Benefusion Sp5: (This Manual Is Also Applicable To Benefusion Sp5 Ex, Benefusion Sp5 Tiva Syringe Pumps)
Benefusion Sp5: (This Manual Is Also Applicable To Benefusion Sp5 Ex, Benefusion Sp5 Tiva Syringe Pumps)
Benefusion Sp5: (This Manual Is Also Applicable To Benefusion Sp5 Ex, Benefusion Sp5 Tiva Syringe Pumps)
Syringe Pump
Operator’s Manual
© Copyright 2012-2017 Shenzhen Mindray Scientific Co., Ltd. All rights reserved.
Version: 13.0
I
Intellectual Property Statement
SHENZHEN MINDRAY SCIENTIFIC CO., LTD. (hereinafter called Mindray
Scientific) owns the intellectual property rights to this product and this manual. This
manual may refer to information protected by copyright or patents and does not
convey any license under the patent rights or copyright of Mindray Scientific, nor
the rights of others.
II
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray Scientific's obligation or liability under this warranty does not include any
transportation or other charges or liability for direct, indirect or consequential
damages or delay resulting from the improper use or application of the product
or the use of parts or accessories not approved by Mindray Scientific or repairs
by people other than Mindray Scientific authorized personnel.
Company Contact
Manufacturer: Shenzhen Mindray Scientific Co., Ltd.
Bldg 2,5, Mindray Guangming facility, 1203 Nanhuan
Address: Avenue, Yutang Block, Guangming District, 518016,
Shenzhen, P.R.China
Website: www.skmedica.com
Tel: +86 755 81888998
Fax: +86 755 26582680
III
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely and
in accordance with its function and intended use. Observance of this manual is a
prerequisite for proper product performance and correct operation and ensures
patient and operator safety.
This manual is based on the maximum configuration and therefore some contents
may not apply to your product. If you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to
the equipment so that it can be obtained conveniently when needed.
Intended Audience
This manual is geared for clinical professionals who are expected to have a
working knowledge of medical procedures, practices and terminology as required
for monitoring of critically ill patients.
Illustrations
All illustrations in this manual serve as examples only. They may not
necessarily reflect the setup or data displayed on your equipment.
Conventions
Italic text is used in this manual to quote the referenced chapters or sections.
[ ] is used to enclose screen texts.
< > is used to enclose the keys.
→ is used to indicate operational procedures.
IV
Contents
1 Safety ..................................................................................................................... 1-1
1.1 Safety Information ............................................................................................... 1-1
1.1.1 Dangers ................................................................................................... 1-1
1.1.2 WARNING ............................................................................................... 1-2
1.1.3 CAUTION ................................................................................................ 1-3
1.1.4 NOTE ....................................................................................................... 1-4
1.2 Equipment Symbols ............................................................................................ 1-5
1
4.2.2 Insert the Syringe .................................................................................... 4-2
4.2.3 Change the Syringe................................................................................. 4-4
4.2.4 Select Syringe Brands ............................................................................. 4-4
4.2.5 Memory Function ..................................................................................... 4-5
4.2.6 Select Infusion Mode ............................................................................... 4-5
4.2.7 Purge ....................................................................................................... 4-5
4.2.8 Set Infusion Parameters .......................................................................... 4-6
4.2.9 Infusion .................................................................................................... 4-7
4.2.10 Infusion Pause ....................................................................................... 4-7
4.2.11 BOLUS ................................................................................................... 4-8
4.2.12 Change the Rate during Operation ....................................................... 4-8
4.2.13 Complete ............................................................................................... 4-8
4.2.14 Standby ................................................................................................. 4-9
4.2.15 Turn off the Pump .................................................................................. 4-9
2
7.3 Nurse Call ............................................................................................................ 7-1
7.4 Wireless Networking (Optional) ........................................................................... 7-2
7.5 Data Export .......................................................................................................... 7-3
7.6 Automatic Multi-channel Relay ............................................................................ 7-3
7.7 Patient Information .............................................................................................. 7-4
7.8 Night Mode .......................................................................................................... 7-4
7.9 Day's Prescription ................................................................................................ 7-5
7.10 WLAN Setting .................................................................................................... 7-5
3
A.3.3 Host LED ................................................................................................ A-3
A.3.4 Auditory Indicator ................................................................................... A-3
A.3.5 External Ports ......................................................................................... A-4
A.3.6 Signal Output Interface .......................................................................... A-4
A.4 Specifications ..................................................................................................... A-5
A.5 A Reference Table Showing Occlusion Alarm Delay and Possible Dose .......... A-7
A.6 Infusion Accuracy Curve .................................................................................... A-8
A.7 Trumpet Curve ................................................................................................... A-8
4
1 Safety
1.1 Safety Information
The safety statements presented in this chapter refer to basic safety information
that the operator must pay attention to and abide by when using the syringe
pump.There are additional safety statements in other chapters or sections, which
may be the same as or similar to the following, or specific to particular operations.
Dangers
WARNING
CAUTION
NOTE
Provides application tips or other useful information to ensure that you get
the most out of the product.
1.1.1 Dangers
This Manual does not contain any information at the "Danger" level.
1-1
1.1.2 WARNING
WARNING
This equipment can only be connected to the socket with ground protection.
Please adopt a rechargeable battery instead of the socket as the power supply if the
socket is not provided with a ground lead.
The patient's clinical condition and the working condition of the syringe
pump must be monitored carefully, and the alarm volume and alarm levels need to be set
according to the actual needs. Operation and performance relying solely on the auditory
alarm system alone is not sufficient, and setting the alarm at a low volume may endanger
the patient. If the alarm volume is less than the surroundings volume, which can further
lead to operators identify alarm mistakenly.
Please carefully install the power line and cables with various accessories to
prevent the patient from choking or suffocation caused by entanglement of the cables or
by electrical disturbance.
Infusion cannula knots, filter coagulation and occlusions arising from needle
insertion can cause the pressure inside the syringe to rise during infusion.When this
occurs, removing the occlusion can cause excessive liquid to be infused into the patient,
so appropriate measures should be taken.
The pump should not be placed more than 100 cm above or below the level of
the patient’s heart. The smaller the height difference between the pump, the more accurate the
pressure test in the infusion cannula will
1-2
be.
Do not touch the patient when connecting the peripheral equipment via the
input/output signal ports to prevent patient leakage current from exceeding the
requirements specified by the standard.
1.1.3 CAUTION
CAUTION
Before the equipment is connected to the power supply, check that the
voltage and frequency of the power supply match the specifications on the label or in this
Operator’s Manual.
Please install and carry the equipment correctly to protect the equipment
from damage from drops, impacts, violent shaking or other external mechanical forces.
Check the built-in battery before use to make sure it has sufficient
power.Recharge the battery if necessary.
1-3
The syringe with the luer taper is recommended for use, which can
effectively prevent patients from under current caused by the occurrence of the cannula
to slip out when under tension.
1.1.4 NOTE
NOTE
Keep this Operator’s Manual near to the equipment for future ease of
reference.
