Optibar DP 7060 Optibar DP 7060 Optibar DP 7060 Optibar DP 7060
Optibar DP 7060 Optibar DP 7060 Optibar DP 7060 Optibar DP 7060
Optibar DP 7060 Optibar DP 7060 Optibar DP 7060 Optibar DP 7060
1 Product features 3
2 Technical data 8
3 Installation 31
4 Electrical connections 39
5 Order information 47
The extremely compact measuring cell has a robust and precise response to temperature
changes and with step response times of just 125 ms, it provides enough measurements for
reliable and stable process control.
Total 3D linearisation
Highlights
• Outstanding temperature stability even under harsh conditions.
• Very good repeatability and long-term stability of the measuring signal.
• Extremely quick step response times < 125 ms.
• Combined DP, SP and T measurements for maximum process reliability.
• Measuring ranges up to 10 mbar / 0.145 psi even without electronic spreading.
• Turn down up to 100:1, higher on request.
• Universal modularity of the entire OPTIBAR process series.
• Quick start-up for all applications.
• Extensive diagnostic and parameterization functions on the display module or the user-
friendly and free DTM.
Industries
• General process technology
• Power generation
• Chemical
• Petrochemicals
• Environmental technology
• Water and wastewater
Applications
• Pressure monitoring of filters and pumps with overload protection of up to 400 bar / 5800 psi.
• Level measurement of liquids in open and pressurised vessels
• Flow measurement of gases, vapours and liquids with differential pressure transmitters.
• Measurement of density and interface of liquids in tanks.
1.2 Options
The OPTIBAR process pressure series allows free choice of pressure sensors, process
connections, electronics and housings - so that each device is perfectly adapted to each
measuring task.
The plastic housing is cost-effective and features a low net weight and high chemical resistance
in corrosive environments.
The standard housing for all pressure transmitters – it is perfectly equipped for industrial use
and can be used in hazardous areas for all protection types.
For applications that place particular demands on the mechanical robustness of the converter.
These housings can be used with all protection types for hazardous areas.
Recommended for applications requiring the corrosion resistance of stainless steel but not the
mechanical robustness of a stainless steel precision casting housing. Also suitable for hygienic
applications that require an IP69K protection class for steam jet cleanings. Converters can only
be used in hazardous areas in intrinsically safe operation.
1 Fill fluid
2 Temperature sensor
3 Absolute pressure sensor
4 Differential pressure sensor
5 Overload system
6 Separating diaphragm
The process pressure is transferred via the separating metallic diaphragms 6 of the high and
low pressure side and the fill fluid 1 to the piezoresistive silicon sensor. Through the prevailing
pressure differential, the silicon diaphragm of the differential pressure sensor 4 is deflected
and changes the resistance value of the four piezoresistive elements in the bridge circuit. The
change in resistance of the bridge circuit is proportional to the differential pressure.
Additionally, the measured cell temperature 2 and the prevailing static pressure 3 on the low
pressure side is measured and then made available to the signal converter for further
processing. If the measurement limit is exceeded, the overload system 5 restricts the prevailing
process pressure at the differential pressure sensor and protects it from damage.
