Urgent Field Safety Notice PHILIPS MX 16-Slice Computed Tomography
Urgent Field Safety Notice PHILIPS MX 16-Slice Computed Tomography
Philips Healthcare
(Suzhou) Co. Ltd. Field Safety Notice
As part of Philips’ continuous focus on reliability and safety we continuously monitor the performance of our
products. During recent evaluations of the Philips MX 16-slice Computed tomography x-ray system, we have
identified a potential issue that may affect the performance of the equipment under certain conditions. This
letter is intended to provide you with information regarding:
• what the issue is, and under what circumstances it may occur
• the actions you can take to avoid or minimize the occurrence of the issue
• the actions planned by Philips to correct the issue
This document contains important information for the continued safe and
proper use of your equipment
Please review the following information with all members of your staff who need to be aware of the
contents of this communication. It is important to understand the implications of this communication.
If you need any further information or support concerning this issue, please contact your local Philips
representative: Technical Support Line: 1-800-722-9377
This notice will be reported to the appropriate Regulatory Agency pursuant to applicable regulations
We apologize for any inconvenience this may cause and trust that this information is adequately addressing
any concerns you may have.
Sincerely,
Dominic Siewko
Sr. Mgr Radiation Health & Product Safety
Customer Support
Philips Healthcare
(Suzhou) Co. Ltd. Field Safety Notice
HOW TO IDENTIFY Identify affected units by the software revision: vi .1.3.1167; vi .1.4.21425;
AFFECTED PRODUCTS vi.1.4.2i424; vl.1.4.21422
How the user can identify:
For Vi .1.3 system, user can always go to “Help” page from the screen top, and
seethe software revision there. It should be 1.1.3.1167
Mx Workspace !
1.1.3.116]
10-Sep-20 10
CE
Ptuhps Mea Sy ec Nd land BV
‘Cbet.
information.
Mb’j1jInfomiatmn —I
Manufacturer
IPNMS
ACTIONS PLANNED BY Philips plans to install updated software in a Service Pack, the product correction
PHILIPS will be issued and will be implemented free of charge. The Service Pack will
also correct the following other issues:
. When a surview which is longer than 557mm is stopped by using
“Abort” function during the scan process, the position of scan box in
surview image is not correct.
. For dental applications, when a patient scan is taken with Gantry tilted,
and the tilt angle falls in the range from 1 to -1 degree (not including 0
degree position), the data post-processing may not operate normally and
the images may not be displayed.
. If the user stops a surview using the “Abort” function during the second
part of the scan, the positions of the scan box are not consistent.
A Philips Field Engineer will contact you shortly to schedule this correction.
FU RTH ER
IN FORMATION AND If you need any further information or support concerning this issue, please
SUPPORT contact your local Philips representative.