DELTA4000, FW Upgrade Procedure: Megger Sweden AB
DELTA4000, FW Upgrade Procedure: Megger Sweden AB
DELTA4000, FW Upgrade Procedure: Megger Sweden AB
2010-09-01
• Most of the bin-files include FW for the HVU as well as the Control Unit. This means that
you need to connect the HVU unit to the Control Unit before you continue with the FW
Upgrade. If you are unsure what type of bin-file you have this is probable how you should
connect.
• If you know that the bin file is intended for the DELTA calibration box. Please connect the
Calibration box to your DELTA unit before you continue with the FW Upgrade.
2010-09-01
• Wait for MIX to detect the DELTA unit (this may take a while depending on how many
COM/RS232 ports you have on your computer)
2010-09-01
• When the DELTA unit has been detected. Click on the Maintenance tab
• Follow the instructions carefully! When asked for it, locate the .bin file that you saved in
step 1 above.
NOTE: Restart the DELTA unit ONLY when prompted by the software!
• If you encounter any problems during the upgrade make sure you have made the correct
unit connections (described in the section above) or contact a Megger representative.
NOTE:
Restarting the DELTA unit at any time during the upgrade process when not asked to restart
may result in a corrupted system and a unit that cannot be used.
If this happens, you will have to send it back to the factory for repair.