VNA User Manual
VNA User Manual
Agilent Technologies
8753ES Option 011
Network Analyzer
Agilent Technologies makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and fitness for a
particular purpose. Agilent Technologies shall not be liable for errors contained herein or
for incidental or consequential damages in connection with the furnishing, performance, or
use of this material.
Certification
Agilent Technologies certifies that this product met its published specifications at the time
of shipment from the factory. Agilent Technologies further certifies that its calibration
measurements are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institute’s calibration facility, and to the
calibration facilities of other International Standards Organization members.
Regulatory Information
The regulatory information is located in Chapter 8 , “Safety and Regulatory Information.”
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT
PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR
IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN,
INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR
INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE OR
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD
AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent Technologies products. For any assistance, contact your nearest Agilent
Technologies sales or service office. See Table 8-1 for the nearest office.
ii
Safety Notes
The following safety notes are used throughout this manual. Familiarize yourself with
each of the notes and its meaning before operating this instrument. All pertinent safety
notes for using this product are located in Chapter 8 , “Safety and Regulatory
Information.”
iii
Documentation Map
The CD-ROM provides the Installation and Quick Start Guide, the
User’s Guide, the Reference Guide, and the Programmer’s Guide in
PDF format for viewing or printing from a PC.
iv
Contents
1. Making Measurements
Using This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
More Instrument Functions Not Described in This Guide . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Making a Basic Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Step 1. Connect the device under test and any required test equipment. . . . . . . . . . . . . .1-4
Step 2. Choose the measurement parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Step 3. Perform and apply the appropriate error-correction. . . . . . . . . . . . . . . . . . . . . . . .1-5
Step 4. Measure the device under test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Step 5. Output the measurement results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Measuring Magnitude and Insertion Phase Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Measuring the Magnitude Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Measuring Insertion Phase Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Using Display Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
Titling the Active Channel Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
Viewing Both Primary Measurement Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-12
Viewing Four Measurement Channels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14
Customizing the Four-Channel Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-17
Using Memory Traces and Memory Math Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-19
Blanking the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-21
Adjusting the Colors of the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-22
Using Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-24
To Use Continuous and Discrete Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-24
To Activate Display Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-25
To Move Marker Information Off the Grids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-26
To Use Delta (∆) Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-28
To Activate a Fixed Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-29
To Couple and Uncouple Display Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-31
To Use Polar Format Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-32
To Use Smith Chart Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-33
To Set Measurement Parameters Using Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-34
Setting the CW Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-38
To Search for a Specific Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-39
To Calculate the Statistics of the Measurement Data . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-42
Measuring Electrical Length and Phase Distortion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-43
Measuring Electrical Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-43
Measuring Phase Distortion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-45
Characterizing a Duplexer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-49
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-49
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-49
Measuring Amplifiers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-53
Measuring Harmonics (Option 002). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-54
Measuring Gain Compression . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-59
Measuring Gain and Reverse Isolation Simultaneously. . . . . . . . . . . . . . . . . . . . . . . . . . .1-63
Using the Swept List Mode to Test a Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-65
Connect the Device Under Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-65
Observe the Characteristics of the Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-66
Choose the Measurement Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-67
Calibrate and Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-69
Using Limit Lines to Test a Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-71
v
Contents
vi
Contents
vii
Contents
viii
Contents
ix
Contents
x
Contents
7. Operating Concepts
Using This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Where to Find More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
System Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
The Built-In Synthesized Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
Test Sets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
The Receiver Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
The Microprocessor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
Required Peripheral Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
Processing Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7
Output Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-10
Power Coupling Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-10
Sweep Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-11
Manual Sweep Time Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-11
Auto Sweep Time Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-11
Minimum Sweep Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-11
Source Attenuator Switch Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-13
Allowing Repetitive Switching of the Attenuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-13
Channel Stimulus Coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-14
Sweep Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-15
Linear Frequency Sweep (Hz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-15
Logarithmic Frequency Sweep (Hz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-15
Stepped List Frequency Sweep (Hz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-15
Swept List Frequency Sweep (Hz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-17
Power Sweep (dBm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-19
CW Time Sweep (Seconds) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-19
Selecting Sweep Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-19
xi
Contents
S-Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
Understanding S-Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
The S-Parameter Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-22
Analyzer Display Formats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
Log Magnitude Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
Phase Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
Group Delay Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-25
Smith Chart Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-26
Polar Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-27
Linear Magnitude Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-27
SWR Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-28
Real Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-29
Imaginary Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-29
Group Delay Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-29
Electrical Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-33
Noise Reduction Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-34
Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-34
Smoothing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-35
IF Bandwidth Reduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-35
Measurement Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-37
What Is Accuracy Enhancement? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-37
What Causes Measurement Errors? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-38
Characterizing Microwave Systematic Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-41
How Effective Is Accuracy Enhancement? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-51
Calibration Routines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-54
Response Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-54
Response and Isolation Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-54
Enhanced Response Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-54
S11 and S22 One-Port Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-55
Full Two-Port Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-55
TRL*/LRM* Two-Port Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-55
E-CAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-55
Modifying Calibration Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-56
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-56
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-57
Modify Calibration Kit Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-57
Verify Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-64
Saving Modified Calibration Kits to a Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-65
Modifying and Saving a Calibration Kit from the Calibration Kit Selection Menu . . . . . 7-65
TRL*/LRM* Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-66
Why Use TRL Calibration? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-66
TRL Terminology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-67
How TRL*/LRM* Calibration Works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-67
Improving Raw Source Match and Load Match for TRL*/LRM* Calibration . . . . . . . . . 7-70
The TRL Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-71
GPIB Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-77
Local Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-77
GPIB STATUS Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-78
System Controller Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-78
xii
Contents
xiii
Contents
xiv
1 Making Measurements
1-1
Making Measurements
Using This Chapter
1-2
Making Measurements
More Instrument Functions Not Described in This Guide
1- 3
Making Measurements
Making a Basic Measurement
CAUTION Damage may result to the device under test (DUT) if it is sensitive to the
analyzer’s default output power level. To avoid damaging a sensitive DUT, be
sure to lower the output power before connecting the DUT to the analyzer.
Step 1. Connect the device under test and any required test
equipment.
Make the connections as shown in Figure 1-1.
1-4
Making Measurements
Making a Basic Measurement
NOTE You could also press the Start and Stop keys and enter the frequency
range limits as start frequency and stop frequency values.
NOTE You could also press POWER RANGE MAN POWER RANGES and select
one of the power ranges to keep the power setting within the defined range.
1- 5
Making Measurements
Making a Basic Measurement
1-6
Making Measurements
Measuring Magnitude and Insertion Phase Response
2. Press Preset and choose the measurement settings. For this example the
measurement parameters are set as follows:
Span 50 M/µ
Power −3 x1
Chan 2
1- 7
Making Measurements
Measuring Magnitude and Insertion Phase Response
If the channels are coupled (the default condition), this calibration is valid for both
channels.
4. Reconnect your test device.
5. To better view the measurement trace, press:
Scale Ref AUTO SCALE
6. To locate the maximum amplitude of the device response, as shown in Figure 1-3, press:
Marker Search SEARCH: MAX
Chan 2
Format PHASE
The channel 2 portion of Figure 1-4 shows the insertion phase response of the device under
test. The analyzer measures and displays phase over the range of −180° to +180°. As phase
changes beyond these values, a sharp 360° transition occurs in the displayed data.
1-8
Making Measurements
Measuring Magnitude and Insertion Phase Response
The phase response shown in Figure 1-5 is undersampled; that is, there is more than 180°
phase delay between frequency points. If the ∆Φ ≥ 180°, incorrect phase and delay
information may result. Figure 1-5 shows an example of phase samples being with
∆Φ less than 180° and greater than 180°.
Undersampling may arise when measuring devices with long electrical length. To correct
this problem, the frequency span should be reduced, or the number of points increased
until ∆Φ is less than 180° per point. Electrical delay may also be used to compensate for
this effect (as shown in the next example procedure).
1- 9
Making Measurements
Using Display Functions
1-10
Making Measurements
Using Display Functions
2. Press ERASE TITLE and enter the title you want for your measurement display.
• If you have a DIN keyboard attached to the analyzer, type the title you want from the
keyboard. Then press ENTER to enter the title into the analyzer. You can enter a
title that has a maximum of 50 characters. (For more information on using a
keyboard with the analyzer, refer to the “Options and Accessories” chapter in the
reference guide.)
• If you do not have a DIN keyboard attached to the analyzer, enter the title from the
analyzer front panel.
a. Turn the front panel knob to move the arrow pointer to the first character of the
title.
b. Press SELECT LETTER .
c. Repeat the previous two steps to enter the rest of the characters in your title. You
can enter a title that has a maximum of 50 characters.
d. Press DONE to complete the title entry.
CAUTION The NEWLINE and FORMFEED keys are not intended for creating display
titles. Those keys are for creating commands to send to peripherals during a
sequence program.
1- 11
Making Measurements
Using Display Functions
2. To view the measurements on separate graticules, press: Set SPLIT DISP to 2X. The
analyzer shows channel 1 on the upper half of the display and channel 2 on the lower
half of the display. The analyzer defaults to measuring S11 on channel 1 and S21 on
channel 2.
1-12
Making Measurements
Using Display Functions
NOTE You can control the stimulus functions of the two channels independent of
each other by pressing Sweep Setup COUPLED CH OFF .
NOTE For dual channel, if channels are uncoupled and you have full 2-port
calibrations on both channels, you will not be able to select a non-ratioed
measurement. For example, you can measure S21 or B/R, but not input B.
1- 13
Making Measurements
Using Display Functions
However, there are situations where the analyzer will not update all measurements
continuously. For analyzers with source attenuators, such situations occur if channel 1
requires one attenuation value and channel 2 requires a different value, or if 2-port cal is
active and the port 1 attenuation value is not equal to the attenuation value of port 2.
Since one attenuator is used for both measurements, this would cause the attenuator to
continuously switch power ranges, which is not allowed.
If one of these conditions exist, the test set hold mode will engage, and the status notation
tsH will appear on the left side of the screen. The hold mode leaves the measurement
function in only one of the two measurement paths. To update both measurements, press
Sweep Setup MEASURE RESTART . Refer to "Source Attenuator Switch Protection" on
page 7-13.
NOTE Channels 1 and 2 are referred to as primary channels and channels 3 and 4
are referred to as auxiliary channels.
Channel 3 or 4 are activated when the Chan 3 or Chan 4 keys are pressed. Alternatively,
you can enable the auxiliary setting AUX CHAN to ON. For example, if channel 1 is
active, pressing AUX CHAN to ON enables channel 3 and its trace appears on the display.
Channel 4 is similarly enabled and viewed when channel 2 is active.
1. Press Format to select the type of display of the data. This example uses the log mag
format.
2. If channel 1 is not active, make it active by pressing Chan 1 .
3. Press Display DUAL | QUAD SETUP , set DUAL CHAN to ON, set AUX CHAN to
ON, and set SPLIT DISP to 4X .
The display will appear as shown in Figure 1-9. Channel 1 is in the upper-left quadrant
of the display, channel 2 is in the upper-right quadrant, and channel 3 is in the lower
half of the display.
1-14
Making Measurements
Using Display Functions
1- 15
Making Measurements
Using Display Functions
5. Press Chan 4 .
Observe that the amber LED adjacent to the Chan 4 key is lit and the CH4 indicator
on the display has a box around it. This indicates that channel 4 is now active and can
be configured.
6. Press Marker MARKER 1 MARKER 2 .
Markers 1 and 2 appear on all four channel traces. Rotating the front panel control
knob moves marker 2 on all four channel traces. Note that the active function, in this
case the marker frequency, is the same color and in the same grid as the active channel
(channel 4).
7. Press Chan 3 .
Observe that the amber LED adjacent to the Chan 3 key is lit. This indicates that
channel 3 is now active and can be configured.
8. Rotate the front panel control knob and notice that marker 2 still moves on all four
channel traces.
1-16
Making Measurements
Using Display Functions
2X/4X Off 3 or 4 2
2X On Don’t Care
4X On 3 or 4 3
4X On Both on 4
1- 17
Making Measurements
Using Display Functions
1-18
Making Measurements
Using Display Functions
NOTE You may not be able to store 31 instrument states if they include a large
amount of calibration data. The calibration data contributes considerably to
the size of the instrument state file and therefore the available memory may
be full prior to filling all 31 registers.
• DATA/MEM (data/memory)
• DATA-MEM (data−memory)
(Note that normalization is DATA/MEM not DATA-MEM .) Memory traces are saved and
recalled and trace math is done immediately after error-correction. This means that any
data processing done after error-correction, including parameter conversion, time domain
transformation (Option 010), scaling, etc., can be performed on the memory trace. You can
also use trace math as a simple means of error-correction, although that is not its main
purpose.
All data processing operations that occur after trace math, except smoothing and gating,
are identical for the data trace and the memory trace. If smoothing or gating is on when a
memory trace is saved, this state is maintained regardless of the data trace smoothing or
gating status. If a memory trace is saved with gating or smoothing on, these features can
be turned on or off in the memory-only display mode.
The actual memory for storing a memory trace is allocated only as needed. The memory
trace is cleared on instrument preset, power on, or instrument state recall.
If sweep mode or sweep range is different between the data and memory traces, trace math
is allowed, and no warning message is displayed. If the number of points in the two traces
is different, the memory trace is not displayed nor rescaled. However, if the number of
points for the data trace is changed back to the number of points in the memory, the
memory trace can then be displayed.
If trace math or display memory is requested and no memory trace exists, the message
CAUTION: NO VALID MEMORY TRACE is displayed.
1- 19
Making Measurements
Using Display Functions
1-20
Making Measurements
Using Display Functions
1- 21
Making Measurements
Using Display Functions
NOTE Preset does not reset or change colors to the default color values. However,
cycling power to the instrument will reset the colors to the default color
values.
TEXT WARNING
To change the color of a display elements, press the softkey for that element (such as
CH1 DATA ). Then press TINT and turn the analyzer front panel knob; use the step keys
or the numeric keypad, until the desired color appears.
1-22
Making Measurements
Using Display Functions
NOTE Maximum viewing with the LCD display is achieved when primary colors or a
combination of them are selected at full brightness (100%). Table 1-2 lists the
recommended colors and their corresponding tint numbers.
NOTE Color changes and adjustments remain in effect until changed again in these
menus or the analyzer is powered off and then on again. Cycling the power
changes all color adjustments to default values. Once the colors are saved,
pressing the Preset key does not affect the color selections.
1- 23
Making Measurements
Using Markers
Using Markers
The Marker key displays a movable active marker on the screen and provides access to a
series of menus to control up to five display markers for each channel. Markers are used to
obtain numerical readings of measured values. They also provide capabilities for reducing
measurement time by changing stimulus parameters, searching the trace for specific
values, or statistically analyzing part or all of the trace.
Markers have a stimulus value (the x-axis value in a Cartesian format) and a response
value (the y-axis value in a Cartesian format). In polar format, the second part of a
complex data pair is also provided as an auxiliary response value. In Smith chart format,
the real and imaginary rectangle are both displayed, and the effective capacitance or
inductance of the imaginary part is also displayed. When a marker is activated and no
other function is active, its stimulus value is displayed in the active entry area and can be
controlled with the knob, the step keys, or the numeric keypad. The active marker can be
moved to any point on the trace, and its response and stimulus values are displayed at the
top right corner of the graticule for each displayed channel, in units appropriate to the
display format. The displayed marker response values are valid even when the measured
data is above or below the range displayed on the graticule.
• If you activate both data and memory traces, the marker values apply to the data trace.
• If you activate only the memory trace, the marker values apply to the memory trace.
• If you activate a memory math function (data/memory or data-memory), the marker
values apply to the trace resulting from the memory math function.
Marker values are normally continuous: that is, they are interpolated between measured
points. They can also be set to read only discrete measured points. Markers normally have
the same stimulus values for all channels, or they can be uncoupled so that each channel
has independent markers, regardless of whether stimulus values are coupled or dual
channel display is on.
• Press Marker Fctn MARKER MODE MENU and select one of the following choices:
1-24
Making Measurements
Using Markers
NOTE Using MARKERS: DISCRETE will also affect marker search and positioning
functions when the value entered in a search or positioning function does not
exist as a measurement point. The marker will be positioned to the closest
adjacent point that satisfies the search or positioning value.
Marker MARKER 1
The active marker is identified on the analyzer display with the following symbol: ∇
The active marker stimulus value is displayed in the active entry area. You can modify
the stimulus value of the active marker, using the front panel knob or numerical
keypad. All of the marker response and stimulus values are displayed in the upper right
corner of the display.
To switch on the corresponding marker and make it the active marker, press:
MARKER 2 MARKER 3 , MARKER 4 , or MARKER 5
All of the markers, other than the active marker, become inactive and are represented on
the analyzer display as ∆. The active and inactive markers are shown in Figure 1-13.
1- 25
Making Measurements
Using Markers
NOTE Observe that the markers appear on all of the grids. To activate markers on
individual grids, press Marker Fctn MARKER MODE MENU , and set
MARKERS: to UNCOUPLED. Then, activate the channel in which you wish
to have markers, press Marker , then select the markers for that channel.
3. Turn off the softkey menu and move the marker information off the grids by pressing
.
The display will be similar to Figure 1-14.
1-26
Making Measurements
Using Markers
Figure 1-14 Marker Information Moved into the Softkey Menu Area
pg654e
4. Restore the softkey menu and move the marker information back onto the graticules:
Press .
The display will be similar to Figure 1-15.
1- 27
Making Measurements
Using Markers
pg655e
You can also restore the softkey menu by pressing a hardkey which opens a menu (such as
Meas ) or pressing a softkey.
1-28
Making Measurements
Using Markers
1- 29
Making Measurements
Using Markers
1-30
Making Measurements
Using Markers
• Choose MARKERS: COUPLED if you want the analyzer to couple the marker stimulus
values for the display channels.
• Choose MARKERS: UNCOUPLED if you want the analyzer to uncouple the marker
stimulus values for the display channels. This allows you to control the marker
stimulus values independently for each channel.
1- 31
Making Measurements
Using Markers
NOTE For greater accuracy when using markers in the polar format, it is
recommended to activate the discrete marker mode. Press Marker Fctn
MKR MODE MENU MARKERS:DISCRETE .
• Choose LIN MKR if you want to view the magnitude and the phase of the active
marker. The magnitude values appear in units and the phase values appear in
degrees.
• Choose LOG MKR if you want to view the logarithmic magnitude and the phase of
the active marker. The magnitude values appear in dB and the phase values appear
in degrees.
• Choose Re/Im MKR if you want to view the real and imaginary pair, where the
complex data is separated into its real part and imaginary part. The analyzer shows
the real part as the first marker value (M cos Θ), and the second value is the
imaginary part (M sin Θ, where M = magnitude).
1-32
Making Measurements
Using Markers
2. Press Marker Fctn MARKER MODE MENU SMITH MKR MENU and turn the front
panel knob, or enter a value from the front panel keypad to read the resistive and
reactive components of the complex impedance at any point along the trace. This is the
default Smith chart marker.
The marker annotation tells that the complex impedance is capacitive in the bottom
half of the Smith chart display and is inductive in the top half of the display.
• Choose LIN MKR if you want the analyzer to show the linear magnitude and the
phase of the reflection coefficient at the marker.
• Choose LOG MKR if you want the analyzer to show the logarithmic magnitude and
the phase of the reflection coefficient at the active marker. This is useful as a fast
method of obtaining a reading of the log magnitude value without changing to log
magnitude format.
• Choose Re/Im MKR if you want the analyzer to show the values of the reflection
coefficient at the marker as a real and imaginary pair.
• Choose R+jX MKR to show the real and imaginary parts of the device impedance
(the series resistance and reactance, in ohms) at the marker. Also shown is the
equivalent series inductance or capacitance.
• Choose G+jB MKR to show the complex admittance values of the active marker in
rectangular form. The active marker values are displayed in terms of conductance
(in Siemens), susceptance, and equivalent parallel circuit capacitance or inductance.
Siemens are the international unit of admittance and are equivalent to mhos (the
inverse of ohms).
1- 33
Making Measurements
Using Markers
1. Press Marker Fctn and turn the front panel knob, or enter a value from the front panel
keypad to position the marker at the value that you want for the start frequency.
2. Press MARKER→START to change the start frequency value to the value of the active
marker.
1-34
Making Measurements
Using Markers
1. Press Marker Fctn and turn the front panel knob, or enter a value from the front panel
keypad to position the marker at the value that you want for the stop frequency.
2. Press MARKER→STOP to change the stop frequency value to the value of the active
marker.
1. Press Marker Fctn and turn the front panel knob, or enter a value from the front panel
keypad to position the marker at the value that you want for the center frequency.
2. Press MARKER→CENTER to change the center frequency value to the value of the
active marker.
1- 35
Making Measurements
Using Markers
NOTE Step 2 can also be performed using MKR ZERO and MARKER 1 . However,
when using this method, it will not be possible to iterate between marker zero
and marker 1.
3. Press Marker Fctn MARKER→SPAN to change the frequency span to the range
between marker 1 and marker 2.
1-36
Making Measurements
Using Markers
1. Press Marker Fctn and turn the front panel knob, or enter a value from the front panel
keypad to position the marker at the value that you want for the analyzer display
reference value.
2. Press MARKER→REFERENCE to change the reference value to the value of the
active marker.
1- 37
Making Measurements
Using Markers
2. Press Marker Fctn and turn the front panel knob, or enter a value from the front panel
keypad to position the marker at a point of interest.
3. Press MARKER→DELAY to automatically add or subtract enough line length to the
receiver input to compensate for the phase slope at the active marker position. This
effectively flattens the phase trace around the active marker. You can use this to
measure the electrical length or deviation from linear phase.
Additional electrical delay adjustments are required on devices without constant group
delay over the measured frequency span.
1-38
Making Measurements
Using Markers
Figure 1-28 Example of Searching for the Maximum Amplitude Using a Marker
1- 39
Making Measurements
Using Markers
Figure 1-29 Example of Searching for the Minimum Amplitude Using a Marker
1-40
Making Measurements
Using Markers
1. Press Marker Search and SEARCH: MAX to place the marker near the center of the
filter passband.
2. Press MKR ZERO if you want the bandwidth relative to the maximum.
3. Press Marker Search to access the marker search menu.
4. Press WIDTHS ON to calculate the center stimulus value, bandwidth, and the Q of a
bandpass or band reject shape on the measurement trace.
5. If you want to change the amplitude value (default is −3 dB) that defines the passband
or reject band, press WIDTH VALUE and enter the new value from the front panel
keypad.
1- 41
Making Measurements
Using Markers
1-42
Making Measurements
Measuring Electrical Length and Phase Distortion
2. Press Preset and choose the measurement settings. For this example, the
measurement settings include reducing the frequency span to eliminate under-sampled
phase response. Press the following keys as shown:
Span 2 M/µ
Format PHASE
1- 43
Making Measurements
Measuring Electrical Length and Phase Distortion
You may also want to select settings for the number of data points, averaging, and IF
bandwidth.
3. Substitute a thru for the device and perform a response calibration by pressing:
Cal CALIBRATE MENU RESPONSE THRU
4. Reconnect your test device.
5. To better view the measurement trace, press:
Scale Ref AUTO SCALE
Notice that in Figure 1-34 the SAW filter under test has considerable phase shift within
only a 2 MHz span. Other filters may require a wider frequency span to see the effects
of phase shift.
The linearly changing phase is due to the device’s electrical length. You can measure
this changing phase by adding electrical length (electrical delay) to compensate for it.
6. To place a marker at the center of the band, press Marker and turn the front panel
knob, or enter a value from the front panel keypad.
7. To activate the electrical delay function, press Marker Fctn MARKER→DELAY .
This function calculates and adds in the appropriate electrical delay by taking a ±10%
span about the marker, measuring the ∆Φ, and computing the delay as the negative of
∆Φ / ∆ frequency.
Alternatively, press Scale Ref ELECTRICAL DELAY and turn the front panel knob
to increase the electrical length until you achieve the best flat line, as shown in Figure
1-35.
1-44
Making Measurements
Measuring Electrical Length and Phase Distortion
The measurement value that the analyzer displays represents the electrical length of
your device relative to the speed of light in free space. The physical length of your device
is related to this value by the propagation velocity of its medium.
NOTE Velocity factor is the ratio of the velocity of wave propagation in a coaxial
cable to the velocity of wave propagation in free space. Most cables have a
relative velocity of about 0.66 the speed in free space. This velocity depends
on the relative permittivity of the cable dielectric (εr) as
1
Velocity Factor = --------
εr
You could change the velocity factor to compensate for propagation velocity by
pressing Cal MORE VELOCITY FACTOR (enter the value) x1 . This
will allow the analyzer to accurately display the equivalent distance that
corresponds to the entered electrical delay.
Figure 1-35 Example Best Flat Line with Added Electrical Delay
1- 45
Making Measurements
Measuring Electrical Length and Phase Distortion
Group Delay
The phase linearity of many devices is specified in terms of group or envelope delay. The
analyzer can translate this information into a related parameter, group delay. Group delay
is the transmission time through your device under test as a function of frequency.
Mathematically, it is the derivative of the phase response which can be approximated by
the following ratio:
−∆Φ /(360 × ∆Φ)
where ∆Φ is the difference in phase at two frequencies separated by ∆F. The quantity ∆F is
commonly called the “aperture” of the measurement. The analyzer calculates group delay
from its phase response measurements.
1-46
Making Measurements
Measuring Electrical Length and Phase Distortion
The default aperture is the total frequency span divided by the number of points across the
display (i.e. 201 points or 0.5% of the total span in this example).
1. Continue with the same instrument settings and measurements as in the previous
procedure, “Deviation From Linear Phase.”
2. To view the measurement in delay format, as shown in Figure 1-37, press:
Format DELAY Scale Ref SCALE DIV
3. To activate a marker to measure the group delay at a particular frequency, press
Marker and turn the front panel knob, or enter a value from the front panel keypad.
Group delay measurements may require a specific aperture (∆)F) or frequency spacing
between measurement points. The phase shift between two adjacent frequency points
must be less than 180°, otherwise incorrect group delay information may result.
4. To vary the effective group delay aperture from minimum aperture (no smoothing) to
approximately 1% of the frequency span, press: Avg SMOOTHING ON .
When you increase the aperture, the analyzer removes fine grain variations from the
response. It is critical that you specify the group delay aperture when you compare
group delay measurements.
1- 47
Making Measurements
Measuring Electrical Length and Phase Distortion
Group delay is calculated by dividing the phase difference between points by the frequency
spacing. Thus, if n equals the number of points, the number of phase difference values (or
frequency segments) will be n−1. The first data point is repeated so that the total number
of points remains n.
1-48
Making Measurements
Characterizing a Duplexer
Characterizing a Duplexer
This measurement example demonstrates how to characterize a 3-port device, in this case
a duplexer, using four-parameter display mode. You must use a test adapter or a special
3-port test adapter to route the signals from the analyzer (a two-port instrument) to the
duplexer (a three-port device). This example procedure is performed using one of the
following test adapters:
❏ 8753D Option K36 Duplexer Test Adapter
❏ 8753D Option K39 3-Port Test Adapter
❏ 8753ES Option H39 3-Port Test Adapter (use the same instructions as those for K39
mode)
Definitions
The following abbreviations are used in reference to a duplexer:
Tx Transmitter port
Ant Antenna port
Rx Receiver port
Procedure
1. Press Preset .
2. Connect the test adapter to the analyzer according to the instructions for your
particular model. Connect any test fixture or cables to the duplexer test adapter. Refer
to Figure 1-40.
NOTE You must have an S-parameter test set connected to your analyzer.
1- 49
Making Measurements
Characterizing a Duplexer
3. Set up channel 1 for the Tx-Ant stimulus parameters (start/stop frequency, power level,
IF bandwidth). In this example, a wide frequency range that covers both the Tx-Ant and
Ant-Rx parameters has been chosen.
4. Uncouple the primary channels from each other and then press Sweep Setup and
toggle COUPLED CH on OFF to OFF .
5. Press System CONFIGURE MENU USER SETTINGS .
6. Set up the desired mode.
• For K36 mode, toggle K36 MODE on OFF to ON. Then, press Meas
SELECT [TX-ANT] .
• For K39 mode, toggle K39 MODE on OFF to ON. Then, press Meas
SELECT PORTS [1-3] .
7. Perform a full two-port calibration on channel 1 (refer to Chapter 6 , “Calibrating for
Increased Measurement Accuracy,” if necessary).
NOTE Make sure you connect the standards to the Tx port of the test adapter (or a
cable attached to it) for FORWARD calibrations, and to the Ant port for
REVERSE calibrations.
1-50
Making Measurements
Characterizing a Duplexer
11.Set up control of the test adapter so that channels 2 and 4 are Rx:
NOTE Make sure you connect the standards to the Rx port of the test adapter (or a
cable attached to it) for FORWARD calibrations, and to the Ant port for
REVERSE calibrations.
Chan 4 Refl: REV S22 (B/R) Chan 1 Trans:FWD S21 (B/R) Chan 3
Refl: FWD S11 (A/R) , then set DUAL CHAN on OFF to ON .
The display will be similar to Figure 1-41.
1- 51
Making Measurements
Characterizing a Duplexer
Normally, a 2-port calibration requires a forward and reverse sweep to complete before the
displayed trace updates. For faster tuning, it is possible to set the number of sweeps for the
active display channel (S11 and S21 for channel 1 in this case) to update more often than
the inactive display channel. In this example we choose 8 updates of the forward
parameters to 1 update of the reverse in channel 1, and 8 updates of the reverse to 1
update of the forward in channel 2 (where the active parameters are S22 and S12).
1-52
Making Measurements
Measuring Amplifiers
Measuring Amplifiers
The analyzer allows you to measure the transmission and reflection characteristics of
many amplifiers and active devices. You can measure scalar parameters such as gain, gain
flatness, gain compression, reverse isolation, return loss (SWR), and gain drift versus time.
Additionally, you can measure vector parameters such as deviation from linear phase,
group delay, complex impedance and AM-to-PM conversion.
When you are measuring a device that is very sensitive to absolute power level, it is
important that you accurately set the power level at either the device input or output. The
analyzer is capable of using an external GPIB power meter and controlling source power
directly. Refer to Chapter 6 , "Calibrating for Increased Measurement Accuracy" for
information on power meter calibration.
This section contains the following measurement examples:
• "Measuring Harmonics (Option 002)" on page 1-54
• "Measuring Gain Compression" on page 1-59
• Measuring Gain and Reverse Isolation Simultaneously
1- 53
Making Measurements
Measuring Amplifiers
Figure 1-43 Absolute Fundamental, 2nd, and 3rd Harmonic Output Levels
1-54
Making Measurements
Measuring Amplifiers
1. Press Chan 1 Meas INPUT PORTS B to measure the power for the fundamental
frequencies.
2. Press Chan 2 Meas INPUT PORTS B to measure the power for the harmonic
frequencies.
3. Set the start frequency to a value greater than 16 MHz.
4. Press Sweep Setup and select COUPLED CH OFF . Uncoupling the channels allows
you to have the separate sweeps necessary for measuring the fundamental and
harmonic frequencies.
5. Press Power and select CHAN POWER [COUPLED] . Coupling the channel power
allows you to maintain the same fundamental frequency power level for both channels.
6. Press Power and set the power level for both channels.
7. Press Display DUAL | QUAD SETUP and select DUAL CHAN ON .
1- 55
Making Measurements
Measuring Amplifiers
To show the second harmonic’s power level relative to the fundamental power in dBc, press
Chan 2 Display MORE and select D2/D1 toD2 ON . This display mode lets you see the
relationship between the fundamental and second or third harmonic in dBc. (Refer to
Figure 1-46.)
1-56
Making Measurements
Measuring Amplifiers
Single-Channel Operation You can view the second or third harmonic alone by using
only one of the analyzer’s channels.
1- 57
Making Measurements
Measuring Amplifiers
Frequency Range The frequency range is determined by the upper frequency range of
the instrument or system (3 or 6 GHz) and by the harmonic being displayed. The 6 GHz
operation requires an 8753ES Option 006. Table 1-3 shows the highest fundamental
frequency for maximum frequency and harmonic mode.
Table 1-3 Maximum Fundamental Frequency using Harmonic Mode
3 GHz 6 GHz
Accuracy and input power Refer to the “Specifications and Characteristics” chapter in
the reference guide. The maximum recommended input power and maximum
recommended source power are related specifications.
Using power levels greater than the recommended values may cause undesired harmonics
in the source and receiver. The recommended power levels ensure that these harmonics are
less than −45 dBc. Use test port power to limit the input power to your test device.
1-58
Making Measurements
Measuring Amplifiers
1. Set up the stimulus and response parameters for your amplifier under test. To reduce
the effect of noise on the trace, press:
Avg IF BW 1000 x1
1- 59
Making Measurements
Measuring Amplifiers
4. To produce a normalized trace that represents gain compression, perform either step 5
or step 6. (Step 5 uses trace math and step 6 uses uncoupled channels and the display
function D1/D2 to D2 ON .)
5. Press Display DATA →MEMORY DATA/MEM to produce a normalized trace.
6. To produce a normalized trace, perform the following steps:
• Press Display DUAL | QUAD SETUP and select DUAL CHANNEL ON to view
both channels simultaneously.
• Press Chan 2 Meas Trans:FWD S21 (B/R)
• To uncouple the channel stimulus so that the channel power will be uncoupled,
press:
Sweep Setup COUPLED CH OFF
This will allow you to separately increase the power for channel 2 and channel 1, so
that you can observe the gain compression on channel 2 while channel 1 remains
unchanged.
• To display the ratio of channel 2 data to channel 1 data on the channel 2 display,
press:
Chan 2 DISPLAY MORE D2/D1 to D2 ON
This produces a trace that represents gain compression only.
7. Press Marker MARKER 1 and position the marker at approximately mid-span.
9. Press Power .
10.Increase the power until you observe approximately 1 dB of compression on channel 2,
using the step keys or the front panel knob.
11.To locate the worst case point on the trace, press:
Marker Search SEARCH:MIN
1-60
Making Measurements
Measuring Amplifiers
12.If COUPLED CH OFF was selected, recouple the channel stimulus by pressing:
Sweep Setup COUPLED CH ON
13.To place the marker exactly on a measurement point, press:
Marker Fctn MARKER MODE MENU MARKERS:DISCRETE
14.To set the CW frequency before going into the power sweep mode, press:
Seq SPECIAL FUNCTIONS MARKER→ CW
1- 61
Making Measurements
Measuring Amplifiers
NOTE A receiver calibration will improve the accuracy of this measurement. Refer
to Chapter 6 , “Calibrating for Increased Measurement Accuracy.”
1-62
Making Measurements
Measuring Amplifiers
Since an amplifier will have high gain in the forward direction and high isolation in the
reverse direction, the gain (S21) will be much greater than the reverse isolation (S12).
Therefore, the power you apply to the input of the amplifier for the forward measurement
(S21) should be considerably lower than the power you apply to the output for the reverse
measurement (S12). By applying low power in the forward direction, you’ll prevent the
amplifier from being saturated. A higher power in the reverse direction keeps noise from
being a factor in the measurement and accounts for any losses caused by attenuators or
couplers on the amplifier’s output needed to lower the output power into the analyzer. The
following steps demonstrate the features that best accomplish these measurements.
3. Press Chan 1 Meas Trans:FWD S21 (B/R) Power and set the power level for
port 1.
4. Press Chan 2 Meas Trans: REV S12 (A/R) Power and set the power level for
port 2.
