Epa Ilm05.3
Epa Ilm05.3
D-1/Introduction ILM05.3
THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.3 D-2/Introduction
Exhibit D - Analytical Methods
Table of Contents
Section Page
1.0 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Inorganic Methods Flow Chart . . . . . . . . . . . . . . . . . 5
1.2 Figure 1 - Inorganic Methods Flow Chart . . . . . . . . . . . . 6
1.3 Glassware Cleaning . . . . . . . . . . . . . . . . . . . . . . 6
1.4 Standard Stock Solutions . . . . . . . . . . . . . . . . . . . 6
1.5 Verification of Aqueous Sample Preservation . . . . . . . . . . 6
1.6 Percent Solids Determination Procedure . . . . . . . . . . . . 7
1.7 Insufficient Sample Volume . . . . . . . . . . . . . . . . . . 8
1.8 Sample Mixing . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.9 Undiluted Analysis . . . . . . . . . . . . . . . . . . . . . . 8
1.10 Dissolved Metals . . . . . . . . . . . . . . . . . . . . . . . 9
1.11 Replicate Exposure . . . . . . . . . . . . . . . . . . . . . . 9
1.12 Raw Data Requirements . . . . . . . . . . . . . . . . . . . . . 9
1.13 Quality Control Samples . . . . . . . . . . . . . . . . . . . . 9
1.14 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.15 Pollution Prevention . . . . . . . . . . . . . . . . . . . . . 10
1.16 Waste Management . . . . . . . . . . . . . . . . . . . . . . . 10
D-3/Introduction ILM05.3
THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.3 D-4/Introduction
Exhibit D -– Section 1
Introduction
1.0 INTRODUCTION
The analytical methods that follow are designed to analyze water and
sediment/soil from hazardous waste sites for the presence of inorganic
analytes contained on the Inorganic Target Analyte List (TAL) (see
Exhibit C). The inorganic methods include alternative analysis
procedures for some analytes, multiple preparation procedures, and
Quality Control (QC) requirements. Analytical techniques in the
inorganic methodologies include Inductively Coupled Plasma - Atomic
Emission Spectroscopy (ICP-AES), Inductively Coupled Plasma - Mass
Spectrometry (ICP-MS), Cold Vapor Atomic Absorption Spectroscopy, and
Spectrophotometry. Graphite Furnace Atomic Absorption (GFAA) may be
requested by the modified analysis clause in the contract.
D-5/Introduction ILM05.3
Exhibit D -– Section 1
Introduction (Con’t)
Lab glassware to be used within the metals analysis must be acid cleaned
according to USEPA's manual, Methods for Chemical Analysis of Water and
Wastes or an equivalent procedure. An electronic version can be found
via USEPA's National Environmental Publications Internet Site (NEPIS) at
http://www.epa.gov/cincl.
1.5.1 At the time of sample receipt, the Contractor shall check the pH of
the sample and note in a preparation log if the pH is less than 2 for
metals. In addition, it should be noted if the pH is greater than 12
for a cyanide sample. Unless instructed by the USEPA Regional CLP
Project Officer (CLP PO), the Contractor shall not perform any pH
adjustment action if the sample has not been properly preserved. If
the sample has not been properly preserved, contact Sample Management
ILM05.3 D-6/Introduction
Exhibit D -– Section 1
Introduction (Con’t)
1.6.2 Place weighing dish plus sample, with the cover tipped to allow for
moisture escape, in a drying oven maintained at 103-105°C. Sample
handling and drying should be conducted in a well-ventilated area.
1.6.3 Dry the sample overnight (12-24 hours) but no longer than 24 hours.
If dried less than 12 hours, it must be documented that constant
weight was attained.1 Remove the sample from the oven and cool in a
desiccator with the weighing dish cover in place before weighing.
Weigh and record weight to nearest 0.01 g. Do not analyze the dried
sample.
1
Drying time is defined as the elapsed time in the oven; thus raw data
must record time in and out of the oven to document the 12-hour drying time
minimum. In the event it is necessary to demonstrate the attainment of
constant weight, data must be recorded for a minimum of two repetitive
weigh/dry/desiccate/weigh cycles with a minimum of 1-hour drying time in each
cycle. Constant weight would be defined as a loss in weight of no greater
than 0.01 g between the start weight and final weight of the last cycle.
D-7/Introduction ILM05.3
Exhibit D -– Section 1
Introduction (Con’t)
1.6.5 For the duplicate percent solids determination, designate one sample
aliquot as the “original” sample and the other aliquot as the
“duplicate” sample. Calculate dry weight using the results of the
“original” sample aliquot.
1.6.6 Calculate percent solids by the formula below. The value thus
obtained will be reported on the appropriate Forms I and, where
applicable, Forms VA-IN and VI-IN. This value will be used for
calculating analytical concentration on a dry weight basis.
1.6.7 If the sample contains less than 50% solids, the Contractor shall
notify SMO immediately of the samples impacted. After notification
to SMO, the Contractor shall proceed with sample analysis and
document the issue in the SDG Narrative.
Unless instructed otherwise by the USEPA Regional CLP PO, all samples
shall be mixed thoroughly prior to aliquoting for digestion. There is
no specific procedure provided herein for homogenization of soil/
sediment samples; however, an effort should be made to obtain a
representative aliquot.
1.9.1 All samples shall be run undiluted for multi-analyte analysis (i.e.,
the final product of the sample preparation procedure) unless the
dilution adjusted detection limits for all analytes are below the
CRQL. When an analyte concentration exceeds the calibrated or linear
range, appropriate dilution (but not below the CRQL) and re-analysis
is required. The Contractor shall use the least dilution necessary
to bring the analyte(s) instrument reading within the upper 75% of
the calibrated or linear range and report the highest valid value for
each analyte as measured from the undiluted and diluted analyses.
Unless the Contractor can submit proof that dilution was required to
obtain valid results, both diluted and undiluted sample measurements
must be contained in the raw data.
1.9.2 For single analyte analysis, a diluted sample analysis may be the
only sample analysis performed if the analyte’s instrument result is
in the upper 75% of the calibrated or linear range. An undiluted
sample analysis does not have to be performed in this case. The
sample and its associated matrix spike and duplicate shall initially
be run at the same dilution.
ILM05.3 D-8/Introduction
Exhibit D -– Section 1
Introduction (Con’t)
1.9.3 All sample dilutions shall be made with reagent water appropriately
acidified (except for cyanide) to maintain constant acid strength.
If the Sampler designated two (or more) samples as QC for the same
matrix, and the QC samples are not specifically labeled with the
analysis they are to be used for (dissolved metals and total metals),
then the Contractor is to contact SMO to report the issue. SMO shall
then contact the Region and notify the Contractor of the Regional
decision.
1.14 Safety
The toxicity or carcinogenicity of each reagent used in this SOW has not
been precisely defined; however, each chemical compound should be
treated as a potential health hazard. From this viewpoint, exposure to
these chemicals must be reduced to the lowest possible level by whatever
means available. The laboratory is responsible for maintaining a
current awareness file of Occupational Safety and Health Administration
(OSHA) regulations regarding the safe handling of chemicals specified in
this method. A reference file of material handling data sheets should
also be made available to all personnel involved in the chemical
analysis.
D-9/Introduction ILM05.3
Exhibit D -– Section 1
Introduction (Con’t)
ILM05.3 D-10/Introduction
EXHIBIT D - PART A
ANALYTICAL METHODS
FOR
D-1/ICP-AES ILM05.3
THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.3 D-2/ICP-AES
Exhibit D - Analytical Methods for ICP-AES
Table of Contents
Section Page
3.0 DEFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4.0 INTERFERENCES . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
5.0 SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
6.1 Glassware/Labware . . . . . . . . . . . . . . . . . . . . . . . 7
(ICP-AES) . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
7.1 Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
7.2 Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
10.0 PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
(CRI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
D-3/ICP-AES ILM05.3
Exhibit D - Analytical Methods for ICP-AES
Section Page
16.0 REFERENCES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
17.0 TABLES/DIAGRAMS/FLOWCHARTS . . . . . . . . . . . . . . . . . . . . . 33
ILM05.3 D-4/ICP-AES
Exhibit D (ICP-AES) -- Sections 1-3
Scope and Application
Water and soil samples are treated with acids and heat or microwave
energy to solubilize the metals present. These digestates are then
analyzed for trace metals by an atomic emission optical spectroscopic
technique. Samples are nebulized and the aerosol that is produced is
transported to a plasma torch where excitation occurs. Characteristic
atomic-line emission spectra are produced by a radio-frequency
inductively coupled plasma. The spectra are dispersed and the
intensities of the lines are monitored by a photosensitive device. The
signals from the photosensitive device are processed by a computer. A
background correction technique is required to compensate for variable
background contribution to the spectra of trace elements. Background
must be measured adjacent to analyte lines on samples during analysis.
The position selected for the background intensity measurement, on
either or both sides of the analytical line, will be determined by the
complexity of the spectrum adjacent to the analyte line. The position
used must be free of spectral interference and reflect the same change
in background intensity as occurs at the analyte wavelength measured.
Background correction is not required in cases of line broadening where
a background correction measurement would actually degrade the
analytical result.
3.0 DEFINITIONS
D-5/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Sections 4 & 5
Interferences
4.0 INTERFERENCES
Another problem which can occur from high dissolved solids is salt
buildup at the tip of the nebulizer. This affects aerosol flow rate
causing instrumental drift. Wetting the argon prior to nebulization,
the use of a tip washer, or sample dilution has been used to control
this problem. Also, it has been reported that better control of the
argon flow rate improves instrument performance. This is accomplished
with the use of mass flow controllers.
5.0 SAFETY
ILM05.3 D-6/ICP-AES
Exhibit D (ICP-AES) -- Section 6
Equipment and Supplies
Brand names, suppliers, and part numbers are for illustrative purposes
only. No endorsement is implied. Equivalent performance may be
achieved using equipment and supplies other than those specified here,
however, a demonstration of equivalent performance meeting the
requirements of this Statement of Work (SOW) is the responsibility of
the Contractor. The Contractor shall document any use of alternate
equipment or supplies in the Sample Delivery Group (SDG) Narrative.
6.1 Glassware/Labware
6.1.3 Funnels
6.1.9.5 The system must use PTFE PFA digestion vessels (120 mL capacity)
capable of withstanding pressure of up to 110 (±10) pounds per
square inch (psi) [7.5 (±0.7 atm)]. These vessels are capable of
controlled pressure relief at pressures exceeding 110 psi.
6.1.10 Balances - Analytical Balance, 300 gram (g) capacity, and minimum
±0.01 g.
D-7/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 7
Reagents and Standards
7.1 Reagents
7.1.1 Reagent water - The purity of this water must be equivalent to ASTM
Type II water (ASTM D1193-77). Use this preparation for all
reagents, standards, and dilutions of solutions.
7.1.4 Hydrochloric acid, (1+1) - Add 500 milliliters (mL) conc. HCl
(specific gravity 1.19) to 400 mL reagent water and dilute to 1 Liter
(L).
7.1.6 Nitric acid, (1+1) - Add 500 mL conc. HNO3 (specific gravity 1.41) to
400 mL reagent water and dilute to 1 L.
7.1.8 Nitric acid, 5% (v/v) - Add 50 mL conc. HNO3 to 500 mL reagent water;
dilute to 1 L.
7.2 Standards
7.2.1 Introduction
(CAUTION: Many metal salts are extremely toxic and may be fatal if
swallowed. Wash hands thoroughly after handling) Typical stock
solution preparation procedures follow.
ILM05.3 D-8/ICP-AES
Exhibit D (ICP-AES) -- Section 7
Reagents and Standards (Con’t)
D-9/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 7
Reagents and Standards (Con’t)
ILM05.3 D-10/ICP-AES
Exhibit D (ICP-AES) -- Section 7
Reagents and Standards (Con’t)
D-11/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 8
The samples must be protected from light and refrigerated at 4°C (±2°C)
from the time of receipt until 60 days after delivery of a complete,
reconciled data package to USEPA. After 60 days the samples may be
disposed of in a manner that complies with all applicable regulations.
The maximum holding time for metals is 180 days from Validated Time of
Sample Receipt (VTSR).
ILM05.3 D-12/ICP-AES
Exhibit D (ICP-AES) -- Section 9
Calibration and Standardization
9.2.1 The calibration procedure is a critical step prior to the use of any
microwave unit. The microwave unit must be calibrated every six
months. The data for each calibration must be available for review
during on-site audits. In order that absolute power settings may be
interchanged from one microwave unit to another, the actual delivered
power must be determined.
9.2.3 The power available for heating is evaluated so that the absolute
power setting (watts) may be compared from one microwave to another.
