SW9.1x Instruction of Use PDF
SW9.1x Instruction of Use PDF
1 Safe handling 1
2 Product description 2
5 Initiating haemodialysis 5
7 Disinfection 7
8 HDF-online/HF-online 8
9 Single-needle procedure 9
10 Use of options 10
11 Configuration 11
14 Accessories 14
15 Technical data 15
16 Appendix 16
Table of contents Dialog+
Table of contents
14 Accessories..........................................................................................14-3
14.1 Options ................................................................................................ 14-3
16 Appendix............................................................................................. 16-3
16.1 Flow diagrams..................................................................................... 16-3
16.1.1 Key to flow diagrams ............................................................................ 16-3
16.1.2 Flow diagram Dialog+............................................................................ 16-5
16.1.3 Flow diagram Dialog+ HDF-online ..................................................... 16-6
16.2 Technical safety inspection and preventive maintenance ............. 16-7
Table of contents 1
1 Safe handling
1
1.1 About these instructions for use
These instructions for use form an integral part of the dialysis machine. They describe
the appropiate and safe use of the dialysis machine.
The dialysis machine must always be used in accordance with the instructions for
use.
Always keep the instructions for use at the dialysis machine for later use.
Pass on the instructions for use to any future user of the dialysis machine.
1.1.1 Validity
Article numbers
These instructions for use apply to Dialog+ dialysis machines with the following article
numbers (REF):
• 710200X
• 710201X
• 710207X
X = Combination of options at the time of delivery.
Software version
These instructions for use apply to software version 9.1x (x = any).
The target group for these instructions for use is specialist medical staff.
The dialysis machine may only be used by persons instructed for its appropiate
operation.
DANGER Imminent danger that can lead to death or serious injury if not
avoided.
The warning notices are highlighted in the following manner (see below example for a
CAUTION warning):
Here, the type and source of the danger are listed and possible consequences if
measures are not followed!
CAUTION This is the list of measures to prevent the hazard.
This is the list of important information, directly or indirectly relating to safety and
the prevention of damage.
1.1.4 Abbreviations
1
ABPM Automatic blood pressure monitoring
BPA Arterial blood pump
BPV Venous blood pump
BSL Bed Side Link
CO Cross-over
HD Haemodialysis
HDF Haemodiafiltration
HF Haemofiltration
HP Heparin pump
ISO UF Isolated ultrafiltration
PA Arterial pressure
PBE Blood-side entry pressure at dialysis machine
PBS Blood pump control pressure for single-needle procedure
PV Venous pressure
RDV Venous red detector
SAD Safety air detector
SAKA Arterial tube clamp
SAKV Venous tube clamp
SN Single-needle
SN-CO Single-needle cross-over
TMP Trans membrane pressure
TSM Technical support and maintenance mode
UF Ultrafiltration
ZKV Central concentrate supply
The dialysis machine can be used for implementing and monitoring haemodialysis
treatments for patients with acute or chronic kidney failure. The system can be used
for hospital, health centre, limited-care or home dialysis.
Depending on the model, the following types of therapy can be carried out with the
system:
• Haemodialysis (HD)
• Isolated ultrafiltration (ISO UF): Sequential therapy (Bergström)
• Haemodiafiltration (HDF)
• Haemofiltration (HF)
1 1.3 Contraindication
The dialysis system may only be operated on doctor´s instructions if the patient suffers
from one of the following conditions:
• Unstable circulation
• Hypokalemia
Dialysis for patients with a body weight of less than 30 kg requires a wider
safety concept than that applicable to heavier patients.
The special risks concern excessive blood loss, excessive blood flow and a
disproportionately large extracorporeal blood volume compared to the blood
volume of the patient!
Special precautions are required, e. g. intensified monitoring of the patient.
WARNING
The dialysis machine must not be employed without qualification and
adjustment of the entire therapy concept to the special conditions of the
patient. Especially, the compatibility of individual consumables with each
other and with the dialysis system must be considered.
It must not be used or connected to mains voltage if the housing or the mains cord is
damaged in any way. A damaged dialysis machine must be submitted for repairs or
disposed of.
The dialysis machine has been developed and tested in accordance with the valid
standards for interference suppression and electromagnetic compatibility (EMC).
However, it cannot be guaranteed that no electromagnetic interaction with other
devices will occur (examples: mobile phones, computer tomograph (CT)).
The operator may only use the device after the manufacturer has trained the
responsible staff based on these instructions for use.
The dialysis machine may only be used by persons instructed for its appropriate
operation.
The operator must ensure that the instructions for use are read and understood by all
operators of the dialysis machine.
Prior to using the dialysis machine, check for safe functioning and correct condition of
the dialysis machine.
1.6.3 Conformity
1
The dialysis machine and the options comply with the requirements of the generally
applicable standards in their respective valid version:
• IEC 60601-1
• IEC 60601-2-30
• DIN EN 1060-1
• DIN EN 1060-3
Additional equipment connected to the analog or digital interfaces of the dialysis
machine must demonstrably meet the relevant IEC specifications (e.g. IEC 60950 for
data processing devices and IEC 60601-1 for electromedical devices). Also, all
configurations must conform with the valid version of System Standard
IEC 60601-1-1.
Persons connecting additional devices to signal input or output components are
performing a system configuration and are, thus, responsible for ensuring that the
valid version of System Standard IEC 60601-1-1 is complied with. In case of queries,
please contact your local specialist dealer or technical service.
In each country the distribution of the machine is carried out provided that the device
is registered and classified according to the local regulations.
The manufacturer, assembler, installer or implementer shall only be responsible for the
effects on the safety, reliability and performance of the device, if
• the assembly, expansion, readjustments, changes or repairs were carried out by a
person authorised by him and
• if the electrical installation of the affected room comply with the valid national
requirements on the equipment of medical treatment rooms
(i. e. VDE 0100 part 710 and/or IEC60364-7-710).
The device may only be operated if the manufacturer or an authorised person acting
on behalf of the manufacturer
• has carried out a functional check on site (initial commissioning),
• if the persons appointed by the operator to use the device have been trained in the
correct handling, use and operation of the medical product with the aid of the
instructions for use, enclosed information and maintenance information and
• if the quality of the water used with the device corresponds to the relevant
standards.
The company B. Braun Avitum AG guarantees the taking back of old B. Braun dialysis
machines.
The dialysis machine has to be disinfected according to regulations before disposal.
Table of contents
2
2 Product description ............................................................................. 2-3
2.1 Basic models.......................................................................................... 2-3
2.1.1 Dialog+ single-pump machine ...............................................................2-6
2.1.2 Dialog+ double-pump machine .............................................................2-7
2.1.3 Dialog+ HDF-online...................................................................................2-8
2.2 Symbols on the dialysis machine ........................................................ 2-9
2.3 Control elements and information on the monitor ........................ 2-10
2.4 Overview of all icons.......................................................................... 2-12
2.5 Entering numerical values ................................................................. 2-19
2.6 Therapy types...................................................................................... 2-22
2.6.1 Haemodialysis (HD) ............................................................................... 2-22
2.6.2 Isolated ultrafiltration (ISO UF) ......................................................... 2-22
2.6.3 Haemofiltration (HF/HF-online)......................................................... 2-23
2.6.4 Haemodiafiltration (HDF/HDF-online) ............................................. 2-23
2.7 Methods of treatment ....................................................................... 2-24
2.7.1 Double-needle procedure .................................................................... 2-24
2.7.2 Single-needle procedure ...................................................................... 2-24
2.7.3 Single-needle cross-over procedure ................................................. 2-24
2.7.4 Single-needle valve procedure........................................................... 2-26
2.8 Effectiveness of dialysis (Kt/V)......................................................... 2-27
2.9 Using the timer/stop watch............................................................... 2-28
2 Product description
Front view
Legend
1 Venous pressure sensor
connection (blue)
2 Arterial pressure sensor
connection (red)
3 Heparin pump
4 Pressure sensor connection for
regulating the venous blood
pump in single-needle cross-
over operating mode (white)
5 Syringe stop
6 Pressure sensor connection for
arterial inlet pressure to dialyser
(red)
7 Blood pump (one or two blood
pumps depending on basic
model)
8 Rinsing chambers for
concentrate rods
9 Connection for central
concentrate supply (option)
10 Connection for the supply and
discharge of substitution
solution (only for Dialog+ HDF-
online)
11 Arterial tube clamp (for Dialog+
single-pump machine: only
present with option ”SN-valve”)
12 Lever for manual opening of the
venous tube clamp
13 Venous tube clamp
14 Safety air detector (SAD) and red
sensor
15 Fixings for the chamber(s) of the
SN blood tube system
16 Fixings for blood tube system
Side views
Legend
2 1 Mains switch
2 Card reader
3 Connections for dialyser
tubes and rinsing bridge
4 Connection for
disinfectant
5 Caster brake
6 Connection for central
concentrate supply
(option)
7 Bicarbonate cartridge
holder (standard for
Dialog+ HDF-online,
optional for Dialog+ single-
pump machine and
double-pump machine)
8 Storage tray (box)
9 Infusion pole (in some
models, pole may not be
adjustable)
Rear view
Legend
1 Type plate
2 Fixture for disinfectant container 2
3 Water inlet
4 Dialysate outlet
5 Mains cord
6 Connection for electrical ground
7 Crank for manual blood return
8 Alternative position of mains
switch
3 Water inlet
4 Dialysate outlet
The Dialog+ single-pump dialysis machine is suitable for hospitals, health centres,
limited-care and home dialysis. It offers the following features as standard:
2 • Colour screen and on-screen operation (colour touch screen)
• Acetate/bicarbonate operation
• Volumetric ultrafiltration device
• Heparin syringe pump
• Fixed or freely selectable profile controls for dialysate composition, temperature
and flow rate, for heparin supply and for ultrafiltration
• Heat exchanger
The following features are available as additional options:
• Automatic blood pressure monitoring (ABPM)
• bioLogic RR® Comfort (Automatic blood pressure stabilisation, only in combination
with ABPM)
• Bicarbonate cartridge holder
• Central concentrate supply (ZKV)
• Dialysis fluid filter
• Emergency power supply
• Data interface (also available for other use):
– Dialog+-computer interface
– Staff call
– BSL (Bed Side Link): Card reader and interface to Nexadia® data management
system
• SN-valve with additional tube clamp
• Rollers 7 x 10 mm
• Adimea
• Crit-Line Interface
• Card reader
Therapy types
The Dialog+ dialysis machine with a single blood pump can be used for the following
therapy procedures:
• Haemodialysis (HD)
• Isolated ultrafiltration (ISO UF)
Methods of treatment
The Dialog+ dialysis machine with a single blood pump can be used for the following
therapy methods:
• Double-needle procedure
• Single-needle valve procedure
The Dialog+ double-pump dialysis machine is suitable for hospital, health centre,
limited-care and home dialysis and offers the following features as standard:
• Colour screen and on-screen operation (colour touch screen)
2
• Double-pump single-needle equipment
• Acetate/bicarbonate operation
• Volumetric ultrafiltration device
• Heparin syringe pump
• Fixed or freely selectable profile controls for dialysate composition, temperature
and flow rate, for heparin supply and for ultrafiltration
• Heat exchanger
The following features are available as additional options:
• Automatic blood pressure monitoring (ABPM)
• bioLogic RR® Comfort (Automatic blood pressure stabilisation, only in combination
with ABPM)
• Bicarbonate cartridge holder
• Central concentrate supply (ZKV)
• Dialysais fluid filter
• Emergency power supply
• Data interface (also available for other use):
– Dialog+-computer interface
– Staff call
– BSL (Bed Side Link): Card reader and interface to Nexadia® data management
system
• Rollers 7 x 10 mm
• Adimea
• Crit-Line Interface
• Card reader
Therapy types
The Dialog+ double-pump dialysis machine can be used for the following therapy
types:
• Haemodialysis (HD)
• Isolated ultrafiltration (ISO UF)
Methods of treatment
The Dialog+ double-pump dialysis machine can be used for the following therapy
methods:
• Double-needle procedure
• Single-needle cross-over procedure
• Single-needle procedure
The Dialog+ HDF-online dialysis machine is suitable for hospital, health centre,
2 limited-care and home dialysis and offers the following features as standard:
• Colour screen and on-screen operation (colour touch screen)
• Double-pump single-needle equipment
• Acetate/bicarbonate operation
• Volumetric ultrafiltration device
• Heparin needle pump
• Fixed or freely selectable profile controls for dialysate composition, temperature
and flow rate, for heparin supply and for ultrafiltration
• Online production of ultra-clean dialysate for haemodialysis with two-stage dialysis
fluid filter system to ensure the cleanness of the substitution solution
• Online production of substitution solutions free from bacteria and pyrogenes as
used for haemofiltration and haemodiafiltration
• Option of adding substitution solutions by pre- or post-dilution
• Bicarbonate cartridge holder
• Heat exchanger
The following features are available as additional options:
• Automatic blood pressure monitoring (ABPM)
• bioLogic RR® Comfort (Automatic blood pressure stabilisation, only in combination
with ABPM)
• Central concentrate supply (ZKV)
• Emergency power supply
• Data interface (also available for other use):
– Dialog+-computer interface
– Staff call
– BSL (Bed Side Link): Card reader and interface to Nexadia® data management
system
• Rollers 7 x 10 mm
• Adimea
• Crit-Line Interface
• Card reader
Therapy types
The Dialog+ HDF-online dialysis machine can be used for the following therapy types:
• Haemodialysis (HD)
• Isolated ultrafiltration (ISO UF)
• Haemofiltration (HF-online)
• Haemodiafiltration (HDF-online)
Methods of treatment
The Dialog+ HDF-online dialysis machine can be used for the following therapy
methods:
• Double-needle procedure
• Single-needle cross-over procedure; only possible with therapy types HD and ISO UF
• Single-needle procedure
Electrical ground
Dialysis machine ON
Alternating current
Schematic illustration on safety air detector (SAD) and air detector of the
substitution line showing the correct way of installing the tube
2 Signal lamps
Signal lamps on the left and right of the monitor light up in three different colours to
indicate the conditions "Operation", "Fault" and "Alarm".
Legend
1 Signal lamps:
Green = operation
Yellow = warning/note
Red = alarm
2 Buttons on the monitor
Legend
1 Battery symbol (display
only): battery charging
2 Reduce blood pump speed
3 Switch on/switch off blood
pump
4 Increase blood pump speed
5 Confirm alarm (when button
is illuminated); switches of
the alarm buzzer
6 Enter button: Confirm
entered data and reset
information (if button is
illuminated)
Fig. 2-5 Buttons on the monitor
Touch screen
Most functions of the dialysis machine are controlled via the touch screen. The screen
displays different contents (windows) depending on the activated program section.
Different parts (fields and icons) of the screen react to touch. By touching one of these
areas, another window is called up or a stored action is triggered. 2
Some windows show a lateral scroll bar. They could be scrolled by moving a finger on
the scroll bar.
Legend
1 Screen
2 Fields
3 Icons
4 Call up help function for
explaining the icons
Icons are control buttons on the touch screen used for operating the dialysis machine.
2 Depending on the displayed window, different icons are available, which all represent
a specific action. By touching an icon, the respective action is carried out. A list of all
icons is provided below.
