X8000 Light Source
X8000 Light Source
X8000 Light Source
220-200-000
X8000 Light Source
User Guide
Contents
Warnings and Cautions ......................................... 2
Symbol Definitions ............................................................................ 4
Product Description and Intended Use ................. 5
Setup and Assembly ............................................. 8
Connecting the AC Power Cable ...................................................... 8
Connecting the Light Cable .............................................................. 8
System Operation ............................................... 10
Powering the System On and Off ................................................... 10
Selecting the Operation Mode ........................................................ 11
Adjusting the Brightness ................................................................. 11
Visual Display .................................................................................. 12
Language Selection ........................................................................ 13
Safety Shutoff ................................................................................. 13
Checking the ESST Feature ............................................................ 14
Using the X8000 with a Voice-Controlled System Interface ........... 15
Using the SFB Serial Interface ........................................................ 15
Troubleshooting .............................................................................. 16
Cleaning and Maintenance ................................. 18
Cleaning the X8000 ......................................................................... 18
Caring for the Bulb Module ............................................................. 18
Replacing the Bulb Module ............................................................ 18
Replacing the Fuses ....................................................................... 20
Disposing of the X8000 ................................................................... 20
Technical Specifications ..................................... 21
Electromagnetic Compatibility ........................................................ 22
Warranty ............................................................. 26
Service and Claims ............................................. 27
1
Warnings and Cautions
Please read this manual and follow its instructions carefully. Certain symbols and
the words warning, caution, and note carry special meanings and should be
carefully reviewed:
2
DOING SO MAY RESULT IN FIRE. ALWAYS PLACE THE LIGHT SOURCE
IN STANDBY MODE WHENEVER THE SCOPE IS REMOVED FROM THE
LIGHT CABLE OR THE DEVICE IS UNATTENDED. THE SCOPE TIP,
SCOPE LIGHT POST, LIGHT CABLE ADAPTER, AND LIGHT CABLE TIP
WILL TAKE SEVERAL MINUTES TO COOL OFF AFTER BEING PLACED
IN STANDBY MODE, AND THEREFORE MAY STILL RESULT IN FIRE OR
BURNS TO THE PATIENT, USER, OR INANIMATE OBJECTS.
3
Symbol Definitions
Equipotentiality
Alternating current
4
Temperature limitation
Pressure limitation
Federal law (USA) restricts this device to use by, or on order of, a
physician
Date of manufacture
Legal manufacturer
Fuse rating
5
WARNING THE SCOPE TIP, SCOPE LIGHT POST, LIGHT CABLE ADAPTER, AND
LIGHT CABLE TIP WILL TAKE SEVERAL MINUTES TO COOL OFF
AFTER BEING PLACED IN STANDBY MODE AND THEREFORE MAY
STILL RESULT IN FIRE AND/OR BURNS TO THE PATIENT, USER, OR
AN INANIMATE OBJECT IF NOT USED PROPERLY. DO NOT PLACE
THE SCOPE OR THE LIGHT CABLE ON THE PATIENT OR ON THE
DRAPES OR OTHER FLAMMABLE MATERIAL, EVEN WHEN THE
DEVICE IS IN STANDBY MODE.
The Stryker X8000 Light Source consists of one of each of the following:
• light source console
• power cord
• xenon bulb module (spare or replacement bulb modules can be
purchased separately as P/N 220-201-000)
The features of the X8000 console are described below (see Figures 1 and 2).
5 6 7
3
2
1 9 8
Figure 1 Front panel of the X8000 console
6
6. Up Button: Increases light intensity in Run mode.
7. Cable Clamp: Grasps the light-source end of an inserted fiberoptic
cable. (When no cable is inserted, the bulb automatically turns off to
save bulb hours.)
8. Jaw Handle: Opens the fiberoptic-cable holder.
9. Down Button: Decreases light intensity in Run mode.
10 11
Stryker Endoscopy
5900 Optical Court
San Jose, CA 95138 USA
SIDNE
U.S. Patents:
5850496, 6110107, 6689050
Other patents pending
MADE IN USA
14 12, 13
Figure 2 Rear panel of the X8000 console
7
Setup and Assembly
Note Your local Stryker Endoscopy sales representative will perform
inservice training at your convenience to instruct you and your
staff on the safe and effective operation and maintenance of the
X8000, and to help set up the equipment. Please contact your
local Stryker Endoscopy sales representative to schedule an
inservice after the equipment has arrived.