This is accomplished by measuring the temperature rise in 1 kilogram
(kg) of water exposed to microwave radiation for a fixed period of
time. The water is placed in a PTFE beaker (or a beaker that is made
of some other material that does not absorb microwave energy) and
stirred before measuring the temperature. Glass beakers absorb
microwave energy and may not be used. The initial temperature of the
water must be between 19 and 25°C. The beaker is circulated
continuously through the field for at least two minutes at full
power. The beaker is removed from the microwave, the water is
stirred vigorously, and the final temperature is recorded. The final
reading is the maximum temperature reading after each energy
exposure. These measurements must be accurate to ±0.1°C and made
within 30 seconds of the end of heating. If more measurements are
needed, do not use the same water until it has cooled down to room
temperature. Otherwise, use a fresh water sample.
D-13/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 9
Calibration and Standardization (Con’t)
The microwave user can now relate power in watts to the percent power
setting of the microwave.
9.3.1 Instruments shall be calibrated daily or once every 24 hours and each
time the instrument is set up. The instrument standardization date
and time shall be included in the raw data.
9.4.1 Immediately after each of the ICP-AES systems have been calibrated,
the accuracy of the initial calibration shall be verified and
documented for every analyte by the analysis of the ICV solution(s)
at each wavelength used.
9.4.2 Only if the ICV solution(s) is(are) not available from USEPA, or
where a certified solution of an analyte is not available from any
source, analyses shall be conducted on an independent standard at a
ILM05.3 D-14/ICP-AES
Exhibit D (ICP-AES) -- Section 9
Calibration and Standardization (Con’t)
9.4.3 The ICV solution(s) shall be run at each wavelength used for
analysis. The values for the ICV shall be reported on Form IIA-IN.
9.5.1 To ensure calibration accuracy during each analysis run, one of the
following standards is to be used for the CCV and shall be analyzed
and reported for every wavelength used for the analysis of each
analyte, at a frequency of 10% or every 2 hours during an analysis
run, whichever is more frequent. The standard shall also be analyzed
and reported for every wavelength used for analysis at the beginning
of the run and after the last analytical sample. The analyte
concentrations in the CCV standard shall be different than the
concentration used for the ICV and shall be one of the following
solutions at or near one-half of the calibration standard:
C USEPA Solutions
C NIST Standards
The same CCV standard shall be used throughout the analysis runs for
a Sample Delivery Group (SDG) of samples received.
9.5.2 Each CCV analyzed shall reflect the conditions of analysis of all
associated analytical samples (the preceding 10 analytical samples or
the preceding analytical samples up to the previous CCV). The
duration of analysis, rinses, and other related operations that may
affect the CCV measured result may not be applied to the CCV to a
greater extent than the extent applied to the associated analytical
samples. For instance, the difference in time between a CCV analysis
and the blank immediately following it, as well as the difference in
time between the CCV and the analytical sample immediately preceding
it, may not exceed the lowest difference in time between any two
consecutive analytical samples associated with the CCV.
NOTE: A CCB shall be analyzed immediately after the last CCV, and the
last CCV shall be analyzed immediately after the last analytical sample
of the run. The results for the calibration blanks shall be reported on
Form III-IN.
D-15/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 10
Procedure
10.0 PROCEDURE
10.1.1 If insufficient sample amount (less than 90% of the required amount)
is received to perform the analyses, the Contractor shall contact the
Sample Management Office (SMO) to inform them of the problem. SMO
will contact the Region for instructions. The Region will either
require that no sample analyses be performed or will require that a
reduced volume be used for the sample analysis. No other changes in
the analyses will be permitted. The Contractor shall document the
Region’s decision in the Sample Delivery Group (SDG) Narrative.
10.1.2.1 If all of the phases are not amenable to analysis (i.e., outside
scope), the Region may require the Contractor to do any of the
following:
ILM05.3 D-16/ICP-AES
Exhibit D (ICP-AES) -- Section 10
Procedure (Con’t)
10.1.3.2.3 The caps with the pressure release valves are placed on the
vessels hand tight and then tightened, using constant torque,
to 12 ft/lbs. The weight of each vessel is recorded to
0.02 gram (g).
10.1.3.2.8 The microwave unit first-stage program must be set to give 545
watts for 10 minutes and the second-stage program to give 344
watts for 10 minutes. This sequence brings the samples to
160°C (±4°C) in 10 minutes and permits a slow rise to 165-170°C
during the second 10 minutes.
10.1.3.2.9 Following the 20 minute program, the samples are left to cool
in the microwave unit for 5 minutes, with the exhaust fan on.
The samples and/or carousel may then be removed from the
microwave unit. Before opening the vessels, let cool until
they are no longer hot to the touch.
10.1.3.2.10 After the sample vessel has cooled, weigh the sample vessel and
compare to the initial weight as reported on the preparation
log. Any sample vessel exhibiting a less than or equal to
0.5 g loss into the overflow vessel must have any excess sample
from the associated collection vessel added to the original
sample vessel before proceeding with the sample preparation.
D-17/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 10
Procedure (Con’t)
10.1.3.2.11 Sample Filtration - The digested samples are shaken well to mix
in any condensate within the digestion vessel before being
opened. The digestates are then filtered into 50 mL glass
volumetric flasks through Whatman No. 41 (or equivalent) filter
paper and diluted to 50 mL (if necessary). The samples are now
ready for analysis. The sample results must be corrected by a
factor of 1.11 in order to report final concentration values
based on an initial volume of 45 mL. Concentrations so
determined shall be reported as “total”.
10.1.3.3.4 Sample Filtration - The digested samples are shaken well to mix
in any condensate within the digestion vessel before being
opened. If necessary, the digestates are then filtered through
filter paper and diluted to 55 mL. The samples are now ready
for analysis. The sample results must be corrected by a factor
of 1.1 in order to report final concentration values based on
an initial volume of 50 mL. Concentrations so determined shall
be reported as “total”.
10.1.4.1.2 Add 10 mL of 1:1 nitric acid (HNO3), mix the slurry, and cover
with a watch glass. Heat the sample to 92-95°C on hot plate or
block digester, and reflux for 10 minutes without boiling.
Allow the sample to cool, add 5 mL of concentrated HNO3,
replace the watch glass, as appropriate, and reflux for 30
minutes. Do not allow the volume to be reduced to less than 5
mL while maintaining a covering of solution over the bottom of
the heating vessel.
10.1.4.1.3 After the second reflux step has been completed and the sample
has cooled, add 2 mL of reagent water and 3 mL of 30% hydrogen
peroxide (H2O2). Return the heating vessel to the heat source
for warming to start the peroxide reaction. Care must be taken
to ensure that losses do not occur due to excessively vigorous
effervescence. Heat until effervescence subsides, and cool the
heating vessel.
ILM05.3 D-18/ICP-AES
Exhibit D (ICP-AES) -- Section 10
Procedure (Con’t)
10.1.4.2.2 Add 10 mL of 1:1 nitric acid (HNO3), mix the slurry, and cover
with a watch glass. Heat the sample to 92-95°C on hot plate,
block digester, or equivalent heating source, and reflux for 10
minutes without boiling. Allow the sample to cool, add 5 mL of
concentrated HNO3, replace the watch glass, as appropriate, and
reflux for 30 minutes. Do not allow the volume to be reduced
to less than 5 mL while maintaining a covering of solution over
the bottom of the heating vessel. Add an additional 5 mL of
concentrated HNO3 and reflux. Repeat this step until sample
oxidation is complete (no brown fumes generated).
10.1.4.2.3 After the reflux steps have been completed and the sample has
cooled, add 2 mL of reagent water and 3 mL of 30% hydrogen
peroxide (H2O2). Return the heating vessel to the heat source
for warming to start the peroxide reaction. Care must be taken
to ensure that losses do not occur due to excessively vigorous
effervescence. Heat until effervescence subsides, and cool the
heating vessel.
D-19/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 10
Procedure (Con’t)
10.1.4.3.4 Connect the sample vessel to the overflow vessel using PTFE PFA
tubing.
10.1.4.3.8 The PB must have 0.5 mL of reagent water and the same amount
(10 mL) of acid that is added to the samples. The PB must
later be diluted to 50 mL in the same manner as the samples.
10.1.4.3.10 This program brings the samples to 175°C in 5.5 minutes; the
temperature remains between 170-180°C for the balance of the 10
minute irradiation period. The pressure should peak at less
than 6 atmospheres (atm) for most samples. The pressure may
exceed these limits in the case of high concentrations of
carbonate or organic compounds. In these cases, the pressure
will be limited by the relief pressure of the vessel to 7.5
(±0.7 atm).
10.1.4.3.12 Weigh each vessel assembly. If the weight of acid plus the
sample has decreased by more than 10% from the original weight,
discard the digests. Determine the reason for the loss.
Losses typically are attributed to use of digestion time longer
than ten minutes, using too large of a sample, or having
improper heating conditions. Once the source of the losses has
been corrected, prepare a new set of samples for digestion.
ILM05.3 D-20/ICP-AES
Exhibit D (ICP-AES) -- Section 10
Procedure (Con’t)
10.1.5.1.1 This code shall be used to report samples that are not digested
prior to analysis (e.g., dissolved metal samples that the
Contractor was instructed not to digest).
10.2.1.1 Prior to first use - new vessels must be annealed before they are
used. A pretreatment/cleaning procedure must be followed. This
procedure calls for heating the vessels for 96 hours at 200°C.
The vessels must be disassembled during annealing and the sealing
surfaces (the top of the vessel or its rim) must not be used to
support the vessel during annealing.
10.2.1.3 Immerse in 1:1 HCl for a minimum of 3 hours after the cleaning
bath has reached a temperature just below boiling.
10.2.1.5 Immerse in 1:1 HNO3 for a minimum of 3 hours after the cleaning
bath has reached a temperature just below boiling.
10.2.1.6 The vessels are then rinsed with copious amounts of reagent water
prior to use for any analyses under this contract.
D-21/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Sections 10 & 11
Data Analysis and Calculations
ILM05.3 D-22/ICP-AES
Exhibit D (ICP-AES) -- Section 11
Calculate the adjusted MDL or adjusted CRQL for soil samples as follows:
D-23/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 12
Quality Control
The ICV standard shall be prepared in the same acid matrix as the
calibration standards and in accordance with the instructions provided
by the supplier. If measurements exceed the control limits of 90% (low)
and 110% (high), the analysis shall be terminated, the problem
corrected, the instrument recalibrated, and the calibration reverified.
Information regarding the ICV shall be reported on Form IIA-IN.
12.3.1 To verify linearity near the CRQL, a standard at the CRQL (CRI) shall
be prepared, in the same acid matrix as the calibration standards,
and analyzed at the beginning (immediately following the ICV/ICB) and
end of each sample analysis run, immediately preceding the
Interference Check Sample (ICS) analyses. In addition, the
Contractor shall analyze the CRI at a frequency of not less than once
per 20 analytical samples1 per analysis run. These analyses of the
CRI sample shall be immediately followed by the ICS analyses. [That
is, the analytical run sequence shall be CRI, ICS Solution A (ICSA),
ICS Solution AB (ICSAB), CCV and Continuing Calibration Blank (CCB),
in that order].
12.3.2 The CRI shall be run for every wavelength used for analysis, except
those for Al, Ba, Ca, Fe, Mg, Na, and K. Information regarding the
CRI shall be reported on Form IIB-IN.
12.3.3 If the percent recovery of the CRI falls outside the control limits
of 70-130% (50-150% for antimony, lead, and thallium) for one or more
analytes, the CRI shall be re-analyzed immediately for those analytes
only. If the results of the re-analysis for those analytes fall
within the control limits, no further corrective action is required.
If the results of the re-analysis for those analytes do not fall
within the control limits, the analysis shall be terminated, the
problem corrected, the instrument recalibrated, the CRI analyzed, and
the samples associated with the CRI re-analyzed.
There are two different types of blanks required by this method. The
calibration blank is used in establishing the analytical curve while the
Preparation Blank is used to monitor for possible contamination.
1
As defined in Exhibit G, CRI is an analytical sample.
ILM05.3 D-24/ICP-AES
Exhibit D (ICP-AES) -- Section 12
Quality Control (Con’t)
The ICB and CCB are prepared with acids and reagent water. If the
absolute value of the calibration blank (ICB/CCB) result exceeds the
CRQL (see Exhibit C), the analysis shall be terminated, the problem
corrected, the instrument recalibrated, the calibration verified, and
re-analysis of the preceding 10 analytical samples or all analytical
samples analyzed since the last compliant calibration blank shall be
performed for the elements affected.
12.4.2.1 The PB shall contain all the reagents and in the same volumes as
used in processing the samples. The PB shall be carried through
the complete procedure and contain the same acid concentration in
the final solution as the sample solution used for analysis.
12.4.2.4 The PB is to be reported for each SDG and used in all analyses to
ascertain whether sample concentrations reflect contamination in
the following manner:
2
A group of samples prepared at the same time.
D-25/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 12
Quality Control (Con’t)
12.5.3 The ICS consists of two solutions: Solution A and Solution AB.
Solution A consists of the interferents, and Solution AB consists of
the analytes mixed with the interferents. An ICS analysis consists
of analyzing both solutions consecutively, starting with Solution A.