Switch off all icon functions for 10 sec to allow cleaning of monitor
Call up overview
2
Return to program selection
Reduce value
Increase value
Red symbol: error symbol during reading of patient data therapy card
2
Call up dialyser rinsing program with simultaneous ultrafiltration
Filter data (only active if option DF filter has been installed or if the
machine is equipped with HDF-online)
Start reinfusion
2
Change to "Therapy end" mode
Activate stand-by
2
Call up exponential profile in case of specified start and end values
2
Start central thermal disinfection
Timer/stop watch
2
Select language of screen text
Crit-Line Table
Crit-Line Trend
The changing of values is based on the same principle for all parameters. We are,
therefore, providing an example at this point. The example refers to the change of the 2
parameter UF quantity on the ultrafiltration data window.
Touch desired icon (here: icon for calling up ultrafiltration data window).
The preset values for the parameter are displayed.
The selected icon lights up in green.
Touch value to be changed on screen (here: value for UF quantity 2000 ml).
A field of icons for changing the value is displayed.
The desired value lights up in green.
Legend
1 Reduce value
2 Increase value
The dialysis machine can be set in the service program in such a way that a keypad
appears immediately after the value to be changed has been touched. In this case,
the keypad has no O.K. icon. To confirm entry, press on monitor.
Reduce value: Touch icon 1 until the desired value has been reached.
Increase value: Touch icon 2 until the desired value has been reached.
Enter different value: Touch icon 3.
2
A keypad is displayed. The permissible setting range is specified in square brackets
below the numerical value (here: 100 ... 20000).
By pressing the icons 1 and 2 permanently, the setting could be adjusted up or down.
Legend
1 Numerical keys
2 Change sign of numerical value
3 Delete set numerical value
4 Leave window and accept data
5 Leave window without accepting
data
To rush the access to the groups of parameters so-called “shortcuts” can be used.
Therefore, touch the parameter which should be changed or a concerning graphic
indicator on the main screen. The corresponding window of the group of parameters
will open as shown in Fig. 2-9. 2
The following screen shows the available shortcut squares in frames.
Legend
1 Help icon, active
Shortcuts
Shortcuts are only active if the corresponding parameters are relevant for the actual
therapy. For example: The setting of the venous limit can only be done by shortcut
within SN therapies.
Some shortcuts directly open the +/- window for changing the setting. For example:
UF-quantity.
Mode of operation
The dialysis machine pumps blood through a patient’s vascular access into the
dialyser.
Inside the dialyser metabolic waste products are separated from the blood. The
dialyser operates as a filter that is divided into two parts by a semi-permeable
membrane. On one side the patient's blood, on the other side the dialysate flows past.
During the therapy the dialysate is prepared by the dialysis machine. It consists of
prepared water to which certain quantities of electrolyte and bicarbonate, depending
on the individual patient’s requirements, are added.
The concentrations of electrolyte and bicarbonate in the dialysate are adjusted in such
a way that certain substances can be removed from the blood through convection,
diffusion and osmosis, while other substances are added at the same time. This is
mainly achieved by diffusive clearance through the semi-permeable membrane of the
dialyser. The dialysate transports the metabolic products from the dialyser into the
discharge line. The cleaned blood is then recycled into the patient.
During treatment the dialysis machine monitors the blood circulation outside of the
body, pumps blood and dialysate in separate circulation systems through the dialyser
and monitors the composition and volume balance of the dialysate.
The heparin pump, which is also part of the dialysis machine, can be used to add
anticoagulants to the blood in order to prevent the formation of blood clots in the
extracorporeal circulation.
In addition to cleaning the blood, the dialysis machine removes water from the blood,
which would be excreted through the kidney in healthy humans.
Isolated ultrafiltration (ISO UF, sequential therapy, Bergström therapy) is used for a
short-term extraction of a higher amount of fluid from the patient.
For further information see section 11.3.
Mode of operation
During isolated ultrafiltration no dialysate flows through the dialyser. This therapy
type serves only for extracting fluid from the patient.
Haemofiltration (HF) is often used with patients suffering from blood circulation
problems for whom this form of therapy is more comfortable.
Middle-molecular substances such as ß-2 microglobulin are better eliminated from
2
the blood by the HF therapy than by the HD therapy.
Mode of operation
During Haemofiltration (HF) the blood is predominantly cleaned by convection on the
dialyser membrane.
In HF therapy, no dialysate is pumped through the dialyser. Instead, a sterile
substitution solution is infused into the blood cycle upstream or downstream of the
dialyser. Depending on whether the infusion takes place upstream or downstream of
the dialyser, the procedure is called predilution (upstream) or postdilution
(downstream).
The filter membrane in the dialyser has a higher water permeability than a HD filter. It
contains a so-called high-flux membrane (haemofilter) and allows ultrafiltration of
considerably more fluid than through HD therapy.
The electrolyte concentrations in the substitution solution and in the dialysate are
identical. The infusion rate is 4–5 l/h. The same volume is ultrafiltered through the
dialyser. In this way the so-called convective clearance is increased so that the
elimination results are the same as with HD therapy.
In HF-online therapy the substitution solution is prepared ”online”. Dialysate is sterile-
filtered so that it can be used as substitution solution. Unlimited substitution solution
is available, allowing higher infusion rates.
Mode of operation
During haemodiafiltration (HDF) the dialysis machine removes more water from the
blood than necessary to replace the natural kidney function. The result is better
cleaning of the blood, although the missing fluid has to be replaced. Therefore,
substitution solution is simultaneously infused into the patient's body. Depending on
whether the infusion takes place upstream or downstream of the dialyser, the
procedure is called predilution (upstream) or postdilution (downstream).
In HDF-online therapy a part of the dialysate is processed as substitution solution.
The single-needle procedure is applied when patients had problems with the
predominantly used double-needle dialysis. In the single-needle procedure, only one
needle (single-needle cannula) or a single-lumen, single-needle catheter is applied to
the patient. The arterial and venous ends of the tube system are connected via a Y-
piece. This procedure allows reducing the number of punctures by half compared to
double-needle dialysis, thus, preserving the patient's shunt.
The following single-needle procedures are available:
• Single-needle cross-over as an alternative to double-needle dialysis
• Single-needle valve as "emergency procedure" for terminating a dialysis in case of
problems with a double-needle dialysis.
The single-needle cross-over procedure with two blood pumps allows a continuous
flow through the dialyser with only one patient connection. During a single-needle
cross-over procedure, the pressure and pulsation conditions within the dialyser are
roughly the same as in a double-needle dialysis.
Mode of operation
Legend
1 Patient connection
2 Arterial tube clamp
2
3 Arterial chamber
4 Arterial pressure sensor
5 Arterial blood pump
6 Heparin pump
7 Dialyser with connection for
dialysate
8 Control pressure for venous
blood pump
9 Venous blood pump
10 Venous pressure sensor
11 Venous chamber
12 Air detector
13 Venous tube clamp
With the arterial tube clamp open and the venous tube clamp closed, the blood pumps
move, at the preset rate, blood from the patient through the dialyser into the venous
chamber. The blood level in the venous chamber rises. The pressure in the venous
chamber is monitored via the venous pressure sensor. Once the set venous switchover
pressure has been reached, the arterial tube clamp closes. Shortly afterwards the
venous tube clamp opens.
The blood from the venous chamber flows back to the patient. The blood pumps
operate simultanously and pump blood from the arterial chamber through the dialyser
into the venous chamber. The pressure in the arterial chamber is monitored via the
arterial pressure sensor. Once the set arterial switchover pressure has been reached,
the venous tube clamp closes and the arterial tube clamp opens. Blood flows again
into the arterial chamber and the process starts again with the withdrawal of blood
from the patient.
Mode of operation
Legend
1 Arterial tube clamp (option)
2 Venous tube clamp
3 Venous red sensor in the housing
of the SAD
4 Safety air detector (SAD)
5 Venous chamber
6 Venous pressure sensor
7 Arterial pressure sensor
8 Arterial blood pump
9 Heparin pump
10 Arterial entry pressure at the
dialyser
11 Arterial chamber
12 Dialyser
2
If the theoretical calculation of the effectiveness is selected, the option Adimea as
described in chapter 10 is not applicable.
The dialysis machine allows optimisation of therapy over many treatments. For this
purpose, the theoretical effectiveness is calculated by the dialysis machine. This
theoretical figure can then be compared with the actual effectiveness determined
from blood samples.
For the actual effectiveness, the urea values before and after dialysis have to be
determined in the laboratory and entered into the dialysis machine.
It cannot be guaranteed that the calculated Kt/V value will actually be reached.
The Dialog+ screen offers a timer or stop watch function for individual use.
2 These functions are offered in the phases:
• Preparing
• Therapy
• End of Therapy
• Selection of disinfection and
• Disinfection.
Legend
1 Adjustment of an absolute
time for a warning sound
2 Adjustment of an interval
time for a warning sound
3 Displays rest or expired time
4 Starts/stops/resets timer or
stop watch
5 Starts/stops the timer for
recurring warnings after
input in 1 or 2
6 Switches off the warning
sound after the chosen time
interval
7 Opens an input window for
reminder
The running timer/stop watch function is shown with a symbol in the date line of the
screen.
Table of contents
Goods-in check
Unpack dialysis machine and check for completeness and damage.
In case of damage, call technical service.
3.2 Storage
3.2.3 Decommissioning
3.3 Transportation
3.3.1 Wheeling
3.3.2 Carrying
For carrying, the dialysis machine can be held at the base, at the rear panel and/or the
protrusion at the front of the machine, as shown in the illustration below.
The ambient conditions of the premises should be according to DIN VDE 0100 Part 710
and IEC 603647-7-10.
3
Ambient conditions
Observe information about ambient conditions, see section 15.2.
The existing mains voltage must correspond with the voltage specified on the rating
plate.
The use of extension cables or adapters with the mains cord or the mains socket is
NOT permitted. Modifications of the mains cord are forbidden! If the mains cord has
to be changed, only the original mains cord listed in the spare parts list must be used.
Electrical installations in the room where the dialysis machine will be operated must
conform with relevant regulations, e.g. VDE 0100 Part 710 and/or IEC-stipulations
(like DIN EN 60309-1/-2 and VDE 0620-1 for example).
Using devices of protection class I the quality of the protective conductor is important.
It is recommended to use a mains socket with additional PE-contact pin according to
CEE 7/7 for cables with safety plug (Schuko). Alternatively, it is recommended to
connect a potential equalisation to the dialysis machine. Regulations and deviations
specific to the individual country must also be observed. For further information, ask
technical service.
The dialysis machine must be properly grounded.
Grounding reliability can only be achieved when equipment is connected to an
equivalent receptacle marked “hospital only” or “hospital-grade”. North American
medical equipment cords and plugs have to be "hospital-grade" or "hospital only",
meaning, they are subject to special requirements contained in relevant applied
standards. It is imperative that the ground connection be reliably maintained to
protect the patient and medical staff. Hospital-grade power cords and cordsets carry
the "green dot" signifying that they have been designed and tested for grounding
reliability, assembly integrity, strength and durability.
We recommend the use of water detectors to protect against any unnoticed water
leaks.
The user must ensure that the water quality is continuously monitored.
The following requirements must be fulfilled:
• The incoming water must be free from Mg++ and Ca++.
3
• pH value of between 5 and 7
Water and dialysate must comply with the country-specific standards, i.e.:
• ISO 13959
Water for haemodialysis and related therapies
• DIN VDE 0753-4
Anwendungsregeln für Hämodialysegeräte
• ANSI/AAMI RD5-03
Hemodialysis systems
• ANSI/AAMI RD61
Concentrates for hemodialysis
• ANSI/AAMI RD62
Water treatment equipment for hemodialysis applications
• AAMI WQD
Water quality for dialysis
• American National Standard for Hemodialysis Systems (RD-5)
• European pharmacopoeia 2002. Official German 4th edition.
Risk of infection due to backflow of contaminated fluids from the drain into the
dialysis machine!
Ensure air clearance between haemodialysis equipment waste connector and
WARNING
the drain (8 cm).
Setting date
Touch field showing date and time 1.
The field containing icons 2, 3 and 4 appears.
There are two setting options:
To increase or decrease the date, change date with icons 2 and 3.
To enter the date using the keypad, touch icon 4.
The numeric keypad appears on the screen.
Enter date using keypad and confirm by selecting OK.
Setting time
Touch field containing date and time 1.
There are two setting options.
To increase or decrease time by minutes, change date with icons 2 and 3.
To enter the time using the keypad, touch icon 4.
The numeric keypad appears on the screen.
Enter date using keypad and confirm by selecting OK.
Touch field containing date and time 1.
The field containing 2, 3 and 4 disappears.
The set date and time are displayed.
• In case of any damage that may put into question the safe use of the machine, the
dialysis machine may not be used. Inform the customer service in charge.
• Only switch on dialysis machine after it has reached room temperature. 3
• Observe requirements on installation site and water supply.
In case of accidentally switching off the dialysis machine by actuating the mains
switch during disinfection, proceed as follows:
Press mains switch again.
The disinfection process is continued.
Table of contents
4
Haemodialysis is the standard dialysis procedure for all system variants.
The procedure is the same for all system variants.
After switch-on, the following main screen is displayed on the dialysis machine:
Touch field 1.
The first preparation screen for haemodialysis appears. The dialysis machine starts an
automatic test sequence.
At the automatic test stage, the dialysis machine automatically checks all control
functions relevant to the safety of the machine.
While the dialysis machine is carrying out automatic tests you can begin entering the
treatment parameters.
If the option „Blood side pressure test with pressure compensation“ is activated in
TSM, the excess pressure in the AV system will be removed via the dialyser after the
pressure test on the blood side.
Depending on the used type of dialyser, this may take up to two minutes.
Legend
1 Status field
2 Operating field
While the dialysis machine goes through the automatic test sequence, messages on a
yellow background appear in field 2 if the machine expects you to carry out actions
such as connecting the concentrate. The test sequence is only continued once this
action has been completed.
Information windows can be hidden, by a touch, for approx. 20 seconds while you use
the screen for other actions, e.g. entering parameters. Upon completion of the entry,
the information window will reappear. Taking over the data with the Enter button
will only be possible after confirming the information window.
Touch icon.
The automatic test sequence is terminated.
The options "Return to therapy selection" and "Repeat blood-side tests" are displayed.
4 Touch the appropriate field.
This screen icon is enabled as soon as the dialysis machine has completed all
automatic tests successfully. The patient can now be connected.
For the user there is a possibility to suppress some warning sounds during preparation,
except only warning sounds which require interaction with the user. For example,
fault removal or on demand for action. Optical alarms and the fault finding are not
affected.
The function “Reduced warning sounds during preparation“ can be used for the
following warnings.
ID Text
1927 Rinsing volume attained
1928 Filling volume is reached
1112 UF Rinse volume for dialyser too high
1153 Repeat self test!
1033 Temperature too low
1034 Temperature too high
1038 Connect acid-/acetate concentrate
1040 Connect bicarbonate
1041 Connect blue concentrate coupling to rinse bridge
1045 Bicarbonate cartridge holder open
Preparing the machine with reduced warning sound could cause a delay of the
following treatment. To schedule planned preparation time requires increased
attention of the staff.
Touch icon
If the function is not active (icon background not coloured in green), it could be
activated by touching the icon. The warning sounds listed in the table above are
automatically suppressed. To indicate this, a crossed out speaker symbol appears at
the date line of the screen.
