Bc-3000plus - Operation - Manual
Bc-3000plus - Operation - Manual
Bc-3000plus - Operation - Manual
Operation Manual
Postfach 1367
D-48543 Steinfurt/Germany
Tel.: 02551-20-97
Tel.: 02551-20-98
Tel.: 02551-40-90
Tel.: 02551-70-3888 oder 3838
Fax.:02551-80-883
Fax.:02551-12-98
e-mail: [email protected] website: www.keul.de
Lieferanschrift:
Von-Langen-Weg 10
D-48565 Steinfurt/Germany
Copyright
© 2003-2005 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
For this Operation Manual, the issued Date is 2005-10 (Version: 1.1).
Release, amendment, reproduction, distribution, rent, adaption and translation of this manual
in any manner whatsoever without the written permission of Mindray is strictly forbidden.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for safety, reliability and performance of this product only in the
condition that:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national and
local requirements;
I
the product is used in accordance with the instructions for use.
II
Warranty
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
any Mindray product which has been subjected to misuse, negligence or accident;
any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to Mindray,
the following procedure should be followed:
1. Obtain return authorization: Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must appear on the
outside of the shipping container. Returned shipments will not be accepted if the Mindray
number is not clearly visible. Please provide the model number, serial number, and a brief
description of the reason for return;
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges);
3. Return address: Please send the part(s) or equipment to the address offered by
Customer Service department.
III
Company Contact
Manufacture: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan, Shenzhen, P.R.China,518057
Phone: +86 755 26582479 26582888
Phone: 0049-40-2513175
Fax: 0049-40-255726
IV
Table of Contents
1
Table of Contents
2
Table of Contents
12 Appendices ..............................................................................................A-1
A Index ......................................................................................................A-1
B Specifications .........................................................................................B-1
C Precautions, Limitations and Hazards ...................................................C-1
D Communication ......................................................................................D-1
3
1 Using This Manual
1.1 Introduction
This chapter explains how to use your BC-3000 Plus operation manual, which is shipped with
your BC-3000 Plus hematology analyzer and contains reference information about the
BC-3000 Plus and procedures for operating, troubleshooting and maintaining the analyzer.
Read this manual carefully before operating your analyzer and operate your analyzer strictly
as instructed in this manual.
1-1
Fehler! Formatvorlage nicht definiert.
This manual contains information written for clinical laboratory professionals to:
1-2
Fehler! Formatvorlage nicht definiert.
This operation manual comprises 11 chapters and 4 appendices. Refer to the table below to
find the information you need.
learn about the intended use and parameters of the BC-3000 Chapter 2 Understanding
Plus Your Analyzer
learn about the hardware and software of the BC-3000 Plus Chapter 2 Understanding
Your Analyzer
learn about how the BC-3000 Plus works Chapter 3 Understanding the
System Principles
learn about how to install the BC-3000 Plus Chapter 4 Installing Your
Analyzer
learn about how to define/adjust system settings Chapter 5 Customizing the
Analyzer Software
learn about how to use the BC-3000 Plus to perform your daily Chapter 6 Operating Your
operating tasks Analyzer
learn about how to review the saved analysis results Chapter 7 Reviewing Sample
Results
learn about how to use the quality control programs Chapter 8 Using the QC
Programs
learn about how to calibrate the BC-3000 Plus Chapter 9 Using the
Calibration Programs
learn about how to maintain/service the BC-3000 Plus Chapter 10 Maintaining Your
Analyzer
learn about the meanings of the error messages and how to Chapter 11 Troubleshooting
correct the problems Your Analyzer
learn about the technical specifications of the BC-3000 Plus Appendix B Specifications
see the summary of all safety messages included in this Appendix C Precautions,
manual Limitations and Hazards
learn about the communication protocol of the BC-3000 Plus Appendix D Communication
1-3
Fehler! Formatvorlage nicht definiert.
This manual uses certain typographical conventions to clarify meaning in the text:
All capital letters enclosed in [ ] indicate a key name (either on the built-in keypad or the
external keyboard), such as [ENTER].
All capital, bold and italic letters indicate a special operation defined in the following
section, such as SELECT.
Bold letters included in “ ” indicate text you can find on the screen, such as “Prepare to
ship”.
Bold letters indicate defined screen areas/fields, such as System Status area, or
chapter titles, such as Chapter 1 Using This Manual.
All illustrations in this manual are provided as examples only. They may not necessarily
reflect your analyzer setup or data displayed.
1-4
Fehler! Formatvorlage nicht definiert.
z This analyzer adopts a fixed decimal point. You can enter the digits without
bothering to look for the [.] on the external keyboard.
1-5
Fehler! Formatvorlage nicht definiert.
1.6 Symbols
You may find the following symbols on the analyzer or the reagents.
EQUIPOTENTIALITY
BIOLOGICAL RISK
HIGH VOLTAGE
ALTERNATING CURRENT
1-6
Fehler! Formatvorlage nicht definiert.
BATCH CODE
USE BY
SERIAL NUMBER
DATE OF MANUFACTURE
TEMPERATURE LIMITATION
IRRITATING SUBSTANCE
1-7
Fehler! Formatvorlage nicht definiert.
1-8
Fehler! Formatvorlage nicht definiert.
5
6
(4)
(1)
(2)
(3)
(1)
Equipotentiality.
(2)
To avoid electric shock, disconnect power cord prior to removing or replacing fuse;
(3)
Biological risk.
(4)
The following definition of the WEEE label applies to EU member states only: The use of this
symbol indicates that this product should not be treated as household waste. By ensuring
that this product is disposed of correctly, you will help prevent bringing potential negative
1-9
Fehler! Formatvorlage nicht definiert.
consequences to the environment and human health. For more detailed information with
regard to returning and recycling this product, please consult the distributor from whom you
purchased the product.
1-10
Fehler! Formatvorlage nicht definiert.
(5)
(5)
To avoid being injured, do not put hand under the motor when the machine is running.
1-11
2 Understanding Your Analyzer
2.1 Introduction
2-1
Fehler! Formatvorlage nicht definiert.
z The purpose of this analyzer is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that
require additional studies.
The analyzer is used for the quantitative determination of the following 19 parameters and 3
histograms of blood samples.
Coefficient of Variation
Red Blood Cell (erythrocyte) Distribution Width RDW-SD
Standard Deviation
Hematocrit HCT
Platelet PLT
Mean Platelet Volume MPV
Platelet Distribution Width PDW*
Plateletcrit PCT*
2-2
Fehler! Formatvorlage nicht definiert.
2-3
Fehler! Formatvorlage nicht definiert.
2-4
Fehler! Formatvorlage nicht definiert.
1
2
3
4
5
6
Figure2-3 Inside front of the analyzer
2-5
Fehler! Formatvorlage nicht definiert.
2
0
1
1
8
2
1
7
3
1
6
1
5
4
1
4
5
1
3
6
1
2
7
1
0
1
1
8
2-6
Fehler! Formatvorlage nicht definiert.
2-7
Fehler! Formatvorlage nicht definiert.
2.3.1 LCD
The LCD is located on the front panel of the analyzer. It displays all alphanumeric and graphic
data.
Aspirate key
The aspirate key is located behind the sample probe, as Figure2-6 shows. You can press the
key to start the selected run cycle or dispense diluent.
2-8
Fehler! Formatvorlage nicht definiert.
Built-in keypad
PS/2 keyboard
The analyzer can also be controlled by an external PS/2 keyboard that should be connected
to the analyzer’s keyboard interface. See the table below for the correspondence between
the keypad keys and the keyboard keys and for their functions.
2-9
Fehler! Formatvorlage nicht definiert.
2.3.3 Recorder
The thermal recorder is located on the front panel. You can use it to print out analysis reports
and other interested information.
2.3.9 Printer(Optional)
An external printer can be connected to the parallel port at the back of the analyzer. You can
use it to print out a detailed report and other interested information.
2.3.10 Scanner(Optional)
A bar-code scanner can be connected to the RS-232 port 1 of the analyzer. You can use it to
scan the bar-coded sample IDs into the analyzer.
2-10
Fehler! Formatvorlage nicht definiert.
2-11
Fehler! Formatvorlage nicht definiert.
Title area
Analysis area
Menu area
Help area
Title Area
The Title area displays the title of the current screen, which, in case of Figure2-8, is “Count”.
Status Area
System Status Area
The System Status area displays whether this analyzer is ready for the next analysis. When
it displays “Ready”, it means this analyzer is ready and you can proceed to analyze the next
sample. When it displays “Waiting”, it means the analyzer is not ready for the next run yet.
When it displays “Running”, it means this analyzer is analyzing a sample.
2-12
Fehler! Formatvorlage nicht definiert.
The Count Mode area displays in which analysis (count) mode, whole blood or prediluted,
the next sample is to be analyzed.
Transmission status
The System Time area displays the system time (in the 24-hour format).
Menu Area
When you press [MENU], this area displays the system menu.
Help Area
The Help area reminds you how to proceed to the next step.
The system menu contains 9 programs. The programs followed by “ ”s have further
sub-menus. See the figure below for the fully expanded menu.
2-13
Fehler! Formatvorlage nicht definiert.
Count
Sample Mode
Gain Child
Auto Clean Time Neonate
Report Title
Parameter Units
Other Setting
Service Maintenance
System Status
Valve Test
System Test
Prepare to Ship
Error Message
2-14
Fehler! Formatvorlage nicht definiert.
2-15
Fehler! Formatvorlage nicht definiert.
Because the analyzer, reagents (diluent, rinse, lyse, probe cleanser and E-Z cleanser),
controls, and calibrators are components of a system, performance of the system depends on
the combined integrity of all components. You should only use the Mindray-specified reagents
(see Appendix B Specifications), which are formulated specifically for the fluidic system of
your analyzer in order to provide optimal system performance. Do not use the analyzer with
reagents from multiple suppliers. In such use, the analyzer may not meet the performance
specified in this manual and may provide unreliable results. All references related to reagents
in this manual refer to the reagents specifically formulated for this analyzer.
Each reagent package must be examined before use. Inspect the package for signs of
leakage or moisture. Product integrity may be compromised in packages that have been
damaged. If there is evidence of leakage or improper handling, do not use the reagent.
z Store and use the reagents as instructed by instructions for use of the
reagents.
z When you have changed the diluent, rinse or lyse, run a background to see
if the results meet the requirement.
z Pay attention to the expiration dates and open-container stability days of all
the reagents. Be sure not to use expired reagents.
z After installing a new container of reagents, keep the reagents still for a
while before using them.
2.5.1 Diluent
The diluent is formulated to meet the following requirements:
To provide the blood cells with an environment similar to the blood plasma;
To maintain the cell volume of each red blood cell and platelet during the count and
sizing portion of the measurement cycle;
To provide a conductive medium for impedance counting of white and red blood cells
and platelets.
2-16
Fehler! Formatvorlage nicht definiert.
2.5.2 Lyse
The lyse is formulated to meet the following requirements:
To rapidly break down red blood cell walls, release the hemoglobin from the cell, and
reduce the size of cellular debris to a level that does not interfere with white blood cell
counting.
2.5.3 Rinse
The rinse is formulated to rinse the baths and metering tubes and to provide proper meniscus
formation in the metering tubes and maintain it during each measurement cycle.
The controls are commercially prepared whole-blood products used to verify that the analyzer
is functioning properly. They are available in low, normal, and high levels. Daily use of all
levels verifies the operation of the analyzer and ensures reliable results are obtained. The
calibrators are commercially prepared whole-blood products used to calibrate the analyzer.
Read and follow the instructions for use to use the controls and calibrators. All references
related to controls and calibrators in this manual refer to the controls and calibrators reagents
specifically formulated for this analyzer. You should buy those controls and calibrators from
Mindrayor Mindray-authorized distributors.
2-17
3 Understanding the System
Principles
3.1 Introduction
the Coulter method for determining the WBC, RBC, and PLT data;
During each analysis cycle, the sample is aspirated, diluted and mixed before the
determination for each parameter is performed.
3-1
Fehler! Formatvorlage nicht definiert.
3.2 Aspiration
This analyzer can process two types of blood samples – whole blood samples and prediluted
blood samples.
If you are to analyze a whole blood sample, you can simply present the sample to the sample
probe and press the aspirate key to aspirate 13μL of the sample into the analyzer.
If you are to analyze a capillary blood sample, you should first manually dilute the sample (20
μL of capillary sample needs to be diluted by 0.7mL of diluent) and then present the
pre-diluted sample to the sample probe and press the aspirate key to aspirate 0.3ml of the
sample into the analyzer.
3-2
Fehler! Formatvorlage nicht definiert.
3.3 Dilution
Usually in blood samples, the cells are too close to each other to be identified or counted. For
this reason, the diluent is used to separate the cells so that they are drawn through the
aperture one at a time as well as to create a conductive environment for cell counting.
Moreover, red blood cells usually outnumber white blood cells by 1,000 times. For this reason,
lyse needs to be added to the sample to eliminate the red cells before the WBC counting.
When analyzing a whole blood sample, this analyzer aspirates 13μL of the sample and
follows the procedure presented in Figure3-1 to dilute it before proceeding to the actual
analysis.
3.5mL of diluent
15.6uL
About 1:269 dilution
3-3
Fehler! Formatvorlage nicht definiert.
When analyzing a prediluted sample, you should first collect 20μL of capillary sample and
dispense 0.7mL of diluent from this analyzer to predilute it. Then the analyzer aspirates 0.3ml
of the pre-diluted sample for further dilution, as Figure3-2 shows.
0.7mL of diluent
1:36 dilution
0.3mL
2.9 mL of diluent
24.8uL
About 1:384 dilution
3-4
Fehler! Formatvorlage nicht definiert.
The metering unit controlling the WBC count cycle consists of a metering tube with two
optical sensors mounted on it. This tube ensures that a precise amount of diluted sample is
measured during each count cycle. The exact amount is determined by the distance between
the two optical sensors. The rinse is used to create a meniscus in the metering tube. The
count cycle starts when the meniscus reaches the upper sensor and stops when the
meniscus reaches the lower sensor. The amount of time required for the meniscus to travel
from the upper sensor to the lower sensor is called the WBC Count Time and is measured in
seconds. At the end of the count cycle, the measured count time is compared to the
pre-defined reference count time (see Chapter 5.3 for details). If the former is less than or
greater than the latter by 2 seconds or more, the analyzer will report a WBC bubble or WBC
Clog error. Seeing the error message, you can refer to Chapter 11 Troubleshooting Your
Analyzer for solutions.
3-5
Fehler! Formatvorlage nicht definiert.
WBC measurement
WBCs are counted and sized by the Coulter method. This method is based on the
measurement of changes in electrical resistance produced by a particle, which in this case is
a blood cell, suspended in a conductive diluent as it passes through an aperture of known
dimensions. An electrode is submerged in the liquid on both sides of the aperture to create an
electrical pathway. As each particle passes through the aperture, a transitory change in the
resistance between the electrodes is produced. This change produces a measurable
electrical pulse. The number of pulses generated signals the number of particles that passed
through the aperture. The amplitude of each pulse is proportional to the volume of each
particle. Each pulse is amplified and compared to the internal reference voltage channels,
which only accepts the pulses of a certain amplitude. If the pulse generated is above the
WBC threshold, it is counted as a WBC.
Figure3-4
HGB measurement
HGB is determined by the colorimetric method. The WBC/HGB dilution is delivered to the
WBC bath where it is bubble mixed with a certain amount of lyse, which converts hemoglobin
to a hemoglobin complex that is measurable at 525 nm. An LED is mounted on one side of
the bath and emits a beam of light, which passes through the sample and a 525nm filter, and
then is measured by a photo-sensor that is mounted on the opposite side. The signal is then
amplified and the voltage is measured and compared to the blank reference reading
3-6
Fehler! Formatvorlage nicht definiert.
(readings taken when there is only diluent in the bath). The HGB is calculated per the
following equation and expressed in g/L.
WBC
WBC (109/ L) is the number of leukocytes measured directly by counting the white blood cells
passing through the aperture. Note that when you observe NRBCs (nucleated red blood
cells), which do not react with the lyse and can be mistaken by the analyzer for white cells, in
the microscope, be sure to correct the system-generated result by the following formula,
100
WBC'=WBC ×
100+NRBC
where WBC represents the system-generated white cell number, NRBC the number of
NRBCs counted in 100 white cells and WBC′ the corrected white cell number.
WBC differentia
With the help of the diluent and lyse, this analyzer can size the white cells into three
sub-populations - lymphocytes, mid-sized cells (including monocytes, basophils and
eosinophils ) and granulocytes. Based on the WBC histogram, this analyzer calculates
Lymph%, Mid% and Gran% as follows and express the results in percents.
PL
Lymph% = × 100
PL + PM + PG
PM
Mid% = × 100
PL + PM + PG
PG
Gran% = × 100
PL + PM + PG
Having achieved the three parameters above, this analyzer proceeds to calculate the
3-7
Fehler! Formatvorlage nicht definiert.
Lymph# , Mid# and Gran# per the following equations and express them in 10 9 / L .
Lymph% × WBC
Lymph# =
100
Mid % × WBC
Mid # =
100
Gran % × WBC
Gran # =
100
WBC histogram
Besides the parameters mentioned above, this analyzer also presents a WBC histogram,
whose x-coordinate represents the cell volume(fL)and y-coordinate represents the number
of the cells. The histogram is presented in the Analysis area of the “Count” screen when the
analysis is done. You can also review the histograms of the stored patient results (see
Chapter 7 Reviewing Sample Results).
The first three discriminators of the WBC histogram can be adjusted in case you are not
satisfied with the result. Note that you cannot adjust them if the WBC result is less than 0.5
or out of the operating range.
3.4.4 HGB
Using the colorimetric method, this analyzer calculates hemoglobin concentration (g/L) as
follows.
