumina Infringement Claim to the extent such infringement arises from: (i) the use of
this Product in any manner or for any purpose outside the scope of research use purposes, (ii) the use of this Product
in any manner not in accordance with its Specifications, its Documentation, the rights expressly granted to Purchaser
hereunder, or any breach by Purchaser of these terms and conditions, (iii) the use of this Product in combination
with any other products, materials, or services not supplied by Illumina, (iv) the use of this Product to perform any
assay or other process not supplied by Illumina, or (v) Illumina's compliance with specifications or instructions for
this Product furnished by, or on behalf of, Purchaser (each of (i) – (v), is referred to as an "Excluded Claim").

MiSeq System User Guide


v
c Indemnification by Purchaser. Purchaser shall defend, indemnify and hold harmless Illumina, its affiliates, their
non-affiliate collaborators and development partners that contributed to the development of this Product, and their
respective officers, directors, representatives and employees against any claims, liabilities, damages, fines, penalties,
causes of action, and losses of any and every kind, including without limitation, personal injury or death claims, and
infringement of a third party's intellectual property rights, resulting from, relating to, or arising out of (i) Purchaser's
breach of any of these terms and conditions, (ii) Purchaser's use of this Product outside of the scope of research use
purposes, (iii) any use of this Product not in accordance with this Product's Specifications or Documentation, or (iv)
any Excluded Claim.
d Conditions to Indemnification Obligations. The parties' indemnification obligations are conditioned upon the
party seeking indemnification (i) promptly notifying the other party in writing of such claim or action, (ii) giving the
other party exclusive control and authority over the defense and settlement of such claim or action, (iii) not admitting
infringement of any intellectual property right without prior written consent of the other party, (iv) not entering into
any settlement or compromise of any such claim or action without the other party's prior written consent, and (v)
providing reasonable assistance to the other party in the defense of the claim or action; provided that, the party
reimburses the indemnified party for its reasonable out-of-pocket expenses incurred in providing such assistance.
e Third-Party Goods and Indemnification. Illumina has no indemnification obligations with respect to any goods
originating from a third party and supplied to Purchaser. Third-party goods are those that are labeled or branded
with a third-party's name. Purchaser's indemnification rights, if any, with respect to third party goods shall be
pursuant to the original manufacturer's or licensor's indemnity. Upon written request Illumina will attempt to pass
through such indemnity, if any, to Purchaser.

vi Part # 15027617 Rev. O


Part # Revision Date Description of Change

15027617 D July 2012 Updated software descriptions to MCS v2.0.


Added the following new information:
• Added a section titled What's New in MCS to describe new
software features, interface changes, and workflow changes
• Added catalog number and description of the MiSeq Reagent Kit
v2, 500 Cycles
• Added Version Compatibility and Requirements section
• Added description of MiSeq Expansion Pack, which is required for
14-tile dual-surface flow cell imaging
• Added description of dual-surface flow cell tile numbering
• Added the PCR Amplicon analysis workflow for Nextera XT
libraries
• Added the use of 10% Tween 20 in wash procedures and expected
wash volumes
• Added the reagent cartridge version to the RFID read failure
procedure
Updated the following information:
• Changed reagent acronyms for IMF, CMF, and AMX to v2
reagent names IMS, CMS, and AMS, respectively
• Changed the PhiX concentration from 8 pM to 12.5 pM
• Changed the maximum recommended NaOH concentration to 1
mM in final solution
• Noted that a maintenance wash is required to remove the
instrument from standby mode and begin the setup steps for a
subsequent run
• Removed Sample Sheet Parameters section and sample sheet
setup step in the workflow; Illumina recommends creating the
sample sheet prior to sample preparation (See the MiSeq Sample
Sheet Quick Reference Guide, part # 15028392 and the Illumina
Experiment Manager User Guide, part # 15031335)

x Part # 15027617 Rev. O


Revision History
Part # Revision Date Description of Change

15027617 C April 2012 Updated software descriptions to MCS v1.2


Added the following new procedures and sections: BaseSpace
overview, Using Custom Primers, Generating FASTQ files,
Troubleshooting Flow Rate Error, Performing a Volume Test,
Performing a Maintenance Wash, and Idling the Instrument, which
includes a standby wash.
Updated the following information:
• Updated name of Amplicon workflow to Custom Amplicon;
updated name of DenovoAssembly workflow to Assembly;
added GenerateFASTQ workflow
• Added descriptions of run folders and files; updated run folder
naming; added output file size
• Listed genome folder as required for amplicon sequencing in
Sample Sheet Parameters
• Added instructions for diluting NaOH to denature libraries
• Updated Resolving RFID Read Failure to include MiSeq Self-
Service instructions
• Listed files and folders used for troubleshooting run performance
15027617 B December Updated software descriptions to MCS v1.1
2011 Added information about anti-virus protection
Updated the following information:
• Instructions to resolve RFID failure
• Preparing libraries—Changed to 0.2 N NaOH
• Run folder naming convention
• Required disk space and storage capacity
• Run setup steps—Added more information to Setting Up the
Sample Sheet
• Run setup steps—Added note to dispose of remaining PR2
• Analysis duration—Added when analysis exceeds two hours
• Analysis Input Requirements—Listed manifest files as required
for TruSeq Custom Amplicon libraries
• Corrected HT1 tube size in MiSeq Reagent Kit Contents
• Changed iCom references to MyIllumina
15027617 A September Initial release
2011

