Barrier-Free Power Exam Table 625: Service and Parts Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 133

Barrier-Free®

Power Exam Table


Model Numbers:
Service and
Parts Manual
625

FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY

004-0763-00 Rev. AA8 (3/14/19)


GENERAL INFORMATIOn access procedures
Symbols............................................................iii Manually Raising / Lowering Table................. C-6

General Info
Ordering Parts..................................................iii Removal / Installation:

Section C
Table of Contents Serial Number Location....................................iii
Specifications.................................................. iv
Model Identification /
Compliance Chart.......................................... vi
Scheduled Maintenance
Cladding................................................... C-2
Table Shrouds.......................................... C-3
Upholstery................................................ C-7
Chair Arms / Brace................................... C-8
Top Cover................................................. C-9
Cleaning Chart...............................................vii

Troubleshooting
Troubleshooting Chart.................................. A-2
WIRING DIAGRAMS

Section D
Power to the Chair: 625-001.......................................................D-2*
120 V models.......................................... A-10 625-003.......................................................D-3*
230 V models...........................................A-11 625-004........................................................ D-4
Section A

Base Up / Down.......................................... A-12 625-005........................................................ D-5


Back Up / Down.......................................... A-14 625-006........................................................ D-6
Home Function........................................... A-16
Drawer Heater System............................... A-18 Exploded VIEWS / PARTS LIST

Section E
Table Receptacle(s).................................... A-19
Scale Information Flow............................... A-20 625 (-001 /-003 /-005 / -006)............................ E-2
625 (-004)........................................................... E-3

tESTING REpair procedures & forms

Digitally Linked Files


Foot / Hand Controls.................................... B-2 Base Actuator.................................003-2072-00
Wireless Controls......................................... B-4 Back Actuator.................................003-2071-00
Base Function Components......................... B-7 Main PC Board...............................003-2073-00
Scale / USB PC Boards..................003-2454-00
Section B

Back Function Components........................B-11


Drawer Heater System............................... B-15 LIN to VSM PC Board.....................003-2455-00
Table Calibration Procedure....................... B-20 Wireless Basestation PC Board.....003-2471-00
Scale Calibration Procedure....................... B-21
Scale Boards / LED Testing........................ B-22 Color Selector.......................www.midmark.com

(*) Indicates multiple pages due to model / serial number break(s).

© Midmark Corporation 2009 SF-1935 [Revised: 9/20/13] ii


Symbols Ordering Parts

General Information
The following information is required when ordering parts:
danger • Serial number & model number
Indicates an imminently hazardous situation which • Part number for desired part
will result in serious or fatal injury if not avoided. (Refer to Section E: Exploded Views & Parts Lists)
This symbol is used only in the most extreme conditions.
Non-warranty part orders may be faxed to Midmark using
warning the Fax Order Form in the back of this manual.
Indicates a potentially hazardous situation which
could result in serious injury if not avoided. For warranty part orders, call Midmark’s Technical Service
Department with the required information.
Hours: 8:00 am to 5:00 p.m. EST (Monday thru Friday)
Caution Phone: 1-800-Midmark (1-800-643-6275)
Indicates a potentially hazardous situation which may
result in minor or moderate injury if not avoided. It may
also be used to alert against unsafe practices.
Serial Number Location
Equipment Alert
Indicates a potentially hazardous situation which
could result in equipment damage if not avoided.
Model / Serial Number Label
Located on the Back Mounting Frame
(exact location may vary)
Note
Used for special instructions or additional information.

The symbols below may be used in this manual to represent


the operational status of table functions and components.

Indicates the function / component is working properly.


No action required.

Indicates the function / component is working,


but a problem exists.

Indicates the function is not working at all, or that the


component is faulty.

© Midmark Corporation 2009 SF-1935 [Revised: 6/15/12] iii


Specifications 625 (-001 /-003 /-005 /-006)
Range of Motion, Weights, Electrical Specifications

Quick Exam™ Height Home Height


Specifications Chart Programmable 20 ± 0.5 in.
(set by the user) (50.8 ± 1.3 cm)
Patient Weight (maximum): 650 lbs (295 kg)
Intended Weight Between 30 and 650 pounds (13.6/295kg)
Weight of Table:
w/upholstery 470 lbs (213 kg)
w/packaging & skid (no uph.) 487 lbs (221 kg) Max.
Uph. w/packaging (shipped separately) 45 lbs (20 kg) 37 ± 0.5 in.
(94 ± 1.3 cm)
Power Cord Length: 8 ft. (244 cm)
Electrical Requirements: (See Regulatory Compliance Chart)
Foot / Hand Control Voltage: 14 VAC, SELV (Safety Extra Low Voltage) Min.
Duplex Receptacle(s) (maximum load) 115 VAC, 5A, 50/60 Hz 18 ± 0.5 in.
(45.7 ± 1.3 cm)
Fuses:
IEC inlet fuses (2) 6.3A, 250V, Type-T, 5 x 20 mm
F1 & F2 (on main PC board): 10A, 250V, Type-T, 5 x 20 mm
F3 (on main PC board): 160mA, 250V, Type-T, 5 x 20 mm
F4 (on main PC board) 6.3A, 250V, Type-T, 5 x 20 mm
Models w/ drawer heaters
distribution board fuses (2) 800mA, 250V, Type-T, 5 x 20mm 0° (horizontal) to
+80° ±5°
Duty Cycle
(Motor Run Time): Intermittent Operation Pelvic Tilt
[30 seconds ON / 5 minutes OFF] Max. 7° ±2°
Receptacle(s) & Drawer Heater: Continuous Operation Min. 1.5°
Classifications: Class 1, Type B Applied Part
Protection against ingress of fluids: IPXO [Foot Control only: IPX1]

MA8023i

warning
Equipment not suitable for use in the presence of a flammable anesthetic mixture.

© Midmark Corporation 2009 SF-1935 [Revised:7/12/18 ] iv


Specifications 625 (-004)
Range of Motion, Weights, Electrical Specifications

Quick Exam™ Height Home Height


Programmable 20.5 ± 0.5 in.
Specifications Chart (set by the user) (52.1 ± 1.3 cm)
Patient Weight (maximum): 650 lbs (295 kg)
Intended Weight Between 30 and 650 pounds (13.6/295kg)
Weight of Table:
w/upholstery 475 lbs (216 kg)
w/packaging & skid (no uph.) 502 lbs (228 kg) Max.
Uph. w/packaging (shipped separately) 45 lbs (20 kg) 37.5 ± 0.5 in.
(95.3 ± 1.3 cm)
Power Cord Length: 8 ft. (244 cm)
Electrical Requirements: (See Regulatory Compliance Chart)
Foot / Hand Control Voltage: 14 VAC, SELV (Safety Extra Low Voltage) Min.
18.5 ± 0.5 in.
Simplex Receptacles (maximum load) 115 VAC, 5A, 50/60 Hz (47 ± 1.3 cm)
Fuses:
IEC inlet fuses (2) 6.3A, 250V, Type-T, 5 x 20 mm
F1 & F2 (on main PC board): 10A, 250V, Type-T, 5 x 20 mm
F3 (on main PC board): 160mA, 250V, Type-T, 5 x 20 mm
F4 (on main PC board) 6.3A, 250V, Type-T, 5 x 20 mm
Models w/ drawer heaters 0° (horizontal) to
+80° ±5°
distribution board fuses (2) 800mA, 250V, Type-T, 5 x 20mm
Duty Cycle
(Motor Run Time): Intermittent Operation Pelvic Tilt
[30 seconds ON / 5 minutes OFF] Max. 7° ±2°
Receptacles & Drawer Heater: Continuous Operation Min. 1.5°

Classifications: Class 1, Type B Applied Part


Protection against ingress of fluids: IPXO [Foot Control only: IPX1]

warning
Equipment not suitable for use in the presence of a flammable anesthetic mixture.

© Midmark Corporation 2009 SF-1935 [Revised: 7/12/18] v


Model Identification / Compliance Chart

Fire Code Ratings:

All upholstery complies with California Bureau of Home Furnishing Technical Bulletin 117 and
California Code of Regulations, Sect. 93120-93120.12, Title 17.Optional upholstery is available
that complies with California Bureau of Home Furnishing Technical Bulletin 133.

Complies To: Electrical Ratings:

Model Description CAN / CSA EN


UL EN NFPA VAC Cycles
22.2, 60601-1-2 Amps
60601-1 60601-1 99 +/- 10% (Hz)
#601.1-M90 (EMC)
625-001 Two-function table
(Base / Back), w/ receptacle(s), • • • • 115 12 50/60
pelvic tilt, & drawer heater
625-003 Two-function table
(Base / Back), w/ receptacle(s),
pelvic tilt, drawer heater, & wire- • • • • 115 12 50/60
less controls
625-004 Two-function table
(Base / Back), w/ receptacles,
pelvic tilt, drawer heater, • • • • 115 12 50/60
wireless controls, IQscale™,
and IQhub™
625-005 Two-function table
(Base / Back), w/ pelvic tilt • • 230 4 50/60

625-006 Two-function table


(Base / Back), w/ pelvic tilt, & • • 230 4.5 50/60
drawer heater

© Midmark Corporation 2009 SF-1935 [Revised: 6/15/12] vi


Scheduled Maintenance / Cleaning Chart

Interval Inspection / Service Description

Clean upholstery with appropriate diluted bleach solution 10:1 (water: bleach)
Weekly
Cleaning
Wipe painted metal & plastic surfaces with a clean soft cloth and mild cleaner.

Obvious Damage Visually inspect components for damage that could result in unsafe operation.

Check all mechanical functions using the foot control. Repeat using the hand control.
Semi-Annually
Mechanical Operation Table shrouds should move smoothly & quietly when base is raised & lowered.

(NOTE: There are plastic glides on the shroud tabs, Missing glides may result in noisy operation.)

Labels / Decals Replace any missing or illegible labels.

Lubricate back hinge with light machine oil.


Lubrication
Lubricate footrest slide with household furniture wax.

Hardware All fasteners must be present and fastened securely.

Inspect power cord and all wiring for damage.


Electrical System
Be sure all electrical connections are tight.

Stirrups Check that stirrups extend easily, and lock securely into each lateral position.

Scale Verify the scale is measuring accurately using a reference weight of known value.

Date of Service: Model Number:

Location: Serial Number:

Service Technician: Notes:

© Midmark Corporation 2009 SF-1935 [Revised: 2/12/18] vii


Troubleshooting Troubleshooting Chart........................A-2
Power to the Chair
120 V models................................A-10
230 V models................................ A-11
Base Up / Down...............................A-12

Section A Back Up / Down................................A-14


Home Function.................................A-16
Drawer Heater System.....................A-18
Table Receptacle(s)..........................A-19

© Midmark Corporation 2009 SF-1935 [Revised: 9/20/13] A-1


Quick Reference for 625 Scale Table (also refer to 625 Troubleshooting Chart section):
If below information does not resolve the issue, then contact technical service at 1-800-MIDMARK (643-6275).

Run the base of the


table all the way to the
bottom by pressing and Is table
holding the base down Yes “seat” higher Table Lo Scale does not work. Yes
button until the table than 22”?
stops and the press the No Readings
weight button again

No
Does the display
turn on (dashed lines,
Press “weight” button etc.) when a scale No Replace batteries,
again being careful not button is pressed verifying correct polarity.
to hit any other buttons (weight or zero)?

Yes

Verify remote is the Does display


one that was shipped come on?
with table
Will remote
cause table to
move using base No
or back buttons?

Check battery compartment


for broken battery terminals
Is remote a
replacement? Yes

Verify all four leveling


feet are securely on the Replace hand control,
Yes floor, no rocking reassociate.

Associate hand control,


with table.
Verify the patient (if
any) is not moving
around on the table.

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-2


Troubleshooting Chart
Problem Symptom Probable Cause Check Correction
No functions will operate. All functions inoperable. Facility supply voltage Power cord connections & Secure power cord connections.
facility circuit breaker. Reset circuit breaker if necessary.
Main PC board fuse(s) blown Test F1, F2, F3, & F4 fuses Refer to:
Section D - Wiring Diagrams
Loose wire connections Wire connections between Secure loose wire connections.
power inlet and main PC
board.
Wireless controls (if applicable) Refer to: Section B - Wireless Perform wireless controls test.
Controls
Overtravel limit switch Perform limit switch test Replace faulty limit switch.
(Later models only) Section B - Limit Switch Test
Base function not operating No Base Up or Base Down. Loose wire connections Wire connections Secure loose wire connections.
properly. (esp. Foot/Hand control(s) and
base up/down limit switches).
Main PC board fuse(s) blown Test F1, F2, F3, & F4 fuses Refer to:
Section D - Wiring Diagrams
Foot / Hand control (refer to: Perform Foot / Hand control Refer to:
Wireless Controls if applicable) test. Section B - Foot / Hand Controls
Base actuator / PC board Perform PC board tests. Refer to: Section B -
PC Board Test
Table stops and beeps Patient weight exceeded 650 lbs If patient weight exceeded the Inform staff that max. patient
650 lb. weight limit. weight is 650 lbs.
No Base Up / Base Down OK Loose wire connections Wire connections Secure loose wire connections.
(esp. Foot/Hand control(s) and
base up limit switch).
Foot / Hand Control (refer to: Perform Foot / Hand control Refer to:
Wireless Controls if applicable) test. Section B - Foot / Hand Controls
Base Up limit switch Perform limit switch test Replace faulty limit switch.
Section B - Limit Switch Test
Base actuator / PC board Perform PC board test. Refer to: Section B -
PC Board Test
No Base Down / Base Up OK Loose wire connections Wire connections Secure loose wire connections.
(esp. Foot/Hand control(s) and
base down limit switch).
Foot / Hand Control (refer to: Perform Foot / Hand control Refer to:
Wireless Controls if applicable) test. Section B - Foot / Hand Controls
Base Down limit switch Perform limit switch test Replace faulty limit switch.
Section B - Limit Switch Test
Base actuator / PC board Perform PC board test. Refer to: Section B -
PC Board Test

© Midmark Corporation 2009 SF-1935 [Revised: 6/23/11] A-3


Troubleshooting Chart
Problem Symptom Probable Cause Check Correction
Base function not operating Base Up / Down noisy, grinding or Actuator threads dry Actuator threads. Clean and lubricate base actuator
properly. - continued squeaking. threads with Lithium grease.
Debris on Base Slides For any debris on base slides. Remove any debris from base
slides.
Note: Do not lubricate base slides.
Base Actuator Run table up and down, if Replace base actuator.
squeaking or grinding continue
check base actuator.
Table Up / Down - Table drifts Foot / Hand Control (refer to: Run table up, then unplug Refer to:
down. Wireless Controls if applicable) power cord. If drifting stops, Section B - Foot / Hand control
perform Foot / Hand control
test.
Base Actuator Run table up, then unplug Replace base actuator.
power cord. If drifting
continues, check base
actuator.
Back function not operating No Back Up or Back Down. Loose wire connections Wire connections Secure loose wire connections.
properly. (esp. Foot/Hand control(s) and
back up/down limit switches.)
Main PC board fuse(s) blown Test F1, F2, F3, & F4 fuses Section D - Wiring Diagrams
Foot / Hand control (refer to: Perform Foot / Hand control Refer to:
Wireless Controls if applicable) test. Section B - Foot / Hand Controls
Back actuator / PC board Perform PC board test. Refer to: Section B -
PC Board Test
Table stops and beeps. Patient weight exceeded 650 lbs If patient weight exceeded the Inform staff that max. patient
650 lb. weight limit. weight is 650 lbs.
No Back Up / Back Down OK. Loose wire connections Wire connections Secure loose wire connections.
(esp. Foot/Hand control(s) and
back up limit switch.)
Foot / Hand Control (refer to: Perform Foot / Hand control Refer to:
Wireless Controls if applicable) test. Section B - Foot / Hand Controls
Back Up Limit Switch Perform limit switch test. Replace faulty limit switch.
Section B - Limit Switch Test
Back actuator / PC board Perform PC board test. Refer to: Section B -
PC Board Test

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-4


Troubleshooting Chart
Problem Symptom Probable Cause Check Correction
Back function not operating No Back Down / Back Up OK. Loose wire connections Wire connections Secure loose wire connections.
properly. - continued (esp. Foot/Hand control(s) and
back down limit switch).
Foot / Hand Control (refer to: Perform Foot / Hand control Refer to:
Wireless Controls if applicable) test. Section B - Foot / Hand Controls
Back Down limit switch Perform limit switch test Replace faulty limit switch.
Section B - Limit Switch Test.
Back actuator / PC board Perform PC board test. Refer to: Section B -
PC Board Test
Back Up / Down - Back drifts Foot / Hand Control (refer to: Run back up, then unplug Refer to:
down. Wireless Controls if applicable) power cord. If drifting stops, Section B - Foot / Hand Controls
perform Foot / Hand control
test.
Back Actuator Run back up, then unplug Replace back actuator
power cord. If drifting
continues, check back
actuator.
Home not functioning No Home / Base Down OK. Loose wire connections. Wire connections Secure loose wire connections.
properly. (esp. Foot/Hand control(s) and
Home limit switches).
Foot / Hand Control (refer to: Perform Foot / Hand control Refer to:
Wireless Controls if applicable) test. Section B - Foot / Hand Controls
Home Limit Switch(es) Perform Limit Switch test. Replace faulty Limit Switch.
Refer to:
Section B - Limit Switch Test
Quick Exam® not functioning Quick Exam® function Quick Exam® function not If Quick Exam® function was Refer to: Quick Exam®
properly. inoperable. programed. properly programed. Operation in User Guide.
Loose wire connections. Wire connections Secure loose wire connections.
(esp. Foot/Hand control(s) and
Travel limit switches).
Foot / Hand Control (refer to: Perform Foot / Hand control Refer to:
Wireless Controls if applicable) test. Section B - Foot / Hand Controls
Travel Limit Switch(es) Perform Limit Switch test. Replace faulty Limit Switch.
Refer to:
Section B - Limit Switch Test

© Midmark Corporation 2009 SF-1935 [Revised: 4/17/13] A-5


Troubleshooting Chart
Problem Symptom Probable Cause Check Correction
Table Receptacle(s) No power at table receptacle(s) Circuit breaker If circuit breaker is tripped. Lift seat section to access circuit
malfunctioning. All other functions work. breaker(s), press to reset.
Loose wire connection Check wire connections Secure loose wire connections.
between table receptacle
and power cord.
Blown isolation transformer Test fuses at IEC inlet. Refer to:
fuse(s) at IEC inlet Section D - Wiring Diagrams
Isolation Transformer Output voltage at isolation trans- Replace faulty isolation
former. transformer.
Table Receptacle Voltage at receptacle. Replace faulty receptacle.