3-8
Fehler! Formatvorlage nicht definiert.
The metering unit controlling the RBC/PLT count cycle consists of a metering tube with two
optical sensors mounted on it. This tube ensures that a precise amount of diluted sample is
measured during each count cycle. The exact amount is determined by the distance between
the two optical sensors. The rinse is used to create a meniscus in the metering tube. The
count cycle starts when the meniscus reaches the upper sensor and stops when the
meniscus reaches the lower sensor. The amount of time required for the meniscus to travel
from the upper sensor to the lower sensor is called the RBC Count Time and is measured in
seconds. At the end of the count cycle, the measured count time is compared to the
pre-defined reference count time (see Chapter 5.3 for details). If the former is less than or
greater than the latter by 2 seconds or more, the analyzer will report a RBC bubble or RBC
Clog error. Seeing the error message, refer to Chapter 11 Troubleshooting for solutions.
3-9
Fehler! Formatvorlage nicht definiert.
3-10
Fehler! Formatvorlage nicht definiert.
RBCs/PLTs are counted and sized by the Coulter method. This method is based on the
measurement of changes in electrical resistance produced by a particle, which in this case is
a blood cell, suspended in a conductive diluent as it passes through an aperture of known
dimensions. An electrode is submerged in the liquid on both sides of the aperture to create an
electrical pathway. As each particle passes through the aperture, a transitory change in the
resistance between the electrodes is produced. This change produces a measurable
electrical pulse. The number of pulses generated signals the number of particles that passed
through the aperture. The amplitude of each pulse is proportional to the volume of each
particle. Each pulse is amplified and compared to the internal reference voltage channels,
which only accepts the pulses of a certain amplitude. If the pulse generated is above the
RBC/PLT lower threshold, it is counted as a RBC/PLT.
Figure3-6
RBC
RBC (1012/L) is the number of erythrocytes measured directly by counting the erythrocytes
3-11
Fehler! Formatvorlage nicht definiert.
3-12
Fehler! Formatvorlage nicht definiert.
MCV
Based on the RBC histogram, this analyzer calculates the mean cell volume (MCV) and
expresses the result in fL .
This analyzer calculates the HCT (%), MCH (pg) and MCHC (g/L) as follows:
This analyzer calculates the HCT (%), MCH (pg) and MCHC(g/L) as follows:
RBC × MCV
HCT =
10
HGB
MCH =
RBC
HGB
MCHC = × 100
HCT
RDW-CV
Based on the RBC histogram, this analyzer calculates the CV (Coefficient of Variation) of the
erythrocyte distribution width.
RDW-SD
RDW-SD (RBC Distribution Width – Standard Deviation, fL) is set on the 20% frequency level
with the peak taken as 100%, Figure3-7 shows.
3-13
Fehler! Formatvorlage nicht definiert.
Figure3-7
RBC Histogram
Besides the parameters mentioned above, this analyzer also presents a RBC histogram,
whose x-coordinate represents the cell volume(fL)and y-coordinate represents the number
of the cells. The histogram is presented in the Analysis area of the “Count” screen when the
analysis is done. You can also review the histograms of the stored patient results (see
Chapter 7 Reviewing Sample Results).
The two discriminators of the RBC histogram can be adjusted in case you are not satisfied
with the result. Note that you cannot adjust them if the RBC result is less than 0.2 or out of the
operating range.
PLT
PLT (109/L) is measured directly by counting the platelets passing through the aperture.。
MPV
Based on the PLT histogram, this analyzer calculates the mean platelet volume (MPV, fL).
PDW
Platelet distribution width (PDW) is the geometric standard deviation (GSD) of the platelet
size distribution. Each PDW result is derived from the platelet histogram data and is reported
as 10(GSD).
PCT
PLT × MPV
PCT = 3-14
10000
Fehler! Formatvorlage nicht definiert.
PLT Histogram
Besides the parameters mentioned above, this analyzer also presents a PLT histogram,
whose x-coordinate represents the cell volume(fL)and y-coordinate represents the number
of the cells. The histogram is presented in the Analysis area of the “Count” screen when the
analysis is done. You can also review the histograms of the stored patient results (see
Chapter 7 Reviewing Sample Results).
The two discriminators of the PLT histogram can be adjusted in case you are not satisfied
with the result. Note that you cannot adjust them if the PLT result is less than 10 or out of the
operating range.
3-15
Fehler! Formatvorlage nicht definiert.
3.6 Wash
3-16
4 Installing Your Analyzer
4.1 Introduction
This chapter introduces the installation procedure of the BC-3000 Plus. To ensure all system
components are functioning correctly and to verify system performance, a
Mindray-authorized representative will handle the installation and initial software setup.
4-1
Fehler! Formatvorlage nicht definiert.
Before installation, you should ensure that the following space, power and environmental
requirements are met.
enough room on or below the countertop to accommodate the diluent, rinse and waste
containers.
a female receptacle;
50/60 Hz.
z Before turning on the analyzer, make sure the input voltage meets the above
requirements.
Operating temperature: 15 ℃ - 35 ℃;
4-2
Fehler! Formatvorlage nicht definiert.
The environment should be as free as possible from dust, mechanical vibrations, loud
noises, and electrical interference;
Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close;
Do not place the analyzer in direct sunlight or in front of a source of heat or drafts.
4-3
Fehler! Formatvorlage nicht definiert.
4.3 Unpacking
1. Place the carton on the floor upright with the arrows on the side upwards;
2. Remove the tape and take out the accessory box. Check the accessories against the
packing list. Notify the Mindray customer service department or your local distributor
immediately if you find anything missing;
3. Open the main box and check the items inside against the packing list. Notify the Mindray
customer service department or your local distributor immediately if you find anything
missing;
4. Remove the top protective foam. Firmly grip the two cardboard handles and lift the
analyzer out of the box and place it on the floor. Carry the analyzer away from the foam
and set it on the countertop.
z Be sure to retain the shipping carton and all the packing materials, as they
can be used for packaging if analyzer must be reshipped.
If your analyzer has been used for a while, do the”Empty Tubing” procedure and shut it
down before moving it;
4-4
Fehler! Formatvorlage nicht definiert.
For a short - distance moving on a smooth ground, you may use a trolley to facilitate the
transportation;
During the moving process, be sure to protect the LCD and the sample probe from
excessive force and from contact with other objects;
Be sure to keep the analyzer upright during the moving process. Do not tilt or incline it;
Do your best to minimize the mechanical shock when moving the analyzer. After a
long-distance moving, check and tune the analyzer before using it.
4-5
Fehler! Formatvorlage nicht definiert.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
Before the analyzer is shipped out, the sample probe is fixed by a plastic cable tie to avoid
damage during shipment. After unpacking the analyzers, you need to release the sample
probe as follows:
1. Push the right door latch in the direction indicated in Figure 4-1 to open the right door;
Figure 4-1
4-6
Fehler! Formatvorlage nicht definiert.
2. Lift up the front panel latch as indicated in Figure 4-2 and open the front panel;
Figure 4-2
3. Cut the plastic cable tie to release the probe, as Figure 4-3 shows;
Figure 4-3
4-7
Fehler! Formatvorlage nicht definiert.
Figure 4-4
5. Lift the front panel latch and close the front panel and then release the latch to lock it.
Finally close the right door.
z If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
4-8
Fehler! Formatvorlage nicht definiert.
z Be sure to keep the reagents still for a while before using them.
1. Push the left door latch in the direction indicated in Figure 4-5 to open the left door;
Figure 4-5
Figure 4-6
4-9
Fehler! Formatvorlage nicht definiert.
3. Take out the lyse pickup tube with an orange connector (see Figure4-7) from the
accessory box. Take out the lyse container, in which there should be enough lyse.
Remove the container cap and insert the double-pronged end of the tube into the
container and turn (clockwise) the tube’s cap onto the lyse container until properly
secured;
Figure 4-7
4. Place the lyse container onto the shelf and connect the black connector on the cap to the
black fitting and the orange connector to the orange fitting, as Figure 4-8 shows.
Figure4-8
4-10
Fehler! Formatvorlage nicht definiert.
1. Take out the diluent pickup tube with a green connector (Figure 4-9) from the accessory
box;
2. Take out the diluent container, in which there should be enough diluent, and place it on or
below the countertop;
3. Remove the container cap and insert the double-pronged end of the tube into the diluent
container and turn the tube’s cap onto the diluent container until properly secured;
Figure4-9
4. Locate the green fitting, marked DILUENT, in the lower right corner of the back of the
analyzer. Plug the green connector of the tube into the fitting and turn it clockwise until
properly secured;
5. Locate the transducer fitting, marked DILUENT, in the lower right corner of the back of
the analyzer. Connect the wire by pushing in and turning until properly secured.
1. Take out the diluent pickup tube with a blue connector (see Figure 4-10) from the
accessory box;
2. Take out the rinse container, in which there should be enough diluent, and place it on or
below the countertop;
3. Remove the container cap and insert the double-pronged end of the tube into the rinse
container and turn clockwise the tube’s cap onto the diluent container until properly
secured;
4-11
Fehler! Formatvorlage nicht definiert.
Figure 4-10
4. Locate the green fitting, marked RINSE, in the lower right corner of the back of the
analyzer. Plug the blue connector of the tube into the fitting and turn it clockwise until
properly secured;
5. Locate the transducer fitting, marked RINSE, in the lower right corner of the back of the
analyzer. Connect the wire by pushing in and turning until properly secured.
1. Take out the waste tube with a red connector from the accessory box;
2. Locate the red fitting, marked WASTE, in the lower right corner of the back of the
analyzer. Plug the red connector of the tube into the fitting and turn it clockwise until
properly secured;
3. Prepare a container to receive the waste and place it on or below the countertop;
4-12
Fehler! Formatvorlage nicht definiert.
z Improper installation of recorder paper may jam the paper and/or result in
blank printouts.
1. Locate the projecting part in the upper right corner of the recorder and press it in the
direction shown in Figure 4-11 to open it;
Figure 4-11
2. Flip the paper tension lever on the left side upwards. Keep the printing side face-down.
Insert the pointed end of the paper into the slot below the paper rod and push the paper
until it comes out from above the rod. Pull the paper out. Keep the paper centered and
place the paper into the paper holder. See Figure 4-12;
4-13
Fehler! Formatvorlage nicht definiert.
Figure 4-12
z The recorder paper is treated on one side for printing. To determine which
side is the printing side, gently scratch both sides with nails and the one
with visible nail trace left is the printing side.
3. Flip the paper tension level downwards to lock the paper in place, as Figure 4-13 shows.
Figure 4-13
4-14
Fehler! Formatvorlage nicht definiert.
Figure 4-14
4-15
Fehler! Formatvorlage nicht definiert.
Take out the power cord from the accessory box. Plug the non-pronged end into the AC input
at the back of the analyzer and the pronged end into an electrical outlet. Place the power
switch at the back of the analyzer in the ON position (1) to turn on the analyzer. The power
indicator light will be illuminated and the screen will display “Initializing…“. The analyzer will
sequentially initialize the file, hardware and fluidic systems and the whole initializing process
lasts about 3 - 4 minutes. When the initialization is finished, the analyzer will automatically
enter the “Count” screen.
4-16
5 Customizing the Analyzer Software
5.1 Introduction
The BC-3000 Plus is a flexible laboratory instrument that can be tailed to your work
environment. You can use the “Setup”program to customize the software options as
introduced in chapters 5.2 - 5.13.
5-1
Fehler! Formatvorlage nicht definiert.
5.2 Print
SELECT “Setup → Print” (Figure 5-1) to enter the “Print” screen(Figure 5-2).
5-2
Fehler! Formatvorlage nicht definiert.
5-3
Fehler! Formatvorlage nicht definiert.
If you prefer the recorder, SELECT “Recorder” from the “Device” pull-down list.
If you prefer the printer, SELECT “Printer” from the “Device” pull-down list.
To choose the desired format, SELECT the desired format from the “Print Format”
pull-down list, as Figure 5-4 shows.
5-4
Fehler! Formatvorlage nicht definiert.
If you have selected the recorder, you can choose any of the following 4 printing formats.
To choose the desired format, SELECT the desired format from the “Print Format”
pull-down list, as Figure 5-5 shows.
5-5
Fehler! Formatvorlage nicht definiert.
5-6
Fehler! Formatvorlage nicht definiert.
The “Count Time” screen is where you view and/or set (if you have the administrator
password) the reference time for the WBC and/or RBC count portion of the measurement
cycle. If the actual WBC or RBC count time (see Chapter 3.4.1 and 3.5.1) deviates from the
reference time by 2 seconds or more, the analyzer will alarm you for clogging or bubbles and
invalidate the results of all related parameters.
settings
Press [MENU] to enter the system menu.
SELECT “Setup → Count Time”(Figure 5-7)to enter the “Count Time” screen (Figure 5-8).
5-7
Fehler! Formatvorlage nicht definiert.
3. ENTER the desired number into the ”WBC Count Time” box or ”RBC Count Time” box
to set the reference WBC or RBC count time.
5-8
Fehler! Formatvorlage nicht definiert.
5.4 Password
The BC-3000 Plus classifies users into two categories: common users (default) and
administrators. You need to enter the administrator password to adjust certain options such
as WBC/RBC Count Time, Gain, etc.
SELECT “Setup → Password” (Figure 5-9) to enter the ”Password” screen (Figure 5-10).
5-9
Fehler! Formatvorlage nicht definiert.
ENTER “3000” and a message box will pop up to remind you of the current password level,
as Figure 5-11 shows.
CLICK “Enter” to confirm the password and exit to the system menu.
Enter the “Password” screen and the default password is the common user password.
Press [MENU] again and a message box will pop up to remind you of the current password
level, as Figure 5-12 shows.
CLICK “Enter” to confirm the password and exit to the system menu.
5-10
Fehler! Formatvorlage nicht definiert.
Man Male
> 12 years
Woman Female
> 12 years
5-11
Fehler! Formatvorlage nicht definiert.
5-12
Fehler! Formatvorlage nicht definiert.
group)
Press [MENU] to exit to the system menu; press [MAIN] to return to the “Count” screen. If
you have made any changes, a message box will pop up to ask you to save the changes, as
Figure 5-16 shows. CLICK “Enter” to save the changes and exit to the system menu or the
main screen; CLICK “Cancel” to abort the changes and exit to the system menu or the
“Count” screen.
5-13
Fehler! Formatvorlage nicht definiert.
z At the “General” screen (or the screen of any other group), you can press
[PRINT] to print out the displayed limits.
z At the “General” screen (or the screen of any other group), you can press
[DEL] to resume the manufacturer-recommended settings.
5-14
Fehler! Formatvorlage nicht definiert.
5.5 Transmission
The “Transmission” screen is where you set communication parameters.
5-18).
5-15
Fehler! Formatvorlage nicht definiert.
5-16
Fehler! Formatvorlage nicht definiert.
5-17
Fehler! Formatvorlage nicht definiert.
5-18
Fehler! Formatvorlage nicht definiert.
5-19
Fehler! Formatvorlage nicht definiert.
The “Date & Time” screen is where you set the system date and time.
SELECT “Setup→Date & Time” (Figure 5-23) to enter the “Date & Time” screen (Figure
5-24).
5-20
Fehler! Formatvorlage nicht definiert.
5-21
Fehler! Formatvorlage nicht definiert.
5-22
Fehler! Formatvorlage nicht definiert.
5.7 Gain
The ”Gain” screen is where you view and/or set (if you have the administrator password)
gains of the “WBC (Whole Blood)”, “WBC (Predilute)”, “RBC” and “HGB” gains.
SELECT “Setup→Gain”(Figure 5-26) shows to enter the ”Gain” screen (Figure 5-27).
5-23
Fehler! Formatvorlage nicht definiert.
5-24
Fehler! Formatvorlage nicht definiert.
When WBC histograms of most samples are similar to Figure 5-28, it implies too small a
WBC gain and you need to increase the gain appropriately.
When WBC histograms of most samples are similar to Figure 5-29, it implies too large a
WBC gain and you need to decrease the gain appropriately.
2. Enter the ”Gain” screen and ENTER the desired gain into the “WBC (Whole) ”, as Figure
5-30 shows, or “WBC (Predilute)”, as Figure 5-31 shows.
5-25
Fehler! Formatvorlage nicht definiert.
Assume the expected MCV result is 90.0fL and the obtained MCV result is 82.0fL.
5-26
Fehler! Formatvorlage nicht definiert.
Then
ExpectedMCV 90.0
× 100%= × 100% = 109.8%
ActualMCV 82.0
2. Enter the “Gain” screen and ENTER a number into the ”RBC” box, as Figure 5-32 shows,
so that RBC “Factor” is as close to 109.8% as possible.
5-27
Fehler! Formatvorlage nicht definiert.
CLICK “Enter” to save the changes and exit to the system menu; CLICK “Cancel” to exit to
the system menu or the “Count” screen without saving the changes.
5-28
Fehler! Formatvorlage nicht definiert.
The “Auto Clean Time” screen is where you set the interval for auto cleaning of the fluidic
lines. The valid interval is 2 - 24 hours and the default interval is 4 hours. To set the interval,
Press [MENU] to enter the system menu.
SELECT “Setup→Auto Clean Time” (Figure 5-35) to enter the “Auto Clean Time” screen
(Figure 5-36).
ENTER the desired interval. Press [MENU] to exit to the system menu or [MAIN] to exit to the
5-29
Fehler! Formatvorlage nicht definiert.
The ”Reagent Exp. Date” screen is where you set expiration dates for the diluent, rinse and
lyse. The analyzer will alarm you for expired reagents when the system time exceeds any of
the three expiration dates.
SELECT ” Setup→ Reagent Exp. Date” (Figure 5-37) to enter the ”Reagent Exp. Date”
5-30
Fehler! Formatvorlage nicht definiert.
