Datasheet: Kit Contents - NI 9263 - NI 9263 Getting Started Guide
Datasheet: Kit Contents - NI 9263 - NI 9263 Getting Started Guide
NI 9263
4 AO, ±10 V, 16 Bit, 100 kS/s/ch Simultaneous
The NI 9263 is an analog output module for any CompactDAQ and CompactRIO systems. It
also features ±30 V overvoltage protection, short-circuit protection, low crosstalk, fast slew
rate, high relative accuracy, and NIST-traceable calibration. The NI 9263 module includes a
channel-to-earth ground double isolation barrier for safety and noise immunity.
• NI 9263
Kit Contents
• NI 9263 Getting Started Guide
Voltage 3 Vrms
NI 9260 2 51.2 kS/s/ch Yes 24-Bit BNC, mini XLR
Output
Voltage Screw-Terminal,
NI 9263 ±10 V 4 100 kS/s/ch Yes 16-Bit
Output Spring-Terminal
Voltage Spring-Terminal,
NI 9264 ±10 V 16 25 kS/s/ch Yes 16-Bit
Output 37-Pin DSUB
Current 0 mA to
NI 9265 4 100 kS/s/ch Yes 16-Bit Screw-Terminal
Output 20 mA
Voltage
NI 9269 ±10 V 4 100 kS/s/ch Yes 16-Bit Screw-Terminal
Output
NI C Series Overview
NI provides more than 100 C Series modules for measurement, control, and communication
applications. C Series modules can connect to any sensor or bus and allow for high-accuracy
measurements that meet the demands of advanced data acquisition and control applications.
• Measurement-specific signal conditioning that connects to an array of sensors and signals
• Isolation options such as bank-to-bank, channel-to-channel, and channel-to-earth ground
• -40 °C to 70 °C temperature range to meet a variety of application and environmental
needs
• Hot-swappable
The majority of C Series modules are supported in both CompactRIO and CompactDAQ
platforms and you can move modules from one platform to the other with no modification.
CompactDAQ
CompactDAQ is a portable, rugged data acquisition platform
that integrates connectivity, data acquisition, and signal
conditioning into modular I/O for directly interfacing to any
sensor or signal. Using CompactDAQ with LabVIEW, you
can easily customize how you acquire, analyze, visualize,
and manage your measurement data.
Software
LabVIEW Professional Development System for Windows
Circuitry
Each channel has a digital-to-analog converter (DAC) that produces a voltage signal. Each
channel also has overvoltage and short-circuit protection.
Figure 1. Output Circuitry for One Channel of the NI 9263
Overvoltage/
Isolated DAC Amp Short-Circuit AO
Protection
COM
NI 9263
NI 9263 Specifications
The following specifications are typical for the range -40 °C to 70 °C unless otherwise noted.
All voltages are relative to COM unless otherwise noted.
Caution Do not operate the NI 9263 in a manner not specified in this document.
Product misuse can result in a hazard. You can compromise the safety protection
built into the product if the product is damaged in any way. If the product is
damaged, return it to NI for repair.
Output Characteristics
Number of channels 4 analog output channels
DAC resolution 16 bits
Table 1. Accuracy
Measurement Conditions Percent of Reading Percent of Range3
(Gain Error) (Offset Error)
Stability
Gain drift 11 ppm/°C
Offset drift 110 μV/°C
Protection
Overvoltage ±30 V
Short-circuit Indefinitely
1 When the module powers on, a glitch occurs for 20 μs peaking at -1.5 V.
2 The power-down voltage peaks at 1.8 V before exponentially discharging to 0 V in 100 μs. You can
add a 10 kΩ load to reduce the peak voltage.
3 Range equals ±10.7 V
4 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes
where the calibration constants stored in the module are not applied to the data.
