Al-Balqa Applied University: Edit by
Al-Balqa Applied University: Edit by
“ventilator”
Edit by :
1-Mohmmad Saad Yousef Bsileh
2-Mahmoud Ali Hasan Qunees
Mechanical ventilator
Actually is According to the definition offered by Chatburn, a
mechanical ventilator is an automated machine in which
A ventilator blows air into the airway through a breathing tube. One
end of the tube is inserted into patient’s windpipe and the other end is
attached to the ventilator. The breathing tube serves as an airway by
letting air and oxygen from the ventilator flows into the lungs.
Depending on the patient’s medical condition, they may be able to use
a respiratory mask instead of the breathing tubes.
Input
Mechanical ventilators are typically powered by electricity or
compressed gas. Electricity, either from wall outlets (e.g., 100 to 240
volts AC, at 50/60 Hz) or from batteries (e.g., 10 to 30 volts DC), is used
to run compressors of various types. Batteries are commonly used as
the primary power source in the home-care environment but are
usually reserved for patient transport or emergency use in hospitals.
These sources provide compressed air for motive power as well as air
for breathing. Alternatively, the power to expand the lungs is supplied
by compressed gas from tanks, or from wall outlets in the hospital (e.g.,
30 to 80 pounds per square inch [psi]). Some transport and emergency
ventilators use compressed gas to power both lung inflation and the
control circuitry. For these ventilators, knowledge of gas consumption
is critical when using cylinders of compressed gas.
Directing flow from the source gas into the patient requires the
coordination of the output flow-control valve and an expiratory valve or
“exhalation manifold” . In the simplest case, when inspiration is
triggered on, the output control valve opens, the expiratory valve
closes, and the only path left for gas is into the patient. When
inspiration is cycled off, the output valve closes and the exhalation
valve opens, flow from the ventilator ceases and the patient exhales
out through the expiratory valve . The most sophisticated ventilators
employ a complex interaction between the output flow-control valve
and the exhalation valve, such that a wide variety of pressure, volume,
and flow waveforms may be generated to synchronize the ventilator
output with patient effort as much as possible.
Controlled system
The ventilator can be scheduled to “breathe” for the patient a set
number of times each minute, or it can be set so the patient can trigger
the machine to deliver air. If the patient fails to trigger the machine
after a set period of time, the ventilator will automatically blow air into
the breathing tube.
Output
Output waveforms are conveniently graphed in groups of three. The
horizontal axis of all three graphs is the same and has the units of time.
The vertical axes are in units of pressure, volume, and flow. For the
purpose of identifying characteristic waveform shapes, the specific
baseline values are irrelevant. What is important is the relative
magnitudes of each of the variables and how the value of one affects or
is affected by the value of the others.
a)Sources of power:
Gas supply
Power supply
Pressure generator
b) Control of gas delivery:
Gas blender
Gas accumulator
Inspiratory flow regulator
Humidification equipment
Patient circuit
Expiratory pressure regulator (i.e PEEP valve)
Monitoring
c) Sensors
Gas concentration
Flow
Pressure
Volume
Safety features
d) Filters
Gas intake particle filters
Pre-circuit bacteria filters
Moisture traps and heat/moisture exchange systems
Expired gas filter
1) Input
Pneumatic
Electric
AC
DC (battery)
2) Power conversion and transmission
External compressor
Internal compressor
Output control valves
Control scheme
Control circuit
Mechanical
Pneumatic Fluidic
Electric
Electronic
Control variables
Pressure
Volume
Time
Phase variables
Trigger
Target
Cycle
Baseline
Modes of ventilation
Control variable
Breath sequence
Targeting schemes
3) Output
Pressure waveforms
Rectangular
Exponential
Sinusoidal
Oscillating
Volume waveforms
Ascending ramp
Sinusoidal
Flow waveforms
Rectangular
Ascending ramp
Descending ramp
Sinusoidal
4) Alarm
Input power alarms
Loss of electric power
Loss of pneumatic power
Control circuit alarms
General systems failure
Incompatible ventilator settings
Warnings (e.g., inverse inspiratory-to-expiratory timing ratio)
Output alarms (high/low conditions)
Pressure
Volume
Flow
Time
Frequency
Inspiratory time
Expiratory time
Inspired gas
Temperature
FIO2
Type ventilator
1) Negative Pressure
2) Positive pressure
*such ventilator replaced the iron lungs as safe endotracheal tubes with
high –volume/flow-pressure cuffs were developed. The popularity of
positive pressure ventilators rose during the palio epidemic in the 1950.
Risk
The main risk of mechanical ventilation is an infection, as the artificial
airway (breathing tube) may allow germs to enter the lung. This risk of
infection increases the longer mechanical ventilation is needed and is
highest around two weeks. Another risk is lung damage caused by
either over inflation or repetitive opening and collapsing of the small air
sacs Alveoli) of the lungs. Sometimes, patients are unable to be weaned
off of a ventilator and may require prolonged support. When this
occurs, the tube is removed from the mouth and changed to a smaller
airway in the neck. This is called a tracheostomy. Using a ventilator may
prolong the dying process if the patient is considered unlikely to
recover.