1.0 BRIEF INTRODUCTION
Weighing Indicator XK3118T1 adopts high anti-jamming Single-chip microprocessor and high
precision A/D conversion technology. It is with stainless steel frame (XK3118T1B is with plastic frame)
and widely applied in electronic pplatform scale, electronic floor scale and other static weighing
systems that equipped with 1-6pcs load cells. AC/DC are both available fort his indicator.
2.0 INSTALLATION
2.1 INDICATOR INTERFACE
2.2 K1 BACK SIDE
3
2.1.2 Loadcell with Connection
Specification
① - Excitation
② - Feedback
⑥ + Excitation
⑦ + Feedback
⑧ - Signal
⑨ + Signal
⑤ Shield
! Connection between load cell and indicator must be reliable, shield wire must be connected
to ground reliably. Connection or disconnection are not allowed when the indicator is on, which
may damage the indicator or load cells.
! Static protection must be properly adopted as the load cell and indicator are all static sensitive
equipments. Welding or other strong electricity operation should be strictly forbidden. During
thunderstorm season, proper lightening protection should be tanken care of to protect the load
cells and indicators from damaging by lightening and to ensure the personal safety and the safely
running of the weighing and related equipments.
4
3.0 OPERASYON
3.1 CİHAZ AÇMA/KAPATMA VE OTOMATİK SIFIRLAMA
The indicator power can be controlled by the on-off key on the front faceplate . The indicator
will perform self-check after turned on. If the scale was found departure from the calibrate zero
however still within the range of turn on auto zero then the indicator will display “0” and the in
dicator light for “zero digit” will be on. If the scale was found departure from the calibrate zero and
out of the range of turn on auto zero then the indicator will display the current read
If the range setting of the turn on zero is”- -“ wihch means zero of last turn off, then zero
operation will not be carried on and the zero of last turn off will be automatically readin and the
current weight will be displayed.
Indıcator will back to zero when pressing “ZERO” key if the gross weight is within the range
of manually zero and stable. Manually zero is not valid under the “NET” displaying mode;
TARE operation can be carried out if both gross weight and net weight are above zero and
stable.
The indicator will display ”0” after pressing the “TARE” key. The tared weight is current gross
weight. And the indicator will enter “NET” displaying mode and the “NET” indicator lights will be on;
The indicator will exit “NET” displaying mode after pressing “TARE” when the gross weight is
“0” and under “NET” displaying mode
1. In case, the net weight is bigger than the minimum measurable value (5 divison) and stable,
When pressing “TOTAL”, the current net weight will be added and the “TOTAL” indicator light will be
on and total value will be displayed , which will be changed to adding times [n ***]
After 3 seconds and exit the total displaying status after another 3 seconds. Next operation will be
only valid when the net weight is smaller than the minimum measurable value.
2. If keep pressing “TOTAL” until buzzer alarms under the weighing status “FUNCTION” and “TOTAL”
lights will be on and you can check the total value by pressing “ ” to see the adding times and again
“” to exit.
3. If you press ““ after keep pressing “ TOTAL” and enter the total displaying status there will be an
notice [CLRAr-] which means to clear the total value or not. Press ““ to clear and “FUNCTION” to
exit and keep the value.
5
OPERATION DIAGRAM
Keep Pressing Display Total Press ”” to see Press ”” to exit
“Total” Value the adding times
Press “FUNCTION”
to cancel
4.0 CALIBRATION
After self checking of turn on, it will enter normal weighing display, warm up for 15-30
minutes, open the calibration cover in the back side of the indicator; touch the calibration switch to
enter calibration status, the detailed steps as below:
6 Return back to 【 End 】 After calibration, 2 second after display “End”, the
weighing status 【 3.000】 current weight will be displayed immediately.
★ “ Fn” tuşu ile kalibrasyonu tamamlamadan çıkabilirsiniz.
6
5.0 OTHER PARAMETER SETTING AND FUNCTION
After enter calibration and display 【CAL】, press “” button, indicator displays 【Zero】, then
press “” enter to other parameter calibration setting, the detailed steps as below:
7
“”to
Power on auto set zero range(Auto Set Zero)
switch
3 【At **】 :
“”to
0,2,4,10,20,100 % of full capacity;
confirm
Filter set(Filter):
“”to 【Stb】:stable algorithm(Stable);
switch 【SEn】:sensitive algorithm(Sensitive);
4 【FL ***】
“”to 【 3】:strong vibration resistance
confirm algorithm;
【 4】:high accuracy algorithm;
8
6.0 ERROR NOTICE
Display Note
Err 01 Exceed the Zero Range
Err 02 Not Meet the Requirement of Total
Err 03 Weight Overloaded
Err 04 Weight not stable during Calibration
Err 05 Load Calibratin Error. Too low load or
Calibration code too small or AD everse
Err 09 Data read verify Error, Data Memory Damaged
Err 10 Boot verift Error, SCM damaged
K1 INBOX
No Name PCS
1 Indicator 1
2 Adapter 1
3 Communication Plug (9 Pin) 1
4 Manual 1
5 Certificate 1