Alia Manual Avf7000

Download as pdf or txt
Download as pdf or txt
You are on page 1of 20

AVF7000 Operator Manual OP7000.1.1.5.

R5ENG

ALIAVTX
®

Vortex Flowmeter
AVF7000 Series
Operator Manual

Alia Group Inc. URL: http://www.alia-inc.com


113 Barksdale Professional Center, Newark, DE 19711, USA E-mail: [email protected]
TEL : +1 - 302 - 213 - 0106 FAX : +1 - 302 - 213 - 0107 OP7000.1.1.5.R5ENG
1
AVF7000 Operator Manual OP7000.1.1.5.R5ENG

Index
1. Flowmeter Check - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3

2. Specification - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3

3.Diameter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4

4. Installation- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5

5. Wiring Chart - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7

5.1 compact version - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7

5.2 separate version - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8

6. Panel Display - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9

7. Function - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9

7.1 Button Function- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9

7.2 Display Function - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9

8. General Parameter Operation Chart- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10

9. Calibration Parameter Operation Chart- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 11

10. General Parameter Setting- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 13

10.1 Parameter and Adjustment information setting - - - - - - - - - - - - - - - - - - - - - - - - - - - - 13

10.2 Totalization Zero Clearing- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 16

10.3 Password Revised- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 16

11. Calibration Parameter Setting- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 17

11.1 Temp. Calibrate Operation - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 17

11.2 Pressure Calibrate Operation - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 18

11.3 Electric current output calibration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 19

11.4 Zero Point Calibration Operate - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 20

2
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
1. Flowmeter Check
A. First check the package if it is good or damage, if broken hard you should notify the meter transport department or
contact the customer service center of Alia immediately
B. Open the package and check for meter and all parts attachment if it is good or damage and shortage
C. Read the operating instruction in detail and comprehends all contents, if any part of that you do not understand full,
fax technical service department of Alia
D. Make sure that the specification of the meter you received is conformed to the operating condition
E. Power on the meter in house, observe the LCD display if it is on normal
F. Select proper meter install site, make sure to meet the installing condition
G. Meter be moved to the field and be mounted on the pipe according to the installing requirements
H. Wiring the power cable, special to care of the shield braid of the cable connected to the grounding terminal of the
converter.
I. Power on the meter in field, first observe if there is any leakage around the meter (care of the personal safety), then
the display if it appears any changes of transient flow rate, if it is not, check it as step above carefully, particular to
the wiring, power supply, shield braid grounding, the specification of the meter and meter surrounding, or contact
technical service center of local agency.

2. Specification

Size (mm) : 10,15,20,25,32,40,65,80,100,125,150, Protection : IP 65.

200,250,300,350,400,450,500 mm Explosion Proof Exd IIC T6

Measuring Range : Steam – 1.6 to 540,000 Kg/Hr Intrinsically safe Exia IIC T4

Gas - 3 to4 6,000 m3/Hr Local display : 4 Line LCD

Liquid - 0.3 to 4950 m3/Hr 4 digit Flowrate

Material : Stainless Steel 304 / 316 8 digit Totalization

Accuracy Output : 4~20 mA (2wire)

Liquid : +/- 0.7% Load : 600 Ω

Gas/Steam : +/- 1.0% Pulse Output : Standard Pulse Output

Repeatability : +/- 0.2% Output Range : 3 to 30VDC, 20mA Max.

Connection : Flange / Wafer Communication : RS485

Operation Parameter, Totalization Figures


Flange Type : JIS 10K / JIS 20K / JIS 40K Data storage :
are stored by EEPROM.

