Benefusion Sp5: (This Manual Is Also Applicable To Benefusion Sp5 Ex, Benefusion Sp5 Tiva Syringe Pumps)

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 127

BeneFusion SP5

(This manual is also applicable to BeneFusion SP5 Ex, BeneFusion SP5


TIVA Syringe Pumps)

Syringe Pump

Operator’s Manual
© Copyright 2012-2017 Shenzhen Mindray Scientific Co., Ltd. All rights reserved.

For this Operator’s Manual, the issue date is 2017-10.

Version: 13.0

I
Intellectual Property Statement
SHENZHEN MINDRAY SCIENTIFIC CO., LTD. (hereinafter called Mindray
Scientific) owns the intellectual property rights to this product and this manual. This
manual may refer to information protected by copyright or patents and does not
convey any license under the patent rights or copyright of Mindray Scientific, nor
the rights of others.

Mindray Scientific intends to maintain the contents of this manual as confidential


information. Disclosure of the information in this manual in any manner whatsoever
without the written permission of Mindray Scientific is strictly forbidden.

Release, amendment, reproduction, distribution, rental, adaptation, translation of


this manual in any manner whatsoever without the written permission of Mindray
Scientific is strictly forbidden.

, , , and are registered


trademarks or trademarks owned by SHENZHEN MINDRAY BIO-MEDICAL
ELECTRONICS CO.,LTD. (hereinafter called Mindray) in China and other
countries. Mindray Scientific is authorized by Mindray to use above registered
trademarks or trademarks.

Responsibility on the Manufacturer


Party
Contents of this manual are subject to changes without prior notice.

All information contained in this manual is believed to be correct. Mindray Scientific


shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance, or use of this manual.

Mindray Scientific is responsible for the effects on safety, reliability and


performance of this product, only if:
 All installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray Scientific authorized personnel;
 The electrical installation of the relevant room complies with the applicable national
and local requirements;
 The product is used in accordance with the instructions for use.

II
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.

Exemptions
Mindray Scientific's obligation or liability under this warranty does not include any
transportation or other charges or liability for direct, indirect or consequential
damages or delay resulting from the improper use or application of the product
or the use of parts or accessories not approved by Mindray Scientific or repairs
by people other than Mindray Scientific authorized personnel.

This warranty shall not extend to:


 Malfunction or damage caused by improper use or man-made failure.
 Malfunction or damage caused by unstable or out-of-range power input.
 Malfunction or damage caused by force majeure such as fire and earthquake.

 Malfunction or damage caused by improper operation or repair by unqualified


orunauthorized service people.
 Malfunction of the instrument or part whose serial number is not legible enough.
 Others not caused by instrument or part itself.

Company Contact
Manufacturer: Shenzhen Mindray Scientific Co., Ltd.
Bldg 2,5, Mindray Guangming facility, 1203 Nanhuan
Address: Avenue, Yutang Block, Guangming District, 518016,
Shenzhen, P.R.China
Website: www.skmedica.com
Tel: +86 755 81888998
Fax: +86 755 26582680

EC-Representative: Shanghai International Holding Corp. GmbH (Europe)


Address: Eiffestraße 80, 20537 Hamburg, Germany
Tel: 0049-40-2513175
Fax: 0049-40-255726

III
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely and
in accordance with its function and intended use. Observance of this manual is a
prerequisite for proper product performance and correct operation and ensures
patient and operator safety.

This manual is based on the maximum configuration and therefore some contents
may not apply to your product. If you have any question, please contact us.

This manual is an integral part of the product. It should always be kept close to
the equipment so that it can be obtained conveniently when needed.

Intended Audience
This manual is geared for clinical professionals who are expected to have a
working knowledge of medical procedures, practices and terminology as required
for monitoring of critically ill patients.

Illustrations
All illustrations in this manual serve as examples only. They may not
necessarily reflect the setup or data displayed on your equipment.

Conventions
 Italic text is used in this manual to quote the referenced chapters or sections.
 [ ] is used to enclose screen texts.
 < > is used to enclose the keys.
 → is used to indicate operational procedures.

IV
Contents
1 Safety ..................................................................................................................... 1-1
1.1 Safety Information ............................................................................................... 1-1
1.1.1 Dangers ................................................................................................... 1-1
1.1.2 WARNING ............................................................................................... 1-2
1.1.3 CAUTION ................................................................................................ 1-3
1.1.4 NOTE ....................................................................................................... 1-4
1.2 Equipment Symbols ............................................................................................ 1-5

2 Overview ................................................................................................................ 2-1


2.1 Description ........................................................................................................... 2-1
2.1.1 Indications for Use................................................................................... 2-1
2.1.2 Contraindications ..................................................................................... 2-1
2.1.3 Appearance, Parts and Features ............................................................ 2-1
2.2 Host ..................................................................................................................... 2-2
2.2.1 Front View ............................................................................................... 2-2
2.2.2 Rear View ................................................................................................ 2-4
2.2.3 Side View ................................................................................................. 2-4
2.2.4 Side View with the Door Opened ............................................................ 2-5
2.2.5 Bottom View ............................................................................................ 2-5
2.3 Screen Display .................................................................................................... 2-6
2.4 Cursor .................................................................................................................. 2-6

3 Installation and Setting ........................................................................................ 3-1


3.1 Installation ........................................................................................................... 3-1
3.1.1 Out of Box Audit (OOBA) ........................................................................ 3-1
3.1.2 Operating Conditions .............................................................................. 3-2
3.1.3 Mount the Clamp ..................................................................................... 3-3
3.1.4 Multi-channel Pumps Combination ......................................................... 3-4
3.1.5 Mount the Handle Set ............................................................................. 3-4
3.1.6 Connect the AC Power Source ............................................................... 3-5
3.2 Conventional Settings ......................................................................................... 3-6
3.2.1 Set Language .......................................................................................... 3-6
3.2.2 Adjust Screen Contrast ........................................................................... 3-6
3.2.3 Set Date and Time................................................................................... 3-6
3.2.4 Adjust Volume .......................................................................................... 3-7
3.3 Restore Factory Default ...................................................................................... 3-7

4 Basic Operation .................................................................................................... 4-1


4.1 Infusion Flow Chart ............................................................................................. 4-1
4.2 Operational Procedures ...................................................................................... 4-2
4.2.1 Turn on the Pump .................................................................................... 4-2

1
4.2.2 Insert the Syringe .................................................................................... 4-2
4.2.3 Change the Syringe................................................................................. 4-4
4.2.4 Select Syringe Brands ............................................................................. 4-4
4.2.5 Memory Function ..................................................................................... 4-5
4.2.6 Select Infusion Mode ............................................................................... 4-5
4.2.7 Purge ....................................................................................................... 4-5
4.2.8 Set Infusion Parameters .......................................................................... 4-6
4.2.9 Infusion .................................................................................................... 4-7
4.2.10 Infusion Pause ....................................................................................... 4-7
4.2.11 BOLUS ................................................................................................... 4-8
4.2.12 Change the Rate during Operation ....................................................... 4-8
4.2.13 Complete ............................................................................................... 4-8
4.2.14 Standby ................................................................................................. 4-9
4.2.15 Turn off the Pump .................................................................................. 4-9

5 Infusion Mode ....................................................................................................... 5-1


5.1 Rate Mode ........................................................................................................... 5-1
5.2 Time Mode ........................................................................................................... 5-1
5.3 Body Weight (BW) Mode ..................................................................................... 5-2
5.4 Intermittent Administration Mode ......................................................................... 5-3
5.5 TIVA Mode (Optional) .......................................................................................... 5-4
5.6 Sequential Mode (Optional) ................................................................................ 5-5
5.7 Loading Dose Mode (Optional) ........................................................................... 5-6
5.8 Ramp Up/Down Mode (Optional) ........................................................................ 5-7
5.9 Micro-infusion Mode (Optional) ........................................................................... 5-8

6 Setting Parameters ............................................................................................... 6-1


6.1 KVO ..................................................................................................................... 6-1
6.2 Drug Library ......................................................................................................... 6-1
6.3 Occlusion Pressure (Unit) ................................................................................... 6-1
6.3.1 Set Occlusion Pressure ........................................................................... 6-1
6.3.2 Set Pressure Unit .................................................................................... 6-2
6.3.3 Dynamic Pressure Scanning (DPS) ........................................................ 6-2
6.3.4 Automatic Pressure Release Function (Anti-Bolus) ................................ 6-2
6.4 Key Lock Function ............................................................................................... 6-3
6.5 Reminder Function .............................................................................................. 6-3
6.6 Time Near End ..................................................................................................... 6-3
6.7 Common Syringe Brands .................................................................................... 6-4
6.8 Bed No. Settings.................................................................................................. 6-4
6.9 View Department ................................................................................................. 6-4

7 Other Functions .................................................................................................... 7-1


7.1 History Record ..................................................................................................... 7-1
7.2 Power-down Save ............................................................................................... 7-1

2
7.3 Nurse Call ............................................................................................................ 7-1
7.4 Wireless Networking (Optional) ........................................................................... 7-2
7.5 Data Export .......................................................................................................... 7-3
7.6 Automatic Multi-channel Relay ............................................................................ 7-3
7.7 Patient Information .............................................................................................. 7-4
7.8 Night Mode .......................................................................................................... 7-4
7.9 Day's Prescription ................................................................................................ 7-5
7.10 WLAN Setting .................................................................................................... 7-5

8 Alarms .................................................................................................................... 8-1


8.1 Alarm Level .......................................................................................................... 8-1
8.2 Alarm Types ......................................................................................................... 8-1
8.2.1 Multi-level Alarm Rules ............................................................................ 8-2
8.3 Alarm Handling Rules .......................................................................................... 8-2
8.4 Alarm Countermeasures ...................................................................................... 8-3

9 Battery ................................................................................................................... 9-1


9.1 Battery Performance Optimization ...................................................................... 9-2
9.2 Check the Battery ................................................................................................ 9-2
9.3 Insert Batteries .................................................................................................... 9-3
9.4 Battery Recycling................................................................................................. 9-3

10 Preservation and Sanitation ........................................................................... 10-1


10.1 Description ...................................................................................................... 10-1
10.2 Cleaning ......................................................................................................... 10-1
10.3 Disinfection ..................................................................................................... 10-2

11 Maintenance ...................................................................................................... 11-1


11.1 Inspection......................................................................................................... 11-1
11.2 Maintenance Plan ............................................................................................ 11-1
11.3 View Information .............................................................................................. 11-2
11.4 Syringe Calibration .......................................................................................... 11-2
11.5 Safe Disposal and Recycling ........................................................................... 11-3

12 Accessories ...................................................................................................... 12-1

A Product Specifications ....................................................................................... A-1


A.1 Safety Specifications .......................................................................................... A-1
A.1.1 Product Classification ............................................................................ A-1
A.1.2 Operating Environment .......................................................................... A-2
A.2 Physical Specifications ...................................................................................... A-2
A.3 Hardware Specifications .................................................................................... A-3
A.3.1 Display .................................................................................................... A-3
A.3.2 Battery .................................................................................................... A-3

3
A.3.3 Host LED ................................................................................................ A-3
A.3.4 Auditory Indicator ................................................................................... A-3
A.3.5 External Ports ......................................................................................... A-4
A.3.6 Signal Output Interface .......................................................................... A-4
A.4 Specifications ..................................................................................................... A-5
A.5 A Reference Table Showing Occlusion Alarm Delay and Possible Dose .......... A-7
A.6 Infusion Accuracy Curve .................................................................................... A-8
A.7 Trumpet Curve ................................................................................................... A-8

B EMC and Radio Regulatory Compliance .......................................................... B-1


B.1 EMC ................................................................................................................... B-1
B.2 Radio Regulatory Compliance ........................................................................... B-7

C Default Factory Settings .................................................................................... C-1


C.1 Alarms ................................................................................................................ C-1
C.2 Interface ............................................................................................................. C-1
C.3 Parameters ........................................................................................................ C-1
C.4 System Time ...................................................................................................... C-2
C.5 Drug Library List................................................................................................. C-2

D Alarm Information ............................................................................................... D-1

E Symbols and Terms ............................................................................................ E-1


E.1 List of Units ........................................................................................................ E-1
E.2 List of Symbols ................................................................................................... E-2
E.3 List of Terms ....................................................................................................... E-2
E.4 List of Unit Conversion ....................................................................................... E-3

F Toxic and Hazardous Substances or Elements ............................................... F-1

G Declaration of Conformity ................................................................................. G-1

4
1 Safety
1.1 Safety Information
The safety statements presented in this chapter refer to basic safety information
that the operator must pay attention to and abide by when using the syringe
pump.There are additional safety statements in other chapters or sections, which
may be the same as or similar to the following, or specific to particular operations.

Dangers

 Indicates an imminent hazard that, if not avoided, could result in death,


serious injury or damage to product/property.

WARNING

 Indicates a potential hazard or unsafe practice that, if not avoided, could


result in death, serious injury or damage to product/property.

CAUTION

 Indicates a potential hazard or unsafe practice that, if not avoided, could


result in minor personal injury, product malfunction or damage to product/property.

NOTE

 Provides application tips or other useful information to ensure that you get
the most out of the product.