The function “Reduced warning sounds during preparation“ could be preset in the
TSM-mode by a technician.
The function “Reduced warning sounds during preparation“ is only available during
program selection and preparation and can be configured during selection of program
and preparation. For all other phases of treatment this function is not available (icon
appears grey). Changing into the next therapy the function automatically set resets to
the TSM pre-adjustment.
After completion of the internal pressure test, the request connect acetate/acid
concentrate appears on a yellow background.
This option allows a rinsing of the dialyser membrane with or without ultrafiltration.
Touch icons.
The rinsing parameters are displayed.
Set intended values for rinsing parameters according to the table below.
The blood pump tube segment in the AV system must have the dimensions
8 x 12 mm (inner/outer dimension) for the standard machine. A roller rotor for pump
segments 7 x 10 mm is available as an option.
4
Risk to patient due to haemolysis or blood loss when using a faulty blood tubing
system!
Check to ensure that the tube system is not damaged.
WARNING Check to ensure that no line is kinked.
Make certain that all connections are tightly seated.
Risk of contamination at the patient connectors on the blood tubing system per
use of a rinsing bucket.
WARNING Ensure hygienic handling of the blood lines.
Legend
1 Venous tube valve
2 Safety air detector with venous
red detector
3 Venous chamber
4 Venous pressure sensor
5 Arterial pressure sensor
4 6 Arterial blood pump
7 Heparin pump
8 Pressure sensor for arterial entry
pressure in front of the dialyser
(optional)
9 Arterial chamber
10 Dialyser
A dialyser holder that can be attached to the infusion pole above the top fixing, is
available as an accessory.
To swivel or shift the dialysis holder, always undo the screw clamp on the infusion
pole so that the latter will not be damaged.
Risk to patient due to blood loss when using a faulty blood tube system!
Check to ensure that the tube system and pump segments are not being
damaged at the insertion.
Ensure that the pump segment is placed in the backmost position of the
pump housing.
CAUTION
When inserting the pump segments do not rotate the rollers against a drag.
If the tube system has been damaged through the insertion, replace it by a
new one.
4
The spacers on the inside of the lid do not serve to place the pump segment in the
right position. They prevent the pump segment from moving out of the right position
during operation, therefore, preventing only damage to the rollers.
Risk of damage to the tube system due to prolonged clamping of the venous
tube by the tube clamp!
CAUTION Only place the venous line into the tube clamp (SAK) on therapy day.
If a tube system without PBE sensor is used, the message "No pressure measuring
connection on PBE" is displayed during the pressure test.
The message automatically disappears after 60 seconds.
4
4.6.2 Rinsing and testing the tube system
The level regulation system allows the user to set saline levels in the blood line
chambers for preparation by screen touch.
• During preparation the levels can only be set while the blood pump is running.
• The user is obligated to check for correct setting of the levels in the chambers.
Touch icon
The level window opens.
Legend
1 PV - Venous chamber
2 PA – Arterial chamber
3 PBE – Arterial blood
entry chamber
4
The adjustment of the PBE chamber is only possible if an AV system with PBE line is
used and the line is connected to the machine.
Level increasing
Touch icon gently with one touch
Observe level
Touch again for the correct setting if necessary
Level decreasing
Touch icon gently with one touch
Observe level
Touch again for the correct setting if necessary
The heparin pump is suitable for tube systems with heparinisation downstream of the
blood pump in the positive pressure region.
4 Legend
1 Syringe bracket
2 Syringe gripping plate
3 Clip
4 Unlocking lever
5 Syringe stop
Set syringe stop 5 in such a way that the syringe size can be read.
Release unlocking lever 4 and pull out drive mechanism.
Lift and turn syringe bracket 1.
Insert syringe in such a way that grip and pressure plate engage in the guide.
If the syringe was inserted correctly, the unlocking mechanism will jump back
automatically. Do not close the unlocking mechanism manually.
Close syringe bracket.
4
4.7.2 Venting the heparin line
With these icons, the following parameter groups can be called up:
The actual temperature at the dialyser may differ marginally from the before adjusted
temperature.
4 3 Ultra filtration
Profile
For selecting an ultrafiltration
profile or choosing sequential
therapy, see section 11.3
4 Minimal 0–500 ml/h Min. ultrafiltration rate
UF rate
5 Up. limit 0–4000 ml/h Max. ultrafiltration rate
UF rate (must be set in TSM)
6 Button to set therapy The therapy time can be set. The
time end of therapy time is calculated.
7 Button to set end of The absolute time to end therapy
therapy time can be set. The effective therapy
time is calculated.
8 End of therapy time The absolute end of therapy time is
indicated.
A keypad will open. The end of therapy time can be set in a time range considering the
ultrafiltration volume, the minimal UF rate and the upper limit UF rate.
The effective therapy time is calculated as the difference between the set end of
therapy time and the current time.
The set end of therapy time will not be extended by Bypass phases.
To avoid alarms, adjust the upper limit for the ultrafiltration rate to value above the
calculated actual ultrafiltration rate.
Selecting low UF-rates with long UF-time can cause deviation between debit value
and actual value. Corresponding warnings will appear on the screen. The deviation
will be indicated and has to be confirmed by the user pressing the Enter button .
Touch icon
The lower TMP-limit will be set to –100 mmHg. Through this the backfiltration
warning when reaching -10 mmHg is not applicable.
While using the function “Extended TMP-limit range” you will have to act on the
assumption of back filtration.
For this reason we recommend the use of dialysis-fluid filter (Diacap® Ultra).
Risk to the patient with high risk of internal bleeding (e. g. at recent surgery,
gastro-intestinal abscess or similar diseases)!
Check for indication of internal bleeding during therapy.
WARNING
Check the process of heparin application during therapy.
If in TSM presetting the heparin pump is set “off”, you’ll have to switch it on
manually!
After rinsing the blood side, an information window appears with the request to
connect the dialyser.
Take dialyser tubes from rinsing bridge and connect to dialyser. Observe colour-
coding.
Connect the dialyser inlet coupling (blue) to the Luer Lock connection of the venous
line.
4 Connect the dialyser outlet coupling (red) to the Luer Lock connection of the
arterial line.
Turn dialyser so that the blue connection is facing downward.
Confirm correct connection of dialyser by pressing the Enter key on the monitor.
The dialyser is filled and rinsed.
Adjust level as follows:
– Fill chamber in front of the dialyser entry (PBE) nearly half full,
– Fill venous drip chamber up to approx. 1 cm from the upper edge.
Once the set rinsing volume has passed the system the blood pump stops running. An
information window appears.
Ensure that the blood tube system and the dialyser are filled and rinsed with
physiological saline solution.
Ensure that all levels in the chambers have been set correctly.
Confirm correct settings by pressing Enter key on the monitor.
The dialysis machine is testing the blood tube system.
This screen icon will be enabled as soon as the dialysis machine has successfully
completed all automatic tests. The patient can now be connected.
The dialysis machine features a stand-by mode for the dialysate side. This mode allows
switching off the dialysate side in order to save on permeate and concentrate when
the machine is being prepared and will not be used immediately.
Depending on the service program setting performed by technical service, there are
the following ways in which the stand-by mode can be activated for an adjustable
period:
• Automatic start after automatic test sequence
• Automatic start after rinsing program
• Manual start after automatic test sequence
• Manual start after rinsing program
The max. duration of the stand-by mode is preset in the service program by technical
service.
Depending on the setting entered by the technical service in the service program,
there are the following options for switching off the stand-by mode:
• Manual switch-off
• Automatic switch-off after expired time
• Automatic switch-off during connection of patient
During a power failure in Preparation phase the status of this phase will be saved. If
the power supply will be restored, only the interrupted work step must be repeated by
the device if necessary.
Already entered treatment parameters will remain unchanged.
The saved data will be stored up to 120 minutes. After that time the device has to be
prepared new.
4
It is possible to exchange a bic cartridge during preparation. (see also chapter 10.4)
Table of contents
5 Initiating haemodialysis
After completion of the preparation work, the icon for connecting the patient is
enabled. The dialysis machine is in bypass mode. The signal lamps on the monitor
change to yellow.
Touch icon in preparation screen.
Two brief acoustic signals are sounded. The Enter key on the monitor is lit up. An
overview of the entered patient data appears on the screen.
Check that patient data corresponds with what has been prescribed by the doctor
and confirm by pressing the Enter key on monitor.
The treatment screen appears.
Risk to patients with central venous catheters due to excessive patient leak
current!
WARNING Connect electrical ground on the dialysis machine, see section 1.5.2.
Before starting therapy after a disinfection the clamps SAKA and SAKV as well as the
5 metallic parts of the substitution inlet/outlet ports may not exceed a temperature of
41°C.
Legend
1 Remaining therapy time,
graphical and in numbers
2 Current UF rate
3 Current UF volume
4 Set UF volume
5 Heparin rate
6 Blood flow
7 Heparin bolus
8 Treatment with min. UF rate
9 Bypass
10 Information bar
11 Display of trans-membrane
pressure (TMP), with limits
12 Display of arterial pressure,
with limits
13 Display of venous pressure, with
limits
Fig. 5-2 "Haemodialysis" treatment screen
During the connection phase, the set limit values are not rigorously monitored.
Therefore, particular care is required during the connection phase.
The blood pump stops automatically if blood is detected on the red sensor in the
safety air detector (SAD).
Connect patient venously.
Start blood pump.
Touch icon.
The dialysis machine switches to main connection and the haemodialysis is running.
The signal lamps on the monitor switch to green.
Risk to patient due to haemolysis if the blood flow setting is too high for the
selected needle (PA pressure too low!)!
WARNING Adapt blood flow taking into consideration the arterial pressure. 5
Risk to patient due to reduced dialysis effectiveness since the actual blood flow
is lower than the displayed flow rate if arterial pressures are highly negative!
Correct the blood flow setting.
WARNING
Extend treatment time.
Risk to patient due to reduced dialysis effectiveness since the blood flow is too
low (for example wrong cross-section of the cannula)!
Ensure that the blood flow is high enough.
WARNING
Ensure that the cross-section of the cannula is big enough.
The level regulation system allows the user to set blood levels in the blood line
chambers in treatment by screen touch.
• During treatment, the blood levels can only be set while the blood pump is running
in double-needle mode. The active chambers depend on the used blood line
system.
• The user is obligated to check for correct setting of the levels in the chambers.
Touch icon
The level window opens.
Level increasing
Touch icon gently with one touch
Observe level
Touch again for the correct setting if necessary
Level decreasing
Touch icon gently with one touch
Observe level
Touch again for the correct setting if necessary
• In case the blood pump is stopped, the level regulation system is not active. A
message is displayed that a previous start of the blood pump is required.
• In case of blood side alarms, level regulation is not possible. Alarms have to be
cleared first.
Risk to patient due to blood loss if cannulas get disconnected or slip out!
Standard monitoring function of the dialysis machine cannot safely detect that
such a situation has arisen! 5
Ensure that the access to the patient always remains fully visible during
therapy.
WARNING
Ensure that cannulas are adequately fixed.
Regularly check patient access.
Venous lower limit should preferably be > 0 mmHg.
The venous lower limit value is automatically adjusted during treatment. This means
that the distance between the lower limit and the actual pressure decreases. This
compensates for the haematocrit increase generally caused by ultrafiltration. The
adjustment is carried out every 5 minutes and adds up to 2.5 mmHg at a time. The
minimum distance of 22.5 mmHg is, however, always maintained.
Check venous lower pressure limit during dialysis. An optimal interval is approx.
35 mmHg between the lower pressure limit and the current value.
By changing the speed of the blood pump for a brief period it is possible to reposition
the limits window. In this connection it is spread to the in TSM adjusted size. This
means that an already adjusted lower limit value is put back to the in TSM adjusted
interval.
When setting the limits window ensure that the upper limit is as negative as possible.
5
Blood-side entry pressure at the dialyser (PBE)
If a PBE pressure sensor is used, the blood-side entry pressure (1) at the dialyser is
controlled by its upper limit. The PBE monitoring function warns or signals a possible
blockage of the dialyser due to a kinked tube or increased clotting within the dialyser.
The PBE measurement allows the operator to monitor the formation of a secondary
membrane layer in the dialyser. A possible filter clotting might be avoided. The limits
can only be set via the Alarm limits screen at the beginning of the therapy.
Additionally to the maximum PBE value (2), a so-called Delta (3) could be adjusted.
Delta represents a limiting value which lies above the average actual value of the PBE.
It serves as monitoring the accumulation of a secondary membrane. The average
actual value of the PBE is determined by the Dialog+ within the first five minutes after
starting the therapy and stored as an reference value in the SW. Changes of pressure
by variation of blood flow are automatically considered (e. g.: Average actual value of
PBE at 155 mmHg, plus Delta 150 mmHg, the outcome of this is a PBE limiting value
of 305 mmHg). Achieving this limiting value a yellow warning text appears.
The therapy time still continues during treatment at the minimum UF rate. Where
necessary, adjust UF volume after a treatment at minimum UF rate.
UF compensation
Technical service can determine in the service program how treatment is to be
continued after a minimum UF rate phase.
UF compensation YES
After temporary treatment with minimum UF rate, the preselected UF volume will be
nearly reached by increasing the UF rate in the set UF time.
UF compensation NO
After temporary treatment at minimum UF rate, the preselected UF quantity will not
be reached in the preset UF time.
Touch icon.
A safety message will be displayed.
Confirm heparin bolus by pressing Enter key on monitor.
The heparin bolus preset in the heparin parameters is activated.
Using the function “Arterial bolus” a defined volume of sodium chloride is infused
from a NaCl bag.
Touch icon.
The set-up window for the arterial bolus is displayed.
Enter bolus volume.
Legend
1 Start bolus
2 Bolus volume
3 Infused bolus quantity
4 Arterial infused volume
5 Total volume infused
Touch icon.
The blood pump stops automatically and a safety message appears on the screen.
Connect bag with physiological saline solution to arterial infusion connector.
Clamp off arterial bolus 1 if necessary.
Confirm arterial bolus by pressing the Enter key on monitor.
The arterial bolus is infused. The values can be monitored in the settings window.
Once the set quantity has been infused or the arterial bolus has been terminated by an
alarm, a window appears to confirm Bolus terminated.
Remove clamp on patient inlet, clamp off infusion line and confirm by pressing the
Enter key on monitor.
The window for the arterial bolus is closed and replaced by the therapy screen.
If the arterial bolus was terminated by an alarm, the entire bolus quantity will be
infused upon reactivation of the arterial bolus.
Touch icon.
A screen with the graphical representation icon appears.
Touch icon.
5 Touch icon
The following screen is displayed.
Legend
1 Trend group
2 Activate pre-adjustment from
TSM
3 Edit a trend group
4 Safe and leave window
5 Safe not and leave window
A standard of six groups with three parameters each is preset in the TSM.
Legend
1 Graphic representation of a
treatment parameter
2 Move reference time period
forwards
3 Move reference time period
backwards
4 Set time for reference period
5 List of all trends 5
1st option:
Directly enter the time (4) in the Time window.
2nd option:
Move the time reference line by using the icons << (2) or >> (3).
Legend
1 Actual therapy
2 All therapies, max. 20
Patient names only appear if they are manually entered before therapy or if a
therapy card is used.