Do not insert devices that are not specified by the manufacturer into the
multifunction interfaces.
During infusion, the syringe pump can accurately control the rate, infusion
volume and infusion time, and monitor the operation in real-time, to effectively prevent
over currents, under currents and instances of backflow.
The device is not in touch with the drugs or patients directly. Thus, there is
no need to process Biocompatibility test on it.
1-4
1.2 Equipment Symbols
The equipment you purchased may not provide you with all the following symbols.
NOTE!
Refer to the
ON/OFF
accompanying document
(This Manual)
Clear/Back Start
Bolus Confirm
Stop Menu
Current pump
communicated
Wireless network is
successfully with
establishing a network
BeneFusion DS5
connection
Infusion Supervision
System
1-5
Non-executed
Executing prescription
prescription
Caution Lock
Stacking limit by
Fragile, handle with care
number
Electronic equipment:
dispose of separately to
Wireless transceiver
avoid polluting the
environment
Recycle CE mark
Environmentally-friendly
The European Union
use periods of electronic
Representative Office
products (20 years)
The Syringe Pump is suitable for adults, children and newborns in clinical
departments.
WARNING
The syringe pump is for clinical use. It must only be used under appropriate
conditions by professional clinicians, medical device technicians, or by suitably trained
nurses. Personnel using this product must receive sufficient training.This product must
not be operated by anyone who has not been authorized to do so or has not received
suitable training.
2.1.2 Contraindications
None
2-1
Anti-bolus Function.
Since some parts and functions are optional, the Syringe Pump you purchased
may not contain these additional parts and their relevant functions.
2.2 Host
2.2.1 Front View
1. Alarm light
The alarm light indicates different alarm levels in different colors and
flash frequencies, please refer to Chapter 8 Alarms for details.
2. Display
Used for displaying infusion parameters and relevant content.
3. <DIRECTION>
Used for adjusting value, change lines and pages.
4. <CLEAR/BACK>
Under non-setting status, indicate to returns to previous menu or operation.
Under the setting status, indicate to clear the current set or cancel the edit.
5. <START>
After installing the syringe correctly and completing setting infusion parameters,
press this key to start the infusion.
6. <BOLUS >
During infusion, press this key to enter the Bolus Settings screen.
When the pump is stopped, press this key to enter the purge settings screen.
2-2
7. <POWER>
Used for turning power on, entering in standby state and turning off operations.
When power off, press and hold (>3 s) the key.
8. AC indicator light
On: The pump is connected to an AC power supply.
Off: The pump is not connected to an AC power supply.
9. Clip
Clamps the syringe plunger.
10. Handle
Controls clip opening/closing and slider action.
11. <OK>
Used for confirming input operation and saving value.
12. <STOP>
During infusion, press this key in any non-setting interface to stop infusion.
Infusion stops caused by alarms, such as occlusion and so on, press this key to
cancel the alarm.
14. <MENU>
interfaces.
Under non-operation status, used for switching [Main Menu] interface and other
Under operation status, press and hold this key to lock;in locked state, press and
hold to unlock.
Light off indicates that there is no battery or the equipment is turned off and not
connected to an AC power supply.
17. Slider
Secures syringe and drives plunger assembly.
2-3
2.2.2 Rear View
1. Battery box
2. Multi-channel pumps connect rail Used for combining multi-channel pumps.
3. Slider
4. Equipotential port
When other device and syringe pump are used together, please adopt wires
to connect other devices and equipotential port of the pump, to eliminate
ground potential difference between different devices and ensure safety.
5. Alternating current power supply (AC) port
Connected by three-core-type power cord and AC power source.
6. Multifunction interface, which combines the following interface functions:
DC power input interface
RS232 interface
Nurse call interface
1. Speaker hole
2. Multi-channel pumps hasp
The multi-channel pumps are connected to secure the equipment.
2-4
2.2.4 Side View with the Door Opened
1. Warning label
2. Speaker hole
3. Door
4. Spindle clamp
5. Flange plate
6. Clip
7. Handle
1. Product label
2. Fixed mounting holes
2-5
2.3 Screen Display
This syringe pump has a built-in LCD screen. The display information comprises
three main parts:
1
3
1. Title bar
Display the name of the current screen.
Display drug information, please refer to 6.2 Drug Library.
Display alarm information, please refer to Chapter 8 Alarms.
Display lock icon, please refer to 6.4 Key Lock Function.
Display wireless icon, please refer to 7.4 Wireless Networking (Optional).
Display battery icon, please refer to Chapter 9 Battery.
2. Parameter area
Display every parameter and the parameter value of the current screen.
3. Prompt bar
Display prompt messages, run icon and so on.The run icon on the screen displays
the running operation:
Green icon indicates normal running. Arrows move from right to
■
left, and the running speed increases as the rate is increased.
■ Blue icon lighting indicates bolus.
■ Steady grey icon indicates infusion stops.
2.4 Cursor
In the main screen and parameter settings screen, when the cursor is located at an
option or at a data value, the grounding of the option or the data value will turn to
white and the font will become black. Press or to move cursor up or
down and confirm the location. Press move the cursor up and down and
confirm the location.
2-6
3 Installation and Setting
3.1 Installation
WARNING
All the analog equipment and digital facilities should be certified according
to the specified IEC standard (such as:IEC60950 Information Technology Equipment
Safety and IEC60601-1 Medical Electrical Equipment Safety), moreover, all equipment
should be connected based on the requirements of the valid version of the IEC60601-1
system.The qualified individual responsible for connecting auxiliary equipment to the
input and output signal ports is also accountable for making the system in accordance
with the IEC60601-1 standard. Please contact the company if you have any queries.
3-1
Please carefully remove the equipment and its accessories from the packaging in a
correct manner, and inspect them against the packing list. Examine the equipment
for any mechanical damage and ensure that the box includes all items on the
packing list. Please contact the company if you have any queries.
NOTE
Keep the packing case and packaging materials for future transportation or
storage.
WARNING
They must be kept out of the reach of children. The packaging materials
must be disposed of in compliance with local laws and regulations or the hospital policy
on waste management.
WARNING
Please use only when the operating environment meets the requirements
specified above. Otherwise, the pump's performance will not match the technical
specifications in A Product Specifications. Device failure and other unexpected
consequences may also result.
3-2
3.1.3 Mount the Clamp
1. Align the mounting holes on
the installation dead plate with
the installation screw holes on
the bottom of the machine, and
tighten the screws.
2. Turn counterclockwise to
loosen the pole clamp until an
IV stand can be inserted in.
3-3
3.1.4 Multi-channel Pumps Combination
A maximum of 2 pumps can be combined (syringe pump and infusion pump can
be cross combined).
When multi-channel pumps are combined, align the rail on the bottom of the upper
pump to the rail slot on the top of the lower pump, slide from backward to
frontward until the alignment is matched, then strain the two pumps by the hasps.
When disconnecting, hold the upper-part of the hasp of the upper pump and pull
out the lower pump.
