Datasheet: Kit Contents - NI 9201 - NI 9201 Getting Started Guide

Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

DATASHEET

NI 9201
8 AI, ±10 V, 12 Bit, 500 kS/s Aggregate

• DSUB, screw-terminal, or spring-terminal connectivity


• 250 Vrms, CAT II, channel-to-earth isolation (screw and
spring terminal); 60 VDC, CAT I, channel-to-earth isolation
(DSUB)
• -40 °C to 70 °C operating range, 5 g vibration, 50 g shock

The NI 9201 is an analog input module for CompactDAQ and CompactRIO systems. The
NI 9201 provides eight channels of ±10 V input with 500 kS/s sample rate.

• NI 9201
Kit Contents
• NI 9201 Getting Started Guide

• NI 9927 Backshell Connector Kit (Screw Terminal)


Accessories • NI 9981 Backshell Connector Kit (Spring Terminal)
• NI 9924 Screw-Terminal Block (DSUB)
C SERIES ANALOG INPUT MODULE COMPARISON

Product Signal Sample


Channels Simultaneous Resolution Connectivity
Name Levels Rate
Screw-Terminal,
NI 9201 ±10 V 8 Single-Ended 500 kS/s No 12-Bit Spring-Terminal,
DSUB
±200 mV, ±1 V, 32 Single-Ended, Spring-Terminal,
NI 9205 250 kS/s No 16-Bit
±5 V, ±10 V 16 differential DSUB

±200 mV, ±1 V, 32 Single-Ended,


NI 9206 250 kS/s No 16-Bit Spring-Terminal
±5 V, ±10 V 16 Differential

NI 9207 ±10 V 8 Differential 500 S/s No 24-Bit DSUB

32 Single-Ended,
NI 9209 ±10 V 500 S/s No 24-Bit DSUB
16 Differential

Screw-Terminal,
NI 9215 ±10 V 4 Differential 100 kS/s/ch Yes 16-Bit Spring-Terminal,
BNC
Spring-Terminal,
NI 9220 ±10 V 16 Differential 100 kS/s/ch Yes 16-Bit
DSUB

Screw-Terminal,
NI 9221 ±60 V 8 Single-Ended 800 kS/s No 12-Bit Spring-Terminal,
DSUB
Screw-Terminal,
NI 9222 ±10 V 4 Differential 500 kS/s/ch Yes 16-Bit
BNC

Screw-Terminal,
NI 9223 ±10 V 4 Differential 1 MS/s/ch Yes 16-Bit
BNC

NI C Series Overview

NI provides more than 100 C Series modules for measurement, control, and communication
applications. C Series modules can connect to any sensor or bus and allow for high-accuracy
measurements that meet the demands of advanced data acquisition and control applications.
• Measurement-specific signal conditioning that connects to an array of sensors and signals
• Isolation options such as bank-to-bank, channel-to-channel, and channel-to-earth ground
• -40 °C to 70 °C temperature range to meet a variety of application and environmental
needs
• Hot-swappable

2 | ni.com | NI 9201 Datasheet


The majority of C Series modules are supported in both CompactRIO and CompactDAQ
platforms and you can move modules from one platform to the other with no modification.

CompactRIO
CompactRIO combines an open-embedded architecture
with small size, extreme ruggedness, and C Series
modules in a platform powered by the NI LabVIEW
reconfigurable I/O (RIO) architecture. Each system
contains an FPGA for custom timing, triggering, and
processing with a wide array of available modular I/O to
meet any embedded application requirement.

CompactDAQ
CompactDAQ is a portable, rugged data acquisition platform
that integrates connectivity, data acquisition, and signal
conditioning into modular I/O for directly interfacing to any
sensor or signal. Using CompactDAQ with LabVIEW, you
can easily customize how you acquire, analyze, visualize,
and manage your measurement data.

Software
LabVIEW Professional Development System for Windows

• Use advanced software tools for large project development


• Generate code automatically using DAQ Assistant and Instrument
I/O Assistant
• Use advanced measurement analysis and digital signal processing
• Take advantage of open connectivity with DLLs, ActiveX,
and .NET objects
• Build DLLs, executables, and MSI installers

NI 9201 Datasheet | © National Instruments | 3


NI LabVIEW FPGA Module

• Design FPGA applications for NI RIO hardware


• Program with the same graphical environment used for desktop and
real-time applications
• Execute control algorithms with loop rates up to 300 MHz
• Implement custom timing and triggering logic, digital protocols, and
DSP algorithms
• Incorporate existing HDL code and third-party IP including Xilinx IP
generator functions
• Purchase as part of the LabVIEW Embedded Control and Monitoring
Suite

NI LabVIEW Real-Time Module

• Design deterministic real-time applications with LabVIEW


graphical programming
• Download to dedicated NI or third-party hardware for reliable
execution and a wide selection of I/O
• Take advantage of built-in PID control, signal processing, and
analysis functions
• Automatically take advantage of multicore CPUs or set
processor affinity manually
• Take advantage of real-time OS, development and debugging
support, and board support
• Purchase individually or as part of a LabVIEW suite

NI 9201 Input Circuitry

AI0
Overvoltage Isolated
Mux Buffer
AI7 Protection ADC

COM

NI 9201

• Input signals are scanned, buffered, conditioned, and then sampled by a single ADC.
• Each AI channel provides an independent signal path and ADC, enabling you to sample
all channels simultaneously.