Note When selecting a setup location for the X8000, consult the
“Electromagnetic Compatibility” section included in this manual
to determine the best location.
To set up the X8000, make the following connections:
• Connect the AC power cable
• Connect the light cable
Note The X8000 Light Source is compatible with all Stryker Light
Cables.
8
1. Lock open the cable clamp by turning the jaw handle clockwise until it
stops (see Figure 3).
WARNING KEEP FINGERS AWAY FROM THE CABLE CLAMP AS THE CLAMP
MAY INADVERTENTLY DEPLOY AND CAUSE INJURY.
WARNING DO NOT LOOK DIRECTLY INTO THE CABLE PORT. THE HIGH-
INTENSITY LIGHT MAY CAUSE DAMAGE TO THE EYES.
2. Insert a clean, dry fiberoptic cable into the cable port until the jaw latch
releases and the jaw clamps the cable in place (see Figure 4). Pull gently
on the fiberoptic cable to test that it is securely seated in the cable port.
9
4. To remove the light cable, press the Mode button to put the unit into Standby
mode. Then turn the jaw handle clockwise until it latches fully open.
Note The Light Source will default to Standby mode when a light cable
is inserted.
System Operation
Note Before operating the X8000, see the “Setup and Assembly” section
of this manual.
10
Selecting the Operation Mode
The X8000 has two operation modes, Run and Standby.
• RUN mode: The Run mode is used during normal operation. It
enables light output to be controlled by the brightness controls on the
front console panel.
• STANDBY mode: The Standby mode is used when the X8000 is
powered on, but not in use. It reduces the light output to a minimum,
thereby reducing the heat generated at the tip of the light cable or
scope when the X8000 is not being used.
To select either the Run or Standby mode, press the Mode Button. The LED for
the selected mode will illuminate.
WARNING THE HIGHER THE BRIGHTNESS, THE MORE HEAT ENERGY THAT
WILL BE GENERATED IN THE SCOPE AND THE TIP OF THE CABLE.
ALWAYS ADJUST THE BRIGHTNESS LEVEL OF THE CAMERA AND
THE MONITOR BEFORE ADJUSTING THE BRIGHTNESS LEVEL OF THE
LIGHT SOURCE. ADJUST THE BRIGHTNESS LEVEL OF THE LIGHT
SOURCE TO THE MINIMUM BRIGHTNESS NECESSARY TO
ILLUMINATE THE SURGICAL SITE. FAILURE TO FOLLOW THIS
11
INSTRUCTION COULD RESULT IN FIRE OR BURNS TO THE PATIENT,
USER OR AN INANIMATE OBJECT. IN ADDITION, ADJUST THE
INTERNAL SHUTTER OF THE CAMERA HIGHER IN ORDER TO RUN
THE LIGHT SOURCE AT A LOWER INTENSITY.
Visual Display
The X8000 provides feedback through the LCD.
• Brightness: The LCD display shows the intensity level of the light as a
percentage between 0 and 100. For example, if the LCD shows “70,”
the light output to the fiberoptic cable is running at 70 percent of
capacity.
• Bulb Hours: The LCD display shows how many total hours the bulb
has operated. For example, if the LCD shows “250,” the bulb has
operated 250 total hours. The bulb hours will flash after 500 hours,
and the LCD will display “Replace Bulb” after 1000 hours.
• The LCD also displays warning and error codes. The table below lists
and defines the warning and error codes displayed.
Code Definition Recommended
Action
E-1 All conditions are met for the bulb to Return the X8000 for
illuminate, yet it remains off. repair.
E-2 All conditions are not met for the bulb Return the X8000 for
to illuminate, yet it remains on. repair.
E-3 The bulb has higher than expected Return the X8000 for
current or voltage applied to it. repair.
E-4 The bulb is kept off because the ballast Return the X8000 for
fan is not working properly. repair.
