Service Manual: Wellwash 4 / 4Mk2
Service Manual: Wellwash 4 / 4Mk2
Service Manual: Wellwash 4 / 4Mk2
Wellwash 4 / 4Mk2
WW004/853
(17.10.2007)
to
ion.
ressure on Vessel 7
imensisning o f t e elcctm-icd lnsulatio
nners
1e Allen K
nor its agents can b
ive.
initiates the was
icates a vacuuln
uttons is confirmed
fill rows sequentially once only wi
fill rows 3 times fo'ollowe
.2.
Panel the sow number of t
then rows 4 to 4
e last wasl~ingcycle oft1
ocedure i s requir
la user protectio
a socket outlet
as often as necessa
nit witc
Air leaks
ubing leaks
e attention or if re
contact
Wellwash®
User Manual
Copyright
Trademarks (® and ™)
Disclaimer
Thermo Fisher Scientific reserves the right to change its products and services
at any time to incorporate technological developments. This manual is subject
to change without prior notice as part of a continuous product development.
Although this manual has been prepared with every precaution to ensure
accuracy, Thermo Fisher Scientific assumes no liability for any errors or
omissions, or for any damages resulting from the application or use of this
information. This manual supersedes all previous editions.
Thermo Fisher Scientific shall not be liable for any damages whatsoever arising
out of the use or inability to use this product.
Contact information
Table of Contents
1 Safety Symbols and Markings....................................................................................... 7
2 About the User Manual .................................................................................................. 8
3 Introduction to the Wellwash 4 Mk 2 ........................................................................... 9
4 Functional Description ................................................................................................. 11
5 Installation .................................................................................................................... 13
6 Operation ...................................................................................................................... 26
7 Maintenance.................................................................................................................. 36
8 Troubleshooting ........................................................................................................... 49
9 Technical Specifications............................................................................................... 55
10 Ordering Information ................................................................................................... 57
11 Warranty Certificate ..................................................................................................... 58
12 Index .............................................................................................................................. 59
13 Glossary and Abbreviations ........................................................................................ 63
Appendix A: Wellwash 4 Mk 2 Quick Reference Guide................................................................... 64
Appendix B: Certificate of Decontamination ................................................................................... 65
Appendix C: Wellwash 4 Mk 2 Feedback Form ............................................................................... 66
Appendix D: Addresses ...................................................................................................................... 67
Notes ............................................................................................................................. 68
Table of Contents
1 Safety Symbols and Markings....................................................................................... 7
2 About the User Manual .................................................................................................. 8
3 Introduction to the Wellwash 4 Mk 2 ........................................................................... 9
3.1 Intended use .................................................................................................................... 9
3.2 Principle of operation...................................................................................................... 9
3.3 Advantages of using Wellwash 4 Mk 2 .......................................................................... 9
4 Functional Description ................................................................................................. 11
4.1 Controls and indicators ................................................................................................. 12
5 Installation .................................................................................................................... 13
5.1 Installation check list ..................................................................................................... 13
5.2 What to do upon delivery ............................................................................................. 13
5.2.1 How to unpack ............................................................................................................ 13
5.2.2 Checking delivery for completeness ......................................................................... 14
5.2.3 Checking for damage during transport ..................................................................... 15
5.2.4 Environmental requirements ..................................................................................... 16
5.2.5 Things to avoid ........................................................................................................... 16
5.2.6 Technical prerequisites .............................................................................................. 16
5.3 Installation procedure ................................................................................................... 17
5.3.1 How to install the pump unit and washer unit connections..................................... 19
5.3.2 Wash head height ....................................................................................................... 23
5.3.3 Mains supply cable ..................................................................................................... 24
5.3.4 Changing wash heads ................................................................................................ 24
5.4 Operational check ......................................................................................................... 24
6 Operation ...................................................................................................................... 26
6.1 Operational checklist..................................................................................................... 26
6.2 Washing a microplate ................................................................................................... 27
6.3 Priming procedure ........................................................................................................ 29
6.4 Adding wash reagent .................................................................................................... 30
6.5 Program selection ......................................................................................................... 31
6.5.1 Two-way switch functions ......................................................................................... 32
6.5.2 Rotary switch functions .............................................................................................. 33
6.6 Resetting ........................................................................................................................ 35
6.7 Shutdown procedure .................................................................................................... 35
6.7.1 Daily shutdown procedure......................................................................................... 35
6.7.2 Long-period shutdown ............................................................................................... 35
6.8 Emergency situations.................................................................................................... 35
7 Maintenance.................................................................................................................. 36
7.1 Regular and preventive maintenance .......................................................................... 36
7.1.1 Service contracts ........................................................................................................ 36
7.1.2 Maintenance check list ............................................................................................... 36
7.1.3 General ........................................................................................................................ 36
7.1.4 Immediate ................................................................................................................... 37
7.1.5 Tubing ......................................................................................................................... 37
7.1.6 Reagent / waste bottles .............................................................................................. 37
7.1.7 Liquid trap bottle......................................................................................................... 38
List of Figures
List of Tables
Power ON
Power OFF
SN Serial number
Date of manufacture
WEEE symbol This product is required to comply with the European Union’s Waste
Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC.
This User Manual has been written for the actual user (e.g., laboratory
technician) and provides information on the Thermo Scientific Wellwash 4 Mk 2
microplate strip washer, including installation and operating instructions.
This User Manual also describes features and specifications of the Wellwash
4 Mk 2 hardware and on-board software.
Chapter 6 explains the operating procedures.
The user should be familiar with the contents of Chapter 7 on maintenance
procedures that may be performed by the user.
For warranty and ordering information, refer to Chapters 10 Ordering
Information and 11 Warranty .
4 Functional Description
6 5
7
3
8
1
9 4
10
2
11
12
13 14
The pump creates pressure to the wash bottle (Fig. 4.1, item 4) allowing the
wash liquid to be dispensed through the pinch valve (Fig. 4.1, item 14) into the
microplate wells and a vacuum to the waste bottle (Fig. 4.1, item 3) to aspirate
the used wash liquid into the waste bottle.
PRIME indicator
This yellow LED (light-emitting diode) located to the right of the PRIME button
flashes to indicate that the PRIME function is required and, when continuously
ON, indicates that the washer is primed and ready for use. The indicator will
turn off after a while, but the instrument remains ready for use.
POWER indicator
This green LED is located at the bottom left of the keypad and, when
illuminated, indicates that the POWER switch is ON.
5 Installation
Tick Item
Caution: DO NOT touch or loosen any screws or parts other than those specially
designated in the instructions. Doing so might cause misalignment and
will invalidate the instrument warranty.