Wireless Controls When any control button is Low battery Refer to: Section B - Wireless Replace batteries. (size: AA)
malfunctioning. pressed, that controller sounds Controls
a single “beep”. (All functions
Faulty PC board in controller Refer to: Section B - Wireless Replace controller PC board.
operate)
Controls
When any control button is Controller not associated to table. Refer to: Section B - Wireless Perform Association Procedure.
pressed, nothing happens. Controls
(No “beeps”, no movement, ect.)
Wireless controls component Refer to: Section B - Wireless Replace malfunctioning
malfunction. Controls component.
Drawer Heater Drawer Heater does not warm up Loose wire connections. Check wire connections Secure loose wire connections.
malfunctioning. Heater Switch does not illuminate. between heater switch and
power cord.
Blown fuse(s) on distribution Test fuses on distribution board. Refer to:
board. Section D - Wiring Diagrams
Drawer heater switch. Test drawer heater switch. Replace drawer heater switch.
Refer to: Section B - Drawer
Heater Switch Test
Drawer Heater does not warm up Loose wire connections. Check wire connections Secure loose wire connections.
Heater Switch-OK (illuminates). between heater switch and
heater plate.
Drawer Heater Plate. Test drawer heater plate. Replace drawer heater plate.
Refer to: Section B - Drawer
Heater Plate Test
Drawer Heater-OK Drawer Heater Switch. Refer to: Section B - Drawer Replace drawer heater switch.
Heater Switch does not illuminate. Heater Switch replacement.

© Midmark Corporation 2009 SF-1935 [Revised: 4/17/13] A-6


Troubleshooting Chart (625-004)
Problem Symptom Probable Cause Check Correction
Digital Scale is Nothing is displayed on the digital Ten second scale time out feature Press Zero button to reactivate Inform operator of the ten second
malfunctioning. display. Table functions OK. activated. display. time out feature.

Faulty PC board in controller Replace PC Board with known Replace controller PC board.
working PC Board.

The measured weight of patient is Scale is not zeroed. Check if operator failed to Refer to: Scale Operation
inaccurate or the display scrolls. zero scale before seating in Users Guide.
patient.
The patient’s feet are making Check that the patient is Refer to: Scale Operation
contact with the floor after weight properly seated. in Users Guide.
button has been pressed.. Check if stop button was Refer to: Scale Operation
depressed while table was in Users Guide.
automatically raising.
Obstruction under or around table Check for any obstruction Remove obstruction.
base. under or around table.
Table not level. Using bubble level supplied Refer to: Leveling Instructions
with table, check each corner in Installation Guide.
of table base for level.
Display on hand control scrolls Hand control not associated with Perform association Refer to: Wireless Controls.
without displaying weight. table. (No table functions). procedure.
Wire connection to external Check wire connections to Refer to: Wireless Controls.
antenna loose or broken. External Antenna.
Digital Display works, but Scale calibration lost. Verify scale is measuring Refer to: Scale Calibration
measured weight stays at zero. accurately using a reference Procedure.
weight of known value.
Loose wire connection. Check wire connections Secure loose wire connection.
from each load cell and scale
control board.
Check wire connections Secure loose wire connection.
between Scale PC Board
and LIN to Scale PC Board.
Check wire connections Secure loose wire connection.
between 625 to USB board
and LIN to Scale board.
Defective Scale PC Board. Replace Scale PC Board with Refer to: Scale / USB PC Boards
known working Scale PC Replacement Instructions.
Board.
Defective LIN to Scale PC Board. Replace LIN to Scale PC Refer to: Scale / USB PC Boards
Board with known working LIN Replacement Instructions.
to Scale PC Board.

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-7


Troubleshooting Chart (625-004)
Problem Symptom Probable Cause Check Correction
No function with hand or When depressing wired or wire- Supply voltage. Facility circuit breaker. Reset circuit breaker..
foot control less hand or foot control, no base
or back up/down function (green Power cord inlet (IEC) Power cord. Ensure inserted securely into inlet.
light D12 on PC Board not illumi-
Export Only: IEC inlet trans- Replace fuses.
nated).
former fuses (6.3A, 250 V
Slo-Blo)
Main PC Board primary fuses (F1 Continuity of fuses (10 A, 250 Replace blown fuse
and F2) V Slo-Blo)

Main PC board transformer fuse Continuity on fuse (160 mA, Replace blown fuse.
(F3) 250V Slo-Blo)
Loose wire connections Wire connections between Check for correct lie voltage at
power intlet and PC board. board

Wire connections on main PC Secure loose wire connection to


Board back/base functions. Ensure har-
ness wires are staggered properly
to prevent any possible harness
disconnection from board or from
actuator.
Overtravel limit switch (only on Perform limit switch test Replace faulty limit switch.
models with serial numbers from (untripped = less than 5 ohm;
V96852 to present) tripped = OL)

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-8


Troubleshooting Chart (625-004)
Problem Symptom Probable Cause Check Correction
No function with hand When depressing wired hand or Hand control Coil cord plugged properly into Properly plug coil cord into receptacle
or foot control foot control, no base or back up/ receptacle on table
down function (green light D12 on Break in coil cord Check for breaks. Replace coil cord as
PC Board illuminated). needed.
Inlet malfunctioning Plug hand control into second port to check
for functionality. Replace defective port.
Cord running from inlet to PC Board Check cord for proper installation. Replace
disconnected or broken. suspect cord with known working cord or
perform continuity check on cord.
Refer to:
Section B - Foot / Hand Controls
Main PC Board is malfunctioning Refer to: Section B -PC Board Test

Main PC Board base and Continuity on fuse (6.3A, 250 V If board is Version 1.5.X or below, then re-
back motor fuse (F4) Slo-Blo) place main PC Board.
If board is Version 1.6.X or above, then
replace fuse.
Replace all 625 boards if fuse blown with se-
rial number before V1309272. Replace fuse
only with serial number after V1365847.
Wire connections on main PC Board Secure loose wire connection to back/
base functions. Ensure harness wires are
staggered properly to prevent any possible
harness disconnection from board or from
actuator.

No function with hand When depressing wireless hand Wireless hand and/or foot Batteries (controller will sound a Replace batteries (size AA).
or foot control. Unit or foot control, no base or back control single beep if batteries are low).
gives audible PC up/down function (green light D12
Association of controller to table Perform Association Procedure. If associa-
Board tones when un- on PC Board illuminated).
tion does not work, then refer to:
plugged and plugged Wireless Controls - Isolating a Malfunction
back in.
Main PC Board base and Continuity on fuse (6.3A, 250 V If board is Version 1.5.X or below, then re-
back motor fuse (F4) Slo-Blo) place main PC Board.
If board is Version 1.6.X or above, then
replace fuse.
Wire connections on main PC Board Secure loose wire connection(s).
Receptacle Power cord inlet (IEC) Continuity of isolation transformer Replace fuses.
fuses (6.3A, 250 V Slo-Blo).

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-9


Power to the Table (120 V models)
No functions will operate....................A-3
This illustration shows only the components that affect ALL TABLE FUNCTIONS.
A detailed description of current flow also appears below.

Facility Supply Voltage


With the chair’s power cord properly connected, facility supply voltage
(115 VAC) is supplied thru the cord to the main PC board.

Note
The two external fuses at the power cord (IEC) inlet do not affect IEC
Base or Back functions. These fuses affect the table receptacle only. Inlet

Overtravel Limit Switch (Later models only)


Disrupts power to main PC board if table is over extended.
Overtravel
Main PC Board Grn/Yel
COM
NC Limit Switch
When proper voltage is supplied to the PC board, the power indicator
light (on PC board) is illuminated. Circuitry on the PC board provides Line Filter
the required voltage to power all of the table’s components: hand / foot
controls, actuators, limit switches. Limit Switch Harness
[Base / Back functions will not operate if
harness connection is loose.]

Fuses

Blue

Brn
COM

The fuses on the PC board provide the following protection: Home NC

Limit Switches COM Blk


NC Blk J13 J12 J15 J14
F1: Protects PC board (supply voltage). Blk
Back Dn
F2: Protects PC board (supply voltage).
COM
J11
Limit Switch NC
Blk J10
F3: Protects PC board transformer. Back Up COM
Wht

F4: Protects Base and Back motors. Limit Switch NC


Wht
F3 F2 F1
Blk
Base Dn COM
Limit Switch NC
Blk
Wht
Base Up COM J9 Main PC
Limit Switch NC
Wht Board
J7 F4

Power Indicator Light


If light is OFF, there is no power to PC board. J3 J2 J1
J6 J5 J4
If light is ON, proper voltage is present at PC board.

MA8117i

Models: 625 (-001 & -003)


Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 2/23/17] A-10 Always verify model & serial number
Power to the Table (230 V models)
No functions will operate....................A-3
This illustration shows only the components that affect ALL TABLE FUNCTIONS.
A detailed description of current flow also appears below.

Facility Supply Voltage


With the chair’s power cord properly connected, facility
Fuse IEC Fuse
supply voltage (230 VAC) is supplied thru the two power Holder Inlet Holder
Main PCB
cord inlet fuses to the export transformer. Export
Brn

Transformer
Export Transformer Brn

The export transformer reduces the line voltage and Blue A Blue
supplies 120V to the main PC board. Grn/Yel

Overtravel Limit Switch (Later models only)


Overtravel
Disrupts power to main PC board if table is over extended.
COM
NC Limit Switch

Main PC Board
When proper voltage is supplied to the PC board, Limit Switch Harness
[Base / Back functions will not operate if
the power indicator light (on PC board) is illuminated. Cir- harness connection is loose.]

Blue
Brn
cuitry on the PC board provides the required voltage COM
to power all of the table’s components: actuators, Home NC

hand / foot controls, limit switches, etc. Limit Switches COM


Blk
NC Blk J13 J12 J15 J14
Back Dn Blk
Fuses Limit Switch
COM
NC J10
J11
Blk
The fuses on the PC board provide the following protection: Back Up COM Wht
Limit Switch NC
Wht F3 F2 F1
F1: Protects PC board (supply voltage). Base Dn
Limit Switch
COM Blk
NC
F2: Protects PC board (supply voltage). Blk
Wht
Base Up COM J9 Main PC
F3: Protects PC board transformer. Limit Switch NC
Wht Board
F4: Protects Base and Back motors. J7 F4

Power Indicator Light


If light is OFF, there is no power to PC board. J3 J2 J1
J6 J5 J4
If light is ON, proper voltage is present at PC board.

MA8119i

Models: 625 (-005 & -006)


Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 2/23/17] A-11 Always verify model & serial number
Base UP / DOWN Function
No Base Up or Base Down................A-3
This illustration shows only the components that affect the Base UP / DOWN function. . Table Stops and Beeps......................A-3
Refer to the following page for a detailed description of current flow during this function. No Base Up. Base Down - OK...........A-3
No Base Down. Base Up - OK...........A-3
115 VAC Noisy Operation..................................A-4
Base Up / Down -Table Drifts.............A-4

Note: 230 VAC models only


Line voltage (230 VAC) is supplied thru two fuses (located by
the IEC inlet) to a transformer that reduces the line voltage to
115 VAC before it reaches the PC Board.

14 VAC

Base Down Limit Switch


[Base Down will not operate if switch is tripped]

Base Up Limit Switch


[Base Up will not operate if switch is tripped]

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-12 Always verify model & serial number
Base UP / DOWN Function Base Up Operation Base Down Operation

Note Note Note


On 230 VAC models, line voltage (230VAC) is If the Base Up limit switch is tripped (open), If the Base Down limit switch is tripped (open),
supplied thru two fuses (located by IEC inlet) the Base Up function will not operate. the Base Down function will not operate.
to a transformer that reduces the line voltage
to 115 VAC before it reaches the PC board.

When the Base Up function is activated, When the Base Down function is activated,
Power to Foot / Hand Control
current (14 VAC) flows thru the UP function current (14 VAC) flows thru the DOWN function
115 VAC is supplied thru the two primary fuses
foot switch (N.O.), then back to the PC board. foot switch (N.O.), then back to the PC board.
(F1 & F2) to the transformer on the PC board.
Circuitry on the PC Board energizes the base Circuitry on the PC Board energizes the base
[F3 fuse protects the transformer].
motor. [F4 fuse protects the base motor]. motor. [F4 fuse protects the base motor].
The transformer reduces the line voltage and
When the base function is energized, current When the base function is energized, current
supplies 14 VAC to the foot / hand control.
flows to the actuator motor. When current flows to the actuator motor. When current
flows to the actuator, the actuator motor flows to the actuator, the actuator motor
runs and raises the table. runs and lowers the table.

Actuator motor runs until: Actuator motor runs until:


1. Foot / Hand control button is released. 1. Foot / hand control button is released.
2. Base Up limit switch is tripped. 2. Base Down limit switch is tripped.
3. Fuse(s) opens (blows). 3. Fuse(s) opens (blows).
[Primary, Base Motor, Transformer fuse] [Primary, Base Motor, Transformer fuse]
4. Over current condition is detected. 4. Over current condition is detected.

Equipment Alert
If an over current condition* is detected the PC board will “beep” and all power functions
will be disabled. Release the function button, remove load, then retry function.
*(weight limit exceeded, mechanical binding, etc),

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-13 Always verify model & serial number
Back UP / DOWN Function No Back Up or Back Down.................A-4
This illustration shows only the components that affect the Back UP / DOWN function. . Table Stops and Beeps......................A-4
Refer to the following page for a detailed description of current flow during this function. No Back Up. Back Down - OK............A-4
No Back Down. Back Up - OK............A-5
115 VAC Back Up / Down - Back Drifts.............A-5

Note: 230 VAC models only


Line voltage (230 VAC) is supplied thru two fuses (located by
the IEC inlet) to a transformer that reduces the line voltage to
115 VAC before it reaches the PC Board.

14 VAC

Back Down Limit Switch


[Back Down will not operate if switch is tripped]

Back Up Limit Switch


[Back Up will not operate if switch is tripped]

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-14 Always verify model & serial number
Back UP / DOWN Function Back Up Operation Back Down Operation

Note Note Note


On 230 VAC models, line voltage (230VAC) is If the Back Up limit switch is tripped (open), If the Back Down limit switch is tripped (open),
supplied thru two fuses (located by IEC inlet) the Back Up function will not operate. the Back Down function will not operate.
to a transformer that reduces the line voltage
to 115 VAC before it reaches the PC board.

When the Back Up function is activated, When the Back Down function is activated,
Power to Foot / Hand Control
current (14 VAC) flows thru the UP function current (14 VAC) flows thru the DOWN function
115 VAC is supplied thru the two primary fuses
foot switch (N.O.), then back to the PC board. foot switch (N.O.), then back to the PC board.
(F1 & F2) to the transformer on the PC board.
Circuitry on the PC Board energizes the back Circuitry on the PC Board energizes the back
[F3 fuse protects the transformer].
motor. [F4 fuse protects the back motor]. motor. [F4 fuse protects the back motor].
The transformer reduces the line voltage and
When the back function is energized, current When the back function is energized, current
supplies 14 VAC to the foot / hand control.
flows to the actuator motor. When current flows to the actuator motor. When current
flows to the actuator, the actuator motor flows to the actuator, the actuator motor
runs and raises the back section. runs and lowers the back section.

Actuator motor runs until: Actuator motor runs until:


1. Foot / Hand control button is released. 1. Foot / Hand control button is released.
2. Back Up limit switch is tripped. 2. Back Down limit switch is tripped.
3. Fuse(s) opens (blows). 3. Fuse(s) opens (blows).
[Primary, Back Motor, Transformer fuse] [Primary, Back Motor, Transformer fuse]
4. Over current condition is detected. 4. Over current condition is detected.

Equipment Alert
If an over current condition* is detected the PC board will “beep” and all power functions
will be disabled. Release the function button, remove load, then retry function.
*(weight limit exceeded, mechanical binding, etc).

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-15 Always verify model & serial number
Home Function
No Home. Base Down - OK...............A-5
This illustration shows only the components that affect the Home function. .
Refer to the following page for a detailed description of current flow during this function.

Note: 230 VAC models only


Line voltage (230 VAC) is supplied thru two fuses (located by
the IEC inlet) to a transformer that reduces the line voltage to
115 VAC before it reaches the PC Board.

Home Home Home


Start Switch Stop Switch Start Switch

Home Limit Switches Home


[Home will not operate if switches are tripped]
Stop Switch

Base Down Limit Switch


[Home will not operate if switch is tripped]

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-16 Always verify model & serial number
Home Function Home Operation

Note
Note If either of the Home limit switches or the
On 230 VAC models, line voltage (230VAC) is Base Down limit switch are tripped (open), the
supplied thru two fuses (located by IEC inlet) Home function will not operate.
to a transformer that reduces the line voltage
to 115 VAC before it reaches the PC board.
When the Home function is activated,
current (14VAC) flows thru the Home foot switch,
Power to Foot / Hand Control (N.O.), then back to the PC Board.
115 VAC is supplied thru the two primary fuses Circuitry on the PC Board energizes the base
(F1 & F2) to the transformer on the PC board. motor. [F4 fuse protects the base motor].
[F3 fuse protects the transformer].
When the Home function is energized,
current flows to the base motor. When
The transformer reduces the line voltage and
current flows to the actuator, the actuator motor
supplies 14 VAC to the foot / hand control.
runs and lowers the table.

When voltage flows to the actuator, the actuator


motor runs lowering the table.

Actuator motor runs until:


1. Stop button is depressed.
2. Home limit switch(es) is tripped.
3. Base Down limit switch is tripped.
4 Fuse(s) opens (blows).
[Primary, Base Motor, Transformer fuse]
5. Over current condition is detected.

Equipment Alert
If an over current condition* is detected the PC board will “beep” and all power functions will be
disabled. Release the function button, remove load, then retry function.
*(weight limit exceeded, mechanical binding, etc).

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-17 Always verify model & serial number
Drawer Heater System (optional)
Drawer Heater Does Not Warm up:
This illustration shows only the components that affect the Drawer Heater System. Switch Does Not Illuminate................A-5
A detailed description of current flow during this function also appears below. Switch Illuminates...............................A-5
Drawer Heater Works - Properly
Switch Does Not Illuminate................A-5

Note: 230 VAC models only


Line voltage (230 VAC) is supplied thru two fuses (located by
the IEC inlet) to a transformer that reduces the line voltage to
115 VAC before it reaches the PC Board.

115 VAC from


Main PC Board

Drawer Heater Operation


115 VAC is supplied directly to the Drawer Heater
switch thru two fuses on the Distribution Board.
[Voltage bypasses the Main PC Board].

When the drawer heater switch is turned ON,


current flows to the heater plate. When the heater
plate is energized, it warms the contents of the Distribution Board Fuses
[Drawer Heater only]
drawer.