5. Perform an error-correction and connect the amplifier to the network analyzer. Refer to
the Chapter 5 , “Optimizing Measurement Results,” for error-correction procedures.
6. Press Display DUAL | QUAD SETUP DUAL CHAN ON .
You can view both measurements simultaneously by using the dual channel display
mode. Refer to Figure 1-51. If the port power levels are in different power ranges, one of
the displayed measurements will not be continually updated and the annotation tsH
will appear on the left side of the display. Refer to "Source Attenuator Switch
Protection" on page 7-13 for information on how to override this state.
NOTE To obtain best accuracy, you should set the power levels prior to performing
the calibration. However, the analyzer compensates for nominal power
changes you make during a measurement, so that the error correction still
remains quite valid. In these cases, the Cor annunciator will change to C∆.
1- 63
Making Measurements
Measuring Amplifiers
1-64
Making Measurements
Using the Swept List Mode to Test a Device
1- 65
Making Measurements
Using the Swept List Mode to Test a Device
• Generally, the passband of a filter exhibits low loss. A relatively low incident power may
be needed to avoid overdriving the next stage of the DUT (if that stage contains an
amplifier) or the network analyzer receiver.
1-66
Making Measurements
Using the Swept List Mode to Test a Device
• Conversely, the stopband of a filter generally exhibits high isolation. To measure this
characteristic, the dynamic range of the system will have to be maximized. This can be
done by increasing the incident power and narrowing the IF bandwidth.
NUMBER of POINTS 51 x1
4. To maximize the dynamic range in the stopband (increasing the incident power and
narrowing the IF bandwidth), press
MORE
RETURN DONE
SPAN 40 M/µ
1- 67
Making Measurements
Using the Swept List Mode to Test a Device
6. To specify a lower power level and a wider IF bandwidth for the passband, press
MORE
SEGMENT IF BW 3700 x1
RETURN DONE
NUMBER of POINTS 51 x1
8. To maximize the dynamic range in the stopband (increasing the incident power and
narrowing the IF bandwidth), press:
MORE
SEGMENT POWER 10 x1
SEGMENT IF BW 300 x1
RETURN DONE
1-68
Making Measurements
Using the Swept List Mode to Test a Device
3. Reconnect the filter and adjust the scale to compare results with the first filter
measurement that used a linear sweep.
In Figure 1-55, notice that the noise level has decreased over 10 dB, confirming that the
noise reduction techniques in the stopbands were successful. Also, notice that the
stopband noise in the third segment is slightly lower than in the first segment. This is
due to the narrower IF bandwidth of the third segment (300 Hz).
1- 69
Making Measurements
Using the Swept List Mode to Test a Device
Figure 1-55 Filter Measurements Using Linear Sweep and Swept List Mode
Using Linear Sweep
(Power: 0 dBm/IF BW: 3700 Hz)
1-70
Making Measurements
Using Limit Lines to Test a Device
2. Press Preset and choose the measurement settings. For this example the
measurement settings are as follows:
• Span 50 M/µ
1- 71
Making Measurements
Using Limit Lines to Test a Device
You may also want to select settings for the number of data points, power, averaging,
and IF bandwidth.
3. Substitute a thru for the device and perform a response calibration by pressing:
Cal CALIBRATE MENU RESPONSE THRU
4. Reconnect your test device.
5. To better view the measurement trace, press:
Scale Ref AUTO SCALE
1. To access the limits menu and activate the limit lines, press:
System LIMIT MENU LIMIT LINE LIMIT LINE ON EDIT LIMIT LINE
CLEAR LIST YES
2. To create a new limit line, press:
ADD
The analyzer generates a new segment that appears on the center of the display.
3. To specify the limit’s stimulus value, test limits (upper and lower), and the limit type,
press:
STIMULUS VALUE 127 M/µ
UPPER LIMIT −21 x1
LOWER LIMIT −27 x1 DONE
NOTE You could also set the upper and lower limits by using the MIDDLE VALUE
and DELTA LIMITS keys. To use these keys for the entry, press:
MIDDLE VALUE −24 x1 DELTA LIMITS 3 x1
This would correspond to a test specification of −24 ±3 dB.
1-72
Making Measurements
Using Limit Lines to Test a Device
• To create a limit line that tests the low side of the filter, press:
ADD
DONE
ADD
DONE
1- 73
Making Measurements
Using Limit Lines to Test a Device
• To create a limit line that tests the high side of the bandpass filter, press:
ADD
DONE
ADD
DONE
1-74
Making Measurements
Using Limit Lines to Test a Device
1. To access the limits menu and activate the limit lines, press:
System LIMIT MENU LIMIT LINE LIMIT LINE ON EDIT LIMIT LINE
CLEAR LIST YES
2. To establish the start frequency and limits for a sloping limit line that tests the low side
of the filter, press:
ADD STIMULUS VALUE 123 M/µ
1- 75
Making Measurements
Using Limit Lines to Test a Device
1-76
Making Measurements
Using Limit Lines to Test a Device
System LIMIT MENU LIMIT LINE LIMIT LINE ON EDIT LIMIT LINE
1- 77
Making Measurements
Using Limit Lines to Test a Device
NOTE Selecting the beep fail indicator BEEP FAIL ON is optional and will add
approximately 50 ms of sweep cycle time. Because the limit test will still work
if the limits lines are off, selecting LIMIT LINE ON is also optional.
The limit test results appear on the right side on the analyzer display. The analyzer
indicates whether the filter passes or fails the defined limit test:
• The message FAIL will appear on the right side of the display if the limit test fails.
• The analyzer beeps if the limit test fails and if BEEP FAIL ON has been selected.
• The analyzer changes the color of the trace to flashing red where the measurement
trace is out of limits.
• A TTL signal on the rear panel BNC connector "LIMIT TEST" provides a pass/fail
(5 V/0 V) indication of the limit test results.
1-78
Making Measurements
Using Limit Lines to Test a Device
1- 79
Making Measurements
Using Ripple Limits to Test a Device
1-80
Making Measurements
Using Ripple Limits to Test a Device
2. Press Preset and choose the measurement settings. For this example, the
measurement settings are as follows:
1- 81
Making Measurements
Using Ripple Limits to Test a Device
Notice that Frequency Band 1 overlaps in frequency the remaining frequency bands.
Whereas, Frequency Bands 2 and 3 are separate bands that cover the same span of
frequency. This can be done to put tighter limits over narrower frequency spans within the
bandpass or to customize the ripple test to meet your specific requirements.
1-82
Making Measurements
Using Ripple Limits to Test a Device
2. To access the ripple test edit menu, press EDIT RIPL LIMIT .
3. Add the first frequency band (Frequency Band 1) to be tested by pressing ADD .
4. Set the lower frequency value of Frequency Band 1 by pressing:
MINIMUM FREQUENCY 500 M/µ
5. Set the upper frequency value of Frequency Band 1 by pressing:
MAXIMUM FREQUENCY 3.2 G/n
6. Set the maximum allowable ripple amplitude value of Frequency Band 1 by pressing:
MAXIMUM RIPPLE 2.0 x1
7. Repeat steps 3 through 6 for the two remaining frequency bands to be tested for
maximum ripple.
The network analyzer allows you to enter up to 12 frequency bands to be tested for
maximum ripple.
8. After you have entered all of the ripple test frequency band parameters, return to the
ripple test menu by pressing DONE .
a. FREQUENCY BAND
b. The numeric key indicating the frequency band number that you are changing.
The frequency band number is located in the left column of the list of frequency
bands.
1- 83
Making Measurements
Using Ripple Limits to Test a Device
a. MINIMUM FREQUENCY and the new value to change the lower frequency of the
frequency band.
b. MAXIMUM FREQUENCY and the new value to change the upper frequency of the
frequency band.
c. MAXIMUM RIPPLE and the new decibel value to change the maximum allowable
ripple of the frequency band. Terminate the new decibel value with the x1 key.
4. Repeat steps 2 and 3 for additional frequency bands.
5. After you have entered the necessary changes to the ripple test frequency band
parameters, return to the ripple test menu by pressing DONE .
1. To access the ripple test edit menu, press EDIT RIPL LIMIT .
2. Create a new frequency band by pressing ADD
3. Set the lower frequency value of the frequency band by pressing:
a. MINIMUM FREQUENCY
b. the numeric keys indicating the minimum frequency value of the frequency band
c. the appropriate frequency key (either G/n , M/µ , or k/m )
4. Set the upper frequency value of the frequency band by pressing:
a. MAXIMUM FREQUENCY
b. the numeric keys indicating the maximum frequency value of the frequency band
c. the appropriate frequency key (either G/n , M/µ , or k/m )
5. Set the maximum allowable ripple amplitude value of the frequency band by pressing:
a. MAXIMUM RIPPLE
b. the decibel value of the frequency band’s maximum allowable ripple
c. x1
6. Repeat steps 2 through 5 for additional frequency bands to be tested for maximum
ripple.
7. After you have added all of the new frequency bands, return to the ripple test menu by
pressing DONE .
1-84
Making Measurements
Using Ripple Limits to Test a Device
1- 85
Making Measurements
Using Ripple Limits to Test a Device
As the analyzer measures the ripple, a message is displayed indicating whether the
measurement passes or fails:
• If the ripple test passes, a RIPLn PASS message (where n = the channel number) is
displayed in the color assigned to Channel 1 Memory. The ripple test must pass in all
frequency bands before the pass message is displayed.
• If the ripple test fails, a RIPLn FAIL message (where n = the channel number) is
displayed in red. The portion of the trace that exceeds the user-specified maximum
ripple value is also displayed in red.
1-86
Making Measurements
Using Ripple Limits to Test a Device
• If the ripple test fails, the ripple limits are drawn on the display for each frequency
band. Within each frequency band, the lower ripple limit is drawn at the lowest point on
the measured trace and the upper ripple limit is drawn at the user-specified maximum
ripple value above the lower ripple limit. The ripple that exceeds the maximum ripple
value extends above the upper limit. This measured trace that extends above the upper
limit is displayed in red.
Figure 1-66 shows the filter pass band tested with the ripple limits activated. Notice that
there are three sets of ripple limits shown. Also notice that the measured trace exceeds the
upper ripple limit only in Frequency Band 3.
Figure 1-66 Filter Pass Band with Ripple Test and Ripple Limits Activated
Changing the Ripple Limits Line Color. The color of the lines that represent the
ripple limits can be changed by:
3. pressing RIPPLE LIM LINES TINT and turning the analyzer front panel knob until
the desired color appears (You may also use the step keys or the numeric keypad
instead of the front panel knob to change the color.)
1- 87
Making Measurements
Using Ripple Limits to Test a Device
To display the ripple value, press RIPL VALUE [ ] . Pressing this softkey toggles
between RIPL VALUE [OFF ] , RIPL VALUE [ABSOLUTE ] , and
RIPL VALUE [MARGIN ] . RIPL TEST on OFF from the Ripple Test Menu until ON is
displayed on the softkey. Pressing this softkey toggles the analyzer between ripple test on
and ripple test off status.
When the Absolute and Margin choices are selected, the frequency band and measurement
value are displayed to the right side of the pass/fail message described previously. This
display is displayed in the same color as the pass/fail message.
The frequency band of the displayed value is displayed as Bn (where n = the frequency
band number). The frequency band may be changed to display the value of each band. To
change the displayed frequency band value, from the Ripple Test Menu press
RIPL VALUE BAND and then use the and keys (or the numerical keypad) to
select the desired frequency band.
1-88
Making Measurements
Using Ripple Limits to Test a Device
Figure 1-67. Filter Pass Band with Absolute Ripple Value for Band 1 Activated
1- 89
Making Measurements
Using Ripple Limits to Test a Device
Figure 1-68 shows the ripple test with margin ripple value displayed for Frequency
Band 2. Notice that Frequency Band 2 passes the ripple test with a margin of 0.097 dB.
The plus sign (+) indicates this band passes the ripple test by the amount displayed. A
minus sign (−) would indicate that the band failed by the displayed amount.
Figure 1-68 Filter Pass Band with Margin Ripple Value for Band 2 Activated
1-90
Making Measurements
Using Bandwidth Limits to Test a Bandpass Filter
1- 91
Making Measurements
Using Bandwidth Limits to Test a Bandpass Filter
2. Press Preset and choose the measurement settings. For this example, the
measurement settings are as follows:
1-92
Making Measurements
Using Bandwidth Limits to Test a Bandpass Filter
3. Substitute a thru for the device and perform a response calibration by pressing:
Cal CALIBRATE MENU RESPONSE THRU
4. Reconnect your test device.
Refer to Figure 1-71.
1- 93
Making Measurements
Using Bandwidth Limits to Test a Bandpass Filter
1. Start the bandwidth test by pressing the BW TEST on OFF softkey until ON is
displayed.
The bandwidth test continues to run until the softkey is returned to the OFF position.
The test displays a message in the upper left corner of the graticule showing that the
bandwidth test is being performed and the channel on which the test is being
performed. For example, BW1: indicates that the bandwidth test is being run on
channel 1. See Figure 1-72.
The test also displays a message indicating whether the filter passes or fails the
bandwidth test. When the filter is passing the test, the message indicates Pass. When
the filter is failing the test, the failure message indicates either Wide (when the pass
band is wider than the maximum bandwidth input) or Narrow (when the pass band is
narrower than the minimum bandwidth input).
When the filter passes the bandwidth test, the color of the bandwidth test Pass message
is green. When the filter fails the bandwidth test, the color of the bandwidth test
Wide/Narrow message is red.
1. Display the bandwidth markers by pressing the BW MARKER on OFF softkey until
ON is displayed on the softkey.
When the bandwidth markers are displayed, a marker is placed on each side of the peak
amplitude at a position equal to the N dB Points value below the peak. The markers are
placed at the 40 dB points on the signal in Figure 1-73. The bandwidth markers
resemble the following symbol: T
1-94
Making Measurements
Using Bandwidth Limits to Test a Bandpass Filter
1. Display the bandwidth value by pressing the BW DISPLAY on OFF softkey until ON
is displayed on the softkey.
When this softkey is set to the ON position, the measured bandwidth value is displayed
in the upper left corner of the display, to the right of the bandwidth Pass/Wide/Narrow
message. This value changes as the analyzer continues measuring the bandwidth. The
bandwidth value is displayed in Figure 1-74.
If the filter is failing the bandwidth test, the color of the bandwidth value is red, the
same color as the failure (Wide) message of Figure 1-72. If the filter is passing the
bandwidth test, the displayed bandwidth value is green (the same color as the
bandwidth test Pass message).
1- 95
Making Measurements
Using Bandwidth Limits to Test a Bandpass Filter
1-96
Making Measurements
Using Test Sequencing
Creating a Sequence
1. To enter the sequence creation mode, press:
Seq NEW SEQ/MODIFY SEQ
As shown in Figure 1-75, a list of instructions appear on the analyzer display to help
you create or edit a sequence.
1- 97
Making Measurements
Using Test Sequencing
Span 50 M/µ
1-98
Making Measurements
Using Test Sequencing
CAUTION When you create a sequence, the analyzer stores it in volatile memory where
it will be lost if you switch off the instrument power (except for sequence #6
which is stored in the analyzer non-volatile memory). However, you may store
sequences to a floppy disk.
Running a Sequence
To run a stored test sequence, press:
Preset and the softkey labeled with desired sequence number.
or, press:
Seq DO SEQUENCE and the softkey labeled with the desired sequence number
Stopping a Sequence
To stop a sequence before it has finished, press Local .
Editing a Sequence
Deleting Commands
1. To enter the creation/editing mode, press:
Seq NEW SEQ/MODIFY SEQ
2. To select the particular test sequence you wish to modify (sequence 1 in this example),
press:
SEQUENCE 1 SEQ1
1- 99
Making Measurements
Using Test Sequencing
3. To move the cursor to the command that you wish to delete, press:
or
• If you wish to scroll through the sequence without executing each line as you do so,
you can press the key and scroll through the command list backwards.
• If you use the key to move the cursor through the list of commands, the
commands are actually performed when the cursor points to them. This feature
allows the sequence to be tested one command at a time.
4. To delete the selected command, press:
(backspace key)
5. Press Seq DONE SEQ MODIFY to exit the modify (edit) mode.
Inserting a Command
1. To enter the creation/editing mode, press:
Seq NEW SEQ/MODIFY SEQ
2. To select the particular test sequence you wish to modify (sequence 1 in this example),
press:
SEQUENCE 1 SEQ1
3. To insert a command, move the cursor to the line immediately above the line where you
want to insert a new command, by pressing:
or
• If you use the key to move the cursor through the list of commands, the
commands are actually performed when the cursor points to them. This feature
allows the sequence to be tested one command at a time.
• If you wish to scroll through the sequence without executing each line as you do so,
you can press the key and scroll through the command list backwards.
4. To enter the new command, press the corresponding analyzer front panel keys. For
example, if you want to activate the averaging function, press:
Avg AVERAGING ON
5. Press Seq DONE SEQ MODIFY to exit the modify (edit) mode.
Modifying a Command
1. To enter the creation/editing mode, press:
Preset Seq NEW SEQ/MODIFY SEQ
2. To select the particular test sequence you wish to modify, (sequence 1 in this example),
press:
SEQUENCE 1 SEQ1
1-100
Making Measurements
Using Test Sequencing
The following list is the commands entered in "Creating a Sequence" on page 1-97.
Notice that for longer sequences, only a portion of the list can appear on the screen at
one time.
Start of Sequence
RECALL PRST STATE
Trans: FWD S21 (B/R)
LOG MAG
CENTER
134 M/u
SPAN
50 M/u
SCALE/DIV
AUTO SCALE
3. To change a command (for example, the span value from 50 MHz to 75 MHz), move the
cursor (→) next to the command that you wish to modify, press:
or
• If you use the key to move the cursor through the list of commands, the
commands are actually performed when the cursor points to them. This feature
allows the sequence to be tested one command at a time.
• If you wish to scroll through the sequence without executing each line as you do so,
you can press the key and scroll through the command list backwards.
4. To delete the current command (for example, span value), press:
6. Press Seq DONE SEQ MODIFY to exit the modify (edit) mode.
1- 101
Making Measurements
Using Test Sequencing
Seq MORE TITLE SEQUENCE and select the particular sequence softkey.
The analyzer shows the available title characters. The current title is displayed in the
upper-left corner of the screen.
2. You can create a new file name in two ways:
• If you have an attached DIN keyboard, you can press the f6 function key on the
keyboard and type the new file name.
• If you do not have an attached DIN keyboard, press ERASE TITLE and turn the
front panel knob to point to the characters of the new file name, pressing
SELECT LETTER as you stop at each character.
The analyzer cannot accept a title (file name) that is longer than eight characters. Your
titles must also begin with a letter, and contain only letters and numbers.
3. To complete the titling, press DONE .
• Save/Recall
• FILE UTILITIES
• SEQUENCE FILENAMING
This menu presents two choices:
• FILE NAME FILE0 supplies a name for the saved state or data file. This also brings
up the Title File Menu.
• PLOT NAME PLOTFILE supplies a name for the plot file generated by a plot-to-disk
command. This also brings up the Title File Menu.
These keys show the current file name in the 2nd line of the softkey.
When titling a file for use in a loop function, you are restricted to only 2 characters in the
file name due to the 6 character length of the loop counter keyword “[LOOP].” When the
file is actually written, the [LOOP] keyword is expanded to only 5 ASCII characters
(digits), resulting in a 7 character file name.
After entering the 2 character file name, press:
LOOP COUNTER DONE
1-102
Making Measurements
Using Test Sequencing
CAUTION The analyzer will overwrite a file on the disk that has the same title.
CAUTION Do not mistake the line switch for the disk eject button.
• If the desired sequence is not among the first six files, press:
READ SEQ FILE TITLS until the desired file name appears.
2. Press the softkey next to the title of the desired sequence. The disk access light should
illuminate briefly.
NOTE If you know the title of the desired sequence, you can title the sequence (1–6)
with the name, and load the sequence. This is also how you can control the
sequence number of an imported titled sequence.
• If the desired sequence is not among the first six files, press:
READ SEQ FILE TITLS until the desired file name appears.
2. Press the softkey next to the title of the desired sequence. The disk access light should
illuminate briefly.
1- 103
Making Measurements
Using Test Sequencing
Printing a Sequence
1. Configure a compatible printer to the analyzer. (Refer to the “Options and Accessories”
chapter of the reference guide.)
2. To print a sequence, press:
Seq MORE PRINT SEQUENCE and the softkey for the desired sequence.
NOTE If the sequence is on a disk, load the sequence (as described in a previous
procedure) and then follow the printing sequence.
(backspace key)
1-104
Making Measurements
Using Test Sequencing
Titles
A title may contain non-printable or special ASCII characters if you download it from an
external controller. A non-printable character is represented on the display as π.
Sequence Size
A sequence may contain up to 2 kbytes of instructions. Typically, this is around 200
sequence command lines. To estimate a sequence's size (in kbytes), use the following
guidelines.
Table 1-5 Guidelines for Determining the Size of a Sequence
Typical command 2
Autostarting Sequences
You can define a sequence to run automatically when you apply power to the analyzer. To
make an autostarting sequence, create a sequence in position six (SEQ6) and title it
“AUTO”. To stop an autostarting sequence, press Local . To stop an autostarting sequence
from engaging at power on, you must clear it from memory or rename it.
NOTE Presetting the instrument does not run the Auto Sequence automatically.
1- 105
Making Measurements
Using Test Sequencing
TESTSET I/O
The TESTSET I/O interconnect on the rear panel was originally intended for use solely
with the HP/Agilent 85046A/B and HP/Agilent 85047B external S-parameter test sets.
Since the introduction of the 8753D, a network analyzer with an internal test set, this test
set I/O port has become a general purpose control port for a variety of external devices,
such as the K36 or K39 test adapters and Option 014 configurations. Refer Table 1-6 on
page 1-110 for the definition of each pin of the test set I/O connector.
CAUTION +22 volts is available on the TESTSET I/O connector. Be careful not to
connect this to a printer port or to sensitive electronic equipment.
This connector, with the limit output, TTL OUT and TESTSET I/O outputs can also be
used with part handlers to provide control interface.
The TESTSET I/O bits are set using the TESTSET I/O FWD and TESTSET I/O REV keys
under the Seq TTL I/O TTL OUT keys. The values of the outputs (pins 11, 22, and 23)
are described in Table 1-6. The value changes with the test port, so if the external control
is required for both test port directions, the settings must be made under both
TESTSET I/O FWD and TESTSET I/O REV . This capability can be used to set different
external conditions in a test requiring changes between the forward and reverse
measurements, as might be needed in a high power test, for example.
1-106
Making Measurements
Using Test Sequencing
TTL Output for Controlling Peripherals Eight TTL compatible output lines can be
used for controlling equipment connected to the parallel port. By pressing Seq TTL I/O ,
you will access the following softkeys that control the individual output bits. Refer to
Figure 1-76 for output bus pin locations.
PARALLEL OUT ALL lets you input a number (0 to 255) in base 10 and outputs it to
the bus as binary.
SET BIT lets you set a single bit (0 − 7) to high on the output bus.
CLEAR BIT lets you set a single bit (0 − 7) to low on the output bus.
TTL Input Decision Making Five TTL compatible input lines can be used for decision
making in test sequencing. For example, if a test fixture is connected to the parallel port
and has a micro switch that needs to be activated in order to proceed with a measurement,
you can construct your test sequence so that it checks the TTL state of the input line
corresponding to the switch. Depending on whether the line is high or low, you can jump to
another sequence. To access these decision making functions, press Seq TTL I/O . Refer
to Figure 1-76 for input bus pin locations.
PARALL IN BIT NUMBER lets you select the single bit (0 − 4) that the sequence will
be looking for.
PARALL IN IF BIT H lets you jump to another sequence if the single input bit you
selected is in a high state.
PARALL IN IF BIT L lets you jump to another sequence if the single input bit you
selected is in a low state.
Pin assignments:
• pin 1 is the data strobe
• pin 16 selects the printer
• pin 17 resets the printer
• pins 18-25 are ground
Electrical specifications for TTL high:
• volts(H) ≥ 2.7 volts (V)
• current = 20 microamps (µA)
Electrical specifications for TTL low:
• volts(L) ≤ 0.4 volts (V)
• current = 0.2 milliamps (mA)
1- 107
Making Measurements
Using Test Sequencing
Figure 1-76 Parallel Port Input and Output Bus Pin Locations in GPIO Mode
1-108
Making Measurements
Using Test Sequencing
Control of the external switch (8762B Option T24) can be done through the test set
interface on the rear panel of the analyzer.
Pin 22 (TTL 1) on the TEST SET-I/O INTERCONNECT connector is a TTL line that
changes from TTL high to TTL low when changing TTL I/O FWD from 7 to 6. Refer to
Figure 1-77. To change from 7 to 6, press the following sequence:
1- 109
Making Measurements
Using Test Sequencing
Pin 2 Sweep delay: holds off sweeps until test set has finished sweeping (85046A/B and 85047B
only)
Pin 4 NC
Pin 5 NC
Pin 6 NC
Pin 7 Ground
Pin 9 NC
Pin 11 TESTSET I/O Bit 2 (most significant bit). +5 V when TESTSET I/O has values of 4, 5, 6, or
7. Otherwise, bit is TTL low.
Pin 12 Ground
Pin 13 NC
Pin 15 NC
Pin 16 NC
Pin 18 Ground
Pin 19 NC
Pin 20 NC
Pin 21 Ground
Pin 22 TESTSET I/O Bit 0 (least significant bit). +5 V when TESTSET I/O has values of 1, 3, 5, or
7. Otherwise, bit is TTL low.
Pin 23 TESTSET I/O Bit 1 (middle bit). +5 V when TESTSET I/O has values of 2, 3, 6, or 7.
Otherwise, bit is TTL low.
Pin 24 Lremtrig: TTL low when TEST- SET I/O pins are valid. This bit can be used to latch these
values.
Pin 25 NC
1-110
Making Measurements
Using Test Sequencing
TTL Out Menu The TTL OUT softkey provides access to the TTL out menu. This menu
allows you to choose between the following output parameters of the TTL output signal:
Having a sequence jump to itself A decision making command can jump to the
sequence it is in. When this occurs, the sequence starts over and all commands in the
sequence are repeated. This is used a great deal in conjunction with loop counter
commands. See the loop counter description that follows.
TTL input decision making TTL input from a peripheral connected to the parallel port
(in the GPIO mode) can be used in a decision making function. Refer to "The GPIO Mode"
on page 1-106.
Limit test decision making A sequence can jump to another sequence or start over
depending on the result of a limit test. When entered into a sequence, the
IF LIMIT TEST PASS and IF LIMIT TEST FAIL commands require you to enter the
destination sequence.
1- 111
Making Measurements
Using Test Sequencing
Loop counter decision making The analyzer has a numeric register called a loop
counter. The value of this register can be set by a sequence, and it can be incriminated or
decremented each time a sequence repeats itself. The decision making commands
IF LOOP COUNTER = 0 and IF LOOP COUNTER <> 0 jump to another sequence if the
stated condition is true. When entered into the sequence, these commands require you to
enter the destination sequence. Either command can jump to another sequence, or restart
the current sequence.
As explained in "Embedding the Value of the Loop Counter in a Title" on page 1-105, the
loop counter value can be appended to a title. This allows customized titles for data
printouts or for data files saved to disk.
1-112
Making Measurements
Using Test Sequencing to Test a Device
Span 50 M/µ
1- 113
Making Measurements
Using Test Sequencing to Test a Device
LOOP COUNTER 10 x1
1-114
Making Measurements
Using Test Sequencing to Test a Device
LOOP COUNTER 7 x1
1- 115
Making Measurements
Using Test Sequencing to Test a Device
Copy PLOT
SEQUENCE 2 SEQ 2
Start of Sequence
FILE NAME
DT[LOOP]
PLOT NAME
PL[LOOP]
SINGLE
SAVE FILE 0
PLOT
DECR LOOP COUNTER
IF LOOP COUNTER <> 0 THEN DO
SEQUENCE 2
Sequence 1 initializes the loop counter and calls sequence 2. Sequence 2 repeats until the
loop counter reaches 0. For each loop, it takes a single sweep, saves the data file and plots
the display.
• The data file names generated by this sequence will be:
DT00007.D1
through
DT000001.D1
1-116
Making Measurements
Using Test Sequencing to Test a Device
1- 117
Making Measurements
Using Test Sequencing to Test a Device
T U N E D E V I C E DONE
1-118
2 Making Mixer Measurements
2-1
Making Mixer Measurements
Using This Chapter
2-2
Making Mixer Measurements
Mixer Measurement Capabilities
NOTE This chapter uses the following 3 terms when referring to mixer signals:
LO Local Oscillator. LO is normally provided by an external source or
internally generated by the frequency converter.
IF Intermediate Frequency. IF is usually the mixer’s output signal.
RF Radio Frequency. RF is usually the mixer’s input signal.
2- 3
Making Mixer Measurements
Measurement Considerations
Measurement Considerations
In mixer transmission measurements, you have RF and LO inputs and an IF output. Also
emanating from the IF port are several other mixing products of the RF and LO signals. In
mixer measurements, leakage signals from one mixer port propagate and appear at the
other two mixer ports. These unwanted mixing products or leakage signals can cause
distortion by mixing with a harmonic of the analyzer’s first down-conversion stage.
To ensure successful mixer measurements, the following measurement challenges must be
taken into consideration:
• Mixer Considerations
❏ “Minimizing Source and Load Mismatches”
❏ "Reducing the Effect of Spurious Responses" on page 2-5
❏ "Eliminating Unwanted Mixing and Leakage Signals" on page 2-6
• Analyzer Operation
❏ "How RF and IF Are Defined" on page 2-7
❏ "Frequency Offset Mode Operation" on page 2-10
❏ "LO Frequency Accuracy and Stability" on page 2-10
❏ "Power Meter Calibration" on page 2-10
2-4
Making Mixer Measurements
Measurement Considerations
2- 5
Making Mixer Measurements
Measurement Considerations
2-6
Making Mixer Measurements
Measurement Considerations
Figure 2-5 Example of Conversion Loss versus Output Frequency with Correct
IF Signal Path Filtering and Attenuation at All Mixer Ports
2- 7
Making Mixer Measurements
Measurement Considerations
In standard mixer measurements, the input of the mixer is always connected to the
analyzer’s RF source, and the output of the mixer always produces the IF frequencies that
are received by the analyzer’s receiver.
However, the ports labeled RF and IF on most mixers are not consistently connected to the
analyzer’s source and receiver ports, respectively. These mixer ports are switched,
depending on whether a down converter or an up converter measurement is being
performed.
It is important to keep in mind that in the setup diagrams of the frequency offset mode, the
analyzer’s source and receiver ports are labeled according to the mixer port to which they
are connected.
2-8
Making Mixer Measurements
Measurement Considerations
2- 9
Making Mixer Measurements
Measurement Considerations
2-10
Making Mixer Measurements
Conversion Loss Using the Frequency Offset Mode
Higher measurement accuracy may be obtained through the use of power meter
calibration. You can use power meter calibration to correct for power offsets, losses, and
flatness variations occurring between the analyzer source and the input to the mixer under
test. Refer to the power meter documentation for its calibration procedures.
The following procedure describes the R channel swept IF frequency conversion loss
measurement of a broadband component mixer with power meter calibration. For this
example, we will use the following example settings. For your measurement, you will need
to use settings specific to your measurement.
TIP For ease of use, the RF frequency range needs to be the same as the network
analyzer’s frequency range limit.
2- 11
Making Mixer Measurements
Conversion Loss Using the Frequency Offset Mode
2. From the front panel of the analyzer, set the desired receiver (RF) frequency and source
output power by pressing:
Meas R
The measurement trace is shown on the display.
4. Select the analyzer as the system controller:
SET ADDRESSES
2-12
Making Mixer Measurements
Conversion Loss Using the Frequency Offset Mode
3. Select the appropriate power meter by pressing POWER MTR [ ] until the correct
model number is displayed (Agilent 436A or Agilent 438A/437).
NOTE The Agilent E4418B and E4419B (EPM) power meters have a “437emulation”
mode which can be used in this procedure by following these steps.
• If you are using an Agilent EPM power sensor and your network analyzer
has firmware revision 7.72 or greater:
a. Select the remote interface command set on the power meter by
pressing the System Inputs key and the following softkeys: Remote
Interface, Command Set, SCPI
b. Skip step 4 and continue at step 5.
The power sensor factors are automatically read by the analyzer.
• If you are using an Agilent 848X-series power sensor or your network
analyzer does not have firmware revision 7.72 or greater:
a. Choose the 438A/437 selection on the network analyzer by pressing:
4. Press Cal PWRMTR CAL LOSS/SENSR LISTS CAL FACTOR SENSOR A and
enter the correction factors as listed on the power sensor. Press ADD FREQUENCY
fff (where fff is the frequency of the calibration factor in MHz) M/µ CAL FACTOR
nnn (where nnn is the calibration factor number) x1 DONE for each correction
factor. When finished, press DONE .
5. To perform a one sweep power meter calibration over the RF frequency range at 0 dBm,
press:
NOTE Because power meter calibration requires a longer sweep time, you may want
to reduce the number of points before pressing TAKE CAL SWEEP . After the
power meter calibration is finished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration.
6. Make sure the power meter calibration is on. When the power meter calibration is on,
“PC” is displayed at the left edge of the display. Refer to Figure 2-15 on page 2-17 for an
example.
2- 13
Making Mixer Measurements
Conversion Loss Using the Frequency Offset Mode
7. From the front panel of the analyzer, set the desired IF start and stop frequencies by
pressing:
❏ The low pass filter is required to limit the range of frequencies passed into the R
channel input port. The filter is selected to pass the IF frequencies for the
measurement but prevent the LO feedthrough and unwanted mixer products from
confusing the phase-lock loop operation.
❏ A pad is used to isolate the filter and improve the IF port match for the mixer.
Once completed, the display should read 0 dBm.
10. Save the power meter and receiver calibration to an instrument state by pressing
Save/Recall SAVE STATE .
2-14
Making Mixer Measurements
Conversion Loss Using the Frequency Offset Mode
The analyzer is now displaying the conversion loss of the mixer calibrated with power
meter accuracy.
2. While the analyzer is still set to the IF frequency range, set the frequency offset mode
LO frequency from the analyzer by pressing:
2- 15
Making Mixer Measurements
Conversion Loss Using the Frequency Offset Mode
2-16
Making Mixer Measurements
Conversion Loss Using the Frequency Offset Mode
5. To view the conversion loss in the best vertical resolution, press Scale Ref
AUTOSCALE .
In this measurement, you set the input power and measured the output power. Figure 2-15
shows the absolute loss through the mixer versus mixer output frequency. If the mixer
under test contained built-in amplification, then the measurement results would have
shown conversion gain.
2- 17
Making Mixer Measurements
High Dynamic Range Swept RF/IF Conversion Loss
Preset
Start 100 M/µ Stop 1 G/n
Power −5 x1
Sweep Setup NUMBER of POINTS 10 x1
2-18
Making Mixer Measurements
High Dynamic Range Swept RF/IF Conversion Loss
Local
SYSTEM CONTROLLER
4. Set the power meter’s address:
SET ADDRESSES
NOTE The E4418B and E4419B power meters have a “437 emulation” mode. This
allows these power meters, with an 848X-series power sensor, to be used with
the network analyzer. In this step, when selecting a power meter, choose the
438A/437 selection.
