Opt. Liquicolor: Design Verification
Opt. Liquicolor: Design Verification
Opt. Liquicolor: Design Verification
opt. liquicolor
1 Introduction .................................................................................................................................................................. 2
2 Imprecision .................................................................................................................................................................... 2
3 Linearity and Sensitivity............................................................................................................................................... 2
3.1 Linearity ....................................................................................................................................................................................... 2
3.1 Sensitivity.................................................................................................................................................................................... 3
4 Comparison of Methods ............................................................................................................................................... 3
5 Stability.......................................................................................................................................................................... 4
5.1 Stability Working Reagent .................................................................................................................................................... 4
Recovery of Control Sera ......................................................................................................................................................................... 4
Linearity......................................................................................................................................................................................................... 5
5.2 Real Time Stability ................................................................................................................................................................... 5
6 Interference ................................................................................................................................................................... 8
7 Traceability .................................................................................................................................................................... 9
2 Imprecision
The imprecision (within-run and day-to-day) of ALKALINE PHOSPHATASE opt. liquicolor was calculated from six
determinations on ten consecutive days. Commercial control sera with low, medium and high alkaline
Phosphatase levels were employed as sample material.
Hitachi 717
High Pool Analytical Data Regressed Data Deviation from Regression Line
Content (%) (U/l) (U/l) (U/l) (%)
0 -1.05 -11.6 10.5 -90.9
10 201 194 7.23 3.72
20 391 400 -9.12 -2.28
30 606 606 0.74 0.12
40 799 812 -12.1 -1.50
50 998 1017 -18.9 -1.86
60 1226 1223 3.11 0.25
70 1432 1429 3.21 0.22
80 1646 1635 11.3 0.69
90 1861 1840 20.3 1.10
100 2030 2046 -16.2 -0.79
Design Verification and Product Data for Alkaline Phosphatase opt. liquicolor 2/9
Rev. 007
Conclusion: The ALKALINE PHOSPHATASE opt. liquicolor is linear up to about 2000 U/l..
3.1 Sensitivity
A 20-fold determination of a 0 calibrator (phys. saline) on a Hitachi 717 analyser revealed an absolute mean of
0.87 U/l and a SD of 1.16 U/l. Sensitivity was calculated on base of mean + 3SD = 3.48 U/l.
4 Comparison of Methods
ALKALINE PHOSPHATASE opt. liquicolor has been compared with a commercially available alkaline Phosphatase
method. Control sera as well as patient samples have been employed in the comparison (N=54).
The results have been evaluated by a main component analysis. The linear regression obtained could be described
as follows:
ALKALINE PHOSPHATASE opt. liquicolor (Y) = 0.995 * Ref. method (X) - 3.506
The coefficient of correlation was determined to r = 0.996.
Evaluation of the method comparison by a non-parametric model according to Bablok&Passing yielded the
following results:
Reference (REF)
Lowest value = 106.0000
Highest value = 331.5000
Arithmetic mean = 188.1111
Median = 178.9500
Design Verification and Product Data for Alkaline Phosphatase opt. liquicolor 3/9
Rev. 007
Standard deviation = 57.3444
Standard error of the mean = 7.8036
HUMAN AP_DGKCH
Lowest value = 104.8000
Highest value = 326.7000
Arithmetic mean = 183.6981
Median = 174.7500
Standard deviation = 57.3240
Standard error of the mean = 7.8008
Y = -1,8810 + 0,9881 X
Intercept A :-1.8810;95% CI : -6.6865 to 3.2489
Slope B : 0.9881;95% CI : 0.9572 to 1.0148
300
AP 250
_D
G
K
C
H,
200
U/l
150
100
100 150 200 250 300 350
REF, U/l
Conclusion: The comparison study confirmed an excellent agreement between both methods.
5 Stability
5.1 Stability Working Reagent
Design Verification and Product Data for Alkaline Phosphatase opt. liquicolor 4/9
Rev. 007
Control serum recovery Working Reagent fresh Working Reagent Working Reagent
5 days 25°C 7 days 25°C
Control LOT Result Result Deviation Result Deviation
Serum U/l U/l (%) U/l (%)
HumaTrol N # 018 203 210 -3.45 197 -3.06
HumaTrol P # 016 513 516 -0.58 498 -3.03
SERODOS # 6868 180 184 -2.23 181 0.55
SERODOS plus # 6794 418 386 7.66 391 -6.5
SERODOS plus # 6795 411 423 2.92 430 4.63
Mean 345 343.8 3.37 339.4 1.48
Linearity
The linearity of the ALKALINE PHOSPHATASE opt. liquicolor was controlled according to the procedure already
described in 2.1. The working reagents were stressed 5 and 7 days at 25°C.
Conclusion: The working reagent is stable for 4 weeks at 4°C and for 5 days at 15…25°C.