MiSeq System User Guide


xi
Chapter 4 Troubleshooting 73
Introduction 74
Bundle Logs for Troubleshooting 75
Software Settings 77
Pause or Stop a Run 81
Resolve Run Setup Errors 83
Resolve RFID Read Failure 85
Troubleshoot Flow Rate Error 87
Perform a Volume Test 88
Measure Expected Wash Volumes 91

Index 93
Technical Assistance 97

xiv Part # 15027617 Rev. O


Getting Started
Introduction
The Illumina MiSeq® system combines proven sequencing by synthesis (SBS) technology
with a revolutionary workflow that enables you to go from DNA to analyzed data in as
little as 8 hours. The MiSeq integrates cluster generation, sequencing, and data analysis on
a single instrument.

Features
} Walk away automation—After setting up your run, which includes loading the pre-
filled reagent cartridge, buffer bottle, and flow cell, no additional hands-on time is
required.
} Pre-filled reagent cartridge—A specially designed single-use pre-filled reagent
cartridge provides reagents for cluster generation and sequencing, including paired-end
sequencing reagents and indexing reagents. Integrated radio-frequency identification
(RFID) tracking enables accurate consumable tracking.
} Interface controls—The MiSeq Control Software (MCS) interface provides controls to
configure the instrument, set up and monitor runs, and perform maintenance
procedures.
} Convenient flow cell loading—A clamping mechanism auto-positions the flow cell as
it is loaded onto the instrument. Integrated radio-frequency identification (RFID)
tracking enables accurate consumable tracking.
} Innovative fluidics architecture—The MiSeq fluidics system enables unmatched
efficiency in chemistry cycle time during sequencing.
} Real-time analysis (RTA)—Integrated primary analysis software performs real-time on-
instrument data analysis during the sequencing run, which includes image analysis
and base calling, and saves valuable downstream analysis time.
} MiSeq Reporter—Integrated secondary analysis software processes data from primary
analysis to perform alignment and provide information about each sample analyzed.

Additional Resources
The following documentation is available for download from the Illumina website.

2 Part # 15027617 Rev. O


MiSeq Reagent Kit Overview
MiSeq Reagent Kit Overview
To perform a run on the MiSeq, you need one single-use MiSeq Reagent Kit, which is
available in different types and sizes. Each type of MiSeq Reagent Kit includes a kit-specific
flow cell type and all reagents required for performing one run.
The flow cell, PR2 bottle, and reagent cartridge provided in the kit use radio-frequency
identification (RFID) for accurate consumable tracking and compatibility.
Always use the reagent cartridge associated with your flow cell type. If the reagent cartridge
is not compatible, a message appears during run setup that prompts you to load a
compatible reagent cartridge.
For a description of available reagent kits, visit the Illumina website at
support.illumina.com/sequencing/sequencing_kits/miseq_reagent_kit.ilmn.
For information about kit contents and requirements, see the reagent prep documentation
for the kit you are using.

MiSeq System User Guide


9
Run Folders
Files
The files that are copied to the output and analysis folders include the following:
} SampleSheet.csv—Provides parameters for the run and subsequent analysis. At the
start of the run, the sample sheet is copied to the root folder and renamed
SampleSheet.csv. Copies are written to Data\Intensities and
Data\Intensities\BaseCalls.
} runParameters.xml—Contains a summary of run parameters and information about
run components, such as the RFID of the flow cell and reagents associated with the
run.
} RunInfo.xml—Contains high-level run information, such as the number of reads and
cycles in the sequencing run, and whether a read is indexed.

Folders
The folders that are copied to the output and analysis folders include the following folders
generated during the sequencing run:
} <Run folder name>\Config—Contains configuration files for the run.
} <Run folder name>\Data—Contains subfolders Intensities, BaseCalls, and Alignment.
Data generated from MiSeq Reporter are located in the Alignment subfolder.
} <Run folder name>\Data\RTA Logs—Contains log files that describe each step
performed by RTA for each Read.
} <Run folder name>\Data\Intensities\BaseCalls—Contains subfolders with base call
(*.bcl) files, matrix files, and phasing files. MiSeq Reporter writes FASTQ files to this
folder during secondary analysis. For more information, see the MiSeq Reporter User
Guide (part # 15042295).
} <Run folder name>\Recipe—Contains the recipe used for the run.
} <Run folder name>\Logs—Contains log files that describe every step performed by the
instrument for each cycle.
} <Run folder name>\InterOp—Contains binary files used by Sequencing Analysis
Viewer (SAV) to summarize various primary analysis metrics such as cluster density,
intensities, quality scores, and overall run quality.
All other files and folders created in the temporary run folder are not copied to the output
and analysis folders. They contain temporary files that are not required for analysis or
troubleshooting.