Measuring system
Measuring principle Piezoresistive differential pressure cell
Application range • Flow measurement (volume or mass flow) with corresponding differential pressure
transmitter in gases, vapours and liquids
• Differential pressure measurement
• Interface and density measurement
• Level, volume and mass measurement of liquids
Measuring range 10 mbar, 30 mbar, 100 mbar, 500 mbar, 3 bar, 16 bar / 0.14 psi, 0.43 psi, 1.45 psi, 7.25 psi,
43.51 psi, 232 psi
Display and User interface
Display on signal • Dot-matrix display 45 x 27 mm / 1.77 x 1.06", illuminated
converter • Display turnable in 90° steps
• Digit size 13 x 7 mm / 0.51 x 0.27"
• Ambient temperatures below -20°C / -4°F may affect the readability of the display
Display values Number of digits 5
Digit size WxH = 7x13 mm
Display function • Display of measured value or derived measured value such as filling height, density,
interface position, volume or mass flow and total counter
• Warning and diagnostic information
• All parameters are accessible via the operating menu
Operating and display German, English, French, Spanish, Portuguese, Italian, Dutch, Russian, Turkish, Polish, Czech,
languages Japanese and Chinese
Operating Local operation via 4 push buttons on the display and adjustment module
Remote control • PACTwareTM, incl. Device Type Manager (DTM)
• HART® Hand Held Communicator from Emerson Process
• AMS® from Emerson Process
• PDM® from Siemens
Integrated clock
Date format Day / Month / Year
Time format 12 hours / 24 hours
Time zone CET (Factory setting)
Rate deviation Maximum 10.5 minutes / year
Measuring accuracy
Differential pressure
Reference conditions • Ambient temperature (constant): +18...+30°C / +64...+86°F
acc. to IEC 60770-1 • Relative humidity (constant): 45...75%
• Ambient pressure (constant): 860...1060 mbar / 12.5...15.4 psi
• Mounting position: vertical
• Rising characteristics
• Measurement start at 0.00 bar / psi
• Process diaphragm: 316L / 1.4404
• Fill fluid: silicone oil
• Material of process flanges: 316L / 1.4404
• Power supply: 24 VDC ±3 VDC
• Load for HART®: 250 Ω
Reference accuracy Includes the linearity, hysteresis and non-repeatability under reference conditions. Applies to
acc. to DIN EN 61298 the digital interfaces (HART®, Profibus PA, Foundation Fieldbus) as well as for the analogue
4…20 mA current output.
[% of the set span]
TD < 5:1 TD > 5:1 TD < 10:1 TD > 10:1
10 mbar / 0.145 psi <± 0.10% <± 0.02% x TD
30 mbar / 0.44 psi
100 mbar / 1.5 psi <± 0.065% <±-0.035% + 0.01% x TD
500 mbar / 7.3 psi <± 0.015% + 0.005% x TD
3 bar / 43.51 psi
16 bar / 232.1 psi <±-0.035% + 0.01% x TD
Effect of ambient Ambient temperature effect on zero and span in relation to the set measuring span. Applies to
temperature the digital interfaces (HART®, Profibus PA, Foundation Fieldbus) as well as for the analogue
4…20 mA current output.
[% of the set span]
-10...+60°C / +14...+140°F -40...+85°C / -40...+185°F
10 mbar / 0.145 psi <± 0.15% + 0.20% x TD <± 0.4% + 0.3% x TD
30 mbar / 0.44 psi <± 0.15% + 0.10% x TD <± 0.2% + 0.15% x TD
100 mbar / 1.5 psi <± 0.15% + 0.15% x TD <± 0.15% + 0.20% x TD
500 mbar / 7.3 psi <± 0.15% + 0.05% x TD <± 0.2% + 0.06% x TD
3 bar / 43.51 psi
16 bar / 232.1 psi <± 0.15% + 0.15% x TD <± 0.15% + 0.20% x TD
Effect of system Ambient temperature effect on zero and span in relation to the set measuring span. Zero-point
pressure offsets can be calibrated out under operating pressure. Applies to the digital interfaces
(HART®, Profibus PA, Foundation Fieldbus) as well as for the analogue 4…20 mA current
output.
