04 User Manual 3000 Systems V2.0 PDF
04 User Manual 3000 Systems V2.0 PDF
04 User Manual 3000 Systems V2.0 PDF
3000 systems
TN/TS/TX
Version: 2.0.2
User Manual
TN-TS 3000 Rev. 2.0.2
Disclaimer
Save to the extent permitted by law, the Customer hereby undertakes not to alter or modify the whole
or any part of the software incorporated in the Products in any way whatever nor permit the whole or
any part of the software to be combined with or become incorporated in any other programs nor
decompile disassemble or reverse engineer the same nor attempt to do any other such things.
In no event shall [Thermo Fisher Scientific] have any obligation to repair or replace in accordance with
any warranties given with the Product in whole or in part, as the result of
The warranty shall not apply to any software which the Customers alters or modifies.
Thermo Fisher Scientific does not warrant that the software is error-free, will accomplish any particular
result or is fit for any particular purpose of or intended use by the Customer and it is for the Customer
to satisfy yourself that the software is so fit.
The photographs used in this manual can differ from the actual apparatus and set-up’s. All
photographs, drawings and diagrams are for illustrative purposes only.
Copyright
All efforts have been made to ensure the accuracy of the contents of this manual. However should any
errors be detected Thermo Fisher Scientific would greatly appreciate being informed of them
The above notwithstanding Thermo Fisher Scientific can assume no responsibility for any errors in this
manual or their consequences.
Any use of this manual other than, intended by the producer, is prohibited without the prior written
permission of Thermo Fisher Scientific
Content
1. GENERAL INFORMATION
This operating manual has been written with the user of the 3000 system in
mind. For this reason the manual mainly gives instructions detailing
operation of the apparatus and does not give detailed directions for the
maintenance or repair of the apparatus.
The organization making use of the 3000 system must ensure itself that the
operator is qualified to use an instrument of this sort and is capable of
following the recommendations for its use.
The operator should closely follow the instructions contained in this operating
manual in order to ensure the proper operation of the instrument.
This manual makes use of special signs and pictograms in order to highlight
remarks and instructions, which require special attention.
These signs and pictograms are listed in section 1.2 of this chapter.
Chapter 2 – “General Safety Instructions" details the safety pictograms
appearing on the outside of / and inside the apparatus. All directions relating
to these pictograms must be complied with at all times.
Should you encounter problems or have questions, which are not discussed
in this manual, please do not hesitate to get in touch with your Thermo Fisher
Scientific supplier.
Voltaweg 22
2627 BC, Delft
The Netherlands
EMail: [email protected]
(for technical questions and remarks)
[email protected]
(for questions and remarks about the user’s guide)
Web: http://www.thermo.com
1.2. Pictograms
Special Attention
Warning
Electrical Hazard
Heat
Before using the 3000 system carefully read through this operating manual. It is
essential that the persons operating the apparatus or who come into direct
contact with it are fully informed of the nature and existence of all possible
hazardous areas in and around the apparatus.
Thermo Fisher Scientific uses the following safety colors and signs on the
analysis apparatus. These are based on the ISO 3864 international
standards for safety colors and signs. These signs indicate those locations
that require special attention and/or constitute a danger to personnel or the
apparatus.
All personnel coming into either direct or indirect contact with the 3000
system should be fully acquainted with these international safety-markings.
The signs and accompanying texts appearing on or inside the 3000 system
are shown below.
HIGH VOLTAGE!
The general safety rules applicable to the apparatus are given below.
Warning
UV-Fluorescence Permapure
detector
Chemiluminescense
detector
Gas outlet
Detection is based on the principle that SO2 molecules absorb ultraviolet (UV)
light and become exited, then decay to a lower energy state emitting UV light at
a specific wavelength.
The sample flows into the reaction chamber, where pulsating light (at a rate of
10 Hz) excites the SO2 molecules. The condensing lens focuses the pulsating
UV light into the mirror assembly. The mirror assembly contains eight selective
mirrors that reflect only the wavelengths, which excite SO2 molecules.
Internal Vibrational
conversion relaxation
Intersystem
crossing
Internal &
external
conversion
SO*2 → SO2 + hυ 2
In seven steps:
NO + O3 NO2* + O2
NO2* NO2 + hν
The emitted light is detected by the PMT and the generated area is integrated and
converted into a value representing the amount of nitrogen in the sample.
In seven steps:
The following figure displays the flow path of the 3000 system.
(Optional extension with a Halogen detection cell requires an exchange of the
permapure dryer for a (heated) glass connection (part number: ECS300009).)
O2 O2
Dryer gas in Ar
Liquid Module
Perma Pure
Dryer gas out
Molybdene Convertor
Vacuum pump
Ozone Killer
Ozone Generator
PMT detector
Temp. Contr.
Capillary
Samples are introduced into the system by the use of different introduction
modules. (refer to application info)
Note This module is not suitable for low level nitrogen analysis. Instead
use the Module 0019.
The only option in running two different elements and still obtaining the best available
result was redesigning the introduction module.
This resulted in the “model 0019”, capable in handling both sulfur and nitrogen,
without compromising any of the two. The actual results are a perfect match with
results obtained from the individual “liquids modules”.
The “liquids module 0019” supports all know ASTM methods in the boiling range
below 400°C
Major advantage is the capability of the temperature controlled inlet zone and
different expansion chambers, which allows full fine-tuning for various application
purposes.
3.6. Furnace
According to ASTM requirements, the furnace is divided into two temperature
zones.
The temperature in each zone can be adjusted individually, up to 1100°C.
Maximum temperature difference between the zones is approximately 500°C.
3.7. Conditioning
3.7.1. Drying
The gas flow coming from the furnace travels through a Perma pure
membrane dryer. The principle of the nafion perma pure membrane
dryer is the diffusion of water out of the gas stream, by co-polymers with
sulfuric acid ligands.
5
3 7
4
1
2
1. PMT housing
2. Gold plated, heated reaction chamber
3. Capillary location (heated)
4. Gas output
5. Ozone input
6. Analyses gas input
7. Heater connection
8. Photo Multiplier connections
9. Cooling block PMT (Peltier driven)
4.1. Introduction
Before running actual samples on the 3000 system an overall check and
installation procedure has to be carried out.
This chapter will guide you through the installation procedure.
The power cable has to be connected to the 3000 system. The 3000
system has two additional power-outputs of 150 VA.
Warning
Before turning the 3000 system on for the first
time, the voltage setting must be checked. The
small plate beneath the power lead should be
positioned in a way that the 230 V indication is in
the normal position (not upside down) and is read
from left to right when the operator is facing the
back of the apparatus.
The 115 V indication will then be upside down.
Should you need to switch from a 230 V to a 115
Volts supply, pull the plate away from the
instrument so that the prongs of the plug are fully
withdrawn from the apparatus and re-insert the
other way up.
Electrical Hazard
Do not open the cabinet when the power cable is
connected. Remove it from the wall outlet and
disconnect the cable from the analyzer
completely.
Special Attention
After supplying power to the analyzer and starting
the program, the furnaces and heating elements
will rise to the standard temperatures required by
the connected module. (initialization procedure)
When no module is connected, the apparatus will
go into standby mode (i.e. no furnace control).
Figure: com-connector
Warning
Wrong connections could lead to malfunction of
the analyzer or computer.
*Notes*
Preferred pressure settings.
For “manual flow meters” advised pressure is 1-3 bar
For “mass flow meters” advised pressure is 1-2 bar
Gas quality/specifications.
For most applications the above mentioned specifications are valid.
For ultra low level Nitrogen analysis, the specifications may differ.
Usually Argon and Helium are of sufficient quality for analysis below
100 ppb.
Background information: Oxygen will most likely contain a small amount of nitrogen
as well. This is normal, considering the process for oxygen production. The better the
oxygen, the lower the expected background is. Oxygen is also the larger part of the
total carrier gas used in the 3000 system. During combustion, NOx may be formed
from the oxidation of mainly hydrocarbons in the presence of nitrogen gas. (the
background in the oxygen) This will produce higher areas for blanks and ultra low
standards, affecting the detection capability of the 3000 system. Most interference’s
are covered by the system itself. The gas quality however should be addressed at
the source. Check with your local gas supplier for the details.
There are a few differences between the introduction modules for the
3000 system. Two of them have to be installed differently. The
introduction modules are easily interchangeable.
Make sure that the furnace tube (when installed) is supported by the
stainless steel fork.
second flow meter from the right! The right flow meter
corresponds with the oxygen connector on the right!
Special Attention
Avoid touching the surface of the furnace tube with your
bare fingers. The traces of grease left by your fingertips will
create hotspots and reduce the service life of the furnace
tube.
The furnace tube and outlet become very hot during the
operation of the 3000 system. Carelessness may result
in burns. Avoid contact with bare hands. Use special
protective gloves and/or tools for “hot tube removal”.
The safest way for removal is allowing the furnace to
cool down completely.