1.1.1 Dangers
This Manual does not contain any information at the "Danger" level.

1-1
1.1.2 WARNING

WARNING

 Device, cables and accessories must be inspected before use to guarantee


their normal and safe operation.

 This equipment can only be connected to the socket with ground protection.
Please adopt a rechargeable battery instead of the socket as the power supply if the
socket is not provided with a ground lead.

 To prevent fire or explosion, do not operate the system in the presence of


anesthetic, flammable or explosive materials.

 Do not open the equipment casing as there is the impending danger of


electric shock. Equipment maintenance and upgrades must be carried out by
maintenance technicians whom are trained and licensed by the manufacturer. Moreover,
the process must be done only after the AC power supply is disconnected. Maintenance
carried out by individuals non-affiliated to the manufacturer or by non-licensed personnel
may affect the safety, performance and function of the product.

 When used with electrosurgery equipment, the safety of patients should be


ensured.

 The patient's clinical condition and the working condition of the syringe
pump must be monitored carefully, and the alarm volume and alarm levels need to be set
according to the actual needs. Operation and performance relying solely on the auditory
alarm system alone is not sufficient, and setting the alarm at a low volume may endanger
the patient. If the alarm volume is less than the surroundings volume, which can further
lead to operators identify alarm mistakenly.

 Please carefully install the power line and cables with various accessories to
prevent the patient from choking or suffocation caused by entanglement of the cables or
by electrical disturbance.

 The packaging materials must be disposed of in compliance with local laws


and regulations or the hospital policy on waste management.They must be kept out of the
reach of children.

 Infusion cannula knots, filter coagulation and occlusions arising from needle
insertion can cause the pressure inside the syringe to rise during infusion.When this
occurs, removing the occlusion can cause excessive liquid to be infused into the patient,
so appropriate measures should be taken.

 The pump should not be placed more than 100 cm above or below the level of
the patient’s heart. The smaller the height difference between the pump, the more accurate the
pressure test in the infusion cannula will

1-2
be.

 This equipment has to be used with professional medical consumables, and


its accuracy cannot be guaranteed when it is used with a syringe which is a non-standard
consumable or a consumable without calibration, please contact the company for
calibration service.

 Do not touch the patient when connecting the peripheral equipment via the
input/output signal ports to prevent patient leakage current from exceeding the
requirements specified by the standard.

 In the process of defibrillation, do not touch patient and other non-


defibrillation equipments to prevent electric shock damage, and defibrillation will not
affect the basic performance (such as infusion accuracy, alarm and signal transmission)
of the pump.

1.1.3 CAUTION

CAUTION

 Use the accessories specified in this Operator’s Manual to guarantee the


patient’s safety.

 When equipment and the accessories exceed their recommended service


life, they must be disposed of in accordance with local statutes or hospital regulations.

 Electromagnetic fields may affect equipment performance. This makes it


necessary for other equipment used in the vicinity of the pump to meet EMC standards.
Mobile phones, X ray and MRI equipment are all potential interference sources because of
their high-intensity electromagnetic radiation.

 Before the equipment is connected to the power supply, check that the
voltage and frequency of the power supply match the specifications on the label or in this
Operator’s Manual.

 Please install and carry the equipment correctly to protect the equipment
from damage from drops, impacts, violent shaking or other external mechanical forces.

 Disposable accessories must be disposed of after use in accordance with


the relevant hospital regulations.
 Avoid direct sunshine, high temperatures and dampness.

 Check the built-in battery before use to make sure it has sufficient
power.Recharge the battery if necessary.

1-3
 The syringe with the luer taper is recommended for use, which can
effectively prevent patients from under current caused by the occurrence of the cannula
to slip out when under tension.

1.1.4 NOTE

NOTE

 Install the equipment in a position where it can be easily accessed for


inspection, operation and maintenance.

 Keep this Operator’s Manual near to the equipment for future ease of
reference.

 The software of the equipment is developed according to the software


development demands of IEC60601-1 standard, which can minimize the possibility of the
risk caused by program error.

 This Operator’s Manual describes the most complete functional


configuration of the system. The product you are using may not have some of the settings
or functions described herein.

 Do not insert devices that are not specified by the manufacturer into the
multifunction interfaces.

 During infusion, the syringe pump can accurately control the rate, infusion
volume and infusion time, and monitor the operation in real-time, to effectively prevent
over currents, under currents and instances of backflow.

 The device is not in touch with the drugs or patients directly. Thus, there is
no need to process Biocompatibility test on it.

1-4
1.2 Equipment Symbols
The equipment you purchased may not provide you with all the following symbols.

NOTE!
Refer to the
ON/OFF
accompanying document
(This Manual)

Alternating current power


Battery
supply (AC)

Alarms AUDIO PAUSED

Clear/Back Start

Bolus Confirm

Stop Menu

Move up/Increase Move down/Decrease

Move left Move right

Syringe Selected drug

Night mode Equipotential port

Wireless networking, Wireless networking,


normal communication no communication with
with BeneFusion CS5 BeneFusion CS5
Infusion Supervision Infusion Supervision
System System

Current pump
communicated
Wireless network is
successfully with
establishing a network
BeneFusion DS5
connection
Infusion Supervision
System

Current pump has Current pump has


downward relayed No.X forward relayed No.X
pump pump
Executed prescription Pause prescription

1-5
Non-executed
Executing prescription
prescription

Caution Lock

Protected against solid


foreign objects with a
Protected against
diameter no less than
IP23 defibrillation CF
12.5mm and protected
applied parts
against spraying liquid
water

Date of manufacture Manufacturer

This side up Keep away from rain

Stacking limit by
Fragile, handle with care
number

Electronic equipment:
dispose of separately to
Wireless transceiver
avoid polluting the
environment

Multifunction interface Serial number

Recycle CE mark

Environmentally-friendly
The European Union
use periods of electronic
Representative Office
products (20 years)

Package shall be kept


Package shall be kept
between 10%–95%
between 50–106 kPa
humidity during
during transport
transport

Package shall be kept between -20–60°C during transport


1-6
2 Overview
2.1 Description
2.1.1 Indications for Use
The Syringe Pump is used in conjunction with the syringe to control the dose
of liquid infused into the patient’s body.

The Syringe Pump is suitable for adults, children and newborns in clinical
departments.

This Syringe Pump is expected to be used in institutes or units with healthcare


capabilities. This includes but is not limited to: outpatient departments, emergency
departments, wards, ICU, operating rooms, observation rooms, clinics, and
nursing hospital.

WARNING

 The syringe pump is for clinical use. It must only be used under appropriate
conditions by professional clinicians, medical device technicians, or by suitably trained
nurses. Personnel using this product must receive sufficient training.This product must
not be operated by anyone who has not been authorized to do so or has not received
suitable training.

2.1.2 Contraindications
None

2.1.3 Appearance, Parts and Features


The Syringe Pump primarily consists of a housing, a motor drive system, an input
system, a memory system, a control system, a display system, a sensor-based
monitor system and an alarm system. By precisely controlling the rotational speed
of the stepping motor, the screw rod is driven to run at the set speed, the dose of
liquid infused into the patient’s body by syringe can be controlled, the syringe pump
can be used for precise and continuous infusion of liquids , and all components of
are suitable for use in patient environment. Optional wireless modules. Optional
functions of the software comprise Rate Mode, Time Mode, Body Weight Mode,
TIVA Mode, Ramp Up/Down Mode, Sequential Mode, Loading Dose Mode, Micro-
infusion Mode, Intermittent Administration Mode, Drug Library, Record, and

2-1
Anti-bolus Function.

Since some parts and functions are optional, the Syringe Pump you purchased
may not contain these additional parts and their relevant functions.

2.2 Host
2.2.1 Front View

1. Alarm light
The alarm light indicates different alarm levels in different colors and
flash frequencies, please refer to Chapter 8 Alarms for details.

2. Display
Used for displaying infusion parameters and relevant content.

3. <DIRECTION>
Used for adjusting value, change lines and pages.

4. <CLEAR/BACK>
 Under non-setting status, indicate to returns to previous menu or operation.
 Under the setting status, indicate to clear the current set or cancel the edit.

5. <START>
After installing the syringe correctly and completing setting infusion parameters,
press this key to start the infusion.

6. <BOLUS >
 During infusion, press this key to enter the Bolus Settings screen.
 When the pump is stopped, press this key to enter the purge settings screen.

2-2
7. <POWER>
 Used for turning power on, entering in standby state and turning off operations.
 When power off, press and hold (>3 s) the key.

8. AC indicator light
 On: The pump is connected to an AC power supply.
 Off: The pump is not connected to an AC power supply.

9. Clip
Clamps the syringe plunger.

10. Handle
Controls clip opening/closing and slider action.

11. <OK>
Used for confirming input operation and saving value.

12. <STOP>
During infusion, press this key in any non-setting interface to stop infusion.
Infusion stops caused by alarms, such as occlusion and so on, press this key to
cancel the alarm.

13. <AUDIO PAUSED> Pauses alarm sound.

14. <MENU>

interfaces.
Under non-operation status, used for switching [Main Menu] interface and other

 Under operation status, press and hold this key to lock;in locked state, press and
hold to unlock.

15. Battery indicator


 Steady green indicates that the battery is charging.
 Flashing indicates that the battery is providing power.

 Light off indicates that there is no battery or the equipment is turned off and not
connected to an AC power supply.

16. Spindle clamp


Used along with flange plate to secure syringe flange and clamps the syringe.

17. Slider
Secures syringe and drives plunger assembly.

2-3
2.2.2 Rear View

1. Battery box
2. Multi-channel pumps connect rail Used for combining multi-channel pumps.
3. Slider
4. Equipotential port
When other device and syringe pump are used together, please adopt wires
to connect other devices and equipotential port of the pump, to eliminate
ground potential difference between different devices and ensure safety.
5. Alternating current power supply (AC) port
Connected by three-core-type power cord and AC power source.
6. Multifunction interface, which combines the following interface functions:
 
DC power input interface
RS232 interface

Nurse call interface

2.2.3 Side View

1. Speaker hole
2. Multi-channel pumps hasp
The multi-channel pumps are connected to secure the equipment.

2-4
2.2.4 Side View with the Door Opened

1. Warning label
2. Speaker hole
3. Door
4. Spindle clamp
5. Flange plate
6. Clip
7. Handle

2.2.5 Bottom View

1. Product label
2. Fixed mounting holes

2-5
2.3 Screen Display
This syringe pump has a built-in LCD screen. The display information comprises
three main parts:
1

3
1. Title bar
 Display the name of the current screen.
 Display drug information, please refer to 6.2 Drug Library.
 Display alarm information, please refer to Chapter 8 Alarms.
 Display lock icon, please refer to 6.4 Key Lock Function.
 Display wireless icon, please refer to 7.4 Wireless Networking (Optional).
 Display battery icon, please refer to Chapter 9 Battery.

2. Parameter area
Display every parameter and the parameter value of the current screen.

3. Prompt bar
Display prompt messages, run icon and so on.The run icon on the screen displays
the running operation:
Green icon indicates normal running. Arrows move from right to

left, and the running speed increases as the rate is increased.
■ Blue icon lighting indicates bolus.
■ Steady grey icon indicates infusion stops.

2.4 Cursor
In the main screen and parameter settings screen, when the cursor is located at an
option or at a data value, the grounding of the option or the data value will turn to

white and the font will become black. Press or to move cursor up or

down and confirm the location. Press move the cursor up and down and
confirm the location.

Note: Press or to "locate" cursor; Press to "Select".

2-6
3 Installation and Setting
3.1 Installation

WARNING

 Equipment assembly and refit (including correct protective grounding


connection) during life period must be carried out by maintenance technicians whom are
trained and licensed by the manufacturer, and evaluated according to the specified
IEC60601-1. Please contact the company if you have any queries.

 The software copyright for this equipment belongs to the manufacturer.


Unless explicitly authorized, any alteration, reproduction or sale by any means or in any
form by any organization or individual is prohibited.

 All the analog equipment and digital facilities should be certified according
to the specified IEC standard (such as:IEC60950 Information Technology Equipment
Safety and IEC60601-1 Medical Electrical Equipment Safety), moreover, all equipment
should be connected based on the requirements of the valid version of the IEC60601-1
system.The qualified individual responsible for connecting auxiliary equipment to the
input and output signal ports is also accountable for making the system in accordance
with the IEC60601-1 standard. Please contact the company if you have any queries.

 When this equipment combining with other electrical equipments forms a


combination with a special function, and the user cannot determine whether there is an
impending danger from each equipment specification (such as a danger of electric shock
due to aggregation of current leakage), please contact the company or a specialist in the
field at the hospital, to guarantee that all equipment in the combination are safe enough
and will not be damaged.

 Please make sure this equipment is securely fixed and positioned.


Positioning changes and severe shock can lead to minor changes in the delivery
accuracy.

3.1.1 Out of Box Audit (OOBA)


Please check the packing case carefully before opening the box. If there is any
damage, please contact the distributor or manufacturer immediately.

3-1
Please carefully remove the equipment and its accessories from the packaging in a
correct manner, and inspect them against the packing list. Examine the equipment
for any mechanical damage and ensure that the box includes all items on the
packing list. Please contact the company if you have any queries.

