Analytical Method For Cleaning Validation of Levothyroxine Sodium in

Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

Acta Chimica and Pharmaceutica Indica

Research | Vol 8 Issue 1

Analytical Method for Cleaning Validation of Levothyroxine Sodium in


Production Area
Md Didarul Islam1*, TM Mohiuddin2, Md Mynul Hassan3, Asheful Latif4, M Mehedi Hasan1 and Dr Papia
Haque1
1
Applied Chemistry and Chemical Engineering, University of Dhaka, Dhaka, Bangladesh
2
Biotechnology and Genetic Engineering, Islamic University, Kushtia, Bangladesh
3
Department of Biotechnology, Bangladesh Agricultural University, Mymensingh, Bangladesh
4
Pharmacy, State University of Bangladesh, Dhaka, Bangladesh
*
Corresponding author: Mohammed Didarul Islam, Applied Chemistry and Chemical Engineering, University of Dhaka,
Dhaka, Bangladesh, E-Mail: [email protected]
Received: February 03, 2018; Accepted: March 07, 2018; Published: March 28, 2018

Abstract
The aim of this present study is to explore a rapid, selective, and sensitive HPLC method for cleaning validation for the analysis of
levothyroxine sodium from production area. Chromatographic separation of levothyroxine sodium was achieved at 6.73 min by using 40: 60
volume of acetonitrile and deionized water that contain 0.05% of orthophosphoric acid mobile phase at a flow rate of 1.5 mL/min and
detection UV wavelength of 225 nm. The method was validated according to USP Category I requirements. Linearity for detector response
was observed in the concentration range of 0.0045–27ppm and coefficient of determination (r2) for calibration curve was found exactly
0.9999 (r2 > 0.99). Percent recovery was found from 83.15 to 83.80% ranging from 0.02265 to 006795mg of L-T4. Precision and
intermediate precision showed that the % RSD of both were 2.80 and 0.98 respectively which were in the range of acceptable limit (< 10) for
cleaning validation. Limit of detection (LOD) and limit of quantification (LOQ) were obtained at 0.002 and 0.006 ppm respectively. The
HPLC method was successfully applied for analysis of levothyroxine sodium samples from different parts of equipment’s in production
area.

Keywords: Levothyroxine sodium; Cleaning validation; HPLC; Limit of detection; Limit of quantification

Introduction

Thyroid stimulating hormone (TSH) is secreted by the pituitary gland which is stimulated by hypothalamic TSH-releasing
hormone (TRH) and inhibited by high serum levels of two hormones produced by the thyroid. This two hormones are tri-
iodothyronine (T3) and tetra-iodothyronine (T4, or thyroxine), which contain three and four atoms of iodine respectively [1].
Both T3 and T4 have two enantiomers of them L-enantiomers (L-T3 and L-T4) are responsible for their biologic effects [1].
Deiodinase enzymes convert T4 to T3, which is a major source of T3 (85%) in peripheral tissues [2]. These two hormones
control energy metabolism, protein synthesis, and body's sensitivity to other hormones. Cells of the fetal brain are a major
target for action of T3 and T4 that play an important role in brain maturation [3-6]. Lower level of TSH hormone may causes
hypothyroidism, one of the most common endocrine disorders occurring in up to 5% of the United States and the United

Citation: Islam DMD, Mohiuddin MT, Mynul Hassan MMD, et al. Analytical Method for Cleaning Validation of Levothyroxine Sodium
in Production Area. Acta Chim Pharm Indica, 2018; 8(1):125
© 2018 Trade Science Inc.
www.tsijournals.com | March-2018