Heater Plate

115 VAC

Heater Switch

115 VAC

Models: 625 (-001 /-003 /-004 /-006


Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-18 Always verify model & serial number
Table Receptacle (optional)
No Power At Table Receptacle...........A-6
This illustration shows only the components that affect the table receptacle.
A detailed description of current flow during this function also appears below.

Table Receptacle
115 VAC is supplied directly to the
receptacle thru two inlet fuses, isolation
transformer, and circuit breakers.
[Voltage bypasses the Main PC Board].

Models: 625 (-001 /-003)


Serial Numbers: V2200 thru V1149713

© Midmark Corporation 2009 SF-1935 [Revised: 11/15/11] A-19 Always verify model & serial number
Table Receptacles (optional)
No Power At Table Receptacle...........A-6
This illustration shows only the components that affect the table receptacles.
A detailed description of current flow during this function also appears below.

Table Receptacles
115 VAC is supplied directly to the
receptacles thru two inlet fuses, isolation
transformer, and circuit breakers.
[Voltage bypasses the Main PC Board].

Models: 625 (-001 /-003)


Serial Numbers: V1149714 thru Present

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-19.1 Always verify model & serial number
Table Receptacles
No Power At Table Receptacle...........A-6
This illustration shows only the components that affect the table receptacles.
A detailed description of current flow during this function also appears below.

Table Receptacles
115 VAC is supplied directly to the
receptacles thru two inlet fuses, isolation
transformer, and circuit breakers.
[Voltage bypasses the Main PC Board].

Models: 625 (-004)


Serial Numbers: All

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-19.2 Always verify model & serial number
Scale Information Flow
This illustration shows only the components that affect the flow of information for weight.
A detailed description of information flow during this function also appears below.

Hand Control
When activated, the hand control transmits the
request through the antenna to the Base Station.

Base Station
When the Base Station receives the request, the “TX” light will flash momentarily.
The Base Station then sends the request to the “LIN to Scale” board.

LIN to Scale Board


The LIN to Scale Board receives request from Base Station.

The LIN to Scale Board then asks for weight from the Scale Board.

Note: These steps happen even if the


Scale Board Hand Control has not requested a weight.
When the Scale Board receives request from LIN to Scale Board, The LIN to Scale Board is constantly
the Red light will flash momentarily. monitoring weight.
The Scale Board reads the weight from the load cells.
The Amber light will flash momentarily when the information is
sent back to the LIN to Scale Board.
Note: Load cells are plugged into the Scale Board.

The LIN to Scale Board sends information back to Base Station.


The Base Station sends information back to Hand Control.
The Hand Control displays information.

Models: 625 (-004)


Serial Numbers: All

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] A-20 Always verify model & serial number
Testing Foot / Hand Controls..........................B-2
Wireless Controls...............................B-4
Base Function Components...............B-7
Base Limit Switch Test.......................B-8
Base PC Board Test...........................B-9

Section B Base Actuator


Motor Resistance Test......................B-10
Back Function Components............. B-11
Back Limit Switch.............................B-12
Back PC Board Test.........................B-13
Back Actuator
Motor Resistance Test......................B-14
Drawer Heater System.....................B-15
Table Calibration Procedure.............B-20
Scale Calibration Procedure.............B-21

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-1


Foot / Hand Controls
Foot / Hand Control Test....................B-3
Access Procedures............................C-1
Isolating a Malfunction Wiring Diagrams.................................D-1

To isolate a malfunction, try activating the inoperable function(s) from hand control and foot control.

If function(s) are inoperable from the foot / hand control...


A) Plug the foot / hand control into the other connection port.
B) Secure Inlet harness connection(s) on main PC board at J12 & J13.

If function(s) are operable:


C) Replace Inlet harness at malfunctioning connection port.

If function(s) still inoperable:


D) Perform the Foot / Hand Control Test.

Connection Connection
Port Port

Hand
Control
Foot
Control

J13 J12 J15 J14

J11
J10

F3 F2 F1
Note
Refer to “Wireless Controls” if applicable.

J9 Main PC
Board
J7 F4

J6 J3 J2 J1
J5 J4

MA8118i

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-2 Always verify model & serial number
Foot / Hand Controls - continued
Access Procedures............................C-1
Wiring Diagrams.................................D-1
Foot / Hand Control Test

Note
Refer to “Wireless Controls” if applicable.
Step 1: Disconnect wire harness from J12 and J13 on main PC board.
*Note: J12 and J13 are interchangeable. Either harness could be plugged
into either J12 or J13.

3
J1
With switch “untripped”...
Meter Reading Required Action

Red (Base Down)


Any resistance Replace control /
reading switch membrane.

Black (com.)
Step 2: Set meter to Ω. OL Switch is good.
Check fuses.

Step 3: Place one meter probe on black wire


(common) of foot / hand control harness.

Place other meter probe on the wire that


corresponds to the switch to be tested.

Switch...............Wire Color
With switch “tripped”...
Base Up...............white
Base Down..........red Meter Reading Required Action
Back Up...............green Less than or = to Switch is good.
Back Down..........orange 5 ohms Check fuses.
Home...................blue
OL Replace control /
Stop.....................brown
switch membrane.
Quick Exam™.....yellow
MA8108i

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-3 Always verify model & serial number
Wireless Controls
Association Procedure.......................B-6

Isolating a Malfunction

Antenna Equipment Alert


Cable Antenna connection is very delicate.
Testing Do not kink wire or apply stress to connection.

• When any control button is pressed, MA771100i


that controller sounds a single “beep”. LEDs Normal Status
(All functions operate)

Cause: 1. Low battery


TX (OFF - except during Association Procedure)

2. Faulty PC board in controller RX (ON - except during Association Procedure)


Solution: 1. Replace batteries (size: AA)
2. Replace controller
COM (ON)

PWR (ON)

• When any control button is pressed, nothing happens.


(No “beeps”, no movement, etc.) Base Station PC Board

Check: 1. Perform Association Procedure


2. LEDs on base station PC board
(lights should quickly flash between “RX” and “TX”
when function is activated.)

If LED status is normal (see illustration):


• Check LEDs on hand and/or foot control (see next page).
Modular
Harness
If no LEDs are illuminated, or if no quick flash between
“RX” and “TX” lights when function is activated, check the following:
• Modular harness (loose connections, damage, etc.)
• Power supply to table
• Replace base station PC board if necessary. MA8052i

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 2/23/17] B-4 Always verify model & serial number
Wireless Controls - continued

Testing - continued

• When any control button is pressed, nothing happens. - continued


(No “beeps”, no movement, etc.)

(see previous page for Checks 1 & 2)


Check: 3. LEDs on hand / foot control PC board
(press any button on hand / foot control)

If LEDs function properly (see illustration):


• Check antenna connection. (see parts list for detail)
• Replace antenna if necessary.

If LEDs do not function properly:


• Replace batteries in controller (size: AA)
• Replace controller touch membrane
• Replace controller PC board if necessary.
LEDs LEDs
Normal Status

D2 (ON - only when a button on controller is pressed) D10


D3 (OFF - except during Association Procedure) D11

Hand / Foot Control PC Board Hand Control / w Scale PC Board


(refer to parts list for location) (625-004 Only)
(refer to parts list for location)

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 2/21/12] B-5 Always verify model & serial number
Wireless Controls - continued

Association Procedure

The wireless controllers will only function with the chair they have been “associated”
with. This prevents unwanted movement in locations with more than one chair. Association Procedure:
A) Disconnect power to the table for at least 3 seconds.
Each wireless controller’s association is preset at the factory. This procedure is only
required in the event of a malfunction, if a new controller is purchased, or if you are B*) Reconnect power, wait for the power up beep sequence from
attempting to use the controller with a chair other then the one it was originally the table, then press & hold the Stop & Home buttons on the
associated with. wireless controller until you hear three “beeps”.
NOTE: The base station PC board only stores the last two associated controllers. To ensure proper * Note: Step B must be performed within 10 seconds of the power up beep
operation, Midmark recommends reassociating both controllers (hand & foot), even if only one new sequence from the table.
controller is purchased.

C) Check for proper operation of wireless controller.

D) If associating a hand and foot control, repeat procedure using


the other wireless controller.

Home Stop

Home

Stop

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 6/19/12] B-6 Always verify model & serial number
Base Function Components
Wiring Diagrams.................................D-1

Isolating a Malfunction Base Actuator /


Motor Replacement........... 003-2072-00
This illustration shows the base limit switches, the serviceable components of the base actuator, and the
fuses on the main PC board. Use the table below to isolate the malfunction.

Problem Required Action


Motor does not run. Check main PC board fuses
(F1, F2, F3, & F4)
Perform Limit Switch Test.
Base function operates, but makes Clean / lube actuator threads.
grinding / squeaking noises Remove any debris from base slides.
Note: Do not lubricate base slides.
Replace actuator if necessary*. J13 J12 J15 J14

Motor runs, but table does not move. Inspect / replace motor coupler. J10 J11

F3 F2 F1
Home Function
Limit Switches
J9 Main PC
Base Down Board
Base Up Actuator Threads
Limit Switch J7 F4
Limit Switch

J6 J3 J2 J1
J5 J4

MA8070i

Fuses
Actuator Motor
(F1, F2, F3, F4)

Motor
Coupler

MA8121i

Models: 625 * Replacement instructions are provided with the


part. They are also available on midmark.com,
Serial Numbers: all or by clicking on the blue link.

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-7 Always verify model & serial number
Base Function Components - continued

Limit Switch Test

Step 1: Disconnect wires from switch.

Step 2: Place meter probes on


COM and NC terminals.
Note: Check switch “tripped” and “untripped”.

With switch “untripped”... With switch “tripped”...


Meter Reading Required Action Meter Reading Required Action
OL Replace limit switch Limit switch - OK
OL
Perform PC Board Test
Limit switch - OK
less than 5 ohms less than 5 ohms Replace limit switch
Perform PC Board Test

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-8 Always verify model & serial number
Base Function Components - continued
Calibration Procedure.......................B-20

PC Board Test PC Board Replacement....... 003-2073-00

Equipment Alert
The back limit switches will not stop movement during this test.
Do not run past max. / min. positions.

Step 1: Move Back section so that it Did BASE move Up and Down in Step 4? Required Action
is approx. halfway between its YES Replace main PC board
maximum & minimum positions. NO Refer to: Base Actuator Motor Resistance Test

Step 4: Using the hand / foot control:


A) Press & hold BACK UP briefly.
B) Press & hold BACK DOWN briefly.

Equipment Alert
Failure to move both connections may
result in damage to PC Board.

Step 3: Move base actuator connection to J4


and base sensor connection to J7.
J7

J4

Step 2: Tag, then disconnect back & base actuator


wire connections at J4 & J5. Tag then disconnect
back & base sensor wire connections at J6 & J7.
Step 5: Calibrate PC Board.

Models: 625 * Replacement instructions are provided with the


part. They are also available on midmark.com,
Serial Numbers: all or by clicking on the blue link.

© Midmark Corporation 2009 SF-1935 [Revised: 8/14/14] B-9 Always verify model & serial number
Base Actuator Motor Resistance Test
Calibration Procedure.......................B-20
Isolating a Malfunction
Base Actuator /
Motor Replacement........... 003-2072-00

Equipment Alert
Anytime actuator wires are disconnected,
the PC Board must be calibrated.

Step 1:
A) Disconnect power to the table.
B) Disconnect base actuator from J6 on PC Board.
C) Set multimeter to Ohms to check resistance.

J6

Step 2:
A) Place the Red meter probe on the Black wire,
and the Black meter probe on the Blue wire.

B) Place the Red meter probe on the Black wire,


and the Black meter probe on the Green wire.

C) Place the Red meter probe on the Black wire,


and the Black meter probe on the White wire.

If any of the readings are less than two Mega Ohms


after three seconds, the motor is defective.

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 7/12/18] B-10 Always verify model & serial number
Back Function Components
Wiring Diagrams.................................D-1

Isolating a Malfunction Back Actuator /


Motor Replacement........... 003-2071-00
This illustration shows the back limit switches, the serviceable components of the back actuator, and the
fuses on the main PC board. Use the table below to isolate the malfunction.

Problem Required Action


Motor does not run. Check main PC board fuses
(F1, F2, F3, & F4)
Perform Limit Switch Test
Back function operates, but makes Clean / lube actuator threads.
grinding / squeaking noises Replace actuator if necessary*.
Motor runs, but table does not move. Inspect / replace motor coupler.

Motor Coupler
J13 J12 J15 J14
Actuator Motor
J10 J11

F3 F2 F1

J9 Main PC
Board
J7 F4

J3 J2 J1
Back Down
J6 J5 J4 Limit Switch

MA8070i
MA8120i

Fuses
(F1, F2, F3, F4) Back Up
Limit Switch

Models: 625 * Replacement instructions are provided with the


part. They are also available on midmark.com,
Serial Numbers: all or by clicking on the blue link.

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-11 Always verify model & serial number
Back Function Components - continued

Limit Switch Test

Step 1: Disconnect wires from switch.

Step 2: Place meter probes on


COM and NC terminals.
Note: Check switch “tripped” and “untripped”.

With switch “untripped”... With switch “tripped”...


Meter Reading Required Action Meter Reading Required Action
OL Replace limit switch Limit switch - OK
OL
Perform PC Board Test.
less than 5 ohms Limit switch - OK
Perform PC Board Test. less than 5 ohms Replace limit switch

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-12 Always verify model & serial number
Back Function Components - continued
Calibration Procedure.......................B-20

PC Board Test PC Board Replacement....... 003-2073-00

Equipment Alert
The base limit switches will not stop movement during this test.
Do not run past max. / min. positions.

Step 1: Move Base section so that it Did BACK move Up and Down in Step 4? Required Action
is approx. halfway between its YES Replace main PC board*
maximum & minimum positions. NO Refer to: Back Actuator Motor Resistance Test

Step 4: Using the hand / foot control:


A) Press & hold BASE UP briefly.
B) Press & hold BASE DOWN briefly.

Equipment Alert
Failure to move both connections may
result in damage to PC Board.

Step 3: Move back actuator connection to J5


and back sensor connection to J6.
J6

J5

Step 2: Tag, then disconnect back & base actuator


wire connections at J4 & J5. Tag then disconnect
back & base sensor wire connections at J6 & J7.
Step 5: Calibrate PC Board.

Models: 625 * Replacement instructions are provided with the


part. They are also available on midmark.com,
Serial Numbers: all or by clicking on the blue link.

© Midmark Corporation 2009 SF-1935 [Revised: 8/14/14] B-13 Always verify model & serial number
Back Actuator Motor Resistance Test
Calibration Procedure.......................B-20
Isolating a Malfunction
Back Actuator /
Motor Replacement........... 003-2071-00

Equipment Alert
Anytime actuator wires are disconnected,
the PC Board must be calibrated.

Step 1:
A) Disconnect power to the table.
B) Disconnect back actuator from J7 on PC Board.
C) Set multimeter to Ohms to check resistance.

J7

Step 2:
A) Place the Red meter probe on the Black wire,
and the Black meter probe on the Blue wire.

B) Place the Red meter probe on the Black wire,


and the Black meter probe on the Green wire.

C) Place the Red meter probe on the Black wire,


and the Black meter probe on the White wire.

If any of the readings are less than two Mega Ohms


after three seconds, the motor is defective.

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-14 Always verify model & serial number
Drawer Heater System
Heater Switch Test...........................B-16
Heater Switch Replacement.............B-17
Distribution Board Test Heater Plate Test..............................B-18
Heater Plate Replacement...............B-19
Chair Arm Brace Removal..................C-8
Wiring Diagrams.................................D-1

Caution
Unplug power cord before Step 3: Set meter to 200 VAC.
removing stirrup guide.

Step 1: Remove stirrup guide.


Step 4: Place meter probes on
Note: If necessary, remove chair arm incoming power wires.
brace (see Access Procedures).

Step 5: Plug in power cord,


then depress switch ON.

Step 2: Disconnect two incoming


power wires.

Meter Reading Required Action


approx.
Perform Heater Switch Test.
115 VAC
Check connections &
0 VAC
distribution board fuses.

Models: 625 (-001 /-003 /-004 /-006)


Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 6/15/12] B-15 Always verify model & serial number
Drawer Heater System - continued
Heater Switch Replacement.............B-17
Heater Plate Test..............................B-18
Heater Switch Test Chair Arm Brace Removal..................C-8

Caution
Unplug power cord before
removing stirrup guide. Step 3: Set meter to 200 VAC.

Step 1: Remove stirrup guide.


Step 4: Place meter probes
Note: If necessary, remove chair arm on two white leads
brace (see Access Procedures). from switch.

Step 7: Replace stirrup guide. Step 2: Disconnect two white


leads from drawer heater
switch to the drawer heater.

Step 5: Plug in power cord,


then depress switch ON.

Caution
Meter Reading Required Action Unplug power cord before
0 VAC Replace switch. re-connecting electrical leads.

approx. Switch - OK
115 VAC Perform Heater Plate test
Step 6: Re-connect electrical leads.

Models: 625 (-001 /-003 /-004 /-006)


Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 2/10/14] B-16 Always verify model & serial number
Drawer Heater System - continued
Access Procedures
Chair Arm Brace Removal..................C-8
Heater Switch Replacement

(This procedure also applies to the upholstery heater switch)

Caution
Unplug power cord before
removing stirrup guide.
Step 2: Disconnect switch harness.

Step 1: Remove stirrup guide.


Note: If necessary, remove chair arm
brace (see Access Procedures).

Connect to
Distribution
Board
J5 J6

Step 4: Remove switch


assembly.

Step 3: Remove switch housing.

Step 5:
A) Install switch assembly.
B) Install switch housing.
C) Re-connect switch harness.
D) Replace stirrup guide.
Models: 625 (-001 /-003 /-004 /-006)
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 3/10/14] B-17 Always verify model & serial number
Drawer Heater System - continued
Access Procedures
Chair Arm Brace Removal..................C-8
Heater Plate Test

Caution
Unplug power cord before
removing stirrup guide.
Step 3: Set meter to 2K Ω.

Step 1: Remove stirrup guide.


Note: If necessary, remove chair arm
brace (see Access Procedures).
Step 4: Place meter probes on
terminals of heater plate wires.

Step 6: Replace stirrup guide.

Step 5: Re-connect electrical


leads.

Step 2: Disconnect two white


leads from drawer
heater switch to the
drawer heater. Meter Reading Required Action
approx.
Heater plate - OK.
360 ohms
0 ohms Replace Heater Plate.

Models: 625 (-001 /-003 /-004 /-006)


Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 6/15/12] B-18 Always verify model & serial number
Drawer Heater System - continued
Access Procedures
Chair Arm Brace Removal..................C-8
Heater Plate Replacement

Caution
Unplug power cord before
removing stirrup guide.

Step 1: Remove stirrup guide.