E-5 The bulb is kept off because the bulb fan Return the X8000 for
is not working properly. repair.
12
Code Definition Recommended
Action
Replace The bulb has reached 1000 hours and Install a new Stryker
Bulb has exceeded its recommended lifetime bulb.
of 500 hours.
Language Selection
The X8000 LCD has the capability of displaying text in the following languages:
Danish German Polish
Dutch Greek Portuguese
English Italian Simplified Chinese
Finnish Japanese Spanish
French Korean Swedish
Safety Shutoff
The X8000 Light Source is equipped with a Safety Shutoff feature which will
temporarily turn off the bulb in the event of excessive heat in the bulb assembly.
WARNING ONCE THE LIGHT SOURCE COOLS DOWN (AFTER 7-10 MINUTES),
POWER WILL RESUME TO THE BULB AND THE UNIT WILL RESTART
IN STANDBY MODE. TO PREVENT FIRES AND ACCIDENTAL BURNS
TO THE PATIENT, USER OR INANIMATE OBJECTS, ALWAYS PLACE
THE SCOPES AND/OR FIBEROPTIC CABLES IN A SAFE PLACE, AND
NOT ON THE PATIENT, DRAPES, OR OTHER FLAMMABLE
MATERIAL, TO ENSURE SAFE RESUMPTION OF LIGHT OUTPUT. IF
THE X8000 LIGHT SOURCE EXPERIENCES A TEMPORARY
SHUTDOWN, IT IS RECOMMENDED THAT THE DEVICE BE
RETURNED FOR SERVICE.
Caution Do not abruptly interrupt power to the unit. This will turn off
the fan and may cause severe damage to the internal cooling
system.
13
Checking the ESST Feature
The X8000 is equipped with Electronic Scope Sensing Technology (ESST), a
special safety feature that helps prevent accidental fires or burns to the patient or
user caused by a light cable not connected to a scope. This feature will only work
if the X8000 is used with an ESST light cable. When operated with an ESST light
cable, the X8000 senses when the scope and light cable are separated, and places
the light source in Standby mode. In Standby mode, the X8000 will reduce light
output to a minimum, preventing the light cable from generating excessive heat.
To verify the ESST feature is active, perform the following test before every
surgical procedure:
1. Set up the X8000 system with an ESST light cable and scope, and then
power on the system.
2. Place the X8000 in Run mode.
3. Remove the light cable from the ESST scope adapter.
The X8000 should return to Standby mode, indicating that the ESST feature is
functioning properly.
WARNING ALWAYS CHECK TO ENSURE THAT THE UNIT HAS SWITCHED INTO
STANDBY MODE, BEFORE ASSUMING ESST SAFETY PROTECTION. IF
THE UNIT FAILS TO RETURN TO STANDBY MODE, THERE MAY BE A
FAULT WITH THE ESST FEATURE. IN THIS CASE, DO NOT ASSUME
ESST SAFETY PROTECTION, AND RETURN THE UNIT FOR SERVICE.
14
Using the X8000 with a Voice-Controlled System
Interface
The X8000 can be used in conjunction with Stryker voice-control systems
(SIDNE®). For more information about using the X8000 with Stryker voice-
control systems, refer to the SIDNE® Operating and Maintenance Manual
(P/N 1000-400-653).
15
Troubleshooting
16
Too much or too • Ensure the light cable is correctly engaged with the
little light output cable port.
• Ensure the bulb has adequate bulb life remaining.
The bulb has a warranty of 500 hours.
• Ensure the X8000 is in Run mode. (The Run LED
should be illuminated.) If necessary, press the Mode
button to switch from Standby to Run. If the unit
remains in Standby:
1. Ensure the light cable is correctly engaged with
the cable port.
2. If an ESST cable is connected to the X8000,
ensure the cable is attached to the scope using
an ESST scope adapter.
• Use the up/down buttons to adjust the brightness. For
details, see the “Adjusting the Brightness” section in
this manual.
• Ensure the fiberoptic cable is transmitting light
properly. Hold the light-source end of the cable up to
an overhead room light and look into the scope end
of the light cable. If the pattern contains any black
spots, the light cable may be worn out and may
require replacement.