If the filter shows any damage or pollution, replace it with a new one.
Note:
If the PC is equipped with 2 interface boards and the proper
software, simultaneous analyses are possible. Data cables
are connected to the “I” sockets of the PC boards.
UV fluorescence detector
Note:
Never use a sulfuric acid scrubber in combination with a TN
Coupling (PMECH0368)
Simultaneous Mode
TN Mode Only
4
1
2
3
6
For most applications the mid-range is considered the “ideal working range”
Warning
Before turning the UV detector module on for the
first time, the voltage setting must be checked.
The small plate beneath the power lead should be
positioned in a way that the 230 V indication is in
the normal position (not upside down) and is read
from left to right when the operator is facing the
back of the apparatus.
The 115 V indication will then be upside down.
Should you need to switch from a 230 V to a 115
V supply pull the plate away from the instrument
so that the prongs of the plug are fully withdrawn
from the apparatus and re-insert the other way
up.
Electrical Hazard
Special Attention
With power supplied to the UV detector, the
heating elements will rise to the standard
temperature required for optimum performance.
After installation, the 3000 system is ready for operation. The following paragraphs
should be read carefully. Please note that the software manual should be read before
the 3000 system is switched on.
Special Attention
Read the software manual for the 3000 system before operating
the 3000 system.
1. Preparation of standards
2. Furnace tube and permapure connection
3. Module connection
4. Powering the system
5. Adjust gas flow
6. Gas leakage test
In below shown table, the mass that has to be weighed for the most commonly
applied standards.
Nitrate Ammonia Mix Glycine
Conc. N mg KNO3 mg (NH4)2SO4 mg KNO3 and mg (NH4)2SO4 mg C2H5NO2
Name Formula MM Molfr N mg N/L per L per L per L per L
Potassium Nitrate KNO3 101.1 0.138 1 7.22 4.72 3.61 + 2.36 5.36
Ammonium Sulfate (NH4)2SO4 132.13 0.212 2 14.44 9.44 7.22 + 4.72 10.72
Glycine C2H5NO2 75.06 0.187 5 36.11 23.59 18.05 + 11.80 26.81
10 72.21 47.19 36.11 + 23.59 53.61
25 180.54 117.97 90.27 + 58.99 134.04
50 361.07 235.95 180.54 + 117.97 268.07
75 541.61 353.92 270.80 + 176.96 402.11
100 722.14 471.89 361.07 + 235.95 536.14
Stock solution >>> 1000 7221.43 4718.93 3610.71 + 2359.46 5361.43
E A
The introduction module has to be installed before the five flow meters
are adjusted.
The different settings for water, organic -liquids and -solids are given at
Appendix I.
Note: The number of manual flow meters can differ per system
depending on the configuration.
Procedure 1
For reliable and accurate analysis the 3000 SYSTEM needs to be free of leakage.
Therefore a gas leak test has to be performed.
The next steps explain the gas leak test:
1. Close the flow meters for the oxygen-supply and lower the Argon-flow to
approximately 5 mm.
2. Disconnect the permapure from the particle filter and insert a separate piece of
tubing into it (length about 50 cm and diameter ¼”)
3. Fill a small beaker with water and insert the outer part of the tubing into the water.
4. Observe the flow in the beaker. There still has to be a low flow of bubbles visible.
5. This demonstrates that even at low flow-rates the gas coming from the permapure
dryer, arrives without significant leakage.
The furnace tube, outlet, modules, joints, etc may become very
hot during the operation of the 3000 SYSTEM. Carelessness may
result in burns.
Procedure 2
1. If there are no bubbles visible, a gas leakage has to be fixed. Adjust flows to the
“normal” levels. (setting according current method)
2. Disconnect the gas stream between furnace output and the “permapure
connection”.
3. Manually obstruct the gas stream on the outlet of the furnace tube (using a
septum) and observe the flow meters. They should drop approximately 5-10 mm
within 10-15 seconds. (do not block the flows longer than that) Removing the
septum from the output should clearly release pressure from the tube. This
indicates that no gas-flow leakage is present in the connectors, Module, Furnace-
tube and other joints. The leakage has to be further down the flow path. (leakage
could be found in the septum of the module, any gas connector, glass joints,
loose clamps, mineralized furnace tube, etc.)
4. Connect the “permapure dryer” again and obstruct the flow, coming from it. The
flow meters should drop approximately 5-10 mm within 10-15 seconds. Removing
the septum from the output should clearly release pressure from the permapure
dryer. This indicates that no gas-flow leakage is present in all connected parts in
User Manual 3000 systems V2.0.doc Page 51 of 169
User Manual
TN-TS 3000 Rev. 2.0.2
the flow path. The leakage has to be further down the flow path. (leakage could
be found in the gas connector going to the particle filter, the glass joints, due to
dirt on the grind surface, the furnace tube ball-joint, etc.
5. Resolve any other leakage problem found in the flow path and test it.
6. Finally connect a separate piece of tubing to the outlet connection of the particle
filter (see picture chapter 4.5.4 )and run the test from “procedure 1” again.
7. The final check will be the number 3-5 step from procedure 1.
Following these steps should clearly lead to the “problem” and solution in the flow
path.
The next steps will guide you trough a gas leak test:
In this paragraph a brief explanation will be given how to start working with
the ThEuS™ software supplied with the analyzer. For more details please
refer to the manual on the ThEuS™ CD or to the interactive manual build into
the help function of ThEuS™.
In the following window, click the New Queue button and enter a name for
the Queue that has to be created.
Next, click the Add button. The following window will appear:
After all parameters have been entered, click the OK button. Above shown
example will generate the following queue:
By clicking the Add button again, the next (set of) sample(s) can be added.
Repeat this action until all desired samples have be added to this queue.
When finished adding samples click the analyze button. A similar window will
appear:
If in the box ‘calibration line’ the option new is selected, a new calibration set
will be created. In that case a new name is requested:
In this menu, the new name the calibration order, signal and through zero
has to be selected/entered. The order is in case of a coulometer (TX and TS)
Is used, the order should be ‘Absolute’, otherwise in most cases a first order
(linear) is applied. Note: Afterwards the order can always be changed. If the
system has more then one (parallel) detector the desired signal has to be
selected. ‘Through zero’ is used to force the calibration line through the zero
point (0,0).
For more details please refer to the manual provided with the software.
Before actual samples are introduced an overall performance of the 3000 system
must be performed. If these test results are within the specification, samples- and
calibration procedures may be carried out, without further system proofing.
Before actual shipment to the final destination of the system a “factory calibration” is
carried out. This information is found in the test report, sent along with the system.
In the report test results and calibration data is found.
The data clearly shows the generated area for a specific standard.
This value may be used as an “absolute reference” for system performance.
By injecting standards with an injection speed of 1µL/s, using the appropriate “liquids
module” manually or with the auto sampler a comparable result should be found.
When no analysis are carried out for a short period of time, (one to several
days) it is recommended to put the 3000 system in standby mode.
Warning
Whenever the vacuum pump is shut down, than do close the
oxygen supply to the ozonator! This will prevent the production
of ozone, which could affect the performance of several
devises in the 3000 system.
Heating Setpoint
Inlet temp 25°C
Furnace 1 700°C
Furnace 2 700°C
Converter 25°C
6. TN Water
6.1. Introduction
This manual is an addition to the TN3000 manual and replaces the paragraphs
in this manual concerning the water configurations.
In contradiction to Total Nitrogen analysis in non-aqueous applications,
additional care of the glassware is essential in order to insure a typical life span.
In addition to this special care, some of the parts will have a relatively short life
span. The main problem with aqueous samples is that often salts are present
(i.e. sea water).
O 2/Ar mixture
Combustion
Filter
Water + Salts Detection
6.2. Specifications
160000
140000
120000
100000
area (mV/s
NH4
80000 NO3
Glycin
60000
y = -16.064x2 + 3083.6x - 538.37
R2 = 0.9996
40000
y = -1.46x2 + 3050x + 228.86
R2 = 0.9999
20000
y = -15.97x2 + 2728.9x + 922.68
R2 = 0.9996
0
0 10 20 30 40 50 60
conc. (mg/l)
80000
70000
60000
50000
area (mV/s
NH4
40000 NO3
Glycin
30000
y = -17.581x2 + 3122.7x - 647.06
R2 = 0.9999
20000
y = -4.8538x2 + 3136.6x - 4.5776
R2 = 0.9999
10000
y = -25.542x2 + 2981.6x + 131.84
R2 = 0.9998
0
0 5 10 15 20 25 30
conc. (mg/l)
6.2.3. Summary/conclusion
The TN-3000 water system with Nickel boat introduction and cold trap works
good within the in NVN-EVN 12260 and DIN 38409 normalization providing
the total Nitrogen concentration is below 25 ppm and the Ammonia-Nitrate
relation is relevant. If this relation is not relevant, concentrations up to 2000
ppm can be measured.
