Laboratory Flotation Rate Test Procedure For PGM, Base Metal Sulphide and Oxide Ores
Laboratory Flotation Rate Test Procedure For PGM, Base Metal Sulphide and Oxide Ores
APPLICATION
This test procedure applies to a single stage MF1 (i.e. primary Mill-Float) rate test. See
DEFINITIONS
A rate test is where a number of concentrates are collected over various time periods in
order to generate recovery-time, grade-time and mass-time curves. The data is used to
estimate flotation kinetics.
A batch test is when a single concentrate is collected for the full duration of the test.
OBJECTIVE
The purpose of this procedure is to generate information to achieve the following;
Characterise the ore
Determine metallurgical variability of samples, ore types and ore mixes from the
deposit
Ascertain what metallurgical changes result in an improvement of
recovery/grade such as changes in grind, reagent type and/or addition, pH, pulp
density
Generate data to determine flotation kinetics as a means of optimising flotation
plant performance
EQUIPMENT
There are a number of standard laboratory flotation machines and cell sizes. The Denver
D1 and D12 machines are the most commonly used (Figure 1) and come with a variety of
cell and rotor/stator sizes from 1.5 to 8.0 l. (Figure 2). Tests using the 8.0 l cell require a
larger rotor/stator. The LabTech Essa machine with 2.5 l cell is shown in Figure 3 and
Figure 4 (note that annotations on these pictures are equally applicable to the Denver
machine).
Figure 1
Figure 2
Figure 3
Level of pulp
10-15 mm
Figure 4
Eurus Mineral Consultants
Laboratory Float Procedure. Update 4 January 2012.
Page 5 of 21
6
SET-UP
Agitation and Air
Usual set-up conditions for 1.5 to 8.0 l cells are shown in Table 1. Cell air flowrates and
cell rpms are plotted in Figure 5. These running conditions are guidelines and are not
“cast in stone”. Cell rpm is chosen, usually on a visual basis, to provide adequate
agitation and suspension of solids but should not be too intense to cause spillage. The
intensity of agitation reduces once air is introduced into the system.
Cell volume and dimensions in Table 1 are detailed in Figure 6. With agitation and
airflow, pulp level should be about 10-15mm below the froth overflow lip. As per the
illustration in Figure 6, this measurement is approximately the length of a thumbnail (the
proverbial “rule of thumb”!). The froth paddles are sized to extend 10mm below the
overflow lip. When air is turned on, pulp level rises and it should rise to 10-15mm below
the froth overflow lip as represented in Figure 4. The cell should be marked and
calibrated for pulp level with agitation and no air so that with airflow, pulp level is 10-
15mm below the froth overflow lip. Froth height should be level with the overflow lip.
Guidelines for air addition are given in Table 1. During a test-run an appropriate air set-
point was established at about 50% on the rotameter indicated by the line and arrow in
Figure 3. Pulp height was about 15mm below the overflow lip (see Figure 4).
Sample Mass
The Denver 2.5 l cell has a live volume of 2.4 l. This is sufficient for 1,0 kg of sample and
gives 32% solids in the cell which is a typical float plant feed density. In this case the
laboratory mill should be set-up for 1.0 kg of sample as it is practical and best to mill the
same mass of sample that will be used in the subsequent flotation test.
The LabTech Essa flotation cell has a live volume of 2.0 l. To float at between 30 and 35%
solids Table 2 below shows that 800 gr of solids meets this requirement and allows about
1.25 l of water to be used to wash the solids from mill to float cell.
From a visual analysis using the 2.5 l cell, the Essa flotation machine provided adequate
agitation at 1200rpm. This is the maximum that the machine can operate at. Usual rotor
speeds for the Denver cell are shown in Table 1.
Eurus Mineral Consultants
Laboratory Float Procedure. Update 4 Jan 2012.
Page 6 of 21
Measurements for the Denver D12 Laboratory Machine and
Cells
1.5 litre Cell 2.5 litre cell 5.0 litre cell 8.0 litre cell
A mm 115 135 155 195
B mm 115 135 155 195
C mm 165 205 240 295
D mm 120 160 215 260
Table 1
FLOAT CELL
Ore SG (t/m³) 2.7 Choice Choice
Lab Cell
2.0 2.0 2.0 2.0 2.5 2.5 2.5
volume (l)
% solids 30.0 35.0 38.0 32.0 30.0 35.0 32.0
Pulp RD 1.233 1.283 1.315 1.252 1.233 1.283 1.252
Pulp (kg) 2.47 2.57 2.63 2.50 3.08 3.21 3.13
Solids (kg) 0.740 0.898 1.000 0.800 0.925 1.122 1.000
Water (l) 1.73 1.67 1.63 1.70 2.16 2.08 2.13
MILL
Mill % Solids 65.0
Pulp RD 1.693 Choice Choice
Pulp (kg) 1.138 1.381 1.538 1.231 1.423 1.727 1.538
Solids (kg) 0.740 0.898 1.000 0.800 0.925 1.122 1.000
Water (l) 0.40 0.48 0.54 0.43 0.50 0.60 0.54
Table 2
Cells Airflow per Active Volume Airflow per Nominal Volume RPM
13.0 2000
12.0
1800
11.0
1600
10.0
Airflow per Active Volume (min^-1)
Usual Air Flowrate (l/min) &
9.0 1400
8.0 1200
Cell rpm
7.0
1000
6.0
5.0 800
4.0 600
3.0
400
2.0
200
1.0
0.0 0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5
Cell Volume (l)
Figure 5
15mm
Level of
pulp
D
A
B
Figure 6
Froth Paddles
Froth paddles should be sized and shaped as per the illustration in Figure 7. The paddles
should be half the width of the cell, extend 10mm below the froth overflow lip and have
handles. When collecting froth the paddle should sit on the cell wall so that froth
collection depth is constant. The paddle can then slide along the cell wall from back to
front ie “A” to “B” in Figure 11.