6. Press Cal PWRMTR CAL LOSS/SENSR LISTS CAL FACTOR SENSOR A and
enter the correction factors as listed on the power sensor. Press ADD FREQUENCY
fff (where fff is the frequency in MHz) M/µ CAL FACTOR nnn (where nnn is the
calibration factor number) x1 DONE for each correction factor. When finished, press
DONE RETURN .
2- 19
Making Mixer Measurements
High Dynamic Range Swept RF/IF Conversion Loss
7. Perform a one sweep power meter calibration over the IF frequency range at −5 dBm:
ONE SWEEP
−5 x1
NOTE Because power meter calibration requires a longer sweep time, you may want
to reduce the number of points before pressing TAKE CAL SWEEP . After the
power meter calibration is finished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration.
Meas B
2-20
Making Mixer Measurements
High Dynamic Range Swept RF/IF Conversion Loss
2- 21
Making Mixer Measurements
High Dynamic Range Swept RF/IF Conversion Loss
3. Perform a one sweep power meter calibration over the RF frequency range at −5 dBm:
ONE SWEEP
−5 x1
Figure 2-19 Connections for a High Dynamic Range Swept IF Conversion Loss
Measurement
4. Set the analyzers LO frequency to match the frequency of the LO source by pressing:
2-22
Making Mixer Measurements
High Dynamic Range Swept RF/IF Conversion Loss
2- 23
Making Mixer Measurements
Fixed IF Mixer Measurements
Sequence 1 Setup
The following sequence initializes and calibrates the network analyzer. It then initializes
the two external sources prior to measurement. This sequence includes:
• putting the network analyzer into tuned receiver mode
• setting up a frequency list sweep of 26 points
• performing a response calibration
• prompting the user to connect a mixer to the test set up
• initializing a loop counter value to 26
• addressing and configuring the two sources
• calling the next measurement sequence
1. Make the following connections as shown in Figure 2-21. Set the GPIB address of the
external RF source to 19 and the external LO source to 21.
2. Confirm that the external sources are configured to receive commands in the SCPI
programming language and that their output power is switched on.
2-24
Making Mixer Measurements
Fixed IF Mixer Measurements
NOTE You may have to consult the user’s guide of the external source being used to
determine how to set the source to receive SCPI commands.
3. Be sure to connect the 10 MHz reference signals of the external sources to the EXT REF
connector on the rear panel of the analyzer (a BNC tee must be used).
NOTE If the 10 MHz reference signals of the external sources are connected
together, then it will only be necessary to connect one reference signal from
one of the sources to the EXT REF connector of the analyzer.
NOTE To enter the following sequence commands that require titling, an external
keyboard may be used for convenience.
2- 25
Making Mixer Measurements
Fixed IF Mixer Measurements
LIST FREQ
Performing a Response Calibration
THRU
Prompting the User to Connect a Mixer to the Test Setup
PAUSE
2-26
Making Mixer Measurements
Fixed IF Mixer Measurements
Scale Ref 2 x1
REFERENCE POSITION 0 x1
TITLE TO PERIPHERAL
TITLE TO PERIPHERAL
TITLE TO PERIPHERAL
Calling the Next Measurement Sequence
2- 27
Making Mixer Measurements
Fixed IF Mixer Measurements
Press Seq NEW SEQ/MODIFY SEQ SEQUENCE 1 SEQ 1 and the analyzer will
display the following sequence commands:
SEQUENCE SEQ1
Start of Sequence
RECALL PRST STATE
SYSTEM CONTROLLER
TUNED RECEIVER
EDIT LIST
ADD
CW FREQ
100M/u
NUMBER OF POINTS
26x1
DONE
DONE
LIST FREQ
B
TITLE
POW:LEV 6DBM
PERIPHERAL HPIB ADDR
19x1
TITLE TO PERIPHERAL
TITLE
FREQ:MODE CW;CW 100MHZ
TITLE TO PERIPHERAL
CALIBRATE: RESPONSE
CAL STANDARD
DONE CAL CLASS
TITLE
CONNECT MIXER
PAUSE
LOOP COUNTER
26x1
SCALE/DIV
2 x1
REFERENCE POSITION
0 x1
REFERENCE VALUE
−20x1
MANUAL TRG ON POINT
TITLE
FREQ:MODE CW;CW 500MHZ;:FREQ:CW:STEP 100MHZ
TITLE TO PERIPHERAL
TITLE
POW:LEV 13DBM
PERIPHERAL HPIB ADDR
21x1
TITLE TO PERIPHERAL
2-28
Making Mixer Measurements
Fixed IF Mixer Measurements
TITLE
FREQ:MODE CW;CW 600MHZ;:FREQ:CW:STEP 100MHZ
TITLE TO PERIPHERAL
DO SEQUENCE
SEQUENCE 2
Sequence 2 Setup
The following sequence makes a series of measurements until all 26 CW measurements
are made and the loop counter value is equal to zero. This sequence includes:
• taking data
• incrementing the source frequencies
• decrementing the loop counter
• labeling the screen
1. Press the following keys on the analyzer to create sequence 2:
NOTE To enter the following sequence commands that require titling, an external
keyboard may be used for convenience.
TITLE TO PERIPHERAL
2- 29
Making Mixer Measurements
Fixed IF Mixer Measurements
Press Seq NEW SEQ/MODIFY SEQ SEQUENCE 2 SEQ2 and the analyzer will
display the following sequence commands:
SEQUENCE SEQ2
Start of Sequence
WAIT x
1 x1
MANUAL TRG ON POINT
TITLE
FREQ:CW UP
PERIPHERAL HPIB ADDR
19x1
TITLE TO PERIPHERAL
PERIPHERAL HPIB ADDR
21x1
TITLE TO PERIPHERAL
DECR LOOP COUNTER
IF LOOP COUNTER <>0 THEN DO
SEQUENCE 2
TITLE
MEASUREMENT COMPLETED
2. Press the following keys to run the sequences:
Seq DONE SEQ MODIFY DO SEQUENCE SEQUENCE 2 SEQ2
When the prompt CONNECT MIXER appears, connect the equipment as shown in Figure
2-22.
2-30
Making Mixer Measurements
Fixed IF Mixer Measurements
Figure 2-22 Connections for a Conversion Loss Using Tuned Receiver Mode
When the sequences are finished you should have a result as shown in Figure 2-23.
The displayed trace represents the conversion loss of the mixer at 26 points. Each point
corresponds to one of the 26 different sets of RF and LO frequencies that were used to
create the same fixed IF frequency.
2- 31
Making Mixer Measurements
Phase or Group Delay Measurements
Phase Measurements
When you are making linear measurements, you must provide a reference for determining
phase by splitting the RF source power and send part of the signal into the reference
channel. (This does not work for frequency offset measurements, since the source and
receiver are functioning at different frequencies.)
To provide a reference signal for the phase measurement, you need a second mixer. This
mixer is driven by the same RF and LO signals that are used to drive the mixer under test.
The IF output from the reference mixer is applied to the reference (R) channel of the
analyzer.
2-32
Making Mixer Measurements
Phase or Group Delay Measurements
1. Set the LO source to the desired CW frequency of 1000 MHz and power level to 13 dBm.
2. Initialize the analyzer by pressing Preset .
3. Set the analyzer’s LO frequency to match the frequency of the LO source by pressing:
2- 33
Making Mixer Measurements
Phase or Group Delay Measurements
Format DELAY
9. To make a response error-correction, press:
Meas B/R
2-34
Making Mixer Measurements
Phase or Group Delay Measurements
10.Replace the "calibration" mixer with the device under test. If measuring group delay, set
the delay equal to the "calibration" mixer delay (for example −0.6 ns) by pressing:
Scale Ref
ELECTRICAL DELAY
−06 G/n
11.Scale the data for best vertical resolution.
Scale Ref
AUTOSCALE
The displayed measurement trace shows the device under test delay, relative to the
"calibration" mixer. This measurement is also useful when the device under test has
built-in filtering, which requires >30 dB range (the maximum of R input). PORT 1 to PORT
2 range is >100 dB.
2- 35
Making Mixer Measurements
Amplitude and Phase Tracking
2. Once the analyzer has displayed the measurement results, press Display
DATA→MEM .
3. Replace the calibration mixer with the mixer under test.
4. Press DATA/MEM .
The resulting trace should represent the amplitude and phase tracking of the two mixers.
2-36
Making Mixer Measurements
Conversion Compression Using the Frequency Offset Mode
Figure 2-27 Conversion Loss and Output Power as a Function of Input Power
Level Example
Notice that the IF output power increases linearly with the increasing RF signal, until
mixer compression begins and the mixer saturates.
The following example uses a ratio of mixer output to input power and a marker search
function to locate a mixer’s 1 dB compression point.
1. Set the LO source to the desired CW frequency of 600 MHz and power level to 13 dBm.
2. Initialize the analyzer by pressing Preset .
3. Set the analyzers LO frequency to match the frequency of the LO source by pressing:
Sweep Setup
Power
Start −10 x1
Stop 10 x1
2- 37
Making Mixer Measurements
Conversion Compression Using the Frequency Offset Mode
Meas R
7. To store a trace of the receiver power versus the source power into memory and view
data/memory, press:
Display
DATA→MEM
DATA/MEM
This removes the loss between the output of the mixer and the input to the receiver, and
provides a linear power sweep for use in subsequent measurements.
8. Make the connections as shown in Figure 2-29.
2-38
Making Mixer Measurements
Conversion Compression Using the Frequency Offset Mode
System
RETURN
UP CONVERTER
11.To select a low-side LO measurement configuration, press:
RF>LO
FREQ OFFS ON
In this low-side LO, up converter measurement, the analyzer source frequency is offset
lower than the receiver frequency. The analyzer source frequency can be determined from
the following equation:
receiver frequency (800 MHz) − LO frequency (600 MHz) = 200 MHz
The measurements setup diagram is shown in Figure 2-30.
2- 39
Making Mixer Measurements
Conversion Compression Using the Frequency Offset Mode
12.To view the mixer’s output power as a function of its input power, press:
VIEW MEASURE
13.To set up an active marker to search for the 1 dB compression point of the mixer, press:
Scale Ref
AUTO SCALE
2-40
Making Mixer Measurements
Conversion Compression Using the Frequency Offset Mode
2- 41
Making Mixer Measurements
Isolation Example Measurements
LO to RF Isolation
LO to RF isolation is the amount the LO power is attenuated when it appears directly at
the RF port.
Start 10 M/µ
Power 0 x1
This source stimulates the mixer’s LO port.
3. To select a ratio B/R measurement, press:
Meas B/R
4. Make the connections as shown in Figure 2-33.
2-42
Making Mixer Measurements
Isolation Example Measurements
NOTE A full 2-port calibration will increase the accuracy of isolation measurements.
Refer to Chapter 5 , “Optimizing Measurement Results.”
CAUTION To get an accurate assessment of the LO-IF isolation, the proper LO power
level must be input to the LO port.
2- 43
Making Mixer Measurements
Isolation Example Measurements
Scale Ref
AUTO SCALE
The measurement results show the mixer’s LO to RF isolation.
2-44
Making Mixer Measurements
Isolation Example Measurements
RF Feedthrough
The procedure and equipment configuration necessary for this measurement are very
similar to those of the previous LO to RF Isolation procedure, with the addition of an
external source to drive the mixer’s LO port as we measure the mixer’s RF feedthrough.
RF feedthrough measurements do not use the frequency offset mode.
1. Select the CW LO frequency and source power from the front panel of the external
source.
CW frequency = 300 MHz
Power = 10 dBm
2. Initialize the analyzer by pressing Preset .
3. To select the analyzer frequency range and source power, press:
Start 10 M/µ
Power 0 x1
This source stimulates the mixer’s LO port.
4. To select a ratio B/R measurement, press:
Meas B/R
NOTE Isolation is dependent on LO power level and frequency. To ensure good test
results, you should choose these parameters as close to actual operating
conditions as possible.
2- 45
Making Mixer Measurements
Isolation Example Measurements
NOTE You may see spurious responses on the analyzer trace due to interference
caused by LO to IF leakage in the mixer. This can be reduced with averaging
or by reducing the IF bandwidth.
2-46
Making Mixer Measurements
Isolation Example Measurements
You can measure the IF to RF isolation in a similar manner, but with the following
modifications:
• Use the analyzer source as the IF signal drive.
• View the leakage signal at the RF port.
2- 47
Making Mixer Measurements
Isolation Example Measurements
V
Γ = -----r
Vi
+ Γ
SWR = 1---------------
1– Γ
NOTE Mixers are three-port devices, and the reflection from any one port depends
on the conditions of the other two ports. You should replicate the operating
conditions the mixer will experience as closely as possible for the
measurement. For all mixer SWR measurements, use the same power level
that the mixer will use during normal operation.
When you measure the RF port SWR, you should have the LO drive level present and set
to the expected frequency and power levels. Different LO drives and frequencies may yield
different values for SWR at the same RF frequencies. The IF port should be terminated in
a condition as close to its operating state as possible.
The measurements of LO port SWR and IF port SWR are very similar. For IF port SWR,
you should terminate the RF port in a matched condition and apply the LO signal at its
normal operating level. For the LO port SWR, the RF and IF ports should both be
terminated in conditions similar to what will be present during normal operation.
2-48
3 Making Time Domain Measurements
3-1
Making Time Domain Measurements
Using This Chapter
3-2
Making Time Domain Measurements
Introduction to Time Domain Measurements
NOTE The analyzer can be ordered with Option 010, or the option can be added at a
later date.
The transform used by the analyzer resembles time domain reflectometry (TDR)
measurements. TDR measurements, however, are made by launching an impulse or step
into the test device and observing the response in time with a receiver similar to an
oscilloscope. In contrast, the analyzer makes swept frequency response measurements,
and mathematically transforms the data into a TDR-like display.
Figure 3-1 illustrates the frequency and time domain reflection responses of a test device.
The frequency domain reflection measurement is the composite of all the signals reflected
by the discontinuities present in the test device over the measured frequency range.
3- 3
Making Time Domain Measurements
Introduction to Time Domain Measurements
Figure 3-1 Device Frequency Domain and Time Domain Reflection Responses
The time domain measurement shows the effect of each discontinuity as a function of time
(or distance), and shows that the test device response consists of three separate impedance
changes. The second discontinuity has a reflection coefficient magnitude of 0.035 (i.e. 3.5%
of the incident signal is reflected). Marker 1 on the time domain trace shows the elapsed
time from the reference plane (where the calibration standards are connected) to the
discontinuity and back: 18.2 nanoseconds.
The analyzer has three frequency-to-time transform modes:
• Time domain bandpass mode simulates the time domain response of an impulse
input and is designed to measure band-limited devices. Although this mode is the
easiest to use, it results in less time domain resolution than low pass mode, and may
result in some magnitude errors at low frequencies when gating is used. For devices
that are not band-limited, one of the low pass modes is recommended.
• Time domain low pass step mode simulates the time domain response of a step
input. As in a traditional TDR measurement, the distance to the discontinuity in the
test device, and the type of discontinuity (resistive, capacitive, inductive) can be
determined.
• Time domain low pass impulse mode simulates the time domain response of an
impulse input (like the bandpass mode). Both low pass modes yield better time domain
resolution for a given frequency span than does the bandpass mode. In addition, when
using the low pass modes, you can determine the type of discontinuity. However, these
modes have certain limitations that are defined in "Time Domain Bandpass Mode" on
page 3-12.
The analyzer has one time-to-frequency transform mode:
• Forward transform mode transforms CW signals measured over time into the
frequency domain, to measure the spectral content of a signal. This mode is known as
the CW time mode.
3-4
Making Time Domain Measurements
Making Transmission Response Measurements
Preset
Meas Trans: FWD S21 (B/R)
Start 119 M/µ
Stop 149 M/µ
Scale Ref AUTO SCALE
3. Substitute a thru for the device under test and perform a frequency response correction.
Refer to Chapter 6 , “Calibrating for Increased Measurement Accuracy.”
4. Reconnect your device under test.
3- 5
Making Time Domain Measurements
Making Transmission Response Measurements
5. To transform the data from the frequency domain to the time domain and set the sweep
from 0 s to 6 µs, press:
System TRANSFORM MENU BANDPASS TRANSFORM ON
Start 0 G/n Stop 6 M/µ
The other time domain modes, low pass step and low pass impulse, are described in
"Time Domain Low Pass Mode" on page 3-15.
6. To better view the measurement trace, press:
Scale Ref REFERENCE VALUE and turn the front panel knob, or enter a value from
the front panel keypad.
7. To measure the peak response from the main path, press:
Marker Search SEARCH: MAX
The three responses shown in Figure 3-3 are the RF leakage near zero seconds, the
main travel path through the filter, and the triple travel path through the filter. Only
the combination of these responses was evident to you in the frequency domain.
3-6
Making Time Domain Measurements
Making Transmission Response Measurements
11.To activate the gating function to remove any unwanted responses, press:
GATE ON
As shown in Figure 3-4, only response from the main path is displayed.
NOTE You may remove the displayed response from inside the gate markers by
pressing SPAN and turning the front panel knob to exchange the "flag"
marker positions.
12.To adjust the gate shape for the best possible time domain response, press
GATE SHAPE and select between minimum, normal, wide, and maximum. Each gate
has a different passband flatness, cutoff rate, and sidelobe levels. A detailed discussion
of gating and gate shape selections is located in "Gating" on page 3-35 and "Selecting
Gate Shape" on page 3-36.
The passband ripple and sidelobe levels are descriptive of the gate shape. The cutoff
time is the time between the stop time (−6 dB on the filter skirt) and the peak of the
first sidelobe, and is equal on the left and right side skirts of the filter. The minimum
gate span is just twice the cutoff time because it has no passband.
3- 7
Making Time Domain Measurements
Making Transmission Response Measurements
3-8
Making Time Domain Measurements
Making Reflection Response Measurements
Preset
Meas Refl: FWD S11 (A/R)
Start 50 M/µ
Stop 3 G/n
2. Perform an S11 1-port correction on PORT 1. Refer to Chapter 5 , "Optimizing
Measurement Results" for a detailed procedure.
3. Connect your device under test as shown in Figure 3-7.
3- 9
Making Time Domain Measurements
Making Reflection Response Measurements
5. To transform the data from the frequency domain to the time domain, press:
System TRANSFORM MENU BANDPASS TRANSFORM ON
6. To view the time domain over the length (<4 meters) of the cable under test, press:
NOTE Most cables have a relative velocity of 0.66 (for polyethylene dielectrics) or 0.7
(for teflon dielectrics). If you would like the markers to read actual one-way
distance rather than return trip distance, enter one-half the actual velocity
factor. Then the markers will read the actual one-way distance to the
reflection of interest rather than the "electrical length" that assumes a
relative velocity of 1.
1
Velocity Factor = --------
εr
where εr is the relative permittivity of the cable dielectric.
3-10
Making Time Domain Measurements
Making Reflection Response Measurements
3- 11
Making Time Domain Measurements
Time Domain Bandpass Mode
NOTE To cause the markers to read the actual one-way distance to a discontinuity,
rather than the two-way distance, enter one-half the actual velocity factor.
3-12
Making Time Domain Measurements
Time Domain Bandpass Mode
3- 13
Making Time Domain Measurements
Time Domain Bandpass Mode
Format Parameter
3-14
Making Time Domain Measurements
Time Domain Low Pass Mode
NOTE If the start and stop frequencies do not conform to the low pass requirement
before a low pass mode (step or impulse) is selected and transform is turned
on, the analyzer resets the start and stop frequencies. If error correction is on
when the frequency range is changed, this turns it off. Therefore, set the
frequency range for time domain low pass before performing a calibration.
3- 15
Making Time Domain Measurements
Time Domain Low Pass Mode
Table 3-2 Minimum Frequency Ranges for Time Domain Low Pass
Number of Points Minimum Frequency Range Number of Points Minimum Frequency Range
3-16
Making Time Domain Measurements
Time Domain Low Pass Mode
3- 17
Making Time Domain Measurements
Time Domain Low Pass Mode
Figure 3-13 Simulated Low Pass Step and Impulse Response Waveforms (Real
Format)
Figure 3-14 shows example cables with discontinuities (faults) using the low pass step
mode with the real format.
3-18
Making Time Domain Measurements
Time Domain Low Pass Mode
Figure 3-14 Low Pass Step Measurements of Common Cable Faults (Real
Format)
3- 19
Making Time Domain Measurements
Time Domain Low Pass Mode
Figure 3-15 Time Domain Low Pass Measurement of an Amplifier Small Signal
Transient Response
Measuring Separate Transmission Paths through the Test Device Using Low
Pass Impulse Mode
The low pass impulse mode can be used to identify different transmission paths through a
test device that has a response at frequencies down to dc (or at least has a predictable
response, above the noise floor, below 30 kHz).
For example, use the low pass impulse mode to measure the relative transmission times
through a multi-path device such as a power divider. Another example is to measure the
pulse dispersion through a broadband transmission line, such as a fiber optic cable. Both
examples are illustrated in Figure 3-16. The horizontal and vertical axes can be
interpreted as already described in "Time Domain Bandpass Mode" on page 3-12.
3-20
Making Time Domain Measurements
Time Domain Low Pass Mode
3- 21
Making Time Domain Measurements
Transforming CW Time Measurements into the Frequency Domain
3-22
Making Time Domain Measurements
Transforming CW Time Measurements into the Frequency Domain
Using the demodulation capabilities of the analyzer, it is possible to view the amplitude or
the phase component of the modulation separately. The window menu includes the
following softkeys to control the demodulation feature:
DEMOD: OFF is the normal preset state, in which both the amplitude and phase
components of any test device modulation appear on the display.
AMPLITUDE displays only the amplitude modulation, as illustrated in Figure 3-19a.
PHASE displays only the phase modulation, as shown in Figure 3-19b.
3- 23
Making Time Domain Measurements
Transforming CW Time Measurements into the Frequency Domain
Number of points – 1
Range = -------------------------------------------------------------
time span
201 – 1
Range = --------------------------
–3
200 × 10
For the example, a 201 point CW time measurement made over a 200 ms time span, choose
a span of 1 kHz or less on either side of the center frequency (see Figure 3-20). That is,
choose a total span of 2 kHz or less.
3-24
Making Time Domain Measurements
Transforming CW Time Measurements into the Frequency Domain
To increase the frequency domain measurement range, increase the span. The maximum
range is inversely proportional to the sweep time, therefore it may be necessary to increase
the number of points or decrease the sweep time. Because increasing the number of points
increases the auto sweep time, the maximum range is 2 kHz on either side of the selected
CW time measurement center frequency (4 kHz total span). To display a total frequency
span of 4 kHz, enter the span as 4000 Hz.
3- 25
Making Time Domain Measurements
Masking
Masking
Masking occurs when a discontinuity (fault) closest to the reference plane affects the
response of each subsequent discontinuity. This happens because the energy reflected from
the first discontinuity never reaches subsequent discontinuities. For example, if a
transmission line has two discontinuities that each reflect 50% of the incident voltage, the
time domain response (real format) shows the correct reflection coefficient for the first
discontinuity (ρ=0.50). However, the second discontinuity appears as a 37.5% reflection
(ρ=0.375) because only some the incident voltage reached the second discontinuity, and
some of that reflected energy is reflected off the first discontinuity as it returns. For two
discrete discontinuities, the apparent reflection of the second discontinuity is appears as
approximately ρ̂ 2 = ( 1 – ρ 2 ) ⋅ ρ , where ρ̂ 2 is the apparent reflection of the second
1 2
discontinuity, ρ 1 is the reflection of the first discontinuity, and ρ2 is the reflection of the
second discontinuity.
NOTE This example assumes a lossless transmission line. Real transmission lines,
with non-zero loss, attenuate signals as a function of the distance from the
reference plane.
As an example of masking due to line loss, consider the time domain response of a 3 dB
attenuator and a short circuit. The impulse response (log magnitude format) of the short
circuit alone is a return loss of 0 dB, as shown in Figure 3-21a. When the short circuit is
placed at the end of the 3 dB attenuator, the return loss is −6 dB, as shown in Figure 3-21b.
This value actually represents the forward and return path loss through the attenuator,
and illustrates how a lossy network can affect the responses that follow it.
3-26
Making Time Domain Measurements
Windowing
Windowing
The analyzer provides a windowing feature that makes time domain measurements more
useful for isolating and identifying individual responses. Windowing is needed because of
the abrupt transitions in a frequency domain measurement at the start and stop
frequencies. The band limiting of a frequency domain response causes overshoot and
ringing in the time domain response, and causes a non-windowed impulse stimulus to have
a sin(kt)/kt shape, where k = π/frequency span and t = time (see Figure 3-22). This has two
effects that limit the usefulness of the time domain measurement:
• Finite impulse width (or rise time). Finite impulse width limits the ability to
resolve between two closely spaced responses. The effects of the finite impulse width
cannot be improved without increasing the frequency span of the measurement (see
Table 3-3).
• Sidelobes. The impulse sidelobes limit the dynamic range of the time domain
measurement by hiding low-level responses within the sidelobes of higher level
responses. The effects of sidelobes can be improved by windowing (see Table 3-3).
Windowing improves the dynamic range of a time domain measurement by filtering the
frequency domain data prior to converting it to the time domain, producing an impulse
stimulus that has lower sidelobes. This makes it much easier to see time domain responses
that are very different in magnitude. The sidelobe reduction is achieved, however, at the
expense of increased impulse width. The effect of windowing on the step stimulus (low pass
mode only) is a reduction of overshoot and ringing at the expense of increased rise time.
To select a window, press System TRANSFORM MENU WINDOW . A menu is
presented that allows the selection of three window types, see Table 3-3.
3- 27
Making Time Domain Measurements
Windowing
Window Type Impulse Sidelobe Low Pass Impulse Step Sidelobe Step Rise Time
Level Width (50%) Level (10 − 90%)
NOTE: The bandpass mode simulates an impulse stimulus. Bandpass impulse width is twice
that of low pass impulse width. The bandpass impulse sidelobe levels are the same as low pass
impulse sidelobe levels.
Choose one of the three window shapes listed or use the knob to select any windowing
pulse width (or rise time for a step stimulus) between the softkey values. The time domain
stimulus sidelobe levels depend only on the window selected.
MINIMUM is essentially no window. Consequently, it gives the highest
sidelobes.
NORMAL (the preset mode) gives reduced sidelobes and is the mode
most often used.
MAXIMUM window gives the minimum sidelobes, providing the
greatest dynamic range.
USE MEMORY on OFF remembers a user-specified window pulse width (or step
rise time) different from the standard window values.
A window is activated only for viewing a time domain response, and does not affect a
displayed frequency domain response. Figure 3-23 shows the typical effects of windowing
on the time domain response of a short circuit reflection measurement.
3-28
Making Time Domain Measurements
Windowing
Figure 3-23 The Effects of Windowing on the Time Domain Responses of a Short
Circuit (Real Format)
3- 29
Making Time Domain Measurements
Range
Range
In the time domain, range is defined as the length in time that a measurement can be
made without encountering a repetition of the response, called aliasing. A time domain
response repeats at regular intervals because the frequency domain data is taken at
discrete frequency points, rather than continuously over the frequency band.
1-
Measurement range = ------
∆F
where ∆F is the spacing between frequency data points
( number of points – 1 )
Measurement range = --------------------------------------------------------------
frequency span ( Hz )
For example:
1- ( number of points – 1 )
Range = ------ or -----------------------------------------------------------------
∆F frequency span
1 ( 201 – 1 )
Range = ------------------------- or ----------------------
9
( 2 × 10 )
6
( 10 × 10 )
ElectricalLength = 30 meters
3-30
Making Time Domain Measurements
Range
In this example, the range is 100 ns, or 30 meters electrical length. To prevent the time
domain responses from overlapping, the test device must be 30 meters or less in electrical
length for a transmission measurement (15 meters for a reflection measurement). The
analyzer limits the stop time to prevent the display of aliased responses.
To increase the time domain measurement range, first increase the number of points, but
remember that as the number of points increases, the sweep speed decreases. Decreasing
the frequency span also increases range, but reduces resolution.
3- 31
Making Time Domain Measurements
Resolution
Resolution
Two different resolution terms are used in the time domain:
• response resolution
• range resolution
Response Resolution
Time domain response resolution is defined as the ability to resolve two closely-spaced
responses, or a measure of how close two responses can be to each other and still be
distinguished from each other. For responses of equal amplitude, the response resolution is
equal to the 50% (−6 dB) impulse width. It is inversely proportional to the measurement
frequency span, and is also a function of the window used in the transform. The
approximate formulas for calculating the 50% impulse width are given in Table 3-3. For
example, using the formula for the bandpass mode with a normal windowing function for a
50 MHz to 13.05 GHz measurement (13.0 GHz span):
0.98
50 percent calculated impulse width = --------------------------- × 2
13.0 ( GHz )
= 0.151 nanoseconds
c m
Electrical length ( in air ) = 0.151 × 10 –9 s × 30 × 10
9
----
s
= 4.53 centimeters
With this measurement, two equal responses can be distinguished when they are
separated by at least 4.53 centimeters. In a measurement with a 20 GHz span, two equal
responses can be distinguished when they are separated by at least 2.94 cm. Using the low
pass mode (the low pass frequencies are slightly different) with a minimum windowing
function, you can distinguish two equal responses that are about 1.38 centimeters or more
apart.
For reflection measurements, which measure the two-way time to the response, divide the
response resolution by 2. Using this example, you can distinguish two faults of equal
magnitude provided they are 0.69 centimeters (electrical length) or more apart.
NOTE Remember, to determine the physical length, the relative velocity factor of the
transmission medium under test must be entered into the electrical length
equation.
3-32
Making Time Domain Measurements
Resolution
For example, a cable with a teflon dielectric (0.7 relative velocity factor), measured under
the conditions stated above, has a fault location measurement response resolution of 0.45
centimeters. This is the maximum fault location response resolution. Factors such as
reduced frequency span, greater frequency domain data windowing, and a large
discontinuity shadowing the response of a smaller discontinuity, all act to degrade the
effective response resolution.
Figure 3-24 illustrates the effects of response resolution. The solid line shows the actual
reflection measurement of two approximately equal discontinuities (the input and output
of an SMA barrel). The dashed line shows the approximate effect of each discontinuity, if
they could be measured separately.
While increasing the frequency span increases the response resolution, keep the following
points in mind:
• The time domain response noise floor is directly related to the frequency domain data
noise floor. Because of this, if the frequency domain data points are taken at or below
the measurement noise floor, the time domain measurement noise floor is degraded.
• The time domain measurement is an average of the response over the frequency range
of the measurement. If the frequency domain data is measured out-of-band, the time
domain measurement is also the out-of-band response.
You may (with these limitations in mind) choose to use a frequency span that is wider than
the test device bandwidth to achieve better resolution.
3- 33
Making Time Domain Measurements
Resolution
Range Resolution
Time domain range resolution is defined as the ability to locate a single response in time. If
only one response is present, range resolution is a measure of how closely you can pinpoint
the peak of that response. The range resolution is equal to the digital resolution of the
display, which is the time domain span divided by the number of points on the display. To
get the maximum range resolution, center the response on the display and reduce the time
domain span. The range resolution is always much finer than the response resolution (see
Figure 3-25).
3-34
Making Time Domain Measurements
Gating
Gating
Gating provides the flexibility of selectively removing time domain responses. The
remaining time domain responses can then be transformed back to the frequency domain.
For reflection (or fault location) measurements, use this feature to remove the effects of
unwanted discontinuities in the time domain. You can then view the frequency response of
the remaining discontinuities. In a transmission measurement, you can remove the effects
of multiple transmission paths.
Figure 3-26a shows the frequency response of an electrical airline and termination. Figure
3-26b shows the response in the time domain. The discontinuity on the left is due to the
input connector. The discontinuity on the right is due to the termination. We want to
remove the effect of the connector so that we can see the frequency response of just the
airline and termination. Figure 3-26c shows the gate applied to the connector
discontinuity. Figure 3-26d shows the frequency response of the airline and termination,
with the connector "gated out."
3- 35
Making Time Domain Measurements
Gating
Gate Shape Passband Ripple Sidelobe Levels Cutoff Time Minimum Gate Span
Gate Span
Note: With 1601 frequency points, gating is available only in the bandpass mode.
The passband ripple and sidelobe levels are descriptive of the gate shape. The cutoff time
is the time between the stop time (−6 dB on the filter skirt) and the peak of the first
sidelobe, and is equal on the left and right side skirts of the filter. As shown in Table 3-4,
the minimum gate span is just twice the cutoff time because it has no passband. Always
choose a gate span wider than the minimum. For most applications, do not be concerned
about the minimum gate span, simply use the knob to position the gate markers around
the desired portion of the time domain trace.
3-36
4 Printing, Plotting, and Saving
Measurement Results
4-1
Printing, Plotting, and Saving Measurement Results
Using This Chapter
4-2
Printing, Plotting, and Saving Measurement Results
Printing or Plotting Your Measurement Results
4- 3
Printing, Plotting, and Saving Measurement Results
Configuring a Print Function
2. Press Local SET ADDRESSES PRINTER PORT PRNTR TYPE until the correct
printer choice appears: TPE
❏ ThinkJet (QuietJet)
❏ DeskJet (This supports most current models such as DeskJet 890C, DeskJet 895C, or
DeskJet 1600C. See also DJ 540 selection.)
❏ LaserJet (only LaserJet models III, 4, 5, and 6)
❏ PaintJet
❏ Epson-P2 (printers that conform to the ESC/P2 printer control language, such as
Epson LQ-570)
❏ DJ 540 (This can be used for printers that do not support 100 dots per inch (dpi) but
do support 300 dpi such as HP DeskJet 540 or 850C.)
NOTE Selecting DJ 540 converts 100 dpi raster information to 300 dpi raster format.
If your DeskJet printer does not support the 100 dpi raster format and your
printing results seem to be smaller than the normal size (approximately
one-half of the page), select DJ 540.
Information regarding a printer compatibility guide (an up-to-date list of printers that
are compatible with the network analyzer) is available in "Printing or Plotting Your
Measurement Results" on page 4-3.
4-4
Printing, Plotting, and Saving Measurement Results
Configuring a Print Function
• Choose PRNTR PORT GPIB if your printer has a GPIB interface, and then
configure the print function as follows:
❏ If you choose PARALLEL [COPY] , the parallel port is dedicated for normal copy
device use (printers or plotters).
❏ If you choose PARALLEL [GPIO] , the parallel port is dedicated for general
purpose I/O, and cannot be used for printing or plotting.
• Choose SERIAL if your printer has a serial (RS-232) interface, and then configure
the print function as follows:
a. Press PRINTER BAUD RATE and enter the printer's baud rate, followed by
x1 .
b. To select the transmission control method that is compatible with your printer,
press XMIT CNTRL (transmit control - handshaking protocol) until the correct
method appears.
❏ If you choose Xon-Xoff , the handshake method allows the printer to control
the data exchange by transmitting control characters to the network analyzer.
❏ If you choose DTR-DSR , the handshake method allows the printer to control
the data exchange by setting the electrical voltage on one line of the RS-232
serial cable.
NOTE Because the DTR-DSR handshake takes place in the hardware rather than
the firmware or software, it is the fastest transmission control method.
4- 5
Printing, Plotting, and Saving Measurement Results
Defining a Print Function
NOTE The print definition is set to default values whenever the power is cycled.