Note: If necessary, remove chair arm
brace (see Access Procedures).
Step 2: Disconnect two white
wires from drawer
Step 8: Replace stirrup guide. heater switch to the
heater plate.
Step 6: Route heater plate wires
thru upperwrap. Re-connect wires.

Step 3: Push LH slide lever UP and RH lever DOWN


Remove drawer.
Note: Slide levers are clear.

Step 4: Remove heater plate.

Step 7: Install drawer.


Step 5: Install heater plate.

Models: 625 (-001 /-003 /-004 /-006)


Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 6/15/12] B-19 Always verify model & serial number
Calibration Procedure

Note
When S2 switch is pressed you will hear a steady
pattern of “beeps”. When calibration is completed
you will hear three short beeps.

To calibrate...
A) Unplug and re-plug power cord.
B) Wait for the power up beep sequence from table,
then within two seconds, press S2 switch.
C) Run Base all the way down.
D) Run Base all the way up.
E) Run Back all the way down.
F) Run Back all the way up.

J6

S2
Switch

© Midmark Corporation 2009 SF-1935 [Revised: 7/6/12] B-20


Scale Calibration Procedure (625-004 only)
Note
If abort message is displayed, unplug table for 60 seconds.
Remove and reinstall hand control batteries, then repeat step 1.

Step 2:
Step 1: A) When “CLrAL” is displayed, firmly press and release the plus button.
A) Remove any load from the patient surface. B) After count down, “Ad50” will be displayed.
B) Press and hold the Weight and Plus buttons until “Pause”
Note: If something other then “Ad50” is displayed, press the weight button until
is displayed for at least one second, then release buttons. “Ad50” is displayed. (Early units may display “Add 50”).

Weight Plus Plus


Button Button Button

Step 3: Step 4:
A) Place fifty pound calibrated weight on patient surface. A) Wait until display clears, then firmly press and release the weight button.
B) Firmly press and release the plus button. Note:“50 lbs” should be displayed.
Note: After count down “donE” will be displayed. B) Remove calibrated weight.

Plus Weight
Button Button
© Midmark Corporation 2009 SF-1935 [Revised: 10/15/13] B-21
Hand Control PC Board LED Test (625-004 only)

When “Zero” or “Weight” is pressed, the Hand Control transmits the request through the antenna to the Base Station.
Hand Control D10 flashes green.

LEDs Normal Status

D9 ON (green) during Hand Control power up

D10 ON (green) flashes when button on Hand Control is pressed

D11 ON (green) during Hand Control power up


(ON (green) flashes when information is received from Base Station)

D12 Not Used

Zero Weight

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-22


Wireless Base Station PC Board LED Test (625-004 only)

When the Wireless Base Station PC board receives request through the antenna, the request is forwarded to the Main PC board through
the LIN to VMS PC board. Main PC board D9 flashes green (if “Zero” button is pressed, D10 also illuminates red).

The Wireless Base Station sends acknowledgement of request to Hand Control PC board through the antenna.
Wireless Base Station PC board D1 (TX) flashes green, Hand Control PC board D11 flashes green.

LEDs Normal Status

D1 (TX) - ON (green) flashes when sending information to wireless Hand Control

D2 (RX) ON (green) except when sending information to wireless Hand Control

D3 (COM) ON (green) when connected to network

D4 (PWR) ON (green) when board has power

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-23


LIN to Scale PC Board LED Test (625-004 only)

The LIN to Scale PC board request weight data from the Scale PC board.
The LIN to Scale PC board sends weight data to the Main PC board through the 625 to USB PC board.

(D1)

(D2)

(D3)

(D4)

LEDs Normal Status

D1 ON (green) flashes when board has power

D2 Not Used

D3 Not Used
LIN to Scale
PC Board
D4 ON (green) communicating (always on)

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-24


Main PC Board LED Test (625-004 only)

The Main PC board forwards request to the LIN to Scale PC board through the 625 to USB PC board.
The Main PC board sends weight data to the Wireless Base Station PC board through the LIN to VMS PC board.

(D9) (D10) (D11) (D12)

LEDs Normal Status

D9 ON (green) flashes while any button is pressed


(With wireless control only)

D10 ON (red) when “Zero” is pressed


(Remains ON until another button is pressed)

D11 Not Used

D12 ON (green) when board has power

MA10000i

© Midmark Corporation 2009 SF-1935 [Revised: 6/617] B-25


625 to USB PC Board Test (625-004 only)

The 625 to USB PC board relays request from the Main PC board to the LIN to Scale PC board.

Out To
LIN to Scale
PC Board

In From
Main PC Board

625 to USB
PC Board

MA100001

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-26


LIN to VMS PC Board LED Test (625-004 only)

Relays information from the Wireless Base Station PC Board to Main PC Board.

LEDs Normal Status

D4 ON (green) when connected to network

D3 ON (amber) when table operated from VMS

In From D2 ON (red) when table operated from VMS


Wireless
Base Station
D1 ON (green) flashes when board has power

(D4)

(D3)

(D2)

(D1)

Out To
Main PC Board

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-27


Scale PC Board LED Test (625-004 only)

Scale PC board receives request from LIN to Scale PC board - DS1 flashes red.
Scale PC board gathers weight data from load cells.
Weight data from load cells is forwarded to the LIN to Scale PC board - DS2 flashes amber.

Load Cell
Connection

(DS1)
Load Cell
Connection

(DS2)

Load Cell
Connection
(DS3)

Load Cell
Connection

LEDs Normal Status

DS1 ON (red) flashes when receiving request from LIN to Scale board

DS2 ON (amber) flashes when sending information to LIN to Scale board


(Occurs approximately two times per second)

DS3 Flash (green) when board has power


Scale
PC Board

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] B-28


Access Cladding Removal / Installation..........C-2
Table Shrouds
Removal / Installation.........................C-3

Procedures Raising / Lowering

Section C Table Manually...................................C-6


Upholstery
Removal / Installation.........................C-7
Chair Arms / Brace
Removal / Installation.........................C-8
Top Cover
Removal / Installation.........................C-9

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] C-1


Cladding

Removal / Installation

End Cap

Tabs

Long Tab

To remove cladding...
A) Press two tabs.
B) Pull bottom of cladding away from table. Upperwrap
C) Lift cladding up to remove.

To install cladding...
A) Insert tabs at head-end of cladding into side cladding holes.
B) Slide cladding over upperwrap.
MA8084i
C) Pull bottom of cladding out, then push down on the top
to engage the long tab onto top edge of upperwrap.
D) Push bottom of cladding in, until two bottom tabs snap Bottom Tab
into place.

Bottom Tab

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 1/11/11] C-2 Always verify model & serial number
Table Shrouds
Cladding Removal / Installation..........C-2

Removal / Installation

Note
Remove all accessories (ex. chair arms) before performing this procedure.

Step 1: Remove cladding.


Note: Refer to Cladding Removal / Installation.

Step 2: Push LH side lever UP and RH side lever DOWN.


Remove drawer.
Note: Side levers are clear plastic.

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 1/11/11] C-3 Always verify model & serial number
Table Shrouds - continued

Removal / Installation

Step 3: Lower table to lowest height.


Disconnect table power cord. Step 4: Remove screws from head-end
& foot-end of outer shrouds.

Step 5: Remove side screws & outer shroud.


Repeat for other side.

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: 1/11/11] C-4 Always verify model & serial number
Table Shrouds - continued
Cladding Removal / Installation..........C-2

Removal / Installation

Step 6: Remove two screws securing each set of shrouds.


Repeat at head-end of table.

To install shrouds...
A) Install six screws securing bottom of inner shroud.
B) Install four screws securing each set of shrouds.
C) Install side screws & outer shrouds.
D) Install screws at head-end & foot-end of outer shrouds.
E) Install drawer. Step 7: Remove six screws securing
F) Install cladding. bottom of inner shroud.

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] C-5 Always verify model & serial number
Manually Raising / Lowering Table
Cladding Removal / Installation..........C-2
Table Shrouds
Equipment Alert Removal / Installation.........................C-3
Do not to move chair past limit switches
or equipment could be damage.

Step 1: Remove cladding & table shrouds.


(Access procedures)

Step 2: Run drill in reverse to raise table.


Run drill forward to lower table.
Note: Requires 1/4” socket w/ extension.

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] C-6 Always verify model & serial number
Upholstered Top
Exploded Views / Part Numbers:
Premium Uph: 28 inch. Wide..............E-3
Removal / Installation Ultra Premium Uph. 28 inch. Wide.....E-4
Premium Uph. 32 inch. Wide..............E-5

Step 2: Remove upholstered top & footrest pad.


Note: Slide upholstery up to remove from back support plate.

Step 3: Position upholstered top & footrest pad on table.


Note: Back support plate must slide under paper roll holder.

Be sure locking knobs align with holes in back / seat


mounting plates.

Locking Knobs
Locked

Step 4: Rotate locking knobs to locked position..

Step 1: Rotate all locking knobs to unlocked position. Unlocked

Note: There are four locking knobs under back section,


and two locking knobs under seat section.

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] C-7 Always verify model & serial number
Chair Arms / Brace (optional)

Removal

It may be necessary to remove the chair arm brace


when performing certain procedures.
Step 1: Squeeze lever.
Remove chair arm.

Step 2: Remove cladding & table shrouds.

Step 4: Remove brace.

Lever

Step 3: Remove two screws.


Repeat on opposite side.

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] C-8 Always verify model & serial number
Top Cover (Back Limit Switch Access)
Cladding Removal / Installation..........C-2
Table Shrouds
Removal / Installation Removal / Installation.........................C-3

Step 1: Remove cladding & table shrouds.

Step 6: Reattach cladding & table shrouds.

Step 2: Remove three screws


securing top cover.

Step 5: Secure top cover with screws


removed in Step 2.

Note: The back limit switches can be accessed without completely


removing the top cover. But if removing the cover is necessary,
perform Step 3.

Step 3: Partially separate the back lift actuator as shown.


Slide top cover off of actuator.

Back Up
Limit Switch
Step 4: Slide top cover over back actuator.
Reattach back actuator.
MA8017i

Back Down
Limit Switch

Models: 625
Serial Numbers: all

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] C-9 Always verify model & serial number
Wiring Diagrams 625-001............................................ D-2*
625-003............................................ D-3*
625-004..............................................D-4
625-005..............................................D-5
625-006..............................................D-6

Section D

* Indicates multiple pages due to a


serial number break for the wiring diagram.

© Midmark Corporation 2009 SF-1935 [Revised: 1/15/11] D-1


Fuse IEC Fuse
Circuit Holder Inlet Holder
Breaker Blue
Isolation
Brn
D Transformer
Brn
Blue
Grn/Yel
G
Blue Brn Brn
E A
Blue

Receptacle
Circuit Grn/Yel
Grn/Yel Line Filter
Breaker F
Hand
Control
A A
Blue
Brn

Blue

Blue

Brn

Brn
I L
COM Foot
Home NC Control
Limit Switches COM Blk
Distribution Board

NC Blk J13 J12 J15 J14


J10
J3
J1
J2
J9

J4

Blk
Back Dn COM
J11
Limit Switch NC
Blk J10
Wht
F1 F2 Back Up COM
Limit Switch NC
J8
J7
J6
J5

Wht
Blk F3 F2 F1
Base Dn COM
Limit Switch B
Blue

NC
Brn

Blk
Wht
Base Up COM J9 Main PC
Limit Switch NC
Wht Board
Wht

Wht

J J7 F4
H
Brn

Wht J6 J3 J2 J1
Blue

J5 J4

Wht
Wiring Diagram: 625 (-001)
Drawer Drawer Item Description Part
Heater Heater Number
K
Switch A Main Wiring Harness 015-1854-00
MA7997i B Main Wiring Harness 015-1854-00
Back Base C Back Motor Extension 015-2501-00
Actuator Actuator
D Receptacle Jumper (brown) 015-1886-00
E Receptacle Jumper (blue) 015-1886-01
F Distribution Harness 015-2005-00
C G Receptacle Harness 015-2531-00
H Back Hall Effect Extension 015-2502-00
I Foot Control Harness 015-1784-00
Models: 625-001 J Drawer Heater Harness with Switch 015-1759-00
K Used with wireless controls (optional) 015-2149-01
Serial Numbers: V2200 thru V968527
L Hand Control Harness 015-3464-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] D-2 Always verify model & serial number
Fuse IEC Fuse
Circuit Holder Inlet Holder
Breaker Blue
Isolation
Brn
D Transformer Blue

Brn
Grn/Yel

Blue
G Brn

E Blue
Grn/Yel
Brn
Circuit COM
Overtravel
NC
Breaker Limit Switch
Line Filter
Grn/Yel
F
Hand Grn/Yel

Control

A
Blue
Brn

Brn
I L

Blue
Foot

Blue

Brn
COM

Home NC Control
Limit Switches COM Blk
Distribution Board

NC Blk J13 J12 J15 J14


J10
J3
J1
J2
J9

J4

Blk
Back Dn COM
J11
Limit Switch NC
Blk J10
Wht
F1 F2 Back Up COM
Limit Switch NC
J8
J7
J6
J5

Wht
Blk F3 F2 F1
Base Dn COM
B
Limit Switch
Blue

NC
Brn

Blk
Wht
Base Up COM J9 Main PC
Limit Switch NC
Wht Board
Wht

Wht

H J7 F4
J
Brn

Wht J6 J3 J2 J1
Blue

J5 J4

Wht
Drawer Drawer Wiring Diagram: 625 (-001)
Heater Heater Item Description Part
Switch K Number
MA7997-1i A Main Wiring Harness 015-2812-00
Back Base B Main Wiring Harness 015-2812-00
Actuator Actuator
C Back Motor Extension 015-2501-00
D Receptacle Jumper (brown) 015-1886-00
C E Receptacle Jumper (blue) 015-1886-01
F Distribution Harness 015-2005-00
G Receptacle Harness 015-2531-00
H Back Hall Effect Extension 015-2502-00
I Foot Control Harness 015-1784-00
Models: 625-001 J Drawer Heater Harness with Switch 015-1759-00
K Used with wireless controls (optional) 015-2149-01
Serial Numbers: V968528 thru V1149713
L Hand Control Harness 015-3464-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] D-2.1 Always verify model & serial number
Fuse IEC Fuse
Circuit Holder Inlet Holder
Breaker Blue
Isolation
Brn Brn
D Transformer Blue

Brn
Grn/Yel

Blue
H Brn

E Grn/Yel

Blue
Blue Brn
Circuit COM
Overtravel
NC
F Breaker Limit Switch
Line Filter
Grn/Yel
G
Hand Grn/Yel

Control

A
Blue
Brn

Brn
Blue
Foot

Blue

Brn
COM

Home NC Control
Limit Switches COM Blk
Distribution Board

NC Blk J13 J12 J15 J14


J10
J3
J1
J2
J9

J4

Blk
Back Dn COM
J11
Limit Switch NC
Blk J10
Wht
F1 F2 Back Up COM
Limit Switch NC
J8
J7
J6
J5

Wht
Blk F3 F2 F1
Base Dn COM
B
Limit Switch
Blue

NC
Brn

Blk
Wht
Base Up COM J9 Main PC
Limit Switch NC
Wht Board
Wht

Wht

I J7 F4
K
Brn

Wht J6 J3 J2 J1
Blue

J5 J4

Wht
Drawer Drawer
Heater Heater
Switch L Wiring Diagram: 625 (-001)
Item Description Part
Base Number
Back
Actuator Actuator A Main Wiring Harness 015-2812-00
B 625 Limits Harness 029-5558-01
C C Back Motor Extension 029-5558-01
D Receptacle Jumper (brown) 015-1886-00
MA7997-2i
E Receptacle Jumper (blue) 015-1886-01
F Receptacle Harness, Interconnect 015-2939-00
G Distribution Harness 029-5558-01
H Receptacle Harness 029-5558-01
I Back Hall Effect Extension 029-5558-01
J Foot Control Harness 015-1784-00
Models: 625-001 K Drawer Heater Harness with Switch 015-1759-00
Serial Numbers: L Used with wireless controls (optional) 029-5558-01
V1149714 thru V1667149
M Hand Control Harness 015-3464-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/16/16] D-2.2 Always verify model & serial number
Fuse IEC Fuse
Circuit Holder Inlet Holder
Breaker Blue
Isolation
Brn Brn
D Transformer Blue

Brn
Grn/Yel

Blue
H Brn

E Grn/Yel

Blue
Blue Brn
Circuit COM
Overtravel
NC
F Breaker Limit Switch
Line Filter
Grn/Yel
G
Hand Grn/Yel

Control

Blue
A
Brn

Brn
Blue
Foot

Blue

Brn
COM

Home NC Control
Limit Switches COM Blk
Distribution Board

NC Blk J13 J12 J15 J14


J10
J3
J1
J2
J9

J4

Blk
Back Dn COM
J11
Limit Switch NC
Blk J10
Wht
F1 F2 Back Up COM
Limit Switch NC
J8
J7
J6
J5

Wht
Blk F3 F2 F1
Base Dn COM
B
Limit Switch
Blue

NC
Brn

Blk
Wht
Base Up COM J9 Main PC
Limit Switch NC
Wht Board
Wht

Wht

I J7 F4
K
Brn

Wht J6 J3 J2 J1
Blue

J5 J4

Wht
Drawer Drawer
Heater Heater
Switch L Wiring Diagram: 625 (-001)
Item Description Part
Back Base Number
Actuator Actuator A Main Wiring Harness 015-2812-00
B 625 Limits Harness 029-5558-01
C C Back Motor Extension 029-5558-01
MA7997-3i D Receptacle Jumper (brown) 015-1886-00
E Receptacle Jumper (blue) 015-1886-01
F Receptacle Harness, Interconnect 015-2939-00
G Distribution Harness 029-5558-01
H Receptacle Harness 029-5558-01
I Back Hall Effect Extension 029-5558-01
Models: 625-001 J Foot / Hand Control Harnesses 015-3464-00
K Drawer Heater Harness with Switch 015-1759-00
Serial Numbers: V1667150 thru Present
L Used with wireless controls (optional) 029-5558-01

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16 D-2.3 Always verify model & serial number
Fuse IEC Fuse
Circuit
Holder Inlet Holder
Breaker Blue
Isolation
Brn
D Transformer Blue