NOTE

 Keep the packing case and packaging materials for future transportation or
storage.

WARNING

 They must be kept out of the reach of children. The packaging materials
must be disposed of in compliance with local laws and regulations or the hospital policy
on waste management.

 The equipment may be contaminated by microbes during storage, transport


and use. Please ensure that the package is undamaged before using, do not use if there is
any damage.

3.1.2 Operating Conditions


The operating environment of this syringe pump must meet the requirements
in A.1.2 Operating Environment.

The operating environment should also be appropriately protected from noise,


vibration, dust, and corrosive, inflammable or explosive substances. If installed
inside the equipment case, a sufficient space before and after the equipment case
should be ensured to facilitate operation, maintenance and repairing work. There
should be a 2" (5 cm) gap around the syringe pump to ensure that air can circulate
freely for a better cooling effect.

When the pump is transferred from one place to another, differences in


temperature and humidity can cause condensation to form inside the pump. If this
occurs, do not switch the pump to the "ON" state until the condensation has gone.

WARNING

 Please use only when the operating environment meets the requirements
specified above. Otherwise, the pump's performance will not match the technical
specifications in A Product Specifications. Device failure and other unexpected
consequences may also result.

3-2
3.1.3 Mount the Clamp
1. Align the mounting holes on
the installation dead plate with
the installation screw holes on
the bottom of the machine, and
tighten the screws.

2. Turn counterclockwise to
loosen the pole clamp until an
IV stand can be inserted in.

3. Tighten the pole clamp


clockwise to firmly fix the
device on the IV stand (round
vertical bar with diameter
size of 15-32mm).

3-3
3.1.4 Multi-channel Pumps Combination
A maximum of 2 pumps can be combined (syringe pump and infusion pump can
be cross combined).

When multi-channel pumps are combined, align the rail on the bottom of the upper
pump to the rail slot on the top of the lower pump, slide from backward to
frontward until the alignment is matched, then strain the two pumps by the hasps.
When disconnecting, hold the upper-part of the hasp of the upper pump and pull
out the lower pump.

WARNING
 The syringe pump must be installed horizontally.

NOTE
 Please ensure the stand is stable before installation.

 When the pumps are overlapped, all fixing clamps except the fixing clamps
of the lowest pump should be removed, otherwise, the pumps can be used overlapped.

3.1.5 Mount the Handle Set


Align the rail on both sides of the handle set to the rail slot on the top of the lower
pump; slide from backward to frontward until the alignment is matched. When
disconnecting, toggle the button of the handle set outward and pull out the
handle set or pump.

3-4
Press the button of pole clamp, horizontally or vertically adjust pole clamp, turn the
handle, pump can be fixed to cross bar or vertical bar.
 Square tube: Width 10mm, length 15-40mm
 Round tube: External diameter ø15~ ø40

NOTE

 Only 2 pumps (syringe pump and infusion pump can be cross combined) can
be carried by the handle set, or the handle of the handle set might be broken. Please refer
to 3.1.4 Multi-channel Pumps Combination for the combination operation of multi-pump.

3.1.6 Connect the AC Power Source


1. Please confirm to use the original three-core-type power cord.
2. Insert one end of the power line into the AC socket on the back panel of the syringe
pump.
3. Insert the other end of the power line into the matched three-plug connector
connecting to the AC power.

The earthing wire in the three-plug connector should be grounded, if there is a


doubt whether the AC power system is grounded or not, please adopt the built-in
battery and contact an electrical technician at the hospital or the company.

3-5
WARNING

 Do not touch the power plug with wet or moist hands! If there is a liquid drug
or residue on or around the power socket or plug, the user should completely clean and
dry the area before plugging into the power supply, or accidents or injuries may result!

NOTE
 Compatible power supply: 100–240 V~, 50/60 Hz.

 The AC power cable should be correctly inserted and secured into the
socket.

 Removing power cord is to disconnect equipment from power supply. Please


ensure suitable clearance around the device to facilitate connect and remove power cord.

3.2 Conventional Settings


This chapter only introduces the general settings for the syringe pump, please refer
to other relative chapters for parameters and other feature settings.

3.2.1 Set Language


1. Select [Main Menu]→[System Option]→[Language].
2. Select [Language] from the [Language] according to actual needs.

3.2.2 Adjust Screen Contrast


1. Select [Main Menu]→[System Option]→[Contrast].
2. Select [Contrast]: 1-8. 8 for the brightest setting, and 1 for the darkest setting.When
operating on battery power, you can set a low Contrast to save the power of the battery.

3.2.3 Set Date and Time


1. Select [Main Menu]→[System Option]→[System Date and Time].
2. Set [Time] and [Date].

3-6
3. Select [Time Format]: [24h] or [12h].
4. Select [Date Format]: [yyyy-mm-dd], [mm-dd-yyyy] or [dd-mm-yyyy].

CAUTION

 Please check the system date and time to keep accurate records in the
History function.

 After changing the time format or date format, the record will update new
format automatically.

 For syringe pump inserted to BeneFusion DS5 Infusion Supervision System,


time and date can’t be changed.

3.2.4 Adjust Volume


1. Select [Main Menu]→[System Option]→[Volume].
2. Select [Volume]: 1-8. 1 for the lowest volume; 8 for the highest volume.

3.3 Restore Factory Default


During operation, you may change some settings in some situations. However, the
changes may not be appropriate or correct, especially when patient or syringe
brands are changed.Therefore, you should restore the system to the default factory
settings during operation according to actual needs, to guarantee that each
configuration of the syringe pump is applicable for clinical use.For some default
factory settings of this equipment, please refer to C Default Factory Settings.

1. Select [Main Menu]→[System Maintenance]→Input User Maintenance


Password→[Restore factory default].
2. Prompt screen [Restore factory default or not?].

3. Press to confirm, some parameters of General Option, System Option

and System Maintenance will be restored to default values; Press to


cancel and exit the screen.

3-7
4 Basic Operation
4.1 Infusion Flow Chart

Turn pump on
Press to activate and turn on the pump, the pump will
run start-up checks and display the start-up screen;
Install syringe
Please refer to 4.2.2 Insert the syringe;

Select brand Select according to the syringe brand currently being used;

Select infusion mode Enter default infusion mode, if needs switch, refer to
4.2.6 Select Infusion Mode;

Purge
Refer to 4.2.7 Purge;

Set infusion parameters


Refer to 4.2.8 Set Infusion Parameters;

Insert cannula
Connect to the patient;

Start infusion
Press ;

Finish
Press ;

Remove syringe Disconnect from the patient, and remove the syringe in
a correct manner;

Turn pump off Hold down (>3 s) , until the Power Off progress bar
complete.

4-1
4.2 Operational Procedures
4.2.1 Turn on the Pump
After completing the syringe pump set up, turn on the device as in the following
steps:
1. Perform a safety inspection referring to 11.1 Inspection before turning on the
pump.

2. Press , the system will initiate the self-test and the screen will display the
[System Self-test] interface:
 First, the system will give out a sound "di" —— indicating the self-testing of the
loudspeaker to be successful.

 Then, the color of the alarm indicator lamp will change from red to yellow, turn on
and off orderly —— indicating the self-testing of the alarm lamp to be successful.
 At last, the system will give out a sound of "didi" —— indicating the self-testing of
the buzzer to be successful.
3. Enter the operation interface after successfully completing the system self-testing,
and now you can manually operate the system through the key board.

WARNING

 Please monitor the self-test process to make sure that the speaker, the alarm
light, and the buzzer are all self-tested successfully. Otherwise, please contact the
company and do not operate the pump until after maintenance is performed.

 Please contact the company if the syringe pump is damaged or cannot


operate properly, and it cannot be used for patient infusion.

4.2.2 Insert the Syringe


Enter the syringe [Installation Guide] interface after completing the self-test. If the

syringe is not required to install, please press to skip the step.

4-2
Install syringe according to the following method:
1. Open the door, squeeze the
handle to open the clip, then move
the slider to the appropriate
position. Pull out the spindle clamp
and turn it to the right, allowing the
flange handle to form a slot.

2. Align the syringe flange with the


slot, then insert it into the slot.
Turn the spindle clamp to the left
for reposition and clamp the
syringe.Tip: Flange location

3. Align the slider to and snugly


against the syringe plunger,
making the clip clamps firm to the
thumb rest.

4. Close the door gently.

4-3
NOTE

 If the syringe pump has been installed with a syringe, then the system will
directly skip [Installation Guide] interface to enter [Syringe selection] interface.

WARNING

 The flange of the syringe should be firmly inserted into the slot, and not
jutting on the outside of the flange plate.

 Before using this syringe pump, the syringe pump, syringe and other
accessories should be installed correctly.

 Before using this syringe pump, the brand and specifications of the syringe
used must be confirmed. The brand of syringe pump should be calibrated on the
equipment. If there are no settings for the syringe used, the rate and the alarms may not
be accurate.

4.2.3 Change the Syringe


Follow the steps below to change the syringe:
1. To prevent patient injury, before changing the syringe during infusion, press

to stop the pump.


2. Open the door, squeeze the handle to open the clip, then move the slider to the
appropriate position. Pull out the spindle clamp and turn it to the right, and take out the installed
syringe.
3. Please refer to 4.2.2 Insert the Syringe to reinstall the syringe.

4.2.4 Select Syringe Brands


After finishing installation of the syringe, [Syringe selection] will appear on the

screen, and now you can press to select the syringe brand and
specifications of the syringe currently being used. Specific brand please refer to
6.7
Common Syringe Brands.

CAUTION

 Please confirm that the current selected brand is the same as the brand
actually used.

4-4
WARNING
 A new syringe brand should be calibrated when used for the first time.

4.2.5 Memory Function


In clinical treatments, the medical staffs need to initiate the infusion as soon as
possible during emergency situations, infuse the liquid drug into the patient's body
in the shortest time possible, and set the detailed parameters later during infusion.

After selecting the syringe brand, the previous infusion screen will appear, the

previous therapy parameters will be loaded and will prompt the users [ Use last
therapy parameter], start the infusion after confirming. If not necessary,
please perform the normal operation process.

4.2.6 Select Infusion Mode

Press to enter [Main Menu]→[Select Mode].On this interface, users can

press and to select. Please refer to Chapter 5 Infusion Mode for


detailed introduction of each infusion mode.

4.2.7 Purge
During infusion, the user should prevent air bubbles from entering the blood with the
liquid drug, which may form an aeroembolism and put patients in serious danger.
Therefore, air bubbles in the syringe and extension cannula should be eliminated

before the infusion. Press to enter [Purge] setting screen, set purge [Purge

rate], then hold down to purge the system at the set rate, release after
the air bubbles are purged, completing the purge.

WARNING
 During the purge, please disconnect the pump from the patientotherwise, the
patient will be in serious danger!

 The default rate when purge is the maximum rate of each specification of
syringe, and the default purge rate of 50ml/60ml syringe is 800ml/h.
 The purge accumulated volume is not added to the total volume.

4-5
4.2.8 Set Infusion Parameters
Under each infusion mode, the users should master the following basic function of
the keys:

 : Under the non-setting status, move the cursor up and down;Under the
setting status, indicate to increase/decrease the data value.

 : Under the non-setting status, move the cursor to the right and left; Under
the setting status, indicate to increase/decrease the editing space.

 : Indicate to confirm the current selection or settings.

 : Under the non-setting status, indicate to return to the previous

menu;Under the setting status, indicate to clear the current set or cancel the
edit.

As shown above, the procedure for setting the parameter values is as follows:

Step 1: Press the or key to move the cursor up or down and select the
parameter that requires setting;

Step 2: Press to enable the currently selected parameter by using the cursor
for making adjustments;

Step 3: Depending on the preset parameter value, press or to select the


editing space;

Step 4: After confirming the editing space, press or again to increase or


decrease the relevant value;

Step 5: Repeat step 3 and step 4 until finishing all value settings, and press for
confirmation after completing both steps. The settings are now complete.

The parameter value should not exceed the parameter range defined by this
equipment, please refer to Chapter 5 Infusion Mode for each parameter range,
otherwise, the parameter value will be modified automatically to the maximum value

defined when the setting exceeds the maximum value set, press again at the
original space or top of the digit to restore the original value.For example, if the
maximum parameter value is 1500, while the current value is 600, once the user

presses at the thousands digit, the value will be changed automatically to

1500, press again at the thousand digit to recover to 600.When the parameter

4-6
reaches the maximum value, press at any digits, the value will not change,
and the prompt bar displays [Value Reached Limit] and provides a range of
the parameter.

4.2.9 Infusion

When ready, connect the extension cannula to the patient. Press to start the
infusion, and the screen will display the green running icon, the arrows will move
from right to left, and the running speed will increase, which will, indicate that the
rate will also increase.

Run screen with [VTBI] set Run screen without [VTBI] set

WARNING

 Users should regularly monitor the connection between the syringe,


extension cannula, pump and patient, and infuse according to the method mentioned in
the manual.

NOTE

 When in running status, if there is no operation in other interface over 2


minutes, it will return to the running screen automatically.

