Guide To NMR Data File Access and Mestrenova (Mnova) Data Processing
Guide To NMR Data File Access and Mestrenova (Mnova) Data Processing
Guide To NMR Data File Access and Mestrenova (Mnova) Data Processing
Guide To NMR Data File Access and MestReNova (MNova) Data Processing
1. Start up MNova: In Windows, go to the start menu and under programs open MNova (or in Mudd 169, in
the Mudd Lab Shortcuts folder). In MacOS, go to the Applications Folder.
2. Access your NMR data:
Ø On a Mac, from the Finder, select Go > Connect to server.
Type the following address, or select from the list of servers:
smb://ChemNMR.ads.carleton.edu/Data.
When prompted, enter your regular Carleton username and password.
Click Connect.
8. Calibrate your spectrum so that your TMS peak is at 0ppm by clicking on the Reference toolbar
button and then selecting the peak you want to be the reference. Finally, type in 0 (or other
relevant value) in the New Shift box in the dialog box that appears. Note: there are two
Reference toolbar buttons. If you slide the cursor over the two buttons, you’ll find that one is “Absolute
Reference” and the other is just “Reference”. You want the latter.
9. Integration: You can have MNova select the peaks for integration
automatically, but it is almost always preferable to select the peaks
yourself. To manually integrate, click on the Integration icon
on the toolbar and select Manual. Then you will be able to
integrate each peak individually by dragging your mouse (left
click and hold while dragging) from left to right across the region you
would like to integrate. You can normalize the integral values to any peak
of your choosing. In order to normalize the integral values, scroll over the
peak of your choosing until it is red, and then Right Click à Edit
Integral. In the dialog box that appears, type the number you would like to
set that peak to in the Normalized box. To see the numerical value of the
integrals look below each integral. To display a list of the integral values
and the regions integrated go to View à Tables à Integrals.
10. (Optional) To identify desired peaks, click on the Peak Picking
icon to obtain an automatic peak picking analysis. Better: if
you would like to choose only certain peaks, choose the Manual
option and drag your mouse from the left side of the desired peak to the
right side in order to mark the peak. Repeat this process for all the peaks you would like to identify.
11. (Optional) You can create inset expansion plots on your spectrum (as shown below), using the
Expansion tool. Click the Expansion icon, then drag across the portion of the spectrum you want
to expand while holding down the left mouse button. When you release the mouse button, the
inset will appear. Left click on the inset and you can Resize the inset spectrum using the boxes in the
corner of the image. Finally, in order to move the spectrum, grab the green box above the middle of the
inset spectrum (red arrow).
2
MestReNova page 3
12. (Optional) To add annotations to your spectrum, use the toolbar at the bottom of the screen to add shapes
or text. To choose which type of annotation you would like, click on the triangle next to the first icon.
13. (Optional) To paste a Table of Parameters on the spectral window, follow the menu to “Parameters”:
14. To Print or Save your spectrum choose File àPrint/Save. You can also convert the spectrum to a PDF:
choose File àExport to PDF. It is strongly recommended that you save your processed spectrum as an
MNova file, so you don’t have start completely over if you want to print again later, or make any changes.
15. Manual Phasing (if required, see step 4): Manual phasing requires two phasing adjustments –
the so-called zero order and first order phase corrections. To manually phase, click on the phase
icon on the toolbar and select Manual Correction. A Phase Correction dialog box appears
(below) and a cursor line (pivot point) appears on the largest peak in the spectrum. If the pivot point is on a
peak near the middle of the spectrum, it is usually best to move the pivot point to a peak on one side of the
spectrum or the other using the Position slide bar in the dialog box.
As noted in the dialog box, apply the zero order correction (PH0) by left clicking (hold click) in the dialog
box and drag up or down to flatten the baseline about the peak that has the pivot point on it. When that
peak looks good, you are ready to apply the first order correction (PH1). To do so, look to a peak at the
other end of the spectrum from the pivot point (do not move the pivot point) and right click (and hold) the
mouse, again, dragging up or down to flatten the baseline around this peak. Once you are happy with your
phasing, click OK in the dialog box.