Used Material
Reagent Manufacturer REF LOT
ALKALINE PHOSPHATASE opt. liquicolor HUMAN 12037 16001; 17003/reference
ALKALINE PHOSPHATASE opt. liquicolor HUMAN 12037 15002 /25 months
ALKALINE PHOSPHATASE opt. liquicolor HUMAN 12037 15004 /25 months
ALKALINE PHOSPHATASE opt. liquicolor HUMAN 12017 16002/25 months
AUTOCAL HUMAN 13160 H015; H016
Sample HUMAN’s and commercial controls (Beckman Coulter)
Design Verification and Product Data for Alkaline Phosphatase opt. liquicolor 5/9
Rev. 007
Results
Control recovery ALKALINE PHOSPHATASE opt. liquicolor,
REF 12037
Reference Test
LOT 16001 LOT 15002
Fresh 22.5 months
Name LOT Target Range Result Result Within range
U/l U/l U/l U/l Y/N
HumaTrol N 0004 202 152 – 253 195 198 Yes
HumaTrol P 0003 952 714 – 1190 898 874 Yes
SERODOS 0003 333 250 – 416 317 314 Yes
SERODOS
plus 0003 595 446 – 744 630 619 Yes
Control serum L1 0037 171 128 – 214 186 202 Yes
Control serum L1 0038 689 517 - 861 673 733 Yes
Dev.%
Mean 490 483 490 0.1
Control recovery for ALKALINE PHOSPHATASE opt. liquicolor Multipurpose reagent are indicative for a real-time
stability of 18 months.
Design Verification and Product Data for Alkaline Phosphatase opt. liquicolor 6/9
Rev. 007
5.2.2 Linearity
The linearity of ALKALINE PHOSPHATASE opt. liquicolor Multipurpose reagent was controlled according to the
procedure already described in section 3 using high concentrated pool serum on an AU 480. The linearity was
controlled for 3 different reagent LOTs.
Criteria
Linearity Acceptance
criteria
Deviation from regression line ≤ 10%
Used Material
Reagent Manufacturer REF LOT
ALKALINE PHOSPHATASE opt. liquicolor HUMAN 12037 15002/20 months
ALKALINE PHOSPHATASE opt. liquicolor HUMAN 12017 15004/25 months
ALKALINE PHOSPHATASE opt. liquicolor HUMAN 12017 16002/25 months
Sample Diluted pool sera, LOT CEN102A; CEN131; BOG190
Results
ALKALINE PHOSPHATASE opt. liquicolor, REF 12037, LOT 15002/20 months
High Pool Content Analytical Data Calculated Data Deviation from Regression Line
(%) (U/l) (U/l) (U/l) (%)
0 -0.31 0.00 -0.31
10 231 250 -19 -7.5
20 544 500 44 8.8
30 714 750 -36 -4.8
40 985 1000 -15 -1.5
50 1260 1250 10 0.8
60 1511 1500 11 0.8
70 1789 1750 39 2.2
80 2067 2000 67 3.3
90 2246 2250 -4 -0.2
100 2495 2500 -5 -0.2
Design Verification and Product Data for Alkaline Phosphatase opt. liquicolor 7/9
Rev. 007
20 684 680 4 0.5
30 1034 1020 13 1.3
40 1361 1360 0 0.0
50 1684 1701 -16 -1.0
60 1992 2041 -49 -2.4
70 2312 2381 -69 -2.9
80 2618 2721 -103 -3.8
90 2935 3061 -126 -4.1
100 3272 3401 -129 -3.8
ALKALINE PHOSPHATASE opt. liquicolor Multipurpose reagent is specified to cover a linear range on an AU 480 as
indicated in section 3: up to 2000 U/l
Conclusion
All real time stability data for linearity confirm the shelf life
of 18 months for ALKALINE PHOSPHATASE opt. liquicolor Multipurpose reagent.
6 Interference
Interference by ascorbate, bilirubin, hemoglobin, triglycerides and lipemic sera have been studied by adding known
amounts of the potentially interfering substance to a known sample.
Recoveries have been analyzed according to the method of Glick et al. (Clin.Cem. 1986, 32 470-5).
Glick 2 1 2 2 1
Glick 1 2 2 2 1
Design Verification and Product Data for Alkaline Phosphatase opt. liquicolor 8/9
Rev. 007
Conclusion: This test is not significantly influenced by ascorbate up to 20 mg/dl, bilirubin up to 40 mg/dl,
hemoglobin up to 500 mg/dl, triglycerides up to 2500 mg/dl and lipemic sera up to 1000 mg/dl.
7 Traceability
The ALKALINE PHOSPHATASE opt. liquicolor concentration has been verified with the reference material SRM 909 B.
The ALKALINE PHOSPHATASE opt. liquicolor revealed recovery well within the reproducibility limits of the method.
The method is traceable to SRM 909 B.
Design Verification and Product Data for Alkaline Phosphatase opt. liquicolor 9/9
Rev. 007