However, you can save the print definition by saving the instrument state.
❏ Choose PRINT: MONOCHROME if you are using a black and white printer, or you
want just black and white from a color printer.
❏ Choose PRINT: COLOR if you are using a color printer.
3. Press AUTO-FEED until the correct choice (ON or OFF) is highlighted.
NOTE Laser printers and some DeskJet printers do not begin to print until a full
page, or a partial page and a form feed, have been received.
NOTE You can set all the print elements to black to create a hardcopy in black and
white.
Since the media color is usually white or clear, you could set a print element
to white if you do not want that element to appear on your hardcopy.
4-6
Printing, Plotting, and Saving Measurement Results
Printing One Measurement Per Page
Auto Feed ON
Printer Colors
Graticule Cyan
Warning Black
Text Black
If you defined the AUTO-FEED OFF , press PRINTER FORM FEED after the
message COPY OUTPUT COMPLETED appears. FRM FEED
4- 7
Printing, Plotting, and Saving Measurement Results
Printing Multiple Measurements Per Page
NOTE This feature will not work for all printers due to differences in printer
resolution.
4-8
Printing, Plotting, and Saving Measurement Results
Configuring a Plot Function
• Choose PRNTR PORT GPIB if your printer has an GPIB interface, and then
configure the print function as follows:
4- 9
Printing, Plotting, and Saving Measurement Results
Configuring a Plot Function
• Choose PARALLEL if your printer has a parallel (Centronics) interface, and then
configure the print function as follows:
Press Local and then select the parallel port interface function by pressing
PARALLEL until the correct function appears.
❏ If you choose PARALLEL [COPY] , the parallel port is dedicated for normal copy
device use (printers or plotters).
❏ If you choose PARALLEL [GPIO] , the parallel port is dedicated for general
purpose I/O, and cannot be used for printing or plotting.
• Choose SERIAL if your printer has a serial (RS-232) interface, and then configure
the print function as follows:
a. Press PRINTER BAUD RATE and enter the printer's baud rate, followed by
x1 .
b. To select the transmission control method that is compatible with your printer,
press XMIT CNTRL (transmit control - handshaking protocol) until the correct
method appears.
❏ If you choose Xon-Xoff , the handshake method allows the printer to control
the data exchange by transmitting control characters to the network analyzer.
❏ If you choose DTR-DSR , the handshake method allows the printer to control
the data exchange by setting the electrical voltage on one line of the RS-232
serial cable.
NOTE Because the DTR-DSR handshake takes place in the hardware rather than
the firmware or software, it is the fastest transmission control method.
4. Press Local SET ADDRESSES PLOTTER PORT and then PLTR TYPE until
PLTR TYPE [HPGL PRT] appears.
• Choose PLTR PORT GPIB if your plotter has a GPIB interface, and then configure
the plot function as follows:
4-10
Printing, Plotting, and Saving Measurement Results
Configuring a Plot Function
• Choose PARALLEL if your plotter has a parallel (Centronics) interface, and then
configure the plot function as follows:
❏ Press Local and then select the parallel port interface function by pressing
PARALLEL until the correct function appears.
— If you choose PARALLEL [COPY] , the parallel port is dedicated for normal
copy device use (printers or plotters).
— If you choose PARALLEL [GPIO] , the parallel port is dedicated for general
purpose I/O, and cannot be used for printing or plotting.
• Choose SERIAL if your plotter has a serial (RS-232) interface, and then configure
the plot function as follows:
a. Press PLOTTER BAUD RATE and enter the plotter's baud rate, followed by
x1 .
b. To select the transmission control method that is compatible with your plotter,
press XMIT CNTRL (transmit control - handshaking protocol) until the correct
method appears.
❏ If you choose Xon-Xoff , the handshake method allows the plotter to control
the data exchange by transmitting control characters to the network analyzer.
❏ If you choose DTR-DSR , the handshake method allows the plotter to control
the data exchange by setting the electrical voltage on one line of the RS-232
serial cable.
NOTE Because the DTR-DSR handshake takes place in the hardware rather than
the firmware or software, it is the fastest transmission control method.
CAUTION Do not mistake the line switch for the disk eject button when you are
removing the disk from the analyzer. If the line switch is mistakenly pushed,
the instrument will be turned off, losing all settings and data that have not
been saved.
b. Press Save/Recall SELECT DISK and select the disk drive that you will plot to.
• Choose INTERNAL DISK if you will plot to the analyzer internal disk drive.
4- 11
Printing, Plotting, and Saving Measurement Results
Configuring a Plot Function
• Choose EXTERNAL DISK if you will plot to a disk drive that is external to the
analyzer. Then configure the disk drive as follows:
1. Press CONFIGURE EXT DISK ADDRESS: DISK and enter the GPIB
address to the disk drive (defaultDSK
is 00) followed by x1 .
2. Press Local DISK UNIT NUMBER and enter the drive where your disk is
located, followed by x1 .
3. If your storage disk is partitioned, press VOLUME NUMBER and enter the
volume number where you want to store the instrument state file.
2. Press Copy PLOT .
The analyzer assigns the first available default filename for the displayed directory. For
example, the analyzer would assign PLOT00FP for a LIF format (PLOT00.FP for a DOS
format) if there were no previous plot files saved to the disk.
Figure 4-4 shows the three parts of the file name that are generated automatically by
the analyzer whenever a plot is requested. The two digit sequence number is
incremented by one each time a file with a default name is added to the directory.
4-12
Printing, Plotting, and Saving Measurement Results
Defining a Plot Function
❏ Choose PLOT DATA ON if you want the measurement data trace to appear on your
plot.
❏ Choose PLOT MEM ON if you want the displayed memory trace to appear on your
plot.
❏ Choose PLOT GRAT ON if you want the graticule and the reference line to appear
on your plot.
❏ Choose PLOT TEXT ON if you want all of the displayed text to appear on your plot.
(This does not include the marker values or softkey labels.)
❏ Choose PLOT MKR ON if you want the displayed markers, and marker values, to
appear on your plot.
Selecting Auto-Feed
• Press AUTO-FEED until the correct choice is highlighted.
❏ Choose AUTO-FEED ON if you want a “page eject” sent to the plotter or HPGL
compatible printer after each time you press PLOT .
❏ Choose AUTO-FEED OFF if you want multiple plots on the same sheet of paper.
4- 13
Printing, Plotting, and Saving Measurement Results
Defining a Plot Function
NOTE The following color assignments are valid for HPGL/2 compatible color
printers only. When using word processor or graphics presentation programs,
different colors may be assigned to the pen numbers.
0 white 4 yellow
1 cyan 5 green
2 magenta 6 red
3 blue 7 black
Channel 1 Channel 2
NOTE You can set all the pen numbers to black for a plot in black and white.
You must define the pen numbers separately for each measurement channel
(channel 1/3 and channel 2/4).
4-14
Printing, Plotting, and Saving Measurement Results
Defining a Plot Function
— Select LINE TYPE DATA to modify the line type for the data trace. Then enter the
new line type (see Figure 4-6), followed by x1 .
— Select LINE TYPE MEMORY to modify the line type for the memory trace. Then
enter the new line type (see Figure 4-6), followed by x1 .
Table 4-4 Default Line Types for Plot Elements
Plot Elements Channel 1 Line Type Numbers Channel 2 Line Type Numbers
Data Trace 7 7
Memory Trace 7 7
NOTE You must define the line types for each measurement channel (channel 1/3
and channel 2/4).
Choosing Scale
• Press SCALE PLOT until the selection appears that you want.
❏ Choose SCALE PLOT [FULL] if you want the normal scale selection for plotting.
This includes space for all display annotations such as marker values and stimulus
values. The entire analyzer display fits within the defined boundaries of P1 and P2
on the plotter, while maintaining the exact same aspect ratio as the display.
❏ Choose SCALE PLOT [GRAT] if you want the outer limits of the graticule to
correspond to the defined P1 and P2 scaling point on the plotter. (Intended for
plotting on preprinted rectangular or polar forms.)
4- 15
Printing, Plotting, and Saving Measurement Results
Defining a Plot Function
Data 2 Data 3
Memory 5 Memory 6
Graticule 1 Graticule 1
Text 7 Text 7
Marker 7 Marker 7
4-16
Printing, Plotting, and Saving Measurement Results
Plotting One Measurement Per Page Using a Pen Plotter
❏ If you defined the AUTO-FEED OFF , press PLOTTER FORM FEED after the
message COPY OUTPUT COMPLETED appears. FRM FEED
4- 17
Printing, Plotting, and Saving Measurement Results
Plotting Multiple Measurements Per Page Using a Pen Plotter
❏ LEFT UPPER
❏ LEFT LOWER
❏ RIGHT UPPER
❏ RIGHT LOWER
The selected quadrant will appear in the brackets under SEL QUAD [ ] .
4. Press PLOT .
5. Make the next measurement that you want to see on your hardcopy.
6. Press Copy and choose another quadrant where you want to place the displayed
measurement.
7. Repeat the previous three steps until you have captured the results of up to four
measurements.
4-18
Printing, Plotting, and Saving Measurement Results
Plotting Multiple Measurements Per Page Using a Pen Plotter
4- 19
Printing, Plotting, and Saving Measurement Results
To View Plot Files on a PC
When viewed in such programs, the color and font size of the plot may vary from the
output of an HPGL/2 compatible color printer. The following table shows the differences
between the color assignments of HPGL/2 compatible printers and Lotus applications. Also
refer to "Selecting Pen Numbers and Colors" on page 4-14.
Table 4-6 Color Assignment Differences between HPGL/2-Compatible Printers
and Lotus Applications
3 blue 3 green
4 yellow 4 yellow
5 green 5 blue
To modify the color or font size, consult the documentation for the particular application
being used.
NOTE Plot files may also be saved to a floppy disk as a JPEG file and used on a
personal computer. Refer to "Saving in Graphical (JPEG) Form" on page 4-45.
4-20
Printing, Plotting, and Saving Measurement Results
To View Plot Files on a PC
NOTE The network analyzer does not use the suffix *.PLT, so you may want to
change the filename filter to *.* or some other pattern that will allow you to
locate the files you wish to import.
NOTE The network analyzer uses pen 7 for text. The default color in Ami Pro for pen
7 is aqua, which is not very readable against the typical white background.
You may want to change pen 7 to black.
5. After all selections have been made, the file is imported and rendered in a small
graphics frame which can be sized to the page by grabbing one of the nodes and
stretching the box as required.
• You will notice that the annotation around the display is not optimum, as the Ami
Pro filter does not accurately import the HPGL command to render text.
Using Freelance
To view plot files in Freelance, perform the following steps:
1. From the FILE pull-down menu, select IMPORT.
2. Set the file type in the dialog box to HGL.
NOTE The network analyzer does not use the suffix *.HGL, so you may want to
change the filename filter to *.* or some other pattern that will allow you to
locate the files you wish to import.
3. Click OK to import the file.
• You will notice that when the trace is displayed, the text annotation will be illegible.
You can easily fix this with the following steps:
a. From the TEXT pull-down menu, select FONT.
b. Select the type face and size. (Fourteen point text is a good place to start.)
c. Click OK to resize the font.
To change the font color, just do it immediately after you resize the font using the
same dialog box.
4- 21
Printing, Plotting, and Saving Measurement Results
Outputting Plot Files from a PC to a Plotter
NOTE This conversion method has been used to convert many measurement
displays. However, this conversion utility is not supported by Agilent
Technologies.
4-22
Printing, Plotting, and Saving Measurement Results
Outputting Plot Files from a PC to an HPGL Compatible Printer
Command Remark
NOTE As shown in Table 4-7, the <esc> is the symbol used for the escape character,
decimal value 27.
4- 23
Printing, Plotting, and Saving Measurement Results
Outputting Single Page Plots Using a Printer
Step 2. Store the exit HPGL mode and form feed sequence.
1. Create a test file by typing in each character as shown in the left column of
Table 4-8. Do not insert spaces or linefeeds.
2. Name the file exithpgl.
Table 4-8 HPGL Test File Commands
Command Remark
Step 5. Send the exit HPGL mode and form feed sequence to the
printer.
Type print exithpgl to send the HPGL mode and form feed sequence to the printer.
4-24
Printing, Plotting, and Saving Measurement Results
Outputting Multiple Plots to a Single Page Using a Printer
echo off
type hpglinit > spooler
for %%i in (%1) do type %%i >> spooler
type exithpgl >> spooler
copy spooler LPT1
del spooler
echo on
4- 25
Printing, Plotting, and Saving Measurement Results
Plotting Multiple Measurements Per Page from Disk
4-26
Printing, Plotting, and Saving Measurement Results
Plotting Multiple Measurements Per Page from Disk
7. Press Copy PLOT . The analyzer will assign PLOT00FP because you renamed the last
file saved.
8. Press Save/Recall and turn the front panel knob to highlight the name of the file that
you just saved.
9. Press FILE UTILITIES RENAME FILE and turn the front panel knob to place the
↑ pointer to the B character.
10.Press SELECT LETTER DONE .
11.Continue defining plots and renaming the saved file until you have saved all the data
that you want to put on the same page. Renaming the files as shown allows you to use
the provided program, that organizes and plots the files, according to the file naming
convention.
4- 27
Printing, Plotting, and Saving Measurement Results
Plotting Multiple Measurements Per Page from Disk
Figure 4-10 shows plots for both the frequency and time domain responses of the same
device.
4-28
Printing, Plotting, and Saving Measurement Results
Plotting Multiple Measurements Per Page from Disk
4. Press PLOT . The analyzer assigns the first available default filename for the selected
quadrant. For example, the analyzer would assign PLOT01LU if there were no other
left-upper quadrant plots on the disk.
5. Make the next measurement that you want to see on your hardcopy.
6. Repeat this procedure for the remaining plot files that you want to see as quadrants on
a page. If you want to see what quadrants you have already saved, press Save/Recall to
view the directory.
4- 29
Printing, Plotting, and Saving Measurement Results
Titling the Displayed Measurement
2. Press ERASE TITLE and enter the title you want for your measurement display.
• If you have a DIN keyboard attached to the analyzer, type the title you want from the
keyboard. Then press ENTER to enter the title into the analyzer. You can enter a
title that has a maximum of 50 characters. (For more information on using a
keyboard with the analyzer, refer to the “Options and Accessories” chapter of the
reference guide.)
• If you do not have a DIN keyboard attached to the analyzer, enter the title from the
analyzer front panel.
a. Turn the front panel knob to move the arrow pointer to the first character of the
title.
b. Press SELECT LETTER .
c. Repeat the previous two steps to enter the rest of the characters in your title. You
can enter a title that has a maximum of 50 characters.
d. Press DONE to complete the title entry.
CAUTION The NEWLINE and FORMFEED keys are not intended for creating display
titles. Those keys are for creating commands to send to peripherals during a
sequence program.
4-30
Printing, Plotting, and Saving Measurement Results
Configuring the Analyzer to Produce a Time Stamp
4- 31
Printing, Plotting, and Saving Measurement Results
Printing or Plotting the List Values or Operating Parameters
• Choose LIST VALUES if you want a tabular listing of the measured data points, and
their current values, to appear on your hardcopy. This list will also include the limit test
information, if you have the limits function activated.
• Choose OP PARMS (MKRS etc) if you want a tabular listing of the parameters for both
measurement channels to appear on your hardcopy. The parameters include: operating
parameters, marker parameters, and system parameters that relate to the control of
peripheral devices.
NOTE If you are printing the list of operating parameters, only the first four pages
are printed. The fifth page, system parameters, is printed by displaying that
page and then pressing PRINT .
4-32
Printing, Plotting, and Saving Measurement Results
Solving Problems with Printing or Plotting
• Make sure that the analyzer address setting for the peripheral corresponds to the
actual GPIB address of the peripheral. The procedure is explained earlier in this
chapter.
• Make sure that the analyzer is in system controller mode, by pressing Local
SYSTEM CONTROLLER , if the analyzer is not connected to an external controller.
CNTROLLER
Otherwise, the analyzer must be in the pass control mode.
• Substitute the interface cable.
• Substitute a different printer or plotter.
4- 33
Printing, Plotting, and Saving Measurement Results
Saving and Recalling Instrument States
NOTE When the ac line power is switched off, the internal non-volatile memory is
retained by a battery. Refer to “Specifications and Characteristics” in the
reference guide for data retention times.
4-34
Printing, Plotting, and Saving Measurement Results
Saving and Recalling Instrument States
4- 35
Printing, Plotting, and Saving Measurement Results
Saving an Instrument State
❏ INTERNAL MEMORY
❏ INTERNAL DISK
❏ EXTERNAL DISK connect an external disk drive to the analyzer’s GPIB connector,
and configure as follows:
a. Connect an external disk drive to the analyzer's GPIB connector, and configure as
follows:
b. Press Local DISK UNIT NUMBER and enter the drive where your disk is
located, followed by x1 .
c. If your storage disk is partitioned, press VOLUME NUMBER and enter the
volume number where you want to store the instrument state file.
d. Press SET ADDRESSES ADDRESS: DISK .
e. Enter the GPIB address of the peripheral, if the default address is incorrect
(default = 00). Follow the entry by pressing x1 .
f. Press Local and select one of the following:
NOTE If you have saved enough files that you have used all the default names
(FILE00 − FILE31 for disk files, or REG1 − REG31 for memory files), you
must do one of the following to save more states:
• use another disk
• rename an existing file to make a default name available
• re-save a file/register
• delete an existing file/register
4-36
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
You can also save data-only. This is saved to disk with default file names
DATA00D1 to DATA31D1 for channel 1
DATA00D2 to DATA31D2 for channel 2
DATA00D3 to DATA31D3 for channel 3
DATA00D4 to DATA31D4 for channel 4
However, these files are not instrument states and cannot be recalled.
4- 37
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
NOTE If the analyzer has an active two-port measurement calibration, all four
S-parameters will be saved with the measurement results. All four
S-parameters may be viewed if the raw data array has been saved.
1. If you want to title the displayed measurement, refer to "Titling the Displayed
Measurement" on page 4-30.
2. Press Save/Recall SELECT DISK .
3. Choose one of the following disk drives:
• INTERNAL DISK
• EXTERNAL DISK (If necessary, refer to the external disk setup procedure in
"Saving an Instrument State" on page 4-36.)
4. Press Save/Recall DEFINE DISK-SAVE .
5. Define the save by selecting from the following choices:
❏ DATA ARRAY ON
❏ RAW ARRAY ON
4-38
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
❏ FORMAT ARY ON
If you select DATA ARRAY ON , RAW ARRAY ON , or FORMAT ARY ON , the data
is stored to disk in IEEE-64 bit real format (for LIF disks), and 32 bit PC format for
DOS disks. This makes the DOS data files half the size of the LIF files.
NOTE Selecting DATA ARRAY ON may store data to disk in the S2P ASCII data
format. See "ASCII Data Formats" on page 4-40.
❏ GRAPHICS ON
If you select GRAPHICS ON , the user graphics area is saved. (Refer to the
programmer’s guide for information on using display graphics.) The measurement
display is not saved with this selection. (Refer to "If You Are Plotting Measurement
Results to a Disk Drive" on page 4-11 to plot measurement displays to disk.)
❏ DATA ONLY ON
If DATA ONLY ON data array is saved along with any other selected array, the
instrument state is not saved, and therefore, cannot be recalled.
6. Choose the type of format you want:
❏ Choose SAVE USING BINARY for all applications except CITIfile, S2P, or CAE
applications.
❏ Choose SAVE USING ASCII for CITIfile, S2P, and CAE applications or when you
want to import the information into a spread sheet using comma-separated values
(CSV) format.
7. Press RETURN SAVE STATE .
4- 39
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
CITIfile
CITIfile (Common Instrumentation Transfer and Interchange file) is an ASCII data format
that is useful when exchanging data between different computers and instruments.
CITIfiles are always saved when the ASCII format has been selected as shown:
Save/Recall
DEFINE DISK-SAVE
Select one of the following choices:
— DATA ARRAY ON
— DATA ONLY ON
— RAW ARRAY ON
— FORMAT ARY ON
SAVE USING ASCII
RETURN
SAVE STATE
4-40
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
CAUTION Using the smoothing feature or saving data displayed in time domain format
may result in invalid S2P data. Avoid using these functions when saving S2P
files.
4- 41
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
4-42
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
4- 43
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
txtcss.csv
c is the indicator of the channel (1−4) on which the measurement data was
taken.
4-44
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
The graphic file may be retrieved from the floppy disk on personal computer and can be
imported into an application that accepts graphics in the JPEG format.
NOTE When saving measurement results graphically, make sure that no onscreen
measurement data is displayed as white. Since media color is often white, any
measurement data printed using white will not be visible.
You may change the analyzer to the factory default color settings by pressing
Copy DEFINE PRINT DEFAULT PRNT SETUP to correct this problem.
However, to maintain your current color settings (except white), check the
measurement color settings by pressing Copy DEFINE PRINT
PRINT COLORS . Press MORE to check the remaining measurement
colors. To modify any of the measurement colors, select the measurement and
then choose another color from the list of colors that is displayed.
1. The network analyzer firmware is based in part on the work of the Independent JPEG Group.
4- 45
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
Files with .10, .11, .12, .1a, .1b, and .1c File Extensions
The following files are only produced if you have an active calibration. FileXX.10 is a
binary file which stores the stimulus state of the instrument as it relates to an active
calibration (specifically, the Power, Sweep Setup, Start, Stop, Center, and Span settings).
The same type of file is produced if Channel 2 is active, but the file extension is .20 instead
of the .10 file extension mentioned in the previous paragraph.
Files FileXX.11 through .12, .1a, .1b, and .1c are binary files which contain the 12 error
correction coefficients for Channel 1. If Channel 2 is active, it will have the same array, but
file extensions are in the form .21, .22, .2a., .2b, and .2c. If you save in ASCII format, only
.10 and .1c are produced, with .1c containing the entire error correction array in a
two-column, real/imaginary (CITIfile) format.
4-46
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
NOTE The DataXX files are much smaller than an entire instrument state, and are
the best way to get just the data you want without saving the entire
instrument state. Selecting more than one disk-save option does not confuse
the analyzer, and simply produces all files associated with the selected
options. The only exception to this is that selecting DATA ONLY on OFF
suppresses all other selected options.
4- 47
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
Binary Files
The size of the data files is very small, about one tenth the size compared to the ASCII
format. Binary is the format to use when you want to store and recall instrument states on
the analyzer quickly, but do not need to read the data in an external computer.
4-48
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
4- 49
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
Raw Arrays
On the analyzer, press the Save/Recall DEFINE DISK-SAVE RAW ARRAY ON
Data created the first time in this manner will be saved as filename “FILE00.r1”. The file
extension .r1 indicates the data was created while channel 1 was active and stored in the
analyzer's raw data array. If you save the data again, but while channel 2 is active, you will
get a new file called “FILE01.r2”. RAW data are not commonly used unless sophisticated
data processing is to be performed in an external PC. As an example, multi-port calibration
is created by exporting raw data to a PC where error-correction for each of the multi-port
paths is applied to them.
Data Arrays
Press Save/Recall DEFINE DISK-SAVE DATA ARRAY ON .
Data created the first time in this manner will be saved as filename “DATA00.d1.” The file
extension .d1 indicates that the data from the analyzer's channel 1 is error corrected data
only if the analyzer's error correction feature is enabled (in other words, you have
performed a calibration). Otherwise, the data is the same as data stored in the analyzer's
raw data arrays. Data stored in the data arrays does not have any formatting applied to it.
Format Arrays
Press Save/Recall DEFINE DISK-SAVE FORMAT ARY ON
Data created the first time in this manner will be saved as filename “FILE00.f1”. The file
extension .f1 indicates the data is formatted per Figure 4-13 using the analyzer's channel
1. Depending on what features you've selected, data in the format arrays includes data in
the data arrays plus one or more of the following features:
• Trace math (i.e data-memory)
• Gating (Option 010)
• Electrical delay
• Conversion (for complex impedance (Z), admittance (Y), etc.)
• Transform (Option 010)
• Format (log, lin, phase, delay, SWR, excluding Smith and Polar)
• Smoothing
In each of these examples, most users will select SAVE USING ASCII , under the
DEFINE DISK-SAVE softkey menu. If GRAPHICS on OFF is turned ON, an additional
file will be created with file extension .g0. This is a Hewlett-Packard Graphics Language
(HPGL) file.
4-50
Printing, Plotting, and Saving Measurement Results
Re-Saving an Instrument State
NOTE You cannot re-save a file that contains data only. You must create a new file.
❏ INTERNAL MEMORY
❏ INTERNAL DISK
❏ EXTERNAL DISK (If necessary, refer to the external disk setup procedure in
"Saving an Instrument State" on page 4-36.)
2. Press RETURN and then use the or key or the front-panel knob to
highlight the name of the file that you want to re-save.
3. Press RE-SAVE STATE YES .
Deleting a File
1. Press Save/Recall SELECT DISK .
2. Choose from the following storage devices:
❏ INTERNAL MEMORY
❏ INTERNAL DISK
❏ EXTERNAL DISK (If necessary, refer to the external disk setup procedure in
"Saving an Instrument State" on page 4-36.)
3. Press RETURN .
4- 51
Printing, Plotting, and Saving Measurement Results
Renaming a File
Renaming a File
1. Press Save/Recall AUTO-FEED OFF .
2. Choose from the following storage devices:
❏ INTERNAL MEMORY
❏ INTERNAL DISK
❏ EXTERNAL DISK (If necessary, refer to the external disk setup procedure in
"Saving an Instrument State" on page 4-36.)
3. Press RETURN and then use the or keys or the front panel knob to
highlight the name of the file that you want to rename.
4. Press RETURN RENAME FILE ERASE TITLE .
FILE UTILITIES
FLE
5. Turn the front panel knob to point to each character of the new file name, pressing
SELECT LETTER when the arrow points to each character. Press BACK SPACE if
you enter an incorrect character. After you have selected all the characters in the new
file name, press DONE .
NOTE Renaming files may also be done by using the optional external keyboard.
Recalling a File
1. Press Save/Recall SELECT DISK .
2. Choose from the following storage devices:
❏ INTERNAL MEMORY
❏ INTERNAL DISK
❏ EXTERNAL DISK (If necessary, refer to the external disk setup procedure in
"Saving an Instrument State" on page 4-36.)
3. Press the or keys or the front panel knob to highlight the name of the file
that you want to recall.
4. Press RETURN RECALL STATE .
4-52
Printing, Plotting, and Saving Measurement Results
Formatting a Disk
Formatting a Disk
1. Press Save/Recall FILE UTILITIES FORMAT DISK .
2. Choose the type of format you want:
❏ FORMAT:LIF
❏ FORMAT:DOS
3. Press FORMAT EXT DISK YES .
4- 53
Printing, Plotting, and Saving Measurement Results
Formatting a Disk
4-54
5 Optimizing Measurement Results
5-1
Optimizing Measurement Results
Using This Chapter
5-2
Optimizing Measurement Results
Taking Care of Microwave Connectors
Visual Inspection
Do Inspect all connectors carefully Do Not Use a damaged connector - ever
Look for metal particles, scratches, and dents
Connector Cleaning
Do Try compressed air first Do Not Use any abrasives
Use isopropyl alcohol Get liquid into plastic support beads
Clean connector threads
Gaging Connectors
Do Clean and zero the gage before use Do Not Use an out-of-spec connector
Use the correct gage type
Use correct end of calibration block
Gage all connectors before first use
Making Connections
Do Align connectors carefully Do Not Apply bending force to connection
Make preliminary connection lightly Over tighten preliminary connection
Turn only the connector nut Twist or screw any connection
Use a torque wrench for final connect Tighten past torque wrench "break" point
5- 3
Optimizing Measurement Results
Increasing Measurement Accuracy
Interconnecting Cables
Cables that connect the device under test (DUT) to the analyzer are often the most
significant contribution to random errors of your measurement. You should frequently
perform the following steps as a precaution against errors caused by cable
interconnections:
• Inspect for lossy cables.
• Inspect for damaged cable connectors.
• Practice good connector care techniques.
• Minimize cable position changes between error-correction and measurements.
• Inspect for cables which dramatically change magnitude or phase response when flexed.
(This may indicate an intermittent problem.)
Connector Repeatability
Connector repeatability is a source of random measurement error. Measurement
error-corrections do not compensate for these errors. For all connectors, you should
frequently perform the following steps as a precaution against errors caused by connector
repeatability:
• Inspect the connectors.
• Clean the connectors.
• Gauge the connectors.
• Use correct connection techniques. (Refer to "Taking Care of Microwave Connectors" on
page 5-3.)
5-4
Optimizing Measurement Results
Increasing Measurement Accuracy
Temperature Drift
Electrical characteristics will change with temperature due to the thermal expansion
characteristics of devices within the analyzer, calibration devices, test devices, cables, and
adapters. Therefore, the operating temperature is a critical factor in their performance.
During a measurement calibration, the temperature of the calibration devices must be
stable and within 25 ±5 °C.
• Use a temperature-controlled environment.
• Ensure the temperature stability of the calibration devices.
• Avoid handling the calibration devices unnecessarily during calibration.
• Ensure the ambient temperature is ±1 °C of measurement error-correction
temperature.
Frequency Drift
Minute changes in frequency accuracy and stability can occur as a result of temperature
and aging (on the order of parts per million). If you require greater frequency accuracy,
override the internal crystal with a high-stability external source, frequency standard, or
(if your analyzer is equipped with Option 1D5) use the internal frequency standard.
Performance Verification
You should periodically check the accuracy of the analyzer measurements, by performing a
measurement verification at least once per year. The service guide includes the
measurement verification procedure.
5- 5
Optimizing Measurement Results
Increasing Measurement Accuracy
Main Effect The end of a cable becomes the Compensates for the electrical
test port plane for all length of a cable. Set the
S-parameter measurements. cable’s electrical length × 1 for
transmission. Set the cable’s
electrical length × 2 for
reflection.
You can activate a port extension by pressing Cal MORE PORT EXTENSIONS
EXTENSIONS ON . Then enter the delay to the reference plane.
5-6
Optimizing Measurement Results
Making Accurate Measurements of Electrically Long Devices
5- 7
Optimizing Measurement Results
Making Accurate Measurements of Electrically Long Devices
5-8
Optimizing Measurement Results
Increasing Sweep Speed
1. To set up a swept list measurement, press Sweep Setup SWEEP TYPE MENU
EDIT LIST ADD .
2. The frequency segments can be defined in any of the following terms:
• start/stop/number of points/power/IFBW
• start/stop/step/power/IFBW
• center/span/number of points/power/IFBW
• center/span/step/power/IFBW
3. When finished, press DONE LIST TYPE: [SWEPT] .
5- 9
Optimizing Measurement Results
Increasing Sweep Speed
• If there is no difference between the measurements in either list mode, then use the
swept list mode.
• If the memory trace indicates that there is more attenuation in swept list mode, it
may be due to IF delay. You can usually remedy this problem by increasing the sweep
time.
NOTE IF bandwidths of 30 to 10 Hz cause the sweep (or that segment of the sweep)
to always be stepped, thus eliminating IF delay.
5-10
Optimizing Measurement Results
Increasing Sweep Speed
• Select LIN FREQ for the fastest sweep for a given number of fixed points.
• Select LIST FREQ for the fastest sweep when specific non-linearly spaced
frequency points are of interest.
• Select LOG FREQ for the fastest sweep when the frequency points of interest are in
the lower part of the frequency span selected.
5- 11
Optimizing Measurement Results
Increasing Sweep Speed
2. Press Chan 1 and Chan 2 to alternately view the two measurement channels.
If you must view both measurement channels simultaneously (with dual channel), use
the chop sweep mode, explained next.
3. If you want to view channel 3 (or channel 4), press Chan 3 (or Chan 4 ). This will
always result in a dual trace display of channel 1 and channel 3 (or channel 2 and
channel 4).
To return to a single-trace display, press Display DUAL | QUAD SETUP
AUX CHAN on OFF .
5-12
Optimizing Measurement Results
Increasing Sweep Speed
• Continuous: In this mode the analyzer will switch between the test ports on every
sweep. Although this type of test set switching provides the greatest measurement
accuracy, it requires a reverse sweep for every forward sweep.
• Number of Sweeps: In this mode there is an initial cycling between the test ports and
then the measurement stays on the active port for a user-defined number of sweeps.
After the specified number of sweeps have been executed, the analyzer switches
between the test ports and begins the cycle again. This type of test set switching can
provide improved measurement accuracy over the hold mode and faster measurement
speeds than continuous mode.
5- 13
Optimizing Measurement Results
Increasing Dynamic Range
CAUTION Do not exceed the maximum test port power level that is printed on
the front panel of your network analyzer. Exceeding this maximum
power level may damage your analyzer.
5-14
Optimizing Measurement Results
Reducing Noise
Reducing Noise
You can use two analyzer functions to help reduce the effect of noise on the data trace:
• activate measurement averaging
• reduce system bandwidth
To Activate Averaging
The noise is reduced with each new sweep as the effective averaging factor increments.
1. Press Avg IF BW .
NOTE Another capability that can be used for effective noise reduction is the marker
statistics function, which computes the average value of part or all of the
formatted trace.
5- 15
Optimizing Measurement Results
Reducing Receiver Crosstalk
5-16
Optimizing Measurement Results
Reducing Recall Time
NOTE Both functions must be turned off to realize the recall time savings.
Refer to “Specifications and Characteristics” chapter in the reference guide for examples of
recall state times with the following functions on or off: raw offsets, spur avoidance, and
blank display. Using blank display may speed up recall times.
5- 17
Optimizing Measurement Results
Reducing Recall Time
5-18
6 Calibrating for Increased
Measurement Accuracy
6-1
Calibrating for Increased Measurement Accuracy
How to Use This Chapter
6-2
Calibrating for Increased Measurement Accuracy
Introduction
Introduction
The accuracy of network analysis is greatly influenced by factors external to the network
analyzer. Components of the measurement setup, such as interconnecting cables and
adapters, introduce variations in magnitude and phase that can mask the actual response
of the device under test.
Error correction is an accuracy enhancement procedure that removes systematic errors
(repeatable measurement variations) in the test setup. The analyzer measures known
standard devices, and uses the results of these measurements to characterize the system.
Measurement accuracy and system characteristics can be affected by the following factors:
• Adapting to a different connector type or impedance.
• Connecting a cable between the test device and an analyzer test port.
• Connecting any attenuator or other such device on the input or output of the test device.
If your test setup meets any of the these conditions, the following system characteristics
may be affected:
• amplitude at device input
• frequency response accuracy
• directivity
• crosstalk (isolation)
• source match
• load match
6- 3
Calibrating for Increased Measurement Accuracy
Calibration Considerations
Calibration Considerations
Measurement Parameters
Calibration procedures are parameter-specific, rather than channel-specific. When a
parameter is selected, the instrument checks the available calibration data, and uses the
data found for that parameter. For example, if a transmission response calibration is
performed for B/R, and an S11 1-port calibration for A/R, the analyzer retains both
calibration sets and corrects whichever parameter is displayed. Once a calibration has
been performed for a specific parameter or input, measurements of that parameter remain
calibrated in either channel, as long as stimulus values are coupled. In the response and
response and isolation calibrations, the parameter must be selected before calibration.
Other correction procedures select parameters automatically. Changing channels during a
calibration procedure invalidates the part of the procedure already performed.
