Monitor Fetal Ge Corometrics Series 170

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

A Powerful Solution for a Full Range of

Antepartum Applications
Corometrics® 170 Series monitors are designed to provide
the information you need to make quick, accurate and
cost-effective clinical assessments. These compact and
lightweight monitors are designed for antepartum applications
in the office and hospital.

Corometrics 170 Series


Fetal Monitors
System Features and Options Corometrics 170 Series Monitors
• Dual, 9-crystal Nautilus™ watertight ultrasound transducer Model 171 – Single ultrasound and UA; external
permits external monitoring of twins using technology Model 172 – Dual ultrasound and UA; external
that minimizes the need for repositioning. Model 173 – Single ultrasound, FECG and UA; external/internal
• Independent volume controls facilitate easy transducer Model 174 – Dual ultrasound, FECG and UA; external/internal
placement.
• Nautilus watertight toco transducer permits quick, external
acquisition of uterine activity. Its flat-surface design
means it won’t leave a mark on your patient.
• Heart rate offset mode makes it easy to visually distinguish
between the heart rates of twins by allowing you to offset
the secondary FHR by +20BPM.
• Heartbeat coincidence recognition provides a visual
indication when it detects synchronous fetal or maternal
heart-rate signals, indicating that you may be monitoring
a duplicate signal.
Monitors Ultrasound Mode
Environmental Specifications Operating Specifications
Operating Storage Technique Pulsed Doppler with auto
Ambient Temperature 50°F to 104°F 14°F to 131°F correlation processing
(10°C to 40°C) (-10°C to 55°C) Transducer Type 9-crystal
Relative Humidity 30% to 70%, 45% to 65%, Pulse Repetition 2 kHz (all modes)
non-condensing non-condensing Frequency
Pulse Duration 92 μs
FECG Mode
Transmitter Frequency 1.151 MHz
Operating Specifications
Spatial-Average Temporal Isata <5 mW/cm2
Technique Peak detecting, beat-to-beat Average Intensity
cardiotachometer
Focal 20 dB Beam Area 16.6 cm2, at a range = 7 cm
Heart Rate Counting Range 30-240 BPM
Peak Instantaneous 1.8 mW/cm2
Heart Rate Resolution 1 BPM Intensity
Artifact Elimination Service selectable, ±25 BPM Heart Rate Counting Range 50-210 BPM
artifact rejection
Leakage Current Complies with IEC 60 1.1
Countable Input 15 μV to 2 m V peak-to-peak and/or IEC 601.1.1 harmonized
Signal Range national standard; Complies with
Offset Voltage Tolerance ±300 mVdc maximum EN 60601-1-2:2002
(Differential)
Uterine Activity Mode
Maximum Common 20 V peak-to-peak
Mode Voltage Performance Specifications
Common Mode Rejection Balanced: >120 dB at mains Strain Gauge Toco transducer
frequency, with patient cable Range 0-100 mmHg 0-100 relative
Unbalanced 5kΩ RA or LA: units
>110 dB at mains frequency Bandwidth DC to 0.5 Hz DC to 0.5 Hz
Input Impedance Differential: >10 MΩ Resolution 1 mmHg 1 relative unit
Common Mode: >20 MΩ Excitation Voltage +4.0 Vdc +4.0 Vdc
Mains Frequency Rejection >40 dB Zero Set Temperature Drift <0.1 mmHg/° (0.013 kPA/°C),
Leakage Current Complies with IEC 60 1.1 excluding transducer
and/or IEC 601.1.1 harmonized Leakage Current Complies with IEC 60 1.1
national standard; Complies with and/or IEC 601.1.1 harmonized
EN 60601-1-2:2002 national standard; Complies with
Isolation, Mains-to-Patient >5656 Vdc EN 60601-1-2:2002
Strip Chart Recorder Strip Chart Paper
Heart Rate Scale Environmental Specifications
Domestic International Operating Storage
Chart Width 7 cm 8 cm Ambient Temperature 50°F to 104°F <80°F (<26.5°C)
Scaling 30 BPM/cm 20 BPM/cm Relative Humidity 30% to 70%, 45% to 65%,
Range 30-240 BPM 50-210 BPM non-condensing non-condensing
Resolution 1 BPM 1 BPM Power Requirements

Uterine Activity Scale Performance Specifications

Strain Gauge Toco transducer Nominal Line Voltage 100-230 VAC

Chart Width 4 cm 4 cm Line Frequency 50/60 Hz (operates over 47-63 Hz)

Scaling 25 mmHg/cm 25 relative Power Consumption ≤30 VA


units/cm Monitor DC Input 12 Vdcat 2.5 A
Range 0-100 mmHg 0-100 relative Physical Specifications
units
Height 5.75 in (14.6 cm)
Resolution 1 mmHg 1 relative unit
Width 16.75 in (42.5 cm)
Recorder Drive Depth 10.0 in (25.4 cm)
Speeds I, 2 and 3 cm/min Weight 8 lbs (3.6 kg) approx.
Speed Accuracy ±2% over 10 minutes

Certification
UL 2601-1: Designed to meet UL-2601.1 Medical electrical
equipment classified by UnderWriter’s Laboratories, Inc.,
with respect to fire, shock and mechanical hazards in
accordance with UL-2601.1.

CUL: Classified with respect to electric shock, fire,


mechanical and other specified hazards only, in accordance
with CAN/CSAC22.2 No. 601.1.

Warranty
Standard warranty is one year.
© 2007 General Electric Company – All rights reserved.

General Electric Company reserves the right to make


changes in specifications and features shown herein,
or discontinue the product described at any time without
notice or obligation. Contact your GE Representative
for the most current information.

GE and GE Monogram are trademarks of General


Electric Company.

Corometrics is a registered trademark of GE Healthcare.

GE Medical Systems Information Technologies, Inc.,


doing business as GE Healthcare.

Healthcare Re-imagined
GE is dedicated to helping you transform healthcare
delivery by driving critical breakthroughs in biology
and technology. Our expertise in medical imaging
and information technologies, medical diagnostics,
patient monitoring systems, drug discovery, and
biopharmaceutical manufacturing technologies is
enabling healthcare professionals around the world to
discover new ways to predict, diagnose and treat
disease earlier. We call this model of care “Early Health.”
The goal: to help clinicians detect disease earlier,
access more information and intervene earlier with
more targeted treatments, so they can help their
patients live their lives to the fullest. Re-think,
Re-discover, Re-invent, Re-imagine.

GE Healthcare
8880 Gorman Road
Laurel, MD 20753-8500
U.S.A.

www.gehealthcare.com

MIC-0291-07.07-EN-US