ANSI 150# / ANSI 300# / ANSI 600# Housing Material : Aluminum Alloy

DIN PN 10 / PN 16 / PN25 / PN 40 Cable Entry : 2x0.5mm2

Wafer Type : 40,65,80,100,125,150, 200,250 Power Supply : 12~36 VDC

4 keys from internal for Programming and


without Temperature / Pressure Sensor Key Pad :
display contro

Temperature : -40 ~ +280℃ ( Standard Type) Option

-4 ~ +420 ℃ (Explosion proof Type) Pressure Transmitter Pressure Compensation

Ambient Temperature : -20 ~+60℃ Signal Input 0~30 mV DC

Pressure : 78 Kg/cm2 (Max.) Temperature sensor Temperature sensor

Signal Input PT100(3 Wire)

3
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
3. Diameter

112mm 154mm Normal Size Dimensions (mm)


mm Inch L1 D1 H1
40 1-1/2'' 68 92 435
RATE
112 . 7m3
50 68 98 438
TOT
TEMP
PRES
25°C
1 . 6Mpa 2''
65 2-1/2'' 68 110 453
80 3'' 80 134 476
100 4'' 68 158 499
125 5" 70 175 520
150 6" 92 200 545
200 8" 112 250 595

H1 250 10" 132 300 645

D1

L1

154mm
mm Inch L2 H2
10 3/8'' 200 475
RATE
TOT
TEMP
112. 7m3
25°C
15 1/2" 200 480
PRES 1 . 6M pa

20 3/4" 200 490


25 1" 200 500
32 1-1/4" 200 515
40 1-1/2" 200 525
50 2" 200 540

H2 65 2-1/2" 200 560


80 3" 225 575
100 4" 250 610
125 5" 275 625
150 6" 300 675
200 8" 350 735
250 10" 400 800
300 12" 450 860
350 14" 500 930
400 16" 550 995
450 18" 600 1045
L2
500 20" 650 1105

4
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
4. Installation

Vortex Flowmeter Pipe installed location is very important, it concerned to measuring accuracy, please left the
multiple space in the upstream and downstream of the straight pipe as you can, if the pipe is confirmed with not enough
straight pipe, please install 2/3 of total straight pipe length on the upstream, 1/3 on downstream, but the Flowmeter can’t
matched factory accuracy. Example

20D 5D

20D 5D

50D 5D

5
AVF7000 Operator Manual OP7000.1.1.5.R5ENG

18D 5D

18D 5D

● Grounding measure

AVF7000 requested the perfect grounding, to erase the interference, the grounding as follow, only need to put the
converter housing connected with grounding. Sensor don’t need to connected grounding again..
Grounding point, ex.: Stairs, metal fence…etc.

6
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
5. Wiring Chart
This meter uses two wires to transmit 4-20 mA output signal to other external equipment, power supply 11 ~ 36VDC,
the maximum loading resistance for output circuit is 600Ω(including resistance of cable wire)
In general condition, 600V PVC isolating wire or cable be used as connecting wire. The two core shielding wire
(RWP2×0.5mm) be used in the place where electricity noise occurred easily, the out layer of shielding wire should be
connected to the grounding screw in the house of amplifier fixedly. Uses appropriate cable to conform with the operating
temperature if the temperature is too high or low.

5.1 compact version

RS485 Output-
RS485 Output+
+24VDC
Pulse Output+
0V (Pulse Output - )

T+ T- T- PV- PI- PI+ B A V+ F V-

ON
K
1 2 34

4-20 mA Output Pulse Output Sensor Input


Temperature Pressure

V+ F V- V+ F V- T+ T- T- PV PV PI PI

+
mA
- A B b
+ + - Current Veltage
+ - 12-36 Pulse 0V
VDC Output
Output
Max. 20uA
Input
0 ~30mVDC
12-36 VDC
Power Supply Power
Supply

ON ON
OFF OFF
PT100 Ω Pressure Transmitter
1 2 3 4 1 2 3 4 RTD (3 Wire) 4 Wire

7
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
5.2 separate version

Converter
Sensor RS485 Output-
RS485 Output+
+24VDC
Pulse Output+
0V (Pulse Output - )

T+ T- T- PV- PI- PI+ B A V+ F V-

PI+ PI+
PI- PI-
PV+ PV+ ON
PV- PV-
T+
K
T+ 1 2 34
TA- TA-
TB- TB-
S1+ S1+
S2+ S2+

4-20 mA Output Pulse Output Sensor Input


Temperature Pressure

V+ F V- V+ F V- T+ T- T- PV PV PI PI

+
mA
- A B b
+ + - Current
Output
Veltage
Input
+ - 12-36 Pulse 0V
VDC Output
Max . 20uA 0 ~30mVDC
12-36 VDC
Power Supply Power
Supply