Observe local data protection opening trend data which are marked with patient
names.
Touch icon.
The dialysis machine switches to the bypass mode. The haemodialysis is interrupted.
The signal lamps on the monitor switch to yellow. The icon changes its display.
5
Depending on the settings in the service program, the change into the bypass mode
must also be confirmed by pressing the Enter key on monitor.
On completion of the treatment, an acoustic signal can be heard and the message
"Treatment time completed" is displayed, the signal lamps on the monitor switch to
yellow.
• The UF rate is set to 50 ml/h.
• The blood pump is still running.
• The time beyond the adjusted treatment time is shown instead of the remaining
time with a minus symbol in front. The graphics will be displayed in red.
Touch icon.
The message "Terminating treatment" is displayed.
Confirm termination of treatment by pressing the Enter key on monitor.
Touch icon.
After entering new treatment parameters, the haemodialysis can be continued.
Table of contents
6.1 Reinfusion
During the reinfusion phase the limits windows are set to their maximum values. The
reinfusion phase, therefore, demands particular care.
The blood pump starts the reinfusion. The reinfusion screen appears.
The Dialysis machine monitors the reinfusion volume and reinfuses until the red
detector (RDV) detects the physiological saline solution. The blood pump stops.
To continue reinfusion, start blood pump by pressing START/STOP button on
monitor.
The blood pump stops automatically after 400 ml have been reinfused or when a
reinfusion time of 5 minutes has elapsed.
The query "Continue reinfusion?" appears on the screen.
To continue the reinfusion process, confirm by pressing the Enter key on monitor.
The dialysis machine will carry out reinfusion of another 400 ml, or reinfusion for
5 minutes.
Disconnect venous patient connection.
The screen "Confirm reinfusion" (Fig. 6-1) appears only if configured accordingly in
the service program. Otherwise, reinfusion must be called up by pressing icon 1
(Fig. 6-2).
The user is obliged to check for correct setting of the levels in the chambers.
Touch icon.
An information window describing the next steps appears.
Follow the instructions on screen and confirm by pressing the Enter key on the
monitor.
The dialyser is emptied.
Once the dialyser has been emptied, connect the second dialyser coupling to the
rinsing bridge.
Remove from the dialysis machine the dialyser and blood tube system and dispose
of both.
The dialysis machine must be disinfected, see chapter 7.
6
Once "Empty dialyser" has been confirmed, the blood pump cannot be started
anymore!
The functions ”Empty dialyser” and ”Empty cartridge” can be started simultaneously.
However, they are carried out one after the other.
The cartridge is emptied as long as both couplings are connected to the dialyser or
the rinsing bridge.
If the blue coupling is connected to the rinsing bridge, the dialyser is emptied.
Touch icon.
An overview with the actual values appears for the following values:
• Treated blood volume
• UF volume of haemodialysis
• UF volume of sequential phases
• Heparin volume
• Substitution volume (only for HDF/HDF online)
• Profile, if set
More parameters can be displayed by actuating the respective icons.
Table of contents
7 Disinfection
Short chemical disinfection approx. 25–45 min (depending Reduced disinfection effectiveness!
Also: decalcification with citric acid 50 % on disinfectant) Also for decalcification with citric acid 50 %, particularly
following a bicarbonate dialysis.
Chemical disinfection with disinfection adjustable Depending on the installed water treatment system
solution from central water supply With the automatic method, the disinfection solution does
manually or automatically not contact the optional DF filter.
Thermal disinfection with hot permeate approx. 30 min Depending on the installed water treatment system
from central water supply
Settings in the service program such as intake volume, disinfection time, temperature
or rinsing time can only be configurated by Technical Service!
Recommended disinfectants
For disinfection we recommend citric acid 50 % or TIUTOL® KF.
Touch field 1.
The screen listing the different disinfection programs appears.
Legend
1 Select disinfectant
2 Thermal disinfection
3 Chemical disinfection
4 Short chemical disinfection
5 Rinse permeate inlet
6 Chemical disinfection with
disinfection solution from
central water supply
7 Thermal disinfection with hot
permeate
The use of water detectors is recommended for detecting potential leaks during
unsupervised operation.
This function allows switching off the dialysis machine automatically after
disinfection. The dialysis machine is switched on automatically at the specified time
and prepares the next dialysis.
Touch icon.
A screen showing the settings of the weekly disinfection program appears:
Sequence
After activation, the chemical disinfection is carried out automatically as follows:
• Automatic rinse-out
• Aspiration of disinfectant and start of heating cycle
• Disinfection phase: Exposure and circulation
• Rinse-out phase
End disinfection
Check if system is free of disinfectant, see section 7.8.
Short chemical disinfection is intended only for decalcification and can be carried out
only with citric acid!
Activate icon.
The short chemical disinfection is carried out.
Check if system is free of disinfectant, see section 7.8.
Use thermal disinfection only in exceptional cases as its germ-reducing effect is not
sufficient for regular application.
Thermal disinfection is not suitable after bicarbonate dialysis as the dialysis machine 7
needs to be decalcified.
After a bicarbonate dialysis, chemical disinfection with citric acid 50 % is
recommended.
Touch icon.
The thermal disinfection is started.
The progress of disinfection cycle is displayed on the screen.
The dialysis machine offers the option of chemical or thermal disinfection of incoming
water supplied through the water treatment system. The water treatment system must
be suitable for this procedure.
The temperature monitoring during this disinfection program refers to the dialysis
machine and not to the supply line.
Removal of fluid from the central water supply influences the temperature.
7 The use of water detectors is recommended for detecting potential leaks during
unsupervised operation.
For information about the disinfection of the water treatment system, refer to the
operating instructions for water treatment system.
Risk of poisoning the patient with disinfectants left in the water supply!
During central disinfection install warning sign on dialysis machine, e.g.
”Disinfectant in water inlet!”
Reuse dialysis machine for dialysis operation only after the water inlet line
WARNING has been adequately rinsed.
Only connect inlet tubes to central water supply after they have been cleared
of disinfectant.
During chemical disinfection of the water inlet, the disinfecting solution is taken from
the central water supply and pumped into the dialysis machine.
Touch icon.
The following screen appears:
Legend
1 Set flow rate for inlet
disinfection
2 Set duration for disinfection
of the inlet line
3 Set flow rate for rinse-out
4 Set duration for disinfection
of the inlet line 7
Once all disinfectant has been rinsed out of the central water supply:
Touch icon.
Rinsing of the dialysis machine supply line is started and stopped after the preset
time.
Check supply line and dialysis machine for disinfectants.
This disinfection method should only be performed by staff who are also trained for
RO equipment.
Due to technical defects, disinfectant or water from central water supply could leak
over the dialysis machine. The use of humidity sensor is recommended.
During the automatic chemical disinfection of the water branch the disinfectant
solution is removed from the central water supply into the dialysis machine. With this
method certain valve positions prevent contact of the disinfectant with the DF-filter.
Legend
1 Day disinfection
2 Set time
3 Night disinfection
4 Select disinfection programm
5 Set date
6 Select disinfectant
7 Activate week progamm
If the inlet volume (1) is reached, the dialysis machine deactivates itself. After end of
the adjusted residence time (2), the machine reactivates and starts rinse phase with
adjusted parameters (3 and 4).
The service technician can preset the machine in TSM so that it will not reactivate
itself. The dwell time ends and the rinse phase starts if the machine is switched ON
manually. If disinfection parameters are put in as night disinfection, the machine
switches OFF after rinsing phase (see chapter 7.3).
Risk of poisoning the patient with disinfectants left in the dialysis machine!
During central disinfection install warning sign on dialysis machine, e.g.
”Disinfectant in dialysis machine!”
Make sure that disinfectant-free water will be available at the beginning of
rinse phase.
Only use the machine for therapy after sufficient rinsing.
WARNING
Check if the dialysis machine is disinfectant-free.
Only switch ON the machine if the RO equipment is switched ON. If the mains
pressure is too low, disinfectant could reach from the inlet into the central
water supply line.
7.7.3 Thermal disinfection with hot permeate from central water supply
During this disinfection program, hot permeate is taken from the central water supply
into the dialysis machine. If necessary, the permeate is heated up to the temperature
which is required for thermal disinfection of the dialysis machine.
Touch icon.
The following screen appears and the program starts.
Ensure that the dialysis machine is switched on and connected to the central water
supply.
Touch icon.
The following screen appears and the program is started.
Risk of poisoning the patient with disinfectant residues left in the machine!
After using disinfectants, check for any disinfectant residues on the dialyser
WARNING couplings and on the discharge outlet!
If citric acid 50 % was used as disinfectant, a check for disinfectant residues is not
necessary.
The following information window appears on the screen after the set rinsing time:
The following indicators can be used to check that the system is free from
disinfectant:
7.9 Decalcification
When using citric acid 50 % for disinfection, decalcification of the dialysis machine
is not required.
When using alkaline disinfectants, a decalcification with citric acid 50 % must first
be carried out. 7
Effective descaling of the DF filter is influenced by the preset contact time and the
temperature used during the cleaning cycle. Dialysis therapies using higher
concentrations of bicarbonate may require longer contact time and higher
temperature.
The automatic descaling function can be enabled in TSM. Instead of citric acid, acid
concentrate used for treatment is drawn in from the machine in high concentration to
decalcify the DF Filter between two bicarbonate therapies. It does not replace
disinfection.
Automatic descaling is required if the machine is equipped with the option DF filter.
After the patient is disconnected from the machine, empty the dialyzer as usual.
Connect the dialyzer couplings to the rinsing bridge.
The bicarbonate cartridge may be left in the holder during the process.
The bicarbonate concentrate coupling may be left connected to the concentrate
source during the process.
After acid is drawn up, the machine enters into acid rinse out.
The following screen appears:
7
Fig. 7-13 ”Descaling” screen – Acid rinse out
As soon as acid rinse out is completed, the machine enters into Preparation and starts
the preparation process if the “Automatic Preparation Start after Disinfection“ has
been enabled in TSM.
Automatic descaling can be interrupted in any phase of the process. The machine will
go to Disinfection main screen and the acid rinse out will be carried out. Afterwards,
disinfection rinsing will start automatically.
If the dialysis machine was configured in the service program in such a way that
disinfection can be terminated, the disinfection program can be terminated at any
time.
Touch icon.
An information window appears.
If disinfectant has already been drawn in, the termination of the program is followed
by a rinse-out phase (e.g. 5 minutes when using citric acid 50 %, or 20 minutes when
using TIUTOL® KF).
If the settings ”Disinfection after every dialysis” and ”Termination of disinfection”
have been configured, a full disinfection must be carried out prior to the next
dialysis.
Table of contents
8 HDF-online/HF-online......................................................................... 8-3
8.1 Preparing for haemodiafiltration/haemofiltration ........................... 8-4
8.1.1 Calling up haemodiafiltration/haemofiltration ................................8-4
8.1.2 Connecting the concentrate ..................................................................8-4
8.1.3 Entering the substitution parameters .................................................8-5
8.1.4 Inserting the tube system.......................................................................8-7
8.1.5 Filling and rinsing the tube system with substitution
solution from the online system...........................................................8-7
8.1.6 Inspecting the tube system................................................................. 8-10
8.2 Preparing for standard haemodialysis with online fluid ................ 8-10
8.3 Carry out haemodiafiltration/haemofiltration................................ 8-11
8.3.1 Connect patient and start
haemodiafiltration/haemofiltration.................................................. 8-11
8.3.2 During haemodiafiltration/haemofiltration.................................... 8-13
8.4 Finish haemodiafiltration/haemofiltration...................................... 8-15
8
8.4.1 Reinfusion with substitution solution.............................................. 8-15
8.4.2 Emptying the dialyser ........................................................................... 8-17
8.5 Disinfection ......................................................................................... 8-17
8.5.1 Regular disinfection .............................................................................. 8-17
8.5.2 Displaying the online filter data........................................................ 8-17
8.5.3 Changing the online filter ................................................................... 8-18
8.5.4 Sampling of substitution fluid ........................................................... 8-21
8 HDF-online/HF-online
Risk to the patient due to contamination and pyrogenic reaction caused by germ
growth during extended idle times for the system between treatments.
Disinfect the dialysis device before a new treatment especially after
WARNING
extended idle times.
b
Fig. 8-1 Main screen ”HD/HDF/HF”
Touch field 1.
The first preparation screen for HD/HDF/HF is displayed. The dialysis machine starts an
automatic test sequence.
See section 4.
Touch icon.
A screen showing the substitution parameters is displayed.
In this way the HDF/HF mode can be activated even during a running haemodialysis.
However, in this case the substitution line will not be tested. Therefore, special care
must be taken when connecting the line. Therefore, pay special attention to ensure
that the direction of the pump operation corresponds with the desired flow direction
of the substitution solution!
2 Nominal 50-250 ml
bolus volume –
Legend
1 Venous pressure sensor
2 Arterial pressure sensor
3 Arterial blood pump
4 Heparin pump
5 Pressure sensor dialyser input
pressure
6 Dialyser
7 Adapter predilution
8 Path of the substitution line
with predilution
9 Path of the substitution line
with postdilution
10 Online substitution pump
11 Substitution port outlet 8
12 Substitution port reflow
13 Venous tube clamp
14 Safety air detector/red sensor
Fig. 8-3 Tube system for HDF/HF therapy, connected for online rinsing
8.1.5 Filling and rinsing the tube system with substitution solution from the
online system
The Dialog+ HDF-online dialysis machine allows filling and rinsing the blood tube
system and the dialyser with the substitution solution prepared by the machine. The
rinsing liquid is taken from the machine and recycled back to it.
The dialysis solution is only available if all DF tests are passed and no DF alarm is due
to be dealt with.
8
Fig. 8-4 Information window for connection
Take dialyser connections from rinsing bridge and connect to dialyser. Observe
colour-coding.
Turn dialyser with blue connection facing upward.
Open substitution port outlet 2 and connect substitution line.
For postdilution, connect the other end of the substitution line to the venous
bubble catcher. For predilution, connect the other end of the substitution line with
the additional adapter upstream from the dialyser.
Connect arterial patient connection to the Luer-lock connector of the substitution
line between substitution port outlet and substitution pump.
Insert pump segment of the substitution line into the venous blood pump.
Connect venous patient connection to substitution port reflow 1.
Confirm correct connection of dialyser and substitution line by pressing Enter key
on the monitor.
The blood tube system and the substitution line will be filled with substitution
solution.
8
Fig. 8-5 Information window for level adjustment
The dialysis machine can also be conditioned with rinsing solution from bags, see
section 4.
It is possible to use the online fluid to prime for a standard HD treatment without
using a substitution line.
Set-up standard double-needle blood line system as usual without connecting the
arterial and venous patient Luer-Lock connectors.
8 In Program Selection, select "HD/HDF/HF".
The confirmation window (figure 8-4) appears.
Take dialyser connections from rinsing bridge and connect to dialyser. Observe
colour-coding.
Turn dialyser with blue connection facing downward.
Connect the arterial line to the substitution port outlet (white).
Connect the venous line to the substitution port reflow (blue).
Confirm correct connection of dialyser by pressing the Enter key on the monitor.
The blood line is filled with saline from the online port. Follow description on page
8-9.