4 | ni.com | NI 9201 Datasheet


NI 9201 Specifications
The following specifications are typical for the range -40 °C to 70 °C unless otherwise noted.
All voltages are relative to COM unless otherwise noted.

Caution Do not operate the NI 9201 in a manner not specified in this document.
Product misuse can result in a hazard. You can compromise the safety protection
built into the product if the product is damaged in any way. If the product is
damaged, return it to NI for repair.

Input Characteristics
Number of channels 8
ADC resolution 12 bits
Type of ADC Successive approximation register (SAR)

Table 1. Sample Rate (Aggregate)


Maximum Sample Rate Maximum Sample Rate
Mode (R Series Expansion Chassis) (All Other Chassis)

Single Channel 475 kS/s 800 kS/s

Scanning 475 kS/s 500 kS/s

Input range ±10 V


Measurement voltage, channel-to-COM (V)
Minimum ±10.3
Typical ±10.53
Maximum ±10.8
Overvoltage protection, channel-to-COM ±100 V

Table 2. NI 9201 Accuracy (Excludes Noise)


Percent of Reading Percent of Range1
Measurement Conditions (Gain Error) (Offset Error)

Typical (25 °C, ±5 °C) ±0.04% ±0.07%


Calibrated
Maximum (-40 °C to 70 °C) ±0.25% ±0.25%

1 Range equals 10.53 V

NI 9201 Datasheet | © National Instruments | 5


Table 2. NI 9201 Accuracy (Excludes Noise) (Continued)
Percent of Reading Percent of Range1
Measurement Conditions (Gain Error) (Offset Error)

Typical (25 °C, ±5 °C) ±0.26% ±0.46%


Uncalibrated2
Maximum (-40 °C to 70 °C) ±0.67% ±1.25%

Stability
Gain drift ±34 ppm/°C
Offset drift ±100 μV/°C
Input bandwidth (-3 dB) 690 kHz min
Input impedance
Resistance 1 MΩ
Capacitance 5 pF
Input noise, code-centered
RMS 0.7 LSBrms
Peak-to-peak 5 LSB
No missing codes 12 bits
DNL -0.9 to 1.5 LSB
INL ±1.5 LSB
Crosstalk, at 10 kHz -75 dB
Settling time, to 1 LSB 2 μs
MTBF 1,092,512 hours at 25 °C; Bellcore Issue 2,
Method 1, Case 3, Limited Part Stress Method

Power Requirements
Power consumption from chassis
Active mode 1 W maximum
Sleep mode 1 mW maximum
Thermal dissipation (at 70 °C)
Active mode 1 W maximum
Sleep mode 32 mW maximum

1 Range equals 10.53 V


2 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes
where the calibration constants stored in the module are not applied to the data.

6 | ni.com | NI 9201 Datasheet


Physical Characteristics
If you need to clean the module, wipe it with a dry towel.

Tip For two-dimensional drawings and three-dimensional models of the C Series


module and connectors, visit ni.com/dimensions and search by module number.

Screw-terminal wiring
Gauge 0.2 mm2 to 2.5 mm2 (26 AWG to 14 AWG)
copper conductor wire
Wire strip length 13 mm (0.51 in.) of insulation stripped from
the end
Temperature rating 90 °C minimum
Torque for screw terminals 0.5 N · m to 0.6 N · m
(4.4 lb · in. to 5.3 lb · in.)
Wires per screw terminal One wire per screw terminal; two wires per
screw terminal using a 2-wire ferrule
Ferrules 0.25 mm2 to 2.5 mm2
Spring-terminal wiring
Gauge 0.2 mm2 to 2.5 mm2 (30 AWG to 12 AWG)
copper conductor wire
Wire strip length 10 mm (0.39 in.) of insulation stripped from
the end
Temperature rating 90 °C minimum
Wires per spring terminal One wire per spring terminal; two wires per
spring terminal using a 2-wire ferrule
Ferrules 0.25 mm2 to 2.5 mm2
Connector securement
Securement type Screw flanges provided
Torque for screw flanges 0.2 N · m (1.80 lb · in.)
Weight
NI 9201 with screw terminal 165 g (5.8 oz)
NI 9201 with spring terminal 152 g (5.4 oz)
NI 9201 with DSUB 142 g (5.0 oz)

NI 9201 Datasheet | © National Instruments | 7


NI 9201 with Screw Terminal and NI 9201 with Spring
Terminal Safety Voltages
Connect only voltages that are within the following limits.