ON ON
OFF OFF
PT100 Ω Pressure Transmitter
1 2 3 4 1 2 3 4 RTD (3 Wire) 4 Wire

8
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
6. Panel Display

RAT E 10.5m3/ h
RAT E 10.5m3/ h
TOT 20.0m3
T EMP 25°C
TOT. 20.0m3
PRES 1. 5Mpa

> SET > CAL


SET CAL

7. Function
7.1 Button Function

Key Button Measure state Function Parameter Setting State Function


Name
SE
SETT Save the parameter setting so far, and go to next
Go to the parameter setting mode
Setting parameter.

Select the contents selected Move


Move
Change the contents selection Revised the present display
Up
CAL Save the calibration setting at present, and go to
Go to the calibration parameter setting
Calibration next parameter.

7.2 Display Function


RATE 10.5 m3/h Display without Temp./ Pressure compensation
If choose the application of “1. Gas” or “2. Liquid ”, only 2 lines will display,
TOT. 20.0 m3 That is flowrate and totalizer.

OUT 12.5mA Row 1 : output content. If choose current output, it shows the output current value;
DEN ----kg/m3 If choose pulse output, it shows frequency output value.
E.T. 23.1°C Row 2 : fluid density.
IN 88hz Row 3 : PCB temperature.
Row 4 : frequency sensor sampled.

RATE 10.5 m3/h Display without Temp./ Pressure compensation


TOT. 20.0 m3 If choose the application of “1. Gas” or “2. Liquid ”, only 2 lines will display,
TEMP 25°C
PRES 1.5MPa That is flowrate and totalizer.

OUT 12.5mA Row 1 : output content. If choose current output, it shows the output current value;
DEN ----kg/m3 If choose pulse output, it shows frequency output value.
E.T. 23.1°C Row 2 : fluid density.
IN 88hz Row 3 : PCB temperature.
Row 4 : frequency sensor sampled.
9
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
8. General Parameter Operation chart
RATE 10.5 m3/h RATE 10.5 m3/h
TOT. 20.0 m3
TEMP 25°C
TOT. 20.0 m3 PRES 1.5MPa
SET
SE T

PARAMETER B
MODIFY A
TOTALIZER C
PASSWORD K

TEMP HI LIMIT
PARAMETER MODIFY TOTALIZER

Password: 1234 TEMP LOW LIMIT Password: **** Password: 1234

APPLICATION PRES HI LIMIT DATE CLEAR TOTAL

SIZE PRES LOW LIMIT CALIBRATION NO.

FLOW CURVE COMP. FACTOR SERIAL NO.

FLOW UNIT FLUID DENSITY MAX VEL GAS PASSWORD

K FACTOR STD. TEMP MIN VEL GAS Password: 1234

OUTPUT FLOW CUTEOFF MAX VEL LIQUID NEW PASSWORD

DAMPING
COMMUNICATION MIN VEL LIQUID

HIGH FLOW
ADDRESS FREQUENCY FILTER

LOW FLOW
SENSOR TYPE SAVE

PULSE OUTPUT
SAVE
PARAMETER B
MODIFY A
TOTALIZER C
PASSWORD K

10
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
9. Calibration Parameter Operation Chart

RATE 10.5 m3/h RATE 10.5 m3/h


TOT. 20.0 m3
TEMP 25°C
TOT. 20.0 m3 PRES 1.5MPa
CAL

TEMPERATURE B
PRESSURE A
mA OUTPUT C
ZERO POINT K
Page 12

TEMPEARTURE PRESSURE

Password: 1234 Password: 1234

TEMP PRE.
CALIBRATION CALIBRATION

TEMP SET HIGH TEMP RES. HIGH PRE. SET HIGH PRE. CAL. HIGH

TEMP SET LOW TEMP RES. LOW PRE. SET LOW PRE. CAL. LOW

SAVE SAVE SAVE

TEMPERATURE B
PRESSURE A
mA OUTPUT C
ZERO POINT K

11
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
TEMPERATURE B
PRESSURE A
mA OUTPUT C
ZERO POINT K