[% of the set span]
up to nominal on zero on span
pressure
10 mbar / 0.145 psi 40 bar / 580 psi <± 0.10% x TD <± 0.10%
30 mbar / 0.44 psi
100 mbar / 1.5 psi 160 bar / 2320 psi 160 bar / 2320 psi: 160 bar / 2320 psi:
<± 0.10% x TD <± 0.10%
500 mbar / 7.3 psi or
3 bar / 43.51 psi 400 bar / 5800 psi 400 bar / 5800 psi: 400 bar / 5800 psi: ≤0.25%
≤0.25% x TD
16 bar / 232.1 psi
Effect of mounting A position-dependent zero offset can be corrected.
position
≤0.35 mbar per 10° inclination around the transverse axis
Long-term stability Applies to the digital interfaces (HART®, Profibus PA, Foundation Fieldbus) as well as for the
acc. to DIN 16086 analogue 4…20 mA current output.
and IEC 60770-1 [% of the set span]
<0.1% x TD (Turn Down) over a period of 5 years
Total performance At a temperature change of -10...+60°C / +14...+140°F, up to the indicated nominal pressure.
in accordance with [% of the set span]
DIN 16086
up to TD Nominal pressure -10...+60°C / +14...+140°F
10 mbar / 0.145 psi 1:1 40 bar / 580 psi <± 0.42%
30 mbar / 0.44 psi <± 0.29%
100 mbar / 1.5 psi 160 bar / 2320 psi <± 0.32%
500 mbar / 7.3 psi <± 0.18%
3 bar / 43.51 psi
16 bar / 232.1 psi <± 0.32%
The details of total performance comprise the reference accuracy, the effect of the ambient
temperature on the zero signal and the measuring span as well as the effect of the static
pressure on the measuring span.
Eperf=√ ((EΔTZ+EΔTS)2+EΔPS2+Elin2)
EΔTZ= Effect of ambient temperature on the zero signal
EΔTS= Effect of ambient temperature on the measuring span
EΔPS= Effect of the static pressure on the measuring span
Elin = Reference accuracy
Dynamic output These parameters depend on the fill fluid, temperature and, if applicable, the diaphragm seal.
behaviour For further information refer to Dynamic behaviour of the current output on page 30
Damping 63% of the input variable 0…999 seconds, adjustable in 0.1 second steps.
Temperature
The evaluation is made using the HART® output signal.
Resolution 1°C / 1.8°F
Accuracy ± 1°K
System pressure
Reference conditions • Ambient temperature (constant): +18...+30°C / +64...+86°F
acc. to IEC 60770-1 • Relative humidity (constant): 45...75%
• Ambient pressure (constant): 860...1060 mbar / 12.5...15.4 psi
• Mounting position: vertical
Reference accuracy Includes the non-linearity, hysteresis and repeatability under reference conditions. Applies to
acc. to DIN EN 61298 the digital interfaces (HART®, Profibus PA, Foundation Fieldbus) as well as for the analogue
4…20 mA current output. [% of URL]
up to nominal TD 1:1
pressure acc. to URL
absolute pressure
10 mbar / 0.145 psi 40 bar / 580 psi <± 0.10%
30 mbar / 0.44 psi
100 mbar / 1.5 psi 160 bar / 2320 psi
or
500 mbar / 7.3 psi 400 bar / 5800 psi
3 bar / 43.51 psi
16 bar / 232.1 psi
Operating conditions
Temperature
Operating -40...+85°C / -40...+176°F
temperature / nominal
temperature range
Ambient -40...+80°C / -40...+176°F
temperature
Storage temperature -40...+80°C / -40...+176°F
Climate category 4K 4H (air temperature: -20...+55°C, humidity: 4...100% according to DIN EN 60721-3-4)
Vibration
Reference conditions • Without mounting bracket
• Process flanges 316L / 1.4404 PN 160
• Single chamber housing, aluminium
Vibration resistance 10...58 Hz, 0.35 mm
acc. to IEC 60770-1 58...1000 Hz, 20 m/s2
1 octave per minute, 10 cycles per axis
Shock resistant 500 m/s2, 6 ms
according to 100 shocks per axis
IEC 60770-1
Noise 10...200 Hz, 1 (m/s2)2/Hz
according to