4.2.10 Infusion Pause


During infusion, if changing the drug solution or swapping syringes is needed,

please press to enter the [Pause] interface to stop the pump. On the [Pause]

screen, press to return to the previous parameters setting on the interface,

and then press to enter the [Pause] screen. Press to continue the
infusion.

4-7
4.2.11 BOLUS

In any run screen in the infusion mode, press to enter the [Bolus] settings
screen. There are two ways to start the bolus:

 Manual Bolus: Set bolus parameters, hold down to manual bolus, and release
to return to the original rate.

 Auto Bolus: Set bolus parameters, press to auto bolus.

NOTE
 The bolus rate should be faster than the current infusion rate.

 The default rate when bolus is the maximum rate of each specification of
syringe, and the default bolus rate of 50ml/60ml syringe is 800ml/h.

 If no operation is performed within 15 seconds, the syringe pump will


automatically exit the Bolus Settings screen and the procedure must be repeated.

 [VTBI Near Done] and [Syringe Nearly Empty] alarm will not be triggered
during bolus.

4.2.12 Change the Rate during Operation

In any run screen in the infusion mode, press , , , or to change


the value of the [Rate] into the adjustable state, thus to set the expected rate,

press or again for confirmation, then start to infuse under the new set
rate.

4.2.13 Complete
If [VTBI] is set during the infusion, when the remaining infusion time is close to the
[Time Near End] set by the users, the [VTBI Near Done] alarm will be triggered. If
no action has been taken, the alarm will not be cancelled automatically until the
infusion is completed, and then switch to [VTBI Done] alarm. Set [Time Near
End], please refer to 6.6 Time Near End.

If [VTBI] is not set during the infusion, when the time in which the remaining
liquid needs to reach [Time Near End], the [Syringe Nearly Empty] alarm will
be triggered, and this alarm cannot be canceled until the syringe is empty.

4-8
When infusion is completed, enter to [KVO] mode, and KVO mode will run for 30
min at most. Infusion will be stopped automatically after the KVO is finished, and
the [KVO Finish] alarm will be triggered. Set KVO rate, please refer to 6.1 KVO.

4.2.14 Standby

Under non-operation status, tap (<3 s) to enter [Standby] interface, default

display the previous standby time, press to modify (range is 00:01-99:59

hh:mm), press for confirmation after modifying. The equipment cannot be put
in standby mode if there is an advanced alarm.

When the standby state is ended, the title bar will display [Standby Time Expired],

press to confirm and quit, until the screen before standby appears. Press
to remain in standby.

4.2.15 Turn off the Pump


Follow the steps below to turn off the syringe pump:
1. Disconnect from the patient;

2. Hold down (>3 s) , until the Power Off progress bar complete, and the power will
turn off.

NOTE

 When powering off normally, the current operating data and saved data will
be autosaved.

4-9
5 Infusion Mode
5.1 Rate Mode

Mode Parameters Parameter Range

5ml syringe: 0.10-100ml/h


10ml syringe: 0.10-200ml/h
Rate 20ml syringe: 0.10-400ml/h
30ml syringe: 0.10-600ml/h
50ml/60ml syringe: 0.10-1500ml/h

0.10-9999.99ml, If necessary maintenance is not carried


Rate Mode VTBI
out, all the liquid in the syringe will be infused out as default.

00:00:01-99:59:59 h:m:s
Set [Rate] and [VTBI], then calculate [Time] automatically;
Time
Set [Rate] and [VTBI], modify [Time], and the [VTBI] will not
change and will automatically calculate [Rate].

5.2 Time Mode


5-1
Mode Parameters Parameter Range

Time 00:00:01-99:59:59 h:m:s

VTBI 0.10-9999.99ml

Not required, and the range is the same with the Rate
Time Mode
Mode.
Rate Set [Time] and [VTBI], then calculate [Rate] automatically.
Modify [Rate], and the [VTBI] will not change and
automatically calculate [Time].

5.3 Body Weight (BW) Mode

Mode Parameters Parameter Range

Weight 0.1-300.0kg/0.2-660.8 lb

Drug amount 0.1-999.9

Drug amount
g/mg/μg/ng/IU
unit

Body
Volume 0.10-9999.99ml

Weight
Dose rate 0.01-999.99
(BW) Mode

ng/kg/h, μg/kg/h, mg/kg/h, IU/kg/h,


Dose rate unit
IU/kg/min, μg/kg/min , mg/kg/min, ng/kg/min

Rate The same with Rate Mode

VTBI 0.10-9999.99ml
NOTE:
1. [Conc.] will be automatically calculated according to the formula
(Drug amount/Volume).
2. [Rate] will be automatically calculated according to the formula (Dose rate
*Weight)/Conc.

5-2
5.4 Intermittent Administration Mode
In Intermittent Administration Mode, by setting the Rate, Intmt.volume and Intmt.time to
control the infusion. It is suitable for the infusion of long-term analgesia drugs.

Mode Parameters Parameter Range

5ml syringe: 0.10-100ml/h


10ml syringe: 0.10-200ml/h
Rate 20ml syringe: 0.10-400ml/h

Intermittent 30ml syringe: 0.10-600ml/h


Administration 50ml/60ml syringe: 0.10-1500ml/h

Mode

Intmt.volume 0.10-9999.99ml

Intmt.time 00:00:01-99:59:59 h:m:s

Note: After a certain intermittent infusion period is finished, system will display the

remaining time for the next startup, press to continue the infusion.
5-3
5.5 TIVA Mode (Optional)
When used for the total intravenous anesthesia clinically, the TIVA mode controls
the intravenous infusion by setting the Induction parameters and by sustaining
those parameters. First set an Induction time to infuse the entire Induction dose,
which will enable the patients to enter narcotism, and then push the drug with a
maintenance speed to maintain anesthesia depth.

Parameter
Mode Parameters
Range
Basic Weight, Drug amount, Drug amount
Parameters unit, Volume The same

Induction Induction dose, Induction dose unit, as Body


TIVA Mode
Parameters Induction time Weight

Maintenance (BW) Mode


Maintenance dose rate
Parameters

NOTE:
1. The [Induction rate] and [Maintenance rate]are calculated according to the
formula automatically, and the range of the values is taken to the same with
Rate Mode.
2. During the Induction stage, when the current accumulative amount reaches the
Induction VTBI, no matter by manual bolus or auto bolus, the equipment will
quit automatically and enter the maintenance stage.
3. If the [Induction time] is null, the equipment will enter the maintenance stage
directly.
5-4
5.6 Sequential Mode (Optional)
Several different sequences (parameter group) can be set in sequential mode,
and the syringe pump infuses according to the set infusion sequence.

5 sequences can be set in this mode. The rate of the current sequence can be
changed during the operation process. In Sequential Mode, the VTBI, Rate, and
Time are settable and the ranges of the set values are taken to be the same
with Rate Mode.

 : A sign denotes the total VTBI and the total time of all sequences.

 : A status sign denotes the need for a voice prompt after completing each

sequence.When setting this parameter, press to activate, and a voice

prompt will activate after the sequence is completed. Press again to cancel
the activation, there is no voice prompt after the sequence is completed,

and the sign is displayed as .

NOTE

 If only set [Time] for a sequence, denoting that infusion of the sequence is
stopped, and the next sequence will be started when arrived at the specific time.
 If only set [Rate] or [VTBI] for a sequence, the sequence is invalid.

 If there is invalid sequence between sequences, the infusion can not be


started.

5-5
5.7 Loading Dose Mode (Optional)
Loading Dose Mode infuses all [VTBI] of loading dose at the speed of [Loading
Rate], then operates at [Main rate] until completion [VTBI] of the primary
parameters. Primary parameters [VTBI] are total VTBI.

Parameter
Mode Parameters
Range
VTBI (Main VTBI=Total VTBI)
Primary Rate (Main rate) The same with
Loading Parameters Time=(Main VTBI-Loading Dose Rate Mode
Dose VTBI)/ Main rate

Mode Loading VTBI (included in Main VTBI)


The same with
Dose Rate (Loading Rate)
Rate Mode
Parameters Time

NOTE:
1. The primary parameters of the [VTBI] should be more than the loading dose
of the [VTBI], and if set over the specified limit, the parameters that exceed
the system limit can not be set.
2. When the [VTBI] or the [Time] of primary parameter are not be set, the
system will autonomously run with the [Rate] of the primary parameters until
being emptied or completed.
3. When any two parameters of the loading dose are null, the system will run
with the [Rate] until being emptied or completed.
5-6
5.8 Ramp Up/Down Mode (Optional)
In Ramp Up/Down Mode, by setting the ramp-up time and ramp-down time, the
syringe pump will increase the rate (divided as 9 stages) automatically during
the ramp-up time until the rate is steady, sustaining for a period of time, and then
decrease the rate during the ramp-down time.

During any stage of the Ramp Up/Down Mode, if the Bolus Function is initiated, the
VTBI for maintaining the infusion of each stage, such as ramp-up, maintenance and
ramp down, will not be changed.When the bolus rate is recovered to its original
rate, the system will start to run at the stage where the current total accumulated is
the amount located.

If the modified rate is in the Ramp Up/Down Mode, the system will be operated
with the modified rate until completing the infusion or until the syringe is emptied.

Mode Parameters Parameter Range

VTBI (Vol) The same with Rate Mode

The range is the same with Rate Mode, ramp up


Ramp Total Time (T)
time + ramp down time < total time

Up/Down
Ramp up time (t1) Inputtable and modifiable, not required.
Mode

Ramp down time


Inputtable and modifiable, not required.
(t2)
5-7
NOTE:
1. [Steady rate], [Up rate], and [Down rate] are calculated automatically which are
uninputtable and unmodifiable, and the range is the same as that of the speed mode.
2. When no ramp-up time is inputted:
a) Total time=Ramp down time: go directly into the ramp down phase until the
infusion is complete. Ramp down initial time=VTBI/total time.
b) Total time=Ramp down time: Sustain time = (total time)-(ramp up time), go
directly into the ramp down phase and then into sustaining phase until the infusion is complete.
3. When no ramp-down time is inputted:
a. Total time=Ramp up time:go directly into the ramp-up phase until the infusion is
complete.
b. Total time>Ramp up time:sustaining time = (total time)-(ramp up time), go directly
into the ramp down phase and then into sustaining phase until the infusion is complete.
4. When both ramp up and ramp down times are null, go directly into the sustaining phase
until the infusion is complete. Steady rate=VTBI/total time.

5.9 Micro-infusion Mode (Optional)


Clinically, the micro-infusion Mode is mainly used in the clinical slow-rate infusion
for children and new-born.

Mode Parameters Parameter Range

0.10-100.0ml/h for syringes of 5ml, 10ml, 20ml,


Rate
30ml, 50ml/60ml.

Micro-infusion
Mode VTBI 0.10-1000.00ml

Time 00:00:01-99:59:59 h:m:s

5-8
6 Setting Parameters
6.1 KVO
KVO (Keep Vein Open) means to keep the vein open, during which the syringe
pump continues infusion at a very low rate after finishing the infusion in order
to prevent blood backflow or vascular occlusion.

1. Select [Main Menu]→[General Option]→[KVO rate].


2. Select [KVO rate]: 0.1~5.0ml/h is adjustable.

6.2 Drug Library


The product is configured with a drug library with a collection of 2,000 kinds of
drugs available for users to select from.

1. Select [Main Menu]→[General Option]→[Drug library].


2. Select [Drug library]→[On]. If [Off] is selected, the following steps cannot be
conducted.
3. In any one of the infusion mode setting interfaces, select [Drug]→[Select Drug].

4. In the [Select Drug] interface, press to turn the pages to browse the whole drug
list, and for further details, see C.5 Drug Library List.
5. After the drug is selected, its name will appear on the Run screen.

6.3 Occlusion Pressure (Unit)


Occlusion pressure is adjustable, which can meet the requirements for occlusion
pressure of different patients during infusion. Pressure in the infusion tube can be
measured by the built-in pressure sensor, pressure can be calculated by the
internal CPU, which is compared with the preset occlusion alarm threshold.
[Occlusion] alarm will be triggered if pressure exceeds the threshold.

6.3.1 Set Occlusion Pressure


1. Select [Main Menu]→[Occlusion pressure].
2. Select [Occlusion pressure]: Occlusion pressure Degree 11, lowest at 150mmHg,
and highest at 975mmHg. Occlusion pressure should be selected according to actual needs.

6-1
CAUTION

 If the patient experiences discomfort at a higher occlusion pressure, monitor


the patient's physical conditions under the higher occlusion pressure closely, and take
measures instantly if any abnormal condition occurs.

6.3.2 Set Pressure Unit


1. Select [Main Menu]→[General Option]→[Unit of pressure].
2. Select [Unit of pressure]: The 4 various forms of pressure units, mmHg, kPa, bar
and psi are converted automatically, and can be selected according to actual needs.

CAUTION
 Carefully confirm the edit when changing the current pressure unit.

6.3.3 Dynamic Pressure Scanning (DPS)


During the infusion, the bottom-right corner of the Run screen demonstrates real-
time pressure changes of the patient in order to find the cannula occlusion at an
earlier time and to prevent the occurrence of further complications.