Brn
Grn/Yel

Blue
G Brn

E Blue
Grn/Yel
Brn
Circuit COM
Overtravel
Breaker NC
Limit Switch

A A
I L

H
J
Wiring Diagram: 625 (-003)
Item Description Part
Number

K A Main Wiring Harness 015-2812-00


B Main Wiring Harness 015-2812-00
C Back Motor Extension 015-2501-00
D Receptacle Jumper (brown) 015-1886-00
E Receptacle Jumper (blue) 015-1886-01
F Distribution Harness 015-2005-00
C G Receptacle Harness 015-2531-00
H Back Hall Effect Extension 015-2502-00
I Foot Control Harness 015-1784-00
Models: 625-003 J Drawer Heater Harness with Switch 015-1759-00
K Modular Harness (straight) 015-2149-01
Serial Numbers: V2200 thru V1149713
L Hand Control Harness 015-3464-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] D-3 Always verify model & serial number
Fuse IEC Fuse
Circuit
Holder Inlet Holder
Breaker Blue
Isolation
Brn Brn
D Transformer Blue

Brn
Grn/Yel
H Brn
Blue
E Grn/Yel

Blue
Blue Brn
Circuit COM
Overtravel
F Breaker NC
Limit Switch

A A

J M

I
K

Wiring Diagram: 625 (-003)


Item Description Part
L Number
A Main Wiring Harness 015-2812-00
B 625 Limits Harness 029-5558-01
C Back Motor Extension 029-5558-01
D Receptacle Jumper (brown) 015-1886-00
E Receptacle Jumper (blue) 015-1886-01
C F Receptacle Harness, Interconnect 015-2939-00
G Distribution Harness 029-5558-01
H Receptacle Harness 029-5558-01
I Back Hall Effect Extension 029-5558-01
J Foot Control Harness 015-1784-00
Models: 625-003 K Drawer Heater Harness with Switch 015-1759-00

Serial Numbers: L Modular Harness (straight) 029-5558-01


V1149714 thru V1667149
M Hand Control Harness 015-3464-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] D-3.1 Always verify model & serial number
Fuse IEC Fuse
Circuit
Holder Inlet Holder
Breaker Blue
Isolation
Brn Brn
D Transformer Blue

Brn
Grn/Yel
H Brn
Blue
E Grn/Yel

Blue
Blue Brn
Circuit COM
Overtravel
F Breaker NC
Limit Switch

A A

J J

I
K

L Wiring Diagram: 625 (-003)


Item Description Part
Number
A Main Wiring Harness 015-2812-00
B 625 Limits Harness 029-5558-01
C Back Motor Extension 029-5558-01
C D Receptacle Jumper (brown) 015-1886-00
E Receptacle Jumper (blue) 015-1886-01
F Receptacle Harness, Interconnect 015-2939-00
G Distribution Harness 029-5558-01
H Receptacle Harness 029-5558-01
I Back Hall Effect Extension 029-5558-01
Models: 625-003 J Foot / Hand Control Harnesses 015-3464-00
K Drawer Heater Harness with Switch 015-1759-00
Serial Numbers: V1149714 thru Present
L Modular Harness (straight) 029-5558-01

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] D-3.2 Always verify model & serial number
Fuse IEC Fuse
Holder Inlet BR Holder
Load Cells Load Cells
Y BL

X BL
BR

BL

BR

BL
W COM Overtravel
BL BL
V
NC
Limit Switch
BR
BR

BR
BL
Circuit
T U C Breaker
BL

BL BL
S D
A A F
R
Q AA B BR
E
BR

BR

Circuit
Breaker

Wiring Diagram: 625 (-004)


Item Description Part
Number
A Main Wiring Harness 015-2812-00
Z
B Distribution Harness 029-5558-00
P
C Receptacle Harness 029-5558-00

BR

BL
D Receptacle Jumper (blue) 015-1886-01
E Receptacle Jumper (brown) 015-1886-00
O F Receptacle Harness, Interconnect 015-2939-00
G Drawer Heater Harness with Switch 015-1759-00
H Back Motor Extension 029-5558-00
I Modular Harness (straight) 015-2931-02
J Modular Harness (straight) 029-5558-00

BR
BL
N K Modular Harness (serial) 029-5558-00
M L Modular Harness (serial) 015-2930-00

L J M Modular Harness (serial) 015-2930-00


N Modular Harness (straight) 015-2931-00
G
K O Back Hall Effect Extension 029-5558-00

BR
P Harness, 625 to USB COMM. PCB 015-2932-00

BL
I Q Foot Control Harness 015-1784-00
R Harness, HUB to USB COMM. PCB 015-2933-00
S Modular Harness (straight) 015-2931-02
T Harness, SER,SCL to SCL COMM. PCB 015-2934-00
U Harness, 625 USB HUB Power 015-2941-00
H
V Cable, USB Type B Panel Mount 015-2881-00
W Cable, USB Type A Panel Mount 015-2458-00
X Cable, USB Type A Panel Mount 029-5558-00
Models: 625-004 Y Cable, USB Type A Panel Mount 029-5558-00
Serial Numbers: V2200 thru V1667149 Z Distribution Harness 029-5558-00
AA Hand Control Harness 015-3464-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/16/16] D-4 Always verify model & serial number
Fuse IEC Fuse
Holder Inlet BR Holder
Load Cells Load Cells
Y BL

X BL
BR

BL

BR

BL
W COM Overtravel
BL BL
V
NC
Limit Switch
BR
BR

BR
BL
Circuit
T U C Breaker
BL

BL BL
S D
A A F
R
Q Q B BR
E
BR

BR

Circuit
Breaker

Wiring Diagram: 625 (-004)


Item Description Part
Number
Z
A Main Wiring Harness 015-2812-00
P B Distribution Harness 029-5558-00

BR

BL
C Receptacle Harness 029-5558-00
D Receptacle Jumper (blue) 015-1886-01

O E Receptacle Jumper (brown) 015-1886-00


F Receptacle Harness, Interconnect 015-2939-00
G Drawer Heater Harness with Switch 015-1759-00
H Back Motor Extension 029-5558-00
I Modular Harness (straight) 015-2931-02

BR
BL
N J Modular Harness (straight) 029-5558-00
M K Modular Harness (serial) 029-5558-00

L J L Modular Harness (serial) 015-2930-00


M Modular Harness (serial) 015-2930-00
G
K N Modular Harness (straight) 015-2931-00

BR
O Back Hall Effect Extension 029-5558-00

BL
I P Harness, 625 to USB COMM. PCB 015-2932-00
Q Foot / Hand Control Harnesses 015-3464-00
R Harness, HUB to USB COMM. PCB 015-2933-00
S Modular Harness (straight) 015-2931-02
T Harness, SER,SCL to SCL COMM. PCB 015-2934-00
H
U Harness, 625 USB HUB Power 015-2941-00
V Cable, USB Type B Panel Mount 015-2881-00
W Cable, USB Type A Panel Mount 015-2458-00
Models: 625-004 X Cable, USB Type A Panel Mount 029-5558-00

V1667150 thru Present Y Cable, USB Type A Panel Mount 029-5558-00


Serial Numbers:
Z Distribution Harness 029-5558-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] D-4.1 Always verify model & serial number
Fuse IEC Fuse
Holder Inlet Holder
Main PCB Brn
Export
Transformer
Brn

Blue
A Blue
Grn/Yel

Hand
Control
A A

Blue
Brn
E H
COM Foot
Home NC Control
Limit Switches COM
Blk
NC Blk J13 J12 J15 J14
Back Dn COM
Blk
J11
Limit Switch NC J10
Blk
Back Up COM Wht
Limit Switch NC
Wht F3 F2 F1
Base Dn COM Blk
Limit Switch B
NC
Blk
Base Up COM Wht Main PC
J9
Limit Switch NC
Wht Board
J7 F4

J6 J3 J2 J1
J5 J4

MA8063i
Back Base
Actuator Actuator

Wiring Diagram: 625 (-005)


Item Description Part
C Number
A Main Wiring Harness 015-2812-00
B Main Wiring Harness 015-2812-00
C Back Motor Extension 015-2501-00
D Back Hall Effect Extension 015-2502-00
E Foot Control Harness 015-1784-00
Models: 625-005 F Power Interrupt Harness 015-2813-00
Serial Numbers: V2200 thru V1667149 G Ground Harness 015-2570-00
H Hand Control Harness 015-3464-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] D-5 Always verify model & serial number
Fuse IEC Fuse
Holder Inlet Holder
Main PCB Brn
Export
Transformer
Brn

Blue
A Blue
Grn/Yel

Hand
Control
A A

Blue
Brn
E E
COM Foot
Home NC Control
Limit Switches COM
Blk
NC Blk J13 J12 J15 J14
Back Dn COM
Blk
J11
Limit Switch NC J10
Blk
Back Up COM Wht
Limit Switch NC
Wht F3 F2 F1
Base Dn COM Blk
Limit Switch B
NC
Blk
Base Up COM Wht Main PC
J9
Limit Switch NC
Wht Board
J7 F4

J6 J3 J2 J1
J5 J4

MA8063-1i
Back Base
Actuator Actuator
Wiring Diagram: 625 (-005)
Item Description Part
Number
C
A Main Wiring Harness 015-2812-00
B Main Wiring Harness 015-2812-00
C Back Motor Extension 015-2501-00
D Back Hall Effect Extension 015-2502-00
Models: 625-005 E Foot / Hand Control Harnesses 015-3464-00
F Power Interrupt Harness 015-2813-00
Serial Numbers: V1667150 thru Present G Ground Harness 015-2570-00

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] D-5.1 Always verify model & serial number
Fuse IEC Fuse
Holder Inlet Holder
Main PCB Blue
Export
Transformer
Blue
Brn

A Brn
Grn/Yel
Brn

Blue

COM Overtravel
NC
Limit Switch
H
Grn/Yel
D
Hand
Control
Blue
Brn

A A
F I
Foot

Blue
Blue
COM

Brn

Brn
Home NC Control
Limit Switches COM
Distribution Board

Blk
NC Blk J13 J12 J15 J14
J10
J3
J1
J2
J9

J4

Back Dn COM Blk


J11
Limit Switch NC
Blk J10
Wht
F1 F2 Back Up COM
Limit Switch NC
J8
J7
J6
J5

Wht
F3 F2 F1
Base Dn COM Blk
Limit Switch B
Blue

NC
Blk
Brn

Base Up COM Wht


J9 Main PC
Limit Switch NC
Wht Board
Wht

Wht

E J7 F4
G
Brn

Wht J6 J3 J2 J1
Blue

J5 J4

Wht
Drawer Drawer
Heater Heater
Switch Wiring Diagram: 625 (-006)
MA8064-3i Item Description Part
Base Number
Back
Actuator Actuator A Main Wiring Harness 015-2812-00
B Main Wiring Harness 015-2812-00
C Back Motor Extension 015-2501-00
C
D Distribution Harness 015-2005-00
E Back Hall Effect Extension 015-2502-00
F Foot Control Harness 015-1784-00
Models: 625-006 G Drawer Heater Harness with Switch 015-1759-00
H Power Interrupt Harness 015-2813-00
Serial Numbers: V2200 thru V1667149
I Hand Control Harness 016-3464-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] D-6 Always verify model & serial number
Fuse IEC Fuse
Holder Inlet Holder
Main PCB Blue
Export
Transformer
Blue
Brn

A Brn
Grn/Yel
Brn

Blue

COM Overtravel
NC
Limit Switch
H
Grn/Yel
D
Hand
Control
Blue
Brn

A A
F F
Foot

Blue
Blue
COM

Brn

Brn
Home NC Control
Limit Switches COM
Distribution Board

Blk
NC Blk J13 J12 J15 J14
J10
J3
J1
J2
J9

J4

Back Dn COM Blk


J11
Limit Switch NC
Blk J10
Wht
F1 F2 Back Up COM
Limit Switch NC
J8
J7
J6
J5

Wht
F3 F2 F1
Base Dn COM Blk
Limit Switch B
Blue

NC
Blk
Brn

Base Up COM Wht


J9 Main PC
Limit Switch NC
Wht Board
Wht

Wht

E J7 F4
G
Brn

Wht J6 J3 J2 J1
Blue

J5 J4

Wht
Drawer Drawer
Heater Heater
Switch
MA8064-4i

Back Base
Actuator Wiring Diagram: 625 (-006)
Actuator
Item Description Part
Number
C
A Main Wiring Harness 015-2812-00
B Main Wiring Harness 015-2812-00
C Back Motor Extension 015-2501-00
D Distribution Harness 015-2005-00
E Back Hall Effect Extension 015-2502-00
Models: 625-006 F Foot \ Hand Control Harnesses 015-3464-00
G Drawer Heater Harness with Switch 015-1759-00
Serial Numbers: V1667150 thru Present
H Power Interrupt Harness 015-2813-00

© Midmark Corporation 2009 SF-1935 [Revised: 5/16/16] D-6.1 Always verify model & serial number
Exploded Views & (-001 /-003 /-005 /-006)....................E-2
(-004)................................................E-3

Parts Lists
Section E
XXX-XXXX-XX

XXX-XXXX-XX

XXX-XXXX-XX

© Midmark Corporation 2009 SF-1935 [Revised: 2/22/12] E-1


625 (-001,-003, -005, & -006) Table Receptacle System............... E-32*
Drawer Heater System.....................E-34

Premium: 28 in. wide..........................E-4


Premium UltraFree: 28 in. wide..........E-5 Base............................................... E-20*
Premium: 32 in. wide..........................E-6 Back............................................... E-16*
Premium UltraFree: 32 in. wide..........E-7

Back Section.................................. E-15*

Seat Section.....................................E-12
Pelvic Tilt..........................................E-14 Electrical Components................... E-23*
Fuses................................................E-31

Stirrups.............................................E-13

Base Scissor / Slide Mechanism.... E-19*

Base Shrouds / Side Cladding....... E-18*

Exam Assistant® Drawer Assy........ E-17*

Foot Control......................................E-25
Hand Control....................................E-26

Base Components.......................... E-21*


MA8106i
Wireless Control Components........ E-27*

Models: 625
* Indicates multiple pages due to a
Serial Numbers: -001, -003, serial number break for the parts illustration.
-005, & -006

© Midmark Corporation 2009 SF-1935 [Revised: 1/23/19] E-2 Always verify model & serial number
625 (-004) 625 (-004 )Table Receptacle System...... E-33
Drawer Heater System............................ E-34

625 (-004 ) Premium: 28 in. wide........................ E-8


625 (-004 ) Premium UltraFree: 28 in. wide........ E-9 Base............................................... E-20*
625 (-004 ) Premium: 32 in. wide........................ E-10 Back............................................... E-16*
625 (-004 ) Premium UltraFree: 32 in. wide........ E-11

Back Section.................................. E-15*

Seat Section.....................................E-12
Pelvic Tilt..........................................E-14
Electrical Components................... E-23*
Fuses................................................E-31

Stirrups.............................................E-13

Base Scissor / Slide Mechanism.... E-19*

Base Shrouds / Side Cladding....... E-18*

Exam Assistant® Drawer Assy........ E-17*


625 (-004 ) Scale / USB Boards.......E-24

Wireless Foot Control.......................E-28


Wireless Hand Control......................E-30
625 (-004 ) Base Components....... E-22*
MA8106i

Models: 625 * Indicates multiple pages due to a


Serial Numbers: -004 serial number break for the parts illustration.

© Midmark Corporation 2009 SF-1935 [Revised: 1/23/19] E-3 Always verify model & serial number
Color Selector............ www.midmark.com

2 Standard) 028-10023-xxx
Cal 133) 028-10025-xxx
Standard) 002-10118-xxx 1
Cal 133) 002-10120-xxx
3 041-0010-01 Item #16 is a special order option.
Note: The footrest shelf will not slide into the
stowed position with this pad installed.

16 002-10218-xxx

4 053-1368-00
3.75 in.
(9.5 cm)
5 Left) 030-1494-51
Right) 030-1494-50

6 040-0010-154
14 002-1427-00

7 042-0059-08 13 MA9113i
7
045-0001-00 11 Premium Upholstery: 28 Inch Wide
9 029-3530-00-216 Item Description Qty.
8 053-1366-00
1 Upholstery Set (includes items 2 thru 13) 1
12 028-0606-00-xxx 10 040-0010-12 2 • Seat / Back Upholstery 1
3 • Nut 4
15 055-0005-00 4 • Handle 2
13 053-0424-00 5 • Handle Bracket 2
6 • Screw 4
7 • Bolt (1/4-20 x .787”) 8

Color insert not req’d for 625 8 • Paper Roll Housing 1


9 • Locking Knob 6
10 Screw 12
11 Washer 12
12 • Footrest Pad (includes item 13 - qty 4) 1
13 • • Footrest Glide 4
14 Tear Strap 1
15 Paper Roll Rod 1
Models: 625 Footrest Pad - special (4.25 in. thick)
16 1
[includes four glides (item #13)]
Serial Numbers: -001, -003,
Kit, Upholstery Packaging (002-1907-00)
-005, & -006 17 (not shown) 1

© Midmark Corporation 2009 SF-1935 [Revised: 2/13/18] E-4 Always verify model & serial number
Color Selector............ www.midmark.com

2 UltraFree) 028-10031-xxx
UltraFree) 002-10156-xxx 1 UltraFree Cal 133) 028-10032-xxx
UltraFree Cal 133) 002-10157-xxx
Item #16 is a special order option.
3 041-0010-01
Note: The footrest shelf will not slide into the
stowed position with this pad installed.

16 002-10218-xxx

4 053-1368-00 3.75 in.


(9.5 cm)

5 Left) 030-1494-51
Right) 030-1494-50

14 002-1427-00 6 040-0010-154
13 MA9113i

7 7 042-0059-08
045-0001-00 11 Premium UltraFree Upholstery:
9 029-3530-00-216 28 Inch Wide
12 028-0606-00-xxx 8 053-1366-00 Item Description Qty.