IEC 60770-1 200...500 Hz, 0.3 (m/s2)2/Hz
4 hours per axis
Materials
Gaskets
EPDM -40...+85°C / -40...+185°F
Copper
FKM -20...+85°C / -4...+185°F
Filling oil
Silicone oil -40...+85°C / -40...+185°F
Halocabon oil
Wetted parts
Process connection, 316L / 1.4404, NACE MR0175 / MR0103, Hastelloy® C-276
screwed flange
Separating diaphragm 316L / 1.4404, NACE MR0175 / MR0103, Hastelloy® C-276
Vent and lock screws 316L / 1.4404, NACE MR0175 / MR0103, Hastelloy® C-276
Internal transmission Silicone oil, halocabon oil
fluid
Non-wetted parts
Electronics housing Plastic PBT (Polyester), powder coated polyester die-cast aluminium, 316L / 1.4404
Housing cover sealing Silicone (Si 850,R), NBR (silicone-free version)
ring
Inspection window in Polycarbonate (UL746-C listed),
housing cover in case of Ex d version: glass
(display, adjustment
module)
Screws and bolts for up to PN160: hexagon screw DIN 931 M8 x 85 A2-70, hexagon nut DIN 934 M8 A2-70
the side flanges PN400: hexagon screw DIN 931 M8 x 85 A2-70, hexagon nut DIN 934 M8 A2-70
Grounding flange 316Ti, 316L / 1.4404
Process connection
Process 1/4-18 NPT (Female), IEC 61518 A
Mounting 7/16 UNF, M10 (up to PN160)
Electrical connections
Mechanical
Cable gland M20 x 1.5 mm
cable diameter 5...9 mm / 0.20...0.35"
6...12 mm / 0.24...0.47"
10...14 mm / 0.39...0.55"
Cable entry 1/2 NPT Blind plug: M20 x 1.5 mm, 1/2-14 NPT
Closing cap: M20 x 1.5 mm, 1/2-14 NPT
Connector option: M12 x 1, Harting HAN 7D,8D, 7/8 FF
Wire cross-section Solid wire with cords: 0.2 mm...2.5 mm2 / AWG 24...14
Cord with wire end sleeve: 0.2 mm...1.5 mm2 / AWG 24...16
Electrical
Operating voltage non-illuminated indication illuminated indication
Non-Ex device: 11...35 VDC Non-Ex device: 16...35 VDC
Ex ia device: 11...30 VDC Ex ia device: 16...30 VDC
Ex d device: 11...35 VDC Ex d device: 16...35 VDC
Ex ia d device: 15...35 VDC No Backlight
Reverse polarity Integrated
protection
Permissible residual Non-Ex devices, for Un 12 VDC (11 < UB < 14 VDC) ≤ 0.7 Veff (16...400 Hz)
ripple
Ex ia devices, Ex ia d devices for Un 24 VDC (18 < UB < 35 VDC) ≤ 1.0 Veff (16...400 Hz)
Load RL,max=(UB-11) / 22 mA
Potential connection Electronics: not electrically isolated
in device
Ground terminal: galvanically connected with process connection
Overvoltage category III
Protection class II
With integrated display and adjustment module the height of the housing increases by 9 mm /
0.35 inch.
With integrated display and adjustment module the height of the housing increases by 9 mm /
0.35 inch.
With integrated display and adjustment module the height of the housing increases by 9 mm /
0.35 inch.
With integrated display and adjustment module the height of the housing increases by 9 mm /
0.35 inch.
e
f
g
Overall height of the differential pressure transmitter = b (process connection) + overall height of
the respective housing
e
f
Overall height of the differential pressure transmitter = b (process connection) + overall height of
the respective housing
e
f
e
f
g
Figure 2-8: 90° vertical 1/4-18 Process connection, side vent (VD)
e
f
g
Figure 2-9: 1/4 NPT without venting, Process connection in Hastelloy® C-276 (HO)
e
f
g
Figure 2-10: 1/4 NPT Process connection in Hastelloy C-276 with venting on the process axis (HR)
f h
Figure 2-13: Mounting bracket for pipe and wall mounting 2¨ / 50.8 mm
h
e
Zero/Span adjustment
Zero: -95...+95%
Span: -120...+120%
Difference between zero and span: max. = 120% of the nominal range.