mA OUTPUT ZERO POINT

Password: 1234 Password: 1234

mA ZERO
CALIBRATION CALIBRATION

mA SET 20mA mA CAL. 20mA SET ZEROVALUE AUTOZEROVALUE

mA SET 4mA mA CAL. 4mA SAVE SAVE

SAVE SAVE

TEMPERATURE B
PRESSURE A
mA OUTPUT C
ZERO POINT K

12
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
10. General Parameter Setting

10.1 Parameter and Adjustment information setting

RATE 10.5 m3/h RATE 10.5 m3/h


TOT. 20.0 m3
TEMP 25°C
TOT. 20.0 m3 PRES 1.5MPa

SET
SE T

PARAMETER B Parameter and adjustment setting


MODIFY A Press button could select: parameter setting, Modify, totalization reset, password revised.
TOTALIZER C SET
SET
PASSWORD K Select PARAMETER and press button to do the operation.

SET
SE T

PASSWORD Input password: 1234 and it will go to parameter setting state


Press button to move position, and press button to change the Value
1234

SET
SE T
Measure the Medium selection: Press button to select the different Medium No.
APPLICATION 1. Gas 2. Liquid 3. Gas+P+T 4.Liquid+P+T 5. OIL+P+T 6.Natual Gas+P+T
7. Heat Steam+P+T 8. Saturated Steam +T 9. Saturated Steam+P
2. LIQUID
If choose the “1. Gas” or “2. Liquid”, only 2 lines will display, Which is flowrate and Totalizer.
If choose others, they will be 4 lines display, That is flowrate, Totalizer, Temp.and Pressure.
SET
SE T

SIZE mm Size Selection


0050 Press button to move the position, Press button to change the digit.
Setting Range 0000 ~ 9999 mm
SET
SE T

FLOW CURVE Flow Curve


1. LINEAR 2. 6 POINT
LINEAR
If choose the LINEAR, will come to next window

SET
SE T

K FACTOR Flowmeter K Factor


Setting Range:0.000000-99999999 Pulse/M3
09075.42

SET
SE T

FLOW UNIT Flowrate Unit


1. m3/h 2. km3/h 3. kg/h 4. t/h
1. m3/h
Press button to do the selection.
SET
SE T

OUTPUT Output Type


4-20mA 1. 4-20mA Output 2. Unscale pulse output 3. Scale pulse output
Press button do the selection.
SET
SE T

13
AVF7000 Operator Manual OP7000.1.1.5.R5ENG

DAMPING Sec. Reaction Time


Setting Range:01-99 seconds
05
Press button to move the position, press button to change the digit.
SET
SE T

HIGH FLOW m3/h Flowrate Up Limit


Setting Range:0.000000-99999999 m3/h
050.00

SET
SE T

LOW FLOW m3/h Flowrate Lower Limit


Setting Range:0.000000-99999999 m3/h
0.000

SET
SE T

PULSE OUTPUT m3 Pulse Output


Setting Range:0.000000-99999999 Pulse/M3
09765.00

SET
SE T

TEMP HI LIMIT Temp. Up Limit


Setting Range:-999 -+9999 ℃
0430 ℃
When Temp. High and low limit are the same, it’s stable Temper
SET
SE T

TEMP LOW LIMIT Temp Lower Limit


Setting Range:-999 -+9999 ℃
0.00 ℃
When Temp. High and low limit are the same, it’s stable Temperature.
SET
SE T

PRES HI LIMIT Pressure high Limit


Setting Range:-98888-+99999999 Mpa
0002.000 Mpa
When pressure high and low limits are the same, it’s stable pressure.
SET
SE T

PRES LOW LIMIT Pressure Low Limit


Setting Range:-98888-+99999999 Mpa
0000. 000 MPa
When pressure high and low limits are the same, it’s stable pressure.
SET
SE T