Fig. 4.1 shows the pump and washer units in their entirety.
Lift out the washer and pump units, removing the shaped packing pieces from
around the washer, and place them on a flat level surface.
To lift the instrument, put your fingers under the bottom on both sides and lift it
with your back straight. The net weight of the instrument is 10 kg (22 lbs.).
Caution: When unpacking the instrument, it is recommended that two people lift
the instrument together, taking proper precautions to avoid injury.
Retain the original packaging and packing material for future transportation. The
packaging is designed to assure safe transport and minimize transit damage.
Use of alternative packaging materials may invalidate the warranty. Also, retain
all instrument-related documentation provided by the manufacturer for future
use.
Note: Before use remember to remove the transit screws at the bottom of the
instrument and save them in the adjoining screw holes (marked with the
text "Transit screw storage"). Remove and save the pinch valve holder
until needed for future transportation of the instrument.
2 6
1
3
5 7
1. Combiplate
2. 4 spare fuses (see Section 7.1.10 Fuse replacement)
3. 2 Allen keys (see Section 5.3 Installation procedure, step 1)
4. 2 spanners (see Section 5.3.2 Wash head height)
5. De-clogging tool (see Section 7.1.9 Unclogging tips)
6. 6 spare O rings for the bottle caps
7. 4 spare O rings (see Section 5.3 Installation procedure, step 5)
Fig. 5.1 Accessories included with the instrument.
1
2
3
4
5 6
Note: The use of the instrument in strong, direct sunlight or next to radiators
and other heat sources may markedly increase the temperature
prevailing at the instrument. This should be taken into account when
deciding where the instrument should be located and operated.
The Wellwash 4 Mk 2 does not produce operating noise at a level that would be
harmful. No sound level measurements are required after installation.
2. Ensure that the pump mains switch is in the OFF position. After checking
that the electrical supply conforms to that specified on the voltage label on
the rear panel of the unit, slide the frequency selector switch (in the pump
model Cat. no. 5160771) to the 50 Hz or 60 Hz position depending on the
mains supply voltage frequency (Fig. 5.4). (Other pump models do not
include a frequency selector.)
2
1
4. Lift the wash arm assembly (the stainless steel cover). Remove the pinch
valve holder (Fig. 5.5) before use. Store the wire holder for future use.
Note: The wash heads are packed separately. Install them before use.
5. Install first the four (4) small O-rings to the wash head holders, two (2) on
both sides (Fig. 5.6).
Note: The wash heads will only fit into the end slots when the raised flange on
the wash head is facing the front.
5.3.1 How to install the pump unit and washer unit connections
1. The wash bottle with the black connections is located on the right-hand side
of the pump unit and the waste bottle to the left. All the tubing and
connections are color coded to ensure correct connection (Table 5.2).
2. Connect the pressure feed from the pump on the right to the wash bottle
using the black-coded tube.
3. Connect the vacuum feed tube (red) from the pump to the waste bottle
(Fig. 5.9, item 2).
4. Connect the liquid level sensor cord from the liquid trap bottle to the waste
bottle. Make sure the black connector is firmly in place (Fig. 5.9, item 1).
Caution: The liquid level sensor does not function if the connector is not properly
inserted: the black connector has to be pushed down all the way. The
waste bottle may otherwise overfill and cause the liquid trap bottle to
start filling. If at that point the liquid trap bottle is not emptied, the
overflowing waste liquid can enter the pump unit and damage it.
1 2
5. Check that the tubes are properly connected to the pinch valve under the
wash arm assembly on the washer unit (Fig. 4.1, item 14). The thin tube
going through the valve should not be pinched by the edges of the valve.
6. Remove the three (3) black plastic caps from the hose connectors and the
pump control at the rear of the washer unit (Fig. 5.10).
7. Slide the two units together until the locating dowel and electrical
connectors are firmly engaged (Fig. 5.11).
4
1
2 6
3a
3b
8. Open the wash bottle (Fig. 4.1, item 4) by rotating the cap ring, while the
white plug remains stationary. Lift the cap ring and open the plug. Fill the
wash bottle up to the shoulder with wash reagent and reconnect ensuring
that the plug mates correctly with the bottle insert (Fig. 5.12) and that the
cap ring is tightly screwed down.
Cap ring
Plug
1. Wash head
2. Priming reservoir
3. Plate carrier
Fig. 5.13 Correct position of the plate carrier
4. Loosen the locknut (Fig. 5.14) by keeping the 6.0-spanner stationary while
turning the 5.5-spanner clockwise.
5. Adjust the height by rotating the domed nut, clockwise to lower the wash
head and anticlockwise to raise the wash head.
6. Set the tips of the wash head to approx. 1.5 mm to 2.0 mm above the
height of a microplate on the plate carrier, when the wash head is in the
raised position.
7. Lock the domed nut into position with the locknut. When locking the domed
nut into position, keep it still with the spanner 6.0 and only rotate the
locknut anticlockwise using the spanner 5.5.
Warning: Never operate your instrument from a power outlet that has no ground
connection. Never use a mains supply cable (cord) other than the
Thermo Scientific mains supply cable designed for your region.
1. Ensure that the mains switch at the rear of the pump unit is in the OFF
position.
2. Connect the mains supply cable (cord) to the mains input socket at the rear
of the pump unit. If you need to use any other type of mains supply cable
than supplied, use only cables certified by the local authorities.
3. Connect the instrument to a correctly installed line power outlet that has a
protective conductor that is grounded (earthed).
Note: The wash heads will only fit into the end slots when the raised flange on
the wash head is facing the front.
A1 A1
Fig. 5.16 Placing the microplate with 12-way and 8-way wash head
8. Select the row number of the last row to be washed on the keypad (e.g.,
when button 4 is pressed, rows 1 to 4 will be washed).
9. Press the START button to set the washing cycle into operation. The plate
carrier will be drawn into the washer, filling and aspirating each row
according to the sequence selected on the program card. At the end of a
sequence, rows will be left substantially dry and the plate carrier will stop
with the priming reservoir under the wash head.
10. The instrument is performing properly, if the wells are dry. If they are not,
check that both bottle caps are firmly closed
6 Operation
Tick Item
Perform
Indicates the actions you should take.
action
Turn the The POWER indicator on the keypad (Fig. 4.2) illuminates and a
instrument countdown sequence will begin on the row indicator LEDs.
ON
The PRIME indicator flashes until you prime the system (Section 6.3).
The priming function is complete when the PRIME indicator turns OFF.