Maximum allowed turn down = unlimited (recommended 20:1)
Nominal range 10 mbar 30 mbar 100 mbar 500 mbar 3 bar 16 bar
Limit 10 mbar 30 mbar 100 mbar 500 mbar 3 bar 16 bar
URL (upper)
Limit -10 mbar -30 mbar -100 mbar -500 mbar -3 bar -16 bar
LRL (lower)
Smallest adjustable 0.5 mbar 1 mbar 1 mbar 5 mbar 30 mbar 160 mbar
measuring span
Turn down 20:1 30:1 100:1 100:1 100:1 100:1
MWP 40 bar 40 bar 160 bar / 160 bar / 160 bar / 160 bar /
(maximum system 400 bar 400 bar 400 bar 400 bar
pressure) 1
Minimum system 1 mbar abs (under reference conditions)
pressure
1 MWP corresponds to the PS designation in the PED (maximum system pressure)
Nominal range 0.15 psi 0.44 psi 1.50 psi 7.30 psi 43.51 psi 232.1 psi
Limit 0.15 psi 0.44 psi 1.50 psi 7.30 psi 43.51 psi 232.1 psi
URL (upper)
Limit -0.15 psi -0.44 psi -1.45 psi -7.25 psi -43.51 psi -232.1 psi
LRL (lower)
Smallest adjustable 0.007 psi 0.015 psi 0.015 psi 0.073 psi 0.44 psi 2.32 psi
measuring span
Turn down 20:1 30:1 100:1 100:1 100:1 100:1
MWP 580 psi 580 psi 2320 psi / 2320 psi / 2320 psi / 2320 psi /
(maximum system 5800 psi 5800 psi 5800 psi 5800 psi
pressure) 1
Minimum system 0.014 psi abs (under reference conditions)
pressure
1 MWP corresponds to the PS designation in the PED (maximum system pressure)
These parameters depend on the fill fluid, temperature and, if applicable, the diaphragm seal
Responsibility for the use of the measuring devices with regard to suitability, intended use and
corrosion resistance of the used materials against the measured fluid lies solely with the
operator.
This device is a Group 1, Class A device as specified within CISPR11:2009. It is intended for use in
industrial environment. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments, due to conducted as well as radiated disturbances.
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
The OPTIBAR DP 7060 is a differential pressure transmitter suitable for measuring flow, level,
differential pressure, density and interface of gases, vapours and liquids. The available
measurement ranges and the respective permissible overloads are indicated on the nameplate.
To observe the intended use, adhere to the following points:
The accuracy of the measurement is only guaranteed if the transmitter and accompanying
impulse line(s), if any, have been correctly installed. In addition, extreme ambient conditions
including large fluctuations in temperature, vibrations and shocks should be kept as far away as
possible from the measuring equipment.
3.4 Venting
The ventilation for the electronics housing is assured via a filter element in the vicinity of the
cable glands, which is permeable to air but moisture-blocking.
In order to ensure effective ventilation, the filter element must be always free of deposits.
Do not use a high-pressure cleaner to clean the housing. The filter element may become
damaged and as a result moisture can penetrate into the housing. The exception to this is the
IP69K single chamber housing.
• Mount the differential pressure transmitter below the lower process connection so that the
impulse lines are always filled with liquid.
• The low-pressure side (L) is open to atmospheric pressure.
• For measurements of fluids with solid content, the installation of separators and drain valves
is recommended to enable collection and removal of debris and sediment.
• Mount the differential pressure transmitter below the lower process connection so that the
impulse line is always filled with liquid.
• The low-pressure side (L) must always be connected above the maximum level.
• For measurements of fluids with solid content, the installation of separators and drain valves
is recommended to enable collection and removal of debris and sediment.