COMP. FACTOR Compress Factor


Setting range:0.000000~99999999
1.000000
Original setting digit:1.000000
SET
SE T

14
AVF7000 Operator Manual OP7000.1.1.5.R5ENG

FLUID DENSITY Density Setting


Setting Range:0.000000~99999999 kg/m3
0740.000 kg/m3
Fluid density setting only for Engineering Unit are Kg/H or T/H
SET
SET

GAS PRESSURE ATM Pressure


Setting Range:0.000000-99999999 Mpa
0.101325 Mpa
Local Pressure original setting value:0.101325 Mpa

SET
SET

STD. TEMP. Standard Temperature


Setting Range:00-99 ℃
00 ℃
Local pressure original setting value:00

SET
SET

FLOW CUTEOFF Low Flowrate Cut Off


Setting Range:0.000000-99999999
00.000 m3/h
When it’s lower than the setting value, it won’t display, Normally setting the Max. flowrate is 5%
SET
SET

COMMUNICATION Communication mode


1. No 2. RS-485 3. RS-232
NO

SET
SET

ADDRESS Communication Address


Setting Range:0000-9999
0000

SET
SET

SENSOR TYPE Sensor Type


1. Standard Type 2. Special 3. Insert Type
STANDRAD

SET
SET

SAVE Parameter Save


Select whether to save the parameter setting, and turn back to main function
YES NO
Press SET for 3 second, then loose, and turn back to main function
SET
SET

PARAMETER B
MODIFY A
TOTALIZER C
PASSWORD K

15
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
10.2 Tantalization Zero Clearing

PARAMETER B Parameter and adjust information setting main function


MODIFY A
TOTALIZER C Press button could select: parameter setting, Modify, totalization reset, password revised.
SET
SET
PASSWORD K Select TOTALIZER and press button to do the operation.
SET
SET

PASSWORD Input Password:1234 and it will go to totalization zero clearing operation

1234

SET
SET

CLEAR TOTAL Totalization Zero Clearing


Select if it needs Totalization Zero Clearing
YES NO

10.3 Password Revised

PARAMETER B Parameter and adjust information setting main function


MODIFY A
TOTALIZER C Press button could select: parameter setting, Modify, totalization reset, password revised.
PASSWORD K Select PASSWORD and press SET
SET
button to do the operation.
SET
SET

NEW PASSWORD Password Revised


Revised to new password
0000

SET
SET

SAVE Save Parameter


Select if it needs Save Parameter Setting, and go back to Main Function
YES NO

16
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
11. Calibration Parameter Setting

11.1 Temp. Calibrate Operation

RATE 10.5 m3/h RATE 10.5 m3/h


TOT. 20.0 m3
TEMP 25°C
TOT. 20.0 m3 PRES 1.5MPa

CAL

TEMPERATURE B Calibration Setting Main Function


PRESSURE A
mA OUTPUT C Press button could select: Temp. Calibration, Pressure Calibration, Electric Current Calibration,
ZERO POINT K Zero Point Calibration
CAL
Select TEMPERATURE and press button to do the operation.
CAL

PASSWORD Input Password : 1234 and go to Temp. Calibrate operation


1234 Press button to move the position, press button to change the digit.

CAL

TEMP Temp. Calibration Function


CALIBRATION
Window Selection : Setting / AUTO / Turn Back
SET RES. BACK Press button to move the position

CAL

TEMP SET HIGH Setting Temp. Calibration Mode


Input Temp. Up Limit
3023.1
Press button to move the position, press button to change the digit.

CAL

TEMP SET LOW Setting Temp. Calibration Mode


0750.2 Input Temp. Lower Limit
Press button to move the position, press button to change the digit.

CAL

SAVE Save Setting


YES NO Select if it needs Save Setting
CAL
Press button to confirm

17
AVF7000 Operator Manual OP7000.1.1.5.R5ENG

11.2 Pressure Calibrate Operation

TEMPERATURE B Calibration Setting Function


PRESSURE A
mA OUTPUT C Press button could select: Temp. Calibration, Pressure Calibration, Electric Current Calibration,
ZERO POINT K Zero Point Calibration.
CAL
Select PRESSURE and then press button to do the operation.
CAL

PASSWORD Input password : 1234 and it will go to pressure calibrate operation


1234 Press button to move the position, press button to change the digit.

CAL

PRE. Pressure Calibration Mode


CALIBRATION
Window selection: Setting / Calibration / Turn back
SET CAL. BACK Press button to move the position

CAL

PRE. SET HIGH Setting Pressure Calibration Up Limit


Input Pressure High Limit
2972.6
Press button to move the position, press button to change the digit.