WARNING
The syringe pump must be installed horizontally.
NOTE
Please ensure the stand is stable before installation.
When the pumps are overlapped, all fixing clamps except the fixing clamps
of the lowest pump should be removed, otherwise, the pumps can be used overlapped.
3-4
Press the button of pole clamp, horizontally or vertically adjust pole clamp, turn the
handle, pump can be fixed to cross bar or vertical bar.
Square tube: Width 10mm, length 15-40mm
Round tube: External diameter ø15~ ø40
NOTE
Only 2 pumps (syringe pump and infusion pump can be cross combined) can
be carried by the handle set, or the handle of the handle set might be broken. Please refer
to 3.1.4 Multi-channel Pumps Combination for the combination operation of multi-pump.
3-5
WARNING
Do not touch the power plug with wet or moist hands! If there is a liquid drug
or residue on or around the power socket or plug, the user should completely clean and
dry the area before plugging into the power supply, or accidents or injuries may result!
NOTE
Compatible power supply: 100–240 V~, 50/60 Hz.
The AC power cable should be correctly inserted and secured into the
socket.
3-6
3. Select [Time Format]: [24h] or [12h].
4. Select [Date Format]: [yyyy-mm-dd], [mm-dd-yyyy] or [dd-mm-yyyy].
CAUTION
Please check the system date and time to keep accurate records in the
History function.
After changing the time format or date format, the record will update new
format automatically.
3-7
4 Basic Operation
4.1 Infusion Flow Chart
Turn pump on
Press to activate and turn on the pump, the pump will
run start-up checks and display the start-up screen;
Install syringe
Please refer to 4.2.2 Insert the syringe;
Select brand Select according to the syringe brand currently being used;
Select infusion mode Enter default infusion mode, if needs switch, refer to
4.2.6 Select Infusion Mode;
Purge
Refer to 4.2.7 Purge;
Insert cannula
Connect to the patient;
Start infusion
Press ;
Finish
Press ;
Remove syringe Disconnect from the patient, and remove the syringe in
a correct manner;
Turn pump off Hold down (>3 s) , until the Power Off progress bar
complete.
4-1
4.2 Operational Procedures
4.2.1 Turn on the Pump
After completing the syringe pump set up, turn on the device as in the following
steps:
1. Perform a safety inspection referring to 11.1 Inspection before turning on the
pump.
2. Press , the system will initiate the self-test and the screen will display the
[System Self-test] interface:
First, the system will give out a sound "di" —— indicating the self-testing of the
loudspeaker to be successful.
Then, the color of the alarm indicator lamp will change from red to yellow, turn on
and off orderly —— indicating the self-testing of the alarm lamp to be successful.
At last, the system will give out a sound of "didi" —— indicating the self-testing of
the buzzer to be successful.
3. Enter the operation interface after successfully completing the system self-testing,
and now you can manually operate the system through the key board.
WARNING
Please monitor the self-test process to make sure that the speaker, the alarm
light, and the buzzer are all self-tested successfully. Otherwise, please contact the
company and do not operate the pump until after maintenance is performed.
4-2
Install syringe according to the following method:
1. Open the door, squeeze the
handle to open the clip, then move
the slider to the appropriate
position. Pull out the spindle clamp
and turn it to the right, allowing the
flange handle to form a slot.
4-3
NOTE
If the syringe pump has been installed with a syringe, then the system will
directly skip [Installation Guide] interface to enter [Syringe selection] interface.
WARNING
The flange of the syringe should be firmly inserted into the slot, and not
jutting on the outside of the flange plate.
Before using this syringe pump, the syringe pump, syringe and other
accessories should be installed correctly.
Before using this syringe pump, the brand and specifications of the syringe
used must be confirmed. The brand of syringe pump should be calibrated on the
equipment. If there are no settings for the syringe used, the rate and the alarms may not
be accurate.
screen, and now you can press to select the syringe brand and
specifications of the syringe currently being used. Specific brand please refer to
6.7
Common Syringe Brands.
CAUTION
Please confirm that the current selected brand is the same as the brand
actually used.
4-4
WARNING
A new syringe brand should be calibrated when used for the first time.
After selecting the syringe brand, the previous infusion screen will appear, the
previous therapy parameters will be loaded and will prompt the users [ Use last
therapy parameter], start the infusion after confirming. If not necessary,
please perform the normal operation process.
4.2.7 Purge
During infusion, the user should prevent air bubbles from entering the blood with the
liquid drug, which may form an aeroembolism and put patients in serious danger.
Therefore, air bubbles in the syringe and extension cannula should be eliminated
before the infusion. Press to enter [Purge] setting screen, set purge [Purge
rate], then hold down to purge the system at the set rate, release after
the air bubbles are purged, completing the purge.
WARNING
During the purge, please disconnect the pump from the patientotherwise, the
patient will be in serious danger!
The default rate when purge is the maximum rate of each specification of
syringe, and the default purge rate of 50ml/60ml syringe is 800ml/h.
The purge accumulated volume is not added to the total volume.
4-5
4.2.8 Set Infusion Parameters
Under each infusion mode, the users should master the following basic function of
the keys:
: Under the non-setting status, move the cursor up and down;Under the
setting status, indicate to increase/decrease the data value.
: Under the non-setting status, move the cursor to the right and left; Under
the setting status, indicate to increase/decrease the editing space.
menu;Under the setting status, indicate to clear the current set or cancel the
edit.
As shown above, the procedure for setting the parameter values is as follows:
Step 1: Press the or key to move the cursor up or down and select the
parameter that requires setting;
Step 2: Press to enable the currently selected parameter by using the cursor
for making adjustments;
Step 5: Repeat step 3 and step 4 until finishing all value settings, and press for
confirmation after completing both steps. The settings are now complete.
The parameter value should not exceed the parameter range defined by this
equipment, please refer to Chapter 5 Infusion Mode for each parameter range,
otherwise, the parameter value will be modified automatically to the maximum value
defined when the setting exceeds the maximum value set, press again at the
original space or top of the digit to restore the original value.For example, if the
maximum parameter value is 1500, while the current value is 600, once the user
1500, press again at the thousand digit to recover to 600.When the parameter
4-6
reaches the maximum value, press at any digits, the value will not change,
and the prompt bar displays [Value Reached Limit] and provides a range of
the parameter.
4.2.9 Infusion
When ready, connect the extension cannula to the patient. Press to start the
infusion, and the screen will display the green running icon, the arrows will move
from right to left, and the running speed will increase, which will, indicate that the
rate will also increase.
Run screen with [VTBI] set Run screen without [VTBI] set
WARNING
NOTE
please press to enter the [Pause] interface to stop the pump. On the [Pause]
and then press to enter the [Pause] screen. Press to continue the
infusion.
4-7
4.2.11 BOLUS
In any run screen in the infusion mode, press to enter the [Bolus] settings
screen. There are two ways to start the bolus:
Manual Bolus: Set bolus parameters, hold down to manual bolus, and release
to return to the original rate.