Kingdom [7-9]. In adults, the incidence of hypothyroidism is estimated to be 3.5 per 1000 per year in women and 0.6 per
1000 per year in men [10]. Most patients with the disease require life time thyroxine replacement for normalizing thyroid
hormone levels in peripheral tissues. Several oral administrated FDA approved T4 tablets are now available such as,
Unithroid®, Synthroid®, L-Thyroxine®, Levoxyl® and Levothyrox® [7]. In last few years some alternative L-T4
formulations have become available soft gel capsule and liquid solution are some of them. Pharmacokinetic studies revealed
that the latter formulation process have more bioavailability compared to classical L-T4 tablet [11]. Softgel formulation
process has some advantages like higher bioequivalent rate and lower exposure in the production environment [12].
Levothyroxine sodium (L-3, 5, 3, 5-tetraiodothyronine sodium salt) pentahydrate (FIG. 1) is a salt of levoisomer of thyroxine
(FIG. 1) and serves as a replacement therapy for the inadequate secretion of T4 in the body [13]. The initial daily dose to
treat hypothyroidism is approximately 1.7 µg/kg, or 100–125 mcg for a 70 kg adult [11]. Daily uptake of levothyroxine
sodium should be controlled strictly otherwise it may cause hyperthyroidism and thyrotoxicosis. Hyperthyroidism refers to
hyper function of the thyroid gland, whereas thyrotoxicosis refers to any state characterized by thyroid hormone excess [9].
The accurate determination of levothyroxine from the environment of production area is a major concern. There are various
analytical methods cited in the literature used for the quantitative determination for levothyroxine like isotope dilution
tandem mass spectroscopy [14], high performance liquid
d chromatography [13-17], inductively coupled plasma (ICP) mass spectrometry [18,19], liquid chromatography/tandem
mass spectrometry (LC-MS/MS) [20] and liquid chromatography using electrochemical and MS detection [16]. This study is
based on to develop a rapid, selective and sensitive HPLC method to accurately and precisely determine L-T4 from the
production area at a very low concentration.
FIG. 1. Structure of (a) thyroxine and (b) levothyroxine sodium pentahydrate.

Materials and Methods


Materials
L-Thyroxine sodium (L-T4) certified reference standard was purchased from the Excella GmbH & Co (Germany).
Chromafil® Xtra PTFE 0.45 µm syringe filters were purchased from the Pall Corporation (Ann Arbor, MI, USA). HPLC
grade acetonitrile, analytical grade phosphoric acid and Sodium Hydroxide were purchased from Fisher Scientific (Fairlawn,
NJ, USA). HPLC ready deionized 18Milli-Q water was obtained, in-house, from a Milli-Q Gradient A-10 water purification
system, Millipore, (Bedford, MA, USA).

2
www.tsijournals.com | March-2018

Determination of the binding of PB to HSA in the absence and presence of endogenous substances
The binding of PB to HSA in the absence and presence of endogenous substances was determined using an ultrafiltration
technique. Ultrafiltration was carried out using an AmiconR Ultra-0.5 mL centrifuge filter unit with an UltracelR-30
membrane (Merck Millipore Company, MA). Samples of 500 µL were centrifuged at 2,500 × g at 25 °C for 5 min. The
concentration of unbound PB in the filtrates was then determined by HPLC.
Recovery calculation
The amount of levothyroxine sodium in the test samples was calculated, as quantity and percent, from the measured peak area
response for the test samples (AU) and compared to peak area response (As) for the standard levothyroxine solution using the
following equations:

Where C is the concentration in ppm of the USP levothyroxine working standard.


Instrumentation and chromatographic conditions
Prominence I HPLC (Shimadzu Corporation, Japan) consisted of a quaternary pump, an automatic injector, variable
wavelength detector, and a column oven. Data were processed by using Lab solution 6.82-ST1 software. Separation was
achieved on a Kromasil L10 column (4.6 mm × 250 mm, 5 µm) and ProntoSIL L10 column (4.6 mm × 250 mm, 5 µm) was
used for intermediate precision analysis. The flow rate was 1.5 mL/min. The chromatographic conditions: Filtered and
degassed 40 volume of acetonitrile and 60 volume of deionized water that contain 0.5ml of orthophosphoric acid were used
as mobile phase and diluent. 10 min as equilibration time. The column temperature was controlled at 25°C and the injection
volume was 100µL. The UV detection wavelength was 225 nm.
Stock solution preparation (45 ppm)
45 mg of levothyroxine sodium working standard was transferred in 1000 ml amber volumetric flasks add 5 ml 1 M NaOH
and 20 ml diluent and sonicate until dissolve. Allow to cool and volume up to the mark with same diluent.
Results and Discussion
The method was validated according to the United States Pharmacopeia Category I requirements. The following validation
characteristics were addressed: specificity, accuracy, precision, linearity, range, LOD and LOQ.
System suitability standard
System suitability standard solution was prepared daily using stock solution, for that purpose 10 ml of stock solution was
transferred in to 50 ml amber volumetric flasks and dilute up to the mark with diluent. System suitability was determined
from five replicate injections of the stock solution before sample analysis. The acceptance criteria were less than 2% relative
standard deviation (RSD) for peak area, greater than 2000 theoretical plates, USP tailing factor less than 2. All critical