PRIME
If no further key entry is made within fifteen seconds of priming, then
the pump will automatically switch OFF. The instrument will then enter
the idle state, waiting for instructions from the keypad and program
card.
Place the
microplate Check that the correct wash head (8-way or 12-way) is fitted.
onto the
plate carrier Check that the plate is placed properly onto the plate carrier (Fig. 5.16).
Turn the
instrument
OFF
Caution: Handle the microplate with care to avoid any contamination of the
instrument and the subsequent specimens. The contamination of the
specimen will lead to incorrect results.
If the contents of the microplate does spill, however, follow the
decontamination procedure presented in Section 7.3. Also, follow the
system supplier's instructions in regard to the reagent kit and specimen
handling in the event of contamination.
Warning: Never open the wash bottle, detach any of the tubing or press the black
top of the pinch valve without first releasing the pressure by slightly
opening the bottle cap. Let the hissing sound end prior to opening the
cap all the way.
Warning: If you are using a washing reagent that is either volatile or splashes
easily, keep yourself at a safe distance from the instrument while it is
operating to avoid exposure. Place the instrument into a fume hood.
Note: If the pump fails to start or if the bottle caps are loose, an audible
warning will be heard and the VAC FAIL indicator will illuminate. This
condition will clear once the pump has started. If the VAC FAIL indicator
does not clear after the pump has started, the RESET key must be
depressed. Also, check that the waste bottle is not full.
Note: The priming procedure must be repeated whenever the POWER switch
has been engaged or if there has been an interruption of the power
supply.
Caution: Make sure that sufficient reagent is present in the wash bottle to
complete the required protocol. See Fluid consumption in Section 9.1
General specifications.
Caution: If you use the programmable switch card, ensure that the 8 WAY –
12 WAY switch is in the correct position (for either the 8-way or the
12-way wash head).
Caution: If removable-well microplates are being used, make sure that there are
no missing wells in the rows to be washed and that all the wells are at
the same level.
Note If you press the RESET button on the keypad at any stage, this will stop
the immediate washing sequence and reset the system to its pre-
programmed state.
Caution: DO NOT smoke, eat or drink while using the Wellwash 4 Mk 2. Wash
your hands thoroughly after handling test plates. Observe normal
laboratory procedures for handling potentially dangerous plates. Use
proper protective clothing. Use disposable gloves. Ensure that the
working area is well ventilated.
Caution: Never spill fluids in or on the equipment. Prevent any liquid from
entering the instrument.
Caution: Immediately wipe away spilled liquids from outer surfaces to prevent
damage and wipe over with a mild detergent.
Caution: Keep the instrument free of dust and other foreign matter. Clean the
plate carrier and the plate carrier slideway and keep them dry to prevent
jamming.
Note: The priming procedure must be repeated whenever the POWER switch
has been operated or if there has been an interruption of the power
supply.
Also, when you are changing the used reagent and you wish to rinse and clean
the tubing, first release the pressure from the tubing and the wash bottle by
slightly opening the wash bottle cap.
Close the wash bottle tightly (Fig. 5.12).
The waste bottle needs to be emptied when the wash bottle is filled.
Note: Before inserting any program card, ensure that it is clean and dry.
Wiping the surfaces with a soft dry cloth will suffice. If any liquid is
present on the card, a malfunction of the program may occur.
Note 1: If a run is started without a program card or any row selection, the
instrument will signal an error by sounding the audible warning device
3 times.
Note 2: If a run is started with the rows selected but with no program card
inserted, the instrument aspirates the rows selected and returns to its
idle state with the wash head lowered into the priming reservoir.
Note 3: If a run is started without any rows selected but with a program card
fitted, the instrument will aspirate all the rows in accordance with the
card configuration, i.e., 8-way or 12-way wash head.
The preprogrammed wash cards will dispense and aspirate the plate row by
row for the number of times indicated on the card. There are four cards which
can be inserted either side up to perform the wash routine indicated on the side
facing up.
Fig. 6.2 Preprogrammed card: three washes with an 8-way wash head
Contained on the programmable switch card are four (4) rotary type and nine
(9) two-way type switches each marked with their function. The setting of these
switches controls the function of the washer, enabling a wide variety of
protocols to be performed.
Settings Number of
washes
F1 OFF F2 OFF 0
FINAL X1 F2 OFF 1
F1 OFF FINAL X2 2
FINAL X1 FINAL X2 3
Settings
PLATE WET No aspiration (wells filled)
PLATE DRY All rows aspirated after the final wash
STRIP WET No aspiration (wells filled)
STRIP DRY Each row aspirated after its final wash
Setting 0 1 2 3 4 5 6 7
Soak time (minutes) 0 0.5 1.0 1.5 2.0 2.5 3.0 3.5
Setting 8 9 A B C D E F
Soak time (minutes) 4.0 4.5 5.0 6.0 7.0 8.0 9.0 10.0
PAUSE
Set the pause between each wash cycle (Table 6.5). During the strip wash
mode, the pause is between each row. During the plate wash mode, the pause
is between each pass over the plate.
Note: The times currently available are subject to approval and may be altered
to suit individual customer requirements.
Setting 0 1 2 3 4 5 6 7
Strip wash mode
0 1 2 3 4 5 6 7
(seconds)
Plate wash mode
0 5 10 15 20 25 30 40
(seconds)
Setting 8 9 A B C D E F
Strip wash mode
8 9 10 15 20 30 45 60
(seconds)
Plate wash mode
50 60 75 85 105 120 180 240
(seconds)
WASHES
Set the number of washes to be performed before a soak period (Table 6.6).
Setting 0 1 2 3 4 5 6 7
Number of washes 0*) 1 2 3 4 5 6 7
Setting 8 9 A B C D E F
Number of washes 8 9 10 11 12 13 14 15
*) No wash, only aspirate
VOLUME
Set the volume dispensed during washing (Table 6.7).
Setting 0 1 2 3 4 5 6 7
Washing volume (µl) 0 50 100 150 200 250 300 350
Setting 8 9 A B C D E F
Washing volume (µl) 400 450 500 550 600 650 700 750
Note: The dispense volume is less than the wash volume to avoid overfilling
the wells.
6.6 Resetting
If you press the RESET button on the keypad at any stage, this will stop the
immediate washing sequence and reset the system to its pre-programmed
state.
Warning: Remove any microplate still on the plate carrier. Dispose of all
microplates and strips as biohazardous waste.
The wash reagents may contain buffer salts, which can crystallize when the
washer is not in use. This may cause clogging of the wash/aspirate tips. To
avoid this, the following procedures should be followed.
