R5ENG
ALIAVTX
®
Vortex Flowmeter
AVF7000 Series
Operator Manual
Index
1. Flowmeter Check - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
2. Specification - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
3.Diameter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4
4. Installation- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5
5. Wiring Chart - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7
6. Panel Display - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9
7. Function - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9
2
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
1. Flowmeter Check
A. First check the package if it is good or damage, if broken hard you should notify the meter transport department or
contact the customer service center of Alia immediately
B. Open the package and check for meter and all parts attachment if it is good or damage and shortage
C. Read the operating instruction in detail and comprehends all contents, if any part of that you do not understand full,
fax technical service department of Alia
D. Make sure that the specification of the meter you received is conformed to the operating condition
E. Power on the meter in house, observe the LCD display if it is on normal
F. Select proper meter install site, make sure to meet the installing condition
G. Meter be moved to the field and be mounted on the pipe according to the installing requirements
H. Wiring the power cable, special to care of the shield braid of the cable connected to the grounding terminal of the
converter.
I. Power on the meter in field, first observe if there is any leakage around the meter (care of the personal safety), then
the display if it appears any changes of transient flow rate, if it is not, check it as step above carefully, particular to
the wiring, power supply, shield braid grounding, the specification of the meter and meter surrounding, or contact
technical service center of local agency.
2. Specification
Measuring Range : Steam – 1.6 to 540,000 Kg/Hr Intrinsically safe Exia IIC T4
ANSI 150# / ANSI 300# / ANSI 600# Housing Material : Aluminum Alloy
3
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
3. Diameter
D1
L1
154mm
mm Inch L2 H2
10 3/8'' 200 475
RATE
TOT
TEMP
112. 7m3
25°C
15 1/2" 200 480
PRES 1 . 6M pa
4
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
4. Installation
Vortex Flowmeter Pipe installed location is very important, it concerned to measuring accuracy, please left the
multiple space in the upstream and downstream of the straight pipe as you can, if the pipe is confirmed with not enough
straight pipe, please install 2/3 of total straight pipe length on the upstream, 1/3 on downstream, but the Flowmeter can’t
matched factory accuracy. Example
20D 5D
20D 5D
50D 5D
5
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
18D 5D
18D 5D
● Grounding measure
AVF7000 requested the perfect grounding, to erase the interference, the grounding as follow, only need to put the
converter housing connected with grounding. Sensor don’t need to connected grounding again..
Grounding point, ex.: Stairs, metal fence…etc.
6
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
5. Wiring Chart
This meter uses two wires to transmit 4-20 mA output signal to other external equipment, power supply 11 ~ 36VDC,
the maximum loading resistance for output circuit is 600Ω(including resistance of cable wire)
In general condition, 600V PVC isolating wire or cable be used as connecting wire. The two core shielding wire
(RWP2×0.5mm) be used in the place where electricity noise occurred easily, the out layer of shielding wire should be
connected to the grounding screw in the house of amplifier fixedly. Uses appropriate cable to conform with the operating
temperature if the temperature is too high or low.
RS485 Output-
RS485 Output+
+24VDC
Pulse Output+
0V (Pulse Output - )
ON
K
1 2 34
V+ F V- V+ F V- T+ T- T- PV PV PI PI
+
mA
- A B b
+ + - Current Veltage
+ - 12-36 Pulse 0V
VDC Output
Output
Max. 20uA
Input
0 ~30mVDC
12-36 VDC
Power Supply Power
Supply
ON ON
OFF OFF
PT100 Ω Pressure Transmitter
1 2 3 4 1 2 3 4 RTD (3 Wire) 4 Wire
7
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
5.2 separate version
Converter
Sensor RS485 Output-
RS485 Output+
+24VDC
Pulse Output+
0V (Pulse Output - )
PI+ PI+
PI- PI-
PV+ PV+ ON
PV- PV-
T+
K
T+ 1 2 34
TA- TA-
TB- TB-
S1+ S1+
S2+ S2+
V+ F V- V+ F V- T+ T- T- PV PV PI PI
+
mA
- A B b
+ + - Current
Output
Veltage
Input
+ - 12-36 Pulse 0V
VDC Output
Max . 20uA 0 ~30mVDC
12-36 VDC
Power Supply Power
Supply
ON ON
OFF OFF
PT100 Ω Pressure Transmitter
1 2 3 4 1 2 3 4 RTD (3 Wire) 4 Wire
8
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
6. Panel Display
RAT E 10.5m3/ h
RAT E 10.5m3/ h
TOT 20.0m3
T EMP 25°C
TOT. 20.0m3
PRES 1. 5Mpa
7. Function
7.1 Button Function
OUT 12.5mA Row 1 : output content. If choose current output, it shows the output current value;
DEN ----kg/m3 If choose pulse output, it shows frequency output value.