12.5.4 The analytical results of ICS Solution A (ICSA) shall fall within the
control limit of ±2 times the CRQL of the analyte’s true value or
±20% of the analyte’s true value, whichever is greater (the true
value shall be zero unless otherwise stated) in the ICSA. For
example, if the analysis result(s) for Arsenic (CRQL = 10 µg/L, ICSA
true value = 0 µg/L) in the ICSA analysis during the run is 19 µg/L,
then the analytical result for Arsenic falls within the ±2 times the
CRQL window for Arsenic in the ICSA. If the analytical results of
the ICSA do not fall within the control limits, the analysis shall be
terminated, the problem corrected, the instrument recalibrated, and
re-analysis of the analytical samples analyzed since the last
compliant ICSA shall be performed. For analytes with CRQLs less than
5000 µg/L, the ICSA results shall be reported from an undiluted
sample analysis.
12.5.5 Results for the ICS Solution AB (ICSAB) during the analytical runs
shall fall within the control limit of ±2 times the CRQL of the true
value or ±20% of the true value, whichever is greater, for the
analytes included in the ICSAB. If the analytical results of the
ICSA do not fall within the control limits, the analysis shall be
terminated, the problem corrected, the instrument recalibrated, and
re-analysis of the analytical samples analyzed since the last
compliant ICSAB shall be performed.
NOTE: The control limits and concentrations for the ICSAB are being
monitored. These may be adjusted to provide greater control of
interferences.
12.5.6 If true values for analytes contained in the ICS are not supplied
with the solutions, the mean shall be determined by initially
analyzing the ICS at least five times repetitively for the particular
analytes. This mean determination shall be made during an analytical
run where the results for the previously supplied ICS met all
contract specifications. Additionally, the results of this initial
mean determination shall be used as the true value for the lifetime
of that solution (i.e., until the solution is exhausted). Only if
the ICS solutions are not available from USEPA, independent Check
Samples shall be prepared with interferent and analyte concentrations
at the levels specified in Table 1 - Interferent and Analyte
Elemental Concentrations Used for ICP-AES Interference Check Sample
(ICS). The mean value and standard deviation shall be established by
3
As defined in Exhibit G, ICSA and ICSAB are analytical samples.
ILM05.3 D-26/ICP-AES
Exhibit D (ICP-AES) -- Section 12
Quality Control (Con’t)
12.6.2 If the spike analysis is performed on the same sample that is chosen
for the duplicate sample analysis, spike calculations shall be
performed using the results of the sample designated as the “original
sample” (see Section 12.7). The average of the duplicate results
cannot be used for the purpose of determining percent recovery.
Samples identified as field blanks and Performance Evaluation (PE)
samples shall not be used for spiked sample analysis. USEPA may
require that a specific sample be used for the spike sample analysis.
12.6.3 The analyte spike shall be added in the amount given in Table 2 -
Spiking Levels for Spike Sample Analysis, for each element analyzed.
12.6.4 If the spike recovery is not at or within the limits of 75-125%, the
data of all samples received and associated with that spike sample
shall be flagged with the letter “N” on Forms IA/IB-IN and VA-IN. An
exception to this rule is granted when the sample concentration
exceeds the spike added concentration by a factor of four or more.
In such an event, the data shall be reported unflagged even if the
percent recovery does not meet the 75-125% recovery criteria.
12.6.5 When the matrix spike recovery falls outside the control limits and
the sample result does not exceed four times the spike added, a post-
digestion spike shall be performed for those elements that do not
meet the specified criteria (exception: Ag). Note that if a post-
digestion spike analysis is required for an analyte, the same EPA
sample that was used for the matrix spike analysis shall be used for
the post-digestion spike analysis. Spike the unspiked aliquot of the
sample at two times the indigenous level or two times the CRQL,
whichever is greater. Results of the post-digestion spike shall be
reported on Form VB-IN.
12.6.6 In the instance where there is more than one spike sample per matrix
per SDG, if one spike sample recovery is not within contract
criteria, flag all the samples of the same matrix and method in the
SDG. Individual component percent recoveries are calculated as
follows:
4
USEPA may require additional spike sample analyses, upon USEPA Regional
CLP Project Officer (CLP PO) request.
D-27/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 12
Quality Control (Con’t)
12.6.7 When sample concentration is less than the Method Detection Limit
(MDL), use SR = 0 only for purposes of calculating percent recovery.
The Spike Sample Results (SSRs), Sample Results (SRs), Spike Added
(SA), and percent recovery (positive or negative) shall be reported
on Form VA-IN.
12.6.8 The units used for reporting SSRs will be identical to those used for
reporting sample results on Form IA-IN.
12.7.1 One duplicate sample shall be analyzed from each group of samples of
a similar matrix type (i.e., water, soil) or for each SDG.5
Duplicates cannot be averaged for reporting on Form IA-IN.
12.7.2 Duplicate sample analyses are required for percent solids. Samples
identified as field blanks and PE samples shall not be used for
duplicate sample analysis. USEPA may require that a specific sample
be used for duplicate sample analysis. The Relative Percent
Difference (RPD) for each component is calculated as follows:
12.7.4 If one result is above five times the CRQL level and the other is
below, use the CRQL criteria to determine if the duplicate analysis
is in control. If both sample and duplicate values are less than the
5
USEPA may require additional duplicate sample analyses, upon USEPA
Regional CLP PO request.
ILM05.3 D-28/ICP-AES
Exhibit D (ICP-AES) -- Section 12
Quality Control (Con’t)
MDL, the RPD is not calculated on Form VI-IN. For solid sample or
solid duplicate results less than five times the CRQL, enter the
value of the CRQL, corrected for sample weight and percent solids,
(i.e., original, not duplicate sample weight and percent solids), in
the “Control Limit” column. If the duplicate sample results are
outside the control limits, flag all the data for samples received
associated with that duplicate sample with an “*” on Forms IA/IB-IN
and VI-IN. In the instance where there is more than one duplicate
sample per SDG, if one duplicate result is not within contract
criteria, flag all samples of the same matrix in the SDG. The
percent difference data will be used by USEPA to evaluate the long-
term precision of the methods for each element. Specific control
limits for each element will be added to Form VI-IN at a later date
based on these precision results.
12.8.1 Water/aqueous and solid LCS shall be analyzed for each analyte using
the same sample preparations, analytical methods, and Quality
Assurance/Quality Control (QA/QC) procedures employed for the EPA
samples received.
12.8.1.1 The aqueous LCS solution (LCSW) shall be obtained from USEPA [if
unavailable, the ICV solution(s) may be used]. One LCSW shall be
prepared and analyzed for every group of aqueous samples in a SDG,
or for each batch of aqueous samples digested, whichever is more
frequent.
12.8.1.2 The USEPA provided solid LCS (LCSS) shall be prepared and analyzed
using each of the procedures applied to the solid samples received
(exception: percent solids determination not required). If the
USEPA LCSS is unavailable, other USEPA QC Check Samples or other
certified materials may be used. The control limits for these
materials and samples must be documented. One LCSS shall be
prepared and analyzed for every group of solid samples in a SDG,
or for each batch of samples digested, whichever is more frequent.
12.8.2 All LCS and percent recovery results shall be reported on Form VII-
IN. If the percent recovery for the LCSW falls outside the control
limits of 80-120% (exception: Ag and Sb), the analyses shall be
terminated, the problem corrected, and the samples associated with
that LCSW redigested and re-analyzed with appropriate new QC.
12.8.3 If the results for the LCSS fall outside the control limits
established by USEPA, the analyses shall be terminated, the problem
corrected, and the samples associated with that LCSS redigested and
re-analyzed with appropriate new QC.
12.9.1 Prior to reporting concentration data for the analyte elements, the
Contractor shall analyze and report the results of the ICP-AES serial
dilution analysis. The ICP-AES serial dilution analysis shall be
performed on a sample from each group of samples of a similar matrix
type (i.e., water, soil) or for each SDG, whichever is more frequent.
Samples identified as field blanks and PE samples shall not be used
for serial dilution analysis.
D-29/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 12
Quality Control (Con’t)
12.9.3 The percent differences for each component are calculated as follows:
12.9.4 In the instance where there is more than one serial dilution per SDG,
if one serial dilution result is not within contract criteria, flag
all the samples of the same matrix in the SDG. Serial dilution
results and “E” flags shall be reported on Form VIII-IN.
12.10.1 Before any field samples are analyzed under this contract, the MDLs
shall be determined for non-prepared analyses (Preparation
Method/Code “NP1”), each digestion procedure and instrument used,
prior to the start of contract analyses, and annually thereafter, and
shall meet the levels specified in Exhibit C.
12.10.1.1 To determine the MDLs, the Contractor shall run MDL studies
following the procedures given in 40 CFR, Part 136. The
Contractor shall prepare the MDL samples by each digestion
procedure used and shall analyze these samples on each instrument
used. The Contractor shall also analyze the non-prepared MDL
samples on each instrument used.
12.10.1.2 The determined concentration of the MDL shall be less than half
the concentration of the CRQL listed in Exhibit C.
12.11.1 Before any field samples are analyzed under this contract, the
interelement correction factors shall be determined prior to the
start of contract analyses and at least quarterly thereafter.
Correction factors for spectral interference due to Al, Ca, Fe, and
Mg shall be determined for all ICP-AES instruments at all wavelengths
ILM05.3 D-30/ICP-AES
Exhibit D (ICP-AES) -- Section 12
Quality Control (Con’t)
12.11.2 If the instrument was adjusted in any way that may affect the ICP-AES
interelement correction factors, the factors shall be redetermined
and the results submitted for use. In addition, all data used for
the determination of the interelement correction factors shall be
available to the USEPA during an on-site laboratory evaluation.
Results from interelement correction factors determination shall be
reported on Form XA-IN and Form XB-IN for all ICP-AES analytes.
12.12.1 Before any field samples are analyzed under this contract, the linear
ranges shall be determined and reported prior to the start of
contract analyses, and at least quarterly thereafter by the analysis
of a linear range verification check standard, for each element on
Form XI-IN. The standard shall be analyzed during a routine
analytical run performed under this contract. The analytically
determined concentration of this standard shall be within 5% of the
true value. This concentration is the upper limit of the ICP-AES
linear range beyond which results cannot be reported under this
contract without dilution of the analytical sample.
S0
ICV
ICB
CRI
ICSA
ICSAB
CCV
CCB
10 samples
CCV
CCB
7 samples
CRI
ICSA
ICSAB
CCV
CCB
10 samples, etc.
D-31/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Sections 13-16
Method Performance
Not applicable.
16.0 REFERENCES
16.3 American Society for Testing and Materials. Standard Practice for Sample
Digestion Using Closed Vessel Microwave Heating Technique for the
Determination of Total Recoverable Metals in Water. D4309-91. October
1991.
ILM05.3 D-32/ICP-AES
Exhibit D (ICP-AES) -- Section 17
Tables/Diagrams/Flowcharts
17.0 TABLES/DIAGRAMS/FLOWCHARTS
D-33/ICP-AES ILM05.3
Exhibit D (ICP-AES) -- Section 17
Tables/Diagrams/Flowcharts
*
No spike required. NOTE: Elements without spike levels, and not
designated with an asterisk, shall be spiked at appropriate levels.
1
The levels shown indicate concentrations in the spike sample when the
wet weight of 1 gram of sample is taken for analysis. Adjustment shall be
made to maintain these spiking levels when the weight of sample taken deviates
by more than 10% of these values. Appropriate adjustment shall be made for
microwave digestion procedures where 0.5 grams of sample or 50 mL (45 mL of
sample plus 5 mL of acid) or 55 mL (50 mL of sample plus 5 mL of acid) of
aqueous sample are required for analysis.