NOTE
The bolus rate should be faster than the current infusion rate.
The default rate when bolus is the maximum rate of each specification of
syringe, and the default bolus rate of 50ml/60ml syringe is 800ml/h.
[VTBI Near Done] and [Syringe Nearly Empty] alarm will not be triggered
during bolus.
press or again for confirmation, then start to infuse under the new set
rate.
4.2.13 Complete
If [VTBI] is set during the infusion, when the remaining infusion time is close to the
[Time Near End] set by the users, the [VTBI Near Done] alarm will be triggered. If
no action has been taken, the alarm will not be cancelled automatically until the
infusion is completed, and then switch to [VTBI Done] alarm. Set [Time Near
End], please refer to 6.6 Time Near End.
If [VTBI] is not set during the infusion, when the time in which the remaining
liquid needs to reach [Time Near End], the [Syringe Nearly Empty] alarm will
be triggered, and this alarm cannot be canceled until the syringe is empty.
4-8
When infusion is completed, enter to [KVO] mode, and KVO mode will run for 30
min at most. Infusion will be stopped automatically after the KVO is finished, and
the [KVO Finish] alarm will be triggered. Set KVO rate, please refer to 6.1 KVO.
4.2.14 Standby
hh:mm), press for confirmation after modifying. The equipment cannot be put
in standby mode if there is an advanced alarm.
When the standby state is ended, the title bar will display [Standby Time Expired],
press to confirm and quit, until the screen before standby appears. Press
to remain in standby.
2. Hold down (>3 s) , until the Power Off progress bar complete, and the power will
turn off.
NOTE
When powering off normally, the current operating data and saved data will
be autosaved.
4-9
5 Infusion Mode
5.1 Rate Mode
00:00:01-99:59:59 h:m:s
Set [Rate] and [VTBI], then calculate [Time] automatically;
Time
Set [Rate] and [VTBI], modify [Time], and the [VTBI] will not
change and will automatically calculate [Rate].
VTBI 0.10-9999.99ml
Not required, and the range is the same with the Rate
Time Mode
Mode.
Rate Set [Time] and [VTBI], then calculate [Rate] automatically.
Modify [Rate], and the [VTBI] will not change and
automatically calculate [Time].
Weight 0.1-300.0kg/0.2-660.8 lb
Drug amount
g/mg/μg/ng/IU
unit
Body
Volume 0.10-9999.99ml
Weight
Dose rate 0.01-999.99
(BW) Mode
VTBI 0.10-9999.99ml
NOTE:
1. [Conc.] will be automatically calculated according to the formula
(Drug amount/Volume).
2. [Rate] will be automatically calculated according to the formula (Dose rate
*Weight)/Conc.
5-2
5.4 Intermittent Administration Mode
In Intermittent Administration Mode, by setting the Rate, Intmt.volume and Intmt.time to
control the infusion. It is suitable for the infusion of long-term analgesia drugs.
Mode
Intmt.volume 0.10-9999.99ml
Note: After a certain intermittent infusion period is finished, system will display the
remaining time for the next startup, press to continue the infusion.
5-3
5.5 TIVA Mode (Optional)
When used for the total intravenous anesthesia clinically, the TIVA mode controls
the intravenous infusion by setting the Induction parameters and by sustaining
those parameters. First set an Induction time to infuse the entire Induction dose,
which will enable the patients to enter narcotism, and then push the drug with a
maintenance speed to maintain anesthesia depth.
Parameter
Mode Parameters
Range
Basic Weight, Drug amount, Drug amount
Parameters unit, Volume The same
NOTE:
1. The [Induction rate] and [Maintenance rate]are calculated according to the
formula automatically, and the range of the values is taken to the same with
Rate Mode.
2. During the Induction stage, when the current accumulative amount reaches the
Induction VTBI, no matter by manual bolus or auto bolus, the equipment will
quit automatically and enter the maintenance stage.
3. If the [Induction time] is null, the equipment will enter the maintenance stage
directly.
5-4
5.6 Sequential Mode (Optional)
Several different sequences (parameter group) can be set in sequential mode,
and the syringe pump infuses according to the set infusion sequence.
5 sequences can be set in this mode. The rate of the current sequence can be
changed during the operation process. In Sequential Mode, the VTBI, Rate, and
Time are settable and the ranges of the set values are taken to be the same
with Rate Mode.
: A sign denotes the total VTBI and the total time of all sequences.
: A status sign denotes the need for a voice prompt after completing each
prompt will activate after the sequence is completed. Press again to cancel
the activation, there is no voice prompt after the sequence is completed,
NOTE
If only set [Time] for a sequence, denoting that infusion of the sequence is
stopped, and the next sequence will be started when arrived at the specific time.
If only set [Rate] or [VTBI] for a sequence, the sequence is invalid.
5-5
5.7 Loading Dose Mode (Optional)
Loading Dose Mode infuses all [VTBI] of loading dose at the speed of [Loading
Rate], then operates at [Main rate] until completion [VTBI] of the primary
parameters. Primary parameters [VTBI] are total VTBI.
Parameter
Mode Parameters
Range
VTBI (Main VTBI=Total VTBI)
Primary Rate (Main rate) The same with
Loading Parameters Time=(Main VTBI-Loading Dose Rate Mode
Dose VTBI)/ Main rate
NOTE:
1. The primary parameters of the [VTBI] should be more than the loading dose
of the [VTBI], and if set over the specified limit, the parameters that exceed
the system limit can not be set.
2. When the [VTBI] or the [Time] of primary parameter are not be set, the
system will autonomously run with the [Rate] of the primary parameters until
being emptied or completed.
3. When any two parameters of the loading dose are null, the system will run
with the [Rate] until being emptied or completed.
5-6
5.8 Ramp Up/Down Mode (Optional)
In Ramp Up/Down Mode, by setting the ramp-up time and ramp-down time, the
syringe pump will increase the rate (divided as 9 stages) automatically during
the ramp-up time until the rate is steady, sustaining for a period of time, and then
decrease the rate during the ramp-down time.
During any stage of the Ramp Up/Down Mode, if the Bolus Function is initiated, the
VTBI for maintaining the infusion of each stage, such as ramp-up, maintenance and
ramp down, will not be changed.When the bolus rate is recovered to its original
rate, the system will start to run at the stage where the current total accumulated is
the amount located.
If the modified rate is in the Ramp Up/Down Mode, the system will be operated
with the modified rate until completing the infusion or until the syringe is emptied.
Up/Down
Ramp up time (t1) Inputtable and modifiable, not required.
Mode
Micro-infusion
Mode VTBI 0.10-1000.00ml
5-8
6 Setting Parameters
6.1 KVO
KVO (Keep Vein Open) means to keep the vein open, during which the syringe
pump continues infusion at a very low rate after finishing the infusion in order
to prevent blood backflow or vascular occlusion.
4. In the [Select Drug] interface, press to turn the pages to browse the whole drug
list, and for further details, see C.5 Drug Library List.