Device Measurements
In calibration procedures that require measurement of several different devices, for
example a short, an open, and a load, the order in which the devices are measured is not
critical. Any standard can be re-measured, until the DONE key is pressed. The change in
trace during measurement of a standard is normal.
“Response” and “response and isolation” calibrations require measurement of only one
standard device. If more than one device is measured, only the data for the last device is
retained.
NOTE Since the 85032F calibration kit offsets are equal for both male and female
connectors, the standard’s sex type is not requested during the calibration.
6-4
Calibrating for Increased Measurement Accuracy
Calibration Considerations
• 90 to 100 dB: Isolation calibration is recommended with test port power greater than
0 dBm. For this isolation calibration, averaging should be turned on with an averaging
factor at least four times the measurement averaging factor. For example, use an
averaging factor of 16 for the isolation calibration, and then reduce the averaging factor
to four for the measurement after calibration.
• Greater than 100 dB: Same as 90 to 100 dB, but alternate mode should be used. See "To
View a Single Measurement Channel" on page 5-12.
Restarting a Calibration
If you interrupt a calibration to go to another menu, such as averaging, you can continue
the calibration by pressing the RESUME CAL SEQUENCE softkey in the correction
menu.
6- 5
Calibrating for Increased Measurement Accuracy
Calibration Considerations
Electrical Offset
Some standards have reference planes that are electrically offset from the mating plane of
the test port. These devices will show a phase shift with respect to frequency. Table 6-1
shows which reference devices exhibit an electrical offset phase shift. The amount of phase
shift can be calculated with the formula:
Φ = (360° × f × l)/c where:
f = frequency
l = electrical length of the offset
Fringe Capacitance
All open circuit terminations exhibit a phase shift over frequency due to fringe
capacitance. Offset open circuits have increased phase shift because the offset acts as a
small length of transmission line. Refer to Table 6-1.
6-6
Calibrating for Increased Measurement Accuracy
Calibration Considerations
Type-N male
2.4-mm female
Type-N female
7-mm Open
2.4-mm female
Type-N female
6- 7
Calibrating for Increased Measurement Accuracy
Calibration Considerations
Cor Cor
START .300 000 MHZ STOP 3 000 . 000 000 MHZ START .300 000 MHZ STOP 3 000 . 000 000 MHZ
Cor Cor
START .300 000 MHZ STOP 3 000 . 000 000 MHZ START .300 000 MHZ STOP 3 000 . 000 000 MHZ
pa5162e
6-8
Calibrating for Increased Measurement Accuracy
Calibration Considerations
NOTE The preset state of the instrument can be configured so that interpolated
error correction is on or off. Press System CONFIGURE MENU
USER SETTINGS PRESET SETTINGS CAL INTERP ON off to configure
the preset state of interpolated error correction.
System performance is unspecified when using interpolated error correction. The quality of
the interpolated error correction is dependent on the amount of phase shift and the
amplitude change between measurement points. If phase shift is no greater than 180° per
approximately five measurement points, interpolated error correction offers a great
improvement over uncorrected measurements. The accuracy of interpolated error
correction improves as the phase shift and amplitude change between adjacent points
decrease. When you use the analyzer in linear frequency sweep, perform the original
calibration with at least 30 points per 1 GHz of frequency span for greatest accuracy with
interpolated error correction.
Interpolated error correction is available in three sweep modes: linear frequency, power
sweep, and CW time.
NOTE If there is a valid correction array for a linear frequency sweep, this may be
interpolated to provide correction at the CW frequency used in power sweep
or CW time modes. This correction is part of the interpolated error-correction
feature.
6- 9
Calibrating for Increased Measurement Accuracy
Procedures for Error Correcting Your Measurements
NOTE If the channels are uncoupled, you must make a correction for each channel.
6-10
Calibrating for Increased Measurement Accuracy
Procedures for Error Correcting Your Measurements
S11 1-port Reflection of any Directivity, source Short, open and load
one-port device or match, frequency or ECal module.
well terminated response.
two-port device.
S22 1-port Reflection of any Directivity, source Short, open and load
one-port device or match, frequency or ECal module.
well terminated response.
two-port device.
6- 11
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
5. If your calibration kit is different than the kit specified under the Cal CAL KIT [ ]
softkey, press:
Cal CAL KIT SELECT CAL KIT (select your type of kit) RETURN
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying
Calibration Kits" on page 7-56.
6. To select a response correction, press:
Cal CALIBRATE MENU RESPONSE
Connect the short or open calibration standard to the port you selected for the test port
(PORT 1 for S11 or PORT 2 for S22).
NOTE Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device to the particular connector where you
will connect your DUT.
6-12
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
7. To measure the standard when the displayed trace has settled, press SHORT or
OPEN .
If the calibration kit you selected has a choice between male and female calibration
standards, remember to select the sex that applies to the test port and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the softkey that you selected after it finishes
the measurement, and computes the error coefficients.
NOTE This calibration allows only one standard to be measured. If you press the
wrong key for a standard, press RESPONSE again and choose the correct
standard. Do not use a thru standard for a reflection response correction.
NOTE You can save or store the measurement correction to use for later
measurements, that use the same measurement parameters. Refer to the
Chapter 4 , “Printing, Plotting, and Saving Measurement Results” for
procedures.
This completes the response correction for reflection measurements. You can connect
and measure your device under test.
6- 13
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
NOTE Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your DUT.
6-14
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
NOTE Do not use an open or short standard for a transmission response correction.
NOTE You can save or store the measurement correction to use for later
measurements. Refer to the Chapter 4 , “Printing, Plotting, and Saving
Measurement Results” for procedures.
7. This completes the response correction for transmission measurements. You can
connect and measure your device under test.
Receiver Calibration
Receiver calibration provides a frequency response error correction for a non-ratioed
measurement that also indicates absolute power in dBm. This calibration is most useful
when performed with a power meter calibration. This calibration is only allowed for
non-ratioed measurements A, B, and R.
This calibration normalizes the trace to the current reference value. Typically, this
reference value is entered to be the same as the current source power.
1. Perform a power meter calibration to the desired level. Refer to step A of Figure 6-4.
Use −10 dBm for this example. (See also, "Power Meter Measurement Calibration" on
page 6-33.) This provides a calibrated power, referenced to the power meter, to use as a
receiver calibration standard.
or
Set the analyzer test port power to the desired level (−10 dBm in this example) by
pressing:
Power −10 x1 . This calibrates the receiver to the approximate accuracy of the
source output power, which is subject to the source power flatness specification.
2. Make a "thru" connection between the points where you will connect your device under
test. Refer to Step B of Figure 6-4.
NOTE Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your device under
test.
6- 15
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
NOTE You can save or store the measurement correction to use for later
measurements. Refer to Chapter 4 , “Printing, Plotting, and Saving
Measurement Results” for procedures.
6. This completes the receiver calibration for transmission measurements. You can
connect and measure your device under test.
NOTE The accuracy of the receiver calibration will be nearly the same as the test
port power accuracy; and the test port power accuracy can be significantly
improved by performing a power meter source calibration, as described later
in "Power Meter Measurement Calibration" on page 6-33.
Calibrations at powers other than 0 dBm are possible. Receiver calibration normalizes the
trace to the value set for the reference level. For example, to do a receiver calibration at
−10 dBm, set the source to −10 dBm, set the reference level to −10 dBm, then perform the
receiver calibration.
6-16
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
NOTE Although you can perform a response and isolation correction for reflection
measurements, we recommend that you perform an S11 one-port error
correction; it is more accurate and just as convenient.
1. Press Preset .
2. Select the type of measurement you want to make.
❏ If you want to make a transmission measurement in the forward direction (S21),
press:
Meas Trans: FWD S21 (B/R)
❏ If you want to make a transmission measurement in the reverse direction (S12),
press:
Meas Trans: REV S12 (A/R)
3. Set any other measurement parameters that you want for the device measurement:
power, number of points, IF bandwidth.
4. To access the measurement correction menus, press:
Cal
5. If your calibration kit is different than the kit specified under the CAL KIT [ ] softkey,
press:
CAL KIT SELECT CAL KIT (select your type of kit) RETURN
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying
Calibration Kits" on page 7-56.
6. To select a response and isolation correction and to start the response portion of the
calibration, press:
CALIBRATE MENU RESPONSE & ISOL’N RESPONSE
6- 17
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
7. Make a "thru" connection between the points where you will connect your device under
test.
NOTE Include any adapters that you will have in the device measurement. That is,
connect the standard device to the particular connector where you will
connect your device under test.
8. To measure the standard, when the displayed trace has settled, press:
THRU
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the THRU softkey after it measures the
calibration standard, and computes the error coefficients.
9. Connect impedance-matched loads to PORT 1 and PORT 2, as shown in Figure 6-5.
Include the adapters that you would include for your device measurement.
NOTE If you will be measuring highly reflective devices, such as filters, use the test
device, connected to the reference plane and terminated with a load, for the
isolation standard.
6-18
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
NOTE You can save or store the measurement correction to use for later
measurements. Refer to Chapter 4 , “Printing, Plotting, and Saving
Measurement Results” for procedures.
14.This completes the response and isolation correction for transmission measurements.
You can connect and measure your device under test.
1. Press Preset .
2. Select the type of measurement you want to make.
❏ If you want to make a reflection measurement on PORT 1 (in the forward direction,
S11), leave the instrument default setting.
❏ If you want to make a reflection measurement on PORT 2 (in the reverse direction,
S22), press:
6- 19
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
5. If your calibration kit is different than the kit specified under the CAL KIT [ ] softkey,
press:
CAL KIT SELECT CAL KIT (select your type of kit) RETURN
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying
Calibration Kits" on page 7-56.
6. To select a response and isolation correction and to start with the response portion of
the calibration, press:
CALIBRATE MENU RESPONSE & ISOL’N RESPONSE
7. Connect the short or open calibration standard to the port you selected for the test port
(PORT 1 for S11 or PORT 2 for S22).
NOTE Include any adapters that you will have in the device measurement. That is,
connect the standard device to the particular connector where you will
connect your device under test.
6-20
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
NOTE You can save or store the error correction to use for later measurements.
Refer to Chapter 4 , “Printing, Plotting, and Saving Measurement Results” for
procedures.
12.This completes the response and isolation error correction for reflection measurements.
You can connect and measure your device under test.
6- 21
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
IMPORTANT Use enhanced reflection error correction only on bilateral devices. A bilateral
device has similar forward and reverse transmission characteristics.
Examples of bilateral devices are passive devices (filters, attenuators, and
switches). Most active devices (amplifiers) and some passive devices (isolators
and circulators) are not bilateral. If this error correction is used for a
non-bilateral device, errors will occur in the resulting measurement.
1. Press Preset .
2. Select the type of measurement you want to make.
❏ If you want to make measurements in the forward direction (S21 S11), press:
5. If your calibration kit is different than the kit specified under the CAL KIT [ ] softkey,
press:
CAL KIT SELECT CAL KIT (select your type of kit) RETURN
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying
Calibration Kits" on page 7-56.
6. To select the correction type, press CALIBRATE MENU ENHANCED RESPONSE
and select the correction type.
6-22
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
NOTE Include any adapters that you will have in the device measurement. That is,
connect the standard to the particular connector where you will connect your
device under test.
8. To measure the standard, when the displayed trace has settled, press:
OPEN
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the OPEN softkey after it measures the
standard.
9. Disconnect the open, and connect a short circuit to the test port.
10.To measure the device, when the displayed trace has settled, press:
SHORT
The analyzer measures the short circuit and underlines the SHORT softkey.
11.Disconnect the short, and connect an impedance-matched load to the test port.
6- 23
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
12.To measure the standard, when the displayed trace has settled, press:
LOADS , select the type of load you are using, and then press DONE: LOADS when
the analyzer has finished measuring the load.
Notice that the LOADS softkey is now underlined.
13.To compute the reflection correction coefficients, press STANDARDS DONE .
14.To start the transmission portion of the correction, press TRANSMISSION .
15.Make a "thru" connection between the points where you will connect your device under
test as shown in Figure 6-7.
NOTE Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your device under
test.
NOTE The thru in most calibration kits is defined with zero length. The correction
will not work properly if a non-zero length thru is used, unless the calibration
kit is modified to change the defined thru to the length used. This is
important for measurements of non-insertable devices (devices having ports
that are both male or both female). The modified calibration kit must be
saved as the user calibration kit, and the USER KIT softkey must be
selected before the calibration is started.
16.To measure the standard, when the trace has settled, press:
FWD TRANS THRU or REV TRANS THRU
STANDARDS DONE
The analyzer underlines the softkey label after it makes each measurement.
17.Press ISOLATION and select from the following two options:
❏ If you will be measuring devices with a dynamic range less than 90 dB, press:
OMIT ISOLATION
❏ If you will be measuring devices with a dynamic range greater than 90 dB, follow
these steps:
a. Connect impedance-matched loads to the test ports. Include the adapters that you
would include for your device measurement.
NOTE If you will be measuring highly reflective devices such as filters, use the test
device, connected to the reference plane and terminated with a load, for the
isolation standard.
6-24
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
b. Activate at least four times more averages than desired during the device
measurement.
c. Press Cal RESUME CAL SEQUENCE ISOLATION FWD or REV
ISOL’N STD DONE .
d. Return the averaging to the original state of the measurement, and press Cal
RESUME CAL SEQUENCE .
18.To compute the error coefficients, press DONE ENH RESP CAL .
The analyzer displays the corrected measurement trace. The analyzer also shows the
notation Cor at the left of the screen, indicating that error correction is on.
NOTE You can save or store the measurement correction to use for later
measurements. Refer to Chapter 4 , “Printing, Plotting, and Saving
Measurement Results” for procedures.
20.This completes the enhanced response correction procedure. You can connect and
measure your device under test.
6- 25
Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
NOTE This is the recommended error correction process for all reflection
measurements, when full two-port correction or enhanced response
calibration is not used.
1. Press Preset .
2. Select the type of measurement you want to make.
❏ If you want to make a reflection measurement on PORT 1 (in the forward direction,
S11), leave the instrument default setting.
❏ If you want to make a reflection measurement on PORT 2 (in the reverse direction,
S22), press:
5. If your calibration kit is different than the kit specified under the CAL KIT [ ] softkey,
press:
CAL KIT SELECT CAL KIT (select your type of kit) RETURN
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying
Calibration Kits" on page 7-56.
6. To select the correction type, press CALIBRATE MENU and select the correction type.
❏ If you want to make a reflection measurement at PORT 1, press:
S11 1-PORT
❏ If you want to make a reflection measurement at PORT 2, press:
S22 1-PORT
7. Connect a shielded open circuit to PORT 1 (or PORT 2 for an S22 measurement).
6-26
Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
NOTE Include any adapters that you will have in the device measurement. That is,
connect the calibration standard to the particular connector where you will
connect your device under test.
Figure 6-8 Standard Connections for a One Port Reflection Error Correction
8. To measure the standard, when the displayed trace has settled, press: OPEN
NOTE If the calibration kit that you selected has a choice between male or female
calibration standards, remember to select the sex that applies to the test port
and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the OPEN softkey after it measures the
calibration standard.
9. Disconnect the open, and connect a short circuit to the test port.
10.To measure the standard when the displayed trace has settled, press:
SHORT
The analyzer measures the short circuit and underlines the SHORT softkey.
11.Disconnect the short, and connect an impedance-matched load to the test port.
12.When the displayed trace has settled, press:
LOADS , select the type of load you are using, and then press DONE: LOADS when
the analyzer has finished measuring the load.
Notice that the LOADS softkey is now underlined.
6- 27
Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
NOTE The open, short, and load could be measured in any order, and need not follow
the order in this example.
NOTE You can save or store the error correction to use for later measurements.
Refer to Chapter 4 , “Printing, Plotting, and Saving Measurement Results” for
procedures.
14.This completes the one-port correction for reflection measurements. You can connect
and measure your device under test.
6-28
Calibrating for Increased Measurement Accuracy
Full Two-Port Error Correction
NOTE This is the most accurate error-correction procedure. Since the analyzer takes
both forward and reverse sweeps to update one measurement trace, this
procedure takes more time than the other correction procedures.
1. Set any measurement parameters that you want for the device measurement: power,
format, number of points, or IF bandwidth.
2. To access the measurement correction menus, press:
Cal
3. If your calibration kit is different than the kit specified under the CAL KIT [ ] softkey,
press:
CAL KIT SELECT CAL KIT (select your type of kit) RETURN
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying
Calibration Kits" on page 7-56.
4. To select the correction type, press:
CALIBRATE MENU FULL 2-PORT REFLECTION
5. Connect a shielded open circuit to PORT 1.
NOTE Include any adapters that you will have in the device measurement. That is,
connect the standard to the particular connector where you will connect your
DUT.
6- 29
Calibrating for Increased Measurement Accuracy
Full Two-Port Error Correction
6. To measure the standard, when the displayed trace has settled, press:
FORWARD: OPEN
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the OPEN softkey after it measures the
standard.
7. Disconnect the open, and connect a short circuit to PORT 1.
8. To measure the device, when the displayed trace has settled, press:
FORWARD: SHORT
The analyzer measures the short circuit and underlines the SHORT softkey.
9. Disconnect the short, and connect an impedance-matched load to PORT 1.
10.To measure the standard, when the displayed trace has settled, press:
FORWARD: LOAD , select the type of load you are using, and then press
DONE: LOADS when the analyzer has finished measuring the load.
6-30
Calibrating for Increased Measurement Accuracy
Full Two-Port Error Correction
14.Make a "thru" connection between the points where you will connect your device under
test as shown in Figure 6-9.
NOTE Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your DUT.
NOTE The thru in most calibration kits is defined with zero length. The correction
will not work properly if a non-zero length thru is used, unless the calibration
kit is modified to change the defined thru to the length used. This is
important for measurements of non-insertable devices (devices having ports
that are both male or both female). The modified calibration kit must be
saved as the user calibration kit, and USER KIT must be selected before the
calibration is started.
15.To measure the standard, when the trace has settled, press:
DO BOTH FWD+REV
The analyzer underlines the softkey label after it makes each measurement.
16.Press ISOLATION and select from the following two options:
❏ If you will be measuring devices with a dynamic range less than 90 dB, press:
OMIT ISOLATION
❏ If you will be measuring devices with a dynamic range greater than 90 dB, follow
these steps:
a. Connect impedance-matched loads to PORT 1 and PORT 2. Include the adapters
that you would include for your device measurement.
NOTE If you will be measuring highly reflective devices, such as filters, use the test
device, connected to the reference plane and terminated with a load, for the
isolation standard.
b. Activate at least four times more averages than desired during the device
measurement.
NOTE If loads can be connected to both port 1 and port 2 simultaneously, then the
following step can be performed using the DO BOTH FWD + REV softkey.
d. Return the averaging to the original state of the measurement, and press Cal
RESUME CAL SEQUENCE .
6- 31
Calibrating for Increased Measurement Accuracy
Full Two-Port Error Correction
NOTE You can save or store the measurement correction to use for later
measurements. Refer to Chapter 4 , “Printing, Plotting, and Saving
Measurement Results” for procedures.
18.This completes the full two-port correction procedure. You can connect and measure
your device under test.
6-32
Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
6- 33
Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
6-34
Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
❏ If you are modifying the frequency, enter the new value, followed by a G/n , M/µ ,
or k/m key.
❏ If you are modifying the correction factor, enter the new value, followed by the x1
key.
4. Press DONE after you have finished modifying the segment.
5. If you want to edit any other segments, press SEGMENT and follow the previous steps,
starting with step 2.
5. Press DONE when you are finished modifying the segment list.
6- 35
Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
NOTE Remember to subtract the through arm loss from the coupler arm loss before
entering it into the power loss table, to ensure the correct power at the output
of the coupler.
4. Repeat the previous two steps to enter up to 12 frequency segments, depending on the
required accuracy.
You may enter multiple segments in any order because the analyzer automatically sorts
them and lists them on the display in increasing order of frequency.
If you only enter one frequency segment, the analyzer assumes that the single value is
valid over the entire frequency range of the correction.
5. After you have entered all the segments, press DONE .
6. Press Cal PWRMTR CAL PWR LOSS ON to activate the power loss compensation.
6-36
Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
Local
SYSTEM CONTROLLER
4. Set the power meter’s address (“XX” represents the address in the following keystrokes:
SET ADDRESSES
ADDRESS: P MTR/GPIB XX x1
5. Select the appropriate power meter by pressing POWER MTR [ ] until the correct
model number is displayed (436A or 438A/437).
NOTE The E4418B and E4419B power meters have a “437 emulation” mode. This
allows these power meters, with an HP/Agilent 848X-series power sensor, to
be used with the network analyzer. In this step, when selecting a power
meter, choose the 438A/437 selection.
NOTE Because power meter calibration requires a longer sweep time, you may want
to reduce the number of points before pressing TAKE CAL SWEEP . After the
power meter calibration is finished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration. Some
accuracy will be lost for the interpolated points.
The analyzer will use the data table for subsequent sweeps to correct the output power
level at each measurement point. Also, the status annunciator PC will appear on the
analyzer display.
NOTE You can abort the calibration sweep by pressing PWRMTR CAL OFF .
10.Remove the power sensor from the analyzer test port and connect your test device.
6- 37
Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
NOTE You may level at the input of a device under test, using a 2-resistor power
splitter or a directional coupler before the device; or level at the output of the
device using a 3-resistor power splitter or a bidirectional coupler after the
device.
1. Connect a power splitter or directional coupler to the port supplying RF power to your
test device, as shown in Figure 6-11.
2. Set test port power to approximate desired leveled power.
3. Press Cal PWRMTR CAL and enter the test port power level that you want the
analyzer to maintain at the input to your test device. Compensate for the power loss of
the power splitter or directional coupler in the setup.
4. If you want the analyzer to make more than one power measurement at each frequency
data point, press NUMBER OF READINGS n x1 (where n = the number of desired
iterations).
If you increase the number of readings, the power meter correction time will
substantially increase.
5. Press Cal PWRMTR CAL EACH SWEEP TAKE CAL SWEEP to activate the
power meter correction.
6-38
Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
NOTE Because power meter calibration requires a longer sweep time, you may want
to reduce the number of points before pressing TAKE CAL SWEEP . After the
power meter calibration is finished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration.
The status notation PC will appear on the analyzer display. Port 1 is now a calibrated
source of power.
4. Connect the test port 1 output to the test port 2 input.
5. Choose a non-ratioed measurement by pressing:
Meas INPUT PORTS B TEST PORT 1
This sets the source at PORT 1, and the measurement receiver to PORT 2, or input port
B.
6. To perform a receiver error correction, press:
Cal CALIBRATE MENU RECEIVER CAL (enter power level) x1
TAKE RCVR CAL SWEEP
The receiver channel now measures power to a characteristic accuracy of 0.35 dB or
better. The accuracy depends on the match of the power meter, the source, and the
receiver.
6- 39
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
6-40
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
NOTE Adapter A1 and Adapter A2 become part of the test setup to allow connection
to the DUT. Adapter A3 is used during calibration only. Its effects will be
removed.
6- 41
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
c. Connect the A3 adapter to "Reference Port 1" as shown in Step B of Figure 6-14.
Attach the short (from the calibration kit for port 2) to the other end of the adapter.
You must know the delay of the short. The delay of the short can be found in the
calibration kit that you are using. Typical delays of shorts are 31.7 ps for the short
from the 85052D calibration kit and 31.8 ps for the short from the 85033D
calibration kit.
d. Measure the delay of the adapter and short together by pressing Format DELAY .
e. Divide the resulting delay measurement by 2 to determine the delay of the thru and
the short in one direction.
f. Subtract the offset delay of the short (determined in step c) from the delay of the thru
and the short in one direction (determined in step e). The result is the electrical
delay of the thru. This value is used in the Step 12.
g. Remove the short from the adapter.
6-42
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
NOTE You must use the floppy disk to store the following calibrations. Select the
floppy disk by pressing Save/Recall SELECT DISK INTERNAL DISK.
3. Connect adapter A3 (same sex and connector type as the DUT) to adapter A2 on port 2
as shown in Figure 6-15.
4. Perform a full 2-port calibration between ports 1 and 2 using calibration standards
appropriate for the connector type at port 1 (the connector type for adapter A1). Save
the calibration by selecting Save/Recall SAVE STATE . Name the file "PORT1."
5. Connect adapter A3 to adapter A1 on port 1 as shown in Figure 6-16.
6- 43
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
6. Perform a full 2-port calibration between ports 1 and 2 using calibration standards
appropriate for the connector type at port 2 (the connector type for adapter A2). Save
the calibration by selecting Save/Recall SAVE STATE . Name the file "PORT2."
NOTE In the following steps, calibration data is recalled, not instrument states.
6-44
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
6- 45
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
Matched Adapters
With this method, you use two precision matched adapters which are "equal." To be equal,
the adapters must have the same match, Z0, insertion loss, and electrical delay. The
adapters in most Agilent calibration kits have matched electrical length, even if the
physical lengths appear different.
To use this method, refer to Figure 6-18 and perform the following steps:
1. Perform a transmission calibration using the first adapter.
2. Remove adapter A, and place adapter B on port 2. Adapter B becomes the effective test
port.
3. Perform a reflection calibration.
4. Measure the test device with adapter B in place.
The errors remaining after calibration with this method are equal to the differences
between the two adapters that are used.
6-46
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
6- 47
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
8. Modify the calibration kit thru definition by entering in the electrical delay of adapter
A3. Save this as a user kit.
For example, if A3 has 100 ps of delay, press:
Cal CAL KIT MODIFY DEFINE STANDARD 4 x1
STD DONE (DEFINED) RETURN KIT DONE (MODIFIED) SAVE USER KIT
9. Perform the desired calibration with this new user kit.
10.Connect the test device as shown in Figure 6-17 and measure the device.
6-48
Calibrating for Increased Measurement Accuracy
Minimizing Error When Using Adapters
6- 49
Calibrating for Increased Measurement Accuracy
Making Non-Coaxial Measurements
Fixtures
Fixtures are needed to interface non-coaxial devices to coaxial test instruments. It may
also be necessary to transform the characteristic impedance from standard 50 Ω
instruments to a non-standard impedance and to apply bias if an active device is being
measured.
For accurate measurements, the fixture must introduce minimum change to the test
signal, not destroy the test device, and provide a repeatable connection to the device.
For information about test fixtures for your measurement systems, ask for literature
number 5962-9723E or contact: Inter-Continental Microwave, 1515 Wyatt Drive, Santa
Clara, CA 95054, USA (Web site: http://www.icmicrowave.com).
6-50
Calibrating for Increased Measurement Accuracy
Making Non-Coaxial Measurements
6- 51
Calibrating for Increased Measurement Accuracy
Calibrating for Non-Coaxial Devices
DEFINE STANDARD 4
DELAY/THRU x1 MODIFY STD DEFINITION
SPECIFY OFFSET OFFSET DELAY 0 x1 STD OFFSET DONE
STD DONE (DEFINED)
5. To define the LINE/MATCH standard, press:
DEFINE STANDARD 6 x1
DELAY/THRU MODIFY STD DEFINITION
SPECIFY OFFSET OFFSET DELAY .08 G/n MAXIMUM FREQUENCY
Enter a frequency greater than the maximum frequency range of the analyzer. For
example, press 10 G/n . Then, press STD OFFSET DONE .
6. For the purposes of this example, change the name of the standard by pressing
LABEL STD and modifying the name to "LINE."
6-52
Calibrating for Increased Measurement Accuracy
Calibrating for Non-Coaxial Devices
TRL THRU 4 x1
12.To complete the specification of class assignments, press:
SPECIFY CLASS DONE
Label the Classes
NOTE To enter the following label titles, an external keyboard may be used for
convenience.
NOTE Refer to "Saving Modified Calibration Kits to a Disk" on page 7-65 for
information about saving modified calibration kits, along with calibration
data and instrument states, to a disk.
6- 53
Calibrating for Increased Measurement Accuracy
Calibrating for Non-Coaxial Devices
1. Press Cal CAL KIT SELECT CAL KIT USER KIT RETURN RETURN
CALIBRATE MENU TRL*/LRM* 2-PORT .
2. To measure the "TRL THRU," connect the "zero length" transmission line between the
two test ports.
3. To make the necessary four measurements, press THRU THRU .
4. To measure the "TRL SHORT," connect the short to PORT 1, and press:
S11 REFL: TRLSHORT
5. Connect the short to PORT 2 and press S22 REFL: TRLSHORT .
6. To measure the "TRL LINE," disconnect the short and connect the TRL line from PORT
1 to PORT 2.
7. Press LINE/MATCH DO BOTH FWD+REV .
8. The line data is measured and the LN/MATCH1 LINE and LN/MATCH2 LINE softkey
labels are underlined.
9. To measure the "ISOLATION" class, press ISOLATION .
❏ You could choose not to perform the isolation measurement by pressing
OMIT ISOLATION DONE TRL/LRM .
NOTE You should perform the isolation measurement when the highest dynamic
range is desired. To perform the best isolation measurements, you should
reduce the system bandwidth or activate the averaging function.
A poorly measured isolation class can actually degrade the overall
measurement performance. If you are in doubt of the isolation measurement
quality, you should omit the isolation portion of this procedure.
NOTE If loads can be connected to both port 1 and port 2 simultaneously, then the
following measurement can be performed using the DO BOTH FWD+REV
softkey.
6-54
Calibrating for Increased Measurement Accuracy
Calibrating for Non-Coaxial Devices
NOTE You can save or store the measurement correction to use for later
measurements. Refer to Chapter 4 , “Printing, Plotting, and Saving
Measurement Results” for procedures.
13.Connect the device under test. The device S-parameters are now being measured.
6- 55
Calibrating for Increased Measurement Accuracy
LRM Error Correction
NOTE LRM with a zero length line is sometimes referred to as TRM (THRU,
REFLECT, MATCH).
DEFINE STANDARD 4 x1
DELAY/THRU MODIFY STD DEFINITION
SPECIFY OFFSET OFFSET DELAY 0 x1 STD OFFSET DONE
STD DONE (DEFINED)
5. To define the LINE/MATCH standard, press:
6-56
Calibrating for Increased Measurement Accuracy
LRM Error Correction
TRL THRU 4 x1
12.To complete the specification of class assignments, press:
SPECIFY CLASS DONE
Label the Classes
NOTE To enter the following label titles, an external keyboard may be used for
convenience.
NOTE Refer to "Saving Modified Calibration Kits to a Disk" on page 7-65 for
information about saving modified calibration kits, along with calibration
data and instrument states, to a disk.
6- 57
Calibrating for Increased Measurement Accuracy
LRM Error Correction
2. Press Cal CAL KIT SELECT CAL KIT USER KIT RETURN RETURN
CALIBRATE MENU TRL*/LRM* 2-PORT .
3. To measure the "LRM THRU," connect the "zero length" transmission line between the
two test ports.
4. To make the necessary four measurements, press:
LRMTHRU
5. To measure the "LRM SHORT," connect the short to PORT 1, and press:
S11 REFL LRMSHORT
6. Connect the short to PORT 2, and press:
S22 REFL LRMSHORT
NOTE If loads can be connected to both port 1 and port 2 simultaneously, then the
following LRM load measurement can be performed using the
DO BOTH FWD+REV softkey.
7. To measure the "LRM LOAD," disconnect the short and connect the LRM load to
PORT1.
8. Press LINE/MATCH LN/MATCH1 LOAD to access the No Loads menu. When the
displayed trace settles, press the softkey corresponding to the load used. If a sliding load
is used, press SLIDING to access the Sliding Load menu. Position the slide and press
SLIDE IS SET .
9. When all the appropriate load measurements are complete, the load data is measured
and the LN/MATCH1 LOAD softkey label is underlined.
10.Connect the load to PORT 2 and press LN/MATCH2 LOAD .
11.Repeat the previous LRM load measurement steps for PORT 2.
12.After the measurement is complete, press:
DONE LINE/MATCH
13.To measure the "ISOLATION" class, press:
ISOLATION
❏ You could choose not to perform the isolation measurement by pressing
OMIT ISOLATION DONE TRL/LRM .
6-58
Calibrating for Increased Measurement Accuracy
LRM Error Correction
NOTE You should perform the isolation measurement when the highest dynamic
range is desired.
To perform the best isolation measurements, you should reduce the system
bandwidth or activate the averaging function.
A poorly measured isolation class can actually degrade the overall
measurement performance. If you are in doubt of the isolation measurement
quality, you should omit the isolation portion of this procedure.
14.You may repeat any of the previous steps. There is no requirement to go in the order of
steps. When the analyzer detects that you have made all the necessary measurements,
the message line will show PRESS ’DONE’ IF FINISHED WITH CAL. Press
DONE TRL/LRM .
The message COMPUTING CAL COEFFICIENTS will appear, indicating that the analyzer is
performing the numerical calculations of error coefficients.
NOTE You can save or store the measurement correction to use for later
measurements. Refer to Chapter 4 , “Printing, Plotting, and Saving
Measurement Results” for procedures.
15.Connect the device under test. The device S-parameters are now being measured.
NOTE When making measurements using the same port with uncoupled channels,
the power level for each channel must fall within the same power range
setting of that single port. An error message will be displayed if you enter two
power levels that do not fall within the same power range.
6- 59
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
6-60
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
CAUTION Only connect the DB25 cable to the Parallel Port connector of the network
analyzer.
If the cable is connected to the Test Set - I/O Interconnect connector, damage
to the PC interface unit could occur.
NOTE For steps 3 through 5, use the three DB25 cables (part number 8120-8710)
shipped with the 85097A Electronic Calibration System. Other cables may
not give reliable results.
4. Connect one end of a DB25 cable to the connector on the PC interface unit labeled
"DB25 Interface to ECal Module A." Connect the other end of the DB25 cable to the
parallel cable connector on the ECal module.
6- 61
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
5. If you need to calibrate with a second ECal module, connect one end of another DB25
cable to the connector on the PC interface unit labeled "DB25 Interface to ECal
Module B". Connect the other end of the DB25 cable to the parallel cable connector on
the ECal module.
6. Using an RF cable (or a microwave cable, as appropriate), connect one port of the ECal
module to test port 1 of the analyzer. Refer to Figure 6-21.
CAUTION RF ECal modules can be damaged if you apply excessive torque to the
connectors. Do not exceed the recommended torque indicated in the
Electronic Calibration Module Reference Guide (part number 85091-90009).
NOTE It is not critical which ECal module port (Port A or Port B) is connected to the
network analyzer test ports. The network analyzer detects where each ECal
module port is connected and uses the appropriate module data. If Port A is
connected to Port 1 of the analyzer, the calibration will be performed slightly
faster.
7. Using an RF cable (or a microwave cable, as appropriate), connect the other port of the
ECal module to test port 2 of the analyzer.
6-62
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
NOTE If you are performing an ECal using two modules, selecting either option sets
up the option to be performed with both modules.
1. To select the manual thru calibration option, press Cal ECal MENU
CONFIGURE [ ] .
1. To select the isolation calibration option, press Cal ECal MENU CONFIGURE [ ] .
NOTE The isolation averaging default is set to take 10 sweeps. This isolation
averaging value is less than the network analyzer default sweep averaging
value of 16.
4. Press RETURN .
5. Continue with step 2 of “Perform the Calibration.”
6- 63
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
NOTE If you are calibrating with two modules, the overlapping frequency span will
be determined by the second module. Therefore, if you want to use the
calibration data of Module A in the overlapping frequency span, calibrate
using Module B first, and then calibrate using Module A.