The pressure icon on the screen indicates the condition of the current pressure:
When the progress bar is white, this indicates that the current

pressure value is below 80% of the occlusion pressure.

When the progress bar is yellow, this indicates that the current
◆ pressure value is above 80% and below 95% of the occlusion
pressure.

When the progress bar is red, this indicates that the current
◆ pressure value is above 95% and below 100% of the occlusion
pressure.

6.3.4 Automatic Pressure Release Function (Anti-Bolus)


When occlusion occurs, infusion will stop and the [Occlusion] alarm will be
triggered. After the alarm is triggered, the motor is reversed, and the cannula
pressure is then released. This prevents an additional aggressive dose to the
patient after the occlusion is eliminated.

6-2
6.4 Key Lock Function

When locked, an icon in the upper-right corner of the screen emerges. The
following are two ways for automatic locking and manual locking:

 Automatic Locking:
1. Select [Main Menu]→[General Option]→[Auto-lock time].
2. Select [Auto-lock time]: Off, 1-5min. After a specific time is set during the running
state, and if there is no operation or high-level alarm within the set
auto-lock time, the key board will be auto-locked. [Off] indicates closing
automatic locking function.
 Manual Locking: In the running interface, under the unlocking lock condition,

press and hold (>3 seconds) to lock the key board.

If unlocking is needed, press and hold (>3 seconds) to unlock, it is


automatically locked during the high-level alarm.

6.5 Reminder Function


1. Select [Main Menu]→[General Option]→[Reminder time].
2. Select [Reminder time]: Off, 1-5min. After a specific time is set, the syringes are to
be inserted. No operations are performed by the pump within the set time (including operations
on the keyboard, the slider and the pull handle), and the [Reminder] alarm will then alert the user
to proceed to the next step. [Off] indicates the closing of the function.

6.6 Time Near End


1. Select [Main Menu]→[General Option]→[Time near end].
2. Select [Time near end]: Off, 1-30min (when <10min, the stair-step is 1min, and
when ≥10min, the stair-step is 5min). After a specific time set, when the remaining time
approaches the specified time for terminating the alarm, the alarm of [VTBI Near Done] will be
triggered. [Off] indicates the closing of the function.

6-3
6.7 Common Syringe Brands
There are multiple commonly used syringe brands installed inside the syringe
pump, making it convenient for the user to select from.Only 5 ml, 10 ml, 20 ml, 30
ml and 50 ml/60ml syringes that comply with national standards should be used
with this syringe pump. For specific syringe brands, please refer to actual syringes
as the standard.

1. Select [Main Menu]→[General Option]→[Commonly used syringes].


2. Select in [Commonly used syringes] according to actual needs.

Note: Recommended syringe brands include B.Braun, B-D,Shuangge,


Jierui,Terumo, SuYun, ShanChuan, HuaFu, QiaoPai, Diprivan and etc.

6.8 Bed No. Settings


1. Select [Main Menu]→[Patient Information]→[Bed No.].
2. Select [Bed No.]: 1-999. [---] indicates invalid values.Beds can be differentiated
through setting bed number.

NOTE
 For syringe pump inserted to BeneFusion DS5 Infusion Supervision
System, Bed No. can’t be changed.

6.9 View Department


Syringe pump has been inserted to BeneFusion DS5 Infusion Supervision System
and the system contains department information, the system will automatically
distribute department information to all syringe pumps when pumps are on.

Select [Main Menu]→[Patient Information]→[Department] to view the


department information.

6-4
7 Other Functions
7.1 History Record
The syringe pump when in use will produce some key data stored in [History Record],
providing foundation for the treatment review and maintenance review at a later
period.The attribute of recording events includes action, time and description.

A record is created once an event occurs. The memory can store up to 2000
records. Once the memory is full, the oldest records will be removed first. History
record will not loss when the syringe pump powers down.

1. Select [Main Menu]→[System Option]→[History Record].


2. Select [History Record]: Each page can demonstrate up to 4 records, and

press to turn the pages.

7.2 Power-down Save


To prevent the loss of patient data when the syringe pump suddenly powers
down, the syringe pump provides the function of the power-down data storage. If
the syringe pump powers down suddenly after it is restarted, the last infusion
parameters will display the alarm information and will remain in consistency with
those before the power-down, and will be reloaded. You can refer to [History
Record] to view such information as infusion parameters and alarm.

7.3 Nurse Call


Select [Nurse call] in [System Option], and set in the open menu:
 Switch

 Signal Type 1. Continuous


indicates that the output nurse call signal type is the same as that of the alarm
existence time, i.e., from the occurrence of the alarm to the end of it.
2. Pulse
indicates the output nurse call signal is a pulse signal with the type of 1 second.
When several alarms exist at the same time, only one pulse signal can be
outputted.If the current alarm is not removed and another alarm occurs, then one

7-1
additional pulse signal is outputted.
 Trigger Type
1. Normally Close: Select when the hospital call system is set as [Nor. Close].
2. Normally Open: Select when the hospital call system is set as [Nor. open].

 Alarm Level: Three options: [High], [Medium] and [Low].The system sends nurse
call signals according to the alarm at the selected alarm level or above.

WARNING
 Non-medical personnel are forbidden to modify the nurse call setting.

 The nurse call function must be used in conjunction with a special cable.

NOTE

 Medical personnel should not consider the nurse call function as the main
alarm notice approach, and rather combine the sound and visual alarms of the syringe
pump and the clinical performances and symptoms of the patient in order to judge the
patient's conditions and take further attention as needed.

7.4 Wireless Networking (Optional)


The syringe pump can be configured with the wireless modules, and be net-
connected with BeneFusion CS5 Infusion Supervision System through the
approach of wireless networking.Through the network:
1. The syringe pump sends real-time infusion parameters, drug information, alarm
information, prompt information, Bed No. and other information to BeneFusion CS5 Infusion
Supervision System.
2. BeneFusion CS5 Infusion Supervision System and the syringe pump operate
synchronously.

For detailed descriptions, please refer to the instructions of BeneFusion CS5


Infusion Supervision System (hereinafter called CIMS).

Normal communication of the pump and CIMS depends on whether the network
connection is successful, operators are unable to observe the operation status of
the pump in real time when the communication is interrupted. After the network
connection settings of the pump and CIMS are modified, operators shall reset the
network connection as required in the manual to ensure the communication of the
pump and CIMS are restored.

7-2
When using wireless modules to connect to the Internet while using the syringe
pump, the wireless icon on the upper-right corner indicates the working condition of
the wireless modules:
■ Wireless modules work in order

■ No icons No wireless modules configured

NOTE
 Wireless security transmission distance is no more than 50 meters.

 2.4 GHz Wi-Fi frequency range, WEP and WPA/WPA2 security modes and
802.11b/g/n wireless standard are supported.

 The settings of the wireless network must be conducted by technicians


approved by the company or maintenance staff designated by the company.

7.5 Data Export


To export the data in the syringe pump, please refer to the following steps:
1. Log on PC tools, and connect the PC to the syringe pump;
2. When the syringe pump is in working communication with the PC, the PC
automatically reads all the data in the pump;
3. Select [History Record] in PC tools;
4. Export data.

7.6 Automatic Multi-channel Relay


Multi-syringe pumps/infusion pumps have been inserted to BeneFusion DS5
Infusion Supervision System and the pumps are on, follow the steps below to
start relay mode:
1. Select [Main Menu]→[Select Mode], then select [Auto. multi-channel relay] in
each mode.

2. Set the slot No. of the “Forward channel”, press to save. If the slot of
“Forward channel” pump is infusion pump, please input VTBI parameter value
of the pump.

3. When the pump with finishes infusion, the current pump with will start
running automatically. When the last relayed pump finishes infusion,
[VTBI Done] alarm will be triggered.

7-3
7.7 Patient Information
Syringe pump has been inserted to BeneFusion DS5 Infusion Supervision System
and the system contains patient information, the system will automatically
distribute patient information to all syringe pumps when pumps are on.

Select [Main Menu]→[Patient Information] to view the patient information; press

to discharge patient.

NOTE
 For syringe pump inserted to BeneFusion DS5 Infusion Supervision
System, patient information can’t be discharged.

7.8 Night Mode


To activate the night mode:
1. Select [Main Menu]→[System Option]→[Night Mode].
2. Select “On” to activate the night mode.

3. Set start time, end time, volume and contrast, and press to save. When
system time reaches the start time of night mode, volume and contrast of
system mode will automatically change to volume and contrast of night mode;
when system time reaches the end time of night mode, volume and contrast of
night mode will automatically change to system volume and contrast.

Note: Syringe pump has been inserted to BeneFusion DS5 Infusion Supervision
System and the system contains night mode information, the system will
automatically distribute night mode information to all syringe pumps when
pumps are on.

NOTE

 Minimum time period of start time and end time is 1 hour, while maximum
time period is 15 hours.

WARNING

 Before entering night mode, confirm the settings of brighness and alarm
volume. Pay attention to the potential risk when the setting value is a bit low.

7-4
7.9 Day's Prescription
Syringe pump has been inserted to BeneFusion DS5 Infusion Supervision System
and the system contains prescription information, the system will automatically assign
all non-executed prescription to all idle syringe pumps when pumps are on.

You will enter the [Prescription detail] interface, press to execute


prescription.

To execute prescription:

1. Select [Main Menu]→[Day's Prescription], press to enter the [Prescription


detail] interface.

2. In the [Prescription detail] interface, press to execute prescription and


enter the infusion parameter setting interface.
 If infusion mode is specified in prescription, the pump wil start infusion as the
specified mode.
 If infusion mode is not specified in prescription, then Rate mode for pedia or adult;
Micro-infusion Mode for neonate.

3. When ready, press to start the infusion.

7.10 WLAN Setting


The pump can be net-connected through built-in Wi-Fi module.

1. Select [Main Menu]→[System Option]→[WLAN Setting], then select [On] to


enable Wi-Fi function.
2. Select [Advanced Settings], there are two ways to distribute IP address:
 DHCP: Check the checkbox to activate DHCP, IP address, subnet mask and
gateway can not be modified, automatically obtain an IP address.
 Manually: Uncheck the DHCP checkbox, enter IP address, subnet mask and
gateway.
3. Available networks shall be displayed:
 If password is required for to-be-connected network, please enter the password.

 If password is not required for to-be-connected network, you can connect the
network directly.

7-5
8 Alarms
The alarm is used in order to alert the medical staff by means of sound and light
when abnormal situations occur during the infusion procedure which can lead to
infusion changes or when the infusion of the patient cannot continue due to the
unexpected breakdown or pause/delay of the syringe pump.

WARNING

 It is potentially hazardous to use the same or similar equipment with different


alarm presets within the same area.

8.1 Alarm Level


According to the severity scale of the alarm, the alarms of the syringe pump can
be classified to be high-level alarms, mid-level alarms and low-level alarms.

8.2 Alarm Types


When an alarm is triggered, the syringe pump will use the following visual
and audible methods to alert the user:
 Visible alarms
 Audible alarm
 Alarm Information

Among the visible alarms and audible alarms, the alarm information will
distinguish alarm levels in different ways.

Color of Audible alarm Flashing light Light/no-lig


Alarm level Alarm Information

alarm light frequency frequency ht ratio

High-level black characters on


Red 10 seconds 2.0±0.6Hz 20%~60%
alarms a red background

Mid-level black characters on


Yellow 15 seconds 0.6±0.2Hz 20%~60%
alarms a yellow background

Low-level white characters on


Yellow 20 seconds Steady 100%
alarms a black background

8-1
8.2.1 Multi-level Alarm Rules
When several alarms occur simultaneously, the alarms proceed according to
the following rules:
 When several alarms at different levels occur, the visible alarms and audible alarms
are consistent with the highest-level alarms.
 When several alarms at different levels occur, only the highest-level alarm is
displayed, and after it is cancelled, the lower-level alarm will then be displayed.
 When several alarms at the same level occur, the alarm information will be
demonstrated in an alternate manner.

The title bar of the syringe pump screen will display the corresponding alarm
information during the alarm blast, to see more details in D Alarm
Information:
 Occlusion
 Battery empty
 VTBI done
 KVO finish
 Syringe empty
 Syringe disengaged
 System error
 Reminder
 Battery Low
 No battery inserted
 System abnomal
 Syringe nearly empty
 VTBI near done
 Standby time expired
 AC power disconnection
 Pump serial number conflict
 Relay invalid
 Communication is interrupted
 Syringe not correctly inserted

NOTE

 The [Communication is interrupted] alarm of the pump and BeneFusion CS5


Infusion Supervision System are delayed for 3 minutes, while other alarms are delayed for
less than 5 seconds.

8.3 Alarm Handling Rules


Under normal working conditions, when an alarm occurs, all the alarm types of
the syringe pump will alert according to their respective alarm levels. In addition,
the user can pause the alarm sound according to demands.

8-2
 For high level (except battery empty) and medium level alarms, press to pause
alarm sound for 2 minutes, no alarm sound is made in any case. When the

alarm pause time expires, the alarm tone will sound. Press to cancel high level
alarms (except battery empty and system error).

 For low level alarms, press , no alarm sound, alarm information and alarm light,
until it is triggered next time.

NOTE

 [Battery Empty] alarm sound is unable to be paused.

