RGA22G Scale Guide: A B C DH

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

RGA22G scale guide efg

Scale applicator tool for RGH22 readhead installations

The RGA22G scale guide is mounted in place of any Renishaw RGH22 readhead and
uses the motion of the axis to apply RGS-S tape scale parallel to the guideway whilst
simultaneously removing the backing paper - a one man, one hit operation.

Scale installation
● Prepare scale and surface
● Ensure scale is cut to correct length and mounting surface has been prepared (refer to
RGS-S scale installation guide for details)

● Install the scale guide


● Mount the scale guide to the readhead bracket as shown in figure 1
● Be careful to check that the readhead and the scale guide have the same orientation, as
the readhead does not lie centrally over the scale
● Move the guide to the start of axis travel. Note the direction of scale application as shown
in figure 2
● The design of the RGA22G allows scale application in either direction

Mounting screws

Readhead
mounting RGA22G scale guide
bracket
Splitter screw
Figure 1
Limit of travel
Limit stop
Direction of scale
application

RGS-S tape scale

Splitter screw

Backing paper

Figure 2 20mm
RGA22G scale guide
(continued overleaf) Start of measuring length

a b c dh
INSTALLATION INSTRUCTION SHEET
Part No. M-9531-1281
● Load the scale into the scale guide
● Remove the splitter screw (see figure 1) from the scale guide body
● Separate the backing paper from the first 40mm of scale and feed it into the scale
guide as shown in figure 2
● Relocate the splitter screw behind the backing paper as shown in figure 2
● Push the scale carefully through to the end of scale mark, ensuring that it does not
stick to the mounting surface until it is in position. NOTE to prevent the scale sticking
to the mounting surface during this operation it may be necessary to re-apply
approximately 20mm of backing paper to the end of the scale before inserting through
scale guide. This can then be removed before to fitting end clamps.
● Attach the end of the scale to the mounting surface with light finger pressure

● Apply the scale


● Traverse the axis through its full travel at a slow, steady speed
● While moving the axis
● Apply a light finger pressure to the scale behind the scale guide to
attach it to the mounting surface
● Gently pull the backing paper away from in front of the scale guide as it
is separated

● Remove the scale guide


● When the axis has reached the limit of its travel, lock the axis in place and unbolt the
scale guide from the readhead mounting bracket
● Withdraw the scale guide along the scale whilst applying the remaining scale using
light finger pressure

● Ensure complete adhesion of the scale to the mounting surface by applying firm finger
pressure along the full length of the scale from the centre outwards to each end

● Fit RGC-F end clamps (refer to RGS-S scale installation guide for details)

Scale handling precautions Scale

Reel for lengths


Plastic extrusion for
over 6m
lengths under 6m
Keep coils tight
during storage

Straighten coils to
withdraw scale

Avoid alcohol and >100mm


solvents
Scale surface
is immune to
handling
No sharp objects Do not twist

DO NOT remove the scale for refitting or use elsewhere once it has been applied to the mounting surface.
The scale can be applied once only.

a b c dh USA, Renishaw Inc France, Renishaw SA Switzerland, Renishaw AG Renishaw’s Representative Offices
Tel +1 847 843 3666 Tel +33 1 64 61 84 84 Tel +41 55 410 66 66 People’s Republic of China, Beijing
Renishaw plc Fax +1 847 843 1744 Fax +33 1 64 61 65 26 Fax +41 55 410 66 69 Tel +86 10 6461 2689
Encoder Products Division Fax +86 10 6461 2687
New Mills, Wotton-under-Edge Japan, Renishaw KK Italy, Renishaw SpA Hong Kong, Renishaw (Hong Kong) Ltd Singapore
Gloucestershire GL12 8JR, UK Tel +81 3 5350 2201 Tel +39 11 9 66 10 52 Tel +852 2753 0638 Tel +65 438 2778
Tel: +44 (0)1453 523 634 Fax +81 3 5350 2207 Fax +39 11 9 66 40 83 Fax +852 2756 8786 Fax +65 438 4780
(07000 RENISHAW) Indonesia, Jakarta
Fax: +44 (0)1453 523 760 Germany, Renishaw GmbH Spain, Renishaw Iberica SA Brazil, Renishaw Latino America Ltda Tel/fax +62 21 424 3934
E mail [email protected] Tel +49 712 797 960 Tel +55 11 7295 2866 Tel +55 11 7295 2866 India, Bangalore (liaison office)
Internet http://www.renishaw.com Fax +49 712 788 237 Fax +55 11 7295 1641 Fax +55 11 7295 1641 Tel/fax +91 80 224 3223

© 1999 Renishaw plc Renishaw reserves the right to change specifications without notice Printed in England 499 Part No M-9531-1281-02

You might also like