5. After the drug is selected, its name will appear on the Run screen.
6-1
CAUTION
CAUTION
Carefully confirm the edit when changing the current pressure unit.
The pressure icon on the screen indicates the condition of the current pressure:
When the progress bar is white, this indicates that the current
◆
pressure value is below 80% of the occlusion pressure.
When the progress bar is yellow, this indicates that the current
◆ pressure value is above 80% and below 95% of the occlusion
pressure.
When the progress bar is red, this indicates that the current
◆ pressure value is above 95% and below 100% of the occlusion
pressure.
6-2
6.4 Key Lock Function
When locked, an icon in the upper-right corner of the screen emerges. The
following are two ways for automatic locking and manual locking:
Automatic Locking:
1. Select [Main Menu]→[General Option]→[Auto-lock time].
2. Select [Auto-lock time]: Off, 1-5min. After a specific time is set during the running
state, and if there is no operation or high-level alarm within the set
auto-lock time, the key board will be auto-locked. [Off] indicates closing
automatic locking function.
Manual Locking: In the running interface, under the unlocking lock condition,
6-3
6.7 Common Syringe Brands
There are multiple commonly used syringe brands installed inside the syringe
pump, making it convenient for the user to select from.Only 5 ml, 10 ml, 20 ml, 30
ml and 50 ml/60ml syringes that comply with national standards should be used
with this syringe pump. For specific syringe brands, please refer to actual syringes
as the standard.
NOTE
For syringe pump inserted to BeneFusion DS5 Infusion Supervision
System, Bed No. can’t be changed.
6-4
7 Other Functions
7.1 History Record
The syringe pump when in use will produce some key data stored in [History Record],
providing foundation for the treatment review and maintenance review at a later
period.The attribute of recording events includes action, time and description.
A record is created once an event occurs. The memory can store up to 2000
records. Once the memory is full, the oldest records will be removed first. History
record will not loss when the syringe pump powers down.
7-1
additional pulse signal is outputted.
Trigger Type
1. Normally Close: Select when the hospital call system is set as [Nor. Close].
2. Normally Open: Select when the hospital call system is set as [Nor. open].
Alarm Level: Three options: [High], [Medium] and [Low].The system sends nurse
call signals according to the alarm at the selected alarm level or above.
WARNING
Non-medical personnel are forbidden to modify the nurse call setting.
The nurse call function must be used in conjunction with a special cable.
NOTE
Medical personnel should not consider the nurse call function as the main
alarm notice approach, and rather combine the sound and visual alarms of the syringe
pump and the clinical performances and symptoms of the patient in order to judge the
patient's conditions and take further attention as needed.
Normal communication of the pump and CIMS depends on whether the network
connection is successful, operators are unable to observe the operation status of
the pump in real time when the communication is interrupted. After the network
connection settings of the pump and CIMS are modified, operators shall reset the
network connection as required in the manual to ensure the communication of the
pump and CIMS are restored.
7-2
When using wireless modules to connect to the Internet while using the syringe
pump, the wireless icon on the upper-right corner indicates the working condition of
the wireless modules:
■ Wireless modules work in order
NOTE
Wireless security transmission distance is no more than 50 meters.
2.4 GHz Wi-Fi frequency range, WEP and WPA/WPA2 security modes and
802.11b/g/n wireless standard are supported.
2. Set the slot No. of the “Forward channel”, press to save. If the slot of
“Forward channel” pump is infusion pump, please input VTBI parameter value
of the pump.
3. When the pump with finishes infusion, the current pump with will start
running automatically. When the last relayed pump finishes infusion,
[VTBI Done] alarm will be triggered.
7-3
7.7 Patient Information
Syringe pump has been inserted to BeneFusion DS5 Infusion Supervision System
and the system contains patient information, the system will automatically
distribute patient information to all syringe pumps when pumps are on.
to discharge patient.
NOTE
For syringe pump inserted to BeneFusion DS5 Infusion Supervision
System, patient information can’t be discharged.
3. Set start time, end time, volume and contrast, and press to save. When
system time reaches the start time of night mode, volume and contrast of
system mode will automatically change to volume and contrast of night mode;
when system time reaches the end time of night mode, volume and contrast of
night mode will automatically change to system volume and contrast.
Note: Syringe pump has been inserted to BeneFusion DS5 Infusion Supervision
System and the system contains night mode information, the system will
automatically distribute night mode information to all syringe pumps when
pumps are on.
NOTE
Minimum time period of start time and end time is 1 hour, while maximum
time period is 15 hours.
WARNING
Before entering night mode, confirm the settings of brighness and alarm
volume. Pay attention to the potential risk when the setting value is a bit low.
7-4
7.9 Day's Prescription
Syringe pump has been inserted to BeneFusion DS5 Infusion Supervision System
and the system contains prescription information, the system will automatically assign
all non-executed prescription to all idle syringe pumps when pumps are on.
To execute prescription:
If password is not required for to-be-connected network, you can connect the
network directly.
7-5
8 Alarms
The alarm is used in order to alert the medical staff by means of sound and light
when abnormal situations occur during the infusion procedure which can lead to
infusion changes or when the infusion of the patient cannot continue due to the
unexpected breakdown or pause/delay of the syringe pump.
WARNING
Among the visible alarms and audible alarms, the alarm information will
distinguish alarm levels in different ways.
8-1
8.2.1 Multi-level Alarm Rules
When several alarms occur simultaneously, the alarms proceed according to
the following rules:
When several alarms at different levels occur, the visible alarms and audible alarms
are consistent with the highest-level alarms.
When several alarms at different levels occur, only the highest-level alarm is
displayed, and after it is cancelled, the lower-level alarm will then be displayed.
When several alarms at the same level occur, the alarm information will be
demonstrated in an alternate manner.
The title bar of the syringe pump screen will display the corresponding alarm
information during the alarm blast, to see more details in D Alarm
Information:
Occlusion
Battery empty
VTBI done
KVO finish
Syringe empty
Syringe disengaged
System error
Reminder
Battery Low
No battery inserted
System abnomal
Syringe nearly empty
VTBI near done
Standby time expired
AC power disconnection
Pump serial number conflict
Relay invalid
Communication is interrupted
Syringe not correctly inserted
NOTE
8-2
For high level (except battery empty) and medium level alarms, press to pause
alarm sound for 2 minutes, no alarm sound is made in any case. When the
alarm pause time expires, the alarm tone will sound. Press to cancel high level
alarms (except battery empty and system error).
For low level alarms, press , no alarm sound, alarm information and alarm light,
until it is triggered next time.
NOTE
WARNING
When an alarm is triggered, please follow these steps and take appropriate action:
1. Check the patient;
2. Check the alarm type and the parameter which triggered the alarm;
3. Determine the reason for the alarm;
4. Eliminate the reason for the alarm;
5. Check whether the alarm has cleared.
NOTE
Please refer to D Alarm Information for specific handling procedures for each
alarm.
The operator position shall be the normal operating position of the syringe
pump (0.5m). Otherwise, operators might identify alarm mistakenly.