6-64
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
5. After you connect the manual thru, press CONTINUE ECal to complete the manual
thru portion of the ECal.
6. If you are calibrating using two ECal modules, a prompt is displayed directing you to
remove the first module and connect the second module. Follow this prompt as shown in
Figure 6-23.
6- 65
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
7. After you connect the second ECal module, press CONTINUE ECal to continue the
ECal.
8. Repeat steps 4 and 5 if you selected to calibrate using the manual thru option.
9. Review the displayed calibration results. Refer to "Perform the Confidence Check" on
page 6-67.
10.Save the calibration results by pressing Save/Recall .
6-66
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
NOTE A confidence check is only valid after an ECal has been performed.
• S11 1-PORT
• S22 1-PORT
6- 67
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
5. Press TRACE TYPE [ ] until the calibration confidence check trace that you want to
view is displayed.
Pressing the TRACE TYPE [ ] softkey toggles between the five trace-type display
options. The confidence check can display the measured ECal results (DATA) and the
premeasured calibration data (MEM) in following five ways.
• DATA&MEM displays two traces representing the measured ECal results and
module's premeasured calibration data trace.
• DATA/MEM displays a single trace representing the ratio of the measured ECal
results to the module's premeasured calibration data.
• DATA-MEM displays a single trace representing the difference between the
measured ECal results and the module's premeasured calibration data.
• DATA displays a single trace representing only the measured ECal results.
• MEM displays a single trace representing only the module's premeasured
calibration data.
6. If you want to change the scale of the display, press AUTO SCALE .
The AUTO SCALE softkey is located in this menu for convenience in viewing the
confidence check data. It acts the same as AUTO SCALE under the Scale Ref key.
7. Review the confidence check display.
8. If you want to check other calibration S-parameters or trace types, repeat steps 4
through 7.
9. When finished, select RETURN to complete the confidence check.
6-68
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
CAUTION The confidence check described in the previous section displays the ECal data
of a single state. This confidence state is a calibrated standard not used
during ECal. It is provided to give an independent assessment of the quality
of a calibration. In the Ecal Service menu, you may also display each of the
calibration standards which are used during an ECal calibration, along with
the analyzer’s measurement of those standards.
You may notice a difference in measurement results when comparing ECal
confidence state data and ECal standard state data. This result may be
related to certain measurement errors with the network analyzer system,
which add uncertainty to measurement results. More detailed information
regarding measurement uncertainty is documented in the “Determining
System Measurement Uncertainties” chapter of the network analyzer's
reference guide. Additional information on improving your measurements
can be found in Chapter 5 , “Optimizing Measurement Results.”
Using the ECal Service menu is not a standard part of the ECal procedure. It is a tool to
allow you to identify problems in the calibration equipment, cables, connectors, or
procedures.
The Confidence Check menu supports the comparison of the measured data versus the
module's premeasured calibration data for the confidence state. The ECal Service menu
supports the comparison of the measured data versus the module's premeasured
calibration data for the remainder of the module states, in addition to the confidence state.
Access the ECal Service menu by pressing ECal SERVICE from the Confidence Check
menu.
The ECal Service menu softkeys are:
ECal STD [ ] Toggles the analyzer to show the data for the following calibration
states:
• CONF (Confidence)
• THRU
• ISOL (Isolation)
• S11 REFL (S11 Reflection)
• S22 REFL (S22 Reflection)
REFL STD [ ] Toggles between the available S11 and S22 reflection states listed in
the ECal STD [ ] softkey described above.
• S11 REFL allows up to 13 reflection states.
• S22 REFL allows up to 13 reflection states.
6- 69
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
NOTE When there is no premeasured calibration data for a given state and
measurement parameter, a warning is displayed indicating that no module
date is available.
PARAMETER [ ] Toggles the analyzer to show the data for the following S-parameters:
• S11
• S21
• S12
• S22
AUTO SCALE Changes scale and reference values to bring the trace data in view on
the display. The analyzer determines the smallest possible scale factor
that will put all displayed data onto 80% of the vertical graticule. The
reference value is chosen to put the trace in center screen.
NOTE When returning to the Confidence Check menu from the ECal Service menu,
press the SET CONF STANDARD softkey on the ECal Confidence Check
menu. If this softkey is not pressed, the confidence check information
displayed may not be accurate.
6-70
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal
The adapter removal calibration technique may be used with the ECal calibration to
accurately measure noninsertable devices. The following adapters (shown in Figure 6-26)
are needed.
• Adapter A1, which mates with port 1 of the DUT, must be installed on Port 1 of the
analyzer.
• Adapter A2, which mates with port 2 of the DUT, must be installed on Port 2 of the
analyzer.
• Adapter A3 must match the connectors on the DUT. The effects of this adapter will be
completely removed with this calibration technique.
NOTE Adapters A1 and A2 become part of the test setup to allow connection to the
DUT. Adapter A3 is used during the calibration only. Its effects will be
removed.
6- 71
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal
6-72
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal
4. Press FULL 2-PORT to perform the first 2-port error correction using the ECal
module.
NOTE When using adapter removal calibration, you must save calibration sets to
the internal disk, not to internal memory.
6- 73
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal
9. Press FULL 2-PORT to perform the second 2-port error correction using the ECal
module.
10.Save the results to disk. Name the file "PORT2."
11.Determine the electrical delay of adapter A3. If you have adapter specifications that
identify the electrical delay, you may use that information and continue with "Remove
the Adapter" on page 6-76. If you do not know the delay of adapter A3, perform the
“Determine the Electrical Delay” procedure on page 6-75.
6-74
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal
6- 75
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal
HELP ADAPT REMOVAL (This Help softkey provides a quick reference guide
to using the adapter removal technique.)
RECALL CAL SETS
ADAPTER DELAY
ADAPTER COAX
ADAPTER WAVEGUIDE
REMOVE ADAPTER
NOTE In the following two steps, calibration data is recalled, not instrument states.
3. From the disk directory, choose the file associated with the port 1 error correction, then
press RECALL CAL PORT 1 .
4. When this is complete, choose the file for the port 2 error correction and press
RECALL CAL PORT 2 .
5. When complete, press RETURN .
6. Enter the value of adapter A3 electrical delay by pressing ADAPTER DELAY and
entering the value.
7. Select the appropriate key: ADAPTER COAX or ADAPTER WAVEGUIDE .
8. Press REMOVE ADAPTER to complete the technique for calculating the new error
coefficients and overwrite the current active calibration set in use.
This process uses up an internal memory register. The calibration in this register is not
the calibration created by adapter removal, rather it is a "scratch" calibration. You may
wish to delete the register, or re-save the new calibration in this register as shown in
the following step.
9. To save the results of the new calibration set, press Save/Recall SELECT DISK
INTERNAL MEMORY RETURN SAVE STATE .
NOTE Adapter removal can leave a residual state in internal memory. This is not a
valid instrument state and should be deleted.
6-76
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal
10.Connect the DUT to the network analyzer as shown in Figure 6-30 to perform
calibrated measurements.
6- 77
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal
6-78
7 Operating Concepts
7- 1
Operating Concepts
Using This Chapter
7-2
Operating Concepts
System Operation
System Operation
Network analyzers measure the reflection and transmission characteristics of devices and
networks. A network analyzer test system consists of the following:
• source
• signal-separation devices
• receiver
• display
The analyzer applies a signal that is transmitted through the test device, or reflected from
its input, and then compares it with the incident signal generated by the swept RF source.
The signals are then applied to a receiver for measurement, signal processing, and display.
The vector network analyzer integrates a high resolution synthesized RF source, test set,
and a dual channel three-input receiver to measure and display magnitude, phase, and
group delay of transmitted and reflected power. With Option 010, the analyzer has the
additional capability of transforming measured data from the frequency domain to the
time domain. Other options are explained in “Options and Accessories” chapter of the
reference guide. A simplified block diagram of the network analyzer system is shown in
Figure 7-1. A detailed block diagram of the analyzer is provided in the service guide
together with a theory of system operation.
7- 3
Operating Concepts
System Operation
Test Sets
A test set provides connections to the test device, as well as to the signal-separation
devices. The signal separation devices are needed to separate the incident signal from the
transmitted and reflected signals. The incident signal is applied to the R channel input.
Meanwhile, the transmitted and reflected signals are applied to the A or B inputs. Port1 is
connected to the A input and port 2 is connected to the B input (does not apply to the
HP/Agilent 85044A/B T/R test sets).
The HP/Agilent 85046A/B and HP/Agilent 85047A S-parameter test sets contain the
hardware required to make simultaneous transmission and reflection measurements in
both the forward and reverse directions. An RF path switch in the test set allows reverse
measurements to be made without changing the connections to the test device.
The Microprocessor
A microprocessor takes the raw data and performs all the required error correction, trace
math, formatting, scaling, averaging, and marker operations, according to the instructions
from the front panel or over GPIB. The formatted data is then displayed. The data
processing sequence is described in “Processing” on page 7-6.
7-4
Operating Concepts
System Operation
7- 5
Operating Concepts
Processing
Processing
The analyzer’s receiver converts the R, A, and B input signals into useful measurement
information. This conversion occurs in two main steps:
• The swept high frequency input signals are translated to fixed low frequency IF signals,
using analog sampling or mixing techniques. (Refer to the service guide for more details
on the theory of operation.)
• The IF signals are converted into digital data by an analog to digital converter (ADC).
From this point on, all further signal processing is performed mathematically by the
analyzer microprocessors.
The following paragraphs describe the sequence of math operations and the resulting data
arrays as the information flows from the ADC to the display. They provide a good
foundation for understanding most of the response functions, and the order in which they
are performed.
Figure 7-2 is a data processing flow diagram that represents the flow of numerical data
from IF detection to display. The data passes through several math operations, denoted in
the figure by single line boxes. Most of these operations can be selected and controlled with
the front panel response block menus. The data, stored in arrays along the way and
denoted by double line boxes, are places in the flow path where data is accessible via GPIB.
7-6
Operating Concepts
Processing
While only a single flow path is shown, two identical paths are available, corresponding to
channel 1 and channel 2. Each channel also has an auxiliary channel for which the data is
processed along with the primary channel’s data. Channel 3 is the auxiliary channel for
channel 1, while channel 4 is the auxiliary channel for channel 2. When the channels are
uncoupled, each channel is processed and controlled independently.
Data point definition: A "data point" or "point" is a single piece of data representing a
measurement at a single stimulus value. Most data processing operations are
performed point-by-point; some involve more than one point.
Sweep definition: A "sweep" is a series of consecutive data point measurements,
taken over a sequence of stimulus values. A few data processing operations require that
a full sweep of data is available. The number of points per sweep can be defined by the
user. The units of the stimulus values (such as power, frequency, and time) can change,
depending on the sweep mode, although this does not generally affect the data
processing path.
Processing Details
The ADC
The ADC (analog-to-digital converter) converts the R, A, and B inputs (already
down-converted to a fixed low frequency IF) into digital words. (The AUX INPUT connector
on the rear panel is a fourth input.) The ADC switches rapidly between these inputs, so
they are converted nearly simultaneously.
IF Detection
This detection occurs in the digital filter, which performs the discrete Fourier transform
(DFT) on the digital words. The samples are converted into complex number pairs (real
plus imaginary, R+jX). The complex numbers represent both the magnitude and phase of
the IF signal. If the AUX INPUT is selected, the imaginary part of the pair is set to zero.
The DFT filter shape can be altered by changing the IF bandwidth, which is a highly
effective technique for noise reduction.
Ratio Calculations
These calculations are performed if the selected measurement is a ratio of two inputs (for
example, A/R or B/R). This is a complex divide operation. If the selected measurement is
absolute (such as A or B), no calculations are performed. The R, A, and B values are also
split into channel data at this point.
Sampler/IF Correction
The next digital processing technique used is sampler/IF correction. This process digitally
corrects for frequency response errors (both magnitude and phase, primarily sampler
rolloff) in the analog down-conversion path.
7- 7
Operating Concepts
Processing
Sweep-To-Sweep Averaging
Averaging is another noise reduction technique. This calculation involves taking the
complex exponential average of several consecutive sweeps. This technique cannot be used
with single-input measurements.
Raw Arrays
Raw arrays contain the pre-raw data which has sampler and attenuator offset applied.
7-8
Operating Concepts
Processing
Conversion
This converts the measured S-parameter data to the equivalent complex impedance (Z) or
admittance (Y) values, or to inverse S-parameters (1/S).
Format
This operation converts the complex number pairs into a scalar representation for display,
according to the selected format. This includes group delay calculations. These formats are
often easier to interpret than the complex number representation. (Polar and Smith chart
formats are not affected by the scalar formatting.) It is impossible to recover the complex
data after formatting, as shown in Figure 7-2.
Smoothing
This noise reduction technique smoothes noise on the trace. Smoothing is also used to set
the aperture for group delay measurements.
When smoothing is on, each point in a sweep is replaced by the moving average value of
several adjacent (formatted) points. The number of points included depends on the
smoothing aperture, which can be selected by the user. The effect is similar to video
filtering. If data and memory are displayed, smoothing is performed on the memory trace
only if smoothing was on when data was stored into memory.
Format Arrays
The data processing results are now stored in the format arrays. Notice that the marker
values and marker functions are all derived from the format arrays in Figure 7-2. Limit
testing is also performed on the formatted data. The format arrays are accessible via GPIB.
Display Memory
The display memory stores the display image for presentation on the analyzer. The
information stored includes graticules, annotation, and softkey labels. If user display
graphics are written, these are also stored in display memory. When a print or plot is
made, the information is taken from display memory.
The display is updated frequently and synchronously with the data processing operations.
7- 9
Operating Concepts
Output Power
Output Power
Source output power can be set over a range of −5 to +20 dBm (−5 to +18 dBm for
Option 006). The power setting can be combined with the test set step attenuator setting
for a wide output power range at the test ports. The actual test power range depends on
the test set.
NOTE After measurement calibration, you can change the power within a range (i.e.
without changing the step attenuator setting) and still maintain nearly full
accuracy. In some cases better accuracy can be achieved by changing the
power within a range. It can be useful to set different power levels for
calibration and measurement to minimize the effects of sampler compression
or noise floor.
If you decide to change the step attenuator setting, the calibration accuracy is
degraded and accuracy is no longer specified. However, the analyzer leaves
the correction on.
The annotation C∆ will be displayed whenever you change the power after
calibration.
Channel coupling
CH PWR [COUPLED] toggles between coupled and uncoupled channel power. With the
channel power coupled, the power levels are the same on each channel. With the channel
power uncoupled, you can set different power levels for each channel. For the channel
power to be uncoupled, the other channel stimulus functions must also be uncoupled
( COUPLED CH OFF ).
7-10
Operating Concepts
Sweep Time
Sweep Time
The SWEEP TIME [ ] softkey selects sweep time as the active entry and shows whether
the automatic or manual mode is active. The following explains the difference between
automatic and manual sweep time:
• Manual sweep time. As long as the selected sweep speed is within the capability of
the instrument, it will remain fixed, regardless of changes to other measurement
parameters. If you change measurement parameters such that the instrument can no
longer maintain the selected sweep time, the analyzer will change to the fastest sweep
time possible.
• Auto sweep time. Auto sweep time continuously maintains the fastest sweep speed
possible with the selected measurement parameters.
Sweep time refers only to the time that the instrument is sweeping and taking data, and
does not include the time required for internal processing of the data, retrace time, or band
switching time. A sweep speed indicator ↑ is displayed on the trace for sweep times longer
than 1.0 second. For sweep times equal to or faster than 1.0 second, the ↑ indicator appears
in the status notations area at the left of the analyzer's display.
7- 11
Operating Concepts
Sweep Time
In addition to the these parameters, the actual cycle time of the analyzer is also dependent
on the following measurement parameters:
• smoothing
• limit test
• trace math
• marker statistics
• time domain (Option 010 only)
Refer to the specifications and characteristics chapter of the reference guide to see the
minimum cycle time values for specific measurement parameters.
7-12
Operating Concepts
Source Attenuator Switch Protection
• MEASURE RESTART causes one measurement to occur before activating the test set
hold mode.
• NUMBER OF GROUPS causes a specified number of measurements to occur before
activating the test set hold mode.
7- 13
Operating Concepts
Channel Stimulus Coupling
7-14
Operating Concepts
Sweep Types
Sweep Types
The following sweep types will function with the interpolated error-correction feature
(described in “Interpolated Error Correction” on page 6-8):
• linear frequency
• power sweep
• CW time
The following sweep types will not function with the interpolated error correction feature:
• logarithmic frequency sweep
• list frequency sweep
7- 15
Operating Concepts
Sweep Types
NOTE Earlier 8753 models allowed a maximum of 1632 points, but this value was
reduced to 1601 to add the 4 channels in the 4-parameter display feature.
One list is common to both channels. Once a frequency list has been defined and a
measurement calibration performed on the full frequency list, one or all of the frequency
segments can be measured and displayed without loss of calibration.
When the LIST FREQ [STEPPED] key is pressed, the network analyzer sorts all the
defined frequency segments into CW points in order of increasing frequency. It then
measures each point and displays a single trace that is a composite of all data taken. If
duplicate frequencies exist, the analyzer makes multiple measurements on identical points
to maintain the specified number of points for each subsweep. Since the frequency points
may not be distributed evenly across the display, the display resolution may be uneven,
and more compressed in some parts of the trace than in others. However, the stimulus and
response readings of the markers are always accurate. Because the list frequency sweep is
a stepped CW sweep, the sweep time is slower than for a continuous sweep with the same
number of points.
Segment Menu
The LIST FREQ [STEPPED] softkey provides access to the segment menu, which allows
you to select any single segment ( SINGLE SEG SWEEP ) in the frequency list or all of the
segments ( ALL SEGS SWEEP ) in the frequency list.
See the following information on how to enter or modify the list frequencies. If no list has
been entered, the message CAUTION: LIST TABLE EMPTY is displayed. A tabular printout
of the frequency list data can be obtained using the LIST VALUES function in the copy
menu.
7-16
Operating Concepts
Sweep Types
The frequency subsweeps, or segments, can be defined in any of the following terms:
• start/stop/number of points
• start/stop/step
• center/span/number of points
• center/span/step
• CW frequency
The subsweeps can overlap, and do not have to be entered in any particular order. The
analyzer sorts the segments automatically and lists them on the display in order of
increasing start frequency, even if they are entered in center/span format. If duplicate
frequencies exist, the analyzer makes multiple measurements on identical points to
maintain the specified number of points for each subsweep. The data is shown on the
display as a single trace that is a composite of all data taken. The trace may appear uneven
because of the distribution of the data points, but the frequency scale is linear across the
total range.
Once the list frequencies have been defined or modified, the list frequency sweep mode can
be selected with the LIST FREQ [STEPPED] softkey in the sweep type menu. The
frequency list parameters can also be saved with an instrument state.
7- 17
Operating Concepts
Sweep Types
The frequency subsweeps, or segments, can be defined in any of the following terms:
• start/stop/number of points/power/IFBW
• start/stop/step/power/IFBW
• center/span/number of points/power/IFBW
• center/span/step/power/IFBW
See “Setting Segment Power” and “Setting Segment IF Bandwidth” on page 7-18 for
information on how to set the segment power and IF bandwidth.
The subsweeps may be entered in any particular order but they cannot overlap. The
analyzer sorts the segments automatically and lists them on the display in order of
increasing start frequency, even if they are entered in center/span format. The data is
shown on the display as a single trace that is a composite of all data taken. The trace may
appear uneven because of the distribution of the data points, but the frequency scale is
linear across the total range.
Once the list frequencies have been defined or modified, the list frequency sweep mode can
be selected with the LIST FREQ [SWEPT] softkey in the sweep type menu. The frequency
list parameters can also be saved with an instrument state.
7-18
Operating Concepts
Sweep Types
Narrow IF bandwidths require more data samples per point and thus slow down the
measurement time. Selectable IF bandwidths can increase the throughput of the
measurement by allowing you to specify narrow bandwidths only where needed.
7- 19
Operating Concepts
S-Parameters
S-Parameters
The Meas key accesses the S-parameter menu which contains softkeys that can be used
to select the parameters or inputs that define the type of measurement being performed.
Understanding S-Parameters
S-parameters (scattering parameters) are a convention used to characterize the way a
device modifies signal flow. A brief explanation of the S-parameters of a two-port device is
provided, however, for additional details, refer to Application Notes 95-1 and 154.
S-parameters are always a ratio of two complex (magnitude and phase) quantities.
S-parameter notation identifies these quantities using the numbering convention:
S out in
where the first number (out) refers to the test-device port where the signal is emerging and
the second number (in) is the test-device port where the signal is incident. For example,
the S-parameter S21 identifies the measurement as the complex ratio of the signal
emerging at the test device’s port 2 to the signal incident at the test device’s port 1.
Figure 7-3 is a representation of the S-parameters of a two-port device, together with an
equivalent flowgraph. In the illustration, "a" represents the signal entering the device and
"b" represents the signal emerging. Note that a and b are not related to the A and B input
ports on the analyzer.
7-20
Operating Concepts
S-Parameters
S-parameters are exactly equivalent to these more common description terms, requiring
only that the measurements be taken with all test device ports properly terminated.
7- 21
Operating Concepts
S-Parameters
Analog In Menu
This menu allows you to monitor voltage and frequency nodes, using the analog bus and
internal counter. For more information, refer to the "Service Menus and Error Messages"
chapter in the service guide.
Conversion Menu
This menu converts the measured reflection or transmission data to the equivalent
complex impedance (Z) or admittance (Y) values. This is not the same as a two-port Y or Z
parameter conversion, as only the measured parameter is used in the equations. Two
simple one-port conversions are available, depending on the measurement configuration.
An S11 or S22 trace measured as reflection can be converted to equivalent parallel
impedance or admittance using the model and equations shown in Figure 7-4.
7-22
Operating Concepts
S-Parameters
NOTE Avoid the use of Smith chart, SWR, and delay formats for display of Z and Y
conversions, as these formats are not easily interpreted.
7- 23
Operating Concepts
Analyzer Display Formats
Phase Format
The PHASE softkey displays a Cartesian format of the phase portion of the data,
measured in degrees. This format displays the phase shift versus frequency. The phase
response of the same filter in a phase-only format is illustrated in Figure 7-7.
7-24
Operating Concepts
Analyzer Display Formats
7- 25
Operating Concepts
Analyzer Display Formats
7-26
Operating Concepts
Analyzer Display Formats
Polar Format
The POLAR softkey displays a polar format as shown in Figure 7-10. Each point on the
polar format corresponds to a particular value of both magnitude and phase. Quantities
are read vectorally: the magnitude at any point is determined by its displacement from the
center (which has zero value), and the phase by the angle counterclockwise from the
positive x-axis. Magnitude is scaled in a linear fashion, with the value of the outer circle
usually set to a ratio value of 1. Since there is no frequency axis, frequency information is
read from the markers.
The default marker readout for the polar format is in linear magnitude and phase. A log
magnitude marker and a real/imaginary marker are available in the polar marker menu.
7- 27
Operating Concepts
Analyzer Display Formats
SWR Format
The SWR softkey reformats a reflection measurement into its equivalent SWR (standing
wave ratio) value. See Figure 7-12. SWR is equivalent to (1 + ρ)/(1 − ρ), where ρ is the
reflection coefficient. Note that the results are valid only for reflection measurements. If
the SWR format is used for measurements of S21 or S12 the results are not valid.
7-28
Operating Concepts
Analyzer Display Formats
Real Format
The REAL softkey displays only the real (resistive) portion of the measured data on a
Cartesian format. See Figure 7-13. This is similar to the linear magnitude format, but can
show both positive and negative values. It is primarily used for analyzing responses in the
time domain, and also to display an auxiliary input voltage signal for service purposes.
Imaginary Format
The IMAGINARY softkey displays only the imaginary (reactive) portion of the measured
data on a Cartesian format. This format is similar to the real format except that reactance
data is displayed on the trace instead of resistive data.
7- 29
Operating Concepts
Analyzer Display Formats
Note, however, that the phase characteristic typically consists of both linear and higher
order (deviations from linear) components. The linear component can be attributed to the
electrical length of the test device, and represents the average signal transit time. The
higher order components are interpreted as variations in transit time for different
frequencies, and represent a source of signal distortion. See Figure 7-15.
The analyzer computes group delay from the phase slope. Phase data is used to find the
phase change, ∆ Φ over a specified frequency aperture, ∆ f, to obtain an approximation for
the rate of change of phase with frequency. Refer to Figure 7-16. This value, (τ)g,
represents the group delay in seconds assuming linear phase change over ∆f. It is
important that ∆ Φ be ≤ 180°, or errors will result in the group delay data. These errors can
be significant for long delay devices. You can verify that ∆ Φ is ≤ 180° by increasing the
number of points or narrowing the frequency span (or both) until the group delay data no
longer changes.
7-30
Operating Concepts
Analyzer Display Formats
When deviations from linear phase are present, changing the frequency step can result in
different values for group delay. Note that in this case the computed slope varies as the
aperture ∆f is increased. See Figure 7-17. A wider aperture results in loss of the fine grain
variations in group delay. This loss of detail is the reason that in any comparison of group
delay data, it is important to know the aperture that was used to make the measurement.
In determining the group delay aperture, there is a trade-off between resolution of fine
detail and the effects of noise. Noise can be reduced by increasing the aperture, but this
will tend to smooth out the fine detail. More detail will become visible as the aperture is
decreased, but the noise will also increase, possibly to the point of obscuring the detail. A
good practice is to use a smaller aperture to assure that small variations are not missed,
then increase the aperture to smooth the trace.
7- 31
Operating Concepts
Analyzer Display Formats
The default group delay aperture is the frequency span divided by the number of points
across the display. To set the aperture to a different value, turn on smoothing in the
average menu, and vary the smoothing aperture. The aperture can be varied up to 20% of
the span swept.
Group delay measurements can be made on linear frequency, log frequency, or list
frequency sweep types (not in CW or power sweep). Group delay aperture varies depending
on the frequency spacing and point density, therefore the aperture is not constant in log
and list frequency sweep modes. In list frequency mode, extra frequency points can be
defined to ensure the desired aperture.
To obtain a readout of aperture values at different points on the trace, turn on a marker.
Then press Avg SMOOTHING APERTURE . Smoothing aperture becomes the active
function, and as the aperture is varied its value in Hz is displayed beneath the active entry
area.
7-32
Operating Concepts
Electrical Delay
Electrical Delay
The ELECTRICAL DELAY softkey adjusts the electrical delay to balance the phase of the
test device. This softkey must be used in conjunction with COAXIAL DELAY or
WAVEGUIDE DELAY (with cut-off frequency) in order to identify which type of
transmission line the delay is being added to. These softkeys can be accessed by pressing
the Scale Ref key.
Electrical delay simulates a variable length lossless transmission line, which can be added
to or removed from a receiver input to compensate for interconnecting cables, etc. This
function is similar to the mechanical or analog "line stretchers" of other network analyzers.
Delay is annotated in units of time with secondary labeling in distance for the current
velocity factor.
With this feature, and with MARKER → DELAY (refer to “Setting the Electrical Delay”
on page 1-37), an equivalent length of air-filled, lossless transmission line is added or
subtracted according to the following formula:
Φ
Length ( meters ) = -------------------------------------------------------------------
( Freq ( MHz ) × 1.20083 )
Once the linear portion of the test device's phase has been removed, the equivalent length
of the lossless, transmission line can be read out in the active marker area. If the average
relative permittivity (εr) of the test device is known over the frequency span, the length
calculation can be adjusted to indicate the actual length of the test device more closely.
This can be done by entering the relative velocity factor for the test device using the
calibrate more menu. The relative velocity factor for a given dielectric can be calculated by:
1
Velocity Factor = --------
εr
7- 33
Operating Concepts
Noise Reduction Techniques
Averaging
Averaging computes each data point based on an exponential average of consecutive
sweeps weighted by a user-specified averaging factor. Each new sweep is averaged into the
trace until the total number of sweeps is equal to the averaging factor, for a fully averaged
trace. Each point on the trace is the vector sum of the current trace data and the data from
the previous sweep. A high averaging factor gives the best signal-to-noise ratio, but slows
the trace update time. Doubling the averaging factor reduces the noise by 3 dB. Averaging
is used for ratioed measurements: if it is attempted for a single-input measurement (e.g. A
or B), the message CAUTION: AVERAGING INVALID ON NON-RATIO MEASURE is displayed.
The effect of averaging on a log magnitude format trace is shown in Figure 7-18.
NOTE If you switch power ranges with averaging on, the average will restart.
7-34
Operating Concepts
Noise Reduction Techniques
Smoothing
Smoothing (similar to video filtering) averages the formatted active channel data over a
portion of the displayed trace. Smoothing computes each displayed data point based on one
sweep only, using a moving average of several adjacent data points for the current sweep.
The smoothing aperture is a percent of the swept stimulus span, up to a maximum of 20%.
Rather than lowering the noise floor, smoothing finds the mid-value of the data. Use it to
reduce relatively small peak-to-peak noise values on broadband measured data. Use a
sufficiently high number of display points to avoid misleading results. Do not use
smoothing for measurements of high resonance devices or other devices with wide trace
variations, as it will introduce errors into the measurement.
Smoothing is used with Cartesian and polar display formats. It is also the primary way to
control the group delay aperture, given a fixed frequency span. Refer to “Group Delay
Principles” on page 7-29. In polar display format, large phase shifts over the smoothing
aperture will cause shifts in amplitude, since a vector average is being computed. The
effect of smoothing on a log magnitude format trace is illustrated in Figure 7-19.
IF Bandwidth Reduction
IF bandwidth reduction lowers the noise floor by digitally reducing the receiver input
bandwidth. It works in all ratio and non-ratio modes. It has an advantage over averaging
as it reliably filters out unwanted responses such as spurs, odd harmonics, higher
frequency spectral noise, and line-related noise. Sweep-to-sweep averaging, however, is
better at filtering out very low frequency noise. A tenfold reduction in IF bandwidth lowers
the measurement noise floor by about 10 dB. Bandwidths less than 300 Hz provide better
harmonic rejection than higher bandwidths.
Another difference between sweep-to-sweep averaging and variable IF bandwidth is the
sweep time. Averaging displays the first complete trace faster but takes several sweeps to
reach a fully averaged trace. IF bandwidth reduction lowers the noise floor in one sweep,
but the sweep time may be slower. The difference in noise floor between a trace measured
with a 3000 Hz IF bandwidth and with a 10 Hz IF bandwidth is illustrated by Figure 7-20.
7- 35
Operating Concepts
Noise Reduction Techniques
NOTE Hints
Another capability that can be used for effective noise reduction is the marker
statistics function, which computes the average value of part or all of the
formatted trace.
7-36
Operating Concepts
Measurement Calibration
Measurement Calibration
Measurement calibration is an accuracy enhancement procedure that effectively removes
the system errors that cause uncertainty in measuring a test device. It measures known
standard devices, and uses the results of these measurements to characterize the system.
This section discusses the following topics:
• definition of accuracy enhancement
• causes of measurement errors
• characterization of microwave systematic errors
• effectiveness of accuracy enhancement
• ensuring a valid calibration
• modifying calibration kits
• TRL*/LRM* calibration
7- 37
Operating Concepts
Measurement Calibration
Directivity
Normally a device that can separate the reverse from the forward traveling waves (a
directional bridge or coupler) is used to detect the signal reflected from the test device.
Ideally the coupler would completely separate the incident and reflected signals, and only
the reflected signal would appear at the coupled output, as shown in Figure 7-21a.
7-38
Operating Concepts
Measurement Calibration
However, an actual coupler is not perfect, as shown in Figure 7-21b. A small amount of the
incident signal appears at the coupled output due to leakage as well as reflection from the
termination in the coupled arm. Also, reflections from the coupler output connector appear
at the coupled output, adding uncertainty to the signal reflected from the device. The
figure of merit for how well a coupler separates forward and reverse waves is directivity.
The greater the directivity of the device, the better the signal separation. System
directivity is the vector sum of all leakage signals appearing at the analyzer receiver input.
The error contributed by directivity is independent of the characteristics of the test device
and it usually produces the major ambiguity in measurements of low reflection devices.
Source Match
Source match is defined as the vector sum of signals appearing at the analyzer receiver
input due to the impedance mismatch at the test device looking back into the source, as
well as to adapter and cable mismatches and losses. In a reflection measurement, the
source match error signal is caused by some of the reflected signal from the test device
being reflected from the source back toward the test device and re-reflected from the test
device. This is illustrated in Figure 7-22. In a transmission measurement, the source
match error signal is caused by reflection from the test device that is re-reflected from the
source. Source match is most often given in terms of return loss in dB: thus the larger the
number, the smaller the error.
The error contributed by source match is dependent on the relationship between the actual
input impedance of the test device and the equivalent match of the source. It is a factor in
both transmission and reflection measurements. Source match is a particular problem in
measurements where there is a large impedance mismatch at the measurement plane.
(For example, reflection devices such as filters with stop bands.)
Load Match
Load match error results from an imperfect match at the output of the test device. It is
caused by impedance mismatches between the test device output port and port 2 of the
measurement system. Some of the transmitted signal is reflected from port 2 back to the
test device as illustrated in Figure 7-23. A portion of this wave may be re-reflected to
port 2, or part may be transmitted through the device in the reverse direction to appear at
port 1. If the test device has low insertion loss (for example a filter pass band), the signal
reflected from port 2 and re-reflected from the source causes a significant error because the
test device does not attenuate the signal significantly on each reflection. Load match is
usually given in terms of return loss in dB: thus the larger the number, the smaller the
error.
7- 39
Operating Concepts
Measurement Calibration
The error contributed by load match is dependent on the relationship between the actual
output impedance of the test device and the effective match of the return port (port 2). It is
a factor in all transmission measurements and in reflection measurements of two-port
devices. The interaction between load match and source match is less significant when the
test device insertion loss is greater than about 6 dB. However, source match and load
match still interact with the input and output matches of the DUT, which contributes to
transmission measurement errors. (These errors are largest for devices with highly
reflective ports.)
Isolation (Crosstalk)
Leakage of energy between analyzer signal paths contributes to error in a transmission
measurement, much like directivity does in a reflection measurement. Isolation is the
vector sum of signals appearing at the analyzer samplers due to crosstalk between the
reference and test signal paths. This includes signal leakage within the test set and in both
the RF and IF sections of the receiver.
The error contributed by isolation depends on the characteristics of the test device.
Isolation is a factor in high-loss transmission measurements. However, analyzer system
isolation is more than sufficient for most measurements, and correction for it may be
unnecessary.
For measuring devices with high dynamic range, accuracy enhancement can provide
improvements in isolation that are limited only by the noise floor. Generally, the isolation
falls below the noise floor, therefore, when performing an isolation calibration you should
use a noise reduction function such as averaging or reduce the IF bandwidth.
7-40
Operating Concepts
Measurement Calibration
To characterize the errors, the reflection coefficient is measured by first separating the
incident signal (I) from the reflected signal (R), then taking the ratio of the two values. See
Figure 7-25. Ideally, (R) consists only of the signal reflected by the test device (S11A, for S11
actual).
However, all of the incident signal does not always reach the unknown. Refer to Figure
7-26. Some of (I) may appear at the measurement system input due to leakage through the
test set or through a signal separation device. Also, some of (I) may be reflected by
imperfect adapters between a signal separation device and the measurement plane. The
vector sum of the leakage and the miscellaneous reflections is the effective directivity, EDF.