ILM05.3 D-34/ICP-AES
EXHIBIT D - PART B
ANALYTICAL METHODS
FOR
INDUCTIVELY COUPLED PLASMA -
MASS SPECTROMETRY
D-1/ICP-MS ILM05.3
THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.3 D-2/ICP-MS
Exhibit D - Analytical Methods for ICP-MS
Table of Contents
Section Page
3.0 DEFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4.0 INTERFERENCES . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.1 Isobaric Elemental Interferences . . . . . . . . . . . . . . . 6
4.2 Abundance Sensitivity . . . . . . . . . . . . . . . . . . . . . 6
4.3 Isobaric Polyatomic Ion Interferences . . . . . . . . . . . . . 6
4.4 Physical Interferences . . . . . . . . . . . . . . . . . . . . 6
4.5 Memory Interferences . . . . . . . . . . . . . . . . . . . . . 7
5.0 SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
10.0 PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
10.1 Sample Preparation . . . . . . . . . . . . . . . . . . . . . . 16
10.2 Sample Analysis . . . . . . . . . . . . . . . . . . . . . . . . 17
D-3/ICP-MS ILM05.3
Exhibit D - Analytical Methods for ICP-MS
16.0 REFERENCES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
17.0 TABLES/DIAGRAMS/FLOWCHARTS . . . . . . . . . . . . . . . . . . . . . 30
Table 1. Isobaric Molecular-Ion Interferences . . . . . . . . . . 30
Table 2. Mass Choices for Elements that Must Be Monitored
During the Analytical Run . . . . . . . . . . . . . . . . 33
Table 3. Recommended Elemental Expressions for Isobaric
Interferences . . . . . . . . . . . . . . . . . . . . . . 34
Table 4. Internal Standards . . . . . . . . . . . . . . . . . . . 35
Table 5. Spiking Levels for Spike Sample Analysis . . . . . . . . 35
ILM05.3 D-4/ICP-MS
Exhibit D (ICP-MS) -- Sections 1-3
Scope and Application
3.0 DEFINITIONS
D-5/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 4
Interferences
4.0 INTERFERENCES
These are caused by ions consisting of more than one atom which have the
same nominal mass-to-charge ratio as the isotope of interest, and which
cannot be resolved by the mass spectrometer. These ions are commonly
formed in the plasma or interface system from support gases or sample
components. Most of the common interferences have been identified and
are listed in Table 1 - Isobaric Molecular-Ion Interferences, with the
target analytes affected. Such interferences must be recognized, and
when they cannot be avoided by the selection of alternative analytical
isotopes, appropriate corrections must be made to the data. Equations
for the correction of data should be established at the time of the
analytical run sequence, since the polyatomic ion interferences will be
highly dependent on the sample matrix and chosen instrument conditions.
These are associated with the physical processes which govern the
transport of the sample into the plasma, sample conversion processes in
the plasma, and the transmission of ions through the plasma-mass
spectrometer interface. These interferences may result in differences
between instrument responses for the sample and the calibration
standards. Physical interferences may occur in the transfer of solution
to the nebulizer (e.g., viscosity effects), at the point of aerosol
formation and transport to the plasma (e.g., surface tension), or during
ILM05.3 D-6/ICP-MS
Exhibit D (ICP-MS) -- Sections 4 & 5
Safety
the excitation and ionization processes within the plasma itself. High
levels of dissolved solids in the sample may contribute to deposits of
material on the extraction and/or skimmer cones. Deposits can reduce
the effective diameter of the orifices and therefore ion transmission.
Dissolved solid levels not exceeding 0.2% (w/v) have been recommended to
reduce such effects. Internal standardization may be effectively used
to compensate for many physical interference effects. Internal
standards ideally should have similar analytical behavior to the
elements being determined.
5.0 SAFETY
D-7/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 6
Equipment and Supplies
Brand names, suppliers, and part numbers are for illustrative purposes
only. No endorsement is implied. Equivalent performance may be
achieved using equipment and supplies other than those specified here,
however, a demonstration of equivalent performance meeting the
requirements of this Statement of Work (SOW) is the responsibility of
the Contractor. The Contractor shall document any use of alternate
equipment or supplies in the Sample Delivery Group (SDG) Narrative.
6.1 Glassware/Labware
6.1.3 Funnels
6.1.9 Balances - Analytical Balance, 300 gram (g) capacity, and minimum
±0.1 milligram (mg).
ILM05.3 D-8/ICP-MS
Exhibit D (ICP-MS) -- Section 7
Reagents and Standards
7.1 Reagents
7.1.1 Reagent Water - The purity of this water must be equivalent to ASTM
Type II water (ASTM D1193-77). Use this preparation for all
reagents, standards, and dilutions of solutions.
7.1.3 Nitric acid (1+1) - Add 500 milliliters (mL) conc. HNO3 to 400 mL of
reagent water and dilute to 1 Liter (L).
7.1.4 Nitric acid (1+9) - Add 100 mL conc. nitric acid to 400 mL of reagent
water and dilute to 1 L.
7.1.6 Hydrochloric acid (1+1) - Add 500 mL conc. HCl to 400 mL of reagent
water and dilute to 1 L.
7.1.7 Hydrochloric acid (HCl) (1+4) - Add 200 mL conc. HCl to 400 mL
reagent water and dilute to 1 L.
7.2 Standards
7.2.1 Introduction
D-9/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 7
Reagents and Standards (Con’t)
repeatedly, rinsed with water, dried and weighed until the desired
weight is achieved.
ILM05.3 D-10/ICP-MS
Exhibit D (ICP-MS) -- Section 7
Reagents and Standards (Con’t)
D-11/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 7
Reagents and Standards (Con’t)
7.3 Blanks
Three types of blanks are required for this method. A calibration blank
is used to establish the analytical calibration curve, the Preparation
Blank (PB) (see Section 12.5.2) is used to assess possible contamination
ILM05.3 D-12/ICP-MS
Exhibit D (ICP-MS) -- Sections 7 & 8
Sample Collection, Preservation, and Storage
7.3.2 Preparation Blank - Must contain all the reagents in the same volumes
as used in preparing the samples. The PB must be carried through the
complete procedure and contain the same acid concentration in the
final solution as the sample solution used for analysis.
The samples must be protected from light and refrigerated at 4°C (±2°C)
from the time of receipt until 60 days after the delivery of a complete,
reconciled data package to USEPA. After 60 days the samples may be
disposed of in a manner that complies with all applicable regulations.
The maximum holding time for metals is 180 days from Validated Time of
Sample Receipt (VTSR).
D-13/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 9
Calibration and Standardization
ILM05.3 D-14/ICP-MS
Exhibit D (ICP-MS) -- Section 9
Calibration and Standardization (Con’t)
9.2.3 Calibration
9.3.1 Immediately after each instrument has been calibrated, the accuracy
of the initial calibration shall be verified and documented for every
analyte by the analysis of the ICV solution(s) for each mass used to
report final results.
9.3.2 Only if the ICV solution(s) is(are) not available from USEPA, or
where a certified solution of an analyte is not available from any
source, analyses shall be conducted on an independent standard at a
concentration other than that used for instrument calibration, but
within the calibration range. An independent standard is defined as
a standard composed of the analytes from a different source other
than those used in the standards for instrument calibration.
9.3.3 The ICV solution(s) shall be run at each mass used for reporting
final results. The values for the ICV shall be reported on Form IIA-
IN.
9.4.1 To ensure calibration accuracy during each analysis run, one of the
following standards shall be used for the CCV for each mass used for
reporting final results for each element, at a frequency of 10% or
every 2 hours during an analysis run, whichever is more frequent.
The standard shall also be analyzed and reported for each mass used
for reporting final results for each element at the beginning of the
run and after the last analytical sample. The analyte concentrations
in the CCV standard(s) shall be different from the concentrations for
the ICV and shall be one of the following solutions at or near one-
half of the calibration standard:
C USEPA Solutions
C NIST Standards
The same CCV standard shall be used throughout the analysis runs for
a Sample Delivery Group (SDG) of samples received.
9.4.2 Each CCV analyzed shall reflect the conditions of analysis of all
associated analytical samples (the preceding 10 analytical samples or
the preceding analytical samples up to the previous CCV). The
duration of analysis, rinses, and other related operations which may
affect the CCV measured result may not be applied to the CCV to a
greater extent than the extent applied to the associated analytical
samples. For instance, the difference in time between a CCV analysis
and the blank immediately following it, as well as the difference in
D-15/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Sections 9 & 10
Procedure
time between the CCV and the analytical sample immediately preceding
it, may not exceed the lowest difference in time between any two
consecutive analytical samples associated with the CCV.
A calibration blank shall be analyzed for each mass used for reporting
final results for each element immediately after every ICV and CCV, at a
frequency of 10% or every 2 hours during the run, whichever is more
frequent. The blank shall be analyzed at the beginning of the run and
after the last analytical sample.
NOTE: A CCB shall be analyzed immediately after the last CCV, and the
last CCV shall be analyzed immediately after the last analytical sample
of the run. The results of the calibration blanks shall be reported on
Form III-IN.
10.0 PROCEDURE
10.1.1 If insufficient sample amount (less than 90% of the required amount)
is received to perform the analyses, the Contractor shall contact the
Sample Management Office (SMO) to inform them of the problem. SMO
will contact the Region for instructions. The Region will either
require that no sample analysis be performed or will require that a
reduced volume be used for the sample analysis. No other changes in
the analysis will be permitted. The Contractor shall document the
Region’s decision in the Sample Delivery Group (SDG) Narrative.
10.1.2.1 If all of the phases are not amenable to analysis (i.e., outside
scope), the Region may require the Contractor to do any of the
following:
ILM05.3 D-16/ICP-MS
Exhibit D (ICP-MS) -- Section 10
Procedure (Con’t)
D-17/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 10
Procedure (Con’t)
10.2.3 The rinse blank should be used to flush the system between samples.
Allow sufficient time to remove traces of the previous sample or a
minimum of one minute. Samples should be aspirated for a sufficient
period of time to obtain a stable response prior to the collection of
data.
10.2.5 All masses which might affect data quality must be monitored during
the analytical run. At a minimum, those masses prescribed in Table 2
- Mass Choices for Elements that Must Be Monitored During the
Analytical Run, must be monitored in the same scan that is used for
the collection of the data. This information should be used to
correct the data for identified interferences.
10.2.6 During the analysis of samples, the laboratory must comply with the
required QC described in Section 12. For the determination of
dissolved analytes when the Region has specified that no preparation
is required, the Preparation Blank (PB) and Laboratory Control Sample
(LCS) are not required.
ILM05.3 D-18/ICP-MS
Exhibit D (ICP-MS) -- Section 11
Data Analysis and Calculations
D-19/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 11
Data Analysis and Calculations (Con’t)
Vf
Concentration (g/L) = C x x DF
Vi
ILM05.3 D-20/ICP-MS
Exhibit D (ICP-MS) -- Section 12
Quality Control
12.0 QUALITY CONTROL (QC)
The Tune Standard shall be prepared in the same acid matrix as the
calibration standards and analyzed at least 5 times consecutively.
Analyses may be carried out as five separate analyses or as a single
analysis with at least five integrations. If the mass calibration is
not within 0.1 amu over the range of 6 to 210 amu, or the percent
Relative Standard Deviation (%RSD)of the absolute signals of the
analytes exceeds 5%, the analysis shall be terminated, the problem
corrected, and the instrument re-tuned. All sample results reported
must be associated with an instrument tune that meets these
requirements.
The ICV Standard shall be prepared in the same acid matrix as the
calibration standards and in accordance with the instructions provided
by the supplier. If measurements exceed the control limits of 90% (low)
and 110% (high), the analysis shall be terminated, the problem
corrected, the instrument recalibrated, and the calibration reverified.
Information regarding the ICV shall be reported on Form IIA-IN.
12.4.1 To verify linearity near the CRQL, a standard at the CRQL (CRI) shall
be prepared, in the same acid matrix as the calibration standards,
and analyzed at the beginning (immediately following the ICV/ICB and
immediately preceding the Interference Check Sample (ICS) analyses).
In addition, the contractor shall analyze the CRI at the end of each
sample analysis run and at a frequency of not less than once per 20
analytical samples1 per analysis run. These subsequent analyses of
the CRI shall be immediately followed by CCV/CCB analyses.
12.4.2 The CRI shall be run for every required isotope used for the analysis
of all Inductively Coupled Plasma - Mass Spectrometry (ICP-MS)
analytes. Information regarding the CRI shall be reported on Form
IIB-IN.
12.4.3 If the percent recovery of the CRI falls outside the control limits
of 70-130% (50-150% for cobalt, manganese, and zinc) for one or more
analytes, the CRI shall be re-analyzed immediately for those analytes
only. If the results of the re-analysis for those analytes fall
within the control limits, no further corrective action is required.
If the results of the re-analysis for those analytes do not fall
within the control limits, the analysis shall be terminated, the
1
As defined in Exhibit G, CRI is an analytical sample.
D-21/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 12
Quality Control (Con’t)
problem corrected, the instrument recalibrated, the CRI analyzed, and
the samples associated with the CRI re-analyzed.
There are two different types of blanks required by this method. The
calibration blank is used in establishing the analytical curve while the
preparation blank is used to monitor for possible contamination.
The ICB and CCB are prepared with acid and reagent water. If the
absolute value of the calibration blank (ICB/CCB) result exceeds the
CRQL (see Exhibit C), the analysis shall be terminated, the problem
corrected, the instrument recalibrated, the calibration verified, and
re-analysis of the preceding 10 analytical samples or all analytical
samples analyzed since the last compliant calibration blank shall be
performed for the elements affected.
12.5.2.1 The PB shall contain all the reagents and in the same volumes as
used in processing the samples. The PB shall be carried through
the complete procedure and contain the same acid concentration in
the final solution as the sample solution used for analysis.