Noise
Updating at 100 kS/s 600 μVrms
Not updating 260 μVrms
Slew rate 4 V/μs
Crosstalk 76 dB
Settling time (100 pF load, to 1 LSB)
Full-scale step 20 μs
1 V step 13 μs
0.1 V step 10 μs
Capacitive drive 1,500 pF minimum
Monotonicity 16 bits
DNL ±1 LSB maximum
INL (endpoint) ±12 LSB maximum
MTBF 1,732,619 hours at 25 °C; Bellcore Issue 2,
Method 1, Case 3, Limited Part Stress Method
Power Requirements
Power consumption from chassis
Active mode (at -40 °C) 500 mW maximum
Sleep mode 25 μW maximum
Thermal dissipation (at 70 °C)
Active mode 750 mW maximum
Sleep mode 25 μW maximum
Screw-terminal wiring
Gauge 0.2 mm2 to 2.5 mm2 (26 AWG to 14 AWG)
copper conductor wire
Wire strip length 13 mm (0.51 in.) of insulation stripped from
the end
Temperature rating 90 °C minimum
Torque for screw terminals 0.5 N · m to 0.6 N · m
(4.4 lb · in. to 5.3 lb · in.)
Wires per screw terminal One wire per screw terminal; two wires per
screw terminal using a 2-wire ferrule
Ferrules 0.25 mm2 to 2.5 mm2
Spring-terminal wiring
Gauge 0.2 mm2 to 2.5 mm2 (30 AWG to 12 AWG)
copper conductor wire
Wire strip length 10 mm (0.39 in.) of insulation stripped from
the end
Temperature rating 90 °C minimum
Wires per spring terminal One wire per spring terminal; two wires per
spring terminal using a 2-wire ferrule
Ferrules 0.25 mm2 to 2.5 mm2
Connector securement
Securement type Screw flanges provided
Torque for screw flanges 0.2 N · m (1.80 lb · in.)
Weight
NI 9263 with screw terminal 150 g (5.3 oz)
NI 9263 with spring terminal 139 g (4.9 oz)
Channel-to-channel None
Channel-to-earth ground
Continuous 250 Vrms, Measurement Category II
Withstand 2,300 Vrms, verified by a 5 s dielectric
withstand test
Division 2 and Zone 2 hazardous locations 60 VDC, Measurement Category I
applications (Channel-to-earth ground)
Caution Do not connect the NI 9263 to signals or use for measurements within
Measurement Categories II, III, or IV.
Note Measurement Categories CAT I and CAT O are equivalent. These test and
measurement circuits are not intended for direct connection to the MAINS building
installations of Measurement Categories CAT II, CAT III, or CAT IV.
Caution Do not connect the NI 9263 to signals or use for measurements within
Measurement Categories III or IV.
Hazardous Locations
U.S. (UL) Class I, Division 2, Groups A, B, C, D, T4;
Class I, Zone 2, AEx nA IIC T4
Canada (C-UL) Class I, Division 2, Groups A, B, C, D, T4;
Class I, Zone 2, Ex nA IIC T4
Europe (ATEX) and International (IECEx) Ex nA IIC T4 Gc
Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for sensitive electrical
equipment for measurement, control, and laboratory use:
• EN 61326 (IEC 61326): Class A emissions; Industrial immunity
• EN 55011 (CISPR 11): Group 1, Class A emissions
• AS/NZS CISPR 11: Group 1, Class A emissions
• FCC 47 CFR Part 15B: Class A emissions
• ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for
use in commercial, light-industrial, and heavy-industrial locations. In Europe,
Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended
for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical
equipment that does not intentionally generate radio frequency energy for the
treatment of material or inspection/analysis purposes.
Note For EMC declarations and certifications, and additional information, refer to
the Online Product Certification section.
CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
• 2014/35/EU; Low-Voltage Directive (safety)
• 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
• 94/9/EC; Potentially Explosive Atmospheres (ATEX)
Operating vibration
Random (IEC 60068-2-64) 5 grms, 10 Hz to 500 Hz
Sinusoidal (IEC 60068-2-6) 5 g, 10 Hz to 500 Hz
Operating shock (IEC 60068-2-27) 30 g, 11 ms half sine; 50 g, 3 ms half sine;
18 shocks at 6 orientations
Environmental
Refer to the manual for the chassis you are using for more information about meeting these
specifications.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible
manner. NI recognizes that eliminating certain hazardous substances from our products is
beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact
web page at ni.com/environment. This page contains the environmental regulations and
directives with which NI complies, as well as other environmental information not included in
this document.
电子信息产品污染控制管理办法(中国 RoHS)
中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物
质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录
ni.com/environment/rohs_china。(For information about China RoHS
compliance, go to ni.com/environment/rohs_china.)
Calibration
You can obtain the calibration certificate and information about calibration services for the
NI 9263 at ni.com/calibration.
Calibration interval 1 year
373781B-02 Mar16