ELECTRIC PULP TESTING - Old
ELECTRIC PULP TESTING - Old
ELECTRIC PULP TESTING - Old
Describe the test to the patient in a way that will reduce anxiety and will
eliminate a biased response
Isolate the area of teeth to be tested with cotton rolls and saliva ejector and air-
dry all the teeth
Check the electric pulp tester for function and determine that current is passing
through electrode
Apply an electrolyte on the tooth electrode and place it against the dried enamel
of the crowns occlussabuccal or incisalabial surface .It is important to avoid
contacting any restoration in the tooth or the adjacent gingival tissue with
electrolyte or the electrode,this would cause a false response.
Electric current must be complete a circuit from the electrode through the
tooth to the body of the patient then back to the electrode. When gloves
are not used the circuit is completed by dentists finger with electrode .
When using rubber gloves one of the 2 methods must be followed-:
A ground attachment may be clipped onto the patient’s lip
The patient may complete the circuit by placing a finger on the metal
electrode handle .The second gives a advantage of giving patient more
control as the patient can simply lift the finger off the electrode handle
when sensation is felt .This will immediately interrupt the current and
terminate the stimulation.
First the test is performed on vital tooth so as know the normal
stimulus as it differs from person to person a