Precision Engineering & Metrology
Precision Engineering & Metrology
[ME3312]
BOOK
Everyday measurements
A measurement system is a tool used for quantifying the measured variable by extend the abilities of the
human senses.
A measurement system is composed of components that work together to accomplish a specific objective.
Example
Sensor and
Measurement
Sensor Stage Transducer Output Stage
System
Stage
MEASUREMENT SYSTEM (ATOMIC FORCE MICROSCOPE)
1. Sensor Stage
Variable Measurement:
Height of the surface
MEASUREMENT SYSTEM (ATOMIC FORCE MICROSCOPE)
3. Output Stage
Examples
Example
General Stage:
Sensor-transducer stage
Output Stage
Feedback-control stage
Filtering Signal-Conditioning
Stage
Linkages
MEASUREMENT SYSTEM
CHOOSING APPROPRIATE MEASURING INSTRUMENTS
An actuator is a transducer that converts an electrical signal to another form of energy such as mechanical
energy.
GOOD SENSOR CHARACTERISTICS
Should not
Influence the measurement
Should
Have output that deviates randomly over
Sensitive to only the measured property
time
Have sensitivity adequate for measurement
Consume high power
Be linear
High Cost
Be huge
Be toxic
SENSOR
CHARACTERISTICS
Calibration
It can be performed by placing thermometer in
ice water tab and temperature sensor in it.
Note down the reading of thermometer and
voltage from thermocouple.
Plot a graph and develop a calibration equation.
Linearity
“Property of calibrated equation being a straight line.”
Linearity makes the sensitivity of sensor constant throughout the input of the sensor.
STATIC CHARACTERISTICS
Sensitivity
Amount of change in output with small change in input.
Identify low and high sensitivity regions
This is response of non-linear sensor.
STATIC CHARACTERISTICS
Sensitivity
Linear Sensors
Input Impedance
It is the measure of the sensors ability to not interfere with the signal it is measuring.
Sensor 1 will influence the temperature of T1 body as it is of same mass. Part of heat will be conducted into the
sensor 2. However, Sensor 2 can measure temperature without influence.
STATIC CHARACTERISTICS
Errors
1. Systematic Errors
Errors which do not change when measurement conditions do not change or change predictably when conditions change
Example
Instrument, Method
2. Random Errors
Errors that remain after eliminating systematic errors.
Positive and negative errors are equally likely to occur
Large errors are less likely
Mean of errors in a number of measurements approaches zero.
Example
Electrical interference, Mechanical variation
STATIC CHARACTERISTICS
Resolution
Minimal change in input necessary to produce a detectable change in output.
STATIC CHARACTERISTICS
Uncertainty
Estimated range of an error
Individual errors are properties of the instruments, the test
method, the analysis and the measurement system.
Uncertainty is a property of the test result
Assigned Uncertainty
Based on available information, the operator might feel
confident that the error is within certain bounds, a plus or
minus range of the indicated reading.
Values assigned to each error is the uncertainty
SEQUENTIAL TEST AND HYSTERESIS
A sequential test applies a sequential variation in the input value over the desired input range.
Upscale
Downscale
Hysteresis error refers to differences between an upscale sequential test and a downscale sequential test.
Sensitivity error
Statistical measure of the random error in the estimate of the slope of the calibration curve.
Reproducibility in instrument specification, refers to the closeness of agreement in results obtained from
duplicate tests carried out under similar conditions of measurement.
The term instrument precision in instrument specifications, refers to a random uncertainty based on the results
of separate repeatability tests.
OVERALL INSTRUMENT ERROR AND UNCERTAINTY
An estimate of the overall instrument error is made by combining the estimates of all known errors into a term
called the instrument uncertainty.
The estimate is computed from the square root of the sum of the squares of all known uncertainty values. For
M known errors, the overall instrument uncertainty, uc, is estimated by
For example, for an instrument having known hysteresis, linearity, and sensitivity errors, the instrument
uncertainty is estimated by
CHARACTERISTICS (DYNAMIC)
How to check dynamic characteristics randomly changing input values and recording output.
Example
Keeping a thermometer in ice and suddenly taking it out and put in boiling water. After this, measure how long
the sensor takes to give an output corresponding to new input which is boiling water.
Jumps observed while travelling in a car (suspension system)
Car can suddenly come to rest of gradually come to rest depending upon type of suspension system.
DYNAMICS CHARACTERISTICS (RESPONSE)
Speed of Response
The rapidity with which an instrument responds to a change in the value of the quantity
being measured.
CHARACTERISTICS (DYNAMIC)
Overshoot
It refers to an output exceeding its final, steady-state value.
Due to mass and inertia of moving part the pointer of the instrument does not immediately come to rest in the
final deflected position.
Rise Time
The time required for the response to rise to its final commanded value.
Delay
Delays the signal for an amount of time or time of response delay before reaching commanded position.
CHARACTERISTICS (DYNAMIC)
Overdamped
Response that does not oscillate about the steady-state value but takes longer to reach steady-state than the critically
damped case
Underdamped
The system oscillates (at reduced frequency compared to the undamped case) with the amplitude gradually decreasing to
zero.
Critically damped
The system returns to equilibrium as quickly as possible without oscillating.