Range
Low High
Injection volume 1 ~ 800 1 ~ 800 µL
Typ. Inj. Vol. 200 200 µL
Within NVN/DIN
Ammonia 0.5 ~ 25 - ppm
Nitrate 0.5 ~ 30 - ppm
Extended
Ammonia 0.2 ~ 75 1 ~ 1000 ppm
Nitrate 0.2 ~ 40 1 ~ 750 ppm
Glycin 0.2 ~ 80 1 ~ 1000 ppm
Particles < 250 < 250 µm
Injection opening
Nickel sample cylinder
Magnet
Quartz holder Quartz boat
Little edge for locking the position Quartz boat with Nickel sample cylinder
Turning the screw clockwise will move the boat more towards the furnace.
Note: Make sure that the boat drive (magnet) is in its home position.
Warning
Salts of Nickel can be toxic. Handle with care.
Looking closely to the boat, a little fixation point can be seen. Make sure that
the Nickel sample cylinder is ‘locked’ properly at this point while the injection
opening is pointed up. As shown in the figure above, a little ball of quartz
wool is put into the quartz boot first before inserting the Nickel sample
cylinder. This extra wool is to better fixate the cylinder.
Fixation point
When to replace the Nickel cylinder: During normal use, the cylinder will
remain more or less the same shape. It will, however, become brittle or
porous. Also the color can change to gray and/or green. If the cylinder is
visually damaged (dents or cracks) this might be the result of a bad
alignment of the quartz boat and/or the auto sampler.
some material is stuck between the ceramic inlay and the quartz tube, just
leave it there. This can do no harm to the system. When some quartz
splinters can not be removed with the scraper tool they can also pushed
further into the furnace. Normally this material will not influence the
Scraper tool
measurements.
Is an analyzer is not used for a longer period of time it is common practice to
lower the furnace temperature and reduce the gas flows. In this case, it is
strongly advised not to set the furnace temperature below 750°C. The gas
flows can be reduced to approx. 15 mL/min. If the furnace outlet is
disconnected from the cold trap, the flows can even be set to zero.
The outlet (quartz tube) from the furnace can easily be removed by releasing
the clamp and the wedge nut.
Warning
Please be very careful, the part that is normally inside the
furnace is around 1000°C.
TN filter
Furnace outlettube
The none return tube (or recoil tube) is to prevent water from the cold trap
going back to the furnace in case the gas flows totally stop.
Warning
This recoil tube is an additional safety device to prevent water
going back to the furnace. However, the used must always
disconnect the cold trap prior to setting the gas flows to zero.
The non return tube must be clean, upright and in the correct direction (gas
must flow from the furnace to the cold trap). If the outlet- or recoil tube is dirty
it can be easily cleaned with tap water. Before rebuilding the system, please
dry these parts for example with compressed air.
Outlet Tube
‘Dead zone’
Gas inlet
Cooled water
(with Salts an d
soot particles)
When the system is running fine and the applied method is sufficiently
optimized, the liquid level in the cold trap should be more or less stable and
small amounts of water will be pushed out. This level can be adjusted by
shifting the siphon tube higher or lower (first loosen the red wedge nut). It
can be necessary to adjust the height after some time, depending on the
pollution of the filter (more pollution, higher pressure, higher siphon tube).
Another point that can be adjusted is the gas inlet
However, normally this inlet has to be installed just above the bottom of the
lower chamber. Make sure, no gas bubbles can escape directly to the
overflow siphon. Place a beaker under the overflow tube.
resulting in constantly pushing out all the water. In this case an alternative for
the siphon tube is used. This tube has a little valve. During normal operation,
this valve has to be closed. Depending on the number of samples and the
injected volume, the cold trap will slowl fill up. Please frequently check the
level and when necessary shortly open the valve so, the excess of water is
pushed out. Note: The height of this tube does not have to be adjusted like
the siphon tube.
If the tubing (from the furnace outlet to the cold trap and from the cold trap to
the green filter housing) is polluted or wet, disconnect it and wash it with
ample water. Dry it with compressed air.
Depending on the type of samples, the special paper filter in the green filter
housing can be polluted with salts or other deposits. If so, replace the filter
according to the instructions in de TN3000 manual supplied with the system.
Warning
The deposits on the filter may be toxic.
Daily:
Check if the boat and Nickel cylinder is still intact and properly outlined. If not,
or in case of any doubt replace it with a new one and clean the furnace tube.
Check the liquid level and temperature (is it cold) of the cold trap. When
necessary add or drain water. If the cold trap should be totally empty, also
add a few drops of a strong acid and check during the first few
measurements if the cold trap is not emptied too fast. If this is the case,
adjust the height of the siphon/overflow tube.
If present, empty the beaker under the overflow tube.
Add extra water if the level is too low (less then half the lower chamber).
Note: Acidification will not be necessary.
Weekly:
Clean the cold trap by rinsing it with ample water. Next, fill it to approx. 1 cm
below the upper chamber with water and a few drops of concentrated acid
like sulfuric or phosphoric acid.
Check and if necessary replace the filter in the green filter holder next to the
furnace outlet.
Clean the outlet tube and the tubing connected of the furnace.
Check, and if necessary replace the blue septum in the injection port
(Aluminum funnel on top of a red screw).
Problem: The boat is not moving anymore, but the boat drive is.
Solutions: Check if the boat is still undamaged. If not, replace it.
If the boat is not damaged, clean the outside of the boat and the
inside of the introduction module with water. Beware that the
quartz wool does not get too wet. Make sure to dry the parts
carefully.
Check for blockage inside the furnace tube. Clean if necessary.
Note: Do not remove the tube from the furnace unless absolutely
necessary. Follow the instructions written above if the tube has to
be removed.
Oxygen
Vacuum pump
NO Convertor
Ozone Killer Argon
Ozone Generator
Coldtrap
PMT detector
Temp. Contr.
Capillary
Rev. 2.0.2
Page 74 of 169
User Manual
TN-TS 3000 Rev. 2.0.2
7. External samplers
Warning:
Always stop the software (ThEuS or ECS) running before
switching off the sampler. In not doing so, it can happen that
commands are still waiting to be executed while the sampler is
switch on again. This can cause malfunctions like a needle crash.
7.2.1. Introduction-Description
The Thermo Fisher Scientific Column sampler model ECA1700 is designed to
increase the throughput of the TOX/AOX sample load.
In combination with the automatic sample pretreatment units from Thermo Fisher
Scientific, a fully automated configuration ensures unattended and trouble free
analysis.
The autosampler has a 37 position carousel, introducing samples in combination with
the ECA-AOX module. For Q.C. purposes, the ECA system software allows control
standards, samples and reference samples to be measured during one run. The
software handles “multiple point calibration” and the traditional absolute method,
which is commonly used with the coulometric technique. Data acquisition,
recalculation, reintegration, report generation are just a few features within the
software. In short: Both 3000 system and ECA1700 are fully software controlled.
ECA1700 autosampler
After a queue is built/created and the start button is pressed a plunger pushes the
carbon plus ceramic wool from the glass column into the sample boat. The carousel
moves to the home position and the plunger moves downward, closing the stopper
and seal. At this point the module plus furnace tube are gas tight. The boat, which
carries the carbon plus ceramic wool, moves into the high temperature combustion
zone. The content is oxidized completely and a signal appears on the screen.
After combustion, the boat returns to the injection area and a scraper takes out the
remaining ashes. The ashes fall into a collection area, which after a full run is
emptied, using a vacuum device. In the mean while the boat is being cooled to
prepare it for the next sample introduction.
When the curve on the screen reaches the baseline again, the analysis may be
terminated and the curve is automatically integrated. A result appears on the screen
and a printout can be made. In the automatic mode, the next sample is being
introduced.
The ECA-AOX module, necessary for running samples in combination with the
ECA1700, differs in a few details from the standard AOX module.
Check if the mounting brackets are already in place. If not position them as explained
in the picture.
Brackets
Brackets
Replace the normal boat pusher with the extended one. (mounted on the carriage in
the boatdriver, which could be pre-installed)
Extended items
Standard Boat items
Note that the threads of the ECA bracket are pointing outwards!
(viewed from the furnace tube)
Install the ECA-AOX module, connect the clamp, which joints the furnace tube and
the module and insert the oxygen connector.
Check if the glass parts are aligned in the correct manner. (see picture next page)
ECA fasteners
Rubber sealing
Fastener ECA
ECA Carousel
Position the ECA1700 sampler over the 3000 system, align it with the module and
lock the sampler, using the supplied fasteners.
PC control cable
Electrical Hazard
Warning
Do not temper with- or bypass the safety switch.
Operate the ECA1700 with closed access door only.
Not complying could result in injury.
The column pusher uses excessive force to empty the columns.
Stay away from the pusher during operation and keep the access door
closed!
Basic settings for the 3000 system in the ECA1700 mode are similar to the “normal”
AOX methods.
System flows, temperatures and coulometric settings are the same as for AOX
according the batch and other manual methods.