10 040-0010-12 1 Upholstery Set (includes items 2 thru 14) 1


2 • Seat / Back Upholstery 1
13 053-0424-00 15 055-0005-00 3 • Nut 4
4 • Handle 2
5 • Handle Bracket 2
6 • Screw 4
7 • Bolt (1/4-20 x .787”) 8
Color insert not req’d for 625 8 • Paper Roll Housing 1
9 • Locking Knob 6
10 • Screw 12
11 • Washer 12
12 • Footrest Pad (includes item 13 - qty 4) 1
13 • • Footrest Glide 4
14 Tear Strap 1
15 Paper Roll Rod 1
Models: 625 Footrest Pad - special (4.25 in. thick)
16 1
[includes four glides (item #13)
Serial Numbers: -001, -003,
Kit, Upholstery Packaging (002-1907-00)
-005, & -006 17 1
(not shown)

© Midmark Corporation 2009 SF-1935 [Revised: 2/13/18] E-5 Always verify model & serial number
Color Selector............ www.midmark.com

Standard) 002-10119-xxx 1
Cal 133) 002-10121-xxx 2 Standard) 028-10024-xxx
Cal 133) 028-10025-xxx

029-0017-12 11
3 053-2045-00-216

042-0059-08 4

045-0001-00 7 5 029-3530-00-216

Premium Upholstery: 32 Inch Wide


9 053-0424-00 Item Description Qty.
6 040-0010-12 055-0005-10 10 1 Upholstery Set (includes items 2 thru 10) 1
8 028-0606-00-xxx 2 • Seat / Back Upholstery 1
3 • Paper Roll Housing 1
4 • Screw 4
MA8001i
5 • Locking Knob 6
6 • Screw 12
7 • Washer 12
8 • Footrest Pad (includes item 9 - qty 4) 1
9 • • Footrest Glide 4
Models: 625 10 • Paper Roll Rod 1
11 Tear Strap 2
Serial Numbers: -001, -003,
Kit, Upholstery Packaging (002-1907-00)
-005, & -006 12 1
(not shown)

© Midmark Corporation 2009 SF-1935 [Revised: 2/13/18] E-6 Always verify model & serial number
Color Selector............ www.midmark.com

Standard) 002-10159-xxx 1
Cal 133) 002-10160-xxx 2 Standard) 028-10033-xxx
Cal 133) 028-10034-xxx

029-0017-12 11
3 053-2045-00-216

042-0059-08 4

045-0001-00 7 5 029-3530-00-216

Premium UltraFree Upholstery:


32 Inch Wide
9 053-0424-00 Item Description Qty.
6 040-0010-12 055-0005-10 10 1 Upholstery Set (includes items 2 thru 10) 1
8 028-0606-00-xxx 2 • Seat / Back Upholstery 1
3 • Paper Roll Housing 1
4 • Screw 4
MA8001i
5 • Locking Knob 6
6 • Screw 12
7 • Washer 12
8 • Footrest Pad (includes item 9 - qty 4) 1
9 • • Footrest Glide 4

Models: 625 10 • Paper Roll Rod 1


11 Tear Strap 2
Serial Numbers: -001, -003,
Kit, Upholstery Packaging (002-1907-00)
-005, & -006 12 1
(not shown)

© Midmark Corporation 2009 SF-1935 [Revised: 2/13/18] E-7 Always verify model & serial number
Color Selector............ www.midmark.com

2 Standard) 028-10027-xxx
Cal 133) 028-10028-xxx
Standard) 002-10122-xxx 1
Cal 133) 002-10123-xxx Item #16 is a special order option.
3 041-0010-01 Note: The footrest shelf will not slide into the
stowed position with this pad installed.

16 002-10028-xxx

3.75 in.
4 053-1368-00 (9.5 cm)

5 Left) 030-1494-51
Right) 030-1494-50

6 040-0010-154
14 002-1427-00
13 MA9113i

7 7 042-0059-08
625 (-004) Premium Upholstery:
045-0001-00 11 28 Inch Wide
9 029-3530-00-216 Item Description Qty.
8 053-1366-00
1 Upholstery Set (includes items 2 thru 13) 1
12 028-0606-00-xxx 10 040-0010-12 2 • Seat / Back Upholstery 1
3 • Nut 4
15 055-0005-00 4 • Handle 2
13 053-0424-00 5 • Handle Bracket 2
6 • Screw 4
7 • Bolt (1/4-20 x .787”) 8
Color insert not req’d for 625 8 • Paper Roll Housing 1
9 • Locking Knob 6
10 Screw 12
11 Washer 12
12 • Footrest Pad (includes item 13 - qty 4) 1
13 • • Footrest Glide 4
14 Tear Strap 1
15 Paper Roll Rod 1
Models: 625 Footrest Pad - special (4.25 in. thick)
16 1
[includes four glides (item #13)]
Serial Numbers: -004 Kit, Upholstery Packaging (002-1907-00)
17 1
(not shown)

© Midmark Corporation 2009 SF-1935 [Revised: 2/16/18] E-8 Always verify model & serial number
Color Selector............ www.midmark.com

UltraFree) 002-10161-xxx 1
2 UltraFree) 028-10035-xxx
UltraFree Cal 133) 028-10036-xxx
UltraFree Cal 133) 002-10162-xxx
Item #16 is a special order option.
3 041-0010-01 Note: The footrest shelf will not slide into the
stowed position with this pad installed.

16 002-10218-xxx

3.75 in.
4 053-1368-00 (9.5 cm)

5 Left) 030-1494-51
Right) 030-1494-50

14 002-1427-00 6 040-0010-154
13 MA9113i

7 7 042-0059-08
045-0001-00 11
625 (-004) Premium UltraFree Upholstery:
9 029-3530-00-216 28 Inch Wide
12 028-0606-00-xxx 8 053-1366-00 Item Description Qty.

10 040-0010-12 1 Upholstery Set (includes items 2 thru 14) 1


2 • Seat / Back Upholstery 1
13 053-0424-00 15 055-0005-00 3 • Nut 4
4 • Handle 2
5 • Handle Bracket 2
6 • Screw 4
7 • Bolt (1/4-20 x .787”) 8
Color insert not req’d for 625 8 • Paper Roll Housing 1
9 • Locking Knob 6
10 • Screw 12
11 • Washer 12
12 • Footrest Pad (includes item 13 - qty 4) 1
13 • • Footrest Glide 4
14 Tear Strap 1
15 Paper Roll Rod 1
Models: 625 Footrest Pad - special (4.25 in. thick)
16 1
[includes four glides (item #13)
Serial Numbers: -004 Kit, Upholstery Packaging (002-1907-00)
17 1
(not shown)

© Midmark Corporation 2009 SF-1935 [Revised: 2/16/18] E-9 Always verify model & serial number
Color Selector............ www.midmark.com

Standard) 002-10124-xxx 1
Cal 133) 002-10125-xxx 2 Standard) 028-10029-xxx
Cal 133) 028-10030-xxx

029-0017-12 11
3 053-2045-00-216

042-0059-08 4

045-0001-00 7 5 029-3530-00-216

625 (-004) Premium Upholstery:


32 Inch Wide
9 053-0424-00 Item Description Qty.
6 040-0010-12 055-0005-10 10 1 Upholstery Set (includes items 2 thru 10) 1
8 028-0606-00-xxx 2 • Seat / Back Upholstery 1
3 • Paper Roll Housing 1
4 • Screw 4
MA10664i
5 • Locking Knob 6
6 • Screw 12
7 • Washer 12
8 • Footrest Pad (includes item 9 - qty 4) 1
9 • • Footrest Glide 4
Models: 625 10 • Paper Roll Rod 1
11 Tear Strap 2
Serial Numbers: -004 Kit, Upholstery Packaging (002-1907-00)
12 1
(not shown)

© Midmark Corporation 2009 SF-1935 [Revised: 2/16/18] E-10 Always verify model & serial number
Color Selector............ www.midmark.com

Standard) 002-10163-xxx 1
Cal 133) 002-10164-xxx 2 Standard) 028-10037-xxx
Cal 133) 028-10038-xxx

029-0017-12 11
3 053-2045-00-216

042-0059-08 4

045-0001-00 7 5 029-3530-00-216

625 (-004) Premium UltraFree Upholstery:


32 Inch Wide
9 053-0424-00 Item Description Qty.
6 040-0010-12 055-0005-10 10 1 Upholstery Set (includes items 2 thru 10) 1
8 028-0606-00-xxx 2 • Seat / Back Upholstery 1
3 • Paper Roll Housing 1
4 • Screw 4
MA10665i
5 • Locking Knob 6
6 • Screw 12
7 • Washer 12
8 • Footrest Pad (includes item 9 - qty 4) 1
9 • • Footrest Glide 4
Models: 625 10 • Paper Roll Rod 1
11 Tear Strap 2
Serial Numbers: -004 Kit, Upholstery Packaging (002-1907-00)
12 1
(not shown)

© Midmark Corporation 2009 SF-1935 [Revised: 2/16/18] E-11 Always verify model & serial number
050-6305-00-216 13

029-4550-00-216 1
040-0010-47 2

053-2146-00-312 3
12 053-0068-06
053-1550-00-216 4

053-1508-00-312 5 11 057-0840-00

053-0220-00-271 6

030-1786-00-216 7
10 030-1819-00-216
Seat Section
Item Description Qty.
Upperwrap Assembly
1 1
(includes items 2 thru 10)
2 • Screw (#10-24 x 3/8”) 6
3 • Stirrup Guide 1
053-1377-00 8 4 • Stem Bumper 4
2 042-0184-01 14 5 • Runner Glide 4
6 • Plastic Treatment Pan 1
7 • Footrest Shelf 1
9 002-1331-02 15 8 • Footrest Trim 1
9 • Refer to: “Stirrup Assembly” 2
10 • Upperwrap Weldment 1
002-1331-00 16 11 Pivot Rod 1
12 Snap Bushing 3
13 Seat Mtg. Frame 1
14 Screw 2
Models: 625 15 Switch Housing and Label Kit (with one switch 1
hole [includes label])
Serial Numbers: all Switch Housing and Label Kit (no holes [includes
16 1
label])

© Midmark Corporation 2009 SF-1935 [Revised: 9/26/17] E-12 Always verify model & serial number
029-2951-06 1

8 053-0387-00
020-0337-00-216 2
7 050-5027-00

6 016-0400-00

5 040-0010-47

4 051-1003-01
042-0001-39 3

MA703900I
Stirrups
Item Description Qty.
Stirrup Assembly
1 2
(includes items 2 thru 4)
2 • Stirrup 1
3 • Roll Pin 1
4 • Stirrup Bar 1
5 Screw (10-24 x 3/8”) 2
Models: 625 6 Guide Bracket Spring 2
Serial Numbers: all 7 Guide Bracket 2
8 Pivot Boss 2

© Midmark Corporation 2009 SF-1935 [Revised: 04/10] E-13 Always verify model & serial number
016-1018-00 2

4 050-6304-00-216
002-1712-00 1
053-1550-00-216 3
5 053-1564-00-312

Pelvic Tilt
MA716200i
Item Description Qty.
Pelvic Tilt Assembly
1 1
(includes items 2 thru 5)
2 • Pivot Pin 2
Models: 625 3 • Bumper 2
Serial Numbers: all 4 • Bracket 1
5 • Handle 2

© Midmark Corporation 2009 SF-1935 [Revised: 3/7/16] E-14 Always verify model & serial number
029-4803-00-216 14
1 050-6494-00-216

2 053-1594-00-312

053-1595-00-312 13 3 053-1998-00-271

053-1565-00-312 12

030-1818-00-216 11
8
4 050-7584-00-216
8
15
5 002-1460-00

6 040-0010-47
Back Section
Item Description Qty.
057-0840-00 10 1 Bracket 1
7 030-1818-01-216 2 Limit Switch Bumper 1
3 Top Cover 1
040-0312-90 9 4 Top Cover Bracket 1
5 Back Actuator Bracket 1
6 Screw (10-24 x 3/8”) 2
8 040-0250-157 7 Support Channel - LH 1
8 Hex Bolt (1/4-20 x 1/2”) 14
MA8103i 9 Hex Bolt (5/16-18 x 1/2) 4
10 Hinge Rod 1
11 Support Channel - RH 1
12 Bumper 2
Models: 625 13 Bumper 1
Serial Numbers: V2200 thru V1083261 14 Back Weldment 1
15 Refer to: “Back Actuator / Limit Switches” Ref

© Midmark Corporation 2009 SF-1935 [Revised: 10/15/15] E-15 Always verify model & serial number
029-4803-00-216 13
1 050-6494-00-216

2 053-1594-00-312

053-1595-00-312 12 3 053-1998-00-271

053-1565-00-312 11

030-1818-00-216 10
7

7
14
4 002-1460-00

5 040-0010-47

Back Section
Item Description Qty.
057-0840-00 9 6 030-1818-01-216
1 Bracket 1
2 Limit Switch Bumper 1
040-0312-90 8 3 Top Cover 1
4 Back Actuator Bracket 1
7 040-0250-157 5 Screw (10-24 x 3/8”) 2
6 Support Channel - LH 1
MA8103-1i 7 Hex Bolt (1/4-20 x 1/2”) 14
8 Hex Bolt (5/16-18 x 1/2) 4
9 Hinge Rod 1
10 Support Channel - RH 1
11 Bumper 2
Models: 625 12 Bumper 1
Serial Numbers: V1083262 thru Present 13 Back Weldment 1
14 Refer to: “Back Actuator / Limit Switches” Ref

© Midmark Corporation 2009 SF-1935 [Revised: mo/dd/yr] E-15.1 Always verify model & serial number
042-0602-33 15
2 053-2140-00

16 053-0858-01

002-1183-01 1 17 042-0004-00

6 016-0662-00

5 053-2127-00

4 002-1408-00
041-0010-00 14
7 041-0250-01
045-0001-00 13
18 015-1055-00
053-2097-00 12

053-2139-01 3 Back Actuator / Limit Switches


Item Description Qty.
1 Back Actuator Assembly (incl. items 2 thru 4) 1
2 • Bushing 2
3 • Bushing 2
4 • Motor Only (includes items 5 thru 14) 1
042-0622-02 23 5 • • Motor Isolator 1
8 050-8239-00-216 6 • • Motor Coupler 1
7 • • Nut 1
042-0007-06 22 8 • • Bracket 1
9 040-0010-158 9 • • Screw (#10 x 32 x 1/4”) 2
10 • • Spring 1
053-2161-00-312 21 11 • • Hex Screw (1/4-20 x 1 1/2”) 1

10 025-0049-03 12 • • Isolation Washer 2


13 • • Washer 2
040-0250-157 20 14 • • Nut 2
11 042-0059-04 MA8105i
15 Clevis Pin 1

050-7793-00-216 19 16 Nylon Washer 2

18 17 Cotter Pin 1
18 Limit Switch 2
19 Bracket 1
20 Hex Screw 1
Models: 625 21 Rubber Washer 2

Serial Numbers: 22 E-ring 2


V2200 thru V1083261
23 Clevis Pin 1

© Midmark Corporation 2009 SF-1935 [Revised: 1/27/17 E-16 Always verify model & serial number
042-0602-33 14
2 053-2140-00

15 053-0858-01

002-1183-00 1 16 042-0004-00

5 016-0662-00

4 053-2127-00

3 002-1408-00
041-0010-00 13
6 041-0250-01
045-0001-00 12
17 015-1055-00
053-2097-00 11

Back Actuator / Limit Switches


Item Description Qty.
1 Back Actuator Assembly (incl. items 2 & 3) 1
2 • Bushing 2
3 • Motor Only (includes items 4 thru 13) 1
042-0622-02 21 4 • • Motor Isolator 1
7 050-8239-00-216 5 • • Motor Coupler 1
6 • • Nut 1

042-0007-06 20 7 • • Bracket 1
8 040-0010-158 8 • • Screw (#10 x 32 x 1/4”) 2
9 • • Spring 1
10 • • Hex Screw (1/4-20 x 1 1/2”) 1
040-0250-157 19 9 025-0049-03 11 • • Isolation Washer 2
12 • • Washer 2
13 • • Nut 2
10 042-0059-04 14 Clevis Pin 1
050-7793-00-216 18 MA8105-1i
15 Nylon Washer 2
17 16 Cotter Pin 1
17 Limit Switch 2
18 Bracket 1
Models: 625 19 Hex Screw 1
Serial Numbers: V1083262 thru V1513012 20 E-ring 2
21 Clevis Pin 1

© Midmark Corporation 2009 SF-1935 [Revised: 01/24/14] E-16.1 Always verify model & serial number
042-0602-33 16 2 053-2140-00

17 053-0858-01

002-1183-00 1 18 042-0004-00

7 016-0662-00 053-2539-00-312 3

6 053-2127-00
042-0211-01 4
041-0010-00 15 5 002-1408-00

045-0001-00 14 8 041-0250-01

053-2097-00 13 19 015-1055-00

Back Actuator / Limit Switches


Item Description Qty.
1 Back Actuator Assembly (incl. items 2 thru 5) 1
2 • Bushing 2
042-0622-02 23 9 050-8239-00-216
3 • Actuator Motor Cover 1
4 • Screw (5.0mm x 20mm) 1
5 • Motor Only (includes items 4 thru 15) 1
6 • • Motor Isolator 1
042-0007-06 22 10 040-0010-158
7 • • Motor Coupler 1
8 • • Nut 1
11 025-0049-03 9 • • Bracket 1
040-0250-157 21 10 • • Screw (#10 x 32 x 1/4”) 2
11 • • Spring 1
12 042-0059-04 12 • • Hex Screw (1/4-20 x 1 1/2”) 1
050-7793-00-216 20 13 • • Isolation Washer 2

19 14 • • Washer 2
15 • • Nut 2
16 Clevis Pin 1
MA8105-2i 17 Nylon Washer 2
18 Cotter Pin 1
19 Limit Switch 2
20 Bracket 1
Models: 625 21 Hex Screw 1
Serial Numbers: V1513013 thru Present 22 E-ring 2
23 Clevis Pin 1

© Midmark Corporation 2009 SF-1935 [Revised: 01/24/14] E-16.2 Always verify model & serial number
1 029-4536-00
17 040-0010-47

13 053-2001-00-216
053-2072-00-216 15

16 050-7552-00-216
029-4535-00 2
6 041-0009-03 18 042-0010-07

051-1008-02 5

016-0945-01 7

19 050-7553-00-216
Exam Assistant® Drawer Assembly
14 016-1186-04 Item Description Qty.
002-1676-00 3
Drawer Assistant® Drawer Assembly
1 1
(includes items 2 thru 19)

029-4534-00-271 4 12 040-0008-106 2 • Drawer Assembly (includes 4 thru 13) 1


• Drawer Front and Hardware Kit
3 (includes items 4 thru 6)
4 •• Drawer Front Assembly 1
11 044-0013-50 5 •• Pivot Rod 1
6 •• Push Nut 2
RH: 016-1186-02 8 7 • Edge Trim 2
LH: 016-1186-03 8 • Inner Slide (RH shown) 2
9 • Drawer Heat Sink 1
10 • Screw (M4-1.79 x 12mm) 7
11 • Weld Nut 2
050-7551-00-216 9 12 • Screw (#8-32 x 1/4”) 2
042-0184-01 10 13 • Drawer 1
14 • Outer & Middle Slide Assembly 2
15 • Drawer Divider 2
16 • Side Bracket 2
Models: 625 17 • Screw 4
Serial Numbers: V2200 thru V1307858 18 • Pop Rivet 4
19 • Drawer Housing 1

© Midmark Corporation 2009 SF-1935 [Revised: 5/22/13] E-17 Always verify model & serial number
1 029-4536-00
18 040-0010-47

