UVGI DISINFECTION OF ROOM AIR. An Evidence Based Guideline For Desing, Implementation and Maintenance

Download as pdf or txt
Download as pdf or txt
You are on page 1of 36

UVGI DISINFECTION OF ROOM AIR:

AN EVIDENCE BASED GUIDELINE FOR


DESIGN, IMPLEMENTATION AND
MAINTENANCE
DISCUSSION DRAFT

Rev 6.1

AUTHORS:
Tanusha Singh (NIOH) (Chair)

Peta de Jager (CSIR)

Mladen Poluta (University of Cape Town)

Tobias van Reenen (CSIR)

Anton Stoltz (University of Pretoria)

ACKNOWLEDGEMENTS:
Edward A. Nardell (Brigham and Women's’ Hospital)

Richard L. Vincent (Mount Sinai Medical Center)

Steve N. Rudnick (Harvard School of Public Health)

Paul Jensen (US Centers for Disease Control)

Lindiwe Mvusi (NDoH)

FW Leuschner (University of Pretoria)

The development of the guidance document was supported by the Presidents Emergency Plan for Aids
Relief (PEPFAR) grant together with the US-CDC. The content of this report does not necessarily reflect
the opinion of the funders or those acknowledged or referenced in its development.
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

TABLE OF CONTENTS
Authors: ........................................................................................................................................................ 1
Acknowledgements: ..................................................................................................................................... 1
PREFACE ........................................................................................................................................................ 8
1 BACKGROUND: evidence based airborne infection controls ................................................................ 9
2 SAFETY................................................................................................................................................... 9
2.1.1 Safe Exposure Limit ............................................................................................................. 10
2.1.2 Reflections into the Lower Room ....................................................................................... 10
3 APPLICATIONS ..................................................................................................................................... 10
3.1 UVGI Application and environments .......................................................................................... 10
3.1.1 Enclosed Ventilated UVGI Room Air Cleaners .................................................................... 11
3.1.2 Efficacy testing .................................................................................................................... 11
3.1.3 Whole room UVGI disinfection ........................................................................................... 11
3.1.4 Upper room devices ............................................................................................................ 11
3.1.5 Upper room devices with integrated fans .......................................................................... 11
4 METROLOGY & EQUIPMENT ............................................................................................................... 11
4.1.1 Types of measurement equipment..................................................................................... 12
4.1.2 Procurement of measurement equipment ......................................................................... 12
5 PLANNING AND PROCUREMENT FOR FACILITIES ............................................................................... 12
5.1 Management Policy .................................................................................................................... 12
5.2 Sourcing specifications................................................................................................................ 12
5.3 Performance data and certification ............................................................................................ 13
6 DESIGN ................................................................................................................................................ 13
6.1 Outline of design lifecycle ........................................................................................................... 14
6.2 Radiometric Design process ........................................................................................................ 15
6.2.1 Characterization of UVGI devices for radiometric design................................................... 15
6.3 Prescriptive design process......................................................................................................... 16
6.3.1 Characterization of luminaires for prescriptive design....................................................... 16
6.4 Lamp aging, seasoning and life ................................................................................................... 17
6.4.1 Acceptance criteria ............................................................................................................. 18
6.5 Safety .......................................................................................................................................... 18
6.6 Design deliverables and reporting .............................................................................................. 19
6.7 Device standards ......................................................................................................................... 20
6.8 Factors affecting effectiveness ................................................................................................... 20

2|Page
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

6.8.1 Temperature ....................................................................................................................... 20


6.8.2 Relative Humidity ................................................................................................................ 20
6.8.3 Ventilation, air mixing and air movement .......................................................................... 20
6.8.4 Sunlight ............................................................................................................................... 21
7 Installation .......................................................................................................................................... 21
8 Commissioning .................................................................................................................................... 22
8.1 Start-up ....................................................................................................................................... 22
8.2 Initial performance verification testing ...................................................................................... 22
9 DOCUMENTATION .............................................................................................................................. 23
9.1 User Requirement Specification: ................................................................................................ 23
9.2 Design Report.............................................................................................................................. 23
9.3 Performance verification report ................................................................................................. 23
9.4 Operation and Maintenance Manual ......................................................................................... 23
9.5 Monitoring record....................................................................................................................... 24
10 Decommissioning and Disposal ...................................................................................................... 24
11 Additional Services .......................................................................................................................... 24
12 Handover......................................................................................................................................... 25
13 Maintenance and monitoring ......................................................................................................... 25
13.1 Maintenance Plan ....................................................................................................................... 25
13.1.1 Cleaning............................................................................................................................... 25
13.1.2 Lamp replacement .............................................................................................................. 25
13.1.3 Labelling .............................................................................................................................. 26
13.2 Corrective actions ....................................................................................................................... 26
13.2.1 Review of pre-existing installations. ................................................................................... 26
14 Training and competence ............................................................................................................... 27
14.1 Training and Responsibility Matrix ............................................................................................. 28
APPENDIX A BASIC PRINCIPLES OF UVGI RADIOMETRY .......................................................................... 32
APPENDIX B UVGI Maintenance Record Template - Example ................................................................ 35
APPENDIX C Example Signage ................................................................................................................. 36

3|Page
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

DEFINITIONS
Air exchange The ratio between the volumetric air supply to a space and the volume of that
rate space. Normally relating to fresh air supply only.
Device (UVGI) UVGI luminaire, classified as either open or closed
Equivalent air Air exchange rate under fully mixed conditions calculated to produce the
exchange rate equivalent effect of rated UVGI device (ACH)
Equivalent Ventilation rate under fully mixed conditions calculated to produce the
ventilation rate equivalent effect of rated UVGI device (ACH)
Irradiated Zone Zone in upper room irradiated to levels adequate to deactivate considered
airborne bacteria.
Rational Design Design process undertaken by a competent person whereby pre-determined
model design objectives are realized though a planned design process involving
scientific and mathematical principles.
Verification. The evaluation of whether or not a product, service, or system complies with a
regulation, requirement, specification, or imposed condition. It is often an
internal process.
Ultra-Violet Ultraviolet germicidal irradiation is a disinfection method that uses short
Germicidal wavelength ultraviolet radiation to inactivate microorganisms by the mutagenic
Irradiation effect of sufficient exposure short wavelength UV light.
Z-Value Relates to the sensitivity of a microbial species to UVGI. Is defined as the ratio of
the inactivation rate normalized for UV irradiance.

ABBREVIATIONS

ACH Air changes per hour


CADR Clean Air Delivery Rate
CADRe Equivalent clean air delivery rate
CDC Centres for Disease Control (USA)
CIE Commission Internationale d'Eclairage (International Commission on Illumination)
eACH Equivalent Air Changes per Hour
ECSA Engineering Council of South Africa
HBA Hazardous Biological Agent
ICNRIP International Commission on Non-Ionizing Radiation Protection
ICU Intensive Care Unit
IEEE Institute of Electrical and Electronics Engineers
IES Illuminating Engineering Society
IESNA Illuminating Engineering Society of North America
IPC Infection Prevention and Control
ISL Inverse Square Law
IUVA International Ultra Violet Association
LMPV Low Pressure Mercury Vapour
MRU Minimum Required Ultraviolet Fluence Rate
O&M Operation and Maintenance
PET Permissible Exposure Time
REL Recommended Exposure Limit
REL Recommended Exposure Limit

4|Page
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

SANS South African National Standard


SATS South African Technical Standard
TB Tuberculosis
URS User Requirement Specification
UV Ultra Violet
UV-b Ultra Violet – (280-315nm)
UV-c Ultra Violet – C (100-280nm)
UV-e Ultra Violet – eye sensitive bands
UV-g Ultra Violet – germicidal bands
UVGI Ultra Violet Germicidal Irradiation
WHO World Health organisation

5|Page
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

GUIDANCE APPLICATION MATRIX


The following matrix is presented to guide role players in the design, implementation and maintenance
of room air UVGI systems to applicable sections of this document. Applicability is classified as either
“required” or “recommended” reading.