• Ensure the light cable is of an adequate size for the
application. The cable diameter may be too small to
provide adequate light transmission for the medical
video camera in the endoscopic application.
17
Cleaning and Maintenance
Cleaning the X8000
WARNING UNPLUG THE X8000 BEFORE CLEANING THE UNIT.
WARNING DURING OPERATION, THE BULB AND THE HOUSING AROUND THE
BULB MAY BE HOT. WAIT AT LEAST THREE MINUTES FOR THE
BULB TO COOL BEFORE HANDLING IT.
18
To replace the bulb module:
1. Power down the X8000.
2. Open the bulb door (Figure 5).
Note If the unit is on, opening the bulb door will turn off the bulb.
3. Rotate the handle to the right, grasp the bulb module by the handle, and
remove the bulb (Figure 6).
WARNING DO NOT REACH INSIDE THE BULB DOOR FOR ANY REASON OTHER
THAN REPLACING THE BULB MODULE. TOUCHING PARTS OTHER
THAN THE BULB MODULE MAY CAUSE BURNS OR PRODUCT
DAMAGE.
WARNING DO NOT TOUCH ANY PART OF THE BULB MODULE EXCEPT FOR
THE HANDLE. THE MODULE MAY BE VERY HOT AND CAUSE
BURNS.
19
4. Insert the new bulb module (Stryker part number 220-201-000) along
the guide rails until it is fully seated on the mating connectors.
5. Rotate the handle to the left to secure the bulb module within the light
source.
6. Close and latch the bulb door.
7. Power on the unit.
8. Power on the bulb (if not already on) and verify that the LCD displays
the bulb hours as “0”.
Note Using a non-Stryker replacement bulb module will result in no
light output.
Caution Do not operate the X8000 with a burned-out bulb or with no
bulb installed.
WARNING TO HELP AVOID THE RISK OF FIRE, USE ONLY FUSES RATED AS
SPECIFIED ON THE CONSOLE.
20
Technical Specifications
Electrical
Primary: 100 - 120 VAC, 50/60 Hz, 450 W
220 - 240VAC, 50/60 Hz, 450W
Fuses (2): T 5.0 AH 250 V
Dimensions
Height: 4.75"(12.1 cm)
Width: 12.5" (31.8 cm)
Depth: 16.8" (42.7 cm)
Weight: 16.0 lbs. (7.3 kg)
Fiberoptic Cable Range:
2 mm to 6.5 mm diameter
Bulb
Type: 300 Watt Xenon (Elliptical)
Life: Approximately 500 hours
Operating Conditions
10 to 40°C
30 to 75% Relative Humidity
700hPa to 1060hPa
Transportation & Storage
-20 to 50°C
10% to 75% Relative Humidity
700hPa to 1060hPa
Classifications
Class 1 Equipment
Type CF applied parts
Water Ingress Protection, IPX0 — Ordinary Equipment
Continuous Operation
21
Electromagnetic Compatibility
Like other electrical medical equipment, the X8000 requires special precautions
to ensure electromagnetic compatibility with other electrical medical devices. To
ensure electromagnetic compatibility (EMC), the X8000 must be installed and
operated according to the EMC information provided in this manual.
Note The X8000 has been designed and tested to comply with IEC
60601-1-2 requirements for EMC with other devices.
Caution Portable equipment may affect the normal function of the
X8000.
X8000 is intended for use in the electromagnetic environment specified below. The customer or the
user of X8000 should ensure that it is used in such an environment.
RF emissions CISPR
Class B
11
X8000 is suitable for use in all establishments,
Harmonic emissions including domestic establishments and those
Class A
IEC61000-3-2 directly connected to the public low-voltage
power supply network that supplies buildings
Voltage Fluctuations/ used for domestic purposes.
flicker emissions Complies
IEC61000-3-3
22
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
X8000 is intended for use in the electromagnetic environment specified below. The customer or
the user of X8000 should ensure that it is used in such an environment.
Electromagnetic
IEC 60601 Test
Immunity Test Compliance Level Environment:
Level
Guidance
Floors should be
wood, concrete, or
Electrostatic Discharge ceramic tile. If floors
±6kV contact ±2,4,6kV contact are covered with
(ESD)
±8kV air ±2,4,8kV air synthetic material,
IEC61000-4-2 the relative humidity
should be at least
30%.
