APP 104 Remote Display

This technical guide provides instructions for configuring a Z series indicator as a master and another as a remote display. The master is set up to broadcast data from port 2 using print format 7. The remote is configured to receive this broadcast data using port 1 and protocol 1 set to remote display. Settings on both indicators are outlined, including enabling the remote display application on the remote unit. Wiring connections between the two units are also specified.

Uploaded by

Robinson Sanchez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
72 views3 pages

APP 104 Remote Display

This technical guide provides instructions for configuring a Z series indicator as a master and another as a remote display. The master is set up to broadcast data from port 2 using print format 7. The remote is configured to receive this broadcast data using port 1 and protocol 1 set to remote display. Settings on both indicators are outlined, including enabling the remote display application on the remote unit. Wiring connections between the two units are also specified.

Uploaded by

Robinson Sanchez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

Guide: APP 104

For more information login:


www.averyweigh-tronix.com

Product: Z Series
Tech Guide: Remote Display
Introduction
This technical guide will provide instructions on how to configure and connect a Z series Master
indicator to another Z series indicator setup as a Remote indicator. The Remote indicator keys
will perform the same functions as the keys on the Master indicator. In the example the default
settings for the Com Ports will be assumed; 9600 Baud, 8 Data Bits, No Parity, 1 Stop Bit. The
Master indicator Com Port 1 and Protocol 1 will be left for use with a connected printer.

In this example the Master indicator will use Protocol 2 set for Broadcast on Com Port 2 using
print format #7, a pre-defined print format for use with remote displays. Inhib (motion inhibit)
and Rate can be left in the default settings. The Remote indicator will use Protocol 1 set for
Remote Display on Com Port 1.
Setup Menu – Master indicator
Setup

Calib Scale System Ports

Serial E-net Protcl

Prot 2

Type 2 Bind 2 Attr 2

B-Cast Port 2 Prnft Inhib Rate

7 Off 1 2 5 10

Configuration Procedure - Master


1. Press and hold F1 until PASS is displayed…release F1 > 0 is displayed
2. Enter 3 0 8 8 and press ZERO > BUSY is displayed and then SETUP
3. Press SELECT > CALIB is displayed
4. Press UNITS until PORTS is displayed
5. Press SELECT > SERIAL is displayed
6. Press UNITS until PROTCL is displayed
7. Press SELECT > PROT 1 is displayed
8. Press UNITS > PROT 2 is displayed
9. Press SELECT > TYPE 2 is displayed
10. Press SELECT > NONE or current setting is displayed
11. Press UNITS until B-CAST is displayed
12. Press ZERO > TYPE 2 is displayed
13. Press UNITS > BIND 2 is displayed
14. Press SELECT > NONE or current setting is displayed
15. Press UNITS until PORT 2 is displayed
16. Press ZERO > BIND 2 is displayed

APP 104 Remote Display.docx Page 1 of 3


Tech Guide: Remote Display continued Guide: APP 104

17. Press UNITS > ATTR 2 is displayed


18. Press SELECT > PRNFT is displayed
19. Press SELECT > 1 or current setting is displayed
20. Press TARE until 7 is displayed
21. Press ZERO > PRNFT is displayed
22. Press TARE until SAVE NO is displayed
23. Press UNITS > SAVEYES is displayed
24. Press ZERO and the indicator will reboot and return to normal weigh mode.

Setup Menu – Remote Display


Setup

Calib System Ports

Site App Serial Protcl

General R-disp Prot 1

Type 1 Bind x

RD Port 1

Configuration Procedure – Remote Display Application & Port Setup


1. Press and hold F1 until PASS is displayed…release F1 > 0 is displayed
2. Enter 3 0 8 8 and press ZERO > BUSY is displayed and then SETUP
3. Press SELECT > CALIB is displayed
4. Press UNITS until SYSTEM is displayed
5. Press SELECT > SITE is displayed
6. Press UNITS until APP is displayed
7. Press SELECT > GENERAL or the active application is displayed
8. Press SELECT until R-DISP is displayed
9. Press ZERO > APP is displayed
10. Press TARE > SYSTEM is displayed
11. Press UNITS > PORTS is displayed
12. Press SELECT > SERIAL is displayed
13. Press UNITS until PROTCL is displayed
14. Press SELECT > PROT 1 is displayed
15. Press SELECT > TYPE 1 is displayed
16. Press SELECT > NONE or current setting is displayed
17. Press UNITS until RD is displayed

APP 104 Remote Display.docx Page 2 of 3


Tech Guide: Remote Display continued Guide: APP 104

18. Press ZERO > TYPE 1 is displayed


19. Press UNITS > BIND 1 is displayed
20. Press SELECT > NONE or current setting is displayed
21. Press UNITS until PORT 1 is displayed
22. Press ZERO > BIND 1 is displayed
23. Press TARE until “SAVE NO is displayed
24. Press UNITS > SAVEYES is displayed
25. Press ZERO > BUSY is displayed. The indicator will reboot and return to normal weighing

The Remote Display Application must be enabled in the Supervisor menu to complete the
configuration
Supervisor Menu – Enable Remote Display
Super

Setpnt R-Disp

Enable

Off On

Configuration Procedure – Remote Display Enable


1. Press and hold F1 until PASS is displayed…release F1 > 0 is displayed
2. Enter 1 7 9 3 and press ZERO > BUSY is displayed and then SUPER
3. Press SELECT > SETPNT is displayed
4. Press UNITS > R-DISP is displayed
5. Press SELECT > ENABLE is displayed
6. Press SELECT > OFF is displayed
7. Press UNITS > ON is displayed
8. Press ZERO > ENABLE is displayed
9. Press TARE until the indicator exits the Supervisor menu
NOTE: The Remote indicator will show the message “OFFLINE” when not receiving data from
the Master indicator

Wiring Configuration
Z series Master Z Series Remote
TB3 – 1 GND to TB3 – 1 GND
TB3 – 3 TXD 2 to TB3 – 4 RXD 1
TB3 – 5 RXD 2 to TB3 – 2 TXD 1

APP 104 Remote Display.docx Page 3 of 3