CHAPTER Planning & Specification Testing & Operation


Design & Maintenance
Procurement
0 BACKGROUND: REQUIRED REQUIRED REQUIRED REQUIRED
evidence based
airborne infection
controls
2 Safety REQUIRED REQUIRED REQUIRED REQUIRED
3.1 UVGI Application REQUIRED REQUIRED NOT REQUIRED REQUIRED
and environments
3.1.1 Enclosed REQUIRED RECOMMENDED NOT REQUIRED NOT REQUIRED
Ventilated UVGI
Room Air Cleaners
3.1.2 Efficacy testing REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
3.1.3 Whole room UVGI REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
disinfection
3.1.4 Upper room REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
devices
3.1.5 Upper room REQUIRED REQUIRED REQUIRED NOT REQUIRED
devices with
integrated fans
4.1.1 Types of REQUIRED REQUIRED NOT REQUIRED
measurement
equipment
4.1.2 Procurement of RECOMMENDED REQUIRED NOT REQUIRED NOT REQUIRED
measurement
equipment
5.1 Management REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
Policy
5.2 Sourcing REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
specifications
5.3 Performance data REQUIRED REQUIRED NOT REQUIRED
and certification
0 Outline of REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
radiometric design
methods vs
prescriptive design
methods
6.2 Radiometric Design REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
process
6.2.1 Characterization of REQUIRED REQUIRED REQUIRED NOT REQUIRED
UVGI devices for
radiometric design
6.3 Prescriptive design REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
process

6|Page
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

CHAPTER Planning & Specification Testing & Operation


Design & Maintenance
Procurement
6.3.1 Characterization of REQUIRED REQUIRED REQUIRED NOT REQUIRED
luminaires for
prescriptive design
6.4 Lamp aging, REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
seasoning and life
6.4.1 Acceptance criteria REQUIRED RECOMMENDED REQUIRED NOT REQUIRED
6.6 Design deliverables REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
and reporting
6.7 Device standards REQUIRED REQUIRED RECOMMENDED NOT REQUIRED
6.8 Factors affecting REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
effectiveness
6.8.1 Temperature REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
6.8.2 Relative Humidity REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
6.8.3 Ventilation, air REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
mixing and air
movement
7 Installation REQUIRED RECOMMENDED NOT REQUIRED NOT REQUIRED
8.1 Start-up REQUIRED RECOMMENDED REQUIRED NOT REQUIRED
8.2 Initial performance REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
verification testing
9.1 User Requirement REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
Specification:
9.2 Design Report REQUIRED REQUIRED REQUIRED NOT REQUIRED
9.3 Performance REQUIRED RECOMMENDED REQUIRED NOT REQUIRED
verification report
9.4 Operation and REQUIRED REQUIRED REQUIRED NOT REQUIRED
Maintenance
Manual
9.5 Monitoring record REQUIRED NOT REQUIRED REQUIRED REQUIRED
10 Decommissioning REQUIRED REQUIRED NOT REQUIRED REQUIRED
and Disposal
11 Additional Services REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
12 Handover REQUIRED REQUIRED NOT REQUIRED REQUIRED
13.1 Maintenance Plan REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
13.1.1 Cleaning REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
13.1.2 Lamp replacement REQUIRED NOT REQUIRED REQUIRED NOT REQUIRED
13.2 Corrective actions REQUIRED NOT REQUIRED REQUIRED REQUIRED
13.2.1 Review of pre- REQUIRED REQUIRED NOT REQUIRED NOT REQUIRED
existing
installations.
0 Training and REQUIRED NOT REQUIRED REQUIRED REQUIRED
Responsibility
Matrix
APPENDIX REQUIRED RECOMMENDED RECOMMENDED RECOMMENDED
A
Principles of UVGI
Radiometry

7|Page
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

PREFACE
This document serves as a compilation of current best practice and review of relevant current
knowledge regarding the design, development and operation of indoor room air ultraviolet germicidal
irradiation (UVGI) systems for the purpose of controlling the transmission of airborne diseases such as
tuberculosis (TB) in institutional and congregate settings in South Africa.

This guidance document is intended for the implementation of UVGI systems for continuous disinfection
of room air. It is not intended for the applications of water or surface disinfection.

This document is intended to provide guidance to all designers, installers, owners and operators of new
or existing UVGI room air disinfection systems. It intends to provide technical information relating to the
effectiveness and safety limits, design and planning, installation, commissioning, operation and record
keeping of UVGI systems.

This guidance document is intended to function as an informative framework from which interested
parties can develop designs, specifications and protocols UVGI systems. This document does not
prescribe to offer solutions for every implementation or scenario and the user is therefore advised to
also rely on proven competencies in the relevant fields of photo-biology, ventilation, infection control,
relevant regulations (e.g. building, occupational health & safety, hazardous waste, etc) and national
policy in TB infection control and management.

To date no national standards on the design and implementation of UVGI systems exist however a
standard is currently in development.

8|Page
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

1 BACKGROUND: EVIDENCE BASED AIRBORNE INFECTION CONTROLS


This section presents a summary of key discoveries and events in the progress towards our current
understanding of UVGI technology.

The current burden of tuberculosis (TB) highlights the importance of controlling the transmission in
institutional and congregate settings. This problem is further exacerbated by the emergence of drug
resistant strains, opportunistic infections and HIV/AIDs.

Infectious TB aerosols, expelled during coughing, talking, sneezing or work procedures can remain
airborne in a room for hours after particle generation. This poses a risk to susceptible persons entering
the contaminated area. While administrative controls such as early detection and directly observed
treatment of infected individuals is foundational for a TB control strategy, there is great advantage to be
gained from the implementation of effective evidence-based environmental control measures. For
example, appropriately designed and functional mechanical ventilation systems should sufficiently dilute
or flush the airborne contaminants from indoor spaces and can be used as the primary environmental
control measure1. In contrast, dysfunctional or poorly designed ventilation systems offer little benefit, or
may further distribute contaminants, increasing the risk of TB transmission.

The WHO policy on TB infection control strongly recommends the application of defined ventilation
rates based on the number of occupants which is proportional to the risk. This approach is preferred to
the universal application of a volumetric room air exchange rate. Based on the recommended WHO
ventilation rate of 12 ACH (air changes per hour)2 in airborne precaution rooms, an adjustment for a
room volume of 24m³ yields an equivalent ventilation rate of 80l/s/person. This resultant value can be
applied to high risk settings with mechanical ventilation. 160l/s/person should be applied to similar
rooms relying on natural ventilation1,3. Where these recommended ventilation rates cannot be
achieved, the implementation of UVGI systems with a deactivation capacity equivalent to this
ventilation rate should be pursued.

The effectivity of UVGI and the susceptibility of the TB bacilli to UV-c (280nm-100nm, with peak
sensitivity at around 265nm) are well understood4,5 and proven to be effective in the disinfection of
microorganisms including M. tuberculosis (TB) in air,6,7. UVGI should therefore be considered as a valid
element in indoor airborne infection control strategy. Such studies also demonstrated the importance of
good vertical air mixing in the room, 8 9 10 11and safety of UVGI application.12

 From 1990s to the present a number of important articles and discoveries have been published
which concurs with previous findings that UVGI technology can reduce the risk of TB infection in
a variety of high-risk environments.”13

Although UVGI systems can greatly reduce the risk of airborne transmission, an inherent risk of
transmission will always remain. Therefore further research is warranted to determine the rates at
which the bacteria are inactivated sufficiently to stop transmission without causing harm to occupants.

2 SAFETY
UV-c has a much shorter wavelength and therefore a lower skin penetration depth thus does not easily
cause skin irritation or cancers when compared to the UV-a and UV-b found in sunlight. UV-c does
cause eye irritation at high exposure levels14.

Since open UVGI devices are principally designed to emit a certain amount of UV-c radiation, these
devices cannot be considered to be intrinsically safe. The UVGI system should be designed to achieve

9|Page
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

the maximum system output and maintain the lowest reasonable exposure levels in the occupied zone
of the room.

The primary safety concern with UVGI in controlling eye exposure to prevent irritation, limits the
irradiance limits achievable in the occupied room.6

2.1.1 Safe Exposure Limit


The occupied rooms considered for UVGI should not exceed an exposure dose below 6 mJ/cm² (for
mercury vapour lamps at 254 nm 15) and 3.8 mJ/cm² (at 265nm) per 8h14,16.

Therefore, in the typical application of 254nm devices, occupant exposure shall not exceed 0.40W/cm2
for a period of time equivalent to 4 hours per work day. If exceeded, steps must be taken to reduce the
effective irradiance dose to less than 6.0mJ/cm2 for the total duration per day.

It becomes important to consider the actual nature and usage of these spaces to determine appropriate
safe exposure rates. In determining these rates, occupancy factors such as the eye level of regular
occupants, position of occupants (seated, standing, lying down), frequency and duration of exposure,
need to be determined and incorporated. Therefore different areas or rooms will require different
safety limits.

Areas such as passages and wards would therefore demand vastly different safety limits from each
other.

The economic considerations of infection control suggest that high risk areas be given priority for air
disinfection. These areas will be determined through the outcome of a comprehensive risk assessment
of the facility. Installation of open UVGI devices in stairwells is not considered safe practice as there is
seldom sufficient soffit height for an effective upper room zone. Where passages are used as return air
plenums in recirculating ventilation systems, these passages shall be considered for air disinfection. For
further information see section 6.5 and refer to the ICNIRP guidelines on limits of exposure to ultraviolet
radiation16

For the safety of staff warning signs should be visible in all areas where UVGI is applied including the
device housings (see APPENDIX C for example).