8.4 Alarm Countermeasures

WARNING

 When an alarm is triggered, the patient's condition should be checked firstly


and operation should only be allowed to proceed after the reason for the triggering of the
alarm is ruled out.

When an alarm is triggered, please follow these steps and take appropriate action:
1. Check the patient;
2. Check the alarm type and the parameter which triggered the alarm;
3. Determine the reason for the alarm;
4. Eliminate the reason for the alarm;
5. Check whether the alarm has cleared.

NOTE

 Please refer to D Alarm Information for specific handling procedures for each
alarm.

 Without AC power when disassembling the battery or when the battery is


abnormal, the buzzer on the alarm will sound out and the alarm blast will continue for 3
minutes.

 The operator position shall be the normal operating position of the syringe
pump (0.5m). Otherwise, operators might identify alarm mistakenly.

8-3
9 Battery
The syringe pump is configured with rechargeable Lithium ion batteries to ensure
that the syringe pump can be used normally under the condition of the patient's
migration within the hospital or during the circumstance of a power failure. When
the syringe pump switches to the AC power, the battery can be charged regardless
of whether the syringe pump is on or off. The battery is chargeable only within the
syringe pump.During charge, the battery icon in the upper-right corner of the screen
floats up and down.If the battery icon stops floating and is completely filled, it
indicates that the battery is fully charged. Under the condition of a sudden power
failure, the system will automatically use the battery to provide power as a backup.

The battery icon on the screen indicates the condition of the battery:
The battery jar of the syringe pump is installed with batteries,
and the white fill area indicates the quantity of electricity.

Low battery electric quantity indicates that charging is needed.

When the battery is empty, charging is needed immediately.

No battery is installed in the syringe pump.

The power supply by the battery can only be sustained for a limited period of time.
The [Battery Empty] alarm will be triggered when the battery voltage is too low,
and red alarm light will flash. The alarm will continue within the remaining time of
the battery’s electric quantity and cannot be paused. Now, the syringe pump should
be switched on to AC power for charging.

WARNING
 Please keep the batteries out of the reach of children.

 Use the battery specified by the manufacturer.

NOTE

 Please remove the batteries before conveying or when out of use for a long
time.

9-1
9.1 Battery Performance Optimization
When the battery is used for the first time, at least two complete optimizing cycles
should be ensured. A complete optimizing cycle entails the following: Charging
incessantly, and then discharging until the power of the syringe pump runs out.
During usage, regularly optimizing the battery performance will extend its lifespan.
It is suggested that the battery should be optimized when in use or in storage for
two months, or when the running time of the battery is significantly shortened.

Please follow the steps below during optimization:


1. Disconnect the pump from the patient and stop the infusion.
2. Insert the battery needing optimization into the battery jar of the syringe pump.
3. Switch the syringe pump on the AC power and charge the battery incessantly for
over 10 hours.
4. Disconnect the AC power and use the battery to charge the syringe pump until the
syringe pump is closed.
5. Switch the syringe pump over to AC power again and charge the battery incessantly
for over 10 hours.
The battery optimization is now complete.

9.2 Check the Battery


The performance of the battery may decrease over time. Please follow the steps
below when checking the battery:
1. Disconnect the pump from the patient and stop the infusion.
2. Switch the syringe pump on the AC power and charge the battery incessantly for
over 10 hours.
3. Disconnect the AC power and use the battery to charge the syringe pump until the
syringe pump is closed.
4. The length of the battery’s lifetime reflects the performance of the battery.
5. If the length of the battery’s lifetime is obviously shorter than that claimed in the
specifications, please consider changing the battery or contact us.

9-2
NOTE

 The lifespan of the battery depends on how frequently it is used and on how
long it has been used.If the maintenance and storage of the battery is appropriate, the
lifespan of the Lithium ion battery is about 5 years.Inappropriate use of the battery will
shorten its lifespan.We recommend replacing the Lithium ion battery every 3 years.

 The length of the battery’s lifetime depends on the device configuration and
operation, for example:Under the condition of the power supply by the battery, frequent
infusion at a high rate will also shorten the length of the battery’s lifetime.

9.3 Insert Batteries


Batteries need to be replaced:
1. It is suggested to close the syringe pump power source, and disconnect the power
source line.
2. Tilt or lay the syringe pump down.
3. Open the battery door, and unlock the battery lock.
4. Remove the old battery, install the new one into the battery box and insert the
battery lock.
5. Close the battery door and return the syringe pump to its upright position.

9.4 Battery Recycling


If there is any obvious damage to the battery or to the battery capacity exhausts, it
should be replaced and recycled appropriately. Please follow the applicable laws
on recycling.

WARNING

 The battery must not be disassembled, burned or short-circuited. Burning,


exploding or leaking batteries can cause personal injury.

9-3
10 Preservation and Sanitation
The pump must be cleaned or disinfected using the materials and methods listed in
this section. The manufacturer will not be responsible for any damage or accident
caused by cleaning and disinfection using other materials and methods.

The manufacturer shall not be held responsible for the efficacy of the following
chemicals or methods for infection control. Please contact your hospital's
infection prevention department or epidemiology specialists for advice on
infection control practices.

10.1 Description
Please make sure that your device and other fittings are clean without dust.In order
to prevent any damage to the device, please abide by the following rules:
 Dilute all cleaning agents and disinfectants in accordance with the manufacturer’s
instructions, or use as low a concentration as possible.
 Do not immerse or submerge the device in liquid.
 Do not pour liquid on the device or accessories.
 Avoid liquid from entering the pump body.

 Do not use abrasive materials (such as steel wool or silver polishes), or any strong
solvent (such as acetone or any detergent containing acetone).

WARNING

 Turn off the pump and disconnect the AC power source line from the socket
before cleaning. Do not clean and disinfect the device, export history record and perform
other operations when patients are using the pump.

10.2 Cleaning
The pump should be cleaned regularly. If operating in dirty or sandy areas,
cleaning should be more frequent. Before cleaning, please consult or refer to the
hospital's specific regulations concerning medical device cleaning. The following
are detergents available for selection:
 Hydrogen peroxide (3%)
 Ethanol (70%)
 Isopropanol (70%)

10-1
When cleaning the device:
1. Turn off the pump and disconnect the AC power source line.
2. Wipe the display screen after soft cotton balls absorb an appropriate amount of
detergent.
3. Use a piece of soft cloth which absorbs a modest amount of cleaning agent to wipe
the surface of the device.
4. When necessary, use a piece of cloth to wipe off any excess cleaning agents.
5. Place the equipment in a cool and ventilated environment to dry.

10.3 Disinfection
The operation of disinfection may cause certain damage to the syringe pump.
You are recommended to disinfect only when it is necessary in your desired
maintenance plan. Clean the equipment before disinfection.

The recommended disinfectants include: ethano 70%, isopropanol 70%,


glutaraldehyde-type 2% liquid disinfectant.

CAUTION
 Never use EtO or formaldehyde for disinfection.

 Do not conduct high pressure or high temperature disinfection for the syringe
pump and its accessories.

10-2
11 Maintenance
WARNING

 The hospital or medical facility using this syringe pump must set up a
comprehensive maintenance plan. Failure to do so may result in equipment failure or
other unexpected consequences, and may even jeopardize personal safety.

 All safety inspections or maintenance work involving the disassembly of the


device must be conducted by professional maintenance personnel. Actions by unqualified
persons may result in device failure and may even jeopardize personal safety.

 Please contact the company immediately if you encounter problems with the
device.

11.1 Inspection
The syringe pump must be given a thorough inspection before use, after 6-12
months of continuous use, and after maintenance or upgrades, to ensure that it is
operating and functioning normally.

The inspection criteria are:


 The environment and power supply meet requirements.
 The equipment and accessories have no mechanical damage.
 The power cord is not damaged and has sufficient electrical insulation.
 Accessories used with the pump are correct.
 The alarm system functions correctly.
 Battery performance.
 Self-checking and pump functions are normal.

If there are any forms of damage or abnormal circumstances, do not use the
syringe pump and contact the company immediately.

11.2 Maintenance Plan


The following tasks must be conducted by professional maintenance personnel
approved by the company. Please contact the company if the following
maintenances are needed. Must clean and disinfect the device before the test
or maintenance.

11-1
Inspection/Maintenance Items Frequency

Once every two years.Perform after the


Perform a safety inspection according
board is changed or the syringe pump is
to the IEC60601-1 standard.
accidentally dropped.

Preventive maintenance (refers to the Once every two years, or when you
Maintenance Manual for pressure suspect the occlusion alarm is abnormal,
calibration, sensor calibration, and the flow volume is inaccurate, or the
pump inspection). syringe is incorrectly identified.

11.3 View Information


Select [Main Menu]→[System Option]→[History Record]. In the [History
Record] interface, you can view the infusion parameters, alarm information and
operation information, etc.

Select [Main Menu]→[System Option]→[Version Information]. In the [Version


Information] interface, you can view the information of the version of the syringe
pump system software or other versions.

11.4 Syringe Calibration


The syringe in the syringe pump needs no daily calibration.But calibrate when the
syringe pump is used for the first time, is replaced with a new syringe brand or
when you suspect the deviation of the flow volume is much larger.

Prepare the following materials before the calibration:


 Syringe: 5 standard syringes of the corresponding full dimensions of 5 ml, 10 ml, 20
ml, 30 ml, and 50 ml/60 ml, without liquid.

The steps for calibration are as follows:


1. Pull the syringes to the full dimensions and install them on the syringe pump."L" in
the fig. below refers to the full dimension of the 50 ml syringe.
2. Open the [Syringe Calibration] interface: Select [Main Menu]→[System
Maintenance]→Input User Maintenance Password→[Syringe Calibration].
3. Select [Brand] and [Size] of the current syringe in the [Syringe Calibration]
interface.

4. Press and the syringe pump starts automatic calibration.


5. The screen prompts [Calibration complete] after the calibration is successful.If
the calibration is failed, the screen prompts [Calibration failed, press to

11-2
restart calibration], then you need to calibrate again.

6. Press to exit the current interface.

Please align the front end of the plunger rod's


sealing plug with the full dimension mark

NOTE
 The full dimension of the 50/60 ml syringe is the distance from 0-50 ml.

11.5 Safe Disposal and Recycling


Please contact the company for related information about safe disposal and
recycling.

11-3
12 Accessories
WARNING

 Use the accessories specified in this chapter only. Other accessories may
cause damage to this syringe pump, or cannot reach the specification in this manual.
 Please do not replace an accessory if its package or itself is damaged.

Materials PN

Li-Ion battery 022-000161-00

009-002755-00
Power cord 009-003358-00
(Select PN according to sales area) 009-003651-00
009-002756-00

Fixing clamp 115-022050-00

Nurse call cable 009-004287-00

DC input cable 009-004285-00

Serial communication cable 009-004286-00

Handle set 115-028601-00

NOTE

 This Operator’s Manual describes the most complete functional


configuration of the system. The device you are using may not have some of the settings
or functions described herein.
12-1
A Product Specifications
A.1 Safety Specifications
A.1.1 Product Classification
Classified according to the China SFDA, this syringe pump is a Type II device.
Classifications of this syringe pump according to the IEC60601-1 standard are as
follows:

Safety

Components Host

IEC protection class I

Protection against
CF Protected against defibrillation
electric shock
Impurities and liquid
IP23
ingress protection
Explosion protection
Unsuitable
level

Operating mode Continuous

Mobile level Portable

NOTE:
 I: Type I devices
 CF: Class CF applied parts, can be directly used in the heart.

 IP23: Protected against solid foreign objects with a diameter no less than 12.5mm
and protected against spraying water.
 Unsuitable: The device is unsuitable to be used in environments containing air
mixed with flammable anesthetic gas, oxygen or nitrous oxide.
 Portable devices: Can be moved from one place to another by one or more persons
or by other means when the devices are in use or being used.
 Portable syringe pump: Used to control the infusion of patients and are devices
which can be carried by the patients continuously.
A-1
A.1.2 Operating Environment

Work environment

Temperature 5-40ºC

Relative Humidity 15-95%, non-condensing

Atmospheric pressure 57-106 kPa

Storage environment

Temperature -20 - 60 ºC

Relative Humidity 10-95%, non-condensing

Atmospheric pressure 50-106 kPa

Storage conditions Corrosive-free and ventilated indoors

AC Power Supply

Voltage 100-240 V~

Frequency 50/60 Hz

Current 0.53-0.28A

Fuse Low interrupting rating, T 1A 250 V~

External DC power supply

Voltage DC 10 V-16 V

Current 2.25-1.5A

A.2 Physical Specifications

Components Weight Size Remark

Less than Less than 295 x 174 x 87(mm) Battery


Host

2.5 kg (length×width×height) included


A-2
A.3 Hardware Specifications
A.3.1 Display

Display

Type Color TFT LCD

Size (diagonal) 3.5 inches, ration 16:9

Differentiation 480 x 272 pixels

A.3.2 Battery

Internal battery

No. of batteries 1

Battery type Li-Ion ion battery

Shutdown delay About 30 mins (new battery, after the first low battery alarm)

Rated battery voltage 7.4 VDC

Battery capacity 4800 mAh

Continuously operate at a rate of 5 ml/h, discharge for at least


Power supply time

10h using a fully charged new battery.