12.5.2.4 The PB is to be reported for each SDG and used in all analyses to
ascertain whether sample concentrations reflect contamination in
the following manner:
2
A group of samples prepared at the same time.
ILM05.3 D-22/ICP-MS
Exhibit D (ICP-MS) –- Section 12
Quality Control (Con’t)
with the blank, shall be redigested and re-analyzed with
appropriate new QC.
12.6.3 The ICS consists of two solutions: Solution A and Solution AB.
Solution A consists of the interferents, and Solution AB consists of
the analytes mixed with the interferents. An ICS analysis consists
of analyzing both solutions consecutively, starting with Solution A.
12.6.4 The analytical results of ICS Solution A (ICSA) shall fall within the
control limit of ±3 times the CRQL of the analyte’s true value or
±20% of the analyte’s true value (the true value shall be zero unless
otherwise stated) in the ICSA, whichever is greater. If not, the
analysis shall be terminated, the problem corrected, the instrument
recalibrated, and re-analysis of the analytical samples analyzed
since the last compliant ICSA shall be performed. The ICSA results
for these analytes shall be reported from an undiluted sample
analysis.
12.6.5 Results for the ICS Solution AB (ICSAB) during the analytical runs
shall fall within the control limit of ±3 times the CRQL of the true
value or ±20% of the true value, whichever is greater, for the
analytes included in the ICSAB. If not, the analysis shall be
terminated, the problem corrected, the instrument recalibrated, and
re-analysis of the analytical samples analyzed since the last
compliant ICSAB shall be performed.
NOTE: The control limits and concentrations for the ICSAB are being
monitored. These may be adjusted to provide greater control of
interferences.
12.6.6 If true values for analytes contained in the ICS are not supplied
with the solutions, the mean shall be determined by initially
analyzing the ICS at least five times repetitively for the particular
analytes. This mean determination shall be made during an analytical
run where the results for a previously supplied ICS met all contract
specifications. Additionally, the results of this initial mean
determination shall be used as the true value for the lifetime of
that solution (i.e., until the solution is exhausted). Only if the
ICS solutions are not available from USEPA, independent Check Samples
shall be prepared with interferent and analyte concentrations at the
levels specified in Sections 7.2.4.4.1 and 7.2.4.4.2. The mean value
and standard deviation shall be established by initially analyzing
the Check Samples at least five times repetitively for each analyte
listed on Form IVB-IN. Results shall fall within the control limit
of ±3 times the CRQL of the established mean value or ±20% of the
D-23/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 12
Quality Control (Con’t)
established mean value, whichever is greater. The mean and standard
deviation shall be reported in the raw data. Results from the ICS
analyses shall be reported on Form IVB-IN for all ICP-MS parameters.
12.7.2 If the spike analysis is performed on the same sample that is chosen
for the duplicate sample analysis, spike calculations shall be
performed using the results of the sample designated “original
sample” (see Section 12.8). The average of the duplicate results
cannot be used for the purpose of determining percent recovery.
Samples identified as field blanks and Performance Evaluation (PE)
samples shall not be used for spiked sample analysis. USEPA may
require that a specific sample be used for the spike sample analysis.
12.7.3 The analyte spike shall be added in the amount given in Table 5 -
Spiking Levels for Spike Sample Analysis, for each element analyzed.
12.7.4 If the spike recovery is not at or within the limits of 75-125%, the
data for all samples received and associated with that spike sample
and shall be flagged with the letter “N” on Forms IA/IB-IN and VA-IN.
An exception to this rule is granted when the sample concentration
exceeds the Spike Added (SA) concentration by a factor of four or
more. In such an event, the data shall be reported unflagged even if
the percent recovery does not meet the 75-125% recovery criteria.
12.7.5 When the matrix spike recovery falls outside the control limits and
the sample result does not exceed four times the spike added, a post-
digestion spike shall be performed for those elements that do not
meet the specified criteria. Note that if a post-digestion spike
analysis is required for an analyte, the same EPA sample that was
used for the matrix spike shall be used for the post-digestion spike
analysis. Spike an unspiked aliquot of the digestate at two times
the indigenous level or two times the CRQL, whichever is greater.
Results of the post-digestion spike shall be reported on Form VB-IN.
12.7.6 In the instance where there is more than one spike sample per matrix
per SDG, if one spike sample recovery is not within contract
criteria, flag all the samples in the SDG. Individual component
percent recoveries are calculated as follows:
SSR − SR
%Recovery = x 100
SA
3
USEPA may require additional spike sample analyses, upon USEPA Regional
CLP Project Officer (CLP PO) request.
ILM05.3 D-24/ICP-MS
Exhibit D (ICP-MS) –- Section 12
Quality Control (Con’t)
12.7.7 When sample concentration is less than the Method Detection Limit
(MDL), use SR = 0 only for purposes of calculating percent recovery.
The Spike Sample Results (SSRs), Sample Results (SRs), Spike Added
(SA), and percent recovery (positive or negative) shall be reported
on Form VA-IN.
12.7.8 The units used for reporting SSRs will be identical to those used for
reporting sample results on Form IA-IN.
12.8.1 One duplicate sample shall be analyzed for each SDG4. Duplicates
cannot be averaged for reporting on Form IA-IN.
12.8.2 Samples identified as field blanks and PE samples shall not be used
for duplicate sample analysis. USEPA may require that a specific
sample be used for duplicate sample analysis. The Relative Percent
Difference (RPD) for each analyte is calculated as follows:
S−D
RPD = x 100
(S + D)/2
12.8.4 If one result is above five times the CRQL level and the other is
below, use the CRQL criteria to determine if the duplicate analysis
is in control. If both sample and duplicate values are less than the
MDL, the RPD is not calculated on Form VI-IN. If the duplicate
sample results are outside the control limits, flag all the data for
samples received associated with that duplicate sample with an “*” on
Forms IA/IB-IN and VI-IN. In the instance where there is more than
one duplicate sample per SDG, if one duplicate result is not within
contract criteria, flag all samples in the SDG. The percent
difference data will be used by USEPA to evaluate the long-term
precision of the methods for each element. Specific control limits
for each element may be added to Form VI-IN at a later date based on
these precision results.
4
USEPA may require additional duplicate sample analyses, upon USEPA
Regional CLP PO request.
D-25/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 12
Quality Control (Con’t)
12.9 Laboratory Control Sample (LCS) Analysis
12.9.1 A water/aqueous LCS (LCSW) shall be analyzed for each analyte using
the same sample preparations, analytical methods, and Quality
Assurance/Quality Control (QA/QC) procedures employed for USEPA
samples received.
12.9.2 The LCSW solution must be obtained from USEPA (if unavailable, the
ICV solution(s) may be used). One aqueous LCS shall be prepared and
analyzed for each group of samples in an SDG, or for each batch of
samples digested, whichever is more frequent.
12.9.3 All LCSW and percent recovery results shall be reported on Form VII-
IN. If the percent recovery for the LCSW falls outside the control
limits of 80-120%, the analyses shall be terminated, the problem
corrected, and the samples associated with that LCSW redigested and
re-analyzed with appropriate new QC.
12.10.1 Prior to reporting concentration data for the analyte elements, the
Contractor shall analyze and report the results of the ICP-MS serial
dilution analysis. The ICP-MS serial dilution analysis shall be
performed on a sample from each SDG. Samples identified as field
blanks and PE samples shall not be used for serial dilution analysis.
12.10.3 The percent differences for each component are calculated as follows:
I− S
%Difference = x 100
I
12.10.4 In the instance where there is more than one serial dilution per SDG,
if one serial dilution result is not within the contract criteria,
flag all samples in the SDG. Serial dilution results and “E” flags
shall be reported on Form VIII-IN.
12.10.5 If the internal standard responses for the field sample chosen for
serial dilution analysis are not within the limits and the
appropriate corrective action (two-fold dilution and reanalysis) is
taken, the following shall apply to the serial dilution analysis: if
the internal standard responses of the field sample reanalysis are
within the limits, the serial dilution results are to be reported
ILM05.3 D-26/ICP-MS
Exhibit D (ICP-MS) –- Section 12
Quality Control (Con’t)
from a five-fold dilution of the reanalyzed sample. If the internal
standard responses of the field sample reanalysis are not within the
limits, the serial dilution results are to be reported from a five-
fold dilution of the original sample.
12.11.1 The analyst shall monitor the responses from the internal standards
throughout the sample set being analyzed. Ratios of the internal
standard responses between isotopes should also be routinely
monitored. This information may be used to correct potential
problems caused by mass dependent drift, errors incurred in adding
the internal standards or increases in the concentrations of
individual internal standards caused by background contributions from
the sample. The absolute response of any one internal standard must
not deviate more than 60-125% of the original response in the
calibration blank. If deviations greater than these are observed in
field samples, matrix spikes, or duplicate samples, the original
sample shall be diluted by a factor of two, internal standards added,
and the sample re-analyzed. If the internal standard responses for
the diluted sample analysis are within the limits, report the results
of this analysis on the appropriate Summary Form. If the internal
standard responses for the diluted sample analysis are not within the
limits, note this in the SDG Narrative and report the results of the
undiluted original sample analysis on the appropriate Summary Form.
12.12.1 Before any field samples are analyzed under this contract, the MDLs
shall be determined for each instrument used, prior to the start of
contract analyses, and annually thereafter, and shall meet the levels
specified in Exhibit C.
12.12.2 To determine the MDLs, the Contractor shall run MDL studies following
the procedures given in 40 CFR, Part 136. The Contractor shall
prepare the MDL samples by each digestion procedure used and shall
analyze these samples on each instrument used. The Contractor shall
also analyze non-prepared MDL samples on each instrument used.
12.12.3 The determined concentration of the MDL shall be less than half the
concentration of the CRQL listed in Exhibit C.
12.12.4 The direct analysis MDL (Preparation Method/Code “NP1”) shall be used
to determine the appropriate concentration qualifier for the results
of instrument QC.
12.13.1 Before any field samples are analyzed under this contract, the upper
limit of the linear calibration range shall be established for each
analyte by determining the signal responses from a minimum of three
different concentration standards, one of which is close to the upper
limit of the linear range, prior to the start of contract analyses
D-27/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 12
Quality Control (Con’t)
and at least quarterly thereafter. The linear calibration range used
for the analysis of samples shall be determined from the resulting
data. The upper LDR limit shall be an observed signal no more than
10% below the level extrapolated from lower standards. Determined
sample analyte concentrations that are greater than 90% of the
determined upper LDR limit must be diluted and re-analyzed. The LDRs
must be verified whenever a change in instrument hardware operating
conditions indicate they should be redetermined, or verified
quarterly.
Tune
S0
S
ICV
ICB
CRI
ICSA
ICSAB
CCV
CCB
10 samples
CCV
CCB
7 samples
CRI
CCV
CCB
10 samples, etc.
ILM05.3 D-28/ICP-MS
Exhibit D (ICP-MS) -- Sections 13-16
Method Performance
13.0 METHOD PERFORMANCE
Not applicable.
16.0 REFERENCES
D-29/ICP-MS ILM05.3
Exhibit D (ICP-MS) -- Section 17
Tables/Diagrams/Flowcharts
17.0 TABLES/DIAGRAMS/FLOWCHARTS
ILM05.3 D-30/ICP-MS
Exhibit D (ICP-MS) -- Section 17
Tables/Diagrams/Flowcharts (Con’t)
D-31/ICP-MS ILM05.3
Exhibit D (ICP-MS) –- Section 17
Tables/Diagrams/Flowcharts (Con’t)
NOTE: The information provided in this table does not indicate that all of the
described interferences need to be tested. However, this table can be
consulted if unusual samples are encountered.
ILM05.3 D-32/ICP-MS
Exhibit D (ICP-MS) -- Section 17
Tables/Diagrams/Flowcharts (Con’t)
Table 2. Mass Choices for Elements that Must Be Monitored
During the Analytical Run
D-33/ICP-MS ILM05.3
Exhibit D (ICP-MS) – Section 17
Tables/Diagrams/Flowcharts (Con’t)
Table 3. Recommended Elemental Expressions for Isobaric Interferences
ILM05.3 D-34/ICP-MS
Exhibit D (ICP-MS) – Section 17
Tables/Diagrams/Flowcharts (Con’t)
Table 4. Internal Standards (must use at least five)
D-35/ICP-MS ILM05.3
EXHIBIT D - PART C
ANALYTICAL METHODS
FOR
D-1/Mercury ILM05.3
THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.3 D-2/Mercury
Exhibit D - Analytical Methods for Cold Vapor Mercury Analysis
Table of Contents
Section Page
3.0 DEFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.0 INTERFERENCES . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.1 Water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.2 Soil/Sediment . . . . . . . . . . . . . . . . . . . . . . . . . 6
5.0 SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Manual Techniques) . . . . . . . . . . . . . . . . . . . . . . 7
7.1 Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
7.2 Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
10.0 PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Standard (CRI) . . . . . . . . . . . . . . . . . . . . . . . . 19
D-3/Mercury ILM05.3
Exhibit D - Analytical Methods for Cold Vapor Mercury Analysis
16.0 REFERENCES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
17.0 TABLES/DIAGRAMS/FLOWCHARTS . . . . . . . . . . . . . . . . . . . . . 24
ILM05.3 D-4/Mercury
Exhibit D (Mercury) -- Sections 1 & 2
Scope and Application
The range of the method may be varied through instrument and/or recorder
expansion. Using a 100 milliliters (mL) sample, a detection limit of
less than 0.1 micrograms per Liter (µg/L) can be achieved.