Mechanism
Froth Paddle
Overflow Lip
A B
Pulp Level
A: 10mm below
overflow lip
B: 15-25mm
below overflow lip
Figure 7
In points 3 and 4 above, flotation concentrates should be timed so that the shape and
position of the fast and slow floating sections of the curve can be well delineated and the
slow floating section should be approaching or have approached a plateau, ie the
recovery-time curve should have two distinct components with a discernable inflection
point as per the bottom curve in Figure 9. If the recovery-time curve trend is described by
a straight line as shown by the mass curve (red line, triangle markers) in Figure 10 then
the test should be redone for a longer time to generate a curve with two distinct fast and
slow sections. An example of this is shown in Figure 10 where the original test had to be
redone for an extended flotation time from 23 to 40 minutes.
From a cost and sample processing point of view the minimum number of concentrates to
adequately describe the recovery-time curve and the ore’s flotation characteristic is four
timed at 1, 3, 10 and 20 minutes. The first concentrate has the smallest mass and if this is a
problem for assay, then the first two concentrate collection times can be moved to 2 and 4
minutes without adversely affecting the accuracy of estimating fast floating kinetics.
The table below recommends concentrate collection times for three categories of ore
according to how fast mineral and gangue float and how quickly the recovery-time curve
forms a plateau. The choice of collection times should suit the slowest floating
component which is usually mass. For example, a 20 minute float would suit the ores in
Figure 9 and the top curve in Figure 10, but only a 40 minute float would be suitable for
an ore with a mass-time curve in Figure 8 and the bottom curve of Figure 10.
A B C
Slow Floating Ores
Average to Fast Average to Slow
and/or linear mass-
Floating Ores Floating Ores
time ores
Time Time Time
Conc Conc Conc
(min) (min) (min)
1 1 1 1 2 1
3 2 3 2 5 2
7 3 15 3 20 3
20 4 30 4 40 4
Alternately: Alternately:
Time Time
Conc Conc
(min) (min)
1 1 2 1
3 2 5 2
7 3 10 3
20 4 25 4
30 5 40 5
The following excerpt from “Using Simulation to Understand Metal and Mineral
Flotation Performance at one of Falconbridge’s Base Metal Operations”; M. P. Hay et al,
Nickel 05 Conference, Cape Town serves to outline the importance of not generating a
linear recovery-time curve.
where kGF=kGS=0.0160 with θ being any value – as described by the dashed line.
30
25
20
15
10
0
0 2 4 6 8 10 12 14 16 18 20 22 24 26
TIME (min)
100
95
Slow Floating
90
85
LABORATORY RECOVERY
80
75
Fast Floating
Slow
70
65
60
55
50
Inflection point
45
Fast
40
35
30
0 2 4 6 8 10 12 14 16 18 20
FLOTATION ROUGHER RATE TIME (min)
Figure 9
70
65
60
55
50
45
40
35 MASS
30
25
20
15
10
5
0
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42
TIME (min)
Figure 10
Air Rate
LabTech Essa
The test is begun by opening the air. The aim throughout the test is to maintain the top of
the froth bed level with the overflow lip of the cell. The set point for air flow has been
marked at about 50% on the rotameter (indicated by the dark line and white arrow in
Figure 3). This setting is suitable for a 2.0/2.5 l cell. If a smaller or larger cell is used then
air rate will need to be adjusted as per Table 1. Air flow should not be increased greater
than the set point because an unstable situation can be created where air begins to blast
through the froth.
Air rate is adjusted by the knob marked “A” and flow is turned on/off by the switch
marked “B”. The ideal is to have previously set knob “A” and on starting the test, switch
“B” is then turned to “on” so that the test begins with the set point flow of air. This may
not always be possible depending on the floatability of the ore. If this is high, concentrate
production will be voluminous and a runaway froth will occur which will be
uncontrollable (ie concentrate overflow without having to use the paddles). For very
floatable ores it is best to begin the test by slowly opening the air and gradually building a
froth bed until it becomes level with the overflow lip. As soon as this is done timed
concentrate collection can begin.
What is required is that the froth is controlled at all times and is kept level with the
overflow lip of the cell. Thus at the start of the test, air valve “A” will need to be adjusted
and then gradually opened to the set point. After this, the level of pulp is topped up with
water to maintain froth level. All the time, froth is collected by paddle and the quantity of
froth removed is always controlled by the paddle design and the frequency of collection.