Depending on the specific application the fan will start cooling at maximum capacity
(100%) as soon as the boat returns. The duration of cooling is approximately 4
minutes. Before the next sample gets introduced, the cooling will have to be shut
down or reduced to 10%. Check if this setting is correct in the “Analysis method”
window.
“Normal cooling” is 10% “Forced cooling” is 100%
Note: The cooling time must be programmed in step 4. This is the “Home”
position of the boat-program. This step controls the cooling time before the
next sample is introduced.
Not doing so will affect the next result because of too high temperature in the
entrance zone.
7.3.1. Introduction-Description
The Thermo Fisher Scientific AOX-Solid auto sampler model ESA2000, is designed
to increase the throughput of the sample load. In combination with the 3000 system,
unattended and trouble free analysis are a reality. The auto sampler has a 48
position carousel, introducing the samples in combination with the special sample
cups, made from quartz glass or metal. A laser beam controls the in- and out coming
sample cups, to make sure that there is no pile up in the furnace tube or module. For
Q.C. purposes, the ESA system software allows control standards, samples and
reference samples to be measured during a run. The software handles “multiple point
calibration” as well as the traditional “absolute method”, which is commonly used with
the coulometric technique. Data acquisition, recalculation, reintegration, report
generation are just a few features within the software. In short: Both 3000 system
and ESA2000 are fully software controlled.
Figure: ESA2000
After a queue is listed and the start button is pressed the arm-hand lifts a Frit or
sample-cup from the carousel and puts it into the ESA module on top of the boat.
The grabber-hand has a neoprene sealing ring, which closes the entry of the module,
creating a gas-tight environment.
The boat, which carries the Frit plus carbon, moves into the high temperature
combustion zone. The content is oxidized completely and a signal appears on the
screen.
After combustion, the boat returns to the injection area and the Frit or sample-cup is
taken out and placed onto the carousel again. A laser beam is used to check if the
sample cups placed in the module are taken out as well. An error message will
appear and prompts the operator to take corrective action. (frit detected yes/no)
The laser control will prevent pile-up and malfunction. In the mean while, the boat is
being cooled as soon as the boat starts to retract. The cooling action is necessary to
lower the temperature of the boat and module, prior to the next sample introduction.
As soon as the curve on the screen reaches the baseline again and maximum
analysis time is reached, the analysis may be terminated and the curve is
automatically integrated and stored. The result appears on the screen and a printout
can be made. In the automatic mode, the next sample is being introduced.
Next a queue is listed and the start button is pressed the arm-hand lifts a Frit or
sample-cup from the carousel and puts it into the ESA module on top of the boat.
The grabber-hand has a neoprene sealing ring, which closes the entry of the module,
creating a gas-tight environment.
The boat, which carries the sample cup with content, moves into the high
temperature combustion zone. The content is oxidized completely and a signal
appears on the screen.
After combustion, the boat returns to the injection area and the sample-cup is taken
out and placed onto the carousel again. A laser beam is used to check if the sample
cups placed in the module are taken out as well. An error message will appear and
prompts the operator to take corrective action. (cup/crucible detected yes/no)
The laser control will prevent pile-up and malfunction. In the mean while, the boat is
being cooled as soon as the boat starts to retract. The cooling action is necessary to
lower the temperature of the boat and module, prior to the next sample introduction.
As soon as the curve on the screen reaches the baseline again and maximum
analysis time is reached, the analysis may be terminated and the curve is
automatically integrated and stored. The result appears on the screen and a printout
can be made. In the automatic mode, the next sample is being introduced.
Mounting brackets
Docking-station
Fastener
Check if the mounting bracket for the Docking-station is already in place. If not
position it as showed in the picture.
Verify if the base plate including gas connections and ID connector is repositioned in
a way as the picture below.
Positioning ESA
Spring
Rear- end support
Boat positioner
The boat positioner is used to adjust the home position of the returning boat.
The exact home position is important for the ESA hand. The hand is adjusted with
both soft- and hardware. The magnet, which controls the movement of the boat, has
some tolerance in the positioning. In order to take away this tolerance, a “mechanical
home” position is created. An adjustable plastic thread is used to dictate the
maximum allowed distance. The boat can not travel beyond this point and all
tolerance is taken out of the motion.
ESA Module
Teflon Boat
positioning
Adjustment screws
ESA position lock
Center hand/arm
This takes care of the “hardware alignment” for the 3000 system ESA2000
configuration. Besides this procedure a “software calibration” must be carried out.
This procedure is explained in the software part of this chapter.
ESA2000 rear-view
Electrical Hazard
Warning
Do not temper with- or modify the laser settings or adjustment.
Laser light can be a threat to your eyesight.
Even though the intensity is not of a high level, do not look directly into
the beam.
Operate the ESA2000 with supplied cover only.
Not complying could result in injury.
Readjustments should be carried out by trained Thermo Fisher Scientific
technicians only!
Basic settings for the 3000 system in the ESA2000 mode are similar for the “normal”
AOX methods.
System flows, temperatures and coulometric settings are the same as for AOX,
according the batch and other manual methods.
However, for solid methods refer to application notes or reports, for correct settings.
Special Attention
Before the power is switched on, move the arm into a “neutral position”.
Manually move the arm in the approximate center of the “left-right, up-
down” position.
When the ESA is positioned in the Docking-station the arm should be positioned into
the approximate center of the unit. (check the above “special attention”)
Pressing the “Init” button with the power supplied and data cable connected to the
ESA2000 and a powered 3000 system, should activate the “Arm-hand” and sample
carousel. They both are now moving to their home positions and perform the
initialization procedure. The ESA checks if position 48 is empty. If this is confirmed it
checks if there is no sample cup left in the boat. If one is found, the ESA picks it up
and places it on position 48. The system will produce a message to the screen,
prompting the user to remove it from 48.
For first time operation a calibration procedure has to be carried out, for optimal
performance. Sensors inside the ESA2000 should be positioned correctly.
This is normally carried out in the factory. In case of doubt about the precision of
sensor positioning, contact a Thermo Fisher Scientific representative to correct
and/or assist with this procedure in case of readjustment.
Warning
Do not temper with the laser beam or it’s adjustment!
This must be carried out by a trained Thermo Fisher Scientific
representative.
Direct Sample positioning is also possible. Just key-in a number in the “Sample
Position” window. (1-48) Pressing “Enter” on the keyboard will trigger the action and
moves the carousel to the desired position.
Note: Movement of stepper motors can be “noisy and/or irregular” when data
traffic on the PC gets heavier or initialization is carried out. This should not
affect the sample handling.
1. To check if this procedure is correct, close the bypass gas flow (only for this
specific test) and reduce the oxygen carrier to approximately 10 mm (manual flow
meters) or 25 mL/min for mass flow meters. Observe the Sulfuric acid in the
scrubber and check for bubbles. If there are no bubbles appearing in the acid,
increase the number of steps, to get more pressure on the sealing between the
rubber and the glass wall. If necessary, repeat the procedure until a gas tight
sealing is reached.
Warning
Warning
Prevent the hand from crashing onto the carousel. The alignment or
mechanical features of the stainless steel hand may be lost, because of
mechanical damage.
Depending on the specific application the fan will start cooling at maximum capacity
(100%) as soon as the boat returns. The duration of cooling is approximately 4
minutes. Before the next sample gets introduced, the cooling will have to be shut
down or reduced to 10%. Check if this setting is correct in the “Analysis method”
window.
“Normal cooling” is 10% “Forced cooling” is 100%
For more details, refer to the Software manual or the Thermo Fisher Scientific web-
site: www.thermo.com
The principles of the design of the modules are based on the requirements for the
standard analyses of organic halogen and sulfur compound/content in samples of a
divergent nature and origin. Various fractions of organic halogen (X) and sulfur (S)
compounds are analyzed by these standard analytical methods. The modules can
also be used for non-standard analyses.
Warning
Do not use this introduction module for samples containing significant
quantities of (combustible) organic components, like oil, waxes etc.. The
sample is introduced in a flow of pure oxygen. Any combustible
component will ignite prematurely.
8.4. Furnace
The sample is placed in a boat in a stream of oxygen and/or argon into the furnace.
The furnace is divided into two temperature zones. (according ASTM guidelines)
The temperature in each zone can be individually adjusted to all temperatures up to
1100°C. The sample incineration time in each zone can also be adjusted.
In the furnace the sample is oxidized into gas. In the event of incomplete incineration
soot deposits may be formed. (refer to chapter 13 “trouble shooting”)
When Ei-r changes, the coulometer passes current (1) through the generator circuit,
thus forming silver ions at the anode. Hydrogen is then formed at the cathode.
The silver ions react with halide ions until these have disappeared or, in other words,
until the concentration of silver ions has returned to the same level as the initial
concentration of roughly 10-7 M.