2.1.2 Reflections into the Lower Room


The potential of high UV intensities being reflected into the occupied room must be considered by
designers and users. Certain materials and surfaces that reflect visible light might also reflect UV-c light;
for example consider reflectors of regular open luminaires, windows, exposed ducting and metallic or
high gloss architectural finishes in the upper room.

3 APPLICATIONS
3.1 UVGI Application and environments
The application of UVGI should form a component of a comprehensive IPC policy and should not be seen
as a substitute thereof. Careful consideration should be given to the selection of the type of UVGI
application in a particular facility. Efficacy data obtained from laboratory settings or controlled
environments should be interpreted with caution as the lab conditions differ from the actual facility
environment. For any room air UVGI system, the dose received by any irradiated droplet nuclei is
dependent on the exposure time. For this reason the ratio between the volumes of the irradiated zone
and the whole room greatly affects the efficacy of the system. For example, doubling the volume of the
irradiated zone for the same room volume would double the average dose received by droplet nuclei.

10 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

3.1.1 Enclosed Ventilated UVGI Room Air Cleaners


Room air cleaners involve application of enclosed UVGI lamps in housings, through which air is recycled,
to achieve some measure of clean air delivery to the room. The advantages of this application are the
safety aspects and lower maintenance costs.

3.1.2 Efficacy testing


Efficacy testing should be uniform for all UVGI applications. Only reputable laboratories with experience
and competence should be selected for the laboratory based performance assessment. The
performance of each UVGI application should be compared when the device is on versus off, and not to
other devices.

3.1.3 Whole room UVGI disinfection


Whole room disinfection is a UVGI process which mainly attempts to sterilise room surfaces by exposing
the entire room volume to high fluence levels.

This process is not appropriate for air disinfection as, in order to perform the process safely and
effectively, the room needs to be unoccupied. Similarly, the surest way to reduce the transmission risk
within an indoor space is to remove all potential infectious and susceptible persons. In this occupancy
state, air disinfection is no longer required as the risk has been removed. As soon as the room occupants
return to the “disinfected” room the risk of transmission returns with them and the whole process is
rendered futile.

The application of room surface disinfection by UVGI is limited to only those surfaces for which the
process can be closely controlled and for which the process can be validated. The primary consideration
with this form of room surface disinfection is that of unintended surface shading. In addition, airborne
organisms in the form of droplet nuclei are more susceptible to UV-c than organisms on surfaces.17

3.1.4 Upper room devices


Upper room devices are fundamentally divided into three different types according to installation
configuration.

Corner type devices produce an irradiation field of 90°, wall type devices produce a 180° field and
pendant devices produce a 360° field. These devices should not be installed in a stagnant zone of the
room as this may affect its effectiveness.

3.1.5 Upper room devices with integrated fans


UVGI devices fitted with an integral fan are designed to aid ventilation around the device. The fans
employed are often 80mm computer fans which should not be considered to aid the overall room air
mixing to a significant measure18. A disadvantage is that these integral fans degenerate and become
increasingly noisy with age. Facilities with these applications should have a regular maintenance
programme to change defective fans and prevent the device being turned off.

4 METROLOGY & EQUIPMENT


Measurements of UV levels are required for characterizing and evaluating the performance of UVGI
systems and installations through two distinct but equally important requirements, namely:

I. to assess the installation’s performance in inactivating airborne microorganisms, such as


TB, based on the acceptable criteria of the lower fluence limit for either the upper or
whole room, dependant on the design methodology. For UVGI effectiveness, cosine
corrected irradiance meters should be used. The meter used should be calibrated within

11 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

one decade of the required range. That implies calibration to the first decimal (0.1
µW/cm²).
II. to ensure occupational safety and health of the occupants the lower portion of the
room, referred to as the “occupied zone” must be evaluated. For eye safety, radiance
measurements are most appropriate and the meter used should be calibrated to the
second decimal (0.01 µW/cm²).

The methodology for the characterisation of UVGI devices falls outside of the current scope of this
guidance document. The reader is referred to SATS 1706: UVGI Luminaires - Safety and performance
requirements 19.

4.1.1 Types of measurement equipment


Two types of devices are ideally required for the evaluation of the installation’s UVGI efficacy and eye
safety.

UV meters used for safety assessments should be able to detect UV in both the UV-b and UV-c
spectrums, as both of these spectrums are harmful to eyes.

However, installed UVGI devices fitted with good quality low pressure mercury vapour arc lamps
(dominant at 254nm) generally does not emit UV-b and may not be of concern. These safety
measurements should form part of the characterisation and commissioning of the fixture as the human
eye is sensitive to the UV-b spectrum (280-315nm). The onus is on the user to confirm or have these
safety limits confirmed14.

4.1.2 Procurement of measurement equipment


Facilities which utilise UVGI shall have access to dedicated, calibrated measurement equipment for
routine monitoring of installation performance to ensure appropriate maintenance planning and
response which has financial implications. The procurement of dedicated UVGI irradiance meters may
be included in the system supplier’s brief and its calibration must be considered on a regular basis
(usually annually) according to the manufacturer’s instructions.

5 PLANNING AND PROCUREMENT FOR FACILITIES


5.1 Management Policy
The uncontrolled use of UVGI without a comprehensive management policy can result in fruitless
expenditure and create a false sense of protection, thereby exposing building occupants to an increased
risk.

The management policy should align with the regional and national IPC policies and should include
sourcing and procurement, maintenance, training, decommissioning and disposal of UVGI devices.

5.2 Sourcing specifications


In the absence of norms and standards with respect to UVGI devices and installations, careful
consideration must be taken in the selection process which should include designer competence and
liability. When sourcing UVGI system design vendors, consideration should be given to their track record
and their proven or tested competence. Negligence in this regard could result in medico-legal or civil
claims or even criminal charges against persons involved in the implementation of malfunctioning
systems. The procured services must include supply, installation, maintenance and disposal of UVGI
devices. The procurer must ensure that any consumable parts/components of the device are available
from more than one reputable local supplier to avoid the device becoming obsolete. Where single
vendors are used for more than one service it should be ensured that no conflicts of interest are

12 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

created. In addition, the user is advised to procure the services of a vendor who utilises designers that
are registered with an engineering professional council such as the Engineering Council of South Africa
(ECSA). Cost comparisons can be done with vendors that comply with the minimum performance
standard and maintenance requirements (South African Technical Standard 1706: 2015 SOUTH AFRICAN
TECHNICAL STANDARD UVGI Luminaires - Safety and performance requirements.20).

5.3 Performance data and certification


When UVGI devices are selected for procurement, the provision of standardised performance data must
be included in the device specification. This independently validated performance data provided should
include but not be limited to:

 The total maintained UVGI radiant flux value per device model (mW)
 The device’s effective clean air delivery rate (l/s)
 Maximum 1m, 2m and 3m maintained and lifetime deprecated irradiance measurement values
 Three or more recommended lamp models
 Rated lamp life
 Device cleaning and maintenance instructions

6 DESIGN
The fundamental objective of UVGI system design is an airborne transmission reduction of 80% or
better, which can be achieved by the criteria of two approaches recommended:

 comprehensive rational or radiometric design process – two methods are described


 prescriptive design process - design based on prescriptive occupancy and performance criteria

Although, the prescriptive route offers a simplified design process the rational design approach may
yield more economical long term solutions. The selection between them will depend on the availability
of data and design tools.

The following decision tree describes the design path.

13 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

6.1 Outline of design lifecycle

Figure 1

14 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

6.2 Radiometric Design process


Two methods are proposed for the radiometric design approach, both of which should be undertaken by
suitably competent persons.

The first rational design method identified follows the guidance of CIE 155:200315 and the design
completion will need quantification and verification of the following:

 Selected UVGI device’s radiometric data


 Considered pathogen’s UV-c sensitivity (Z-value)21
 Considered pathogen’s infectious dose
 Air exchange rates and ventilation efficiency parameters

The design objective of this method is a prescribed reduction in transmission of 80% or higher. A
comprehensive discussion of this design methodology is currently beyond the scope of this guide.

A second rational design process is available through application of design software targeting prescribed
whole-room radiant flux levels. However, this method is only appropriate for open type UVGI devices.
CAD based design tools have the capacity to evaluate photometric data and resolve the irradiance
levels, radiance levels and probable dose distribution throughout the modelled space, thereby
optimising UVGI system solutions. The software would be distinct from conventional lighting design
software in that it is intended for the calculation of UV-c irradiance levels and not radiance or
illuminance. The modelled room needs to accurately represent the actual room, number of occupants,
ventilation capacity and regime, building finish materials, UV-c sensitivity of the considered pathogens
and the UVGI device’s capacity and characteristics. The software should additionally have the capability
of modelling UV-c reflection off room surfaces.