E.T. 23.1°C Row 2 : fluid density.
IN 88hz Row 3 : PCB temperature.
Row 4 : frequency sensor sampled.
OUT 12.5mA Row 1 : output content. If choose current output, it shows the output current value;
DEN ----kg/m3 If choose pulse output, it shows frequency output value.
E.T. 23.1°C Row 2 : fluid density.
IN 88hz Row 3 : PCB temperature.
Row 4 : frequency sensor sampled.
9
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
8. General Parameter Operation chart
RATE 10.5 m3/h RATE 10.5 m3/h
TOT. 20.0 m3
TEMP 25°C
TOT. 20.0 m3 PRES 1.5MPa
SET
SE T
PARAMETER B
MODIFY A
TOTALIZER C
PASSWORD K
TEMP HI LIMIT
PARAMETER MODIFY TOTALIZER
DAMPING
COMMUNICATION MIN VEL LIQUID
HIGH FLOW
ADDRESS FREQUENCY FILTER
LOW FLOW
SENSOR TYPE SAVE
PULSE OUTPUT
SAVE
PARAMETER B
MODIFY A
TOTALIZER C
PASSWORD K
10
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
9. Calibration Parameter Operation Chart
TEMPERATURE B
PRESSURE A
mA OUTPUT C
ZERO POINT K
Page 12
TEMPEARTURE PRESSURE
TEMP PRE.
CALIBRATION CALIBRATION
TEMP SET HIGH TEMP RES. HIGH PRE. SET HIGH PRE. CAL. HIGH
TEMP SET LOW TEMP RES. LOW PRE. SET LOW PRE. CAL. LOW
TEMPERATURE B
PRESSURE A
mA OUTPUT C
ZERO POINT K
11
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
TEMPERATURE B
PRESSURE A
mA OUTPUT C
ZERO POINT K
mA ZERO
CALIBRATION CALIBRATION
SAVE SAVE
TEMPERATURE B
PRESSURE A
mA OUTPUT C
ZERO POINT K
12
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
10. General Parameter Setting
SET
SE T
SET
SE T
SET
SE T
Measure the Medium selection: Press button to select the different Medium No.
APPLICATION 1. Gas 2. Liquid 3. Gas+P+T 4.Liquid+P+T 5. OIL+P+T 6.Natual Gas+P+T
7. Heat Steam+P+T 8. Saturated Steam +T 9. Saturated Steam+P
2. LIQUID
If choose the “1. Gas” or “2. Liquid”, only 2 lines will display, Which is flowrate and Totalizer.
If choose others, they will be 4 lines display, That is flowrate, Totalizer, Temp.and Pressure.
SET
SE T
SET
SE T
SET
SE T
13
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
SET
SE T
SET
SE T
SET
SE T
14
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
PARAMETER B
MODIFY A
TOTALIZER C
PASSWORD K
15
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
10.2 Tantalization Zero Clearing
1234
SET
SET
SET
SET
16
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
11. Calibration Parameter Setting
CAL
CAL
CAL
CAL
CAL
17
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
CAL
CAL
CAL
CAL
18
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
11.3 Electric current output calibration
PASSWORD Input Password : 1234 and it will go to the electric current calibrate operation
1234 Press button to move the position, press button to change the digit.
CAL
CAL
19
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
11.4 Zero Point Calibration Operate
PASSWORD Input password: 1234 and it will go to the Zero calibrate operation
1234 Press button to move the position, press button to change the digit.
CAL
CAL
● Zero Adjustment
When sensor installed in the pipe, the pipe with no flowrate, and shows the low flow rate in display, then please does the Zero
Adjustment.
Situation 1: Zero Adjustment could through the shortcut key of electrical board as follow.
Press the ZERO button about 5 seconds, the light is turned on, then loosen the button, when light about 5 seconds ,it start glitter,
and will enter into zero adjustment status, the glittering time about 3 seconds, the zero adjustment is completed once the light turn off.
Situation 2: Through the window to do the zero adjustment. It will search the zero position. Please check 11.4 Zero Point Calibration
Operate. When you finished the auto zero adjustment, the small flowrate still exist, please press CAL button at Zero Point Calibration
selected MENU (hand zero adjustment), if the value is 3,please add the zero point value. Change to 4, whether the status improved, if
it still exists, please adds one more point. The bigger the value is, the better the anti-interference. But when the value too big, it will
cause no signal, please select carefully.
RATE 15.2m3/h
TOT 112.7m3
TEMP 25ºC
PRES 1.6MPa
ZERO
20