8-3
9 Battery
The syringe pump is configured with rechargeable Lithium ion batteries to ensure
that the syringe pump can be used normally under the condition of the patient's
migration within the hospital or during the circumstance of a power failure. When
the syringe pump switches to the AC power, the battery can be charged regardless
of whether the syringe pump is on or off. The battery is chargeable only within the
syringe pump.During charge, the battery icon in the upper-right corner of the screen
floats up and down.If the battery icon stops floating and is completely filled, it
indicates that the battery is fully charged. Under the condition of a sudden power
failure, the system will automatically use the battery to provide power as a backup.
The battery icon on the screen indicates the condition of the battery:
The battery jar of the syringe pump is installed with batteries,
and the white fill area indicates the quantity of electricity.
The power supply by the battery can only be sustained for a limited period of time.
The [Battery Empty] alarm will be triggered when the battery voltage is too low,
and red alarm light will flash. The alarm will continue within the remaining time of
the battery’s electric quantity and cannot be paused. Now, the syringe pump should
be switched on to AC power for charging.
WARNING
Please keep the batteries out of the reach of children.
NOTE
Please remove the batteries before conveying or when out of use for a long
time.
9-1
9.1 Battery Performance Optimization
When the battery is used for the first time, at least two complete optimizing cycles
should be ensured. A complete optimizing cycle entails the following: Charging
incessantly, and then discharging until the power of the syringe pump runs out.
During usage, regularly optimizing the battery performance will extend its lifespan.
It is suggested that the battery should be optimized when in use or in storage for
two months, or when the running time of the battery is significantly shortened.
9-2
NOTE
The lifespan of the battery depends on how frequently it is used and on how
long it has been used.If the maintenance and storage of the battery is appropriate, the
lifespan of the Lithium ion battery is about 5 years.Inappropriate use of the battery will
shorten its lifespan.We recommend replacing the Lithium ion battery every 3 years.
The length of the battery’s lifetime depends on the device configuration and
operation, for example:Under the condition of the power supply by the battery, frequent
infusion at a high rate will also shorten the length of the battery’s lifetime.
WARNING
9-3
10 Preservation and Sanitation
The pump must be cleaned or disinfected using the materials and methods listed in
this section. The manufacturer will not be responsible for any damage or accident
caused by cleaning and disinfection using other materials and methods.
The manufacturer shall not be held responsible for the efficacy of the following
chemicals or methods for infection control. Please contact your hospital's
infection prevention department or epidemiology specialists for advice on
infection control practices.
10.1 Description
Please make sure that your device and other fittings are clean without dust.In order
to prevent any damage to the device, please abide by the following rules:
Dilute all cleaning agents and disinfectants in accordance with the manufacturer’s
instructions, or use as low a concentration as possible.
Do not immerse or submerge the device in liquid.
Do not pour liquid on the device or accessories.
Avoid liquid from entering the pump body.
Do not use abrasive materials (such as steel wool or silver polishes), or any strong
solvent (such as acetone or any detergent containing acetone).
WARNING
Turn off the pump and disconnect the AC power source line from the socket
before cleaning. Do not clean and disinfect the device, export history record and perform
other operations when patients are using the pump.
10.2 Cleaning
The pump should be cleaned regularly. If operating in dirty or sandy areas,
cleaning should be more frequent. Before cleaning, please consult or refer to the
hospital's specific regulations concerning medical device cleaning. The following
are detergents available for selection:
Hydrogen peroxide (3%)
Ethanol (70%)
Isopropanol (70%)
10-1
When cleaning the device:
1. Turn off the pump and disconnect the AC power source line.
2. Wipe the display screen after soft cotton balls absorb an appropriate amount of
detergent.
3. Use a piece of soft cloth which absorbs a modest amount of cleaning agent to wipe
the surface of the device.
4. When necessary, use a piece of cloth to wipe off any excess cleaning agents.
5. Place the equipment in a cool and ventilated environment to dry.
10.3 Disinfection
The operation of disinfection may cause certain damage to the syringe pump.
You are recommended to disinfect only when it is necessary in your desired
maintenance plan. Clean the equipment before disinfection.
CAUTION
Never use EtO or formaldehyde for disinfection.
Do not conduct high pressure or high temperature disinfection for the syringe
pump and its accessories.
10-2
11 Maintenance
WARNING
The hospital or medical facility using this syringe pump must set up a
comprehensive maintenance plan. Failure to do so may result in equipment failure or
other unexpected consequences, and may even jeopardize personal safety.
Please contact the company immediately if you encounter problems with the
device.
11.1 Inspection
The syringe pump must be given a thorough inspection before use, after 6-12
months of continuous use, and after maintenance or upgrades, to ensure that it is
operating and functioning normally.
If there are any forms of damage or abnormal circumstances, do not use the
syringe pump and contact the company immediately.
11-1
Inspection/Maintenance Items Frequency
Preventive maintenance (refers to the Once every two years, or when you
Maintenance Manual for pressure suspect the occlusion alarm is abnormal,
calibration, sensor calibration, and the flow volume is inaccurate, or the
pump inspection). syringe is incorrectly identified.
11-2
restart calibration], then you need to calibrate again.
NOTE
The full dimension of the 50/60 ml syringe is the distance from 0-50 ml.
11-3
12 Accessories
WARNING
Use the accessories specified in this chapter only. Other accessories may
cause damage to this syringe pump, or cannot reach the specification in this manual.
Please do not replace an accessory if its package or itself is damaged.
Materials PN
009-002755-00
Power cord 009-003358-00
(Select PN according to sales area) 009-003651-00
009-002756-00
NOTE
Safety
Components Host
Protection against
CF Protected against defibrillation
electric shock
Impurities and liquid
IP23
ingress protection
Explosion protection
Unsuitable
level
NOTE:
I: Type I devices
CF: Class CF applied parts, can be directly used in the heart.
IP23: Protected against solid foreign objects with a diameter no less than 12.5mm
and protected against spraying water.
Unsuitable: The device is unsuitable to be used in environments containing air
mixed with flammable anesthetic gas, oxygen or nitrous oxide.
Portable devices: Can be moved from one place to another by one or more persons
or by other means when the devices are in use or being used.
Portable syringe pump: Used to control the infusion of patients and are devices
which can be carried by the patients continuously.
A-1
A.1.2 Operating Environment
Work environment
Temperature 5-40ºC
Storage environment
Temperature -20 - 60 ºC
AC Power Supply
Voltage 100-240 V~
Frequency 50/60 Hz
Current 0.53-0.28A
Voltage DC 10 V-16 V
Current 2.25-1.5A
Display
A.3.2 Battery
Internal battery
No. of batteries 1
Shutdown delay About 30 mins (new battery, after the first low battery alarm)
Charging time When the pump is off, the charging time is not longer than 6h
A.3.3 Host LED
Host LED
A-3
A.3.5 External Ports
Ports
AC power supply
1 AC power supply port
port
Multifunction 1 multifunction interface, which combines the following interface
interface functions:
DC power input interface
RS232 interface
Nurse call interface
Output impedance 50 Ω
Electric
≤60 W, ≤2 A ≤36 VDC, ≤25 VAC
specification
Parameters Specifications
single use.