7 Maintenance
Item
As required
Monthly
Weekly
Yearly
Daily
Wipe the case of the instrument.
•
Cover the instrument with the dust cover after use.
•
Clean the instrument and the keyboard with a cloth
dampened with mild detergent, followed by deionized
•
distilled aqua. See Sections 7.1.3 and 7.1.4
Check the condition of the tubing and the wash and
waste bottles. See Sections 7.1.5 and 7.1.6
•
Check the pinch valve. See Section 7.1.8.
•
Unclog the wash head tips. See Section 7.1.9.
•
Change the fuses when blown. See Section 7.1.10.
•
Decontaminate the instrument if any biohazardous
material has been spilled on it or when shipping it to
•
service. See Section 7.3.
Perform the operational check regularly. See
Section 5.4.
•
Arrange for the yearly servicing of the instrument
(Section 7.1.1 above).
•
7.1.3 General
Routine and service procedures must be performed by the user to prevent
unnecessary wear or hazards and are described below at the frequency with
which they should be applied.
For reliable daily operation, keep the instrument free of dust and spills from
liquids. It is also advisable to cover the instrument with the dust cover supplied
when not in use. In the event of any damage, contact your local Thermo Fisher
Scientific representative for service.
Abrasive cleaning agents are not recommended, because they are likely to
damage the paint finish.
It is recommended that you clean the case of the instrument periodically to
maintain its good appearance. A soft cloth dampened in a warm, mild detergent
solution will be sufficient.
Caution: Painted surfaces can be cleaned with most laboratory detergents. Dilute
the cleaning agent as recommended by the manufacturer. DO NOT
expose painted surfaces to concentrated acids or alcohols for prolonged
periods of time as damage may occur.
Warning: If, for any reason, you need to move the instrument, release first the
pressure from the tubing by slightly opening the bottle cap. Let the
hissing sound end and then detach the yellow and blue coded tubes
from the instrument.
7.1.4 Immediate
Although the Wellwash 4 Mk 2 is constructed from high-quality materials, you
must immediately wipe away spilled saline solutions, solvents, acids or alkaline
solutions from outer surfaces to prevent damage and wipe with deionized
distilled aqua.
Spillages of liquid on the carrier slideway should be wiped off promptly, as this
may cause the plate carrier to drag during operation.
7.1.5 Tubing
Inspect the condition of the tubing regularly. When you notice that a tube is
becoming harder and it is no longer clear, it should be replaced. Only use
tubing supplied by Thermo Fisher Scientific Oy. The instrument is supplied with
spare tubes (Fig. 5.2). See Chapter 10 Ordering Information.
Warning: Never detach a tube without first releasing the pressure by slightly
opening the bottle cap. Let the hissing sound end prior to detaching any
of the tubes.
3. Release the cover by slightly lifting the tube and slide the cover then
upwards along the tube (Fig. 7.2)
4. Pull the thick vacuum tube (with the 90° angle) out of the liquid trap bottle.
Fig. 7.3 Pulling the thick tube out of the liquid trap bottle
Fig. 7.4 Pulling the thin tube out of the liquid trap bottle
6. Let the red-coded tube remain connected to the bottle cap and take the
bottle out, unscrew the liquid trap bottle cap and empty the bottle.
7. Close the bottle. Make sure the end of the red-coded tube is about 2 cm
deeper inside the bottle than the thin tube will be. Put the bottle back to its
place, but don't push it yet all the way down.
8. Insert the thin tube back through the liquid trap bottle cap so that the mark
on the tube (Fig. 7.4, item 1) is at the level of the cap.
9. Push the bottle all the way down. Push the thick vacuum tube as far as it
goes. Its end should now be approximately at the same level as the red-
coded tube's end.
10. Slide the cover back down and screw the black screw back on. You can pull
the red-coded tube slightly out to make sure the cover is tightly on.
Warning: Never press the black top of the pinch valve without first releasing the
pressure by slightly opening the wash bottle cap. Let the hissing sound
end prior to pushing the valve top.
Warning: Never disconnect the pressure tubing without first releasing the
pressure by slightly opening the wash bottle cap. Let the hissing sound
end prior to disconnecting the tubing.
3. Locate the fuse holders at the rear of the washer and pump units.
4. On the washer unit, use a screwdriver to remove the fuse holder cap. Turn
the screwdriver counterclockwise about 45° to release it and then pull it out
(Fig. 7.8).
Fuse holder
6. Ensure that the replacement has the same rating as appears on the rating
label on the rear panel of the washer unit. Only use fuses specified by
Thermo Fisher Scientific Oy.
7. Insert the new fuse to the fuse holder and push the holder into its place.
Turn the holder clockwise about 45° with the screwdriver.
8. On the pump unit, use the screwdriver to gently pry the fuse folder open.
Fuse holder
9. There are two fuses inside the holder. The outer one is a spare fuse and the
inner one is the one in use.
Used fuse
Spare fuse
Note: Ensure that only fuses with the required rated current and of the
specified type are used for replacement. The use of makeshift fuses and
the short-circuiting of fuse holders are prohibited.
Example of decontaminants
• Ethanol 70%
• Virkon solution 1 – 3%
• Glutaraldehyde solution 4%
• Chloramine T
• Microcide SQ™ 1:64
Caution: Do not use sodium hypochlorite as a decontaminant. The wash head will
be damaged by it.
Warning: Always use disposable gloves and protective clothing and operate in a
well-ventilated area.
2. Prime the unit with distilled water from the wash bottle.
3. Lift the wash head and fill the prime tray with glutaraldehyde solution.
4. Aspirate the prime tray using the wash head, and then reprime the prime
tray with distilled water.
5. Remove the plate carrier and the wash head and place into a glutaraldehyde
solution for 24 hours.
6. Empty the fluid containers.
7. Switch OFF the power and disconnect the mains supply cable.
8. Decontaminate the outside of the unit using a wad of cotton wool soaked in
70% alcohol.
9. Place the entire instrument into a large plastic bag. Ensure that the reagent
containers are disconnected and open.
10. Place a wad of cotton wool soaked in 10% formaldehyde solution into the
bag. Ensure that the wad does not make contact with the instrument.
11. Close the bag firmly and leave the instrument in the bag for at least
24 hours.
12. Remove the instrument from the bag.
13. Clean the instrument using a mild detergent.
14. Remove any stains using 70% alcohol.
15. Take the plate carrier and the wash head out of the glutaraldehyde solution
and rinse with distilled water. Reposition them on the unit.
16. Flush the fluid path of the instrument with deionized distilled aqua using the
priming procedure.