D-5/Mercury ILM05.3
Exhibit D (Mercury) -- Sections 3 & 4
Definitions
3.0 DEFINITIONS
4.0 INTERFERENCES
4.1 Water
4.1.1 Some sea waters and wastewaters high in chlorides have shown a
positive interference, and require additional permanganate [as much
as 25 milliliters (mL)]. During the oxidation step, chlorides are
converted to free chlorine which will also absorb radiation at 253
nanometers (nm). Care must be taken to assure that free chlorine is
absent before the mercury is reduced and swept into the cell. This
may be accomplished by using an excess of hydroxylamine sulfate
reagent (25 mL). Both inorganic and organic mercury spikes have been
quantitatively recovered from the sea water using this technique.
4.1.5 Samples containing solids must be blended and then mixed while being
sampled if total mercury values are to be reported.
4.2 Soil/Sediment
4.2.1 The same types of interferences that may occur in water samples are
also possible with sediments (i.e., sulfides, high copper, high
chlorides, etc.).
ILM05.3 D-6/Mercury
Exhibit D (Mercury) -- Sections 5 & 6
Safety
5.0 SAFETY
Brand names, suppliers, and part numbers are for illustrative purposes
only. No endorsement is implied. Equivalent performance may be
achieved using equipment and supplies other than those specified here,
however, a demonstration of equivalent performance meeting the
requirements of this Statement of Work (SOW) is the responsibility of
the Contractor. The Contractor shall document any use of alternate
equipment or supplies in the Sample Delivery Group (SDG) Narrative.
6.1 General Information for Water and Soils (Automated and Manual
Techniques)
NOTE: All cold vapor mercury analyzers shall be equipped with all
manufactured required equipment (i.e., dryers) to ensure that the
specified CRQLs are met.
6.2.1.2 Manifold
6.3.2 Air Pump - Any device capable of delivering 1 Liter (L) of air per
minute may be used.
6.3.4 Aeration Tubing - Tygon tubing is used for transporting the mercury
vapor from the sample bottle to the absorption cell and for its
return.
D-7/Mercury ILM05.3
Exhibit D (Mercury) -- Sections 6 & 7
Reagents and Standards
6.3.5 Drying Tube - 6" X 3/4" diameter tube containing 20 grams (g) of
magnesium perchlorate.
7.1 Reagents
ILM05.3 D-8/Mercury
Exhibit D (Mercury) -- Section 7
Reagents and Standards (Con’t)
7.2 Standards
7.2.1 Introduction
D-9/Mercury ILM05.3
Exhibit D (Mercury) -- Sections 7 & 8
Sample Collection, Preservation, and Storage
The samples must be protected from light and refrigerated at 4°C (±2°C)
from the time of receipt until 60 days after delivery of a complete,
reconciled data package to USEPA. After 60 days the samples may be
disposed of in a manner that complies with all applicable regulations.
The maximum holding time for mercury is 26 days from Validated Time of
Sample Receipt (VTSR).
ILM05.3 D-10/Mercury
Exhibit D (Mercury) -- Section 9
Calibration and Standardization
9.1.1 Instruments shall be calibrated daily or once every 24 hours and each
time the instrument is set up. The instrument standardization date
and time shall be included in the raw data.
9.1.2 The date and time of preparation and analysis shall be given in the
raw data.
9.1.4 Aspirate the standards and record the readings. Results for these
standards shall be within 5% of the true value. Each standard
concentration and the calculations to show that the 5% criteria has
been met shall be given in the raw data. If the values do not fall
within this range, recalibration is necessary. The 5% criteria does
not apply to the calibration standard at the CRQL. The acceptance
criteria for the initial calibration curve is a correlation
coefficient more than or equal to 0.995.
9.2.1 Immediately after the AA system has been calibrated, the accuracy of
the initial calibration shall be verified and documented for mercury
by the analysis of the ICV solution at the wavelength used for
analysis.
9.2.2 Only if the ICV solution is not available from USEPA, or where a
certified solution of the analyte is not available from any source,
analyses shall be conducted on an independent standard at a
concentration other than that used for instrument calibration, but
within the calibration range. An independent standard is defined as
a standard composed of the analyte from a different source than that
used in the standards for the instrument calibration. The value for
the ICV shall be reported on Form IIA-IN.
9.3.1 To ensure calibration accuracy during each analysis run, one of the
following standards is to be used for the CCV and shall be analyzed
and reported at a frequency of 10% or every 2 hours during an
analysis run, whichever is more frequent. The standard shall also be
analyzed and reported at the beginning of the run and after the last
analytical sample. The analyte concentration in the CCV standard
shall be different than the concentration used for the ICV and shall
be one of the following solutions at or near the mid-range level of
the calibration curve:
D-11/Mercury ILM05.3
Exhibit D (Mercury) -- Section 9
Calibration and Standardization (Con’t)
C USEPA Solutions
C NIST Standards
The same CCV standard shall be used throughout the analysis runs for
a Sample Delivery Group (SDG) of samples received.
9.3.2 Each CCV analyzed shall reflect the conditions of analysis of all
associated analytical samples (the preceding 10 analytical samples or
the preceding analytical samples up to the previous CCV). The
duration of analysis, rinses, and other related operations that may
affect the CCV measured result may not be applied to the CCV to a
greater extent than the extent applied to the associated analytical
samples. For instance, the difference in time between a CCV analysis
and the blank immediately following it, as well as the difference in
time between the CCV and the analytical sample immediately preceding
it, may not exceed the lowest difference in time between any two
consecutive analytical samples associated with the CCV.
NOTE: A CCB shall be analyzed immediately after the last CCV, and the
last CCV shall be analyzed immediately after the last analytical sample
of the run. The results for the calibration blanks shall be reported on
Form III-IN.
ILM05.3 D-12/Mercury
Exhibit D (Mercury) -- Section 10
Procedure
10.0 PROCEDURE
10.1.1 If insufficient sample amount (less than 90% of the required amount)
is received to perform the analyses, the Contractor shall contact the
Sample Management Office (SMO) to inform them of the problem. SMO
will contact the Region for instructions. The Region will either
require that no sample analyses be performed or will require that a
reduced volume be used for the sample analysis. No other changes in
the analyses will be permitted. The Contractor shall document the
Region’s decision in the Sample Delivery Group (SDG) Narrative.
10.1.2.1 If all of the phases are not amenable to analysis (i.e., outside
the scope), the Region may require the Contractor to do any of the
following:
D-13/Mercury ILM05.3
Exhibit D (Mercury) -- Section 10
Procedure (Con’t)
ILM05.3 D-14/Mercury
Exhibit D (Mercury) -- Section 10
Procedure (Con’t)
D-15/Mercury ILM05.3
Exhibit D (Mercury) -- Section 10
Procedure (Con’t)
10.2.4.2 Feed all the reagents through the system with acid wash solution
(see Section 7.1.1.3) through the sample line, adjusting the
heating bath to 105°C.
10.2.4.5 After the analysis is complete, put all lines except the H2SO4
line in reagent water to wash out system. After flushing, wash
out the H2SO4 line. Also flush the coils in the high temperature
heating bath by pumping stannous sulfate (see Section 7.1.1.4)
through the sample lines followed by reagent water. This will
prevent build-up of oxides of manganese.
ILM05.3 D-16/Mercury
Exhibit D (Mercury) -- Section 11
11.1.2 If samples were diluted for analysis, multiply the results from the
linear regression by the dilution factor.
11.2.1 Determine the instrumental response of the unknown and determine the
mercury value from the standard curve.
11.3.1 Measure the instrumental response of the unknown and determine the
mercury value from the standard curve.
D-17/Mercury ILM05.3
Exhibit D (Mercury) -- Section 11
Data Analysis and Calculations (Con’t)
ILM05.3 D-18/Mercury
Exhibit D (Mercury) -- Section 12
Quality Control
The ICV Standard shall be prepared in the same acid matrix as the
samples and carried through the entire preparation and analysis
procedure. If measurements exceed the control limits of 80% (low) and
120% (high), the analysis shall be terminated, the problem corrected,
the instrument recalibrated, and the calibration reverified.
Information regarding the ICV shall be reported on Form IIA-IN.
12.3.1 To verify linearity near the CRQL, the Contractor shall analyze a CRI
at the beginning and end of each sample analysis run, immediately
following the ICV/ICB. In addition, the Contractor shall analyze and
report the results for the CRI at a frequency of not less than once
per 20 analytical samples1 per analysis run. The CRI analysis shall
be run immediately followed by the CCV and Continuing Calibration
Blank (CCB) analyses. The CRI shall be prepared by spiking an
aliquot of reagent water with mercury at the CRQL. The CRI shall be
taken through the same process used to digest and analyze the
associated samples.
12.3.2 CRI and percent recovery results shall be reported on Form IIB-IN.
If the percent recovery falls outside the control limits of 70-130%,
the CRI shall be re-analyzed immediately. If the result of the re-
analysis falls within the control limits, no further corrective
action is required. If the result of the re-analysis does not fall
within the control limits, the analysis shall be terminated, the
problem corrected, the instrument recalibrated, or the CRI and
associated samples redigested and analyzed.
There are two different types of blanks required by this method. The
calibration blank is used in establishing the analytical curve while the
preparation blank is used to monitor for possible contamination.
The ICB and CCB are prepared with acids and reagent water and carried
through the entire preparation and analysis procedure. If the
absolute value of the calibration blank (ICB/CCB) result exceeds the
CRQL (see Exhibit C), the analysis shall be terminated, the problem
corrected, the instrument recalibrated, the calibration verified, and
re-analysis of the preceding 10 analytical samples or all analytical
1
As defined in Exhibit G, CRI is an analytical sample.
D-19/Mercury ILM05.3
Exhibit D (Mercury) -- Section 12
Quality Control (Con’t)
12.4.2.1 The PB shall contain all the reagents and in the same volumes as
used in processing the samples. The PB shall be carried through
the complete procedure and contain the same acid concentration in
the final solution as the sample solution used for analysis.
12.4.2.4 The PB is to be reported for each SDG and used in all analyses to
ascertain whether sample concentrations reflect contamination in
the following manner:
2
A group of samples prepared at the same time.
3
USEPA may require additional spike sample analyses, upon USEPA Regional
CLP Project Officer (CLP PO) request.
ILM05.3 D-20/Mercury
Exhibit D (Mercury) -- Section 12
Quality Control (Con’t)
and its associated spike sample shall initially be run at the same
dilution.
12.5.2 If the spike analysis is performed on the same sample that is chosen
for the duplicate sample analysis, spike calculations shall be
performed using the results of the sample designated as the “original
sample” (see Section 12.6). The average of the duplicate results
cannot be used for the purpose of determining percent recovery.
Samples identified as field blanks and Performance Evaluation (PE)
samples shall not be used for spiked sample analysis. USEPA may
require that a specific sample be used for the spike sample analysis.
12.5.3 The analyte spike shall be added at 1 µg/L (water) or 0.5 mg/kg
(soil). Adjustment shall be made to maintain these spiking levels
when the weight of sample taken deviates by more than 10% of these
values.
12.5.4 If the spike recovery is not at or within the limits of 75-125%, the
data of all samples received and associated with that spike sample
and determined by the same analytical method shall be flagged with
the letter “N” on Forms IA-IN and VA-IN. An exception to this rule
is granted when the sample concentration exceeds the spike added
concentration by a factor of four or more. In such an event, the
data shall be reported unflagged even if the percent recovery does
not meet the 75-125% recovery criteria.
12.5.5 In the instance where there is more than one spike sample per matrix,
per method, per SDG, and one spike sample recovery is not within
contract criteria, flag all the samples of the same matrix and method
in the SDG. Individual component percent recoveries (%R) are
calculated as follows:
12.5.6 When sample concentration is less than the Method Detection Limit
(MDL), use SR = 0 only for purposes of calculating percent recovery.
The Spike Sample Results (SSRs), Sample Results (SRs), Spike Added
(SA), and percent recovery (positive or negative) shall be reported
on Form VA-IN.
12.5.7 The units used for reporting SSRs will be identical to those used for
reporting sample results on Form IA-IN.
12.6.1 One duplicate sample shall be analyzed from each group of samples of
a similar matrix type (i.e., water, soil) or for each SDG.4
Duplicates cannot be averaged for reporting on Form IA-IN. The
4
USEPA may require additional duplicate sample analyses, upon USEPA
Regional CLP PO request.