Between the overdamped and underdamped cases, there exists a certain level of damping at which the system will just fail to
overshoot and will not make a single oscillation. This case is called critical damping
Steady-state error
The difference between the desired value and the actual value of a system output
CHARACTERISTICS (DYNAMIC)
Damping Ratio
The ratio of the damping coefficient in the system's differential equation to the critical damping coefficient.
1. Independent Variable
2. Dependent Variable
4. Controlled Variable
Constant value or some prescribed condition during a measurement
5. Extraneous Variables
That are not or cannot be controlled but have effect
Introduce differences in repeated measurements under identical operating conditions
Impose a false trend onto the behavior of measured variable
Superimpose signal as noise and drift
TEMPERATURE SENSORS
Most physical, electronic, chemical, mechanical, and biological systems are affected by temperature.
Certain chemical reactions, biological processes, and even electronic circuits perform best within limited temperature ranges
Temperature sensing can be done either through
Direct contact with the heating source
Remotely (using radiated energy)
Sensor Types
Thermocouples
Thermometer
Resistance Temperature Detectors (RTDs)
Thermistors
Infrared
Semiconductor Sensors
THERMOMETER
Devices that measure temperature on the basis of this thermoelectric principle are called thermocouples (TCs).
Working principle (Thermoelectric Principle/Seebeck Effect)
Temperature difference between two dissimilar electrical conductors or semiconductors produces a voltage
difference between the two substances.
A current flows in a circuit consisting of two dissimilar metals, when one junction is heated, while other
junction is kept constant.
The joined end is referred to as the HOT JUNCTION. The other end of these dissimilar metals is referred to as the COLD END or
COLD JUNCTION.
Self powered
WORKING PRINCIPLE
When a temperature differential is maintained across a given metal, the vibration of atoms and motion of
electrons is affected so that a difference in potential exists across the material.
This potential difference is related to the fact that electrons in the hotter end of the material have more
thermal energy than those in the cooler end, and thus tend to drift toward the cooler end.
Drift varies for different metals at the same temperature because of differences in their thermal conductivities.
If a circuit is closed by connecting the ends through another conductor, a current is found to flow in the closed
loop.
The proper description of such an effect is to say that an emf has been established in the circuit and is causing
the current to flow. (Seebeck effect)
THERMOCOUPLE
This equation, which describes the Seebeck effect, shows that the emf produced is proportional to the
difference in temperature.
Thus, if the metals are the same, the emf is zero, and if the temperatures are the same, the emf is also zero.
PROBLEM
PELTIER EFFECT
Peltier effect
The reverse of the Seebeck effect
In this case, we construct a closed loop of two different metals, A and B, as before.
An external voltage is applied to the system to cause a current to flow in the circuit.
Because of the different electrothermal transport properties of the metals, it is found that one of the junctions
will be heated and the other cooled. ( the device is a refrigerator)
Applications
Cooling small electronic parts
One temperature must already be known because the Seebeck voltage is proportional to the difference
between junction temperatures.
The voltage produced has a magnitude dependent on
Absolute magnitude of the temperature difference
Polarity dependent on which temperature is larger, reference or measurement junction
In most industrial environments, this would be difficult to achieve if the measurement junction and reference
junction were close.
It is possible to move the reference junctions to a remote location without upsetting the measurement process
by the use of extension wires.
THERMOCOUPLE TYPES
Standard configurations of thermocouples using specific metals (or alloys of metals) have been adopted.
Each type has its particular features, such as range, linearity, inertness to hostile environments, sensitivity, and
so on, and is chosen for specific applications accordingly.
THERMOCOUPLE POLARITY
The voltage produced by a TC is differential in the sense that it is measured between the two metal wires. As
noted in the footnote to thermocouple type Table.
A type J thermocouple is called iron-constantan. This means that if the reference temperature is less than the
measurement junction temperature, the iron will be more positive than the constantan.
Example
A type J with a reference will 0oC reference will produce +5.27mV for a measurement junction of 100oC,
meaning that the iron is more positive than the constantan.
For a measurement junction of -100oC, the polarity changes, and the voltage will be -4.63mV, meaning that the
iron is less positive than the constantan.
THERMOCOUPLE TABLE
Tables simply give the voltage that results for a particular type of thermocouple when the reference junctions
are at a particular reference temperature, and the measurement junction is at a temperature of interest.
For a type J thermocouple at with a reference of 0oC, the voltage is
THERMOCOUPLE TABLE
CHANGE OF TABLE REFERENCE
THERMOCOUPLE SENSOR
Sensitivity
Construction
Joints welded or even twisted junction between two metals
TC is sheathed in a protective covering or even sealed in glass (hostile environment)
Size of the TC wire: Determined by the application and can range from #10 wire in rugged environments to fine #30
AWG wires or even 0.02-mm microwire in refined biological measurements of temperature.
Range (Refer to range figure)
Time Response
How long it takes the TC system to reach thermal equilibrium with the environment
Signal Conditioning
Amplifier (high gain)
THERMOCOUPLE SENSOR
Reference Compensation
Controlled temperature reference block (temperature-controlled box)
Noise
Twisted and then wrapped with a grounded foil Sheath
Instrumentation amplifier that has excellent common-mode rejection is employed for measurement.
REFERENCE PICTURES