6.2.1 Characterization of UVGI devices for radiometric design


Only UVGI devices that have been independently characterised for radiometric output of new stabilised
lamps should be considered for selection. Radiometric data should be published in the form of IES
format files22.

Radiant flux characterization could take the form of either:

 Total Maintained UVGI Radiant Flux (ФT, mW), measured using an integrating sphere in
accordance with the SATS 1706: UVGI Luminaires - Safety and performance requirements 20 and
UVGI MEASUREMENT AND INSTRUMENTATION CALIBRATION PLAN 23
 Near field gonioradiometry24, which should be used understanding the limitations where noise
to signal ratio of measuring equipment at the assessment field depth might significantly affect
the measurement results.

6.2.1.1 Design objective


CAD based solutions should achieve a reduction in transmission of better than 80%. These conditions
should be achieved while maintaining a maximum uniformity of irradiance and the prescribed levels of
lower room eye safety.

6.2.1.2 Design data


The following UVGI device performance data based on independent characterisation should be obtained
by the device manufacturers for the purposes of a rational design.

 Total Maintained UVGI Radiant Flux (ФT, mW),


 The device’s equivalent clean air delivery rate (l/s)

15 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

 Maximum 1m, 2m and 3m maintained and lifetime depreciated irradiance measurement values
(µW/cm²)
 Radiometric data in the IES file format.

The designer needs knowledge of the diffuse and specular UV-c reflective properties of the materials
and coatings present in the lower and upper room as well as the performance indices of any existing
ventilation systems.

Where the design objective is a prescribed reduction in transmission of better than 80% the designer
should obtain the relative sensitivity to UVGI of the considered airborne infectious agents (Z-value21).

6.3 Prescriptive design process


The prescriptive design method follows simplified design codes. This design process is applicable for well
mixed room air, which can be achieved by installing mixing fans as part of the UVGI system. In addition,
the prescriptive design process assumes that there are no UV-reflective surfaces in the upper room
significant enough to affect eye safety.

6.3.1 Characterization of luminaires for prescriptive design


Apart from the total maintained radiant flux, the selected device’s equivalent clean air delivery rate is
required (CADRe) which can be determined by a competent and experienced testing facility through a
static decay or constant generation test in a 4m x 4m x 3.5m high bio-aerosol chamber with fully mixed
air18. The evaluation represents the equivalent ventilation rate, in litres per second, which the
considered device can provide under prescribed laboratory conditions.

6.3.1.1 Design objective


The prescriptive design process aims to determine the required number of UVGI devices for the
considered indoor space in accordance with the WHO recommended minimum ventilation rate of 80 l/s
per person. If the considered room has an existing and functional mechanical ventilation system, the
outside air portion of that system’s ventilation rate should be included in the calculation to reduce the
number of UVGI devices required.

6.3.1.2 Design data


The prescriptive design process requires the following set of device and application data:

 Total Maintained UVGI Radiant Flux (ФT, mW), measured using an integrating sphere in
accordance with SATS 1706:201520 and the UVGI MEASUREMENT AND INSTRUMENTATION AND
CALIBRATION PLAN 23
 The selected device’s minimum equivalent clean air delivery rate (CADRe, l/s).
 Volume of considered room (Vr, m³).
 The occupancy profile of the considered room (Np, number of persons).
 The effective mechanical ventilation rate of the considered room (Qv, l/s).

6.3.1.3 Method
The following calculations should be made:
(80×𝑁𝑝 )−𝑄𝑣
𝑁𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑑𝑒𝑣𝑖𝑐𝑒𝑠CADR = CADR𝑒
-(7.1)

Where,

 CADRe (l/s) is the selected device’s minimum equivalent clean air delivery rate
 Vr (m³) is the volume of considered room

16 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

 Np (number of persons) is the occupancy profile of the considered room


 Qv (l/s) is the normal mechanical ventilation rate of the considered room

In addition to the CADRe a minimum required UVGI fluence rate (MRU) of 20mW/m³ should be
maintained. In order to determine the number of UVGI devices required to achieve the MRU, the
following equation should be used.
20×𝑉𝑟
𝑁𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑑𝑒𝑣𝑖𝑐𝑒𝑠Ф = Ф𝑇
-(7.2)

Where an ideal design requires that a number of devices be installed with various performance
characteristics, the weighted averages of the CADRe and total maintained radiant flux should be
calculated and used according to the distribution of devices selected. The devices shall be distributed
such that the space served by each device is proportional to its declared total maintained radiant flux.
The orientation and positioning of upper room devices shall also be such that the irradiated volume and
length of the rays cast is maximised without significantly reducing the uniformity of irradiation in that
volume.
(𝑁1 ×Ф1 )+(𝑁2 ×Ф2 )+ … +(𝑁𝑛 ×Ф𝑛 )
𝑀𝑅𝑈𝑎𝑣𝑔 = 𝑁1 +𝑁2 + … +𝑁𝑛
-(7.3)

(𝑁1 ×Ф1 )+(𝑁2 ×Ф2 )+ … +(𝑁𝑛 ×Ф𝑛 )


𝐶𝐴𝐷𝑅𝑒 𝑎𝑣𝑔 = -(7.4)
𝑁1 +𝑁2 + … +𝑁𝑛

Where,

 Nx is the number of fittings of type x


 Ф𝑥 is the total maintained radiant flux of device type x

For example:

Assuming 3 devices are required with a CADRe of 150 l/s and 2 devices of 250 l/s respectively. The
weighted performance average is calculated as follows:
(3 × 150) + (2 × 250) 950
𝐶𝐴𝐷𝑅𝑒 𝑎𝑣𝑔 = = = 190 𝑙/𝑠
3+2 5
In designs incorporating a number of similar UVGI devices, the devices should be evenly distributed
within the room.

6.4 Lamp aging, seasoning and life


UV lamp output deteriorates with lamp age with rapid decline and instability during the initial 100 hours
(about four days) of service thereafter stabilizing into a gradual rate of decline. Therefore, a 100 hour
burn-in is recommended prior to device characterisation or system verification tests. In the case of
dimmable systems, the lamps should be burnt-in at full output for the initial seasoning period before the
lamps are dimmed for the first time. Refer to the IESNA for
further information on lamp seasoning25. Guidance
documents and standards for predicting lamp life and
performance decay of visible light should not be used to
interpret loss factors of UV-c, as the penetration rate of UV-
c through accumulated dirt is not comparable to that of
visible light.

It is recommended that all designs be conducted using the

Figure 2
17 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

total maintained radiant flux.

A lifetime lamp aging depreciation factor of 0.7 (30% reduction) from total maintained radiant flux, or
new stabilised output, over the life of the lamp is consistent with good engineering practice26.
Considering the lifetime degradation of the UV lamps, the final capacity is referred as the lifetime
deprecated UV-c radiant flux. Systems shall be designed to remain effective even with this level of
degradation.

SANS 50285:201028 defines the rated average lamp life as the average time in hours, declared by the
responsible vendor, in which half a representative group of lamps will fail. The actual average life shall
not be less than 90% of this declared value. The IESNA Approved Method for Life Testing of Fluorescent
Lamps (IESNA LM-40-01) defines the standardised operating cycle for these tests as 3 hours on and 20
minutes off. This type of cycle is not appropriate for most UVGI installations as UVGI systems often run
continuously during occupied hours.

As seen from the graph in Figure 2 27, extended operating cycle lengths can deliver better lamp life than
predicted by the standardised tests. System designers and planners are therefore advised to request
additional 8h and 12h predicted lamp life figures for the considered lamps.

6.4.1 Acceptance criteria


6.4.1.1 Effectiveness
Design requirements for effectiveness can be assessed through compliance with the following:

 MRU minimum limit of 20mW/m³


 A UVGI equivalent ventilation rate of the product of 80 l/s per person times the peak number of
room occupants, as recommended by WHO.

Any shortfall between the required and the actual ventilation rate or MRU respectively should therefore
be corrected for by the number of UVGI devices installed as determined using the formulae in section
6.3. The number of devices required should not be less than the minimum number determined, or
exceed this value by more than 1 unit.

The final location of fixtures must take into consideration possible interference from obstructions to
airflow around the device or radiation from the device. Obstructions to avoid include light fixtures,
beams, ductwork, piping, furniture, equipment or non-prescribed lighting diffusers.

It is important to note that compliance with the SABS specification for UVGI luminaires, does not include
microbiological efficiency test to verify effectiveness. This can only be done by a competent laboratory
with experience in efficacy testing. While various percentages of effectiveness have been reported in
the scientific literature and technical reports, the UV dose intervention should inactivate airborne
infectious agents by at least 90% (D90) or a percentage equivalent of the ventilation rate of 80l/s/per
person as recommended by the WHO.