The solubility product indicates the relationship between the concentrations of silver
ions and halide ions in a saturated solution. For silver chloride this is:
As long as Ei-r continues to change current will pass through the generator circuit and
silver ions will be generated. As Ei-r approaches its initial potential the current will fall
off. Once Ei-r has reached its initial value again, i.e. when all the halide ions have
been converted, the generator curve (1) will again be zero.
Ei-r + Ebias
Ebias is a fixed potential of -0.315 V that serves, among other things to compensate
the electrical resistance of the cell. The total of Ei-r + Ebias is equal to about 0.5 V.
The quantity of Ag+ ions which is generated by the flow of the current through the cell
and consequently the quantity of deposited halide ions, behave in accordance with
Faraday's Law:
t
1 Q
m= ∫
n. f 0
I . dt =
n. f
where: Q = the quantity of charge in Coulomb passing through the electrode surface in time t.
m= mass of converted substance [mol]
F= Faraday's constant (96487 C/mol)
n= number of electrons involved in the electrode reaction (here n = 1)
t= titration time
I= current (as a function of time)
Integrating the current over time provides the quantity of charge used for titrating the
halide. The quantity of charge, which has been titrated is shown in µC in the right
hand column in the measurement display. The quantity of halide present in the cell
can then be calculated from this figure. The calculations are carried out automatically
by the Thermo Fisher Scientific 3000 SYSTEM software.
The cathode is contained in a glass tube, and is separated from the measurement solution by
means of a frit. The purpose of this frit is to exclude the effect of disruptive reactions around
this electrode.
0= Outlet
G= Gas inlet tube
Acid
The titration cell contains a solution of 75% acetic acid in
water. The concentration of silver ions in solution is
maintained at a constant level of about 10-7 M.
Where: V = Voltmeter
Amp = Ammeter
RΩ = Variable resistance
R = Reference electrode
I = Indicator electrode
A = Anode
C = Cathode
Generation path:
- Anode A (silver)
- Cathode C (platinum)
Acid
The titration cell contains a solution of 75% acetic acid in
water. The concentration of silver ions in solution is
maintained at a constant level of about 10-7 M.
When Ei-r changes the coulometer passes current (1) through the generator circuit.
This results in the conversion of I- into I2 at the platinum anode, followed by the
formation of I3- as indicated below:
2 I- -------- > I2 + 2e
This process will continue until Ei-r no longer changes, i.e. until the concentration of I3-
has returned to the value it had before the addition of S02. The generator current will
then be equal to zero and all the S02 will have been converted into sulfate.
Here too the difference in potential between anode and cathode of:
Ei-r + Ebias = about 0.5 V
Ebias is in this case equal to + 0.145 V
The quantity of I- converted can be calculated from the charge needed for the
conversion. Then the quantity of S02 present in the cell can be determined according
to Faraday's Law.
The same remarks apply here to the titration curve as were made with respect to
halide analysis.
Acid
The titration cell contains a solution of 0.1% KI, 0.1% NaN3 and
0.5 % acetic acid in deionized water.
The NaN3 (sodium azide) is added to prevent possible disturbances to the
measurement caused by nitrous oxides forming in the cell.
Warning
Where: V = Voltmeter
Amp = Ammeter
RΩ = Variable resistance
R = Reference electrode
I = Indicator electrode
A = Anode
C = Cathode
Generation path:
- Anode A (platinum)
- Cathode C (platinum)
Acid
The titration cell contains a solution of 0.1% KI, 0.1% NaN3 and
0.5 % acetic acid in deionized water.
The NaN3 (sodium azide) is added to prevent possible disturbances to the
measurement caused by nitrous oxides forming in the cell.
In total there are four introduction modules for the 3000 SYSTEM. Each of them has
to be installed differently. The introduction modules are easily interchangeable.
Warning
Do not use this introduction module for samples containing significant
quantities of (combustible) organic components, like oil, waxes etc.. The
sample is introduced in a flow of pure oxygen. Any combustible
component will ignite prematurely.
Special Attention
Avoid touching the surface of the furnace tube with your bare fingers. The
traces of grease left by your fingertips will create hotspots and reduce the
service life of the furnace tube.
The furnace tube and outlet become very hot during the operation
of the 3000 SYSTEM. Carelessness may result in burns.
1. Remove the Permapure scrubber and install the stainless steel bracket. (A)
2. Suspend the scrubber on the bracket. (B)
3. Place the “connection furnace scrubber”. (C)
4. Connect the inlet of the scrubber to the outlet of the heated connecting tube (ball-
joint), using clamp no. 12. (D)
5. Insert the connector (electrical) into the socket. (E)
6. Adjust the temperature of the outlet. (typical 300°C)
7. Adjust the gasses and fill the scrubber with sulfuric acid.
D
E
9.1.7. NOx-scrubber
Sometimes it is necessary to use a NOx-scrubber. Make sure that a clean and dry
drip tray is placed in the “Cell compartment”. The NOx-scrubber is connected directly
onto the gas inlet of the titration cell. Follow the next steps for installing the scrubber:
TOP VIEW
HALOGEN CELL SULFUR CELL
G
A
D
C
1. Insert the connector through the hole of the Titration compartment (5-pole DIN-
connector (A)
2. Place the “Titration Cell” on the stirring motor (B)
3. Connect the inlet coming from the scrubber onto the gasinlet (without NOx
scrubber) (E)
After installation the 3000 SYSTEM is ready for operation. The following paragraphs
should be read carefully, otherwise the measurement is worthless. Please note that
the software manual should be read before switching on the 3000 SYSTEM.
Special Attention
Read the Thermo Fisher Scientific software manual for TOC systems
before operating the 3000 SYSTEM.
Acids
Read the chemical safety sheet before working with sulfuric acid and
follow the safety instructions.
AOX
Warning
Do not use this introduction module for samples containing significant
quantities of (combustible) organic components, like oil, waxes etc.. The
sample is introduced in a flow of pure oxygen. Any combustible
component will ignite prematurely.
AOX sampling
Sampling
Storage at 4°C.
DOC analysis
Choose method
pH 2-3 pH 2-3
AOX analysis
Washing by 25 ml water
AOX analysis
Storage at 4 degr. C
pH 2-3
POX analysis
TOX sampling
Sampling
POX method
AOX batchmethod
EOX
Sampling
50 g sample
100ml extract in a
separating funnel
EOX analysis
Sampling
Store at 4 degr. C
pH = 2
Water layer pH = 9
Separation of organic phase Extract with 100 ml petroleum ether
Separation of organic phase
Collect extracts
Concentrate to 5 ml by evaporation
Concentrate to 1 ml
EOX analysis
Sampling
EOX analysis
Close all flow meters by turning the black knob in the clockwise direction. (do not
over- tighten them)
1. Open the gas headers on the gas supply line.
2. Set the oxygen bypass flow meter to 10 mm (35 mL/min) by turning the black
knob counter clockwise. (left flow meter)
The introduction module has to be installed before the next two steps can be
executed.
3. Set the oxygen flow meter to 50 mm (330 mL/min), minimum 40 mm (200
mL/min) (middle flow meter)
4. Set the argon flow meter to 40 mm (200 mL/min), minimum 25 mm (90 mL/min)
(right flow meter)
(Only operating together with the liquids and solids module.)
Note: Instead of manual flow (Rota) meters, digital mass flow controllers can be
installed. In this case, all flow settings will be automatically done by the ThEuS
software.
Special Attention
Make sure that the drip tray is clean and dry and has been placed in
the scrubber compartment.
Place the small beaker under the scrubber.
1. Install the selected “sample introduction module” connect it and open the gas
flow.
2. Open the scrubber compartment and position the drip-tray.
3. Shut both stopcocks of the scrubber, the red handles in horizontal position
4. Remove upper glass stopper from the scrubber
5. Fill the scrubber buffer with 95-98 % sulfuric acid
6. Remove any spilled sulfuric acid.
7. Replace upper glass stopper.
8. Check that the lower stopcock of the scrubber is shut. Fill the absorber of the
scrubber to the red line by opening the upper stopcock (red handle in vertical
position) and lift up glass stopper until scrubber is filled.
9. If the gas is connected take care that a flow is running through the scrubber
immediately after filling.
10. Close upper stopcock (red handle in horizontal position)
11. Close the scrubber compartment.
10.1.8.2. NOx-scrubber
Fill the scrubber in the following order:
1. Open the titration cell compartment and place the drip-tray.
2. Disconnect the outlet of the splash trap with the NOx-scrubber
3. Fill the NOx scrubber with 95-98 % sulfuric acid till the mark on the glass.
4. Remove any spilled sulfuric acid.
5. Reconnect the outlet of the splash bulb to the NOx-scrubber, using clamp no. 12.
6. If the gas is connected take care that a flow is running through the scrubber
immediately after filling.
7. Close the compartment.
Fill the electrode with the halide electrolyte solution and add the four electrodes like
described in paragraph 8.6.1.1. Fill the cell with the correct electrolyte solution, up to
the level mark. Before and during cell replenishing, the gain must be set to zero in
order to prevent heavy pollution of the cell and/or damage to the coulometer
electronics. Once the cell is filled and the electrodes have been inserted, the cell can
be placed onto the stirrer motor.