5-31
Fehler! Formatvorlage nicht definiert.
z For any reagent, the entered expiration date should be either the expiration
date printed on the labeling or the open-container expiration date, whichever
is earlier.
CLICK “Enter” to save the changes and exit to the system menu or the “Count” screen;
CLICK “Cancel” to abort the changes and exit to the system menu or the “Count”
screen.
5-32
Fehler! Formatvorlage nicht definiert.
The “Report Title” screen is where you set the title of the report to be printed. To set the
report title,
Press [MENU] to enter the system menu.
SELECT “Setup→Report Title”(Figure 5-40) to enter the ”Report Title” screen (Figure 5-41).
5-33
Fehler! Formatvorlage nicht definiert.
ENTER the desired title in the ”Report Title (by recorder) ” or ” Report Title (by printer) ”
box, depending on the printing device you choose to print out the report. Press [MENU] or
[MAIN] to save the changes and exit to the system menu or the system menu or the
“Count” screen.
5-34
Fehler! Formatvorlage nicht definiert.
The “Parameter Units” screen is where you view and/or set (if you have the administrator
password) the reporting units of the parameters.
See Table 5-2 for the available units for every parameter groups. Note that if you choose g/L
or g/dL for the HGB/MCHC group, the MCH unit will automatically change to pg and its
reporting format will be ***.*; if you choose mmol/L for the HGB/MCHC group, the MCH unit
will automatically change to fmol and its reporting format will be **.**.
5-35
Fehler! Formatvorlage nicht definiert.
**.* um3
PDW **.* None Default
PCT .*** % Default
*.** mL/L
settings
Press [MENU] to enter the system menu.
SELECT ”Setup→ Parameter Units” (Figure 5-42), to enter the “Parameter Units” screen
(Figure 5-43 ).
5-36
Fehler! Formatvorlage nicht definiert.
5-37
Fehler! Formatvorlage nicht definiert.
3. SELECT the desired unit from the corresponding pull-down list (e.g. RBC in Figure
5-44).
5-38
Fehler! Formatvorlage nicht definiert.
5.12 Other
5-39
Fehler! Formatvorlage nicht definiert.
5-40
Fehler! Formatvorlage nicht definiert.
5-41
Fehler! Formatvorlage nicht definiert.
To set the display time, ENTER the desired time into the “Alarm time (s)” box.
5-42
Fehler! Formatvorlage nicht definiert.
5-43
Fehler! Formatvorlage nicht definiert.
5-44
6 Operating Your Analyzer
6.1 Introduction
This chapter provides step-by-step procedures for operating your analyzer on a daily basis.
A flow chart indicating the common daily operating process is presented below.
Initial Checks
Power on
No
Run Prediluted Samples
Whole Blood Mode?
Yes
Shutdown
6-1
Fehler! Formatvorlage nicht definiert.
Check and make sure the diluent, rinse and waste tubes are properly connected and not
bent;
Check and make sure the power cord of the analyzer is properly plugged into an
electrical outlet.
Check and make sure enough printer or recorder paper is installed. Check and make sure the
power cord of the printer is properly plugged into an electrical outlet. Check and make sure
the printer cable is properly connected to the analyzer.
Check and make sure the keyboard is properly connected to the keyboard interface (marked
KB) of the analyzer.
6-2
Fehler! Formatvorlage nicht definiert.
6.3 Power-on
Place the power switch at the back of the analyzer in the ON position (1) to turn on the
analyzer. The power indicator light will be illuminated and the screen will display
“Initializing…“.
The analyzer will sequentially initialize the file, hardware and fluidic systems and the whole
initializing process lasts 3 to 5 minutes, depending on how the analyzer was previously shut
down.
If any error occurs during the initialization, the analyzer will display the error messages in the
upper left corner of the screen. You should remove all the errors before running any sample.
See Chapter 11 Troubleshooting Your Analyzer for solutions.
z Running samples with the abnormal background error present will lead to
misleading results.
6-3
Fehler! Formatvorlage nicht definiert.
Before running any samples, run the controls. See Chapter 8 Using the QC Programs for
details.
6-4
Fehler! Formatvorlage nicht definiert.
z For the whole blood samples to be used for WBC differential or PLT count,
you shall store them at the room temperature and run them within 8 hours
after collection.
z If you do not need the PLT, MCV and WBC differential results, you can store
the samples in a refrigerator (2℃ - 8℃) for 24 hours. You need to warm the
z Be sure to mix any sample that has been prepared for a while before running
6-5
Fehler! Formatvorlage nicht definiert.
it.
5. Press [DILUENT] and a message box will pop up to instruct you how to dispense the
diluent into the sample tube, as Figure 6-1shows;
6. Present a clean sample tube to the sample probe and make sure the tube is tilted
towards the probe, as Figure 6-2 shows, to avoid spills and bubbles. Press the aspirate
key to dispense 0.7ml of diluent (the dispensing volume is controlled by the analyzer) into
the tube;
7. When the dispensing is finished, press [ENTER] to close the message box;
8. Add 20μL of capillary blood to the diluent and shake the tube to mix the sample.
6-6
Fehler! Formatvorlage nicht definiert.
z After mixing the capillary sample with the diluent, be sure to wait 3 minutes
before running the sample.
z Be sure to run the prediluted samples within 30 minutes after the mixing.
z Be sure to mix any sample that has been prepared for a while before running
it.
6-7
Fehler! Formatvorlage nicht definiert.
Press [MENU] and SELECT ”Mode” to enter the ”Sample Mode” screen, as Figure 6-4
shows.
Press [MENU] and SELECT ”Count” to enter the ”Count” screen, as Figure 6-5 shows.
6-8
Fehler! Formatvorlage nicht definiert.
z When switching from the predilute mode to the whole blood mode, the
analyzer will automatically wash the fluidic system.
6-9
Fehler! Formatvorlage nicht definiert.
Entering sample ID
ENTER the ID number in the “ID” box. If you have the bar-code scanner installed, you can
simple scan the sample ID into the analyzer.
SELECT the desired item from the “ Gender” pull-down list, as Figure 6-7 shows. Note
that you can select blank in case you are not aware of the patient gender.
6-10
Fehler! Formatvorlage nicht definiert.
This analyzer provides three ways for you to enter the patient age – in years, in months and
in days. The first way is designed for the adult or pediatric patients no younger than one year;
the second for the infant patients one month to one year; the third for the neonatal patients no
older than 28 days. You can choose only one of the three ways to enter the patient age.
To enter the patient age in years:ENTER the desired number, an integer from 0 to 200, into
To enter the patient age in months:ENTER the desired number, an integer from 0 to 12, into
To enter the patient age in days:ENTER the desired number, an integer from 0 to 31, into the
“Days” box.
ENTER the number of the patient’s medical chart into the “Chart No.” box.
ENTER the number of the patient’s bed into the “Bed No.” box.
You can either directly ENTER the name of the department, from which the sample came,
into the “Department” box or SELECT the desired department from the “Department”
pull-down list (if there are previously saved departments in the list, as Figure 6-8 shows).
6-11
Fehler! Formatvorlage nicht definiert.
Figure 6-8 How to select department name from the pull-down list
6-12
Fehler! Formatvorlage nicht definiert.
To enter the name of the person who sent the sample for analysis, ENTER the name into the
“Sender” box or SELECT the desired name from the “Sender” pull-down list (if there are
previously saved names in the list, as Figure 6-9 shows) ; to enter the name of the person
who is to run (or has run) the sample, ENTER the name into the “Tester” box or SELECT the
desired name from the “Tester” pull-down list (if there are previously saved names in the
list) ; to enter the name of the person who is to review the sample results, ENTER the name
into the “Reviewer” box, or SELECT the desired name from the “Reviewer” pull-down list
(if there are previously saved names in the list). All the three pull-down lists are capable of
z To correct an erroneous entry, DELETE the wrong character and ENTER the
correct one.
z After entering all the desired information, you may press [F4] on the external
keyboard to save the changes and exit to the “Count” screen.
“Enter” button
6-13
Fehler! Formatvorlage nicht definiert.
When you have finished entering the all the interested sample information, CLICK the “Enter”
button to save the changes and return to the “Count” screen.
“Cancel” button
If you do not want to save the entered information, CLICK the “Cancel” button to return to
the ”Count” screen without saving the changes.
ID Only
At the ”Count” screen, press [ID] and an ID window will pop up, as Figure 6-10 shows.
ENTER the sample ID into the ID box and press [ENTER] to save the changes and close the
window. If you have the bar-code scanner installed, you can simply scan the sample ID into
the analyzer.
6-14
Fehler! Formatvorlage nicht definiert.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
z Be sure to keep the sample probe tip away from the tube bottom, otherwise
the aspiration volume may be inaccurate.
z When the aspiration is done, remove the sample tube only when the sample
probe is out of the tube.
1. At the “Count” screen, be sure the System Status area displays “Ready“ and Count
Mode area displays “Whole“.
2. Present the mixed sample to the sample probe so that the tip is well into the tube, and
press the aspirate key. The System Status area will display “Running” and the analyzer
will start aspirating sample.
3. When you hear the beep and the sample probe is out of the tube, remove the sample
tube. The sample probe will retract into the analyzer and the analysis progress will be
displayed on the screen.
4. When the analysis is finished, the result will be displayed on the screen and the sample
ID will automatically increase by 1 and the sample probe will be replaced. And if the auto
print function is enabled, the analysis result will be automatically printed out.
6-15
Fehler! Formatvorlage nicht definiert.
Parameter flags
If the analysis result is followed by an ”H” or “L”, it means the analysis result has
exceeded the upper or lower limit of the reference range.
If you see *** as opposed to the result, it means the result is either unreliable or out of the
operating range.
If the WBC result is less than 0.5 × 109/L, this analyzer will not perform the differential
analysis and all the related parameter values will be non-numeric (***).
Histogram flags
The system will flag abnormal histograms.
Abnormal WBC histograms will be flagged by one of the markings: R1,R2,R3,R4 and
Rm.
R1:indicates abnormality on the left side of the lymphocyte hump and possible presence of
platelets coagulate, large platelet, nucleated red cell, insolvable red cell, protein, lipoid debris
in sample, or electrical noise.
R2: indicates abnormality between the lymphocyte hump and the mononuclear area and
possible presence of atypical lymphocyte, original cell in the sample and increased eosinophil
or increased basophil.
R4 :indicates abnormality on the right side of the neutrophilic granulocytes hump and
6-16
Fehler! Formatvorlage nicht definiert.
Abnormal PLT histograms will be flagged by one of the markings: Pm,PS and PL.
Pm:indicates blur demarcation between the platelet and red blood cell area and possible
presence of large platelet, platelet coagulation, small red blood cell, cell debris or fibrin.
z When the PLT value is less than 100 × 109 / L, a manual count by the
microscope is recommended.
Screen saver
This analyzer will enter the screen saver if it has been idle at the “Count” screen for 10
minutes. When it happens, the sample probe will retract into the analyzer, the LCD will turn
dark and the power indicator will be flickering. You can press any key to resume the display
and replace the sample probe.
6-17
Fehler! Formatvorlage nicht definiert.
Press [MENU] and SELECT ”Sample Mode” (Figure 6-11) to enter the ”Sample Mode”
(Figure6-12) screen.
6-18
Fehler! Formatvorlage nicht definiert.
Press [MENU] and SELECT ”Count” to enter the ”Count” screen,as Figure6-13 shows.
You can enter the sample information one of the two ways, ID or All Info., depending on the
configuration of your analyzer (see Chapter 5.13 for how to select the entering mode).
At the “Count” screen, press [ID] and an edit window will pop up, as Figure6-14 shows.
6-19
Fehler! Formatvorlage nicht definiert.
Entering sample ID
ENTER the ID number in the “ID” box, or if you have the bar-code scanner installed, you
can simply scan the sample ID into the analyzer.
SELECT the desired item from the “Gender” pull-down list, as Figure6-15 shows.
Note that you can select blank in case you are not aware of the patient gender.
6-20
Fehler! Formatvorlage nicht definiert.
This analyzer provides three ways for you to enter the patient age –in years, in months and in
days. The first way is designed for the adult or pediatric patients no younger than one year;
the second for the infant patients one month to one year; the third for the neonatal patients no
older than 28 days s. You can choose only one of the three ways to enter the patient age.
To enter the patient age in years:ENTER the desired number, an integer from 0 to 200, into
To enter the patient age in months: ENTER the desired number, an integer from 0 to 12,
To enter the patient age in days:ENTER the desired number, an integer from 0 to 31, into
ENTER the number of the patient’s medical chart into the “Chart No.” box.
ENTER the number of the patient’s bed into the”Bed No. “box.
You can either directly ENTER the name of the department, from which the sample came,
into the “Department” box or SELECT the desired department from the “Department”
pull-down list (if there are previously saved departments in the list, as Figure6-16 shows).
6-21
Fehler! Formatvorlage nicht definiert.
To enter the name of the person who sent the sample for analysis, enter the name into the
“Sender” box or SELECT the desired name from the “Sender” pull-down list (if there are
previously saved names in the list, as Figure6-17 shows) ; to enter the name of the person
who is to run (or has run) the sample, enter the name into the “Tester” box or SELECT the
desired name from the “Tester” pull-down list (if there are previously saved names in the
list) ; to enter the name of the person who is to review the sample results, enter the name into
the “Reviewer” box, or SELECT the desired name from the “Reviewer” pull-down list (if
there are previously saved names in the list). All the three pull-down lists are capable of
6-22
Fehler! Formatvorlage nicht definiert.
6-23
Fehler! Formatvorlage nicht definiert.
“Enter” button
When you have finished entering the all the interested sample information, CLICK the “Enter”
button to save the changes and return to the “Count” screen.
“Cancel” button
If you do not want to save the entered information, CLICK the “Cancel” button to return to
the ”Count” screen without saving the changes.
ID Only
At the ”Count” screen, press [ID] and an ID window will pop up, as Figure6-18 shows.
Figure6-18 ID window
ENTER the sample ID into the ID box and press [ENTER] to save the changes and close the
window.
6-24
Fehler! Formatvorlage nicht definiert.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
z Be sure to keep the sample probe tip away from the tube bottom, otherwise
the aspiration volume may be inaccurate.
z When the aspiration is done, remove the sample tube only when the sample
probe is out of the tube.
1. At the “Count” screen, be sure the System Status area displays “Ready“ and Count
Mode area displays “Predilute“.
2. Present the mixed sample to the sample probe so that the tip is well into the tube, and
press the aspirate key. The System Status area will display “Running” and the analyzer
will start aspirating sample.
3. When you hear the beep and the sample probe is out of the tube, remove the sample
tube. The sample probe will retract into the analyzer and the analysis progress will be
displayed on the screen.
4. When the analysis is finished, the result will be displayed on the screen and the sample
ID will automatically increase by 1 and the sample probe will be replaced. And if the auto
print function is enabled, the analysis result will be automatically printed out.
6-25
Fehler! Formatvorlage nicht definiert.
Parameter flags
If the analysis result is followed by an ”H” or “L”, it means the analysis result has
exceeded the upper or lower limit of the reference range.
If you see *** as opposed to the result, it means the result is either unreliable or out of the
operating range.
If the WBC result is less than 0.5 × 109/L, this analyzer will not perform the differential
analysis and all the related parameter values will be non-numeric (***).
Histogram flags
The system will flag abnormal histograms.
Abnormal WBC histograms will be flagged by one of the markings: R1,R2,R3,R4 and
Rm.
R1:indicates abnormality on the left side of the lymphocyte hump and possible presence of
platelets coagulate, large platelet, nucleated red cell, insolvable red cell, protein, lipoid debris
in sample, or electrical noise.
R2: indicates abnormality between the lymphocyte hump and the mononuclear area and
possible presence of atypical lymphocyte, original cell in the sample and increased eosinophil
or increased basophil.
R4 :indicates abnormality on the right side of the neutrophilic granulocytes hump and
6-26
Fehler! Formatvorlage nicht definiert.
Abnormal PLT histograms will be flagged by one of the markings: Pm,PS and PL.
Pm:indicates blur demarcation between the platelet and red blood cell area and possible
presence of large platelet, platelet coagulation, small red blood cell, cell debris or fibrin.
z When the PLT value is less than 100 × 109 / L, a manual count by the
microscope is recommended.
Screen saver
This analyzer will enter the screen saver if it has been idle at the “Count” screen for 10
minutes. When it happens, the sample probe will retract into the analyzer, the LCD will turn
dark and the power indicator will be flickering. You can press any key to resume the display
and replace the sample probe.
6-27
Fehler! Formatvorlage nicht definiert.
6.8 Shutdown
1. Press [MENU] to enter the system menu and SELECT ”Shutdown”, as Figure 6-19
shows;
2. A message box will pop up to ask you to confirm the shutdown, as Figure 6-20shows;
6-28
Fehler! Formatvorlage nicht definiert.
3. CLICK “Enter”and a window will pop up to instruct you how to shut down the analyzer, as
Figure6-21shows;
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
z The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
4. Present the E-Z cleanser to the sample probe and press the aspirate key. The analyzer
will aspirate the E-Z cleanser and automatically clean the fluidic lines and the baths. The
cleaning progress will be displayed on the screen, as Figure 6-22 shows;
5. When the cleaning is finished, place the switch at the back of the analyzer to OFF (O) to
turn off the analyzer;
6-29
Fehler! Formatvorlage nicht definiert.
6-30
7 Reviewing Sample Results
7.1 Introduction
The analyzer automatically saves analysis results. Totally 35,000 results can be saved. You
can either browse all the saved sample results in general (see “Browsing All Sample Results”)
or search for the results of a particular sample or samples (see “Searching for Interested
Sample Results”).