6.5 Safety
To ensure safety, the installation design must be aligned to the specific design of the UVGI device or
system and the manufacturer’s recommendations. Tilting of the device under normal operation or
conditions must be avoided. This also applies to self-aligning device mounting systems under the
influence of air currents. The minimum installation height of the lower horizontal reference plane of any
open UVGI device is 2.2m. The highest eye level of the occupied zone is dependent on the room
function but is considered by default to be 1.7 meters, unless otherwise motivated by room function

18 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

analysis. Acceptance criteria for room safety assessments are the upper limits of eye and skin exposure
as described in section 2 above.

The permissible exposure times (PET) can be calculated for various irradiation levels using the following
equation:
𝑅𝐸𝐿
𝑃𝐸𝑇 (𝑠𝑒𝑐𝑜𝑛𝑑𝑠) = 𝐼
- 6.1

Where:

REL is the recommended exposure limit for a considered wavelength14, 25

I is the measured irradiance level and the considered wavelength

The recommended exposure limit for the 254 nm wavelength is readily available. The safe exposure
limit for LED devices with 265nm is reduced by a factor of 0.62.

The strategy for determining acceptable exposure levels in the lower room in typical healthcare settings
is proposed in Table 1.

Table 1

ZONE EYE ESTIMATED DAILY IRRADIANCE


LEVEL EXPOSURE TIME LIMIT
(m) (h) (uW/cm²)
Corridors (no waiting) 1.7 1.0 1.6
Indoor Waiting Areas 1.7 3 0.5
ICU 1.7 4 0.4
1.0 6 0.3
Bedrooms/Offices 1.7 2 0.8
1.2 4 0.4
Wards/Dormitories 1.7 2 0.8
1.2 4 0.4
Isolation Rooms 1.7 2 0.8
1.2 4 0.4
Consultation Rooms 1.7 2 0.8
Bronchoscopy rooms 1.7 2 0.8
Autopsy Rooms 1.7 2 0.8
Other areas 1.7 4 0.4

6.6 Design deliverables and reporting


On completion of the design the designer should submit a design report to the user for approval before
installation commences so that the designer is obligated to his/her professional responsibility or liability.
Refer to section 9 for components of a design report.

19 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

6.7 Device standards


The SABS has published technical standards for performance and safety and installation and
maintenance of UVGI air disinfection devices in South Africa. SATS 1706: UVGI Luminaires - Safety and
performance requirements 20

6.8 Factors affecting effectiveness


6.8.1 Temperature
Extreme low and high temperatures can both affect the effectiveness of UVGI due to reduced output of
both low pressure mercury vapour (LPMV) lamps and luminaire control gear. In addition,
microorganisms show a reduced sensitivity to UVGI at low temperatures. Heating and cooling
interventions are controlled at limits of occupancy comfort which, fortunately fall outside the
temperature extremes affecting effectiveness.

6.8.2 Relative Humidity


Many studies have shown that high humidity reduces the effectiveness of UVGI due to a combination of
factors including the DNA conformation, self-repair mechanisms and protection from presence of
moisture26,27,28. There is some indication that UVGI effectiveness may also be reduced at low RH levels25.
The recommended room relative humidity range of 25-60% is consistent with limits for occupancy
comfort. In climatic regions with a prevailing high humidity, it may be prudent to overdose the upper
room although the appropriate values for such a dosing strategy are not readily available. It should be
noted here that the recommendation of 20mW/m³ are already on the upper end of proven guidance
values33.

6.8.3 Ventilation, air mixing and air movement


Air UVGI disinfection performance is dependent on air movement rates. Similarly, upper room UVGI
effectiveness is dependent on maximising air mixing either through natural convection or mechanically
aided (e.g. paddle fan) however, not all air distribution regimes are equally effective.29 The effectiveness
of UVGI may be reduced if the mechanical ventilation rates in a room are increased together with
equivalent room air mixing. Ventilation is a primary environmental control against indoor airborne
transmission; therefore the ventilation capacity of the considered area should be determined before
deciding to implement a UVGI solution. Although, reduced UVGI effectiveness has been reported where
mechanical ventilation rates exceed 6 air changes per hour30,26, this should be assessed per area as
facility profiles differ and the increased ventilation rate may adequately serve to address the airborne
contaminants.

Mixing ventilation

While UVGI systems rely on air movement and good air mixing, it has been shown that not all room air
distribution regimes are equally effective for supporting upper room UVGI29. Mixing room air
distribution systems where both supply and extract air terminals are positioned in the upper room is
least suited to complementing upper room UVGI systems29 as is does not always force a vertical air
exchange in the room. In winter heating modes this air terminal arrangement can fundamentally negate
the effectiveness of upper room UVGI as the buoyancy of supplied warm air remains in the upper room,
increasing the contaminant removal effectiveness of the upper room while decreasing the ventilation
efficiency of the occupied space, or lower room. This may reduce the contaminant removal
effectiveness, rendering the UVGI system ineffective.

Displacement ventilation

20 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

Well-designed stratified room air distribution systems improve room air quality and reduce the mean
age of air. This system moves contaminants to the upper room for extraction while the occupied zone is
flooded with relatively clean air. Displacement ventilation can be incompatible with upper room UVGI
systems as the upper room air is only decontaminated before it is extracted. In these instances, if the air
is to be recirculated, in-duct UVGI or enclosed room UVGI devices could be considered for greater
efficiency.

Air mixing

Where no mechanical ventilation is present in the room, it is strongly recommended to include mixing
fans with the UVGI installation.

The integrated application of low speed ceiling fans can ensure a constant vertical air exchange and
improved UVGI effectiveness. The direction of the fan is not thought to significantly affect the
performance of upper room UVGI; however, it is recommended that fans blow down in spaces with high
ceilings to prevent lower room stagnancy. For acceptable occupant comfort, the fan direction may need
to be changed between summer and winter to reduce the negative effect of drafts during winter36. The
designed rate of vertical air exchange between the upper and lower room should not be less than the
installed UVGI system’s equivalent air change rate (eACH), as derived from the CADRe.

Electrical interlocking between mixing fans and the UVGI system is recommended to ensure both
operate simultaneously. Fans and UV light controls should be located away from public access.

Ceiling fan installations require earth leakage protection only where they are low enough to be touched
by room occupants, and should be installed on protected circuits accordingly. Motor control gear and
wiring shall be able to withstand expected rotor overload current (LRA) and heat.

6.8.4 Sunlight
Many microbiological organisms exhibit the ability to repair damaged DNA through a process of photo-
reactivation. Since the photo-reactivation rate counters the deactivation rate from UVGI, it reduces the
system’s overall effectiveness. The potential reactivation rate in a room can be reduced by shading the
indoor space from direct sunlight.

7 INSTALLATION
The installation of any UVGI device or system should comply with the requirements of the local building
and electrical regulations and shall be supported with the required certificates of compliance. Lines of
sight need to be considered when planning device locations to avoid aligning with commonly viewed
items such as notice boards, monitors, televisions etc. When planning the location and installation of the
UVGI devices, the load bearing capacity of structural elements (e.g. ceilings, walls) should be considered.

7.1.1.1 Open UVGI Device mounting height


The effectiveness of a UVGI installation is a function of the volumetric ratio between the irradiated
upper room zone and the whole room and should be installed as low in the room as possible to
maximises the upper room irradiated volume yet maintain safe levels in the occupied zone. The
following table (

Table 2) is adapted from the Guidelines for the utilisation of ultraviolet germicidal irradiation (UVGI)
technology in controlling the transmission of tuberculosis in health care facilities in South Africa31.

Table 2

21 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

Ceiling Recommended Mounting Height:


Height Finished floor to lower horizontal reference plane
(m) (m)
Corner type (90°) Wall type (180°) Pendant type
(360°)
2.4 2.2 2.2 2.4
2.7 2.3 2.3 2.4
3.0 2.4 2.4 2.4

7.1.1.2 Upper room UVGI switching and operation


7.1.1.3 Labelling
For better management and traceability, each installed device should be clearly marked with a unique
device identification (ID). The device ID shall include the supplier ID, installation year, installation
contract identifier, area/location number and device number. The following format is proposed:

ABC 14/01/01/02

7.1.1.4 Completion
The UVGI installation phase shall only be deemed complete when the following have been satisfactorily
achieved:

1. system adjustments made (to rectify any unforeseen problems)


2. performance test (including safety) reports handed over and accepted
3. Compliance documentation issued.

8 COMMISSIONING
Commissioning which includes start-up and verification tests should be conducted safely by a suitably
qualified service provider who is independent of the system installer and designer. The results of the
verification test results are only valid if these were witnessed and approved by a competent
representative of the client. This would typically take the form of the signed-off design report (see
below section 9.2).

8.1 Start-up
Prior to start-up and the initial verification testing, each device and lamp must be cleaned according to
the manufacturer’s instructions. The system start-up should commence after the installation process is
complete to allow stabilisation and lamp seasoning as described above. For larger installations a planned
concurrent start-up process can be followed assuming this is well managed and documented to ensure
sufficient time for each lamp to stabilise.