Legend
1 Venous pressure sensor
2 Arterial pressure sensor
3 Arterial blood pump
4 Heparin pump
5 Pressure sensor dialyser input
pressure
6 Dialyser
7 Online substitution pump
(not used)
8 Substitution port outlet
9 Substitution port reflow
10 Venous tube clamp
11 Safety air detector/red sensor
Risk to patient due to blood loss caused by wrong positioning of the substitution
line!
Inspect the substitution line for its flow direction prior to every treatment.
Position substitution line always prior to selftest.
DANGER
It is recommended to use only substitution lines produced by B. Braun.
Other systems than those produced by B. Braun must have a non-return valve.
As soon as the patient data have been confirmed, the therapy screen is displayed with
a window requesting "Connect patient".
Fig. 8-7
8 Therapy screen ”HDF/HF”
Remove the arterial line from the substitution line and connect it to the patient.
Start blood pump by pressing START/STOP button on monitor.
The blood pump operates automatically at the preset rate.
Fill blood tube system with blood.
The blood pump stops automatically if blood is detected at the red sensor
downstream from the safety air detector.
Remove the venous line from substitution port reflow and connect it to the patient.
Close the substitution port.
If the blood pumps are stopped manually or deactivated, the patient connection will
be interrupted (venous pressure rises).
In the same way as during haemodialysis, the following additional functions are
available during haemodiafiltration/haemofiltration:
• Treatment at minimum UF rate
• Administration of a heparin bolus
• Administration of an arterial bolus 8
• Halting the haemodiafiltration/haemofiltration
Administering an infusion bolus is also possible.
Infusion bolus
Loss of blood pressure because of missing volume for the stabilisation of the
circulation if the DF-flow is interrupted by a technical defect or by switching
into bypass.
CAUTION
Keep an NaCl-bag ready for infusion or reinfusion.
Touch icon.
The substitution parameters and infusion bolus screen is displayed.
Fig. 8-8
8 ”Substitution parameters HDF” screen
Adjust bolus volume if necessary. To end this, touch field bolus Volume and enter
new setting.
Touch field Inf. bolus and acknowledge information by pressing on the monitor.
The infusion of the bolus is started. The blood pump operates at 100 ml/min, the
substitution pump at 200 ml/min. The infused volume is added up in the field Total of
Inf.volume.
As soon as the bolus has been administered completely, blood flow and substitution
flow automatically reset to their original values.
The bolus volume is not added to the ultrafiltration volume automatically so that it
remains with the patient.
At the end of the therapy an acoustic signal is sounded. The message ”Therapy time
over” is displayed. The dialysis machine reduces the UF rate to 50 ml/h.
Touch icon.
A query window ”End of Therapy” pops up.
Confirm end of treatment by pressing Enter key on the monitor.
The treatment is finished.
During the reinfusion phase, the limit windows are set to their maximum values. The
reinfusion phase, therefore, demands particular care.
Touch icon.
The following screen is displayed:
The ”Confirm reinfusion” screen (Fig. 8-9) appears only if configured accordingly in
the service program. Otherwise, reinfusion must be called up by pressing icon 1
(Fig. 8-10).
The dialysis machine monitors the reinfusion volume and reinfuses until the red
detector recognises a preset dilution level of the blood. The blood pump stops.
To continue reinfusion, press START/STOP button on the monitor.
The blood pump also stops automatically after 400 ml have been reinfused or once
a reinfusion time of 5 minutes has been achieved.
Disconnect venous patient connection.
It is also possible to connect the arterial line directly to the port outlet without an
adapter line. A long disinfection is mandatory directly after this procedure.
8.5 Disinfection
The regular disinfection after a dialysis and in the morning prior to the first dialysis is 8
described in chapter 7.
Scalding or alkali burn risk to users due to disinfectant solution escaping from
the substitution port or the filter holders!
The substitution connection and the filter holder heat up during disinfection.
WARNING Do not open the substitution port or the filter holder during disinfection.
Make sure that the substitution port is closed correctly.
Touch icon.
The remaining operating time and the number of dialyses performed are displayed.
The online filters must be changed at the latest when the following information
window pops up on the screen:
Touch icon.
The selection menu appears.
Touch icon.
The following window appears:
For complete emptying, the emptying button should remain active for about 3 to 5
minutes. Minor residues remaining in the filters are unavoidable.
For the HDF filter, open filter cover 1; for the DF filter, open filter cover 2. Use a
suitable screwdriver and open the cover.
Remove used-up filters and replace them with new ones.
Close the filter cover(s) and relock them with the screwdriver.
To complete the filter change, touch the middle icon "EMPTY FILTERS" once more.
Close the substitution port.
Reset the filter data using the icons "DF" and "HDF".
Fill and rinse the filter.
Carry out disinfection with citric acid 50 %.
For hygienically faultless sampling of the above mentioned fluid proceed as follows.
Prepare the device as usual.
Insert the substitution line.
Start therapy (without patient).
Adjust substitution rate to 200 ml/min.
Extract the required amount for your sample from the infusion connection of the
substitution line.
End therapy.
Start disinfection.
Table of contents
9 Single-needle procedure
In the following we describe the single-needle procedure only as far as it differs from
double-needle dialysis. For detailed operating information see chapter 4., 5. and 6.
Insert tubes
The following is required:
• SN-CO tube system
• Dialog+ with two blood pumps
Risk of blood loss due to arterial line leakage upstream of the tube clamp!
9 Ensure that there are no connection leaks and that the tube system is fully
WARNING intact.
Legend
1 Set SN-CO parameters
2 Activate SN-CO parameters
3 Connect patient
4 Call up single-needle selection
The level regulation system allows the user to set blood levels in the blood line
chambers in single-needle cross-over mode by screen touch.
The user is obligated to check for correct setting of the levels in the chambers.
In SN-CO mode, the blood level regulation requires a previous blood pump stop
automatically performed by the machine.
Touch icon
The level window opens. All chambers are paled (inactive).
9 Touch button
A supervisor window opens.
Level increasing
Touch icon gently with one touch and observe level.
Touch again for the correct setting if necessary.
Level decreasing
Touch icon gently with one touch and observe level.
The level regulation is performed with the preset blood flow speed, but with a
maximum of 400 ml/min.
Touch icon.
The dialysing machine switches to connection mode.
Confirm patient data, see section 5.1.
Connect arterial tube.
Insert tube segment of venous blood pump. Ensure that the pressure gauge line for
the blood pump control pressure is located upstream of blood pump inlet.
Start blood pumps.
Fill tube system with blood. 9
Stop blood pumps.
Connect venous tube to patient.
Restart blood pumps.
– At 150 ml/min for central catheter
– At approx. 100 to 120 ml/min for fistula connection
The dialysis is started.
Increase blood pumping rate, taking into account the phase volume.
Alternatively, the double-needle mode could be used. Therefore, the venous pump
tube segment will be inserted after connecting the venous patient access.
Then change into SN-mode
Confirm window on the screen.
Start blood pump.
Dialysis starts.
Central catheter
up to -200 mmHg 360 to 390 mmHg
Good fistula
During dialysis
Observe levels in arterial and venous chamber.
If necessary, change levels in field SN chamber level, see below.
The therapy ends automatically or after touching the respective icon. Also observe the
following steps.
Leave tube segment of venous blood pump in venous blood pump.
Always start reinfusion by pressing the appropriate icon.
Disconnect patient, see chapter 6.
In the following we describe the single-needle valve procedure only as far as it differs
from double-needle dialysis. For detailed operating information see 4., 5. and 6.
Inserting tubes
The following is required:
• AV set for SN-valve (venous chamber 100 ml) or normal AV set for Dialog (venous
chamber 30 ml)
• For Dialog+ single-pump machine: Option SN-valve with arterial tube clamp
(without arterial tube clamp increased recirculation will occur)
Low effectiveness due to high recirculation ratio at small phase volumes when
single-pump machines without SN-valve option are used!
WARNING Set phase volume >12 ml.
9
Insert standard arterial tube.
Push arterial tube through arterial tube clamp (if present).
Insert venous tube.
Place venous tube through venous tube clamp.
Connect pressure sensors PA, PBE, PV. Check for secure seat.
Legend
1 Set min. arterial control
pressure
2 Set venous control pressure
3 Activate SN-valve parameters
4 Connect patient
5 Call-up single-needle selection
9
Fig. 9-5 Single-needle valve (SN-valve)
It is possible to set a lower max. limit to protect the arterial pressure limit.
In order to achieve the highest effective blood flow at minimum recirculation, the
control pressures must be set for an optimum phase volume.
Touch icon.
The dialysis machine switches to therapy mode.
Confirm patient data, see section 5.1.
Connect patient, see section 5.2.
Fill tube system with blood. Fill level in venous chamber to only approx. 35 % in
order to achieve a good phase volume.
Start blood pump and slowly increase speed depending on the vascular condition of
the patient.
The dialysis is started.
During dialysis, the following phase volume should be reached:
• For standard-AV set with 30 ml chamber: 12–18 ml
• For AV set for SN-valve with 100 ml chamber: 15–25 ml
In order to change the phase volume, the control pressures can be set within certain
limits depending on the patient's connection conditions.
The level regulation system (if present) allows the user to set blood levels in the blood
line chambers in single-needle valve mode by screen touch. See section 9.1.2.
9
Recommendation
Lower venous control pressure min PV Upper venous control pressure max PV
If necessary, change phase volume through control pressures min PV and max PV:
– To increase the phase volume: Increase interval between min PV and max PV.
– To decrease the phase volume: Decrease interval between min PV and max PV.
Make certain the phase volume does not drop below 12 ml.
The phase volume reacts to:
– Changes in the blood flow
– Changes in the control pressures
– Blood levels in venous chambers
– Pressure changes in shunt
Observe level in venous chamber.
If necessary, change level via field SN chamber level.
If necessary, adjust min. PV and max. PV, see section 4.7.
The optimum return flow time is set automatically.
The therapy ends automatically or after touching the respective icon, see section 5.4.
Also, observe the following steps.
Leave tube segment of venous blood pump in venous blood pump.
Always start reinfusion by pressing the appropriate icon.
Disconnect patient, see chapter 6.
Table of contents
10.9.5 Reading of the data from the patient therapy card ..................10-46
10
10 Use of options
The automatic blood pressure monitoring function does not release operators from
the obligation to regularly monitor the patient. 10
The information transmitted and displayed by the option may not be used as only
source of information for the medical indication.
10.1.1 Cuff
The following cuffs can be used for ABPM blood pressure monitoring:
• Small (upper arm size 18 – 26 cm)
• Medium (25 – 35 cm)
• Large (33 – 47 cm)
A ”Medium” size cuff is supplied with every system delivered.
10
Apply cuff tightly in a suitable place around upper arm of the patient.
Place marking on inside of cuff over artery.
Ensure that cuff tube is not kinked.
If applicable, set measuring cycle to the desired time interval (1 - 60 min depending
on the clinical situation).
10.1.2 Settings
Touch icon.
The settings menu appears.
Touch icon.
The ABPM main view appears: 10
Legend
1 Set the alarm limits
2 Set the cycle time, in minutes
3 Activate/deactivate cyclical
measurement
4 Start/Stop ABPM
The TSM allows a preset to determine whether the cyclic measurements are
terminated by changing to Disinfection mode.
10
Option 2: To set the alarm limit based on the last measuring result:
Touch field Individual limit adaption.
New limit settings are suggested on a coloured background.
Confirm limit settings by pressing Enter key on monitor.
10
After an initial measurement, the alarm limits should be set closer around the current
blood pressure values.
The suggested alarm limits normally range around ± 30,
in critical areas at ± 10 mmHg around the last measured value.
To ensure best possible measurements, the cuff should be at heart level so that the
measured blood pressure does not differ from the actual blood pressure due to the
height difference.
Touch field 1 on the ”Therapy” screen (Fig. 10-4) and field 4 in the ”ABPM main
10 menu” window (Fig. 10-2).
The last values measured for the systolic and diastolic pressure and the pulse rate are
being displayed.
To stop an ongoing blood pressure measurement, touch the respective field again.
Important!
Erroneous measurements are marked by an asterisk in the first position. By activating
a line with an asterisk a window with the measuring results and an error description
is called up.
When a measurement was canceled, the display field appears in yellow and shows
”---/---”.
The display field is also highlighted in yellow when the limits are exceeded.
In the measurement overview, all results are displayed with the respective time
information. Values shown in red indicate that limits have been exceeded.
Legend
1 Measured values at the time
selected in field 4.
2 Cursor
3 Arrow fields for moving the cursor
4 Selected time
5 Switch on/off graphic display
10
10.2.2 Setting of systolic blood pressure lower limit (SLL) and maximum
ultrafiltration rate
10
Set the parameters for bioLogic RR® Comfort according to the following table.
1 Max. UF rate in % 100–200 % Enter max. UF rate for treatment with bioLogic RR® Comfort
either as absolute value or as percentage of the average UF rate.
To guarantee an effective regulation of the UF rate and the
2 bioLogic RR® Comfort 100–4000 ml/h optimal saving of blood pressure measurements, we recommend a
max. UF rate max. UF rate of 160 %.
3 Set suggested SLL as 65-80 mmHg Suggests an SLL which has been determined from previous
required therapies.
The setting is performed by pressing this button.
(Can be optionally set in TSM)
4 SYS lower limit 65–130 mmHg Lower limit for systolic blood pressure
Recommendation:
To guarantee an effective regulation of the UF rate and the
optimal saving of blood pressure measurements, we recommend
the assumption of the suggestion value.
If this function is deactivated, the value should be determined by
the attending physician.
11 Total of inf. volume Displays the accumulated infusion volume of the current therapy
–
The SLL and max. UF rate/bioLogic RR® Comfort max. UF rate parameters should be
10
determined by the attending physician. The SLL should be set as low as possible in
the range of the known tolerance of the patient
The reaching of the dry patient weight within the set therapy period can conflict
with stable blood pressure behaviour.
The attending physician can decide to:
• Adapt the UF volume
• Extend the therapy period
• Accept that the UF target was not achieved.
For medical reasons, all values can also be adapted during the therapy
Touch the "Set suggested SLL as required" field. The value shown on button 4, Fig.
10-7, is accepted.
Optionally, set a value with button 4.
When accepting an SLL that was suggested by the system, the user should assure the
plausibility of this value relative to the patient’s tolerance.
In the first five minutes after the therapy start the bioLogic RR® Comfort functions
can be switched off again. As from the sixth minute a prompt must be additionally
confirmed: Are you sure ...?. After confirmation of this prompt the functions cannot
be activated again!
As from the sixth minute the text in the fields is displayed in blue to point out that
the first five minutes have expired.
The "Set suggested SLL as..." function can be activated/deactivated in the TSM.
10
10.2.4 Activating/deactivating bioLogic RR® Comfort
The bioLogic RR® Comfort option does not have the “basic” bioLogic RR® function:
unlike previous versions the guideline mode cannot be switched off.
Touch the icon in the "ABPM Main Overview" screen (Fig. 10-2).
An overview with the icon for graphical representation is displayed.
Touch icon.
The following screen is displayed:
Legend
1 Measured values for the systolic
and diastolic blood pressures as
well as for the pulse at the
selected time; also functions as 10
button for switching between
different graphical
representations
2 Graphical progression of systolic
blood pressure and
ultrafiltration rate
3 ON/OFF button for the screen
with the graphic bioLogic RR®
Comfort representations
Touch icon.