Understandably, the measurement is distorted when the directivity signal combines
vectorally with the actual reflected signal from the unknown, S11A.
7- 41
Operating Concepts
Measurement Calibration
Since the measurement system test port is never exactly the characteristic impedance
(50 ohms), some of the reflected signal bounces off the test port, or other impedance
transitions further down the line, and back to the unknown, adding to the original incident
signal (I). This effect causes the magnitude and phase of the incident signal to vary as a
function of S11A and frequency. Leveling the source to produce a constant incident signal
(I) reduces this error, but since the source cannot be exactly leveled at the test device input,
leveling cannot eliminate all power variations. This re-reflection effect and the resultant
incident power variation are caused by the source match error, E SF as shown in Figure
7-27.
7-42
Operating Concepts
Measurement Calibration
These three errors are mathematically related to the actual data, S 11A, and measured
data, S11M, by the following equation:
S
11A E RF
S 11M = EDF + -----------------------------------------
1 – E S
SF 11A
If the value of these three "E" errors and the measured test device response were known
for each frequency, this equation could be solved for S11A to obtain the actual test device
response. Because each of these errors changes with frequency, their values must be
known at each test frequency. These values are found by measuring the system at the
measurement plane using three independent standards whose S11A is known at all
frequencies.
The first standard applied is a "perfect load," which makes S11A = 0 and essentially
measures directivity. See Figure 7-29. "Perfect load" implies a reflectionless termination at
the measurement plane. All incident energy is absorbed. With S11A = 0 the equation can be
solved for E DF, the directivity term. In practice, of course, the "perfect load" is difficult to
achieve, although very good broadband loads are available in the compatible calibration
kits.
7- 43
Operating Concepts
Measurement Calibration
Since the measured value for directivity is the vector sum of the actual directivity plus the
actual reflection coefficient of the "perfect load," any reflection from the termination
represents an error. System effective directivity becomes the actual reflection coefficient of
the near "perfect load" as shown in Figure 7-30. In general, any termination having a
return loss value greater than the uncorrected system directivity reduces reflection
measurement uncertainty.
Next, a short circuit termination whose response is known to a very high degree is used to
establish another condition as shown in Figure 7-31.
7-44
Operating Concepts
Measurement Calibration
The open circuit gives the third independent condition. In order to accurately model the
phase variation with frequency due to fringing capacitance from the open connector, a
specially designed shielded open circuit is used for this step. (The open circuit capacitance
is different with each connector type.) Now the values for EDF, directivity, ESF, source
match, and ERF, reflection frequency response, are computed and stored. See Figure 7-32.
Device Measurement
Now the unknown is measured to obtain a value for the measured response, S11M, at each
frequency. Refer to Figure 7-33.
7- 45
Operating Concepts
Measurement Calibration
This is the one-port error model equation solved for S 11A. Since the three errors and S11M
are now known for each test frequency, S11A can be computed as follows:
( S11M – E DF )
S 11A = ----------------------------------------------------------------
-
ESF ( S 11M – EDF ) + E RF
For reflection measurements on two-port devices, the same technique can be applied, but
the test device output port must be terminated in the system characteristic impedance.
This termination should have as low a reflection coefficient as the load used to determine
directivity. The additional reflection error caused by an improper termination at the test
device’s output port is not incorporated into the one-port error model.
7-46
Operating Concepts
Measurement Calibration
The transmission coefficient is measured by taking the ratio of the incident signal (I) and
the transmitted signal (T). Refer to Figure 7-35. Ideally, (I) consists only of power delivered
by the source, and (T) consists only of power emerging at the test device output.
As in the reflection model, source match can cause the incident signal to vary as a function
of test device S11A. Also, since the test setup transmission return port is never exactly the
characteristic impedance, some of the transmitted signal is reflected from the test set
port 2, and from other mismatches between the test device output and the receiver input,
to return to the test device. A portion of this signal may be re-reflected at port 2, thus
affecting S21M, or part may be transmitted through the device in the reverse direction to
appear at port 1, thus affecting S11M. This error term, which causes the magnitude and
phase of the transmitted signal to vary as a function of S22A, is called load match, E LF. See
Figure 7-36.
7- 47
Operating Concepts
Measurement Calibration
The measured value, S21M, consists of signal components that vary as a function of the
relationship between ESF and S11A as well as ELF and S22A, so the input and output
reflection coefficients of the test device must be measured and stored for use in the S21A
error-correction computation. Thus, the test setup is calibrated as described for reflection
to establish the directivity, EDF, source match, ESF, and reflection frequency response,
ERF, terms for reflection measurements on both ports.
Now that a calibrated port is available for reflection measurements, the thru is connected
and load match, E LF, is determined by measuring the reflection coefficient of the thru
connection.
Transmission signal path frequency response is then measured with the thru connected.
The data is corrected for source and load match effects, then stored as transmission
frequency response, E TF.
NOTE It is very important that the exact electrical length of the thru be known.
Most calibration kits assume a zero length thru. For some connection types
such as Type-N, this implies one male and one female port. If the test system
requires a non-zero length thru, for example, one with two male test ports,
the exact electrical delay of the thru adapter must be used to modify the
built-in calibration kit definition of the thru.
Isolation, EXF, represents the part of the incident signal that appears at the receiver
without actually passing through the test device. See Figure 7-37. Isolation is measured
with the test set in the transmission configuration and with terminations installed at the
points where the test device will be connected. Since isolation can be lower than the noise
floor, it is best to increase averaging by at least a factor of four during the isolation portion
of the calibration. The RESUME CAL SEQUENCE softkey under the Cal menu allows a
calibration sequence to resume after a change to the averaging factor.
If the leakage falls below the noise floor, it is best to increase averaging before calibration.
7-48
Operating Concepts
Measurement Calibration
In this case, omitting isolation is better than measuring the isolation standards without
increasing the averaging factor.
Thus there are two sets of error terms, forward and reverse, with each set consisting of six
error terms, as follows:
• Directivity, EDF (forward) and E DR (reverse)
• Isolation, EXF and E XR
• Source Match, ESF and E SR
• Load Match, ELF and ELR
• Transmission Tracking, ETF and ETR
• Reflection Tracking, ERF and ERR
The analyzer's test set can measure both the forward and reverse characteristics of the test
device without you having to manually remove and physically reverse the device. A full
two-port error model illustrated in Figure 7-38. This illustration depicts how the analyzer
effectively removes both the forward and reverse error terms for transmission and
reflection measurements.
7- 49
Operating Concepts
Measurement Calibration
A full two-port error model equations for all four S-parameters of a two-port device is
shown in Figure 7-39. Note that the mathematics for this comprehensive model use all
forward and reverse error terms and measured values. Thus, to perform full
error-correction for any one parameter, all four S-parameters must be measured.
Applications of these error models are provided in the calibration procedures described in
Chapter 5 , “Optimizing Measurement Results.”
7-50
Operating Concepts
Measurement Calibration
7- 51
Operating Concepts
Measurement Calibration
Figure 7-40a shows a measurement in log magnitude format with a response calibration
only. Figure 7-40b shows the improvement in the same measurement using an S11
one-port calibration. Figure 7-41a shows the measurement on a Smith chart with response
calibration only, and Figure 7-41b shows the same measurement with an S11 one-port
calibration.
Figure 7-40 Response versus S11 1-Port Calibration on Log Magnitude Format
7-52
Operating Concepts
Measurement Calibration
The response of a device in a log magnitude format is shown in Figure 7-42. Figure 7-42a
shows the response using a response calibration and Figure 7-42b the response using a full
two-port calibration.
(a) (b)
7- 53
Operating Concepts
Calibration Routines
Calibration Routines
There are twelve different error terms for a two-port measurement that can be corrected
by accuracy enhancement in the analyzer. These are directivity, source match, load match,
isolation, reflection tracking, and transmission tracking, each in both the forward and
reverse direction. The analyzer has several different measurement calibration routines to
characterize one or more of the systematic error terms and remove their effects from the
measured data.
The calibrate menu allows you to perform the measurement calibration routines. These
procedures range from a simple frequency response calibration to a full two-port
calibration that effectively removes all twelve error terms.
Response Calibration
The response calibration, activated by pressing the RESPONSE softkey within the
calibrate menu, provides a normalization of the test setup for reflection or transmission
measurements. This calibration procedure may be adequate for measurement of well
matched devices. This is the simplest error-correction to perform, and should be used when
extreme measurement accuracy is not required.
7-54
Operating Concepts
Calibration Routines
The enhanced reflection calibration effectively removes load match error from the
enhanced response calibration performed on a bilateral device. A bilateral device has an
identical forward (S21) and reverse transmission (S12) response. Most passive devices (such
as filters, attenuators, or switches) are bilateral. Some passive devices (circulators,
isolators) and most active devices do not have identical forward and reverse transmission
responses and enhanced reflection calibration will not work with these devices.
E-CAL
The E-Cal calibration menu is activated by pressing E-CAL MENU in the calibration
menu.
The E-Cal (Electronic Calibration) system determines systemic errors of the analyzer
through a one-time connection of an E-Cal module to the network analyzer ports. The
random error of connector repeatability is reduced substantially through a one-time
connection when compared to frequent connections and disconnections of the conventional
short/open/load methods.
7- 55
Operating Concepts
Modifying Calibration Kits
Definitions
The following are definitions of terms:
• A "standard" (represented by a number 1-8) is a specific, well-defined, physical device
used to determine systematic errors. For example, standard 1 is a short in the 3.5 mm
calibration kit. Standards are assigned to the instrument softkeys as part of a class.
• A standard "type" is one of five basic types that define the form or structure of the
model to be used with that standard (short, open, load, delay/thru, and arbitrary
impedance); standard 1 is of the type short in the 3.5 mm calibration kit.
• Standard "coefficients" are numerical characteristics of the standards used in the model
selected. For example, the offset delay of the short is 32 ps in the 3.5 mm calibration kit.
• A standard "class" is a grouping of one or more standards that determines which of the
eight standards are used at each step of the calibration. For example, standard number
2 and 8 usually makes up the S11A reflection class, which for type-N calibration kits are
male and female shorts.
7-56
Operating Concepts
Modifying Calibration Kits
Procedure
The following steps are used to modify or define a user kit:
1. Select the predefined kit to be modified. (This is not necessary for defining a new
calibration kit.)
2. Define the standards:
• Define which "type" of standard it is.
• Define the electrical characteristics (coefficients) of the standard.
3. Specify the class where the standard is to be assigned.
4. Store the modified calibration kit.
• DEFINE STANDARD makes the standard number the active function, and brings up
the define standard menus. Before selecting a standard, a standard number must be
entered. This number (1 to 8) is an arbitrary reference number used to reference
standards while specifying a class. The standard numbers for the predefined calibration
kits are as follows:
1 short (m)
2 open (m)
3 broadband load
4 thru
5 sliding load
6 lowband load
7 short (f)
8 open (f)
NOTE Although the numbering sequences are arbitrary, confusion can be minimized
by using consistency. However, standard 5 is always a sliding load.
• SPECIFY CLASS leads to the specify class menu. After the standards are modified,
use this key to specify a class to group certain standards.
• LABEL CLASS leads to the label class menu, to give the class a meaningful label for
future reference.
7- 57
Operating Concepts
Modifying Calibration Kits
• LABEL KIT leads to a menu for constructing a label for the user-modified cal kit. If a
label is supplied, it will appear as one of the five softkey choices in the select cal kit
menu. The approach is similar to defining a display title, except that the kit label is
limited to ten characters.
• TRL/LRM OPTION brings up the TRL Option menu.
• KIT DONE (MODIFIED) terminates the calibration kit modification process, after all
standards are defined and all classes are specified. Be sure to save the kit with the
SAVE USER KIT softkey, if it is to be used later.
STANDARD b C0c C1c C2c C3c FIXED d TERMe OFFSET FREQ (GHz) COAX STND
or LABEL
×10−15 ×10−27 ×10−36 ×10−45 SLIDING IMPED
F F/Hz or Ω WG
NO. TYPE F/Hz2 F/Hz3 DELAY Z0 LOSS MIN MAX
OFFSET
s Ω Ω/s
Each standard must be identified as one of five "types": open, short, load, delay/thru, or
arbitrary impedance.
7-58
Operating Concepts
Modifying Calibration Kits
After a standard number is entered, selection of the standard type will present one of five
menus for entering the electrical characteristics (model coefficients) corresponding to that
standard type, such as OPEN . These menus are tailored to the current type, so that only
characteristics applicable to the standard type can be modified.
The following is a description of the softkeys located within the define standard menu:
• OPEN defines the standard type as an open, used for calibrating reflection
measurements. Opens are assigned a terminal impedance of infinite ohms, but delay
and loss offsets may still be added. Pressing this key also brings up a menu for defining
the open, including its capacitance.
As a reflection standard, an open termination offers the advantage of broadband
frequency coverage. At RF and microwave frequencies, however, an open rarely has
perfect reflection characteristics because fringing capacitance effects cause phase shift
that varies with frequency. This can be observed in measuring an open termination
after calibration, when an arc in the lower right circumference of the Smith chart
indicates capacitive reactance. These effects are impossible to eliminate, but the
calibration kit models include the open termination capacitance at all frequencies for
compatible calibration kits. The capacitance model is a cubic polynomial, as a function
of frequency, where the polynomial coefficients are user-definable. The capacitance
model equation is:
— C0 allows you to enter the C0 term, which is the constant term of the cubic
polynomial and is scaled by 10−15 Farads.
— C1 allows you to enter the C1 term, expressed in F/Hz (Farads/Hz) and scaled by
10−27.
— C2 allows you to enter the C2 term, expressed in F/Hz2 and scaled by 10−36.
— C3 allows you to enter the C3 term, expressed in F/Hz3 and scaled by 10−45.
• SHORT defines the standard type as a short, for calibrating reflection measurements.
Shorts are assigned a terminal impedance of 0 ohms, but delay and loss offsets may still
be added.
• LOAD defines the standard type as a load (termination). Loads are assigned a
terminal impedance equal to the system characteristic impedance Z0, but delay and loss
offsets may still be added. If the load impedance is not Z0, use the arbitrary impedance
standard definition.
7- 59
Operating Concepts
Modifying Calibration Kits
• DELAY/THRU defines the standard type as a transmission line of specified length, for
calibrating transmission measurements.
• ARBITRARY IMPEDANCE defines the standard type to be a load, but with an
arbitrary impedance (different from system Z0).
7-60
Operating Concepts
Modifying Calibration Kits
• OFFSET DELAY allows you to specify the one-way electrical delay from the
measurement (reference) plane to the standard, in seconds (s). (In a transmission
standard, offset delay is the delay from plane to plane.) Delay can be calculated from the
precise physical length of the offset, the permittivity constant of the medium, and the
speed of light.
In coax, group delay is considered constant. In waveguide, however, group delay is
dispersive, that is, it changes significantly as a function of frequency. Hence, for a
waveguide standard, offset delay must be defined as though it were a TEM wave
(without dispersion).
• OFFSET LOSS allows you to specify energy loss, due to skin effect, along a one-way
length of coax offset. The value of loss is entered as ohms/nanosecond (or Giga
ohms/second) at 1 GHz. (Such losses are negligible in waveguide, so enter 0 as the loss
offset.)
• OFFSET Z0 allows you to specify the characteristic impedance of the coax offset.
(Note: This is not the impedance of the standard itself.) For waveguide, the offset
impedance as well as the system Z0 must always be set to 1Ω.
• MINIMUM FREQUENCY allows you to define the lowest frequency at which the
standard can be used during measurement calibration. In waveguide, this must be the
lower cutoff frequency of the standard, so that the analyzer can calculate dispersive
effects correctly (see OFFSET DELAY ).
• MAXIMUM FREQUENCY allows you to define the highest frequency at which the
standard can be used during measurement calibration. In waveguide, this is normally
the upper cutoff frequency of the standard.
• COAX defines the standard (and the offset) as coaxial. This causes the analyzer to
assume linear phase response in any offsets.
• WAVEGUIDE defines the standard (and the offset) as rectangular waveguide. This
causes the analyzer to assume a dispersive delay (see OFFSET DELAY ).
7- 61
Operating Concepts
Modifying Calibration Kits
S11A
S11B
S11C
S22A
S22B
S22C
Forward Transmission
Reverse Transmission
Forward Match
Reverse Match
Response
TRL thru
TRL reflect
The number of standard classes required depends on the type of calibration being
performed, and is identical to the number of error terms corrected. A response calibration
requires only one class, and the standards for that class may include an open, or short, or
thru. A 1-port calibration requires three classes. A full 2-port calibration requires 10
classes, not including two for isolation.
7-62
Operating Concepts
Modifying Calibration Kits
The number of standards that can be assigned to a given class may vary from none (class
not used) to one (simplest class) to seven. When a certain class of standards is required
during calibration, the analyzer will display the labels for all the standards in that class
(except when the class consists of a single standard). This does not, however, mean that all
standards in a class must be measured during calibration. Unless band-limited standards
are used, only a single standard per class is required.
NOTE It is often simpler to keep the number of standards per class to the bare
minimum needed (often one) to avoid confusion during calibration.
Each class can be given a user-definable label as described under label class menus.
Standards are assigned to a class simply by entering the standard’s reference number
(established while defining a standard) under a particular class. The following is a
description of the softkeys located within the specify class menu:
• S11A allows you to enter the standard numbers for the first class required for an S11
1-port calibration. (For default calibration kits, this is the open.)
• S11B allows you to enter the standard numbers for the second class required for an
S11 1-port calibration. (For default calibration kits, this is the short.)
• S11C allows you to enter the standard numbers for the third class required for an S11
1-port calibration. (For default calibration kits, this is the load.)
• S22A allows you to enter the standard numbers for the first class required for an S22
1-port calibration. (For default calibration kits, this is the open.)
• S22B allows you to enter the standard numbers for the second class required for an
S22 1-port calibration. (For default calibration kits, this is the short.)
• S22C allows you to enter the standard numbers for the third class required for an S22
1-port calibration. (For default calibration kits, this is the load.)
• FWD TRANS allows you to enter the standard numbers for the forward transmission
thru calibration. (For default calibration kits, this is the thru.)
• REV TRANS allows you to enter the standard numbers for the reverse transmission
(thru) calibration. (For default calibration kits, this is the thru.)
• FWD MATCH allows you to enter the standard numbers for the forward match (thru)
calibration. (For default calibration kits, this is the thru.)
• REV MATCH allows you to enter the standard numbers for the reverse match (thru)
calibration. (For default calibration kits, this is the thru.)
• RESPONSE allows you to enter the standard numbers for a response calibration. This
calibration corrects for frequency response in either reflection or transmission
measurements, depending on the parameter being measured when a calibration is
performed. (For default kits, the standard is either the open or short for reflection
measurements, or the thru for transmission measurements.)
7- 63
Operating Concepts
Modifying Calibration Kits
• RESPONSE &ISOL’N allows you to enter the standard numbers for a response &
isolation calibration. This calibration corrects for frequency response and directivity in
reflection measurements, or frequency response and isolation in transmission
measurements.
• TRL THRU allows you to enter the standard numbers for a TRL thru calibration.
• TRL REFLECT allows you to enter the standard numbers for a TRL reflect
calibration.
• TRL LINE OR MATCH allows you to enter the standard numbers for a TRL line or
match calibration.
Verify Performance
Once a measurement calibration has been generated with a user-defined calibration kit, its
performance should be checked before making device measurements. To check the
accuracy that can be obtained using the new calibration kit, a device with a well-defined
frequency response (preferably unlike any of the standards used) should be measured. The
verification device must not be one of the calibration standards: measurement of one of
these standards is merely a measure of repeatability.
To achieve more complete verification of a particular measurement calibration, accurately
known verification standards with a diverse magnitude and phase response should be
used. National standard traceable or Agilent standards are recommended to achieve
verifiable measurement accuracy.
NOTE The published specifications for this network analyzer system include
accuracy enhancement with compatible calibration kits. Measurement
calibrations made with user-defined or modified calibration kits are not
subject to the analyzer specifications, although a procedure similar to the
system verification procedure may be used.
7-64
Operating Concepts
Modifying Calibration Kits
Cal CAL KIT SELECT CAL KIT MODIFY KIT DONE (MODIFIED)
To save the modified calibration kit, press:
Cal CAL KIT SELECT CAL KIT USER KIT SAVE USER KIT or
Save/Recall SAVE STATE .
Ensure that USER KIT is underlined before saving the modified user kit.
7- 65
Operating Concepts
TRL*/LRM* Calibration
TRL*/LRM* Calibration
The network analyzer has the capability of making calibrations using the "TRL"
(thru-reflect-line) method. This section contains information on the following subjects:
• Why Use TRL Calibration?
• TRL Terminology
• How TRL*/LRM* Calibration Works
• Improving Raw Source Match and Load Match for TRL*/LRM* Calibration
• The TRL Calibration Procedure
— Requirements for TRL Standards
— TRL Options
7-66
Operating Concepts
TRL*/LRM* Calibration
TRL Terminology
Notice that the letters TRL, LRL, LRM, etc. are often interchanged, depending on the
standards used. For example, "LRL" indicates that two lines and a reflect standard are
used; "TRM" indicates that a thru, reflection and match standards are used. All of these
refer to the same basic method.
TRL* calibration is a modified form of TRL calibration. It is adapted for a receiver with
three samplers instead of four samplers. The TRL* calibration is not as accurate as the
TRL calibration because it cannot isolate the source match from the load match, so it
assumes load match and source match are equal.
7- 67
Operating Concepts
TRL*/LRM* Calibration
For the analyzer TRL* 2-port calibration, a total of 10 measurements are made to quantify
eight unknowns (not including the two isolation error terms). Assume the two
transmission leakage terms, EXF and EXR, are measured using the conventional technique.
The eight TRL error terms are represented by the error adapters shown in Figure 7-43.
Although this error model is slightly different from the traditional Full 2-port 12-term
model, the conventional error terms may be derived from it. For example, the forward
reflection tracking (E RF) is represented by the product of ε10 and ε01. Also notice that the
forward source match (ESF) and reverse load match (E LR) are both represented by ε11,
while the reverse source match (ESR) and forward load match (ELF) are both represented
by ε22. In order to solve for these eight unknown TRL error terms, eight linearly
independent equations are required.
The first step in the TRL* 2-port calibration process is the same as the transmission step
for a Full 2-port calibration. For the thru step, the test ports are connected together
directly (zero length thru) or with a short length of transmission line (non- zero length
thru) and the transmission frequency response and port match are measured in both
directions by measuring all four S-parameters.
For the reflect step, identical high reflection coefficient standards (typically open or short
circuits) are connected to each test port and measured (S11 and S22).
For the line step, a short length of transmission line (different in length from the thru) is
inserted between port 1 and port 2 and again the frequency response and port match are
measured in both directions by measuring all four S-parameters.
In total, ten measurements are made, resulting in ten independent equations. However,
the TRL error model has only eight error terms to solve for. The characteristic impedance
of the line standard becomes the measurement reference and, therefore, has to be assumed
ideal (or known and defined precisely).
At this point, the forward and reverse directivity (EDF and EDR), transmission tracking
(ETF and ETR), and reflection tracking (ERF and ERR) terms may be derived from the TRL
error terms. This leaves the isolation (EXF and EXR), source match (ESF and ESR) and load
match (ELF and ELR) terms to discuss.
Isolation
Two additional measurements are required to solve for the isolation terms (EXF and EXR).
Isolation is characterized in the same manner as the Full 2-port calibration. Forward and
reverse isolation are measured as the leakage (or crosstalk) from port 1 to port 2 with each
port terminated. The isolation part of the calibration is generally only necessary when
measuring high loss devices (greater than 70 dB).
7-68
Operating Concepts
TRL*/LRM* Calibration
Figure 7-44 8-term TRL (or TRL*) Error Model and Generalized Coefficients
7- 69
Operating Concepts
TRL*/LRM* Calibration
If the device measurement requires bias, it will be necessary to add external bias tees
between the fixed attenuators and the fixture. The internal bias tees of the analyzer will
not pass the bias properly through the external fixed attenuators. Be sure to calibrate with
the external bias tees in place (no bias applied during calibration) to remove their effect
from the measurement.
Because the bias tees must be placed after the attenuators, they essentially become part of
the fixture. Their mismatch effects are the same for source match and load match, so the
TRL CAL routine will correct for their effects. Although the fixed attenuators improve the
raw mismatch of the network analyzer system, they also degrade the overall measurement
dynamic range.
This effective mismatch of the system after calibration has the biggest effect on reflection
measurements of highly reflective devices. Likewise, for well matched devices, the effects
of mismatch are negligible. This can be shown by the following approximation:
7-70
Operating Concepts
TRL*/LRM* Calibration
Types Requirements
THRU
(Zero length) • No loss. Characteristic impedance (Z0 ) need not be known.
• S21= S12= 1 ∠0°
• S11= S22 = 0
THRU
(Non-zero
length) • Z0 of the thru must be the same as the line. (If they are not the same,
the average impedance is used.)
• Attenuation of the thru need not be known.
• If the thru is used to set the reference plane, the insertion phase or
electrical length must be well-known and specified. If a non-zero length
thru is specified to have zero delay, the reference plane is established in
the middle of the thru, resulting in phase errors during measurement of
devices.
REFLECT • Reflection coefficient Γ magnitude is optimally 1.0, but need not be
known.
• Phase of Γ must known and specified to within ± 1/4 wavelength or
± 90°. During computation of the error model, the root choice in the
solution of a quadratic equation is based on the reflection data. An error
in definition would show up as a 180° error in the measured phase.
7- 71
Operating Concepts
TRL*/LRM* Calibration
7-72
Operating Concepts
TRL*/LRM* Calibration
The line standard must meet specific frequency related criteria, in conjunction with the
length used by the thru standard. In particular, the insertion phase of the line must not be
the same as the thru. The optimal line length is 1/4 wavelength (90 degrees) relative to a
zero length thru at the center frequency of interest, and between 20 and 160 degrees of
phase difference over the frequency range of interest. (Note: these phase values can be
±N × 180 degrees where N is an integer.) If two lines are used (LRL), the difference in
electrical length of the two lines should meet these optimal conditions. Measurement
uncertainty will increase significantly when the insertion phase nears zero or is an integer
multiple of 180 degrees, and this condition is not recommended.
For a transmission media that exhibits linear phase over the frequency range of interest,
the following expression can be used to determine a suitable line length of one-quarter
wavelength at the center frequency (which equals the sum of the start frequency and stop
frequency divided by 2):
( 15000 × V F )
Electrical length ( cm ) = --------------------------------------------------
f1 ( MHz ) + f2 ( MHz )
-
let:
f1 = 1000 MHz
f2 = 2000 MHz
VF = Velocity Factor = 1 (for this example)
Thus, the length to initially check is 5 cm.
Next, use the following to verify the insertion phase at f1 and f2:
( 360 × f × l -)
Phase ( degrees ) = ----------------------------
v
where:
f = frequency
l = length of line
v = velocity = speed of light × velocity factor
which can be reduced to the following using frequencies in MHz and length in centimeters:
0.012 × f ( MHz ) × l ( cm )
Phase ( degrees ) approx = ----------------------------------------------------------
VF
-
7- 73
Operating Concepts
TRL*/LRM* Calibration
So for an air line (velocity factor approximately 1) at 1000 MHz, the insertion phase is
60 degrees for a 5 cm line; it is 120 degrees at 2000 MHz. This line would be a suitable line
standard.
For microstrip and other fabricated standards, the velocity factor is significant. In those
cases, the phase calculation must be divided by that factor. For example, if the dielectric
constant for a substrate is 10, and the corresponding "effective" dielectric constant for
microstrip is 6.5, then the "effective" velocity factor equals 0.39 (1 ÷ square root of 6.5).
Using the first equation with a velocity factor of 0.39, the initial length to test would be
1.95 cm. This length provides an insertion phase at 1000 MHz of 60 degrees; at 2000 MHz,
120 degrees (the insertion phase should be the same as the air line because the velocity
factor was accounted for when using the first equation).
Another reason for showing this example is to point out the potential problem in
calibrating at low frequencies using TRL. For example, one-quarter wavelength is
7500 × VF-
Length ( cm ) = -------------------------
fc
where:
• fc = center frequency
Thus, at 50 MHz,
7500
Length ( cm ) = ----------------------
50 ( MHz )
- = 150 cm or 1.5 m
Such a line standard would not only be difficult to fabricate, but its long term stability and
usability would be questionable as well.
Thus at lower frequencies or very broad band measurements, fabrication of a "match" or
termination may be deemed more practical. Since a termination is, in essence, an infinitely
long transmission line, it fits the TRL model mathematically, and is sometimes referred to
as a "TRM" calibration.
The TRM calibration technique is related to TRL with the difference being that it bases the
characteristic impedance of the measurement on a matched Z O termination instead of a
transmission line for the third measurement standard. Like the TRL thru standard, the
TRM THRU standard can either be of zero length or non-zero length. The same rules for
thru and reflect standards used for TRL apply for TRM.
TRM has no inherent frequency coverage limitations which makes it more convenient in
some measurement situations. Additionally, because TRL requires a different physical
length for the thru and the line standards, its use becomes impractical for fixtures with
contacts that are at a fixed physical distance from each other.
7-74
Operating Concepts
TRL*/LRM* Calibration
For information on how to modify calibration constants for TRL*/LRM*, and how to
perform a TRL or TRM calibration, refer to “Calibrating for Non-Coaxial Devices” on
page 6-52.
TRL Options
The TRL/LRM OPTION softkey accesses the TRL/LRM options menu. There are two
selections under this menu:
7- 75
Operating Concepts
TRL*/LRM* Calibration
7-76
Operating Concepts
GPIB Operation
GPIB Operation
This section contains information on the following topics:
• local key
• GPIB controller modes
• instrument addresses
• using the parallel port
Local Key
This key is allows you to return the analyzer to local (front panel) operation from remote
(computer controlled) operation. This key will also abort a test sequence or hardcopy
print/plot. In this local mode, with a controller still connected on GPIB, you can operate the
analyzer manually (locally) from the front panel. This is the only front panel key that is
not disabled when the analyzer is remotely controlled over GPIB by a computer. The
exception to this is when local lockout is in effect: this is a remote command that disables
the Local key, making it difficult to interfere with the analyzer while it is under
computer control.
In addition, the Local key accesses the GPIB menu, where you can set the controller
mode, and to the address menu, where you can enter the GPIB addresses of peripheral
devices and select plotter/printer ports. You can also set the mode of the parallel port here.
The GPIB menu consists of the following softkeys:
• SYSTEM CONTROLLER
• TALKER/LISTENER
• USE PASS CONTROL
• SET ADDRESS
• PARALLEL [ ]
• GPIB DIAG on OFF
• DISK UNIT NUMBER
• VOLUME NUMBER
The analyzer is factory-equipped with a remote programming interface using the General
Purpose Interface Bus (GPIB). This enables communication between the analyzer and a
controlling computer as well as other peripheral devices. This menu indicates the present
GPIB controller mode of the analyzer. Three GPIB modes are possible: system controller,
talker/listener, and pass control.
7- 77
Operating Concepts
GPIB Operation
Talker/Listener Mode
The TALKER/LISTENER softkey activates the talker/listener mode, which is the mode of
operation most often used. In this mode, a computer controller communicates with the
analyzer and other compatible peripherals over the bus. The computer sends commands or
instructions to and receives data from the analyzer. All of the capabilities available from
the analyzer front panel can be used in this remote operation mode, except for control of
the power line switch and some internal tests.
7-78
Operating Concepts
GPIB Operation
Address Menu
This menu can be accessed by pressing the SET ADDRESS softkey within the GPIB
menu.
In communications through the General Purpose Interface Bus (GPIB), each instrument
on the bus is identified by a GPIB address. This decimal-based address code must be
different for each instrument on the bus.
This menu lets you set the GPIB address of the analyzer, and enter the addresses of
peripheral devices so that the analyzer can communicate with them.
Most of the GPIB addresses are set at the factory and need not be modified for normal
system operation. The standard factory-set addresses for instruments that may be part of
the system are as follows:
Analyzer 16
Plotter 05
Printer 01
Controller 21
Power Meter 13
The address displayed in this menu for each peripheral device must match the address set
on the device itself. The analyzer does not have a GPIB switch: its address is set only from
the front panel.
These addresses are stored in non-volatile memory and are not affected by preset or by
cycling the power.
7- 79
Operating Concepts
GPIB Operation
7-80
Operating Concepts
Limit Line Operation
NOTE The limit test output has three selectable modes. For more information, refer
to “Using Limit Lines to Test a Device” on page 1-71.
Limit lines and limit testing can be used simultaneously or independently. If limit lines are
on and limit testing is off, the limit lines are shown on the display for visual comparison
and adjustment of the measurement trace. However, no pass/fail information is provided.
If limit testing is on and limit lines are off, the specified limits are still valid and the
pass/fail status is indicated even though the limit lines are not shown on the display.
Limits are entered in tabular form. Limit lines and limit testing can be either on or off
while limits are defined. As new limits are entered, the tabular columns on the display are
updated, and the limit lines (if on) are modified to the new definitions. The complete limit
set can be offset in either stimulus or amplitude value.
Limits are checked only at the actual measured data points. It is possible for a device to be
out of specification without a limit test failure indication if the point density is insufficient.
Be sure to specify a high enough number of measurement points in the stimulus menu.
Limit lines are displayed only on Cartesian formats. In polar and Smith chart formats,
limit testing of one value is available: the value tested depends on the marker mode and is
the magnitude or the first value in a complex pair. The message NO LIMIT LINES
DISPLAYED is shown on the display in polar and Smith chart formats.
The list values feature in the copy menu provides tabular listings to the display or a
printer for every measured stimulus value. These include limit line or limit test
information if these functions are activated. If limit testing is on, an asterisk is listed next
to any measured value that is out of limits. If limit lines are on, and other listed data
allows sufficient space, the upper limit and lower limit are listed, together with the margin
by which the device data passes or fails the nearest limit.
7- 81
Operating Concepts
Limit Line Operation
If limit lines are on, they are plotted with the data on a plot. If limit testing is on, the PASS
or FAIL message is plotted, and the failing portions of the trace that are a different color on
the display are also a different color on the plot. If limits are specified, they are saved in
memory with an instrument state.
7-82
Operating Concepts
Knowing the Instrument Modes
Primary Applications
External source mode is useful in several applications:
• when your test device is a mixer or other frequency translation device
• in automated test applications where a source is already connected to the system, and
you do not want to switch between the system source and the analyzer's internal source.
7- 83
Operating Concepts
Knowing the Instrument Modes
7-84
Operating Concepts
Knowing the Instrument Modes
• The incoming signal should not have large spurs or sidebands, as the analyzer may
phase lock on a spur or not phase lock at all.
• The frequency of the incoming signal should be within −0.5 to +5.0 MHz of the selected
frequency or the analyzer will not be able to phase lock to it.
CW Frequency Range in External Source Mode 300 kHz to 3 GHz (6 GHz for Option
006)
Compatible Sweep Types The external source mode will only function in CW time
sweep. If the instrument is in any other sweep type when external source is activated, the
warning message CHANGED TO CW TIME MODE will appear on the display.
External Source Requirements The external source mode has spectral purity and
power input requirements, which are described in the specifications and characteristics
chapter of the reference guide.