7-1
Fehler! Formatvorlage nicht definiert.
To browse all the saved sample results, you can choose either of the following modes:
The “Sample Table Review” mode. In this mode, the sample results are presented in a
columnar fashion without histograms (namely you can only see the parameter values). One
screen displays maximum 6 sample results.
The “Sample Histogram Review” mode. In this mode, you can review both parameter
values and histograms of the saved sample results. One screen displays one sample result.
The sample results are sequentially displayed on the screen, the earliest on the utmost left.
The “Location/Total” displayed in the lower right corner of the screen indicates the location
of the current sample result (the one whose “ID” is backlit) and the total number of the saved
sample results.
7-2
Fehler! Formatvorlage nicht definiert.
ENTER the location into the “Location” box and press [ENTER] to jump to the desired
sample result.
7-3
Fehler! Formatvorlage nicht definiert.
Figure7-4 “Find”window
ENTER the sample ID into the “ID” box and press [↑] to search backward or [↓] to search
forward. If the desired sample result is found, the analyzer will jump to it; if not, a message
box will pop up, as Figure7-5 shows. Press [ENTER] to close the message box.
Press [←] or [→] to move the cursor to the interested sample result and press [ENTER] to
select it. The selected sample result will be marked with a “*”, as sample “75” in Figure7-6
shows.
7-4
Fehler! Formatvorlage nicht definiert.
Press [ENTER] again to deselect the sample result. Once the sample is deselected, the “*”
will disappear, as sample “75” in Figure7-7 shows.
Example1:To select the sample results of locations 1 – 5 (sample ID:75, 77, 78, 84, 95 in
7-5
Fehler! Formatvorlage nicht definiert.
4. CLICK “Select” and the lower left corner of the “Select” window will display “Results
selected”, as Figure7-9 shows;
5. CLICK “Quit” to return to the “Sample Table Review” screen. The selected sample
results will be marked with “*”, as Figure7-10 shows.
7-6
Fehler! Formatvorlage nicht definiert.
Example2:To deselect the sample results of locations 1 – 5 (sample ID: 75, 77, 78, 84, 95 in
2. CLICK “Deselect” and the lower left corner of the “Select” window will display “Results
deselected”, as Figure7-11 shows;
3. CLICK “Quit” to return to the “Sample Table” screen. The “*” above those sample results
7-7
Fehler! Formatvorlage nicht definiert.
Example3: To select the sample results of locations 1 to 5 and 7 to 8, follow the procedure
below to do so:
3. CLICK “Quit” to return to the “Sample Table Review” screen. The selected sample
results will be marked with “*”, as Figure7-13 shows.
7-8
Fehler! Formatvorlage nicht definiert.
Example4:To deselect the sample results of locations 1 to 5 and 7 to 8, follow the procedure
below to do so:
3. CLICK “Quit” to return to the “Sample Table Review” screen. The “*” above those
sample results will disappear, as Figure7-14 shows.
7-9
Fehler! Formatvorlage nicht definiert.
To transmit the selected sample results to a host, CLICK “Selected”;to transmit all the
sample results, CLICK “All”;to stop a transmission, CLICK “Stop”; to return to the “Sample
Select the sample results you want to delete and press [DEL]. A message box will pop up to
confirm the deletion, as Figure7-16 shows. CLICK “Enter” to delete the selected results;
CLICK “Cancel” to abort the deletion.
Press [5] and a message box will pop up to ask you to confirm the deletion, as Figure7-17
shows.
CLICK ”Enter” to delete all the sample results; CLICK “Cancel” to abort the deletion.
7-10
Fehler! Formatvorlage nicht definiert.
Calculating reproducibility
This analyzer provides three reproducibility indices Mean,SD(Standard Deviation) and CV%
(Coefficient of Variation).
∑x i
Mean= i=1
n
∑ (X − Mean )
2
SD =
i
n −1
SD
CV% = × 100
Mean
Where n represents how many sample results are selected and Xi is the result of the ith
analysis.
To check the reproducibility of the selected sample results, select at least three sample
results and press [7] to view the reproducibility. If any selected result contains invalid
parameter value (s), the reproducibility indices of that parameter(s) will also be non-numeric
(***).
To print out the displayed indices, press [PRIINT]. To exit the “Reproducibility” screen, press
[MENU] to exit the “Reproducibility”screen.
7-11
Fehler! Formatvorlage nicht definiert.
7-12
Fehler! Formatvorlage nicht definiert.
ENTER the location into the “Location” box and press [ENTER] to jump to the desired
sample result.
ID
7-13
Fehler! Formatvorlage nicht definiert.
SELECT the desired item from the “Gender” pull-down list. Note that you can select blank
in case you are not aware of the patient gender.
This analyzer provides three ways for you to enter the patient age – in years, in months and
in days. The first way is designed for the adult or pediatric patients no younger than one year;
the second for the infant patients one month to one year; the third for the neonatal patients
younger than one month. You can choose only one of the three ways to enter the patient age.
To enter the patient age in years:ENTER the desired number, an integer from 0 to 200, into
To enter the patient age in months:ENTER the desired number, an integer from 0 to 12, into
To enter the patient age in days:ENTER the desired number, an integer from 0 to 31, into the
“Days” box.
ENTER the number of the patient’s medical chart into the “Chart No.” box.
ENTER the number of the patient’s bed into the “Bed No.” box.
You can either directly ENTER the name of the department, from which the sample came,
into the “Department” box or SELECT the desired department from the “Department”
pull-down list (if there are previously saved departments in the list.
To enter the name of the person who sent the sample for analysis, ENTER the name into the
“Sender” box or SELECT the desired name from the “Sender” pull-down list (if there are
previously saved names in the list) ; to enter the name of the person who ran the sample,
7-14
Fehler! Formatvorlage nicht definiert.
ENTER the name into the “Tester” box or SELECT the desired name from the “Tester”
pull-down list (if there are previously saved names in the list) ; to enter the name of the
person who reviewed the sample results, ENTER the name into the “Reviewer” box, or
SELECT the desired name from the “Reviewer” pull-down list (if there are previously
saved names in the list). All the three pull-down lists are capable of saving 30 entered names.
“Enter” button
When you have finished entering the all the interested sample information, CLICK the “Enter”
button (or press [F4] of the external keyboard) to save the changes and return to the “Sample
Histogram Review” screen.
“Cancel” button
If you do not want to save the entered information, CLICK the “Cancel” button to return to
the ”Sample Histogram Review” screen without saving the changes.
Adjusting histograms
If you are not satisfied with the obtained histograms, you can adjust them manually, provided
you have the administrator password.
The first three discriminators of the WBC histogram are adjustable. Note that if the WBC
result is less than 0.5 or non-numeric (***), the WBC histogram is not adjustable. The first two
discriminators of the RBC histogram are adjustable. Note that if the RBC result is less than
0.2 or non-numeric (***), the RBC histogram is not adjustable. The first two discriminators of
the PLT histogram are adjustable. Note that if the PLT result is less than 10 or non-numeric
(***), the PLT histogram is not adjustable.
Example 5:To move the third discriminator of the following WBC histogram to 100fL, follow
1. Press [ENTER] and the discriminator will become adjustable. See Figure7-23;
7-15
Fehler! Formatvorlage nicht definiert.
5. Press [ENTER] and a message box will pop up, as Figure7-26 shows.
7-16
Fehler! Formatvorlage nicht definiert.
CLICK “Enter” to save the changes and return to the “Sample Histogram Review” screen;
CLICK ”Cancel” to abort the changes and return to the “Sample Histogram Review” screen.
7-17
Fehler! Formatvorlage nicht definiert.
To include a search condition, press [↑] or [↓] to move the cursor to the desired condition and
press [ENTER] to tick the condition, as Figure7-28 shows.
SELECT the desired item from the “Gender”pull-down list. Note that you can select blank
7-18
Fehler! Formatvorlage nicht definiert.
You can either directly ENTER the name of the department, from which the sample came,
into the “Department” box or SELECT the desired department from the “Department”
pull-down list (if there are previously saved departments in the list).
Entering sample ID
ENTER the number of the patient’s bed into the “Bed No.” box.
ENTER the number of the patient’s medical chart into the “Chart No.” box.
ENTER the start date into the “Start” box; ENTER the end date into the “End” box.
CLICK “Enter” to start the search. The analyzer will search the saved sample results for
matches and report the conclusion, as Figure7-29 shows. CLICK “Enter” to return to the
“Sample Table Review”screen.
7-19
Fehler! Formatvorlage nicht definiert.
Mode
z For every search, the analyzer can display maximum 500 matches.
z The matches will be deleted if you have run another sample (including
background check), or deleted a sample result, or restarted the analyzer
after the search.
The sample results are sequentially displayed on the screen, the earliest on the utmost left.
The “Location/Total” displayed in the lower right corner of the screen indicates the location
of the current sample result (the one whose “ID” is backlit) and the total number of the sample
results matching the search conditions.
7-20
Fehler! Formatvorlage nicht definiert.
ENTER the location into the “Location” box and press [ENTER] to jump to the desired
sample result.
7-21
Fehler! Formatvorlage nicht definiert.
Press [←] or [→] to move the cursor to the interested sample result and press [ENTER] to
select it. The selected sample result will be marked with a “*”, as sample “75” in Figure7-33
shows.
Press [ENTER] again to deselect the sample result. Once the sample is deselected, the “*”
will disappear, as sample “75” in Figure7-34 shows.
7-22
Fehler! Formatvorlage nicht definiert.
Example 6:To select the sample results of locations 1 – 5 (sample ID: 75, 77, 78, 84, 95, 106
7-23
Fehler! Formatvorlage nicht definiert.
4. CLICK “Select” and the lower left corner of the “Select” window will display “Results
selected”, as Figure7-36 shows;
5. CLICK “Quit” to return to the “Sample Table Review” screen. The selected sample
results will be marked with “*”, as Figure7-37 shows.
7-24
Fehler! Formatvorlage nicht definiert.
Example 7:To deselect the sample results of locations 1 – 5 (sample ID: 75, 77, 78, 84, 95,
2. CLICK “Deselect” and the lower left corner of the “Select” window will display “Results
deselected”, as Figure7-38 shows;
3. CLICK “Quit” to return to the “Sample Table” screen. The “*” above those sample results
will disappear, as Figure7-39 shows.
7-25
Fehler! Formatvorlage nicht definiert.
Example 8: To select the sample results of locations 1 to 3 and 5 to 6, follow the procedure
below to do so:
3. CLICK “Quit” to return to the “Sample Table Review” screen. The selected sample
results will be marked with “*”, as Figure7-40 shows.
Example 9:To deselect the sample results of locations 1 to 3 and5 to 6, follow the procedure
below to do so:
3. CLICK “Quit” to return to the “Sample Table Review” screen. The “*” above those
sample results will disappear, as Figure7-41 shows.
7-26
Fehler! Formatvorlage nicht definiert.
Calculating reproducibility
This analyzer provides three reproducibility indices Mean, SD(Standard Deviation)and CV%
(Coefficient of Variation).
∑x i
Mean= i=1
n
7-27
Fehler! Formatvorlage nicht definiert.
∑ (X − Mean )
2
SD =
i
n −1
SD
CV% = × 100
Mean
Where n represents how many sample results are selected and Xi is the result of the ith
analysis.
To check the reproducibility of the selected sample results, select at least three sample
results and press [7] to view the reproducibility. If any selected result contains invalid
parameter value (s), the reproducibility indices of that parameter(s) will also be invalid (***).
To print out the displayed indices, press [PRIINT]. To exit the “Reproducibility” screen, press
[MENU] to exit the “Reproducibility”screen.
Review” Mode
z For every search, the analyzer can display maximum 500 matches.
z The matches will be deleted if you have run another sample (including
background check), or deleted a sample result, or restarted the analyzer
after the search.
7-28
Fehler! Formatvorlage nicht definiert.
7-29
Fehler! Formatvorlage nicht definiert.
ENTER the location into the “Location” box and press [ENTER] to jump to the desired
sample result.
ID
SELECT the desired item from the “Gender” pull-down list. Note that you can select
blank in case you are not aware of the patient gender.
7-30
Fehler! Formatvorlage nicht definiert.
This analyzer provides three ways for you to enter the patient age –in years, in months and in
days. The first way is designed for the adult or pediatric patients no younger than one year;
the second for the infant patients one month to one year; the third for the neonatal patients no
older than 28 days. You can choose only one of the three ways to enter the patient age.
To enter the patient age in years:ENTER the desired number, an integer from 0 to 200, into
To enter the patient age in months: ENTER the desired number, an integer from 0 to 12,
To enter the patient age in days:ENTER the desired number, an integer from 0 to 31, into the
“Days” box.
ENTER the number of the patient’s medical chart into the “Chart No.” box.
ENTER the number of the patient’s bed into the “Bed No.” box.
You can either directly ENTER the name of the department, from which the sample came,
into the “Department” box or SELECT the desired department from the “Department”
pull-down list (if there are previously saved departments in the list).
To enter the name of the person who sent the sample for analysis, ENTER the name into the
“Sender” box or SELECT the desired name from the “Sender” pull-down list (if there are
previously saved names in the list) ; to enter the name of the person who ran the sample,
ENTER the name into the “Tester” box or SELECT the desired name from the “Tester”
pull-down list (if there are previously saved names in the list) ; to enter the name of the
person who reviewed the sample results, ENTER the name into the “Reviewer” box, or
SELECT the desired name from the “Reviewer” pull-down list (if there are previously
7-31
Fehler! Formatvorlage nicht definiert.
saved names in the list). All the three pull-down lists are capable of saving 30 entered names.
7-32
Fehler! Formatvorlage nicht definiert.
“Enter” button
When you have finished entering all the interested sample information, CLICK the “Enter”
button (or press [F4] of the external keyboard) to save the changes and return to the “Search
Histogram Review” screen.
“Cancel” button
If you do not want to save the entered information, CLICK the “Cancel” button to return to
the ”Search Histogram Review” screen without saving the changes.
Adjusting histograms
If you are not satisfied with the obtained histograms, you can adjust them manually, provided
you have the administrator password.
The first three discriminators of the WBC histogram are adjustable. Note that if the WBC
result is less than 0.5 or non-numeric (***), the WBC histogram is not adjustable. The first two
discriminators of the RBC histogram are adjustable. Note that if the RBC result is less than
0.2 or non-numeric (***), the RBC histogram is not adjustable. The first two discriminators of
the PLT histogram are adjustable. Note that if the PLT result is less than 10 or non-numeric
(***), the PLT histogram is not adjustable.
Example10:To move the third discriminator of the following WBC histogram to 100fL, follow
1. Press [ENTER] and the discriminator will become adjustable. See Figure7-47;
7-33
Fehler! Formatvorlage nicht definiert.
5. Press [ENTER] and a message box will pop up, as Figure7-50 shows.
CLICK “Enter” to save the changes and return to the “Search Histogram Review” screen;
CLICK ”Cancel” to abort the changes and return to the “Search Histogram Review” screen.
7-34
8 Using the QC Programs
8.1 Introduction
Quality Control (QC) consists of strategies and procedures that measure the precision and
stability of the analyzer. The results imply the reliability of the sample results. QC involves
measuring materials with known, stable characteristics at frequent intervals. Analysis of the
results with statistical methods allows the inference that sample results are reliable. Mindray
recommends you run the QC program daily with low, normal and high level controls. A new
lot of controls should be analyzed in parallel with the current lot prior to their expiration dates.
This may be accomplished by running the new lot of controls twice a day for five days using
any empty QC files. The QC files calculate the mean, standard deviation and coefficient of
variation for each selected parameter. The instrument-calculated means of these ten runs
should be within the expected ranges published by the manufacturer.
The BC-3000 Plus provides 4 QC programs: L-J Analysis, X Analysis, X -R Analysis and
X-B Analysis.
8-1
Fehler! Formatvorlage nicht definiert.
Using the “L-J Analysis” program, you can provide quality control for maximum 12
parameters. The analyzer provides 9 QC files for you to save QC settings and results. Every
QC file can save results of maximum 31 QC runs. When the saved QC results have reached
the maximum number, the newest result will overwrite the oldest. The following introduction
will use “File 1” as the example.
SELECT “Quality Control→ L-J Analysis → L-J Edit → File 1” (Figure8-1)to enter the
8-2
Fehler! Formatvorlage nicht definiert.
If there are saved L-J results and settings, you need to delete them first. Press [DEL] and a
message box will pop up to confirm the deletion, as Figure8-3 shows.
CLICK “Enter”to confirm the deletion; CLICK “Cancel”to abort the deletion.
ENTER the lot number of the control to be used into the “Lot No.” box.
ENTER the expiration date of the control to be used into the “Exp. Date” box.
ENTER the expected results (mean) and limits (range) respectively into the “Mean” and
“Range” boxes of the parameters to be included in the L-J analysis.
8-3
Fehler! Formatvorlage nicht definiert.
z Refer to the instructions of use of the control for information on the lot
number, expiration date, open-vial stability days, expected results and limits.
z The entered expiration date should be either the expiration date printed on
the labeling or the open-vial expiration date, whichever is earlier.
z The open-vial expiration date is calculated as follows: the date that vial is
opened + the open-vial stability days.
z At the “L-J Edit” screen, if you want to correct an erroneous entry, MODIFY
the wrong digit.
Deleting settings
Printing settings
Press [MENU] to exit to the system menu; press [MAIN] to exit to the “Count” screen.
A message box shown in Figure8-4 will pop up, if :
There is a parameter for which you have entered only the expected result or the limit;
or
There is a parameter whose expected result is less than or equal to the limit.
CLICK “Enter” to close the box and clear the erroneous entries. Re-enter the correct values
before trying to exit the screen again. The settings can be saved only when both the expected
result and limit are valid.