17. The wash heads can be sterilized in 70% ethanol.
18. Also, wash and decontaminate the wash, rinse and waste bottles.
See Section 9.1 for details on storage and transportation temperatures. See also
Section 8.2 Service request protocol.
7.4.1 Packing
The Wellwash 4 Mk 2 comprises two units that are packed in a single box. For
packing the washer and pump units, follow the procedure below:
1. Place the box on the floor and open the top. Put the bottom part of the
package shock absorber at the bottom of the box.
2. Wrap the washer and pump units into plastic. Drop them into their own
places onto the bottom part of the package shock absorber.
3. Put the top part of the package shock absorber into the box. Ensure correct
positioning and the correct location for the manifold support piece.
4. Put the bottles into Minigrip plastic bags and drop them into their own
locations in two corners of the box.
5. Put the two small carton boxes onto the top part of the package shock
absorber. DO NOT place the small boxes on top of each other, but place
them instead into their own specific locations.
6. Put the rest of the items (user manual, labels, notice, packing list, etc.) into a
Minigrip plastic bag and place it into its own position on the top part of the
package shock absorber.
7. Close and seal the box.
Warning: Decontaminate the instrument prior to disposal. See Section 7.3 and
Appendix B on decontamination. Also, wash and decontaminate the
bottles prior to disposal.
Thermo Fisher Scientific has contracted with one or more recycling/disposal companies
in each EU Member State European Country, and this product should be disposed of or
recycled through them. Further information on Thermo Fisher Scientific’s compliance
with these Directives, the recyclers in your country, and information on Thermo Fisher
Scientific products which may assist the detection of substances subject to the RoHS
Directive are available at www.thermo.com/WEEERoHS.
Regarding the original packaging and packing materials, use the recycling
operators known to you.
For further information, contact your local Thermo Fisher Scientific
representative.
8 Troubleshooting
Warning: DO NOT use the instrument if it appears that it does not function
properly.
If you were able to correct the error without having to turn off the instrument,
you may continue instrument operation by pressing start.
8.3.2 Electrical
Ensure that the mains supply cable supplied with the unit is always used. If a
correct type of mains cable is not provided, use only cables certified by the
local authorities.
The mains plug should only be inserted into a socket outlet provided with a
protective ground contact. The protective action must not be negated by use of
an extension cable without a protective ground wire.
Warning: DO NOT replace fuses without first disconnecting the mains supply
cable. Only use fuses specified by Thermo Fisher Scientific Oy.
Ensure that only fuses with the required rated current and of the specified type
are used for replacement. The use of makeshift fuses and the short-circuiting of
fuse holders are prohibited.
Warning: When the instrument is connected to the mains supply, the opening of
covers or the removal of components is likely to expose live parts.
Before the instrument is opened for any adjustment, replacement,
maintenance or repair, disconnect it from all voltage sources by
disconnecting the mains supply cable.
Environmental
Infectious clinical samples and corrosive fluids are commonly used with this
equipment. The "hands-off" nature of the system allows the user to wash the
reaction wells without getting into direct contact with these fluids. However,
wells that have been in contact with potentially hazardous fluids must be
handled before and after the wash process, and this should be done with
utmost care. Hand and eye protections should always be worn as well as
corrosive resistant laboratory coats.
If you are using a washing reagent that is either volatile or easily splashing, or
when you are washing microplates with samples of human or animal origin,
keep yourself at a safe distance from the instrument while it is operating to
avoid exposure. Place the instrument into a fume hood.
When spills get into contact with the carrier slideway, it should be
decontaminated with a hard surface disinfectant (see Section 7.3).
Observe Good Laboratory Practices (GLP) when handling potentially hazardous
samples.
8.5.1 Installation
The instrument should be unpacked and installed in accordance with Chapter 5.
Particular attention should be paid to the correct fitting of the tubing to obviate
leaks (Fig. 5.11). Always ensure that the electrical supply in the laboratory
conforms to that specified on the voltage label on the rear panel of the pump
unit.
8.5.2 In operation
Ensure that the waste bottle is emptied regularly and at least every time that a
wash bottle is loaded. The liquid trap bottle that prevents liquid from entering
the pump should be kept dry at all times.
8.6 Cleanliness
Clean the instrument regularly as stated below.
1. It is recommended that the case of the instrument is cleaned periodically to
maintain its good appearance. It is particularly essential that the carrier
slideway be clean and dry to prevent jamming. A soft cloth dampened in
warm, mild detergent solution will suffice.
2. If any surfaces have been contaminated with biohazardous material, a mild
sterilizing solution should be used. See Section 7.3 Decontamination
procedure.
9 Technical Specifications
General specifications
Overall dimensions Washer unit Pump unit
41 cm x 22 cm x 12 cm 30 cm x 18 cm x 16 cm
[16.1" x 8.7" x 4.7"] [11.8" x 7.1" x 6.3"]
Weight Washer unit Pump unit Total
4.5 kg [10 lbs.] 5.5 kg [12 lbs.] 10 kg [22 lbs.]
Mains power supply 220 ― 240 Vac, 50 Hz
110 ― 120 Vac, 60 Hz
100 Vac, 50/60 Hz
Power consumption 300 VA
Heat dissipation 1024 BTU
Operating cycle Continuous
Fuse requirements 220/240 Vac 50 Hz 100/110/120 Vac
Washer unit: 2 x 0.5 A/250 V UL 198G Time Delay 2 x 0.5 A/250 V UL 198G Time Delay
Pump unit: 2 x 1.0 A/250 V UL 198G Time Delay 2 x 2.0 A/250 V UL 198G Time Delay
Operating conditions Indoor use
Temperature Operating range +5°C to +40°C
Full specification range +15°C to +25°C
Storage temperature range –10°C to +70°C, packed in transport packaging
Transport temperature range –40°C to +70°C, packed in transport packaging
Relative humidity 90% max. non-condensing
Pressure Max. pressure on vessel 7 psi (0.5 bar) (self-limiting)
Plate type 96-well plate
Containers Material: HDPE (high density polyethylene)
Wash bottle capacity 2 liters
Waste bottle capacity 2 liters
Additional wash bottle capacity 2 liters
Fluid consumption Wash fluid consumption approx. 150 ml/complete microplate
Priming consumption approx. 15 – 20 ml
Residual aspiration volume < 5 µl per well, using 0.02% Tween solution
Dispensing precision 5%, using 0.02% Tween solution
Wash heads 8-way and 12-way
Material: POM (polyoxymethylene, polyacetals)
Wash program cards 1 – 4 washes
Programmable card Volume 0 – 750 µl
Washes 1 – 15
Soak time 0 – 10 minutes
Pause 0 – 4 minutes
Wash time 120 s, 3 x 350 µl with 12-way wash head
165 s, 3 x 350 µl with 8-way wash head
Safety performance:
EN 61010-1:1993 + A2:1995/IEC 61010-1:1990 + A1:1992 + A2:1995, including CENELEC Common
Modifications, US and CA National differences
EN 61010-1:2001 (Ed. 2)
EMC performance:
EN 50081-1:1992 Generic emission standard. Residential, commercial and light industry.