The following screen with the representation of the bioLogic RR® Comfort parameters
is displayed:
Legend
1 Graphical display of the
bioLogic RR® Comfort values
2 UF rate at the selected time
3 Reference value for the UF rate
at the selected time
4 Systolic blood pressure at the
selected time
5 Systolic blood pressure over the
period of the treatment
6 Cursor
7 Reference line for the UF rate
over the treatment period,
probable progression
8 Arrow keys for selection of a
point in time
9 Graphical display of delivered
UF rate within an interval.
10
10.3 Adimea
The option Adimea is an exact measuring method for precise monitoring of the dialysis
dose during a complete dialysis therapy.
The system works with spectroscopy (UV light) in order to measure the reduction of
urinary waste products in the dialysate outlet of the dialysis machine. Due to
continuously recorded measurements the Kt/V and the urea reduction ratio (URR) can
be determined and displayed. Besides the mentioned parameters, it is also possible to
show the behaviour of the UV light absorption, allowing to evaluate the reduction of
substances during the complete therapy.
Two calculation methods are available for calculation of Kt/V and URR:
• Single pool Kt/V (spKt/V)
• Equilibrated Kt/V (eKt/V)
The selection is performed once in TSM mode. The calculation method set is displayed
on the screen.
Input of patient weight before dialysis (Fig. 10-10, 1). Setting the parameter
enables the calculation and display of Kt/V, URR and UV absorption.
10
Input/adaptation of the target Kt/V value (Fig. 10-10, 2).
Enable/disable target warning (Fig. 10-10, 3). If the target warning is enabled, the
system informs the user whether the target value will possibly not be reached at
the end of therapy. In that case the user can adapt the parameters in order to reach
the determined dialysis dose.
The user can directly access three parameters which influence the Kt/V without
changing the menu. Those are therapy time, blood flow and dialysate flow.
Influences on curves and values will be displayed after a short processing time.
The introduction of the patient weight and, therefore, the activation of the Kt/V
measuring function can be done at any time during therapy. The displayed Kt/V and
URR and UV light absorption always consider the already achieved dialysis time.
Legend
1 Input of patient weight before
dialysis
2 Input/adaptation of the target
Kt/V value
3 Enable/disable target warning
Red dashed line Prediction that target value will not be reached
(extension of blue line)
Blue dashed line Prognosis
(extension of blue line)
Legend
1 Chose Kt/V and/or URR and/or
UV absorption
2 Actual progress line of the
Kt/V (graphical display) and
actual Kt/V value (numerical
display)
3 Orientation line of complete
therapy
In haemodialysis (HD) mode the user gets a “prognosis” of the estimated Kt/V value at
the end of therapy. It is displayed numerically (Fig. 10-12, 1) and graphically (Fig. 10-
12, 2). The blue actual progress line will be extended from the actual therapy status in
10
order to predict the therapy progress.
This functionality is not available in HDF and Single-needle mode.
If the “Target warning” is enabled, the machine will inform the user showing a yellow
warning on the screen in case that either Kt/V resp. URR will not be reached at the
end of therapy.
The warning is displayed if either the blue actual progress line (Fig. 10-13, 1) has
already been below the green dashed orientation line (Fig. 10-13, 2) or if it could fall
below it within the remaining therapy time (Fig. 10-14).
10
Fig. 10-13 Graphical display of blue actual progress line below green dashed
orientation line
In the second case the blue line (Fig. 10-14, 1) will be extended by a red dashed line
(Fig. 10-14, 2) predicting that the target value will not be reached.
2 Therapy time 1 h – 10 h
4 Blood flow 30 ml/min – 600 ml/min Adjustment via the +/- keys on the monitor
Using the patient therapy card allows to store the patient’s individual Kt/V parameters
and graphic Kt/V or URR therapy progresses. Therefore, data are still available for the
user after the end of therapy. It is possible to store up to 12 completed therapies and
to compare them graphically or to evaluate Kt/V and URR values of up to 50
completed therapies. Trends or unusual therapies can be identified and analysed if
necessary.
By touching one of the displayed therapies the screen showing the black dashed
progress line (Fig. 10-16, 1) opens:
10
Text Description
1 Therapy date and time [dd.mm.yyyy, h:min] Date and time of performed therapies
5 Average blood flow [ml/min] Average blood flow over the therapy period
6 Average dialysate flow [ml/min] Average dialysate flow over the therapy period
Press in lateral button at top fixing and pull up top fixing as far as possible.
Using your left hand place cartridge between top and bottom fixings. At the same
time, place inlet and outlet necks of cartridge into their respective recesses at the
top and bottom fixings.
The lever at the top fixing is automatically pushed back in the process.
Crushing hazard!
When closing the cartridge holder do not bruise hands or fingers!
CAUTION
To close the cartridge holder, press top fixing centrally onto the cartridge.
The cartridge is pierced, automatically vented and filled with permeate.
When the cartridge is empty, the bicarbonate conductivity alarm is triggered and an
information window appears. An almost empty cartridge can be replaced before an
alarm is activated.
With draining
Touch icon.
10
Without draining
Touch icon.
The functions ”Empty dialyser” and ”Empty cartridge” can be started simultaneously,
although, they are carried out successively.
The cartridge is emptied as long as both couplings are connected to the dialyser or
the rinsing bridge.
If the blue coupling is connected to the rinsing bridge, the dialyser is emptied.
10
When using a dialysis machine equipped with the ”Central concentrate supply” option,
the concentrate (acetate or bicarbonate components) does not have to be provided in
containers but can be obtained from the central supply. Both components can be
obtained centrally or individually from containers. Another option is combining a
bicarbonate cartridge with acid components from central concentrate supplies.
To connect the concentrate, place couplings of suction rods onto connections of
the central concentrate supply, which are located on the dialysis machine below
the rinsing quivers. Observe colour-coding!
The concentrate connections of the dialysis machine are, thus, directly connected to
the wall connections of the central concentrate supply.
The dialysis fluid filter is a hollow-fiber filter. It is used for performing haemodialysis
therapy with ultra pure dialysate. Even if the machine has been correctly cleaned and
disinfected, the permeate and the bicarbonate concentrate, which, unlike the acid-
containing concentrate, is not autosterile, can be the source of possible
contamination.
The dialysis fluid filter may only be operated with permeate or dialysate.
The dialysis fluid filter must be changed at the latest when the following warning
window is displayed on the screen:
The dialysis machine supervises remaining operating hours of the filter as well as
performed therapies. Operating hours are hours in Therapy as well as hours in
Preparation and Disinfection.
When either the operating hours or the number of therapies are reached, a warning
window will be displayed on the screen. It informs the user about the upcoming filter
change. The warning window appears in case that 60 operating hours or 10 therapies
remain. It is displayed when the user changes from Program Selection to Preparation
and it lasts for 1 minute.
The number of hours or therapies is set by the technician in the TSM. It is
recommended to change the filter after 150 therapies or 900 operating hours.
Preconditions
• No patient connected to dialysis machine
• Dialysis machine switched on
• Screen Disinfection selection is displayed, no disinfection program started (machine
is in Rinse out, see figure 10-25)
Touch icon.
A screen is displayed.
Touch icon.
The following screen is displayed:
10
The remaining dialysis time and the number of performed dialyses are displayed.
Touch the "FILTER DRAINING" icon.
A prompt to remove the dialyser feed connector is displayed.
Remove all couplings (red and blue) from the filter. Catch the fluid escaping in the
process.
Hold old filter centrally and remove it from the clamping brackets of the filter
holder.
Hold new filter centrally and press it into the clamping brackets of the filter holder.
Push blue couplings onto dialysate couplings on the filter caps.
Push red couplings onto lateral dialysate connections.
Reset data with the dialysis machine switched on.
Preconditions
• The dialysis machine is switched on.
• The disinfection selection screen is displayed.
Touch icon.
A screen appears.
Touch icon.
The following screen appears:
10
Fig. 10-27 ”Select Disinfection” screen with active window ”Filter change”
10.6.4 Disinfection
The dialysis fluid filter is a fixed part of the dialysis machine for the entire duration of
its use. It is cleaned and disinfected together with the dialysis machine.
Suitable disinfectants
The following agents are suitable for disinfecting the dialysis fluid filter Diacap® Ultra:
• Citric acid 50 % (hot disinfection)
• TIUTOL® KF (only two times during runtime and immediately prior to a filter change)
Unsuitable disinfectants
The following substances may not be used for disinfecting the dialysis fluid filter:
• Chlorine-containing fluids and organic solvents, e.g. chloroform, acetone, ethyl
10 alcohol.
• Watery solutions, e.g. sodium hypochlorite (bleaching lye) or soda lye.
The manufacturer will not accept any liability if unsuitable disinfectants are used.
Samples can be taken at the dialyser feed connector (blue) via the piercing
membrane.
As an option, a sample port can be installed in the dialyser feed line. A sample can be
taken directly there with a suitable syringe with Luer connector.
pH 7,2-7,5
pCO2 B B 40-60 mmHg
-
HCO3 B PB
P
25-38 mmol/l
The battery operation mode serves for maintaining the extracorporeal blood
circulation in case of a mains power failure.
In such event, the dialysis machine automatically switches to battery operation.
• ”Battery/bypass” will be displayed in the status line.
• The remaining battery life is shown in the patient name field.
• This is followed by an acoustic signal.
• The alarm message ”Power failure – battery operation” is displayed.
This message must be confirmed.
An indicator light in the keyboard membrane of the screen indicates that the battery
is being charged while the system is operating on mains power. Battery charging
continues even when the machine is switched off. The indicator light goes dark as
soon as the battery is fully charged.
During the automatic machine test performed at each call-up for dialysis, the battery
function is tested, too. In case of an unsuccessful automatic test, an information
message appears. The test could be unsuccessful for the following reasons:
Cause Action
Battery not fully charged, e.g. because the machine has Charge battery.
not been connected to mains power for some time.
Battery change
To maintain the battery functionality, we recommend replacing the battery at least
every 5 years.
For correct disposal of the battery please refer to service manual.
As soon as the mains power supply has been restored, the battery operation is
automatically ended. The dialysate treatment is activated again. Once the unit has
adjusted itself to the set values, dialysis is automatically continued. User intervention
is not required.
The dialysis machine has a RS232 communication interface for communicating with
other information systems. Only one of the systems described below can be installed.
With the BSL, the dialysis machine can be connected to the data management system
Nexadia®. For further information see Nexadia® instructions for use.
The staff call is used for integrating the dialysis machine into an existing staff call
system.
The user may not solely rely on the function of the staff call option during the event
of an alarm. Regular checking of the status of the patient is still required.
10.9.1 Function
The Crit-Line unit from Hema Metrics™ is an external measuring instrument which
measures non-invasively various parameters in the blood via optical sensors. For this
purpose, a suitable cuvette (disposable measuring chamber) must be placed on the
dialyser inlet line.
The following parameters of the blood are measured or calculated:
The serial interface of the Hema Metrics™ Crit-Line device is connected to the rear of
the Dialog+ DSI (Dialog Serial Interface) using its serial connection cable.. The
P
P
P
P
measured data of the Crit-Line unit are, thus, shown on the display of the Dialog+, P
P
B. Braun provides the serial DSI for the Dialog+ to the Crit-Line unit.
P
P
B. Braun is neither responsible for the correct functioning of the Crit-Line unit or
10 guaranteeing the correctness of the displayed values.
Dialog+ displays the data measured by Crit-Line on the monitor.
P P
Do not connect the Crit-Line mains adapter to a different voltage phase than
the Dialog+ is connected to.
P
P
DANGER When operating the Crit-Line with the Dialog+ do not connect the serial data
P
P
adapter to a different interface (for example a PC) than that of the Dialog+. P
P
Do not connect ANY other serial device to the DSI interface than the Hema
Metrics™ Crit-Line device or other devices released from B. Braun for
operation with the DSI interface.
10
Risk to patient due to missing alarms because of incorrect limit values settings!
The reaching of alarm limits is indicated by Dialog+, however, this has no
P P
communication. A single fault condition with the Dialog+ software may result
P
P
Attach the cuvette between the arterial line and the dialyser.
Place the Crit-Line unit on the Combi Shelf Holder (Art. No. 7102871) behind the
monitor or on a safe, dry place.
In case catheter patients are treated:
- Connect potential equalisation cable (Art. No. 7106605) to potential equalisation
bolt on the back of the Crit-Line unit.
- Connect other end of the potential equalisation cable to the facility potential
equalisation connector.
Connect power plug of Crit-Line unit to wall or operate the unit on battery.
Plug in and tighten the Crit-Line unit serial adapter RS232 socket with the DSI
interface RS232 plug of the Dialog+.
P
P
For further information on Crit-Line set-up please read carefully the Crit-Line
operation manual.
After the blood pump starts in therapy mode, measurements on the Crit-Line unit
should start immediately. For all other settings on the Crit-Line unit, follow the
instructions in the associated instructions for use from Hema Metrics™.
10.9.3 Setting
With the Crit-Line option enabled by the technician in TSM and the Crit-Line
connected to the Dialog+, the message "CL enabled" is displayed in the upper part of
P
P
the screen.
10
10
Touching button 1 activates the function and the data transmission process from the
Crit-Line unit to the Dialog+ begins. The transferred parameters are displayed in fields
P
P
4 to 8, exclusively in therapy mode. The oxygen saturation limit is set with button 9
and the haematocrit limit with button 10. The default values can be accepted or
changed. An alarm is triggered when the values are exceeded.
The values are updated every 6 seconds. To obtain information concerning the portion
of the recirculation of the vascular access, a saline bolus must be administered. See
the relevant chapters in the Crit-Line instructions for use from Hema Metrics™.
When button 2 in Fig. 10-29 touched, the trends of haematocrit, blood volume change
in percent, recirculation, access blood flow and oxygen saturation can be displayed. A 10
trend group consisting of three trends is displayed on the screen. The trend groups can
be edited by touching the trend button (5) (see chapter 11.10).
The haematocrit and Delta BV limits are also displayed. The HCT limit (1) corresponds
to the value set in Fig. 10-29nd the BV limit (2) is calculated from the HCT start value
and the HCT limit. Line 3 shows the oxygen saturation limit.
The trend history (4) shows the actual progression as well as the last 20 saved trends.
1st option:
Directly enter the time (7) in the Time window.
2nd option:
Move the time reference line by using the icons << or >> (6).
Touch button 3 in Fig. 10-29 The saved data are read from the patient therapy card
and displayed. The table saves up to 50 Crit-Line progressions and includes the
following parameters:
A yellow background field states an abnormal deviation during the use of the Crit-Line
device. This can be the case in particular when the unit is not used during the
complete therapy. 10
Parameter Condition for abnormal deviation
Date/Time The first valid date/time information is not available until
5 minutes after the blood pump start.
HCT Start The first valid HCT value is not available until 5 minutes after
the blood pump start.
HCT Max. At least 5 successive HCT values are missing or are incorrect.
HCT End If the therapy will end in more than 5 minutes.
DeltaBV Min. At least 5 successive DeltaBV values are missing or are incorrect.
DeltaBV End If the therapy will end in more than 5 minutes.
Sat Min. At least 5 successive saturation values are missing or are
incorrect.
The table can also be called up in the End of Therapy phase by touching the
"Parameter" and "Folder" icons.