3
www.tsijournals.com | March-2018

parameters were tested before sample run and it was found that all parameter met the acceptable criteria throughout all days
which is shown in the TABLE 1.
TAB.1. System Suitability Test Results (n=5).
Parameter Specifications Observed results
Analyst 01 Analyst 02 Analyst 03
Retention Time (% RSD) ≤ 2.0 0.08 0.02 0.09
Area (% RSD) ≤ 2.0 0.09 0.19 0.67
Tailing Factor ≤ 2.0 1.51 1.44 1.52
Theoretical plates ≥ 2000 7135 ± 64 7996 ± 75 7222 ± 92
n: number of replicates per concentration levels and per series.

Specificity
Specificity of the method was determined by comparing the system suitability standard and diluent of the sample, examined
active component specifically and accurately measured without any interference of diluent peaks. The acceptance criteria
were peak of active should be pure that means purity index should be higher than purity threshold and diluent should not
show any interfere at the retention time (6.73 min) of Levothyroxine Sodium peak at LOD and LOQ concentration.
Specificity was established by determining that peak of levothyroxine sodium standard and diluent were completely
segregated FIG. 2 and 3.
FIG. 2. Specificity of levothyroxine sodium from other peaks of diluent.

FIG. 3. Peak comparison of LOD and LOQ from diluent.

4
www.tsijournals.com | March-2018

Precision and intermediate precision


The precision of test method was evaluated by analyzing six swab sample solutions which were prepared and analyzed on
same day by HPLC. 1.0 ml (45 ppm) of levothyroxine stock solution was spread out on the 10 × 10 cm surface area of a 316
grade SS plate, and dried it at ambient temperature. Dried sample was collected by using swab stick which was initially rinse
into diluent. Swab stick was transferred in test tube, 5 ml diluent was added and then sample was sonicate for 10 minutes.
Total six (6) replicate samples were prepared by the same procedure. The method of precision was established by injecting
six replicate samples after running five standard samples. Intermediate precision was determined by applying same procedure
with different analyst, HPLC and column. Critical value and results of precision and intermediate precision were summarized
in TABLE 2. From the TABLE 2 it was found that all parameter of cleaning validation were in permeable limit.
TABLE. 2. Precision and intermediate precision Results (n=6).
Parameter Specifications Observed results
Precision Intermediate precision
Area of Sample - 2157021 ± 2.80 1892832 ± 0.98
Tailing Factor ≤ 2.0 1.50 1.44
Theoretical plates ≥ 2000 7215 ± 79 6521 ± 56
n: number of replicates per concentration levels and per series.

Recovery or accuracy
Accuracy Study was performed by spreading Levothyroxine Sodium solution on the 10x10 cm surface area of a 316 grade SS
plate ranging from 50 to 150% (50%, 80%, 100%, 120% and 150%) which were prepared from stock solution. For this
purpose 0.5, 0.8, 1, 1.2 and 1.5 ml of stock solution was spread in to the SS plate and then dried at ambient temperature. After
complete dry sample were collected through swab stick which were initially rinse with diluent. Swab stick was transferred in
test tube, 5 ml diluent was added and then sample was sonicate for 10 minutes. For cleaning validation recovery should be
higher than 80%, and RSD of the each sample should be less than 10% [21]. Results for recovery of levothyroxine sodium

5
www.tsijournals.com | March-2018

from SS plate by swab stick and there specifications were summarized in TABLE 3. Recovery factor was 1.20 that means
this cleaning procedure was satisfactory for accurately analysis of levothyroxine sodium from production area.
Linearity and range
Standard calibration curves were prepared with sixteen calibrators over a concentration range of 0.006-27 ppm (0.006, 0.009,
0.018, 0.027, 0.054, 0.09, 0.45, 0.9, 2.25, 4.5, 7.2, 9, 10.8, 13.5, 18 and 27 ppm) for levothyroxine. The data of peak area
versus active concentration were treated by linear least square regression analysis (TABLE 4). Coefficient of determination
for cleaning validation sample should be within 0.99 to 1.0. Data from the FIG. 4 complies that coefficient of determination
value was 0.9999.
LOD and LOQ
LOD means the lowest concentration of the standard solution (six replicate samples) that can be detected but not necessarily
quantitated under stated experimental conditions [22,23]. ICH and USP states that signal-to-noise ratio should be ≥3 and
RSD of six replicates should be within 10 to 33 [24-27]. Six replicate sample concentration of 0.002 ppm was used to
examine the LOD concentration. At that concentration S/N was found 4.11 and RSD was 12.51 TABLE 5.
LOQ means the lowest concentration of the standard solution (six replicate samples) that can be quantitatively determined
with suitable precision and accuracy [22]. In that case signal to noise ratio should be ≥10 and RSD of six replicates should be
≤ 10 [22, 26, 27]. Six replicate sample concentration of 0.006 ppm was used to examine the LOQ concentration. At that
concentration S/N was founded 11.19 and RSD was 2.37 TABLE 5 that means at that concentration levothyroxine sodium
quantitatively determined precisely.