EN 50082-1:1997 Generic immunity standard. Residential, commercial and light industry.
EN 61326-1:1997 + A1:1998 Product family standard.
10 Ordering Information
Contact your local Thermo Fisher Scientific representative for ordering and
service information.
Code Instrument
5160770 Wellwash 4 Mk 2 220 – 240 V AC/50 Hz
5160772 Wellwash 4 Mk 2 110 – 120 V AC/60 Hz
Code Product
N03283 2 l wash bottle
1514313 2 l waste bottle
WW004/1 12-way wash head
WW004/11 8-way wash head
WW004/48 Wash height adapter
Code Product
006/612; Spare parts bag for 220 V; (see Fig. 5.1)
006/613; Spare parts bag for 110 V;
006/613J Spare parts bag for 100 V
004/023D Spare tubes bag (see Fig. 5.2)
1210840; Spare fuses 0.5 A;
1210890; Spare fuses 1 A (for 220 V);
1210920 Spare fuses 2 A (for 100 V and 110 V)
11 Warranty Certificate
12 Index
A
Accessories........................................................................................................................................... 56
inspecting.......................................................................................................................................... 14
unpacking.......................................................................................................................................... 12
Accessory pack..................................................................................................................................... 13
Application............................................................................................................................................ 65
Audible warning ................................................................................................................................... 30
B
Bottle caps ............................................................................................................................................ 10
Bottles ................................................................................................................................................... 10
emptying ........................................................................................................................................... 37
replacement interval ......................................................................................................................... 36
Buttons.................................................................................................................................................. 11
C
Certificate of Decontamination ............................................................................................................ 64
Check list............................................................................................................................................... 12
maintence.......................................................................................................................................... 35
Clogging................................................................................................................................................ 42
prevention ......................................................................................................................................... 34
Color coding of the tubes .................................................................................................................... 18
Complaint order.................................................................................................................................... 51
Consumables
warranty limitations .......................................................................................................................... 57
Contamination
specimen ........................................................................................................................................... 27
Control panel ........................................................................................................................................ 11
Countdown sequence of row indicators ............................................................................................. 23
D
Decontamination .................................................................................................................................. 45
Certificate of ...................................................................................................................................... 64
procedure.......................................................................................................................................... 45
Dimensions ........................................................................................................................................... 54
Directives .............................................................................................................................................. 55
Disinfectants ......................................................................................................................................... 45
Disinfection procedure......................................................................................................................... 34
Dispensed volume................................................................................................................................ 33
Dispensing precision............................................................................................................................ 54
Disposal
instrument ......................................................................................................................................... 47
materials............................................................................................................................................ 45
E
Environmental conditions .................................................................................................................... 54
F
Feedback Form ..................................................................................................................................... 65
Frequency selector......................................................................................................................... 10, 16
Fuses ..................................................................................................................................................... 54
replacing............................................................................................................................................ 43
H
Hose connectors................................................................................................................................... 20
How to disinfect the instrument .......................................................................................................... 34
I
Idle state.......................................................................................................................................... 24, 30
Indicators .............................................................................................................................................. 11
Installation............................................................................................................................................. 12
check list............................................................................................................................................ 12
site ..................................................................................................................................................... 15
Installing
liquid level sensor............................................................................................................................. 19
pump and washer units together..................................................................................................... 20
Instrument
decontamination ............................................................................................................................... 45
functional description ....................................................................................................................... 10
inspecting.......................................................................................................................................... 14
markings............................................................................................................................................ 55
safety symbols and markings ............................................................................................................ 7
unpacking.......................................................................................................................................... 12
K
Keypad ............................................................................................................................................ 10, 11
cleaning ............................................................................................................................................. 36
L
Liquid level sensor ............................................................................................................................... 10
installation ......................................................................................................................................... 19
Liquid trap............................................................................................................................................. 10
Liquid trap bottle ............................................................................................................................ 10, 37
overfilling precautions ...................................................................................................................... 19
M
Mains supply cable............................................................................................................................... 23
Mains switch......................................................................................................................................... 23
Maintenance ......................................................................................................................................... 35
check list............................................................................................................................................ 35
Manifold pipes ...................................................................................................................................... 20
Microplate
handling............................................................................................................................................. 27
size differences ................................................................................................................................. 22
N
Number keys......................................................................................................................................... 11
Number of washes ............................................................................................................................... 33
O
Operating conditions............................................................................................................................ 15
Operational check................................................................................................................................. 23
Ordering information............................................................................................................................ 56
Overfilling of wells................................................................................................................................ 49
P
Packing.................................................................................................................................................. 47
for service.......................................................................................................................................... 46
list ...................................................................................................................................................... 13
Pause between washes........................................................................................................................ 33
Tubes
color codes........................................................................................................................................ 18
unpinching ........................................................................................................................................ 41
Two-way switches................................................................................................................................ 31
U
Unpacking ............................................................................................................................................. 12
Using after prolonged shutdown......................................................................................................... 34
V
Vacuum fail indicator............................................................................................................................ 11
Vacuum Fail indicator............................................................................................................... 27, 28, 49
Voltage .................................................................................................................................................. 15
Volume.................................................................................................................................................. 33
W
Warning signal...................................................................................................................................... 30
Warranty ................................................................................................................................... 12, 16, 57
Wash arm assembly ............................................................................................................................. 10
Wash bottle........................................................................................................................................... 10
installation ......................................................................................................................................... 21
Wash head
height................................................................................................................................................. 22
holder ................................................................................................................................................ 17
installation ................................................................................................................................... 18, 23
unclogging tips ................................................................................................................................. 42
Wash mode
plate ................................................................................................................................................... 31
Wash reagent
bottle ................................................................................................................................................. 37
no flow......................................................................................................................................... 41, 49
unequal dispensing or aspirating..................................................................................................... 49
Washer unit........................................................................................................................................... 10
Waste bottle.................................................................................................................................... 10, 37
overfilling precautions ...................................................................................................................... 19
Weight....................................................................................................................................... 13, 15, 54
Working area ........................................................................................................................................ 15
EN European Norm.