Charging time When the pump is off, the charging time is not longer than 6h
A.3.3 Host LED

Host LED

Alarm light 1 (tow colors: red and yellow)

AC indicator light 1 (green)

Battery indicator light 1 (green)

A.3.4 Auditory Indicator


Produce an alarm, the sound pressure is 53-73 dB(A) and key
Speaker beep; Support multi-level volume functions; The alarm sound
meets the requirements of the IEC60601-1-8 standard.

A-3
A.3.5 External Ports

Ports

AC power supply
1 AC power supply port
port
Multifunction 1 multifunction interface, which combines the following interface
interface functions:

DC power input interface
 RS232 interface
 Nurse call interface

A.3.6 Signal Output Interface

Auxiliary output interface


Comply with the Meet the requirements of EN 60601-1 standard about short circuit
standards protection and current leakage.

Output impedance 50 Ω

Nurse call signal output

Driving mode Relay drive

Electric
≤60 W, ≤2 A ≤36 VDC, ≤25 VAC
specification

Isolation voltage >1500 VAC

Action mode Normally open or normally closed (optional)


A-4
A.4 Specifications

Parameters Specifications

Syringe used in conjunction with syringe pump should meet the

Syringe standard requirements of ISO 7886-1: Sterile hypodermic syringes for

single use.
Compatible syringe
5 ml, 10 ml, 20 ml, 30 ml, 50 ml /60 ml
sizes (ml)
 5ml syringe: 0.10-100ml/h
 10ml syringe: 0.10-200ml/h

Rate range  20ml syringe: 0.10-400ml/h

 30ml syringe: 0.10-600ml/h


 50ml/60ml syringe: 0.10-1500ml/h
Minimum increment
0.01 ml/h
of rate

The same with rate range


Bolus rate range

Note: Bolus accuracy is not declared

Purge rate range The same with rate range

VTBI range 0.10-9999.99 ml, the increment is 0.01 ml

Volume range 0.01-9999.99 ml, the increment is 0.01 ml

Time display range 00:00:01-99:59:59 h:m:s

Standby time range 00:01-99:59 hh:mm

Basic modes: Rate Mode,Time Mode, Body Weight Mode and

Intermittent Administration Mode


Infusion mode

Optional modes: TIVA Mode, Loading Dose Mode, Sequential


Mode, Ramp Up/Down Mode and Micro-infusion Mode
KVO rate 0.1 - 5.0 ml/h, the increment is 0.1ml/h

Drug library On, Off

Anti-bolus switch On, Off

1-11, respectively are (150, 225,300, 375, 450, 525, 600, 750,

Occlusion pressure 825, 900 and 975) mmHg

Maximum occlusion pressure is about 1300mmHg.

Unit of pressure mmHg, kPa, bar and psi

Auto-lock time Off, 1-5 min, step for 1min

Reminder time Off, 1-5 min, step for 1min

Off, 1-30 min

Time near end when the time is <10min, step for 1min, and step for 5 min when

the time is ≥10 min

A-5
Bed No. 1-999

Volume 1-8

Contrast 1-8

Time:_ _:_ _

Date:_ _ _ _-_ _-_ _


System date and

time
Time format:12h, 24h

Date format: yyyy - mm - dd, mm - dd - yyyy or dd - mm - yyyy

System language You can select language according to actual needs.

History Record Can store up to 2000 records.

Nurse call On, Off

Infusion rate ≥1 ml/h, infusion accuracy error ≤±2%


Infusion accuracy

Infusion rate <1 ml/h, infusion accuracy error ≤±5%


Mechanical
Mechanical accuracy error ≤±1%
accuracy
Occlusion, battery empty, VTBI Done, KVO Finish, syringe
empty, syringe disengaged, system error, reminder, battery low,

no battery inserted, system abnormal, syringe nearly empty,


Alarm Information

VTBI near done, standby time expired, AC power disconnection,


pump serial number conflict, relay invalid, communication is
interrupted, syringe not correctly inserted

Status indicators Stop, infusion, bolus, KVO, pause, standby, alarm and purge
Dose of single fault About 1.66ml

A-6
A.5 A Reference Table Showing Occlusion Alarm
Delay and Possible Dose

Occlusion Time of
Rate Bolus
pressure occlusion alarm
(ml/h) (ml)

(Level)
(hh:mm:ss)

0.1 04:10:07 0.081

Level 1 1 00:38:38 0.055

5 00:14:52 0.097

1 01:54:45 0.086

Level 6

5 00:20:36 0.080

0.1 26:41:27 0.075

Level 11 1 02:55:17 0.226

5 00:32:41 0.204

NOTE
 Test conditions:
 FLUKE IDA4 PLUS tester
 Syringe brand: Jierui
 Syringe specification: 50ml
 Test temperature: 20±2℃
 Extension tube length: 1 meter

 Occlusion alarm pressure, alarm delays and bolus volume may vary depending
on test conditions, temperature and tube length.

 The above data are only typical values under normal test conditions. The actual
data may vary as test conditions change. Please refer to the test data for the product you
have purchased.Under the same standard occlusion value and rate, the higher the value
of the tested pressure is, the longer the alarm time will be delayed.
A-7
A.6 Infusion Accuracy Curve
The following typical infusion accuracy table expresses performance after infusion
has started and infusion fluctuations occurring within a certain period of time after
normal infusion flow volumes have been reached. The infusion accuracy table is
for reference only; detailed infusion accuracy curve is in accordance with the final
device.

Plotted on the basis of data collected over a two-hour measurement period.

Syringe brand: Jierui


Syringe specification: 50ml
Sampling quantity of pump: 3
Sampling quantity of syringe: 3
Sampling rate: 1ml/h
Sampling interval: △ t =0.5 min

Test period: t =120 mins


Infusion rate: Q (m/h)

Syringe brand: Jierui


Syringe specification: 50ml
Sampling quantity of pump: 3
Sampling quantity of syringe: 3
Sampling rate: 5ml/h
Sampling interval: △ t =0.5 min

Test period: t =120 mins


Infusion rate: Q (m/h)

NOTE

 Infusion accuracy may be influenced by the pump's environment (such as


pressure, temperature, humidity and any infusion consumables used).

A.7 Trumpet Curve


Flow rate deviation over time (p△ t)

A-8
Syringe brand: Jierui
Syringe specification: 50ml
Sampling quantity of pump: 3
Sampling quantity of syringe: 3
Sampling rate: 1ml/h
Sampling interval: △ t =0.5 min
Observation windows: p△ t = 2, 5, 11, 19, 31 mins

Maximum deviation over the course of a full observation window: EPmax (%)
Minimum deviation over the course of a full observation window: EPmin (%)
Average deviation: A (%)

Syringe brand: Jierui


Syringe specification: 50ml
Sampling quantity of pump: 3
Sampling quantity of syringe: 3
Sampling rate: 5ml/h
Sampling interval: △ t =0.5 min
Observation windows: p△ t = 2, 5, 11, 19, 31 mins

Maximum deviation over the course of a full observation window: EPmax (%)
Minimum deviation over the course of a full observation window: EPmin (%)
Average deviation: A (%)

A-9
B EMC and Radio Regulatory Compliance
B.1 EMC
This pump complies with EMC standard IEC60601-1-2 and IEC60601-2-24.

NOTE

 Use of accessories, sensors or cables outside of the specified scope of this


pump may increase electromagnetic emissions and/or lower the electromagnetic
immunity of the pump.

 This pump may not be used in close proximity to or stacked with other
equipment. If necessary, closely observe the pump to ensure that it is able to operate
normally in its environment.

 Special care must be taken to protect the pump from electromagnetic


interference. The following requirements describe the conditions in which the pump must
be installed and maintained.

 The syringe pump should not be used at the same time as any MRI (Magnetic
Resonance Imaging) or other similar equipment in order to avoid the possibility of the
pump malfunctioning or crashing due to electromagnetic interference.

 Even if other equipment is compliant with CISPR emission requirements, it


may still interfere with pump operation.

 Where electromagnetic signals are weaker than the measuring device's


sensitivity range, measurements may be inaccurate.

 This pump is intended to be used only by qualified medical professionals.


Operation of the device/system may cause radio interference or disturbance of other
equipment within the pump's vicinity. Mitigation measures may be necessary, such as the
reorientation and the re-placement of the surrounding equipment or by shielding the
appropriate venue.

 Portable and mobile RF communications equipment can affect the


performance of measuring devices.

 Devices of A Type are intended to be used in hospital. This pump conducts


emission and radiation disturbance, so there may be potential difficulty in guaranteeing
EMC under other circumstances.

 User shall install and use the device according to the EMC information of
random file.

B-1
Guidance and statements regarding electromagnetic emissions

This pump should be used in the electromagnetic environment for which it was
designed. The customer or user should ensure that the pump is used in an
electromagnetic environment that complies with the following conditions.

Emission test Standard Electromagnetic environment - guidance

RF emissions Group 1 The pump only generates radio frequency


CISPR 11 energy incidentally from its internal functions.

The pump's radio emissions are therefore very


low and will not cause any electromagnetic
interference in nearby electronic equipment.

RF emissions Class A The device is suitable for use in all


CISPR 11 establishments other than domestic and those

directly connected to the public low-voltage

power supply network that supplies buildings


Harmonic emissions N/A
used for domestic purposes.
IEC61000-3-2

Voltage fluctuations N/A


and scintillation
IEC 61000-3-3
B-2
Guidance and statements regarding electromagnetic immunity
This pump should be used in the electromagnetic environment for which it was designed.
The customer or user should ensure that the pump is used in an electromagnetic
environment that complies with the following conditions.

Electromagnetic
Immunity test IEC60601 test level Standard

environment - guidance
Electrostatic ±8 kV contact ±8 kV contact Flooring must be wood,
discharge discharge discharge concrete or ceramic tile. If
(ESD) ±15 kV air ±15 kV air the floor is lined with
IEC 61000-4-2 discharge discharge synthetic materials there
must be a relative
humidity of at least 30%.
Electrical fast ±2 kV power cord ±2 kV power cord The network power source
transient ±1 kV I/O cable must be of typical
(EFT) commercial or hospital
IEC 61000-4-4 quality.
Surge ±1 kV differential ±1 kV differential
IEC 61000-4-5 mode mode

±2 kV common ±2 kV common
mode mode

Voltage drops, <5% UT (drop > <5% UT (drop > The network power source
short 95% UT) 0.5 cycle 95% UT) 0.5 cycle must be of typical
interruptions commercial or hospital
and changes 40% UT (drop 60% 40% UT (drop 60% quality. If the pump needs
IEC 61000-4-11 UT) 5 cycles UT) 5 cycles to run continuously, we
recommend using an
70% UT (drop 30% 70% UT (drop 30% uninterruptible power
UT) 25 cycles UT) 25 cycles supply (UPS) in case of
interruptions to the main
<5% UT (drop > <5% UT (drop > power supply.
95% UT) 5 seconds 95% UT) 5 seconds
Power 400 A/m 400 A/m, 50/60Hz The power frequency
frequency magnetic field must be at
magnetic field a typical level for typical
(50/60 Hz) commercial or hospital
IEC 61000-4-8 environments.

NOTE: UT refers to the voltage of the AC power network before voltage testing.
B-3
Guidance and statements regarding electromagnetic immunity
This pump should be used in the electromagnetic environment for which it was designed.
The customer or user should ensure that the pump is used in an electromagnetic
environment that complies with the following conditions.

IEC 60601 Electromagnetic environment -


Immunity test Standard

Test level guidance


Conducted 10 Vrms 10 Vrms Portable and mobile radio frequency
immunity 150 k–80 MHz communications devices must be used at
IEC61000-4-6 the stipulated distance away from the
Radiated 10 V/m 10 V/m syringe pump or any system components
immunity 80 M–2.7 GHz (including the cables). This isolation
IEC61000-4-3 distance is calculated using appropriate
formulas selected on the basis of the
frequency of the radiation. Recommended
calculation formulas for isolation distance
are:

d  0.4 P

d  0.2 P 80 M–800 MHz

d  0.4 P 800 M–2.7 GHz


Where P is the rated maximum output
power of the transmitter, in W.d is the
recommended distance, in m.
The field strength of the radio frequency
transmitter obtained by measuring the
c
electromagnetic field must be within
compliant levels for every frequency
d
range .
May cause interference if used in close
proximity to equipment with the following

symbols:
Note 1: Between 80 MHz–800 MHz, use a formula for higher frequency bands.
Note 2: The above guidance is not suitable for use in all conditions. Material structures,
objects and people can absorb and reflect electromagnetic waves, affecting
electromagnetic propagation.
B-4
a. 150k ~ 80MHz ISM (Industrial Scientific Medical) Band refers to 6.765MHz ~
6.795MHz, 13.553MHz ~ 13.567MHz, 26.957 MHz ~ 27.283MHz, 40.66MHz ~ 40.70
MHz.
b. Coincidentdet level within the frequency band range of 150kHz ~ 80MHz and
frequency range of 80MHz~2.5GHz of ISM (Industrial Scientific Medical), is intented to
reduce the possibility of interference caused by portable and mobile radio frequency
communications devices being taken to patient areas accidentally. Therefore, accessory
factor 10/3 is used in frequency range of transmitter corresponding to recommended
calculation formulas for isolation distance.
c. Field strengths for radio (honeycomb and wireless) handset base stations and
terrestrial mobile radio receiving apparatus, antenna reception apparatus, and FM and
AM radio/television broadcasts cannot be accurately estimated using a purely theoretical
approach.
The use of electromagnetic field measuring methods should be considered when
estimating the electromagnetic environment produced by a fixed radio frequency
transmitter. If the measured field strengths in the environment in which this pump is used
exceed stipulated RF levels, the pump must be observed to check whether normal
operation is possible. If any abnormal performance is observed, action must be taken
immediately: for example, by changing the position of the pump or moving it to another
environment.
d. In a frequency range of 150 kHz–80 MHz, the field strength should be less than 3 V/m.