FIG. 4. Linearity for levothyroxine sodium.

TABLE 3. Accuracy Results (n=6).

Parameter Specifications 50% 80% 100% 120% 150%

Area of Sample - 927047 ± 0.19 1503889 ± 0.23 1852183 ±1.69 2155581±0.43 2750230 ± 0.20

6
www.tsijournals.com | March-2018

Declared
- 0.0225 0.036 0.045 0.054 0.0675
amount (mg)

Recovery (mg) - 0.01887 ± 0.19 0.0306 ± 0.23 0.03769 ± 1.69 0.04385 ± 0.43 0.05731 ± 0.20

Recovery (%) ≥ 80 83.85 85 83.76 81.2 82.9

n: number of replicates per concentration levels and per series.

TABLE 4. LOD and LOQ results (n=6).


Parameter LOD LOQ
Critical value Observed value Critical value Observed value
Concentration - 0.002 ppm - 0.006 ppm
Signal to noise ratio Close to 3 4.11 Close to 10 11.19
% RSD 10.0-33.0 12.51 ≤10.0 2.37
n: number of replicates per concentration levels and per series.

Determination of levothyroxine sodium in production area


The validated method was successfully applied for the determination of levothyroxine sodium from different parts of the
production environment. In production area after cleaning with IPA levothyroxine sodium was found in very low
concentration which is shown in the TABLE 5. Maximum levothyroxine sodium was found in internal surface of the
granulation and it was 2.69 ppm, which was within limit of cleaning process validation (less than 9 ppm).
TABLE 5. Swab samples from different parts of production area.
Sample ID Concentration (ppm)
Inside the bin of Blister 0.13
Feeding disk of Blister 0.05
Blister hopper 1.27
Bin lid of Granulator 0.38
Internal surface of Granulator 2.69
Surface gate of Compressor 0.14
Inner O Matic of Compressor BDL
Inner chute of Compressor 0.25

7
www.tsijournals.com | March-2018

Window flap 0.09


Metal detector BDL
Surface disk of deduster 0.04
BDL: Below detection limit.

Conclusion
A simple and efficient HPLC method was developed and validated for levothyroxine sodium. The method addressed each of
the analytical validation characteristi Determination of levothyroxine sodium in production area cs such as accuracy,
precision, specificity, linearity, LOD and LOQ met the USP acceptance criteria. The usefulness of this method is
demonstrated by successful application for the analysis of levothyroxine sodium in different parts of production area.
.
REFERENCES
1. Biondi B, Filetti S, Schlumberger M. Thyroid-hormone therapy and thyroid cancer:a reassessment. Nat Clin Pract
Endocrinol Metab, 2005;1:32.
2. Braverman LE, Ingbar SH, Sterling K. Conversion of thyroxine (T4) to triiodothyronine (T3) in athyreotic human
subjects. J Clin Invest, 1970. 49:855.
3. Kester MH, Martinez de Mena R, Obregon MJ, et al. Iodothyronine levels in the human developing brain:major
regulatory roles of iodothyronine deiodinases in different areas. J Clin Endocrinol Metab, 2004;89:3117-3128.
4. Fallahi P, Ferrari SM, Ruffilli I, et al. Advancements in the treatment of hypothyroidism with L-T4 liquid
formulation or soft gel capsule:an update. Expert Opin Drug Deliv, 2017;14:647-655.
5. Gika H, Lämmerhofer M, Papadoyannis I, et al. Direct separation and quantitative analysis of thyroxine and
triiodothyronine enantiomers in pharmaceuticals by high-performance liquid chromatography. J Chromatogr B
Analyt Technol Biomed Life Sci, 2004;800:193-201.
6. Patel H, Stalcup A, Dansereau R, et al. The effect of excipients on the stability of levothyroxine sodium
pentahydrate tablets. Int J Pharm, 2003;264:35-43.
7. Agu RU, Mactavish J, Yeung KP, et al, Thyroid hormone (levothyroxine) replacement via the respiratory route by
inhalation:in vitro exploratory studies. Expert Opin Drug Deliv, 2016;13:195-205.
8. Clyde PW, Harari AE, Getka EJ, et al, Combined levothyroxine plus liothyronine compared with levothyroxine
alone in primary hypothyroidism:a randomized controlled trial. JAMA, 2003;290:2952-2958.
9. Woeber KA. Update on the management of hyperthyroidism and hypothyroidism. Arch Intern Med,
2000;160:1067-1071.
10. Espaillat R, Jarvis MF, Torkelson C, et al. Gluten and Aluminum Content in Synthroid®(levothyroxine sodium
tablets). Adv Ther, 2017;34:1764-1769.
11. Yue CS, Scarsi C, Ducharme MP. Pharmacokinetics and potential advantages of a new oral solution of
levothyroxine vs. other available dosage forms. Arzneimittelforschung, 2012;62:631-636.