CAL

PRE. SET LOW Setting Pressure Calibration Lower Limit


Input Pressure Low Limit
0967.4
Press button to move the position, press button to change the digit.

CAL

SAVE Save Setting


Select if it needs Save Setting
YES NO
CAL
Press button to confirm

18
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
11.3 Electric current output calibration

TEMPERATURE B Calibration Setting Function


PRESSURE A
mA OUTPUT C Press button could select: Temp. Calibration, Pressure Calibration, Electric current calibration,
ZERO POINT K Zero point calibration.
CAL
Select mA OUTPUT and press button to do the operation.
CAL

PASSWORD Input Password : 1234 and it will go to the electric current calibrate operation
1234 Press button to move the position, press button to change the digit.

CAL

mA Electric Current Calibration


CALIBRATION
Window Selection : Setting, AUTO, Turn back
SET CAL. BACK Press button to move the position,

CAL

mA SET 20mA Adjust 20mA Output


20 mA electric current output value amended, if mA output value too low, please increase the value,
004235 if the value too high, please decrease the value.
Press button to move the position, press button to change the digit.
CAL

mA SET 4mA Adjust 4mA Output


4 mA electric current output value amended, if mA output value too low, please increase the value, if
000711 the value too high, please decrease the value.
Press button to move the position, press button to change the digit.
CAL

SAVE Save Setting


Select if it needs Save Setting
YES NO
CAL Press CAL
button to confirm

19
AVF7000 Operator Manual OP7000.1.1.5.R5ENG
11.4 Zero Point Calibration Operate

TEMPERATURE B Calibration Setting Main Function


PRESSURE A
mA OUTPUT C Press button could select: Temp. Calibration, Pressure Calibration, Electric current calibration,
ZERO POINT K Zero point calibration.
CAL
Press ZERO POINT and then press button to do the operation.
CAL

PASSWORD Input password: 1234 and it will go to the Zero calibrate operation
1234 Press button to move the position, press button to change the digit.

CAL

ZERO Zero Point Calibration


CALIBRATION
Window Selection: By manual / Auto / Turn back. Press button could select.
MENU AUTO BACK If choose the MENU, will come to next window. Then set zero manually.

CAL

SET ZERO VALUE Zero Point Calibration by manual


4 Zero Point Adjust Range: :0 - 7
Press button to change the digit.
CAL

SAVE Save Setting


Select if it needs Save Setting
YES NO
CAL
Press button to confirm

● Zero Adjustment
When sensor installed in the pipe, the pipe with no flowrate, and shows the low flow rate in display, then please does the Zero
Adjustment.
Situation 1: Zero Adjustment could through the shortcut key of electrical board as follow.
Press the ZERO button about 5 seconds, the light is turned on, then loosen the button, when light about 5 seconds ,it start glitter,
and will enter into zero adjustment status, the glittering time about 3 seconds, the zero adjustment is completed once the light turn off.

Situation 2: Through the window to do the zero adjustment. It will search the zero position. Please check 11.4 Zero Point Calibration
Operate. When you finished the auto zero adjustment, the small flowrate still exist, please press CAL button at Zero Point Calibration
selected MENU (hand zero adjustment), if the value is 3,please add the zero point value. Change to 4, whether the status improved, if
it still exists, please adds one more point. The bigger the value is, the better the anti-interference. But when the value too big, it will
cause no signal, please select carefully.

RATE 15.2m3/h
TOT 112.7m3
TEMP 25ºC
PRES 1.6MPa

ZERO

20

You might also like