B-5
It is recommended that the pump be kept away from
portable/mobile RF communications equipment
The pump can be used in electromagnetic environments in which RF interference is
controlled. In order to avoid electromagnetic interference, the customer or user should
maintain the minimum recommended distance between the pump and portable/mobile RF
communication equipment. The following recommended isolation distances are calculated
on the basis of the maximum output power of the communications equipment.
Calculate isolation distances on the

Transmitter ratings basis of transmitter frequencies (m)


Maximum output 150 k–80 MHz 80 M–800 MHz 800 M–2.7 GHz

power (W)

d  0.4 P d  0.2 P d  0.4 P

0.01 0.04 0.02 0.04

0.1 0.13 0.06 0.13

1 0.4 0.2 0.4

10 1.26 0.63 1.26

100 4 2 4

If the rated maximum output power of the transmitter is not included above, the isolation
distance may be estimated by using the formula in the corresponding column. In the
formula, P is the rated maximum output power of the transmitter as provided by the
manufacturer, in W.
Note 1: Between 80M-800MHz, use a formula for higher frequency bands.
Note 2: 150k~80MHz ISM (Industrial Scientific Medical) Band refers to 6.765MHz~
6.795MHz, 13.553MHz~13.567MHz, 26.957 MHz~27.283MHz, 40.66MHz~40.70 MHz.
Note 3: Accessory factor 10/3 is used in transmitter recommended calculation formulas for
isolation distance within the frequency band range of 150kHz~80MHz and frequency
range of 80MHz~2.5GHz of ISM (Industrial Scientific Medical), to reduce the possibility of
interference caused by portable and mobile radio frequency communications devices being
taken to patient areas accidentally.
Note 4: The above guidance is not suitable for use in all conditions. Material structures,
objects and people can absorb and reflect electromagnetic waves, affecting
electromagnetic propagation.
B-6
B.2 Radio Regulatory Compliance

RF Parameter
Radio devices 2.4GHz Wi-Fi devices
Operating frequency 2412MHz to 2472MHz
Modulation mode DSSS and OFDM
Output power ≤20dBm

The radio device used in this product is in compliance with the essential
requirements and other relevant provisions of Directive 2014/53/EU.

B-7
C Default Factory Settings
This chapter presents some default factory settings.User cannot change the
factory default, but may restore to the default factory settings when it is necessary.

C.1 Alarms

Alarm Setting Factory Default

Volume 4

Alarm sound Sound2

C.2 Interface

UI Factory Default

Contrast 4

C.3 Parameters

Setting Parameters Factory Default

KVO rate 1.0ml/h

Drug library Off

Unit of pressure mmHg


Occlusion pressure 450mmHg

Auto-lock time Off

Reminder time 2min

Time near end 3min

Domestic: Jierui
Commonly used syringes

International: B-D

Bed No. ---

Standby time range 24:00

C-1
C.4 System Time

System Date and Time Factory Default

Time 00:00

Date 01/01/2013

Time format 24h

Domestic: yyyy-mm-dd
Date format

International: dd-mm-yyyy

C.5 Drug Library List

No. Drug No. Drug

1 Aminophylline 21 Magnesium sulfate

2 Amiodarone 22 Mannitol

3 Atropine 23 Metaraminol Bitartrate

Dexmedetomidine
4 24 Metoprolol Tartrate Tablets
Hydrochloride Injection

5 Diazepam 25 Midazolam

6 Diazoxide 26 Morphine

7 Digoxin 27 Nicardipine

8 Diltiazem 28 Nimodipine

9 Dobutamine Hydrochloride 29 Nitroglycerin

10 Dopamine Hydrochloride 30 Noradrenaline Bitartrate

11 Epinephrine 31 Pancuronium
12 Esmolol 32 Phentolamine

13 Fentanyl 33 potassium chloride

14 Flumazenil 34 Propafenone Hydrochloride

15 Furosemide 35 Propofol

16 Heparin Sodium 36 Remifentani

17 Insulin 37 Sodium Nitroprusside

18 Isoprenaline Hydrochloride 38 Suxamethonium

19 Isosorbide dinitrate 39 Urapidil

20 Lidocaine Hydrochloride 40 vasopressin

C-2
D Alarm Information
This chapter presents the alarm information of the syringe pump.Prompt
information for operation guidance will not be presented in this chapter.

The table shows the appropriate countermeasures for each piece of information
related to alarm triggering. If the problem still exists after operating according to the
countermeasures, please contact the company.

Alarm Alarm
Reason Countermeasure

Information level

Press to cancel alarm,


clean the cannula occlusion,
Infusion cannula blocked.
and then press to
[Occlusion] High continue the infusion.

Occlusion pressure Increase the occlusion


threshold is too low. pressure threshold.

Please contact the company


Pressure sensor error. to carry out pressure
adjustment.

Connect to the AC power

[Battery empty] High Battery Empty.


source, press or
to cancel alarm.

Preset infusion volumes


[VTBI done] High
complete. Press to cancel alarm.

Alarm is triggered when


[KVO finish] High KVO mode runs 30 minutes
Press to cancel alarm.
unattended.

The liquid in the syringe is


[Syringe empty] High
emptied. Press to cancel alarm.

Syringe is disengaged
[Syringe
High during the operation of
disengaged] Press to cancel alarm.
syringe pump.

Motor operation error, data Alarm cannot be cancelled.


[System error] High communication error, Please stop operation and
sensor failure etc. contact the company.

D-1
Alarm Alarm
Reason Countermeasure

Information level
The syringe pump performs
no operation during the set Operate the pump to cancel
[Reminder] Low
reminder time after the alarm.
syringe is installed.

Connect to the power source


[Battery low] Low Battery charge insufficient.
to cancel alarm automatically.

Alarm cannot be cancelled.


Charging circuit error or
[System abnomal] Mid-level Please stop operation and
supplying circuit error.
contact the company.

Required time for the


[Syringe nearly remaining liquid within the Cannot be cancelled until
Low
empty] syringe almost reaches empty.
empty time.

1. The alarm will not be


cancelled automatically until
Required time for the the infusion is completed, and
remaining VTBI volumes then switch to [VTBI Done]
[VTBI near done] Low
almost reaches the time for alarm.
completing alarm.
2. Or press to cancel
alarm.

Press to cancel alarm,


[Standby time and then exit standby or
Low Standby complete.
expired] continue standby by pressing

After inserted to
BeneFusion DS5 Infusion 1. Modify pump serial number
Supervision System, the and ensure different serial
[Pump serial pump (no Wi-Fi) has the numbers in BeneFusion DS5
Low
number conflict] same serial number as Infusion Supervision System.
other pump. The alarm is 2. Pull out the pumps with the
triggered by pumps with the conflicting serial number.
conflicting serial number.
Not insert the syringe or
[Syringe not syringe is not correctly
Low Correctly insert the syringe.
correctly inserted] inserted, and starts
infusion.

D-2
Alarm Alarm
Reason Countermeasure

Information level
Communication is
abnormally interrupted for 3
minutes (due to pump been
pulled out or internal
abnormality) after pump
inserted to BeneFusion
DS5 Infusion Supervision
System and communicated
successfully with the Press or restore the
system (no pump serial communication between
[Communication number conflict or version syringe pump(s) and
Low
is interrupted] number compatible) or BeneFusion DS5 Infusion
WLAN interruption for 3 Supervision System or
minutes after pump BeneFusion CS5 Infusion
communicated successfully Supervision System.
with BeneFusion CS5
Infusion Supervision
System. After the alarm is
triggered, infusion of the
pump will not be influenced,
and the pump continues
infusion.

1. After the Auto.


multi-channel relay is
started, the “Forward
channel” pump is pulled out
while infusion is unfinished.
The alarm is triggered by
the relayed pump.
2. After the Auto.
multi-channel relay is
Press , or insert
started, the “Forward
[Relay invalid] Low syringe pump to BeneFusion
channel” pump not finishes
DS5 Infusion Supervision
infusion, and the relayed
System.
pump is pulled out. The
alarm is triggered by the
pulled pump.
3. After the Auto.
multi-channel relay is
started, the “Forward
channel” pump finishes
infusion, the relayed pump

D-3
Alarm Alarm
Reason Countermeasure

Information level
is not in parameter setting
interface, or not set valid
parameters for starting
infusion in parameter
setting interface. The alarm
is triggered by the relayed
pump.

[No battery No battery inserted or


Low Insert battery.
inserted] battery disengaged.

Power cord disconnected


when the network power Connect to the network
[AC power
Low source supply is powered. power source to cancel the

disconnection]
Not connected to the alarm automatically.
network power source.

NOTE

 All alarm sounds can be paused by pressing , except for the


circumstance of [Battery Empty].
D-4
E Symbols and Terms
E.1 List of Units

Abbreviation Meaning

A ampere

℃ centigrade

cm centimeter

dB decibel

g gram

hr hour

Hz hertz

inch inch

k kilo

kg kilogram

kPa kilopascal

l litre

lb pound

m meter

mg milligrams

min minute

ml milliliter

mm millimeters

mmHg millimeters of mercury

s second

μg Microgram
V volt

VA volt ampere

W watt

E-1
E.2 List of Symbols

Symbols Meaning

minus
-

% percent

/ Per; divide; or

~ to

^ power

plus

= equal to

< less than

> greater than

≤ less than or equal to

≥ greater than or equal to

plus or minus
±

× multiply

© copyright

E.3 List of Terms


Abbreviation Meaning

AC Alternating current

Anti-Bolus Anti-Bolus

BOLUS Bolus

CCU(CICU) Cardiac Intensive Care Unit

CE Conformité Européenne

CISPR International Special Committee on Radio Interference

CPU central processing unit


DC Direct current

DPS Dynamic Pressure System

ECU(EICU) Emergency Intensive Care Unit

E-2
Abbreviation Meaning

EEC European Economic Community

EMC Electromagnetic compatibility

EMI Electromagnetic interference

EtO C2H4O

ICU Intensive Care Unit

ID Identification

IEC International Electrotechnical Commission

IEEE Institute of Electrical and Electronic Engineers

ISO International organization for Standardization

KVO Keep vein open

LED light emitting diode

Max Maximum

MDD Medical Device Directive

Min Minimum

MRI magnetic resonance imaging

N/A not applied

NICU Newborn Intensive Care Unit

OR operating room

SN Series Number

TIVA Total Intra Venous Anesthesia

VTBI Volume To Be Infused

E.4 List of Unit Conversion


Unit Symbols Unit Conversion

kPa 1kPa=7.5mmHg=0.145psi=0.01bar

psi 1psi=51.724mmHg=6.897kPa=0.069bar

bar 1bar=750mmHg=14.5psi=100kPa

lb 1 lb=0.454kg

E-3
F Toxic and Hazardous Substances or
Elements

Pb Hg Cd Cr(VI) PBB PBDE


Name of the Parts

Pb Hg Cd Cr(VI) PBB PBDE


○ ○ ○ ○ ○ ○
Front housing

○ ○ ○ ○ ○ ○
Back housing

Device
○ ○ ○ ○ ○ ○
Keys
housing

○ ○ ○ ○ ○ ○
Facing

○ ○ ○ ○ ○ ○
Labels

○ ○ ○ ○ ○ ○
Display Display

○ ○ ○ ○ ○ ○
Host hardware

○ ○ ○ ○ ○ ○
Host Internal cables

○ ○ ○ ○ ○ ○
PCBA

Cartons (K=K ○ ○ ○ ○ ○ ○
crimp paper)

Foam packages ○ ○ ○ ○ ○ ○
Packaging

(EPE)
○ ○ ○ ○ ○ ○
Plastic bag (PE)

Connecting pieces ○ ○ ○ ○ ○ ○

General Power cord ○ ○ ○ ○ ○ ○


○ ○ ○ ○ ○ ○
Battery Battery

○ ○ ○ ○ ○ ○
Accessories Accessories

○: Indicates that this toxic or hazardous substance contained in all of


the homogeneous materials for this part is below the limit
requirement in SJ/T11363-2006.
Remark
╳: Indicates that this toxic or hazardous substance contained in at
least one of the homogeneous materials used for this part is above
the limit requirement in SJ/T11363-2006.

F-1
G Declaration of Conformity

G-1
P/N: 046-005902-00(13.0)

You might also like