8
www.tsijournals.com | March-2018

12. Colucci P, D'Angelo P, Mautone G, et al. Pharmacokinetic equivalence of a levothyroxine sodium soft capsule
manufactured using the new food and drug administration potency guidelines in healthy volunteers under fasting
conditions. Ther Drug Monit, 2011;33:355-361.
13. Collier JW, Shah RB, Bryant AR, et al. Development and application of a validated HPLC method for the analysis
of dissolution samples of levothyroxine sodium drug products. J Pharm Biomed Anal, 2011;54:433-438.
14. Gu J, Soldin OP, Soldin SJ. Simultaneous quantification of free triiodothyronine and free thyroxine by isotope
dilution tandem mass spectrometry. Clin Biochem, 2007;40:1386-1391.
15. Dalal PS, Albuquerque P, Bhagat HR, Development and standardization of levothyroxine analysis by high-
performance capillary electrophoresis. Anal Biochem, 1993;211:34-36.
16. Kazemifard AG, Moore DE, Aghazadeh A, Identification and quantitation of sodium-thyroxine and its degradation
products by LC using electrochemical and MS detection. J Pharm Biomed Anal, 2001;25:697-711.
17. Shah RB, Bryant A, Collier J, et al. Stability indicating validated HPLC method for quantification of
levothyroxine with eight degradation peaks in the presence of excipients. Int J Pharm, 2008;360:77-82.
18. Pabla D, Akhlaghi F, Ahmed A, et al. Development and validation of an inductively coupled plasma mass
spectrometry method for quantification of levothyroxine in dissolution studies. Rapid Commun Mass Spectrom,
2008;22:993-996.
19. Pabla D, Akhlaghi F, Zia HA. Comparative pH-dissolution profile study of selected commercial levothyroxine
products using inductively coupled plasma mass spectrometry. Eur J Pharm Biopharm, 2009;72:105-110.
20. Piehl S, Thomas H, Gabor B, et al. Development of a validated liquid chromatography/tandem mass spectrometry
method for the distinction of thyronine and thyronamine constitutional isomers and for the identification of new
deiodinase substrates. Rapid Commun Mass Spectrom, 2008;22:3286-3296.
21. World Health Organization. Supplementary guidelines on good manufacturing practices (GMP):Validation. World
Health Organization, 2003. Geneva.
22. Shabir GA. Step-by-step analytical methods validation and protocol in the quality system compliance industry. J
Valid Technol, 2005;10:314-325.
23. Shrivastava A, Gupta V. Methods for the determination of limit of detection and limit of quantitation of the
analytical methods. Chronicles of Young Scientists, 2011;2:21.
24. Shabir GA. Validation of high-performance liquid chromatography methods for pharmaceutical analysis:
Understanding the differences and similarities between validation requirements of the US Food and Drug
Administration, the US Pharmacopeia and the International Conference on Harmonization. J Chromatogr A,
2003;987:57-66.
25. IHT Guideline. Validation of analytical procedures: text and methodology. Q2 (R1), 2005.
26. Peters FT, Drummer OH, Musshoff F. Validation of new methods. Forensic Sci Int, 2007;165:216-224.
27. Kollipara S, Bende G, Agarwal N, et al. International guidelines for bioanalytical method validation:a comparison
and discussion on current scenario. Chromatographia, 2011;73:201-217.

You might also like