Power-frequency
magnetic fields
Power frequency (50/60Hz) should be at levels
magnetic field characteristic of a
3 A/m N/A
typical location in a
IEC 61000-4-8 typical commercial
or hospital
environment.
23
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
X8000 is intended for use in the electromagnetic environment specified below. The customer or
the user of X8000 should ensure that it is used in such an environment.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast,
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the X8000 system is used exceeds the applicable RF compliance level
above, the X8000 system should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the X8000 unit.
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
24
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the X8000 System
The X8000 system is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the X8000 system can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the X8000 system as recommended below, according to the maximum
output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
25
Warranty
Stryker Endoscopy warrants the X8000 Light Source against defects in both
materials and workmanship to the registered owner at the time of purchase.
All components except the bulb are covered by the warranty for a period of one
year from the date of purchase. The bulb module is covered by the warranty for
a period of 60 days from the date of purchase.
This warranty does not apply to any unit that has been subject to misuse, abuse,
neglect, improper installation or operation, or that has been altered, adjusted, or
tampered with by any person other than Stryker Endoscopy authorized service
personnel.
The customer is responsible for returning the defective equipment to the factory
at his or her own expense. Stryker Endoscopy or its representative will service the
unit, repair or replace any defective parts thereof, and return the unit to the
customer.
If, upon examination, it is determined that the fault has been caused by misuse or
abnormal conditions of operation, the repairs will be billed to the customer in the
same manner as out-of-warranty repairs.
Instruments repaired by Stryker Endoscopy will be issued a thirty-day warranty
against defects in both materials and workmanship, provided the original
warranty period has passed. Instruments submitted due to defects in materials
and workmanship during the warranty period will be repaired at no charge to the
customer.
The warranty as set forth herein is exclusive and in lieu of all other warranties,
remedies, obligations, and liabilities of Stryker Endoscopy Inc., expressed or
implied, including the implied warranties of merchantability and fitness for use
and of consequential damages. These products are sold only for the purpose
described herein, and such warranty runs only to the purchaser. In no event shall
Stryker Endoscopy be liable for any breach of warranty in any amount exceeding
the purchase price of the product.
No agent, employee, or representative of Stryker Endoscopy has the authority to
bind the Company to any other warranty, affirmation, or representation
concerning this instrument.
This warranty is valid only to the original purchaser of Stryker Endoscopy
products obtained directly from Stryker Endoscopy or from a Stryker Endoscopy
authorized agent. The warranty cannot be transferred or assigned by the original
purchaser.
The X8000 Light Source warranty is void if any WARNINGS, CAUTIONS, or
NOTES are disregarded.
26
Service and Claims
This equipment is carefully packaged to prevent damage during transit and is
shipped freight outbound (F.O.B.) from San Jose, CA. Therefore, Stryker’s
responsibility for ensuring damage-free delivery totalled ends upon remission of
the equipment to the carrier. Examine the shipment promptly upon receipt.
If the product packaging appears damaged at the time of delivery, immediately
check the device for damage. If there is a malfunction, return the device to the
distributor.
To return equipment damaged during shipment:
1. File a claim with both the carrier and the distributor.
2. Repack the equipment in its original shipping container and ship it
prepaid and insured to:
Stryker Endoscopy
Customer Service Department
5900 Optical Court
San Jose, CA 95138
USA
If service is needed either during or after the warranty period:
1. Contact Stryker Endoscopy at 1-800-624-4422 or phone your local
Stryker Endoscopy sales representative.
2. Package all the components carefully in the original shipping container
if possible.
3. Ship the light source, prepaid and insured to:
27
28
Stryker Endoscopy
5900 Optical Court
San Jose, CA 95138 USA
1-408-754-2000, 1-800-624-4422
www.stryker.com
Stryker Corporation or its divisions or other corporate affiliated entities own, use or have applied for the following
trademarks or service marks: the Stryker logo. All other trademarks are trademarks of their respective owners.
2014/01
1000-400-885 F