In case of any invalid entries of expiration dates, a message box will pop up to remind you of
the error, as Figure8-5 shows. CLICK “Enter” to close the box and clear the erroneous
entries. Re-enter the correct values before trying to exit the screen again.
8-4
Fehler! Formatvorlage nicht definiert.
If all the entries are correct, a message box will pop up to remind you to save the changes, as
Figure8-6 shows. CLICK “Enter” to save the changes and exit to the system menu (or the
“Count” screen); CLICK “Cancel” to abort the changes and exit to the system menu (or
the “Count” screen).
Press [MENU] and SELECT “Mode” to enter the “Sample Mode”screen. SELECT “Whole
Blood” from the “Sample Mode” pull-down list.
Press [MENU] to enter the system menu. SELECT “Quality Control→L-J Analysis → L-J
Count →File 1” to enter the “L-J Count”screen, as Figure 8-7 shows.
8-5
Fehler! Formatvorlage nicht definiert.
z Be sure to use the Mindray - specified controls. Using controls other than
the specified will lead to misleading results.
z Refer to the instructions of use of the controls for how to store and use the
controls.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
8-6
Fehler! Formatvorlage nicht definiert.
z When switching from the Predilute mode to the Whole Blood mode, the
analyzer will automatically flush the fluidic system.
z Be sure to keep the sample probe tip away from the tube bottom, otherwise
the aspiration volume may be inaccurate.
z When the aspiration is done, remove the sample tube only when the sample
probe is out of the tube.
1. Be sure the System Status area displays “Ready“ and Count Mode area displays
“Whole“.
2. Present a vial of control to the sample probe so that the tip is well into the vial, and press
the aspirate key. The System Status area will display “Running” and the analyzer will
start aspirating sample.
3. When you hear the beep and the sample probe is out of the vial, remove the vial. The
sample probe will retract into the analyzer and the analysis progress will be displayed on
the screen.
4. When the analysis is finished, the result will be displayed on the screen and the
“NO./Total” in the upper left corner of the screen will automatically increase by 1 and the
sample probe will be replaced.
To browse the result of the preceding or following L-J analysis, press [PgUp] or [PgDn].
8-7
Fehler! Formatvorlage nicht definiert.
To delete the current result, press [DEL] and a message box will pop up, as Figure8-8 shows.
CLCIK “Enter” to confirm the deletion; CLICK “Cancel” to abort the deletion.
Press [MENU] to exit to the system menu, or press [MAIN] to exit to the “Count” screen.
8-8
Fehler! Formatvorlage nicht definiert.
SELECT “Quality Control→ L-J Analysis→ L-J Graph→ File 1”(Figure8-9) to enter the
“L-J Graph” screen (Figure 8-10).
The 12 parameters are displayed on three screens, 4 parameters on every screen, as Figure
8-10 to Figure 8-12 show. The saved QC results are sequentially displayed in the L-J graph,
the latest on the utmost left (No.1).
The x-coordinate represents the number of the L-J analyses performed; the y-coordinate
represents the results of the L-J analyses.
For every parameter, its L-J graph can display maximum 31 points.
For every parameter, the upper dash line represents the expected result + limit.
For every parameter, the lower dash line represents the expected result – limit.
For every parameter(e.g. WBC), the three numbers to the left of the graph are:
8-9
Fehler! Formatvorlage nicht definiert.
For every parameter,the three numbers to the right of the L-J graph are defined and
calculated as follows:
Mean – the average of the saved QC runs;
SD – Standard Deviation;
CV% – Coefficient of Variation.
8-10
Fehler! Formatvorlage nicht definiert.
∑x i
Mean= i=1
n
∑ (X − Mean )
2
SD =
i
n −1
SD
CV% = × 100
Mean
Where, n is the number of the saved L-J analyses and Xi is the result of the ith L-J analysis.
If the saved L-J analyses are less than 3, only the “mean” will be displayed. For a parameter,
if any of the saved results is non-numeric (*), the “mean”, “SD” and “CV%” are all empty.
If you see any points fallen outside the control range, do the following steps until the problem
is solved. If all the steps have failed, contanct Mindray customer service department or your
local distributor for assistance.
1. Check the upper left corner of the screen for error messages. Refer to Chapter 11
Troubleshooting Your Analyzer for solutions to any displayed error messages;
8-11
Fehler! Formatvorlage nicht definiert.
Press [↑] or [↓] to review the preceding or following screen; press [←] or [→] to review the
preceding or following result. The parameter value of the current point (the one the cursor is
located at) is displayed below the parameter box. The location of the current point is
displayed in the “No.” field. The analysis time is displayed in the “Time” field.
Press [MENU] to exit to the system menu, or press [MAIN] to exit to the “Count” screen.
SELECT “Quality Control → L-J Analysis →L-J Table → File 1” (Figure8-13) to enter the
“L-J Table” screen (Figure 8-14). Every screen displays 5 results. The parameter values
fallen outside the expected range will be flagged “H” (higher than the upper limit) or “L” (lower
than the lower limit).
8-12
Fehler! Formatvorlage nicht definiert.
Press [DLE] and a message box will pop up to ask you whether to delete all the QC results
saved in this file, as Figure8-15 shows. CLICK “Enter” to confirm the deletion; CLICK
“Cancel”to abort the deletion.
If you want to transmit all the L-J analysis results to an external computer (a host), press [1]
and a message box will pop up to confirm the transmission, as Figure8-16 shows. CLICK
8-13
Fehler! Formatvorlage nicht definiert.
Press [MENU] to exit to the system menu, or press [MAIN] to exit to the “Count” screen.
8-14
Fehler! Formatvorlage nicht definiert.
Using the “ X Analysis” program, you can provide quality control for maximum 12
parameters. The analyzer provides 9 QC files for you to save QC settings and results. Every
QC file can save maximum 31 QC run results. When the saved QC results have reached the
maximum number, the newest result will overwrite the oldest. The following introduction will
use “File 1” as the example.
8-15
Fehler! Formatvorlage nicht definiert.
If there are saved X analysis results and settings, you need to delete them first. Press [DEL]
and a message box will pop up to confirm the deletion, as shows Figure8-19 shows.
CLICK “Enter” to confirm the deletion; CLICK “Cancel”to abort the deletion.
ENTER the lot number of the control to be used into the “Lot No.” box.
ENTER the expiration date of the control to be used into the “Exp. Date” box.
ENTER the expected results (mean) and limits (range) respectively into the “Mean” box and
“Range” boxes of the parameters to be included in the X analysis.
8-16
Fehler! Formatvorlage nicht definiert.
z Refer to the instructions of use of the control for information on the lot
number, expiration date, open-vial stability days, expected results and limits.
z The entered expiration date should be either the expiration date printed on
the labeling or the open-vial expiration date, whichever is earlier.
z The open-vial expiration date is calculated as follows: the date that vial is
opened + the open-vial stability days.
Deleting settings
Printing settings
CLICK “Enter” to close the box and clear the erroneous entries. Re-enter the correct values
before trying to exit the screen again.
In case of any invalid entries of expiration dates, a message box will pop up to remind you of
the error, as Figure8-21shows. CLICK “Enter” to close the box and clear the erroneous
entries. Re-enter the correct values before trying to exit the screen again. The settings can be
8-17
Fehler! Formatvorlage nicht definiert.
saved only when both the expected result and limit are valid.
If all the entries are correct, a message box will pop up to remind you to save the changes, as
Figure 8-22 shows. CLICK “Enter” to save the changes and exit to the system menu (or the
“Count” screen); CLICK “Cancel” to abort the changes and exit to the system menu (or the
“Count” screen).
Press [MENU] and SELECT “Mode” to enter the “Mode”screen. SELECT “Whole Blood”
from the“Sample Mode” pull-down list.
Press [MENU] to enter the system menu. SELECT “Quality Control → X Analysis → X
Count → File 1” to enter the “ X Count”screen, as Figure8-23 shows.
8-18
Fehler! Formatvorlage nicht definiert.
Figure8-23 “ X Count”screen
z Be sure to use the Mindray - specified controls. Using controls other than
the specified will lead to misleading results.
z Refer to the instructions of use of the controls for how to store and use the
controls.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
8-19
Fehler! Formatvorlage nicht definiert.
z When switching from the Predilute mode to the Whole Blood mode, the
analyzer will automatically flush the fluidic system.
z Be sure to keep the sample probe tip away from the tube bottom, otherwise
the aspiration volume may be inaccurate.
z When the aspiration is done, remove the sample tube only when the sample
probe is out of the tube.
1. Be sure the System Status area displays “Ready“ and Count Mode area displays
“Whole“;
2. Present a vial of control to the sample probe so that the tip is well into the vial, and press
the aspirate key. The System Status area will display “Running” and the analyzer will
start aspirating control;
3. When you hear the beep and the sample probe is out of the vial, remove the control vial.
The sample probe will retract into the analyzer and the analysis progress will be
displayed on the screen;
4. When the analysis is finished, the sample probe is replaced, the analysis result is
displayed on the screen, and a message box pops up to confirm the validity of the
analysis result, as Figure8-24 shows;
5. CLICK “Enter” to save the result and the “NO./Total” in the upper left corner of the
screen will automatically increase by 1; CLICK “Cancel” to abort the result;
6. Follow the above steps to have another QC run. When you have obtained two valid QC
results, the analyzer will calculate the average and take it as an X analysis result. The
average will be flagged “H” or “L” if it falls outside the expected range.
8-20
Fehler! Formatvorlage nicht definiert.
To browse the result of the preceding or following X analyses, press [PgUp] or [PgDn].
Press [PRINT] to print out the current X analysis result by the printer.
8-21
Fehler! Formatvorlage nicht definiert.
“ X Graph” mode
Figure8-27“ X Graph”screen
8-22
Fehler! Formatvorlage nicht definiert.
The x-coordinate represents the number of the X analyses performed; the y-coordinate
represents the results of the X analyses;
For every parameter, the lower dash line represents the expected result – limit;
For every parameter(e.g. WBC), the three numbers to the left of the graph are:
8-23
Fehler! Formatvorlage nicht definiert.
For every parameter,the three numbers to the right of the X graph are defined and
calculated as follows:
Mean – the average of the saved X analyses;
SD – Standard Deviation;
CV% – Coefficient of Variation.
n
∑x i
Mean= i=1
n
∑ (X − Mean )
2
SD =
i
n −1
SD
CV % = × 100
Mean
Where, n is the number of the X analyses performed and Xi is the result of the ith X
analysis.
If the saved X analyses are less than 3, only the “mean” will be displayed. For a parameter,
if any of the saved results is non-numeric *, the “mean”, “SD” and “CV%” are all empty.
8-24
Fehler! Formatvorlage nicht definiert.
The “■”points fallen outside the upper and lower dash lines are out of the expected ranges ;
If you see any points fallen outside the control range, do the following steps until the problem
is solved. If all the steps have failed, contact Mindray customer service departmentor your
local distributor for assistance.
1. Check the upper left corner of the screen for error messages. Refer to Chapter 11
Troubleshooting Your Analyzer for solutions to any displayed error messages;
Printing X graphs
“ X Table” mode
8-25
Fehler! Formatvorlage nicht definiert.
SELECT “Quality Control →Analysis →Table→ File 1” (Figure 8-30) to enter the “ X
Table” screen (Figure8-31). Every screen displays 5 results. The parameter value fallen
outside the expected range will be flagged “H” (higher than the upper limit) or “L” (lower than
the lower limit).
Figure8-31“ X Table”screen
Deleting QC results
Press [DLE] and a message box will pop up to ask you whether to delete all the QC results
8-26
Fehler! Formatvorlage nicht definiert.
saved in this file, as Figure8-32 shows. CLICK “Enter” to confirm the deletion; CLICK
“Cancel” to abort the deletion.
Printing QC results
Press [PRINT] to print out all the QC results saved in this file.
8-27
Fehler! Formatvorlage nicht definiert.
Using the “ X -R Analysis” program, you can provide quality control for maximum 12
parameters. The analyzer provides 9 QC files for you to save QC settings and results. Every
QC file can save maximum 31 QC run results. When the saved QC results have reached the
maximum number, the newest result will overwrite the oldest. The following introduction will
use “File 1” as the example.
8-28
Fehler! Formatvorlage nicht definiert.
If there are saved QC results and settings, you need to delete them first. Press [DEL] and a
message box will pop up to confirm the deletion, as Figure8-35 shows.
CLICK “Enter” to confirm the deletion; CLICK “Cancel” to abort the deletion.
ENTER the lot number of the control to be used into the “Lot No.” box.
ENTER the expiration date of the control to be used into the “Exp. Date” box.
8-29
Fehler! Formatvorlage nicht definiert.
z Refer to the instructions of use of the control for information on the lot
number, expiration date, open-vial stability days, expected results and limits.
z The entered expiration date should be either the expiration date printed on
the labeling or the open-vial expiration date, whichever is earlier.
z The open-vial expiration date is calculated as follows: the date that vial is
opened + the open-vial stability days.
Deleting settings
If all the entries are correct, a message box will pop up to remind you to save the changes, as
Figure 8-37shows. CLICK “Enter” to save the changes and exit to the system menu (or the
“Count” screen); CLICK “Cancel” to abort the changes and exit to the system menu (or the
“Count” screen).
8-30
Fehler! Formatvorlage nicht definiert.
Press [MENU] and SELECT “Mode” to enter the “Mode”screen. SELECT “Whole Blood”
from the “Sample Mode” pull-down list.
Press [MENU] to enter the system menu. SELECT “Quality Control → X -R Analysis→ X -R
Count → File 1” to enter the “ X -R Count” screen (Figure8-38).
Figure8-38 “ X -R Count”screen
z Be sure to use the Mindray - specified controls. Using controls other than
the specified will lead to misleading results.
z Refer to the instructions of use of the controls for how to store and use the
controls.
8-31
Fehler! Formatvorlage nicht definiert.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
z When switching from the Predilute mode to the Whole Blood mode, the
analyzer will automatically flush the fluidic system.
z Be sure to keep the sample probe tip away from the tube bottom, otherwise
the aspiration volume may be inaccurate.
z When the aspiration is done, remove the sample tube only when the sample
probe is out of the tube.
1. Be sure the System Status area displays “Ready“ and Count Mode area displays
“Whole“;
2. Present a vial of control to the sample probe so that the tip is well into the vial, and press
the aspirate key. The System Status area will display “Running” and the analyzer will
start aspirating sample;
3. When you hear the beep and the sample probe is out of the vial, remove the control vial.
The sample probe will retract into the analyzer and the analysis progress will be
displayed on the screen;
4. When the analysis is finished, the sample probe is replaced, the analysis result is
displayed on the screen, and a message box pops up to confirm the validity of the
analysis results, as Figure8-39 shows;
8-32
Fehler! Formatvorlage nicht definiert.
5. CLICK “Enter” to save the result and the “NO./Total” in the upper left corner of the
screen will automatically increase by 1; CLICK “Cancel” to abort the result;
6. Follow the above steps to run the control again. When you have obtained two valid QC
results, the analyzer will calculate the average X and the difference R. The calculated
X and R will be respectively displayed on the screen.
Press [PgUp] or [PgDn] to browse the result of the preceding or following X -R analysis.
To delete the current X -R analysis result, press [DEL] and a message box will pop up, as
Figure8-40 shows. CLCIK “Enter” to confirm the deletion; CLICK “Cancel” to abort the
deletion.
8-33
Fehler! Formatvorlage nicht definiert.
Press [PRINT] to print out the current X -R analysis result by the printer.
You can review the X -R analysis results in either the “ X -R Graph” mode or “ X -R Table”
mode.
“ X -R Graph” mode
8-34
Fehler! Formatvorlage nicht definiert.
Figure8-42“ X -R Graph”screen
For every parameter, the center dash line represents the X (average of all the X -R
analyses performed);
For every parameter, the upper dash line represents the upper control limit = X +A× R ;
For every parameter, the lower dash line represents the upper control limit = X -A× R ;
For every parameter(e.g. WBC), the three numbers to the left of the graph are:
10.2 – X +A× R ;
10.1 – X ;
10.0 – X -A× R .
8-35
Fehler! Formatvorlage nicht definiert.
y-coordinate represents the difference between the two runs of every X -R analysis;
For every parameter,the center line of its R-graph represents the average of all the
differences R ;
For every parameter,the upper dash line represents the upper control limit B× R ;
For every parameter,the lower dash line represents the lower control limit of the
expected range C× R ;
For every parameter(e.g. WBC) the three numbers to the left of its R graph are defined
as follows:
0.3 – B× R ;
0.1 – R ;
0.0 – C× R .
If you see any points fallen outside the control range, do the following steps until the problem
is solved. If all the steps have failed, contact Mindray customer service department or your
local distributor for assistance.
1. Check the upper left corner of the screen for error messages. Refer to Chapter 11
Troubleshooting Your Analyzer for solutions to any displayed error messages;
8-36
Fehler! Formatvorlage nicht definiert.
Browsing QC results
Press [↑] or [↓] to review the preceding or following screen; press [←] or [→] to review the
preceding or following result. The X or R value of the current point (the one the cursor is
located at) is displayed below the X or R. The location of the current point is displayed in the
“No.” field. The analysis time is displayed in the “Time” field.
8-37
Fehler! Formatvorlage nicht definiert.
“ X -R Table” mode
8-38
Fehler! Formatvorlage nicht definiert.
Press [PRINT] to print out all the X -R analysis results saved in this file by the printer.
8-39
Fehler! Formatvorlage nicht definiert.
The X-B analysis is a weighted moving average analysis that uses values obtained from
patient samples. It was proposed by Brian Bull, M.D. using the 3 red cell indices, MCV, MCH
and MCHC to indicate the hematology instrument performance. Effective use of X-B requires
randomization of samples and a normal cross section of patients to prevent skewing of
indices.