If the cell replenishment was carried out correctly, with the bias for the halide-
analysis at -315 mV and for the sulfur-analysis at 135-145 mV, a positive signal
should be shown. The current has to have a value between approximately 200 and
600 µA.
Depending on the speed of the cell (condition of the electrodes) and gain level, the
signal should return within 3 to 5 minutes. If upon return, the signal at first drops to
negative and then returns to a positive signal to arrive at a base line between
approximately -0.5 to +1 µA, you may have a case of overshoot due to the excess
gain.
For the halide-cell approximately 5 µL of 0.1 to 0.2 M NaCl has to be injected directly
into the cell before decreasing the gain. While doing this, make sure the needle tip is
fully immersed in the electrolyte in order to obtain a regular signal. This procedure
must be carried out twice. If any overshoot occurs, the gain must be decreased.
After carrying out the procedure of above, the recovery must be tested.
Halide-analysis:
A 2 mM solution of NaCl can be used. Inject 10 µL of the solution. In theory, this
should result in a yield of 1.92 mC, but in practice this value could be exceeded with
approximately 10% due to the contribution of light that enters the cell when the cover
is opened, a statically charge coming from the syringe and operator etc.
If the test results are correct, the cell, coulometer and PC have been tested and are
working properly.
If the subsequent connection of the gases results in insufficient yield or
reproducibility, the cell need not be included in any further trouble-shooting
procedures.
Add the four electrodes like described in paragraph 8.6.2.1. Fill the cell with the correct
electrolyte solution, up to the level mark. Before and during cell replenishing, the gain
must be set to zero in order to prevent heavy pollution (incorrect titration) of the cell
and/or damage to the coulometer electronics. Once the cell is filled and the
electrodes have been inserted, the cell is placed onto the stirrer motor. Connect the
sulfur cell to the output of the splash bulb with the gasses turned on. Turn the gain
back on. (level set at 10%) If the cell replenishment was carried out correctly, with the
bias for the sulfur-analysis at 135-145 mV, a positive signal should show. If the signal
goes negative, put the gain to zero again and inject a sample/standard first. This
procedure is carried out, using the “liquids module” and an organic sulfur standard. If
the sample introduction is finished the gain may be turned to the previous level again.
Check if the cell runs “positive” now. Depending on the speed of the cell (condition of
the electrodes) and gain level, the signal should return within 3 to 5 minutes. If upon
return, the signal at first drops to negative and then returns to a positive signal to
arrive at a base line between approximately -0.5 to +1 µA, you may have a case of
overshoot due to the excess gain.
After carrying out the above procedure, the recovery must be tested.
Procedure 1
For reliable and accurate analysis the 3000 SYSTEM needs to be free of leakage.
Therefore a gas leak test has to be performed.
The next steps explain the gas leak test:
1. Close the flow regulator for the oxygen-supply and lower the Argon-flow to
approximately 5 mm.
2. Observe the flow in the Sulfuric Acid of the scrubber. There still has to be a low
flow visible.
3. Shut down the stirrer of the Titration Cell and observe if a similar bubble appears.
This demonstrates that even at low flow-rates the gas arrives into the Cell and at
the same time proves that all other connected items are gas-tight.
The furnace tube, outlet, modules, joints, etc may become very
hot during the operation of the 3000 SYSTEM. Carelessness may
result in burns.
Procedure 2
1. If there are no bubbles visible, a gas leakage has to be fixed. Adjust flows to the
“normal” levels. (setting according current method)
2. Disconnect the gas stream between furnace output and the “connection furnace
scrubber”.
3. Manually obstruct the gas stream on the outlet (using a septum) and observe the
flow meters. They should drop approximately 5-10 mm within 10-15 seconds.
Removing the septum from the output should clearly release pressure from the
tube. This indicates that no gas-flow leakage is present in the connectors,
Module, Furnace-tube and other joints. The leakage has to be further down the
flow path. (leakage could be found in the septum of the module, any gas
connector, glass joints, loose clamps, mineralized furnace tube, etc.)
4. Connect the “connection furnace scrubber” and “scrubber” again and obstruct the
flow, coming from the scrubber. The flow meters should drop approximately 5-10
mm within 10-15 seconds. Removing the septum from the output should clearly
release pressure from the scrubber. This indicates that no gas-flow leakage is
present in all connected parts in the flow path. The leakage has to be further
down the flow path. (leakage could be found in the gas connector going to the
scrubber, the glass joints, due to dirt on the grind surface, the furnace tube,
connection furnace scrubber, scrubber-inlet, -outlet etc.
5. Install the splash bulb again including the tubing and repeat the “obstruct”
procedure.
6. The final check will be the number 3 step from procedure 1.
Following these steps should clearly lead to the “problem” and solution in the flow
path.
During normal operation small maintenance has to be carried out regularly. The next
paragraphs will describe the following subjects:
10.2.2.1. Scrubber
If the capillary in the scrubber (final gas inlet into the sulfuric acid) appears black, this
may cause blockage or serious restriction to the flow. In addition, the "soothing" may
cause memory effects to occur. Therefore, a clean scrubber is necessary.
Cleaning procedure:
1. Remove all H2SO4 from the scrubber.
2. Remove the drip tray from the system.
3. Disconnect the connecting plug to the scrubber.
4. Remove the bypass gas connection from the glassware.
5. Remove the scrubber from the system and remove all “stop cocks”.
6. Rinse off as much of the pollution as possible (taking care not to damage the
heater ribbon).
7. Rinse out the scrubber with water.
8. Heat the scrubber in an oven at 500 oC for approx. 30 minutes (leaving the
plastic connection plug to protrude through the oven door!).
9. After re-installation, make sure the heater ribbon reaches the set temperature
again (300°C) and carry out the gas test.
Check
1. Set the 02 flow at 10 mm (35 mL/min) and close the bypass.
2. Check gas flow through the scrubber.
3. Insert an empty sample bottle with a septum in good condition (in case of
doubt, use a new one) and insert both needles into the empty sample bottle.
4. Press the analyze button at the control-window and check the gas flow that
should be equal to that of point 2.
If both gas flows are not equal, a leaking septum or the tubing not fitting properly into
the couplings, blocked needles, etc may cause this.
(further info on gas-test, trouble shooting 10.1.13)
1. EOX / EOS Extractable organic halogen or organic Sulfur analysis method for
liquid applications
2. AOX / TOX Adsorbable or Total organic halogen analysis method for liquid
applications
3. POX purgable organic halogen analysis method for liquid applications
4. TX / TS-liquids Total halogen or sulfur analysis method for liquid applications
5. TX / TS-gas Total halogen or sulfur analysis method for gas applications
6. TX / TS solids Total halogen or sulfur analysis method for solid applications
Other methods are custom made by Thermo Fisher Scientific or created by customers
themselves. The software manual describes how these methods can be
created/changed.
An example of setpoints is given in “Appendix I : Settings”. For more information see
software manual.
1. Disconnect the gas supply to the NOx-scrubber. (remove the clamp between the
splash-bulb and the NOx-scrubber)
2. Drain the sulfuric acid scrubber into the small beaker.
3. Close header gas supply or close flow meters by turning the black knob
clockwise. (observe if all acid has left the scrubber before the header is closed)
4. Set the following temperatures:
Heating Setpoint
Furnace 1 700°C
Furnace 2 700°C
Outlet 25°C
This chapter gives a brief description of the actual procedure for analysis.
Basically there are two possibilities for running samples: Manually or automatically.
Once the unit is switched on and all setting like temperature, flows etc have reached
their desired levels, the actual analysis may be started.
Some system configuration could come without a “boat driver”. This device is
responsible for the actual movement of the sample boat and automatic plunger
pushing for syringe injection (controlled by stepper motor and software).
In case of not having a “boat driver,” the sample boat and plunger from the syringe
are controlled by hand.
The following steps should take you quickly through the procedure and run an actual
sample.
1. Open a new queue list and key-in the necessary parameters (for more details,
refer to the software manual)
2. Open “analysis” in the “controller window” and select the box “methods”. Choose
the appropriate POX analysis method according the application.
3. Open “evaluation” and select the “absolute” calculation method (the calibration
field at this point is not used)
4. Place the sample in the POX module (for sample prep refer to: 10.1.6.2) and
allow the first sample to heat up to approximately 60°C, which usually takes
about 8 minutes. In the mean while the second sample is placed to avoid waiting
time for the next analysis.
5. Lower the injection head of the POX module and observe if both needles have
penetrated the septum sufficiently.
6. Select or add in the queue list the sample to be analyzed and press “analyze”.
Depending on program settings the analysis terminates when sample is
completely purged and the baseline reaches its normal baseline level again.
Manual termination is possible by keying in a lower number of minutes/seconds
analysis time or putting the cursor at the finishing baseline and click on the right
mouse key. This will automatically key in the selected time.