D-21/Mercury ILM05.3
Exhibit D (Mercury) -- Section 12
Quality Control (Con’t)
12.6.2 Duplicate sample analyses are required for percent solids. Samples
identified as field blanks and PE samples shall not be used for
duplicate sample analysis. USEPA may require that a specific sample
be used for duplicate sample analysis. The Relative Percent
Difference (RPD) is calculated as follows:
12.6.4 If one result is above five times the CRQL level and the other is
below, use the CRQL criteria to determine if the duplicate analysis
is in control. If both sample and duplicate values are less than the
MDL, the RPD is not calculated on Form VI-IN. For solid sample or
solid duplicate results less than five times the CRQL, enter the
value of the CRQL, corrected for sample weight and percent solids
(i.e., original, not duplicate sample weight and percent solids), in
the “Control Limit” column. If the duplicate sample results are
outside the control limits, flag all the data for samples received
associated with that duplicate sample with an “*” on Forms IA-IN and
VI-IN. In the instance where there is more than one duplicate sample
per SDG, if one duplicate result is not within contract criteria,
flag all samples of the same matrix and method in the SDG. The
percent difference data will be used by USEPA to evaluate the long-
term precision of the method. Specific control limits for each
element will be added to Form VI-IN at a later date based on these
precision results.
12.7.1 A solid LCS (LCSS) shall be analyzed using the same sample
preparations, analytical methods, and Quality Assurance (QA)/QC
procedures employed for the EPA samples received.
12.7.2 The USEPA provided LCSS shall be prepared and analyzed using the
procedures applied to the solid samples received (exception: percent
solids determination not required). If the USEPA LCSS is
unavailable, other USEPA QC Check samples or other certified
materials may be used. In such a case, control limits for the LCSS
must be documented and provided. One LCSS shall be prepared and
ILM05.3 D-22/Mercury
Exhibit D (Mercury) -- Section 12
Quality Control (Con’t)
analyzed for every group of solid samples in a SDG, or for each batch
of samples digested, whichever is more frequent.
12.7.3 All LCSS and percent recovery results will be reported on Form VII
IN. If the results for the LCSS fall outside the control limits
established by USEPA, the analyses shall be terminated, the problem
corrected, and the samples associated with that LCSS redigested and
re-analyzed with appropriate new QC.
12.8.1 Before any field samples are analyzed under this contract, the MDLs
shall be determined for each digestion procedure and instrument used,
prior to the start of contract analyses, and annually thereafter, and
shall meet the levels specified in Exhibit C.
12.8.2 To determine the MDLs, the Contractor shall run MDL studies following
the procedures given in 40 CFR, Part 136. The Contractor shall
prepare the MDL samples by each digestion procedure used and shall
analyze these samples on each instrument used.
12.8.3 The determined concentration of the MDL shall be less than half the
concentration of the CRQL listed in Exhibit C.
S0
S0.2
S0.5
S1.0
S2.0
S5.0
S10.0
ICV
ICB
CRI
CCV
CCB
10 samples
CCV
CCB
9 samples
CRI
CCV
CCB
10 samples, etc.
D-23/Mercury ILM05.3
Exhibit D (Mercury) -- Sections 13-17
Method Performance
Not applicable.
16.0 REFERENCES
17.0 TABLES/DIAGRAMS/FLOWCHARTS
Not applicable.
ILM05.3 D-24/Mercury
EXHIBIT D - PART D
ANALYTICAL METHODS
FOR
TOTAL CYANIDE ANALYSIS
D-1/Cyanide ILM05.3
THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.3 D-2/Cyanide
Exhibit D - Analytical Methods for Total Cyanide Analysis
Table of Contents
Section Page
3.0 DEFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4.0 INTERFERENCES . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.1 Sulfides . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.2 Surfactants . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.3 Oxidizing Agents . . . . . . . . . . . . . . . . . . . . . . . 6
5.0 SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
10.0 PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
10.1 Sample Preparation . . . . . . . . . . . . . . . . . . . . . . 14
10.2 Water and Soil Preparation of Standards and Samples . . . . . . 15
10.3 Sample Analysis . . . . . . . . . . . . . . . . . . . . . . . . 19
D-3/Cyanide ILM05.3
Exhibit D - Analytical Methods for Total Cyanide Analysis
Section Page
16.0 REFERENCES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
17.0 TABLES/DIAGRAMS/FLOWCHARTS . . . . . . . . . . . . . . . . . . . . . 30
ILM05.3 D-4/Cyanide
Exhibit D (Cyanide) –- Sections 1-3
Scope and Application
This analytical method includes the use of acid and heat to remove
cyanide from the sample.
3.0 DEFINITIONS
D-5/Cyanide ILM05.3
Exhibit D (Cyanide) -- Sections 4 & 5
Interferences
4.0 INTERFERENCES
4.1 Sulfides
4.2 Surfactants
5.0 SAFETY
ILM05.3 D-6/Cyanide
Exhibit D (Cyanide) -- Section 6
Equipment and Supplies
Brand names, suppliers, and part numbers are for illustrative purposes
achieved using equipment and supplies other than those specified here,
nanometers (nm) with a 1.0 centimeter (cm) cell or larger (for manual
spectrophotometric method).
method) including:
6.1.3.1 Sampler
6.1.3.2 Pump
6.1.3.4 Colorimeter with 15 millimeters (mm) flow cells and 580 nm filters
6.1.3.5 Recorder
6.2.3.1 Sampler
6.2.3.2 Pump
D-7/Cyanide ILM05.3
Exhibit D (Cyanide) – Section 7
Reagents and Standards
7.1 Reagents
7.1.1 Reagent water - The purity of this water must be equivalent to ASTM
Type II water (ASTM D1193-77). Use this preparation for all
reagents, standards, and dilutions of solutions.
ILM05.3 D-8/Cyanide
Exhibit D (Cyanide) – Section 7
Reagents and Standards (Con’t)
7.2 Standards
7.2.1 Introduction
7.2.2.2 Stock Standard Reagents for Midi Distillation of Water and Soils
D-9/Cyanide ILM05.3
Exhibit D (Cyanide) – Section 7
Reagents and Standards (Con’t)
7.2.4.2.1 The concentration of the CRI for cyanide shall be at the CRQL.
Information regarding the CRI shall be reported on Form IIB-IN.
ILM05.3 D-10/Cyanide
Exhibit D (Cyanide) -- Section 8
Sample Collection, Preservation, and Storage
Samples shall be kept at 4°C (±2°C) from the time of collection until
distillation.
8.2.1 The samples must be protected from light and refrigerated at 4°C
(±2°C) from the time of receipt until 60 days after delivery of a
complete, reconciled data package to the USEPA. After 60 days the
samples may be disposed of in a manner that complies with all
applicable regulations.
The maximum sample holding time for cyanide is 12 days from Validated
Time of Sample Receipt (VTSR).
D-11/Cyanide ILM05.3
Exhibit D (Cyanide) -- Section 9
Calibration and Standardization
9.3.2 The date and time of preparation and analysis shall be given in the
raw data.
9.4.1 Immediately after each cyanide system has been calibrated, the
accuracy of the initial calibration shall be verified and documented
for cyanide by the analysis of the ICV Solution at the wavelength
used for analysis.
9.4.2 Only if the ICV Solution is not available from USEPA, or where a
certified solution of the analyte is not available from any source,
analyses shall be conducted on an independent standard at a
concentration other than that used for instrument calibration, but
ILM05.3 D-12/Cyanide
Exhibit D (Cyanide) -- Section 9
Calibration and Standardization (Con’t)
9.4.3 The ICV shall be distilled. This means that an ICV must be distilled
with each batch of samples analyzed and that the samples distilled
with an ICV must be analyzed with that particular ICV.
9.4.4 The value for the ICV shall be reported on Form IIA-IN.
9.5.1 To ensure calibration accuracy during each analysis run, one of the
following standards is to be used for the CCV and shall be analyzed
and reported at a frequency of 10% or every 2 hours during an
analysis run, whichever is more frequent. The standard shall also be
analyzed and reported at the beginning of the run and after the last
analytical sample. The analyte concentration in the CCV standard
shall be different than the concentration used for the ICV and shall
be one of the following solutions at or near the mid-range level of
the calibration curve:
C USEPA Solutions
C NIST Standards
The same CCV standard shall be used throughout the analysis runs for
a Sample Delivery Group (SDG) of samples received.
9.5.2 Each CCV analyzed shall reflect the conditions of analysis of all
associated analytical samples (the preceding 10 analytical samples or
the preceding analytical samples up to the previous CCV). The
duration of analysis, rinses, and other related operations that may
affect the CCV measured result may not be applied to the CCV to a
greater extent than the extent applied to the associated analytical
samples. For instance, the difference in time between a CCV analysis
and the blank immediately following it, as well as the difference in
time between the CCV and the analytical sample immediately preceding
it, may not exceed the lowest difference in time between any two
consecutive analytical samples associated with the CCV.
NOTE: A CCB shall be analyzed immediately after the last CCV, and the
last CCV shall be analyzed immediately after the last analytical sample
of the run. The results for the calibration blanks shall be reported on
Form III-IN.
D-13/Cyanide ILM05.3
Exhibit D (Cyanide) -- Section 10
Procedure
10.0 PROCEDURE
10.1.1 If insufficient sample amount (less than 90%, of the required amount)
is received to perform the analyses, the Contractor shall contact
Sample Management Office (SMO) to inform them of the problem. SMO
will contact the Region for instructions. The Region will either
require that no sample analyses be performed or will require that a
reduced volume be used for the sample analysis. No other changes in
the analyses will be permitted. The Contractor shall document the
Region’s decision in the Sample Delivery Group (SDG) Narrative.
10.1.2.1 If all of the phases are not amenable to analysis (i.e., outside
scope), the Region may require the Contractor to do any of the
following:
10.1.3 Soil samples are not dried prior to analysis. A separate percent
solids determination must be made in accordance with the procedure in
Exhibit D - Introduction to Analytical Methods, Section 1.6.
ILM05.3 D-14/Cyanide
Exhibit D (Cyanide) -- Section 10
Procedure (Con’t)
D-15/Cyanide ILM05.3
Exhibit D (Cyanide) -- Section 10
Procedure (Con’t)
NOTE: The bubble rate will not remain constant after the
reagents have been added and while heat is being applied to the
flask. It will be necessary to re-adjust the air rate
occasionally to prevent the solution in the boiling flask from
backing up into the air inlet tube.
10.2.2.1.5 Drain the solution from the absorber into a 250 mL volumetric
flask and bring up to volume with reagent water washings from
the absorber tube.
10.2.3.1.5 Turn on the vacuum and adjust the gang (Whitney) valves to give
a flow of three bubbles per second from the impingers in each
reaction vessel.
ILM05.3 D-16/Cyanide
Exhibit D (Cyanide) -- Section 10
Procedure (Con’t)
10.2.3.1.8 Turn on the heating block and set for 123-125°C. Heat the
solution to boiling, taking care to prevent solution backup by
periodic adjustment of the vacuum flow.
10.2.3.1.9 After one and a half hours of refluxing, turn off the heat and
continue the vacuum for an additional 15 minutes. The flasks
should be cool at this time.
10.2.3.1.10 After cooling, close off the vacuum at the gang valve and
remove the absorber. Seal the receiving solutions and store
them at 4°C until analyzed. The solutions must be analyzed for
cyanide within the 12 day holding time specified in Section
8.3.
NOTE: The bubble rate will not remain constant after the
reagents have been added and while heat is being applied to the
flask. It will be necessary to re-adjust the air rate
occasionally to prevent the solution in the boiling flask from
backing up into the air inlet tube.
D-17/Cyanide ILM05.3
Exhibit D (Cyanide) -- Section 10
Procedure (Con’t)
10.2.4.1.6 Drain the solution from the absorber into a 250 mL volumetric
flask and bring up to volume with reagent water washings from
the absorber tube.
10.2.4.2.2 Weigh 1.0 g of sample (to the nearest 0.01 g) into the
distillation flask and dilute to 50 mL with reagent water. Add
2 or 3 boiling chips.
10.2.4.2.5 Turn on the vacuum and adjust the gang (Whitney) valves to give
a flow of three bubbles per second from the impingers in each
reaction vessel.
10.2.4.2.8 Turn on the heating block and set for 123-125°C. Heat the
solution to boiling, taking care to prevent solution backup by
periodic adjustment of the vacuum flow.
10.2.4.2.9 After one and a half hours of refluxing, turn off the heat and
continue the vacuum for an additional 15 minutes. The flasks
should be cool at this time.
10.2.4.2.10 After cooling, close off the vacuum at the gang valve and
remove the absorber. Seal the receiving solutions and store
them at 4°C until analyzed. The solutions must be analyzed for
cyanide within the 12 day holding time specified in
Section 8.3.