It is recommended the X-B analysis be enabled when the sample volume of your laboratory is
greater then 100 samples per day.
SELECT “Quality Control → X-B Analysis → Limit” (Figure 8-46) to enter the “Limit”
screen (Figure8-47).
8-40
Fehler! Formatvorlage nicht definiert.
If there are saved QC results and settings, you need to delete them first. Press [DEL] and a
message box will pop up to confirm the deletion, as Figure8-48 shows.
CLICK “Enter” to confirm the deletion; CLICK “Cancel”to abort the deletion.
The expected results vary depending on laboratories. It is recommended they are obtained
by calculating the averages of at least 500 random patient samples. The recommended limit
is 3% - 5%.
8-41
Fehler! Formatvorlage nicht definiert.
ENTER the expected results (mean) and limits (range) respectively into the “Mean” box and
“Range” boxes of the parameters to be included in the QC run.
Deleting settings
Printing settings
Press [MENU] (or [MAIN] if you want to go directly to the “Count” screen) to exit the “Limit”
screen.
CLICK “Enter” to save the changes and exit to the system menu (or the “Count” screen);
CLICK “Cancel” to abort the changes and exit to the system menu (or the “Count” screen).
8-42
Fehler! Formatvorlage nicht definiert.
Figure8-52 “Samples/Batch”screen
8-43
Fehler! Formatvorlage nicht definiert.
Setting Samples/Batch
Press [MENU] to exit to the system menu; press [MAIN] to exit to the “Count” screen.
8-44
Fehler! Formatvorlage nicht definiert.
Random samples are required for the X-B analysis. In case of known samples of a particular
type (oncology, neonatal and so forth) that will seriously interfere with the X-B results, disable
the X-B analysis.
Press [MENU] (or [MAIN] if you want to go directly to the “Count” screen) to exit the “Limit”
screen and a message box will pop up to remind you to save the changes, as Figure8-55
shows. CLICK “Enter” to save the changes and exit to the system menu (or the “Count”
screen); CLICK “Cancel” to abort the changes and exit to the system menu (or the “Count”
screen).
8-45
Fehler! Formatvorlage nicht definiert.
SELECT “Quality Control → X-B Analysis → X-B Graph” (Figure8-56) to enter the “X-B
Graph” screen (Figure8-57).
8-46
Fehler! Formatvorlage nicht definiert.
The saved X-B analysis results are sequentially displayed in the X-B graph, the latest on the
utmost left (No.1).
The x-coordinate represents the number of X-B analyses performed; the y-coordinate
represents the results of the X-B analyses;
For every parameter, its X-B graph can display maximum 500 points, 30 points per
screen;
For every parameter, the upper dash line represents the expected result + limit;
For every parameter, the upper dash line represents the expected result – limit;
For every parameter (e.g. MCV), the three numbers to the left of the X-BFigure are
defined as follows:
90 – expected result;
80 – expected result – limit.
The “■”points fallen between the upper and lower dash lines are within the expected ranges;
The “■”points fallen outside the upper and lower dash lines are out of the expected ranges
If you see any points fallen outside the control range, do the following steps until the problem
is solved. If all the steps have failed, contact Mindray customer service department or your
local distributor for assistance.
1. Check the upper left corner of the screen for error messages. Refer to Chapter 11
8-47
Fehler! Formatvorlage nicht definiert.
Press [↑] or [↓] to review the preceding or following screen; press[←] or [→] to review the
preceding or following result. The parameter value of the current point (the one the cursor is
located at) is displayed below the parameter. The location of the current point is displayed in
the “No.” field. The analysis time is displayed in the “Time” field.
Press [MENU] to exit to the system menu, or press [MAIN] to exit to the “Count” screen.
8-48
Fehler! Formatvorlage nicht definiert.
SELECT “Quality Control → X-B Analysis → X-B Table → File 1” (Figure8-58) to enter the
“X-B Table” screen (Figure8-59). Every screen displays 5 results. The parameter value fallen
outside the expected range will be flagged “H” (higher than the upper limit) or “L” (lower than
the lower limit).
Press [DLE] and a message box will pop up to ask you whether to delete all the X-B analaysis
results saved in this file, as Figure8-60 shows. CLICK “Enter” to confirm the deletion; CLICK
“Cancel” to abort the deletion.
8-49
Fehler! Formatvorlage nicht definiert.
Press [MENU] to exit to the system menu; press [MAIN] to exit to the “Count” screen.
8-50
9 Using the Calibration Programs
9.1 Introduction
Purpose of the calibration is to maintain system accuracy. Quality of the calibration depends
on the calibration materials and reagents used. You should only use the calibrators and
reagents specified by Mindray for the calibration. Be sure to store and use the calibrators and
reagents as instructed by their instructions for use.
9-1
Fehler! Formatvorlage nicht definiert.
9-2
Fehler! Formatvorlage nicht definiert.
The analyzer provides 3 calibration programs: manual calibration, auto calibration using
commercial calibrators and auto calibration using fresh blood samples. Two sets of
calibration factors are prepared respectively for the whole blood mode and the predilute
mode.
Check and make sure enough reagents have been prepared for the calibration. You need to
start over the calibration if the reagents run out during the process.
Do the background check. If the analyzer alarms for abnormal background results, see
Chapter 11 Troubleshooting Your Analyzer for solutions.
Enter the “Count” screen and run a vial of normal control 11 consecutive times. Enter the
“Review” screen to check the reproducibility of the second to eleventh runs and make sure
they meet the following requirements.
At the “Count” screen, run a vial of high control three consective times and then immediately
run the diluent three consective times, calculate the carryover per the following equation.
The calculated carryovers shall meet the following requirements: WBC, RBC and HGB
shall be no greater than 0.5 % ; PLT shall be no greater than 1%.
9-3
Fehler! Formatvorlage nicht definiert.
It is recommended that you create a log table for your analyzer. This log table should contain
all necessary information that is pertinent to your analyzer. Suggested items that you may
want to include in the log table are:
Calibration date
Supplier of calibrator
Lot number
Enter the administrator password instructed in Chapter 5.4.1 and then choose one or several
parameters among WBC, RBC, HGB, MCV and PLT for calibration.
9-4
Fehler! Formatvorlage nicht definiert.
z When switching from the predilute mode to the whole blood mode, the
analyzer will automatically wash the fluidic system.
ENTER the lot number of the calibrator to be used into the “Lot No.” box.
ENTER the expiration date of the calibrator to be used into the “Exp. Date” box.
ENTER the expected results (mean) and limits (range) respectively into the “Mean” and
“Range” boxes of the parameters to be included in the calibration.
9-5
Fehler! Formatvorlage nicht definiert.
z Refer to the instructions of use of the calibrators for information on the lot
number, expiration date, expected results and limits.
When you have finished editing the interested settings, press [ENTER] to deactivate the edit
boxes.
z Be sure to use the Mindray- specified calibrator. Using calibrator other than
the specified will lead to misleading results.
z Refer to the instructions of use of the controls for how to store and use the
calibrator.
z Be sure to use fused silica glass/plastic test tubes and 20μL borosilicate
glass capillary tubes.
z Be sure to keep the sample probe tip away from the tube bottom, otherwise
the aspiration volume may be inaccurate.
z When the aspiration is done, remove the sample tube only when the sample
probe is out of the tube.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
9-6
Fehler! Formatvorlage nicht definiert.
1. Be sure the System Status area displays “Ready“ and Count Mode area displays
“Whole“;
2. Present a vial of mixed calibrator to the sample probe so that the tip is well into the tube,
and press the aspirate key. The System Status area will display “Running” and the
analyzer will start aspirating sample;
3. When you hear the beep and the sample probe is out of the vial, remove the calibrator.
The sample probe will retract into the analyzer and the analysis progress will be
displayed on the screen;
4. When the analysis is finished, the result will be displayed on the screen and the sample
probe will be replaced.
1. Be sure the System Status area displays “Ready“ and Count Mode area displays
“Predilute“;
2. Press [DILUENT] and a message box will pop up to instruct you how to dispense the
diluent into the sample tube, as Figure 9-3 shows;
3. Present a clean sample tube to the sample probe and make sure the tube is tilted
towards the probe, as Figure 9-4 shows, to avoid spills and bubbles. Press the aspirate
key to dispense 0.7ml of diluent (the dispensing volume is controlled by the analyzer) into
the tube;
9-7
Fehler! Formatvorlage nicht definiert.
5. Add 20μL of calibrator to the diluent and shake the tube to mix the sample.
z After mixing the calibrator with the diluent, be sure to wait 3 minutes before
running it.
z Be sure to run the prediluted calibrators within 30 minutes after the mixing.
z Mix any pre-diluted calibrator that has been prepared for a while before
running it.
6. Present the mixed calibrator to the sample probe so that the tip is well into the tube, and
press the aspirate key. The System Status area will display “Running” and the analyzer
will start aspirating sample;
7. When you hear the beep and the sample probe is out of the tube, remove the calibrator.
The sample probe will retract into the analyzer and the analysis progress will be
displayed on the screen;
8. When the analysis is finished, the result will be displayed on the screen and the sample
probe will be replaced.
9-8
Fehler! Formatvorlage nicht definiert.
Figure 9-5 A message box to warn you about the invalid results
If all the parameter values obtained are numeric, a message box will pop up to confirm the
validity of the results, as Figure 9-6 shows.
9-9
Fehler! Formatvorlage nicht definiert.
CLICK “Enter” to save the results; CLICK “Cancel” to abort the result. The saved results will
be displayed on the screen.
Repeat the above steps to run the calibrator 3 – 5 times (5 is recommended) and the analyzer
will automatically calculate the CVs and calibration factors, as Figure 9-7 shows. Be sure the
CVs meet the requirements of Table 9-1.
The calculated calibration factor should be within the 75% - 125%. Any calculated value that
falls between 0%-75% or 125%-9999% will be flagged with a “*”. Other values will not be
displayed. In case of an empty calibration factor, try to find out the reason and if necessary,
contact Mindray customer service department or your local distributor for assistance.
9-10
Fehler! Formatvorlage nicht definiert.
CLICK ”Enter” to save the new calibration factors to the “Manual Calibration” screen and
enter the “Count” screen.
At the “Count” screen, run the calibrator or a normal control material at least 5 consecutive
times and calculate the mean of the results. The means should be within the expected ranges
supplied by the manufacturer. If not, contact Mindray customer service department or your
local distributor for assistance.
9-11
Fehler! Formatvorlage nicht definiert.
SELECT “Calibration →“Fresh Blood” (Figure 9-9) to enter the “Fresh Blood” screen
(Figure9-10).
9-12
Fehler! Formatvorlage nicht definiert.
z Once switching from the predilute mode to the whole blood mode, the
analyzer will automatically wash the fluidic system.
z You should prepare 3 – 5 normal fresh blood samples for the calibration.
If all the parameter values obtained are numeric, a message box will pop up to confirm the
validity of the results, as Figure9-12 shows.
9-13
Fehler! Formatvorlage nicht definiert.
CLICK “Enter” to save the results to the “Auto – fresh blood” screen; CLICK “Cancel” to
abort the result.
z If you press [MENU] to enter the system menu before the average is
obtained at the “Calculate” screen, the next time you enter the “Auto-fresh
blood” screen a message box will pop up to ask you whether to clear the
data of the last calibration; If you press [MAIN] to exit to the “Count”screen
before the average is obtained at the “Calculate” screen, the next time you
enter the “Auto-fresh blood” screen a message box will pop up to ask you
whether to clear the data of the last calibration
Repeat the above steps to run the sample 3 – 5 times (5 is recommended) and the analyzer
will automatically calculate the CV and calibration factor, as Figure9-13 shows. Be sure the
CVs meet the requirements of Table 9-1.
9-14
Fehler! Formatvorlage nicht definiert.
The calculated calibration factors should be within the 75% - 125%. Any calculated value
that falls between 0%-75% or 125%-9999% will be flagged with a “*”. Other values will not
be displayed. In case of an empty calibration factor, try to find out the reason and if necessary,
contact Mindray customer service department or your local distributor.
You can enter the “Auto – fresh blood” screen of “Sample 2”... “Sample 5” by pressing [2] ...
[5]. Follow the calibration steps of sample 1 to run at least another two fresh blood samples.
When you have obtained the calibration factors of at least 3 fresh blood samples, you can
press [6] to enter the “Calculate” screen as Figure9-14 shows. At the screen, the digits “1”, “2”,
“3”, “4”and “5”respectively correspond to the calibration factors of samples 1 - 5.
The “Calculate” screen can maximum display 5 sets of calibration factors. The calculated
calibration factors should be within the 75% - 125%. Any calculated factor that falls between
0% - 75% or 125% - 9999% will be flagged with a “*”. Other values will not be displayed. In
case of an empty calibration factor, try to find out the reason and if necessary, contact
Mindray customer service department or your local distributor. For every parameter, the
analyzer will calculate the average calibration factor, which serves as the new calibration
factor, only when there are at least 3 valid calibration factors(e.g. RBC in Figure9-14).
Otherwise, the average calibration factor will be empty (e.g. WBC in Figure9-14).
9-15
Fehler! Formatvorlage nicht definiert.
CLICK ”Enter” to save the new calibration factors to the “Manual Calibration” screen and
enter the “Count” screen.
Method one:
Prepare 3-5 normal fresh blood samples and run each one of them on a reference analyzer at
least 3 consecutive times. Calculate the mean (MEAN 1) and SD (SD 1) of every sample.
Run the same samples on your analyzer for the same number of times and calculate the
mean (MEAN 2). The MEAN 2 should be within MEAN 1 ± 2SD. If any of the sample fails to
reach the criterion, call Mindray customers service department or your local distributor.
Method two:
At the “Count” screen, run the calibrator at least 5 consecutive times and calculate the
means of the results. The means should be within the expected ranges supplied by the
manufacturer. If not, contact Mindray customer service department or your local distributor.
9-16
Fehler! Formatvorlage nicht definiert.
Press [MENU] to enter the system menu. SELECT “Count” (Figure7-1) to enter the “Count”
screen (Figure9-17).
z Once switching from the predilute mode to the whole blood mode, the
analyzer will automatically wash the fluidic system.
9-17
Fehler! Formatvorlage nicht definiert.
SELECT “Review → Sample Review → Sample Table Review” to enter the “Sample Table
Review” screen, as Figure 9-19 shows.
9-18
Fehler! Formatvorlage nicht definiert.
Check the reproducibility as instructed in Chapter 7.11. If the reproducibility meets the
requirements listed in Table 9-1, record the Mean of the 10 runs. If the means of any
parameter falls outside the expected range (see the instructions for use of the calibrator),
calibrate the analyzer as instructed below, otherwise the calibration is not necessary.
If the reproducibility of the calibrated parameter does not meet the requirements of Table 9-1,
you must try to find out the reason and re-run the calibrator after you have solved the problem.
If necessary, contact Mindray customer service department or your local distributor for
assistance.
It is recommended that you create a log table for your analyzer. This log table should contain
all necessary information that is pertinent to your analyzer. Suggested items that you may
want to include in the log table are:
Calibration date
Supplier of calibrator
Lot number
Enter the administrator password instructed in Chapter 5.4.1 and then choose one or several
parameters among WBC, RBC, HGB, MCV and PLT for calibration.
9-19
Fehler! Formatvorlage nicht definiert.
The calculated new calibration factor should be within 75%-125%. If not, try to find out the
reason and if necessary, call Mindray customer service department or your distributor for
assistance.
SELECT “Calibration → Manual” (Figure 9-20) to enter the “Manual Calibration” screen
(Figure 9-21).
9-20
Fehler! Formatvorlage nicht definiert.
ENTER the calculated calibration factor into the corresponding boxes. To correct an
erroneous entry, DELETE the wrong digit and enter the correct digit.
CLICK ”Enter” to save the new calibration factors to the “Manual Calibration” screen and
enter the “Count” screen.
If you have used the calibrator for the manual calibration, verify the new calibration as
instructed in Chapter 9.2.2; If you have used a fresh blood sample for the manual calibration,
verify the new calibration as instructed in Chapter 9.2.3.
9-21
Fehler! Formatvorlage nicht definiert.
9-22
10 Maintaining Your Analyzer
10.1 Introduction
Routine preventive maintenance and cleaning are required to keep the BC-3000 Plus in good
operating condition. Cleanliness is important in keeping your analyzer operating efficiently
and accurately. The analyzer has automatic cleaning functions that are performed during
normal operation. These built-in functions keep the fluidic system clean.
In spite of the automatic cleaning functions, Mindray encourages you to routinely perform the
required maintenance to lengthen the operational life of your analyzer and to minimize
system problems that lead to imprecision and inaccuracy. This chapter describes the
recommended preventive maintenance procedures and provides instructions for preparing
the analyzer for an extended period of inactivity.
z Do not perform any maintenance procedures that are not described in this
chapter. Performing unauthorized maintenance procedures can damage
your analyzer.
10-1
Fehler! Formatvorlage nicht definiert.
Everyday If you are to use this analyzer 24 hours a day, be sure to perform
the “E-Z cleanser cleaning” procedure everyday.
Run the QC program everyday. See Chapter 7 Using Quality
Control Programs for details.
Every three days If you are to use this analyzer 24 hours a day, be sure to perform
the “Probe cleanser cleaning” procedure every three days.
Every Week If you shut down your analyzer every day and follow the specified
shutdown procedure to do that, you need to perform the “Probe
cleanser cleaning” procedure every week.
Every Month You should use the supplied probe localizer to calibrate the
position of the probe to that of the probe wipe. The analysis result
is sensitive to their alignment.
As needed When you think the baths might be contaminated, perform the
“Clean the baths” procedure.
When the analyzed whole blood samples add up to 300 or
prediluted samples add up to 150, the analyzer will remind you to
perform the “Probe cleanser cleaning” procedure.