EU European Union.
Microplate A rigid or framed polystyrene plate with microwells in different well formats (e.g., 6,
12, 24, 48, 96, 384, 864, etc., wells) for ease of use in performing multiple tests
through techniques such as EIA or ELISA.
Priming Completely filling the pump tubing and syringe with bubble-free fluid to allow
sustained, reproducible pumping action. The air in an unprimed line acts as a
spring, adversely affecting accuracy and precision.
RS-232C EIA approved standard used in serial data transmission, covering voltage and
control signals. The data is transferred serially (one digital bit at a time) via one path,
but some control signals can be transferred simultaneously via parallel paths.
13.2 Literature
Newman, D.J. and C.P. Price (1997) Separation Techniques. In: Principles and
Practice of Immunoassay 2nd ed. (eds. Price, C.P. and Newman, D.J.) pp. 154 –
172 (Stockton Press, New York)
Coupe, N.B. and A.M. Cain. Microplate washing – the most important step in the
ELISA process. Labsystems-Denley application note.
Beumer T., Stoffelen, E., Smits, J. and Carpay, W. (1992). Microplate washing:
process description and improvements. J. Immunol. Methods 154, 77 – 87.
Parry, J.V., Mortimer, P.P., Friederich, P. and Connell, J.A. (1997). Faulty
washers and soiled micropipettors may generate false positive serological
results. Clinical and Diagnostic Virology 7, 173 – 181.
Turn the The POWER indicator on the keypad (Fig. 4.2) illuminates and a
instrument countdown sequence will begin on the row indicator LEDs.
ON
The PRIME indicator flashes until you prime the system.
The priming function is complete when the PRIME indicator turns OFF.
PRIME
If no further key entry is made within fifteen seconds of priming, then
the pump will automatically switch OFF. The instrument will then enter
the idle state, waiting for instructions from the keypad and program
card.
Place the
microplate Check that the correct wash head (8-way or 12-way) is fitted.
onto the
plate carrier Check that the plate is placed properly onto the plate carrier (Fig. 5.16).
Remove the
microplate
from the
plate carrier
Turn the
instrument
OFF
A)
I confirm that the returned items have not been contaminated by body fluids, toxic,
carcinogenic or radioactive materials or any other hazardous materials.
B)
I confirm that the returned items have been decontaminated and can be handled
without exposing the personnel to health hazards.
Specific information
about contaminants:
Decontamination
procedure1:
*) The signature of a Radiation Safety Officer is also required when the unit has been
used with radioactive materials.
1
Please include decontaminating solution used.
Company/Organization Distributor
Department
Address Address
Tel. Tel.
Date of purchase
Would you like to receive information about other Thermo Fisher Scientific products?
Appendix D: Addresses
For the latest information on products and services, visit our worldwide web
sites on the Internet at:
http://www.thermo.com
Declogging Tool
3mm Allen Key
2.5mm Allen Key
2.0mm Allen Key
Spanner
Feeler Gauges
WW004/010 Programmable Card
Pressure Gauge Range 0-15 p.s.i.
Vacuum Flow Gauge Range 0-8 bar
SECTION 4 DIAGRAMS
4.1 Diagrams.
This section contains the following diagrams:
004/050 Washer Unit General Assembly
004/050 Items List (1 Sheet)
004/051 Washer Chassis Assembly
004/051 Items Lists ( 3 Sheets)
004/052 Washer Cover Assembly
004/052 Items List (1 Sheet)
004/053 Pump Unit General Assembly
004/053 Items List (2 Sheets)
004/054 Pump Unit Chassis Assembly
004/054 Items Lists ( 2 Sheets)
004/066 Control Logic PCB Assembly
004/066 Items Lists (4 Sheets)
004/701 Control Logic Circuit Diagram
004/750 Washer Unit Wiring Diagram
004/751 Pump Unit Wiring Diagram
004/752 Plumbing Diagram
004/752 Items List ( 1 Sheet)
SECTIOX 5 TECHNICAL BULLETINS
Spare parts
Wellwash 4 / 4Mk2
Spare parts
Contents
SUBJECT PAGE
NOTE! All Spare parts that are considered as Recommended Spare parts to service offices have in
theyre stock are marked with *.
Spare parts
Main parts
7
6
1
3 2
Spare parts
CONTROL ASSEMBLY
6 2 5 3
11
4
1
10
8
Spare parts
Control assembly
8
SILENCER
FILTER
6
9
SILENCER
2 5
Spare parts
Spare parts
Miscellaneous
ITEM CAT.
Spare parts
Accessories
ITEM CAT.
I I I I
t
10 1, 00089100 1 1 EARTH SYMBOL
I
1 1
I
1
I
I 1
0041646 1
'0' RING -
SILICONE RUBBER 1 1 1
1
18 ( 000804076 WASHER M3 W A V E 1 8 1
BILLINCSHURST.
SUSSEX.ENCLAND. .
RH14 9EY
ECKED
D .NAPPER
TITLE
DENLEY
D. NAPPER PWSR ACCV fUCCUEDI
10 / 934/330 1 BOTTLE CAP MOD 12 1
2 1 '
1 1
.-.-'.
lfw- 1
,2
;
i
I
BILLINGSHURST,
SUSSEX.ENGLAND.
YII
TITLE SHEET DEPT I/LIST FOR DRC No.
DRAUN GENERAL ASSEHBLY - PUHP UNIT No. t A4
220-240V 50HZ OF 2 22 004/053A
USED ON W W 004 -"-%
ITEM DRAWING NO TITLE OTY R€rlAmS
- -
t-
UITHOUT THE WRITTEN CONSENT OF
CHECKED
J.FOX
THE COMPANY. @ 1993 ISS DATE ECN No. I CHK 1 ISS 1 DATE 1 ECN No. 1 Ct
TITLE SHEET DEPT IlLlST FOR DRG No.
DRAWN GENERAL ASSEHBLY - PUHP UNIT No. 2 A,
D. NAPPER OF 2 22 OO4/05%
ITEn DAAYIK No T I N OTI aWYI
I
1
I , ,
I 004/308 ( BOTTLE CAP MOD 1 1
1 000543901 1
13 CABLE T I E BASE 1 1 I
114 1 000839003 / BOTTLE 12 1
I 15 000839009 BOTTLE (150.1) 1
16 000921006 SWITCH COVER 1
'17 000800686 SCREW H4 x 10 BUTTON HD 1
18 000802085 SCREW M4 x 8 PAN HD POZI 6
19 000800885 SCREW M4 x 8 CAP HD 2
BILLINGSWURST.