7. The next sample should have reached operational temperature and is ready for
analysis. Select the sample to be analyzed and repeat the steps starting from 5.
Warning
Do not use this introduction module for samples containing significant
quantities of (combustible) organic components, like oil, waxes etc.. The
sample is introduced in a flow of pure oxygen. Any combustible
component will ignite prematurely.
The following steps should take you quickly through the procedure and run an actual
sample.
1. Open a new queue list and key-in the necessary parameters (for more details,
refer to the software manual)
2. Open “analysis” in the “controller window” and select the box “methods”. Choose
the appropriate AOX analysis method (batch or column) according the application.
(for more details see chapter 10.1.6.1)
User Manual 3000 systems V2.0.doc Page 136 of 169
User Manual
TN-TS 3000 Rev. 2.0.2
3. Open “evaluation” and select the “absolute” calculation method (the calibration
field at this point is not used)
4. Place the prepared carbon-sample in the AOX module.
5. Select or add in the queue list the sample to be analyzed and press “analyze”.
Depending on program settings the analysis terminates when sample is
completely oxidized and the baseline reaches its starting point again. Manual
termination is possible by keying in a lower number of minutes/seconds analysis
time or putting the cursor at the finishing baseline and click on the right mouse
key. This will automatically key in the selected time. Note that a certain “cooling
time” is needed for the sample boat to reach an acceptable temperature again. A
hot boat could oxidize the sample in an early stage, producing erratic numbers.
(typical cooling time is 4 minutes)
6. Now the 3000 SYSTEM is ready for the next analysis. Select the sample to be
analyzed and repeat the mentioned steps starting from 4.
1. Open a new queue list and key-in the necessary parameters (for more details,
refer to the software manual)
2. Open “analysis” in the “controller window” and select the box “methods”. Choose
the appropriate “Solids analysis method” according the application. (for more
details check your application note or check the web: www.thermo.com )
3. Open “evaluation” and select the “absolute” calculation method (the calibration
field at this point is not used)
4. Place the sample in the Solids module. (by weight in a sample cup or via syringe
by volume injection in a sample cup)
5. Select or add in the queue list the sample to be analyzed and press “analyze”.
Depending on program settings the analysis terminates when sample is
completely oxidized and the baseline reaches its starting point again. Manual
termination is possible by keying in a lower number of minutes/seconds analysis
time or putting the cursor at the finishing baseline and click on the right mouse
key. This will automatically key in the selected time. Note that a certain “cooling
time” is needed for the sample boat to reach an acceptable temperature again. A
hot boat could oxidize the sample in an early stage, producing erratic numbers.
(typical cooling time is 4 minutes)
6. Now the 3000 SYSTEM is ready for the next analysis. Select the sample to be
analyzed and repeat the mentioned steps starting from 5.
1. Open a new queue list and key-in the necessary parameters (for more details,
refer to the software manual)
2. Open “analysis” in the “controller window” and select the box “methods”. Choose
the appropriate EOX/TX analysis method according the application. (for more
details see chapter10.1.6.4)
3. Open “evaluation” and select the “absolute” calculation method (the calibration
field at this point is not used)
4. Take a 100µL syringe and fill it with the prepared EOX or TX (light hydrocarbon)
sample.
5. Select or add in the queue list the sample to be analyzed and press “analyze”.
6. Insert the needle through the septum and let the syringe rest on the support. The
plunger pusher in the “boat driver” performs the injection itself. Injection speed
and volume are software controlled and application dependable.
7. The analysis terminates when sample is completely oxidized and the baseline
reaches its starting point again. Manual termination is possible by keying in a
lower number of minutes/seconds analysis time or putting the cursor at the
finishing baseline and click on the right mouse key. This will automatically key in
the selected time. Now the 3000 SYSTEM is ready for the next analysis. Select
the sample to be analyzed and repeat the mentioned steps starting from 4.
7. the unit waits until the curve has returned to baseline and terminates the analysis
as soon as the maximum time has elapsed or the appropriate baseline level is
reached. (depending on software settings)
8. the calculated result appears on the screen and the sampler starts the next
sample injection/introduction.
The 3000 system has been designed to achieve a high level of reliability. In case
of problems or failure, these troubleshooting guidelines will be helpful.
Generally too high baseline Cell problems Check whether the baseline
drops to an acceptable level
when the cell is disconnected
from the scrubber
CONSUMABLES TS 3000
PGENE0052 Glass fiber filters, 10 pcs
PANAL0025 Septum, diam. 12mm, 10 pieces
PINJE0012 Quartz sample boat, solids
PANAL0211* Set of Standards; 0, 1, 5, 10 mg S/L di-n-butyl sulfide in
toluene dibenzothiophene, 1 mg S/L
PANAL0125 Combustion tube universal model 3000
PANAL0174 Ceramic needle, 150 mm
CONSUMABLES TN 3000
PANAL0161 Paper filters 5µm, 10 pcs
PGENE0029 Catalyst Pt on ceramic base, 100 gr
PANAL0025 Septum, diam. 12mm, 10 pieces
PANAL0167* Ammoniumsulphate, 50 gr
PANAL0168* Potassiumnitrate, 50 gr
CONSUMABLES EOX
PANAL0025 Septum, diam. 12mm, 10 pieces
PANAL0095* Standard Aldrin in water, 0 and 5 mg CI/L 10 x 2 mL
PANAL0097* Blank to standard aldrin, 5 x 2 mL
PGENE0002 Syringe 100 µL, 71 mm needle, p.s. 5
PGENE0004 Injection needle Hamilton, 71mm, p.s. 5, 3 pcs
CONSUMABLES POX
PINJE0003 POX samplebottle 100 mL complete
PINJE0002 Septum, 24mm, 10 pieces
PANAL0102* Standard DiChloroMethane 10 mg Cl/L; 10 mL
Method Coolforce CoolNOrm Furn. 1 Furn. 2 Inlet Outlet Argon/Hel Oxygen Ext. Oxygen Ozonator Bypass Needle Fast. Fast Injection Injection Measure
. NOrm/std NOrm/std NOrm/std NOrm/std turbo Position Speed Position Speed
sort % % celsius celsius celsius celsius mm mm mm mm mm mm mm mm/sec mm mm/sec sec.
Liq-Petro 10 10 850/700 1000/700 --------- 60/10 65 40 20 40/50 10/15 120 65.0 15 110 0.6 190
Liq-Water 10 10 900/700 900/700 --------- 60/10 50/65 40 20 40/50 05 120 55.0 15 110 1.2 240
Method Coolforce CoolNOrm Furn. 1 Furn. 2 Inlet Outlet Argon/Hel Oxygen Ext. Oxygen Bypass Needle inj. Speed
. NOrm/std NOrm/std NOrm/std NOrm/std turbo NOrm/turbo
sort % % celsius celsius celsius celsius mm mm mm mm mm mm/sec
Tips: 1) The temperature for the solids method can be higher, but that depends of the sort of sample you want to measure.
2) When you use a 75 mm needle (100 μl) the next value can be used for the injection speed
0.1 mm/sec = 0.167 μl/sec
TYPES OF
ACUTE HAZARDS/ FIRST AID/
HAZARD/ PREVENTION
SYMPTOMS FIRE FIGHTING
EXPOSURE
Highly flammable. no open flames, no sparks, Powder, AFFF, foam, carbon
FIRE and no smoking. dioxide.
Vapour/air mixtures are Closed system, ventilation, In case of fire: keep drums,
explosive. explosion-proof electrical etc., cool by spraying with
equipment and lighting. water.
Prevent build-up of
EXPLOSION electrostatic charges (e.g., by
grounding). Do NOT use
compressed air for filling,
discharging, or handling.
STRICT HYGIENE! AVOID
EXPOSURE EXPOSURE OF
(PREGNANT) WOMEN!
Dizziness. Drowsiness. Ventilation, local exhaust, or Fresh air, rest. Artificial
INHALATION Headache. Nausea. breathing protection. respiration if indicated. Refer
Unconsciousness. for medical attention.
Dry skin. Redness. Protective gloves. Remove contaminated
clothes. Rinse and then wash
SKIN
skin with water and soap.
Refer for medical attention.
Redness. Pain. Safety goggles or face First rinse with plenty of
shield. water for several minutes
EYES (remove contact lenses if
easily possible), then take to
a doctor.
Abdominal pain. Burning Do not eat, drink, or smoke Rinse mouth. Give a slurry of
sensation (further see during work. activated charcoal in water to
INGESTION Inhalation). drink. Do NOT induce
vomiting. Refer for medical
attention.
