Digital Weighing Indicator VP 100: User Manual
Digital Weighing Indicator VP 100: User Manual
Digital Weighing Indicator VP 100: User Manual
VP 100
User Manual
VP 100-51007 V1.0
VECTOR ®
Content
CHAPTER 1 MAIN SPECIFICATION …………………………………………….……......................-2-
Chapter 2 Installation
2.1 FRONT AND BACK VIEW OF THE INDICATOR
Front View
[ON/OFF ] Press this button to start the indicator when it is off; and press it to shut off upon on.
■3. Indicator must be reliably connected to Load cell and shielded-cable of load cell must be
reliably connect ed to underground. If indicator is powered on, the user should not insert or
withdraw the plug in order to protect the indicator and load cell.
■4. Sensor and indicator are static sensitive devices; you must adopt anti-static measures. The
electric wielding operation and other strong electric operation are prohibited. In order to protect
the operator, indicator, and relevant devices, you should install lightning rod in the
thunderstorm frequently happening area
Chapter 3 Operation
3.1 POWER ON AND AUTO ZERO-SETTING
3.1.1 The indicator will perform “999999-000000” to self-checking when turning on. Then it will enter
weighing mode.
3.1.2 When power on, if loading weight on the scale deviates from the zero point, but still within zero
set range, the indicator will set zero automatically; if out of range, it is necessary to adjust the
zero point or recalibrate or reset.
1、P1 x kg Lb change
X=1: kg display
X=2: Lb display
X=3: 20 minutes
X=4: 30 minutes
X=1: 9600
X=2: 4800
X=3: 2400
X=4: 1200
X=6: Keep(Printer)
X=1: No backlight
X=2: Automatic backlight
X=3: Keep lighting
X=1: 0.5e
X=2: 1.0e
X=3: 1.5e
X=4: 2.0e
X=5: 2.5e
X=6: 3.0e
X=7: 5.0e
X=1: 2%FS
X=2: 4%FS
X=3: 10%FS
X=4: 20%FS
X=1: 2%FS
X=2: 4%FS
X=3: 10%FS
X=4: 20%FS
X=1: high
X=2: middle
X=3: low
With RS232(optional) serial communication interface, the -VP 100 indicator can be connected to
computer for communication.
1. All data are ASCII code, every set of which is composed of 10 bits: the 1st is starting bit, the 10th is
stop bit, the middle in between are 8 data bits.
Communication mode as follows:
(1). In continuous mode:
The data transmitted is weight (Gross weight or net weight)
The format of G.W.:ww000.000kg or ww000.000lb
The format of N.W: wn000.000kg or wn000.000lb
Note:The position of above decimal is decided by the decimal set on the indicator.
(2). In command mode:
The indicator performs the corresponding operation according to the command transmitted from
the indicator.
Command R The indicator receives and sends weight data once time (the format is the same as
the continuous mode)
Command T The indicator receive the command and tare (the same as tare key); if no receipt of
the command. The indicator returns CR LF
Command Z The indicator receives the command and zero (the same as zero key); if no receipt
of the command, the indicator returns CR LF.
Chapter 4 Calibration
4.1 Connect load cell properly, then turn on the indicator, press [#] key while it is initialization, it will
enter into the calibration mode and calibrate as following:
Press [TARE] for Select decimal point optional: 0~3, press [#] for
selection of DECIMAL
2 POINT selection confirm
[P X ]
Example:3
Set the full range Press [TARE] for selection of the digit bit; Press
[FULL ]
3 [ZERO] for selection of the digit;
Press [#] for confirm the input of full range
Zero point calibration: Assure there is no load
4 Press [#] when the stable [nOLOAD]
signal is on
Full range point calibration: While inputting the loaded weight,
Press [#] when the value
input is the same as the Press [TARE] for selection of the digit bit; Press
loaded weight and the [ZERO] for selection of the digit; when the input
5 stable signal is on
[AdLOAD] value is the same as the loaded weight and the
digit bit is at the highest bit, press [#] when the
stable signal is on
6 [ End]
Press the calibration It saves the calibration parameter and back to
switch under the lead
sealing board at the back the weighing status. Attention: if no pressing the
7
of the indicator calibration switch, all the parameters won’t be
saved.
4.2 FAST CALIBRATION FOR ZERO POINT AND FULL RANGE POINT
Press [#] while it is initialization, it enters into the calibration mode.
6.1 Turing on the AC power, the indicator will charge the battery automatically. So if you don’t use
battery frequently, you should take battery out.
■ Note: red end is +, black end is -. Wrong connection will destroy indicator.
■ Note: The built-in battery should be fully charged before it is used for the first time.
6.2 Only when you turn off the AC power, and push start key, battery works. Displaying [LouoL] means
the insufficient of voltage, it needs charge.
6.3 When you use the battery first time, you should charge the battery for 20 hours in order to prevent
low voltage resulted from the self leakage of the battery.
6.4 If you don’t use battery for a long time, you should charge the battery for 10-12 hours for each 2
month to prolong using life of battery.
6.5 The battery is easily exhausted products. And it is not granted free guarantee.
VP 100-51007 V1.0 - 10 -
VECTOR ®
Chapter 7 Maintenance
7.1 To guarantee indicator clarity and using life, the indicator shouldn’t be placed directly under
sunshine and should be set in the plain space.
7.2 The indicator can’t be placed into the place where the dust pollution and vibration are serious.
7.3 Load cell should connect with indicator reliably, and the system should be well connected into
ground. The indicator must be protected from high electrical field and high magnetic field.
■ In order to protect the operator, indicator and relevant device, you should mount lightning rod in
thunderstorm frequently happening area.
■ Load cell and indicator are static sensitive device, you must adopt anti static measures.
7.4 It is strictly forbidden to clean the case of indicator with intensive solvents (for example: benzene
and nitro oils)
7.5 Liquid and conducting particle should not be poured into the indicator, otherwise the electronic
components will be damaged and electric shock is likely to happen.
7.6 You should cut off power supply of indicator and relevant device before you pull-in and out the
connecting line of indicator and external device.
■You must cut off power supply of indicator, before pulling out connecting line of load cell.
7.7 During operation, if trouble occurs, operator must pull off the power supply plug immediately, and
user should return this indicator to our company for repair. Non-weighing manufacturer should not
repair it, or by yourself, otherwise further destruction may happen.
7.8 The storage is not granted the free repair guarantee, because it is easily exhausted products.
■ In order to prolong using life, please charge the cell fully before using it. If you don’t use the
indicator for a long time, you must charge the cell every two month and for eight hours/each
charging time.
■ Moving or installation must be carefully taken and must avoid strong vibration, impact and bump in
order to protect the storage cell from being damaged.
7.9 From invoice date, the indicator has a one-year free repair period. If any non-artificially obstacle
about the indicator happens under correct using conditions within this period, the user is allowed
to send the product with its guarantee card (of the correct number) back to our corporation for free
repair. The indicator shouldn’t be taken apart, otherwise free guarantee will be cancelled.
VP 100-51007 V1.0 - 11 -