APT2000 Series 2-Wire Contacting Conductivity Transmitters Specification
APT2000 Series 2-Wire Contacting Conductivity Transmitters Specification
APT2000 Series 2-Wire Contacting Conductivity Transmitters Specification
2/00
2-Wire Contacting Page 1 of 8
Description
Figure 1 APT2000 Transmitter
The Honeywell APT 2000 series of
transmitters offer the widest available
selection of advanced features in a Quick Problem Assessment constant feedback to the operator on
reliable and economical instrument. The APT2000 has a large front whether or not there is a problem
display for quick recognition of with the cell. These easily learned
Reliability First process parameters and diagnostics and recognized symbols make the
The advanced features of the even at a distance. Only the APT2000 an easy-to-use instrument
APT2000 transmitter guarantee APT2000 employs visual feedback to in any language.
complete reliability. The APT2000 quicken setup and maintenance
continuously monitors sensor and times and to minimize errors made Foolproof Calibrations
transmitter electronics and during calibrations. Visual feedback Calibration is easy with automatic
immediately displays diagnostic refers to pictograph type characters buffer recognition, utilizing a variety
information at the onset of a that appear on the display both to of buffer standards recognized
problem. If an error or diagnostic is prompt and respond to operator and worldwide. Foolproof calibrations
found, the transmitter will indicate the process changes. are ensured with special on-board
appropriate error code or pictograph diagnostics that monitor the step-by-
(see Figure 2), blink a red LED and Pictograph type characters also step progress of each calibration.
adjust the error current to 22 mA if appear during problem conditions to Simple messages with pictographs
desired. A manual loop-back check report diagnostics for easy trouble- provide feedback on the status of the
is available to test the integrity of the shooting. There is even a calibration. Manual one or two-point
4-20mA output. Sensoface pictograph that provides calibrations can also be performed
for complete flexibility.
Conductivity
Temperature Cell Check
Loop Alarm Calibration
Display/ Current Setting Interval/
Process Variable
Response Time
Conductivity
Continue with
Cell Data
ENTER
Features Applications
• Large display with easy-to-read • Robust, tightly sealed plastic The APT2000 CC transmitter is
0.75 inch measured value enclosure designed to meet the measurement
• Simple operator interface with • Wall, pipe or panel mounting needs of a number of industries,
basic pictographs • Easy installation with pre- including:
• Application in hazardous and assembled empty enclosure and
safe areas plug-in terminals • Power
• HART bi-directional • Optical alarm signaling by • Chemical
communications protocol blinking red LED • Pharmaceutical
• Continuous diagnostics for • Integrated current source for • Pulp and Paper
monitoring calibration, cell simple checking of peripheral • Water Quality
devices
health, and transmitter self-test • Metals
• Quick Response Time (less than
• Manual loopback check for
five seconds per step change)
integrity of 4-20 mA output
70-82-03-33
Page 3 of 8
Specifications
Conductivity Input
Display Range (0.2 µS/cm * Cell Constant) to (1000 mS/cm * Cell Constant)
Accuracy Less than 1% of Measured Value or +/- (0.4 µS/cm * Cell Constant), whichever is greater
Step Change Response Less than 5 seconds
Time
Process Variable/Range 0.000 to 9.999 µS/cm, 00.00 to 99.99 µS/cm, 000.0 to 999.9 µS/cm, 0000 to 9999 µS/cm
0.000 to 9.999 mS/cm, 00.00 to 99.99 mS/cm, 000.0 to 999.9 mS/cm
0.000 to 9.999 MΩ-cm, 00.00 to 99.99 MΩ-cm, 000.0 to 999.9 MΩ-cm
Salinity 0.0 to 45.0% (0 to 35 °C)
Diagnostics
Sensocheck Polarization detection and monitoring of cable capacitance (can be switched off)
Sensor Standarization - Entry of cell calibration factor with display of conductivity and temperature