• Mount the differential pressure transmitter below the lower process connection so that the
impulse lines are always filled with liquid.
• The low-pressure side (L) must always be connected above the maximum level.
• For measurements of fluids with solid content, the installation of separators and drain valves
is recommended to enable collection and removal of debris and sediment.
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
Cables may only be connected when the power is switched off! Since the transmitter has no
switch-off elements, overcurrent protection devices, lightning protection and/or energy isolating
devices need to be provided by the customer.
Check which outer diameter is suitable for the cable gland in order to ensure the sealing effect
according to the specified IP protection class.
The terminals in the terminal compartment are designed for wire cross-sections of up to
1.5 mm². To ensure a proper connection, you should strip the cable 40…50 mm / 1.6...2".
Figure 4-5: Connector, Harting HAN 8D (left) and Harting HAN 7D (right)
1 VS-
2 VS+
In the device, the cable shield must be connected directly to the internal ground terminal.
The ground terminal outside on the housing must be connected to the grounding potential with
low impedance.
In hazardous areas, the grounding is carried out according to the installation instructions.
Significant potential differences exist inside galvanization plants as well as on vessels with
cathodic corrosion protection. A two-sided shield grounding can cause unacceptably high shield
currents as a result.
The metallic and wetted parts (process connection, cap flange, measuring cell and separating
diaphragm etc.) are conductive connected with the inner and outer ground terminal on the
housing.
Procedure
• Unscrew the housing cover.
• If present, remove the display and adjustment module by turning it to the left.
• Loosen union nut of the cable gland.
• For preparation of connection cable refer to Cable preparation on page 41.
• Push the cable through the cable gland into the terminal compartment.
• Insert the wire ends into the open terminal connection according to the wiring plan. Flexible
cores with wire end sleeves as well as solid cores can be inserted directly into the terminal
openings. In case of flexible cores, press the spring terminal with a small screwdriver to open
the terminal opening.
• Check the proper hold of the wires in the terminals by lightly pulling on them.
• Connect the cable shield to the internal ground terminal, connect the outer ground terminal to
the customer/plant equipotential bonding.
• Tighten the union nut of the cable gland. The sealing ring must completely enclose the cable.
• Screw the housing cover back on.
The following illustration applies to both the non-Ex as well as the the Ex ia, and the Ex d version.
Electronics compartment
The following illustration applies to both the non-Ex as well as the Ex ia, and the Ex d version.
Electronics compartment
Terminal compartment
OPTIBAR DP 7060
VGK7 4 Approval
AX Ex-free zone Europe
AC ATEX II 1G, 1/2G, 2G Ex ia IIC T6
AE ATEX II 1/2G, 2G Ex d IIC T6
AR ATEX II 1D, 1/2D, 1/3D, 2D Ex t IIC T* 1
AH ATEX II 1G, 1/2G 2G Ex ia IIC + II 1D, 1/2D, 1/3D, 2D IP66
A7 ATEX II 1/2G, 2G Ex d IIC + II 1D, 1/2D, 1/3D, 2D IP66