Compatible syringe
5 ml, 10 ml, 20 ml, 30 ml, 50 ml /60 ml
sizes (ml)
5ml syringe: 0.10-100ml/h
10ml syringe: 0.10-200ml/h
1-11, respectively are (150, 225,300, 375, 450, 525, 600, 750,
Time near end when the time is <10min, step for 1min, and step for 5 min when
A-5
Bed No. 1-999
Volume 1-8
Contrast 1-8
Time:_ _:_ _
time
Time format:12h, 24h
Status indicators Stop, infusion, bolus, KVO, pause, standby, alarm and purge
Dose of single fault About 1.66ml
A-6
A.5 A Reference Table Showing Occlusion Alarm
Delay and Possible Dose
Occlusion Time of
Rate Bolus
pressure occlusion alarm
(ml/h) (ml)
(Level)
(hh:mm:ss)
5 00:14:52 0.097
1 01:54:45 0.086
Level 6
5 00:20:36 0.080
5 00:32:41 0.204
NOTE
Test conditions:
FLUKE IDA4 PLUS tester
Syringe brand: Jierui
Syringe specification: 50ml
Test temperature: 20±2℃
Extension tube length: 1 meter
Occlusion alarm pressure, alarm delays and bolus volume may vary depending
on test conditions, temperature and tube length.
The above data are only typical values under normal test conditions. The actual
data may vary as test conditions change. Please refer to the test data for the product you
have purchased.Under the same standard occlusion value and rate, the higher the value
of the tested pressure is, the longer the alarm time will be delayed.
A-7
A.6 Infusion Accuracy Curve
The following typical infusion accuracy table expresses performance after infusion
has started and infusion fluctuations occurring within a certain period of time after
normal infusion flow volumes have been reached. The infusion accuracy table is
for reference only; detailed infusion accuracy curve is in accordance with the final
device.
NOTE
A-8
Syringe brand: Jierui
Syringe specification: 50ml
Sampling quantity of pump: 3
Sampling quantity of syringe: 3
Sampling rate: 1ml/h
Sampling interval: △ t =0.5 min
Observation windows: p△ t = 2, 5, 11, 19, 31 mins
Maximum deviation over the course of a full observation window: EPmax (%)
Minimum deviation over the course of a full observation window: EPmin (%)
Average deviation: A (%)
Maximum deviation over the course of a full observation window: EPmax (%)
Minimum deviation over the course of a full observation window: EPmin (%)
Average deviation: A (%)
A-9
B EMC and Radio Regulatory Compliance
B.1 EMC
This pump complies with EMC standard IEC60601-1-2 and IEC60601-2-24.
NOTE
This pump may not be used in close proximity to or stacked with other
equipment. If necessary, closely observe the pump to ensure that it is able to operate
normally in its environment.
The syringe pump should not be used at the same time as any MRI (Magnetic
Resonance Imaging) or other similar equipment in order to avoid the possibility of the
pump malfunctioning or crashing due to electromagnetic interference.
User shall install and use the device according to the EMC information of
random file.
B-1
Guidance and statements regarding electromagnetic emissions
This pump should be used in the electromagnetic environment for which it was
designed. The customer or user should ensure that the pump is used in an
electromagnetic environment that complies with the following conditions.
Electromagnetic
Immunity test IEC60601 test level Standard
environment - guidance
Electrostatic ±8 kV contact ±8 kV contact Flooring must be wood,
discharge discharge discharge concrete or ceramic tile. If
(ESD) ±15 kV air ±15 kV air the floor is lined with
IEC 61000-4-2 discharge discharge synthetic materials there
must be a relative
humidity of at least 30%.
Electrical fast ±2 kV power cord ±2 kV power cord The network power source
transient ±1 kV I/O cable must be of typical
(EFT) commercial or hospital
IEC 61000-4-4 quality.
Surge ±1 kV differential ±1 kV differential
IEC 61000-4-5 mode mode
±2 kV common ±2 kV common
mode mode
Voltage drops, <5% UT (drop > <5% UT (drop > The network power source
short 95% UT) 0.5 cycle 95% UT) 0.5 cycle must be of typical
interruptions commercial or hospital
and changes 40% UT (drop 60% 40% UT (drop 60% quality. If the pump needs
IEC 61000-4-11 UT) 5 cycles UT) 5 cycles to run continuously, we
recommend using an
70% UT (drop 30% 70% UT (drop 30% uninterruptible power
UT) 25 cycles UT) 25 cycles supply (UPS) in case of
interruptions to the main
<5% UT (drop > <5% UT (drop > power supply.
95% UT) 5 seconds 95% UT) 5 seconds
Power 400 A/m 400 A/m, 50/60Hz The power frequency
frequency magnetic field must be at
magnetic field a typical level for typical
(50/60 Hz) commercial or hospital
IEC 61000-4-8 environments.
NOTE: UT refers to the voltage of the AC power network before voltage testing.
B-3
Guidance and statements regarding electromagnetic immunity
This pump should be used in the electromagnetic environment for which it was designed.
The customer or user should ensure that the pump is used in an electromagnetic
environment that complies with the following conditions.
d 0.4 P
symbols:
Note 1: Between 80 MHz–800 MHz, use a formula for higher frequency bands.
Note 2: The above guidance is not suitable for use in all conditions. Material structures,
objects and people can absorb and reflect electromagnetic waves, affecting
electromagnetic propagation.
B-4
a. 150k ~ 80MHz ISM (Industrial Scientific Medical) Band refers to 6.765MHz ~
6.795MHz, 13.553MHz ~ 13.567MHz, 26.957 MHz ~ 27.283MHz, 40.66MHz ~ 40.70
MHz.
b. Coincidentdet level within the frequency band range of 150kHz ~ 80MHz and
frequency range of 80MHz~2.5GHz of ISM (Industrial Scientific Medical), is intented to
reduce the possibility of interference caused by portable and mobile radio frequency
communications devices being taken to patient areas accidentally. Therefore, accessory
factor 10/3 is used in frequency range of transmitter corresponding to recommended
calculation formulas for isolation distance.
c. Field strengths for radio (honeycomb and wireless) handset base stations and
terrestrial mobile radio receiving apparatus, antenna reception apparatus, and FM and
AM radio/television broadcasts cannot be accurately estimated using a purely theoretical
approach.
The use of electromagnetic field measuring methods should be considered when
estimating the electromagnetic environment produced by a fixed radio frequency
transmitter. If the measured field strengths in the environment in which this pump is used
exceed stipulated RF levels, the pump must be observed to check whether normal
operation is possible. If any abnormal performance is observed, action must be taken
immediately: for example, by changing the position of the pump or moving it to another
environment.
d. In a frequency range of 150 kHz–80 MHz, the field strength should be less than 3 V/m.