- Temperature probe adjustment
Sensoface - Provides information on the electrode state via Sensocheck
- Monitors asymmetry potential, slope, and response time during calibration
Temperature Input
Range 8550 Ω Thermistor: -10.0 to +130.0 °C / -14 to +266 °F
Pt100/1000 Ω RTD: -20.0 to +150.0 °C / +4 to +302 °F
Resolution 0.1 °C or 1 °F
Accuracy < 0.5 K
Temperature Automatic Compensation using Pt 100Ω or 1000Ω RTD, 8550Ω Thermistor or manual adjust
Compensation
Display LCD display 76 mm x 48 mm dimensions (3” x 1 7/8”), 7-segment
Cond Value: character height 17 mm (.66") , meas. symbol 10 mm (.4")
Temperature: character height 10 mm (.4"), meas. symbol 7 mm (.33")
Sensoface with three states, 5 status bars, 16 pictographs / symbols, Red Alarm LED
Security protection with four-digit mode codes to access calibration and configuration options
Conductivity Range
10 Cell Constant
1 Cell Constant
0.1 Cell Constant
0.01 Cell Constant
Resistivity Range
0.1 Cell Constant
0.01 Cell Constant
100 10 1.0 0.1 0.01
MΩ−cm
Figure 3 Optimal Cell Constant for Conductivity or Resistivity Ranges
70-82-03-33
Page 4 of 8
Supply/Output
Output current 4 to 20 mA (22 mA for error notification) current loop, floating (3.8 to 20.5 mA)
Supply voltage 12 to 42 V; Imax = 100 mA; Pmax = 0.8 W
Overrange 22 mA for error messages
Current error <0.3 % of current value +0.05 mA
Current source 3.80 to 22.0 mA
20 0 0
17 5 0
15 0 0 1360 Ohm s
at 42 VDC
12 5 0
10 0 0
750
Operati ng
500 Regi on
250
Wi thout HART Communi c ati on
0
12. 0 V DC 18. 0 24. 0 30. 0 36. 0 42. 0
Power Supply Voltage
Communications
HART Protocol - Digital communication via FSK modulation of the loop current
- Point-to-point connection
- Reading of measured values, status, messages, and multidrop unit identification
- Read and write parameters
Physical
Enclosure Plastic enclosure made of PBT (polybutylene terephthalate) bluish-gray RAL 7031
Mounting Wall, Pipe, or Panel Mount
Dimensions H 144 mm, W 144 mm, D 105 mm (H 5.67", W 5.67", D 4.13")
Protection - IP 65: Dust-tight, protection from powerful water jets (USA, Canada: indoor use only)
- Corrosion-proof housing
Cable glands 3 breakthroughs for Pg 13.5
2 breakthroughs for NPT 1/2” or Rigid metallic conduit
Weight Approx. 1 kg (2.2 lbs)
70-82-03-33
Page 5 of 8
11 10 9 8 7 6 5 4 3 2 1
+
No Connection
4 to 20 mA
shield
No Connection
HART®
GR
WH
RD
BL
Supply/
Output Contacting
Conductivity Cell
Instructions
APT2000 Transmitter Offers:
Power Requirements - 2 Wire, 12 to 42 VDC
Standard - All models CE Compliant
Standard - IP65 Housing; design meets corrosion and waterproof
requirements of NEMA 4X
Optional - HART communications
Approval Options:
1. FM/CSA Class I, Div. 2, Groups A-D (Non-Incendive), T4
2. FM/CSA Class I, Div. 1, Groups A-D (Intrinsically Safe), T4
& CENELEC EEx ib [ia] IIC T6/T4 (Intrinsically Safe)
Select the desired key number. The arrow to the right marks the selection available.
Make one selection from Tables using the column below the proper arrow.
A dot ( ) denotes unrestricted availability.
TABLE II - Approvals
General Purpose: FM/CSA Class I, Div. 2, Groups A - D 00
Intrinsically Safe: FM/CSA Class I, Div. 1, Groups A - D IS
& CENELEC EEx ib [ia] IIC T6/T4
Figure 6 Dimension Drawing for APT2000 and P/N 51205990-001 panel mounting kit
70-82-03-33
Page 8 of 8
WARRANTY/REMEDY
Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local
sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell
will repair or replace without charge those items it finds defective. The foregoing is Buyer’s sole remedy and is in lieu of all
other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. Specifications
may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we
assume no responsibility for its use.
While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer
to determine the suitability of the product in the application.