8.2 Initial performance verification testing


Once the start-up process has been completed each device’s performance should be verified
independently. This is achieved by first switching all devices off then turning each device on to measure
its output individually. This procedure is to ensure that only the output of the considered device is
measured during the tests. The output must concur with the initial specification stipulated in the
procurement bid. Prior to verification testing, the initial performance criteria shall have been agreed on.

During any work in the upper room or with open UVGI devices, eye and skin protection should be worn.

22 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

Each individual device is to have its 1m and 2m maximum irradiance measured and compared with the
manufacturer’s declared depreciated performance values. Each new or refurbished device will be
independently accepted for service when it is proven to exceed these depreciated performance values
by at least 43%26 (30/70). Non-compliant devices shall be repaired or replaced. Records of the device,
performance and safety measurement location, date, direction, irradiance meter, results and technician
are to be recorded (e.g. monitoring template, APPENDIX B).

Eye safety measurements should be made once all UVGI devices are installed, switched-on and
stabilised. Radiance levels at locations and heights with long duration occupancy, such as work stations,
beds, waiting locations for seated or standing persons should be evaluated for compliance with the
recommendations of section 2. If the maximum reading found exceeds the safety limit, corrective action
should be taken by the supplier and supported by evidence.

9 DOCUMENTATION
A number of reports and records are required during the design and implementation of room air UVGI
systems.

9.1 User Requirement Specification:


The user requirement specification (URS) includes the initial risk assessment outcome of the facility and
identifies the areas requiring UVGI intervention. This specification must comply with relevant technical
standards and guidelines. The URS should address the user’s requirements for all aspects of the system’s
lifecycle from design to decommissioning and disposal.

9.2 Design Report


The system designer will present a final technical design as prescribed by the URS. The design report
shall include a description of the design methodology including performance criteria and design
assumptions. Prior to commencement with the installation, the user should review and sign acceptance
of the design report. Acceptance of this report does not absolve the designer of any professional
responsibilities of liability.

9.3 Performance verification report


On completion of testing, a performance verification report must be approved by the designer and user.
The report should reference the acceptance criteria established in the design report.

9.4 Operation and Maintenance Manual


Two copies of an operation and maintenance manual (O&M) must be issued to the user on completion
of the verification testing as stipulated in the URS. One copy should remain with the facility’s
maintenance office on site while the other could be retained as an off-site copy. The O&M should
include copies of the following:

a) Designer’s trade name and contact details.


b) Designer’s professional or technical registration details.
c) Room selection methodology; alternatively, where this selection was not made by the design
consultant, a copy of the design brief.
d) Electrical wiring drawings.
e) Electrical compliance certificate covering all building electrical work relate to the UVGI
installation.

23 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

f) Plan and elevation drawings of considered rooms indicating device installation details,
supporting structures and obstructions to radiation and airflow.
g) Design data, including assumptions such as room occupancy and ventilation levels.
h) Selected device performance criteria.
i) Selected device data sheets.
j) Selected lamp data sheets including rate lamp life and details of 3 alternate lamps and suppliers.
k) Start up and shut down procedures.
l) Safety instructions.
m) Maintenance and monitoring protocol and plan.
n) Maintenance and monitoring log-book template.
o) Initial performance and safety verification acceptance criteria.
p) Performance verification reports
q) Device cleaning instructions.
r) Spare parts list with recommended stock levels.
s) Decommissioning and disposal instructions

9.5 Monitoring record


In every workplace in which hazardous biological agents are deliberately produced, processed, used,
handled, stored or transported; or which may result in persons being exposed to hazardous biological
agents (HBA) in the performance of their duties, the R1390 Government Gazette of 27 December 2001
No. 22956 is applicable. This legislation requires that the employer of persons in that workplace shall
ensure that all control measures, equipment and facilities provided for transmission-based precautions
are maintained in good working order; and that thorough examinations and tests of engineering control
measures - such as UVGI - are carried out at intervals not exceeding 24 months by an approved
inspection authority or their verified delegate.

An employer is required to keep a record of the examinations and tests carried out in terms of
regulation and of any repairs resulting from these investigations and tests for at least three years.

A record of routine monitoring needs to be established as per suggested template (APPENDIX B) and
retained on site.

A responsible person shall be appointed by the IPC committee to establish and maintain the
maintenance and monitoring record.

10 DECOMMISSIONING AND DISPOSAL


Mercury vapour lamps are a source of environmental contamination and can pose a health and safety
risk, they must therefore be included in the facility’s hazardous waste disposal plan and records should
be kept. A decommissioning plan should be established for the safe decommissioning of unserviceable
devices and systems. The user is responsible for the safe decommissioning and disposal of hazardous
waste; however, a suitably qualified service provider can be contracted to complete this work.

11 ADDITIONAL SERVICES
The inclusion of at least a 3 year duration maintenance, monitoring and training contract within the
design and implementation service contract as defined in sections 12 and 13 is recommended. The
contract should include a maintenance and training component before final handover.

24 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

12 HANDOVER
The system should only be handed over to the user once the operation and maintenance manual has
been compiled and issued with the required documentation.

13 MAINTENANCE AND MONITORING


13.1 Maintenance Plan
Since room air UVGI systems are predominantly passive installations, the maintenance activities and
performance monitoring are fundamental. If the maintenance and monitoring is done in-house it should
be conducted by trained and competent personnel. All monitoring should be conducted using an
irradiance meter and protective equipment should be worn (e.g. eye and skin).

13.1.1 Cleaning
Accumulation of dust on the reflective surfaces and lamps of UVGI devices accounts for a significant
decline in device performance. Device cleaning and maintenance shall be performed in accordance with
the device manufacturer’s recommendations. In the absence of manufacturer’s instructions the
following guidance can be followed:

 Don protective personal equipment before entering irradiated zone or opening UVGI devices.
 Turn off fixture and allow lamps to cool
 Use a dry lint free cloth to remove any dust from external surfaces and between louvers
 Open unit as stipulated by manufacturer
 Handle lamps with clean cotton gloves to prevent oil deposits on lamps and reflective surfaces
 Change lamps according to a prescribed schedule or if indicated by the routine performance
monitoring.
 Change failed ballasts of flickering lamps
 Use a cloth dampened with alcohol to clean lamps and reflectors (NOT SOAPY WATER). Carefully
apply pressure to remove stubborn dirt
 Wipe lamps, reflectors, louvers and external surfaces with a clean dry soft cotton cloth
 Close fixtures properly
 Turn on the system and verify lamp operation using eye protection
 Record inspection, cleaning and replacement in maintenance logbook as recommended in
APPENDIX B

The maximum reasonable maintenance interval for cleaning and monitoring interventions would
typically 3 months; however this will vary between installation locations. This interval will differ
depending on the type of ventilation (mechanical or natural) as well as environmental dust levels. The
maintenance plan can be adjusted based on performance data obtained during routine monitoring.

13.1.2 Lamp replacement


UV lamps or their control gear can malfunction/fail at any time or can degrade gradually with time
affecting the lamp effectiveness, thus requiring replacement.

Economical replacement of lamps due to gradual degradation can be achieved through the routine
performance monitoring using a UVGI irradiance meter. For facilities with a large number of installed
lamps and with the technical capacity to conduct a monitoring program, spot re-lamping based on
individual device performance against the minimum functional criteria is recommended (70% of new
seasoned output26).

25 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

Where UVGI installations are not sufficiently large to warrant the procurement and maintenance of a
UVGI meter, a group re-lamping strategy should be considered.

The unexpected failure of mercury vapour lamps can be identified through the visual inspection of
lamps when they are turned on. While the characteristic blue glow of these lamps is not a reliable
indicator of UV-c output, it will indicate whether a lamp or its control gear has unexpectedly failed. In
devices with lamp clusters, special consideration should be given to safely inspecting individual lamps
for failure.

13.1.3 Labelling
A writable label could be attached to the device in a visible and easily accessible location for the
purposes of recording the last maintenance, lamp replacements and cleaning dates.

It is advisable to physically record the irradiance measurement locations and orientations with discreet
and durable markers in the room. This would ensure repeatability between measurements. Such a
marker could be in the form of an arrow with an included number. The direction of the arrow should
unambiguously indicate the direction in which the reading should be taken.

13.2 Corrective actions


13.2.1 Review of pre-existing installations.
For the first evaluation of pre-existing installations (i.e. installed prior to adoption of this guideline or
SATS 1706: UVGI Luminaires - Safety and performance requirement 20), each device model needs to be
identified by user’s appointed responsible person and the following manufacturer’s declared
performance data is to be obtained for each device.