SPILLAGE DISPOSAL STORAGE PACKAGING & LABELLING
Collect leaking liquid in sealable Fireproof. Separated from strong
containers. Absorb remaining oxidants. F symbol
liquid in sand or inert absorbent Xn symbol
and remove to safe place. Do R: 11-20
NOT wash away into sewer S: (2-)16-25-29-33
(extra personal protection: self- UN Hazard Class: 3
contained breathing apparatus). UN Packing Group: II
SEE IMPORTANT INFORMATION ON BACK
Prepared in the context of co-operation between the International Program on Chemical Safety &
ICSC: 0078 the Commission of the European Communities © IPCS CEC 1993
EFFECTS OF LONG-TERM OR
REPEATED EXPOSURE:
D
Repeated or prolonged contact with
skin may cause dermatitis. The
A
substance may have effects on the
central nervous system , resulting in
T
decreased learning ability and
psychological disorders. Animal tests
A show that this substance possibly
causes toxic effects upon human
reproduction.
Boiling point: 111°C Relative density of the vapour/air-
Melting point: -95°C mixture at 20°C (air = 1): 1.06
Relative density (water = 1): 0.87 Flash point: 4°C c.c.°C
PHYSICAL
Solubility in water: SO2ne Auto-ignition temperature: 480°C
PROPERTIES
Vapour pressure, kPa at 20°C: 2.9 Explosive limits, vol% in air: 1.1-7.1
Relative vapour density (air = 1): 3.2 OctaSO2l/water partition coefficient
as log Pow: 2.69
ENVIRONMENTAL
DATA
NOTES
Depending on the degree of exposure, periodic medical examination is indicated.
Transport Emergency Card: TEC (R)-31
NFPA Code: H 2; F 3; R 0;
Neither the CEC or the IPCS nor any person acting on behalf of the CEC or the IPCS is
IMPORTAN responsible for the use which might be made of this information. This card contains the
T LEGAL collective views of the IPCS Peer Review Committee and may not reflect in all cases all the
NOTICE: detailed requirements included in national legislation on the subject. The user should verify
compliance of the cards with the relevant legislation in the country of use.
ARGON
(liquefied, cooled)
Ar
Atomic mass: 39.95
CAS # 7440-37-1
RTECS # CF2300000
ICSC # 0154
UN # 1951
TYPES OF
ACUTE HAZARDS/ FIRST AID/
HAZARD/ PREVENTION
SYMPTOMS FIRE FIGHTING
EXPOSURE
not combustible. Heating will For cylinders, see notes. In
cause rise in pressure with case of fire in the
FIRE
risk of bursting. surroundings: all
extinguishing agents allowed.
EXPLOSION
EXPOSURE
Asphyxia. Dizziness. Ventilation. Fresh air, rest. Artificial
INHALATION Unconsciousness. respiration if indicated. Refer
for medical attention.
ON CONTACT WITH Cold-insulating gloves. ON FROSTBITE: rinse with
LIQUID: FROSTBITE. Protective clothing. plenty of water, do NOT
SKIN
remove clothes. Refer for
medical attention.
Pain. Blurred vision. Severe Safety goggles, or face First rinse with plenty of
deep burns. shield. water for several minutes
EYES (remove contact lenses if
easily possible), then take to
a doctor.
INGESTION
SPILLAGE DISPOSAL STORAGE PACKAGING & LABELLING
Ventilation. NEVER direct water jet Store outside or in a separate well-
on liquid (extra personal ventilated building. Cool. UN Hazard Class: 2.2
protection: self-contained
breathing apparatus).
TYPES OF
ACUTE HAZARDS/ FIRST AID/
HAZARD/ PREVENTION
SYMPTOMS FIRE FIGHTING
EXPOSURE
not combustible but no open flames, no sparks, In case of fire in the
enhances combustion of and no smoking. No contact surroundings: all
other substances. Many with flammable substances. extinguishing agents allowed.
reactions may cause fire or no contact with fuels and
FIRE
explosion. Heating will cause other combustible materials.
rise in pressure with risk of
bursting.
EXPOSURE
INHALATION
SKIN
Redness. Safety goggles. First rinse with plenty of
water for several minutes
(remove contact lenses if
EYES
easily possible), then take to
a doctor.
INGESTION
SPILLAGE DISPOSAL STORAGE PACKAGING & LABELLING
Evacuate danger area! Consult an Fireproof. Separated from
expert! Ventilation. combustible and reducing O symbol
substances. Cool. R: 8-34
S: 21
UN Hazard Class: 2.2
SEE IMPORTANT INFORMATION ON BACK
Prepared in the context of co-operation between the International
ICSC: 0138 Programme on Chemical Safety & the Commission of the European
Communities © IPCS CEC 1993
ENVIRONMENTAL
DATA
NOTES
Do NOT use in the vicinity of
a fire or a hot surface, or
during welding. Also consult
ICSC # 0880.
Transport Emergency Card:
TEC (R)-842
Neither the CEC or the IPCS SO2r any person acting on behalf of the CEC or the IPCS is
IMPORTAN responsible for the use which might be made of this information. This card contains the
T LEGAL collective views of the IPCS Peer Review Committee and may not reflect in all cases all the
NOTICE: detailed requirements included in national legislation on the subject. The user should verify
compliance of the cards with the relevant legislation in the country of use.
OCCUPATIONAL EXPOSURE
LIMITS (OELs):
D TLV (as (ceiling values)): 0.11 ppm
as hydrazoic acid vapour ppm; 0.29
A mg/m3 as sodium azide (ACGIH
1996).
T
Neither the CEC or the IPCS nor any person acting on behalf of the CEC or the IPCS is
IMPORTANT responsible for the use which might be made of this information. This card contains the
LEGAL collective views of the IPCS Peer Review Committee and may not reflect in all cases all the
NOTICE: detailed requirements included in national legislation on the subject. The user should verify
compliance of the cards with the relevant legislation in the country of use.
Neither the CEC or the IPCS nor any person acting on behalf of the CEC or the IPCS is
IMPORTANT responsible for the use which might be made of this information. This card contains the
LEGAL collective views of the IPCS Peer Review Committee and may not reflect in all cases all
NOTICE: the detailed requirements included in national legislation on the subject. The user should
verify compliance of the cards with the relevant legislation in the country of use.
PREVENT DISPERSION OF
DUST! STRICT HYGIENE!
EXPOSURE
AVOID EXPOSURE OF
(PREGNANT) WOMEN!
(see Ingestion). Ventilation (not if powder). Fresh air, rest. Artificial
INHALATION respiration if indicated. Refer
for medical attention.
MAY BE ABSORBED! See Protective gloves. Protective Remove contaminated
SKIN Ingestion. clothing. clothes. Rinse and then wash
skin with water and soap.
Safety goggles or face First rinse with plenty of
shield. water for several minutes
EYES (remove contact lenses if
easily possible), then take to
a doctor.
Dizziness. Headache. Do not eat, drink, or smoke Do NOT induce vomiting.
INGESTION Nausea. Vomiting. during work. Rest. Refer for medical
Weakness. Muscle twitching. attention.
SPILLAGE DISPOSAL STORAGE PACKAGING & LABELLING
Sweep spilled substance into Separated from food and Do not transport with food and
containers. Carefully collect feedstuffs. Cool. Dry. feedstuffs.
remainder, then remove to safe T symbol
place. Do not let this chemical R: 24/25-40-48
enter water courses or sewers S: 22-36/37-44
(extra personal protection: UN Hazard Class: 6.1
complete protective clothing Severe marine pollutant.
including self-contained breathing
apparatus).
Prepared in the context of cooperation between the International
Programme on Chemical Safety & the Commission of the European
ICSC: 0774 Communities © IPCS CEC 1993
NOTES
Other melting points: 40-60°C (technical grade). Technical aldrin is a tan to dark brown waxy solid.
Carrier solvents used in commercial formulations may change physical and toxicological properties. Do
NOT take working clothes home. The recommendations on this Card also apply to ICSC # 0787
(dieldrin). Aldrine, Aldrex, Aldrite, Aldrosol, Drinox, Seedrin, Octalene are trade names.
Transport Emergency Card:
TEC (R)-61G53b
NFPA Code: H2; F0; R0;
Neither the CEC or the IPCS nor any person acting on behalf of the CEC or the IPCS is
IMPORTANT responsible for the use which might be made of this information. This card contains the
LEGAL collective views of the IPCS Peer Review Committee and may not reflect in all cases all
NOTICE: the detailed requirements included in national legislation on the subject. The user should
verify compliance of the cards with the relevant legislation in the country of use.
Operating procedure
Starting procedure:
4. Connect Argon (white) and Oxygen (blue) connectors of the introduction module.
5. Open Gas supplies with a pre-pressure of approximately 3 bar for Argon and
Oxygen.
Standby procedure:
2. When using an ELS 3000 take furnace needle out of the introduction module
For example:
10 mm Oxygen is equal to a (mass) flow of 37.4 mL/min
200 mL/min Argon (massflow) is equal to 40-41 mm on the Rotameter scale
45 mm Helium is equal to 301 mL/min
Storage temperature 0 – 50 ºC
Operating temperature 10 – 30 ºC
Humidity 20 – 85% (non–condensing)
Safety class Class I**
Furnace 42V; 800W
Please visit us at
www.thermo.com