When the analyzed whole blood samples add up to 2,000 or
prediluted samples add up to 4,000, the analyzer will remind you to
perform the “Clean wipe block” procedure.
When this analyzer is not to be used for two weeks, be sure to
perform the “Prepare to ship” procedure to empty and wash the
fluidic lines and then wipe the analyzer dry and wrap it up for
storage.
To obtain reliable analysis results, this analyzer needs to work in a
normal status. Be sure to run the “System Test” items regularly to
check the status of this analyzer.
When this analyzer gives alarms for clogging, you can perform the
“Flush Apertures” or “Zap Apertures” procedure, or press
[FLUSH] to unclog the apertures.
10-2
Fehler! Formatvorlage nicht definiert.
10-3
Fehler! Formatvorlage nicht definiert.
Press [MENU] to enter the system menu. SELECT “Service → Maintenance” (Figure10-1)
to enter the “Maintenance” screen (Figure10-2).
10-4
Fehler! Formatvorlage nicht definiert.
Diluent Prime
Rinse Prime
Lyse Prime
Zap Apertures
Flush Apertures
Lyse Test
Clean Baths
Empty Baths
Empty Tubing
z Be sure to keep the reagents still for a while before using them.
You should perform the “Diluent Prime” procedure to prime the diluent tubing with diluent
when
the old diluent ran out and a new container of diluent is installed.
10-5
Fehler! Formatvorlage nicht definiert.
At the “Maintenance” screen, SELECT “Diluent Prime” to prime the tubing with diluent and
the priming progress will be displayed at the bottom of the screen, Figure10-3 shows. When
the priming is done, the screen displays “Diluent Prime End”.
z Be sure to keep the reagents still for a while before using them.
You should perform the “Rinse Prime” procedure to prime the rinse tubing with rinse when
the old rinse ran out and a new container of rinse is installed.
10-6
Fehler! Formatvorlage nicht definiert.
At the “Maintenance” screen, SELECT “Rinse Prime” to prime the tubing with rinse and the
priming progress will be displayed at the bottom of the screen, as Figure10-4 shows. When
the priming is done, the screen displays “Rinse Prime End”.
z Be sure to keep the reagents still for a while before using them.
You should perform the “Lyse Prime” procedure to prime the lyse tubing with lyse when
the old lyse ran out and a new container of lyse is installed.
10-7
Fehler! Formatvorlage nicht definiert.
At the “Maintenance” screen, SELECT “Lyse Prime” to prime the tubing with lyse and the
priming progress will be displayed at the bottom of the screen, as Figure10-5 shows. When
the priming is done, the screen will display “Lyse Prime End”.
At the “Maintenance” screen, SELECT “Zap Aperture” to zap the apertures and the zapping
progress will be displayed at the bottom of the screen, as Figure10-6 shows. When the
zapping is done, the screen will display “Zap Apertures End”.
10-8
Fehler! Formatvorlage nicht definiert.
At the “Maintenance” screen, SELECT “Flush Aperture” to flush the aperture and the
flushing progress will be displayed at the bottom of the screen, as Figure10-7 shows. When
the flushing is done, the screen will display “Flush Apertures End”.
10-9
Fehler! Formatvorlage nicht definiert.
You can soak the baths and fluidic lines with the probe cleanser, an alkaline detergent, by
performing the “Probe cleanser cleaning” procedure. If your analyzer is to run 24 hours a
day, you should perform this procedure every 3 days. If you follow the shutdown procedure to
turn off your analyzer everyday, you should perform this procedure every week.
2. Present the cleanser to the probe and press [ENTER] to aspirate the cleanser. When
you hear the beep and the sample probe is out of the bottle, remove the cleanser.
10-10
Fehler! Formatvorlage nicht definiert.
3. When the screen reminds you for the second aspiration, present the cleanser to the
probe again and press [ENTER]. When you hear the beep and the sample probe is out
of the bottle, remove the cleanser and the priming progress is displayed on the screen,
as Figure10-9 shows.
4. The cleaning process will last about 15 minutes and you may press [ENTER] to stop it
any time. Note that a shortened priming process may not be as effective as a complete
one.
5. When the cleaning is done, press [ENTER] to flush the bath and tubing, after which
screen will display “Probe Cleanser Cleaning End”.
10-11
Fehler! Formatvorlage nicht definiert.
To make sure this analyzer functions normally, every time the accumulated analyzed whole
blood samples reach 300 or prediluted blood samples reach 150, a message box will pop up
to remind you to perform the “Probe cleanser cleaning” procedure, as Figure10-11 shows.
CLICK “Enter” to proceed with the cleaning; CLICK “Cancel” to cancel the cleaning.
You can use the E-Z cleanser, an enzyme based, isotonic cleaning solution and wetting agent,
to clean the tubing and bath by performing the “E-Z Cleanser Cleaning” procedure.
Follow the steps given below to perform the procedure:
10-12
Fehler! Formatvorlage nicht definiert.
2. Present the cleanser to the probe and press [ENTER] to aspirate the cleanser.
When you hear the beep and the sample probe is out of the bottle, remove the
cleanser. This analyzer will automatically prime the baths and fluidic lines with the
aspirated cleanser and the progress is displayed on the screen, as Figure10-12
shows;
3. When the priming is done, the cleaning process begins, as Figure10-13 shows. The
default cleansing time is 8 hours and you may press [ENTER] to stop the process
any time;
4. When the cleaning is done, press [ENTER] to drain the baths and fluidic lines, as
Figure10-14 shows. When the draining is done, the screen will display “E-Z
Cleanser Cleaning End”.
10-13
Fehler! Formatvorlage nicht definiert.
10-14
Fehler! Formatvorlage nicht definiert.
In case of any abnormal WBC counts or histograms, you can perform the “Lyse Test”
procedure to check whether the lyse can be dispensed properly.
Follow the steps given below to do so:
Figure10-15
2. Lift up the front panel latch as indicated in Figure10-16 and open the front panel.
Figure10-16
10-15
Fehler! Formatvorlage nicht definiert.
3. Remove the screws fixing the shielding box of the bath and lift the shielding box, as
Figure10-17 shows.
10-16
Fehler! Formatvorlage nicht definiert.
5. At the “Maintenance” screen, SELECT “Lyse Test”. Press [ENTER] and the
analyzer will automatically drain the WBC bath and then dispense 2ml of lyse into
the WBC bath.
6. Check the scale to see whether the lyse has reached the expected line (the first
from bottom).If so, press [ENTER] and the analyzer will automatically flush the bath
and dispense lyse and the test is done.
7. If not, repeat steps 5 and 6 several times. If all the tries have failed, check whether
the lyse has run out or the lyse pickup tube is not properly connected to this
analyzer. If the lyse is still enough and the tube is well connected to the analyzer,
contact the Mindray or your local distributor for repair.
3. When the cleaning is done, the screen displays “Clean Baths End”.
10-17
Fehler! Formatvorlage nicht definiert.
3. When the draining is done, check the baths and the tubing below them for residual fluid. If
there is no fluid, press [ENTER] to prime the baths with diluent, as Figure10-21 shows.
When the priming is done, the screen displays “Empty Baths End”.
10-18
Fehler! Formatvorlage nicht definiert.
If there is fluid left, turn off the analyzer and call Mindray customer service department or your
local distributor for assistance.
10-19
Fehler! Formatvorlage nicht definiert.
If this analyzer is not to be used for a long time or it is to be maintained, be sure to perform
the “Empty Tubing” procedure to drain the fluidic lines.
1. At the “Maintenance” screen, press the appropriate arrow keys ([←][→] [↑][↓]) to move
the cursor to “Empty Tubing”.
2. Follow the displayed instructions to remove the diluent, rinse and lyse pickup tubes from
this analyzer and then press [ENTER] to start the draining process, Figure10-22.
3. When the draining is done, the screen will display “Turn off this analyzer” and you
10-20
Fehler! Formatvorlage nicht definiert.
2. Present the probe cleanser to the sample probe and press [ENTER] to aspirate the
cleanser. When you hear the beep and the sample probe is out of the bottle, remove the
cleanser.
Figure10-23
4. Lift up the front panel latch as indicated in Figure10-24 and open the front panel.
10-21
Fehler! Formatvorlage nicht definiert.
Figure10-24
10-22
Fehler! Formatvorlage nicht definiert.
5. Follow the instructions displayed on the screen to place an empty cup, whose diameter
should be no less than 8cm, below the sample probe.
6. Press [ENTER] to soak the wipe block with the aspirated cleanser. The soaking progress
will be displayed on the screen, as Figure10-25 shows.
z Spills are possible during the soaking process. Keep a minimum 30cm
distance from the analyzer.
7. When the soaking is done, wipe the bottom of the wipe block with a probe
cleanser-dipped cloth that does not leave debris.
8. Press [ENTER] to flush the block and the interior of the probe and the flushing progress is
displayed on the screen as Figure10-26 shows .
9. After the flushing is done, the screen returns to the initial state.
When the accumulated analyzed whole blood samples reach 2,000, or prediluted samples
reach 4,000, a message box will pop up to remind to clean the probe wipe, as Figure10-27
shows. CLICK “Enter” to do the procedure; CLICK “Cancel” to abort the procedure.
10-23
Fehler! Formatvorlage nicht definiert.
10-24
Fehler! Formatvorlage nicht definiert.
The items displayed in the “System Status” screen reflect how the analyzer is functioning
and contribute significantly to diagnosing analyzer errors. You may follow the instructions
given below to check those items.
Press [MENU] to enter the system menu. SELECT “Service” → “System Status”
(Figure10-28) to enter the “System Status” screen (Figure10-29).
10-25
Fehler! Formatvorlage nicht definiert.
Note that you can only view the displayed status items without changing them. If any of the
displayed item exceeds the given range, see Chapter 11 Troubleshooting for solutions.
Press [MENU] to exit to the system menu and the screen will display “Resetting” and the
system menu will pop up later.
10-26
Fehler! Formatvorlage nicht definiert.
Malfunctioning valves will lead to fluidic system malfunctions. Therefore, testing the valves is
a major way to remove fluidic errors.
Press [MENU] to enter the system menu. SELECT “Service → Valve Test” (Figure10-30) to
enter the “Valve Test” screen (Figure10-31).
SELECT the valve you want to check and press [ENTER] to test it. If the valve goes through
the Off-On-Off sequence without making any abnormal sound, it passes the test. Otherwise,
something may be wrong with the valve.
10-27
Fehler! Formatvorlage nicht definiert.
Press [MENU] to enter the system menu. SELECT “Service → System Test” ( Figure10-32)
to enter the “System Test” screen (Figure10-33), where 19 test items are available. Note that
you need to enter the administrator password to test the motors.
SELECT the desired item to perform the corresponding test. The test result will be displayed
later. In case of any abnormal test result, see Chapter 11 Troubleshooting Your Analyzer
for solutions and if necessary, contact Mindrayor your local distributor for assistance.
Press [MENU] to exit to the system menu and the screen will display “Resetting” and the
system menu will pop up later.
10-28
Fehler! Formatvorlage nicht definiert.
Use the “Prepare to ship” program to prepare your analyzer for a prolonged period of
non-use or for shipping.
1. Remove the diluent, rinse and lyse pickup tubes from their containers and press [ENTER].
A message box will pop up to confirm the operation, as Figure10-36 shows.
10-29
Fehler! Formatvorlage nicht definiert.
3. The analyzer starts to drain the fluidic lines and the progress is displayed on the screen,
as Figure10-37 shows.
4. When the draining is done, place the diluent, rinse and lyse pickup tubes into a container
filled with distilled water and press [ENTER], as the Figure10-38 shows.
10-30
Fehler! Formatvorlage nicht definiert.
5. When the washing is done, remove the diluent, rinse and lyse pickup tubes from the
distilled water and press [ENTER] to drain the fluidic lines.
6. When the draining is done and the screen displays “You can turn off the analyzer now”,
turn off the analyzer as instructed.
10-31
Fehler! Formatvorlage nicht definiert.
The analyzer can store maximum 1,000 latest error messages. When the maximum number
has reached, the latest overwrites the earliest.
Press [MENU] to enter the system menu. SELECT “Service → Error Message”
(Figure10-39) to enter the “Error Message” screen (Figure10-40).
Press [↑] or [↓] to browse the error messages. Press [PRINT] to print out the displayed error
messages.
10-32
Fehler! Formatvorlage nicht definiert.
For the displayed error messages, see Chapter 11 Troubleshooting Your Analyzer for
solutions.
10-33
Fehler! Formatvorlage nicht definiert.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
The relative position between the sample probe and probe wipe block has influence on the
analysis results. In the accessory box, there is a sample probe localizer, as Figure10-41
shows. You need to use the localizer to adjust the position of the sample probe if you have
replaced wipe block, or observed motor error, or wrong analysis result. Also, as required by
regular maintenance, you should use the localizer to adjust the position of the sample probe
monthly.
10-34
Fehler! Formatvorlage nicht definiert.
Figure10-42
3. Lift up the front panel latch as indicated in Figure10-43 and open the front panel;
Figure10-43
4. SELECT “Service” →“System Test” to enter the “System Test” screen and SELECT
“Elevator motor”;
5. Press [↑] to move the sample probe to its highest position, as Figure10-44 shows;
10-35
Fehler! Formatvorlage nicht definiert.
Figure10-44
Figure10-45
7. Remove the probe from the wipe block and insert the localizer into the wipe block from
the bottom, as Figure10-46 shows;
Figure10-46
8. Insert the probe into the wipe block until it reaches the localizer, as Figure10-47 shows;
10-36
Fehler! Formatvorlage nicht definiert.
Figure10-47
9. Tighten the fixing screws and put away the localizer to finish the work.
10-37
Fehler! Formatvorlage nicht definiert.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
2. Pull the loosen probe wipe upward to remove the wipe block and disconnect the tubes
from the wipe block (pay attention to the correspondence between the tubes and the
connectors), as Figure10-48 shows;
Figure10-48
3. Install a new block and connect the tubing end with the black marking to the connector
below the block;
4. Refer to Chapter 10.9 and do the steps 7 - 9 to fix the sample probe.
10-38
Fehler! Formatvorlage nicht definiert.
You need to replace the filter of the vacuum chamber when there is an air filter error. Follow
the steps below to do so:
Figure10-49
3. Remove the filter and take a new one from the accessory kit and install it.
10-39
Fehler! Formatvorlage nicht definiert.
Press the [Flush] key to unclog the apertures when the analyzer alarms you for clogged WBC
or RBC aperture.
10-40
Fehler! Formatvorlage nicht definiert.
Press the [Startup] key to flush the fluidic lines and check the background.
10-41
11 Troubleshooting Your Analyzer
11.1 Introduction
The BC-3000 Plus continuously monitors the status of the system and displays pertinent
information in the upper left corner of the “Count” screen (the Error Message area). If a
problem is detected, the Error Message area displays the corresponding error message. This
chapter contains information that is helpful in locating and correcting problems that may occur
during operation of your analyzer.
z Unless otherwise instructed, always turn off the power before trying to fix
the error.
11-1
Fehler! Formatvorlage nicht definiert.
Liquid drips from Damaged pump hose or 1. Turn off the power and wipe the
analyzer inside. blocked filter. analyzer dry.
11-2
Fehler! Formatvorlage nicht definiert.
The analyzer can provide 41 error messages. See the tables below for the error messages
and their probable causes and recommended action. If the problem still remains after you
have tried the recommended solutions, contact Mindray customer service department or your
local distributor.
Vacuum Low The vacuum degree does not 1. Check the tubes connected to the
reach the expected value back of the analyzer and make sure
within the given time. they are not pressed;
11-3
Fehler! Formatvorlage nicht definiert.
Pressure 1 low The pressure inside the 1. Enter the “Service → System Test”
vacuum chamber does not screen and do the “Pressure 1”
reach the expected value procedure as instructed in Chapter
within the given time. 10.6. The error will be removed if the
test result is normal;
11-4
Fehler! Formatvorlage nicht definiert.
11-5
Fehler! Formatvorlage nicht definiert.
2.5ml&50ul Motor
1. Poor contact of the signal 1. Enter the “Service → System
Error
line; Test” screen and check the motor
Elevator Motor
1. Jammed sample probe; 1. Open the front panel and check if
Error
11-6
Fehler! Formatvorlage nicht definiert.
the drive board and the error will be removed if the test
Rotation Motor
1. Jammed sample probe; 1. Open the front panel and check if
Error
2. Poor contact of the signal the sample probe is jammed;
11-7
Fehler! Formatvorlage nicht definiert.
Chapter 10.6;
PLT Interrupt Error Something is wrong with the 1. Enter the “Service → System
A/D part of the CPU board. Test” screen and check the PLT
AD interrupt as instructed in
Chapter 10.5;
2. The error will be removed if the test
result is normal;
3. If the problem remains, contact
Mindray customer service
department or your local distributor.
11-8
Fehler! Formatvorlage nicht definiert.
“3.3V” voltage;
the analyzer are pressed. 3. Enter the “Count” screen and press
[STARTUP] (or [F3] of the external
keyboard) to do the startup
procedure;
11-9
Fehler! Formatvorlage nicht definiert.
WBC Clog
1. Clogged WBC aperture; 1. Enter the “Service →
11-10
Fehler! Formatvorlage nicht definiert.
11-11
Fehler! Formatvorlage nicht definiert.
WBC bubbles
1. Diluent or rinse running 1. Check if the diluent or rinse has run
out; out. If so, change a new container
RBC clog
1. Clogged RBC aperture; 1. Enter the “Service →
11-12
Fehler! Formatvorlage nicht definiert.
RBC bubbles
1. Diluent or rinse running 1. Check if the diluent or rinse has run
out; out. If so, change a new container
11-13
Fehler! Formatvorlage nicht definiert.