[[I SUSSEXeENGLAND.
RH14 9EY
BARBED CONNECTOR 1
FEET
LABEL 7 U V TEST MARK"
SCREW M3 x 8 RSD CSK HD POZl
SCREW No6 x 318" PN HD SITAP PO2
WASHER M3 WAVE 7
WASHER M4 WAVE 4
DENLEY Billingshurst, Sussex RH14 9EY England
51 000803008 NUT M 3 FULL HEX 4
52
53
54
55 000803010 NUT M4 FULL HEX 6
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
MAINS SOCKET I
FRICTION BUSH
M4 EARTH TAG SIPROOF
EARTHSYMBOL
SLEEVING HZ0 x 20
WlRE 20 AWG 10130 GREY mrn
WlRE 18 AWG 16/30 GREYNLLW mrn
WIRE 20 AWG 10130 BLUE mm
WlRE 20 AWG 10130 BROWN mm
WIRE 20 AWG 10130 RED mrn
BARBED CONNECTOR 1
FEET 4
WASHER M3 WAVE
1 1 9 APR 1996
WASHER M4 WAVE
I
D.Napper
P.J.F.x / A.Partridge S.Downes
I CHASSIS ASSEMBLY -
PUMP UNIT 110- IIOV 6 O U
I
ffl DENLEY Billingshunt, Sussex RH14 9EY England
51 000803008 NUT M3 FULL HEX 4
52
53
54
55 000803010 N U T M4 FULL HEX 6
SSiS ASSEMBLY -
IflDENLEY Billingshursc Sussen RH14 9EY England
PCB
LEO SUPPORT
INTERGRATED CIRCUITS
MICRO PRO PO 8749 VERSION 4
IC 74 LS 165
IC 74 LS 374
IC 74 LS 04
7 WAY DARL DRIVER ULN 2003N
OPTO TRIAC DRIVER MOC 3041
.- .
IC 7404
HALL EFFECT 1C UGN 3120U
IC 74 LS 125
CAPACITORS
2.2 uF 35V SOLID TANTALUM C3
10 uF 35V SOLID TANTALUM C16
100 uF 16V ELECTROLITIC C4
2200 uF 63V ELECTROLITIC C61
0.47 uF lOOV MIN LAYER C6
0.22 uF IOOV MIN LAYER C5
22 pF 100V PLATE CERAMIC C1,Z
0.1 uF 25V DISC CERAMIC C7 TO C15. C20. C22
0.01 uF 400V POLYESTER C17TO C19
820 gF lOOv SINGLE LAYER CERAMIC C23 TO C27
RESISTORS
56R 0.25W R1.2.3
10OR O.25W R4.22.23.58
l2OR O.25W R5 TO R10. R46 TO R55.
R71 TO R74
R56
R62
R11.12.13
R15
R16 TO R21, R57
R24,61.75
R25.26. R76 TO R79
R32.38
R33. 34
R27
RESISTOR NEWORK 5X1K RP4
CERAMIC W M I RESISTOR 4.7R 4W R45
VOLTAGE REGUMTOR uA 7805UC REG1
VOLTAGE REG L78SlZCV 12V 2A REG2
HIGH SPEED SUPPRESSOR 6.8V 0 1 $2
QUARTZ CRYSTAL HC18U 6MHZ XT1
TRIAC MAC 97-6 SCR1.2.3
RESISTOR NETWORK 8x47K RP1.2.3.5.6
RECTIFIER KBL 02 RECl
-
SLOlTEO OPTO LTH 301 (LITE ON) OSW1,2
1 004/801 PCB 1
2 0041325 LED SUPPORT 1
3
4
5 INTERGRATED CIRCUITS
6 000703065 MICRO PRO PD 8749 VERSION 4 1 IC 1
7 000703017 IC 74 LS 165 4 lC2 TO IC5
8 000703018 IC 74 LS 374 2 IC7. IC20
9 000703019 IC 74 LS 04 1 IC9
10 000703021 7 WAY DARL DRIVER ULM 2003N 1 El0
11 000703022 OPT0 TRlAC DRIVER MOC 3041 3 IC11.12.13
12 000703023 IC 7404 1 ICE
13 000703030 HALL EFFECT IC UGN 3120U 1 IC14
14 000703066 IC 74 LS 125 1 IC22
15 000703067 IC74LS 139 1 IC21
16 CAPACITORS
17 000101787 2.2 UF 35V SOLID TANTALUM 1 C3
18 000102702 10 uF 35V SOLID TANTALUM 1 C16
19 000104477 100 UF 16V ELECTROLITIC 1 C4
20 000103749 2200 uF 63V ELECTROLITIC 1 C61
21 000114424 0.47 UF 1OOV MIN LAYER 1 C6
22 000114418 0.22 uF lOOV MIN LAYER 1 C5
23 000126124 22 pF 100V PLATE CERAMIC 2 C1.2
24 000702009 0.1 uF 25V DISC CERAMIC 11 C7 TO C15. C20, C22
25 000702005 0.01 uF 400V POLYESTER 3 C17 TO C19
26 000702172 820 uF lOOv SINGLE LAYER CERAMIC 5 C23 TO C27
27
28 000131104 3 R1.2.3
29 000131132 4 : R4.22.23.58
30 000131148 20 R5 TO R10, R46 TO R55.
R71 TO R74
31 000131160 1 R56
32 000131176 1 R62
33 000131211 3 R11.12.13
34 000131227 1 R15
35 000131255 7 R16 TO R21. R57
36 000131334 3 R24.61.75
37 000131378 6 R25,26, R76 TO R79
38 000131457 2 R32.38
39 000131520 2 R33.34
40 000842769 1 R27
41 000842776 RESISTOR NETWORK 5X1K 1 RP4
42 000155352 CERAMIC W M I RESISTOR 4.7R 4W 1 R45
43 000305888 VOLTAGE REGULATOR uA 7805UC 1 REG1
44 000703015 VOLTAGE REG L78S12CV 12V 2A 1 REG2
45 000283255 HIGH SPEED SUPPRESSOR 6.8V 2 D l2
46 000302198 QUARTZ CRYSTAL HCl8U 6MHZ I XTI
47 000785008 TRlAC MAC 97-6 3 SCRf.2.3
48 000842011 RESISTOR NETWORK 8x47K 5 RP1.2.3.5.6
49 0002621 13 RECTlFlER KBL 02 1 RECl
50 000711010 SLOTTED OPT0 LTH 301 (LITE -ON) 2 OSW1,2
ffl f j H L E Y Billingshurst, Sussen RH14 9EY England