1. Total maintained UVGI radiant flux (mW)


2. The device’s equivalent clean air delivery rate (l/s)
3. Maximum 1m, 2m and 3m maintained and lifetime depreciated irradiance measurement values
(µW/cm²)
4. Device cleaning and maintenance instructions
5. Lamp replacement details

Where data is not available, the devices cannot be evaluated and should be declared unsuitable for
service.

The number of devices installed shall be checked against the method prescribed in section 6.3 above.
Where performance data is available and the correct number of devices is installed, a monitoring and
maintenance plan needs to be established in accordance with section 13.1 above.

The devices should be cleaned before the initial performance assessment in accordance with the
manufacturer’s instructions or as described in section 13.1.1 and considering the declared maximum 1m
and 2m depreciated irradiance measurement values. Note that the acceptance criterion is different
from that of new or refurbished installations. (See section 13.1.2)

For devices that pass the initial performance tests, the devices shall enter into a maintenance and
monitoring program as described above.

If individual devices fail the initial performance measurement, lamp re-cleaning and ballast replacement,
respectively are recommended, and if these still fail, the devices should be marked clearly with a red
non-conformance sticker and scheduled for decommissioning and disposal. Any new ballasts and lamps
installed in failed devices should be immediately salvaged for re-use.

26 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

DEVICE UNSERVICEABLE
DO NOT RE-USE OR REPAIR

Figure 3

Where the number of devices that are declared unserviceable within a considered space and system
leaves an insufficient number of devices functional (as required by the evaluation described in section
6.3) the entire system serving that space shall be declared as non-compliant. Individual devices declared
functional within a non-functional system may be salvaged for re-use and this should be done by a
competent service provider (see section 6.3) to ensure suitable application.

14 TRAINING AND COMPETENCE


Competence and training of both system owners and service providers are critical to the successful and
sustainable implementation of UVGI air disinfection systems. Professional services providers can be
training either by professional institutions or through tertiary or post graduate education. System
owners can be trained by the service provider or can outsource training from a reputable and accredited
training institution. An example responsibility matrix including the minimum training topics to be
covered is presented in Table 3.

27 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

14.1 Training and Responsibility Matrix

Training Required/Completed
UVGI Specification
Task Responsibility Electrical
Principles and Safety Testing Cleaning
Maintenance
& IPC Procurement
Engineering
&
Maintenance
Monitoring &
Reporting
Cleaning
Infection
Control &
Reporting
Technical
Procurement
Table 3

28 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

ADDITIONAL REFERENCES
IUVA Draft Guideline IUVA-G02A-2005 International Ultraviolet Association Guideline for Design and
Installation of UVGI Air Disinfection Systems in New Building Construction Copyright 2005 International
Ultraviolet Association32

Fundamental Factors Affecting Upper-Room Ultraviolet Germicidal Irradiation—Part II. Predicting


Effectiveness33

BIBLIOGRAPHY

1. World Health Organisation. WHO Policy on TB infection control in Health-Care Facilities,


Congregate Settings and Households. Geneva, Switzerland; 2004.

2. World Health Organisation. Infection prevention and control of epidemic- and pandemic-prone
acute respiratory diseases in health care. Geneva, Switzerland; 2007.

3. World Health Organisation. Natural ventilation for infection control in health-care settings.
Geneva, Switzerland; 2009. Available at:
http://www.who.int/entity/water_sanitation_health/publications/natural_ventilation.pdf.
Accessed April 30, 2013.

4. Downes A, Blunt T. Researches on the effect of light upon bacteria and other organisms. Proc R
Soc London. 1877;26:179-184. Available at:
http://rspl.royalsocietypublishing.org/content/26/179-184/488.full.pdf. Accessed May 28, 2014.

5. Gates FL. A study of the bactericidal action of ultra violet light: III. The absorption of ultra violet
light by bacteria. J Gen Physiol. 1930;14:31-42. Available at:
http://jgp.rupress.org/content/14/1/31.full.pdf.

6. Riley R. Guidelines for the Application of Upper-Room Ultraviolet Germicidal Irradiation for
Preventing Transmission of Airborne Contagion — Part I : Basic Principles. 1999.

7. Wells WF, Fair MG. Viability of B. coli exposed to ultra-violet radiation in air. Science (80- ).
1935;82:280-1.

8. Riley RL, Permutt S. Room air disinfection by ultraviolet irradiation of upper air. Arch Environ Heal
An Int J. 1971;22(2):208-19.

9. Riley RL, Permutt S. Convection, air mixing, and ultraviolet air disinfection in rooms. Arch Environ
Heal An Int J. 1971;22:200-7.

10. Riley RL, Permutt S, Kaufman JE. Room air disinfection by ultraviolet irradiation of upper air:
further analysis of convective air exchange. Arch Environ Heal An Int J. 1971;23(2):35-39.

11. Escombe AR, Moore DAJ, Gilman RH, et al. Upper-room ultraviolet light and negative air
ionization to prevent tuberculosis transmission. Wilson P, ed. PLoS Med. 2009;6(3):e43.
doi:10.1371/journal.pmed.1000043.

29 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

12. Nardell EA, Bucher SJ, Brickner PW, et al. Safety of upper-room ultraviolet germicidal air
disinfection for room occupants: results from the Tuberculosis Ultraviolet Shelter Study. Public
Health Rep. 2008;123(1):52-60. Available at:
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=2099326&tool=pmcentrez&rendert
ype=abstract.

13. Nardell EA. Environmental control of tuberculosis. Med Clin North Am. 1993;77(6):1315-34.
Available at: http://europepmc.org/abstract/MED/8231415. Accessed June 9, 2014.

14. NIOSH US Department of Health, Education and Welfare PHS. Criteria for a recommended
standard Occupational Exposure to Ultraviolet Radiation.; 1972. doi:(HSM) 73-110009.

15. CIE Technical Division 6. CIE 155: 2003 Ultraviolet Air Disinfection. Vienna, Austria; 2003. doi:ISBN
978 3 901906 25 1.

16. The International Commission on Non-Ionizing Radiation Protection. Guidelines on limits of


exposure to ultraviolet radiation of wavelengths between 180 nm and 400 nm (incoherent optical
radiation).; 2004:171-186. doi:10.1097/00004032-200408000-00006.

17. R.L. Riley, M K, G M. Ultraviolet susceptibility of BCG and virulent tubercle bacilli. Am Rev Respir
Dis. 1976;113:413-18.

18. Mundt E, Nielsen P V. Ventilation Effectiveness.

19. SATS 1706: 2015 SOUTH AFRICAN TECHNICAL STANDARD UVGI Luminaires - Safety and
performance requirements.

20. SATS 1706: 2015 SOUTH AFRICAN TECHNICAL STANDARD UVGI Luminaires - Safety and
performance requirements.

21. Beggs CB, Noakes CJ, Sleigh P a., Fletcher L a., Kerr KG. Methodology for determining the
susceptibility of airborne microorganisms to irradiation by an upper-room UVGI system. J Aerosol
Sci. 2006;37(7):885-902. doi:10.1016/j.jaerosci.2005.08.002.

22. Illuminating Engineering Society of North America (IESNA) Testing Procedures Committee.
Standard File Format for Electronic Transfer of Photometric Data and Related Information
(ANSI/IESNA LM-63-02). New York; 2002.

23. F.W. Leuschner. UVGI MEASUREMENTS, INSTRUMENTATION & FACILITIES PLAN. Pretoria;
2014:1-17.

24. Zhang J, Levin R, Angelo R, et al. A radiometry protocol for UVGI fixtures using a moving-mirror
type gonioradiometer. J Occup Environ Hyg. 2012;9(3):140-8.
doi:10.1080/15459624.2011.648569.

25. Illuminating Engineering Society of North America (IESNA) Testing Procedures Committee. IES
Guide for Lamp Seasoning (LM-54-91). New York; 1991.

26. First MW, Banahan KF, Dumyahn TS. Performance of ultraviolet germicidal irradiation and
luminaries in long-term service. Leukos. 2007;3:181-188. doi:10.1582/LEUKOS.2007.03.03.005.

30 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

27. Vorlander FJ, Raddin EH. The effect of operating cycles on flourescent lamp performance. Illum
Eng. 1950.

28. South African Bureau of Standards. SANS 50285 : 2010 SOUTH AFRICAN NATIONAL STANDARD
Energy efficiency of electric lamps for household use Measurement methods.; 2010.

29. Miller S.L., Hernandez M., Fennelly K, et al. Environmental Control for Tuberculosis: Basic Upper-
Room Ultraviolet Germicidal Irradiation Guidelines for Healthcare Settings. Cincinnati, OH; 2009.
Available at: http://www.cdc.gov/niosh/nas/rdrp/appendices/chapter6/a6-23.pdf.

30. Xu P, Peccia J, Fabian P, et al. Efficacy of ultraviolet germicidal irradiation of upper-room air in
inactivating airborne bacterial spores and mycobacteria in full-scale studies. 2002;37(2003):405-
419.