Chapter 5.3;
Barcode Error
1. Poor connection between 1. Check if the analyzer is well
the scanner and the connected to the analyzer;
analyzer; 2. Check if the bar-code is valid;
2. Invalid bar-code. 3. If the problem remains, contact
Mindray customer service
department or your local distributor.
Printer Offline Poor connection between the Check if the printer is well connected to
printer and the analyzer. the analyzer.
Recorder Com Shut down the analyzer and contact
1. Poor connection between
Error Mindray customer service department.
the recorder and the
analyzer;
2. Damaged recorder.
11-14
Fehler! Formatvorlage nicht definiert.
Recorder too Hot Recorder head too hot. Stop using the recorder. If the problem
repeats, contact Mindray customer
service department.
Press Bar Up Tension lever not replaced.
1. Press the tension lever as
instructed in Chapter 4.4.3;
11-15
Fehler! Formatvorlage nicht definiert.
unreliable;
Dynamic Memory Something is wrong with the Shut down the analyzer and contact
Error system memory. Mindray customer service department
or your local distributor.
11-16
12 Appendices
A Index
A
calibrator, 2-15
analyzer
clog, 5-36
name, 2-1
RBC, 11-12
intended use, 2-2
WBC, 11-10
aperture
control, 2-15
flush, 10-9
Coulter Principle, 3-6, 3-10
zap, 10-8
count
principle, 3-6, 3-10
B procedure, 6-8, 6-17
Bath CV
A-1
Fehler! Formatvorlage nicht definiert.
A-2
Fehler! Formatvorlage nicht definiert.
granulocyte, 2-2
lymphocyte, 2-2 N
mid-sized cell, 2-2
NRBC, 3-7
linearity range, 12-4
Lymph#
definition, 2-2 P
formula, 3-8 parameter
Lymph% WBC, 2-2
definition, 2-2 RBC, 2-2
formula, 3-7 HGB, 2-2
lyse PLT, 2-2
connection, 4-9 WBC Histogram, 2-2
definition, 2-16 RBC Histogram, 2-2
prime, 10-7 PLT Histogram, 2-2
Lymph%, 2-2
Mid%, 2-2
M
Gran%, 2-2
maintenance, 10-1
MCV, 2-2
MCH
RDW-CV, 2-2
definition, 2-2, 3-10
RDW-SD, 2-2
formula, 3-10
MPV, 2-2
MCHC
PDW, 2-2
definition, 2-2
A-3
Fehler! Formatvorlage nicht definiert.
A-4
Fehler! Formatvorlage nicht definiert.
S U
sample unpacking, 4-4
samples collection and handling, 6-5
run, 6-8, 6-17
V
sample probe
valve
calibrate, 10-33
test, 10-26
sample probe localizer, 10-33
volumetric metering, 3-5
setup
autoclean time, 5-5
count time, 5-6 W
date & time, 5-19 WBC
gain, 5-21 definition, 2-2, 3-7
password, 5-8 formula, 3-7
parameter units, 5-31 linearity range, 12-4
printing & communication, 5-2 reproducibility, 12-5
reagents Exp. Date, 5-27 weight, 12-7
report title, 5-29
other, 5-34
shutdown, 6-26
X
specifications, 12-2 X-B analysis
edit, 8-39
review, 8-45
T run, 8-45
table X Analysis, 8-15
sample table, 7-2 edit, 8-15
search table, 7-19 review, 8-22
Temperature, 4-2, 12-7 run, 8-18
throughput, 12-3 X -R Analysis, 8-28
A-5
Fehler! Formatvorlage nicht definiert.
Z
zap apertures, 10-8
A-6
B Specifications
B.1 Classification
According to the CE classification, the BC-3000 Plus is an In Vitro Diagnostic device.
B.2 Reagents
Diluent M-30D DILUENT
Rinse M-30R RINSE
Lyse M-30CFL LYSE
M-30E E-Z CLEANSER
E-Z Cleanser(Enzyme cleanser)
B.3 Parameters
B-1
Fehler! Formatvorlage nicht definiert.
B.4.5 Throughput
Less than 1 minute / analysis
B-2
Fehler! Formatvorlage nicht definiert.
B-3
Fehler! Formatvorlage nicht definiert.
B.5.4 Reproducibility
These reproducibility requirements applies only to the situation in which 11 normal-level
controls have been run and the results of the 2nd to 11th runs are used to calculate the
reproducibilities.
Parameter Condition
Reproducibility(CV%)
B.5.5 Carryover
Parameter Carryover
WBC ≤ 0.5 %
RBC ≤ 0.5 %
HGB ≤ 0.5 %
PLT ≤1%
B.6.1 Display
B.6.2 Keypad
23-key keypad.
B.6.3 Keyboard
PS/2 keyboard.
B-4
Fehler! Formatvorlage nicht definiert.
B.6.5 Recorder
Built-in thermal recorder that supports two printing formats and auto printing.
B.6.6 Printer(optional)
EPSON LQ-300K+.
B.6.7 Interfaces
A keyboard interface.
A power supply for the floppy disk drive(only to be used with the power cable supplied by
Mindray ).
Fuse: AC 250 V T4 A.
B-5
Fehler! Formatvorlage nicht definiert.
B.9 Sound
Maximal sound: 77 dB.
Operating temperature: 15 ℃ - 35 ℃;
Relative humidity: 30 % - 85 %;
Relative humidity: 10 % - 93 %
B.12 Dimensions
Length Width Height
40 cm 39 cm 46 cm
B-6
Fehler! Formatvorlage nicht definiert.
B.13 Weight
21 kg
B.14 Contraindications
None.
B-7
C Precautions, Limitations and Hazards
C.1 Introduction
You will find the following symbols in this manual.
C.1.2 Limitations
Whenever the results are outside the normal limits, it is recommended that the laboratory
following whatever written protocol is in place for validating results.
If an error occurs, the analyzer displays the corresponding error message In case of errors
related to the fluidic system (such as clogging or bubbles), it is recommnended that you
re-run the sample after removing the error.
If the PLT value is less than 100 × 109 / L, it is recommended the result be verified by a
microscope.
C.1.3 Maintenance
The maintenance instructions in Chapter 10 decribe corrective and preventive procedures
that must be followed to ensure proper operation and performance of your analyzer.
C-1
Fehler! Formatvorlage nicht definiert.
C.2 Warnings
z Before turning on the analyzer, make sure the input voltage meets the above
requirements.
z When moving the analyzer, be sure to face the front of the analyzer and
carry it from the bottom with hands!
z The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
z If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
z To avoid personal injury, be sure to keep your clothes, hair and hand away
from such moving parts as the sample probe.
C-2
Fehler! Formatvorlage nicht definiert.
C.3 Cautions
z Liquid ingression may damage the analyzer. Do not place any bottles on the
analyzer.
z Improper installation of recorder paper may jam the paper and/or result in
blank printouts.
z When dispensing or aspirating liquids, remove the bottle or tube away only
after the sample probe is out of it.
z Do not perform any maintenance procedures that are not described in this
chapter. Performing unauthorized maintenance procedures can damage
your analyzer.
C-3
Fehler! Formatvorlage nicht definiert.
C.4 Notes
z This analyzer adopts a fixed decimal point. You can enter the digits without
bothering to look for the [.] on the external keyboard.
z The purpose of this analyzer is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that
require additional studies.
z Store and use the reagents as instructed by instructions for use of the
reagents
z When you have changed the diluent, rinse or lyse, run a background to
ensure that the system is primed immediately prior to running any samples.
z Pay attention to and record the expiration date and open-container stability
of all the reagents. Be sure not to use expired reagents
z Be sure to retain the shipping carton and all the packing materials, as they
can be used for packaging if analyzer must be reshipped.
z Pay attention to and record the expiration dates and open-container stability
days of all the reagents. Be sure not to use expired reagents
z For any reagent, the entered expiration date should be either the expiration
date printed on the labeling or the open-container expiration date, whichever
is earlier. The open-container expiration date is calculated as follows: the
date that container is opened + the open-container stability days.
z For the whole blood samples to be used for WBC differential or PLT count,
you shall store them at the room temperature and run them within 8 hours
after collection.
C-4
Fehler! Formatvorlage nicht definiert.
z If you do not need the PLT, MCV and WBC differential results, you can store
the samples in a refrigerator (2℃ - 8℃) for 24 hours. You need to warm the
z Be sure to mix any sample that has been prepared for a while before running
it.
z After mixing the capillary sample with the diluent, be sure to wait 3 minutes
before running the sample. Be sure to run the prediluted samples within 30
minutes after the mixing.
z When switching from the predilute mode to the whole blood mode, the
analyzer will automatically wash the fluidic system.
z After entering all the desired information, you may press [F4] on the external
keyboard to save the changes and exit to the “Count” screen.
z Be sure to keep the sample probe tip away from the tube bottom, otherwise
the aspiration volume may be inaccurate.
z If the PLT value is less than 100 × 109 / L, it is recommended the result be
verified by a microscope.
z To ensure the stability of the analyzer and the acuuracy of the results, be
sure to do the Shutdwon procedure if your analyzer has been working for 24
consecutive hours.
z After entering all the desired information, you may press [F4] on the external
keyboard to save the changes and exit to the “Sample (or Search) Histogram
C-5
Fehler! Formatvorlage nicht definiert.
Review” screen.
z Refer to the instructions of use of the control for information on the lot
number, expiration date, open-vial stability days, expected results and limits.
z The entered expiration date should be either the expiration date printed on
the labeling or the open-vial expiration date, whichever is earlier. The
open-vial expiration date is calculated as follows: the date that vial is opened
+ the open-vial stability days.
z Refer to the instructions of use of the controls for how to store and use the
controls.
z Refer to the instructions of use of the calibrator for information on the lot
number, expiration date, open-vial stability days, expected results and limits.
z The entered expiration date should be either the expiration date printed on
the labeling or the open-vial expiration date, whichever is earlier. The
open-vial expiration date is calculated as follows: the date that vial is opened
+ the open-vial stability days.
z Refer to the instructions of use of the calibrator for how to store and use the
calibrator.
z After mixing the calibrator with the diluent, be sure to wait 3 minutes before
running it. Be sure to run the prediluted calibrator within 30 minutes after the
mixing.
z Be sure to mix any prediluted calibrator that has been prepared for a while
before running it.
z You should prepare 3 – 5 normal fresh blood samples for the calibration.
z If you press [MENU] to enter the system menu before the average is
obtained at the “Calculate” screen, the next time you enter the “Auto-fresh
blood” screen a message box will pop up to ask you whether to clear the
data of the last calibration; If you press [MAIN] to exit to the “Count”screen
before the average is obtained at the “Calculate” screen, the next time you
enter the “Auto-fresh blood” screen a message box will pop up to ask you
whether to clear the data of the last calibration.
C-6
Fehler! Formatvorlage nicht definiert.
z Spills are possible during the soaking process. Keep a minimum 30cm
distance from the analyzer.
z Running sample with the background abnormal error present will lead to
unreliable results.
z Unless otherwise instructed, always turn off the power before trying to fix
the error.
z Be sure to use the printer and scanner of the specified model only.
C-7
Fehler! Formatvorlage nicht definiert.
C.5 Biohazard
z All the analyzer components and surfaces are potentially infectious, take
proper protective measures for operation or maintenance.
C-8
Fehler! Formatvorlage nicht definiert.
Parameter flags
If the analysis result is followed by an ”H” or “L”, it means the analysis result has exceeded
the upper or lower limit of the reference range.
If you see *** as opposed to the result, it means the result is either unreliable or out of the
operating range.
If the WBC result is less than 0.5 × 109/L, this analyzer will not perform the differential
analysis and all the related parameter values will be non-numeric (***).
Histogram flags
The system will flag abnormal histograms.
Abnormal WBC histograms will be flagged by one of the markings: R1,R2,R3,R4 and
Rm.
R1:indicates abnormality on the left side of the lymphocyte hump and possible presence of
platelets coagulate, large platelet, nucleated red cell, insolvable red cell, protein, lipoid debris
in sample, or electrical noise.
R2: indicates abnormality between the lymphocyte hump and the mononuclear area and
possible presence of atypical lymphocyte, original cell in the sample and increased eosinophil
or increased basophil.
R4 :indicates abnormality on the right side of the neutrophilic granulocytes hump and
Abnormal PLT histograms will be flagged by one of the markings: Pm,PS and PL.
C-9
Fehler! Formatvorlage nicht definiert.
Pm:indicates blur demarcation between the platelet and red blood cell area and possible
presence of large platelet, platelet coagulation, small red blood cell, cell debris or fibrin.
C-10
Fehler! Formatvorlage nicht definiert.
In case of any abnormal QC results, do the following steps until the problem is solved. If all
the steps have failed, contanct Mindray customer service department or your local distributor
for assistance.
Check the upper left corner of the screen for error messages. Refer to Chaper 11
Troublshooting Your Analyzer for solutions to any displayed error messages;
C-11
D Communication
D.1 Introduction
The BC-3000 Plus can transmit the sample data and QC data to an external computer (a host)
through its RS-232 serial port. The transmission can be conducted either automatically or
through the command of the operator after the completion of the sample analysis. This
section gives detailed discussion about the setup of transmission parameter, RS-232 serial
port and the data transmission format, therefore, providing detailed information for the
software engineers to program and for the user to conveniently perform transmission.
z When the communication symbol in the upper right corner of the screen
appears animated, it indicates the communication is in process.
D-1
Fehler! Formatvorlage nicht definiert.
D.2 Connection
The BC-3000 Plus can be connected with an external computer through a DB9 connector.
The pins of the DB9 connector are shown in Figure12-1.
Pin description:
DCD:Carrier Detect
RXD:Receive Data
TXD:Transmit Data
GND:Signal Ground
RTS:Request to Send
CTS:Clear to Send
RI:Ring Indicator
The BC-3000 Plus communicates with a host through serial port 2, using Pin2, Pin 3 and Pin
5. The maximum transmission distance is 12 meters.
D-2
Fehler! Formatvorlage nicht definiert.
D.3.1 Description
Symbols
[ENQ] 0x05
[STX] 0x02
[EOT] 0x04
[EOF] 0x1A
[ETX] 0x03
[ACK] 0x06
[NACK] 0x15
"A" 0x41
"B" 0x42
"C" 0x43
"#" 0x30-0x39
"*" 0x2A
If the Lot No., Month, Day, Year are empty in QC Edit menu, the “*” (2A Hex) will be
transmitted to the host.
For all the data formats, if the data are marked “*”, then “*” (2A Hex) will be transmitted to the
host.
D-3
Fehler! Formatvorlage nicht definiert.
Programming
If the Handshake is off, BC-3000 Plus will transmit the body of the text without acknowledging
the presence of an external computer.
If the Handshake is on, BC-3000 Plus will communicate with the external computer in
following procedures:
1. BC-3000 Plus sends an ENQ (05 Hex), then waits up to 4 seconds for the external
computer to respond. If the external computer does not respond, then one more ENQ (05
Hex) is tried. If it fails again, the analyzer aborts the transmission and reports a
transmission error;
2. The external computer must respond by sending an ACK (06 Hex). If any other response
is received, another ENQ (05 Hex) will be sent by the analyzer (maximum two ENQ [05
Hex] will be sent);
Body of text
EOT (04 Hex)
ETX (03 Hex)
4. Disconnection.
BC-3000 Plus sends an ETX 03 Hex), then waits 4 seconds for the external computer to
respond. If no response is received, one more ETX (03 Hex) is sent, BC-3000 Plus waits 4
D-4
Fehler! Formatvorlage nicht definiert.
D-5
Fehler! Formatvorlage nicht definiert.
If the external compute responds ACK, the transmission is done successfully. If the external
computer responds NACK(15 Hex), the analyzer repeat the transmission from step 3. If the
received response from the computer is neither ACK(06 Hex) nor NACK(15 Hex), the
D-6
Fehler! Formatvorlage nicht definiert.
PDW ##.#
PCT[%] .###
RDW-SD[fL] ###.#
Reserved ############
Rm #
R1 #
R2 #
R3 #
R4 #
Pm #
Ps #
Pl #
L1 Region ###
L2 Region ###
L3 Region ###
L4 Region ###
L5 Region ###
L6 Region ###
L7 Region ###
L8 Region ###
Reserved ################
WBC Histo (256 channels) ###
RBC Histo (256 channels) ###
PLT Histo (256 channels) ###
Body of the text end
If handshake is enabled [EOT]
If handshake is disabled [EOF]
D-7
Fehler! Formatvorlage nicht definiert.
Lymph#[109/L] ###.#
Lymph%[%] ##.#
9
Gran#[10 /L] ###.#
Gran%[%] ##.#
HCT[%] ##.#
MCV[fL] ###.#
MCH[pg] ###.#
MCHC[g/L] ####
9
WBC Limit[10 /L] ###.#
RBC Limit[1012/L] #.##
HGB Limit[g/L] ###
9
PLT Limit[10 /L] ####
9
Lymph# Limit[10 /L] ###.#
Lymph% Limit[%] ##.#
9
Gran# Limit[10 /L] ###.#
Gran% Limit[%] ##.#
HCT Limit[%] ##.#
MCV Limit[fL] ###.#
MCH Limit[pg] ###.#
MCHC Limit[g/L] ####
Body of the text end
If handshake is enabled [EOT]
If handshake is disabled [EOF]
If handshake is enabled [ETX]
D-8
Fehler! Formatvorlage nicht definiert.
Gran%[%] ##.#
HCT[%] ##.#
MCV[fL] ###.#
MCH[pg] ###.#
MCHC[g/L] ####
Body of the text end
If handshake is enabled [EOT]
If handshake is disabled [EOF]
If handshake is enabled [ETX]
D.4 Transmission
D-9
P/N:3003-20-34858