Channel-to-COM ±60 VDC maximum


Channel-to-channel None
Channel-to-earth ground
Continuous 250 Vrms, Measurement Category II
Withstand 2,300 Vrms, verified by a 5 s dielectric
withstand test

Measurement Category II is for measurements performed on circuits directly connected to the


electrical distribution system. This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.

Caution Do not connect the NI 9201 with screw terminal or NI 9201 with spring
terminal to signals or use for measurements within Measurement Categories III or
IV.

NI 9201 with DSUB Safety Voltages


Connect only voltages that are within the following limits.

Channel-to-COM ±60 VDC maximum


Channel-to-channel None
Channel-to-earth
Continuous 60 VDC, Measurement Category I
Withstand 1,000 Vrms, verified by a 5 s dielectric
withstand test

Measurement Category I is for measurements performed on circuits not directly connected to


the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live
electrical supply system that powers equipment. This category is for measurements of voltages
from specially protected secondary circuits. Such voltage measurements include signal levels,
special equipment, limited-energy parts of equipment, circuits powered by regulated low-
voltage sources, and electronics.

Caution Do not connect the NI 9201 with DSUB to signals or use for
measurements within Measurement Categories III or IV.

Note Measurement Categories CAT I and CAT O are equivalent. These test and
measurement circuits are not intended for direct connection to the MAINS building
installations of Measurement Categories CAT II, CAT III, or CAT IV.

8 | ni.com | NI 9201 Datasheet


Hazardous Locations
U.S. (UL) Class I, Division 2, Groups A, B, C, D, T4;
Class I, Zone 2, AEx nA IIC T4
Canada (C-UL) Class I, Division 2, Groups A, B, C, D, T4;
Class I, Zone 2, Ex nA IIC T4
Europe (ATEX) and International (IECEx) Ex nA IIC T4 Gc

Safety and Hazardous Locations Standards


This product is designed to meet the requirements of the following electrical equipment safety
standards for measurement, control, and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
• EN 60079-0:2012, EN 60079-15:2010
• IEC 60079-0: Ed 6, IEC 60079-15; Ed 4
• UL 60079-0; Ed 5, UL 60079-15; Ed 3
• CSA 60079-0:2011, CSA 60079-15:2012

Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.

Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for sensitive electrical
equipment for measurement, control, and laboratory use:
• EN 61326 (IEC 61326): Class A emissions; Industrial immunity
• EN 55011 (CISPR 11): Group 1, Class A emissions
• AS/NZS CISPR 11: Group 1, Class A emissions
• FCC 47 CFR Part 15B: Class A emissions
• ICES-001: Class A emissions

Note For EMC declarations and certifications, and additional information, refer to
the Online Product Certification section.

Note For EMC compliance, operate this device with double-shielded cables.

NI 9201 Datasheet | © National Instruments | 9


CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
• 2014/35/EU; Low-Voltage Directive (safety)
• 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
• 94/9/EC; Potentially Explosive Atmospheres (ATEX)

Online Product Certification


Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance
information. To obtain product certifications and the DoC for this product, visit ni.com/
certification, search by model number or product line, and click the appropriate link in the
Certification column.

Shock and Vibration


To meet these specifications, you must panel mount the system.

Operating vibration
Random (IEC 60068-2-64) 5 grms, 10 Hz to 500 Hz
Sinusoidal (IEC 60068-2-6) 5 g, 10 Hz to 500 Hz
Operating shock (IEC 60068-2-27) 30 g, 11 ms half sine; 50 g, 3 ms half sine;
18 shocks at 6 orientations

Environmental
Refer to the manual for the chassis you are using for more information about meeting these
specifications.

Operating temperature -40 °C to 70 °C


(IEC 60068-2-1, IEC 60068-2-2)
Storage temperature -40 °C to 85 °C
(IEC 60068-2-1, IEC 60068-2-2)
Ingress protection IP40
Operating humidity (IEC 60068-2-78) 10% RH to 90% RH, noncondensing
Storage humidity (IEC 60068-2-78) 5% RH to 95% RH, noncondensing
Pollution Degree 2
Maximum altitude 2,000 m

Indoor use only.

10 | ni.com | NI 9201 Datasheet


Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible
manner. NI recognizes that eliminating certain hazardous substances from our products is
beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact
web page at ni.com/environment. This page contains the environmental regulations and
directives with which NI complies, as well as other environmental information not included in
this document.

Waste Electrical and Electronic Equipment (WEEE)


EU Customers At the end of the product life cycle, all NI products must be
disposed of according to local laws and regulations. For more information about
how to recycle NI products in your region, visit ni.com/environment/weee.

电子信息产品污染控制管理办法(中国 RoHS)
中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物
质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录
ni.com/environment/rohs_china。(For information about China RoHS
compliance, go to ni.com/environment/rohs_china.)

Calibration
You can obtain the calibration certificate and information about calibration services for the
NI 9201 at ni.com/calibration.
Calibration interval 1 year

NI 9201 Datasheet | © National Instruments | 11


Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and
company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI
products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the
National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs)
and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/
legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other
import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION
CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in
this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth
in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2016 National Instruments. All rights reserved.

373783A-02 Mar16