31. Ko G, First MW, Burge H a. Influence of relative humidity on particle size and UV sensitivity of
Serratia marcescens and Mycobacterium bovis BCG aerosols. Tuber Lung Dis. 2000;80(4-5):217-
28. doi:10.1054/tuld.2000.0249.

32. Miller S, Hernandez M, Fennelly K. Efficacy of ultraviolet irradiation in controlling the spread of
tuberculosis. Atlanta Centers Dis …. 2002. Available at:
http://scholar.google.com/scholar?hl=en&btnG=Search&q=intitle:Efficacy+of+ultraviolet+irradiat
ion+in+controlling+the+spread+of+tuberculosis#0. Accessed August 15, 2014.

33. Mphahlele M, Dharmadhikari AS, Jensen PA, et al. Institutional Tuberculosis Transmission:
Controlled Trial of Upper Room Ultraviolet Air Disinfection - A Basis for New Dosing Guidelines.
Am J Respir Crit Care Med. 2015:150504122726005. doi:10.1164/rccm.201501-0060OC.

34. Noakes CJ, Beggs CB, Sleigh PA, Kerr KG. Effect of Room Mixing and Ventilation Strategy on the
Performance of Upper Room Ultraviolet Germicidal Irradiation Systems. In: Proceedings of
ASHRAE IAQ 2004.; 2004:1-13.

35. First M, Rudnick SN, Banahan KF, Vincent RL, Brickner PW. Fundamental factors affecting upper-
room ultraviolet germicidal irradiation - part I. Experimental. J Occup Environ Hyg. 2007;4(5):321-
31. doi:10.1080/15459620701271693.

36. ASHRAE. ASHRAE 55-2013; Thermal environmental conditions for human occupancy. Am Soc
Heating, Refrig Air Cond. 2013.

37. Coker I, Nardell E, Fourie B. Guidelines for the utilisation of ultraviolet germicidal irradiation
(UVGI) technology in controlling the transmission of tuberculosis in health care facilities in South.;
2001. Available at: http://www.sahealthinfo.com/tb/guidelines.pdf. Accessed June 30, 2014.

38. Kowalski W, Bahnfleth W. Proposed standards and guidelines for UVGI air disinfection. IUVA
News. 2004;6(1):20-25. Available at: http://www.aerobiologicalengineering.com/stds.pdf.

39. Rudnick SN, First MW. Fundamental factors affecting upper-room ultraviolet germicidal
irradiation - part II. Predicting effectiveness. J Occup Environ Hyg. 2007;4(5):352-62.
doi:10.1080/15459620701298167.

31 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

APPENDIX A BASIC PRINCIPLES OF UVGI RADIOMETRY


1) Flux: Is the rate of flow of a physical property such as energy or mass.
2) Solid Angle: is the analogue of a dimensionless three dimensional angle formed at the intersection
of three of more planes or formed at the vertex of a cone.

The unit is: Steradian

The symbol is: Ω

The three dimensional solid angle Ω at a point P is defined as the projected spherical surface area dA on
a sphere with radius R divided by the radius R squared. An area on a sphere equalling the square of its
radius and observed from the centre subtends one steradian.

∴ ω = dA / R2 in steradian or sr.

∴ For a sphere the solid angle ω = 4πR2 / R2 = 4π steradian

3) Radiant Flux or Flux: Radiant Flux is the amount/quantity of power in Watts flowing from one area
to another. There is always a source area and a receiving area, with the flux flowing between them.

The symbol is Фe or Ф

The unit is Watt (W)

The radiant flux is also defined as the rate of flow of radiant energy:

Equation Фe = dQe / dt
- 4.1

Radiant Flux is sometimes called Radiation

4) Radiant Intensity or Intensity: Radiant Intensity is the density of


Figure 1 Graphical
radiant flux per solid angle. This quantity is measured in a specific
Representation of
direction and is therefore a vector quantity. 1 Steradian

The symbol is Ie or I

The unit is Watt per steradian (W/sr)

Equation Ie = dФe / dω - 4.2

5) Irradiance: Irradiance is the density of radiant flux per receiving surface area (flux falling onto the
surface area).

The symbol is Ee or E

The unit is Watt per square meter (W/m2)

Equation Ee = dФe / dA - 4.3

Irradiance at a distance R from the source can be calculated from Ee = Ie / R2 - 1.4

This is called the Inverse Square Law – ISL (For a surface facing the source)

32 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

This can be extended to Ee = Ie cosθ/ R2 - 4.5

This is called the Inverse Square and Cosine Law (For a point at an angle θ on a horizontal surface below
the source)

Irradiance is sometimes called Irradiation Density or UV Intensity (I) - not to be confused with Intensity
(I) as defined in 1.3

6) Radiance: Radiance is the density of radiant flux per solid angle per projected surface
area.

The symbol is Le or L

The unit is Watt per steradian per square meter (W/sr/m2)

Equation Le = dIe / dAprojected = d2Фe / dω·dAprojected - 4.6

Radiance is analogous to the “brightness” of the source or surface and is the measure that should ideally
be used to determine eye exposure levels.

7) Radiation Dose or Dose: Irradiation or radiation dose is the radiant energy delivered to an area by a
source.

The symbol is De or D

The unit is Watt·second per square meter (W·s/m2) or Joule per square meter

(J/m2)

Equation D e = Ee · t - 4.7

Also called the Effective Dose Heff

8) Fluence Rate and Fluence: Fluence rate and fluence are used in sophisticated studies, where the
internal surface dose with backscatter is included. These quantities are used correctly most often in
theoretical studies of dose distribution and where photochemistry at the molecular level in tissue is
enhanced as a result of multiple scattering events in tissue.

Fluence rate (spherical irradiance) at a point is the quotient of the radiant flux of all radiation incidents
on the outer surface of an infinitely small sphere centred at the given point, by the area of the
diametrical cross-section of that sphere. This is the “surface dose”-quantity for the irradiation of micro-
organisms

The symbol is Ee,o

The unit is Watt per square meter (W/m2)

Equation Ee,o = dФe,o / dA' - 4.9

Fluence (radiant spherical exposure) at a point is the time integral of the fluence rate E at the given
point over the given duration ∆t. This is the “dose”-quantity for the irradiation of micro-organisms

The symbol is He,o

The unit is Joule per m2 (J/m) or Watt·seconds per m2 (W·s)/m2)

Equation Ee,o = dФe,o / dA' - 4.10

33 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

9) Radiant Efficiency of lamp or fixture (“wall-plug efficiency”):

Radiant Efficiency is defined as the output UV-c radiant flux in Watt divided by the input electrical power
in Watt for either the lamp (+ ballast) or the fixture and can be expressed as a per-unit value or a
percentage.

10) Other units frequently used

Radiant Flux or Flux:

Microwatt (µW) 1 µW = 10-6 W

Irradiance: Microwatt per square centimetre (µW/cm2)

1 µW / cm2 = 10-2 Watt / m2 and 1 mW / cm2 = 10 Watt / m2

Fluence (Dose):

Microwatt·second per square centimetre (µW·s/cm2)

or Microjoule per square centimeter (µJ/cm2)

1 µJ/cm2 = 1 µW·s/cm2 = 10-2 W·s / m2 = 10-2 J / m2 and

1 mJ/cm2 = 1 mW·s/cm2 = 10 W·s / m2 = 10 J / m2

34 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

APPENDIX B UVGI MAINTENANCE RECORD TEMPLATE - EXAMPLE


Facility: Room Name: Year:
Record Monthly Readings
Acceptance
(Pre and Post Cleaning)
Fluence rate Criteria
(µW/cm²) (µ W/cm² Quarter Quarter Quarter Quarter
@ 1m) 1 2 3 4

Device ID: ≥ 30
ABC 14/01/01/01

Device ID: ≥ 30
ABC 14/01/01/02

Device ID: ≥ 20
ABC 14/01/01/03

Lower room: ≤ 6

Serial number of UV-c


meter:
Approved by:
Date:
Lamps replaced/Date:

Comments:

SKETCH-PLAN OF MEASUREMENT LOCATIONS:

ABC 14/01/01/03 @ 1m

ABC 14/01/01/01 @ 1m ABC 14/01/01/02 @ 1m

35 | P a g e
UVGI Disinfection of Room Air: An Evidence Based Guideline for Design, Implementation and
Maintenance

APPENDIX C EXAMPLE SIGNAGE

CAUTION
HIGH INTENSITY ULTRAVIOLET ENERGY:

PROTECT EYES AND SKIN

CAUTION
HIGH INTENSITY ULTRAVIOLET ENERGY:

TURN OFF LAMPS BEFORE ENTERING UPPER


ROOM

DEVICE UNSERVICEABLE
DO NOT RE-USE OR REPAIR

36 | P a g e

You might also like