• For all the set conditions of alarms/warnings the monitor must be enabled (Crit-
Line button pressed in CL Parameter Window).
• If the monitor will be disabled (Crit-Line button released in CL Parameter
Window) all the alarms/warnings are reset.
• The occurrence of “Crit-Line communication failed” warning resets all the other
alarms and warnings.
• The “Set/check HCT Limit!” warning is immediately withdrawn if the HCT limit
button is pressed – even if the value is not modified. The limit may also be set
prior to the enabling: in this case the related warning does not occur at all.
10
Table of contents
11
11 Configuration
Legend
1 Select disinfectant
2 Thermal disinfection
3 Chemical disinfection
4 Short chemical disinfection
5 Rinse permeate inlet
6 Chemical disinfection with
disinfection solution from
central water supply
7 Thermal disinfection with hot
permeate
11
The weekly program “Weekly disinfection program” simplifies the configuration of the
operations.
11
11 The auto switch-off and the weekly disinfection program have to be activated in
TSM.
The dialysis machine can be configured in such a way that it automatically switches
on, carries out a disinfection and switches off again. The parameters of the automatic
disinfection can be set for one week.
Fig. 11-4 shows the set-up for the following disinfection modes:
Day/time Description
Monday
0.00 hrs. A thermal, central disinfection is carried out. The machine switches 11
on automatically and off again after disinfection.
6.30 hrs. A chemical disinfection with citric acid 50 % is carried out. The
machine remains switched on after disinfection.
Tuesday
0.00 hrs. The unit is rinsed from the central water supply. The machine
switches itself on automatically and off again after rinsing.
Wednesday
6.30 hrs. A chemical disinfection with citric acid 50 % is carried out. The
machine remains switched on after disinfection.
The automatic start of the preparation of the dialysis machine in the morning must
be activated in the service program.
With the setting ”Day/rinsing”, the dialysis machine changes to the ”Preparation/test”
mode after switch-on.
Upon completion of an automatic night action, the dialysis machine switches itself
off again.
Upon completion of an automatic day action, the dialysis machine remains in ”Rinse-
out” mode.
11
The setting of the parameters is explained using the conductivity (Na+) profile as an
example.
Legend
1 Profile settings
2 Enter therapy parameters
3 Heparinisation data
4 Pressure limits 11
5 Ultrafiltration data
6 Dialysate parameters
Touch icon 2.
Touch icon 6.
Touch icon 1.
Legend
1 Linear profile
2 Exponential profile
3 Parameter bar
4 Duration of parameter bar
5 Therapy time setting
6 Manual input of the total value
= resetting the profile to
horizontal shape
7 Value for the selected
parameter bar
11.5 UF profiles
Apart from the individual settings, the dialysis machine offers standardised
11
ultrafiltration profiles. As another option, an individual UF profile can be preselected
at any time and stored on the patient therapy card or via BSL after the dialysis
therapy. The profile table contains descriptions to the different profiles.
Touch icon.
The "UF parameters" screen appears.
Touch icon.
The "UF profile" screen appears.
The UF rate setting is specified above each parameter bar.
Legend
1 Profile number
2 Next profile number
3 Previous profile number
4 UF without dialysate
(sequential therapy)
5 UF with dialysate
Risk of dehydration!
A sequential therapy for a period of over 2 hours may only be set up on the
WARNING instruction of a doctor.
Risk of hyperpotassemia/hypercalcaemia!
Carrying out a whole therapy in “sequential” mode may lead to increased blood
values of the patient.
WARNING A sequential therapy for a period of over 2 hours may only be set up on the
instruction of a doctor.
During automatic calculation, the last bar is adjusted depending on the total value.
0 Standard profile -
Constant UF rate over entire therapy time.
11
1 Particularly gentle therapy start and end phases, due to stepped UF rate
increase/reduction.
11
11
7 Alternating stressful and gentle phases during the initial treatment period, with
constant fluid removal at the end of the treatment.
11
9 Alternating stressful and gentle phases during the second treatment period, with
constant fluid removal at the beginning of the treatment.
11
10 Freely settable.
When a given profile is changed, the machine assigns a name to it and stores it as
profile 10. When the profile is changed again, the previous profile is overwritten.
11 – 30 Profiles no. 11 to 30 can be preset by the technical service according to customer
requirements.
The patient therapy card offers the option of individually storing nearly all pre-
settings for a therapy and calling them up again at the preparation stage.
Also, treatment results from up to 50 therapies can be stored after a therapy.
The patient therapy card should be ordered from B. Braun to have a certified quality
standard.
Touch icon.
The selection menu appears.
Insert therapy card into drive.
Touch icon
11 All data on the patient therapy card will be erased!
Enter patient name in field 1 using the keyboard, and confirm with icon O.K.
When working with the patient therapy card, an additional field ”Patient number” is
displayed in the ”Patient Overview” screen. This helps to differentiate between
patients with the same name.
Patient data can only be read in the Therapy selection and Preparation modes.
By inserting the patient therapy card in Therapy selection or Preparation, data is read
automatically.
11
11.6.4 Storing patient data (parameter settings)
Machines operated with option Nexadia® BSL have other saving options. These are
described in the respective instructions for use.
Ensure that patient therapy card has been inserted into the dialysis machine.
Touch icon.
A screen for entering patient data for calculation of the theoretical effectiveness
opens.
11
Fig. 11-11 Input window for calculating the effectiveness (Kt/V values)
Fig. 11-12 Input window patient data after touching “Watson“ icon (Kt/UV)
Show table
Touch icon (11).
The Kt/V results are transferred from the patient therapy card and shown in a
screen:
Show graphics
Touch icon
A graphical display of the projected and actual Kt/V progression is shown.
11
3 Actual and current Kt/V Display of actual and current Kt/V progression
progression
6 Projected Kt/V aim Display of projected Kt/V result (O.K., aim Kt/V will be reached, has been
reached)
7 Projected Kt/V aim Display of projected Kt/V result (not O.K., aim Kt/V won’t be reached,
hasn’t been reached)
8 Cursor line Cursor line shows the current therapy moment
11
Kt/V will not be calculated in the therapy mode “Seq.” and “HF”.
End Kt/V
Touch icon on the ”Table Kt/V value” screen.
The screen is closed. All entered data are stored on the patient therapy card. When
closing the screen by touching the icon CANCEL, no data are stored.
Procedure
Touch icon.
The data management screen appears.
Touch icon.
The screen for adjusting the brightness is displayed.
To set the brightness manually:
Adjust brightness using the slide displayed on the screen.
”Manual” is displayed at the centre of the screen.
To set the brightness for daytime:
Activate field Daytime settings.
”Daytime” is displayed at the centre of the screen.
To set the brightness for nighttime:
Activate the field Nighttime settings.
”Nighttime” is displayed at the centre of the screen.
Screensaver
11 To activate the screensaver:
Touch the field Yes next to the field Screensaver on.
To deactivate the screensaver:
Touch field No.
To close the screen, touch the ”Brightness adjustment” icon.
Depending on the languages available in the TSM, you can choose the language for
the screen text.
Procedure
Touch icon
The data management screen appears.
Touch icon
The screen with all available languages appears.
11
You can edit the combination of parameters within the trend group.
Legend
1 Field trend group
2 Button “edit group”
3 Choose TSM presettings
4 Leave screen and save changes
5 Leave screen without saving
11
Fig. 11-18 Screen “Overview trend groups“
Single groups could be edited individually with parameters of your own choice.
Touch favoured button.
The following screen appears.
Legend
1 Field “trend group parameters”
2 Field “list of parameters”
3 Scroll bar “trend groups”
4 Scroll bar “list of parameters 2”
5 Leave screen
Touch icon.
11
Table of contents
12
12
Touch icon.
12
The touch foil will be deactivated for 10 seconds and can now be cleaned.
Do not wipe the monitor too moistly. If necessary, dry with smooth cloth afterwards.
Do not repeatedly put the blood pump rollers into a disinfectant bath, otherwise the
return safety device can be destroyed.
12
If the set maintenance interval is reached, the above mentioned window appears at
every change to disinfection mode.
The battery should be replaced after 5 years to maintain the full functionality of this
option. For correct disposal of the battery please refer to service manual.
The technical safety inspection shall be performed and documented every 12 months,
according to the specified check list in the service manual and with reference to the
instructions for use.
The dialysis machine should be checked by persons who have been appropriately
trained or have the required expertise or experience and do not require instructions
for the check.
Results of the technical safety inspection shall be documented, e.g. by applying an
inspection label to the dialysis machine.
The technical safety inspection must be kept by the responsible organisation (user)
as part of their documentation.
Only accessories, disposable items and spare parts may be used that do not pose a
technical safety risk and demonstrably comply with Medical Devices Directive
93/42/EEC (MDD).
To ensure the full functionality of the dialysis machine, we recommend using products
manufactured by B. Braun Avitum AG.
For the dialysis machine, B. Braun Avitum AG provides a warranty in line with
statutory requirements. The warranty includes the repair or replacement of parts that
have been damaged because of design, manufacturing or material faults.
The warranty becomes void if the owner or third parties carry out modifications or
repairs to the dialysis machine.
The warranty does not include the remedying of faults caused by manipulation,
incorrect treatment or normal wear.
12
Table of contents
13
13
Legend
1 Comment field
2 Information field
3 Alarm field
4 Alarm list
5 Call up comments
13
The service technician can activate an alternative alarm sound in the TSM, which
differ from the continuing alarm sound in an alternating melody.
At failure or disturbance of the loudspeakers, the security system will activate the
power supply buzzer to report an alarm acoustically.
Please inform your service technician.
Alarms are shown in the alarm list in the order of their occurrence. The triggering
alarm is shown in the alarm field.
Upon resetting the triggering alarm all subsequent alarms are also deleted.
The user is responsible for the reset of an alarm and subsequently for the monitoring
of the suppressed parameters of the dialysis machine.
Arterial pressure upper limit Patient access malfunction. Check patient access.
(code 1050)
The arterial suction pressure exceeds Incorrect limits setting. Re-adjust limit.
the preset arterial upper limit.
Incorrect position of cannula. Correct position of cannula.
Arterial pressure lower limit Excessive pump speed. Adjust blood flow to patient
(code 1051) conditions.
The arterial suction pressure is lower
than the preset arterial lower limit. Incorrect limits setting. Re-adjust limit.
Check AV line for PA monitoring A connection of the arterial line was If an AV line for pressure
(SUP) (code 1980) not detected at PA. measuring is present, please
connect it to the PA pressure
sensor.
PBE upper limit (code 1048) Excessive pump speed. Adjust blood flow to dialyser
The blood-sided dialyser inlet pressure and tube conditions.
exceeds the set limit. Readjust limit.
PBE lower limit (code 1049) Tube leaking. Check tubing system.
The blood-side dialyser inlet pressure
dropped below 10 mmHg. Tube kinked downstream of blood
pump.
Venous pressure upper limit Pump speed too high. Adjust blood flow to patient
(code 1052) conditions.
Venous pressure exceeds upper limit.
Incorrect position of cannula. Correct position of cannula.
Venous pressure lower limit – Check Pump speed too low. Adapt blood flow to patient
access (code 1053) conditions.
The venous pressure is lower than the
lower limit. Pressure measuring leaking, leading Produce leak-free connection,
to blood level rise up to the push back fluid column with
syringe.
hydrophobic filter.
Phase volume too low (code 1067) Phase volume significantly lower Reduce blood flow.
Pressure rise at venous pressure sensor than average. If necessary, continue adjustment
too high during pump phase. of control pressures.
Check position of needle/catheter.
Alarm limits: Min. 12 ml
Phase volume too high (code 1064) Phase volume significantly lower Check that all arterial patient
Pressure rise at venous pressure sensor than average. tube is in place in the arterial
too low during pump phase. clamp.
Increase blood flow.
Possibly set closer control
Alarm limits: Min. 12 ml pressures.
Check tube system for leaks and
entering air.
DF and/or HDF filter holder open Filter holder open at rear: Close filter holder. A filter change
(code 1047) is only intended to happen in the
- during Preparation
disinfection phase.
- in Therapy or End of Therapy
mode
- during on-going disinfection
HDF-online filter test failed! Leakage. Repeat filter test. Automatic test
(code 1151) for Online must be repeated by
13 the machine and must be
successful.
Check filter for leakage.
If there is no leakage, change
filter.
SAD (SUP) (code 1958) Air in SAD region. Remove air, see text on screen.
SAD – Air! (code 1058)
Level drop in bubble catcher Correct level in bubble catcher.
Blood pressure too high (foam
formation).
SAD: sensor error (SUP) (code 1961) External sound pulses. Disconnect patient.
SAD - sensor error (code 1059)
Measuring frequency smaller than
600 Hz.
SAD function ref. (SUP) (code 1962) Technical defect. Contact technical service.
The alarm level is outside the calibration
value +/-50 mV.
Pump cover open (arterial) Blood pump cover opened. Close blood pump cover.
(code 1062)
The reed contact in the blood pump Technical defect. Contact technical service. 13
housing has detected that the blood
pump lid was opened with the pump
running.
Pump cover open (SN/Subst) Blood pump cover open. Close blood pump cover.
(code 1063)
The reed contact in the blood pump Technical defect. Contact technical service.
housing has detected that the blood
pump cover was opened with the pump
running.
Substitution disturbed - leakage? Substitution volume on scales differs Check tube system for leaks and
(code 1089) from the total substitution volume. kinks.
Water supply disturbed (code 1022) Water pressure too low. Check water inlet pressure
15 s after activation of the lower reed (minimum pressure 0.5 bar).
contact, the middle reed contact has
not been reached again. The inlet valve Water tap closed. Open shut-off valve.
is open during this time.
Water inlet tube kinked. Check inlet tube.
Technical defect.
13 Degassing insufficient (code 1111) Technical defect. If alarm cannot be reset, contact
The degasification pressure does not technical service.
reach the set value.
Temperature too high (code 1034) Irregular dialysate flow. If alarm cannot be reset, contact
The mean temperature over a filling technical service.
chamber cycle (250 ms sampling time) Technical defect.
measured at TSD exceeds 41 °C.
Temperature too low (code 1033) Irregular dialysate flow. If alarm cannot be reset, contact
The mean temperature over a filling technical service.
chamber cycle (250 ms sampling time) Technical defect.
measured at TSD was 1 °C lower than
the set value.
Bicarbonate conductivity limit Suction rod not correctly inserted in Check position of rod in
(code 1028) container. container.
The average value during a filling
chamber cycle (250 ms sampling time) Concentrate container empty. Replace container.
measured at BICLF deviates by more
than +/-5 % from the preselected value. Defective suction line. Replace suction line.
Bicarbonate mixing ratio (code 1030) Incorrect concentrate used. Connect correct concentrate.
The mixing ratio H2O to bicarbonate
concentrate is outside of the +/-7 Incorrect composition of Where mix is produced on site,
tolerance around the preset ratio. concentrate. observe mixing ratio
powder/water.
Bicarbonate mixing ratio (SUP) Incorrect concentrate used. Connect correct concentrate.
(code 1950)
The mixing ratio H2O to bicarbonate Incorrect composition of Where mix is produced on site,
concentrate is outside of the +/-7 concentrate. observe mixing ratio
tolerance around the preset ratio. powder/water.
Final conductivity limit (code 1029) Suction rods not correctly inserted in Check position of rods in
container.