14 053-2001-00-216
053-2072-00-216 16

029-4535-00 2 17 050-7552-00-216

6 041-0009-03

19 042-0010-07

051-1008-02 5

053-2429-00-216 7

016-0945-01 8
15 016-1186-04 20 050-7553-00-216

Exam Assistant® Drawer Assembly


Item Description Qty.
002-1676-00 3 Drawer Assistant® Drawer Assembly
1 1
(includes items 2 thru 20)
2 • Drawer Assembly (includes 4 thru 14) 1
029-4534-00-271 4 13 040-0008-106 3 • Drawer Front & Hardware Kit (includes 4 thru 7)
4 •• Drawer Front Assembly 1
5 •• Pivot Rod 1
12 044-0013-50 6 •• Push Nut 2
7 •• Fold Down Drawer Stop 2
RH: 016-1186-02 9 8 • Edge Trim 2
LH: 016-1186-03 9 • Drawer Slide Assembly (RH shown) 2
050-7551-00-216 10 10 • Drawer Heat Sink 1
042-0184-01 11 11 • Screw (M4-1.79 x 12mm) 7
12 • Weld Nut 2
13 • Screw (#8-32 x 1/4”) 2
14 • Drawer 1
15 • Outer & Middle Slide Assembly 2
16 • Drawer Divider 2
17 • Side Bracket 2
Models: 625 18 • Screw 4
Serial Numbers: V1307859 thru Present 19 • Pop Rivet 4
20 • Drawer Housing 1

© Midmark Corporation 2009 SF-1935 [Revised: 6/15/15] E-17.1 Always verify model & serial number
1 002-1608-00 053-2301-00 10

053-0752-00 11
2 040-0008-106

3 LH: 029-4548-00-216
RH: 029-4547-00-216

4 040-0010-47
042-0184-01 19

053-1553-00-271 18

002-1717-00 9 13 LH: 053-2059-00-271


RH: 053-2059-01-271

029-3050-00 8 5 016-0140-05

Base Shrouds & Side Cladding


Item Description Qty.
1 Kit, Shroud Assembly (includes items 2 thru 11) 1
2 Screw (#8-32 x 1/4”) 12
Outer Shroud Assembly - LH shown
3 2
(includes items 11)
12 a) LH: 002-1744-00 4 Screw (#10-24 x 3/8”) 10

040-0008-78 7 4 b) RH: 002-1744-01 5 Trim Lock 4


c) RH: 002-1744-02 6 Inner Shroud Assembly - LH shown 2
(includes items 10 & 11)
7 Screw (#8-32 x 1/4”) 4
LH: 029-4789-00-312 6 Mid-Inner Shroud Assembly - LH shown
H: 029-4790-00-312
R
14 053-2564-00-271
8 (includes items 5, 10, & 11) 2

Mid-Outer Shroud Assembly - LH shown


9 2
(includes items 10 & 11)
10 Stem Bumper 14
11 Snap-On Glide 32
053-2070-00-271 17 15 061-0860-01 Side Cladding Assembly
a) LH with receptacle (incl. items 13 thru 17)
12 2
b) RH without receptacle (incl. items 13 thru 19)
c) RH with receptacle (incl. items 13 thru 17)
13 • End Cap - LH shown 2

16 053-2008-00-271 MA8092i
14 • Hole Plug 2
15 • Midmark Nameplate 2
16 • Hole Plug 2
17 • Knee Crutch Access Cover 2
Models: 625 Items 18 & 19 apply only to models w/o receptacle
Serial Numbers: V2200 thru V1077078 18 • Receptacle Hole Cover 1
19 • Screw (M4 x 12) 2

© Midmark Corporation 2009 SF-1935 [Revised: 11/21/14] E-18 Always verify model & serial number
053-2301-00 10
1 002-1608-00
053-0752-00 11
2 040-0008-106

3 LH: 029-4548-00-216
RH: 029-4547-00-216 053-2213-00 12

4 040-0010-47 042-0184-01 20

053-1553-00-271 19

14 LH: 053-2059-00-271
002-1717-00 9 RH: 053-2059-01-271

029-3050-00 8 Base Shrouds & Side Cladding


5 016-0140-05
Item Description Qty.
1 Kit, Shroud Assembly (includes items 2 thru 12) 1
2 Screw (#8-32 x 1/4”) 12
Outer Shroud Assembly - LH shown
3 2
(includes items 11)
4 Screw (#10-24 x 3/8”) 10
5 Trim Lock 4
Inner Shroud Assembly - LH shown
6 2
13 a) LH: 002-1744-00 (includes items 10 & 11)
040-0008-78 7 4 b) RH: 002-1744-01 7 Screw (#8-32 x 1/4”) 4
c) RH: 002-1744-02 8 Mid-Inner Shroud Assembly - LH shown 2
(includes items 5, 10, 11, & 12)
Mid-Outer Shroud Assembly - LH shown
9 2
LH: 029-4789-00-312 6 (includes items 10, 11, & 12)

H: 029-4790-00-312
R 15 053-2564-00-271 10 Stem Bumper 14
11 Snap-On Glide 32
12 Shroud Isolator 8
Side Cladding Assembly
a) LH with receptacle (incl. items 14 thru 18)
13 2
b) RH without receptacle (incl. items 14 thru 20)
053-2070-00-271 18 16 061-0860-01 c) RH with receptacle (incl. items 14 thru 18)
14 • End Cap - LH shown 2
15 • Hole Plug 2
16 • Midmark Nameplate 2
17 053-2008-00-271 MA8092-1i 17 • Hole Plug 2
18 • Knee Crutch Access Cover 2
Models: 625 Items 19 & 20 apply only to models w/o receptacle
19 • Receptacle Hole Cover 1
Serial Numbers: V1077079 thru V1089389 20 • Screw (M4 x 12) 2

© Midmark Corporation 2009 SF-1935 [Revised: 11/21/14] E-18.1 Always verify model & serial number
1 002-1608-00 053-2301-00 10

053-0752-00 11
2 040-0008-106

3 LH: 029-4548-00-216
RH: 029-4547-00-216

4 040-0010-47
042-0184-01 19

053-1553-00-271 18

002-1717-00 9 13 LH: 053-2059-00-271


RH: 053-2059-01-271

029-3050-00 8 5 016-0140-05

Base Shrouds & Side Cladding


Item Description Qty.
1 Kit, Shroud Assembly (includes items 2 thru 11) 1
2 Screw (#8-32 x 1/4”) 12
Outer Shroud Assembly - LH shown
3 2
(includes items 11)
12 a) LH: 002-1744-00 4 Screw (#10-24 x 3/8”) 10

040-0008-78 7 4 b) RH: 002-1744-01 5 Trim Lock 4


c) RH: 002-1744-02 6 Inner Shroud Assembly - LH shown 2
(includes items 10 & 11)
7 Screw (#8-32 x 1/4”) 4
LH: 029-4789-00-312 6 Mid-Inner Shroud Assembly - LH shown
H: 029-4790-00-312
R
14 053-2564-00-271
8 (includes items 5, 10, & 11) 2

Mid-Outer Shroud Assembly - LH shown


9 2
(includes items 10 & 11)
10 Stem Bumper 14
11 Snap-On Glide 32
053-2070-00-271 17 15 061-0860-01 Side Cladding Assembly
a) LH with receptacle (incl. items 13 thru 17)
12 2
b) RH without receptacle (incl. items 13 thru 19)
c) RH with receptacle (incl. items 13 thru 17)
13 • End Cap - LH shown 2

16 053-2008-00-271 MA8092-2i
14 • Hole Plug 2
15 • Midmark Nameplate 2
16 • Hole Plug 2
17 • Knee Crutch Access Cover 2
Models: 625 Items 18 & 19 apply only to models w/o receptacle
Serial Numbers: V1089390 thru V1099004 18 • Receptacle Hole Cover 1
19 • Screw (M4 x 12) 2

© Midmark Corporation 2009 SF-1935 [Revised: 11/21/14] E-18.2 Always verify model & serial number
053-2301-00 10
1 002-1608-00
053-0752-00 11
2 040-0008-106

3 LH: 029-4548-00-216
RH: 029-4547-00-216 053-2213-00 12

4 040-0010-47 042-0184-01 20

053-1553-00-271 19

14 LH: 053-2059-00-271
002-1717-00 9 RH: 053-2059-01-271

029-3050-00 8
5 016-0140-05 Base Shrouds & Side Cladding
Item Description Qty.
1 Kit, Shroud Assembly (includes items 2 thru 12) 1
2 Screw (#8-32 x 1/4”) 12
Outer Shroud Assembly - LH shown
3 2
(includes items 11)
4 Screw (#10-24 x 3/8”) 10
5 Trim Lock 4
13 a) LH: 002-1744-00 6 Inner Shroud Assembly - LH shown 2
040-0008-78 7 4 b) RH: 002-1744-01
(includes items 10 & 11)
7 Screw (#8-32 x 1/4”) 4
c) RH: 002-1744-02
Mid-Inner Shroud Assembly - LH shown
8 2
(includes items 5, 10, 11, & 12)
Mid-Outer Shroud Assembly - LH shown
LH: 029-4789-00-312 6 9 (includes items 10, 11, & 12) 2
H: 029-4790-00-312
R 15 053-2564-00-271 10 Stem Bumper 14
11 Snap-On Glide 32
12 Shroud Isolator 8
Side Cladding Assembly
a) LH with receptacle (incl. items 14 thru 18)
053-2070-00-271 18 16 061-0860-01 13 b) RH without receptacle (incl. items 14 thru 20) 2
c) RH with receptacle (incl. items 14 thru 18)
14 • End Cap - LH shown 2
15 • Hole Plug 2
17 053-2008-00-271 MA8092-3i
16 • Midmark Nameplate 2
17 • Hole Plug 2
18 • Knee Crutch Access Cover 2
Models: 625 Items 19 & 20 apply only to models w/o receptacle
Serial Numbers: V1099005 - V1563949 19 • Receptacle Hole Cover 1
20 • Screw (M4 x 12) 2

© Midmark Corporation 2009 SF-1935 [Revised: 11/21/14] E-18.3 Always verify model & serial number
053-2301-00 10
1 002-1608-00
053-0752-00 11
2 040-0008-106

3 LH: 029-4548-00-216
RH: 029-4547-00-216 053-2213-00 12

4 040-0010-47 042-0184-01 20

053-1553-00-271 19

14 LH: 053-2059-00-271
002-1717-00 9 RH: 053-2059-01-271

029-3050-00 8
5 016-0140-05 Base Shrouds & Side Cladding
Item Description Qty.
1 Kit, Shroud Assembly (includes items 2 thru 12) 1
2 Screw (#8-32 x 1/4”) 12
Outer Shroud Assembly - LH shown
3 2
(includes items 11)
4 Screw (#10-24 x 3/8”) 10
5 Trim Lock 4
13 a) LH: 002-1744-00 6 Inner Shroud Assembly - LH shown 2
040-0008-78 7 4 b) RH: 002-1744-01
(includes items 10 & 11)
7 Screw (#8-32 x 1/4”) 4
c) RH: 002-1744-02
Mid-Inner Shroud Assembly - LH shown
8 2
(includes items 5, 10, 11, & 12)
Mid-Outer Shroud Assembly - LH shown
LH: 029-4789-00-312 6 9 (includes items 10, 11, & 12) 2
H: 029-4790-00-312
R 15 053-2564-00-271 10 Stem Bumper 14
11 Snap-On Glide 32
12 Shroud Isolator 8
Side Cladding Assembly
a) LH with receptacle (incl. items 14 thru 18)
053-2070-00-271 18 16 061-0860-01 13 b) RH without receptacle (incl. items 14 thru 20) 2
c) RH with receptacle (incl. items 14 thru 18)
14 • End Cap - LH shown 2
15 • Hole Plug 2
17 053-2008-00-271 MA8092-4i
16 • Midmark Nameplate 2
17 • Hole Plug 2
18 • Knee Crutch Access Cover 2
Models: 625 Items 19 & 20 apply only to models w/o receptacle
Serial Numbers: V1563950 - Present 19 • Receptacle Hole Cover 1
20 • Screw (M4 x 12) 2

© Midmark Corporation 2009 SF-1935 [Revised: 11/21/14] E-18.4 Always verify model & serial number
2 040-0250-157
RH: 030-1582-00-216 1
LH: 030-1582-01-216
3 050-5454-50

4 002-1493-00
7
042-0606-05 13

025-0049-02 12

050-7608-00-216 11
2
045-0001-29 10

042-0007-03 9
5
030-1940-00-216 8

016-0392-11 7

Base Scissor / Slide Mechanism


4 Item Description Qty.
1 Scissor Channel Weldment (RH shown) 2

6 057-0849-00 2 Flat Head Screw (1/4-20 x 1/2”) 4


3 Top Cover Support Channel 1
4 Kit, Scissor Rollers 1
5 Refer to: “Base Components” Ref
6 Pivot Shaft 1
MA8099i 7 Flange Bearing 4
8 Scissor Arm Weldment 1
9 E-clip 1
10 Washer 1
Models: 625 11 Auto Return (“Home”) Switch Strike 1
Serial Numbers: V2200 thru V1158421 12 Spring 1
13 Clevis Pin 1

© Midmark Corporation 2009 SF-1935 [Revised: 12/6/11] E-19 Always verify model & serial number
2 040-0250-157
RH: 030-1582-00-216 1
LH: 030-1582-01-216
3 050-5454-50

4 021-0086-02
8
042-0606-05 14 5 053-2225-00

025-0049-02 13

050-7608-00-216 12
2
045-0001-29 11

042-0007-03 10
6
030-1940-00-216 9

016-1570-09 8
Base Scissor / Slide Mechanism
Item Description Qty.
4
1 Scissor Channel Weldment (RH shown) 2
2 Flat Head Screw (1/4-20 x 1/2”) 4

7 057-0849-00 3 Top Cover Support Channel 1


4 Extrusion Slide 2
5 Bearing Guide 4
6 Refer to: “Base Components” Ref
7 Pivot Shaft 1
MA8384i 8 Flange Bearing 4
9 Scissor Arm Weldment 1
10 E-clip 1
11 Washer 1
Models: 625 12 Auto Return (“Home”) Switch Strike 1
Serial Numbers: V1158422 thru Present 13 Spring 1
14 Clevis Pin 1

© Midmark Corporation 2009 SF-1935 [Revised: 2/24/15] E-19.1 Always verify model & serial number
040-0375-86 18 19 053-2153-00

20 045-0000-05

053-2143-00 17 21 041-0375-10

030-1894-00 16 2 041-0010-10

15 050-5671-50 3 045-0001-00

14 015-1055-00
4 053-2097-00

5 016-0662-00

002-1273-00 1
6 053-2127-00

7 015-2385-00
13 040-0250-157 Base Actuator / Limit Switches
Item Description Qty.
1 Base Actuator Kit (includes items 2 thru 7) 1
2 • Nut 2
3 • Shim Washer 2
4 • Isolation Washer 2
5 • Motor Coupler 1
6 • Motor Isolator 1
7 • Motor Assembly 1
8 Isolator Clamp 1
8 050-7947-00 9 Groove Pin 1
10 Actuator Pin Isolator 1
11 Bushing 2
9 042-0009-19 12 Base Actuator Pin 1
057-1161-00 12 13 Hex Bolt (1/4-20 x 1/2”) 3
10 053-2155-00 14 Limit Switch 4
15 Switch Bracket 1
053-2139-01 11 MA8101i 16 Cradle Weldment 1
17 Pivot Block Isolator 1
18 Hex Screw (3/8-16 x 3 1/2”) 2
Models: 625 19 Washer 2
Serial Numbers: V2200 thru V968527 20 Washer 2
21 Nut 2

© Midmark Corporation 2009 SF-1935 [Revised: 01/05/11] E-20 Always verify model & serial number
18 053-2153-00
040-0375-86 17
19 045-0000-05
053-2143-00 16 20 041-0375-10
030-1894-00-216 15 2 041-0010-10
22 050-8226-00-000
3 045-0001-00

23 015-2811-00 4 053-2097-00

12 040-0250-157 5 016-0662-00
002-1273-00 1
6 053-2127-00
040-0250-89 11
7 015-2385-00
13a 015-1055-00
Base Actuator / Limit Switches
Item Description Qty.

14 050-8225-00-000 1 Base Actuator Kit (includes items 2 thru 7) 1


2 • Nut 2
3 • Shim Washer 2
4 • Isolation Washer 2
057-1190-00 10 5 • Motor Coupler 1
6 • Motor Isolator 1
7 • Motor Assembly 1
20 8 Inner Actuator Mount 1
9 Base Actuator Isolator 1
19 10 Base Actuator Pin 1
11 Screw (1/4-20 x 1”) 2
13b 18 12 Hex Bolt (1/4-20 x 1/2”) 4
a) Base Up Limit Switch
13 b) Home Function Limit Switches 4
c) Base Down Limit Switch
13c 8 030-1939-00-000
14 Switch Bracket 1
15 Cradle Weldment 1
21 040-0375-87 16 Pivot Block Isolator 1
17 Hex Screw (3/8-16 x 3 1/2”) 2
9 053-2180-00 MA8196i 18 Washer 4
19 Washer 4
20 Nut 4
Models: 625 21 Hex Washer Head Screw (3/8-16 x 1 1/8”) 2
Serial Numbers: V968528 thru present 22 Switch Bracket 1
23 Over Travel Limit Switch 1

© Midmark Corporation 2009 SF-1935 [Revised: 9/4/12] E-20.1 Always verify model & serial number
3 1 040-0250-157

2 020-0364-00-312
3 053-0050-00

4 LH: 020-0365-01-312
RH: 020-0332-00-312

5 040-0250-65

6 051-1232-00

7 016-1196-00

1
11 015-3464-00
MA8100i
Base Components
8 030-1816-00-312 Item Description Qty.
1 Hex Screw (1/4-20 x 1/2”) 11
9 015-1784-00 2 Front Base Casting 1
3 Hole Plug 4

10 050-7631-00-312 4 Back Base Cover - LH shown 2


5 Screw (1/4-20 x 1/2”) 6
6 Leveling Foot Mount 2
7 Leveling Foot 2
8 Base Weldment 1
Models: 625 (-001 /-003 /-005 /-006) 9 Foot Control Harness 1

Serial Numbers: 10 Bracket 1


V2200 thru V968527
11 Hand Control Harness 1

© Midmark Corporation 2009 SF-1935 [Revised: 1/27/17] E-21 Always verify model & serial number
3 1 040-0250-157

2 020-0364-00-312
3 053-0050-00

4 LH: 020-0365-01-312
RH: 020-0365-00-312

5 040-0250-65

6 051-1232-00

7 016-1196-00

1 11 015-3464-00

Base Components
MA8100-1i
8 030-1938-00-312 Item Description Qty.
1 Hex Screw (1/4-20 x 1/2”) 11
9 015-1784-00 2 Front Base Casting 1
3 Hole Plug 4
10 050-7631-00-312 4 Back Base Cover - LH shown 2
5 Screw (1/4-20 x 1/2”) 6
6 Leveling Foot Mount 2
7 Leveling Foot 2
8 Base Weldment 1
Models: 625 (-001 /-003 /-005 /-006) 9 Foot Control Harness 1
Serial Numbers: V968528 thru V1182355 10 Bracket 1
11 Hand Control Harness 1