WX Ex-free zone International
WC IEC Ex ia IIC T6 Ga, Ga/Gb, Gb
WE IEC Ex d IIC T6 Ga/Gb, Gb
WR IEC Ex t IIIC T* 1
WH IEC Ex ia IIC T6 + IEC Ex t IIIC T... IP66
W7 IEC Ex d IIC T6 + IEC Ex t IIIC T... IP66
Material; Process connection / Material; Venting
S0 316L; 1/4-18 NPT IEC A; UNF 7/16 / 316L; without
SR 316L; 1/4-18 NPT IEC A; UNF 7/16 / 316L; on process axis
SD 316L; 1/4-18 NPT IEC A; UNF 7/16 / 316L; at side
A0 316L; Adpt. 1/2-14 NPT; UNF 7/16 / 316L; without
AR 316L; Adpt. 1/2-14 NPT UNF 7/16 / 316L; on process axis
AD 316L; Adpt. 1/2-14 NPT UNF 7/16 / 316L; at side
M0 316L; 1/4-18 NPT IEC A; M10 / 316L; without
MR 316L; 1/4-18 NPT IEC A; M10 / 316L; on process axis
VD 316L; 90°-vertical 1/4-18 NPT IEC A; UNF 7/16 / 316L; at side
H0 C-276; 1/4-18 NPT IEC A; UNF 7/16 / C-276; without
HR C-276; 1/4-18 NPT IEC A; UNF 7/16 / C-276; on process axis
D1 316L; diaphragm seal; H-side: capillary; L-side: open 1/4"-18NPT
DS 316L; diaphragm seal; H-side: direct attachment; L-side: open 1/4"-18NPT
D2 316L; diaphragm seal; H-side: capillary; L-side: capillary
DT 316L; diaphragm seal; H-side: direct attachment; L-side: capillary
Diaphragm / Filling oil
S 316L (1.4404); Silicone oil
R 316L (1.4404); Halocarbon oil
H Hastelloy C-276; silicone oil
Gasket / Temperature
E EPDM; -40...+85°F / -40...+185°F
F FKM; -20...+85°C / +5...+185°F
C Copper; -40...+85°C / -40...+185°F
Measuring range PN Accuracy class
1 10 mbar; 1 kPa; 0.14 PSI 40 bar / 580 psi 0.1%
2 30 mbar; 3 kPa; 0.44 psi 40 bar / 580 psi 0.1%
3 100 mbar; 10 kPa; 160 bar / 2321 psi 0.065%
1.45 psi
4 500 mbar; 50 kPa; 160 bar / 2321 psi 0.065%
7.25 psi;
5 3 bar; 300 kPa; 43.5 psi 160 bar / 2321 psi 0.065%
6 16 bar; 1.6 MPa; 232 psi 160 bar / 2321 psi 0.065%
D 500 mbar; 50 kPa; 400 bar / 5800 psi 0.065%
7.25 psi;
E 3 bar; 300 kPa; 43.5 psi 400 bar / 5800 psi 0.065%
F 16 bar; 1.6 MPa; 232 psi 400 bar / 5800 psi 0.065%
Adjustment (customer specific measuring range)
0 Nominal range %
1 Nominal range mbar
2 Nominal range bar
3 Nominal range psi
4 Nominal range Pa
5 Nominal range kPa
6 Nominal range Mpa
7 Nominal range mmH20
8 Nominal range inH20
A Nominal range mmHG
B Nominal range inHG
C Nominal range mm (density 1)
D Nominal range cm (density 1)
E Nominal range m (density 1)
F Nominal range in (density 1)
G Nominal range ft (density 1)
L Customer specific adjustment for level measurement
S Customer specific adjustment for flow measurement
Y Customer specific adjustment for pressure
Accuracy class
E < 0.1%
M < 0.065%
Electronics
H 2-wire 4...20 mA / HART®
A 2-wire 4...20 mA / HART® with SIL 2/3
F Foundation Fieldbus
P Profibus PA
Supplementary electronics
X without
Housing
K Plastic (PBT), DIN single chamber
R Plastic (PBT), DIN double chamber
A Aluminium, DIN single chamber
D Aluminium, DIN double chamber
8 316 L (electro-polished), DIN single chamber
V 316 L, DIN single chamber
W 316 L, DIN double chamber
Housing version / Ingress protection
1 compact; IP 66 / 67, NEMA 6P
M compact, IP69K
Identification tag
0 without
F Foil (40 x 20 mm)
S Stainless steel (40 x 20 mm)
Device settings
0 without
Further certificates
X without
M Observe VZPA / B code
Accessories
X without
M Mounting bracket;
316L
VGK7 4
1 depends on the equipment category
• Pressure
• Process Analysis
• Services