Denver
If the Denver machine is fitted with an air supply and rotameter then the above
description for the LabTech Essa machine applies. If this is not fitted, air is controlled by
the small “tap” shaped valve at the top of the stainless steel shaft. The same control is
required via this valve for very floatable and voluminous froths.
Froth Collection
Froth is collected every 15 seconds and consists of two sweeps of the paddles one
immediately after the other from the back of the cell to the front (“A” to “B”, Figure 11).
As the paddles moved forward they are angled around the mechanism and brought back
together again before moving to the front of the cell (Figure 12). At all times the paddles
must be operated in unison as a pair so that froth is removed in a uniform manner across
the full width of the cell.
An example of a 4 and 7 concentrate collection test is shown in Table 3. The timing and
sequence of collections is illustrated in Table 4. Note that the test, test conditions and
pulp level are set-up followed by air rate so that froth is level with the overflow lip. As
soon as this is done the test begins and froth collection can begin immediately (ie at 5
seconds), followed by a second collection at 15 seconds. Thereafter collection is at every
15 second interval.
B
Overflow Lip
Front of Cell
Figure 11
Back of Cell
B
Front of Cell
Figure 12
Table 3
Min Secs
Before timing, set-up test, air
rate and froth level
0 0 Begin test
0 5 Two sweep froth collection
0 15 Two sweep froth collection
Conc 1
0 30 Two sweep froth collection
0 45 Two sweep froth collection
1 00 Two sweep froth collection
1 15 Two sweep froth collection
1 30 Two sweep froth collection
1 45 Two sweep froth collection
Conc 2
2 00 Two sweep froth collection
2 15 Two sweep froth collection
2 30 Two sweep froth collection
2 45 Two sweep froth collection
3 00 Two sweep froth collection
3 15 Two sweep froth collection
Conc 3
3 30 Two sweep froth collection
etc to end of test
Table 4
When air rate is at the set point and froth level begins to fall below the level of the
overflow lip, pulp level should be topped with water.
Treatment of Products
All concentrate and tailings are weighed (dish+pulp), then filtered, dried, dry weight
recorded and sent for assay. This records dry mass and contained water for each product.
If the sample contains sulphur/sulphides drying must be done in a low temperature
oven (40C) to avoid burning off any sulphur/sulphides.
A few batch tests should be done to determine this particle size relationship for the
various ore types and ore mixes. This is essential data when checking the grind of the
laboratory mill. After a test, a representative sample of final tailings can be sent for assay
and a further representative sample subjected to screening (the balance of the tailings
material can be stored in case a re-assay is necessary). The particle size distribution and %
-75µm of float feed is then determined by combining the mass and grading of tailings and
concentrate streams.
REAGENT ADDITION
Reagents should be fresh and made-up on the day. Large additions of dilute reagent such
as depressant should be added by pipette or a large barrel syringe.
In cases where it is necessary to add a reagent neat – such as a frother – the required
addition can be just one drop. Small additions of reagent need to be added using a small
barrel medical syringe with a thin hypodermic needle. One drop has a volume of
0.02cm3. In a test using 1kg in a 2.5l cell this is equivalent to 20g/t which is a sizeable
dose in a plant.
o Washing material on the sides of the cell and mechanism back into the cell
and
o Checking and topping up pulp level.
TEST TECHNIQUE
The set-up and execution of a rate test needs to be more organised. Remember what I
said when we did some practice tests during my visit on 24th-27th October. Conducting a
rate test is an intensive activity where the operator is continuously on the go for the
duration of the test. In a period of 15 seconds the operator needs to do some or all of,
a. Two collections of froth from back to front of the cell,
b. Wash down the paddles into the concentrate dish,
c. If needs be wash down froth adhered to the insides of the cell and around the
shaft,
d. Check pulp level,
e. Top up pulp level with water,
f. Add reagents if stage adding say collector,
g. If needs be add extra frother,
h. Change concentrate dishes and
i. Check the time.
To do all this four times a minute for a 30 minute test requires much practice in order for
the test to run smoothly. The intensity of the test requires everything to be at hand with
the items logically arranged around the cell.
For example once the test has started there is no time to fill up a wash bottle so there must
be 4 full ones at the start of the test. Label the concentrate collection dishes beforehand, as
it is easy to forget which is which if they are moved around after the test. Practice adding
one drop of reagent from a syringe. This is not easy to do within the time schedule of a-i
above. It is easy to press the plunger too hard and a stream of reagent gets added at
which point the test is ruined. Thus, if the froth is thin and it is necessary to add more
frother, one (maximum two) drop(s) will suffice. Getting this wrong and adding a stream
of frother might add (say) 120g/t in which case the froth will be voluminous, mass pull
will be unrealistically high and the test must be abandoned.
A supply of syringes and small needles is important. Each should be labelled and used
for a certain reagent. Each should be thorough washed at the end of the day and kept
overnight in a beaker of water.
Wash bottles
Concentrate
Syringes of
collection dish 3
reagents
Concentrate
collection dish 2
Clipboard to
make notes of
froth condition
etc
Edge of bench
Figure 13