ILM05.3 D-18/Cyanide
Exhibit D (Cyanide) -- Section 10
Procedure (Con’t)
10.2.5.1.1 This code shall be used to report samples that are not
distilled prior to analysis.
10.3.2.1 Set up the manifold. Pump the reagents through the system until a
steady baseline is obtained.
10.3.2.3 When a steady reagent baseline is obtained and before starting the
sampler, adjust the baseline using the appropriate knob on the
colorimeter. Aspirate a calibration standard and adjust the
colorimeter until the desired signal is obtained. Establish the
baseline and proceed to analyze calibration standards, blanks,
control standards, distilled samples, and distilled Quality
Control (QC) samples.
D-19/Cyanide ILM05.3
Exhibit D (Cyanide) -- Section 11
Data Analysis and Calculations
WHERE,
A = µg CN read from standard curve (per 250 mL)
B = mL of original sample for distillation (see Section
10.2.2.1.1)
C = mL taken for colorimetric analysis (see Section
10.3.1.1)
50 mL = volume of original sample aliquot (see Section
10.3.1.1)
1000 mL/L = conversion mL to L
ILM05.3 D-20/Cyanide
Exhibit D (Cyanide) -- Section 11
WHERE,
A = µg CN read from standard curve (per 250 mL).
B = mL of distillate taken for colorimetric
determination (see Section 10.3.1.1).
C = wet weight of original sample in g (see Section
10.2.4.1.1).
50 mL = standard volume taken for colorimetric
determination (see Section 10.3.1.1)
% solids = percent solids (see Exhibit D - Introduction to
Analytical Methods, Section 1.6).
WHERE,
A = µg/L determined from standard curve.
C = wet weight of original sample in g (see Section
10.2.4.1.1).
.25 = conversion factor for distillate final volume
(see Section 10.2.4.1.6).
% solids = percent solids (see Exhibit D - Introduction to
Analytical Methods, Section 1.6).
The minimum value that can be substituted for A is the MDL value.
D-21/Cyanide ILM05.3
Exhibit D (Cyanide) -- Section 11
Data Analysis and Calculations (Con’t)
WHERE,
A = µg/L CN of sample from regression analysis
B = volume of original sample for distillation (0.050 L)
(see Section 10.2.3.1.2)
D = any dilution factor necessary to bracket sample value
within standard values
F = sample receiving solution volume (0.050 L)
The minimum value that can be substituted for A is the MDL value.
WHERE,
A = µg/L CN of sample from regression analysis curve
B = wet weight of original sample (see Section
10.2.4.2.2)
D = any dilution factor necessary to bracket sample
value within standard values
E = % solids/100 (see Exhibit D - Introduction to
Analytical Methods, Section 1.6)
F = sample receiving solution volume (0.050 L)
To calculate the adjusted aqueous MDL or adjusted aqueous CRQL for the
manual colorimetric method, multiply the MDL (µg/L) or CRQL (µg/L) by
ILM05.3 D-22/Cyanide
Exhibit D (Cyanide) -- Section 11
0.25 and substitute the result for the “A” term in Equation 1. To
calculate the adjusted aqueous MDL or adjusted aqueous CRQL for all
other methods, follow the instructions in Section 11.1.1 or substitute
the MDL (µg/L) or CRQL (µg/L) for the “A” term in Equation 4, as
appropriate.
The adjusted soil MDL or adjusted soil CRQL for all methods shall be
calculated as follows:
D-23/Cyanide ILM05.3
Exhibit D (Cyanide) -- Section 12
Quality Control
The ICV standard shall be prepared in the same matrix as the calibration
standards and in accordance with the instructions provided by the
supplier. The ICV standard shall be distilled. If measurements exceed
the control limits of 85% (low) and 115% (high), the analysis shall be
terminated, the problem corrected, the instrument recalibrated, and the
calibration reverified. Information regarding the ICV shall be reported
on Form IIA-IN.
12.3.1 To verify linearity near the CRQL, a standard at the CRQL (CRI) shall
be prepared, in the same matrix as the calibration standards, and
analyzed at the beginning and at the end of each sample analysis run,
immediately following the ICV/ICB. In addition, the Contractor shall
analyze the CRI at a frequency of not less than once per 20
analytical samples1 per analysis run. The CRI analysis shall be run
immediately followed by the CCV and Continuing Calibration Blank
(CCB) analyses. The CRI shall be prepared by spiking an aliquot of
reagent water with cyanide to yield a concentration in the final
solution equal to the CRQL.
12.3.2 CRI and percent recovery results shall be reported on Form IIB-IN.
If the percent recovery falls outside the control limits of 70-130%,
the CRI shall be re-analyzed immediately. If the result of the re-
analysis falls within the control limits, no further corrective
action is required. If the result of the re-analysis does not fall
within the control limits, the analysis shall be terminated, the
problem corrected, the instrument recalibrated, the CRI analyzed, and
the samples associated with the CRI re-analyzed.
There are two different types of blanks required by this method. The
calibration blank is used in establishing the analytical curve while the
preparation blank is used to monitor for possible contamination.
The ICB and CCB are prepared with reagents and reagent water. If the
absolute value of the calibration blank (ICB/CCB) result exceeds the
CRQL (see Exhibit C), the analysis shall be terminated, the problem
corrected, the instrument recalibrated, the calibration verified, and
re-analysis of the preceding 10 analytical samples or all analytical
1
As defined in Exhibit G, CRI is an analytical sample.
ILM05.3 D-24/Cyanide
Exhibit D (Cyanide) -- Section 12
Quality Control (Con’t)
12.4.2.1 The PB shall contain all the reagents and in the same volumes as
used in processing the samples. The PB shall be carried through
the complete procedure and contain the same concentration in the
final solution as the sample solution used for analysis.
12.4.2.4 The PB is to be reported for each SDG and used in all analyses to
ascertain whether sample concentrations reflect contamination in
the following manner:
2
A group of samples prepared at the same time.
D-25/Cyanide ILM05.3
Exhibit D (Cyanide) -- Section 12
Quality Control (Con’t)
matrix type (i.e., water, soil) or for each SDG.3 The sample and its
associated spike sample shall initially be run at the same dilution.
12.5.2 If the spike analysis is performed on the same sample that is chosen
for the duplicate sample analysis, spike calculations shall be
performed using the results of the sample designated as the “original
sample” (see Section 12.6). The average of the duplicate results
cannot be used for the purpose of determining percent recovery.
Samples identified as field blanks and Performance Evaluation (PE)
samples shall not be used for spiked sample analysis. USEPA may
require that a specific sample be used for the spike sample analysis.
12.5.3.1 For soil samples, the final sample solution prepared for analysis
(i.e., the distillate) shall contain cyanide spiked at a
concentration of 100 µg/L regardless of the distillation procedure
employed, or the amount of sample used for distillation. The
final sample volume after distillation shall be used as the basis
for the amount of cyanide to add as the spike. The units for
reporting soil sample cyanide results shall be mg/kg. To convert
from µg/L to mg/kg, the equation below shall be used:
12.5.4 If the spike recovery is not at or within the limits of 75-125%, the
data of all samples received and associated with that spike sample
and determined by the same analytical method shall be flagged with
the letter “N” on Forms IA-IN and VA-IN. An exception to this rule
is granted when the sample concentration exceeds the spike added
concentration by a factor of four or more. In such an event, the
data shall be reported unflagged even if the percent recovery does
not meet the 75-125% recovery criteria.
12.5.5 When the matrix spike recovery falls outside the control limits and
the sample result does not exceed 4 times the spike added, a post-
distillation spike shall be performed. Note that if a post-
distillation spike analysis is required, the same USEPA sample that
was used for the matrix spike analysis shall be used for the post
digestion spike analysis. Spike the unspiked aliquot of the sample
at 2 times the indigenous level or 2 times CRQL, whichever is
greater. Results of the post-distillation spike shall be reported on
Form VB-IN.
3
USEPA may require additional spike sample analyses, upon USEPA Regional
CLP Project Officer (CLP PO) request.
ILM05.3 D-26/Cyanide
Exhibit D (Cyanide) -- Section 12
Quality Control (Con’t)
12.5.6 In the instance where there is more than one spike sample per matrix,
per method, per SDG, if one spike sample recovery is not within
contract criteria, flag all the samples of the same matrix and method
in the SDG. Individual component percent recoveries are calculated
as follows:
WHERE,
SSR = Spiked Sample Result
SR = Sample Result
SA = Spike Added
12.5.7 When the sample concentration is less than the Method Detection Limit
(MDL), use SR = 0 only for purposes of calculating percent recovery.
The Spike Sample Results (SSRs), Sample Results (SRs), Spike Added
(SA), and percent recovery (positive or negative) shall be reported
on Form VA-IN.
12.5.8 The units used for reporting spike sample results will be identical
to those used for reporting sample results on Form IA-IN.
12.6.1 One duplicate sample shall be analyzed from each group of samples of
a similar matrix type (i.e., water, soil) or for each SDG.4
Duplicates cannot be averaged for reporting on Form IA-IN. The
sample and its associated duplicate sample shall initially be run at
the same dilution.
12.6.2 Duplicate sample analyses are required for percent solids. Samples
identified as field blanks and PE samples shall not be used for
duplicate sample analysis. USEPA may require that a specific sample
be used for duplicate sample analysis. The Relative Percent
Difference (RPD) is calculated as follows:
WHERE,
RPD = Relative Percent Difference
4
USEPA may require additional duplicate sample analyses, upon USEPA
Regional CLP PO request.
D-27/Cyanide ILM05.3
Exhibit D (Cyanide) -- Section 12
Quality Control (Con’t)
12.6.4 If one result is above five times the CRQL level and the other is
below, use the CRQL criteria to determine if the duplicate analysis
is in control. If both sample and duplicate values are less than the
MDL, the RPD is not calculated on Form VI-IN. For solid sample or
solid duplicate results less than five times the CRQL, enter the
value of the CRQL, corrected for sample weight and percent solids,
(i.e., original, not duplicate sample weight and percent solids), in
the “Control Limit” column. If the duplicate sample results are
outside the control limits, flag all the data for samples received
and associated with that duplicate sample with an “*” on Forms IA-IN
and VI-IN. In the instance where there is more than one duplicate
sample per SDG, if one duplicate result is not within contract
criteria, flag all samples of the same matrix and method in the SDG.
The percent difference data will be used by USEPA to evaluate the
long-term precision of the method. Specific control limits for each
element will be added to Form VI-IN at a later date based on the
precision results.
12.7.1 A solid LCS (LCSS) shall be analyzed using the same sample
preparations, analytical methods, and Quality Assurance (QA)/QC
procedures employed for the EPA samples received. For cyanide, a
distilled ICV shall be used as the aqueous LCS (LCSW).
12.7.2 The USEPA provided LCSS shall be prepared and analyzed using each of
the procedures applied to the solid samples received (exception:
percent solids determination not required). If the USEPA LCSS is
unavailable, other USEPA QC Check samples or other certified
materials may be used. In such a case, the control limits for LCSS
must be documented and provided. One LCSS shall be prepared and
analyzed for every group of solid samples in a SDG, or for each batch
of samples distilled, whichever is more frequent.
12.7.3 All LCSS and percent recovery results will be reported on Form VII
IN. If the results for the LCSS fall outside the control limits
established by USEPA, the analyses shall be terminated, the problem
corrected, and the samples associated with that LCSS reprepared and
re-analyzed with appropriate new QC.
12.8.1 Before any field samples are analyzed under this contract, the MDLs
shall be determined for non-distilled analyses (Preparation
Method/Code “NP1”) and for each distillation procedure and instrument
used, prior to the start of the contract analyses, and annually
thereafter, and shall meet the levels specified in Exhibit C.
12.8.2 To determine the MDLs, the Contractor shall run MDL studies following
the procedures given in 40 CFR, Part 136. The Contractor shall
ILM05.3 D-28/Cyanide
Exhibit D (Cyanide) -- Section 12
Quality Control (Con’t)
prepare the MDL samples by each distillation procedure used and shall
analyze these samples on each instrument used. The Contractor shall
also analyze the non-distilled MDL samples on each instrument used.
12.8.3 The determined concentration of the MDL shall be less than half the
concentration of the CRQL listed in Exhibit C.
12.8.5 The results of the MDL determination study shall be forwarded to the
USEPA Regional CLP PO, Sample Management Office (SMO), and Quality
Assurance Technical Support (QATS).
S0
S10.0
S50.0
S100.0
S200.0
S400.0
ICV (distilled)
ICB
CRI
CCV
CCB
MIDRANGE
9 samples
CCV
CCB
9 samples
CRI
CCV
CCB
10 samples, etc.
D-29/Cyanide ILM05.3
Exhibit D (Cyanide) – Sections 13-17
Method Performance
Not applicable.
16.0 REFERENCES
17.0 TABLES/DIAGRAMS/FLOWCHARTS
Not applicable.
ILM05.3 D-30/Cyanide