© Midmark Corporation 2009 SF-1935 [Revised: 1/27/17] E-21.1 Always verify model & serial number
3 1 040-0250-157

2 020-0364-00-312
3 053-0050-02

4 LH: 020-0365-01-312
RH: 020-0365-00-312

5 040-0250-65

6 051-1232-00

7 016-1196-00

1
11 015-3464-00

MA8100-2i Base Components


8 030-1938-00-312 Item Description Qty.
1 Hex Screw (1/4-20 x 1/2”) 11
9 015-1784-00 2 Front Base Casting 1
3 Hole Plug 4

10 050-7631-00-312 4 Back Base Cover - LH shown 2


5 Screw (1/4-20 x 1/2”) 6
6 Leveling Foot Mount 2
7 Leveling Foot 2
8 Base Weldment 1
Models: 625 (-001 /-003 /-005 /-006) 9 Foot Control Harness 1

Serial Numbers: 10 Bracket 1


V1182356 thru V1667149
11 Hand Control Harness 1

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] E-21.2 Always verify model & serial number
3 1 040-0250-157

2 020-0364-00-312
3 053-0050-02

4 LH: 020-0365-01-312
RH: 020-0365-00-312

5 040-0250-65

6 051-1232-00

7 016-1196-00

1 9 015-3464-00

8 030-1938-00-312 Base Components


Item Description Qty.
1 Hex Screw (1/4-20 x 1/2”) 11
2 Front Base Casting 1
10 050-7631-00-312 3 Hole Plug 4
4 Back Base Cover - LH shown 2
5 Screw (1/4-20 x 1/2”) 6
6 Leveling Foot Mount 2
7 Leveling Foot 2
Models: 625 (-001 /-003 /-005 /-006) 8 Base Weldment 1

Serial Numbers: 9 Hand / Foot Control Harness 2


V1667150 thru Present
10 Bracket 1

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] E-21.3 Always verify model & serial number
1 029-5391-00

2 040-0250-157
5
5 053-0050-02
3 020-0364-00-312

4 040-0312-25 6 LH: 020-0365-01-312


RH: 020-0365-00-312

12 015-3464-00
2
7 030-1938-00-312
625 (-004) Base Components
Item Description Qty.
1 Bubble Level Plug 1
8 015-2469-00
2 Hex Screw (1/4-20 x 1/2”) 11
3 Front Base Casting 1
050-7631-00-312 11 4 Hex Screw (5/16-18 x 5/8”) 8
5 Hole Plug 4
9 029-5612-00 6 Back Base Cover - LH shown 2
015-1784-00 10 7 Base Weldment 1
MA8274 8 Load Cell 4
9 Leveling Feet Set (includes four leveling feet) 1
Models: 625-004 10 Foot Control Harness 1
11 Bracket 1
Serial Numbers: V2200 thru V1667149
12 Hand Control Harness 1

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] E-22 Always verify model & serial number
1 029-5391-00

2 040-0250-157
5
5 053-0050-02
3 020-0364-00-312

4 040-0312-25 6 LH: 020-0365-01-312


RH: 020-0365-00-312

2
10 015-3464-00
7 030-1938-00-312 Base Components
Item Description Qty.
1 Bubble Level Plug 1
2 Hex Screw (1/4-20 x 1/2”) 11
050-7631-00-312 11 8 015-2469-00 3 Front Base Casting 1
4 Hex Screw (5/16-18 x 5/8”) 8
5 Hole Plug 4
9 029-5612-00 6 Back Base Cover - LH shown 2
7 Base Weldment 1
8 Load Cell 4
9 Leveling Feet Set (includes four leveling feet) 1
Models: 625-004 10 Hand / Foot Control Harness 2
Serial Numbers: V1667150 thru Present 11 Bracket 1

© Midmark Corporation 2009 SF-1935 [Revised: 5/13/16] E-22.1 Always verify model & serial number
1 040-0010-47

2 050-6013-00

Attention:
3 002-1274-00 PC board is not compatible with ALL units.
ALWAYS VERIFY MODEL & SERIAL NUMBER.

4 015-2622-00

5 a) 015-1829-00
b) 015-1604-00

1
1 6 015-1259-03

053-1664-00-000 11
7 015-0639-00

053-0068-06 10 8 015-1338-00
*See Note

Electrical Components
Item Description Qty.
002-1394-00 9 1 Screw (1/4-20 x 1/2”) 8
2 Cover 1
3 PC Board 1
4 Power Line Filter (Domestic models only) 1
a) Isolation Transformer (Domestic models)
5 1
b) Transformer (Export models)
6 Fuse Holder 2
7 Power Cord Inlet 1
MA8102i
Power Cord (N. American, 120V, 8 ft)
8 *Note: Export models require accessory: 9A152 1
International Cordsets.
Models: 625 (-001 /-003) 9 Housing (includes nutserts) 1
Serial Numbers: V2200 thru V968527 10 Snap Bushing 3
11 Standoff 6

© Midmark Corporation 2009 SF-1935 [Revised: 4/15/13] E-23 Always verify model & serial number
1 040-0010-47

2 050-6013-00

Attention:
3 002-1422-00 PC board is not compatible with ALL units.
ALWAYS VERIFY MODEL & SERIAL NUMBER.

4 015-2622-00

5 a) 015-1829-00
b) 015-1604-00

1
1 6 015-1259-03

053-1664-00-000 11
7 015-0639-00

053-0068-06 10 8 015-1338-00
*See Note

Electrical Components
Item Description Qty.
002-1394-00 9 1 Screw (1/4-20 x 1/2”) 8
2 Cover 1
3 PC Board 1
4 Power Line Filter (Domesic models only) 1
a) Isolation Transformer (Domestic models)
5 b) Transformer (Export models) 1

6 Fuse Holder 2
7 Power Cord Inlet 1
MA8102i
Power Cord (N. American, 120V, 8 ft)
8 *Note: Export models require accessory: 9A152 1
International Cordsets.
Models: 625 (-001 /-003) 9 Housing (includes nutserts) 1
Serial Numbers: V968528 thru V1266165 10 Snap Bushing 3
11 Standoff 6

© Midmark Corporation 2009 SF-1935 [Revised: 4/15/13] E-23.1 Always verify model & serial number
1 040-0010-47

2 050-6013-00

3 002-1610-00 Attention:
Contact Midmark Technical Service before ordering
any PC board for 625-004 units with serial numbers
V1179275 thru V1260356

4 015-2622-00

5 a) 015-1829-00
b) 015-1604-00

1
1 6 015-1259-03

053-1664-00-000 11
7 015-0639-00

053-0068-06 10 8 015-1338-00
*See Note

Electrical Components
Item Description Qty.
002-1394-00 9 1 Screw (1/4-20 x 1/2”) 8
2 Cover 1
3 PC Board 1
4 Power Line Filter (Domesic models only) 1
a) Isolation Transformer (Domestic models)
5 b) Transformer (Export models) 1

6 Fuse Holder 2
7 Power Cord Inlet 1
MA8102i
Power Cord (N. American, 120V, 8 ft)
8 *Note: Export models require accessory: 9A152 1
International Cordsets.
Models: 625 (-001 /-003) 625-004 9 Housing (includes nutserts) 1
Serial Numbers: V1266166 thru present all 10 Snap Bushing 3
11 Standoff 6

© Midmark Corporation 2009 SF-1935 [Revised: 9/14/12] E-23.2 Always verify model & serial number
2 050-7905-00
040-0010-145 1

3 040-0010-129
002-1513-00 4

5 002-1516-00

6 053-1664-00-000

6
7 002-1512-00

8 050-7904-00

053-0068-06 11

Attention:
Contact Midmark Technical Service before ordering
any PC board for 625-004 units with serial numbers
V1179275 thru V1260356
002-1515-00 10

002-1514-00 9 625 (-004) Scale / USB Boards


Item Description Qty.
1 Screw (#24 x 1/2” (Sink)) 2
2 Cover 1
3 Screw (#24 x 1/2” (MAG525B)) 1
4 LIN to VSM PC Board (includes standoffs) 1
5 USB HUB PC Board (includes standoffs) 1
6 Standoff AR
7 Scale PC Board (includes standoffs) 1
8 Housing 1
Models: 625-004 9 625 to USB PC Board (includes standoffs) 1
Serial Numbers: all 10 LIN to Scale PC Board (includes standoffs) 1
11 Snap Bushing 3

© Midmark Corporation 2009 SF-1935 [Revised: 11/11/15] E-24 Always verify model & serial number
002-1300-04 1 2 053-1903-01-312

3 015-2389-01

4 053-1905-00-312

5 053-1904-00-312

Note
Refer to “Wireless Control Components”
if applicable.
6 050-7256-00-312
015-1675-04 9

053-1439-00 8 7 042-0168-00 Foot Control Assembly


Item Description Qty.
1 Foot Control Assembly (incl. items 2 thru 9) 1
2 • Keypad 1
MA8056-1i 3 • Switch Membrane 1
4 • Retainer 1
5 • Base 1
6 • Baseplate 1
Models: 625 (-001 /-005 /-006) 7 • Screw 10
Serial Numbers: all 8 • Stem Bumper 7
9 • Cord 1

© Midmark Corporation 2009 SF-1935 [Revised: 10/17/11] E-25 Always verify model & serial number
1 002-0911-07

2 015-2391-00

3 053-1381-00

9 029-6031-00-216

4 015-1675-04
050-5868-00 7

Note
Refer to “Wireless Control Components”
if applicable.

10 040-0010-47
5 053-1382-00

Hand Control Assembly


Item Description Qty.
1 Hand Control Assembly (incl. items 2 thru 7) 1

002-1473-00 8 2 • Switch Membrane 1


3 • Housing - Top 1
6 042-0168-00 4 • Cord 1
5 • Housing - Bottom 1
6 • Screw (ATF PT K35 x 10 cross recess pan hd.) 4
7 • Weight 1
Models: 625 (-001 /-005 /-006) 8 Hand Control Holster 1
Serial Numbers: all 9 Hand Control Bracket Assembly (incl. item 10) 1
10 • Screw (#10-24 x 3/8” pan head phillips) 2

© Midmark Corporation 2009 SF-1935 [Revised: 6/6/17] E-26 Always verify model & serial number
2
1

053-0068-00 9
3 002-1564-00

015-2131-00 8

015-2177-00 4

5 050-7644-00-216

040-0010-47 7

015-0013-00 6

Wireless Control Components


8 Item Description Qty.

MA8451i 1 Refer to: Wireless Hand Control 1


2 Refer to: Wireless Foot Control 1
3 Base Station PC Board 1
4 Batteries (size: AA) 4
5 Antenna Bracket 1
6 Cable Tie 1
Models: 625-003 7 Screw (#10-24 x 3/8*) 1
Serial Numbers: V968528 thru V1266165 8 Antenna (includes wire, lockwasher, & nut) 1
9 Snap Bushing 1

© Midmark Corporation 2009 SF-1935 [Revised: 10/23/13] E-27 Always verify model & serial number
2
1 3

053-0068-00 10
4 002-1612-00

015-2131-00 9 Attention:
Contact Midmark Technical Service before ordering
any PC board for 625-004 units with serial numbers
V1179275 thru V1260356
015-2177-00 5

6 050-7644-00-216

040-0010-47 8

015-0013-00 7

Wireless Control Components


Item Description Qty.
9 MA8451.1i
1 Refer to: Wireless Hand Control 1
2 Refer to: Wireless Foot Control 1
3 Refer to: Wireless Hand Control (w/ Scale) 1
4 Base Station PC Board 1
5 Batteries (size: AA) 4
6 Antenna Bracket 1
7 Cable Tie 1
Models: 625-003 625-004 8 Screw (#10-24 x 3/8*) 1
Serial Numbers: V1266166 thru present all 9 Antenna (includes wire, lockwasher, & nut) 1
10 Snap Bushing 1

© Midmark Corporation 2009 SF-1935 [Revised: 10/23/13] E-27.1 Always verify model & serial number
002-1290-07 1

Models: 625 (-003 /-004) Wireless Foot Control Assembly


Item Description Qty.
Serial Numbers: all
1 Foot Control Assembly 1

© Midmark Corporation 2009 SF-1935 [Revised: 5/24/13] E-28 Always verify model & serial number
002-1076-02 1

Models: 625 -003 Wireless Hand Control Assembly


Item Description Qty.
Serial Numbers: all
1 Hand Control Assembly 1

© Midmark Corporation 2009 SF-1935 [Revised: 5/24/13] E-29 Always verify model & serial number
1 002-1076-03

Models: 625-004 Wireless Hand Control w/ Scale Assembly


Item Description Qty.
Serial Numbers: all
1 Hand Control Assembly 1

© Midmark Corporation 2009 SF-1935 [Revised: 5/24/13] E-30 Always verify model & serial number
1

Fuse: Location: Function affected Rating Midmark Part Number

6.3A, 250V,
Isolation Transformer IEC Inlet Electrical Receptacle 015-0346-20
Slo-Blo, 5 x 20 mm
Fuses (2)

10A, 250V,
Primary Fuses (2) Main PC Board [F1 & F2] Base & Back Functions 015-0346-42
Slo-Blo, 5 x 20 mm

160mA, 250V,
Transformer Fuse Main PC Board [F3] Base & Back Functions 015-0346-38
Slo-Blo, 5 x 20 mm

6.3A, 250V,
Base & Back Motor Fuse Main PC Board [F4] Base & Back Functions 015-0346-40
Slo-Blo, 5 x 20 mm

800mA, 250V,
Drawer Heater Fuses (2) Distribution Board Drawer Heater 015-0346-25
Slo-Blo, 5 x 20 mm

Transformer 230V / 115V 6.3A, 250V,


IEC Inlet All 015-0346-20
(Export models only) Slo-Blo, 5 x 20 mm

Models: 625 Fuses


Serial Numbers: Item Description Qty.
all
1 Refer to chart for detailed descriptions -

© Midmark Corporation 2009 SF-1935 [Revised: 1/13/11] E-31 Always verify model & serial number
6 040-0010-47

7 053-1528-00-216 1 a) 002-1935-00-271
b) 002-1935-02-271
8 a) 015-0083-02
b) 015-2183-00

015-1596-00 5

2 042-0184-01

Table Receptacle System


Item Description Qty.
053-1529-00-271 4 Table Receptacle Assy (includes items 2 thru 8)
1 a) Standard Receptacle (black) 1
b) Tamper Resistant Receptacle (white)

040-0006-23 3 MA716100i
2 • Screw (.138-32 x 1/2”) 4
3 • Screw (#6-32 x 1/2”) 1
4 • Receptacle Cover 1
5 • Circuit Breaker (5 amp) 2
6 • Screw (#10-24 x 3/8”) 1
Models: 625 (-001 /-003) 7 • Circuit Breaker Housing 1
• Duplex Receptacle:
Serial Numbers: V2200 thru V1149713 8 a) Standard (black) 1
b) Tamper Resistant (white)

© Midmark Corporation 2009 SF-1935 [Revised: 4/30/15] E-32 Always verify model & serial number
8 040-0010-47
015-1596-00 9
7 053-1528-00-216 1 a) 029-5453-00
b) 029-5453-01
6 a) 015-0083-02
b) 015-2183-00
2 a) 002-1935-00-271
b) 002-1935-02-271

053-0050-19 10

3 042-0184-01

4 a) 053-1529-00-271
b) 053-1529-01-271

Table Receptacle System


Item Description Qty.
Receptacle Assy R.H.(includes items 3 thru 8 & 10)
1 a) Standard Receptacle (black) 1
b) Tamper Resistant Receptacle (white)
Receptacle Assy L.H.(includes items 3 thru 9)
2 a) Standard Receptacle (black) 1
b) Tamper Resistant Receptacle (white)
3 • Screw (.138-32 x 1/2”) 8
• Receptacle Cover
4 a) Standard Receptacle (black) 2
040-0006-23 5 MA716101i b) Tamper Resistant Receptacle (white)
5 • Screw (#6-32 x 1/2”) 2
• Duplex Receptacle:
6 a) Standard (black) 2
b) Tamper Resistant (white)
7 • Circuit Breaker Housing 2
Models: 625 (-001 /-003) 8 • Screw (#10-24 x 3/8”) 2
Serial Numbers: V1149714 thru Present 9 • Circuit Breaker (5 amp) 2
10 • Hole Plug 2

© Midmark Corporation 2009 SF-1935 [Revised: 4/30/15] E-32.1 Always verify model & serial number
8 040-0010-47
015-1596-00 7
9 053-1528-00-216
1 a) 029-5400-00 LH
029-5400-01 RH
b) 029-6282-00 LH
10 a) 015-2940-00 029-6282-01 RH
b) 015-3362-00

053-0050-19 6

2 042-0184-01

625 (-004)Table Receptacle System


Item Description Qty.
015-2924-00 5 Receptacle Assy (includes items 2 thru 10)
1 a) Standard (black) 2
b) Tamper Resistant (white)
053-2278-00-271 4 2 • Screw (.138-32 x 1/2”) 8
3 • Screw (#6-32 x 1/2”) 4
4 • Receptacle Cover 2
040-0006-23 3 5 • Serial Coupler Port 2
6 • Hole Plug 2
7 • Circuit Breaker (5 amp) 2
8 • Screw (#10-24 x 3/8”) 2
Models: 625-004 9 • Circuit Breaker Housing 2
• Simplex Receptacle
Serial Numbers: all 10 a) Standard (black) 2
b) Tamper Resistant (white)

© Midmark Corporation 2009 SF-1935 [Revised: 4/17/15] E-33 Always verify model & serial number
015-1593-00 2 3 015-0346-25

4 053-1664-00-000

015-1759-00 9 1 029-4537-00

5 029-4716-00
041-0010-12 8

053-0858-02 7

040-0010-144 6 Drawer Heater System


Item Description Qty.
1 Drawer Heater Assy. (includes items 2 thru 11) 1
2 • Distribution Board (includes item 3) 1
3 •• Fuse (800mA, 250V, Slo-Blo, 5 x 20mm) 2
4 • PC Board Standoff 4
5 • Heater Plate Assembly (incl. items 6 thru 8) 1
6 •• Screw (10-32 x 1/2”) 4
Models: 625 (-001 /-003 /-004 /-006) 7 •• Nylon Washer 4
Serial Numbers: all 8 •• Nylon Nut 4
9 • Drawer Heater Switch 1

© Midmark Corporation 2009 SF-1935 [Revised: 04/10] E-34 Always verify model & serial number
SUBJECT TO CHANGE WITHOUT NOTICE

Midmark Corporation
For contact information, go to:
www.midmark.com

You might also like