13
The average value during a filling container.
chamber cycle (250 ms sampling time)
measured at ENDLF deviated by more Concentrate container empty. Connect new container.
than +/-5 % from the preset value.
Suction line defective. Replace suction line.
Final conductivity limit (SUP) Suction rods not correctly inserted in Check position of rods in
(code 1951) container. container.
The average value during a filling
chamber cycle (250 msec. sampling Concentrate container empty. Connect new container.
time) measured at ENDLF deviates by
more than +5 % from end conductivity Suction line defective. Replace suction line.
confirmed as ”maximum”, or by more
than -5 % from the level confirmed as Technical defect. Contact technical service.
”minimum”.
Concentrate mixing ratio (code 1031) Incorrect concentrate used. Connect correct concentrate.
The mixing ratio H2O plus possibly BIC
concentrate to acetate/acid Concentrate container empty. Connect new container.
concentrates was outside the permitted
range over two filling chamber cycles. Suction line defective. Replace suction line.
Bicarbonate cartridge not correctly Bicarbonate cartridge probably not Check if it fits the holder
connected lanced correctly. correctly.
(code 1018)
The cartridge could not be filled
UF balance? Air leakage in dialyser Dialyser not filled without air Vent dialyser (water side).
couplings (code 1026) inclusions.
During the dialysis, the valve (VLA) had
to be opened more than 20 times due to Dialyser connection leaking. Check dialyser connections.
air (level below bottom electrode) in the
bubble catcher. Technical defect. Contact technical service.
DF pressure < –400 mmHg Dialyser UF factor too small for set Use dialyser with larger UF factor.
(code 1020) UF rate.
The dialysate pressure downstream of
the dialyser (PDA) is below –400 mmHg. Too high UF volume set. Reduce UF volume/increase UF
time.
Max. UF rate exceeded (SUP) UF volume too high. Decrease UF volume, increase UF
(code 1953) time
The required UFR is 20 ml/h higher than
the specified max. UFR. Dialyser factor too small. Use dialyser with higher factor.
Blood leak, sensor dirty (code 1044) Sensor dirty. Call technical service.
The blood concentration measured at
the sensor (BL) is negative. Air on dialysate-side. Reset alarm.
System restored after power failure Technical defect. The system was Reestablish power supply.
(code 600) restored.
Power failure during
Preparation/Treatment
Blood recognised in Patient connected in If self test isn’t ended, disconnect
Preparing/Disinfection Preparing/Disinfection. patient.
The red detector has recognised blood in If self test is ended, change to
Preparing/Disinfection. Therapy.
Art. bolus vol. > 300 ml (SUP) (code Please call a technician.
13 2026)
Arterial bolus volume exceeded 300
ml.
ABPM: Systolic pressure too Systolic pressure exceeds the set Repeat measurement.
high (code 9100) upper limit. Select individual limit adjustment.
Manually change individual limits.
Inform doctor.
ABPM: Systolic pressure too Systolic pressure is below the set Repeat measurement.
low (code 9101) limit. Select individual limit adjustment.
Manually change individual limits.
Inform doctor.
ABPM: Diastolic pressure too Diastolic pressure exceeds a set Repeat measurement.
high (code 9103) upper limit. Select individual limit adjustment.
Manually change individual limits.
Inform doctor.
ABPM: Diastolic pressure too Diastolic pressure drops below a set Repeat measuring.
low (code 9104) lower limit. Select individual limit adjustment.
Manually change individual limits.
Inform doctor.
ABPM: Internal communication ABPM not operational; no further Carry out pulse measuring with separate
disturbed (code 9138) measurements possible. RR device, or manually.
ABPM: Pulse rate too high Pulse frequency exceeds upper limit. Repeat measurement.
(code 9169) Select individual limit adjustment.
Manually change individual limits.
Inform doctor.
ABPM: Pulse rate too low Pulse frequency lower than lower Repeat measurement.
(code 9170) limit. Select individual limit adjustment.
Manually change individual limits.
Inform doctor.
ABPM: Air leak – check cuff Blood pressure module has Check connections to ABPM and cuff.
connection (code 9300) performed a safety switch-off. Switch dialysis machine off and on
again; all data remain stored.
ABPM: Module failure please Blood pressure module has Switch dialysis machine off and on
switch off/on (code 9301) performed a safety switch-off. again; all data remain stored.
Message 9172 remains on the
display after confirmation.
ABPM: Inflation pressure not reached Check cuff for correct position.
(code 9302) Re-apply cuff if necessary.
—
Repeat measurement.
ABPM: Systolic BP > max. cuff pressure Considerable increase in blood Carry out pulse measuring with
(code 9305) pressure since last measurement. separate RR device, or manually.
ABPM: Irregular pulse (code 9307) Check cuff for correct position.
— Carry out pulse measuring with
separate RR device, or manually.
ABPM: Reading took too long The max. measuring time of 110 Carry out pulse measurement with
(code 9308) seconds is exceeded. separate RR device, or manually.
ABPM: Pulse beat over 100 beats The max. measuring time of 110 Carry out pulse measurement with
(code 9309) seconds is exceeded. separate RR device, or manually.
ABPM: Cuff pressure > 320 mmHg Cuff pressure has exceeded the Check patient's arm position.
(code 9310) cuff pressure limit. Carry out pulse measurement with
separate RR device, or manually.
ABPM: Pulse signal very low Check cuff for correct position.
(code 9311) — Carry out pulse measurement with
separate RR device, or manually.
13
13
bioL. RR SYS Lower Limit Max. SYS Lower Limit of 130 mmHg Press bioLogic RR® button.
reduced (code 3102) exceeded (value 130 mmHg is valid Reduce SYS Lower Limit to max.
130 mmHg.
for bioLogic RR®).
bioL. RR Missing reading 3 minutes without successful blood Press bioLogic RR® button. Alarm
(code 3103) pressure reading since the request disappears automatically.
of bioLogic RR® algorithm.
13
HDFO: VSB or VSAA opened (SUP) Technical defect. Press ”Reset alarm” button to reset.
(code 2021) If not possible, call technician.
Substitution connection outlet Disconnect patient.
Make a disinfection.
valve (VSAA) is open.
HDFO: VBE opened (SUP) Technical defect. Press ”Reset alarm” button to reset.
(code 2022) If not possible, call technician.
Filter vent valve (VBE) is open. Disconnect patient.
Online dialysis not possible.
HDFO: DF system not rinsed (SUP) Technical defect. Press “Reset alarm” button to reset.
(code 2023) If not possible, call technician.
Water part not rinsed after Disconnect patient.
disinfection.
HDF Inf. bolus volume too high Supervisor detected too high Press “Reset alarm” button to reset.
(SUP) (code 2030) bolus rate. If not possible, call technician.
Disconnect patient.
Ratio gross UF/blood flow too low UF rate exceeds a defined Increase blood flow.
(code 2059) percentage of the blood flow. Reduce substitution flow.
HDF/UF alarm limits expanded Any of the UF TLC or LLS alarm Leave Therapy or End of Therapy.
(code 2070) limits have been expanded.
13
In case of air in the area of the SAD, the tube clamp (SAKV) is closed due to the alarm
action. Due to the reaction time of the system, a small amount of air could be below
the SAD, too, in case of SAD alarms.
In case of a power failure during dialysis and where no emergency power supply is
available, the blood must be returned manually to the patient.
The crank for manual blood return may be one of two alternatives (see following
pictures).
13
Observe, using the crank on the picture above, to insert it in the external hole to ease
the manual rotation.
13
ID Text
600 System restored
ID Text
1900 The selected interval is over
1903 Selected UF volume too high
1904 Selected UF volume too low
1905 Selected UF time too high
1906 Selected UF time too low
1907 Interval cannot be modified
1908 Max. UF ratio has profile modified
1911 Selected Heparin rate too high
1912 Selected Heparin rate too low
1922 UF volume has been decreased
1934 Rinsing time too long
1935 Rinsing time too short
1936 UF rinsing volume too high
1937 UF rinsing volume too low
1942 Acknowledge data before connecting patient
2056 No Heparin bolus
2060 Please press longer EQ button again
2066 UF + HDF rate > 5500 ml/h please reduce!
13 2073 Rinsing rate too low
2074 Rinsing rate too high
1093 Pump cover open (substitution)
1054 Preparation of new Bic Cartridge-bypass
Table of contents
14
14
14 Accessories
14.1 Options
bioLogic RR® Comfort for card reader: Automatic blood pressure 7105324
stabilisation with template method (only with option ABPM)
Bicarbonate cartridge holder 7105171
DF filter 7102102
Adimea 7102233
14
* Because of defined cable length the marked items are only allowed to be used in
order to meet the standard according to EMC (electromagnetic compatibility)
directive.
Box 7107320
Universal base storage (max. 30 kg, e.g. single bed - RO unit) 7105500
Table of contents
15
15
15 Technical data
Description Values
Current-voltage-rating 20 A at 120 V
16 A at 230 V
Minimum cross section of 3.3 mm2 (AWG12) for 120 V 1.5 mm2 for 230 V
single strand
Withstand voltage of single min. 2 kV, 50 Hz, ≥1 min
strand (L-N, L-PN, N-PE)
Average energy consumption approx. 1.5 kW/h
Protection class IP21 (Protection against foreign bodies > 12 mm and vertically
falling drip water)
DIN EN 60529
Electrical ground via optional cable 15
Dimensions (W × H × D) approx. 510 × 1678 × 637 mm
Description Values
For information regarding fuse ratings and battery specifications please refer to
service manual.
Description Values
Operation
Relative humidity 15 % – 70 %
Relative humidity 15 % – 80 %
15
The dialysis machine Dialog+ is made for the use in ambient conditions with controlled High-
Frequency disturbance variables. The user can avoid electromagnetic disturbances by keeping
the distance between Dialog+ and HF-telecommunication devices following the values in the
table below in dependency to the output power of those devices.
Nominal output P of Safe distance (d) depending on transmitting frequency
transmitter
(Watt)
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
d = 1.2 P d = 1.2 P d = 2.33 P
For transmitters with other output power ratings, the recommended safety distance
(m) can be calculated with the above formulas. Heed the max. power rating (W), in
accordance to the manufacturer’s information, to use the formula from above.
Remark 1: For 80 MHz and 800 MHz use the higher frequency range.
Remark 2: This guideline may be not practicable in some cases. The propagation of
electromagnetic quantity will be influenced by adsorption and reflexion of the
building, equipment and human.
15
Description Values
Excessive-temperature 41 °C
protection
Protection system Temperature sensor
Description Values
Tolerance 10 %
15
* The overall accuracy F is the sum of two different errors:
F = Fbal +FUF
Fbal = balance chamber deviation (measures per chamber cycles and depends on the
dialysate flow)
FUF = UF pump error
Description Values
Disinfection
During disinfection processes dialysis is blocked.
Reports on the effectiveness of the individual disinfection programs can be obtained from the
manufacturer
Disinfection/cleaning Automatic program with enforced rinse-out
The parameters for the disinfectant used can be set in the service
program.
HDF-online and dialysis fluid filter option: Only disinfectants
cleared for the dialysis fluid filter can be used.
Thermal disinfection Automatic program cycle at approx. 85 °C at the dialyser
couplings
15
Description Values
Blood pump 2-roller pump with automatic motor switch-off when lid is
opened, backstop, low haemolysis.
For 8/12 mm or (optional) 7/10 mm pump tubes
Pumping rate 50 – 600 ml/min (8/12 mm)
50 - 400 ml/min (7/10 mm)
Adjustable in 10 ml steps
Tolerance interval < 10 % for blood pressure up to -150 mmHg
Tolerance between 10% and 25% for blood pressure up to -200
mmHG
Working pressure range Intake pressure up to –390 mm Hg
Pumping pressure 0 – 1725 mm Hg
Heparin pump Syringe pump for 10 – 30 ml syringes
Pumping rate 0.1 – 10 ml/h in steps of 0.5 ml/h or 0.1 ml/h, can be turned off,
bolus: 600 ml/h
Tolerance interval < ±10 %
Sensitivity Air bubbles at > 50 µl, micro foam with cumulated volume
Limit values double-needle:
0.2 ml at 0 – 200 ml/min SAD flow
0.3 ml at 200 – 400 ml/min SAD flow
0.5 ml at > 400 ml/min SAD flow
Limit values single-needle:
0.7 ml at 1200 ml/min constant SAD flow
Protection system Ultrasound detector, automatic cyclical checks during entire
operating phase
Bridging time of protective Not bridgeable during dialysis
system only in end phase for approx. 30 ml (has to be activated at TSM)
Red sensor in SAD housing
15
Function Detects blood in tube system
Modes of operation 1st mode of operation: The blood pump is stopped as soon as the
red sensor detects blood when connected at this site. Alarm
caused
2nd mode of operation: If the red sensor detects blood at this
point, a heparin bolus is administered. This function can be
disabled by technical service in the service program.
3rd mode of operation: If no blood is detected at this site in the
End Therapy mode, the blood pump is stopped.
4th mode of operation: If blood is detected during preparation or
disinfection, the blood pump is stopped Alarm caused
Description Values
Arterial inlet pressure (PA) Electronic pressure sensor with digital quasi-analogue display
measurement
Operating range –400 to +400 mmHg
Ceramics
Glass
Graphite
Polyester
Polyetheretherketone PEEK
Polyetherimid PEI
Polyethylene PE
Polyisoprene
Polymethylmethacrylate PMMA
Polyoxymethylene POM
Polyphenylsulfone PPSU
Polypropylene PP
Polytetrafluoroethylene PTFE
Silicone
15
Stainless steel
Description Values
HDF (Haemodiafiltration)
HF (Haemofiltration)
Online filter
Operating time 150 treatments (900 h), see filter instructions for use
15
Description Values
15
The intended use of the DSI interface is to connect the Crit-Line III TQA device of
Hema Metrics™ to the Dialog+ or other devices released by B. Braun for operation.
It is prohibited to connect any other device.
Description Values
Maximum voltage level (all pins in relation to GND (ground)-level) ±25 VDC
15
Table of contents
16 Appendix............................................................................................. 16-3
16.1 Flow diagrams..................................................................................... 16-3
16.1.1 Key to flow diagrams ............................................................................ 16-3
16.1.2 Flow diagram Dialog+............................................................................ 16-5
16.1.3 Flow diagram Dialog+ HDF-online ..................................................... 16-6
16.2 Technical safety inspection and preventive maintenance ............. 16-7
16
16
16 Appendix
EP Degassing pump
FPE Inlet flow pump
FPA Outlet flow pump
PA
PBE
Arterial pressure sensor
Blood inlet pressure sensor 16
PBS Blood control pressure sensor
PDA Dialysate outlet pressure sensor
PE Degassing pressure sensor
PV Venous pressure sensor
VABK1/2, VBE,
VBICP, VBKO,
VBKS, VBP,
VDA,
VDABK1/2,
Solenoid valve
VDE,
VDEBK1/2,
VEBK1/2, VLA,
VSAE, VSB,
VVB, VVBE, VZ
16
16
16
The technical safety inspection shall be performed and documented every 12 months
according to the specified check list in the service manual and with reference to the
instructions for use.
The preventive maintenance is recommended every 12 months according to the
specified check list in the service manual and with reference to the instructions for
use.
16
16