B-5
It is recommended that the pump be kept away from
portable/mobile RF communications equipment
The pump can be used in electromagnetic environments in which RF interference is
controlled. In order to avoid electromagnetic interference, the customer or user should
maintain the minimum recommended distance between the pump and portable/mobile RF
communication equipment. The following recommended isolation distances are calculated
on the basis of the maximum output power of the communications equipment.
Calculate isolation distances on the
power (W)
100 4 2 4
If the rated maximum output power of the transmitter is not included above, the isolation
distance may be estimated by using the formula in the corresponding column. In the
formula, P is the rated maximum output power of the transmitter as provided by the
manufacturer, in W.
Note 1: Between 80M-800MHz, use a formula for higher frequency bands.
Note 2: 150k~80MHz ISM (Industrial Scientific Medical) Band refers to 6.765MHz~
6.795MHz, 13.553MHz~13.567MHz, 26.957 MHz~27.283MHz, 40.66MHz~40.70 MHz.
Note 3: Accessory factor 10/3 is used in transmitter recommended calculation formulas for
isolation distance within the frequency band range of 150kHz~80MHz and frequency
range of 80MHz~2.5GHz of ISM (Industrial Scientific Medical), to reduce the possibility of
interference caused by portable and mobile radio frequency communications devices being
taken to patient areas accidentally.
Note 4: The above guidance is not suitable for use in all conditions. Material structures,
objects and people can absorb and reflect electromagnetic waves, affecting
electromagnetic propagation.
B-6
B.2 Radio Regulatory Compliance
RF Parameter
Radio devices 2.4GHz Wi-Fi devices
Operating frequency 2412MHz to 2472MHz
Modulation mode DSSS and OFDM
Output power ≤20dBm
The radio device used in this product is in compliance with the essential
requirements and other relevant provisions of Directive 2014/53/EU.
B-7
C Default Factory Settings
This chapter presents some default factory settings.User cannot change the
factory default, but may restore to the default factory settings when it is necessary.
C.1 Alarms
Volume 4
C.2 Interface
UI Factory Default
Contrast 4
C.3 Parameters
Domestic: Jierui
Commonly used syringes
International: B-D
C-1
C.4 System Time
Time 00:00
Date 01/01/2013
Domestic: yyyy-mm-dd
Date format
International: dd-mm-yyyy
2 Amiodarone 22 Mannitol
Dexmedetomidine
4 24 Metoprolol Tartrate Tablets
Hydrochloride Injection
5 Diazepam 25 Midazolam
6 Diazoxide 26 Morphine
7 Digoxin 27 Nicardipine
8 Diltiazem 28 Nimodipine
11 Epinephrine 31 Pancuronium
12 Esmolol 32 Phentolamine
15 Furosemide 35 Propofol
C-2
D Alarm Information
This chapter presents the alarm information of the syringe pump.Prompt
information for operation guidance will not be presented in this chapter.
The table shows the appropriate countermeasures for each piece of information
related to alarm triggering. If the problem still exists after operating according to the
countermeasures, please contact the company.
Alarm Alarm
Reason Countermeasure
Information level
Syringe is disengaged
[Syringe
High during the operation of
disengaged] Press to cancel alarm.
syringe pump.
D-1
Alarm Alarm
Reason Countermeasure
Information level
The syringe pump performs
no operation during the set Operate the pump to cancel
[Reminder] Low
reminder time after the alarm.
syringe is installed.
After inserted to
BeneFusion DS5 Infusion 1. Modify pump serial number
Supervision System, the and ensure different serial
[Pump serial pump (no Wi-Fi) has the numbers in BeneFusion DS5
Low
number conflict] same serial number as Infusion Supervision System.
other pump. The alarm is 2. Pull out the pumps with the
triggered by pumps with the conflicting serial number.
conflicting serial number.
Not insert the syringe or
[Syringe not syringe is not correctly
Low Correctly insert the syringe.
correctly inserted] inserted, and starts
infusion.
D-2
Alarm Alarm
Reason Countermeasure
Information level
Communication is
abnormally interrupted for 3
minutes (due to pump been
pulled out or internal
abnormality) after pump
inserted to BeneFusion
DS5 Infusion Supervision
System and communicated
successfully with the Press or restore the
system (no pump serial communication between
[Communication number conflict or version syringe pump(s) and
Low
is interrupted] number compatible) or BeneFusion DS5 Infusion
WLAN interruption for 3 Supervision System or
minutes after pump BeneFusion CS5 Infusion
communicated successfully Supervision System.
with BeneFusion CS5
Infusion Supervision
System. After the alarm is
triggered, infusion of the
pump will not be influenced,
and the pump continues
infusion.
D-3
Alarm Alarm
Reason Countermeasure
Information level
is not in parameter setting
interface, or not set valid
parameters for starting
infusion in parameter
setting interface. The alarm
is triggered by the relayed
pump.
disconnection]
Not connected to the alarm automatically.
network power source.
NOTE
Abbreviation Meaning
A ampere
℃ centigrade
cm centimeter
dB decibel
g gram
hr hour
Hz hertz
inch inch
k kilo
kg kilogram
kPa kilopascal
l litre
lb pound
m meter
mg milligrams
min minute
ml milliliter
mm millimeters
s second
μg Microgram
V volt
VA volt ampere
W watt
E-1
E.2 List of Symbols
Symbols Meaning
minus
-
% percent
/ Per; divide; or
~ to
^ power
plus
+
= equal to
plus or minus
±
× multiply
© copyright
AC Alternating current
Anti-Bolus Anti-Bolus
BOLUS Bolus
CE Conformité Européenne
E-2
Abbreviation Meaning
EtO C2H4O
ID Identification
Max Maximum
Min Minimum
OR operating room
SN Series Number
kPa 1kPa=7.5mmHg=0.145psi=0.01bar
psi 1psi=51.724mmHg=6.897kPa=0.069bar
bar 1bar=750mmHg=14.5psi=100kPa
lb 1 lb=0.454kg
E-3
F Toxic and Hazardous Substances or
Elements
○ ○ ○ ○ ○ ○
Back housing
Device
○ ○ ○ ○ ○ ○
Keys
housing
○ ○ ○ ○ ○ ○
Facing
○ ○ ○ ○ ○ ○
Labels
○ ○ ○ ○ ○ ○
Display Display
○ ○ ○ ○ ○ ○
Host hardware
○ ○ ○ ○ ○ ○
Host Internal cables
○ ○ ○ ○ ○ ○
PCBA
Cartons (K=K ○ ○ ○ ○ ○ ○
crimp paper)
Foam packages ○ ○ ○ ○ ○ ○
Packaging
(EPE)
○ ○ ○ ○ ○ ○
Plastic bag (PE)
Connecting pieces ○ ○ ○ ○ ○ ○
○ ○ ○ ○ ○ ○
Accessories Accessories
F-1
G Declaration of Conformity
G-1
P/N: 046-005902-00(13.0)