Capture Range In either automatic or manual mode, you can enter the frequency of the
external CW signal using the CW FREQ softkey (located under the Stimulus Menu key).
The actual signal must be within a certain frequency capture range as shown in Table 7-3.
Table 7-3 8753ES Option 011 External Source Capture Ranges
If the incoming signal is not within the capture range, the analyzer will not phase lock
correctly.
7- 85
Operating Concepts
Knowing the Instrument Modes
Using the analyzer’s tuned receiver mode is useful for automated test applications where
an external synthesized source is available and applications where speed is important.
Although the tuned receiver mode can function in all sweep types, it is typically used in
CW applications.
7-86
Operating Concepts
Knowing the Instrument Modes
Single-Channel Operation
You can view the second or third harmonic alone by using only one of the analyzer's two
channels.
Dual-Channel Operation
To make the following types of measurements, uncouple channels 1 and 2, and switch on
dual channel.
• The analyzer measures the fundamental on one channel while measuring the second or
third harmonic on the other channel.
• The analyzer measures the second harmonic on one channel while measuring the third
harmonic on the other channel.
• Using the COUPLE PWR ON off feature, the analyzer measures the fundamental on
channel 1 while measuring the second or third harmonic in dBc on channel 2.
• Using the COUPLE PWR ON off feature, the analyzer couples power between
channels 1 and 2. This is useful when you are using the D2/D1 to D2 feature because
you can change fundamental power and see the resultant change in the harmonic
power.
7- 87
Operating Concepts
Knowing the Instrument Modes
The analyzer shows the fundamental frequency value on the display. However, a marker in
the active entry area shows the harmonic frequency in addition to the fundamental. If you
use the harmonic mode, the annotation H=2 or H=3 appears on the left-hand side of the
display. The measured harmonic cannot not exceed the frequency limitations of the
network analyzer’s receiver.
Frequency Range
The frequency range is determined by the upper frequency range of the instrument or
system (3 or 6 GHz) and by the harmonic being displayed. The 6 GHz operation requires an
8753ET/ES Option 002 and Option 011 with Option 006. Table 7-4 shows the highest
fundamental frequency for maximum frequency and harmonic mode.
Table 7-4 Maximum Fundamental Frequency using the Harmonic Mode
7-88
Operating Concepts
Differences between 8753 Network Analyzers
a. For this network analyzer, the feature is dependent on the test set being used.
b. 300 kHz to 3 GHz without Option 006; 30 kHz to 6 GHz, with Option 006.
c. 90 dB from 30 kHz to 50 kHz; 100 dB from 300 kHz to 16 MHz.
7- 89
Operating Concepts
Differences between 8753 Network Analyzers
7-90
Operating Concepts
Differences between 8753 Network Analyzers
7- 91
Operating Concepts
Differences between 8753 Network Analyzers
Table 7-7 Comparing the 8753D/E/ES Option 011 Network Analyzers (Continued)
a. For this network analyzer, the feature is dependent on the test set being used.
b. 300 kHz to 3 GHz without Option 006; 30 kHz to 6 GHz, with Option 006.
7-92
8 Safety and Regulatory Information
8-1
Safety and Regulatory Information
General Information
General Information
Maintenance
Clean the cabinet, using a dry or damp cloth only.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
8-2
Safety and Regulatory Information
General Information
8- 3
Safety and Regulatory Information
Safety Symbols
Safety Symbols
The following safety symbols are used throughout this manual. Familiarize yourself with
each of the symbols and its meaning before operating this instrument.
Instrument Markings
The instruction documentation symbol. The product is marked with this
symbol when it is necessary for the user to refer to the instructions in the
documentation.
8-4
Safety and Regulatory Information
Safety Considerations
Safety Considerations
NOTE This instrument has been designed and tested in accordance with IEC
Publication 1010, Safety Requirements for Electronics Measuring Apparatus,
and has been supplied in a safe condition. This instruction documentation
contains information and warnings which must be followed by the user to
ensure safe operation and to maintain the instrument in a safe condition.
CAUTION Always use the three-prong AC power cord supplied with this product. Failure
to ensure adequate earth grounding by not using this cord may cause product
damage.
CAUTION Install the instrument so that the ON/OFF switch is readily identifiable and
is easily reached by the operator. The ON/OFF switch or the detachable power
cord is the instrument disconnecting device. It disconnects the mains circuits
from the mains supply before other parts of the instrument. Alternately, an
externally installed switch or circuit breaker (which is readily identifiable
and is easily reached by the operator) may be used as a disconnecting device.
CAUTION Before switching on this instrument, make sure that the analyzer line voltage
selector switch is set to the voltage of the power supply and the correct fuse is
installed. Assure the supply voltage is in the specified range.
8- 5
Safety and Regulatory Information
Safety Considerations
Servicing
WARNING These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you
are qualified to do so.
WARNING The power cord is connected to internal capacitors that may remain
live for 10 seconds after disconnecting the plug from its power
supply.
WARNING For continued protection against fire hazard replace line fuse only
with same type and rating (115V operation: T 5A 125 V UL/CSA;
230V operation: T 4A H 250V IEC). The use of other fuses or material
is prohibited.
8-6
Safety and Regulatory Information
Safety Considerations
General
CAUTION This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 1010 and 664 respectively.
8- 7
Safety and Regulatory Information
Safety Considerations
LpA<70 dB Lpa<70 dB
8-8
Safety and Regulatory Information
Declaration of Conformity
Declaration of Conformity
8- 9
Safety and Regulatory Information
Declaration of Conformity
8-10
Index
Index-1
Index
calibration, receiver, 6-15 comma-separated values. See decision making functions, 1-111
calibration, TRL*/LRM*, 7-66 CSV format decoupled
calling the next measurement compensating for directional channel power, 1-13
sequence, 2-27 coupler response, 6-35 stimulus, 1-13
capabilities compliance with German noise decreasing
mixer measurement, 2-3 requirements, 8-8 frequency span, 5-10
capacitance, fringe, 6-6 computer, what you can save, 4-35 sweep rate, 5-8
cause of measurement problems, confidence check, 6-67 time delay, 5-8
5-7 configuring decrementing the loop counter,
center frequency, setting , 1-35 plot function, 4-9 2-29
changing configuring two sources , 2-27 default colors, 1-22
sequence title, 1-102 connecting defining
system bandwidth, 5-15 device under test, 1-4 plot function, 4-13
changing the ripple limits color, required test equipment, 1-4 print function, 4-6
1-87 connector defining a print function
channel coupling, 7-10 care, 5-3 color printer, using, 4-6
channel position softkey, 1-17 repeatability, 5-4 resetting print parameters to
channel stimulus coupling, 7-14 continuous correction mode, default value, 4-7
characteristics of the filter, 1-66 using, 6-38 delay block, electrical, 7-8
characterizing a duplexer, 1-49 continuous markers, 1-24 delay, determining electrical, 6-75
definitions, 1-49 conversion, 7-9 delay, electrical, 7-33
procedure, 1-49 conversion compression using delay, group, 1-46, 2-32
characterizing microwave frequency offset mode, 2-37 deleting
systematic errors, 7-41 conversion loss, 2-18 commands, 1-99
device measurement, 7-45 conversion loss using frequency frequency signals, 6-35
one-port error model, 7-41 offset mode, 2-11 line segments, 1-77
two-port error model, 7-46 conversion menu, 7-22 deleting a file, 4-51
choosing copy mode, 7-79 all files, 4-51
display elements, 4-13 correction, sampler/IF, 7-7 instrument state file, 4-51
measurement parameters, 1-4 coupling display markers, 1-31 delta markers, 1-28
plot speed, 4-16 coupling power between channels demodulating the results of the
scale, 4-15 1 and 2, 1-58, 7-88 forward transform , 3-23
chop sweep mode, activating, 5-12 coupling, channel stimulus, 7-14 designing your own fixture, 6-51
CITIfile, 4-40 creating detecting IF delay, 5-10
clarifying type-N connector sex, flat limit lines, 1-72 detection, IF, 7-7
6-4 sequence, 1-97 deviation from linear phase, 1-46
clearing a sequence from memory, sloping limit line, 1-74 device measurement, 7-45
1-101 user-defined TRL calibration device measurements, 6-4
color kit, 6-52 device under test
changing the ripple limits, 1-87 user-defined TRM calibration measuring , 1-5
ripple test message, 1-86 kit, 6-56 device under test, connecting , 1-4
color printer, using, 4-6 creating single point limits, 1-76 device, bilateral, 6-22, 6-25
colors of the display, adjusting, crosstalk , 7-40 device, noninsertable, 6-71
1-22 CSV format, saving measurement directional coupler response,
default, 1-22 results in, 4-43 compensating for, 6-35
intensity , 1-22 CW frequency range in external discrete markers, 1-24
modified, 1-23 source mode, 7-85 disk
modify colors menu, 1-22 CW frequency, setting, 1-38 formatting, 4-53
command CW time measurements, 3-22 disk, plotting a measurement to,
deleting, 1-99 CW time sweep, 7-19 4-11
inserting, 1-100 display elements, choosing , 4-13
modifying, 1-100 D display functions, 1-10
commands that require clean data formats, ASCII, 4-40 active channel display, 1-11
sweep, 1-105 data processing, 7-6 titling, 1-11
commands that sequencing processing details, 7-7 adjusting colors of the display,
completes before next data trace, 1-19 1-22
command, 1-104 saving to display memory, 1-19 blanking the display, 1-21
Index-2
Index
Index-3
Index
prompting user to connect mixer frequency offset mode, conversion moving marker information off
test setup, 2-26 loss, 2-11 the grids, 1-26
response calibration, 2-26 frequency offset operation, 7-87 group delay, 1-46
sequence 1 setup, 2-24 frequency range, 1-58, 7-88 group delay format, 7-25
sequence 2 setup, 2-29 frequency range, setting, 1-5 group delay measurements, 2-32
taking data, 2-29 frequency range, setting for time group delay principles, 7-29
tuned receiver mode, 2-26 domain low pass, 3-15
fixed markers, 1-29 frequency response, 7-40 H
fixtures, 6-50 frequency response and isolation harmonic measurements
designing your own fixture, 6-51 error corrections, 6-17 additional, 1-56
flat limit lines, 1-72 reflection measurements, 6-19 making, 1-55
floppy disk, what you can save, transmission measurements, harmonic operation, 7-87
4-35 6-17 accuracy, 1-58
form feed sequence, 4-24 frequency response error accuracy and input power, 7-88
sending to the printer, 4-24 corrections, 6-12 coupling power between
format, 7-9 receiver calibration, 6-15 channels 1 and 2, 1-58, 7-88
format arrays, 7-9 response error correction for dual-channel operation, 1-57,
format markers, polar, 1-32 reflection measurements, 7-87
formats, analyzer display, 7-24 6-12 frequency range, 1-58, 7-88
formatting a disk, 4-53 response error correction for input power, 1-58
forward stepping in edit mode, transmission single-channel operation, 1-57,
1-105 measurements, 6-14 7-87
forward transform frequency response of calibration test setup, typical, 7-87
measurements, 3-22 standards, 6-6 understanding , 1-57
demodulating the results of the electrical offset, 6-6 harmonics, measuring, 1-54
forward transform , 3-23 fringe capacitance, 6-6 high dynamic range
forward transform range, 3-24 fringe capacitance, 6-6 measurement, 2-22
interpreting the forward full two-port calibration, 7-55 swept RF/IF conversion loss,
transform horizontal axis, full two-port error correction, 6-29 2-18
3-23 high dynamic range
interpreting the forward G measurement, 2-22
transform vertical axis, 3-22 gain, 1-63 high dynamic range swept RF/IF
forward transform mode, 3-4 gain compression conversion loss
forward transform range, 3-24 measuring, 1-59 measurement parameters for IF
four-channel display using linear sweep, 1-61 range, 2-18
4 Param Displays softkey, 1-18 gating, 3-35, 7-8 power meter calibration over IF
Channel Position softkey , 1-17 selecting gate shape, 3-36 range, 2-18
customizing, 1-17 setting the gate, 3-35 power meter calibration over RF
viewing, 1-14 general, 8-7 range, 2-21
Freelance, using , 4-21 generating files in a loop counter receiver calibration over IF
frequency example, 1-115 range, 2-20
segments, editing, 6-34 German FTZ emissions RF frequency range, 2-21
signals, deleting, 6-35 requirements, 8-8 horizontal axis, 3-13, 3-14, 3-16,
span, decreasing , 5-10 gosub sequence command, 1-106 3-20, 3-23
span, setting, 1-36 GPIB operation how RF and IF are defined, 2-7
frequency bands address menu, 7-79 HPGL
adding, 1-84 GPIB STATUS indicators, 7-78 compatible printer, 4-19
changing, 1-83 local key, 7-77 initialization sequence, sending
deleting, 1-85 pass control mode, 7-78 to the printer, 4-24
setting, 1-82 system controller mode, 7-78 initialization sequence, storing,
frequency drift, 5-5 talker/listener mode, 7-78 4-23
frequency list sweep of 26 points, using the parallel port, 7-79 HPGL compatible printer, 4-23
2-26 GPIO mode, 1-106, 7-80 HPGL/2 compatible printer, 4-9
frequency offset mode operation, graphic files
2-10 saving measurement results as, I
frequency offset mode, conversion 4-45
compression, 2-37 IF bandwidth reduction, 7-35
grids
Index-4
Index
IF bandwidth, setting segment, in-depth sequencing information, interpolated error correction, 6-8
7-18 1-104 interpolation in power meter
IF delay, detecting, 5-10 autostarting sequences, 1-105 calibration, 6-34
IF detection, 7-7 commands completed before interpreting
IF range next sequence begins, 1-104 bandpass reflection response
measurement parameters, 2-18 commands that require a clean horizontal axis, 3-13
power meter calibration, 2-18 sweep, 1-105 bandpass reflection response
receiver calibration, 2-20 decision making functions, vertical axis, 3-13
IF, defining, 2-7 1-111 bandpass transmission response
imaginary format, 7-29 embedding loop counter value in horizontal axis, 3-14
improving measurement results, title, 1-105 bandpass transmission response
5-7 features that operate differently vertical axis, 3-14
decreasing the sweep rate, 5-8 in a sequence, 1-104 forward transform horizontal
decreasing time delay, 5-8 forward stepping in edit mode, axis, 3-23
improving raw source match and 1-105 forward transform vertical axis,
load match for TRL*/LRM* gosub sequence command, 1-106 3-22
calibration, 7-70 GPIO mode, 1-106 low pass response horizontal
increase sweep speed limit test decision making, axis, 3-16
using fast 2-port calibration, 1-111 low pass response vertical axis,
5-12 loop counter decision making, 3-17
increasing dynamic range, 5-14 1-112 low pass step transmission
increasing test port input power, sequence decision making response horizontal axis,
5-14 menu, 1-111 3-20
reducing receiver crosstalk, 5-14 sequence size, 1-105 low pass step transmission
reducing the receiver noise floor, sequence that jumps to itself, response vertical axis, 3-20
5-14 1-111 introduction to time domain
increasing measurement sequencing special functions measurements, 3-3
accuracy, 5-4 menu, 1-111 isolation, 7-40, 7-68
connector repeatability, 5-4 titles, 1-105 averaging, 6-63
frequency drift, 5-5 TTL I/O menu, 1-107 calibrating using ECal, 6-63
interconnecting cables, 5-4 TTL input decision making, calibration, omitting, 6-4
performance verification, 5-5 1-107, 1-111 error corrections and frequency
reference plane and port TTL out menu, 1-111 response, 6-17
extensions, 5-5 TTL output for controlling isolation example measurements,
temperature drift, 5-5 peripherals, 1-107 2-42
increasing sweep speed, 5-9 indicators, GPIB STATUS, 7-78 LO to RF isolation, 2-42
activating chop sweep mode, initializing loop counter value to RF feedthrough, 2-45
5-12 26, 2-27 SWR/return loss, 2-48
decreasing the frequency span, input ports menu, 7-23
5-10 input power, 1-58 J
reducing the averaging factor, inserting a command, 1-100 jpeg files, saving results as, 4-45
5-11 insertion phase response, 1-7, 1-8
reducing the number of instrument markings, 8-4
K
measurement points, 5-11 instrument modes, 7-83
setting the auto sweep time external source mode, 7-83 knowing the instrument modes,
mode, 5-11 frequency offset operation , 7-87 7-83
setting the sweep type, 5-11 harmonic operation, 7-87
using external calibration, 5-12 network analyzer mode, 7-83 L
using swept list mode, 5-9 tuned receiver mode, 7-85 labeling the screen, 2-30
viewing a single measurement instrument state leakage signals, eliminating
channel, 5-12 file, deleting, 4-51 unwanted, 2-6
widening the system bandwidth, files, 4-46 limit line operation, 7-81
5-11 re-saving , 4-51 edit limits menu, 7-82
increasing test port input power, saving, 4-36, 7-65 edit segment menu, 7-82
5-14 saving and recalling, 4-34 offset limits menu, 7-82
incrementing the source interconnecting cables, 5-4 limit lines
frequencies, 2-29 internal memory, 4-34 creating flat limit lines, 1-72
Index-5
Index
creating single point limits, 1-76 connecting the device under isolation example, 2-42
editing line segments, 1-77 test, 1-4 non-coaxial, 6-50
measurement parameters, 1-71 error-correction, 1-5 plotting to a disk, 4-11
offsetting limit lines, 1-79 frequency range, setting, 1-5 printing or plotting results, 4-3
running a limit test, 1-77 measurement, setting, 1-5 reflection response, 3-9
sloping limit line, 1-74 measuring the device under results, outputting, 1-6
using to test a device, 1-71 test, 1-5 results, saving, 4-37
limit test outputting measurement results, saving graphically, 4-45
decision making, 1-111 results, 1-6 setting, 1-5
example sequence, 1-117 source power, setting, 1-5 time domain, 3-3
limit test, running , 1-77 making harmonic measurements, transmission measurements in
activating the limit test, 1-78 1-55 time domain low pass, 3-19
reviewing the limit line making non-coaxial transmission response, 3-5
segments, 1-77 measurements, 6-50 measurement calibration, 7-37
line segments, editing, 1-77 fixtures, 6-50 accuracy enhancement, 7-37,
deleting line segments, 1-77 making reflection response 7-51
line types, selecting, 4-15 measurements, 3-9 characterizing microwave
linear frequency sweep, 7-15 making transmission response systematic errors, 7-41
linear magnitude format, 7-27 measurements, 3-5 measurement errors, 7-38
linear phase, deviation, 1-46 manual sweep time mode, 7-11 measurement considerations, 2-4
linear sweep, 1-61 manual thru, 6-62 eliminating unwanted mixing
linearity, phase, 2-32 margin and leakage signals, 2-6
list values, printing or plotting, ripple test value, 1-87, 1-90 frequency offset mode operation,
4-32 markers, 1-24 2-10
LO frequency accuracy and calculating statistics of how RF and IF are defined, 2-7
stability, 2-10 measurement data, 1-42 LO frequency accuracy and
LO to RF isolation, 2-42 continuous, 1-24 stability , 2-10
load match, 7-39, 7-70 coupling display markers, 1-31 minimizing source and load
load mismatches, minimizing, 2-4 CW frequency, setting, 1-38 mismatches, 2-4
loading a sequence from a disk, delta markers, 1-28 power meter calibration, 2-10
1-103 discrete, 1-24 reducing the effect of spurious
local key, 7-77 display markers responses, 2-5
locking onto a signal with activating, 1-25 measurement data, 1-20
frequency modulation fixed marker dividing by the memory trace,
component, 7-85 activating, 1-29 1-20
log magnitude format, 7-24 moving marker information off viewing, 1-20
logarithmic frequency sweep, 7-15 of the grids, 1-26 measurement data trace, 1-20
loop counter polar format markers, 1-32 subtracting memory trace, 1-20
decision making, 1-112 setting measurement measurement error
example sequence, 1-114 parameters, 1-34 crosstalk, 7-40
value, 2-27 smith chart markers, 1-33 frequency response, 7-40
loss of power meter calibration specific amplitude, searching isolation, 7-40
data , 6-33 for, 1-39 load match, 7-39
low pass impulse mode, 3-20 uncoupling display markers, measurement errors
lower stopband parameters, 1-67 1-31 directivity, 7-38
masking, 3-26 source match, 7-39
M matched adapters, 6-46 measurement parameters, 1-67,
magnitude maximum amplitude, searching 6-4
measuring , 1-7 for, 1-39 center frequency, setting , 1-35
measuring response, 1-7 maximum and minimum, 1-93 choosing, 1-4
maintenance, 8-2 maximum bandwidth, 1-93 display reference value, setting,
making a basic measurement, 1-4 measurement 1-37
choosing measurement accuracy, increasing, 5-4 electrical delay, setting, 1-37
parameters, 1-4 calibration, power meter, 6-33 for IF range, 2-18
connecting required test fault location using low pass, frequency span, setting, 1-36
equipment, 1-4 3-18 lower stopband parameters,
high dynamic range, 2-22 1-67
Index-6
Index
markers, setting with, 1-34 input ports menu, 7-23 calibration kit menu, 7-57
passband parameters, 1-67 offset limits, 7-82 saving to a disk, 7-65
start frequency, setting, 1-34 segment, 7-16 verifying performance, 7-64
stop frequency, setting, 1-35 S-parameter, 7-22 module information, 6-66
upper stopband parameters, stepped edit list, 7-16 multiple
1-68 stepped edit subsweep, 7-16 measurements per page,
measurements swept edit list, 7-17 plotting from a disk, 4-26
basic, 1-4 swept edit subsweep, 7-17 measurements, plotting on a full
phase or group delay, 2-32 microprocessor, 7-4 page, 4-27
measuring microwave connector care, 5-3 plots, outputting to a single page
device under test, 1-5 microwave systematic errors, using a plotter, 4-25
gain and reverse isolation characterizing, 7-41
simultaneously, 1-63 minimizing N
insertion phase response, 1-7, error when using adapters, 6-49 N dB Point, 1-93
1-8 source and load mismatches, 2-4 names for CSV files, 4-44
separate transmission paths minimum allowable stop naming files generated by a
through the test device frequencies, 3-16 sequence, 1-102
using low pass impulse minimum amplitude, searching network analyzer mode, 7-83
mode, 3-20 for, 1-39 noise reduction techniques, 7-34
small signal transient response minimum bandwidth , 1-93 averaging, 7-34
using low pass step, 3-19 minimum sweep time, 7-11 IF bandwidth reduction, 7-35
measuring amplifiers, 1-53 mixer smoothing, 7-35
harmonic operation, fixed IF measurements, 2-24 non-coaxial
understanding, 1-57 measurement, 2-3 making measurements, 6-50
harmonics, measuring, 1-54 measurement diagram, using, non-coaxial devices, calibrating
measuring gain and reverse 2-15, 2-21 for, 6-52
isolation simultaneously, mixing signals, eliminating noninsertable device, 6-71
1-63 unwanted, 2-6 noninsertable devices, calibrating
measuring gain compression, mode for, 6-40
1-59 auto sweep time, 5-11, 7-11
measuring electrical length, 1-43 chop sweep, 5-12
O
measuring gain compression, 1-59 continuous correction, 6-38
linear sweep, 1-61 copy, 7-79 offset and scale, 7-9
using linear sweep, 1-61 external source, 7-83 offset limits menu, 7-82
measuring harmonics, 1-54 frequency offset, 2-37 offset, electrical, 6-6
additional harmonic GPIO , 7-80 offsetting limit lines, 1-79
measurements, 1-56 low pass impulse, 3-20 omitting isolation calibration, 6-4
making harmonic manual sweep time, 7-11 one-port calibration, S11 and S22,
measurements, 1-55 network analyzer, 7-83 7-55
measuring magnitude, 1-7 pass control, 7-78 one-port error model, 7-41
magnitude response, 1-7 system controller, 7-78 one-port reflection error
measuring phase distortion, 1-43, talker/listener, 7-78 correction, 6-26
1-45 time domain bandpass, 3-12 operating parameters, printing or
deviation from linear phase, time domain low pass, 3-15 plotting, 4-32
1-46 tuned receiver, 7-85 operation
group delay, 1-46 model dual-channel, 1-57
memory math functions, 1-19 one-port error, 7-41 single-channel, 1-57
memory trace, 1-19, 1-20 two-port error, 7-46 operation, frequency offset, 7-87
viewing, 1-20 modified colors, 1-23 operation, GPIB, 7-77
memory, display, 7-9 recalling, 1-23 operation, harmonic, 7-87
menu saving, 1-23 operation, limit line, 7-81
address, 7-79 modify colors menu, 1-22 operation, system, 7-3
analog in, 7-22 modifying operation, trace math, 7-8
calibration kit, 7-57 cal kit through definition, 6-47 output power, 7-10
conversion, 7-22 command, 1-100 power coupling options, 7-10
edit limits, 7-82 standard definitions, 6-52 outputting
edit segment, 7-82 modifying calibration kits, 7-56 measurement results, 1-6
Index-7
Index
multiple plots to a single page phase linearity and group delay, power meter
using a plotter, 4-25 2-32 calibration, 2-10
plot files, 4-12 phase measurements, 2-32 calibration over IF range, 2-18
plot files from a PC to a plotter, places where you can save, 4-34 calibration over RF range, 2-21
4-22 plot power meter measurement
single page plots using a printer, aborting a process, 4-31 calibration, 6-33
4-24 plot files calibrating the analyzer receiver
outputting plot files from a PC to outputting from a PC to a to measure absolute power,
an HPGL compatible printer, plotter, 4-22 6-39
4-23 outputting from a PC to an compensating for directional
sending the exit HPGL mode HPGL compatible printer, coupler response, 6-35
and form feed sequence to 4-23 entering the power sensor
the printer, 4-24 sending to the printer, 4-24 calibration data, 6-34
sending the HPGL initialization to output, 4-12 interpolation in power meter
sequence to the printer, to view on a PC, 4-20 calibration, 6-34
4-24 plot function, configuring, 4-9 loss of power meter calibration
sending the plot file to the plotting to a pen plotter, 4-10 data, 6-33
printer, 4-24 plotting to an HPGL/2 using continuous correction
storing the exit HPGL mode and compatible printer, 4-9 mode, 6-38
form feed sequence, 4-24 plot function, defining, 4-13 using sample-and-sweep
storing the HPGL initialization choosing display elements, 4-13 correction mode, 6-36
sequence, 4-23 choosing plot speed, 4-16 power sensor calibration data,
choosing scale, 4-15 entering, 6-34
P resetting plotting parameters to power sweep, 7-19
page quadrants, plotting default values, 4-16 power, output, 7-10
measurements in, 4-28 selecting auto-feed, 4-13 primary measurement channels,
parameters selecting line types, 4-15 viewing, 1-12
lower stopband, 1-67 selecting pen numbers and principles, group delay, 7-29
measurement, 1-4, 1-67, 6-4 colors, 4-14 print
passband, 1-67 plot speed, choosing, 4-16 aborting a process, 4-31
upper stopband, 1-68 plotting print function
pass control mode, 7-78 measurement results, 4-3 configuring, 4-4
passband parameters, 1-67 measurement to a disk, 4-11 defining, 4-6
PC interface unit, 6-61 measurements in page printer
PC, to view files on, 4-20 quadrants, 4-28 color printer, using, 4-6
pen numbers and colors, selecting, multiple measurements on a full HPGL compatible printer, 4-19,
4-14 page, 4-27 4-23
pen plotter, 4-17, 4-18 multiple measurements per HPGL/2 compatible printer, 4-9
plotting to, 4-10 page from a disk, 4-26 printing
performance verification, 5-5 multiple measurements per measurement results, 4-3
performance, verifying , 7-64 page using a pen plotter, multiple measurements per
performing 4-18 page, 4-8
2-port error corrections, 6-42, one measurements per page one measurement per page, 4-7
6-73 using a pen plotter, 4-17 parameters, resetting to default
TRL calibration, 6-54 parameters, resetting to default values, 4-7
TRM calibration, 6-58 values, 4-16 sequence, 1-104
peripheral equipment, required, plotting to an HPGL compatible solving problems, 4-33
7-5 printer, 4-19 printing or plotting the list values
per-raw data arrays, 7-8 solving problems, 4-33 or operating parameters, 4-32
phase plotting a measurement to a disk entire list of values, 4-32
linearity, 2-32 to output the plot files, 4-12 single page of values, 4-32
measurements, 2-32 polar format, 7-27 procedure, TRL calibration, 7-71
tracking , 2-36 polar format markers, 1-32 procedures for error correcting
phase distortion, 1-43, 1-45 port extensions, 5-5 measurements, 6-10
phase format, 7-24 power coupling options, 7-10 types of error correction, 6-10
phase or group delay channel coupling, 7-10 processing details, 7-7
measurements, 2-32 test port coupling, 7-10 accuracy enhancement, 7-8
Index-8
Index
Index-9
Index
Index-10
Index
linear frequency sweep, 7-15 test port input power, increasing , time domain low pass step mode,
logarithmic frequency sweep, 5-14 3-4
7-15 test sequencing, 1-97 time domain measurements,
power sweep, 7-19 changing the sequence title, introduction, 3-3
selecting sweep modes, 7-19 1-102 forward transform mode, 3-4
stepped list frequency sweep, clearing a sequence from time domain bandpass mode,
7-15 memory, 1-101 3-4
swept list frequency sweep, 7-17 creating a sequence, 1-97 time domain low pass impulse
sweep-to-sweep averaging, 7-8 editing a sequence, 1-99 mode, 3-4
swept edit list menu, 7-17 generating files in a loop counter time domain low pass step
swept edit subsweep menu, 7-17 example, 1-115 mode, 3-4
swept list frequency sweep, 7-17 in-depth sequencing time stamp, 4-31
setting segment IF bandwidth, information , 1-104 title, 1-105
7-18 inserting a command, 1-100 title, display, 1-11
setting segment power, 7-18 limit test example sequence, titling the displayed
swept edit list menu, 7-17 1-117 measurement, 4-30
swept edit subsweep menu, 7-17 loading a sequence from a disk, to produce a time stamp, 4-31
swept list mode 1-103 trace math operation, 7-8
calibrate, 1-69 loop counter example sequence, trace noise, reducing, 5-15
characteristics of the filter, 1-66 1-114 tracking, 7-40
device under test, connect, 1-65 modifying a command, 1-100 tracking the amplitude, 1-41
measure, 1-69 naming files generated by a tracking, amplitude and phase,
measurement parameters, 1-67 sequence, 1-102 2-36
stepped list mode, 1-65 printing a sequence, 1-104 transform, 7-9
to test a device, 1-65 purging a sequence from a disk, transforming CW time
swept list mode, using, 5-9 1-103 measurements into the
swept RF/IF conversion loss, high running a sequence, 1-99 frequency domain, 3-22
dynamic range, 2-18 stopping a sequence, 1-99 forward transform
switch protection, source storing a sequence on a disk, measurements, 3-22
attenuator, 7-13 1-103 transmission measurements in
SWR format, 7-28 using to test a device, 1-113 time domain low pass, 3-19
SWR/return loss, 2-48 text file, saving measurements as interpreting the low pass step
synthesized source, built-in, 7-4 a, 4-43 transmission response
system thru, manual, 6-62 horizontal axis, 3-20
bandwidth, changing, 5-15 time delay, decreasing, 5-8 interpreting the low pass step
bandwidth, widening, 5-11 time domain bandpass mode, 3-4, transmission response
system controller mode, 7-78 3-12 vertical axis, 3-20
system operation, 7-3 adjusting the relative velocity measuring separate
built-in synthesized source, 7-4 factor, 3-12 transmission paths through
microprocessor, 7-4 reflection measurements using the test device using low
receiver block, 7-4 bandpass mode, 3-12 pass impulse mode, 3-20
required peripheral equipment, transmission measurements measuring small signal
7-5 using bandpass mode, 3-14 transient response using
systematic errors, 7-41 time domain low pass impulse low pass step, 3-19
mode, 3-4 transmission measurements
T time domain low pass mode, 3-15 using bandpass mode, 3-14
taking care of microwave fault location measurements interpreting the bandpass
connectors, 5-3 using low pass, 3-18 transmission response
talker/listener mode, 7-78 minimum allowable stop horizontal axis, 3-14
target amplitude, searching for, frequencies, 3-16 interpreting the bandpass
1-40 reflection measurements in time transmission response
temperature drift, 5-5 domain low pass, 3-16 vertical axis, 3-14
terminology, TRL, 7-67 setting frequency range for time transmission measurements,
test domain low pass, 3-15 response and isolation error
bandwidth, 1-91–1-96 transmission measurements in correction, 6-17
ripple limit, 1-80–1-90 time domain low pass, 3-19
test port coupling, 7-10
Index-11
Index
transmission measurements, in-depth description, 7-86 finite impulse width (or rise
response error correction, test setup, typical, 7-86 time), 3-27
6-14 two sources sidelobes, 3-27
transmission response addressing and configuring,
measurements, making, 3-5 2-27
TRL calibration, performing, 6-54 two-port calibration, full, 7-55
TRL error correction two-port calibration, TRL*/LRM*,
assigning standards to various 7-55
TRL classes, 6-53 two-port error model, 7-46
label the calibration kit, 6-53 type-N connector sex
label the classes, 6-53 clarifying, 6-4
performing the TRL calibration, types of error correction, 6-10
6-54
TRL error-correction, 6-52 U
creating a user-defined TRL uncoupling display markers, 1-31
calibration kit, 6-52 understanding
TRL options, 7-75 harmonic operation, 1-57
TRL terminology, 7-67 spur avoidance, 5-17
TRL* error model, 7-67 understanding S-parameters,
TRL*/LRM* calibration, 7-66 7-20
fabricating and defining upper stopband parameters, 1-68
calibration standards for using continuous correction mode,
TRL/LRM, 7-72 6-38
how TRL*/LRM* works , 7-67 using external calibration, 5-12
improving raw source match and using fast 2-port calibration, 5-12
load match for TRL*/LRM* using sample-and-sweep
calibration, 7-70 correction mode, 6-36
isolation , 7-68 using swept list mode, 5-9
source match and load match, detecting IF delay, 5-10
7-69 using the parallel port, 7-79
TRL calibration procedure, 7-71 copy mode, 7-79
TRL options, 7-75 GPIO mode, 7-80
TRL standards, requirements,
7-71
V
TRL terminology, 7-67
TRL* error model, 7-67 vector error-correction, 7-8
TRL*/LRM* two-port calibration, verification, performance, 5-5
7-55 verifying performance, 7-64
TRM error correction, 6-56 vertical axis, 3-13, 3-14, 3-17,
assigning standards to various 3-20, 3-22
TRM classes, 6-57 viewing a single measurement
creating a user-defined TRM channel, 5-12
calibration kit, 6-56 viewing plot files on a PC, 4-20
labeling the calibration kit, 6-57 using AmiPro, 4-21
labeling the classes, 6-57 using Freelance, 4-21
modifying the standard
definitions, 6-56 W
performing the TRM what you can save to a computer,
calibration, 6-58 4-35
TTL what you can save to a floppy disk,
I/O menu, 1-107 4-35
input decision making, 1-107 what you can save to the
out menu, 1-111 analyzer’s internal memory,
output for controlling 4-34
peripherals, 1-107 widening the system bandwidth,
tuned receiver mode, 2-24, 2-26, 5-11
7-85 windowing, 3-27
Index-12