MiSeq System User Guide


29
Performing a Run
NOTE
As the flow cell latch is closed, two alignment pins near the hinge of the flow cell latch
properly align and position the flow cell. An audible click indicates that the flow cell latch
is secure.

Figure 19 Close Flow Cell Latch

5 Check the lower-left corner of the screen to confirm that the flow cell RFID was
successfully read.
NOTE
If the RFID cannot be read, the software prompts you through the steps to obtain a
temporary bypass code and proceed with setting up the run. For more information, see
Resolve RFID Read Failure on page 85.
6 Close the flow cell compartment door.
7 Select Next on the Load Flow Cell screen. The Load Reagents screen opens.

44 Part # 15027617 Rev. O


Performing a Run
6 Slowly lower the sipper handle. Make sure that the sippers lower into the PR2 and
waste bottles.
Figure 21 Lower Sipper Handle

7 Check the lower-left corner of the screen to confirm that the RFID of the PR2 bottle was
read successfully.
NOTE
If the RFID cannot be read, the software prompts you through the steps to obtain a
temporary bypass code and proceed with setting up the run. For more information, see
Resolve RFID Read Failure on page 85.
8 Select Next on the Load Reagents screen.

Load the Reagent Cartridge


NOTE
Do not leave the reagent chiller door open for extended periods of time.
1 Open the reagent chiller door.
2 Hold the reagent cartridge on the end with the Illumina label, and slide the reagent
cartridge into the reagent chiller until the cartridge stops.

46 Part # 15027617 Rev. O


Loading Reagents
Figure 22 Load Reagent Cartridge

3 Close the reagent chiller door.


4 Check the lower-left corner of the screen to confirm that the RFID of the reagent
cartridge was read successfully.
NOTE
If the RFID cannot be read, the software prompts you through the steps to obtain a
temporary bypass code and proceed with setting up the run. For more information, see
Resolve RFID Read Failure on page 85.
If the reagent cartridge is not compatible with the flow cell, a message appears. Select
Back to load a compatible cartridge, or select Exit to return to the Welcome screen.
5 Close the reagent compartment door.
6 Select Next on the Load Reagents screen. The Review screen opens.

Change Sample Sheet


Every run must have a sample sheet. By default, the software looks for a sample sheet file
with a name matching the barcode number of the reagent cartridge loaded on the
instrument. If a sample sheet is not found, a message appears that prompts you to browse
to the location of the correct sample sheet for your run.
To prevent the software from searching unsuccessfully, use the Change Sample Sheet
command on the Load Reagents screen to direct the software to the appropriate sample
sheet.

MiSeq System User Guide


47
Resolve RFID Read Failure
Resolve RFID Read Failure
If the system cannot read the RFID of a consumable, you can obtain a temporary bypass
code from the Illumina website. A temporary bypass code expires in seven days.
1 Always select Retry before proceeding. If the RFID failed a second time, select Get
Code.
2 From a computer with internet access, go to my.illumina.com and log in to your
MyIllumina account.
3 From the MyIllumina page, click Account. In the Resources column, click MiSeq Self-
Service.
4 On the MiSeq Self-Service page, enter the MiSeq serial number.
5 From the Type of Override Code drop-down list, select RFID Override.
Figure 33 MiSeq Self-Service Page

6 To generate the code, select Get Code.


7 Return to the MCS interface and select Enter Code.
8 Enter the temporary bypass code using the on-screen keyboard, and then select Next.
9 Enter the barcode number of the flow cell, PR2 bottle, or reagent cartridge.

MiSeq System User Guide


85
idling the instrument 68 reagents
Index

initialization 84 kitted 9
instrument setup 23 loading 45
InterOp folder 29 real-time analysis 2
IP address 77 results 53
run folder 28
L template generation 51
laboratory-grade water guidelines 10 rebooting software 23
Live Help 80 recipes, managing 18
loading reagents reference genome
cartridge 46 file format 13
PR2 45 RFID
PR2 45
M reagent cartridge 46
tracking 2
maintenance wash 58, 65 troubleshooting 85
manifest file RTAcomplete.txt 53
copying to instrument 18 run duration 24
definition 13 run folder
MiSeq Reporter 2 definition 13
credentials 78 run folders
overview 26 contents 28
MiSeq Self-Service 85 managing 18
monitoring the run 50 naming 28
moving files and folders 18 primary analysis files 53
temp, output, analysis 28
N run options 20
network connection 84 run setup screens 17, 39
network settings 77 RunInfo.xml 29, 53
runParameters.xml 29
O S
optics module 5
sample sheet
P changing 47
copying to instrument 18
password 12 definition 13
password, changing 78 in run folder 53
pausing a run 81 not found 84
PGS 40 secondary analysis 26
post-run wash 60 sensor indicators 16
power switch 12 sequencing 35
PR2, loading 45 Sequencing Analysis Viewer 24, 50
sequencing screen 50
R shutting down the instrument 23, 72
read length 24 sipper handle 8
reagent cartridge 8 software
reagent chiller, temperature 16 disk space checking 31
reagent compartment 5, 8 initialization 12

94 Part # 15027617 Rev. O